AT-500N-1, AT-500N-2 - Cannon Instrument Company

ISO 9001
for quality assurance
H i g h P e r f o m a n c e : AT- 5 1 0
L o w C o s t : AT- 5 0 0 N - 1 , AT- 5 0 0 N - 2
CE
marked for
product safety
GLP/GMP
conformed
ISO 14001
environmental
management
systems
M u l t i - c h a n n e l S o f t w a r e : AT- W i n
F D A 2 1 C F R P a r t 1 1 C o m p l i a n t S o f t w a r e : AT- W i n / E R
KYOTO ELECTRONICS
KEM Product Line
A wide selection of KEM Automatic Potentiometric Titrators
models is available from process or quality control in
production to research and development works in laboratory.
High Perfomance : AT-510
CHA-500 Multiple Sample Changer as an option.
Low Cost : AT-500N-1, AT-500N-2
1
KEM Product Line
Multi-Channel Software : AT-Win
FDA 21 CFR Part 11 Compliant Software : AT-Win/ER
Security
Integrity
Audit Trail
Electronic Signatures
2
Features of AT Product Line
AT-510
AT-500N-1 , 2
AT-Win
AT-Win/ER
Main unit
Main unit
PC
PC
Language
English / Japanese
English
English / Japanese
English / Japanese
Data entry
Key-in
Key-in
Keyboard and Mouse
Keyboard and Mouse
1 channel
1 channel
4 channels
4 channels
126 x 70mm LCD
with backlight
16 digits x 2 lines
LCD with backlight
Real time titration curve
Yes
No
Yes
(Multi-line)
Yes
(Multi-line)
GLP/GMP conformed
Yes
Yes
Yes
Yes
10 APB-510 to
AT-510 (one built-in)
10 APB-510 to
AT-510 (one built-in)
40 APB-510 to 4
AT-510 (4 built-in)
40 APB-510 to 4
AT-510 (4 built-in)
-
-
Controlling unit
Multitasking
Display
Connectable number of
APB-510
10 (one built-in)
Burette capacity
50mL, 20mL, 10mL, 5mL, 1mL
Data storage media
PC card
No
Yes
Yes
-
-
FDA 21 CFR Part 11
Need AT-Win/ER
Need AT-Win/ER
No
Yes
Electronic balance
Yes
Yes
Yes
Yes
Preamplifier
STD-510, POT-510
PTA-510, CMT-510
TET-510
STD-510
PTA-510
POT-510
SOFT-CAPE
Yes
Yes
Sample changer
3
No
Screen display on PC Screen display on PC
HD and other devices HD and other devices
Depending on a titrator
-
-
AT-510 Automatic Potentiometric Titrator
The AT-510 Automatic potentiometric titrator is a new version of AT series
offering better operational maneuverability and compactness in size while
inheriting KEM’s many years’ advanced proven technology. This new model
can perform potentiometric titration including acid-base, redox, precipitation,
chelatometric and non-aqueous titrations. With optional peripherals, it can
also perform photometric, polarization and surfactant titrations. When the
multiple sample changer CHA-500 is connected, it can automate measurements
of a number of samples unattended and save both time and man-power with
reliable results and simpler operations, thus applicable to many fields where
quality and inspection control, analytical works and R&D activities are
demanded and conducted.
Features
1. Oversized LCD provides clearer readout of
information such as Titration Curve and Messages
2. Increase in the types of endpoint detection
3. Anti-diffusion structure on titration nozzle of burette
4. GLP/GMP fully conforming
5. Data storing / recalling with PC card
6. KF (volumetric) moisture titration unit as an option
7. COD titration unit as an option
8. Printing on A4/Letter size paper as an option
4
AT-500N Automatic Potentiometric Titrator
Features
1. Low cost but High performance
2. Easy-to-read LCD with a backlight
3. Anti-diffusion titration nozzle tip
4. Safety cover to protect burette
5. Selection of one or two burettes
6. GLP/GMP fully conforming
7. CE-marking level of safety and noise immunity
8. Easy-to-enter colored key pad
9. Three (3) interface ports : Balance, PC and Printer
Configuration of Function Keys
Example of printout
APB
Enter Purge mode of burette or dose volume
Up key
Deliver Titrant
Down key
Fill Titrant and exchange burette unit
Purge key
Purge Titrant
Print
Print key
Print out
Stirrer
Stirrer key
Turn on/off stirrer and select stirring speed
Push this key for two seconds, and the display shows
setting stirring speed.
