PAD M2720A, M2725A, M2726A & M2727A
Data Sheet
Avalon CTS
Cordless Fetal Transducer System
Features
The Avalon CTS Cordless Fetal Transducer System
measures and transmits standard fetal parameters to
a compatible fetal monitor using radio transmission
(RF) technology, thus eliminating transducer cables.
• Reliable parameter signal quality, with a line-ofsight operating range of up to 100m/300 ft1using
the standard antenna2
• High capacity, rechargeable Lithium-Ion batteries,
combining an operating time of more than 16
hours per charge cycle with a short recharge time
of less than 3 hours
• Smart battery management ensures rapid “top-up”
charging at any time, even when the base station is
in stand-by, without the memory effects typically
associated with Ni-Cd batteries
• Color-coded transducers and docking slots
making placement of transducers easy
• Intelligent, flexible channel management for
automatic radio frequency allocation, ensuring
optimum use of the radio frequency band and
allowing transducer exchange between different
base stations
• Connects to any fetal sensor socket on the Philips
Avalon FM20 and FM30 fetal monitors3 via
dedicated interface cable
• Interface to the Series 50 family of fetal monitors
from Philips, Agilent and Hewlett-Packard
It is designed for use in the antepartum period and
during labor and delivery.
The Avalon CTS combines maximum convenience
and versatility with ease of use.
It offers:
• Cordless, waterproof, light-weight transducers,
that are easy to clean
• Higher clinical confidence, by allowing expectant
mothers to be continuously monitored, whether in
the bath or shower, in the bed, or while
ambulating
• Increased comfort, convenience and flexibility for
patients with normal labor and deliveries
1. The effective operating range can vary depending on the
geographical and constructional characteristics of the building where
the system is installed, and whether underwater monitoring is
performed.
2. There may be circumstances where you need to connect the Avalon
CTS to an antenna system. Contact Philips Support regarding site
preparation, antenna system design and assistance with your specific
requirements.
3. Requires Avalon FM20/FM30 interface cable M2731-60001, and
monitor software revision D.00.00 or higher.
Description
The Avalon CTS consists of a base station and the
following transducers:
• Cordless US transducer measures fetal heart rate*
and fetal movement profile
• Cordless Toco transducer measures uterine activity
externally
• Cordless ECG transducer measures either DECG*
or MECG.
*Only one fetal heart rate can be measured at a time, using
either ultrasound or DECG.
The transducers are powered by rechargeable
batteries. These are charged when the transducers
are docked in a base station connected to AC power.
Radio Frequency Transmission
The Avalon CTS consists of transducers that convert
measurements to radio signals and send them to the
base station which processes the signals for input to
the fetal monitor.
The transducers have programmable radio frequency
(RF) generators, operating within the country
specific frequency range.
There are two operating modes for the base station:
automatic channel allocation, and configurable fixed
frequency.
Transmission frequency range: 10 MHz hardware
RF tuning range.
The exact frequency range covered is country
specific, using only a section of the entire hardware
frequency range, and the starting point of the tuning
range depends on the selected country option. Be
sure to comply with any local regulations regarding
approved frequencies, and contact Philips for
assistance with your specific requirements.
Frequency Ranges
Country/Country Group
Japan
• when a transducer is ready for use
• status information about the transmission link
between the transducer and base station
• when there is an error condition
Transducer
Each transducer has an LCD display showing:
• a bed symbol and an associated two-digit numeric
bed label
• a battery capacity status symbol with three
elements
• a warning symbol to indicate incorrect use or
technical problems
• the transducer slot position indicator
Alerting
Alerts are available for loss of RF signal and low
battery capacity.
