Ohaus A71P15DTNUS is a versatile weighing scale designed for various applications, including industrial, laboratory, and retail settings. It boasts a large, easy-to-read LCD display, ensuring clear visibility of weighing results. The scale provides precise measurements with a capacity of 15000g and a readability of 0.1g, making it suitable for both small and large items. Its durable construction and overload protection ensure longevity and reliability.
advertisement
Addendum Sheet: (CALIBRATION)
Entering Scale Configuration Setup for Calibration
1. When unit price is 0, press the Menu key for around 3 seconds.
2. After hearing a long beep, the display will show "SET 1 OP".
3. Press the Tare key to go down to set 3 CAL, then press ON/OFF key.
4. Display messages: Press Seal.
5. Locate the Calibration switch under the scale (you may need to unscrew one of the sealing screws).
6. Press the calibration button (within the second hole from the front) to enter the configuration, display will show: SET.
Continue with calibration steps below.*
Important Note:
Assuming the user is to use the (non-automatic) scales/balance in the legally regulated field, the user will be responsible for notifying the competent calibration authorities of the repaired scales/balance, so that the latter can take the appropriate measures (calibration/recalibration).
Calibration Switch Position:
The calibration switch is located under the scale next to the adapter housing (within the second hole from the front). Use a thin rod to press the calibration switch when system displays message “Press seal”.
Calibration Steps:
Operation steps on scale Information displayed on screen
Turn on scale, Press calibration switch (hear a beep) Set1 OP
Press [->T<-] for 2 times
Press [On/Off]
* Press [->T<-]
Press [On/Off] to choose GEO (if current GEO code is what you need, please directly go to next step)
Set3 CAL
Set
GEO xx (xx means GEO code)
Press [->T<-]
Press [On/Off] to toggle to yes
Press [->T<-]
Press [On/Off] to choose Unit LB
Press [->T<-]
Press [->T<-]
Press [On/Off] to choose capacity
Press [->T<-]
Empty weighing pan
Press [->T<-]
Put 2/3 capacity weight on weighing pan
Press [->T<-]
Press [On/Off] can toggle between Abort and Save
CAL N
CAL Y
Kg or LB (depend on selected last time)
Dual
CAP 30
- - - -
Read 0 (count from 5 to 0), then display
ADD XX (2/3 capacity)
Read the weight, then display SAVE
When display Save, press [->T<-] to save calibration Scale automatically turn off and then on
30127897
Page 1 of 1