advertisement
icountLCM20
g
icountLCM20 Condition
Monitoring Handbook
P.849101_CM20_GB_VerA
© 2010, Parker Hannifin Corporation www.parkerhfde.com
Laser Information
This product contains an invisible infrared 5mW laser.
Any dismantling of the product may result in dangerous exposure to laser radiation.
DANGER
INVISIBLE LASER RADIATION
WHEN OPEN. AVOID DIRECT
EXPOSURE TO BEAM.
g
The internal protective housing label which is mounted on laser module contains the following information:
‘This product is a Class I laser product which complies with both USA21 CFR 1040.10 & 1040.11 and (BS) EN 608285-1’
NOTE: Users are not required to access the laser radiation source and should never do so.
2
Parker Hannifin icountLCM20
Contents
Laser Information .....................................................................................2
Introduction ............................................................................. 4
Features ....................................................................................................4
Benefits .....................................................................................................5
Powering the icountLCM20 ......................................................................8
Installing paper and ribbon into the printer ..........................................10
Getting started ...................................................................... 12
Setting the time ......................................................................................12
Setting the date ......................................................................................13
Basic operation ...................................................................... 14
Independent flow test .............................................................................16
ISO mode ................................................................................................17
NAS mode ...............................................................................................18
SAE mode ...............................................................................................19
GOST mode .............................................................................................20
Entering a test identification code .........................................................21
Bar code wand ........................................................................................22
Printing test results................................................................................23
Viewing and printing a stored test by ID code or test number ..............24
Printing a stored test as a graph ............................................................25
Alarm levels ............................................................................................26
Automatic testing ...................................................................................28
Calibration ..............................................................................................29
Adjusting screen brightness ..................................................................29
Additional (optional) features .................................................................29
Reference .............................................................................. 30
Data interpretation .................................................................................30
ISO contamination charts .......................................................................37
Component cleaning guidelines .............................................................40
Fluid flow schematic ..............................................................................41
Logic diagram .........................................................................................41
Peripherals and serial interface ............................................................42
Measurement accuracy ..........................................................................42
Diagnostic codes.....................................................................................43
Technical specifications .........................................................................45
Installation dimensions ..........................................................................46
Dual fluids monitor - icountLCM202061 ................................................47
Operation checklist ................................................................................48
Ordering information ..............................................................................49
g
3
Parker Hannifin icountLCM20
INtRoDuctIoN
Introduction
The icountLCM20 Contamination Monitor from Parker represents the most up-to-date technology in solid particle contamination analysis, and is the first truly portable monitor.
The icountLCM20 is a complex instrument, but at the same time has reliability, simplicity and ease of operation designed-in.
This owner’s manual has been carefully prepared to guide you, the user, step-by-step through how to ‘get started’, how to obtain measurements and how to interpret the results. Information on the ‘Dual fluids monitor – icountLCM202061’ is also included on page 47. g
The real benefits to be gained from the icountLCM20 are achieved through regular use, particularly as an effective comparator.
With a typical test taking only two minutes, the icountLCM20 is ideal for use as your standard fluid contamination monitoring instrument.
Above all, the icountLCM20 has been designed for practical use.
Features
Test time
Repeat test time
Principle of operation
Particle counts
International Codes
Data entry
Data retrieval
Calibration
Recalibration
Max. working pressure
Max. flow rate
Working conditions
Memory store
Computer compatibility
Portability
Power requirement
System connection
Printer facility
Leak-free sampling
Certification
Rechargeable battery icountLCM20 cover
Handset cover
2 minutes
Every 2 minutes
Optical scanning analysis and measurement of actual particulates
MtD: 4+, 6+, 14+, 21+, 38+ and 70+ microns
AcFtD: 2+, 5+, 15+, 25+, 50+ and 100+ microns
ISO 7–22; NAS 0–12; SAE 0–12
32 character two line dot matrix LCD. Full alphanumeric entry facility on keypad
Memory access gives test search facility
By recognised on-line methods confirmed by the relevant International Standard
Organisation procedures.
MtD – Via a certified primary ISO 11171 automatic particle counter using ISO 11943 principles, with particle distribution reporting to ISO 4406:1999.
AcFtD – Conforming to ISO 4402 principles with particle distribution reported to ISO
4406:1996 with the additional 2µ channel included for reference.
Consult Parker
420 bar
400 l/min when used with System 20 Sensors. Higher with Single Point Sampler
(Consult Parker)
The icountLCM20 will operate with the system working normally
300 test (scrolling memory) capacity
Interface via RS232 connection @ 9600 baud rate
Weighs only 8kg with its own battery pack
Battery module or via the 12Vdc input
Via System 20 Inline Sensors or the Single Point Sampler
Integral 16-column printer for hard copy data
System 20 Sensors ensure sealed fluid extraction and no contamination ingress
This product complies with all relevant EC declarations of conformity.
Standard
Weatherproof cover (Standard)
Weatherproof cover (Optional extra)
4
Parker Hannifin icountLCM20
INtRoDuctIoN
Failsafe features
Special ‘Diagnostics’ are incorporated into the icountLCM20 microprocessor control to ensure effective testing.
Circuitry
Adequate flow
Incorporates an internal diagnostic programme to ensure integrity of results
Flow test facility ensures adequate flow
Adverse conditions On-board trace heating enabling testing in adverse conditions can be offered as an optional extra g
Data management
A specially designed DATµm software package is supplied to allow test results to be downloaded onto a computer.
Benefits
■ ■
■ ■
■ ■
■ ■
■ ■
■ ■
■ ■
■ ■
Routine contamination monitoring of oil systems with the icountLCM20 saves time and saves money.
Contamination monitoring is now possible while machinery is working – icountLCM20 saves on production downtime.
Instant, accurate results are available to international standards in hard copy form. That means system maintenance decisions can be taken immediately. icountLCM20 ensures that machinery hydraulic systems are tested in manufacture to ISO cleanliness standards.
Data entry allows individual equipment test log details to be recorded.
Data retrieval of test results from memory via hand set display.
Data graphing of up to 30 test results can be selected for integral printing.
Automatic test cycle logging of up to 300 tests can be selected via hand set display.
■ ■
■ ■
■ ■
■ ■
■ ■
■
■
■
■
User-friendly instrument improves familiarity and awareness of service and maintenance personnel.
Manufactured from lightweight Lexan expanded structural foam which is both durable and strong.
Totally portable, the icountLCM20 can be used as easily in the field as in the laboratory.
Computer interfacing available for downloading data on to compatible computer, through icountLCM20’s
RS232 serial port at 9600 baud.
An internal diagnostic feature ensures the icountLCM20 will always work accurately and reliably.
Supplied in an Astraboard carrying case.
Automatic calibration reminder.
5
Parker Hannifin icountLCM20
INtRoDuctIoN
Monitor – front view
On/Off switch
Printer cover
Handset
Display
Handset
Carrying strap
Limit switch socket
Main case
Battery module
Operation control valve
Function keys
Hydraulic hose
Monitor – rear view
Carrying strap
On/Off switch Printer cover g
Hose tidy
RS232 output port
(covered)
6
Parker Hannifin icountLCM20
INtRoDuctIoN
Handset
g
7
Function Keys
The functions assigned to each of these three keys is displayed on the bottom line of the alphanumeric display
Select option on right, step forward through linked screen displays, etc.