Connectors
•
•
•
•
•
BNC antenna connector
Female D-sub25 monitor interface
Earth ground pin IEC 320 male power connector
3.5mm stereo plug service connector
Power inlet
Safety Features
• Optical and acoustic advance alert one hour before
battery is empty
• Unique transducer ID code to prevent the base
station from accepting the wrong signal
• Automatic resume after power failure
Interface Capabilities
420 - 430
Avalon FM20/FM30 Fetal Monitors
608 - 618
Displays
Base station
The base station has the following displays and
indicators:
• a two-digit numeric LED display showing
– the unique base station ID (the Bed Label)
– configuration settings
– error codes
2
Each transducer docking slot has three backlit
indicators showing:
Hardware RF
Tuning Range
(MHz)
Most European Countries
433 - 443
(covers Industrial, Scientific and Medical
(ISM) band)
USA, Canada, Australia
(covers Wireless Medical Telemetry
Service (WMTS) band)
• a speaker on/off symbol showing whether the
acoustic alert is on or off
Docking slots
The Avalon CTS interfaces to the Avalon family4 of
fetal monitors via the Avalon FM20/FM30 interface
cable (M2731-60001). The cable connects to any of
the four fetal sensor sockets on the monitor.
Series 50 Fetal Monitors
US and Toco parameters interface to all Series 50
fetal monitors. Cordless DECG interfaces with
M1350A/B/C and M1353A. Cordless MECG can
only be measured by M1350A/B/C.
M1350B/C with Tele/System Interface M135066536 (supports FMP measurements).
4. Requires Avalon FM20/FM30 monitor software revision D.00.00 or
higher.
M1350A5 with Tele/System Interface M1350-66536
Safety, Performance
or Telemetry Interface M1350-66531.
FMP measurements only possible with M135066536.
M1351A5 with Combined Interface Module M135366531E. FMP measurements possible only with
board Rev. E and higher (Option J10, J13 or J14).
• EN 60601-1:1990+A1:1993+A2:1995
(IEC 60601-1:1988+ A1:1991+A2:1995)
• EN 60601-1-1:2001 (IEC 60601-1-1:2000)
• EN/IEC 60601-2-37:2001
• EN/IEC 60601-2-49:2001
• EN 60601-1-4:1996
• UL2601-1
• CAN/CSA C22.2#601.1-M90
• JIS T 1001-1992
• JIS T 1002-1992
• AS 3200.1.0-1998
Electromagnetic Compatibility
• EN/IEC 60601-1-2:2001
• FCC 47 CFR Part 15 Subpart B
• ICES-001:1988
M2720A is classified as electromagnetic emissions
Class B.
M1353A5 with Combined Interface Module M135366531E. FMP measurements possible only with
board Rev. E and higher (Option J10, J13 or J14).
Standards Compliance
Federal Law in the USA restricts this device to use by
or on the order of a physician.
The Avalon CTS complies with the essential
requirements of the Medical Devices Directive (93/
42/EEC) and the Radio and Telecommunications
Terminal Equipment Directive (1999/5/EC), and
with the following standards:
Radio
• EN 300 220-3:2000, EN 300 220-1:2000
• FCC 47 CFR Part 15 Subpart C and Part 95
(WMTS)
• RSS-210
IEEE C95.1-1999
5. Telemetry interface upgrade available.
Environmental Specifications
The monitor may not meet the given performance specifications if stored and used outside the specified
temperature and humidity ranges.
Environmental Specifications (Transducers and Base Station)
Temperature Range
Humidity Range
Altitude Range
Charging
0°C to 45°C (32°F to 113°F)
Operating
0°C to 45°C (32°F to 113°F)
Storage (without battery)
-20°C to +60°C (-4°F to 158°F)
Storage with battery
Depends on initial charge level and temperature (storage time
decreases significantly at high (> 45°C/113°F) temperatures
Operating
5% to 95% relative humidity @ 40°C/104°F
Storage
5% to 85% relative humidity @ 50°C/122°F
Operating
≤ 3000 m/9800 ft.
Storage
≤ 15000 m/49000 ft.
