TRANSMITTER TR485-103D-150, TR485-

103D-700

Programmable transmitter of temperature, relative humidity and other derived humidity values with RS485 serial output

Instruction Manual

Instruction Manual for use of temperature and relative humidity transmitter TR485-103D-150, TR485-103D-700

Transmitter is designed for measurement of air temperature at °C or °F and relative humidity of air without aggressive ingredients with computing one of the following values: dew point temperature, absolute humidity, specific humidity, mixing ratio and specific enthalpy. Please read instruction manual before the first transmitter connection. Transmitter communicates via link RS485. Supported communication protocols are Modbus RTU, protocol compatible with standard Advantech-ADAM and ARION. Temperature and relative humidity sensors are nonremovable instrument parts. Measured and computed values are displayed on dual line LCD display. The first line displays temperature. Value displayed on the second line is selectable among relative humidity and computed value. It is also possible to display both readings with cyclic overwriting in 4 seconds interval. Display can be also switched OFF totally. Output link RS485 is

galvanic isolated. Please read instruction manual before the first device connection.

Use user’s software Tsensor for setting of all device parameters (recommended). It supports make the adjustment of the device too. This procedure is described at file „Calibration

manual.pdf“ which is installed commonly with the software. Change of some parameters is possible to do without user’s software with Windows hyperterminal (change of communication protocol, its parameters, LCD display setting). It is described in file “Description of communication protocols of

Txxxx series” which is free to download at the same address.

Transmitter setting from the manufacturer

If special setting was not required in the order, transmitter is set from the manufacturer to the following parameters: communication protocol: transmitter address: preset computed value:

Modbus RTU

01H communication speed: display:

9600Bd, without parity, 2 stop bits

switched ON value displayed at lower line: relative humidity only

dew point temperature

Transmitter installation

Transmitter is designed for air-conditioning duct – metal stem fixed by means of the optional gland. Also optional installation flanges PP4 or PP90 are possible to use (not included in delivery). Measuring temperature and relative humidity sensors are non-removable instrument parts.

It is NOT recommended to use the device for long time under condensation conditions. It could be the cause of water steam condensation inside the sensor’s cover into water phase. This liquid phase stays inside sensor’s cover and can’t escape from the cover easily. It can dramatically increase response time to relative humidity change. If water condensation occurs for longer time it can cause sensor damage. Similar effect can occur under water aerosol conditions. If this effect may occur, it is necessary to use the device at operation position with sensor cover downwards. Don’t connect transmitter while power supply voltage is on. Interconnection terminals are accessible after unscrewing four screws and removing the lid. Lace the cable through a gland at the case wall.

Connect the cable to terminals with respecting the signal polarity (see figure). Terminals are selfclamping and can be opened by a suitable screwdriver. For the opening, insert the screwdriver to smaller terminal hole and lever by him. Do not forget to tighten glands and case lid with inserted packing after cables connecting. It is necessary for warranting of protection IP65.

2

It is recommended to use shielded twisted copper cable (e.g. SYKFY ). Outside diameter of the cable must be from 3 to 6.5 mm, maximal length 1200m. The cable must be located at indoor rooms.

Nominal cable impedance should be 100 Ω, loop resistance max. 240 Ω, cable capacity max.

65 pF/m. Suitable cable is e.g. SYKFY 2x2x0,5 mm 2 , where one wire pair serves for transmitter powering and the other pair for communication link.

The cable should be led in one line, i.e. NOT to „tree“ or „star“. Termination resistor should be located at the end. For short distance other topology is allowed. Terminate the network by a termination resistor. The value of the resistor is recommended about 120 Ω . For short distance termination resistor can be left out.

The cable should not be led in parallel along power cabling. Safety distance is up to 0.5 m, otherwise undesirable induction of interference signals can appear.

Electrical system (wiring) may do only worker with required qualification by rules in operation.

Dimensions

3

Typical application wiring, connection of terminals

Info mode

If in doubt of setting of installed transmitter, verification of its address is enabled even without using computer. Power should be connected.

