Manual for connection of load cells to junction box and weighing indicators
2014-06-26 Ver 2
Junction Box YZ-J4
Load cell Connect to Junction Box Connect to Indicator
Each load cell cable, up to 4 pcs
EXC+
EXC-
SIG+
SIG-
GND
=>
=>
=>
=>
=>
From junction box one cable, min 0,25mm² to indicator
EXC+
EXC-
SIG+
SIG-
GND
=>
=>
=>
=>
=>
EXC+ *
EXC*
SIG+ *
SIG*
GND *
Important!
Turn all potentiometers max clockwise if you do not whish to make corner adjustment
Corner adjustment is in almost all application not necessary, without adjusting corner the difference could be max ±0,1% between the corners.
Junction Box PT100SBE-4
Load cell Connect to
Each load cell cable, up to 4 pcs
EXC+ =>
EXC-
SIG+
=>
=>
SIG-
GND
SEN+
SEN-
=>
=> not used not used
Junction Box Connect to Indicator
From junction box one cable, min 0,25mm² to indicator
EXC+ => EXC+ *
EXC-
SIG+
=>
=>
EXC-
SIG+
*
*
SIG-
GND
SEN+
SEN-
=>
=>
=>
=>
SIG*
GND *
SEN+ **
SEN**
Important!
All Jumpers J1-J4 must be short circuit if you do not whish to make corner adjustment
Corner adjustment is in almost all application not necessary, without adjusting corner the difference could be max ±0,1% between the corners.
Junction Box PT100SB-6 and PT100SB-8 (with corner adjustment)
Load cell Connect to Junction Box Connect to
Each load cell cable, up to 8 pcs
EXC+ =>
EXC-
SIG+
=>
=>
SIG-
GND
SEN+
SEN-
=>
=>
=>
=>
Indicator
From junction box one cable, min 0,34mm² to indicator
EXC+
EXC-
SIG+
SIG-
GND
SEN+
SEN-
=>
=>
=>
=>
=>
=>
=>
EXC+ *
EXC-
SIG+
*
*
SIG*
GND *
SEN+ **
SEN**
Important!
Follow the junction box manual how to make corner adjustments.
Vetek Weighing AB Hantverksv. 15 76040 Väddö Sweden Tel +46 (0)176208920 Fax +46 (0)176208929 [email protected] www.vetek.com
Junction Box PT100SB-6 and PT100SB-8 (without corner adjustment)
Load cell Connect to Junction Box Connect to Indicator
Each load cell cable, up to 8 pcs
EXC+ =>
EXC-
SIG+
SIG-
GND
=>
=>
=>
=>
From junction box one cable, min 0,34mm² to indicator
EXC+
EXC-
SEN+
SEN-
GND
=>
=>
=>
=>
=>
EXC+ *
EXC*
SIG+ *
SIG*
GND *
2014-06-26 Ver 2
Junction Box LC-CB
Load cell Connect to
Each load cell cable, up to 4 pcs
EXC+
EXC-
SIG+
SIG-
GND
=>
=>
=>
=>
=>
Junction Box Connect to Indicator
From junction box one cable, min 0,25mm² to indicator
EXC+
EXC-
SIG+
SIG-
GND
=>
=>
=>
=>
=>
EXC+ *
EXC*
SIG+ *
SIG*
GND *
* On LCA-D SIG+ and SIG- are named I+ and I-
* EXC and SIG are often marked as E and S only.
** If you have more than 10m distance from junction box to indicator we recomend you to choose a
6 wires cable and just connect the wires between SEN+ and SEN- between junction box and indicator
If cable is shorter than 10 meters it´s ok not to use SEN function, in that case you must connect a short circuit wire on the indicator between +EXC to +SEN and -EXC to -SEN. However on many indicators there is a short circuit jumpers to be used on the electronic board, please check indicator manual.
WARNINGS!!!
* Never overload a cell with more than 50%
* Never perform welding work near the load cell. The cell(s)
must be mechnaically removed before welding work can be done.
* Always if possible make sure the cells does not stand in water,
not even the IP68 models.
Vetek Weighing AB Hantverksv. 15 76040 Väddö Sweden Tel +46 (0)176208920 Fax +46 (0)176208929 [email protected] www.vetek.com
2014-06-26 Ver 2
EXC+
EXC-
SIG+
SIG-
Color code;
EXC+
EXC-
SIG+
SIG-
Color code;
EXC+
EXC-
SIG+
SIG-
Color code;
EXC+
EXC-
SIG+
SIG-
SEN+
SEN-
Color code;
EXC+
EXC-
SIG+
SIG-
SEN+
SEN-
Vetek load cell range, color codes:
Color code;
EXC+
EXC-
SIG+
SIG-
For all below models (series)
Red
Black
White
Yellow
C2S, C8S, CLS,
CLT, D100, D200
T20, TCA, TC4
TCE, TS
Color code; For all below models (series)
Red
Black
Green
White
108 and VZ (all models),
202WA, 535, 651,
AR, ARL, BM-CFS, CP,
PA-6181, SBS, SC
For all below models (series)
DT-101, Red
Blue
Green
Yellow
For all below models (series)
FXC, F60X Red
White
Black
Blue
For all below models (series)
SPSC, DSB Red
Black
Green
White
Purple
Grey
For all below models (series)
AG, AH Brown
Green
Yellow
White
Grey
Pink
Vetek Weighing AB Hantverksv. 15 76040 Väddö Sweden Tel +46 (0)176208920 Fax +46 (0)176208929 [email protected] www.vetek.com
2014-01-10 Ver 1
Example: Loadcells VZ563 with Junction Box YZ-J4 and indicator DFWL
Load cell VZ Junction Box YZ Cable from junction box to => Indicator DFWL
Each load cell cable, up to 4 pcs
Red
Black
Green
White
Shield
EXC+
EXC-
SIG+
SIG-
GND
Any color code can be used on this cable of course.
EXC+
EXC-
SIG+
SIG-
GND
=>
=>
=>
=>
=>
EXC+
EXC-
SIG+
SIG-
GND
Make sure J2 and J3 jumber are on (short circuit)
Example: Loadcells VZ563, Junction Box YZ-J4 and indicator (transmitter) DGT1/4
Load cell VZ Junction Box YZ Cable from junction box to => Indicator DFWL
Each load cell cable, up to 4 pcs
Red EXC+
Black
Green
White
EXC-
SIG+
SIG-
Shield GND
Any color code can be used on this cable of course.
EXC+
EXC-
SIG+
SIG-
GND
=>
=>
=>
=>
=>
EXC+
EXC-
SIG+
SIG-
SEN+ => EXC+
SEN => EXC-
GND
One should short-circuiting - SEN with -EXC and +SEN with +EXC by separate wires.
Vetek Weighing AB Hantverksv. 15 76040 Väddö Sweden Tel +46 (0)176208920 Fax +46 (0)176208929 [email protected] www.vetek.com
Vetek indicator range terminal connections
DGT1 model
DGT4, DGT20 and DGT60 models (4 channels)
DGTQ models
DGTP and DGTPK models
DFW
DFWL
3590 all models
LCA-D
On LCA-D SIG+ and SIG- are named I+ and I-
All other models are very easy to see the load cells connection markings directly on the product.