Base Station Specifications
Base Station Specifications
Power
Dimensions and Weight
Supply Voltages
100 VAC to 240 VAC ± 10%
Supply Frequency Range
50 Hz to 60 Hz
Consumption
15 VA
Size mm/(in): width x depth x height
350 x 240 x 75 (13.8 x 9.5 x 3.0 in)
Weight
2.5 kg/5.5 lbs without transducers
Type of Protection Against
Electrical Shock
Class I equipment
Input Sensitivity
Input Sensitivity
-110 dBm @ 30 dB Signal-to-Noise Ratio
Image Rejection
Image Rejection
> 80 dB
Operating Range
Receiving Range (line of sight)
approximately 100 m/300 ft.a
Antenna
Input Impedance
50 Ω
Water Ingress Protection
Code
IP X1 (protection only against vertically falling water drops)
a. The effective operating range can vary depending on the geographical and constructional characteristics of the building where the system is installed, and
whether underwater monitoring is performed.
3
Transducer Specifications
Transducer Specifications
General
Dimensions and Weight
Battery
Size (diameter)
< 10 cm/3.94 in
Weight
< 140 g/4.8 oz.
Type
Lithium Ion
Capacity
> 16 hours
Life
> 500 charge/discharge cycles
(with new battery, at 25°C/77°F)
Transducer Storage Time
≥ 1 year at 25°C/77°F
(battery full)
(depends on initial battery charge level
and temperature. Storage time is
shortened significantly by high
temperatures)
≥ 1 month at 25°C/77°F
(battery empty)
Recharging Time
100% charged ≤ 2,5 h
66% charged ≤ 1 h
Degree of Protection
Against Electrical Shock
For US, ECG (DECG/MECG), and
Type CF
Toco parameters. Transducers are
battery operated.
Water Ingress Protection
Code
IP 68 (0.5m immersion for 5 hours)
Usability Underwater
0.5m
Shock Resistance
Withstands a 1m drop to concrete surface with possible cosmetic damage
RF Unit
Nominal RF Output Power
0.1 mW ERP (typical)
Channel spacing
25 kHz (12.5 kHz Japan)
Data rate
200 bits/s
Modulation type
Analog
frequency modulation
Digital
FSK 1.6 kHz and 2.4 kHz
Performance Specifications
Frontends
US
US Intensity
Average output power
P = (3.3 ± 0.4) mW
Peak-negative acoustic pressure
p_ = (27.4 ± 4.6) kPa
Output beam intensity (Iob)
Isata = (2.64 ± 0.83) mW/cm2
(= spatial average - temporal average
intensity)
TOCO
ECG
4
Spatial-peak temporal average intensity
Ispta = (7.0 ± 2.3) mW/cm2
Effective radiating area @ -6 dB
1.25 cm2
US Frequency
1 MHz
Signal Range
0 to 127 units
Measurement Range
-100 to 300 units
Resolution
0.25 units
Baseline Drift due to Temperature Changes
1 unit/min/°C (free air)
(When monitoring patients under water, Toco baseline drift may be
higher than usual and require frequent manual adjustment of the
baseline.)
5 units/min/°C (underwater)
Type
Two Lead ECG
Input Impedance
> 10MΩ @ 35 Hz
Input Voltage Range ECG
20 μVpp to 4 mVpp (66dB)
Ordering Guide and Accessory Options
You can order the Avalon CTS Cordless fetal Transducer System under the product number M2720A. To
order, precede the required option number with M2720A (for example, you would order a base station
together with one Toco and one US transducer using M2720A #A02). “K” options let you modify your
order. Below you will find tables giving you a quick overview of the various options. Refer to “Standard
Accessories Included” on page 5 for a list of items shipped as part of the standard configuration.
Order upgrade options by using the prefix M2720AU.
System Hardware Components
System Hardware Components
Product Number
Base Station
M2720A
Cordless Toco Transducer
M2725A
Cordless Ultrasound Transducer
M2726A
Cordless ECG Transducer
M2727A
Antenna
-
Mounting brackets
-
M2720A System Hardware Configuration Options
The system comes in three possible configurations to match individual requirements.