Select option on left, step back through linked screen displays, etc.
Parker Hannifin icountLCM20
Alphanumeric display
Paper feed
Alphanumeric keys
INtRoDuctIoN g
Powering the icountLCM20
The icountLCM20 may be powered via:
■ ■ a re-chargeable battery pack. This is supplied as an accessory with its battery charger – see
‘Ordering information’ on page 49.
■ ■ a replaceable battery pack. The battery compartment requires six LR20 ALKALINE batteries.
LR20 batteries are readily available and are also known as Type D or 13A batteries.
Charging the rechargeable battery pack
Step
1 Fully loosen the two retained screws and remove the rechargeable battery compartment.
Warning: Take care not to shortcircuit the terminal plates.
2 To charge the unit, attach the flying lead of the
12Vdc supply to the input power socket of the rechargeable battery pack. The jack plug should be tip positive.
Attach the mains plug to a mains socket and switch on.
The red LED on the rechargeable battery pack flashes to indicate that the unit is charging.
3 When fully charged, the red LED remains steadily on.
Disconnect the 12Vdc supply, replace the battery compartment and refit the two screws.
8
Note: When the Particle Counter displays a Low Battery Warning, you should recharge the batteries immediately. Battery packs in storage or fitted to units should be recharged every three months. The capacity of the battery pack is severely reduced if batteries are left in a discharged condition for long periods.
Parker Hannifin icountLCM20
INtRoDuctIoN
Using the replaceable battery pack
Step
1 Remove the battery compartment
2 Take note of the diagram on the underside of the battery pack. This shows which way up the batteries need to be in the pack (i.e. the polarity of the batteries)
3 Insert a complete set of fresh batteries in the sequence shown. The first two batteries are facing down: the next four batteries face up.
g
4
Replace the battery compartment
9
Parker Hannifin icountLCM20
INtRoDuctIoN
Installing paper and ribbon into the printer
Step
1
Unscrew the two retaining screws and remove the printer cover and reel axle.
g
2 Place the new or replacement ribbon cassette in the printer.
Press down until the cassette is sitting squarely, checking that the ribbon is not twisted.
3 Feed the end of the paper reel under the printer cassette and printer roller.
10
Parker Hannifin icountLCM20
4
INtRoDuctIoN
Press the paper feed button on the handheld unit until the paper appears through the printer’s paper feed slot.
5 Place the paper roll in the paper well. Check that the metal reel axle is correctly aligned in the slots.
6 Feed the paper through the the paper slot in the cover (using the paper feed button on the hand-held unit).
Refit the cover to the icountLCM20 by fastening the two screws.
g
11
Parker Hannifin icountLCM20
GEttING StARtED
Getting started
Setting the icountLCM20 to record the time and date of tests
Setting the time
g
Press Step
1 Switch on the icountLCM20 from the button on the top of the casing.
2
Remove the handset and check that the start-up screen is displayed.
3
4
5
6
7
Press and hold
–
MODE shown on the display. for five seconds.
RESET DATA Y/N
is
To clear all stored test results from the icountLCM20 memory, press
to move the cursor under Y and press
–
MODE
.
If you do not wish to clear the icountLCM20 memory, check that the cursor is under N (press
if it isn’t) and press
–
MODE
.
The time entry screen is displayed with the first digit shown blinking.
MODE PRINT
Change
/ Std
RESET DATA
Y/N?
1 :
TIME
Enter the correct time using the numeric keypad. Enter the time using the 24-hour clock. For example, to enter 9:30 am:
Press
0 9 : 1
TIME
Press .
When you have entered the last digit, the first digit is again shown blinking. If you make a mistake, simply key the time in again, overtyping the previous entry.
When the time is shown correctly, press
–
MODE
to accept it.
0 9 : 3 0
TIME
12
Parker Hannifin icountLCM20
GEttING StARtED
Setting the date
Step
1 When you have confirmed the time by pressing
–
MODE
, the date entry screen is displayed with the first digit shown blinking.
2
3
Enter the correct date using the numeric keypad. For example, to enter 21st August 2010:
Press
Press
4
Press
1 : :
DATE
21 : 0 :
DATE
21 : 08 : 0
DATE
21 : 08 : 10
DATE
5
When the correct date is displayed, press
–
MODE
.
g
13
Parker Hannifin icountLCM20
BASIc opERAtIoN g
Basic operation
the icountLcM20 is now ready to check the cleanliness of the hydraulic oil in your system.
You can take readings at full working pressure (max. 420 bar).
NOTE: The icountLCM20 is supplied filled with hydraulic oil and may need to be flushed prior to use.
Step
1 Disconnect the hydraulic hoses from the ‘hose tidy’.
2
3
The icountLCM20 is designed for use in connection with System 20 size 0, 1 and 2 Industrial Sensors or the Single Point Sampler. The sensors have the following dimensions:
Size
0
1
2
Diameter
30.0mm
41.0mm
66.7mm
NOTE: ‘Dual fluids monitors’ must be connected to a special System 20 Sensor or Single
Point Sampler. See page 45.
Ensure that the Sensor is installed with the arrow matching the direction of flow.
Working viscosity is 2–100 cSt. Ensure a minimum working pressure of 2 bar
Ensure adequate oil flow through the sensor:
Size
0
1
2
Flow rate
12 litres per minute
40 litres per minute
160 litres per minute
Unscrew protection caps 1 and 2 only.
14
Parker Hannifin icountLCM20
4
BASIc opERAtIoN
Connect red hose 1 loosely to the sensor inlet.
g
5
Connect yellow hose 2 loosely to the sensor outlet.
6 Simultaneously tighten both couplings finger-tight.
15
The icountLCM20 is now connected to the fluid to be checked.
NOTE: Parker Hannifin recommend that the icountLCM20 is connected to the operating system sensor for five minutes to allow the fluid condition to stabilise before starting a test.
We also recommended that you carry out an independent flow check if there is high viscosity oil and low flow through the System 20 sensor, or the unit is being operated without the trace heating option in cold ambient temperatures.
Parker Hannifin icountLCM20
BASIc opERAtIoN
Independent flow test
7
8
Step
1
2
3
Switch on the icountLCM20.
Press and hold for five seconds.
The flow test begins and the display shows:
4
5
6
After about 30 seconds the flow test is completed and the test results are displayed.
Check the ΔT (Delta T value – the change in temperature).
A ΔT of 03.60°C or less is required for a successful test.
If the ΔT is more than 03.60°C, then the flow rate through the System 20 sensor should be increased or the system oil temperature raised. Repeat from step 2.
Select ISO or NAS standard. This determines the format of the test data that is stored.
ISO -
MANUAL FLOW TEST
IN PROGRESS
Flowrate Test
Flowrate : Good
Delta T : 2.45C
Press MODE
to Continue
MODE PRINT
Change
/Std
ç
Check the display.
