Nikon OPTIPHOT-POL Instructions Manual


Add to my manuals
28 Pages

advertisement

Nikon OPTIPHOT-POL Instructions Manual | Manualzz

Nikon

Polarizing Microscope

OPTIPHOT -POL

INSTRUCTIONS

NIPPON KOGAKU K. K.

CAUTIONS

o

Avoid sharp knocks!

Handle the microscope gently, taking care to avoid sharp knocks.

o

Exchanging the lamp bulb and fuse

Before replacing the lamp bulb or fuse, turn 0 F F the power switch and disconnect the plug of the power source cord.

In such cases as of replacement, do not touch the lamp bulb with bare hands, immed iately after putting out the lamp.

f}When carrying the microscope

When carrying the microscope, hold its arm with one hand, supporting the bottom of the microscope base with the other. The instrument weighs about

10.5

kg. Do not have the lamp housing carry any load.

8

Place for using

Avoid the use of the microscope in a dusty place, where it is subject to vibrations or exposed to high temperatures, moisture or direct sunlight.

o

Power source voltage

In every case, make sure of the power source voltage by means of the input voltage change-over switch on the bottom of the microscope base.

o

Light source

Halogen lamp bulb to be used is

12V-

50W.

Do not use

12V-l OOW

halogen lamp bulb.

If the lamp bulb of over-rated wattage is used, light adjusting circuit will damage.

Never connect the lamp housing cord to the house current socket directly.

o

I

n lighting the lamp

Take care not to touch the lamp housing being lighted, and don't bring inflammable substances such as gasoline, thinner, and alcohol near to the lamp housing, as some parts of the lamp housing may take a high temperature while the lamp is being lighted.

«;)

Dirt on the lens

Do not leave dust, dirt or finger marks on the lens surfaces.

They will prevent you from clear observation of the specimen image.

o

Strain-free glasses

The optical elements of this microscope being constructed of strain-free glasses, take particular caution in handling the objectives and condenser lenses not to cause strain to them.

4D>

Focus knobs

Never attempt to adjust the tightness of the rightand lefthand focus knobs by turning the one, while holding the other in th is model microscope, because of causi ng disorder.

CARE AND MAINTENANCE

o

Cleaning the lenses

To clean the lens surfaces, remove dust using a soft brush or gauze. Only for removing finger marks or grease, should soft cotton cloth, lens tissue or gauze lightly moistened with absolute alcohol

(ethanol or methanol) be used.

For cleaning the objectives and immersion oil use only xylene.

Observe sufficient caution in handling alcohol and xylene.

8

Cleaning the painted surfaces

Avoid the use of any organic solvent (for example, thinner, ether, alcohol, xylene etc.) for cleaning the painted surfaces and plastic parts of the instrument.

8

Never attempt to dismantle!

Never attempt to dismantle the instrument so as to avoid the possibility of impairing the operational efficiency and accuracy.

o

When not in use

When not in use, cover the instrument with the accessory vinyl cover, and store it in a place free from moisture and fungus.

It is especially recommended that the objectives and eyepieces be kept in an airtight container containing desiccant.

o

Periodical checking

To maintain the performance of the instrument, we recommend to check the instrument periodically.

(For details of this check, contact our agency.)

CONTENTS

I.

NOMENCLATURE

II.

ASSEMBLY.

III.

PREPARATION

1. Centering the Lamp

3. Diopter Adjustment

·············0

....... 0

·······8

2. Interpupillary Distance Adjustment

8

0

0

4. Optical Path Change-over in the

Trinocular Eyepiece Tube "TP"

5. Centering the Objectives

0

0

6. Centering the Condenser Lens 4@

IV. MICROSCOPY

1. Operating Procedure

2. Manipulation of Each Element

1) Focusing

2) Condenser aperture diaphragm

3) Field diaphragm

4) Circular graduated stage

5) Objectives

6) Eyepieces

7) Achromat strain-free condenser

8) Bertrand lens

9) 1/4 A & tint plate

10) Dia-polarizer and analyzer

11) Filters

12) Lowering the substage

13) Illumination system

4J)

4J)

4J)

~

~

~

~

41

41

~

4D

4D

4D

G)

G)

G)

V.

PHOTOMiCROGRAPHy 4D

VI.

ACCESSORIES

1. Senarmont Compensator

~

~

2. Quartz Wedge

3. Monocular Eyepiece Tube "AP"

~

W

4. Universal epi-illuminator

W

5. Attachable Mechanical Stage Type "E".

~

6. Universal Stage ~

VII. TROUBLE SHOOTING TABLE ~

REFERENCE

ELECTRIC SPECIFICATIONS

~

0

I .

