Mettler Toledo PG-S For precision balances PG-S Operating instructions

Add to my manuals
68 Pages

advertisement

Mettler Toledo PG-S For precision balances PG-S Operating instructions | Manualzz

Operating instructions

METTLER TOLEDO

PG-S balances (0.1 g, 1 g)

PG5001-S

Min

Max

5100g

0.5g

FACT

METTLER TOLEDO

Overview of your PG-S balance

Front view PG-S

5

PG5001-S

Max

Min

5100g

0.5g

FACT

METTLER TOLEDO

Bottom view PG-S

11

1

Display PG-S

17

16

10

18 19 20 21

15 14

9 8 7 6

13

12

2

3

4

Rear view PG-S

Display, controls and connections of your PG-S balance

Front view

No.

Designation

1 Display

2 Weighing pan

0 3 Leveling foot

0 4 Control keys

0 5 Model-specific data and type designation

Rear/bottom view

No.

Designation

0 6 RS232C interface connection

0 7 Leveling control

0 8 Eye for antitheft device

0 9 Connection for AC adapter

10 Hanger for below-the-scale weighings

11 Fastening for auxiliary display

Display

No.

Designation

12 DeltaTrac

13 Weighing units

14 Alphanumeric display

15 Symbol of the stability detector

16 Symbol for calculated result

17 Status indicator of the vibration adapter

18 Status indicator of the weighing process adapter

19 Status indicator of the repeatability

20 Special symbols (e.g. for net weight)

21 Display of adjustment mode

Contents

4

Contents

3.

3.1

3.2

3.3

3.4

3.5

3.6

3.7

3.8

2.

2.1

2.2

2.3

2.4

2.5

1.

1.1

1.2

1.3

1.4

Getting to know your PG-S balance␣ ..........................................................................................

Introduction ..............................................................................................................................

Overview of the PG-S balances ...................................................................................................

What you should know about these instructions ...........................................................................

Safety has priority .....................................................................................................................

Putting the balance into operation␣ ...........................................................................................

Unpacking and checking the standard equipment .........................................................................

Power supply ...........................................................................................................................

6

6

6

7

8

9

9

Location selection or location change..........................................................................................

10

Leveling the balance .................................................................................................................

11

11

Adjusting (calibrating) the balance .............................................................................................

12

Weighing made simple␣ ...........................................................................................................

13

Switching the balance on and off ................................................................................................

13

Zeroing the scale ......................................................................................................................

14

Taring the balance ....................................................................................................................

14

Performing a simple weighing ....................................................................................................

15

Weighing with the analog display – DeltaTrac ..............................................................................

15

DeltaRange ® balances with movable fine range ............................................................................

16

Switching weighing units ...........................................................................................................

16

Printing out the weighing result and transferring data ....................................................................

17

4.4

4.5

4.6

4.7

4.8

4.

4.1

4.2

4.3

4.9

4.10

4.11

4.12

The menu␣ ...............................................................................................................................

18

What is the menu? ....................................................................................................................

18

Menu operation ........................................................................................................................

19

Reset .......................................................................................................................................

21

Selection of the adjustment (calibration) and test function .............................................................

21

Switching automatic adjustment call-up on or off..........................................................................

22

Preselecting a function ..............................................................................................................

23

Setting the vibration adapter .......................................................................................................

24

Setting the weighing process adapter ..........................................................................................

24

Selecting repeatability ................................................................................................................

25

Selecting weighing unit 1 ...........................................................................................................

26

Selecting weighing unit 2 ...........................................................................................................

27

Switching automatic zero correction (Autozero) on or off ...............................................................

28

Contents

5

5.

5.1

5.2

5.3

5.4

5.5

5.6

5.7

5.8

6

6.1

6.2

6.3

6.4

6.5

8.

8.1

8.2

8.3

8.4

7.

7.1

7.2

7.3

4.18

4.19

4.20

4.21

4.22

4.23

4.13

4.14

4.15

4.16

4.17

Preselecting automatic shutdown ...............................................................................................