You can proceed if the valve turn symbol or is displayed.
Turn the valve 90° in the indicated direction. The valve position can be checked from the top of the monitor.…
… or from the front of the monitor.
IMporTanT: only turn the valve when starting a test and only when the valve turn symbol:
or is displayed.
testing commences immediately
Each test takes approximately two minutes. A progress bar is shown on the display indicating how far the test has progressed.
During each test, the appropriate test number is displayed. This example shows test number 021 in progress.
TEST NUMBER 021
[
± ± ± ± ± ± ± ± ± ± ± ±
]
ISO
MODE PRINT Change
/Std g
16
Parker Hannifin icountLCM20
BASIc opERAtIoN
ISO mode
NOTE: The following screens show MTD calibrated data. During all tests, an interim count is displayed.
g
Step
1
When the test is complete, the ISO code is automatically displayed.
22 : 20 : 14
ISO
2 particle count
All particle counts are indicated per 100 ml sample.
To display a count for the 4µ particles press
–
MODE
.
>4
µ
3642544
ISO
3
Each press of
displays the particle counts for 6µ, 14µ, 21µ,
38µ and ≥70µ particle sizes respectively.
>70
0
µ
ISO
4
Press to display the particle count for a previous class.
>38
µ
11
ISO
17
Parker Hannifin icountLCM20
BASIc opERAtIoN
NAS mode
When the test is complete, the NAS code is automatically displayed.
Step
1
Press
–
Change/Std
to choose NAS mode. The NAS class code is displayed on the screen.
11
2
3
4
NAS particle count
All particle counts are based on 100 ml samples.
To display the NAS count for the 4µ to 6µ particles, press
–
MODE
.
4
µ
-6
µ
2774621
NAS
For MtD, each press of
displays the particle counts for 6–14µ,
4–21µ, 21–38µ, 38–70µ, and ≥70µ particle sizes respectively.
(AcFtD displays the particle counts for 5–15µ, 15–25µ, 25–50µ,
50–100µ and ≥100µ particle sizes respectively.)
>70
0
µ
Press to display the particle count for a previous class.
38
µ
-70
µ
11
NAS
NAS g
18
Parker Hannifin icountLCM20
BASIc opERAtIoN
SAE mode
When the test is complete, the SAE (AS4059) code is automatically displayed.
Step
1
Press
–
Change/Std
to choose SAE mode. The AS4059 class code is displayed on the screen.
2
3
4 particle count
All particle counts are based on 100 ml samples.
To display the AS4059 count for the 4µ particles, press
–
MODE
.
For MtD, each press of
displays the particle counts for 6µ,
14µ, 21µ, 38µ and ≥70µ particle sizes respectively.
(AcFtD displays the particle counts for 5µ, 15µ, 25µ, 50µ and
≥100µ particle sizes respectively.)
Press to display the particle count for a previous class.
11A 11B 8C 6D
AS4059
>4
µ
3642522
AS4059
>70
0
µ
AS4059
>38
11
µ
AS4059 g
19
Parker Hannifin icountLCM20
BASIc opERAtIoN
GOST mode
When the test is complete, the GOST code is automatically displayed.
Step
1
Press
–
Change/Std
to choose GOST mode. The GOST class code is displayed on the screen.
2
3
11
GOST particle count
All particle counts are based on 100 ml samples.
To display the GOST count for the 5µ to 10µ particles, press
–
MODE
.
5
µ
-10
µ
180319
Each press of
displays the particle counts for 10–25µ, 25–50µ,
50–100µ and >100µ particle sizes respectively.
>100
11
µ
GOST
GOST
4
Press to display the particle count for a previous class.
50
µ
-100
µ
440
GOST g
20
Parker Hannifin icountLCM20
5
6
BASIc opERAtIoN
Entering a test identification code
You can enter a test identification code of up to 32 characters to go with a test number.
Step
1
2
Press
–
MODE
three times from idle to step through ISO/NAS and
Count modes to the Serial Number Entry mode.
PLEASE ENTER A TEST ID
CODE
is displayed.
Each key on the keypad has four characters associated with it (similar to a mobile phone). Press each key repeatedly, up to four times to obtain the desired character. For example:
PLEASE ENTER A
TEST ID CODE
1
Press
A
Press
again
B
Press
again
C
Press
again
3
4
Press another key to enter the next character.
To input two characters using the same key, press character. For example to enter 11:
between each
Press
1 then press
1 then press
again
1 1
When you have completed entering the identification code, turn the valve control knob.
The code is entered into the icountLCM20’s memory and the test begins.
The previously entered identification code is displayed automatically at the beginning of each new test.
If you need the identification code to be unique for each test, simply start typing the new code. The current code is cleared as soon as you enter the first character.
Press to delete the character just entered if you make a mistake.
the identification code is printed with the test number on each test printout.
g
21
Parker Hannifin icountLCM20
BASIc opERAtIoN g
Bar code wand
The bar code wand is available as an optional extra (although it is standard on icountLCM20.2023). It allows you to enter the 32-character serial identification or test data against the test number from a standard 39 bar code. Consult Parker if you require further information.
Step
1
2
3
Connect the bar code wand into the socket .
Press
–
MODE
three times from idle to step through ISO/NAS and
Count modes to the Serial Number Entry mode.
PLEASE ENTER A
TEST ID CODE
is displayed.
Place the bar code wand to the left of the bar code, and with a continuous motion, swipe the wand across the bar. A successful entry is confirmed by a single beep.
NOTE: If you wish to add data manually press
again, or wait 50 seconds until the cursor appears in the top left hand corner
–
MODE
PLEASE ENTER A
TEST ID CODE
4 Once the identification code is displayed, turn the valve control knob.
The code is entered into the icountLCM20’s memory and the test begins.
5
The identification code is printed with the test number on each test printout.
22
Parker Hannifin icountLCM20
BASIc opERAtIoN
Printing test results
Step
1
2
Press
–
once to print out the most recent result.
The printer prints the test number, unit number, date and time, with the appropriate standards code (ISO, NAS or SAE (AS4059)), followed by the particle counts for each range of particle size.
to print the results of all tests completed to date:
Press the print button twice
–
–
Press
–
a third time to cancel printing.
Example test printouts (for MTD)
ISO mode NAS mode
g
SAE mode GOST mode
23
NOTE: If the test count is outside the range for a particular standard, then ‘00’
(below range) or ‘99’ (above range) is displayed.
Parker Hannifin icountLCM20
BASIc opERAtIoN g
Viewing and printing a stored test by ID code or test number
6
7
Step
1
2
Press and hold
for two seconds
Press to select
Y
3
4
Press
–
MODE
Choose
S
to select a test by serial number, or
T
to select a test by test number.
Press to select
T
.
5
Press
–
MODE
DATA RETRIEVAL-
Y/N?
DATA RETRIEVAL-
Y/N?
SERIAL OR TEST
NUMBER- S/T?
SERIAL OR TEST
NUMBER- S/T?
DATA RETRIEVAL-
ENTER TEST NO ID
Enter the test number required.
Press
–
MODE
to view the test.