Method key
Set Methods
Titration key
Set titration parameters
Control key
Set control parameters
Result key
Set calculation or print format
Sample key
Set parameters on a sample including the
number of sample, sample size, etc.
Function key
Set up functions as below:
Key to return to the main screen
display (standby state)
0. Method Copy 3. Changer File 6. Memory Clear
1. Data File
4. Set Check
2. Sample File 5. Periodic Check
Key to switch displays or key functions
and also used for auto input of sample
size measured by an electronic balance
APB
Method
Titration
Control
Result
Sample
Function
5
Key to enter sample number, lot number
and sample size in weight
Parameters can be set with this key.
Functions include:
0.Method File
1.Data File
2.Sample File
3.Changer
4.Re-Titr.
5.Check Titr.
6.Set Check
7.Periodic
8.Buret Check
9.Auto Start
10.Auto Statis.
11.Common Blank
12.pH Cal.Table
13.Card Utility
14.Memory Clear
Time and date can be set with this key.
Setup items include:
0.Interface
1.Date&Time
2.Regist Operator
3.Serial/Version No.
4.LCD Contrast
5.Language
6.Lock
7.Beep
8.pH Polarity
Features of KEM titrators
1. An endpoint can be detected precisely.
350.0
A precise endpoint, which is the most important target
in titration, can be obtained from two detecting factors
of dE (endpoint detecting potential) and dE/dmL (Max
differential potential) without taking undesirable noise
for endpoints.
Potential
mV
Differential
-100.0
2. The endpoint whether single or
multiple can be clearly indicated
and separated in titration.
For example, Level Stop mode
can determine the detected
single endpoint as EP2.
350.0
350.0
mV
Level
mV
-100.0
0.0000
mL
13.0000
-100.0
0.0000
mL
13.0000
7.0000
mL
12.0000
0.0000
mL
12.0000
14.00
3. EP detection range can be set.
User can define titration volume and potential range,
most useful in measurement under the influence of
noise and carbon dioxide.
pH
0.00
4. Delivery speed in titration can be automatically controlled.
Where titration potential change (differential) is smaller, the speed increases; while it
becomes larger, the speed slows down automatically. Optimal control method in chemical
reaction comes from the related combination of delivery speed and control speed : C.S.
Delivery speed
s/mL
2
100
200
300
400
500
0
C.S.=4 (Fast)
Fast : Control Speed (4.0)
C.S.=2 (Medium)
Medium : Control Speed (2.0)
C.S.=0.5 (Slow)
100
200
300
Differential
Slow : Control Speed (0.5)
400
500
600
6
EP detection forms
Various detection forms can be selected suitable for users’ titration needs.
Full
EP stop
Level Stop
320.0
350.0
380.0
mV
mV
mV
-80.0
4.0000
mL
-100.0
12.0000
Titration will last to the preset
volume and finds endpoints. This
is useful for a sample of which
endpoint number is unknown.
Up to 3 EPs can be detected by
AT-500N and 5 EPs by AT-510.
0.0000
mL
13.0000
-130.0
mL
Auto Intersect
12.00
12000
pH
2.00
2.5000
mL
0
7.0000
Two-tangent lines crossing point
is determined as the endpoint,
which is adopted in photometric
titration or the like.
0.0000
mL
0:00:00
0:30:00
0.0000
Reagent is dosed to maintain control potential in Stat titration, and
titration volume will be displayed
at given interval.
7
7.4302
pKa
5.0000
pH
0.00
mL
Two-tangent lines crossing point
is determined as the endpoint,
which is adopted in conductometric
titration or the like.
pH Stat
14.00
12.00
pH
0.00
0.0000
12.8393
Titration will last up to the preset
potential level (up to 2 levels) before stopping.
Titration will last until it finds
preset number of endpoints defined by user on control parameter setup.
Up to 3 EPs can be detected by
AT-500N and 5 EPs by AT-510.
Auto Intersect
0.0000
mL
8.5000
A pKa value can be determined
by finding pH at the half of titration volume in neutralization titration.