NOMENCLATURE

Interpupillary distance scale

Diopter ring

CF eyepiece

Eyeguard

Bertrand lens ring

Centering nosepiece

CF Achromat P objective

(Stra in-free)

Vernier

Circular graduated stage

Achromat strain-free condenser

Orientation plate

Field lens

Brightness control dial

(including power switch)

Lamp centering tool Dust cap

Microflex clamp screw

Trinocular eyepiece tube "TP"

Optical path change-over knob

Eyepiece tube clamp screw

111'1

,

----------Analyzer knob I

Intermediate tube clamp screw

Substage clamp screw

Socket sleeve

Lamp vertical centering ring

Lamp lateral centering screw

Lamp input plug

Filter receptacle

~

(NCB10.

ND2.

ND16&GIFI

Filter frame

Fig. 1

Analyzer rotation ring

Analyzer clamp screw

Intermediate tube "P"

Nosepiece clamp screw

45° click stop lever

Condenser focus knob

Coarse focus knob

Fine focus knob

Lamp housing

Lamp housing clamp screw

Arm rest

-..

Compensator slot

1 /4 A

&

tint plate

Nosepiece centering screw

Specimen clip

Condenser aperture diaphragm control ring

------Dia-polarizer

Condenser clamp screw

Brightness indicator

Field diaphragm control ring

X-POL stand

Fig.2

II. ASSEMBLY

• To assemble the microscope, follow the procedure in the order given:

Bottom of the Base

CF eyepiece

Insert the eyepiece CFW 10XCM into the right-hand sleeve, fitting the pin of the eyepiece in the right-hand notch of the sleeve.

Into the left-hand sleeve, insert theCFW10X.

Power source voltage

Set the input voltage to the power source voltage by means of the change-over switch on the bottom of the base.

Bottom of the Base

~

~

-------

@ 1/4 A

&

tint plate

Remove the screw by the side of the 1/4

"A.

plate of the 1/4

"A.

& tint plate and insert it into the compensator slot of the intermediate tube "P", facing the positioning groove toward the operator side.

Reattach the removed screw.

@

Centering the nosepiece sufficiently.

Releasing the nosepiece clamp screw on the left side of the microscope arm, insert the nosepiece.

Making sure of positive fitting of the pin on the microscope arm into the nosepiece groove, refasten the clamp screw.

Leveling foot screw

For stable installation of the microscope, manipulate the adjusting screw at one foot on the bottom of the microscope base.

Bottom of the Base

@6

CF Achromat P objective (Strain-free) piece in such positions that their magnifying power increases as the nosepiece is revolved clockwise.

@ Specimen clip

Place the clip on the stage using holes on the stage surface.

Aperture number plate lowest power

If the illumination is found too bright or unstable, when the switch is turned ON, make adjustment of the lowest voltage in the following way:

1) Turn the brightness control dial

to

OFF.

2) Turn the lowest voltage adjusting screw on the bottom of the microscope base counterclockwise to the limit, using a screw driver.

3) Turn the brightness control dial to ON_ At this time, the lamp voltage will be highest, immediately after lighting.

4) In this condition, gently turn the lowest voltage adjustingsc;:ew clockwise, to set the voltage to nearly 4 on the ind icator.

Achromat strain-free condenser

Insert the condenser into the condenser carrier, facing the aperture number plate toward the operator.

Fasten the clamp screw on the left side of the carrier.

@ Circular graduated stage

Release the substage clamp screw using a driver, and sl ide the substage on the dovetail fitting.

In such a position that the top ends of the substage and the dovetail are at the same level, fasten the clamp screw.

Substage clamp scre,

@

Eyepiece tube

Attach the eyepiece tube on the intermediate tube "P", fitting the notch of the circular dovetail on the end of the clamp screw.

Fasten the clamp screw.

@

Intermed iate tube "P"

Attach the intermediate tube "P" on the arm of the X-POL stand, fitting the notch of the circular dovetail on the end of the clamp screw. Fasten the clamp screw.

Filter frame

Eyepiece tube clamp screw

@

Filter

Place the filter into the filter frame.

After centering the lamp, put the filter with the frame into the filter receptacle. No frame is provided for the diffuser.

®

Power input plug

Connect the plug to the receptacle on the fi Iter receptacle's right side.

®

Lamp housing lamp housing into the collector lens, and fasten the clamp screw on the left side of the housing.

Lamp input plug

Connect the plug to the receptacle on the filter receptacle's left side.

Socket sleeve

Insert the socket sleeve into the lamp housing, and fasten it firmly with the clamp screw.

~

f3\

Halogen lamp bulb

Insert the lamp bulb with its pins into the accepting holes in the socket.

Note: Don't touch the bulb surface directly with the fi nger.

Socket sleeve clamp screw

Fig. 3

III.

PREPARATION

1. Centering the Lamp

1) Connect the power source cord to the socket.

2) Turn the brightness control dial to switch

ON and adjust the voltage to 6 on the indicator.

3) Place the specimen on the stage, and focus on the specimen using

lOx

objective.

