advertisement
REF 12-656
Analyzer
User Manual
Analyzer
User Manual
Refer to the CD in the analyzer package for instructions in English. The instructions are available from your local distributor.
Consulte el CD incluido en el envase del analizador para obtener instrucciones en español. También puede pedir las instrucciones a su distribuidor local.
Anweisungen auf Deutsch befinden sich auf der CD in der Verpackung des Analysegeräts. Die Anleitung ist von Ihrem Händler erhältlich.
Fare riferimento al CD nella confezione dell'analizzatore per istruzioni in italiano. Le istruzioni sono disponibili presso il distributore di zona.
Le CD contenu dans l'emballage de l'analyseur inclut les directives d'utilisation en français. Le mode d'emploi est disponible auprès du distributeur local.
Consulte o CD no pacote do analisador para instruções em Português. As instruções estão disponíveis junto do seu distribuidor local.
Der henvises til den vedlagte CD i analysatorpakken for instruktioner på dansk. Instruktionerne fås hos den lokale forhandler.
Se CD:n i analysatorförpackningen beträffande instruktioner på svenska. Anvisningarna kan erhållas från en lokal distributör.
Lue suomenkieliset ohjeet analysaattorin pakkauksessa olevalta CD:ltä. Ohjeet ovat saatavissa paikalliselta jälleenmyyjältä.
Türkçe talimat için analizör paketindeki CD’ye bak›n›z. Talimat yerel da¤›t›c›n›zdan elde edilebilir.
∞Ó·ÙÚ¤ÍÙ ÛÙÔ CD ÛÙË Û˘Û΢·Û›· ÙÔ˘ ·Ó·Ï˘Ù‹ ÁÈ·
Ô‰ËÁ›Â˜ ÛÙ· ∂ÏÏËÓÈο. √È Ô‰ËÁ›Â˜ Â›Ó·È ‰È·ı¤ÛÈ̘ ·fi
ÙÔÓ ÙÔÈÎfi ‰È·ÓÔ̤· Û·˜.
TABLE OF CONTENTS
Cholestech GDX Analyzer User Manual
Table of Contents
General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
Unpacking the Cholestech GDX Analyzer . . . . . . . . . . . . . . . . . . . .1
Storage Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2
Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2
Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
Principles of the Cholestech GDX A1C Testing System . . . . . . . . . .4
Cholestech GDX Analyzer Basic Operation . . . . . . . . . . . . . . . . . . .5
To Run an Optics Check Cartridge . . . . . . . . . . . . . . . . . . . . . . . . .6
To Run a Cholestech GDX A1C Test . . . . . . . . . . . . . . . . . . . . . . . .8
Displaying Results in Memory . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
To Use the 10 µL MiniPet ™ Pipette . . . . . . . . . . . . . . . . . . . . . . . .9
Performance Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
Contacting Technical Service . . . . . . . . . . . . . . . . . . . . . . . . . . .13
Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
Limitations of Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
Shutdown Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
TABLE OF CONTENTS
Cholestech GDX Analyzer User Manual
Table of Contents
Specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
References . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
GENERAL INFORMATION / UNPACKING THE CHOLESTECH GDX ANALYZER
Cholestech GDX Analyzer User Manual
1
General Information
The Cholestech GDX Analyzer is a small, portable analyzer for use with Cholestech GDX A1C Test Cartridges. The analyzer performs a self-check before each test is run. You may also check that the analyzer is working correctly by using the Cholestech GDX Optics
Check Cartridge and Cholestech GDX A1C Controls.
The Cholestech GDX Analyzer is powered by a 12-volt regulated DC supply which is supplied with the analyzer. Never use any other power supply transformer as this may give incorrect results or permanently damage the analyzer.
This user manual will give general information about the performance of the analyzer and test cartridge. You must also use the package insert in each pack of Cholestech GDX A1C Test Cartridges.
Please keep this guide for future use.
Unpacking the Cholestech GDX Analyzer
Carefully look at the outer box for any damage. If you see any damage to the analyzer, please contact Cholestech Technical
Service or your local distributor.