Applications of Automatic Titrators
Type of Industry
Object of Analysis
Total acid number of petroleum products
Titration Types
Acid-Bass Titration
Total base number of petroleum products
Petroleum
Foods
Fats and Oils
KOH
98100H171&98100R172
HCIO4 / HCI
or 98100C173
Precipitation Titration
AgNO3
98100C373
Bromine Number
Polarization Titration
NBr, KBrO3
98100M511 or 98100M512 / POT-510
Mercaptan
Precipitation Titration
AgNO3
98100M214&98100H171
Determination of Iodine
Oxdation-Reduction Titration
Na2S2O3
98100C272
HCIO4
98100H171&98100R172
Na(B(C6H5)4)
or 98100C171
Acid-Base Titration
Determination of Benzethonium chloride
Ca and Mg in Dialysis solution
Chelatometric Titration
EDTA
Photo Sensor / PTA-510
Chloride in pharmaceutical products
Precipitation Titration
AgNO3
98100C373
Acid value of beverages
Acid-Base Titration
NaOH
98100C171
Salts in seasoning/dressing
Precipitation Titration
AgNO3
98100C373
Vitamin C
Oxdation-Reduction Titration
indophenol
98100C272
NaOH
98100H171&98100R172
KOH
or 98100C172
Boric acid concentration in nuclear power plant
Electronic Power
Electrode / Preamplifier
Chloride
Determination of Sodium acetate
Medicine
Titrant
Acid-Base Titration
Total Acid Number of insulating oils
Acid Value
Acid-Base Titration
KOH
98100C173
Peroxide Value
Oxdation-Reduction Titration
Na2S2O3
98100C272
Saponification value
Acid-Base Titration
Hydroxy Value
Metal Plating Solution Ni
HCI
98100C173
KOH
Chelatometric Titration
EDTA
Photo Sensor / PTA-510
Precipitation Titration
AgNO3
98100C373
Acid-Base Titration
NaOH
98100C171
Chromic acid in chrome plating bath
Oxdation-Reduction Titration
Na2S2O3
98100C272
Total hardness
Chelatometric Titration
EDTA
Photo Sensor / PTA-510
Chloride
Precipitation Titration
AgNO3
98100C373
Alkalinity
Acid-Base Titration
H2SO4
98100C171
Metal Plating Solution Cu
Electronic and
Metal Planting
Free cyanide concentration
Boric acid in plating baths
Etching Solution H2SO4
City water/
Industrial water
Anionic Surfactants
Detergents
Chemicals
Colloid Titration
Cationic Surfactants
98100S173
SDS
Determination of Isocyanate
Acid-Base Titration
HCI
98100C173
Monomer residues in polyethylene
Oxdation-Reduction Titration
Na2S2O3
98100C272
KOH
98100C173
HCI
98100C171
KOH
98100C173
Carboxyl group in PET resin
Sodium hydroxide
Acid-Base Titration
Purity of acrylic acid
Iron and steel/
Ceramics
Hyamine
Ferric iron in hydrochloric acid
Oxdation-Reduction Titration
Na2S2O3
98100C272
Fluoric acid and nitric acid in pickling solution
Acid-Base Titration
NaOH
98100C171
Trace chlorine in cements
Precipitation Titration
AgNO3
98100I 271&98100R272
Aluminum oxides in cements
Chelatometric Titration
EDTA
Photo Sensor / PTA-510
8
Specifications
Type and Model
Titration mode
Method
Type of titration
Titration form
Preamplifiers
Display
Displayed contents
External input/output
Printer connection
Data storage
GLP/GMP conformity
Structure
Burettes
Steps for a full stroke
Burette precision
Switching for discharge or suction
Liquid contacting materials
Ambient conditions
Power/Consumption
Dimension
Weight
Standard accessories
9
AT-510 Automatic Potentiometric Titrator
AT-500N Automatic Potentiometric Titrator
Blank, Auto control, Auto cut off,
Blank, Auto control, Auto cut off,
Continuous/Intermittent constant rate, Petroleum Continuous/Intermittent constant rate
Products neutralization number, KF titration
control, COD titration control
STD:4, User:16; Total 20
STD:7, User:38, Combined:5; Total 50
Acid-base, Redox, Photometric and Polarization
Acid-base, Redox, Photometric, Polarization,
titration
Combined, pH stat, pKa, Learning, KF moisture
and COD titration
Auto: Max inflection point, Level stop, Intersection Auto: Max inflection point(Full or EP stop titration),
Manual: EP stop, Intersection
Level stop
STD-510, (Option:TET-510, PTA-510, POT-510, CMT-510) STD-510, (Option:PTA-510, POT-510)
LCD with back light, 126 x 70 mm, 30 digits by 13 lines LCD with back light, 16 digits by 2 lines
Titration curves, Titration parameters (titer, potential,
Titration parameters (titer, potential, temperature, etc.),
Messages, etc.
temperature, etc.), Messages, etc.