In this case, open the condenser aperture and field diaphragms to the largest extent.

4) Roughly center the condenser lens using lOX objective, given on P. 10- 6.

following the procedures

5) Put the lamp centering tool on the field lens and onto the tool place a NO filter.

(Fig. 4)

ND filter

Lamp centering tool

Fig. 4

6) Stop down the condenser aperture diaphragm, release the lamp housing clamp screw, and move the lamp housing back and forth (Fig. 5), until a sharp image of the lamp filament appears on the aperture diaphragm surface, which can be seen by the reflection from the NO filter.

7) Release the socket sleeve clamp screw (Fig.

6). Turning the lamp lateral centering screw and vertical centering ring, bring the filament image to the center, as shown in

Fig.7.

Socket sleeve clamp screw

Lamp vertical centering ring

Lamp lateral centering screw

Filament image

Fig. 6

Condenser aperture diaphragm

Fig. 7

8) As shown in Fig. 8, put the diffuser, with its matte surface faced toward the microscope stand, into the filter receptacle which is the closest to the microscope stand.

Lamp housing

Lamp housing clamp screw

/

Fig. 8

The above centering procedure should be carried out, when replacing the lamp bulb.

Fig.5

2.

Interpupillary DistanceAdjustment I

Place a specimen on the stage, and focus on the specimen.

As shown in Fig. 9, adjust the interpupillary distance, so that both the right and left viewfields become one.

'

4. Optical Path Change-over in the

Trinocular Eyepiece Tube "TP"

(Fig. 11)

Fig. 9

3. Diopter Adjustment

Rotate the diopter ring on the eyepiece CFW lOx CM until the cross lines are seen clear.

(Fig. 10)

Vertical photo tube: 86%

Observation Observation tube: 100% tube: 14%

+---+

~OPtical path change-over knob

Fig.11

*Since the CF eyepieces are of high eyepoint type, it is not necessary for the user putting on his spectacles to remove them.

Only fold down the eyeguard rubber.

(Fig.13)

Fig. 10

(For b inocu lar observation)

1) Mount the specimen on the stage. Swing the objective 10 x into position, and bring the specimen image into focus looking into the right-hand eyepiece.

2) Witll.Out

manipulating the coarse-and-fine focus knob, turn the diopter ring on the left-hand eyepiece to focus on the specimen.

Fig. 12 Fig. 13

5. Centeringthe Objectives

1) Place the specimen on the stage, and focus on the specimen. Bring an appropriate target to the center of the cross lines in the eyepiece.

2) Insert the centering tools in the centering screws on the nosepiece. (F ig. 14)

Fig. 14

3) Rotate the stage about 1800, and the target is displaced from the center of the cross lines. Move the objective using the centering tools so that the center of the cross

Iines comes to one half position of the displacement of the target. (F ig. 15)

Rotate

1800

Target-

A half position of the displacement

Fig. 15

• Repeat the above procedure two or three times, and the rotation center of the stage coincides with the cross Iines center .

• Carry out centering for each objective.

6.

Centering the Condenser Lens

1) Close the field diaphragm in the microscope base to its smallest size by means of the field diaphragm control ring. Rotate the condenser focus knob to move the condenser vertically so that a sharp image of the field diaphragm is formed on the specimen surface.

2) Bring the field diaphragm image to the center of the field of view by means of the condenser centering screws.

(Fig.

16-1IJ)

3) Change over to the objective 40 x, and adjust the field diaphragm so that the image of the diaphragm is about the same as the eyepiece viewfield stop, as shown in

Fig.

16-~.

If not centered, use the condenser centering screws again.

Image of field

[~:a~~~~~~

~'\

Eyepiece viewfield stop

":/

,,>J

)

Fig. 16

IV.

MICROSCOPY

11.

Operating Pr0Cf}dur~

1) Turn the brightness control dial (including power switch) to light the lamp.

2) Bring the analyzer and the Bertrand lens out of the optical path.

3) Place the specimen on the stage and swing the

10X objective into position. Focus on specimen.

4) Adjust the interpupillary distance and diopter.

(Refer to P. 9)

5) Make certain of correct illumination.

(Refer to P. 8)

6) Put the filters necessary into the filter receptacle.

Fully opened

Circumscribed the

7) Carry out the centering procedure for the objective. (Refer to P. 9)

8) Carry out the centering procedure for the condenser. (Refer to P. 10)

9) Bring the analyzer into the optical path.

10) Swing in the objective to be used and refocus on specimen.

11) Brightness is adjusted by selecting ND filters or by changing the lamp voltage to 6 ~ 12.

Table 1

IN

IN

field of view the circumfer-

IN

2. Manipulation of Each Element

1) Focusing

• The relation between the direction of rotation of the focus knobs and that of vertical movement of the stage is as indicated in

Fig. 17.