Contents:
1 Cholestech GDX Analyzer
1 Power Supply Transformer and Plug for European Power Supply
1 Analyzer User Manual
1 Quick Reference Guide
1 Instructional CD-ROM
1 Insert CD-ROM
1 Warranty Card
1 10 µL MiniPet Pipette
50 Pipette Tips
1 Cholestech GDX Optics Check Cartridge (packaged separately)
2 STORAGE INSTRUCTIONS / FEATURES
Cholestech GDX Analyzer User Manual
Storage Instructions
The Cholestech GDX Analyzer must be stored at a temperature between 10 and 35ºC (50–95ºF). It is important to prevent dust and dirt from getting into the analyzer. So keep the analyzer covered.
Make sure the Cholestech GDX Optics Check Cartridge is protected from sunlight during storage. Store the Optics Check Cartridge at a humidity of less than 60% and at a temperature between 15 – 25°C
(59° – 77°F).
Features
Positional Lights
Power Connection
Analyzer
Left and Right
Arrows
Enter Button
Icon and Results Display
Cartridge
Test Platen
DISPLAY
Cholestech GDX Analyzer User Manual
3
Display
Cartridge
Hourglass
Rotate Cartridge
Error Message
This icon will be used with “up” and “down” arrow icons to prompt the user to insert or remove a cartridge from the analyzer.
This indicates that the operator must wait for an action to be performed before proceeding. If a number is displayed, it is a countdown timer.
A single icon prompts the user to rotate the cartridge. This will be used in conjunction with flashing lights placed around the edge of the cartridge to indicate the position to rotate to.
The number displayed is an error message.
Pour Solution
Insert/Remove/Mix
Keypad
Result
Blood Drop
A single icon to prompt the user to pour the solution into the A1C Test Cartridge funnel.
These two icons will be used in combination with either the Cartridge icon or the Solution Tube icon to prompt the user to perform an operation.
The keypad icons comprise “left” and “right” arrows and “enter” button. They perform a function when they are solid (as shown). When they are present in outline only, they are not active, and pressing them will not produce a response from the analyzer.
The number displayed is the percentage of hemoglobin HbA
1c
(A1C) in the test sample.
This icon will be used to prompt the user to take a blood sample.
Identification Number The number displayed is the unique test identification code.
Solution Tube This icon will be used along with the “up” and
“down” arrows to prompt the user either to remove a tube or mix a blood sample in the tube.
Computer Connection A single icon to indicate a link-up with an external device such as a printer or computer, shown during download.
4 PRINCIPLES OF THE CHOLESTECH GDX A1C TESTING SYSTEM
Cholestech GDX Analyzer User Manual
Principles of the Cholestech GDX
A1C Testing System
The Cholestech GDX A1C Testing System uses boronate affinity chromatography to separate the glycated hemoglobin fraction.
The Cholestech GDX Analyzer, which is a single wavelength spectrophotometer (440 nms), measures both fractions and an algorithm converts the result into the percentage A1C in the sample.
After a Cholestech GDX Test Cartridge has been placed in the analyzer, a small sample of blood is added to the first sample tube. The blood is instantly lysed to release the hemoglobin and the boronate affinity resin binds the glycated hemoglobin. After a short incubation step, the liquid is poured into the central funnel of the test cartridge and the nonglycated fraction is collected in an optical chamber, where the hemoglobin concentration is photometrically measured. A combination of microswitches is used by the logic of the analyzer to detect the exact position of a test or control cartridge during operation, and Dynamic
Meniscus Sensing (DMS) is used to detect when liquids enter the cartridge reading wells.
The glycated hemoglobin remains bound to the boronate affinity resin, which sits at the bottom of the test cartridge funnel. The boronate affinity resin is then washed with the contents of the second buffer tube and the final step is the elution of the glycated hemoglobin off the boronate affinity resin using the third buffer tube. The glycated hemoglobin concentration is measured and the
A1C concentration in the sample is calculated and reported by the analyzer. The algorithm for the calculation is:
% A1C= M
A Glycated X 100
A Glycated + A Nonglycated
+ C
Where “M” and “C” are slope and intercept factors to correct the value for DCCT calibration (see Calibration, page 8).