RS-232C, Electronic balance, Ext. printer, Multiple sample changer
Standard accessory: Impact dot printer (Ink-ribbon Options: Impact dot printer (Ink-ribbon style), IDP-100
style), IDP-100
Options: A4/letter-size printers
Internal: 100 samples
Internal: 100 samples
External: 100 samples per PC card
Calibration history for electrodes, Setting of periodic Setting of periodic inspection interval, Inspection
history for titration performance (one point),
inspection interval, Checking of burette precision,
Alarm for replacing electrodes, Alarm for shortage of File for operator's name (one person)
titrants, Inspection history for titration performance
(one point), File for operator's name (up to 10)
Integrated construction of main unit and burette, Stirrer(984300018A) with a built-in preamplifier
One built-in APB-510 and
One built-in APB-510 (AT-500N-1) or
up to 9 external APB-510 connectable
two built-in APB-510 (AT-500N-2)
Burette capacity: 20mL
Burette capacity: 20mL
Optional burettes (mL): 1, 5, 10, 50
Optional burettes (mL): 1, 5, 10, 50
20,000 steps
20,000 steps
50mL burette
±0.05mL,
Repeatability ±0.02mL
20mL burette
±0.02mL,
Repeatability ±0.01mL
10mL burette
±0.015mL,
Repeatability ±0.005mL
5mL burette
±0.01mL,
Repeatability ±0.003mL
1mL burette
±0.005mL,
Repeatability ±0.001mL
Automatic with ceramic cock
PTFE, glass and ceramics
Temperature: 5 to 35°C; Humidity: less than 85%RH (No condensing)
AC100-240 V, 50/60 Hz; approx. 30W
Main unit: 288(W) x 468(D) x 481(H) mm; Stirrer unit: 118(W) x 225(D) x 336(H) mm
Approx. 10kg
Instruction Manual
1
Instruction Manual
1
Rotor for stirrer(25mm)
985003141
1
Rotor for stirrer(25mm)
985003141
1
Piston removing stick
985515002
1
Piston removing stick
985515002
1
1
1
Combined glass electrode
98100C171
Combined glass electrode
98100C171
Temperature comp. electrode 98100T111
1
Temperature comp. electrode 98100T111
1
Impact dot printer
IDP-100
1
Electrode cable
984290012
1
Electrode cable
984290012
1
System Configuration
Temperature compensation sensor
984300054
SOFT-CAPE
984300036
Propeller stirrer
MS-510
Magnetic stirrer
APB-510
Automatic piston
burette
Standard system
KF-510
Karl Fischer
titration unit
CHA-500
Multiple Sample
Changer
MW-510
Manual washing unit
connecting
cable
980305030
COD-510
COD titration unit
APB-510
Automatic
piston
burette
Magnetic stirrer
984300018A
APB-510
Automatic
piston
burette
connecting cable
980305030
MS-510
Magnetic stirrer
connecting
cable
980305030
MS-510
Magnetic stirrer
MTA-118
Microtitration cell unit
Burette Units
APB-510-01B (1mL)
APB-510-05B (5mL)
APB-510-10B (10mL)
APB-510-20B (20mL) Standard
APB-510-50B (50mL)
Sealed cell unit
SCU-118
Preamplifier
M o d e l
Application
Equipped as
Detected properties
STD-510
POT-510
PTA-510
CMT-510
TET-510
Potentiometric
Polarization
Photometric
Conductometric
3-way titration
Acid-base titration
Redox titration
Precipitation titration
Non-aqueous titration
Potentiometric titration at constant current
Amperometric titration at constant voltage
Acid-base titration
Redox titration
Non-aqueous titration
Chelate titration
Acid-base titration
Redox titration
Precipitation titration
Non-aqueous titration
Conductometric titration
Acid-base titration
Redox titration
Precipitation titration
Non-aqueous titration
Acid-base titration
Redox titration
Precipitation titration
Non-aqueous titration
Option
Standard
Option
Option
Option
1
pH
pH
pH
pH
pH
2
mV
mV
mV
mV
mV
Polar Potential/Current
Transmittance
Conductivity
3
Detection range
1
2
3
Remarks
0
14pH
2000mV
0
14pH
0
2000mV
0 2000mV
0 100 A
Twin Pt. electrode
not included
14pH
2000mV
0
120%T
Photo sensor (530nm)
included
0
14pH
pH
0
2000mV
0 100 S
0 1000 S
0 10000 S
Conductivity sensor included
Exclusive for AT-510
14pH
2000mV
0
14pH
Exclusive for AT-510
10
Options for AT-510 and AT-500-1,2
Features:
Automatic Piston Burette
Titration nozzle of burette is made of PTFE and two
kinds of nozzle, anti-diffusion type and normal type, are
given as standard accessories.