Torque adjustment ring

Fine focus knob

Coarse focus knob

Fig. 17

• One rotation of the fine focus knob moves the stage 0.1 mm.

The graduation on th is focus knob ~ divided into 111m.

One rotation of the coarse focus knob moves the stage 4.7mm.

• Tightness of the coarse-fine focus knob having been properly adjusted by the manufacturer, it should never be readjusted in this model microscope by turning the one knob while holding the other.

2) Condenser aperture diaphragm (A diaphragm)

(1) Orthoscopic microscopy

• The condenser aperture diaphragm is provided for adjusting the numerical aperture (N.A.) of the illuminating system of microscope.

In genera I, when it is stopped down to 70

~ 80% of the numerical aperture of the objective, a good image of appropriate contrast will be obtained. (Fig. 18)

Exit pupil of objective

Aperture diaphragm

Size of the condenser aperture diaphragm

Fig. 18

• Remove the eyepiece from the eyepiece tube, adjust the size of the diaphragm, observing the image of the diaphragm which is visible on the bright circle of exit pupil of objective inside.

• When swinging out the top lens of the condenser (for microscopy using

4x

or lower objective), fully open the condenser aperture diaphragm.

(2) Conoscopic microscopy

• In conoscopic microscopy, the condenser aperture diaphragm works as a field diaphragm on the conoscopic image surface.

Stop down the diaphragm to such an extent that it circumscribes the circumference of the field of view of the conoscopic image (exit pupil of objective) to shut out the stray light.

3) Field diaphragm (F diaphragm)

• The field diaphragm is used for determining the illuminated area on the specimen surface in relation to the field of view of the microscope.

Generally, it is stopped down to such an extent that the circumference of the illuminated area circumscribes that of the eyepiece field of view.

[Note] This diaphragm does not work as the field diaphragm when the condenser top lens is swung out of the optical path.

In this case the diaphragm is recommended to be fully opened because the numerical aperture of the illuminator will be cut off when this diaphragm is excessively stopped down.

4) Circular graduated stage

• The rotation angle of the stage is readable with the accuracy of 0.10

via a paired vern ier scales.

When the reading with one of the vernier scales is interrupted by the attachable mechanical stage type "E", read the other vernier and add ±90° to the reading .

• The 450 click-stop device comes to act at every 450 rotation, starting from a position where the click stop lever has been pulled toward the operator, giving convenience in switching over the observation from a crossed Nicols position to a diagonal position. (Fig. 19)

Clamp screw

45° click-stop lever

Fig. 19

For the click-stop release, turn the lever toward the microscope body.

• The stage can be clamped at any position using the stage rotation clamp screw on the right-hand vernier.

5) Objectives

• The CF Achromat P objectives (Strain-free) and CF eyepieces adopted in the Nikon

POLARIZING MICROSCOPE OPTIPHOT-

PO L are designed on the basis of a concept

"Chromatic Aberration Free".

In every case use the CF objectives in combination with the CF eyepieces.

(1) Oil immersion objectives (Oil)

• Objective CF Achromat P 100X (Oil), an oil-immersion type, is to be immersed in oil between the specimen and front of the objective.

To see if air bubbles are present in the immersion oil, which deteriorate the image quality, pull out the eyepiece from the eyepiece tube to examine the objective exit pupil inside the tube. To remove air bubbles, revolve the nosepiece slightly to and fro several times, apply additional oil, or replace the oil. Be careful not to rotate the nosep iece too far as to soil the ends of the other objectives with oil.

• To clean off the oil, pass lens tissue or soft cloth moistened with xylene lightly two or three times over the lens. It is essential at this time to avoid touching the lens with the part of tissue or cloth once used.

(2) Coverglass

• With the objectives engraved "160/0.17", use a coverglass of O.17mm in thickness.

• The indication "160/-" on the objective means that no matter whether a coverglass is used or not, no decrease of image definition or of contrast will result.

6) Eyepieces

• To take full advantage of the CF eyepieces, use them in combination with the CF objectives.

• By inserting the eyepiece with cross lines and graduation (CFW 10XCM) into the eyepiece sleeve fitting the protractor pin into the right-hand side groove of the sleeve, the O-direction of the analyzer and dia-polarizer are al igned with the cross lines direction.

If the protractor pin is fitted to the upper right side groove of the sleeve, the cross lines will be aligned with the diagonal position of the polarizaiton.

• CF PL Projection lenses are exclusively designed for photom icrography.

Do not use them for observation.

---

• For focusing with the observation tube of the trinocular eyep iece tube for photomicrography, use the eyepiece incorporating the photo mask.

7) Achromat strain-free condenser

• The top lens of the condenser is to be placed in the optical path for the orthoscopic and conoscopic microscopy provided that it is to be swung out when an objective of 4X or lower magnification is in use.

