Rosemount 848T Specifications

Add to my manuals
136 Pages

advertisement

Rosemount 848T Specifications | Manualzz
Reference Manual
00809-0100-4697, Rev GA
November 2014
Rosemount 848T High Density Temperature
Transmitter with FOUNDATION™ fieldbus
Device Revision 8
Title Page
Reference Manual
00809-0100-4697, Rev GA
November 2014
Rosemount 848T High Density Temperature
Transmitter with FOUNDATION™ fieldbus
Read this manual before working with the product. For personal and system safety, and for
optimum product performance, make sure to thoroughly understand the contents before
installing, using, or maintaining this product.
The United States has two toll-free assistance numbers and one international number
Customer Central
1-800-999-9307 (7:00 a.m. to 7:00 p.m. CST)
National Response Center
1-800-654-7768 (24 hours a day)
Equipment service needs
International
1-(952) 906-8888
The products described in this document are NOT designed for nuclear-qualified
applications.
Using non-nuclear qualified products in applications that require nuclear-qualified
hardware or products may cause inaccurate readings.
For information on Rosemount nuclear-qualified products, contact an Emerson Process
Management Sales Representative.]
i
Title Page
November 2014
ii
Reference Manual
00809-0100-4697, Rev GA
Reference Manual
Contents
00809-0100-4697, Rev GA
November 2014
1Section 1: Introduction
1.1 Safety messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.1.1 Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.2 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.2.1 Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.2.2 Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.3 Service support. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2Section 2: Installation
2.1 Safety messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.1.1 Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.2 Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.2.1 Mounting to a DIN rail without an enclosure . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.2.2 Mounting to a panel with a junction box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.2.3 Mounting to a 2-in. pipe stand . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.3 Wiring. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.3.1 Connections. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.3.2 Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
2.3.3 Surges/transients . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
2.4 Grounding . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
2.5 Switches. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
2.6 Tagging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
2.7 Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
2.7.1 Using cable glands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
2.7.2 Using conduit entries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
3Section 3: Configuration
3.1 Safety messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19
3.1.1 Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19
3.2 Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20
3.2.1 Standard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20
3.2.2 Transmitter configuration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20
3.2.3 Custom configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20
3.2.4 Methods . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20
3.2.5 Alarms. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21
3.2.6 Damping. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21
3.2.7 Configure the differential sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21
3.2.8 Configure measurement validation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22
Contents
iii
Reference Manual
Contents
00809-0100-4697, Rev GA
November 2014
3.3 Common configurations for high density applications. . . . . . . . . . . . . . . . . . . . . .22
3.3.1 Interfacing analog transmitters to Foundation fieldbus . . . . . . . . . . . . . . .25
3.4 Block configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25
3.4.1 Resource block . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25
3.4.2 Field Diagnostics & PlantWeb Alerts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .31
3.4.3 PlantWeb alerts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .32
3.4.4 Recommended actions for PlantWeb alerts . . . . . . . . . . . . . . . . . . . . . . . . .35
3.4.5 Field Diagnostics Alerts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .36
3.4.6 Recommended actions for field diagnostics alerts. . . . . . . . . . . . . . . . . . . .40
3.4.7 Transducer blocks. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .41
3.4.8 Transducer block sub-parameter tables . . . . . . . . . . . . . . . . . . . . . . . . . . . . .47
4Section 4: Operation and Maintenance
4.1 Safety messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .53
4.1.1 Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .53
4.2 Foundation™ fieldbus information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .53
4.2.1 Commissioning (addressing) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .54
4.3 Hardware maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .54
4.3.1 Sensor check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .54
4.3.2 Communication/power check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .54
4.3.3 Resetting the configuration (RESTART) . . . . . . . . . . . . . . . . . . . . . . . . . . . . .55
4.4 Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .55
4.4.1 Foundation fieldbus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .55
4.4.2 Resource block . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .56
4.4.3 Transducer block troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .56
AAppendix A: Reference Data
A.1 Functional specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .57
A.1.1 Inputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .57
A.1.2 Outputs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .57
A.1.3 Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .57
A.1.4 Ambient temperature limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .57
A.1.5 Isolation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .57
A.1.6 Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .57
A.1.7 Transient protection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .57
A.1.8 Update time. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .58
A.1.9 Humidity limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .58
A.1.10Turn-on time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .58
iv
Contents
Reference Manual
Contents
00809-0100-4697, Rev GA
November 2014
A.1.11Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .58
A.1.12Backup Link Active Scheduler (LAS) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .58
A.2 Physical specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .58
A.2.1 Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .58
A.2.2 Entries for optional junction box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .58
A.2.3 Environmental ratings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .59
A.3 Function blocks. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .59
A.3.1 Analog input (AI). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .59
A.3.2 Input selector (ISEL) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .59
A.3.3 Multiple analog input block (MAI) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .59
A.4 Performance specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .60
A.4.1 Stability. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .60
A.4.2 Self calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .60
A.4.3 Vibration effect . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .60
A.4.4 Electromagnetic compatibility compliance testing . . . . . . . . . . . . . . . . . . .60
A.4.5 Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .60
A.4.6 Differential configuration notes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .61
A.4.7 Accuracy for differential configurations . . . . . . . . . . . . . . . . . . . . . . . . . . . . .61
A.4.8 Analog sensors 4–20mA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .61
A.4.9 Ambient temperature effect . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .62
A.4.10Ambient temperature effects . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .62
A.4.11Ambient temperature notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .63
A.5 Dimensional drawings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .64
A.5.1 Mounting options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .66
A.6 Ordering information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .67
BAppendix B: Product Certificates
B.1 North American Approvals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .69
B.1.1 European Approvals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .72
B.2 Intrinsically Safe and Non-Incendive installations . . . . . . . . . . . . . . . . . . . . . . . . . .78
B.3 Installation drawings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .79
CAppendix C: Foundation™ fieldbus Technology
C.1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .89
C.2 Function blocks. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .89
C.3 Device descriptions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .90
C.4 Block operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .91
C.4.1 Instrument- specific function blocks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .91
Contents
v
Reference Manual
Contents
00809-0100-4697, Rev GA
November 2014
C.4.2 Alerts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .91
C.5 Network communication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .91
C.5.1 LAS. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .92
C.5.2 Addressing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .93
C.5.3 Scheduled transfers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .94
C.5.4 Unscheduled transfers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .95
C.5.5 Function block scheduling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .96
DAppendix D: Function Blocks
D.1 Analog Input (AI) function block . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .97
D.1.1 Functionality . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
D.1.2 AI block troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105
D.2 Multiple analog input (MAI) function block. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106
D.2.1 Functionality . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 108
D.2.2 MAI block troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 112
D.3 Input selector function block . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113
D.3.1 Functionality . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115
D.3.2 ISEL block troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119
vi
Contents
Section 1: Introduction
Reference Manual
November 2014
00809-0100-4697, Rev GA
Section 1
1.1
Introduction
Safety messages
Instructions and procedures in this section may require special precautions to ensure the safety
of the personnel performing the operations. Information that potentially raises safety issues is
indicated by a warning symbol ( ). Refer to the following safety messages before performing
an operation preceded by this symbol.
1.1.1
Warnings
Failure to follow these installation guidelines could result in death or serious injury.
Make sure only qualified personnel perform the installation.
Process leaks could result in death or serious injury.

Do not remove the thermowell while in operation. Removing while in operation may
cause process fluid leaks.

Install and tighten thermowells and sensors before applying pressure, or process
leakage may result.
Electrical shock could cause death or serious injury.



If the sensor is installed in a high voltage environment and a fault condition or
installation error occurs, high voltage may be present on transmitter leads and
terminals.
Use extreme caution when making contact with the leads and terminals.
1.2
Overview
1.2.1
Transmitter
The Rosemount 848T is optimal for process temperature measurement because of its ability to
simultaneously measure eight separate and independent temperature points with one
transmitter. Multiple temperature sensor types may be connected to each transmitter. In
addition, the 848T can accept 4-20 mA inputs. The enhanced measurement capability of the
848T allows it to communicate these variables to any FOUNDATION™ fieldbus host or
configuration tool.
1.2.2
Manual
This manual is designed to assist in the installation, operation, and maintenance of the
Rosemount 848T Temperature Transmitter.
Introduction
1
Reference Manual
Section 1: Introduction
00809-0100-4697, Rev GA
November 2014
Section 1: Introduction

Overview

Considerations

Return of Materials
Section 2: Installation

Mounting

Installation

Wiring

Power Supply

Commissioning
Section 3: Configuration

FOUNDATION fieldbus Technology

Configuration

Function Block Configuration
Section 4: Operation and Maintenance

Hardware Maintenance

Troubleshooting
Appendix A: Reference Data

Specifications

Dimensional Drawings

Ordering Information
Appendix B: Product Certificates

Hazardous Locations Certificates

Intrinsically Safe and Non-Incendive Installations

Installation Drawings
Appendix C: Foundation™ fieldbus Technology

Device Descriptions

Block Operation
Appendix D: Function Blocks
2

Analog Input (AI) Function Block

Multiple Analog Input (MAI) Function Block

Input Selector Function Block
Introduction
Reference Manual
Section 1: Introduction
00809-0100-4697, Rev GA
1.3
November 2014
Service support
To expedite the return process in North America, call the Emerson Process Management
National Response Center toll-free at 800-654-7768. This center, available 24 hours a day, will
assist with any needed information or materials.
The center will ask for the following information:

Product model

Serial numbers

The last process material to which the product was exposed
The center will provide the following:

A Return Material Authorization (RMA) number

Instructions and procedures that are necessary to return goods that were exposed to
hazardous substances
For other locations, contact an Emerson Process Management sales representative.
Note
If a hazardous substance is identified, a Material Safety Data Sheet (MSDS), required by law to be
available to people exposed to specific hazardous substances, must be included with the
returned materials.
Introduction
3
Section 1: Introduction
November 2014
4
Reference Manual
00809-0100-4697, Rev GA
Introduction
Section 2: Installation
Reference Manual
November 2014
00809-0100-4697, Rev GA
Section 2
Installation
Safety messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Grounding . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Switches . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Tagging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2.1
page 5
page 5
page 8
page 11
page 14
page 14
page 16
Safety messages
Instructions and procedures in this section may require special precautions to ensure the safety
of the personnel performing the operations. Information that potentially raises safety issues is
indicated by a warning symbol ( ). Refer to the following safety messages before performing
an operation preceded by this symbol.
2.1.1
Warnings
Failure to follow these installation guidelines could result in death or serious injury.
Make sure only qualified personnel perform the installation.
Process leaks could result in death or serious injury.

Do not remove the thermowell while in operation. Removing while in operation may
cause process fluid leaks.

Install and tighten thermowells and sensors before applying pressure, or process
leakage may result.
Electrical shock could cause death or serious injury.



2.2
If the sensor is installed in a high voltage environment and a fault condition or
installation error occurs, high voltage may be present on transmitter leads and
terminals.
Use extreme caution when making contact with the leads and terminals.
Mounting
The 848T is always mounted remote from the sensor assembly. There are three mounting
configurations.
Installation

To a DIN rail without an enclosure

To a panel with an enclosure

To a 2-in pipe stand with an enclosure using a pipe mounting kit
5
Reference Manual
Section 2: Installation
00809-0100-4697, Rev GA
November 2014
2.2.1
Mounting to a DIN rail without an enclosure
To mount the 848T to a DIN rail without an enclosure, follow these steps:
1.
Pull up the DIN rail mounting clip located on the top back side of the transmitter.
2.
Hinge the DIN rail into the slots on the bottom of the transmitter.
3.
Tilt the 848T and place onto the DIN rail. Release the mounting clip. The transmitter
should be securely fastened to the DIN rail.
Figure 2-1. Mounting the 848T to a DIN Rail
B
C
A
A. DIN Rail Mounting Clip
B. 848T without installed enclosure
C. DIN Rail
2.2.2
Mounting to a panel with a junction box
When inside of a plastic or aluminum junction box, the 848T mounts to a panel using four 1/4-20
x 1.25-in. screws.
When inside of a stainless steel junction box, the 848T mounts to a panel using two 1/4-20 x
/2-in. screws.
1
Figure 2-2. Mounting the 848T Junction Box to a Panel
Aluminum/plastic
Stainless steel
A
E
B
C
C
D
D
A. 848T with aluminum or plastic box
B. Cover screws (4)
C. Mounting screws (4 for aluminum/plastic mounting, 2 for stainless steel mounting)
D. Panel
E. 848T with a stainless steel box
6
Installation
Reference Manual
Section 2: Installation
00809-0100-4697, Rev GA
2.2.3
November 2014
Mounting to a 2-in. pipe stand
Use the optional mounting bracket (option code B6) to mount the 848T to a 2-in. pipe stand
when using a junction box.
Figure 2-3. Mounting the 848T to a 2-in. Pipe Stand Using a Junction Box
Aluminum/plastic junction box
(styles JA and JP)
Front view
A
Side view
Stainless steel junction box
(style JS)
Front view
Side view
B
C
D
E
Dimensions are in inches (millimeters).
A. 5.1 (130)
B. 10.2 (260)
C. 6.6 (167) fully assembled
D. 4.7 (119)
E. 7.5 (190) fully assembled
Aluminum/plastic junction box
mounted on a vertical pipe
Installation
Stainless steel junction box
mounted on a vertical pipe
7
Reference Manual
Section 2: Installation
00809-0100-4697, Rev GA
November 2014
2.3
Wiring
If the sensor is installed in a high-voltage environment and a fault condition or installation error
occurs, the sensor leads and transmitter terminals could carry lethal voltages. Use extreme
caution when making contact with the leads and terminals.
Note
Do not apply high voltage (e.g. AC line voltage) to the transmitter terminals. Abnormally high
voltage can damage the unit (bus terminals are rated to 42.4 VDC).
Figure 2-4. 848T Transmitter Field Wiring
B
A
C
D
F
F
G
E
H
A. Integrated Power Conditioner and Filter
B. 6234 ft (1900 m) max (depending on cable characteristics)
C. Terminators (Trunk)
D. Power supply
E. FOUNDATION™ fieldbus host or configuration tool
F. Spur
G. Signal wiring
H. Devices 1-16 (Intrinsically safe installations may allow fewer devices per I.S. barrier)
2.3.1
Connections
The 848T transmitter is compatible with 2 or 3-wire RTD, thermocouple, ohm, and millivolt
sensor types. Figure 2-5 shows the correct input connections to the sensor terminals on the
transmitter. The 848T can also accept inputs from analog devices using the optional analog
input connector. Figure 2-6 shows the correct input connections to the analog input connector
when installed on the transmitter. Tighten the terminal screws to ensure proper connection.
8
Installation
Reference Manual
Section 2: Installation
00809-0100-4697, Rev GA
November 2014
Figure 2-5. Sensor Wiring Diagram
1 2 3
A
1
2
3
B
1
2
C
3
1
2
3
D
A. 2-wire RTD and ohms
B. 3-wire RTD and ohms (Emerson provides 4-wire sensors for all single-element RTDs; use these RTDs in 3-wire
configurations by clipping the fourth lead or leaving it disconnected and insulated with electrical tape.)
C. Thermocouples/ohms and Millivolts
D. 2-wire RTD with compensation loop (transmitter must be configured for a 3-wire RTD in order to recognize an RTD
with a compensation loop)
RTD or ohm inputs
Various RTD configurations, including 2-wire and 3-wire are used in industrial applications. If the
transmitter is mounted remotely from a 3-wire RTD, it will operate within specifications,
without recalibration, for lead wire resistances of up to 60 ohms per lead (equivalent to 6,000
feet of 20 AWG wire). If using a 2-wire RTD, both RTD leads are in series with the sensor element,
so errors can occur if the lead lengths exceed one foot of 20 AWG wire. Compensation for this
error is provided when using 3-wire RTDs.
Thermocouple or millivolt inputs
Use appropriate thermocouple extension wire to connect the thermocouple to the transmitter.
Make connections for millivolt inputs using copper wire. Use shielding for long runs of wire.
Analog inputs
The analog connector converts the 4–20 mA signal to a 20–100 mV signal that can be read by
the 848T and transmitted using FOUNDATION fieldbus.
Use the following steps when installing the 848T with the analog connector:
1.
The 848T, when ordered with option code S002, comes with four analog connectors.
Replace the standard connector with the analog connector on the desired channels.
2.
Wire one or two analog transmitters to the analog connector according to Figure 2-6.
There is space available on the analog connector label for identification of the analog
inputs.
Note
Power supply should be rated to support the connected transmitter(s).
3.
If the analog transmitters can communicate using HART® protocol, the analog
connectors are supplied with the ability to switch in a 250 ohm resistor for HART
communication (see Figure 2-7).
One switch is supplied for each input (top switch for “A” inputs and bottom switch for
“B” inputs). Setting the switch in the “ON” position (to the right) bypasses the 250 ohm
resistor. Terminals are provided for each analog input to connect a field communicator
for local configuration.
Installation
9
Reference Manual
Section 2: Installation
00809-0100-4697, Rev GA
November 2014
Figure 2-6. 848T Analog Input Wiring Diagram
A
B
C
A. Analog input connectors
B. Analog transmitters
C. Power supply
Figure 2-7. 848T Analog Connector
B
A
C
D
A. HART Channel A
B. 250 ohm resistor in the loop when switched to the left
C. HART Channel B
D. Space available for identification of inputs
10
Installation
Reference Manual
Section 2: Installation
00809-0100-4697, Rev GA
2.3.2
November 2014
Power supply
Connections
The transmitter requires between 9 and 32 VDC to operate and provide complete functionality.
The DC power supply should provide power with less than 2% ripple. A fieldbus segment
requires a power conditioner to isolate the power supply filter and decouple the segment from
other segments attached to the same power supply.
All power to the transmitter is supplied over the signal wiring. Signal wiring should be shielded,
twisted pair for best results in electrically noisy environments. Do not use unshielded signal
wiring in open trays with power wiring or near heavy electrical equipment.
Use ordinary copper wire of sufficient size to ensure the voltage across the transmitter power
terminals does not go below 9 VDC. The power terminals are polarity insensitive. To power the
transmitter:
1.
Connect the power leads to the terminals marked “Bus,” as shown in Figure 2-8.
2.
Tighten the terminal screws to ensure adequate contact. No additional power wiring is
necessary.
Figure 2-8. Transmitter Label
NOT USED
SECURITY
SIMULATE ENABLE
A
B
A. Ground (required with T1 option)
B. Connect power leads here
2.3.3
Surges/transients
The transmitter will withstand electrical transients encountered through static discharges or
induced switching transients. However, a transient protection option (option code T1) is
available to protect the 848T against high-energy transients. The device must be properly
grounded using the ground terminal (see Figure 2-8).
2.4
Grounding
The 848T transmitter provides input/output isolation up to 620 V rms.
Note
Neither conductor of the fieldbus segment can be grounded. Grounding out one of the signal
wires will shut down the entire fieldbus segment.
Installation
11
Reference Manual
Section 2: Installation
00809-0100-4697, Rev GA
November 2014
Shielded wire
Each process installation has different requirements for grounding. Use the grounding options
recommended by the facility for the specific sensor type or begin with grounding option 1
(most common).
Ungrounded thermocouple, mV, and RTD/ohm inputs
Option 1:
1.
Connect signal wiring shield to the sensor wiring shield(s).
2.
Ensure shields are tied together and electrically isolated from transmitter enclosure.
3.
Only ground shield at the power supply end.
4.
Ensure sensor shield(s) is electrically isolated from the surrounding grounded fixtures.
B
C
A
D
A. Sensor wires
B. 848T
C. Power supply
D. Shield ground point
Option 2:
1.
Connect sensor wiring shield(s) to the transmitter enclosure (only if enclosure is
grounded).
2.
Ensure sensor shield(s) is electrically isolated from surrounding fixtures that may be
grounded.
3.
Ground signal wiring shield at the power supply end.
B
C
A
D
A. Sensor wires
B. 848T
C. Power supply
D. Shield ground points
12
Installation
Reference Manual
Section 2: Installation
00809-0100-4697, Rev GA
November 2014
Grounded thermocouple inputs
1.
Ground sensor wiring shield(s) at the sensor.
2.
Ensure that the sensor wiring and signal wiring shields are electrically isolated from the
transmitter enclosure.
3.
Do not connect the signal wiring shield to the sensor wiring shield(s).
4.
Ground signal wiring shield at the power supply end.
B
C
A
D
A. Sensor wires
B. 848T
C. Power supply
D. Shield ground points
Analog device inputs
1.
Ground analog signal wire at the power supply of the analog devices.
2.
Ensure the analog signal wire and the fieldbus signal wire shields are electrically isolated
from the transmitter enclosure.
3.
Do not connect the analog signal wire shield to the fieldbus signal wire shield.
4.
Ground fieldbus signal wire shield at the power supply end.
A
C
D
G
B
E
F
G
A. 4-20 mA loop
B. FOUNDATION fieldbus bus
C. Analog device power supply
D. Analog device
E. 848T
F. Power supply
G. Shield ground points
Transmitter enclosure (optional)
Ground the transmitter in accordance with local electrical requirements.
Installation
13
Reference Manual
Section 2: Installation
00809-0100-4697, Rev GA
November 2014
2.5
Switches
Figure 2-9. Switch Location on the Rosemount 848T
NOT USED
SECURITY
SIMULATE ENABLE
Security
After configuring the transmitter, the data can be protected from unwarranted changes. Each
848T is equipped with a security switch that can be positioned “ON” to prevent the accidental
or deliberate change of configuration data. This switch is located on the front side of the
electronics module and is labeled SECURITY.
See Figure 2-9 for switch location on the transmitter label.
Simulate enable
The switch labeled SIMULATE ENABLE is used in conjunction with the Analog Input (AI) and
Multiple Analog Input (MAI) function blocks. This switch is used to simulate temperature
measurement.
Not used
The switch is not functional.
2.6
Tagging
Commissioning tag
The 848T has been supplied with a removable commissioning tag that contains both the Device
ID (the unique code that identifies a particular device in the absence of a device tag) and a space
to record the device tag [the operational identification for the device as defined by the Piping
and Instrumentation Diagram (P&ID)].
When commissioning more than one device on a fieldbus segment, it can be difficult to identify
which device is at a particular location. The removable tag, provided with the transmitter, can
aid in this process by linking the Device ID to its physical location. The installer should note the
physical location of the transmitter on both the upper and lower location of the commissioning
tag. The bottom portion should be torn off for each device on the segment and used for
commissioning the segment in the control system.
14
Installation
Reference Manual
Section 2: Installation
00809-0100-4697, Rev GA
November 2014
Figure 2-10. Commissioning Tag
A
B
A. Device ID
B. Device Tag to denote physical location
Transmitter tag
Hardware

Tagged in accordance with customer requirements

Permanently attached to the transmitter
Software

The transmitter can store up to 32 characters.

If no characters are specified, the first 30 characters of the hardware tag will be used.
Sensor tag
Hardware

A plastic tag is provided to record identification of eight sensors.

This information can be printed at the factory upon request.

In the field, the tag can be removed, printed onto, and reattached to the transmitter.
Software
Installation

If sensor tagging is requested, the Transducer Block SERIAL_NUMBER parameters will
be set at the factory.

The SERIAL_NUMBER parameters can be updated in the field.
15
Reference Manual
Section 2: Installation
00809-0100-4697, Rev GA
November 2014
2.7
Installation
2.7.1
Using cable glands
Use the following steps to install the 848T with cable glands:
1.
Remove the junction box cover by unscrewing the four cover screws.
2.
Run the sensor and power/signal wires through the appropriate cable glands using the
pre-installed cable glands (see Figure 2-11).
3.
Install the sensor wires into the correct screw terminals (follow the label on the
electronics module).
4.
Install the power/signal wires onto the correct screw terminals. Power is polarity
insensitive, allowing the user to connect positive (+) or negative (–) to either fieldbus
wiring terminal labeled “Bus.”
5.
Replace the enclosure cover and securely tighten all cover screws.
Figure 2-11. Installing the 848T with Cable Glands
A
C
F
E
D
G
B
H
I
J
K
A. Enclosure cover screw (4)
B. Cable gland
C. Sensor 1
D. Sensor 3
E. Sensor 5
F. Sensor 7
G. Power/signal
H. Sensor 2
I. Sensor 4
J. Sensor 6
K. Sensor 8
16
Installation
Reference Manual
Section 2: Installation
00809-0100-4697, Rev GA
2.7.2
November 2014
Using conduit entries
Use the following steps to install the 848T with conduit entries.
1.
Remove the junction box cover by unscrewing the four cover screws.
2.
Remove the five conduit plugs and install five conduit fittings (supplied by the installer).
3.
Run pairs of sensor wires through each conduit fitting.
4.
Install the sensor wires into the correct screw terminals (follow the label on the
electronics module).
5.
Install the power/signal wires into the correct screw terminals. Power is polarity
insensitive, allowing the user to connect positive (+) or negative (–) to either fieldbus
wiring terminal labeled “Bus.”
6.
Replace the junction box cover and securely tighten all cover screws.
Figure 2-12. Installing the 848T with Conduit Entries
A
B
C
D
E
F
A. Enclosure cover screw
B. Sensors 1 and 2 conduit
C. Sensors 3 and 4 conduit
D. Sensors 5 and 6 conduit
E. Sensors 7 and 8 conduit
F. Power/signal
Installation
17
Section 2: Installation
November 2014
18
Reference Manual
00809-0100-4697, Rev GA
Installation
Section 3: Configuration
Reference Manual
November 2014
00809-0100-4697, Rev GA
Section 3
Configuration
Safety messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Common configurations for high density applications . . . . . . . . . . . . . . . . . . . . . . . . . . .
Block configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.1
page 19
page 20
page 22
page 25
Safety messages
Instructions and procedures in this section may require special precautions to ensure the safety
of the personnel performing the operations. Information that potentially raises safety issues is
indicated by a warning symbol ( ). Refer to the following safety messages before performing
an operation preceded by this symbol.
3.1.1
Warnings
Failure to follow these installation guidelines could result in death or serious injury.

Make sure only qualified personnel perform the installation.
Process leaks could result in death or serious injury.
Do not remove the thermowell while in operation. Removing while in operation may
cause process fluid leaks.

Install and tighten thermowells and sensors before applying pressure, or process
leakage may result.
Electrical shock could cause death or serious injury.



Configuration
If the sensor is installed in a high voltage environment and a fault condition or
installation error occurs, high voltage may be present on transmitter leads and
terminals.
Use extreme caution when making contact with the leads and terminals.
19
Reference Manual
Section 3: Configuration
00809-0100-4697, Rev GA
November 2014
3.2
Configuration
3.2.1
Standard
Each FOUNDATION™ fieldbus configuration tool or host system has a different way of displaying
and performing configurations. Some will use Device Descriptions (DDs) and DD Methods to
make configuration and displaying of data consistent across host platforms.
Unless otherwise specified, the 848T will be shipped with the following configuration (default):
Table 3-1. Standard Configuration Settings
Sensor Type(1)
Type J Thermocouple
(1)
Damping
5 seconds
Measurement Units(1)
°C
Output
(1)
Linear with Temperature
(1)
Line Voltage Filter
60 Hz
Temperature Specific Blocks
FOUNDATION fieldbus Function Blocks
• Transducer Block (1)
• Analog Input (8)
• Multiple Analog Input (2)
• Input Selector (4)
(1) For all eight sensors
Refer to that systems documentation to perform configuration changes using a FOUNDATION
fieldbus host or configuration tool.
Note
To make configuration changes, ensure that the block is Out of Service (OOS) by setting the
MODE_BLK.TARGET to OOS, or set the SENSOR_MODE to Configuration.
3.2.2
Transmitter configuration
The transmitter is available with the standard configuration setting. The configuration settings
and block configuration may be changed in the field with the Emerson Process Management
Systems DeltaV™, with AMS®inside, or other FOUNDATION fieldbus host or configuration tool.
3.2.3
Custom configuration
Custom configurations are to be specified when ordering.
3.2.4
Methods
For FOUNDATION fieldbus hosts or configuration tools that support device description (DD)
methods, there are two configuration methods available in the transducer block. These
methods are included with the DD software.