Safety cover to protect glass burette.
Easy replacement of valve with newly designed flowcontroll valve for titrants.
APB-510
Specifications:
Type and model
Display
Burette capacity
Reagent bottle
Power source
Power consumption
Dimension
Weight
APB-510, Automatic Piston Burette
LCD , capable six digits display
1)20mL glass burette (standard)
2)Unit with 50/10/5/1mL glass burette (option)
1000mL brown glass bottle
1) AC100~120 V or AC200~240 V
2) 50/60 Hz
20W
120(W) x 365(D) x 450(H) mm
Approx. 5kg
Features:
Multiple Sample Changer
Can select washing sequences for electrodes suitable for titration applications:
1.Acid-base titration
2.Precipitation titration
3.Standard non-aqueous titration
4.Neutralization number titration on petroleum products
Optional Power Shut-off Unit POU-310 can be connected with CHA-500.
CHA-500-12/CHA-500-18
Specifications:
Type and model
Number of samples
Sample vessels
Rinsing method
Rinsing solution
Drain capabilities
Dosing of pretreatment solution
Data Capture Software
SOFT-CAPE
SOFT-CAPE the Data Capture Software
for AT-510/AT-500N Automatic titrators
Data Capture Software can control Automatic
titrators.
Measurement results of Automatic titrators
can be downloaded directly to workbook in
Microsoft® Excel or filed in CSV format,
such as Lotus1-2-3.
Microsoft® Excel must be installed on the
connected personal computer.
11
Combination Surfactant Electrode
CHA-500 Multiple Sample Changer
12 or 18 samples
Standard: 200mL beaker or 300mL tall beaker
Optional: 50, 100mL beaker or 200mL Erlenmeyer flask
Standard: Shower (equipped with rinse pot)
Optional: Dip; Dip plus shower
Standard: Pure water; Ethanol or IPA (Isopropanol)
Optional: Organic solvents such as toluene; Aqueous acid/alkali solution
Standard: Natural falling drain from rinsing pot
Optional: Mixing/Separating drain by drain pump
Standard: N/A
Optional: APB-510 or filling pump can be connected to CHA-500.
98100S173
Surfactants used to be determine by the Epton's
method with the toxic solvent of Chloroform,
but with this electrode one can measure the
concentration of cationic or anionic surfactants
quickly without Chloroform.
Main specifications
Item
Concentration range
pH range
Temperature range
Specifications
1x10-2M 1x10-5M
pH2 12
0 50 C
Stream Potential Titration Unit
PCD-500
Titration instrument for measuring on surfactants
and polymer coagulants.
A number of samples can be automatically
measured using the Multiple Sample Changer.
Main specifications (not CE conformed)
Item
Specifications
Measured objects Surfactants, polymer coagulants and the colloid solution
with dispersing charged particles in Food industry
Measuring range 1x10-5M or more
Output power -2 to +2 VDC
Power source 100 to 240 VAC, 10W
Options for AT-510 and AT-500N-1,2
Micro Titration Assembly
MTA-118
This assembly is useful for measuring trace
amount of samples by titration.
The structure features the sealed jacket of
thermostatic bath.
Type
Cell shape
MTA-118-1 Tube
MTA-118-5 Cone
MTA-118-50 Cone
Min. sample vol.
1mL
5mL
25mL
Max. sample vol.
2.5mL
35mL
100mL
SCU-118
Sealed Cell Flask Assembly
This flask assembly with an electrode supporting
holder is designed for samples that cannot be
measured under normal ambient conditions.