CHOLESTECH GDX ANALYZER BASIC OPERATION
Cholestech GDX Analyzer User Manual
5
Cholestech GDX Analyzer Basic Operation
WARNING: When using the Cholestech GDX
Analyzer, do not operate within an electromagnetic field. Do not operate any
VHF emitting equipment within 1 meter
(39.4 inches).
The analyzer will run through a check stage. All the red positional indicator lights on the analyzer will flash, the beep will sound, and the display will show all icons and characters momentarily.
The Cholestech GDX Analyzer is designed to be simple and easy to operate. This user manual describes some basic functions, but must be used with the package insert in each pack of Cholestech GDX A1C Test Cartridges.
A number will appear on the display. This is the version of software in the analyzer, for example 2.06.
The Hourglass icon will flash to tell you to wait.
Please read this user manual carefully before use.
1. Ensure the Analyzer is placed on a flat surface. Operate the Analyzer between 17 and 30ºC (63–86ºF).
2. Remove the power pack. Check that the input power matches the local voltage supplied. If the input voltage does not match the local power supply, contact Cholestech Technical Service or your local distributor.
Note: The Hourglass icon may flash for some time if the analyzer has been stored at extreme temperature/humidity. Normally the time required for warm-up is 10 minutes. If the analyzer has been moved from a high or low temperature or humid area just prior to warm-up, the warm-up period may be extended to up to 40 minutes. If the analyzer fails to warm-up an error message will appear on the display.
(See Troubleshooting, pages 12–13.) 3. Plug the power pack into the main supply. Connect the jack plug into the back of the Analyzer indicated by the following marking.
When the analyzer is ready for use, the
Insert and Cartridge icons will flash. The
“left” and “right” arrows will be solid.
You can now use the analyzer to run an Optics Check Cartridge or an A1C
Test Cartridge.
6 TO RUN AN OPTICS CHECK CARTRIDGE
Cholestech GDX Analyzer User Manual
To Run an Optics Check Cartridge
Intended Use
The Optics Check Cartridge is supplied with the Cholestech GDX Analyzer to check that it is working correctly.
Replacement Optics Check Cartridges are available from Cholestech.
Summary and Explanation
The measurement of A1C is recommended for monitoring the long-term care of people with diabetes.
1,2 The concentration of A1C within red blood cells reflects the average level of blood sugar (glucose) over the previous 2–3 months. The higher the blood sugar level, the higher A1C.
Principles of the Procedure
The Optics Check Cartridge is reusable.
It is designed to check that the analyzer is working correctly. The cartridge contains a resin with a dye that acts like your blood sample. The analyzer will automatically recognize when an Optics Check Cartridge is being used.
The Optics Check Cartridge should be run every day the Cholestech GDX
Analyzer is used. If you are concerned that your test result is incorrect, also run an Optics Check Cartridge.
Running an Optics Check Cartridge
1.
Plug in the analyzer and allow to warm up.
2.
Place the cartridge in the analyzer when the insert cartridge icon flashes.
• “ ”appears on the display and the Hourglass icon is displayed.
• The Rotate Cartridge icon is displayed. The Analyzer will beep and the light at Position 1 will flash.
3.
Turn the cartridge to Position 1.
• “ ” and the Hourglass icon will show on the display.
• The Rotate Cartridge icon will be displayed. The analyzer will beep and the light at Position 2 will flash.
4.
Turn the cartridge to Position 2.
• “ ” and the Hourglass icon will show on the display.
• The Rotate Cartridge icon will be displayed. The analyzer will beep and the light at Position 3 will flash.
5.
Turn the cartridge to Position 3.
• “ ” and the Hourglass icon will show on the display.
• The Rotate Cartridge icon will be displayed.
6.
Turn the cartridge to Position 0.
• The Remove and Cartridge icons will be displayed.
