85320A/B-H50 Mixer Modules Operating and Service Supplement

Add to my manuals
26 Pages

advertisement

85320A/B-H50 Mixer Modules Operating and Service Supplement | Manualzz
$JLOHQW$%+0L[HU
0RGXOHVWR*+]
2SHUDWLQJDQG6HUYLFH0DQXDO6XSSOHPHQW
Agilent 85320A/B H50
Mixer Modules
2.0 to 50 GHz
Operating and Service
Manual Supplement
Use this modification with manuals
p/n 85310-90001 and 85320-90001
Agilent p/n: 85320-90028
Printed in USA
25 August 1998 Revision 4.0
Notice
The information contained in this document is subject to change without
notice.
Agilent makes no warranty of any kind with regard to this material,
including, but not limited to, the implied warranties of merchantability and
fitness for a particular purpose. Agilent shall not be liable for errors
contained herein or for incidental or consequential damages in connection
with the furnishing, performance, or use of this material.
Agilent assumes no responsibility for the use or reliability of its software on
equipment that is not furnished by Agilent.
This document contains proprietary information which is protected by
copyright. All rights are reserved. No part of this document may be
photocopied, reproduced, or translated to another language without prior
written consent of Agilent Technologies.
Restricted Rights Legend
Use, duplication, or disclosure by the U.S. Government is subject to
restrictions as set forth in subparagraph (c)(1)(ii) of the Rights in Technical
Data and Computer Software clause at DFARS 252.227-7013 for DOD
agencies, and subparagraphs (c)(1) and (c)(2) of the Commercial Computer
Software Restricted Rights clause at FAR 52.227-19 for other agencies.
Agilent Technologies
Santa Rosa Systems Division
1400 Fountaingrove Parkway
Santa Rosa, CA 95403-1799, U.S.A.
© Copyright Agilent Technologies 1997–2001
2 Agilent 85320A/B H50 Operating and Service Manual Supplement
Warranty
Certification
Agilent Technologies certifies that this product met its published
specifications at the time of shipment from the factory. Agilent further
certifies that its calibration measurements are traceable to the United States
National Institute of Standards and Technology (NIST, formerly NBS), to the
extent allowed by the Institute’s calibration facility, and to the calibration
facilities of other International Standards Organization members.
Warranty
This Agilent system product is warranted against defects in materials and
workmanship for a period corresponding to the individual warranty periods
of its component products. Instruments are warranted for a period of one
year. During the warranty period, Agilent Technologies will, at its option,
either repair or replace products that prove to be defective.
Warranty service for products installed by Agilent and certain other products
designated by Agilent will be performed at Buyer’s facility at no charge
within Agilent service travel areas. Outside Agilent service travel areas,
warranty service will be performed at Buyer’s facility only upon Agilent’s
prior agreement and Buyer shall pay Agilent’s round trip travel expenses. In
all other areas, products must be returned to a service facility designated by
Agilent.
For products returned to Agilent for warranty service, Buyer shall prepay
shipping charges to Agilent and Agilent shall pay shipping charges to return
the product to Buyer. However, Buyer shall pay all shipping charges, duties,
and taxes for products returned to Agilent from another country.
Agilent warrants that its software and firmware designated by Agilent for
use with an instrument will execute its programming instructions when
properly installed on that instrument. Agilent does not warrant that the
operation of the instrument, or software, or firmware will be uninterrupted or
error free.
LIMITATION OF WARRANTY. The foregoing warranty shall not apply
to defects resulting from improper or inadequate maintenance by Buyer,
Buyer-supplied software or interfacing, unauthorized modification or
misuse, operation outside of the environmental specifications for the
product, or improper site preparation or maintenance.
NO OTHER WARRANTY IS EXPRESSED OR IMPLIED. AGILENT
SPECIFICALLY DISCLAIMS THE IMPLIED WARRANTIES OR
MERCHANTABILITY AND FITNESS FOR A PARTICULAR
PURPOSE.
Agilent 85320A/B H50 Operating and Service Manual Supplement 3
EXCLUSIVE REMEDIES. THE REMEDIES PROVIDED HEREIN ARE
BUYER’S SOLE AND EXCLUSIVE REMEDIES. AGILENT SHALL
NOT BE LIABLE FOR ANY DIRECT, INDIRECT, SPECIAL,
INCIDENTAL, OR CONSEQUENTIAL DAMAGES, WHETHER BASED
ON CONTRACT, TORT, OR ANY OTHER LEGAL THEORY.