Sensor Configuration

Sensor Input Trim (user input trim)
See the host system documentation for information on running DD methods from the host
system. If the FOUNDATION fieldbus host or configuration tool does not support DD methods,
refer to “Block configuration” on page 25 for information on how to modify sensor
configuration parameters.
20
Configuration
Reference Manual
Section 3: Configuration
00809-0100-4697, Rev GA
3.2.5
November 2014
Alarms
Use the following steps to configure the alarms, which are located in the Resource Function
Block:
1.
Set the resource block to OOS.
2.
Set WRITE_PRI to the appropriate alarm level (WRITE_PRI has a selectable range of
priorities from 0 to 15, see “Alarm Priority Levels” on page 31. Set the other block alarm
parameters at this time.
3.
Set CONFIRM_TIME to the time, in 1/32 of a millisecond, that the device will wait for
confirmation of receiving a report before trying again (the device does not retry if
CONFIRM_TIME is 0).
4.
Set LIM_NOTIFY to a value between zero and MAX_NOTIFY. LIM_NOTIFY is the
maximum number of alert reports allowed before the operator needs to acknowledge
an alarm condition.
5.
Enable the reports bit in FEATURES_SEL. (When Multi-bit alerts is enabled, every active
alarm is visible for any of the eight sensors, generated by a PlantWeb® and Field
Diagnostics alert. This is different than only viewing the highest priority alarm.)
6.
Set the resource block to AUTO.
For modifying alarms on individual function blocks (AI or ISEL blocks), refer to Appendix D:
Function Blocks.
3.2.6
Damping
Use the following steps to configure the damping, which is located in the transducer function
block:
3.2.7
1.
Set Sensor Mode to Out of Service.
2.
Change DAMPING to the desired filter rate (0.0 to 32.0 seconds).
3.
Set Sensor Mode to In Service.
Configure the differential sensors
Use the following steps to configure the differential sensors:
Configuration
1.
Set Dual Sensor Mode to Out of Service.
2.
Set Input A and Input B to the sensor values that are to be used in the differential
equation diff = A–B. (Note: Unit types must be the same.)
3.
Set the DUAL_SENSOR_CALC to either Not Used, Absolute, or INPUT A minus INPUT B.
4.
Set Dual Sensor Mode to In Service.
21
Reference Manual
Section 3: Configuration
00809-0100-4697, Rev GA
November 2014
3.2.8
Configure measurement validation
Use the following steps to configure measurement validation:
1.
Set mode to Disabled for specific sensor.
2.
Select sample rate. 1-10 sec/sample is available. 1 second/sample is preferred for
sensor degradation. The higher the number of seconds between samples, the more
emphasis put on process variation.
3.
Select Deviation Limit from 0 to 10 units. If deviation limit is exceeded, a status event
will be triggered.
4.
Select Increasing Limit. Sets the limit for increasing rate of change. If limit is exceeded, a
status event will be triggered.
5.
Select Decreasing Limit. Sets the limit for decreasing rate of change. If limit is exceeded,
a status event will be triggered.
Note
The decreasing limit selected is required to be a negative value.
3.3
6.
Set the Deadband from 0 to 90%. This threshold is used to clear the PV status.
7.
Set Status Priority. This determines what happens when the specific limit has been
exceeded. No Alert - Ignores limit settings. Advisory - Sets Advisory Plant Web Alert,
but does not do anything with PV status. Warning - Sets a Maintenance Plant Web Alert
and sets PV status to uncertain. Failure - Sets A Failure Plant Web Alert and sets PV
status to Bad.
8.
Set mode to Enabled for specific sensor.
Common configurations for high density
applications
For the application to work properly, configure the links between the function blocks and
schedule the order of their execution. The Graphical User Interface (GUI) provided by the
FOUNDATION fieldbus host or configuration tool will allow easy configuration.
The measurement strategies shown in this section represent some of the common types of
configurations available in the 848T. Although the appearance of the GUI screens will vary from
host to host, the configuration logic is the same.
Note
Ensure the host system or configuration tool is properly configured before downloading the
transmitter configuration. If configured improperly, the FOUNDATION fieldbus host or
configuration tool could overwrite the default transmitter configuration.
22
Configuration
Reference Manual
Section 3: Configuration
00809-0100-4697, Rev GA
November 2014
Typical profiling application
Example: Distillation column temperature profile where all channels have the same sensor units
(°C, °F, etc.).
Out_1
Out_2
Out_3
Out_4
Out_5
MAI
Function
Block
Out_6
Out_7
Out_8
1.
Place the Multiple Analog Input (MAI) function block in OOS mode (set
MODE_BLK.TARGET to OOS).
2.
Set CHANNEL= “channels 1 to 8.” Although the CHANNEL_X parameters remain
writable, CHANNEL_X can only be set = X when CHANNEL=1.
3.
Set L_TYPE to direct or indirect.
4.
Set XD_SCALE (transducer measurement scaling) to the appropriate upper and lower
range values, the appropriate sensor units, and display decimal point.
5.
Set OUT_SCALE (MAI output scaling) to the appropriate upper and lower range values,
the appropriate sensor units, and display decimal point.
6.
Place the MAI Function Block in auto mode.
7.
Verify that the function blocks are scheduled.
Monitoring application with a single selection
Example: Average exhaust temperature of gas and turbine where there is a single alarm level for
all inputs.
IN_1
Out
Out_2
IN_2
Out_D
Out_3
IN_3
Out_1
MAI
Function
Block
Configuration
Out_4
IN_4
Out_5
IN_5
Out_6
IN_6
Out_7
IN_7
Out_8
IN_8
ISEL
Function
Block
1.
Link the MAI outputs to the ISEL inputs.
2.
Place the Multiple Analog Input (MAI) function block in OOS mode (set
MODE_BLK.TARGET to OOS).
3.
Set CHANNEL= “channels 1 to 8.” Although the CHANNEL_X parameters remain
writable, CHANNEL_X can only be set = X when CHANNEL=1.
4.
Set L_TYPE to direct or indirect.
5.
Set XD_SCALE (transducer measurement scaling) to the appropriate upper and lower
range values, the appropriate sensor units, and display decimal point.
23
Reference Manual
Section 3: Configuration
00809-0100-4697, Rev GA
November 2014
6.
Set OUT_SCALE (MAI output scaling) to the appropriate upper and lower range values,
the appropriate sensor units, and display decimal point.
7.
Place the MAI function block in auto mode.
8.
Place the Input Selector (ISEL) function block in OOS mode by setting
MODE_BLK.TARGET to OOS.
9.
Set OUT_RANGE to match the OUT_SCALE in the MAI block.
10.
Set SELECT_TYPE to the desired function (Maximum Value, Minimum Value, First Good
Value, Midpoint Value, or Average Value).
11.
Set the alarm limits and parameters if necessary.
12.
Place the ISEL function block in auto mode.
13.
Verify that the function blocks are scheduled.
Measuring temperature points individually
Example: Miscellaneous monitoring of temperature in a “close proximity” where each channel
can have different sensor inputs with different units and there are independent alarm levels for
each input.
Out
AI
Function
Block 1
Out_D
Out
AI
Function
Block 8
24
Out_D
1.
Place the first Analog Input (AI) function block in OOS mode (set MODE_BLK.TARGET to
OOS).
2.
Set CHANNEL to the appropriate channel value. Refer to “Alarm Priority Levels” on
page 31 for a listing of channel definitions.
3.
Set L_TYPE to direct.
4.
Set XD_SCALE (transducer measurement scaling) to the appropriate upper and lower
range values, the appropriate sensor units, and display decimal point.
5.
Set OUT_SCALE (AI output scaling) to the appropriate upper and lower range values,
the appropriate sensor units, and display decimal point.
6.
Set the alarm limits and parameters if necessary.
7.
Place the AI function block in auto mode.
8.
Repeat steps 1 through 7 for each AI function block.
9.
Verify that the function blocks are scheduled.
Configuration
Reference Manual
Section 3: Configuration
00809-0100-4697, Rev GA
3.3.1
November 2014
Interfacing analog transmitters to FOUNDATION fieldbus
Transducer block configuration
Use the sensor configuration method to set the sensor type to mV – 2-wire for the applicable
transducer block or follow these steps.
1.
Set the MODE_BLK.TARGET to OOS mode, or set the SENSOR_MODE to configuration.
2.
Set the SENSOR to mV.
3.
Set the MODE_BLK.TARGET to AUTO, or set the SENSOR_MODE to operation.
Multiple analog input or analog input block configuration
Follow these steps to configure the applicable block.
1.
Set the MODE_BLK.TARGET to OOS mode, or set the SENSOR_MODE to configuration.
2.
Set CHANNEL to the transducer block configured for the analog input.
3.
Set XD_SCALE.EU_0 to 20
Set XD_SCALE.EU_100 to 100
Set XD_SCALE.ENGUNITS to mV
4.
SET OUT_SCALE to match the desired scale and units for the connected analog
transmitter.
Flow Example: 0 – 200 gpm
OUT_SCALE.EU_0 = 0
OUT_SCALE.EU_100 = 200
OUT_SCALE.ENGUNITS = gpm
5.
Set L_TYPE to INDIRECT.
6.
Set the MODE_BLK.TARGET to AUTO, or set the SENSOR_MODE to operation.
3.4
Block configuration
3.4.1
Resource block
The resource block defines the physical resources of the device including type of measurement,
memory, etc. The resource block also defines functionality, such as shed times, that is common
across multiple blocks. The block has no linkable inputs or outputs and it performs
memory-level diagnostics.
Table 3-2. Resource Block Parameters
Number
Parameter
01
ST_REV
02
TAG_DESC
The user description of the intended application of the block.
03
STRATEGY
The strategy field can be used to identify grouping of blocks.
04
ALERT_KEY
The identification number of the plant unit.
05
MODE_BLK
The actual, target, permitted, and normal modes of the block. For further
description, see the Mode parameter formal model in FF-890.
06
BLOCK_ERR
This parameter reflects the error status associated with the hardware or software
components associated with a block. Multiple errors may be shown. For a list of
enumeration values, see FF-890, Block_Err formal model.
Configuration
Description
The revision level of the static data associated with the function block.
25
Reference Manual
Section 3: Configuration
00809-0100-4697, Rev GA
November 2014
Table 3-2. Resource Block Parameters
Number
Parameter
07
RS_STATE
State of the function block application state machine. For a list of enumeration
values, see FF-890.
08
TEST_RW
Read/write test parameter - used only for conformance testing.
09
DD_RESOURCE
String identifying the tag of the resource which contains the Device Description for
the resource.
10
MANUFAC_ID
Manufacturer identification number - used by an interface device to locate the DD
file for the resource.
11
DEV_TYPE
Manufacturer's model number associated with the resource - used by interface
devices to locate the DD file for the resource.
12
DEV_REV
Manufacturer revision number associated with the resource - used by an interface
device to locate the DD file for the resource.
13
DD_REV
Revision of the DD associated with the resource - used by the interface device to
locate the DD file for the resource.
14
GRANT_DENY
Options for controlling access of host computer and local control panels to
operating, tuning and alarm parameters of the block.
15
HARD_TYPES
The types of hardware available as channel numbers. The supported hardware
type is: SCALAR_INPUT
16
RESTART
Allows a manual restart to be initiated.
1) Run: This is passive state of this parameter.
2) Restart resource: To clear up the problems like garbage collection.
3) Restart with defaults: reset all configurable function block application objects to
their initial value i.e. their value before any configuration was done by the user.
This will also remove appended serial numbers of function block tags
4) Restart processor: provides a way to hit the reset button on the processor
associated with the resource.
5) Restart to append serial number: Appends serial number to function block tags.
11) Restart default blocks: defaults manufacturer pre-instantiated blocks.
17
FEATURES
Used to show supported resource block options. The supported features are:
SOFT_WRITE_LOCK_SUPPORT, HARD_WRITE_LOCK_SUPPORT, REPORTS,
UNICODE, MULTI_BIT_ALARM_SUPPORT and FB_ACTION_RESTART_RELINK
18
FEATURE_SEL
Used to select resource block options.
19
CYCLE_TYPE
Identifies the block execution methods available for this resource. The supported
cycle types are: SCHEDULED, and COMPLETION_OF_BLOCK_EXECUTION
20
CYCLE_SEL
26
Description
Used to select the block execution method for this resource.
21
MIN_CYCLE_T
Time duration of the shortest cycle interval of which the resource is capable.
22
MEMORY_SIZE
Available configuration memory in the empty resource. To be checked before
attempting a download.
23
NV_CYCLE_T
Minimum time interval specified by the manufacturer for writing copies of NV
parameters to non-volatile memory. Zero means it will never be automatically
copied. At the end of NV_CYCLE_T, only those parameters which have changed
need to be updated in NVRAM.
24
FREE_SPACE
Percent of memory available for further configuration. Zero in preconfigured
resource.
25
FREE_TIME
Percent of the block processing time that is free to process additional blocks.
26
SHED_RCAS
Time duration at which to give up on computer writes to function block RCas
locations. Shed from RCas will never happen when SHED_RCAS = 0.
27
SHED_ROUT
Time duration at which to give up on computer writes to function block ROut
locations. Shed from ROut will never happen when SHED_ROUT = 0.
28
FAULT_STATE
Condition set by loss of communication to an output block, fault promoted to an
output block or physical contact. When FAIL_SAFE condition is set, then output
function blocks will perform their FAIL_SAFE actions.
29
SET_FSTATE
Allows the FAIL_SAFE condition to be manually initiated by selecting Set.
30
CLR_FSTATE
Writing a Clear to this parameter will clear the device FAIL_SAFE if the field
condition has cleared.
31
MAX_NOTIFY
Maximum number of unconfirmed notify messages possible.
Configuration
Reference Manual
Section 3: Configuration
00809-0100-4697, Rev GA
November 2014
Table 3-2. Resource Block Parameters
Number
Parameter
Description
32
LIM_NOTIFY
Maximum number of unconfirmed alert notify messages allowed.
33
CONFIRM_TIME
The time the resource will wait for confirmation of receipt of a report before trying
again. Retry will not happen when CONFIRM_TIME=0.
34
WRITE_LOCK
If set, all writes to static and non-volatile parameters are prohibited, except to clear
WRITE_LOCK. Block inputs will continue to be updated.
35
UPDATE_EVT
This alert is generated by any change to the static data.
36
BLOCK_ALM
The BLOCK_ALM is used for all configuration, hardware, connection failure or
system problems in the block. The cause of the alert is entered in the subcode field.
The first alert to become active will set the Active status in the Status attribute. As
soon as the Unreported status is cleared by the alert reporting task, another block
alert may be reported without clearing the Active status, if the subcode has
changed.
37
ALARM_SUM
The current alert status, unacknowledged states, unreported states, and disabled
states of the alarms associated with the function block.
38
ACK_OPTION
Selection of whether alarms associated with the block will be automatically
acknowledged.
39
WRITE_PRI
Priority of the alarm generated by clearing the write lock.
40
WRITE_ALM
This alert is generated if the write lock parameter is cleared.
41
ITK_VER
Major revision number of the interoperability test case used in certifying this
device as interoperable. The format and range are controlled by the fieldbus
FOUNDATION.
42
FD_VER
This parameter's value equals the value of the major version of the Field
Diagnostics specification that this device was designed to.
43
FD_FAIL_ACTIVE
Reflects the error conditions that are being detected as active as selected for this
category. It is a bit string, so that multiple conditions may be shown.
44
FD_OFFSPEC_ACTIVE
Reflects the error conditions that are being detected as active as selected for this
category. It is a bit string, so that multiple conditions may be shown.
45
FD_MAINT_ACTIVE
Reflects the error conditions that are being detected as active as selected for this
category. It is a bit string, so that multiple conditions may be shown.
46
FD_CHECK_ACTIVE
Reflects the error conditions that are being detected as active as selected for this
category. It is a bit string, so that multiple conditions may be shown.
47
FD_FAIL_MAP
Maps conditions to be detected as active for this alarm category. Thus the same
condition may be active in all, some, or none of the 4 alarm categories.
48
FD_OFFSPEC_MAP
Maps conditions to be detected as active for this alarm category. Thus the same
condition may be active in all, some, or none of the 4 alarm categories.
49
FD_MAINT_MAP
Maps conditions to be detected as active for this alarm category. Thus the same
condition may be active in all, some, or none of the 4 alarm categories.
50
FD_CHECK_MAP
Maps conditions to be detected as active for this alarm category. Thus the same
condition may be active in all, some, or none of the 4 alarm categories.
51
FD_FAIL_MASK
Allows the user to suppress any single or multiple conditions that are active, in this
category, from being broadcast to the host through the alarm parameter. A bit
equal to ‘1’ will mask i.e. inhibit the broadcast of a condition, and a bit equal to ‘0’
will unmask i.e. allow broadcast of a condition.
52
FD_OFFSPEC_MASK
Allows the user to suppress any single or multiple conditions that are active, in this
category, from being broadcast to the host through the alarm parameter. A bit
equal to ‘1’ will mask i.e. inhibit the broadcast of a condition, and a bit equal to ‘0’
will unmask i.e. allow broadcast of a condition.
53
FD_MAINT_MASK
Allows the user to suppress any single or multiple conditions that are active, in this
category, from being broadcast to the host through the alarm parameter. A bit
equal to ‘1’ will mask i.e. inhibit the broadcast of a condition, and a bit equal to ‘0’
will unmask i.e. allow broadcast of a condition.
54
FD_CHECK_MASK
Allows the user to suppress any single or multiple conditions that are active, in this
category, from being broadcast to the host through the alarm parameter. A bit
equal to ‘1’ will mask i.e. inhibit the broadcast of a condition, and a bit equal to ‘0’
will unmask i.e. allow broadcast of a condition.
Configuration
27
Reference Manual
Section 3: Configuration
00809-0100-4697, Rev GA
November 2014
Table 3-2. Resource Block Parameters
Number
Parameter
55
FD_FAIL_ALM
Used primarily to broadcast a change in the associated active conditions, which are
not masked, for this alarm category to a Host System.
56
FD_OFFSPEC_ALM
Used primarily to broadcast a change in the associated active conditions, which are
not masked, for this alarm category to a Host System.
57
FD_MAINT_ALM
Used primarily to broadcast a change in the associated active conditions, which are
not masked, for this alarm category to a Host System.
58
FD_CHECK_ALM
Used primarily to broadcast a change in the associated active conditions, which are
not masked, for this alarm category to a Host System.
28
Description
59
FD_FAIL_PRI
Allows the user to specify the priority of this alarm category.
60
FD_OFFSPEC_PRI
Allows the user to specify the priority of this alarm category.
61
FD_MAINT_PRI
Allows the user to specify the priority of this alarm category.
62
FD_CHECK_PRI
Allows the user to specify the priority of this alarm category.
63
FD_SIMULATE
Allows the conditions to be manually supplied when simulation is enabled. When
simulation is disabled both the diagnostic simulate value and the diagnostic value
track the actual conditions. The simulate jumper is required for simulation to be
enabled and while simulation is enabled the recommended action will show that
simulation is active.
64
FD_RECOMMEN_ACT
A device enumerated summarization of the most severe condition or conditions
detected. The DD help should describe by enumerated action, what should be
done to alleviate the condition or conditions. 0 is defined as Not Initialized, 1 is
defined as No Action Required, all others defined by manufacturer.
65
FD_EXTENDED_ACTIVE_1
An optional parameter or parameters to allow the user finer detail on conditions
causing an active condition in the FD_*_ACTIVE parameters.
66
FD_EXTENDED_MAP_1
An optional parameter or parameters to allow the user finer control on enabling
conditions contributing to the conditions in FD_*_ACTIVE parameters.
67
COMPATIBILITY_REV
68
HARDWARE_REVISION
Manufacturer hardware revision
69
SOFTWARE_REV
Manufacturer hardware revision
Optionally used when replacing field devices. The correct usage of this parameter
presumes the COMPATIBILITY_REV value of the replacing device should be equal or
lower than the DEV_REV value of the replaced device.
70
PD_TAG
71
DEV_STRING
PD tag description of device
Used to load new licensing into the device. The value can be written but will always
read back with a value of 0.
72
DEV_OPTIONS
Indicates which miscellaneous and diagnostic device licensing options are enabled.
It also indicates Transducer options.
73
OUTPUT_BOARD_SN
Output board serial number
74
FINAL_ASSY_NUM
75
DOWNLOAD_MODE
Same final assembly number placed on the neck label
76
HEALTH_INDEX
Parameter shall be set based on the active FD alarms or PWA alarms.
HEALTH_INDEX will show 100 if target mode of block is OOS or there are no active
alarms in device. The table below represents HEALTH_INDEX value when FD or
PWA alarms are active in a device.
77
FAILED_PRI
Designates the alarming priority of the FAILED_ALM and also used as switch b/w FD
and legacy PWA. If value is greater than or equal to 1 then PWA alerts will be active
in device else device will have FD alerts.
78
RECOMMENDED_ACTION
79
FAILED_ALM
Alarm indicating a failure within a device which makes the device non-operational
80
MAINT _ALM
Alarm indicating the device needs maintenance soon. If the condition is ignored,
the device will eventually fail.
81
ADVISE _ALM
Alarm indicating advisory alarms. These conditions do not have a direct impact on
the process or device integrity.
Gives access to the boot block code for over the wire downloads
Enumerated list of recommended actions displayed with a device alert
Configuration
Reference Manual
Section 3: Configuration
00809-0100-4697, Rev GA
November 2014
Table 3-2. Resource Block Parameters
Number
Parameter
Description
82
FAILED_ENABLE
83
FAILED_MASK
Mask of Failure Alarm. Corresponds bit for bit to the FAILED_ACTIVE. A bit on
means that the failure is masked out from alarming. This parameter is the Read
Only copy of FD_FAIL_MASK.
84
FAILED_ACTIVE
Enumerated list of failure conditions within a device. All open bits are free to be
used as appropriate for each specific device.This parameter is the Read Only copy
of FD_FAIL_ACTIVE.
85
MAINT_PRI
86
MAINT_ENABLE
Enabled MAINT_ALM alarm conditions. Corresponds bit for bit to the
MAINT_ACTIVE. A bit on means that the corresponding alarm condition is enabled
and will be detected. A bit off means the corresponding alarm condition is disabled
and will not be detected.This parameter is the Read Only copy of
FD_OFFSPEC_MAP.
87
MAINT_MASK
Mask of Maintenance Alarm. Corresponds bit for bit to the MAINT_ACTIVE. A bit on
means that the failure is masked out from alarming.This parameter is the Read
Only copy of FD_OFFSPEC_MASK.
88
MAINT_ACTIVE
Enumerated list of maintenance conditions within a device. All open bits are free to
be used as appropriate for each specific device. This parameter is the Read Only
copy of FD_OFFSPEC_ACTIVE.
Enabled FAILED_ALM alarm conditions. Corresponds bit for bit to the
FAILED_ACTIVE. A bit on means that the corresponding alarm condition is enabled
and will be detected. A bit off means the corresponding alarm condition is disabled
and will not be detected. This parameter is the Read Only copy of FD_FAIL_MAP.
Designates the alarming priority of the MAINT_ALM
89
ADVISE_PRI
90
ADVISE_ENABLE
91
ADVISE_MASK
Mask of Advisory Alarm. Corresponds bit for bit to the ADVISE_ACTIVE. A bit on
means that the failure is masked out from alarming.This parameter is the Read
Only copy of FD_MAINT_MASK & FD_CHECK_MASK.
92
ADVISE_ACTIVE
Enumerated list of advisory conditions within a device. All open bits are free to be
used as appropriate for each specific device. This parameter is the Read Only copy
of FD_MAINT_ACTIVE & FD_CHECK_ACTIVE.
Configuration
Designates the alarming priority of the ADVISE_ALM
Enabled ADVISE_ALM alarm conditions. Corresponds bit for bit to the
ADVISE_ACTIVE. A bit on means that the corresponding alarm condition is enabled
and will be detected. A bit off means the corresponding alarm condition is disabled
and will not be detected.This parameter is the Read Only copy of FD_MAINT_MAP
& FD_CHECK_MAP.
29
Reference Manual
Section 3: Configuration
00809-0100-4697, Rev GA
November 2014
Block errors
The table below lists conditions reported in the BLOCK_ERR parameter.
Table 3-3. BLOCK_ERR Conditions
Number
Name and description
0
Other
1
Block Configuration Error: A feature in CYCLE_SEL is set that is not supported by
CYCLE_TYPE.
3
Simulate Active: This indicates that the simulation jumper is in place. This is not an
indication that the I/O blocks are using simulated data.
6
Device needs maintenance soon
7
Input failure/process variable has bad status
9
Memory Failure: A memory failure has occurred in FLASH, RAM, or EEPROM memory.
10
Lost Static Data: Static data that is stored in non-volatile memory
has been lost.
11
Lost NV Data: Non-volatile data that is stored in non-volatile memory
has been lost.
13
Device Needs Maintenance Now
14
Power Up: The device was just powered-up.
15
OOS: The actual mode is out of service.
Modes
The resource block supports two modes of operation as defined by the MODE_BLK parameter:
Automatic (Auto)
The block is processing its normal background memory checks.
Out of service (OOS)
The block is not processing its tasks. When the resource block is in OOS, all blocks within
the resource (device) are forced into OOS. The BLOCK_ERR parameter shows Out of
Service. In this mode, changes can be made to all configurable parameters. The target
mode of a block may be restricted to one or more of the supported modes.
Alarm detection
A block alarm will be generated whenever the BLOCK_ERR has an error bit set. The types of
block error for the resource block are defined above. A write alarm is generated whenever the
WRITE_LOCK parameter is cleared. The priority of the write alarm is set in the following
parameter:

30
WRITE_PRI
Configuration
Reference Manual
Section 3: Configuration
00809-0100-4697, Rev GA
November 2014
Table 3-4. Alarm Priority Levels
Number
Description
0
The priority of an alarm condition changes to 0 after the condition that caused the
alarm is corrected.
1
An alarm condition with a priority of 1 is recognized by the system, but is not reported
to the operator.
2
An alarm condition with a priority of 2 is reported to the operator, but does not require
operator attention (such as diagnostics and system alerts).
3-7
Alarm conditions of priority 3 to 7 are advisory alarms of increasing priority.
8-15
Alarm conditions of priority 8 to 15 are critical alarms of increasing priority.
Status handling
There are no status parameters associated with the resource block.
3.4.2
Field Diagnostics & PlantWeb Alerts
The 848T ITK6 has two mechanisms for alarms, one is Field Diagnostics (FD) and other is
PlantWeb Alerts (PWA) for backward compatibility only.
The flexible capability has been added to allow user to select any alarm to be in the PWA
FAILED/FD FAILED group or PWA MAINTENANCE/FD OFFSPEC group or PWA ADVISE/FD
MAINTENANCE group or PWA ADVISE/FD CHECK group.
In PlantWeb Alerts, the alarms can be represented in three groups i.e. FAILED, MAINT & ADVISE.
In Field Diagnostic, the alarms can be represented in four groups i.e. FAILED, OFFSPEC, MAINT &
CHECK.
Parameter FAILED_PRI is used as a switch for using Field Diagnostic and PlantWeb Alerts.
How to use Field Diagnostic Alarm
If FAILED_PRI is equal to 0, Field Diagnostic alarms are supported and PlantWeb alarms are not.
Field Diagnostic functionality includes four different Field Diagnostic Alarms such as
FD_FAIL_ALM, FD_OFFSPEC_ALM, FD_MAINT_ALM and FD_CHECK_ALM. For these alarms,
there are corresponding alarm priority parameter, masking parameter alarm active and alarm
mapping parameter such as FD_*_PRI, FD_*_MASK & FD_*_ACTIVE & FD_*_MAP.
How to use PlantWeb Alarm
If FAILED_PRI is greater than 0, PlantWeb alarms are supported and Field Diagnostic are not.
PlantWeb functionality includes three different PlantWeb Alarms FAILED_ALM, MAINT_ ALM
and ADVISE_ ALM. For PlantWeb Alerts, there are corresponding alarm masking parameter,
alarm active parameter and alarm mapping parameter such as *_MASK, *_ACTIVE & *_ENABLE.
These parameters have Read only access and are duplicated from corresponding FD parameters.
So for example, in case of PWA alarms, if user wishes to change the mapping of any PlantWeb
Alerts then the new value is written to the corresponding FD_*_MAP parameter. *_ENABLE shall
reflect whatever is being written to FD_*_MAP parameter. The same applies for *_MASK
parameters.
Note
Here * implies all 4 categories of FD alerts for e.g. FD_*_ACTIVE resembles FD_FAIL_ACTIVE,
FD_OFFSPEC_ACTIVE, FD_MAINT_ACTIVE & FD_CHECK ACTIVE.
The similar notation is also applicable for PWA alarms for e.g. FD_*_ACTIVE resembles
FAIL_ACTIVE, MAINT_ACTIVE & ADVISE ACTIVE.
Configuration
31
Reference Manual
Section 3: Configuration
00809-0100-4697, Rev GA
November 2014
3.4.3
PlantWeb alerts
The alerts and recommended actions should be used in conjunction with “Operation and
Maintenance” on page 53.
The Resource Block will act as a coordinator for PlantWeb alerts. There will be three alarm
parameters (FAILED_ALARM, MAINT_ALARM, and ADVISE_ALARM) which will contain
information regarding some of the device errors which are detected by the transmitter
software. There will be a RECOMMENDED_ACTION parameter which will be used to display the
recommended action text for the highest priority alarm and a HEALTH_INDEX parameters (0 100) indicating the overall health of the transmitter. FAILED_ALARM will have the highest
priority followed by MAINT_ALARM and ADVISE_ALARM will be the lowest priority.
FAILED_ALARMS
A failure alarm indicates a failure within a device that will make the device or some part of the
device non-operational. This implies that the device is in need of repair and must be fixed
immediately. There are five parameters associated with FAILED_ALARMS specifically, they are
described below.
FAILED_ENABLED
This parameter contains a list of failures in the device which makes the device
non-operational that will cause an alert to be sent. Below is a list of the failures with the
highest priority first.
Table 3-5. Failure Alarms
Alarm
Priority
ASIC Failure
1
Electronics Failure
2
Hardware/Software Incompatible
3
Memory Failure
4
Body Temperature Failure
5
Sensor 1 Failure
6
Sensor 2 Failure
7
Sensor 3 Failure
8
Sensor 4 Failure
9
Sensor 5 Failure
10
Sensor 6 Failure
11
Sensor 7 Failure
12
Sensor 8 Failure
13
FAILED_MASK
This parameter will mask any of the failed conditions listed in FAILED_ENABLED. A bit on
means that the condition is masked out from alarming and will not be reported.
32
Configuration
Reference Manual
Section 3: Configuration
00809-0100-4697, Rev GA
November 2014
FAILED_PRI
Designates the alerting priority of the FAILED_ALM, see Table 3-4 on page 31. The default is
0 and the recommended value are between 8 and 15.
FAILED_ACTIVE
This parameter displays which of the alarms is active. Only the alarm with the highest
priority will be displayed. This priority is not the same as the FAILED_PRI parameter
described above. This priority is hard coded within the device and is not user configurable.
FAILED_ALM
Alarm indicating a failure within a device which makes the device non-operational.
MAINT_ALARMS
A maintenance alarm indicates the device or some part of the device needs maintenance soon.
If the condition is ignored, the device will eventually fail. There are five parameters associated
with MAINT_ALARMS, they are described below.
MAINT_ENABLED
The MAINT_ENABLED parameter contains a list of conditions indicating the device or some
part of the device needs maintenance soon.
Table 3-6. Maintenance Alarms/Priority Alarm
Alarm
Priority
CJC Degraded
1
Body Temperature Out of Range
2
Sensor 1 Degraded
3
Sensor 2 Degraded
4
Sensor 3 Degraded
5
Sensor 4 Degraded
6
Sensor 5 Degraded
7
Sensor 6 Degraded
8
Sensor 7 Degraded
9
Sensor 8 Degraded
10
MAINT_MASK
The MAINT_MASK parameter will mask any of the failed conditions listed in
MAINT_ENABLED. A bit on means that the condition is masked out from alarming and will
not be reported.
MAINT_PRI
MAINT_PRI designates the alarming priority of the MAINT_ALM, Table 3-4 on page 31. The
default is 0 and the recommended values is 3 to 7.
Configuration
33
Reference Manual
Section 3: Configuration
00809-0100-4697, Rev GA
November 2014
MAINT_ACTIVE
The MAINT_ACTIVE parameter displays which of the alarms is active. Only the condition
with the highest priority will be displayed. This priority is not the same as the MAINT_PRI
parameter described above. This priority is hard coded within the device and is not user
configurable.
MAINT_ALM
An alarm indicating the device needs maintenance soon. If the condition is ignored, the
device will eventually fail.
Advisory alarms
An advisory alarm indicates informative conditions that do not have a direct impact on the
device's primary functions. There are five parameters associated with ADVISE_ALARMS, they are
described below.
ADVISE_ENABLED
The ADVISE_ENABLED parameter contains a list of informative conditions that do not have
a direct impact on the device's primary functions. Below is a list of the advisories with the
highest priority first.
Alarm
Priority
Excessive Deviation
1
Excessive Rate of Change
2
Check
3
Note
Alarms are only prioritized if Multi-Bit Alerts are disabled. If MBA is enabled, all alerts are visible.
ADVISE_MASK
The ADVISE_MASK parameter will mask any of the failed conditions listed in
ADVISE_ENABLED. A bit on means the condition is masked out from alarming and will not
be reported.
ADVISE_PRI
ADVISE_PRI designates the alarming priority of the ADVISE_ALM, see Table 3-4 on page 31.
The default is 0 and the recommended values are 1 or 2.
ADVISE_ACTIVE
The ADVISE_ACTIVE parameter displays which of the advisories is active. Only the advisory
with the highest priority will be displayed. This priority is not the same as the ADVISE_PRI
parameter described above. This priority is hard coded within the device and is not user
configurable.
ADVISE_ALM
ADVISE_ALM is an alarm indicating advisory alarms. These conditions do not have a direct
impact on the process or device integrity.
34
Configuration
Reference Manual
Section 3: Configuration
00809-0100-4697, Rev GA
3.4.4
November 2014
Recommended actions for PlantWeb alerts
RECOMMENDED_ACTION
The RECOMMENDED_ACTION parameter displays a text string that will give a recommended
course of action to take based on which type and which specific event of the PlantWeb alerts are
active.
Table 3-7. RB.RECOMMENDED_ACTION
Configuration
Alarm type
Active event
Recommended action
None
None
No action required
Advisory
Excessive Deviation
Verify the process temperature, sensor wiring, and check
sensor integrity.
Advisory
Excessive Rate of
Change
Verify sensor wiring is appropriate in each junction point and
check sensor integrity.
Maintenance
CJC Degraded
If T/C sensors are being used, restart the device. If condition
persists, replace the device.
Maintenance
Body Temperature
Out of Range
Verify the ambient temperature is within operating limits.
Maintenance
Sensor 1 Degraded
Confirm the operating range of Sensor 1 and/or verify the
sensor connection and device environment.
Maintenance
Sensor 2 Degraded
Confirm the operating range of Sensor 2 and/or verify the
sensor connection and device environment.
Maintenance
Sensor 3 Degraded
Confirm the operating range of Sensor 3 and/or verify the
sensor connection and device environment.
Maintenance
Sensor 4 Degraded
Confirm the operating range of Sensor 4 and/or verify the
sensor connection and device environment.
Maintenance
Sensor 5 Degraded
Confirm the operating range of Sensor 5 and/or verify the
sensor connection and device environment.
Maintenance
Sensor 6 Degraded
Confirm the operating range of Sensor 6 and/or verify the
sensor connection and device environment.
Maintenance
Sensor 7 Degraded
Conform the operating range of Sensor 7 and/or verify the
sensor connection and device environment.
Maintenance
Sensor 8 Degraded
Confirm the operating range of Sensor 8 and/or verify the
sensor connection and device environment.
Failed
Sensor 1 Failure
Verify the Sensor 1 Instrument process is within the Sensor
range and/or confirm sensor configuration and wiring.
Failed
Sensor 2 Failure
Verify the Sensor 2 Instrument process is within the Sensor
range and/or confirm sensor configuration and wiring.
Failed
Sensor 3 Failure
Verify the Sensor 3 Instrument process is within the Sensor
range and/or confirm sensor configuration and wiring.
Failed
Sensor 4 Failure
Verify the Sensor 4 Instrument process is within the Sensor
range and/or confirm sensor configuration and wiring.
Failed
Sensor 5 Failure
Verify the Sensor 5 Instrument process is within the Sensor
range and/or confirm sensor configuration and wiring.
Failed
Sensor 6 Failure
Verify the Sensor 6 Instrument process is within the Sensor
range and/or confirm sensor configuration and wiring.
Failed
Sensor 7 Failure
Verify the Sensor 7 Instrument process is within the Sensor
range and/or confirm sensor configuration and wiring.
35
Reference Manual
Section 3: Configuration
00809-0100-4697, Rev GA
November 2014
Table 3-7. RB.RECOMMENDED_ACTION
Alarm type
Active event
Recommended action
Failed
Sensor 8 Failure
Verify the Sensor 8 Instrument process is within the Sensor
range and/or confirm sensor configuration and wiring.
Failed
Body Temperature
Failure
Verify the ambient temperature is within the operating limits
of this device. If condition persists, replace the device.
Failed
Hardware/Software
Incompatible
Contact Service Center to verify the Device Information
(RESOURCE.HARDWARE_REV, AND
RESOURCE.RB_SFTWR_REV_ALL).
Failed
Memory Error
Restart the transmitter by writing the RESTART parameter to
4 - Restart Processor. If condition persists, replace the
transmitter.
Failed
Electronics Failure
Electronics failure has occurred. Restart the transmitter. If
condition persists, replace the transmitter.
Failed
ASIC Failure
ASIC failure has occurred. Restart the transmitter. If condition
persists, replace the transmitter.
Note
If status is set up to flag failure/warning you will see associated sensor degraded or failure alert.
3.4.5
Field Diagnostics Alerts
The Resource Block will act as a coordinator for Field Diagnostic Alerts. There will be four alarm
parameters (FD_FAILED_ALARM, FD_OFFSPEC_ALARM, FD_MAINT_ALARM, and
FD_CHECK_ALARM) that will contain information regarding some of the device errors that are
detected by the transmitter software. There will be a RECOMMENDED_ACTION parameter that
will be used to display the recommended action text for the highest priority alarm and a
HEALTH_INDEX parameters (0 - 100) indicating the overall health of the transmitter.
FD_FAILED_ALARM will have the highest priority followed by FD_OFFSPEC_ALARM,
FD_MAINT_ALARM and FD_CHECK_ALARM will be the lowest priority.
FD failed alarms
A failure alarm indicates a failure within a device that will make the device or some part of the
device non-operational. This implies that the device is in need of repair and must be fixed
immediately. There are five parameters associated with FD_FAILED_ALARMS specifically, they
are described below.
FD_FAILED_MAP
FD_FAIL_MAP parameter maps conditions to be detected as active for FD_FAIL_ALARM
category. Thus the same condition may be active in all, some, or none of the 4 alarm categories.
Below is a list of the failures with the highest priority first.
Table 3-8. FD Failure Alarms
Alarm
36
Priority
ASIC Failure
1
Electronics Failure
2
Hardware/Software Incompatible
3
Memory Failure
4
Body Temperature Failure
5
Configuration
Reference Manual
Section 3: Configuration
00809-0100-4697, Rev GA
November 2014
Table 3-8. FD Failure Alarms
Alarm
Priority
Sensor 1 Failure
6
Sensor 2 Failure
7
Sensor 3 Failure
8
Sensor 4 Failure
9
Sensor 5 Failure
10
Sensor 6 Failure
11
Sensor 7 Failure
12
Sensor 8 Failure
13
FD_FAILED_MASK
FD_FAIL_MASK parameter will mask any of the failed conditions listed in FD_FAILED_MAP. A bit
on means the condition is masked out from alarming and will not be reported.
FD_FAILED_PRI
Designates the alerting priority of the FD_FAILED_ALM, see Table 3-4 on page 31. The default is
0 and the recommended value are between 8 and 15.
FD_FAILED_ACTIVE
FD_FAIL_ACTIVE parameter displays the active alarms is active that are being selected for this
category. Only the alarm with the highest priority will be displayed. This priority is not the same
as the FD_FAILED_PRI parameter described above. This priority is hard coded within the device
and is not user configurable.
FD_FAILED_ALM
FD_FAIL_ALM indicates a failure within a device which makes the device non-operational.
FD_FAIL_ALM parameter is used primarily to broadcast a change in the associated active
conditions, which are not masked, for this alarm category to a Host System.
FD OFFSPEC ALARMS
An offspec alarm indicates that the device or some part of the device needs maintenance soon,
if the condition is ignored the device will eventually fail. There are five parameters associated
with FD OFFSPEC ALARMS, they are described below.
FD_OFFSPEC_MAP
FD_OFFSPEC_MAP parameter maps conditions to be detected as active for FD_ OFFSPEC
_ALARM category. Thus the same condition may be active in all, some, or none of the 4 alarm
categories. Below is a list of the failures with the highest priority first.
Table 3-9. FD Offspec Alarms
Alarm
Configuration
Priority
CJC Degraded
1
Body Temperature Out of Range
2
Sensor 1 Degraded
3
Sensor 2 Degraded
4
37
Reference Manual
Section 3: Configuration
00809-0100-4697, Rev GA
November 2014
Table 3-9. FD Offspec Alarms
Alarm
Priority
Sensor 3 Degraded
5
Sensor 4 Degraded
6
Sensor 5 Degraded
7
Sensor 6 Degraded
8
Sensor 7 Degraded
9
Sensor 8 Degraded
10
FD_ OFFSPEC _MASK
The FD_OFFSPEC_MASK parameter will mask any of the failed conditions listed in
FD_OFFSPEC_MAP. A bit on means the condition is masked out from alarming and will not be
reported.
FD_ OFFSPEC _PRI
FD_OFFSPEC_PRI designates the alarming priority of the FD_OFFSPEC _ALM, see Table 3-4 on
page 31. The default is 0 and the recommended values are 3 to 7.
FD_ OFFSPEC _ACTIVE
FD_ OFFSPEC _ACTIVE parameter displays the active alarms is active that are being selected for
this category. Only the alarm with the highest priority will be displayed. This priority is not the
same as the FD_ OFFSPEC _PRI parameter described above. This priority is hard coded within the
device and is not user configurable.
FD_ OFFSPEC _ALM
An alarm indicating the device needs maintenance soon. If the condition is ignored, the device
will eventually fail. FD_ OFFSPEC _ALM parameter is used primarily to broadcast a change in the
associated active conditions, which are not masked, for this alarm category to a Host System.
FD MAINT ALARMS
A maintenance alarm indicates informative conditions that do not have a direct impact on the
device's primary function(s).There are five parameters associated with MAINT_ALARMS, they
are described below.
FD_MAINT_MAP
The FD_MAINT_MAP parameter contains a list of conditions that do not have a direct impact on
the device's primary function(s).
Table 3-10. Maintenance Alarms/Priority Alarm
Alarm
Priority
Excessive Deviation
1
Excessive Rate of Change
2
FD_MAINT_MASK
The FD_MAINT_MASK parameter will mask any of the failed conditions listed in
FD_MAINT_ENABLED. A bit on means that the condition is masked out from alarming and will
not be reported.
38
Configuration
Reference Manual
Section 3: Configuration
00809-0100-4697, Rev GA
November 2014
FD_MAINT_PRI
FD_MAINT_PRI designates the alarming priority of the MAINT_ALM, Table 3-4 on page 31. The
default is 0 and the recommended value is greater than 2.
FD_MAINT_ACTIVE
FD_ MAINT_ACTIVE parameter displays the active alarms is active that are being selected for
this category. Only the alarm with the highest priority will be displayed. This priority is not the
same as the FD_ MAINT_PRI parameter described above. This priority is hard coded within the
device and is not user configurable.
FD_MAINT_ALM
FD_ MAINT_ALM indicates advisory alarms. These conditions do not have a direct impact on the
process or device integrity.
FD CHECK ALARMS
An advisory alarm indicates informative conditions that do not have a direct impact on the
device's primary functions. There are five parameters associated with ADVISE_ALARMS, they are
described below.
FD_CHECK_MAP
The FD_CHECK_MAP parameter contains a list of informative conditions that do not have a
direct impact on the device's primary functions. Below is a list of the advisories with the highest
priority first.
Table 3-11. Check Alarms
Alarm
Priority
Check
1
FD_ CHECK _MASK
The FD_CHECK _MASK parameter will mask any of the failed conditions listed in
FD_CHECK _MAP. A bit on means the condition is masked out from alarming and will not be
reported.
FD_ CHECK _PRI
FD_CHECK _PRI designates the alarming priority of the ADVISE_ALM, see Table 3-4 on page 31.
The default is 0 and the recommended values is 1.
FD_ CHECK _ACTIVE
The FD_CHECK _ACTIVE parameter displays which of the advisories is active. Only the advisory
with the highest priority will be displayed. This priority is not the same as the FD_CHECK _PRI
parameter described above. This priority is hard coded within the device and is not user
configurable.
FD_ CHECK _ALM
FD_CHECK _ALM is an alarm indicating advisory alarms. These conditions do not have a direct
impact on the process or device integrity.
Configuration
39
Reference Manual
Section 3: Configuration
00809-0100-4697, Rev GA
November 2014
3.4.6
Recommended actions for field diagnostics alerts
RECOMMENDED_ACTION
The RECOMMENDED_ACTION parameter displays a text string that will give a recommended
course of action to take based on which type and which specific event of the Field Diagnostics
alerts are active.
Table 3-12. RB.RECOMMENDED_ACTION
40
Alarm type
Active event
Recommended action
None
None
No action required
Failed
ASIC Failure
ASIC Failure has occurred. Restart the transmitter. If condition
persists, replace the transmitter
Failed
Electronics Failure
Electronics Failure has occurred. Restart the transmitter. If
condition persists, replace the transmitter.
Failed
Hardware/Software
Incompatible
Contact a Service Center and verify the Device Information
(RESOURCE.HARDWARE_REV and
RESOURCE.RB_SFTWR_REV)
Failed
Memory Failure
Restart the transmitter by writing the RESTART parameter to
4 - Restart Processor. If condition persists, replace the
transmitter.
Failed
Body Temperature
Failure
Verify the ambient temperature is within the operating limits
of this device. If condition persists, replace the device
Failed
Sensor 1 Failure
Verify the Instrument process for Sensor 1 is within the Sensor
range and/or confirm sensor configuration and wiring.
Failed
Sensor 2 Failure
Verify the Instrument process for Sensor 2 is within the Sensor
range and/or confirm sensor configuration and wiring.
Failed
Sensor 3 Failure
Verify the Instrument process for Sensor 3 is within the Sensor
range and/or confirm sensor configuration and wiring.
Failed
Sensor 4 Failure
Verify the Instrument process for Sensor 4 is within the Sensor
range and/or confirm sensor configuration and wiring.
Failed
Sensor 5 Failure
Verify the Instrument process for Sensor 5 is within the Sensor
range and/or confirm sensor configuration and wiring.
Failed
Sensor 6 Failure
Verify the Instrument process for Sensor 6 is within the Sensor
range and/or confirm sensor configuration and wiring.
Failed
Sensor 7 Failure
Verify the Instrument process for Sensor 7 is within the Sensor
range and/or confirm sensor configuration and wiring.
Failed
Sensor 8 Failure
Verify the Instrument process for Sensor 8 is within the Sensor
range and/or confirm sensor configuration and wiring.
Off Spec
CJC Degraded
If T/C sensors are being used, restart the device. If condition
persists, replace the device.
Off Spec
Body Temperature
Out of Range
Verify the ambient temperature is within operating limits
Off Spec
Sensor 1 Degraded
Confirm the operating range of Sensor 1 and/or verify the
sensor connection and device environment.
Off Spec
Sensor 2 Degraded
Confirm the operating range of Sensor 2 and/or verify the
sensor connection and device environment.
Off Spec
Sensor 3 Degraded
Confirm the operating range of Sensor 3 and/or verify the
sensor connection and device environment.
Configuration
Reference Manual
Section 3: Configuration
00809-0100-4697, Rev GA
November 2014
Table 3-12. RB.RECOMMENDED_ACTION
3.4.7
Alarm type
Active event
Recommended action
Off Spec
Sensor 4 Degraded
Confirm the operating range of Sensor 4 and/or verify the
sensor connection and device environment.
Off Spec
Sensor 5 Degraded
Confirm the operating range of Sensor 5 and/or verify the
sensor connection and device environment.
Off Spec
Sensor 6 Degraded
Confirm the operating range of Sensor 6 and/or verify the
sensor connection and device environment.
Off Spec
Sensor 7 Degraded
Confirm the operating range of Sensor 7 and/or verify the
sensor connection and device environment.
Off Spec
Sensor 8 Degraded
Confirm the operating range of Sensor 8 and/or verify the
sensor connection and device environment.
Maintenance
Excessive Deviation
Verify the process temperature, sensor wiring, and check
sensor integrity.
Maintenance
Excessive Rate of
Change
Verify sensor wiring is appropriate in each junction point and
check sensor integrity
Check
Check
Transducer block under maintenance
Transducer blocks
The transducer block allows the user to view and manage the channel information. There is one
transducer block for the eight sensors that contains specific temperature measurement data,
including:

Sensor Type

Engineering Units

Damping

Temperature Compensation

Diagnostics
Transducer block channel definitions
The 848T supports multiple sensor inputs. Each input has a channel assigned to it allowing an AI
or MAI Function Blocks to be linked to that input. The channels for the 848T are as follows:
Table 3-13. Channel Definitions for the 848T
Channel
Configuration
Description
Channel
Description
1
Sensor One
16
Sensor 3 Deviation
2
Sensor Two
17
Sensor 4 Deviation
3
Sensor Three
18
Sensor 5 Deviation
4
Sensor Four
19
Sensor 6 Deviation
5
Sensor Five
20
Sensor 7 Deviation
6
Sensor Six
21
Sensor 8 Deviation
7
Sensor Seven
22
Sensor 1 Rate Change
8
Sensor Eight
23
Sensor 2 Rate Change
9
Differential Sensor 1
24
Sensor 3 Rate Change
41
Reference Manual
Section 3: Configuration
00809-0100-4697, Rev GA
November 2014
Table 3-13. Channel Definitions for the 848T
Channel
Description
Channel
Description
10
Differential Sensor 2
25
Sensor 4 Rate Change
11
Differential Sensor 3
26
Sensor 5 Rate Change
12
Differential Sensor 4
27
Sensor 6 Rate Change
13
Body Temperature
28
Sensor 7 Rate Change
14
Sensor 1 Deviation
29
Sensor 8 Rate Change
15
Sensor 2 Deviation
Channel
1
Units/Ranging
A/D
Signal
Conversion
Temperature
Compensation
Linearization
Figure 3-1. Transducer Block Data Flow
Diagnostics
M
Damping
CJC
Channel
2
Channel
3
Channel
4
Channel
5
Channel
6
Channel
7
Channel
8
Channel
9
Channel
10
Channel
11
Channel
12
Channel
13
S1
S2
S3
S4
S5
S6
S7
S8
DS1
DS2
DS3
DS4
BT
Transducer block errors
The following conditions are reported in the BLOCK_ERR and XD_ERROR parameters.
BLOCK_ERR
Table 3-14. Block/Transducer Error
Condition number, name, and description
0
Other(1)
7
Input failure/process variable has bad status
15
Out of service: The actual mode is out of service
(1) If BLOCK_ERR is “other,” then see XD_ERROR.
Transducer block modes
The transducer block supports two modes of operation as defined by the MODE_BLK
parameter:
42
Configuration
Reference Manual
Section 3: Configuration
00809-0100-4697, Rev GA
November 2014
Automatic (Auto)
The block outputs reflect the analog input measurement.
Out of Service (OOS)
The block is not processed. Channel outputs are not updated and the status is set to Bad: Out of
Service for each channel. The BLOCK_ERR parameter shows Out of Service. In this mode,
changes can be made to all configurable parameters. The target mode of a block may be
restricted to one or more of the supported modes.
Transducer block alarm detection
Alarms are not generated by the transducer block. By correctly handling the status of the
channel values, the down stream block (AI or MAI) will generate the necessary alarms for the
measurement. The error that generated this alarm can be determined by looking at BLOCK-ERR
and XD_ERROR.
Transducer block status handling
Normally, the status of the output channels reflect the status of the measurement value, the
operating condition of the measurement electronics card, and any active alarm conditions. In a
transducer, PV reflects the value and status quality of the output channels.
Table 3-15. Transducer Block Parameters
Configuration
Number
Parameter
Description
0
BLOCK
1
ST_REV
The revision level of the static data associated with the
function block.
2
TAG_DESC
The user description of the intended application of the
block.
3
STRATEGY
The strategy field can be used to identify grouping of
blocks.
4
ALERT_KEY
The identification number of the plant unit.
5
MODE_BLK
The actual, target, permitted, and normal modes of the
block.
6
BLOCK_ERR
This parameter reflects the error status associated with
the hardware or software components associated with a
block. Multiple errors may be shown. For a list of
enumeration values, see FF-890, Block_Err formal model.
7
UPDATE_EVENT
This alert is generated by any change to the static data.
8
BLOCK_ALM
The BLOCK-ALM is used for all configuration, hardware,
connection failure or system problems in the block. The
cause of the alert is entered in the subcode field. The
first alert to become active will set the Active status in
the Status attribute. As soon as the Unreported status is
cleared by the alert reporting task, another block alert
may be reported without clearing the Active status, if
the subcode has changed.
9
TRANSDUCER_DIRECTORY
10
TRANSDUCER_TYPE
A directory that specified the number and stating
indices of the transducers in the transducer block.
Identifies the transducer that follows 101 – Standard
Temperature with Calibration.
43
Reference Manual
Section 3: Configuration
00809-0100-4697, Rev GA
November 2014
Table 3-15. Transducer Block Parameters
44
Number
Parameter
Description
11
XD_ERROR
Provides additional error codes related to transducer
blocks. For a list of enumeration values, see FF-902. See
tables below for a list of sub-parameters that pertain to
XD_ERROR messages.
12
COLLECTION_DIRECTORY
A directory that specifies the number, starting indices,
and DD Item ID’s of the data collections in each
transducer block.
13
SENSOR_1_CONFIG
Sensor Configuration Parameters. See tables below for a
list of sub-parameters that pertain to Sensor
Configuration functions.
14
PRIMARY_VALUE_1
The measured value and status available to the function
block.
15
SENSOR_2_CONFIG
Sensor Configuration parameters. See tables below for a
list of sub-parameters that pertain to Sensor
Configuration functions.
16
PRIMARY_VALUE_2
The measured value and status available to the function
block.
17
SENSOR_3_CONFIG
Sensor Configuration Parameters. See tables below for a
list of sub-parameters that pertain to Sensor
Configuration functions.
18
PRIMARY_VALUE_3
The measured value and status available to the function
block
19
SENSOR_4_CONFIG
Sensor Configuration Parameters. See tables below for a
list of sub-parameters that pertain to Sensor
Configuration functions.
20
PRIMARY_VALUE_4
The measured value and status available to the function
block.
21
SENSOR_5_CONFIG
Sensor Configuration Parameters. See tables below for a
list of sub-parameters that pertain to Sensor
Configuration functions.
22
PRIMARY_VALUE_5
The measured value and status available to the function
block.
23
SENSOR_6_CONFIG
Sensor Configuration Parameters. See tables below for a
list of sub-parameters that pertain to Sensor
Configuration functions.
24
PRIMARY_VALUE_6
The measured value and status available to the function
block.
25
SENSOR_7_CONFIG
Sensor Configuration Parameters. See tables below for a
list of sub-parameters that pertain to Sensor
Configuration functions.
26
PRIMARY_VALUE_7
The measured value and status available to the function
block.
27
SENSOR_8_CONFIG
Sensor Configuration Parameters. See tables below for a
list of sub-parameters that pertain to Sensor
Configuration functions.
28
PRIMARY_VALUE_8
The measured value and status available to the function
block
29
SENSOR_STATUS
Status of each individual sensor. See tables below for a
list of possible status messages.
Configuration
Reference Manual
Section 3: Configuration
00809-0100-4697, Rev GA
November 2014
Table 3-15. Transducer Block Parameters
Configuration
Number
Parameter
Description
30
SENSOR_CAL
Parameter structure to allow for calibration of each
sensor. See tables below for a list of sub-parameters that
pertain to Sensor Calibration functions.
31
CAL_STATUS
Status of the calibration that was previously performed.
See tables below for a list of possible Calibration
Statuses.
32
ASIC_REJECTION
33
BODY_TEMP
34
BODY_TEMP_RANGE
The range of the body temperature including the units
index.
35
TB_SUMMARY_STATUS
Overall summary status of the sensor transducer. See
tables below for a list of possible transducer statuses.
36
DUAL_SENSOR_1_CONFIG
Parameter structure to allow for calibration of each
differential measurement. See tables below for a list of
sub-parameters that pertain to Dual Sensor Calibration
functions.
37
DUAL_SENSOR_VALUE_1
The measured value and status available to the function
block.
38
DUAL_SENSOR_2_CONFIG
Parameter structure to allow for calibration of each
differential measurement. See tables below for a list of
sub-parameters that pertain to Dual Sensor Calibration
functions.
39
DUAL_SENSOR_VALUE_2
The measured value and status available to the function
block.
40
DUAL_SENSOR_3_CONFIG
Parameter structure to allow for calibration of each
differential measurement. See tables below for a list of
sub-parameters that pertain to Dual Sensor Calibration
functions.
41
DUAL_SENSOR_VALUE_3
The measured value and status available to the function
block.
42
DUAL_SENSOR_4_CONFIG
Parameter structure to allow for calibration of each
differential measurement. See tables below for a list of
sub-parameters that pertain to Dual Sensor Calibration
functions.
43
DUAL_SENSOR_VALUE_4
The measured value and status available to the function
block.
44
DUAL_SENSOR_STATUS
Status of each individual differential measurement. See
tables below for a list of possible Dual Sensor statuses.
45
VALIDATION_SNSR1_CONFIG
Validation configuration parameters. See tables below
for a list of sub-parameters that pertain to Validation
Configuration functions.
46
VALIDATION_SNSR1_VALUES
Validation value parameters. See tables below for a list
of sub-parameters that pertain to Validation values.
47
VALIDATION_SNSR2_CONFIG
Validation configuration parameters. See tables below
for a list of sub-parameters that pertain to Validation
Configuration functions.
48
VALIDATION_SNSR2_VALUES
Validation value parameters. See tables below for a list
of sub-parameters that pertain to Validation values.
A configurable power line noise rejection setting.
Body Temperature of the device.
45
Reference Manual
Section 3: Configuration
00809-0100-4697, Rev GA
November 2014
Table 3-15. Transducer Block Parameters
Number
Parameter
Description
49
VALIDATION_SNSR3_CONFIG
Validation configuration parameters. See tables below
for a list of sub-parameters that pertain to Validation
Configuration functions.
50
VALIDATION_SNSR3_VALUES
Validation value parameters. See tables below for a list
of sub-parameters that pertain to Validation values.
51
VALIDATION_SNSR4_CONFIG
Validation configuration parameters. See tables below
for a list of sub-parameters that pertain to Validation
Configuration functions.
52
VALIDATION_SNSR4_VALUES
Validation value parameters. See tables below for a list
of sub-parameters that pertain to Validation values.
53
VALIDATION_SNSR5_CONFIG
Validation configuration parameters. See tables below
for a list of sub-parameters that pertain to Validation
Configuration functions.
54
VALIDATION_SNSR5_VALUES
Validation value parameters. See tables below for a list
of sub-parameters that pertain to Validation values.
55
VALIDATION_SNSR6_CONFIG
Validation configuration parameters. See tables below
for a list of sub-parameters that pertain to Validation
Configuration functions.
56
VALIDATION_SNSR6_VALUES
Validation value parameters. See tables below for a list
of sub-parameters that pertain to Validation values.
57
VALIDATION_SNSR7_CONFIG
Validation configuration parameters. See tables below
for a list of sub-parameters that pertain to Validation
Configuration functions.
58
VALIDATION_SNSR7_VALUES
Validation value parameters. See tables below for a list
of sub-parameters that pertain to Validation values.
59
VALIDATION_SNSR8_CONFIG
Validation configuration parameters. See tables below
for a list of sub-parameters that pertain to Validation
Configuration functions.
60
VALIDATION_SNSR8_VALUES
Validation value parameters. See tables below for a list
of sub-parameters that pertain to Validation values.
61
SENSOR_GRAPH_LIMIT
62
DIFFERENTIAL_GRAPH_LIMIT
Sensor graph limit parameters
Differential graph limit parameters
Changing the sensor configuration in the transducer block
If the FOUNDATION fieldbus configuration tool or host system does not support the use of DD
methods for device configuration, the following steps illustrate how to change the sensor
configuration in the transducer block:
46
1.
Set the MODE_BLK.TARGET to OOS, or set the SENSOR_MODE to configuration.
2.
Set SENSOR_n_CONFIG.SENSOR to the appropriate sensor type, and then set
SENSOR_n_CONFIG.CONNECTION to the appropriate type and connection.
3.
In the Transducer Block, set MODE_BLK.TARGET to AUTO, or set the SENSOR_MODE to
operation.
Configuration
Reference Manual
Section 3: Configuration
00809-0100-4697, Rev GA
3.4.8
November 2014
Transducer block sub-parameter tables
Table 3-16. XD_ERROR Sub-parameter Structure
XD ERROR
Description
0
No Error
17
General Error
An error has occurred that could not be classified as one of the
errors listed below.
18
Calibration Error
An error occurred during calibration of the device or a calibration
error has been detected during operation of the device.
19
Configuration Error
An error occurred during configuration of the device or a
configuration error has been detected during operation of the
device.
20
Electronics Failure
An electronic component has failed.
22
I/O Failure
An I/O failure has occurred.
23
Data Integrity Error
Indicates that data stored within the system may no longer be valid
due to non-volatile memory checksum failure, data verify after
write failure, etc.
24
Software Error
The software has detected an error. This could be caused by an
improper interrupt service routine, an arithmetic overflow, a
watchdog timer, etc.
25
Algorithm Error
The algorithm used in the transducer block produced an error. This
could be due to an overflow, data reasonableness.
Table 3-17. SENSOR_CONFIG Sub-parameter Structure
Configuration
Parameter
Description
SENSOR_MODE
Disables or enables a sensor for configuration.
SENSOR_TAG
Sensor description.
SERIAL_NUMBER
Serial number for the attached sensor.
SENSOR
Sensor Type and Connection. MSB is the sensor type and LSB is the
connection.
DAMPING
Sampling Interval used to smooth output using a first order linear filter.
A value entered between 0 and the Update_Rate, will result in a
damping value equal to the Update_Rate.
INPUT_TRANSIENT_FILTER
Enables or Disables the option for reporting fast changing sensor inputs
without temporary holdoff. 0 = Disable, 1 = Enabled.
RTD_2_WIRE_OFFSET
User entered value for constant lead-wire resistance correction in a
2-wire RTD and ohm sensor types.
ENG_UNITS
The engineering units used for reporting measured sensor values.
UPPER_RANGE
The upper sensor limit for the selected sensor is displayed using
Units_Index sub parameter.
LOWER_RANGE
The lower sensor limit for the selected sensor is displayed using
Units_Index sub parameter.
47
Reference Manual
Section 3: Configuration
00809-0100-4697, Rev GA
November 2014
Table 3-18. SENSOR_STATUS Sub-parameter Structure
Sensor status table
0x00
Active
0x01
Out of Service
0x02
Inactive
0x04
Open
0x08
Short
0x10
Out of Range
0x20
Beyond Limits
0x40
Excess EMF Detected
0x80
Other
Table 3-19. SENSOR_CAL Sub-parameter Structure
Parameter
Description
SENSOR_NUMBER
The sensor number to calibrate
CALIB_POINT_HI
The High calibration point for the selected sensor
CALIB_POINT_LO
The Low calibration point for the selected sensor
CALIB_UNIT
The engineering units used for calibrating the sensor
CALIB_METHOD
The method of the last calibration for sensor
103 - factory trim standard calibration
104 - user trim standard calibration
CALIB_INFO
Information regarding the calibration
CALIB_DATE
Date that the calibration was completed
CALIB_MIN_SPAN
The minimum calibration span value allowed. This minimum span
information is necessary to ensure that when calibration is done, the two
calibrated points are not too close together
CALIB_PT_HI_LIMIT
The High calibration unit
CALIB_PT_LO_LIMIT
The Low calibration unit
Table 3-20. CAL_STATUS Structure
Cal status
48
0
No Command Active
1
Command Executing
2
Command Done
3
Command Done: Errors
Configuration
Reference Manual
Section 3: Configuration
00809-0100-4697, Rev GA
November 2014
Table 3-21. Transducer Status Sub-parameter Structure
Transducer status table
0x01
A/D Failure
0x02
Sensor Failure
0x04
Dual Sensor Failure
0x08
CJC Degraded
0x10
CJC Failure
0x20
Body Temp Failure
0x40
Sensor Degraded
0x80
Body Temperature Degraded
Table 3-22. DUAL_SENSOR CONFIG Sub-parameter Structure
Parameter
Description
DUAL_SENSOR_MODE
Disables or enables a sensor for configuration
DUAL_SENSOR_TAG
Differential description
INPUT_A
Sensor to be used in DUAL_SENSOR_CALC
INPUT_B
Sensor to be used in DUAL_SENSOR_CALC
DUAL_SENSOR_CALC
Equation used for the dual sensor measurement including: Not Used,
Difference (Input A - Input B), and Absolute Difference (Input A - Input B)
ENG_UNITS
Units used to display sensor parameter
UPPER_RANGE
Upper Differential Limit (Input A High - Input B Low)
LOWER_RANGE
Lower Differential Limit (Input A Low - Input B High)
Table 3-23. DUAL_SENSOR_STATUS Sub-parameter Structure
Dual sensor status table
Configuration
0x00
Active
0x01
Out of Service
0x02
Inactive
0x04
Component Sensor Open
0x08
Component Sensor Short
0x10
Component Sensor Out of Range or Degraded
0x20
Component Sensor Out of Limits
0x40
Component Sensor Inactive
0x80
Configuration Error
49
Reference Manual
Section 3: Configuration
00809-0100-4697, Rev GA
November 2014
Table 3-24. Validation Value Sub-parameter Structure
Validation value sub-parameter structure
Parameter
Description
VALIDATION_STATUS
State of the channel specific measurement validation measurement
DEVIATION_VALUE
Deviation output value
DEVIATION_STATUS
Status of the deviation output
RATE_OF_CHANGE_VALUE
Rate of change value output
RATE_OF_CHANGE_STATUS
Status of Rate of change output
Table 3-25. Validation Config Sub-parameter Structure
50
Parameter
Description
VALIDATION_MODE
Activates the measurement validation data gathering process
0 = Disable
1 = Enable
SAMPLE_RATE
Number of seconds per sample used for measurement validation
data collection. This shouldn't exceed 10 seconds per sample, but
currently there are no upper limits.
DEVIATION_LIMIT
Sets the limit for the deviation diagnostic. DD limits the upper range
to 10.
DEVIATION_ENG_UNITS
Units tied to the deviation output value
DEVIATION_ALERT_SEVERITY
Advisory, Maintenance, Failure
0 = Disabled = Does not use the limits, but provides an output
1 = Advisory = No effect on sensor status, sets advisory PWA
2 = Maint = Sets sensor status to uncertain, sets advisory PWA
3 = Failure = Sets sensor status to Bad, sets advisory PWA
DEVIATION_PCNT_LIM_HYST
Deviation Hysteresis Limit = (1 - DEVIATION_PCNT_LIM_HYST/100)
* DEVIATION_LIMIT
RATE_INCREASING_LIMIT
Increasing Rate of Change limit set point
RATE_DECREASING_LIMIT
Decreasing Rate of Change limit set point
RATE_ENG_UNITS
Units tied to the rate of change output value
RATE_ALERT_SEVERITY
Advisory, Maintenance, Failure
0 = Disabled = Does not use the limits, but provides an output
1 = Advisory = No effect on sensor status, sets advisory PWA
2 = Maint = Sets sensor status to uncertain, sets advisory PWA
3 = Failure = Sets sensor status to Bad, sets advisory PWA
RATE_PCNT_LIM_HYST
Rate of Change Increasing Hysteresis Limit = (1 RATE_PCNT_LIM_HYST/100) * RATE_INCREASING_LIMIT
Configuration
Reference Manual
Section 3: Configuration
00809-0100-4697, Rev GA
November 2014
Sensor calibration in the sensor transducer block
If the FOUNDATION fieldbus configuration tool or host system does not support the use of DD
methods for device configuration, the following steps illustrate how to calibrate the sensor from
the sensor transducer block.
Note
Active calibrators should not be used in conduction with RTDs on any multiple input
temperature transmitter such as the 848T.
Configuration
1.
Under SENSOR_CALIB, the SENSOR_NUMBER to the number of the sensor to calibrate.
2.
Set CALIB_UNIT to calibration unit.
3.
Set CALIB_METHOD to User Trim (seeTable on page 41 for valid values).
4.
Set the input value of the sensor simulator to be within the range defined by
CALIB_LO_LIMIT and CALIB_HI_LIMIT.
5.
Set CALIB_POINT_LO (CALIB_POINT_HI) to the value set at the sensor simulator.
6.
Read CALIB_STATUS and wait until it reads “Command Done”
7.
Repeat steps 3 to 5 if performing a two-point trim. Note that the difference in values
between CALIB_POINT_LO and CALIB_POINT_HI must be greater than
CALIB_MIN_SPAN.
51
Section 3: Configuration
November 2014
52
Reference Manual
00809-0100-4697, Rev GA
Configuration
Section 4: Operation and Maintenance
Reference Manual
November 2014
00809-0100-4697, Rev GA
Section 4
Operation and Maintenance
Safety messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Foundation™ fieldbus information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Hardware maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.1
page 53
page 53
page 54
page 55
Safety messages
Instructions and procedures in this section may require special precautions to ensure the safety
of the personnel performing the operations. Information that potentially raises safety issues is
indicated by a warning symbol ( ). Refer to the following safety messages before performing
an operation preceded by this symbol.
4.1.1
Warnings
Failure to follow these installation guidelines could result in death or serious injury.

Make sure only qualified personnel perform the installation.
Process leaks could result in death or serious injury.
Do not remove the thermowell while in operation. Removing while in operation may
cause process fluid leaks.

Install and tighten thermowells and sensors before applying pressure, or process
leakage may result.
Electrical shock could cause death or serious injury.



4.2
If the senor is installed in a high voltage environment and a fault condition or
installation error occurs, high voltage may be present on transmitter leads and
terminals.
Use extreme caution when making contact with the leads and terminals.
FOUNDATION™ fieldbus information
FOUNDATION fieldbus is an all-digital, serial, two-way, multi-drop communication protocol that
interconnects devices such as transmitters and valve controllers. It is a local area network (LAN)
for instruments that enable basic control and I/O to be moved to the field devices. The Model
848T uses FOUNDATION fieldbus technology developed and supported by Emerson Process
Management and the other members of the independent Fieldbus Foundation.
Operation and Maintenance
53
Reference Manual
Section 4: Operation and Maintenance
00809-0100-4697, Rev GA
November 2014
Table 4-1. Block Diagram for the Rosemount 848T
Function Blocks
•AI, MAI, and ISEL
FOUNDATION
fieldbus
Communications
Stack
Resource
Block
•physical
device
information
Analog-to-Digital
Signal Conversion
(8 sensors)
Transducer Block
Measurement Sensor
•sensor and differential
temp
•terminal temp.
•sensor configuration
•calibration
•diagnostics
Cold Junction
Input-to-Output
Isolation
4.2.1
Commissioning (addressing)
To be able to setup, configure, and have it communicate with other devices on a segment, a
device must be assigned a permanent address. Unless requested otherwise, it is assigned a
temporary address when shipped from the factory.
If there are two or more devices on a segment with the same address, the first device to start up
will use the assigned address (ex. Address 20). Each of the other devices will be given one of the
four available temporary addresses. If a temporary address is not available, the device will be
unavailable until a temporary address becomes available.
Use the host system documentation to commission a device and assign a permanent address.
4.3
Hardware maintenance
The 848T has no moving parts and requires a minimal amount of scheduled maintenance. If a
malfunction is suspected, check for an external cause before performing the diagnostics
presented below.
4.3.1
Sensor check
To determine whether the sensor is causing the malfunction, connect a sensor calibrator or
simulator locally at the transmitter. Consult an Emerson Process Management
representative for additional temperature sensor and accessory assistance.
4.3.2
Communication/power check
If the transmitter does not communicate or provides an erratic output, check for adequate
voltage to the transmitter. The transmitter requires between 9.0 and 32.0 VDC at the terminals
to operate with complete functionality. Check for wire shorts, open circuits, and multiple
grounds.
54
Operation and Maintenance
Reference Manual
Section 4: Operation and Maintenance
00809-0100-4697, Rev GA
4.3.3
November 2014
Resetting the configuration (RESTART)
There are two types of restarts available in the Resource Block. The following section outlines
the usage for each of these. For further information, see RESTART in Table 3-2 on page 3-25.
Restart processor (cycling)
Performing a Restart Processor has the same effect as removing power from the device and
reapplying power.
Restart with defaults
Performing a Restart with Defaults resets the static parameters for all of the blocks to their initial
state. This is commonly used to change the configuration and/or control strategy of the device,
including any custom configurations done at the Rosemount factory.
4.4
Troubleshooting
4.4.1
FOUNDATION fieldbus
Symptom
Possible cause
Corrective action
Device does not show
up in the live list
Network configuration
parameters are incorrect
Set the network parameters of the LAS (host system) according to
the FF Communications Profile
ST: 8
MRD: 4
DLPDU PhLO: 4
MID: 7
TSC: 4 (1 ms)
T1: 96000 (3 seconds)
T2: 9600000 (300 seconds)
T3: 480000 (15 seconds)
Network address is not in
polled range
Set first Unpolled Node and Number of UnPolled Nodes so that the
device address is within range
Power to the device is
below the 9 VDC minimum
Increase the power to at least 9V
Noise on the power /
communication is too high
Verify terminators and power conditioners are within specification
Verify that the shield is properly terminated and not grounded at
both ends. It is best to ground the shield at the power conditioner
Device that is acting
as a LAS does not send
out CD
LAS Scheduler was not
downloaded to the Backup
LAS device
Ensure that all of the devices that are intended to be a Backup LAS
are marked to receive the LAS schedule
All devices go off live
list and then return
Live list must be
reconstructed by Backup
LAS device
Current link setting and configured links settings are different. Set
the current link setting equal to the configured settings.
Operation and Maintenance
55
Reference Manual
Section 4: Operation and Maintenance
00809-0100-4697, Rev GA
November 2014
4.4.2
Resource block
Symptom
Possible causes
Corrective action
Mode will not leave
OOS
Target mode not set
Set target mode to something other than OOS.
Memory Failure
BLOCK_ERR will show the lost NV Data or Lost Static Data bit set.
Restart the device by setting RESTART to Processor. If the block
error does not clear, call the factory.
Features
FEATURES_SEL does not have Alerts enabled. Enable the report bit.
Notification
LIM_NOTIFY is not high enough. Set equal to MAX_NOTIFY.
Block Alarms Will not
work
4.4.3
Transducer block troubleshooting
Symptom
Possible causes
Corrective action
Mode will not leave
OOS
Target mode not set
Set target mode to something other than OOS.
A/D board check sum error
The A/D board has a checksum error.
Resource block
The actual mode of the Resource block is in OOS. See Resource
Block Diagnostics for corrective action.
Transducer Block
The actual mode of the Transducer Block is OOS.
The primary value is
BAD
56
Measurement
Look at the SENSOR_STATUS parameter (See Table 3-16 on
page 3-47)
Operation and Maintenance
Appendix A: Reference Data
Reference Manual
November 2014
00809-0100-4697, Rev GA
Appendix A
Reference Data
Functional specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Physical specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Performance specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Function blocks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Dimensional drawings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Ordering information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
A.1
Functional specifications
A.1.1
Inputs
page 57
page 58
page 60
page 59
page 64
page 67
Eight independently configurable channels including combinations of 2- and 3-wire RTDs,
thermocouples, mV, 2- and 3-wire and ohm inputs.
4–20 mA inputs using optional connector(s).
A.1.2
Outputs
Manchester-encoded digital signal that conforms to IEC 61158 and ISA 50.02.
A.1.3
A.1.4
Status

600 Vdc channel to channel isolation(1)

10 Vdc channel to channel isolation for all operating conditions with maximum 150 m.
(500 ft) of sensor lead length 18 AWG.
Ambient temperature limits
–40 to 185 °F (–40 to 85 °C)
A.1.5
Isolation
Isolation between all sensor channels is rated to 10Vdc over all operating conditions. No
damage will occur to the device with up to 600 Vdc between any sensor channel.
A.1.6
Power supply
Powered over FOUNDATION™ fieldbus with standard fieldbus power supplies. The transmitter
operates between 9.0 and 32.0 V dc, 22 mA maximum (transmitter power terminals are rated
to 42.4 V dc).
A.1.7
Transient protection
The transient protector (option code T1) helps to prevent damage to the transmitter from
transients induced on the loop wiring by lightning, welding, heavy electrical equipment, or
switch gears. This option is installed at the factory for the Rosemount 848T and is not intended
for field installation.
(1)
Reference conditions are -40 to 60 °C (-40 to 140 °F) with 30 m. (100 ft) of sensor lead length 18 AWG wire.
Reference Data
57
Reference Manual
Appendix A: Reference Data
00809-0100-4697, Rev GA
November 2014
A.1.8
Update time
Approximately 1.5 seconds to read all 8 inputs.
A.1.9
Humidity limits
0–99% non-condensing relative humidity
A.1.10
Turn-on time
Performance within specifications is achieved in less than 30 seconds after power is applied to
the transmitter.
A.1.11
Alarms
The AI and ISEL function blocks allow the user to configure the alarms to HI-HI, HI, LO, or LO-LO
with a variety of priority levels and hysteresis settings.
A.1.12
Backup Link Active Scheduler (LAS)
The transmitter is classified as a device link master, which means it can function as a Link Active
Scheduler (LAS) if the current link master device fails or is removed from the segment.
The host or other configuration tool is used to download the schedule for the application to the
link master device. In the absence of a primary link master, the transmitter will claim the LAS and
provide permanent control for the H1 segment.
FOUNDATION fieldbus parameters
Schedule Entries
Links
Virtual Communications Relationships (VCR)
A.2
Physical specifications
A.2.1
Mounting
20
30
20
The Rosemount 848T can be mounted directly onto a DIN rail or it can be ordered with an
optional junction box. When using the optional junction box, the transmitter can be mounted
onto a panel or a 2-in. pipe stand (with option code B6).
A.2.2
Entries for optional junction box
No entry

Used for custom fittings
Cable Gland

9 x M20 nickel-plated brass glands for 7.5–11.9 mm unarmored cable
Conduit

58
5 plugged 0.86-in. diameter holes suitable for installing 1/2-in. NPT fittings.
Reference Data
Appendix A: Reference Data
Reference Manual
November 2014
00809-0100-4697, Rev GA
Materials of construction for optional junction box
Junction box type
Paint
Aluminum
Epoxy Resin
Plastic
NA
Stainless Steel
NA
Aluminum Explosion-proof
NA
Weight
Weight
Assembly
Rosemount 848T only
Aluminum
Plastic
(1)
(1)
Stainless Steel
(1)
Aluminum Explosion-proof
oz
lb
kg
7.5
.47
.208
78.2
4.89
2.22
78.2
4.89
2.22
77.0
4.81
2.18
557
34.8
15.5
(1) Add 35.2 oz. (2.2 lb., 0.998 kg) for nickel-plated brass glands.
A.2.3
Environmental ratings
NEMA Type 4X and IP66 with optional junction box. JX3 Explosion-proof enclosure rated to -4 °F
(-20 °C).
A.3
Function blocks
A.3.1
Analog input (AI)
A.3.2
A.3.3

Processes the measurement and makes it available on the fieldbus segment.

Allows filtering, alarming, and engineering unit changes.
Input selector (ISEL)

Used to select between inputs and generate an output using specific selection
strategies such as minimum, maximum, midpoint, or average temperature.

Since the temperature value always contains the measurement status, this block allows
the selection to be restricted to the first “good” measurement.
Multiple analog input block (MAI)

Reference Data
The MAI block allows the eight AI blocks to be multiplexed together so they serve as
one function block on the H1 segment, resulting in greater network efficiency.
59
Reference Manual
Appendix A: Reference Data
00809-0100-4697, Rev GA
November 2014
A.4
Performance specifications
A.4.1
Stability
A.4.2

±0.1% of reading or 0.1 °C (0.18 °F), whichever is greater, for 2 years for RTDs

±0.1% of reading or 0.1 °C (0.18 °F), whichever is greater, for 1 year for thermocouples.
Self calibration
The transmitter’s analog-to-digital circuitry automatically self-calibrates for each temperature
update by comparing the dynamic measurement to extremely stable and accurate internal
reference elements.
A.4.3
Vibration effect
Transmitters are tested to high pipeline vibration specification per IEC 60770-1 1999 with no
effect on performance.
A.4.4
A.4.5
Electromagnetic compatibility compliance testing

Meets the criteria under IEC 61326:2006

Meets the criteria under European Union Directive 2004/108/EC
Accuracy
Table A-1. Input Options/Accuracy
Sensor option
Sensor reference
Input ranges
Accuracy over
range(s)
°C
°F
°C
°F
2- and 3-Wire RTDs
Pt 50 ( = 0.00391)
GOST 6651-94
–200 to 550
–328 to 1022
± 0.57
± 1.03
Pt 100 ( = 0.00391)
GOST 6651-94
–200 to 550
–328 to 1022
± 0.28
± 0.50
Pt 100 ( = 0.00385)
IEC 751; = 0.00385, 1995
–200 to 850
–328 to 1562
± 0.30
± 0.54
Pt 100 ( = 0.003916)
JIS 1604, 1981
–200 to 645
–328 to 1193
± 0.30
± 0.54
Pt 200 ( = 0.00385)
IEC 751; = 0.00385, 1995
–200 to 850
–328 to 1562
± 0.54
± 0.98
Pt 200 ( = 0.003916)
JIS 1604; = 0.003916, 1981
–200 to 645
–328 to 1193
± 0.54
± 0.98
Pt 500
IEC 751; = 0.00385, 1995
–200 to 850
–328 to 1562
± 0.38
± 0.68
Pt 1000
IEC 751; = 0.00385, 1995
–200 to 300
–328 to 572
± 0.40
± 0.72
Ni 120
Edison Curve No. 7
–70 to 300
–94 to 572
± 0.30
± 0.54
Cu 10
Edison Copper Winding No. 15
–50 to 250
–58 to 482
± 3.20
± 5.76
Cu 100 (a=428)
GOST 6651-94
-185 to 200
-365 to 392
± 0.48
±0.86
Cu 50 (a=428)
GOST 6651-94
-185 to 200
-365 to 392
± 0.96
±1.73
Cu 100 (a=426)
GOST 6651-94
-50 to 200
-122 to 392
± 0.48
±0.86
Cu 50 (a=426)
GOST 6651-94
-50 to 200
-122 to 392
± 0.96
±1.73
100 to 300
301 to 1820
212 to 572
573 to 3308
± 6.00
± 1.54
± 10.80
± 2.78
Thermocouples—Cold Junction Adds + 0.5 °C to Listed Accuracy
NIST Type B (Accuracy varies
according to input range)
60
NIST Monograph 175
Reference Data
Appendix A: Reference Data
Reference Manual
November 2014
00809-0100-4697, Rev GA
Table A-1. Input Options/Accuracy
Sensor option
Input ranges
Sensor reference
Accuracy over
range(s)
°C
°F
°C
°F
NIST Type E
NIST Monograph 175
–200 to 1000
–328 to 1832
± 0.40
± 0.72
NIST Type J
NIST Monograph 175
–180 to 760
–292 to 1400
± 0.70
± 1.26
NIST Type K
NIST Monograph 175
–180 to 1372
–292 to 2501
± 1.00
± 1.80
NIST Type N
NIST Monograph 175
–200 to 1300
–328 to 2372
± 1.00
± 1.80
NIST Type R
NIST Monograph 175
0 to 1768
32 to 3214
± 1.50
± 2.70
NIST Type S
NIST Monograph 175
0 to 1768
32 to 3214
± 1.40
± 2.52
NIST Type T
NIST Monograph 175
–200 to 400
–328 to 752
± 0.70
± 1.26
DIN L
DIN 43710
–200 to 900
–328 to 1652
± 0.70
± 1.26
DIN U
DIN 43710
–200 to 600
–328 to 1112
± 0.70
± 1.26
w5Re26/W26Re
ASTME 988-96
0 to 2000
32 to 3632
± 1.60
± 2.88
GOST Type L
GOST R 8.585-2001
-200 to 800
-392 to 1472
± 0.71
± 1.28
-50 to 85
-58 to 185
±3.50
± 6.30
Terminal Temperature
Ohm Input
0 to 2000 ohms
± 0.90 ohms
Millivolt Input
-10 to 100 mV
± 0.05 mV
-10 to 1000 mV
± 1.0 mA
4–20 mA
± 0.01 mA
4–20 mA
± 0.01 mA
1000 mV
4–20 mA (Rosemount)
4–20 mA (NAMUR)
(1)
(1)
Multi-point Sensors(2)
(1) Requires the S002 option code.
(2) Multi-point (up to 8 points) thermocouples and RTDs are available for purchase with the Rosemount 848T. Input ranges and accuracy for these sensors will depend
on the specific multi-point sensor chosen. For more information, contact your local Emerson representative.
A.4.6
Differential configuration notes
Differential capability exists between any two sensor types.
For all differential configurations, the input range is X to Y where:
A.4.7

X = Sensor A minimum - Sensor B max.