TO RUN AN OPTICS CHECK CARTRIDGE
Cholestech GDX Analyzer User Manual
7
7.
Remove the cartridge.
• The analyzer will display the result: e.g. .
Note: “ ” stands for Optics Check Cartridge.
8.
Press the “enter” button to clear the
Optics Check Cartridge result.
Shelf Life Stability
Optics Check Cartridges have a shelf life date. This is printed on the top of the Optics
Check Cartridge.
If the result does not fall within the range displayed, do the following:
• Clean the clear plastic surfaces of the cartridge with a soft, lint-free cloth.
Do this on the inside and outside of the cartridge. Follow the test from point 2.
• If the result is still not in the correct range, turn the analyzer off. Restart the analyzer and run the Optics
Check Cartridge.
• If the result is still not correct, contact
Cholestech Technical Service.
• If the cartridge is physically scratched, dirty or damaged, replace the cartridge with a new Optics Check Cartridge.
Storage
Warning: For In Vitro Diagnostic Use
Always store the Optics Check Cartridge at room temperature in its box. This will help to protect it from damage such as scratches which may affect the result.
The cartridge must not be left in direct sunlight for long periods while not in use.
Do not use the Optics Check Cartridge after the end of the shelf life date. When the test is finished, the Optics Check Cartridge should be returned to its protective packaging for storage. Do not use the Optics Check
Cartridge if it has been stored incorrectly and is scratched or damaged in any way.
The Optics Check Cartridge must be stored at relative humidity of less than 60%.
Results
For technical help and troubleshooting, please refer to the Cholestech GDX A1C
Test Cartridge package insert and this user manual. If technical help is required, please contact your local distributor.
The result displayed on the analyzer at the end of Optics Check Cartridge test must be in the range on the cartridge label.
8 TO RUN A CHOLESTECH GDX AIC TEST / DISPLAYING RESULTS IN MEMORY
Cholestech GDX Analyzer User Manual
/ CALIBRATION
To Run a Cholestech GDX A1C Test
Refer to the Cholestech GDX A1C Test Cartridge package insert which is in each pack of test cartridges or use the Cholestech GDX
Quick Reference Guide supplied with the analyzer.
Displaying Results in Memory
Press the “left” arrow to select the unique identification (ID) number for the result required.
Press the “enter” button. The stored result will be displayed.When this value has been written down, press the “enter” button once more.
You may find another result from the memory in the same way.
Return to the Insert and Cartridge icons by moving up the stored results until the last ID number has been shown on the display.
Note: To exit rapidly from the memory, press the “left” and “right” arrow buttons simultaneously.
Calibration
The Cholestech GDX Analyzer and A1C Test Cartridges have been manufactured to deliver an A1C result. This is calibrated to the recommendations of the Diabetes Control and Complications
Trial (DCCT). This result is traceable to the NGSP (National
Glycohemoglobin Standardization Program), an internationally accepted method of standardization. The DCCT was a landmark multicenter clinical study which conclusively linked elevated A1C levels to the complications associated with diabetes. The result of the test can be used by your doctor to monitor your disease.
Traceability of Values
The Cholestech GDX Analyzer and A1C Test Cartridges have been manufactured to deliver an A1C result. Values have also been assigned using materials supplied by the IFCC network.
3 Users may
CALIBRATION / TO USE THE 10 µL MINIPET PIPETTE
Cholestech GDX Analyzer User Manual
9 convert their DCCT values to IFCC values using the following equation:
GDX IFCC values = 1.1218 x GDX DCCT value - 2.684.
This equation was derived by Provalis Diagnostics Ltd. specifically for the GDX test using IFCC calibrated reference materials and was shown to to have a correlation coefficient of r = 0.99.
To Use the 10 µL MiniPet
™
Pipette
Use this procedure to apply a venous whole blood sample or control sample to the red-capped sample tube of the Cholestech GDX A1C
Test Cartridge.
• Firmly attach a pipette tip to the end of the 10 µL
MiniPet Pipette.
• To fill the pipette, push the plunger down as far as you can.