Assistance
Product maintenance agreements and other customer assistance agreements
are available for Agilent products.
For assistance, call your local Agilent Sales and Service Office, refer to
Contacting Agilent, page -5.
4 Agilent 85320A/B H50 Operating and Service Manual Supplement
Contacting Agilent
By internet, phone, or fax, get assistance with all your test and
measurement needs.
Online assistance: www.agilent.com/find/assist
United States
(tel) 1 800 452 4844
Japan
(tel) (+81) 426 56 7832
(fax) (+81) 426 56 7840
New Zealand
(tel) 0 800 738 378
(fax) (+64) 4 495 8950
Canada
(tel) 1 877 894 4414
(fax) (905) 282 6495
Latin America
(tel) (305) 269 7500
(fax) (305) 269 7599
Australia
(tel) 1 800 629 485
(fax) (+61) 3 9210 5947
Europe
(tel) (+31) 20 547 2323
(fax) (+31) 20 547 2390
Asia Call Center Numbers
Country
Phone Number
Fax Number
Singapore
1-800-375-8100
(65) 836-0252
Malaysia
1-800-828-848
1-800-801664
Philippines
(632) 8426802
1-800-16510170
(PLDT Subscriber Only)
(632) 8426809
1-800-16510288
(PLDT Subscriber Only)
Thailand
(088) 226-008 (outside Bangkok)
(662) 661-3999 (within Bangkok)
(66) 1-661-3714
Hong Kong
800-930-871
(852) 2506 9233
Taiwan
0800-047-866
(886) 2 25456723
People’s Republic of China
800-810-0189 (preferred)
10800-650-0021
10800-650-0121
India
1-600-11-2929
000-800-650-1101
Agilent 85320A/B H50 Operating and Service Manual Supplement 5
Safety and Regulatory Information
Review this product and related documentation to familiarize yourself with
safety markings and instructions before you operate the instrument. This
product has been designed and tested in accordance with international
standards.
WARNING
The WARNING notice denotes a hazard. It calls attention to a procedure,
practice, or the like, that, if not correctly performed or adhered to, could result
in personal injury. Do not proceed beyond a WARNING notice until the
indicated conditions are fully understood and met.
CAUTION
The CAUTION notice denotes a hazard. It calls attention to an operating
procedure, practice, or the like, which, if not correctly performed or adhered
to, could result in damage to the product or loss of important data. Do not
proceed beyond a CAUTION notice until the indicated conditions are fully
understood and met.
Instrument Markings
!
When you see this symbol on your instrument, you should refer to the instrument’s
instruction manual for important information.
This symbol indicates hazardous voltages.
The laser radiation symbol is marked on products that have a laser output.
This symbol indicates that the instrument requires alternating current (ac) input.
The CE mark is a registered trademark of the European Community. If it is
accompanied by a year, it indicates the year the design was proven.
The CSA mark is a registered trademark of the Canadian Standards Association.
1SM1-A
This text indicates that the instrument is an Industrial Scientific and Medical Group 1
Class A product (CISPER 11, Clause 4).
This symbol indicates that the power line switch is ON.
This symbol indicates that the power line switch is OFF or in STANDBY position.
6 Agilent 85320A/B H50 Operating and Service Manual Supplement
Safety Earth
Ground
Before Applying
Power
This is a Safety Class I product (provided with a protective earthing
terminal). An uninterruptible safety earth ground must be provided from the
main power source to the product input wiring terminals, power cord, or
supplied power cord set. Whenever it is likely that the protection has been
impaired, the product must be made inoperative and secured against any
unintended operation.
Verify that the product is configured to match the available main power
source as described in the input power configuration instructions in this
manual. If this product is to be powered by autotransformer, make sure the
common terminal is connected to the neutral (grounded) side of the ac power
supply.
Agilent 85320A/B H50 Operating and Service Manual Supplement 7
Typeface Conventions
•
Used to emphasize important information:
Use this software only with the Agilent xxxxxX system.
•
Used for the title of a publication:
Refer to the Agilent xxxxxX System-Level User’s Guide.
•
Used to indicate a variable:
Type LOAD BIN filename.
Instrument Display
•
Used to show on-screen prompts and messages that you will see on the
display of an instrument:
The Agilent xxxxxX will display the message CAL1 SAVED.
[Keycap]
•
Used for labeled keys on the front panel of an instrument or on a
computer keyboard:
Press [Return].
{Softkey}
•
Used for simulated keys that appear on an instrument display:
Press {Prior Menu}.