Titration can be performed with minimal
influence of atmosphere by purging the inside of
the flask with Nitrogen gas or filling the
desiccant tube with Soda lime or the like.
Capacity of titration cell
200mL
Minimum sample required
At least 50mL
Propeller stirrer
984300036
This unit features stirring titrated sample
solution with rotating propeller at the tip of a
motoring rod as the substitute for a magnetic
stirrer. (Exclusive for AT-510 and AT-500N)
Options Exclusive for AT-510
PC Card
The following three types of PC cards are available:
1.Memory card (984340039)
Can store, read and write measurement methods, titration
conditions and other parameters. Further, can perform
recalculation or statistical calculations by recalling
measured data.
2.Application card (984340173)
Can use 70 various measurement conditions (parameters
for such as titration, control or calculation) prepared by us.
3.pH temperature compensation card (984340175)
Can make temperature compensation on pH, based on
international standards, with pH temperature compensation
tables of ASTM, DIN, BS, NF and JIS.
PC Card Read Software
SOFT-PCRE
KF titration unit
KF-510
KF titration unit KF-510
With this unit and the Karl Fischer reagents, one
can measure water in samples.
Features:
1.Can store all data in PC card for backed up on a personal
computer. In reverse, data stored in personal computer
can be transferred to Memory card, meaning that instrument
can be easily recovered.
2.Measurement results can be directly captured to
Microsoft Excel workbook.
3.Measured data can be directly stored as files in CSV format.
This CSV file is compatible with Microsoft® Excel
spreadsheet software package.
4.Personal computer can separately store and manage
measured data, measurement conditions or sample conditions
stored in Memory card. In addition, data stored in personal
computer can be transferred to Memory card.
COD titration unit
COD-510
COD titration unit COD-510
This unit is used for the titration with Potassium
permanganate in the last step of COD measurement.
12
Multi-Channel software perfectly
Software for Titrators
supporting titration
Can run four measurements simultaneously and concurrently.
Can display four titration curves simultaneously.
AT-510
AT-500N-1
AT-500N-2
Examples of connection of AT-510/AT-500N Automatic Potentiometric Titrators
Features
1. Auto settings of measurement conditions with the Navigating feature:
Just by entering user’s intended main criteria including international
standard (e.g. JIS, ASTM, ISO), sample type, measured object and
measuring range, the complete titration conditions can be set up
automatically. Additional powerful help by the Wizard enhances
setting up measurement conditions through dialogue interactions.
2. Operational procedures displaying feature:
Simply select what you want to do-- the necessary procedures for user’s
desired measurement will be displayed sequentially prompting the entry
step by step, allowing operators to avoid unintended or ignorant entry of
numerical constants such as sample size and the like.
3. IT Support:
Details of AT-Win can be viewed at http://www.kem.jp/itsupport/,
where the user can see and download data files on measurement
conditions plus the up-to-date information on titration technique.
13
Features
4. GLP/GMP fully conforming:
Electrode and sensor
Notice of periodic calibration date
Display of electrode calibration records
Burette validation
Notice of periodic check date
Display of burette validation record
Notice of piston head replacement date
Check with standard substance
Notice of periodic check date
Display of check records
Others
Operational level authentication by user
Display of operational records
Alarm for remaining reagent volume
5.Operational management with security measures:
User ID and password security support operational management allowing
solely managerial person to change parameters in order to avoid illegal or
unauthorized alteration of measurement conditions or the like.
<The range of authentication of operational level>
Create, edit, copy or delete titration conditions of Methods.
GLP/GMP requirements including electrode calibration and burette validation.
Analysis of measurement data like re-calculation or re-detection of endpoints.
6. Standard-equipped Viewer linked to Network:
Not only measurement results can be viewed, but also the simulation
of measurements, recalculation or analysis can be manipulated by
other computers installed with Viewer via LAN linked. Of course,
the user can create report and print it out with the Microsoft Word.
The Viewer can be installed on plural personal computers.
7. Creating Report:
Since titration data are now converted to CSV format, they can be
manipulated in other application software like Microsoft ® Excel
spread sheet or Microsoft Access database software on the word-processing Microsoft Word.
Therefore, the user can create Report easily and faster than before in one's
original form, for example, using Microsoft Word's auto function, making
use of its template and macro.
8. Barcode reader:
Sample ID or lot number can be easily input with a barcode reader sold on
the market connectable to the keyboard of user’s personal computer.