Place the pipette tip into the blood or quality control sample and slowly release the plunger. Do not allow air to be drawn into the tip.
• Place the pipette tip into the liquid of the red-capped sample tube. Dispense the sample into the sample tube by pressing the plunger again. Move the pipette tip out of the liquid before releasing the plunger.
• Remove the pipette tip and discard it in a biohazard waste container.
Note: If the plunger is released before the pipette tip is out of the liquid, it will remove some of the sample just dispensed.
10 PERFORMANCE CHARACTERISTICS
Cholestech GDX Analyzer User Manual
Performance Characteristics
Clinical Trial Results
Clinical studies were carried out to assess the performance of the
Cholestech GDX System for the measurement of A1C in diabetic patients. The results were compared with those for a number of laboratory-based systems.
System Manufacturer Patient Number (n) Correlation
Variant II HPLC
Mono S HPLC
Biomen 8121 HPLC
Total GHb Test
Glycomat 765 HPLC
Glycotest II
DCA 2000
Bio-Rad
Pharmacia
Menarini
Sigma
Corning
Pierce
Bayer
62
62
24
94
95
95
56 r=0.98
r=0.98
r=0.97
r=0.96
r=0.94
r=0.97
r=0.96
The equations for the GDX test compared to the Menarini Biomen
HA8140 and the Bayer DCA 2000 are as follows:
GDX = 0.9167 X DCA 2000 - 0.1323
GDX = 0.8968 X Menarini HA8140 + 0.7074
Precision
In comparative studies using three different batches of Cholestech GDX
Test Cartridges, the overall standard deviation was 0.23 (4.1% CV) for a sample having a level of 5.7% A1C and 0.36 (3.8% CV) for a sample having a level of 9.4% A1C.
Linearity and Temperature Effects
The Cholestech GDX System has been shown to give a linear response from 4-15% A1C using patient samples. Values above
15% A1C are outside the validated range. Caution should be exercised if results outside the validated range of 4-15% are obtained. Cholestech GDX test results were shown not to be affected by operating temperatures of 17-30°C (63-86°F).
PERFORMANCE CHARACTERISTICS
Cholestech GDX Analyzer User Manual
11
Abnormal Hemoglobins
The Cholestech GDX assay uses boronate affinity chromatography to determine the level of A1C in a blood sample. The presence of hemoglobin variants in a sample, such as HbS, HbC and HbD, which are known to affect other, non-boronate-based methods of
A1C measurement, do not interfere with the Cholestech GDX test.
In addition, the presence of fetal hemoglobin (HbF) in a sample does not interfere with the Cholestech GDX test result.
5 Uremic patients with high levels of carbamylated hemoglobin can also be tested with the Cholestech GDX System.
Hemoglobin and Hematocrit
The Cholestech GDX System performs acceptably over a hemoglobin a range of 8–20 g/dL and a hematocrit range of 30–60%.
Interfering Substance
The following substances were shown to have no effect on the
Cholestech GDX test result at the levels indicated: bilirubin 145
µmol/L, acetylsalicylic acid 300 µg/mL, caffeine 300 µg/mL, acetoaminophenol 300 µg/mL and hydroxyzine dihydrochloride
300 µg/mL. Triglycerides at 8.6 mmol/L showed no effect on the
Cholestech GDX test result; however, elevated lipid concentrations may interfere and cause low results in this type of assay.
Venous Sample Use/Storage
Venous samples may be stored at 2–8°C (36–46°F) for up to 4 days.
Store samples as whole blood, not as hemolysates or packed red blood cells. Venous samples may use anticoagulants potassium-EDTA, sodium citrate, sodium heparin, lithium heparin and fluoride.
12 TROUBLESHOOTING
Cholestech GDX Analyzer User Manual
Troubleshooting
The Cholestech GDX Analyzer has been programmed to detect any analyzer or test problems. An error code will appear on the display if any mistakes arise during the test.
Look at the following table to see the cause of the error code and the possible correction.