Press
–
to print the test (test 012 in this example).
To print test 012 through to test 001, press the print button again during printing of test 012
RANGE IS 001-025
012
>4
µ
(c)
8548760
8
Press
–
MODE
To view and print more results select
Y
. Select
N
to return to the idle screen.
RETRIEVE MORE
RESULTS Y/N?
24
Parker Hannifin icountLCM20
BASIc opERAtIoN
Printing a stored test as a graph
to print a range of stored test results as a graph by ID code or test number:
Step
1
2
Press and hold
for two seconds
Press to select
Y
Press
–
MODE
TREND PRINTOUT
Y/N?
SERIAL OR TEST
NUMBER- S/T?
3
Choose
S
to select a test by serial number, or
T
to select a test by test number.
Press to select
T
.
SERIAL OR TEST
NUMBER- S/T?
4
5
6
Press
–
MODE
Enter the first test number required.
Press
–
MODE
Enter the last test number required.
Press
–
MODE
to start the graph printout.
ENTER FIRST TEST
NUMBER:
ENTER LAST TEST
NUMBER:
GENERATING TREND
PLEASE WAIT . . .
7 a. Graphs print down the page rather than across. b. Adjacent points on the graph are joined by a continuous line. c. Printouts for 6µ(c) and 14µ(c) in ISO show separate graphs for each. d. Printouts in NAS show one graph. e. A maximum of 30 tests may be printed as graphs from one request.
Once completed, the following message is shown:
To print more graphs, select
Y
. To return to the idle screen select
N
.
PRINT MORETRENDS
Y/N?
g
25
Parker Hannifin icountLCM20
BASIc opERAtIoN g
Alarm levels
The icountLCM20 includes a built-in limit switch relay which can be triggered when a preset alarm level is reached. The relay contacts can be used to switch on or off an external device connected to the limit switch socket. The socket connections are shown in the following wiring diagram:
N/O
1
3
N/C
2
The contact rating is 5A at 220V ac or dc.
NOTE: the user is responsible for selecting an appropriately rated connector and cable for the externally controlled device.
3
4
Step
1
2
Editing alarm levels (MTD)
Alarm levels are set individually for 4µ, 6µ and 14µ particle sizes.
Press and hold for two seconds.
Press to select
Y
EDIT ALARM LEVEL
Y/N?
EDIT ALARM LEVEL
Y/N?
Enter the code level for 4µ above which the alarm is to be triggered.
Press
–
MODE
to repeat this process for 4µ und 14µ.
When all three levels have been entered, the display shows:
4
µ
LIMIT
21
LIMIT LEVELS SET
1
2
3
Activating alarm level testing
When activated, if any of the pre-set alarm levels are exceeded during a manual or automatic test, the alarm relay is triggered and any externally connected device is activated.
Press and hold for two seconds.
LEVEL TESTING
Y/N?
Press to select
Y
LEVEL TESTING
Y/N?
Press
–
MODE
.
Press to select Y. The unit will run indefinitely whether the alarm has been triggered or not.
Press
–
MODE
to select N. The unit will stop running tests when the alarm is triggered.
CONTINUOUS
Y/N?
26
Parker Hannifin icountLCM20
BASIc opERAtIoN g
3
Step
1
2
Operating the limit switch relay manually
You can manually operate the limit switch relay for test purposes, for example before running a test.
Press and hold for two seconds
Press to select
Y
LIMIT SWITCH
RELAY ON Y/N?
LIMIT SWITCH
RELAY ON Y/N?
Press
–
MODE
.
The limit switch relay is energised (i.e. contact is made between pins 1 and 2 of the level test relay socket).
Choose
N
in step 2 to turn off the relay.
27
Parker Hannifin icountLCM20
BASIc opERAtIoN g
Automatic testing
The automatic testing function is used to run a series of tests at predefined time intervals. Note that automatic testing cannot be used when the icountLCM20 is connected to the Offline Bottle Sampler.
Step
1
AUTOMATIC
TESTING Y/N?
2
4
5
6
7
8
9
Press and hold for two seconds.
Press to select
Y
Press
–
MODE
.
Enter the start time required.
Press
–
MODE
.
Enter the test interval (the number of minutes between each test).
Press
–
MODE
.
Enter the number of tests to be run.
Press
–
MODE
.
Press to select
Y
if a printout of each test result is required.
Press
–
MODE
.
Press to select
Y
to start testing.
Press
–
MODE
.
Testing starts when the two times become equal.
This screen is displayed just before the test takes place.
ENTER TEST START
TIME: 13:48
TEST INTERVAL
(6-999 min):
ENTER NUMBER OF
TESTS TO RUN:
PRINT RESULTS
Y/N?
START TESTING
Y/N?
TIME NOW: 13:42
START AT: 13:45
TEST NUMBER 020
FLUSHING
This is followed immediately by the test. Interim counts are displayed as the test progresses.
TEST NUMBER 020
COUNTING
10
Press
–
MODE
to abort the current test cycle.
WAITING . . .
MODE TO ABORT
When all tests are complete the following screen is displayed.
AUTOMATIC
TESTING COMPLETE
11
Press
–
MODE
to return to the idle screen.
28
Parker Hannifin icountLCM20
BASIc opERAtIoN g
Calibration
When switching the unit on, it will check the date as stored in the real-time clock against the date stored as the next calibration due date.
If the date is within four weeks of the calibration due date, then the following message is displayed when the unit is switched on:
CALIBRATION DUE
20-09-2011
(The example above is for a calibration due date of 20th September 2011)
To ignore the message for now and continue with testing press
–
MODE
If the unit passes the calibration due date, the following message is displayed when the unit is switched on:
CALIBRATION
OVERDUE
To ignore the message for now and continue with testing press
–
MODE
Adjusting screen brightness
The icountLCM20 is supplied with factory default settings for backlight and brightness settings.
These settings can be altered using hot key functions as follows:
Press and hold and press to brighten the screen
Press and hold and press to darken the screen
Press and hold and press to switch backlighting on
Press and hold and press to switch backlighting off
NOTE: When the icountLCM20 is switched off, the backlight function is reset. If you still require the backlight when the icountLCM20 is switched on again, it will need to be reactivated.
Saving screen brightness settings
Once the screen brightness has been set to the required level, you can save the setting by pressing and holding
and
simultaneously and following the on-screen instructions.
Additional (optional) features
When the icountLCM20 is used with the DATµm software, the unit can perform additional functions including:
■
Automatic testing
■
Route mapping.
29
Parker Hannifin icountLCM20
REFERENcE
Reference
Data interpretation
Solid contaminants in fluid power systems vary in size, shape, form and quantity. The most harmful contaminants are normally between 6 microns and 14 microns. The
ISO code is the preferred method of reporting quantity of contaminants.
The ISO code number corresponds to contamination levels pertaining to three sizes.
The first scale number represents the number of particles larger than 4µ(c) per 100 millilitres of fluid, the second number for particles larger than 6µ(c) per 100 millilitres of fluid and the third number for particles larger than 14µ(c) per 100 millilitre of fluid.