User Entry
•
Used to indicate text that you will enter using the computer keyboard;
text shown in this typeface must be typed exactly as printed:
Type LOAD PARMFILE
•
Used for examples of programming code:
Italics
#endif // ifndef NO_CLASS
Path Name
•
Used for a subdirectory name or file path:
Edit the file usr/local/bin/sample.txt
Computer Display
•
Used to show messages, prompts, and window labels that appear on a
computer monitor:
The Edit Parameters window will appear on the screen.
•
Used for menus, lists, dialog boxes, and button boxes on a computer
monitor from which you make selections using the mouse or keyboard:
Double-click EXIT to quit the program.
8 Agilent 85320A/B H50 Operating and Service Manual Supplement
Contents
Notice . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Restricted Rights Legend . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Certification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Assistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
Safety and Regulatory Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Safety Earth Ground . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
Before Applying Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
Typeface Conventions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Operating Note . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Internal Crystal Detector Output Voltage Label . . . . . . . . . . . . . . . . . 1-2
Specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Operating Frequency Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Maximum Input Levels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Table 1. LO Input Levels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Table 2. Conversion Loss (LO at +12 dBm). . . . . . . . . . . . . . . . . . . . . 3
General Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Connector Types . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Weight (Approximate) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Size (Excluding Connectors) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
85310A System Performance Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
Typical System Performance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
Table 3. Fundamental and 3rd Harmonic Performance. . . . . . . . . . . . 4
Table 4. 5th Harmonic Mode Performance . . . . . . . . . . . . . . . . . . . . . 5
Table 5. 7th Harmonic Mode Performance . . . . . . . . . . . . . . . . . . . . . 5
Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Replaceable Parts and Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Rebuild/Exchange Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Replaceable Parts, 85320A H50 . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
85320A H50 Component Layout . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
Replaceable Parts, 85320B H50 . . . . . . . . . . . . . . . . . . . . . . . . . 1-9
85320B H50 Component Layout . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-10
Mechanical Mounting Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-11
Appendix A . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1
Optimizing LO Drive Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1
Determining Detector Voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1
Figure A-1. Equipment Configuration . . . . . . . . . . . . . . . . . . . . . . . . . 1
Agilent 85320A/B H50 Operating and Service Manual Supplement Contents-1
New Voltage Marking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-2
ADD New Voltage Marking to Agilent 85320B Label . . . . . . . A-2
Figure A-2. Agilent 85320B Millimeter Scale Value Marking . . . . . . . 2
Contents-2 Agilent 85320A/B H50 Operating and Service Manual Supplement
Introduction
Introduction
This manual supplement describes the differences in the 85320A H50 and
the 85320B H50 as compared to the standard 85320A/B. It also describes the
manual changes necessary to document the Agilent 85320A/B option H50
Description
The 85320A H50 and the 85320B H50 are identical in function to the
standard 85320A and 85320B except that the frequency range of operation is
2 to 50 GHz instead of 2 to 26.5 GHz.
The 85320A H50 test mixer and the 85320B H50 reference mixer are
designed for use with the 85309A distributed frequency converter as
described in the 85320A/B Mixer Module User’s Guide. These mixers allow
operation over the full 2 to 50 GHz frequency range. See the System
Performance Notes section for more details.
Agilent 85320A/B H50 Operating and Service Manual Supplement 1
Operating Note
Operating Note
Internal Crystal
Detector Output
Voltage Label
The 85320B H50 reference mixer module has a label attached to it; the
bottom half of the label has a millivolt value scale marking. The marking
indicates what the internal crystal detector output voltage is when
approximately +16 dBm, at 6 GHz, is being applied to the 85320B LO Input
port.
The marked reading value can be used to set the 85309A front panel display
reading. This value should provide the required LO drive level to the
85320 mixers when used in an 85310 system configuration.
If the LO drive coax cables used to conduct this 2-18 GHz drive signal from
the 85309A to 85320 mixers has an excessive amount of loss, either by
being extremely long or lossy, the label reading will not be achievable over
the full 2 to 18 GHz range.
A failure to achieve this setting on the 85309A is indicated by the “Red”
front panel indicator coming on during some portion of the 2-18 GHz LO
frequency band. If the “Red” indicator flashes ON, refer to Appendix A for
an alternate L.O. drive power setting.
NOTE
The new drive power setting will require that you lower the drive power to
as low as the minimum +12 dBm level.