Error Cause
Code
Explanation
1 Test abandoned by the user.
2 Test cartridge is not in the analyzer.
3 Cartridge has not been turned in time.
4 The cartridge is in the wrong position.
5 Too long has been taken to do the step.
6 The sample has not been added.
7 Enter button has been pressed too early.
8 Internal analyzer check is incorrect.
9 Internal analyzer check is incorrect.
10 Internal analyzer check is incorrect.
11 Internal analyzer check is incorrect.
12 Internal analyzer check is incorrect.
13 Internal analyzer check is incorrect.
14 Internal analyzer check is incorrect.
N/A
Analyzer has failed to detect a test cartridge. This may be due to an analyzer problem, or the cartridge is damaged.
It is essential to follow the instructions without long delays between steps. The analyzer measures time intervals between steps and will error if the time taken is too long.
The cartridge has been moved when not told to do so. It is essential to follow the instructions for use.
The analyzer measures time intervals between steps and will error if the time taken is too long.
The analyzer measures time intervals between steps and will error if the time taken is too long. It is essential to follow the instructions for use.
The analyzer measures time intervals between steps and will error if the time taken is too long or too short. It is essential to follow the instructions for use.
The cartridge may be damaged or dirty. To check that the analyzer is running correctly, run the Optics Check Cartridge.
The cartridge may be damaged or dirty. To check that the analyzer is running correctly, run the Optics Check Cartridge.
This may be due to an insufficient blood sample, or failure of the liquid to flow through the funnel. Do the test again and make sure that the pipette is filled correctly.
This may be due to too much blood. Do the test again and make sure that the pipette is filled correctly.
This may be due to an insufficient blood sample, or failure of the liquid to flow through the funnel. Do the test again and make sure that the pipette is filled correctly.
Do the test again and make sure that the pipette is filled correctly.
Do the test again.
CONTACTING TECHNICAL SERVICE
Cholestech GDX Analyzer User Manual
13
Contacting Technical Service
For technical assistance, please contact your local distributor.
Error Cause
Code
Explanation
15 Internal analyzer check is incorrect.
16 Internal analyzer check is incorrect.
17 Internal analyzer check is incorrect.
18 Internal analyzer check is incorrect.
19 Internal analyzer check is incorrect.
20 Internal analyzer check is incorrect.
21 Internal analyzer check is incorrect.
22 Internal analyzer check is incorrect.
23 Internal analyzer check is incorrect.
24 The supply voltage is not at the correct level.
25 Internal analyzer check is incorrect.
26 Warm-up cycle has not been completed in time.
27 Programmed analyzer shutdown.
Do the test again.
If the problem persists, contact your local distributor.
Do the test again.
Remove the cartridge and do the test again.
The cartridge may be damaged or dirty. To check that the analyzer is running correctly, run the Optics Check Cartridge.
The cartridge may be damaged or dirty. To check that the analyzer is running correctly, run the Optics Check Cartridge.
The cartridge may be damaged or dirty. To check that the analyzer is running correctly, run the Optics Check Cartridge.
The cartridge may be damaged or dirty. To check that the analyzer is running correctly, run the Optics Check Cartridge.
The cartridge may be damaged or dirty. To check that the analyzer is running correctly, run the Optics Check Cartridge.
The cartridge may be damaged or dirty. To check that the analyzer is running correctly, run the Optics Check Cartridge.
The analyzer must be run with 11.4 –12.6 volts DC. Check the type of transformer being used.
If this problem continues, contact your local distributor.
Turn the analyzer off. Wait a few seconds and turn the analyzer on again. If the problem continues, contact your local distributor.
The analyzer has reached the limit of 9,999 tests. Contact your local distributor.
14 SAFETY / LIMITATIONS OF USE
Cholestech GDX Analyzer User Manual
Safety
The Cholestech GDX Analyzer requires a 12-volt (±5%) DC supply regulated at greater than 250 milliamps. This is provided. There are no user-serviceable parts.