[Note] For the orthoscopic microscopy, a lower numerical aperture illumination with the top lens swung out condenser was used to be recommended, however, this method is not effective especially for high magn ification observation because of the lowered resolution.

Hence, for the latter case, use of the top lens may rather be recommendable except the retardation measurement or the interference color observation for wh ich it is necessary to make the illumination light flux as parallel as possible to the optical axis by swinging out the top lens or stopping down the aperture diaphragm.

• Thickness of the glass slide must be 1.7mm

or less, otherwise, the field diaphragm

might fail to focus its image on the specimen.

8) Bertrand lens

(with the trinocular eyepiece tube "TP" or the binocular eyepiece tube "BP" in use)

• Bring the Bertrand lens into the optical path by turning the Bertrand lens ring leftward to observe the conoscopic image.

(Fig. 20)

In the above simultaneous observation of the conoscopic and orthoscopic images, the former image may appear deviated from the orthoscopic view field center, however, the deviated image represents the conoscopic light flux that covers the central part of the orthoscopic view field to the extent of about 1/18.

9) 1/4

'A

& tint plate

• Removing the 1/4 'A plate side screw, hold the 1/4 'A & tint plate, the cI ick stop groove facing the operator and insert it forward into the compensator slot.

Then screw-in the above screw as it was.

(Fig.23)

Bertrand lens ring

Fig. 20

• The conoscopic view field is as large as about 1/4 of the orthoscopic view field.

(Fig. 21)

Conoscopic image of this area can be observed

--Orthoscopic viewfield

Fig. 21

• The conoscopic image may also be observed overlapping on the orthoscopic image through the binocular observation, one of the paired eyepieces being replaced with the accessory pin hole eyepiece and without the Bertrand lens in the optical path.

(Fig.22)

Fig. 23

• The tint plate has an empty hole at the center. By pushing it through the slot, the sensitive tint plate (530nm) is brought into the optical path and by pu IIing it out the 1/4 'A p late is brought into the optical path.

10) Dia-polarizer and analyzer

• When the both are set at ~ read ing on the protractor scale, position of the polarization plane coincides with the orientation plate, which shows that the indication "P" of the X-direction is for dia-polarizer and the

"A"

of the V-direction is for analyzer, on the microscope base. (Fig. 24)

[Note] Some of the polarizing microscope's reference books or special works available in the market may explain that the X-direction is for analyzer and the V-direction, for polarizer.

Fig. 22

Orientation plate

~

Diffuser

(Without frame)

Fig. 24

• The dia-polarizer rotates 3600, and can be detached from the substage by pulling downward.

(Fig. 25)

For attaching the dia-polarizer, push it with the pin on the dia-polarizer into coincidence with the groove at the position o of 0 on the bottom of the substage.

11) Filters

• Put the filter with the frame into the filter receptacle between the microscope base and the lamp housing. The accessory filters are as shown below:

Table 2.

Use of Filters

2 filter (T=50%) ment and contrast-up in

For retardation measure-

Dia-polarizer

Fig. 25

• The analyzer rotates 1800 via the rotation ring, the left-hand side clamp being released.

The rotation angle readable with accuracy of 0.10

via the vernier.

The analyzer can be taken out of the optical path by pulling out by the analyzer knob. (Fig. 26)

1 2) Lowering the substage

• Releasing the clamp screw using a screw driver, as shown in Fig. 27, permits lowering the substage as far as 32mm from the observing position beyond the moving stroke of the focusing device.

So, the microscope makes it possible to examine the thicker specimens (mainly in episcopic polarizing microscopy) and to use the universal stage.

Substage clamp screw

Analyzer clamp screw

Analyzer rotation ring

Analyzer knob

Fig. 26

~

rtY

Fig. 27

13) Illumination system

• The optical system for illumination in the

OPTIPHOT-POL microscope is constructed to fulfill the Koehler illumination requirements perfectly, and offers a bright, uniform field without any change-over manipulation.

• Halogen lamp 12V-50W (OSRAM 64610 or PH I LIPS 7027) is used as a Iight source.

V.PHOTOMICROGRAPHY

Prepare the following equipments in addition to the OPTIPHOT-POL microscope main body.

*

Nikon Microflex

*

Trinocular eyepiece tube "TP"

*

CF PL Projection lens

1. CF PI:.Projection Lenses

The combined use of the CF P objectives and

CF PL Projection lenses is essential.

For the same total magnification, select a combination of the highest possible objective power and lowest possible projection lens power to achieve contrast.

the utmost image definition and

Monochrome

Daylight film film

2.

Illumination

1) Checking the illu mination

Unevenness in the illumination will show up more conspicuously in photomicrography than in observation.

Consequently, before taking a photograph, recheck the positioning and centering of the lamp and the correct adjustment of the condenser.

2) Selection of voltage and filter

The color temperature of the light source varies with the voltage being used. Therefore, in color photomicrography, the selection of voltage and filter is essential

(for the result to be obtained).