M2720A Options
Option No.
Description
A01
Base station
A02
Base station including 1 Toco and 1 US transducer, 2 belts, and 1 bottle of US gel
A03
Base station including 1 Toco, 1 US and 1 ECG transducer, 3 belts, and 1 bottle of US gel
M2720A Modification Options
M2720A Modification Options
Option No.
K01
Description
Antenna Extension Kit
(see “Antenna Extension Kit” on page 6)
K30
Avalon FM20/FM30 interface cable M2731-60001, for connection to Avalon series fetal monitors (monitor
requires software revision D.00.00 or higher)
K50
Series 50 interface cable
Additional Transducers
Additional Transducers
Part No.
Description
M2725A
Cordless Toco transducer
M2726A
Cordless US transducer
M2727A
Cordless ECG transducer
Standard Accessories Included
The following accessories are supplied with the Avalon CTS:
Standard Accessories
Antenna with rectangular BNC connector
Quantity
1
Power cable
1
Service cable (PC Interface cable for Service Support Tool (M1360-61675))
1
Instructions for Use
1
5
Standard Accessories
Quantity
Documentation DVD-ROM (including Instructions for Use, Service Guide, and Service Support Tool)
1
Training Video (region-free DVD, English only, available in selected geographies)
1
System Accessories and Supplies
Accessory
Order Number
Belts (contain latex)
M1562A
Waterproof Belts
M1562B
Disposable abdominal belts (case of 100)
M2208A
Ultrasound gel
40404-001
DECG Accessories:
DECG reusable legplate adapter cable (with flushing port)
9898 031 37651
New Philips DECG Solution
DECG leg attachment electrode for DECG legplate adapter cable
9898 031 39771
DECG fetal scalp electrode: single spiral, worldwide availability
9898 031 37631
DECG fetal scalp electrode: double spiral, Europe only. Not for USA
9898 031 37641
DECG Accessories:
DECG reusable legplate adapter cable (QwikConnect Plus™)
M1362B
QwikConnect Plus™ Solution
DECG leg attachment electrode for DECG legplate adapter cable
M1349A
DECG fetal scalp electrode: single spiral, worldwide availability
15133E
DECG fetal scalp electrode: double spiral, Europe only. Not for USA
15133D
MECG adapter cable
M1363A
MECG electrodes
40493E
ECG/AUX Transducer
M2727A
Telemetry interface cable
For connection to Avalon fetal monitors
M2731-60501
(cable part number: M2731-60001)
For connection to Series 50 fetal monitors
M2720-61603
Antenna: all bands
0950-2028
Rectangular BNC connector
1250-0076
Battery exchange kit
M2720-64001
Transducer ventilation knob kit
M2720-64002
Upgrade Options
Upgrade Options
Product
Option No.
Description
M2720AU
K01
Antenna Extension Kit
M2720AU
K30
Avalon FM20/FM30 interface cable M2731-60001, for connection to Avalon series
fetal monitors (monitor requires software revision D.00.00 or higher)
M2720AU
K50
Series 50 interface cable
M2720AU
U01
Mounting Kit for Series 50 CL Cart
M2740A
U01
Mounting Kit for Avalon FM20/FM30 Cart
(see “Antenna Extension Kit” on page 6)
Antenna Extension Kit
Antenna Extension Kit
Kit Contents
Description
Antenna extension cable with BNC connectors (1.0 m).
For use with carts or other mounting solutions where the
standard antenna cannot be attached directly to the base station,
or does not provide sufficient transmission range.
Mounting bracket including fixings.
Antenna holder, with screws to attach to mounting bracket. Can
also be fitted on walls or other flat surfaces.