14
System Configuration
Standard components
KEM
Titration software
AT-Win
Key Code
When purchasing AT-Win/2nd software, you need
RS-232C EXT. Board (984040008).
CH1
Key Code Card
CH2
CH3
CH4
CH5
CH6
CH7
CH8
Connecting cable (2m)
980300002
AT-Win/2nd(Add-on Set)
KEM
Printer
Key Code
Connecting cable (2m)
980300002
AT-510
Automatic Potentiometric Titrator
APB-510
Automatic Piston Burette
APB-510
Automatic Piston Burette
(Max 9 connectable)
MS-510
Magnetic Stirrer
AT-510
Automatic Potentiometric Titrator
Key Code Card
AT-500N
Automatic Potentiometric Titrator
APB-510
Automatic Piston Burette
APB-510
Automatic Piston Burette
(Max 9 connectable)
CHA-500
Multiple Sample Changer
APB-510
Automatic Piston Burette
980305030
connecting cable
980303388
connecting cable
CHA-500
Multiple Sample Changer
15
APB-510
Automatic Piston Burette
980300013
connecting cable
MS-510
Magnetic Stirrer
CH9
Specifications
Standard components:
Name of software
Compatible model
Media
Connecting cable
AT-Win
AT-Win/2nd
Base set
Add-on set
AT-510/AT-500N
AT-510/AT-500N
CD-ROM 1 pc.
No
Connecting cable(2m)
980300002
1
Key Code Card
Operation manual
Connecting cable(2m)
980300002
1
1*
1*
1 copy
No
* For the purchase of a Key Code Card, the serial number of user's unit is required. This serial number must be notified by the user when
used in combination with the unit already in use.
System requirements:
Device
Computer
Description
Personal computer with CPU of 500MHz or better
* Intel Pentium/Celeron family, AMD K6-2/Athlon/Duron family or
Intel compatible CPU’s
Operating system
Windows 98SE/Me/2000/NT4.0(SP5 or later)/XP
Other programs
Microsoft Internet Explorer (Version 5.01 or later)
Adobe Acrobat Reader (Version 4.0 or later)
RAM
Free hard disk space
128MB or more
50MB or more free disk space
Display
Resolution of 1024x768 or better
Disk drive
Serial port
CD-ROM or DVD-ROM
RS-232C D-sub (male) 9 pin or USB port: 1 port *2
*2 1) Each titrator has to have an extra serial port. Likewise, when a balance or barcode is connected to a PC, a separate serial port will be
required in the computer accordingly.
2) With the optional RS-232C expanded serial port, up to four (4) titrators can be linked through the interface. Note, however, that at
least one slot must be free on PCI bus. Additionally, Low Profile type of PCI bus may not be compatible.
3) If a PC has a built-in modem, one serial port may be already assigned.
Options:
RS-232C EXT. Board
Extension cable
(For AT-510/AT-500N)
Extension cable
(For AT-420N)
984040008
RS-232C expanded serial port (8ch, converting cable included)
980080003
980080004
980305289
983500005
983500006
Connecting cable set (120 VAC) 10m
Connecting cable set (230 VAC) 10m
TCU cable (10m)
Optical cable (30m)
Optical cable (50m)
980305287
980305288
980305289
983500005
983500006
TCU cable (1m)
TCU cable (5m)
TCU cable (10m)
Optical cable (30m)
Optical cable (50m)
16
FDA 21 CFR Part 11 Compliant Software
Security
Integrity
AT-Win/ER Titration Software featuring Electronic Records and
Signatures compliant with FDA 21 CFR Part 11, inheriting
proven user-friendly and competent features of AT-Win
Software for titrators.
Audit Trail
Electronic Signatures
Software specifications:
Item
Description
Name of software
User authentication
Electronic signatures
FDA 21CFR Part 11
compliant features
Data integrity
Audit trail
Backup function
GLP/GMP compliant features
Network link
Compatible models
AT-Win/ER, FDA 21 CFR Part 11 Compliant Software
Login ID and Password security
User name and title registration
Group management for up to 11 user groups
Password valid period
Detection of illegal access and lockout (access number counting and
its time length)
Signatures according to users’ authentication
Verified data is stored in database with comments
Auto recording raw data and meta data in database
Auto recording measurement parameters and results
(Database:compatible with the Microsoft SQL Server/MSDE, Oracle)
Can see stored and compressed data with Viewer
Record created or changed Methods with comments
Auto recording of results and re-evaluated data with all
changes noted.