Unscrew transmitter cover and shortly press button next to connection terminals (jumper should be opened). Actual adjusted address of transmitter is displayed on LCD display at decimal base. Next press of button exits info mode and actual measured values are displayed.

Note: No measurement and communication is possible during info mode. If transmitter stays in info mode for longer than 15 s, transmitter automatically returns to measuring cycle.

Description of communication protocols

Detailed description of each communication protocols including examples of communication is available in individual document “Description of communication protocols of

Txxxx series”.

Note: After switching ON the power of the device it can last up to 2 s before the device starts to communicate and measure!

Modbus RTU

Control units communicate on master-slave principle in half-duplex operation. Only master can send request and only addressed device responds. During sending of request no other slave station should respond. During communication, data transfer proceeds in binary format. Each Byte is sent as eight bit data word in format: 1 start bit, data word 8 bit (LSB first), 2 stop bits 1

, without

parity. Transmitter supports communication speed from 110Bd to 115200Bd.

1 Transmitter sends two stop bits, for receive one stop bit is enough.

4

Sent request and response have syntax: ADDRESS OF DEVICE – FUNCTION – Modbus CRC

Supported functions

03 (0x03): Reading of 16-bit registers (Read Holding Registers)

04 (0x04): Reading of 16-bit input gates (Read Input Registers)

16 (0x10): Setting of more 16-bit registers (Write Multiple Registers)

Jumper and button

Jumper and button are located next to connection terminals. If communication protocol

Modbus is selected the function of jumper and button is as follows:

• Jumper opened – transmitter memory is protected from writing, from transmitter side it is only enabled to read measured value, writing to memory is disabled (no change of transmitter address, communication speed and LCD setting is enabled)

• Jumper closed – writing to transmitter memory is enabled by means of User’s software

• Jumper closed and button pressed for longer than six seconds – causes restoring of manufacturer setting of communication protocol, i.e. sets Modbus RTU communication protocol, device address sets to 01h and communication speed to 9600Bd (after button press there is “dEF” message blinking at LCD display. Six seconds later message “dEF” stays shown, it means manufacturer setting of communication protocol is done).

• Jumper opened and button shortly pressed – transmitter goes to Info mode, see chapter „Info

mode

Modbus registers of the device

Variable Unit Size Status

Measured temperature

Measured relative humidity

Computed value *

Address of transmitter

[°C] 0x0031 Int*10 BIN16 R

[%] 0x0032 Int*10 BIN16 R

[°C] 0x0033 Int*10 BIN16 R

[-] 0x2001 Int BIN16 R/W*

Code of communication speed [-]

Serial number of transmitter Hi [-]

Serial number of transmitter Lo [-]

Version of Firmware Hi

Version of Firmware Lo

[-]

[-]

0x2002

0x1035

0x1036

0x3001

0x3002

Int

BCD

BCD

BCD

BCD

BIN16

BIN16

BIN16

BIN16

BIN16

R/W*

R

R

R

R

Explanation:

• * to choice of computed value type use User’s software

• Int*10 register is in format integer*10

• R register

• W* register is designed for writing, for more details see chapter description of

• X communication protocols register addresses are indexed from zero – register 0x31 is physically sent as value

0x30, 0x32 as 0x31 (zero based addressing)

Note: In case there is a need for reading of measured values from transmitter with higher resolution than one decimal, measured values in transmitter are stored also in „Float“ format, which is not directly compatible with IEEE754.

Protocol compatible with Advantech-ADAM standard

Control units communicate on master-slave principle in half-duplex operation. Only master can send requests and only addressed device responds. During sending request any of slave devices should respond. During communication data is transferred in ASCII format (in characters). Each

Byte is sent as two ASCII characters. Transmitter supports communication speed from 1200Bd to

5

115200Bd, parameters of communication link are 1 start bit + eight bit data word (LSB first) +

1 stop bit, without parity.