2 Agilent 85320A/B H50 Operating and Service Manual Supplement
Specification
Specification
Operating Frequency
Range
2 to 18 GHz (fundamental mixing mode)
Maximum Input
Levels
Maximum RF level at RF and LO inputs (damage level): +20 dBm
18 to 50 GHz (3rd harmonic mode)
Table 1
LO Frequency
Minimum Power
Typical Power
Maximum Power
2 - 18 GHz
+12 dBm
+14 dBm
+17 dBm
Table 2
General Specifications
LO Input Levels
Conversion Loss (LO at +12 dBm)
RF Frequency
LO Harmonic
LO Frequency
Typical Loss
Maximum Loss
2 - 18 GHz
fundamental
2 - 18 GHz
-6 dB
15 dB
18 - 50 GHz
3 rd.
6 - 16.7 GHz
-24 dB
30 dB
Connector Types
Type-N female except for R.F. Input (2.4 mm male).
Weight (Approximate)
Net Weight:
85320A H50: 794 g (1.75 lb)
85320B H50: 1021 g (2.25 lb).
Size (Excluding Connectors)
NOTE
85320A H50:
width = 97 mm (3.8 in.)
length = 122 (4.8 in.)
height = 34 mm (1.3 in.)
85320B H50:
width = 97 mm (3.8 in.)
length = 186 mm (7.3 in.)
height = 31 mm (1.2 in.)
Mechanical mounting information can be found on page 10 of this
document.
Agilent 85320A/B H50 Operating and Service Manual Supplement 3
85310A System Performance Notes
85310A System Performance Notes
System performance data (when the 85309A, 85320A H50,
85320B H50, and the 8510/8530 are combined in a system) is shown below.
Refer to “Table 5-3” in the General Information section of the
85310A Operating and Service manual as a reference for the following
tables.
Typical System
Performance
Table 3
Fundamental and 3rd Harmonic Performance
System Parameter
2 to 3
3 to 18
-dBm
120
118
103
100
-dBm
18
19
15
25
Dynamic Range
dB
102
99
88
76
Channel Isolation
dB
118
118
110
95
Input Return Loss
-dB
6
6
6
6
1
Sensitivity
2
Compression Level
3
4
(S/N=1, 0 average)
(at 0.1 dB)
GHz
18 to 40*
40 to 50*
* 3rd Harm. Mode
1. Sensitivity is the calculated difference between IF noise and RF/IF conversion gain/loss. Averaging will improve
sensitivity by 10 log (avg. factor).
2. RF level for 0.1 dB compression: the RF input level where the RF and the IF levels are no longer tracking each other
linearly within 0.1dB.
3. Dynamic range is the calculated difference between 0.1dB compression and sensitivity.
4. Crosstalk is the coherent RF leakage from the reference channel to the test channel with 1024 averages.
4 Agilent 85320A/B H50 Operating and Service Manual Supplement
85310A System Performance Notes
Table 4
5th Harmonic Mode Performance
1
System Performance (Typical)
GHz
Sensitivity
-dBm
85
83
83
83
Compression (@ 0.1 dB)
-dBm
22
26
21
21
dB
95
95
95
90
Channel Isolation
10 to 14
14 to 30
30 to 33
33 to 50
1. 14 to 50 GHz is the maximum recommended range of operation.
Table 5
7th Harmonic Mode Performance
1
System Performance (Typical)
GHz
20 to 33
33 to 50
Sensitivity
-dBm
72
75
Compression (@ 0.1 dB)
-dBm
17
26
Channel Isolation
dBm
88
88
1. 33 to 50 GHz is the maximum recommended range of operation.
Agilent 85320A/B H50 Operating and Service Manual Supplement 5
Service
Service
Replaceable Parts and
Accessories
Accessories
Description
Part Number
Qty
RF adapter, 2.4 (f) / (f), for each module
1250-2188
1
Mounting bolts, 1/4-28x1/2”. for each module
2940-0287
4
Manual, Operating and Service
85320-90001
1
Manual, Operating and Service Manual Supplement
85320-90028
1
Rebuild/Exchange Parts
This program does not apply to the 85320A/B H50.
6 Agilent 85320A/B H50 Operating and Service Manual Supplement
Service
Replaceable Parts, 85320A H50
The 85320A H50 consists of the following replaceable parts and assemblies.
See Figure 1.