Search the recorded data
Direct backup recording in CD-R or other device
Backup data can be viewed directly.
Electrode calibration/Check with standard substances/Validation of
burette dosing volume
Notify Periodic calibration/check/validation (by setting date or its interval)
Auto recording of the results of calibration/check/validation in database
Interface to the Shimadzu CLASS-Agent System
AT-510 and AT-500N
* “Windows, Internet Explorer, MSDE and SQL Server”, “Oracle”, “Acrobat” and “CLASS-Agent” are registered trademarks of Microsoft,
Oracle, Adobe Systems and Shimadzu Ltd., respectively.
Standard components:
Name of software
Media
Connecting cable
Key Code Card
Operation manual
Titration software
Database controlling software
AT-Win/ER
AT-Win/2nd
Base set
CD-ROM 1 pc.
CD-ROM 1 pc.
980300002 1piece
1*
One copy
Add-on set
No
No
980300002 1piece
1*
No
* For the issue of a Key Code Card, report of the serial number on a user’s AT unit is a must. When this software is used with the formerly
delivered AT units, be sure to let us know the serial numbers of the concerned AT’s.
Note: Hardware validation is required for user’s computers on which this software is installed. For more details, please contact our local
sales representatives in your country.
17
Features
1. Electronic signatures with ID and Password security
User authentication is secured by login/logout with ID and
password. The password length and validity period can be
precisely set for each user. Illegal access is prevented by
lockout system which records the attempt.
2. User group registration
Up to 11 separate user groups can be registered, each with
its own password entry and individual level of access.
3. Audit trail
The required audit trail feature is fully compliant by
recording data both before and after the change, which is
traceable by the function of log file self-generation.
4. Secured data integrity
Database management of user information (name, title,
e-mail address, etc.) can be secured at higher level by
linking the access-controlled database such as Microsoft
SQL Server/MSDE or Oracle.
5. Electronic records and signatures
Data such as measurement results or conditions, electrode
calibration or check with standard substances are
automatically recorded onto the database, which can be
electronically signed according to an approved workflow.
6. Network compatibility
The network-interfacing feature, also compatible with the
LAN on Client/Server system, allows the users to control
their login records. In addition, the feature includes
the interface to the Shimadzu LC/GC Network.
Titration result
7. Backup system
The backup copies of data and log records together with
the reasons can be made on commercial media such as
CD-R using a CD-R/RW drive and a CD writing software.
Database
18
System Configuration
Database
LC/GC,etc.
Server PC
Client PC
Client PC
Ethernet LAN (Labo)
Ethernet LAN (Office)
When used as stand alone PC,
the database should be installed
on a local PC.
Database
Standard components
FDA21CFR Part11
Compliant software
When purchasing AT-Win/2nd software, you need
RS-232C EXT. Board (984040008).
CH1
KEM
Key Code
Key Code Card
Database
management
software
CLASS-Agent
Connecting cable (2m)
980300002
CH2
CH3
CH4
CH5
CH6
CH7
CH8
CH9
AT-Win/2nd(Add-on Set)
KEM
Key Code
Key Code Card
Connecting cable (2m)
980300002
Printer
AT-510
Automatic Potentiometric Titrator
AT-510
Automatic Potentiometric Titrator
APB-510
Automatic Piston Burette
APB-510
Automatic Piston Burette
(Max 9 connectable)
MS-510
Magnetic Stirrer
AT-500N
Automatic Potentiometric Titrator
APB-510
Automatic Piston Burette
APB-510
Automatic Piston Burette
(Max 9 connectable)
CHA-500
Multiple Sample Changer
APB-510
Automatic Piston Burette
MS-510
Magnetic Stirrer
980305030
connecting cable
980303388
connecting cable
CHA-500
Multiple Sample Changer
APB-510
Automatic Piston Burette
980300013
connecting cable
Distributed by :
KYOTO ELECTRONICS
MANUFACTURING CO.,LTD.
Overseas Division : 8-3 Niban-cho Chiyoda-ku TOKYO 102-0084,JAPAN
Fax : +81-3-3237-0537, Phone : +81-3-3239-7333
URL : http://www.kyoto-kem.com
Specifications and design subject to change for improvements without notice. Printed in Japan.
58IK33C