Below is a table of actual results obtained, of contamination within a hydraulic pump endurance test rig.
particle size
4µ
6µ
14µ
21µ
38µ
No. of particles per 100ml oil
7950100
280500
1500
700
150
ISO-Code: 23/19/11 g
10 7
1,6
8
4
2
10 6
10
10
5
4
5
2,5
1,3
6,4
3,2
1,6
8
4
2
10 3
10
10
2
5
2,5
1,3
6,4
3,2
1,6
8
4
2
1
2
24
19
6
Particle size, µm
14
Note that interpolation (i.e. estimation within the measured range) is acceptable; extrapolation (i.e. estimation outside of the measured range) is not.
17
16
15
14
13
12
24
23
22
21
20
19
18
11
10
9
8
7
21
1
3
2
6
5
4
30
Parker Hannifin icountLCM20
10 7
8
6
4
2
10 4
8
6
4
2
10 3
8
6
4
2
10 6
8
6
4
2
10
5
8
6
4
2
10 2
8
6
4
2
10
8
6
4
2
1
REFERENcE
ISO4406 particle distribution chart
Including various ISO level contamination grades
6
9
8
7
6
5
4
3
2
1
14
13
12
11
10
ISO No.
23
22
21
20
19
18
17
16
15
14 particle size µm
21
ISO 18/15
ISO 17/14
ISO 16/13
ISO 15/12
ISO 14/11
ISO 13/10
38
10
5
10 4
3
10
10 2
10 1
1
10 -1 g
31
Parker Hannifin icountLCM20
REFERENcE
18
19
20
21
22
23
24
12
13
14
15
16
17
8
9
6
7
10
11
3
4
1
2
5
ISO contamination numbers
Range number
64
130
250
500
1 × 10
3
2 × 10
3
4 × 10
3
8 × 10
3
16 × 10
3
32 × 10
3
64 × 10
3
130 × 10
3
250 × 10
3
500 × 10
3
1 × 10
6
2 × 10
6
4 × 10
6
8 × 10
6
1
2
4
8
16
32
Number of particles per 100ml
More than up to and including
2
4
8
16
32
64
130
250
500
1 × 10
3
2 × 10
3
4 × 10
3
8 × 10
3
16 × 10
3
32 × 10
3
64 × 10
3
130 × 10
3
250 × 10
3
500 × 10
3
1 × 10
6
2 × 10
6
4 × 10
6
8 × 10
6
16 × 10
6 g
For example: code 20/18/13 indicates that there are between 500,000 and 1,000,000 particles larger than 2 microns, and between 130,000 and 250,000 particles larger than 5 microns, and between 4000 and 8000 particles larger than 15 microns.
32
Parker Hannifin icountLCM20
REFERENcE
NAS 1638 chart
Size range
00
0
1
4
5
2
3
6
7
10
11
8
9
12 g
5–15µm
125
250
500
1000
2000
4000
8000
16,000
32,000
64,000
128,000
256,000
512,000
1,024,000
15–25µm
22
44
89
178
356
712
1425
2850
5700
11,400
22,800
45,600
91,000
182,400
25–50µm
4
8
16
32
63
126
253
506
1012
2025
4050
8100
16,200
32,400
50–100µm
1
2
3
6
11
22
45
90
180
360
720
1440
2880
5760
8
16
2
4
32
64
>100µm
0
0
1
1
128
256
512
1024
33
Parker Hannifin icountLCM20
REFERENcE
SAE (AS4059 rev E) chart
7
8
5
6
3
4
1
2
MtD
AcFtD
Size code
000
00
0
9
10
11
12
>4µm(c)
>2µm
A
195
390
780
1560
3120
6250
12,500
25,000
50,000
100,000
200,000
400,000
800,000
160,000
320,000
1220
2430
4860
9730
19,500
38,900
77,900
156,000
Maximum contamination limits (particles/ml)
>6µm(c)
>5µm
B
76
152
304
609
>14µm(c)
>15µm c
14
27
54
109
>21µm(c)
>25µm
D
3
5
10
20
311,000
623,000
1,250,000
217
432
864
1730
3460
6920
13,900
27,700
55,400
111,000
222,000
39
76
152
306
612
1220
2450
4900
9800
19,600
39,200
106
212
424
848
7
13
26
53
>38µm(c)
>50µm
E
1
1
2
4
1700
3390
6780 g
18
32
64
128
4
8
1
2
>70µm(c)
>100µm
F
0
0
0
1
256
512
1020
MTD:
ACFTD:
ISO11171 (Calibration or optical microscope count – particle size based on projected area equivalent diameter)
ISO4402 (Calibration or optical microscope count – particle size based on longest dimension)
34
Parker Hannifin icountLCM20
REFERENcE
GOST 17216-2001 chart
Size range
00
0
1
4
5
2
3
6
7
13
14
15
16
17
10
11
8
9
12
5–10µm
8
16
32
63
125
250
500
1000
2000
4000
8000
16000
31500
63000
–
–
–
–
– g
10–25µm
4
8
16
32
63
125
250
500
1000
2000
4000
8000
16000
31500
63000
125000
–
–
–
50–100µm
0
0
0
1
4
6
2
3
12
25
50
100
200
400
800
1600
3150
6300
125000
25–50µm
1
2
3
4
8
12
25
50
100
200
400
800
1600
3150
6300
12500
25000
50000
–
1
2
0
0
4
6
>100µm
0
0
0
0
400
800
1600
3150
12
25
50
100
200
The GOST standard is developed by the Technical Committee of Standardization TK 184 “Ensuring
Industrial Cleanliness” introduced by the Government of Russia.
Adopted by the Inter-governmental Committee of Standardization, Metrology and Certification
(Protocol No. 19, dated 24 May 2001).
35
Parker Hannifin icountLCM20
REFERENcE
ISO/NAS/SAE comparison chart
BS 5540/4
ISo/DIS 4406 code
11/8
12/9
13/10
14/9
14/11
15/9
15/10
18/12
18/13
18/15
19/13
19/16
20/13
15/12
16/10
16/11
16/13
17/11
17/14
20/17
21/14
21/18
22/15
23/17
Def. Std 05/42 table A
2000
–
–
4400
–
6300
–
800
–
–
1300
–
–
15,000
–
21,000
100,000
–
–
–
–
–
400
– table B
–
–
–
400F
–
–
800F
–
–
1300F
–
2000
–
–
4400F
–
6300F
–
–
–
–
–
–
–
SAE 749 class
6
–
–
–
–
–
–
–
–
–
–
–
4
3
–
–
5
2
–
1
–
–
–
0
NAS 1638 class
9
–
–
–
10
–
11
–
12
–
–
–
7
6
–
–
8
5
–
4
–
–
2
3 g
The above comparisons relate to particle count data only. To conform to any particular standard, reference should be made to the recommended experimental procedure.
36
Parker Hannifin icountLCM20
REFERENcE
ISO contamination charts
g
Typical system applications and code numbers
These typical applications and ISO code numbers are taken from the UK Contamination and Control
Research Programme (1980–1984).