Y = Sensor A maximum - Sensor B min.
Accuracy for differential configurations
If sensor types are similar (for example, both RTDs or both thermocouples), the accuracy = 1.5
times worst case accuracy of either sensor type. If sensor types are dissimilar (for example, one
RTD and one thermocouple), the accuracy = Sensor 1 Accuracy + Sensor 2 Accuracy.
A.4.8
Analog sensors 4–20mA
Two types of 4–20 mA sensors are compatible with the Rosemount 848T. These types must be
ordered with the S002 option code complete with an analog connector kit. The alarm levels,
accuracy for each type are listed in Table A-2.
Reference Data
61
Reference Manual
Appendix A: Reference Data
00809-0100-4697, Rev GA
November 2014
Table A-2. Analog Sensors
A.4.9
Sensor option
Alarm levels
Accuracy
4–20mA (Rosemount Standard)
3.9 to 20.8 mA
± 0.01mA
4–20mA (NAMUR)
3.8 to 20.5 mA
± 0.01mA
Ambient temperature effect
Transmitter may be installed in locations where the ambient temperature is between -40 and
85 °C (-40 and 185 °F).
A.4.10
Ambient temperature effects
NIST type
Accuracy per 1.0 °C (1.8 °F) change in ambient temperature(1) °C
Temperature range (°C)
RTD
Pt 50 (α = 0.00391)
• 0.004 °C (0.0072 °F)
N/A
Pt 100 (α = 0.00391)
• 0.002 °C (0.0036 °F)
N/A
Pt 100 (α = 0.00385)
• 0.003 °C (0.0054 °F)
N/A
Pt 100 (α = 0.003916)
• 0.003 °C (0.0054 °F)
N/A
Pt 200 (α = 0.003916)
• 0.004 °C (0.0072 °F)
N/A
Pt 200 (α = 0.00385)
• 0.004 °C (0.0072 °F)
N/A
Pt 500
• 0.003 °C (0.0054 °F)
N/A
Pt 1000
• 0.003 °C (0.0054 °F)
N/A
Cu 10
• 0.03 °C (0.054 °F)
N/A
Cu 100 (a=428)
• 0.002 °C (0.0036 °F)
N/A
Cu 50 (a=428)
• 0.004 °C (.0072 °F)
N/A
Cu 100 (a=426)
• 0.002 °C (0.0036 °F)
N/A
Cu 50 (a=426)
• 0.004 °C (.0072 °F)
N/A
Ni 120
• 0.003 °C (0.0054 °F)
N/A
Thermocouple (R = the value of the reading)
62
Type B
• 0.014 °C
• 0.032 °C - (0.0025% of (R - 300))
• 0.054 °C - (0.011% of (R - 100))
• R ≥ 1000
• 300 ≤ R < 1000
• 100 ≤ R < 300
Type E
• 0.005 °C + (0.00043% of R)
• All
Type J, DIN Type L
• 0.0054 °C + (0.00029% of R)
• 0.0054 °C + (0.0025% of |R|)
• R≥0
• R<0
Type K
• 0.0061 °C + (0.00054% of R)
• 0.0061 °C + (0.0025% of |R|)
• R≥0
• R<0
Type N
• 0.0068 °C + (0.00036% of R)
• All
Reference Data
Appendix A: Reference Data
Reference Manual
November 2014
00809-0100-4697, Rev GA
NIST type
Accuracy per 1.0 °C (1.8 °F) change in ambient temperature(1) °C
Temperature range (°C)
Type R, Type S
• 0.016 °C
• 0.023 °C - (0.0036% of R)
• R ≥ 200
• R < 200
Type T, DIN Type U
• 0.0064 °C
• 0.0064 °C + (0.0043% of |R|)
• R≥0
• R<0
GOST Type L
• 0.007 °C
• 0.007 °C + (0.003% of IRI)
• R≥0
• R<0
Millivolt
• 0.0005 mV
N/A
2- and 3-wire ohm
• 0.0084 ohms
N/A
4–20 mA (Rosemount)
• 0.0001 mA
N/A
4-20 mA (NAMUR)
• 0.0001 mA
N/A
(1) Change in ambient is in reference to the calibration temperature of the transmitter (20 °C (68 °F) typical from the factory).
A.4.11
Ambient temperature notes
Examples
When using a Pt 100 (α = 0.00385) sensor input at 30 °C ambient temperature:
Reference Data

Digital Temperature Effects: 0.003 °C x (30 - 20) = 0.03 °C

Worst Case Error: Digital + Digital Temperature Effects = 0.3 °C + 0.03 °C = 0.33 °C

Total Probable Error
0.30 2 + 0.03 2 = 0.30°C
63
Reference Manual
Appendix A: Reference Data
00809-0100-4697, Rev GA
November 2014
A.5
Dimensional drawings
Junction boxes with no entries (option codes JP1, JA1, and JS1)– external dimensions are the
same as those outlined for the other junction box materials in this section.
Rosemount 848T
Top View
Side View
3-D View
1.7
(43)
Simulation Switch
Security Switch
6.7
(170
3.7
(93)
Removable Wiring
Connection
Dimensions are in inches (millimeters).
Aluminum/Plastic Junction Box—Cable Gland (Option Codes JA2 and JP2)
Top View
3-D View
10.24 (260)
Ground Screw
Side View
Front View
7.84 (199.2)
2.44 (62)
1.57 (40)
2.28 (58)
1.10 (28)
6.30 (160)
4.41 (112)
1.73 (44)
3.78 (96)
Dimensions are in inches (millimeters).
64
Reference Data
Appendix A: Reference Data
Reference Manual
November 2014
00809-0100-4697, Rev GA
Stainless-Steel Junction Box—Cable Gland (Option Code JS2)
Top View
3-D View
9.91 (231)
7.7 (196)
Ground Screw
Side View
Front View
9.14 (232.2)
1.8 (46)
1.1 (28)
7.72 (196)
6.61 (168)
1.73 (44)
4.0 (102)
2.4 (62)
1.8 (47)
1.2 (30)
Dimensions are in inches (millimeters)
Aluminum/Plastic Junction Box—Conduit Entry (Option Codes JA3 and JP3)
Top view
3-D view
10.2 (260)
Front view
Side view
157 (40)
2.44 (62)
10.2 (260)
3.5 (89)
1.7 (42)
Five Plugged 0.86-in. diameter holes
suitable for installing 1/2-in. NPT
fittings
Dimensions are in inches (millimeters).
Reference Data
65
Reference Manual
Appendix A: Reference Data
00809-0100-4697, Rev GA
November 2014
Stainless-Steel Junction Box—Conduit Entry (Option Code JS3)
Top view
3-D view
9.1 (231)
7.7 (196)
Ground Screw
Front view
Side view
1.4 (35)
1.1 (27)
2.8 (70)
1.2 (30)
4.0 (102)
2.4 (62)
1.6 (42)
1.8 (4.7)
0.06 (1.5)
Five Plugged 0.86-in. diameter holes suitable for installing 1/2-in. NPT fittings
Dimensions are in inches (millimeters).
A.5.1
Mounting options
Aluminum/Plastic Junction Box
(Styles JA and JP)
Front view
5.1
(130)
10.2
(260)
Side view
6.6 (167) fully
assembled
Stainless-Steel Junction Box
(Style JS)
Front view
Side view
4.5
(114)
7.5 (190) fully
assembled
Dimensions are in inches (millimeters).
Aluminum/Plastic Junction Box
Mounted on a Vertical Pipe
66
Stainless-Steel Junction Box
Mounted on a Vertical Pipe
Reference Data
Reference Manual
Appendix A: Reference Data
November 2014
00809-0100-4697, Rev GA
A.6
Ordering information
Table A-3. Rosemount 848T FOUNDATION fieldbus Ordering Information
★ The Standard offering represents the most common options. The starred options (★) should be selected for best delivery.
__The Expanded offering is subject to additional delivery lead time.
Model
848T
Product description
High Density Temperature Measurement Family
Transmitter output
F
FOUNDATION fieldbus digital signal (includes AI, MAI, and ISEL function blocks, and Backup Link
Active Scheduler)
Product certifications(1)
I1
I3
I4
H4
I5(2)
I6(2)
I7
IA
IE
IF(2)
IG
N1
N5
N6
N7
NC
ND
NJ
NK
NA
E6
ATEX Intrinsic Safety
NEPSI Intrinsic Safety
TIIS Intrinsically Safety (FISCO) Type “ia”
TIIS Intrinsic Safety (FISCO) Type “ib”
FM Intrinsically Safe
CSA Intrinsically Safe
IECEx Intrinsic Safety
ATEX FISCO Intrinsic Safety
FM FISCO Intrinsically Safe
CSA FISCO Intrinsically Safe, Division 2
IECEx FISCO (Intrinsic Safety)
ATEX Type n (enclosure required)
FM Class I, Division 2, and Dust Ignition-proof (enclosure required)
CSA Class I, Division 2
IECEx Type n (enclosure required)
ATEX Type n Component (Ex nA nL)
ATEX Dust (enclosure required)
IECEx Type n Component (Ex nA nL)
FM Class 1, Division 2
No Approval
CSA Explosion-proof, Dust Ignition-proof, Division 2 (JX3 enclosure required)
★
Rosemount
junction
box required?
No
No
No
No
No
No
No
No
No
No
No
Yes
Yes
No
Yes
No(3)
Yes
No(3)
No
No
Yes(4)
★
★
★
★
★
★
★
★
★
★
★
★
★
★
★
★
★
★
★
★
Options (include with selected model number)
Input types
S001
S002(5)
RTD, Thermocouple, mV, ohm Inputs
RTDs, Thermocouple, mV, ohm and 4–20 mA Inputs
★
★
PlantWeb advanced diagnostics
D04
Measurement Validation Diagnostic
★
Transient protection
T1
Integral Transient Protector
★
Mounting bracket
B6
Mounting Bracket for 2-in. pipe mounting – SST bracket and bolts
★
Enclosure options
JP1
JP2
Plastic Junction Box; No Entries
Plastic Box, Cable Glands (9 x M20 nickel-plated brass glands for 7.5–11.9 mm unarmored cable)
Reference Data
★
★
67
Reference Manual
Appendix A: Reference Data
00809-0100-4697, Rev GA
November 2014
Table A-3. Rosemount 848T FOUNDATION fieldbus Ordering Information
★ The Standard offering represents the most common options. The starred options (★) should be selected for best delivery.
__The Expanded offering is subject to additional delivery lead time.
JP3
JA1
JA2
JA3
JS1
JS2
JS3
JX3(6)
Plastic Box, Conduit Entries (5 Plugged Holes, suitable for installing 1/2-in. NPT fittings)
Aluminum Junction Box; No Entries
Aluminum Cable Glands (9 x M20 nickel-plated brass glands for 7.5–11.9 mm unarmored cable)
Aluminum Conduit Entries (5 Plugged Holes, suitable for installing 1/2-in. NPT fittings)
Stainless Steel Junction Box; No Entries
Stainless Steel Box, Cable Glands (9 x M20 nickel-plated brass glands for 7.5–11.9 mm unarmored
cable)
Stainless Steel Box, Conduit Entries (5 Plugged Holes, suitable for installing 1/2-in. NPT fittings)
Explosion-proof Box, Conduit Entries (4 Plugged Holes, suitable for installing 1/2-in. NPT fittings)
★
★
★
★
★
★
★
★
Software configuration
C1
Custom Configuration of Date, Descriptor, Message and Wireless Parameters (Requires CDS with
Order)
★
50 Hz Line Voltage Filter
★
Line filter
F5
Calibration certificate
Q4
★
Calibration Certificate (3-Point Calibration)
Shipboard certification
SBS
SLL
★
★
American Bureau of Shipping (ABS) Type Approval
Lloyd's Register (LR) Type Approval
Special temperature test
LT
Test to -60 °F (-51.1 °C)
Conduit electrical connector
GE(7)
GM(7)
Typical model number:
(1)
(2)
(3)
(4)
(5)
(6)
(7)
68
★
★
M12, 4-pin, Male Connector (eurofast®)
A size Mini, 4-pin, Male Connector (minifast®)
848T
F
I5
S001
T1
B6
JA2
Consult factory for availability.
Available only with S001 option.
The Rosemount 848T ordered with component approval is not approved as a stand-alone unit. Additional system certification is required.
Enclosure Option JX3 must be ordered with Product Certification Code E6. (O-ring for the JX3 enclosure rated to -20 °C).
S002 is only available with Product Certification N5, N6, N1, NC, NK, and NA.
JX3 Explosion-proof enclosure rated to -4 °F (-20 °C).
Available with no approval or Intrinsically Safe approvals only. For FM Intrinsically Safe (option code I5), install in accordance with Rosemount drawing 00848-4402.
Reference Data
Appendix B: Product Certifications
Reference Manual
November 2014
00809-0100-4697, Rev GA
Appendix B
Product Certificates
European Directive Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 69
Intrinsically Safe and Non-Incendive installations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 81
Installation drawings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 82
B.1
European Directive Information
A copy of the EC Declaration of Conformity can be found at the end of the Quick Start Guide. The
most recent revision of the EC Declaration of Conformity can be found at www.rosemount.com.
B.2
Ordinary Location Certification from FM
Approvals
As standard, the transmitter has been examined and tested to determine that the design meets
the basic electrical, mechanical, and fire protection requirements by FM Approvals, a nationally
recognized test laboratory (NRTL) as accredited by the Federal Occupational Safety and Health
Administration (OSHA).
B.3
Installing Equipment in North America
The US National Electrical Code (NEC) and the Canadian Electrical Code (CEC) permit the use of
Division marked equipment in Zones and Zone marked equipment in Divisions. The markings
must be suitable for the area classification, gas, and temperature class. This information is
clearly defined in the respective codes.
B.3.1
USA
I5
FM Intrinsically Safe and Nonincendive
Certificate: 3011568
Standards: FM Class 3600:1998, FM Class 3610:2010, FM Class 3611:2004, FM Class
3810:2005, ANSI/ISA 60079-0:2009, ANSI/ISA 60079-11:2009, NEMA
250:1991, IEC 60529:2011
Markings: IS CL I, DIV 1, GP A, B, C, D; T4 (-50 °C ≤ Ta ≤ +60 °C); NI CL I, DIV 2, GP A, B, C, D;
T4A (-50 °C ≤ Ta ≤ +85 °C); T5 (-50 °C ≤ Ta ≤ +70 °C) when installed per
Rosemount drawing 00848-4404
Note
Transmitters marked with Nonincendive CL I, DV 2 can be installed in Division 2 locations using
general Division 2 wiring methods or Nonincendive Field Wiring (NIFW). See Drawing
00848-4404.
Product Certificates
69
Reference Manual
Appendix B: Product Certifications
00809-0100-4697, Rev GA
November 2014
IE
FM FISCO
Certificate: 3011568
Standards: FM Class 3600:1998, FM Class 3610:2010, FM Class 3611:2004, FM Class
3810:2005, ANSI/ISA 60079-0:2009, ANSI/ISA 60079-11:2009, NEMA
250:1991, IEC 60529:2011
Markings: IS CL I, DIV 1, GP A, B, C, D; T4 (-50 °C ≤ Ta ≤ +60 °C); NI CL I, DIV 2, GP A, B, C, D;
T4A (-50 °C ≤ Ta ≤ +85 °C); T5 (-50 °C ≤ Ta ≤ +70 °C) when installed per
Rosemount drawing 00848-4404
N5
Nonincendive and Dust-Ignition-proof
Certificate: 3011568
Standards: FM Class 3600:1998, FM Class 3611:2004, FM Class 3810:2005, ANSI/ISA
60079-0:2009, NEMA 250:1991, IEC 60529:2011
Markings: NI CL I, DIV 2, GP A, B, C, D; DIP CL II/III, DIV 1, GP E, F, G; T4A (-50 °C ≤ Ta ≤ +85
°C); T5 (-50 °C ≤ Ta ≤ +70 °C) when installed per Rosemount drawing
00848-4404; Type 4X
NK
Nonincendive
Certificate: 3011568
Standards: FM Class 3600:1998, FM Class 3611:2004, FM Class 3810:2005, ANSI/ISA
60079-0:2009, NEMA 250:1991, IEC 60529:2001
Markings: NI CL I, DIV 2, GP A, B, C, D; T4A (-50 °C ≤ Ta ≤ +85 °C); T5 (-50 °C ≤ Ta ≤ +70 °C)
when installed per Rosemount drawing 00848-4404
Note
Only the N5 and NK are valid with the S002 option.
Table B-1. Entity Parameters
B.3.2
FIELDBUS (Input)
FISCO (Input)
Nonincendive
(Input)
Sensor Field Terminal
(Output)
Vmax = 30 V
Vmax = 17.5 V
Vmax = 42.4 V
VOC = 12.5 V
Imax = 300 mA
Imax = 380 mA
Ci = 2.1 nF
ISC = 4.8 mA
Pi = 1.3 W
Pi = 5.32 W
Li = 0 H
Po = 15 mW
Ci = 2.1 nF
Ci = 2.1 nF
-
CA = 1.2 mF
Li = 0 H
Li = 0 H
-
LA = 1 H
Canada
E6
70
CSA Explosion-proof, Dust-Ignition-proof, Division 2 (JX3 Enclosure Required)
Certificate: 1261865
Standards: CAN/CSA C22.2 No. 0-M91 (R2001), CSA Std. C22.2 No. 25.1966, CSA Std.
C22.2 No. 30-M1986, CAN/CSA C22.2 No. 94-M91, CSA Std. C22.2 No.
142-M1987, CSA Std. C22.2 No. 213-M1987, CSA Std. C22.2 No. 60529:05
Markings: Explosion-proof for Class I, Division 1, Groups B, C, and D; T4 (-40 °C ≤ Ta ≤ +40
°C) when installed per Rosemount drawing 00848-1041; Dust-Ignition-proof
for Class II, Division 1, Groups E, F, and G; Class III; Class I, Division 2, Groups A,
B, C, and D; T3C (-50 °C ≤ Ta ≤ +60 °C) when installed per Rosemount drawing
00848-4405; Conduit Seal Required
Product Certificates
Reference Manual
Appendix B: Product Certifications
00809-0100-4697, Rev GA
November 2014
I6
CSA Intrinsically Safe and Division 2
Certificate: 1261865
Standards: CAN/CSA C22.2 No. 0-M91 (R2001), CAN/CSA C22.2 No. 94-M91, CSA Std.
C22.2 No. 142-M1987, CSA Std. C22.2 No. 157-92, CSA Std. C22.2 No.
213-M1987, CSA Std. C22.2 No. 60529:05
Markings: Intrinsically Safe for Class I, Division 1, Groups A, B, C, and D; T3C (-50 °C ≤ Ta ≤
+60 °C) when installed per Rosemount drawing 00848-4405; Class I, Division
2, Groups A, B, C, D; T3C (-50 °C ≤ Ta ≤ +60 °C) when installed per Rosemount
drawing 00848-4405
IF
CSA FISCO
Certificate: 1261865
Standards: CAN/CSA C22.2 No. 0-M91 (R2001), CAN/CSA C22.2 No. 94-M91, CSA Std.
C22.2 No. 142-M1987, CSA Std. C22.2 No. 157-92, CSA Std. C22.2 No.
213-M1987, CSA Std. C22.2 No. 60529:05
Markings: Intrinsically Safe for Class I, Division 1, Groups A, B, C, and D; T3C (-50 °C ≤ Ta ≤
+60 °C) when installed per Rosemount drawing 00848-4405; Class I, Division 2,
Groups A, B, C, D; T3C (-50 °C ≤ Ta ≤ +60 °C) when installed per Rosemount
drawing 00848-4405
N6
CSA Division 2 and Dust-Ignition-proof (enclosure required)
Certificate: 1261865
Standards: CAN/CSA C22.2 No. 0-M91 (R2001), CSA Std. C22.2 No. 30-M1986, CAN/CSA
C22.2 No. 94-M91, CSA Std. C22.2 No. 142-M1987, CSA Std. C22.2 No.
213-M1987, CSA Std. C22.2 No. 60529:05
Markings: Class I, Division 2, Groups A, B, C, and D; T3C (-50 °C ≤ Ta ≤ +60 °C) when
installed per Rosemount drawing 00848-4405; Dust-Ignition-proof for Class II,
Division 1, Groups E, F, and G; Class III; Conduit Seal Required
B.4
Europe
I1
ATEX Intrinsic Safety
Certificate: Baseefa09ATEX0093X
Standards: EN 60079-0:2012, EN60079-11:2012
Marking
II 1 G Ex ia IIC T4 Ga (-50 °C ≤ Ta ≤ +60 °C) when installed per drawing
00848-4406
Special Conditions for Safe Use (X):
Product Certificates
1.
The equipment must be installed in an enclosure that provides a degree of
protection of at least IP20. Non-metallic enclosures must be suitable to prevent
electrostatic hazards and light allow or zirconium enclosures must be protected
from impact and friction when installed.
2.
The equipment is note capable of withstanding the 500V insulation test required
by EN 60079-11:2012, clause 6.3.13. This must be taken into account when
installing the equipment.
71
Reference Manual
Appendix B: Product Certifications
00809-0100-4697, Rev GA
November 2014
Table B-2. ATEX Entity Parameters
IA
Fieldbus (Input)
Sensor Field Terminal
(Output)
Ui = 30 V
Uo = 12.5 V
Ii = 300 mA
Io = 4.8 mA
Pi = 1.3 W
Po = 15 mW
Ci = 2.1 nF
Co = 1.2 μF
Li = 0
Lo = 1 H
ATEX FISCO Intrinsic Safety
Certificate: Baseefa09ATEX0093X
Standards: EN 60079-0:2012, EN60079-11:2012
Marking
II 1 G Ex ia IIC T4 Ga (-50 °C ≤ Ta ≤ +60 °C) when installed per drawing
00848-4406
Special Conditions for Safe Use (X):
1.
The equipment must be installed in an enclosure that provides a degree of
protection of at least IP20. Non-metallic enclosures must be suitable to prevent
electrostatic hazards and light allow or zirconium enclosures must be protected
from impact and friction when installed.
2.
The equipment is note capable of withstanding the 500V insulation test required
by EN 60079-11:2012, clause 6.3.13. This must be taken into account when
installing the equipment.
Table B-3. ATEX FISCO Entity Parameters
N1
Power/bus
Sensor
Ui = 17.5 V
Uo = 12.5 V
Ii = 380 mA
Io = 4.8 mA
Pi = 5.32 W
Po = 15 mW
Ci = 2.1 nF
Co = 1.2 mF
Li = 0
Lo = 1 H
ATEX Type n (with enclosure)
Certificate: Baseefa09ATEX0095X
Standards: EN 60079-0:2006, EN60079-15:2005
Marking
II 3 G Ex nA nL IIC T5 (– 40 °C ≤ Ta ≤ + 65 °C)
Special Conditions for Safe Use (X):
NC
72
1.
Provision must be made, external to the apparatus, to ensure the rated voltage of
the apparatus supply is not exceeded by transient disturbances of more than 40.
2.
The electrical circuit is connected directly to earth; this must be taken into
account when installing the apparatus.
ATEX Type n (without enclosure)
Certificate: Baseefa09ATEX0094U
Standards: EN 60079-0:2006, EN60079-15:2005
Marking
II 3 G Ex nA nL IIC T4 (-50 °C ≤ Ta ≤ +85 °C), T5(-50 °C ≤ Ta ≤ +70 °C)
Product Certificates
Reference Manual
Appendix B: Product Certifications
00809-0100-4697, Rev GA
November 2014
Special Conditions for Safe Use (X):
1.
The component must be installed in a suitable component certified enclosure
that provides a degree of protection of at least IP54 and meets the relevant
material and environmental requirements of EN 60079-0:2006 and EN
60079-15:2005.
2.
Provision must be made, external to the apparatus, to ensure the rated voltage of
the apparatus supply is not exceeded by transient disturbances of more than
40%.
3.
The electrical circuit is connected directly to earth; this must be taken into
account when installing the apparatus.
Note
The 848T may also be installed in an external energy limited circuit as Ex nL IIC. In this case the
following parameters apply
Table B-4. Baseefa Entity Parameters
ND
Power/Bus (Input)
Sensor Field Terminal
(Output)
Ui = 42.4 V
Uo = 12.5 V
Ci = 2.1 nF
Io = 2.5 mA
Li = 0
Co = 1000 mF
-
Lo = 1 H
ATEX Dust
Certificate: BAS01ATEX1315X
Standards: EN 50281-1-1:1998
Marking
II 1 D T90 (-40 °C ≤ Ta ≤ +65 °C); IP66
Special Conditions for Safe Use (X):
B.5
1.
The user must ensure that the maximum rated voltage and current (42.4 volts,
22 milliamp DC) are not exceeded. All connections to other apparatus or
associated apparatus shall have control over this voltage and current equivalent
to a category “ib” circuit according to EN 50020.
2.
Component approved EEx e cable entries must be used which maintain the
ingress protection of the enclosure to at least IP66.
3.
Any unused cable entry holes must be filled with component approved EEx e
blanking plugs.
4.
The ambient temperature range of use shall be the most restrictive of the
apparatus, cable gland or blanking plug.
International
I7
Product Certificates
IECEx Intrinsic Safety
Certificate: IECEx BAS 09.0030X
Standards: IEC 60079-0:2011, IEC60079-11:2011
Markings: II 1 G Ex ia IIC T4 Ga (– 50 °C ≤ Ta ≤ + 60 °C)
73
Reference Manual
Appendix B: Product Certifications
00809-0100-4697, Rev GA
November 2014
Special Conditions of Safe Use (X):
IG
1.
The equipment must be installed in an enclosure that provides a degree of
protection of at least IP20. Non-metallic enclosures must be suitable to prevent
electrostatic hazards and light allow or zirconium enclosures must be protected
from impact and friction when installed.
2.
The equipment is note capable of withstanding the 500V insulation test required
by EN 60079-11:2012, clause 6.3.13. This must be taken into account when
installing the equipment.
IECEx FISCO Intrinsic Safety
Certificate: IECEx BAS 09.0030X
Standards: IEC 60079-0:2011, IEC60079-11:2011
Markings: II 1 G Ex ia IIC T4 Ga (-50 °C ≤ Ta ≤ +60 °C)
Special Conditions of Safe Use (X):
1.
The equipment must be installed in an enclosure that provides a degree of
protection of at least IP20. Non-metallic enclosures must be suitable to prevent
electrostatic hazards and light allow or zirconium enclosures must be protected
from impact and friction when installed.
2.
The equipment is note capable of withstanding the 500V insulation test required
by EN 60079-11:2012, clause 6.3.13. This must be taken into account when
installing the equipment.
Table B-5. IECEx Entity Parameters
N7
FISCO (Input)
Sensor Field Terminal
(Output)
Ui =17.5 Vdc
Uo = 12.5 Vdc
Ii = 380 mA
Io = 4.8 mA
Pi = 5.32 W
Po = 15 mW
Ci = 2.1 nF
Co = 1.2 mF
Li = 0
Lo = 1 H
IECEx Type n Approval
Certificate: IECEx BAS 09.0032X
Standards: IEC 60079-0:2004, IEC 60079-15:2005
Markings: Ex nA nL IIC T5(-40 °C ≤ Ta ≤ +65 °C)
Special Conditions of Safe Use:
NC
74
1.
Provision must be made, external to the apparatus, to ensure the rated voltage of
the apparatus supply is not exceeded by transient disturbances of more than
40%.
2.
The electrical circuit is connected directly to earth; this must be taken into
account when installing the apparatus.
ATEX Type n (without enclosure)
Certificate: IECEx BAS 09.0031U
Standards: IEC 60079-0:2004, IEC 60079-15:2005
Markings: Ex nA nL IIC T4 (-50°C ≤ Ta ≤ +85 °C), T5(-50 °C ≤ Ta ≤ +70 °C)
Product Certificates
Reference Manual
Appendix B: Product Certifications
00809-0100-4697, Rev GA
November 2014
Special Conditions of Safe Use:
B.5.1
1.
The component must be installed in a suitable component certified enclosure
that provides a degree of protection of at least IP54 and meets the relevant
material and environmental requirements of EN 60079-0:2006 and EN
60079-15:2005.
2.
Provision must be made, external to the apparatus, to ensure the rated voltage of
the apparatus supply is not exceeded by transient disturbances of more than
40%.
3.
The electrical circuit is connected directly to earth; this must be taken into
account when installing the apparatus.
Brazil
I2
INMETRO Intrinsic Safety
Certificate: NCC 12.1156X
Standards: ABNT NBR IEC 60079-0:2008 Versão corrigida 2011, ABNT NBR IEC
60079-11:2009, ABNT NBR IEC 60079-26:2008 Versão corrigida 2009, ABNT
NBR IEC 60079-27:2010
Markings: Ex ia IIC T4 (-50 °C ≤ Ta ≤ +60 °C)
Special Conditions of Safe Use:
1.
The equipment must be installed in an enclosure that provides a degree of
protection of at least IP20 and which is appropriate to the application specified in
ABNT NBR IEC60079-0.
2.
The equipment is not capable of withstanding the dielectric strength test of
500V according to item 6.3.12 of ABNT NBR IEC60079-1, this should be
considered in the installation, see installation manual.
Table B-6. INMETRO-2 Entity Parameters
IB
Product Certificates
FISCO (Input)
Sensor Field Terminal
(Output)
Ui = 30 Vdc
Uo = 12.5 Vdc
Ii = 300 mA
Io = 4.8 mA
Pi = 1.3 W
Po = 15 mW
Ci = 2.1 nF
Co = 1.2 mF
Li = 0
Lo = 1 H
INMETRO Intrinsic Safety
Certificate: NCC 12.1156X
Standards: ABNT NBR IEC 60079-0:2008 Versão corrigida 2011, ABNT NBR IEC
60079-11:2009, ABNT NBR IEC 60079-26:2008 Versão corrigida 2009, ABNT
NBR IEC 60079-27:2010
Markings: Ex ia IIC T4 (-50 °C ≤ Ta ≤ +60 °C)
75
Reference Manual
Appendix B: Product Certifications
00809-0100-4697, Rev GA
November 2014
Special Conditions of Safe Use:
1.
The equipment must be installed in an enclosure that provides a degree of
protection of at least IP20 and which is appropriate to the application specified in
ABNT NBR IEC60079-0.
2.
The equipment is not capable of withstanding the dielectric strength test of
500V according to item 6.3.12 of ABNT NBR IEC60079-1, this should be
considered in the installation, see installation manual.
Table B-7. INMETRO-B Entity Parameters
N2
FISCO (Input)
Sensor Field Terminal
(Output)
Ui =17.5 Vdc
Uo = 12.5 Vdc
Ii = 380 mA
Io = 4.8 mA
Pi = 5.32 W
Po = 15 mW
Ci = 2.1 nF
Co = 1.2 mF
Li = 0
Lo = 1 H
INMETRO Intrinsic Safety Zone 2
Certificate: NCC 12.1182X
Standards: ABNT NBR IEC 60079-0:2008 Versao corrigida 2011, ABNT NBR IEC
60079-11:2009
Markings: Ex ic IIC T5 (-40°C ≤ Ta ≤ +65 °C) Gc
Special Conditions of Safe Use:
B.5.2
1.
The equipment must be mounted within an enclosure that meets at least the
degree of protection IP54, with material and manufacture covered by a quality
certificate. If the enclosure is non-metallic, the enclosure must have a surface
resistance less than 1 GÙ. If the enclosure is made of zirconium alloy, the
enclosure must be protected against impact and friction when installed.
2.
Provision must be made, external to the equipment, to ensure that the supply
voltage (42.2Vdc) is not exceeded by transient disturbances of more than 40%.
3.
The maximum ambient temperature will be restricted to lowest temperature
rating of the equipment, cables, cables glands or plugs.
4.
The electric circuit is connected directly to ground, this should be taken into
account when installing the equipment.
China
I3
NEPSI Intrinsic Safety
Certificate: GYJ111365X
Standards: GB3836.1-2000, GB3836.4-2000
Markings: Ex ia IIC T4
Special Conditions for Safe Use (X):
1.
76
Only when temperature transmitter is installed in IP 20(GB4208-2008) housing,
can it be used in a hazardous location. Metallic housing should observe the
requirements of GB3836.1-2000 Clause 8. Non-metallic housing should observe
the requirements of GB3836.1-2000 Clause 7.3.
Product Certificates
Reference Manual
Appendix B: Product Certifications
00809-0100-4697, Rev GA
November 2014
2.
This apparatus is not capable of withstanding the 500 V rms insulation test
required by Clause 6.4.12 of GB3836.4-2000.
3.
The ambient temperature range of the equipments is T4 (– 50 °C ≤ Ta ≤ + 60 °C).
4.
Parameters:
Terminals of power/loop (1-2):
Maximum
Maximum
Maximum
Output output voltage: output current: output power:
Uo (V)
Io (mA)
Po (mW)
Maximum external
parameters
Co (mF)
Lo (H)
F
30
300
1.3
2.1
0
F
(FISCO)
17.5
380
5.32
2.1
0
Note
Non-FISCO parameters listed above must be derived from a linear supply with a resistance
limited output.
Terminals of sensor:
Output
F
Terminals
1-8
Maximum
output
voltage:
Uo (V)
12.5
Maximum
output
current:
Io (mA)
4.8
Maximum
output
power:
Po (mW)
15
Maximum
external
parameters
Co (mF)
Lo (H)
1.2
1
5.
The product complies to the requirements for FISCO field devices specified in
IEC60079-27: 2008. For the connection of an intrinsically safe circuit in
accordance FISCO model, FISCO parameters of this product are as above.
6.
The product should be used with Ex-certified associated apparatus to establish
explosion protection system that can be used in explosive gas atmospheres.
Wiring and terminals should comply with the instruction manual of the product
and associated apparatus.
7.
The cables between this product and associated apparatus should be shielded
cables (the cables must have insulated shield). The shielded cable has to be
grounded reliably in non-hazardous area.
8.
End users are not permitted to change any component’s insides, but to settle the
problem, in conjunction with manufacturer to avoid damage to the product.
9.
During installation, use and maintenance of this product, observe following
standards:
GB3836.13-1997 “Electrical apparatus for explosive gas atmospheres Part 13:
Repair and overhaul for apparatus used in explosive gas atmospheres.”
GB3836.15-2000 “Electrical apparatus for explosive gas atmospheres Part 15:
Electrical installations in hazardous area (other than mines).”
Product Certificates
77
Reference Manual
Appendix B: Product Certifications
00809-0100-4697, Rev GA
November 2014
GB3836.16-2006 “Electrical apparatus for explosive gas atmospheres Part 16:
Inspection and maintenance of electrical installation (other than mines).”
GB50257-1996 “Code for construction and acceptance of electric device for
explosion atmospheres and fire hazard electrical equipment installation”
N3
NEPSI Type n
Certificate: GYJ12.1035U
Standards: GB3836.1-2010, GB3836.8-2003
Markings: Ex nA nL IIC T4/T5 Gc
Special Conditions for Safe Use (X):
1.
This component is not capable of withstanding the 500V electrical strength test
defined in Clause 8.1 of GB3836.8-2003. The must be taken into account during
installation.
2.
This component must be housed in a suitable component certified enclosure
that provides a degree of protection of at least IP54 and meets the relevant
material and environmental requirements of GB3836.1-2010 and
GB3836.8-2003.
3.
Provision must be made, external to the component, to ensure the rated voltage
of the component supply is not exceeded by transient disturbances of more the
40%.
4.
The ambient temperature range is:
T code
Ambient Temperature
T4
-50 °C ≤ Ta ≤ +85 °C
T5
-50 °C ≤ Ta ≤ +70 °C
5.
Maximum input voltage: 42.4V.
6.
End users are not permitted to change any components inside, but to settle the
problem in conjunction with manufacturer to avoid damage to the product.
7.
During installation, use and maintenance of this product, observe following
standards:
GB3836.13-1997 “Electrical apparatus for explosive gas atmospheres Part 13:
Repair and overhaul for apparatus used in explosive gas atmospheres”
GB3836.15-2000 “Electrical apparatus for explosive gas atmospheres Part 15:
Electrical installations in hazardous area (other than mines)”
GB3836.16-2006 “Electrical apparatus for explosive gas atmospheres Part 16:
Inspection and maintenance of electrical installation (other than mines)”
GB50257-1996 “Code for construction and acceptance of electric device for
explosion atmospheres and fire hazard electrical equipment installation
engineering”
B.5.3
78
Japan
I4
TIIS FISCO Intrinsic Safety (ia)
Certificate: TC19713
Markings: IIC T4
H4
TIIS FISCO Intrinsic Safety (ib)
Certificate: TC20737
Markings: IIC T4
Product Certificates
Reference Manual
Appendix B: Product Certifications
00809-0100-4697, Rev GA
B.6
Combinations
KG
B.7
November 2014
Combination of I1/IA, I5/IE, I6/IF, and I7/IG
Conduit Plugs and Adapters
ATEX Flameproof and Increased Safety
Certificate: FM13ATEX0076X
Standards: EN 60079-0:2012, EN 60079-1:2007, IEC 60079-7:2007
Marking
2 G Ex de IIC Gb
Special Conditions for Safe Use (X):
1.
When the thread adapter or blanking plug is used with an enclosure in type of
protection increased safety “e” the entry thread shall be suitably sealed in order
to maintain the ingress protection rating (IP) of the enclosure.
2.
The blanking plug shall not be used with an adapter.
3.
Blanking Plug and Threaded Adapter shall be either NPT or Metric thread forms.
G½ and PG 13.5 thread forms are only acceptable for existing (legacy)
equipment installations.
IECEx Flameproof and Increased Safety
Certificate: IECEx FMG 13.0032X
Standards: IEC 60079-0:2011, IEC 60079-1:2007, IEC 60079-7:2006-2007
Markings: Ex de IIC Gb
Special Conditions for Safe Use (X):
1.
When the thread adapter or blanking plug is used with an enclosure in type of
protection increased safety “e” the entry thread shall be suitably sealed in order
to maintain the ingress protection rating (IP) of the enclosure.
2.
The blanking plug shall not be used with an adapter.
3.
Blanking Plug and Threaded Adapter shall be either NPT or Metric thread forms.
G½ and PG 13.5 thread forms are only acceptable for existing (legacy)
equipment installations.
Conduit plug thread sizes:
Product Certificates
Thread
Identification Mark
M20 x 1.5
M20
½ - 14 NPT
½ NPT
G½
G½
79
Reference Manual
Appendix B: Product Certifications
00809-0100-4697, Rev GA
November 2014
Thread adapter thread sizes:
B.8
Male Thread
Identification Mark
M20 x 1.5 – 6H
M20
½ - 14 NPT
½ - 14 NPT
¾ - 14 NPT
¾ - 14 NPT
Female Thread
Identification Mark
M20 x 1.5 – 6H
M20
½ - 14 NPT
½ - 14 NPT
PG 13.5
PG 13.5
Additional Certifications
SBS
American Bureau of Shipping (ABS) Type Approval
Certificate: 011-HS771994C-1-PDA
ABS Rules: 2013 Steel Vessels Rules 1-1-4/7.7, 1-1-Appendix 3, 4-8-3/1.7, 4-8-3/13.1
SBV Bureau Veritas (BV) Type Approval
Certificate: 26325/A1 BV
Requirements: Bureau Veritas Rules for the Classification of Steel Ships
Application: Class notations: AUT-UMS, AUT-CCS, AUT-PORT and AUT-IMS
SDN Det Norske Veritas (DNV) Type Approval
Certificate: A-13246
Intended Use: Det Norske Veritas’ Rules for Classification of Ships, High Speed & Light
Craft and Det Norske Veritas’ Offshore Standards
Applications:
Location Classes
I3
80
Temperature
D
Humidity
B
Vibration
A
EMC
B
Enclosure
B/IP66: Al
C/IP66: SST
Lloyds Register (LR) Type Approval
Certificate: 11/60002 (E2)
Application: Environmental categories ENV1, ENV2, ENV2 and ENV5
Product Certificates
Reference Manual
Appendix B: Product Certifications
00809-0100-4697, Rev GA
B.9
November 2014
Intrinsically Safe and Non-Incendive installations
Approval
Safe area
Zone 2
(category 3)
Division 2
Zone 1
(category 2)
Zone 0
(category 1)
Division 1
GAS INSTALLATIONS
I5, I6, I1,
I7, IE, IA
Approved I.S. or
FISCO barrier
848T without
enclosure
N1, N7
Non-approved
power supply
848T with enclosure
N5
Non-approved
power supply
848T with enclosure
NC
Approved
non-incendive
power supply
or barrier
848T without enclosure
DUST INSTALLATIONS
N5, ND
Non-approved
power supply
848T with enclosure
Standard cable
Division 2 wiring
Product Certificates
81
Reference Manual
Appendix B: Product Certifications
00809-0100-4697, Rev GA
November 2014
B.10
Installation drawings
The installation guidelines presented by the drawings must be followed in order to maintain
certified ratings for installed transmitters.
Rosemount Drawing 00848-4404, 3 Sheets
Factory Mutual Intrinsic Safety/ FISCO Installation Drawing
Rosemount Drawing 00848-4405, 2 Sheets
Canadian Standards Association Intrinsic Safety/FISCO Installation Drawing
82
Product Certificates
Reference Manual
00809-0100-4697, Rev GA
Appendix B: Product Certifications
November 2014
Figure B-1. FM Intrinsic Safety/ FISCO
Product Certificates
83
Appendix B: Product Certifications
November 2014
84
Reference Manual
00809-0100-4697, Rev GA
Product Certificates
Reference Manual
00809-0100-4697, Rev GA
Product Certificates
Appendix B: Product Certifications
November 2014
85
Appendix B: Product Certifications
November 2014
Reference Manual
00809-0100-4697, Rev GA
El t
i M t
PRINTED COPIES ARE UNCONTROLLED R
tP
it
Figure B-2. CSA Intrinsic Safety/ FISCO
86
Product Certificates
Reference Manual
00809-0100-4697, Rev GA
Product Certificates
Appendix B: Product Certifications
November 2014
87
Appendix B: Product Certifications
November 2014
88
Reference Manual
00809-0100-4697, Rev GA
Product Certificates
Reference Manual
00809-0100-4697, Rev GA
Product Certificates
Appendix B: Product Certifications
November 2014
89
Appendix B: Product Certifications
November 2014
90
Reference Manual
00809-0100-4697, Rev GA
Product Certificates
Appendix C: FOUNDATION Fieldbus Technology
Reference Manual
November 2014
00809-0100-4697, Rev GA
FOUNDATION™ fieldbus
Technology
Appendix C
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Function blocks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Device descriptions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Block operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Network communication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
C.1
page 89
page 89
page 90
page 91
page 91
Overview
FOUNDATION fieldbus is an all-digital, serial, two-way, multi-drop communication protocol that
interconnects devices such as transmitters, sensors, actuators, and valve controllers. Fieldbus is
a Local Area Network (LAN) for instruments that are used in both process and manufacturing
automation, having the built-in capability to distribute the control applications across the
network. The fieldbus environment is the base level group of digital networks and the hierarchy
of plant networks.
The FOUNDATION fieldbus retains the desirable features of the 4–20 mA analog system, including
standardized physical interface to the wire, bus-powered devices on a single pair of wires, and
intrinsic safety options. It also enables the following capabilities:

Increased capabilities due to full digital communication.

Reduced wiring and wire terminations due to multiple devices on one pair of wires.

Increased supplier selection due to interoperability

Reduced loading on control room equipment due to the distribution of some control
and input/output functions to field devices.
FOUNDATION fieldbus devices work together to provide I/O and control for automated processes
and operations. The Fieldbus Foundation provides a framework for describing these systems as
a collection of physical devices interconnected by a fieldbus network. One of the ways that the
physical devices are used is to perform their portion of the total system operation by
implementing one or more function blocks.
C.2
Function blocks
Function blocks perform process control functions, such as analog input (AI) and analog output
(AO) functions as well as proportional-integral-derivative (PID) functions. The standard function
blocks provide a common structure for defining function block inputs, outputs, control
parameters, events, alarms, and modes, and combining them into a process that can be
implemented within a single device or over the fieldbus network. This simplifies the
identification of characteristics that are common to function blocks.
The Fieldbus Foundation has established the function blocks by defining a small set of
parameters used in all function blocks called universal parameters. The FOUNDATION has also
defined a standard set of function block classes, such as input, output, control, and calculation
blocks. Each of these classes has a small set of parameters established for it. They have also
published definitions for transducer blocks commonly used with standard function blocks.
Examples include temperature, pressure, level, and flow transducer blocks.
FOUNDATION Fieldbus Technology
89
Reference Manual
Appendix C: FOUNDATION Fieldbus Technology
00809-0100-4697, Rev GA
November 2014
The Fieldbus Foundation specifications and definitions allow vendors to add their own
parameters by importing and subclassing specified classes. This approach permits extending
function block definitions as new requirements are discovered and as technology advances.
Figure C-1 illustrates the internal structure of a function block. When execution begins, input
parameter values from other blocks are snapped-in by the block. The input snap process ensures
that these values do not change during the block execution. New values received for these
parameters do not affect the snapped values and will not be used by the function block during
the current execution.
Figure C-1. Function Block Internal Structure
Input Events
Input
Parameter
Linkages
Input
Snap
Execution
Control
Output Events
Processing
Algorithm
Output
Snap
Status
Output
Parameter
Linkages
Status
Once the inputs are snapped, the algorithm operates on them, generating outputs as it
progresses. Algorithm executions are controlled through the setting of contained parameters.
Contained parameters are internal to function blocks and do not appear as normal input and
output parameters. However, they may be accessed and modified remotely, as specified by the
function block.
Input events may affect the operation of the algorithm. An execution control function regulates
the receipt of input events and the generation of output events during execution of the
algorithm. Upon completion of the algorithm, the data internal to the block is saved for use in
the next execution, and the output data is snapped, releasing it for use by other function blocks.
A block is a tagged logical processing unit. The tag is the name of the block. System
management services locate a block by its tag. Thus the service personnel need only know the
tag of the block to access or change the appropriate block parameters.
Function blocks are also capable of performing short-term data collection and storage for
reviewing their behavior.
C.3
Device descriptions
Device descriptions (DD) are specified tool definitions that are associated with the Resource and
Transducer Blocks. Device descriptions provide the definition and description of the function
blocks and their parameters.
To promote consistency of definition and understanding, descriptive information, such as data
type and length, is maintained in the device description. Device Descriptions are written using
an open language called the Device Description Language (DDL). Parameter transfers between
function blocks can be easily verified because all parameters are described using the same
language. Once written, the device description can be stored on an external medium, such as a
CD-ROM or diskette. Users can then read the device description from the external medium. The
use of an open language in the device description permits interoperability of function blocks
90
FOUNDATION Fieldbus Technology
Reference Manual
Appendix C: FOUNDATION Fieldbus Technology
00809-0100-4697, Rev GA
November 2014
within devices from various vendors. Additionally, human interface devices, such as operator
consoles and computers, do not have to be programmed specifically for each type of device on
the bus. Instead their displays and interactions with devices are driven from the device
descriptions.
Device descriptions may also include a set of processing routines called methods. Methods
provide a procedure for accessing and manipulating parameters within a device.
C.4
Block operation
In addition to function blocks, fieldbus devices contain two other block types to support the
function blocks. These are the resource block and the transducer block.
C.4.1
Instrument- specific function blocks
Resource blocks
Resource blocks contain the hardware–specific characteristics associated with a device; they
have no input or output parameters. The algorithm within a resource block monitors and
controls the general operation of the physical device hardware. The execution of this algorithm
is dependent on the characteristics of the physical device, as defined by the manufacturer. As a
result, the algorithm may cause the generation of events. There is only one resource block
defined for a device. For example, when the mode of a resource block is “Out of Service (OOS),”
it impacts all of the other blocks.
Transducer blocks
Transducer blocks connect function blocks to local input/output functions. They read sensor
hardware and write to effector (actuator) hardware. This permits the transducer block to
execute as frequently as necessary to obtain good data from sensors and ensure proper writes
to the actuator without burdening the function blocks that use the data. The transducer block
also isolates the function block from the vendor–specific characteristics of the physical I/O.
C.4.2
Alerts
When an alert occurs, execution control sends an event notification and waits a specified period
of time for an acknowledgment to be received. This occurs even if the condition that caused the
alert no longer exists. If the acknowledgment is not received within the pre-specified time-out
period, the event notification is retransmitted, assuring that alert messages are not lost.
Two types of alerts are defined for the block: events and alarms. Events are used to report a
status change when a block leaves a particular state, such as when a parameter crosses a
threshold. Alarms not only report a status change when a block leaves a particular state, but also
report when it returns back to that state.
C.5
Network communication
Figure C-2 illustrates a simple fieldbus network consisting of a single segment (link).
FOUNDATION Fieldbus Technology
91
Reference Manual
Appendix C: FOUNDATION Fieldbus Technology
00809-0100-4697, Rev GA
November 2014
Figure C-2. Simple, Single-Link Fieldbus Network
B
A
C
D
A. Link Active Scheduler (LAS)
B. Fieldbus Link
C. Link Master
D. Basic Device and/or Link Master Devices
C.5.1
LAS
All links have one LAS that operates as the bus arbiter for the link. The LAS does the following:

recognizes and adds new devices to the link

removes non-responsive devices from the link

distributes Data Link Time (DL) and Link Scheduling Time (LS) on the link
–
DL is a network-wide time periodically distributed by the LAS to synchronize all
device clocks on the bus.
–
LS time is a link-specific time represented as an offset from DL. It is used to
indicate when the LAS on each link begins and repeats its schedule. It is used by
system management to synchronize function block execution with the data
transfers scheduled by the LAS

polls devices for process loop data at scheduled transmission times

distributes a priority-driven token to devices between scheduled transmissions
Any device on the link may become the LAS. The devices that are capable of becoming the LAS
are called Link Master devices (LM). All other devices are referred to as basic devices. When a
segment first starts up, or upon failure of the existing LAS, the link master devices on the
segment bid to become the LAS. The link master that wins the bid begins operating as the LAS
immediately upon completion of the bidding process. Link masters that do not become the LAS
act as basic devices. However, the link masters can act as LAS backups by monitoring the link for
failure of the LAS and then bidding to become the LAS when a LAS failure is detected.
Only one device can communicate at a time. Permission to communicate on the bus is
controlled by a centralized token passed between devices by the LAS. Only the device with the
token can communicate. The LAS maintains a list of all devices that need access to the bus. This
list is called the “Live List.”
Two types of tokens are used by the LAS. A time-critical token, Compel Data (CD), is sent by the
LAS according to a schedule. A non-time critical token, pass token (PT), is sent by the LAS to each
device in ascending numerical order according to address.
There may be many LM devices on a segment but only the LAS is actively controlling
communication traffic. The remaining LM devices on the segment are in a stand-by state, ready
to take over if the primary LAS fails. This is achieved by constantly monitoring the
communication traffic on the bus and determining if activity is not present. Since there can be
multiple LM devices on the segment when the primary LAS fails, the device with the lowest node
address will become the primary LAS and take control of the bus. Using this strategy, multiple
LAS failures can be handled with no loss of the LAS capability of the communications bus.
92
FOUNDATION Fieldbus Technology
Appendix C: FOUNDATION Fieldbus Technology
Reference Manual
November 2014
00809-0100-4697, Rev GA
LAS parameters
There are many bus communication parameters but only a few are used. For standard RS-232
communications, the configuration parameters are baud rate, start / stop bits, and parity. The
key parameters for H1 FOUNDATION fieldbus are as follows.

Slot Time (ST) – Used during the bus master election process. It is the maximum
amount of time permitted for device A to send a message to device B. Slot time is a
parameter which defines a worst case delay which includes internal delay in the sending
device and the receiving device. Increasing the value of ST slows down bus traffic
because a LAS device must wait longer prior to determining that the LM is down.

Minimum Inter-PDU Delay (MID) – The minimum gap between two messages on the
fieldbus segment or it is the amount of time between the last byte of one message and
the first byte of the next message. The units of the MID are octets. An octet is 256 μs,
hence the units for MID are approximately 1/4 ms. This would mean an MID of 16 would
specify approximately a minimum of 4 ms between messages on the fieldbus.
Increasing the value of MID slows down bus traffic because a larger “gap” between
messages occurs.

Maximum Response (MRD) – Defines the maximum amount of time permitted to
respond to an immediate response request, e.g. CD, PT. When a published value is
requested using the CD command, the MRD defines how long before the device
publishes the data. Increasing this parameter will slow down the bus traffic by
decreasing how fast CDs can be put onto the network. The MRD is measured in units of
ST.