Table 3.

Standard Selection

8

type

Contrast fi Iter(green),

Remove NCB 10

Table 3 shows the standard combination.

Depending upon the make of the film, different color renditions may result. It is recommended that in addition to the NCB

10 filter a color compensation filter (CC filter), available from the film manufactu rer, be used.

3.

Shutter Speed

Desirable shutter speeds for least vibration are

1/4 ~ 1/15 sec. Adjustment of the image brightness for color photomicrography made by means of the NO filters.

should be

4. Manipulation of Field and Aperture

Diaphragm

In photomicrography, the adjustment of the field diaphragm is important for the purpose of limiting extraneous light which causes flare in the microscope image.

Stop down the diaphragm so as to get an illuminated area slightly larger than that of the picture field. By adjusting the aperture diaphragm, a change of depth of focus, contrast and resolution of image is attainable.

Select a size suited to the purpose.

Generally speaking, the aperture diaphragm, is properly stopped down to 70 ~ 80% of the aperture of the objective being used.

5.

Focusing

Focusing for photomicrography can be done with the observation tube of the trinocular eyepiece tube "TP" or by using the Microflex finder.

1) Adjust diopter

• Using binocular of eyepiece tube:

Use 4X or 1 OX objective.

Insert the mask eyepiece into either of right or left eyepiece sleeve that is accustomed to usual use. Adjust the diopter ring to bring the double cross line in the view field center into focus. (Fig. 28)

Then focus the specimen image also on the central area of the mask by means of the focus knob of the microscope.

The diopter of another eyepiece is to be adjusted by focusing specimen rotating the diopter ring without using the microscope focus knob.

Rotate the mask eyepiece so as the mask positions as shown in Fig. 32.

• Using ocular finder:

Adjust the diopter ring so as the double cross line in the view field center can be seen clear and each line separated. (Fig. 29)

. _

II

of the mask eyepiece

/il111

II

II

Fig. 28

Double cross Ii.ne

of the ocular finder

«> 4)7 q)P ~

-+ ~

'l

Fig. 29

2) Make focusing according to the magnification of objective to be used.

• Using 40X or higher objective:

With diopter adjusted eyepiece make the specimen image sharp by rotating the microscope fine focus knob and make sure that both of the double cross line and the speci men image are seen crisp Iy at the same time.

• Using medium magnification objective

10X, 20X, etc.:

After focusing the same way as above, bring the specimen image to coincide with the double cross line so as their relative position is fixed and unchanged under observation by swinging your eye laterally.

(Focusing by parallax method.)

• Using 4X or lower objective:

Attach the focusing magnifier to the ocular finder. (Fig. 30)

The focusing magnifier is to be adjusted beforehand for viewing infinit distance

(magnifier is set at the red line).

Viewing through the attached focusing magnifier, move it back and forth until the double cross line is seen clear. Then, focus the double cross line and the specimen image by rotating the fine focus knob as sharp as possible.

I

6.

Picture composing

Compose the picture within the mask in the ocular finder corresponding to the film size in use by driving the microscope stage by lateral and longitudinal movement and rotation. (Fig. 31)

For 35mm film

For 4"x5"

Polaroid film

For 3~" X 4X;"

Polaroid film

For 6X9 roll film

Double cross line

Finder mask

Fig. 31

When the mask eyepiece is used, select one out of masks in the view field suitable to the film size relative to CF PL Projection lens in use, in reference with Fig. 32 and

Table 4.

Inner frame

Intermediate frame

Outer frame

Mask of the mask eyepiece

Fig. 32 o

Fig. 30

mediate

Outer lens

CF PL Mask

3%"X frame

-

---

-

---

--

-

-

-.

-

-

Film size

Table 4

Note: Framing for picture composing will be more accurate by the ocular finder than the mask eyepiece.

7. Others

• As the intermediate tube "P" of OPTI-

PHOT-POL microscope builts in the depolarizer, it's not necessary to give care to the relation between the orientation of the polarizer, analyzer and the position of the Microflex.

• When using the 2 X objective, it is recommended to remove the swing-out achromat condenser.

• For photomicrography, when focusing with the binocular observation tube, use the CF eyepiece, CF Photo eyepiece and

CF Photo Mask eyepiece, with the magnification and other indications engraved in yellow, or in white with a white dot in addition.

• For the use of other photomicrographic attachments refer to the pertinent instruction manuals.

VI.

ACCESSORIES

1.

Senarmont Compensator

To be inserted into the compensator slot of the intermediate tube "P" in place of the 1/4 A & tint plate to measure the retardation with the accuracy of the A unit. (Fig. 33)

2.