28

Selecting the power-up mode .....................................................................................................

29

Setting display of the icons ........................................................................................................

30

Selecting peripheral device .........................................................................................................

30

Selecting data transfer mode ......................................................................................................

30

Selecting data transfer format .....................................................................................................

31

Setting baud rate (data transfer rate) ...........................................................................................

32

Setting parity/bit ........................................................................................................................

32

Setting handshake ....................................................................................................................

33

Printing out or saving menu settings ...........................................................................................

33

Canceling secure function ..........................................................................................................

34

Special applications and functions␣ ..........................................................................................

LocalCAN universal interface ......................................................................................................

35

Piece counting .........................................................................................................................

35

Percent weighing ......................................................................................................................

37

Formula weighing .....................................................................................................................

38

Dynamic weighing of unstable weighing samples.........................................................................

42

Below-the-balance weighings ....................................................................................................

44

Adjustment (calibration) with internal weight ................................................................................

45

Adjustment (calibration) with external weights (VariCal) ................................................................

47

Test of the balance with internal weight or external weights ............................................................

49

Further important information␣ ..................................................................................................

52

Error messages ........................................................................................................................

52

Preventive maintenance and care ...............................................................................................

53

Changing the protective cover ....................................................................................................

54

RS232C interface ......................................................................................................................

54

55

Technical data and optional equipment␣ ....................................................................................

56

Technical data of the PG-S balances ...........................................................................................

56

Dimensions .............................................................................................................................

Optional equipment ...................................................................................................................

58

59

Appendix␣ ...............................................................................................................................

61

Overview of menu .....................................................................................................................

61

Conversion table for weight units ................................................................................................

62

SOP (Standard Operating Procedure) ..........................................................................................

63

Index .......................................................................................................................................

65

Getting to know your PG-S balance

6

1. Getting to know your PG-S balance

This section provides you with detailed information on your PG-S balance. Please read through the section carefully even if you already have experience with METTLER TOLEDO balances and scales and be sure to familiarize yourself with the safety notes!

1.1

Introduction

Thank you for deciding to purchase a balance from METTLER TOLEDO.

The precision balances of the PG-S line combine a wide range of weighing functions and setting possibilities with exceptional ease of operation.

Please read through these operating instructions thoroughly so that you can exploit all the possibilities offered by your balance to the full.

These operating instructions apply to all balances of the PG-S line with a readability in the range 0.1 g/1 g. However, the various models have different equipment and performance features. Where this is important for the operation, special mention is made in the text.

1.2

Overview of the PG-S balances

The PG-S balance family comprises various precision balances which differ according to their weighing range, resolution and equipment features.

The models of the PG-S line have the following features:

– Extremely rugged and chemically resistant construction.

– Convenient keypad for one-hand operation and large size, easily readable display.

– FACT (Fully Automatic Calibration Technology), fully automatic, motorized adjustment (calibration) with internal weight.

– Built-in functions for piece counting, percent weighing, formula weighing and dynamic weight determination.

– Built-in RS232C interface.

– Optional LocalCAN universal interface allows the attachment of up to 5 peripheral devices.

– Integrated short-form operating instructions to facilitate your daily work.

– Dynamic graphic indicator (DeltaTrac) for analog display of the weighing range still remaining.

A brief word regarding standards, directives and procedures for quality assurance: Your PG-S balance conforms with all common standards and directives. It supports standard procedures, handicaps, work techniques and records as required by GLP (Good Laboratory Practice) and SOP (Standard Operating Procedure). Recording of the sequences of operations and adjustment work is highly important in this connection: we recommend use of the METTLER TOLEDO LC-P45 Printer here. Your PG-S balance has a CE declaration of conformity and METTLER TOLEDO as the manufacturer has been awarded ISO 9001 certification.

Certified versions of PG-S balances are also available, please ask your nearest METTLER TOLEDO dealer.