Ref. AHEM Guide to Contamination Control in Hydraulic Power Systems – 1985
SOLID CONTAMINANT CODE NO 15/9
Application: Machine tools
10 5
10 4
8
6
4
2
10 3
8
6
4
2
10
2
8
6
4
2
10
8
6
4
2
10 7
8
6
4
10 6
8
6
4
2
10 5
8
6
4
2
2
1 6
ISO No.
23
22
21
20
19
18
17
5
4
3
2
1
14 particle size µm
21 38
10
4
10 3
10 2
10 1
1
10 -1
SOLID CONTAMINANT CODE NO 13/10
Application: Aircraft test stands
10
5
10 3
8
6
4
2
10 2
8
6
4
2
10
8
6
4
2
2
10 7
8
6
4
10 6
8
6
4
2
10 5
8
6
4
2
10
4
8
6
4
2
1 6
ISO No.
19
18
17
16
15
23
22
21
20
5
4
3
2
1
7
6
14 particle size µm
21 38
10 4
10 3
10 2
10 1
1
10 -1
37
Parker Hannifin icountLCM20
REFERENcE
SOLID CONTAMINANT CODE NO 18/11
Application: Mobile systems
SOLID CONTAMINANT CODE NO 17/12
Application: Marine installations
10
5
8
6
4
2
10 4
8
6
4
2
10 3
8
6
4
2
10 2
8
6
4
2
10 7
8
6
4
10 6
8
6
4
2
10
8
6
4
2
2
1
10 7
8
6
4
10 6
8
6
4
2
10
5
8
6
4
2
10
4
8
6
4
2
10 3
8
6
4
2
10 2
8
6
4
2
10
8
6
4
2
2
1
6
6
ISO No.
23
22
21
20
4
3
2
1
7
6
5
14 particle size µm
21
ISO No.
23
22
21
20
19
8
7
6
5
4
3
2
1
14 particle size µm
21
38
10 5
10 4
10 3
10 2
10
1
1
10 -1
38
10 4
10 3
10 2
10 1
1
10 -1 g
10 5
38
Parker Hannifin icountLCM20
REFERENcE
SOLID CONTAMINANT CODE NO 18/13
Application: Mechanical handling
SOLID CONTAMINANT CODE NO 16/11
Application: Injection moulding;
Metalworking;
Unused commercial-grade oil
10
8
6
4
2
2
10 4
8
6
4
2
10 3
8
6
4
2
10 2
8
6
4
2
10 7
8
6
4
10
6
8
6
4
2
10 5
8
6
4
2
1 6
10 5
8
6
4
2
10 4
8
6
4
2
10 3
8
6
4
2
10 2
8
6
4
2
10 7
8
6
4
10 6
8
6
4
2
10
8
6
4
2
2
1 6
ISO No.
23
22
21
ISO No.
23
22
21
20
19
10
9
8
7
4
3
2
1
6
5
14 particle size µm
21
4
3
2
1
6
5
8
7
14 particle size µm
21
38
10 5
10 4
10 3
10 2
10 1
1
10
-1
38
10 4
10 3
10 2
10 1
1
10 -1 g
10 5
39
Parker Hannifin icountLCM20
REFERENcE
Component cleaning guidelines
The following table gives suggested acceptable contamination levels for various hydraulic systems.
g
Sensitivity type of system typical components
Super critical
Critical
Very important
Silt sensitive control system with very high reliability.
Laboratory or aerospace.
High performance servo and high pressure long life systems, e.g. aircraft, machine tools etc..
High quality reliable systems. General machine requirements.
High performance servovalves
Industrial servovalves
Piston pumps, proportional valves, compensated flow controls
Vane pumps, spool valves Important General machinery and mobile systems. Medium pressure, medium capacity.
Average Low pressure heavy industrial systems, or applications where long life is not critical.
Main protection
Low pressure systems with large clearances.
Gear pumps, manual and poppet valves, cylinders
Ram pumps target contamination class to ISo 4406
6µm
13
14µm
9
Suggested maximum particle level
6µm
4,000
14µm
250
15
16
18
19
21
11
13
14
15
17
16,000
32,000
130,000
250,000
1,000,000
1,000
4,000
8,000
16,000
64,000
40
Parker Hannifin icountLCM20
REFERENcE
Fluid flow schematic
1
Monitor block
2
Laser diode
3
Optical scanner
4
Changeover valve
5
Dual direction syringe pump
6
Flow probe device
P2
System 20 sensor
P1
Direction of flow
4
6
3
1
2
5
icountLCM20
Logic diagram
Optical scanner
Area calculator
Unobstructed light calculator
Reference voltage generator
≥ 4µ comparator
Counter
≥ 6µ comparator
Counter
≥ 14µ comparator
Counter
≥ 21µ comparator
Counter
≥ 38µ comparator
Counter
≥ 70µ comparator
Counter
Display
Count-to-code converter
Switch
Memory PC g
41
Parker Hannifin icountLCM20
REFERENcE
Peripherals and serial interface
Back panel
12V DC
FUSE
Fast blow fuse 1.25A
RS 232 g
9-way connector to RS232 output to download all test data stored in the icountLCM20 memory
(see DATµm software package for data management information).
Measurement accuracy
CALIBRATION
Every monitor is individually calibrated using an accurately determined gravimetric level of test dust media.
Each channel within the instrument is adjusted to read the number of particles in a prescribed size range as specified in ISO procedures, thus guaranteeing calibration accuracy.
REPEATABILITy
The icountLCM20 measures and quantifies individual particles using advanced electronics and laser technology to ensure a high degree of repeatability.
Repeatability is typically better than 5%.
42
Parker Hannifin icountLCM20
REFERENcE g
Diagnostic codes
Explanations of icountLCM20 diagnostic codes and possible remedies are given in the following table.
DIAGNOSTIC CODE 1
a. At beginning of test
DIAG 1A. LIGHT
SOURCE DEVIATION b. at the end of the test
DIAG 1B. LIGHT
SOURCE DEVIATION
Oil is too dark or it is cloudy.
Unstable fluid opacity may be caused by aeration, water sludge or an amount of cold oil passing through icountLCM20.
Check sample of oil visually. This can be done as follows:
Dark oils: Wet your thumb and forefinger in the oil and press together. Release and look at your thumb.
If you can see through the film of oil it should work in the icountLCM20. If you cannot, then you may have problems. (This is most common with engine oils or very highly contaminated oils above ISO 24).
Emulsions: Put sample in a clear container and hold up to the light. This will show cloudy or clear. If cloudy, check the type of oil and change until the oil is clean, then retry icountLCM20.
Allow machine to work up to operating temperature before performing condition monitoring. Run tests with a stable system and ensure that a minimum line pressure of 2 bar is available at the monitor to reduce the possibility of aeration.
DIAGNOSTIC CODE 2
The changeover valve and syringe pump are out of phase
DIAG 2. VALVE
OPERATION ERROR a. Control knob turned, either before monitor switched on, before valve symbol displayed on handset or during a test.
b. Time taken to turn valve fully to next position is too long (20 seconds).
c. Microswitch setting fault.