If the Cholestech GDX Analyzer is dirty, clean it carefully with a lint-free cloth dampened with aqueous solutions containing 5% hypochlorite or alcohol. Take care not to get liquid in the space for the cartridge. Do not use more aggressive chemicals such as acetone.
The Cholestech GDX Analyzer has been fully tested according to
Electrical Safety Regulations (BS EN 61326:1998).
Limitations of Use
• For professional use only.
• For in vitro diagnostic use only.
• Do not use any part of the test components beyond the marked expiry date.
• Store the test components according to the instructions on the box.
• Always wear gloves when handling blood samples.
• In patients where the lifespan of the red blood cell is significantly shortened, e.g. hemolytic anemia, new A1C reference ranges should be established.
• The test solutions (first and second tube) contain 0.1% sodium azide as preservative, which is very toxic by inhalation, in contact with skin and if swallowed. After contact with skin, wash immediately with plenty of water.
• Cholestech is not responsible for any injury or damage caused by the use of the Cholestech GDX Analyzer for purposes other than for which it is intended or by modifications of the Cholestech GDX Analyzer.
SHUTDOWN INFORMATION / DISPOSAL
Cholestech GDX Analyzer User Manual
15
Shutdown Information
Following testing, the instrument may be left on ready for the next test or switched off at the main socket. If unused for a period it should be placed in its carton to protect the optical window.
Disposal
The analyzer has been designed for a long and fault-free life and has been validated for 9,999 tests. As long as the Optics Check
Cartridge is giving acceptable results, the Analyzer may be used with confidence. Excessive wear on the mechanisms will result in error conditions, and may indicate that the Analyzer has reached the end of its useful life. The Analyzer must then be disposed of according to local regulations concerning the disposal of electronic items in general.
The Cholestech GDX Analyzer has been preprogrammed to shut down after 9,999 tests have been performed. From test ID 9900 onwards the ID number will flash momentarily to remind the user that the analyzer is approaching the end of its programmed life.
16 SPECIFICATION
Cholestech GDX Analyzer User Manual
Specification
Precision:
<1%CV
Wavelength:
440 nm (nominal)
Molded materials:
Keypad – Silicone elastomer
Casework – ABS
Window – Acrylic
Central well features – Delrin
Feet – Evoprene-G
Signal to noise:
>1,000:1
Power:
12-volt (±5%) DC at >250 mA
Data download:
RS232 serial port
REFERENCES
Cholestech GDX Analyzer User Manual
17
References
1. American Diabetes Association. Diabetes Care 2002;
25(Suppl.1):S33–S49.
2. Goldstein DE et al. Diabetes Care 1995; 18:896–909.
3. Jeppsson JO, Kobold U, Barr J, Finke A, Hoelzel W, Hishino T,
Miedema K, Mauri P, Mosca A, Paroni R, Thienpont L,
Umemoto M, Weykamp C. Approved IFCC reference method for the measurement of A1C in human blood. Clin Chem Lab
Med 2002;40:78-89.
4. The Diabetes Control and Complications Trial Research
Group. N Engl J Med 1993; 329:977–986.
5. John WG. Glycated hemoglobin analysis.
Ann Clin Biochem 1997; 34:7–31.
18 NOTES
Cholestech GDX Analyzer User Manual
Notes Notes
NOTES
Cholestech GDX Analyzer User Manual
19
Analyzer User Manual
Cholestech GDX is a trademark of Cholestech Corporation.
MiniPet is a trademark of TriContinent Scientific, Inc.
This product is sold under license from Provalis Diagnostics Ltd.
Distributed by Cholestech Corporation. U.S. Patent 6,300,142.
Provalis Diagnostics Ltd, Newtech Square,
Deeside Industrial Park, Deeside, Flintshire, UK. CH5 2NT.
T: + 44 1244 288888 F: + 44 1244 833441.
201412E Rev. 01/06
003 12710 Rev. B
© 2006 Cholestech Corporation.
Distributed by
3347 Investment Blvd.
Hayward, CA 94545 U.S.A.
U.S. Tel 1.510.732.7200
Fax 1.510.732.7227
www.cholestech.com
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project