6
Mounting Solutions
There are several ways to mount the Avalon CTS:
• On a dedicated mounting shelf, available for
Avalon FM Cart and Series 50 CL Cart (see
“Upgrade Options” on page 6)
• On top of Series 50 IX/XM/XMO fetal monitors
using mounting brackets
• In a standard cart drawer - base station with
docked transducers fits into Cart CL, CX, CM
(requires Antenna Extension Mounting Kit)
• On top of carts or desks
• GCX mounting adapter for mounting the base
station in a wide variety of situations (order
directly from GCX, part number PH-0042-80)
Avalon CTS mounted
on top of a Series 50
XM monitor
Avalon FM20/FM30
Cart with Avalon CTS
mounting kit
Installation Considerations
Delivery Configuration
Special Installation Considerations
Country
Description
Australia, Austria, Bahrain, Belgium,
Canada, Croatia, Czech Republic, Estonia,
Finland, France, Germany, Greece,
Hungary, Iceland, India, Ireland, Israel,
Italy, Jordan, Kazakhstan, Latvia,
Lebanon, Lithuania, Luxembourg,
Netherlands, New Zealand, Poland,
Portugal, Romania, Russia, Saudi Arabia,
Slovakia, Slovenia, South Africa, Spain,
Sweden, Switzerland, Taiwan, Thailand,
Tunisia, Turkey, United Arab Emirates,
United Kingdom, USA, Venezuela
In its default delivery configuration,
the Avalon CTS is customer
installable, and delivered with
automatic frequency allocation and a
standard antenna, giving a potential
operating range of up to 100m/300ft.
Japan
As above, except that the Avalon
CTS uses fixed frequencies as the
default system configuration.
Norway, Singapore and All Others
For availability, contact your local
Philips Medical sales representative.
The method of installation may vary
according to country-specific
requirements. Contact Philips
Support for assistance.
The default delivery configuration may be
unsuitable under certain circumstances, for
instance:
• you need to set fixed frequencies when
there are other telemetry systems
installed in the same institution (always
applies to Japan). The configuration should
be carried out either by the hospital’s
biomedical department, or by Philips
Support.
• you need to connect the Avalon CTS to
an antenna system because the standard
antenna is not sufficient to cover the area
intended for cordless monitoring. Contact
Philips Support regarding site preparation,
antenna system design and assistance with
your specific requirements.
Antenna System Guidelines
• Avalon CTS is compatible with the M2600-series dual band antenna system components.
• The same site preparation and antenna system design rules for the Philips M2600A adult telemetry system
should be used for the Avalon CTS. Existing antenna systems used for M1310A fetal telemetry may not
comply and may need to be redesigned to work with the Avalon CTS. For further details, contact Philips
Support.
• Avalon CTS systems can share an antenna system with other telemetry devices, such as M1310A or
M2600A. Note that frequency converters used with M2600A adult telemetry cannot be used with the
Avalon CTS. For further details, contact Philips Support.
Contact Philips Support regarding site preparation, antenna system design and assistance with your specific
requirements.
7
Philips Medical Systems is part of
Royal Philips Electronics
Interested?
Asia
Tel: +852 2821 5888
Europe, Middle East, Africa
Tel: +49 7031 463 2254
Would you like to know more about our
imaginative products? Please do not hesitate to
contact us. We would be glad to hear from you.
Latin America
Tel: +55 11 2125 0764
On the web
North America
Tel: +1 800 229 6417
www.medical.philips.com
Via e-mail
medical@philips.com
By fax
© 1995-2006 Koninklijke Philips Electronics N.V.
All Rights Reserved.
Printed in The Netherlands.
+31 40 27 64 887
By postal service
Philips Medical Systems
Global Information Center
P.O Box 1168
5602 BD Eindhoven
The Netherlands
0366
M2720A complies with the requirements of the
Council Directives 93/42/EEC (Medical Device
Directive) and 1999/5/EC (Radio and Telecommunications Terminal Equipment Directive).
S