Time Synchronization Class (TSC) – A variable that defines how long the device can
estimate its time before drifting out of specific limits. The LM will periodically send out
time update messages to synchronize devices on the segment. Decreasing the
parameter number increases the number of times that time distribution messages
must be published, increasing bus traffic and overhead for the LM device. See Figure
C-3.
Figure C-3. LAS Parameter Diagram
FB 1
C
D
MID
Data
MID
FB 2
MID x ST
Backup LAS
An LM device is one that has the ability to control the communications on the bus. The LAS is the
LM capable device that is currently in control of the bus. While there can be many LM devices
acting as backups, there can only be one LAS. The LAS is typically a host system but for
stand-alone applications, a device may be providing the role of primary LAS.
C.5.2
Addressing
To setup, configure, and communicate with other devices on a segment, a device must be
assigned a permanent address. Unless requested otherwise, it is assigned a temporary address
when shipped from the factory.
FOUNDATION Fieldbus Technology
93
Appendix C: FOUNDATION Fieldbus Technology
Reference Manual
00809-0100-4697, Rev GA
November 2014
FOUNDATION fieldbus uses addresses between 0 and 255. Addresses 0 through 15 are reserved for
group addressing and for use by the data link layer.
If there are two or more devices on a segment with the same address, the first device to start up
will use the assigned address. Each of the other devices will be given one of the four temporary
addresses. If a temporary address is not available, the device will be unavailable until a
temporary address is available.
Use the host system documentation to commission a device and assign a permanent address.
C.5.3
Scheduled transfers
Information is transferred between devices over the FOUNDATION fieldbus using three different
types of reporting.
Publisher/subscriber
This type of reporting is used to transfer critical process loop data, such as the process variable.
The data producers (publishers) post the data in a buffer that is transmitted to the subscriber,
when the publisher receives the Compel Data (CD). The buffer contains only one copy of the
data. New data completely overwrites previous data. Updates to published data are transferred
simultaneously to all subscribers in a single broadcast. Transfers of this type can be scheduled
on a precisely periodic basis.
Report distribution
This type of reporting is used to broadcast and multi-cast event and trend reports. The
destination address may be predefined so that all reports are sent to the same address, or it may
be provided separately with each report. Transfers of this type are queued. They are delivered to
the receivers in the order transmitted, although there may be gaps due to corrupted transfers.
These transfers are unscheduled and occur between scheduled transfers at a given priority.
Client/server
This type of reporting is used for request/response exchanges between pairs of devices. Like
Report Distribution reporting, the transfers are queued, unscheduled, and prioritized. Queued
means the messages are sent and received in the order submitted for transmission, according to
their priority, without overwriting previous messages. However, unlike Report Distribution,
these transfers are flow controlled and employ a retransmission procedure to recover from
corrupted transfers.
Figure C-4 diagrams the method of scheduled data transfer. Scheduled data transfers are
typically used for the regular cyclic transfer of process loop data between devices on the
fieldbus. Scheduled transfers use publisher/ subscriber type of reporting for data transfer. The
LAS maintains a list of transmit times for all publishers in all devices that need to be cyclically
transmitted. When it is time for a device to publish data, the LAS issues a CD message to the
device. Upon receipt of the CD, the device broadcasts or “publishes” the data to all devices on
the fieldbus. Any device that is configured to receive the data is called a “subscriber.”
94
FOUNDATION Fieldbus Technology
Appendix C: FOUNDATION Fieldbus Technology
Reference Manual
November 2014
00809-0100-4697, Rev GA
Figure C-4. Scheduled Data Transfer
LAS
Schedule
X, Y, Z
DT(A)
CD(X,A)
A
B
C
A
D A
P
S
P
S
P
Device X
Device Y
S
Device Z
LAS = Link Active Scheduler
P = Publisher
S = Subscriber
CD = Compel Data
DT = Data Transfer Packet
C.5.4
Unscheduled transfers
Figure C-5 diagrams an unscheduled transfer. Unscheduled transfers are used for things like
user-initiated changes, including set point changes, mode changes, tuning changes, and
upload/download. Unscheduled transfers use either report distribution or client/server type of
reporting for transferring data.
All of the devices on the FOUNDATION fieldbus are given a chance to send unscheduled messages
between transmissions of scheduled data. The LAS grants permission to a device to use the
fieldbus by issuing a pass token (PT) message to the device. When the device receives the PT, it is
allowed to send messages until it has finished or until the “maximum token hold time” has
expired, whichever is the shorter time. The message may be sent to a single destination or to
multiple destinations.
Figure C-5. Unscheduled Data Transfer
LAS
PT(Z)
Schedule
X, Y, Z
DT(M)
A
B
C
A
D
M
P
S
Device X
P = Publisher
FOUNDATION Fieldbus Technology
A
M
S = Subscriber
P
S
Device Y
P
S
Device Z
PT = Pass Token
M = Message
95
Reference Manual
Appendix C: FOUNDATION Fieldbus Technology
00809-0100-4697, Rev GA
November 2014
C.5.5
Function block scheduling
Figure C-6 shows an example of a link schedule. A single iteration of the link-wide schedule is
called the macrocycle. When the system is configured and the function blocks are linked, a
master link-wide schedule is created for the LAS. Each device maintains its portion of the
link-wide schedule, known as the Function Block Schedule. The Function Block Schedule
indicates when the function blocks for the device are to be executed. The scheduled execution
time for each function block is represented as an offset from the beginning of the macrocycle
start time.
Figure C-6. Example Link Schedule Showing Scheduled and Unscheduled Communication
Macrocycle Start Time
Offset from macrocycle start
time = 0 for AI Execution
Sequence Repeats
Device 1
AI
Scheduled
Communication
AI
Offset from macrocycle start time =
20 for AI Communication
Unscheduled
Communication
Offset from macrocycle start
time = 30 for PID Execution
Device 2
PID
AO
PID
AO
Offset from macrocycle start time =
50 for AO Execution
Macrocycle
To support synchronization of schedules, periodically Link Scheduling (LS) time is distributed.
The beginning of the macrocycle represents a common starting time for all Function Block
schedules on a link and for the LAS link-wide schedule. This permits function block executions
and their corresponding data transfers to be synchronized in time.
96
FOUNDATION Fieldbus Technology
Appendix D: Function Blocks
Reference Manual
November 2014
00809-0100-4697, Rev GA
Appendix D
Function Blocks
Analog Input (AI) function block . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 97
Multiple analog input (MAI) function block . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 106
Input selector function block . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 113
D.1
Analog Input (AI) function block
OUT_D
AI
OUT
Out = The block output value and status
Out_D = Discrete output that signals a selected alarm condition
The Analog Input (AI) function block processes field device measurements and makes them
available to other function blocks. The output value from the AI block is in engineering units and
contains a status indicating the quality of the measurement. The measuring device may have
several measurements or derived values available in different channels. Use the channel number
to define the variable that the AI block processes.
The AI block supports alarming, signal scaling, signal filtering, signal status calculation, mode
control, and simulation. In Automatic mode, the block’s output parameter (OUT) reflects the
process variable (PV) value and status. In Manual mode, OUT may be set manually. The Manual
mode is reflected on the output status. A discrete output (OUT_D) is provided to indicate
whether a selected alarm condition is active. Alarm detection is based on the OUT value and
user specified alarm limits. The block execution time is 30 ms.
Table D-1. Analog Input Function Block Parameters
Number
Function Blocks
Parameter
Units
Description
01
ST_REV
None
The revision level of the static data associated with the
function block. The revision value will be incremented
each time a static parameter value in the block is
changed.
02
TAG_DESC
None
The user description of the intended application of the
block.
03
STRATEGY
None
The strategy field can be used to identify a grouping
of blocks. This data is not checked or processed by the
block.
04
ALERT_KEY
None
The identification number of the plant unit. This
information may be used in the host for sorting
alarms, etc.
97
Reference Manual
Appendix D: Function Blocks
00809-0100-4697, Rev GA
November 2014
Table D-1. Analog Input Function Block Parameters
Number
98
Parameter
Units
Description
05
MODE_BLK
None
The actual, target, permitted, and normal modes of
the block.
Actual: The mode the “block is currently in”
Target: The mode to “go to”
Permitted: Allowed modes that target may take on
Normal: Most common mode for target
06
BLOCK_ERR
None
This parameter reflects the error status associated
with the hardware or software components
associated with a block. It is a bit string, so that
multiple errors may be shown.
07
PV
08
OUT
09
SIMULATE
EU of
XD_SCALE
EU of
OUT_SCALE
or XD_SCALE
if in direct
L_TYPE
The process variable used in block execution.
The block output value and status.
None
A group of data that contains the current transducer
value and status, the simulated transducer value and
status, and the enable/disable bit.
10
XD_SCALE
None
The high and low scale values, engineering units code,
and number of digits to the right of the decimal point
associated with the channel input value. The
XD_SCALE units code must match the units code of
the measurement channel in the transducer block. If
the units do not match, the block will not transition to
MAN or AUTO.
11
OUT_SCALE
None
The high and low scale values, engineering units code,
and number of digits to the right of the decimal point
associated with OUT when L_TYPE is not direct.
12
GRANT_DENY
None
Options for controlling access of host computers and
local control panels to operating, tuning, and alarm
parameters of the block. Not used by device.
13
IO_OPTS
None
Allows the selection of input/output options used to
alter the PV. Low cutoff enabled is the only selectable
option.
14
STATUS_OPTS
None
Allows the user to select options for status handling
and processing. The options supported in the AI block
are the following:
Propagate fault forward
Uncertain if limited
Bad if limited
Uncertain if Manual mode.
15
CHANNEL
None
The CHANNEL value is used to select the
measurement value. Configure the CHANNEL
parameter before configuring the XD_SCALE
parameter. Refer to Table 3-5 on page 11.
None
Linearization type. Determines whether the field
value is used directly (Direct), is converted linearly
(Indirect), or is converted with the square root
(Indirect Square Root).
%
If percentage value of transducer input fails below
this, PV = 0.
16
L_TYPE
17
LOW_CUT
Function Blocks
Appendix D: Function Blocks
Reference Manual
November 2014
00809-0100-4697, Rev GA
Table D-1. Analog Input Function Block Parameters
Number
Units
Description
18
PV_FTIME
Seconds
The time constant of the first-order PV filter. It is the
time required for a 63% change in the PV or OUT
value.
19
FIELD_VAL
Percent
The value and status from the transducer block or
from the simulated input when simulation is enabled.
20
UPDATE_EVT
21
Function Blocks
Parameter
BLOCK_ALM
None
This alert is generated by any change to the static
data.
None
The block alarm is used for all configuration,
hardware, connection failure or system problems in
the block. The cause of the alert is entered in the
subcode field. The first alert to become active will set
the Active status in the Status parameter. As soon as
the Unreported status is cleared by the alert reporting
task, another block alert may be reported without
clearing the Active status, if the subcode has changed.
22
ALARM_SUM
None
The summary alarm is used for all process alarms in
the block. The cause of the alert is entered in the
subcode field. The first alert to become active will set
the Active status in the Status parameter. As soon as
the Unreported status is cleared by the alert reporting
task, another block alert may be reported without
clearing the Active status, if the subcode has changed.
23
ACK_OPTION
None
Used to set auto acknowledgment of alarms.
24
ALARM_HYS
Percent
25
HI_HI_PRI
None
26
HI_HI_LIM
EU of
PV_SCALE
27
HI_PRI
None
28
HI_LIM
EU of
PV_SCALE
29
LO_PRI
None
30
LO_LIM
EU of
PV_SCALE
31
LO_LO_PRI
None
32
LO_LO_LIM
EU of
PV_SCALE
33
HI_HI_ALM
None
The HI HI alarm data, which includes a value of the
alarm, a timestamp of occurrence and the state of the
alarm.
34
HI_ALM
None
The HI alarm data, which includes a value of the
alarm, a timestamp of occurrence and the state of the
alarm.
35
LO_ALM
None
The LO alarm data, which includes a value of the
alarm, a timestamp of occurrence and the state of the
alarm.
36
LO_LO_ALM
None
The LO LO alarm data, which includes a value of the
alarm, a timestamp of occurrence and the state of the
alarm.
The amount the alarm value must return within the
alarm limit before the associated active alarm
condition clears.
The priority of the HI HI alarm.
The setting for the alarm limit used to detect the HI HI
alarm condition.
The priority of the HI alarm.
The setting for the alarm limit used to detect the HI
alarm condition.
The priority of the LO alarm.
The setting for the alarm limit used to detect the LO
alarm condition.
The priority of the LO LO alarm.
The setting for the alarm limit used to detect the LO
LO alarm condition.
99
Reference Manual
Appendix D: Function Blocks
00809-0100-4697, Rev GA
November 2014
Table D-1. Analog Input Function Block Parameters
Number
D.1.1
Parameter
Units
Description
37
OUT_D
None
Discrete output to indicate a selected alarm
condition.
38
ALM_SEL
None
Used to select the process alarm conditions that will
cause the OUT_D parameter to be set.
39
STDDEV
% of OUT
Range
Standard deviation of the measurement for 100
macrocycles.
40
CAP_STDDEV
% of OUT
Range
Capability standard deviation, the best deviation that
can be achieved.
Functionality
Simulation
To support testing, either change the mode of the block to manual and adjust the output value,
or enable simulation through the configuration tool and manually enter a value for the
measurement value and its status. In simulation, the ENABLE jumper must be set on the field
device.
Note
All FOUNDATION™ fieldbus instruments have a simulation jumper. As a safety measure, the jumper
has to be reset every time there is a power interruption. This measure is to prevent devices that
went through simulation in the staging process from being installed with simulation enabled.
With simulation enabled, the actual measurement value has no impact on the OUT value or the
status.
Figure D-1. Analog Input Function Block Timing Diagram
OUT (mode in man)
OUT (mode in auto)
PV
63% of Change
FIELD_VAL
Time (seconds)
PV_FTIME
100
Function Blocks
Appendix D: Function Blocks
Reference Manual
November 2014
00809-0100-4697, Rev GA
Figure D-2. Analog Input Function Block Schematic
Analog
Measurement
ALM_SEL
Access
Analog
Meas.
HI_HI_LIM
HI_LIM
LO_LO_LIM
LO_LIM
ALARM_HYS
CHANNEL
Alarm
Detection
OUT_D
LOW_CUT
OUT
Cutoff
Convert
SIMULATE
L_TYPE
FIELD_VAL
Filter
PV_FTIME
IO_OPTS
PV
Status
Calc.
MODE
STATUS_OPTS
F
OUT_SCALE
XD_SCALE
OUT = Block output value and status
OUT_D = Discrete output that signals a selected alarm condition
Filtering
The filtering feature changes the response time of the device to smooth variations in output
readings caused by rapid changes in input. Adjust the filter time constant (in seconds) using the
PV_FTIME parameter. Set the filter time constant to zero to disable the filter feature.
Signal conversion
Set the signal conversion type with the Linearization Type (L_TYPE) parameter. View the
converted signal (in percent of XD_SCALE) through the FIELD_VAL parameter.
100 ¥ ( Channel Value – EU*@0% )
FIELD_VAL = ----------------------------------------------------------------------------------------( EU*@100% – EU*@0% )
* XD_SCALE values
Choose from direct, indirect, or indirect square root signal conversion with the L_TYPE
parameter.
Direct
Direct signal conversion allows the signal to pass through the accessed channel input value (or the
simulated value when simulation is enabled).
PV = Channel Value
Indirect
Indirect signal conversion converts the signal linearly to the accessed channel input value (or the
simulated value when simulation is enabled) from its specified range (XD_SCALE) to the range
and units of the PV and OUT parameters (OUT_SCALE).
FIELD_VAL
PV =  ---------------------------- ¥ ( EU**@100% – EU**@0% ) + EU**@0%
100
** OUT_SCALE values
Function Blocks
101
Reference Manual
Appendix D: Function Blocks
00809-0100-4697, Rev GA
November 2014
Indirect square root
Indirect Square Root signal conversion takes the square root of the value computed with the
indirect signal conversion and scales it to the range and units of the PV and OUT parameters.
PV =
 FIELD_VAL
---------------------------- ¥ ( EU**@100% – EU**@0% ) + EU**@0%

100 
** OUT_SCALE values
When the converted input value is below the limit specified by the LOW_CUT parameter, and
the Low Cutoff I/O option (IO_OPTS) is enabled (True), a value of zero is used for the converted
value (PV). This option eliminates false readings when the differential pressure measurement is
close to zero and it may be useful with zero-based measurement devices such as flowmeters.
Note
Low Cutoff is the only I/O option supported by the AI block. Set the I/O option when the block is
OOS.
Block errors
Table lists conditions reported in the BLOCK_ERR parameter. Conditions in bold are inactive for
the AI block and are given here for reference.
Table D-2. BLOCK_ERR Conditions
Number
102
Name and description
0
Other
1
Block Configuration Error: the selected channel carries a measurement that is
incompatible with the engineering units selected in XD_SCALE, the L_TYPE parameter is
not configured, or CHANNEL = zero.
2
Link Configuration Error
3
Simulate Active: Simulation is enabled and the block is using a simulated value in its
execution.
4
Local Override
5
Device Fault State Set
6
Device Needs Maintenance Soon
7
Input Failure/Process Variable has Bad Status: The hardware is bad, or a bad status is
being simulated.
8
Output Failure: The output is bad based primarily upon a bad input.
9
Memory Failure
10
Lost Static Data
11
Lost NV Data
12
Readback Check Failed
13
Device Needs Maintenance Now
14
Power Up
15
Out of Service: The actual mode is out of service.
Function Blocks
Appendix D: Function Blocks
Reference Manual
November 2014
00809-0100-4697, Rev GA
Modes
The AI Function Block supports three modes of operation as defined by the MODE_BLK
parameter.
Manual (Man)
The value of the block output (OUT) may be set manually
Automatic (Auto)
OUT reflects the analog input measurement or the simulated value when simulation is
enabled.
Out of service (OOS)
The block is not processed. FIELD_VAL and PV are not updated and the OUT status is set to
Bad: Out of Service. The BLOCK_ERR parameter shows Out of Service. In this mode,
changes can be made to all configurable parameters.
Alarm detection
A block alarm will be generated whenever the BLOCK_ERR has an error bit set. The types of
block error for the AI block are defined above.
Process alarm detection is based on the OUT value. Configure the alarm limits of the following
standard alarms:

High (HI_LIM)

High high (HI_HI_LIM)

Low (LO_LIM)

Low low (LO_LO_LIM)
To avoid alarm chatter when the variable is oscillating around the alarm limit, an alarm
hysteresis in percent of the PV span can be set using the ALARM_HYS parameter. The priority of
each alarm is set in the following parameters:

HI_PRI

HI_HI_PRI

LO_PRI

LO_LO_PRI
Table D-3. Alarm Priority Levels
Number
Function Blocks
Description
0
The priority of an alarm condition changes to 0 after the condition that caused the alarm
is corrected.
1
An alarm condition with a priority of 1 is recognized by the system, but is not reported to
the operator.
2
An alarm condition with a priority of 2 is reported to the operator, but does not require
operator attention (such as diagnostics and system alerts).
3-7
Alarm conditions of priority 3 to 7 are advisory alarms of increasing priority.
8-15
Alarm conditions of priority 8 to 15 are critical alarms of increasing priority.
103
Appendix D: Function Blocks
Reference Manual
00809-0100-4697, Rev GA
November 2014
Status handling
Normally, the status of the PV reflects the status of the measurement value, the operating
condition of the I/O card, and any active alarm condition. In Auto mode, OUT reflects the value
and status quality of the PV. In Man mode, the OUT status constant limit is set to indicate that
the value is a constant and the OUT status is Good.
If the sensor limit exceeds the high or low range, PV status is set high or low and EU range status
is set to uncertain.
In the STATUS_OPTS parameter, select from the following options to control the status
handling.
BAD if limited
Sets the OUT status quality to Bad when the value is higher or lower than the sensor limits.
Uncertain if limited
Sets the OUT status quality to Uncertain when the value is higher or lower than the sensor
limits.
Uncertain if in manual mode
The status of the Output is set to Uncertain when the mode is set to Manual.
Notes
1. The instrument must be in OOS mode to set the status option.
2. The AI block only supports the BAD if Limited option, uncertain if limited, and uncertain if
manual.
Advanced features
The AI function block provided with Rosemount® fieldbus devices provides added capability
through the addition of the following parameters:
ALARM_TYPE
Allows one or more of the process alarm conditions detected by the AI function block to be
used in setting its OUT_D parameter.
OUT_D
Discrete output of the AI function block based on the detection of process alarm
condition(s). This parameter may be linked to other function blocks that require a discrete
input based on the detected alarm condition.
STD_DEV and CAP_STDDEV
Diagnostic parameters that can be used to determine the variability of the process.
Application information
The configuration of the AI function block and its associated output channels depends on the
specific application. A typical configuration for the AI block involves the following parameters:
CHANNEL
The device supports more than one measurement, so verify that the selected channel
contains the appropriate measurement or derived value. Refer to Table 3-8 on page 36 for a
listing of available channels on the 848T.
104
Function Blocks
Appendix D: Function Blocks
Reference Manual
November 2014
00809-0100-4697, Rev GA
L_TYPE
Select Direct when the measurement is in the desired engineering units for the block
output. Select Indirect when converting the measured variable into another, for example,
pressure into level or flow into energy.
SCALING
XD_SCALE provides the range and units of the measurement and OUT_SCALE provides the
range and engineering units of the output. OUT_SCALE is only used when in indirect or
indirect square root.
D.1.2
AI block troubleshooting
Symptom
Possible causes
Corrective action
Mode will not leave
OOS
Target mode not set
Set target mode to something other than OOS.
Configuration error
BLOCK_ERR will show the configuration error bit
set. The following are parameters that must be set
before the block is allowed out of OOS:
• CHANNEL must be set to a valid value and cannot be
left at initial value of 0.
• XD_SCALE.UNITS_INDEX must match the units in the
transducer block channel value. Setting the units in
the AI block automatically sets them in the
XD_BLOCK.
• L_TYPE must be set to Direct, Indirect, or Indirect
Square Root and cannot be left at initial value of 0.
Process and/or block
alarms will not work.
Value of output does
not make sense
Resource block
The actual mode of the Resource block is OOS. See
Resource Block Diagnostics for corrective action.
Schedule
Block is not scheduled and therefore cannot execute
to go to Target Mode. Typically, BLOCK_ERR will
show “Power-Up” for all blocks that are not
scheduled. Schedule the block to execute.
Features
FEATURES_SEL does not have Alerts enabled. Enable
the Alerts bit.
Notification
LIM_NOTIFY is not high enough. Set equal to
MAX_NOTIFY. Alarm not linked to host.
Status Options
STATUS_OPTS has Propagate Fault Forward bit set.
This should be cleared to cause an alarm to occur.
Linearization Type
L_TYPE must be set to Direct, Indirect, or Indirect
Square Root and cannot be left at initial value of 0.
Scaling
Scaling parameters are set incorrectly:
• XD_SCALE.EU0 and EU100 should match that of the
transducer block channel value.
• OUT_SCALE.EU0 and EU100 are not set properly.
• Both STB on each ASIC used must by in auto.
Cannot set HI_LIMIT,
HI_HI_LIMIT, LO_LIMIT,
or LO_LO_LIMIT Values
Function Blocks
Scaling
Limit values are outside the OUT_SCALE.EU0 and
OUT_SCALE.EU100 values. Change OUT_SCALE or
set values within range.
105
Reference Manual
Appendix D: Function Blocks
00809-0100-4697, Rev GA
November 2014
D.2
Multiple analog input (MAI) function block
OUT_1
OUT_2
MAI
OUT_3
OUT_4
OUT_5
OUT_6
OUT_7
OUT_8
Out1 = The block output value and status for the first channel.
The MAI function block has the ability to process up to eight field device measurements and
make them available to other function blocks. The output values from the MAI block are in
engineering units and contain a status indicating the quality of the measurement. The
measuring device may have several measurements or derived values available in different
channels. Use the channel numbers to define the variables that the MAI block processes.
The MAI block supports signal scaling, signal filtering, signal status calculation, mode control,
and simulation. In Automatic mode, the block’s output parameters (OUT_1 to OUT_8) reflects
the process variable (PV) values and status. In Manual mode, OUT may be set manually. The
Manual mode is reflected on the output status. Table D-4 lists the MAI block parameters and
their units of measure, descriptions, and index numbers. The block execution time is 30 ms.
Table D-4. Multiple Analog Input Function Block Parameters
Number
106
Parameter
Units
Description
1
ST_REV
None
The revision level of the static data associated with the
input selector block. The revision value will be
incremented each time a static parameter value in the
block is changed.
2
TAG_DESC
None
The user description of the intended application of the
block.
3
STRATEGY
None
The strategy field can be used to identify grouping of
blocks. This data is not checked or processed by the
block.
4
ALERT_KEY
None
The identification number of the plant unit. This
information may be used in the host for sorting alarms,
etc.
5
MODE_BLK
None
The actual, target, permitted, and normal modes of the
block.
Actual: The mode the “block is currently in”
Target: The mode to “go to”
Permitted: Allowed modes that target may take on
Normal: Most common mode for target
6
BLOCK_ERR
None
This parameter reflects the error status associated with
the hardware or software components associated with a
block. It is a bit string, so that multiple errors may be
shown.
Function Blocks
Appendix D: Function Blocks
Reference Manual
November 2014
00809-0100-4697, Rev GA
Table D-4. Multiple Analog Input Function Block Parameters
Number
7
8, 9, 10,
11, 12,
13, 14,
15
Parameter
Units
Description
CHANNEL
None
Allows for custom channel setting. Valid values include:
0: Unitialized
1: Channels 1 to 8 (index values 27 to 34 can only be set
to their corresponding channel number, i.e.
CHANNEL_X=X)
2: Custom settings (index values 27 to 34 can be
configured for any valid channel as defined by the DD)
OUT (1, 2, 3, 4,
5, 6, 7, 8)
EU of
OUT_SCALE
The block output value and status
16
UPDATE_EVT
None
This alert is generated by any change to the static data
17
BLOCK_ALM
None
The block alarm is used for all configuration, hardware
connection feature, or system problems in the block.
The cause of the alert is entered in the subcode field.
The first alert to become active will set the Active status
in the Status parameter. As soon as the Unreported
status is cleared by the alert reporting task, another
block may be reported without clearing the Active
status, if the subcode has changed.
18
SIMULATE
None
A group of data that contains the current sensor
transducer value and status, and the enable/disable bit.
19
XD_SCALE
None
The high and low scale values, engineering units code
and number of digits to the right of the decimal point
associated with the channel input value. The XD_SCALE
units code must match the units code of the
measurement channel in the transducer block. If the
units do not match, the block will not transition to MAN
or AUTO. It will automatically change units in the STB
block to the last one written. Multiple blocks reading the
same channel may conflict (only one unit type per
channel).
20
OUT_SCALE
None
The high and low scale values, engineering unit code
and number of digits to the right of the decimal point
associated with OUT.
21
GRANT_DENY
None
Options for controlling access of host computers and
local control panels for operating, tuning, and alarm
parameters of the block. Not used by device.
22
IO_OPTS
None
Allows the selection of input/output options used to
alter the PV. Low cutoff enabled is the only selectable
option.
23
STATUS_OPTS
None
Allows the user to select options for status handling and
processing. The options supported in the MAI block are
the following:
• Propagate fault forward
• Uncertain if limited
• Bad if limited
• Uncertain if manual mode
Function Blocks
24
L_TYPE
None
25
LOW_CUT
%
26
PV_FTIME
Seconds
Linearization type. Determines whether the field value is
uses directly (Direct), is converted linearly (Indirect), or
is converted with the square root (Indirect Square Root)
If percentage value of the sensor transducer input falls
below this, PV = 0
The time constant of the first-order PV filter. It is the
time required for a 63% change in the IN value.
107
Reference Manual
Appendix D: Function Blocks
00809-0100-4697, Rev GA
November 2014
Table D-4. Multiple Analog Input Function Block Parameters
D.2.1
Number
Parameter
Units
Description
27, 28,
29, 30,
31, 32,
33, 34
CHANNEL_(1,
2, 3,4 5, 6, 7,
8)
None
The CHANNEL (1, 2, 3, 4, 5, 6, 7, 8) value is used to select
the measurement value. See Table D-4 on page 106
for available channels. Configure the CHANNEL
parameters to custom (2) before configuring the
CHANNEL parameters.
35, 36,
37, 38,
39, 40,
41, 42
STDDEV_(1, 2,
3, 4, 5, 6, 7, 8)
% of OUT
Range
Standard deviation of the corresponding measurement.
43, 44,
45, 46,
47, 48,
49, 50
CAP_STDDEV_
(1, 2, 3, 4, 5, 6,
7, 8)
% of OUT
Range
Capability standard deviation, the best deviation that
can be achieved.
Functionality
Simulation
To support testing, either change the mode of the block to manual and adjust the output value
or enable simulation through the configuration tool and manually enter a value for the
measurement value and its status (this single value will apply to all outputs). In both cases, first
set the ENABLE jumper on the field device.
Note
All FOUNDATION fieldbus instruments have a simulation jumper. As a safety measure, the jumper
has to be reset every time there is a power interruption. This measure is to prevent devices that
went through simulation in the staging process from being installed with simulation enabled.
With simulation enabled, the actual measurement value has no impact on the OUT value or the
status. The OUT values will all have the same value as determined by the simulate value.
Figure D-3. Multiple Analog Input Function Block Timing Diagram
OUT (mode in man)
OUT (mode in auto)
PV
63% of Change
FIELD_VAL
Time (seconds)
PV_FTIME
108
Function Blocks
Appendix D: Function Blocks
Reference Manual
November 2014
00809-0100-4697, Rev GA
Figure D-4. Multiple Analog Input Function Block Schematic
XD_SCALE
OUT_SCALE
OUT_1
OUT_2
Ch 1
Ch 2
Ch 3
Ch 4
Ch 5
Ch 6
Ch 7
Ch 8
Simulate
Mode Logic
OUT_3
OUT_4
L_TYPE &
Filter
OUT_5
OUT_6
OUT_7
OUT_8
Filtering
The filtering feature changes the response time of the device to smooth variations in output
readings caused by rapid changes in input. Adjust the filter time constant (in seconds) using the
PV_FTIME parameter (same value applied to eight channels). Set the filter time constant to zero
to disable the filter feature.
Signal conversion
Set the signal conversion type with the Linearization Type (L_TYPE) parameter. Choose from
direct, indirect, or indirect square root signal conversion with the L_TYPE parameter.
Direct
Direct signal conversion allows the signal to pass through the accessed channel input value
(or the simulated value when simulation is enabled).
PV = Channel Value
Indirect
Indirect signal conversion converts the signal linearly to the accessed channel input value
(or the simulated value when simulation is enabled) from its specified range (XD_SCALE) to
the range and units of the PV and OUT parameters (OUT_SCALE).
Channel Value
PV =  -------------------------------------- ¥ ( EU**@100% – EU**@0% ) + EU**@0%