Description
Part Number
Qty
Screw, LID, SMM 3.5, T10 Torx
0515-0458
8
O-ring, 0.364 ID, (not shown)
0900-0012
2
Connector, 2.4 (m) / (f), RF IN port
5062-6615
1
Diplexer (test mixer)
5062-7578
1
*Mixer, 2 to 50 GHz
5086-7601
1
Connector, SMA (f) / N (f), LO / RF IN port
08340-60221
1
*Gasket, housing lid
85320-00002
1
*Housing
85320-20014
1
*Housing lid
85320-20015
1
*RF cable assembly, 0.086 rigid
85320-20018
1
*RF cable assembly, 0.086 rigid
85320-20019
1
Cable assembly
85320-60003
1
*These special parts are Agilent manufacturing division controlled
Agilent 85320A/B H50 Operating and Service Manual Supplement 7
Service
85320A H50
Component Layout
Figure 1
85320A H50 Internal Parts Locations
8 Agilent 85320A/B H50 Operating and Service Manual Supplement
Service
Replaceable Parts, 85320B H50
The 85320B H50 consists of the following replaceable parts and assemblies.
Description
Part Number
Qty
Screw, lid, SMM 3.5, T10 Torx
0515-2038
10
Connector, SMA (f) / N (f)
08340-60221
3
O-ring, 0.364 ID (not shown)
0900-0012
4
Coupler, 1 to 20 GHz, SMA
0955-0510
1
Diode detector 0.1 to 26.5 GHz
33330C
1
Connector assembly, 2.4 (m) / (f), RF IN port
5062-6615
1
*Mixer, 2 to 50 GHz
5086-7601
1
*Gasket, housing lid
85320-00001
1
*Housing
85320-20012
1
*Housing lid
85320-20013
1
*RF cable assembly, 0.086 rigid
85320-20016
1
*RF cable assembly, 0.086 rigid
85320-20017
1
Cable assembly
85320-60004
1
Cable assembly
85320-60005
1
*These special parts are Agilent manufacturing division controlled
Agilent 85320A/B H50 Operating and Service Manual Supplement 9
Service
85320B H50
Component Layout
Figure 2
85320B H50 Internal Parts Locations
10 Agilent 85320A/B H50 Operating and Service Manual Supplement
Mechanical Mounting Information
Mechanical Mounting Information
Figure 3
Mechanical Mounting Information
Agilent 85320A/B H50 Operating and Service Manual Supplement 11
Mechanical Mounting Information
12 Agilent 85320A/B H50 Operating and Service Manual Supplement
Appendix A
Appendix A
Optimizing LO Drive Power
There are special system configurations in which the recommended
“detector voltage” reading on the Agilent 85320B mixer module is not
compatible. The “detector voltage” value that is on the mixer module
assumes that the LO coax cables (B2, B3) used to interface the Agilent
85309A to the Agilent 85320 mixers are similar to the Agilent 85381A cable
assembly that is not greater than 5 meters in length or a cable loss of less
than 5 dB at 18 GHz.
Determining Detector
Voltage
Procedure
Figure A-1
To determine the correct “detector voltage” for your special application,
follow this procedure.
1. Configure your equipment as shown in Figure A-1 below.
Equipment Configuration
2. Calibrate and Zero the power meter and sensor
3. Connect the power sensor to the end of the “B3” cable.
Agilent 85320A/B H50 Operating and Service Manual Supplement A-1
Appendix A
4. Set the Agilent 85309A front panel display reading to the value
indicated on the Agilent 85320B mixer.
5. Manually set the RF source so that it will slowly sweep across the LO
frequency range that you wish to operate in. Note power meter settling
time requirement.
6. If the Red front panel indicator light on the Agilent 85309A does not
flash ON and the minimum power measured by the power sensor (minus
the 10 dB attenuator loss) is > +12 dBm, then use this detector voltage
setting.
7. If the Red indicator light flashes, readjust the Agilent 85309A front
panel adjustment to a lower value until the Red indicator light no longer
flashes. Now insure that the minimum measured power is > +12 dBm.
If you are not able to achieve a > +12 dBm power reading with no Red
indicator, you have one of these possible problems:
a. B2 and B3 cable length are too lossy, find alternative.
b. Cable B3 or B2 are not equal or one is defective.
c. The Agilent 85309A is defective. Insure that the power levels at
ports J2 and J3 (without the Agilent 85320B connected) exceed their
specified levels.
New Voltage Marking
ADD New Voltage Marking to Agilent 85320B Label
For future reference in this special system configuration, it is recommended
that you re-mark the Agilent 85320B millivolt scale to indicate the new
system detector voltage value.
Next to the new marked value, note the minimum LO power level measured
during the measurement portion of this procedure.
Figure A-2
Agilent 85320B Millimeter Scale Value Marking
A-2 Agilent 85320A/B H50 Operating and Service Manual Supplement