Quartz Wedge

The quartz wedge is used instead of the 1/4 A

& tint plate that is in the compensator slot of the intermediate tube "P". (Fig. 34)

With this wedge the retardation in the range of

1 A ~ 6 A can roughly be measured.

Fig. 33

1)

Detecting of extinction position

Rotate the stage with the specimen under the crossed Nicols to find out the direction where the specimen part for measurement appears darkest.

2) Detecting of subtraction position

Rotate the stage 45° to bring it to the diagonal position from the extinction position and confirm that the interference color of the specimen part for measurement changes toward the lower order side by inserting the 1/4 A

&

tint plate into the optical path.

If the color changes toward higher order side, rotate the stage further by 90°.

3) Measurement

Inserting the filter GIF into the filter receptacle, replace the 1/4 A

&

tint plate by the compensator.

Rotate the analyzer so as the specimen part for measurement becomes as dark as possible.

Let the angle of the above analyzer rotation be eO then the retardation R (nm) will be obtained as follows: where A wave length of the

Iight used for the measu rement

When the fi Iter G I F is used: A = 546nm

-"~

Fig. 34

1) Detecting of extinction position

Detect the position where the specimen part for measurement becomes darkest by rotating the stage under the crossed Nicols.

2) Detecting of subtraction position

Rotate the stage 45° to bring it to the diagonal position from the extinction position and confirm that the interference color of the specimen part for measu rement changes toward the lower order side by inserting the quartz wedge into the optical path.

If the color changes toward the higher order side, rotate the stage further by 900

3) Measurement

By sliding the quartz wedge along the slot, the interference color changes consequently.

The wedge sliding is to be stopped when the specimen part for measurement comes under the dark stripe, then compare the interference color of the view field beyond the specimen but under the same dark stripe with the Interference Color Chart to assume the amount of retardation.

If the view field is entirely filled with the speci men arou nd the part to be measu red, restrict the illumination of the view field except around the part for measurement by means of the field diaphragm, remove the specimen away the optical path and then compare the interference color with the chart.

I

3.

Monocu,l,ar Eyepiece Tube

II

AP"

Bertrand lens I~ turret

Bertrand lens focus turret ~.

Fig. 35

1) Bertrand lens

The Bertrand lens is brought in and out of the optical path by turning the Bertrand lens turret.

The lens is in the optical path when the indication on the turret is B.

The Bertrand lens can be focused by turning the focus turret located under the

Bertrand lens turret.

2) Pin hole knob

The pin hold can be put in or out of the optical path by operating the pin hole knob located right-hand side of the eyepiece sleeve.

By means of the pin hole, the conoscopic observation of the specimen area with in

10Mm¢ (when a 100 X objective is used) is possib Ie.

4.

Universal epi- illuminator

Used for episcopic polarizing microscopy, mou nted between the X -PO L stand and the intermediate tube "P".

1) Nomenclature

• Referring to Fig. 36, assemble in the order

CD given.

Remove the eyepiece tube and the intermediate tube "P" from the X-POL stand.

(2) Mount the universal epi-illuminator on the microscope arm, positioning the illuminator nearly parallel to the arm. Clamp the screw.

QJ After releasing sufficiently the clamp screw on the lamp housing, to which the lamp bulb (12V -50W Halogen lamp) and socket is attached, insert the lamp housing into the universal epi-illuminator clamp the screw.

and

@ Connect the lamp cord to the transformer.

®

Remove the accessory ND32 filter slider from the illuminator.

Push in the polarizer

(§)

(J) slider until it clicks twice.

Place the filters.

Mount the intermediate tube "P" on the illuminator, fitting the notch of the circular dovetail on the end of the clamp screw. Fasten the clamp screw.

®

Referring to p.7, mount the eyepiece tube on the intermediate tube "P".

Universal epi-illuminator

Optical-path change-over knob

Dust-tight slider

Intermediate tube clamp screw

Socket sleeve clamp screw

Lamp lateral centeri ng screw

--+

I Transformer I

Lamp vertical centering ring

Field diaphragm ring

Polarizer rotation ring

Polarizer sl ider

Fig. 36

2) Preparation

(1) Centering the lamp

CD

Make certain that the optical- path changeover knob is pushed to the limit.

~ Turn ON the power switch on the transformer, set the voltage to 6V.

Q) If the L900C filter is in the optical-path, remove this.

@

Fully open the aperture diaphragm.

(5) Place the ND filter on the stage and focus on it using objective 10 X.

®

Remove the eyepiece from the sleeve, looking into the exit pupil of objective, move the lamp housing back and forth to form a sharp image of the lamp filament on the diffuser of exit pupil.

(J) Manipulate the lamp centering screws to center the filament image on the exit pupil.

®

Place the L900C filter.

If the image is found too dark with an objective of 40 X or higher, remove the

L900C filter.

(2) Orientation of polarizer (intermediate tube

"P" )

CD

Nearly focus on the ND filter on the stage using objective 40X.

~ Set the polarizer graduation to "0 ".

Q) Remove one eyepiece from the observation tubes.