Switch off the monitor, then switch on and wait for monitor to reset its position (diagnostic screen 6 is displayed).
Return to Parker for repair.
DIAGNOSTIC CODE 3
Uncontrolled power down Wait for monitor to reset itself.
DIAG 3. POWER
INTERRUPTION a. Power to monitor has been disrupted by removing power supply without switching monitor off.
b. Battery power too low. Battery level warning ignored.
c. Battery contact disconnected by excessive vibration.
d. External power supply connected (12Vdc) while unit is on.
Replace batteries with 6 x 1.5V Alkaline D Cells (or recharge battery pack if a rechargeable pack is fitted).
Relocate monitor on a sound surface (May also be hung from a carrying strap).
Leave unit to re-set itself.
DIAGNOSTIC CODE 4
Insufficient flow rate of oil from P1 hose into monitor block to fill syringe pump.
Results are suspect and not made available.
DIAG 4. LOW FLOW
IN BYPASS LINE a. Inadequate differential pressure across P1 and
P2 connections to provide sufficient bypass flow.
b. Air lock in monitor block or high viscosity slug of oil in bypass hoses.
i) Select smaller sized sensor ii) Reduce viscosity of system oil by increasing test temperature where possible.
iii) Use Single Point sampler connected to P1. Purge the system pressure with P2 hose disconnected from system.
Care should be taken to allow oil to discharge safely and should only be performed by a competent operator.
Re-test and if fault reoccurs, return monitor to Parker for repair.
43
Parker Hannifin icountLCM20
REFERENcE
DIAGNOSTIC CODE 5
Test time too short or too long.
Results are suspect and are not made available.
a. Malfunction of Opto-Tacho control, causing flow to stop before particle counting phase completed. Pump drive slipping or failed.
b. Differential pressure too high due to lack of control of flow through icountLCM20.
DIAG 5A. TEST
TIME TOO SHORT
Use SPS or sensor to control flow through icountLCM20.
DIAG 5B. TEST
TIME TOO SHORT
DIAGNOSTIC CODE 6
Unit trying to reset from last error.
Displayed after switching on, while monitor is resetting itself from the previous error condition.
Leave the icountLCM20 alone until it has reset.
If it does not reset, or switches itself off, contact
Parker.
DIAG 6. LCM IN
RESET MODE
DIAGNOSTIC CODE 7 AND ABOVE
All faults which can only be rectified by Parker and are normally software related.
DIAG 7. REFER
TO LCM20 SUPPLIER g
DIAG 8. REFER
TO LCM20 SUPPLIER
DIAG 9. REFER
TO LCM20 SUPPLIER
DIAG 10. LASER
TEMP TOO HIGH
LOw BATTERy
Displayed if the monitor block has reached a temperature above 60°C.
Remove icountLCM20 from system connections and allow to cool down.
If unit does not reset, contact Parker.
Replace or recharge batteries.
The icountLCM20 will not complete a measurement if power is insufficient.
é
ISO
44
Parker Hannifin icountLCM20
REFERENcE g
Technical specifications
Construction
Mechanical components
Seals
Hoses
Hose length
Flow rate
Case: Lexan structural foam and ABS
Hand-held display: ABS
Keypad: fluorosilicone rubber
Carrying case: Astraboard
Brass, plated steel, stainless steel and aluminium
Viton
Nylon (Kevlar braided microbore). Stainless steel armoured ends
Fluid connection hose: 1.2 metres (1 metre extensions can be used)
Hand-held display cable length: 1 metre
Up to 400 l/min (System 20 sensors)
Higher with Single Point Sampler – consult Parker
Up to 420 bar (System 20 sensors)
Mineral oil and petroleum based fluids. For other fluids consult Parker
Max. working pressure
Fluid compatibility
Power Replaceable Battery Pack: requires 6 x 1.5V D cells.
Rechargeable Battery Pack: requires transformer supply voltage of 9–12Vdc at 1.5A,
9mm jack plug with positive centre connection:
– +
Fuse
LCM20 technology
1.25A fast blow fuse included for overload protection (spare supplied)
Unique optical scanning system
Size, measurement & ranges MtD: 4µ+, 6µ+, 14µ+, 21µ+, 38µ+ and 70µ+
AcFtD: 2µ+, 5µ+, 15µ+, 25µ+, 50µ+ and 100µ+
Analysis range
Calibration
Repeatability/Accuracy
ISO 7 to 22 inclusive; NAS 0 to 12 inclusive; SAE 0 to 12 inclusive
Each unit is individually tested and calibrated in accordance with ISO procedures
Better than 5% (typical)
Viscosity range
Oil temp. range
Operating temp. range
Test completion time
Computer Interface
Weight
Commissioning kit
2–100 centistokes (500cSt with SPS)
+5°C to +80°C
+5°C to +40°C
2 minutes
RS232 at 9600 baud icountLCM20: 8kg, Carrying case: 5kg
6-off D cell batteries; 1-off print rolls (shrink wrapped); 2-off printer ribbons;
DATµm software plus cable; weather protector cover; 12Vdc power supply;
Rechargeable battery pack
45
Parker Hannifin icountLCM20
REFERENcE
Installation dimensions
g
(All dimensions are mm)
46
Parker Hannifin icountLCM20
REFERENcE g
Dual fluids monitor - icountLCM202061
To expand further the applications possible with Parker’s icountLCM20 particle counting technology,
Parker now offers a version of the icountLCM20, using Perfluoroelastomer seal technology, which is the only sealing elastomer universally accepted by the civil aviation hydraulic manufacturers.
Calibrated in accordance with ISO procedures, the icountLCM202061 offers the user all the features and benefits of the thoroughly proven icountLCM20 technology accepted throughout the world.
Using the icountLCM202061, civil aviation hydraulic servicing operations can now match the cost and time saving already demonstrated in the military aviation industries.
FEATURES:
■ ■
Perfluoroelastomer seals throughout
■ ■
■ ■
■ ■
■ ■
■ ■
■ ■
Totally compatible with aggressive phosphate esters (e.g. SKYDROL LD4/500B), as well as mineral or synthetic-based fluids.
Red control valve knob and handset keys – the icountLCM202061 is easily distinguished from standard icountLCM20 units
High Technology, fused optical windows for high particle definition
Available with complete range of sample extraction options (e.g. System 20 Sensors or the
Single Point Sampler)
5/8” BSF HSP Hose fitting
Parylene treated for additional protection.
WarnInG: any oil splashed onto this product should be removed immediately, to avoid long-term damage.
NOTE: If you have doubts about fluid/seal compatibility, consult Parker.