100
** OUT_SCALE values
Indirect square root
Indirect Square Root signal conversion takes the square root of the value computed with
the indirect signal conversion and scales it to the range and units of the PV and OUT
parameters.
PV =
Value
 Channel
- ¥ ( EU**@100% – EU**@0% ) + EU**@0%
 ------------------------------------100
** OUT_SCALE values
Function Blocks
109
Reference Manual
Appendix D: Function Blocks
00809-0100-4697, Rev GA
November 2014
When the converted input value is below the limit specified by the LOW_CUT parameter,
and the Low Cutoff I/O option (IO_OPTS) is enabled (True), a value of zero is used for the
converted value (PV). This option is useful to eliminate false readings when the differential
temperature measurement is close to zero, and it may also be useful with zero-based
measurement devices such as flowmeters.
Note
Low Cutoff is the only I/O option supported by the MAI block. Set the I/O option in Manual or
Out of Service mode only.
Block errors
Table lists conditions reported in the BLOCK_ERR parameter. Conditions in bold are inactive for
the MAI block and are given for reference.
Table D-5. BLOCK_ERR Conditions
Number
Name and description
0
Other
1
Block Configuration Error: the selected channel carries a measurement that is incompatible
with the engineering units selected in XD_SCALE, the L_TYPE parameter is not configured,
or WRITE_CHECK = zero.
2
Link Configuration Error
3
Simulate Active: Simulation is enabled and the block is using a simulated value in its
execution.
4
Local Override
5
Device Fault State Set
6
Device Needs Maintenance Soon
7
Input Failure/Process Variable has Bad Status: The hardware is bad, or a bad status is being
simulated.
8
Output Failure: The output is bad based primarily upon a bad input.
9
Memory Failure
10
Lost Static Data
11
Lost NV Data
12
Readback Check Failed
13
Device Needs Maintenance Now
14
Power Up
15
Out of Service: The actual mode is out of service.
Modes
The MAI Function Block supports three modes of operation as defined by the MODE_BLK
parameter.
Manual (Man)
The block output (OUT) may be set manually.
110
Function Blocks
Appendix D: Function Blocks
Reference Manual
November 2014
00809-0100-4697, Rev GA
Automatic (Auto)
OUT_1 to OUT_8 reflects the analog input measurement or the simulated value when
simulation is enabled.
Out of Service (OOS)
The block is not processed. PV is not updated and the OUT status is set to Bad: Out of
Service. The BLOCK_ERR parameter shows Out of Service. In this mode, changes can be
made to all configurable parameters. The target mode of a block may be restricted to one
or more of the supported modes.
Status handling
Normally, the status of the PV reflects the status of the measurement value, the operating
condition of the I/O card, and any active alarm condition. In Auto mode, OUT reflects the value
and status quality of the PV. In Man mode, the OUT status constant limit is set to indicate that
the value is a constant and the OUT status is Good.
If the sensor limit exceeds the high or low side range, PV status is set high or low and EU range
status is set to uncertain.
In the STATUS_OPTS parameter, select from the following options to control the status
handling.
BAD if limited
Sets the OUT status quality to Bad when the value is higher or lower than the sensor limits.
Uncertain if limited
Sets the OUT status quality to Uncertain when the value is higher or lower than the sensor
limits.
Uncertain if in manual mode
The status of the Output is set to Uncertain when the mode is set to Manual.
Notes
1. The instrument must be OOS to set the status option.
2. The MAI block only supports the BAD if Limited option.
Application information
The intended use for this type of function block is for applications where the sensor types and
functionality of each channel (i.e. the simulate, scaling, filtering, alarms type, and options) are
the same.
The configuration of the MAI function block and its associated output channels depends on the
specific application. A typical configuration for the MAI block involves the following parameters:
Function Blocks
111
Reference Manual
Appendix D: Function Blocks
00809-0100-4697, Rev GA
November 2014
CHANNEL
If the device supports more than one measurement, verify that the selected channel
contains the appropriate measurement or derived value. Refer to Table D-4 on page 106 for
a listing of available channels on the 848T.
L_TYPE
Select Direct when the measurement is already in the desired engineering units for the
block output. Select Indirect when converting the measured variable into another, for
example, pressure into level or flow into energy. Select Indirect Square Root when the block
I/O parameter value represents a flow measurement made using differential pressure, and
when square root extraction is not performed by the transducer.
SCALING
XD_SCALE provides the range and units of the measurement and OUT_SCALE provides the
range and engineering units of the output.
D.2.2
MAI block troubleshooting
Symptom
Possible causes
Corrective action
Mode will not leave
OOS
Target mode not set
Set target mode to something other than OOS.
Configuration error
BLOCK_ERR will show the configuration error bit set. The following are
parameters that must be set before the block is allowed out of OOS:
• Initial value is 1.
• XD_SCALE.UNITS_INDEX must match the units in all the corresponding
sensor transducer blocks.
• L_TYPE must be set to Direct, Indirect, or Indirect Square Root and cannot be
left at initial value of 0.
Process and/or block
alarms will not work.
Value of output does
not make sense
Resource block
The actual mode of the Resource block is OOS. See Resource Block
Diagnostics for corrective action.
Schedule
Block is not scheduled and therefore cannot execute to go to Target
Mode. Typically, BLOCK_ERR will show “Power-Up” for all blocks that are
not scheduled. Schedule the block to execute.
Features
FEATURES_SEL does not have Alerts enabled. Enable the Alerts bit.
Notification
LIM_NOTIFY is not high enough. Set equal to MAX_NOTIFY.
Status Options
STATUS_OPTS has Propagate Fault Forward bit set. This should be
cleared to cause an alarm to occur.
Linearization Type
L_TYPE must be set to Direct, Indirect, or Indirect Square Root and
cannot be left at initial value of 0.
Scaling
Scaling parameters are set incorrectly:
• XD_SCALE.EU0 and EU100 should match that of the corresponding sensor
transducer block.
• OUT_SCALE.EU0 and EU100 are not set properly.
• Both STBs in an ASIC must be set to auto. Best in 1, 2, 7, 8, ASICs in Auto for
thermocouples
112
Function Blocks
Appendix D: Function Blocks
Reference Manual
November 2014
00809-0100-4697, Rev GA
D.3
Input selector function block
IN_1
IN_2
IN_3
IN_4
IN_5
IN_6
IN_7
IN_8
DISABLE_1
DISABLE_2
DISABLE_3
DISABLE_4
DISABLE_5
DISABLE_6
DISABLE_7
DISABLE_8
OP_SELECT
OUT
OUT_D
SELECTED
ISEL
IN (1-8) = Input
DISABLE (1-8) = Discrete input used to disable the associated input channel
SELECTED = The selected channel number
OUT = The block output and status
OUT_D = Discrete output that signals a selected alarm condition
The Input Selector (ISEL) function block can be used to select the first good, Hot Backup™,
maximum, minimum, or average of as many as eight input values and place it at the output. The
block supports signal status propagation. There is process alarm detection in the Input Selector
function block. Table D-6 lists the ISEL block parameters and their descriptions, units of
measure, and index numbers. The block execution time is 30 ms.
Table D-6. Input Selector Function Block Parameters
Number
Function Blocks
Parameter
Units
Description
1
ST_REV
None
The revision level of the static data associated with the
input selector block. The revision value will be
incremented each time a static parameter value in the
block is changed.
2
TAG_DESC
None
The user description of the intended application of the
block.
3
STRATEGY
None
The strategy field can be used to identify groupings of
blocks. This data is not checked or processed by the
block.
4
ALERT_KEY
None
The identification number of the plant unit. This
information may be used in the host for sorting alarms,
etc.
5
MODE_BLK
None
The actual, target, permitted, and normal modes of the
block.
Actual: The mode the “block is currently in”
Target: The mode to “go to”
Permitted: Allowed modes that target may take on
Normal: Most common mode for target
113
Reference Manual
Appendix D: Function Blocks
00809-0100-4697, Rev GA
November 2014
Table D-6. Input Selector Function Block Parameters
Number
114
Parameter
Units
None
Description
6
BLOCK_ERR
7
OUT
8
OUT_RANGE
EU of OUT
9
GRANT_DENY
None
Options for controlling access of host computers and
local control panels to operating, tuning, and alarm
parameters of the block. Not used by device.
10
STATUS_OPTS
None
Allows the user to select options for status handling and
processing.
11,1 2,
13, 14,
25, 26,
27, 28
IN_(1, 2, 3, 4,
5, 6, 7, 8)
Determined
by source
15, 16,
17, 18,
29, 30,
31, 32
DISABLE_(1,
2, 3, 4, 5, 6, 7,
8)
None
A connection from another block that disables the
associated input from the selection.
19
SELECT_TYPE
None
Specifies input selection method. Methods available
include: First Good, Minimum, Maximum, Middle,
Average, or Hot Backup.
20
MIN_GOOD
None
The minimum number of good inputs.
21
SELECTED
None
The selected input number (1 to 8) or the number of
input used for the average output.
22
OP_SELECT
None
Overrides the algorithm to select 1 of the 8 inputs
regardless of the selection type.
23
UPDATE_EVT
None
This alert is generated by any change to the static data
24
BLOCK_ALM
None
The block alarm is used for all configuration, hardware,
connection failure, or system problems in the block.
The cause of the alert is entered in the subcode field.
The first alert to become active will set the Active status
in the Status parameter. As soon as the Unreported
status is cleared by the alert reporting task, another
block may be reported without clearing the Active
status, if the subcode has changed.
33
AVG_USE
None
Number of parameters to use in the averaging
calculation. For example, if AVG_USE is 4 and the
number of connected inputs is 6, then the highest and
lowest values would be dropped prior to calculating the
average. If AVG_USE is 2 and the number of connected
inputs is 7, then the two highest and lowest values
would be dropped prior to calculating the average and
the average would be based on the middle three
inputs.
34
ALARM_SUM
None
The current alert status, unacknowledged states, and
disabled states of the alarms associated with the
function block.
35
ACK_OPTION
None
Used to set automatic acknowledgment of alarms.
OUT_RANGE
This parameter reflects the error status associated with
the hardware or software components associated with
a block. It is a bit string, so that multiple errors may
be shown.
The primary analog value calculated as a result of
executing the function block.
The engineering units code to be used in displaying the
OUT parameter and parameters which have the same
scaling as OUT.
A connection input from another block
Function Blocks
Appendix D: Function Blocks
Reference Manual
November 2014
00809-0100-4697, Rev GA
Table D-6. Input Selector Function Block Parameters
Number
D.3.1
Parameter
Units
Description
Percent
The amount the alarm value must return within the
alarm limit before the associated active alarm condition
clears
36
ALARM_HYS
37
HI_HI-PRI
None
38
HI_HI_LIM
Percent
39
HI_PRI
None
40
HI_LIM
EU of IN
41
LO_PRI
None
42
LO_LIM
EU of IN
43
LO_LO_PRI
None
44
LO_LO_LIM
EU of IN
The setting for the alarm limit sued to detect the LO LO
alarm condition
45
HI_HI_ALM
None
The HI HI alarm data, which includes a value of the
alarm, a timestamp of occurrence and the state of the
alarm
46
HI_ALM
None
The HI alarm data, which includes a value of the alarm,
a timestamp of occurrence and the state of the alarm
47
LO_ALM
None
The LO alarm data, which includes a value of the alarm,
a timestamp of occurrence and the state of the alarm
48
LO_LO_ALM
None
The LO LO alarm data, which includes a value of the
alarm, a timestamp of occurrence and the state of the
alarm
49
OUT_D
None
Discrete output to indicate a selected alarm value
50
ALM_SEL
None
Used to select the process alarm conditions that will
cause the OUT_D parameter to be set.
The priority of the HI HI alarm
The setting for the alarm limit used to detect the HI HI
alarm condition.
The priority of the HI alarm
The setting for the alarm limit used to detect the HI
alarm condition
The priority of the LO alarm
The setting of the alarm limit used to detect the LO
alarm condition
The priority of the LO LO alarm
Functionality
Figure D-5. Input Selector Function Block Schematic
IN_n
OUT
Mode Logic
SELECTED
OP_SELECT
OUT_D
SELECT_TYPE
MIN_GOOD
STATUS_OPTS
Selection
Logic
Alarm
DISABLE_n
Function Blocks
115
Reference Manual
Appendix D: Function Blocks
00809-0100-4697, Rev GA
November 2014
Block errors
Table D-7 lists conditions reported in the BLOCK_ERR parameter. Conditions in bold are inactive
for the ISEL block and are given for reference.
Table D-7. BLOCK_ERR Conditions
Number
Name and description
0
Other: The output has a quality of uncertain.
1
Block Configuration Error: Select type is not configured
2
Link Configuration Error
3
Simulate Active
4
Local Override
5
Device Fault State Set
6
Device Needs Maintenance Soon
7
Input Failure/Process Variable has Bad Status: One of the inputs is Bad.
8
Output Failure
9
Memory Failure
10
Lost Static Data
11
Lost NV Data
12
Readback Check Failed
13
Device Needs Maintenance Now
14
Power Up: The device was just powered-up.
15
Out of Service: The actual mode is out of service.
Modes
The ISEL function block supports three modes of operation as defined by the MODE_BLK
parameter:
Manual (Man)
The block output (OUT) may be set manually.
Automatic (Auto)
OUT reflects the selected value.
Out of Service (OOS)
The block is not processed. The BLOCK_ERR parameter shows Out of Service. The target
mode of a block may be restricted to one or more of the supported modes. In this mode,
changes can be made to all configurable parameters.
116
Function Blocks
Appendix D: Function Blocks
Reference Manual
November 2014
00809-0100-4697, Rev GA
Alarm detection
A block alarm will be generated whenever the BLOCK_ERR has an error bit set. The type of block
errors for the ISEL block are defined above.
Process Alarm detection is based on the OUT value. The alarm limits of the following standard
alarms can be configured.

High (HI_LIM)

High high (HI_HI_LIM)

Lo (LO_LIM)

Lo low (LO_LO_LIM)
In order to avoid alarm chattering when the variable is oscillating around the alarm limit, an
alarm hysteresis in percent of the PV span can be set using the ALARM_HYS parameter. The
priority of each alarm is set in the following parameters:

HI_PRI

HI_HI_PRI

LO_PRI

LO_LO_PRI
Table D-8. Alarm Priority Levels
Number
Description
0
The priority of an alarm condition changes to 0 after the condition that caused the alarm is
corrected.
1
An alarm condition with a priority of 1 is recognized by the system, but is not reported to
the operator.
2
An alarm condition with a priority of 2 is reported to the operator, but does not require
operator attention (such as diagnostics and system alerts).
3-7
Alarm conditions of priority 3 to 7 are advisory alarms of increasing priority.
8-15
Alarm conditions of priority 8 to 15 are critical alarms of increasing priority.
Block execution
The ISEL function block reads the values and status of up to eight inputs. To specify which of the
six available methods (algorithms) is used to select the output, configure the selector type
parameter (SELECT_TYPE) as follows:

Max selects the maximum value of the inputs.

Min selects the minimum value of the inputs.

Avg calculates the average value of the inputs.

Mid calculates the update for eight sensors.

1st Good selects the first available good input.
If the DISABLE_N is active, the associated input is not used in the selection algorithm.
If an input is not connected, it is also not used in the algorithm.
Function Blocks
117
Reference Manual
Appendix D: Function Blocks
00809-0100-4697, Rev GA
November 2014
If the OP_SELECT is set to a value between 1 and 8, the selection type logic is overridden and the
output value and status is set to the value and status of the input selected by OP_SELECT.
SELECTED will have the number of selected input unless the SELECT_TYPE is mid, in which case
it will take the average of the two middle values. Then SELECTED will be set to “0” if there is an
even number of inputs.
Status handling
In Auto mode, OUT reflects the value and status quality of the selected input. If the number of
inputs with Good status is less than MIN_GOOD, the output status will be Bad.
In Man mode, the OUT status high and low limits are set to indicate that the value is a constant
and the OUT status is always Good.
In the STATUS_OPTS parameter, select from the following options to control the status
handling:
Use uncertain as good
Sets the OUT status quality to Good when the selected input status is Uncertain.
Uncertain if in manual mode
The status of the Output is set to Uncertain when the mode is set to manual.
Note
The instrument must be to OOS to set the status option.
Application information
Use the ISEL function block to:

Select the maximum temperature input from eight inputs and send it to another
function block (see <HotXRef>Figure D-6)

Calculate the average temperature of the eight inputs (see <HotXRef>Figure D-7)

Use only six of the eight inputs to calculate the average temperature.
Figure D-6. Input Selector Function Block Application Example (SEL_TYPE = max)
IN1 = 126 °F
IN2 = 118 °F
118
Input Selector
(ISEL) Function
Block
IN3 = 104 °F
IN4 = 107 °F
OUT = 140 °F
IN5 = 112 °F
IN6 = 115 °F
IN7 = 130 °F
IN8 = 140 °F
SEL_TYPE = max
To Another
Function Block
Function Blocks
Appendix D: Function Blocks
Reference Manual
November 2014
00809-0100-4697, Rev GA
Figure D-7. Input Selector Function Block Application Example (SEL_TYPE = average)
AVG_USE = 6
IN1 = 126 °F
IN2 = 118 °F
Input Selector
(ISEL) Function
Block
IN3 = 104 °F
IN4 = 107 °F
IN5 = 112 °F
IN6 = 115 °F
IN7 = 130 °F
IN8 = 140 °F
OUT = 118 °F
To Another
Function Block
AVG_USE = 6
SEL_TYPE = avg
To determine OUT for a 6-input reading, read all eight, sort in numerical order, drop
the highest and lowest values, and calculate the average.
107 + 112 + 115 + 118 + 126 + 130
----------------------------------------------------------------------------------------------- = 118°F
6
D.3.2
ISEL block troubleshooting
Symptom
Possible causes
Corrective action
Mode will not leave
OOS
Target mode not set
Set target mode to something other than OOS.
Configuration error
BLOCK_ERR will show the configuration error bit set. SELECT_TYPE must be
set to a valid value and cannot be left at 0.
Resource block
The actual mode of the Resource block is OOS. See Resource Block
Diagnostics for corrective action.
Schedule
Block is not scheduled and therefore cannot execute to go to Target Mode.
Typically, BLOCK_ERR will show “Power-Up” for all blocks that are not
scheduled. Schedule the block to execute.
Inputs
All inputs have Bad status
OP selected
OP_SELECT is not set to 0 (or it is linked to an input that is not 0), and it points
to an input that is Bad.
Min good
The number of Good inputs is less than MIN_GOOD.
Block is in OOS
mode
Change mode to Auto
Features
FEATURES_SEL in the resource block does not have Alerts enabled. Enable
the Alerts bit.
Notification
LIM_NOTIFY in the resource block is not high enough. Set equal to
MAX_NOTIFY.
Status Options
STATUS_OPTS has Propagate Fault Forward bit set. This should be cleared to
cause an alarm to occur.
Scaling
Limit values are outside the OUT_SCALE.EU0 and OUT_SCALE.EU100 values.
Change OUT_SCALE or set values within range.
Status of output is
BAD
Block alarms will
not work.
Cannot set
HI_LIMIT,
HI_HI_LIMIT,
LO_LIMIT,
LO_LO_LIMIT
Function Blocks
119
Appendix D: Function Blocks
November 2014
120
Reference Manual
00809-0100-4697, Rev GA
Function Blocks
Reference Manual
Index
00809-0100-4697, Rev GA
November 2014
Index
Numerics
2-Inch Pipe Stand
Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
A
Alarms
Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Analog Input
Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Ground . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Analog Input Function Block. . . . . . . . . . . . . . . . . . . . . . 93
Advanced Features . . . . . . . . . . . . . . . . . . . . . . . . 100
Alarm Detection . . . . . . . . . . . . . . . . . . . . . . . . . . . 99
Application Information . . . . . . . . . . . . . . . . . . . . 100
Block Errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98
Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Direct. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
Filtering . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
Functionality . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
Indirect . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
Indirect Square Root . . . . . . . . . . . . . . . . . . . . . . . . 98
Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98
Automatic . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99
Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99
Out of Service . . . . . . . . . . . . . . . . . . . . . . . . . 99
Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
Signal Conversion . . . . . . . . . . . . . . . . . . . . . . . . . . 97
Simulation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
Status Handling . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . 101
Wiring Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
B
Block Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
Alerts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
Instrument Specific Blocks . . . . . . . . . . . . . . . . . . . 87
C
Cable Glands
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Tag . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Conduit Entries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Index
17
48
16
18
Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Analog Transmitters . . . . . . . . . . . . . . . . . . . . . . . . 25
Analog Input Block . . . . . . . . . . . . . . . . . . . . . 25
Multiple Analog Input Block . . . . . . . . . . . . . . 25
Block . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Custom . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Damping . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Differential Sensor Blocks . . . . . . . . . . . . . . . . . . . . 21
Methods . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Monitoring Applications
Single Selection . . . . . . . . . . . . . . . . . . . . . . . . 23
Typical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Reset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Restart Processor . . . . . . . . . . . . . . . . . . . . . . 49
Restart with Defaults. . . . . . . . . . . . . . . . . . . . 49
Resource Block . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Standard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
Analog Inputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Millivolt Inputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Ohm Inputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
RTD Inputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Thermocouple Inputs . . . . . . . . . . . . . . . . . . . . . . . 10
D
Damping
Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Device Descriptions . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
Differential Sensor Blocks
Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Differential Transducer Block
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
DIN Rail
Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
Drawing
Switch Location . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Drawings
848T Analog Connector . . . . . . . . . . . . . . . . . . . . . 11
Analog Input Wiring . . . . . . . . . . . . . . . . . . . . . . . . 11
Block Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Block Internal Structure . . . . . . . . . . . . . . . . . . . . . 86
Cable Gland Installation . . . . . . . . . . . . . . . . . . . . . 17
Commissioning Tag . . . . . . . . . . . . . . . . . . . . . . . . 16
Conduit Entries Installation . . . . . . . . . . . . . . . . . . 18
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
Sensor Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Transmitter Label . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Transmitter Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . .9
Index-1
Reference Manual
Index
00809-0100-4697, Rev GA
November 2014
F
FOUNDATION fieldbus . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Addressing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
Block Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
Alerts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
Instrument- Specific Blocks . . . . . . . . . . . . . . . 87
Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Device Descriptions . . . . . . . . . . . . . . . . . . . . . . . . 86
Function Block Scheduling . . . . . . . . . . . . . . . . . . . 92
Function Blocks . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Link Active Scheduler . . . . . . . . . . . . . . . . . . . . . . . 88
Network Communication . . . . . . . . . . . . . . . . . . . . 87
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Scheduled Transfers . . . . . . . . . . . . . . . . . . . . . . . . 90
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Unscheduled Transfers . . . . . . . . . . . . . . . . . . . . . . 91
Function Blocks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Analog Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
Input Selector Function Block . . . . . . . . . . . . . . . . 110
Multiple Analog Input . . . . . . . . . . . . . . . . . . . . . . 102
Scheduling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
G
Grounding. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Analog Device . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Grounded Thermocouple . . . . . . . . . . . . . . . . . . . .
Shielded Wire . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Transmitter Enclosure . . . . . . . . . . . . . . . . . . . . . . .
Ungrounded mV . . . . . . . . . . . . . . . . . . . . . . . . . . .
Ungrounded RTD/Ohm . . . . . . . . . . . . . . . . . . . . . .
Ungrounded Thermocouple . . . . . . . . . . . . . . . . . .
12
14
14
13
14
13
13
13
H
Hardware
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Communication Check . . . . . . . . . . . . . . . . . .
Power Check . . . . . . . . . . . . . . . . . . . . . . . . . .
Reset Configuration . . . . . . . . . . . . . . . . . . . .
Sensor Check . . . . . . . . . . . . . . . . . . . . . . . . . .
48
48
48
49
48
I
Input Selector Function Block . . . . . . . . . . . . . . . . . . . .
Alarm Detection . . . . . . . . . . . . . . . . . . . . . . . . . .
Application Information . . . . . . . . . . . . . . . . . . . .
Block Execution . . . . . . . . . . . . . . . . . . . . . . . . . . .
Errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Functionality . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Automatic . . . . . . . . . . . . . . . . . . . . . . . . . . .
Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Out of Service . . . . . . . . . . . . . . . . . . . . . . . .
Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Status Handling . . . . . . . . . . . . . . . . . . . . . . . . . . .
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . .
Index-2
110
114
116
115
113
113
114
114
114
114
111
115
117
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Intrinsically Safe . . . . . . . . . . . . . . . . . . . . . . . . . . .
Non-Incendive . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Using Cable Glands . . . . . . . . . . . . . . . . . . . . . . . . .
Using Conduit Entries . . . . . . . . . . . . . . . . . . . . . . .
17
75
75
17
18
J
Junction Box
Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
L
Link Active Scheduler . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
Backup LAS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
LAS Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
M
Maintenance
Hardware . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Communication Check . . . . . . . . . . . . . . . . . . 48
Power Check . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Reset Configuration . . . . . . . . . . . . . . . . . . . . 49
Sensor Check . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Measurement Transducer Block
Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Monitoring Applications
Common Configurations
Single Selection . . . . . . . . . . . . . . . . . . . . . . . . .23
Typical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23
Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
2-Inch Pipe Stand . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
DIN Rail Without an Enclosure . . . . . . . . . . . . . . . . . .6
Panel with a Junction Box . . . . . . . . . . . . . . . . . . . . . .6
Multiple Analog Input
Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Multiple Analog Input Block
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . 109
Multiple Analog Input Function Block . . . . . . . . . . . . . 102
Application Information . . . . . . . . . . . . . . . . . . . . 108
Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Errors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106
Filtering . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105
Functionality . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107
Automatic . . . . . . . . . . . . . . . . . . . . . . . . . . . 107
Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107
Out of Service . . . . . . . . . . . . . . . . . . . . . . . . 107
Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
Signal Conversion . . . . . . . . . . . . . . . . . . . . . . . . . 105
Direct. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106
Indirect . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106
Indirect Square Root . . . . . . . . . . . . . . . . . . . 106
Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107
Simulation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
Status Handling . . . . . . . . . . . . . . . . . . . . . . . . . . 108
Index
Reference Manual
Index
00809-0100-4697, Rev GA
November 2014
N
Network Communication . . . . . . . . . . . . . . . . . . . . . . . .
Addressing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Function Block Scheduling . . . . . . . . . . . . . . . . . . .
Link Active Scheduler . . . . . . . . . . . . . . . . . . . . . . .
Scheduled Transfer . . . . . . . . . . . . . . . . . . . . . . . . .
Unscheduled Transfer . . . . . . . . . . . . . . . . . . . . . . .
87
89
92
88
90
91
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
FOUNDATION fieldbus . . . . . . . . . . . . . . . . . . . . . . . . 85
Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
P
Performance specifications . . . . . . . . . . . . . . . . . . . . . . 54
Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
R
30
26
30
30
30
30
26
34
31
33
31
32
50
Index
54
12
15
15
15
Tagging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Transducer Block
Alarm Detection . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Channel Definitions . . . . . . . . . . . . . . . . . . . . . . . . 36
Errors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Automatic . . . . . . . . . . . . . . . . . . . . . . . . . . . . .37
Out of Service . . . . . . . . . . . . . . . . . . . . . . . . . .37
Status Handling . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Transients . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Transmitter
Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Tag . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Transmitter Wiring Diagram . . . . . . . . . . . . . . . . . . . . . . .9
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Analog Input Function Block. . . . . . . . . . . . . . . . . 101
Differential Transducer Block . . . . . . . . . . . . . . . . . 50
FOUNDATION fieldbus . . . . . . . . . . . . . . . . . . . . . . . . 49
Input Selector Function Block . . . . . . . . . . . . . . . . 117
Multiple Analog Input Block . . . . . . . . . . . . . . . . . 109
Resource Block . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
U
Unscheduled Transfers . . . . . . . . . . . . . . . . . . . . . . . . . . 91
W
S
Scheduled Transfers . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Client . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Publisher . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Report Distribution . . . . . . . . . . . . . . . . . . . . . . . . .
Server . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Subscriber . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Security Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Senor
Connection Check . . . . . . . . . . . . . . . . . . . . . . . . . .
Sensor
Tag . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Sensor Transducer Block
Change Sensor Configuration . . . . . . . . . . . . . . . . .
Sensor Calibration . . . . . . . . . . . . . . . . . . . . . . . . . .
Sensor Wiring Diagram. . . . . . . . . . . . . . . . . . . . . . . . . .
Shield Wire
Ground . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
15
T
O
Resource Block
Alarm Detection . . . . . . . . . . . . . . . . . . . . . . . . . . .
Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Automatic . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Out of Service (OOS) . . . . . . . . . . . . . . . . . . . .
Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
PlantWeb Alerts
Recommended Actions . . . . . . . . . . . . . . . . . .
PlantWeb™ Alerts . . . . . . . . . . . . . . . . . . . . . . . . . .
advisory alarms . . . . . . . . . . . . . . . . . . . . . . . .
failed_alarms . . . . . . . . . . . . . . . . . . . . . . . . . .
maint_alarms . . . . . . . . . . . . . . . . . . . . . . . . .
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . .
Simulate Enable Switch . . . . . . . . . . . . . . . . . . . . . . . . .
Specifications
performance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Surges . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Switches . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Security . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Simulate Enable . . . . . . . . . . . . . . . . . . . . . . . . . . .
90
90
90
90
90
90
15
Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
Communication Check . . . . . . . . . . . . . . . . . . . . . . 48
Power Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
48
16
45
45
10
13
Index-3
Index
November 2014
Index-4
Reference Manual
00809-0100-4697, Rev GA
Index
Reference Manual
00809-0100-4697, Rev GA
November 2014
Standard Terms and Conditions of Sale can be found at www.rosemount.com/terms_of_sale
The Emerson logo and DeltaV are trademarks of Emerson Electric Co.
Rosemount. PlantWeb, and the Rosemount logotype are registered trademarks of Rosemount Inc.
HART is a registered trademark of the HART Communication Foundation.
FOUNDATION fieldbus is a registered trademark of the Fieldbus Foundation.
minifast and eurofast are registered trademarks of TURCK.
All other marks are the property of their respective owners.
© 2014 Rosemount, Inc. All rights reserved.
Emerson Process Management
Emerson Process Management
Rosemount Measurement
GmbH & Co.
8200 Market Boulevard
Argelsrieder Feld 3
Chanhassen MN 55317 USA
82234 Wessling
Tel (USA) 1 800 999 9307
Germany
Tel (International) +1 952 906 8888 Tel 49 (8153) 9390
Fax +1 952 906 8889
Fax 49 (8153) 939172
Emerson Process Management
Latin America
1300 Concord Terrace, Suite 400
Sunrise Florida 33323 USA
Tel +1 954 846 5030
Emerson Process Management Asia
Pacific Private Limited
1 Pandan Crescent
Singapore 128461
T (65) 6777 8211
F (65) 6777 0947
[email protected]
Beijing Rosemount Far East
Instrument Co., Limited
No. 6 North Street,
Hepingli, Dong Cheng District
Beijing 100013, China
T (86) (10) 6428 2233
F (86) (10) 6422 8586

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Related manuals

Download PDF

advertisement