Jumper

Jumper is located next to connection terminals. If communication protocol compatible with standard Advantech-ADAM is selected, its function is the following:

• If jumper during switching ON the power is closed, transmitter always communicates with following parameters regardless stored setting in the transmitter: communication speed 9600 Bd, without check sum, transmitter address 00h

• If jumper during switching ON the power is not closed, transmitter communicates in accordance with stored setting.

• If jumper is closed during transmitter operation, transmitter temporarily changes its address to 00h, it will communicate in the same communication speed as before closing jumper and will communicate without check sum. After jumper is opened setting of address and check sum is reset in accordance with values stored in transmitter.

• Communication speed and check sum are possible to change only if jumper is closed.

• Jumper closed and button pressed for longer than six seconds – causes restoring of manufacturer setting of communication protocol, i.e. sets Modbus RTU communication protocol, device address sets to 01h and communication speed to 9600Bd (after button press there is “dEF” message blinking at LCD display. Six seconds later message “dEF” stays shown, it means manufacturer setting of communication protocol is done).

For communication with device which measure more than one value, there is necessary to add at the end of command, number of communication channel, where the measured value is mapped.

Command for value reading is #AAx(CRC) CR, where AA is device address and x is number of communication channel by next table:

Measured value Number of communication channel

Temperature 0

Relative humidity 1

Computed value 2

ARION communication protocol

The device supports communication protocol ARiON version 1.00. For more details see file

Description of communication protocols of Txxxx series”.

Error States of the device

Device continuously checks its state during operation. In case error is found LCD displays corresponding error code:

Error 0

First line displays „Err0“.

Check sum error of stored setting inside device’s memory. This error appears if incorrect writing procedure to device’s memory occurred or if damage of calibration data appeared. At this state device does not measure and calculate values. It is a serious error, contact distributor of the instrument to fix.

6

Error 1

Measured (calculated) value is over upper limit of allowed full scale range. There is a reading

„Err1“ on LCD display. Value read from the device is +999,9.

This state appears in case of:

• Measured temperature is higher than approximately 600°C (i.e. high non-measurable resistance of temperature sensor, probably opened circuit).

• Relative humidity is higher than 100%, i.e. damaged humidity sensor, or humidity calculation of humidity is not possible (due to error during temperature measurement)

• Computed value – calculation of the value is not possible (error during measurement of temperature or relative humidity or value is over range)

Error 2

There is a reading „Err2“ on LCD display. Measured (calculated) value is below lower limit of allowed full scale range. Value read from the transmitter is -999,9.

This state appears in case of:

• Measured temperature is lower than approximately -210°C (i.e. low resistance of temperature sensor, probably short circuit).

• Relative humidity is lower than 0%, i.e. damaged sensor for measurement of relative humidity, or calculation of humidity is not possible (due to error during temperature measurement)

• Computed value – calculation of computed value is not possible (error during measurement of temperature or relative humidity)

Error 3

There is a reading „Err3“ on LCD display upper line.

Error of internal A/D converter appeared (converter does not respond, probably damage of A/D converter). No measurements proceed. It is a serious error, contact distributor of the instrument.

Readings on LCD display

°C, °F

Reading next to this symbol is measured temperature or error state of value.

%RH

Reading next to this symbol is measured relative humidity or error state of value.

°C / °F DP

Reading next to this symbol is calculated dew point temperature or error state of value. g/m

3

Reading next to this symbol is calculated absolute humidity or error state of value. g/kg

Reading next to this symbol is calculated specific humidity or mixing ratio (depends on device setting) or error state of value.

If specific enthalpy is selected, there is shown only value (number) without corresponding unit! symbol 3 near by left display margin

Is on if jumper is closed.