Looking into the exit pupil of the objective, rotate the polarizer rotation ring to form the dark cross image on the exit pupil.

Note:

(Refer to Fig. 37)

Take care not to touch the polarizer rotation ring while observing the specimen, or the orientation of the polarizer will get out of order.

If it is touched by mistake, readjust the orientation.

I

6.

Universal Stage

4) For manipulation and microscopy, refer to diascopic polarizing microscopy.

5.

Attachable Mechanical Stage Type

To attach the attachable stage on the graduated stage, fit the two positioning pins on the rear side of the attachable stage into the two pin holes on the graduated stage surface, and clamp the screw using a driver or a coin.

Attachable mechanical stage is equipped with point counters, whose pitch is O.2mm

or

O.3mm.

The counter can be replaced by releasing the head of the point counter by means of a coin and removing the milled part of the counter.

To release the click-stop of the point counter, release the click spring nut. (Fig. 38)

Point----spring

~I

~o

~

I

,

L~L

Clamp screw

I

,I Positioning pins

Cgy

Fig. 38

Bottom hemispherical lens

Dark cross image

Fig. 37

3) Objectives

Use the objectives CF M Plan Achromat P series (Strain-free, 210/45).

Fig. 39

When using the universal stage, lower the substage beforehand to face the wh ite dot.

with the mark ~ T on the microscope stand, referring to P. 15 12).

For using the universal stage, refer to the separate instructions on "Universal Stage".

periphery or

No good image attached positioned

Dirt or dust in

VII.

TROUBLE SHOOTING TABLE

Although nowhere you can find any disorder or derangement in the instrument, if you encounter some difficulty or dissatisfaction, recheck the use, referring to the table below:

1.

Optical

,

P. 6)

while being

Oneside dim-

• Dirt or dust on the entrance lens

• Incorrect illumination focused

,

) Use ND filter (Refer to P. 12)

No focused from low power low power objective 2·X)

2.

Manipulation

High power ob-

)

) Turn over the slide

Insufficient serving eyes of

Actions not tightly , Adjustment (Refer to P. 9)

----->

Use ND filter or change power

No fusion of voltage

Lamp does not light even

Unstable

3.

Electrical

Actions

Causes obtained

,

)

, ,

---->

Make adjustment (Refer to P. 7)

Strong glare

---->

Make adjustment (Refer to P. 7)

Lamp bulb

Insufficient lamp bulb

Actions

, Replacement

,

.

• Lamp bulb going to be blown (Refer to P. 12)

• Lamp bulb insufficiently inserted

REFERENCE

This manual instructs only how to manipulate the OPTIPHOT-

POL microSCDpe.

For the practical explanation on polarizing microscopy, refer to the following special works:

• "AN INTRODUCTION TO THE METHODS OF OPTICAL

CRYSTALLOGRAPHY"

F. Donald Bloss

Holt, Rinehart and Winston

• "ORE MICROSCOPY"

Eugene N. Cameron

John Wiley & Sons. Inc.

• "THE POLARIZING MICROSCOPE"

F.A. Hallimond -

Vickers Instruments

220/240V

Halogen lamp

Fuse

ELECTRIC SPECIFICATIONS

Power source

100/120V

1A/250V

50/60Hz

12V-50W

[OSRAM 64610J

We reserve the right to make such alterations in design' as we may consider necessary in the light of experience.

For this reason, particulars and illustrations in this handbook may not conform in every detail to models in current production.

:'

(Nikon)

NIPPON KOGAKU K.K.

Fuji Bldg., 2-3, 3 chome, Marunouchi,

Chiyoda-ku, Tokyo 100, Japan ft03-214-5311

Telex: J22601 (NIKON)

Printed in 1apan

(86.4.e)H . E -5r

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Key Features

  • Polarizing optics for detailed observation of birefringent materials
  • Strain-free optical elements for clear and distortion-free images
  • Trinocular eyepiece for versatile observation and documentation options
  • Halogen lamp illumination with adjustable brightness for optimal lighting conditions
  • Achromat strain-free condenser for sharp and high-contrast images
  • Large circular graduated stage for precise specimen positioning
  • Bertrand lens for interference figure observation
  • 1/4 lambda & tint plate for retardation analysis
  • Multiple filter options for precise color balance and contrast

Related manuals

Frequently Answers and Questions

How do I center the objectives?
Refer to section III.5 in the instructions for detailed steps on centering the objectives.
What is the purpose of the condenser aperture diaphragm?
The condenser aperture diaphragm controls the amount of light passing through the condenser, affecting the contrast and resolution of the image.
How do I adjust the field diaphragm?
Use the field diaphragm control ring to adjust the diameter of the illuminated field of view.
What is the function of the analyzer knob?
The analyzer knob is used to rotate the analyzer, which is essential for polarizing observations.
Download PDF

advertisement