47
Parker Hannifin icountLCM20
REFERENcE
Operation checklist
Always ensure:
■ ■
Sensors are installed correctly
■
■ ■
■
Sensor connectors are correctly tightened there is adequate oil flow
■
■
■
■ ■
■
■
■
Steady state pressure conditions oil viscosity is within working range trend monitoring is performed under similar working conditions correctly stowed hoses to avoid fluid spillage
■
■
■
■ ■
■
■
■
Sufficient paper supplies for hard copy printer icountLcM20 is handled with care – it is an instrumentation product
Spares are reordered in advance
the icountLcM20 is calibrated according to parker recommendations.
g
48
Parker Hannifin icountLCM20
REFERENcE
Ordering information
part number
LCM202021/2023
LCM202022/2024
ACC6NE015
ACC6NE014
ACC6ND000
SPS2021
ACC6NE013
ACC6ND002
ACC6NE008
ACC6NE009
ACC6NE010
Description
LCM 6 channel (including carrying case and kit). ACFTD calibrated
LCM 6 channel MTD calibrated
Printer paper (to suit paper reel 44mm wide x 45mm diameter). 5 rolls
Printer ribbon
Cable link package
Single Point Sampler
Rechargeable battery pack
Weather protector cover
UK power supply
Euro power supply
US power supply
System 20 sensors
INDUSTRIAL SENSORS – SIzES 0, 1 AND 2
part number
STI.0144.100
STI.0344.100
STI.1144.100
STI.1344.100
STI.2144.100
STI.2344.100
2
2
1
1
Size
0
0
Flow range
6–25 l/min
0.5–7 US GPM
20–100 l/min
5–26 US GPM
80–380 l/min
21–100 US GPM thread
G
3
8
–
–
3
4
UNF
G
3
4
–
SAE 1
–
– 12UN-2B
G1
1
4
–
SAE 1
5
8
–
– 12UN-2B
Qty
Qty g
49
Parker Hannifin icountLCM20
Parker Worldwide
AE – uAE, Dubai
Tel: +971 4 8875600 [email protected]
AR – Argentina, Buenos Aires
Tel: +54 3327 44 4129
At – Austria, Wiener Neustadt
Tel: +43 (0)2622 23501-0 [email protected]
At – Eastern Europe,
Wiener Neustadt
Tel: +43 (0)2622 23501 970 [email protected]
Au – Australia, Castle Hill
Tel: +61 (0)2-9634 7777
AZ – Azerbaijan, Baku
Tel: +994 50 2233 458 [email protected]
BE/Lu – Belgium, Nivelles
Tel: +32 (0)67 280 900 [email protected]
BR – Brazil, Cachoeirinha RS
Tel: +55 51 3470 9144
BY – Belarus, Minsk
Tel: +375 17 209 9399 [email protected]
cA – canada, Milton, Ontario
Tel: +1 905 693 3000 cH – Switzerland, Etoy
Tel: +41 (0) 21 821 02 30 [email protected]
cN – china, Shanghai
Tel: +86 21 5031 2525 cZ – czech Republic, Klecany
Tel: +420 284 083 111 [email protected]
DE – Germany, Kaarst
Tel: +49 (0)2131 4016 0 [email protected]
DK – Denmark, Ballerup
Tel: +45 43 56 04 00 [email protected]
ES – Spain, Madrid
Tel: +34 902 33 00 01 [email protected]
FI – Finland, Vantaa
Tel: +358 (0)20 753 2500 [email protected]
FR – France, Contamine s/Arve
Tel: +33 (0)4 50 25 80 25 [email protected]
GR – Greece, Athens
Tel: +30 210 933 6450 [email protected]
HK – Hong Kong
Tel: +852 2428 8008
Hu – Hungary, Budapest
Tel: +36 1 220 4155 [email protected]
IE – Ireland, Dublin
Tel: +353 (0)1 466 6370 [email protected]
IN – India, Mumbai
Tel: +91 22 6513 7081-85
It – Italy, Corsico (MI)
Tel: +39 02 45 19 21 [email protected]
Jp – Japan, Fujisawa
Tel: +(81) 4 6635 3050
KR – South Korea, Seoul
Tel: +82 2 559 0400
KZ – Kazakhstan, Almaty
Tel: +7 7272 505 800 [email protected]
LV – Latvia, Riga
Tel: +371 6 745 2601 [email protected]
MX – Mexico, Apodaca
Tel: +52 81 8156 6000
MY – Malaysia, Subang Jaya
Tel: +60 3 5638 1476
NL – the Netherlands,
Oldenzaal
Tel: +31 (0)541 585 000 [email protected]
No – Norway, Ski
Tel: +47 64 91 10 00 [email protected]
NZ – New Zealand, Mt Wellington
Tel: +64 9 574 1744 pL – poland, Warsaw
Tel: +48 (0)22 573 24 00 [email protected]
pt – portugal, Leca da Palmeira
Tel: +351 22 999 7360 [email protected]
Ro – Romania, Bucharest
Tel: +40 21 252 1382 [email protected]
Ru – Russia, Moscow
Tel: +7 495 645-2156 [email protected]
SE – Sweden, Spånga
Tel: +46 (0)8 59 79 50 00 [email protected]
SG – Singapore
Tel: +65 6887 6300
SK – Slovakia, Banská Bystrica
Tel: +421 484 162 252 [email protected]
SL – Slovenia, Novo Mesto
Tel: +386 7 337 6650 [email protected]
tH – thailand, Bangkok
Tel: +662 717 8140 tR – turkey, Istanbul
Tel: +90 216 4997081 [email protected]
tW – taiwan, Taipei
Tel: +886 2 2298 8987 uA – ukraine, Kiev
Tel +380 44 494 2731 [email protected]
uK – united Kingdom,
Warwick
Tel: +44 (0)1926 317 878 [email protected]
uS – uSA, Cleveland
Tel: +1 216 896 3000
VE – Venezuela, Caracas
Tel: +58 212 238 5422
ZA – South Africa,
Kempton Park
Tel: +27 (0)11 961 0700 [email protected]
www.parkerhfde.com
European Product Information Centre
(24-hour)
Freephone: +00800 27 27 5374
(from AT, BE, CH, CZ, DE, EE, ES, FI,
FR, IE, IT, PT, SE, SK, UK)
© 2010 Parker Hannifin Corporation.
All rights reserved.
P.849101_CM20_GB_VerA
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Related manuals
advertisement
Table of contents
- 2 Laser Information
- 4 Introduction
- 4 Features
- 5 Benefits
- 8 Powering the icountLCM
- 10 Installing paper and ribbon into the printer
- 12 Getting started
- 12 Setting the time
- 13 Setting the date
- 14 Basic operation
- 16 Independent flow test
- 17 ISO mode
- 18 NAS mode
- 19 SAE mode
- 20 GOST mode
- 21 Entering a test identification code
- 22 Bar code wand
- 23 Printing test results
- 24 Viewing and printing a stored test by ID code or test number
- 25 Printing a stored test as a graph
- 26 Alarm levels
- 28 Automatic testing
- 29 Calibration
- 29 Adjusting screen brightness
- 29 Additional (optional) features
- 30 Reference
- 30 Data interpretation
- 37 ISO contamination charts
- 40 Component cleaning guidelines
- 41 Fluid flow schematic
- 41 Logic diagram
- 42 Peripherals and serial interface
- 42 Measurement accuracy
- 43 Diagnostic codes
- 45 Technical specifications
- 46 Installation dimensions
- 47 Dual fluids monitor - icountLCM
- 48 Operation checklist
- 49 Ordering information