7

Technical parameters of the instrument:

RS 485 Interface:

Receiver-Input Resistance: 96 kΩ

Devices on bus: max. 256 (1/8 Unit Receiver Load)

Measuring parameters:

Ambient temperature (RTD sensor Pt1000/3850ppm):

Measuring range: -30 to +125 °C

Resolution of display: 0.1 °C

Accuracy: ± 0.4 °C

Relative humidity (reading is temperature compensated at entire temperature range):

Measuring range: 0 to 100 %RH (see Transmitter installation)

Resolution display: 0.1 %RH

Accuracy: ± 2.5 %RH from 5 to 95 %RH at 23 °C

Measuring temperature and humidity range is limited in accordance with the graph below

8

The value computed from ambient temperature and relative humidity:

Display resolution: 0,1 °C

You can choice one of the next value:.

Dew point temperature

Accuracy: ±1,5 °C at ambient temperature T < 25°C and RH >30%

Range: -60 to +80 °C

Absolute humidity

Accuracy: ±3g/m 3 at ambient temperature T < 40°C

Range: 0 to 400 g/m 3

Specific humidity

2

Accuracy:

Range:

±2g/kg at ambient temperature T < 35°C

0 to 550 g/kg

Mixing ratio

Accuracy:

Range:

±2g/kg at ambient temperature T < 35°C

0 to 995 g/kg

Specific enthalpy

Accuracy:

Range:

± 3kJ/kg at ambient temperature T < 25°C

0 to 995 kJ/kg

3

Response time with stainless steel mesh sensor cover (F5200) and bronze sensor cover

(F0000 - selectable option), air flow approximately 1 m/s: temperature: t90 < 9 min (temperature step 20 °C) relative humidity: t90 < 30 s (humidity step 65 %RH, constant temperature)

Recommended calibration interval: 1 year

Measuring interval and LCD display refresh: 0.5 s

Power: 9 to 30 V dc

Consumption: max. 0.5W

Protection of the case with electronics and terminals: IP65

Protection of the sensor cover: IP40

Filtering ability of the sensor cover: 0.025 mm

Operating conditions:

Operating temperature range of case with electronics: -30 to +80 °C, over +70°C switch

LCD display OFF

Operating temperature range measuring tip with sensors: -30 to +125 °C

Operating relative humidity range: 0 to 100 %RH

Outer influence in accordance with EN 33-2000-3: normal environment with those specifications: AE1, AN1, AR1, BE1

Working position: in air-conditioning duct arbitrary, in free space the steel stem downwards

(see Transmitter installation)

Electromagnetic compatibility: complies EN 61326-1

2 This value depends on the atmospheric pressure. For computing is used constant value stored inside device memory. Default value preset by manufacturer is 1013hPa and can be changed by

3 user’s software.

This maximum is reached under conditions about 70°C/100%RH or 80°C/70%RH

9

Not allowed manipulation

It is not allowed to operate the device under conditions other than specified in technical parameters.

Devices are not designed for locations with chemically aggressive environment. Temperature and humidity sensors must not be exposed to direct contact with water or other liquids. It is not allowed to remove the sensor cover—to avoid any mechanical damage of the sensors.

Storing conditions: temperature -30 to +80 °C, humidity 0 to 100 %RH without condensation

Dimensions: case with electronics 89x73x37 mm, length of stem 150mm (TR485-103D-

150)/700mm (TR485-103D-700)

Weight: TR485-103D-150 approximately 250 g, TR485-103D-700 approximately 600 g

Material of the case: ABS, stem from stainless steel

Optional accessory

Mounting flanges:

PP4 PP90

End of operation

Device itself (after its life) is necessary to liquidate ecologically!

Technical support and service

Technical support and service is provided by distributor.

10

Appendix A

Connection of ELO E06D converter (RS232/RS485)

(optionally accessory of the transmitter TR485-103D-700)

The ELO E06D converter is an optional accessory for connection of transmitter/transmitters with

RS485 interface to the PC via serial port RS232. Connect connector marked RS232 directly to the

PC, connect power to connector marked RS485. Power voltage +6V DC from an external acdc adapter connect to pin 9, 0V connect to pin 5. Also mutually connect pin 2 and pin 7. Link RS485 is connected across pin 3 (A+) and pin 4 (B-).

11