Getting to know your PG-S balance

7

1.3

What you should know about these instructions

These instructions contain orientation aids which facilitate your search for the desired information:

Ç

Cal

Key designations are enclosed in double angle brackets

(e.g. «On/Off» or « ±»).

The keys of your PG-S balance have two assignments. The first function of a key

(e.g. « Ç») is always available by pressing the key briefly, whereas the second function (e.g. «Cal.») is called up by pressing and holding the key:

This symbol indicates a brief keystroke.

This symbol indicates a long, sustained keystroke (approx. 2␣ seconds).

long

This representation symbolizes the current display of your balance.

50=1 g

This representation symbolizes a flashing element in the display of your balance.

These symbols indicate safety and hazard instructions which must be complied with. Nonconformance with such instructions can lead to personal injuries to the user, damage to the balance or other tangible assets or to malfunctions.

This symbol indicates additional information and instructions which facilitate your handling of the balance and contribute to proper and economical use.

Getting to know your PG-S balance

8

1.4

Safety has priority

Please note the following instructions for safe and problem-free operation of your PG-S balance.

Read through these operating instructions carefully even if you already have experience with METTLER

TOLEDO balances and scales.

It is essential to note the instructions in section 2 when putting your new balance into operation.

PG-S balances may be used only indoors in closed rooms.

The PG-S balances may not be operated in a hazardous environment and only when attached to receptacle outlets with a ground connection.

Use only the included PG-S power cable.

Operate and use your PG-S balance only in accordance with the instructions in these operating instructions and short-form operating instructions.

Use only optional equipment and peripherals supplied by METTLER TOLEDO with your PG-S balance; these have been optimally matched to your balance.

Your PG-S balance has a very rugged construction, but it is still a precision instrument. Treat it with the appropriate care and it will thank you with years of trouble-free operation.

Never operate the keypad of your PG-S balance with sharp objects!

Never open the balance, it contains no parts which can be maintained, repaired or replaced by the user.

In the unlikely event you experience problems with your balance, please contact your responsible

METTLER TOLEDO dealer.

Putting the balance into operation

9

2. Putting the balance into operation

In this section you will learn how you unpack and set up your new balance and prepare it for operation. On completion of the steps described in this section, your balance is ready for operation.

2.1

Unpacking and checking the standard equipment

PG-S balances are supplied in an environmentally harmless package.

Please check the standard equipment of your balance for completeness:

1

2

– Operating instructions

– Weighing pan (1)

– Pan mount (2)

– Protective cover (3)

– Power cable

– Description of interface commands

(Reference manual MT-SICS, available in English only)

3

PG5001-S

Min

Max

5100g

0.5g

FACT

METTLER TOLEDO

XKD 02

Store all parts of the packaging in a safe place. This packaging guarantees the best possible protection for the transport of your balance.

Putting the balance into operation

10

2.2

Location selection or location change

Your balance is a precision instrument and will thank you for an optimum location with high accuracy and dependability:

Never operate scale in a hazardous environment.

Firm, vibration-free position as horizontal as possible. The foundation must be capable of safely supporting the weight of the fully loaded scale.

No excessive fluctuations in temperature.

No direct sunlight.

No powerful drafts (e.g. from fans).

Putting the balance into operation

11

2.3

Leveling the balance

To assure repeatable weighing results at all times, the balance must be exactly horizontal. To compensate minor unevenness at its location, the balance can be leveled:

Turn the leveling feet of the balance housing until the air bubble is in the center of the level control.

The balance must be releveled each time its location is changed.

2.4

Power supply

On delivery the PG-S balances are set to the line voltage of their country of destination.

First check whether the voltage printed on the model plate of the balance matches your local line voltage. If this is not the case, on no account connect the

balance to the power supply, but contact your METTLER TOLEDO dealer.

Connect the balance to the power supply. Connect the balance only to power receptacles with a ground contact and never use an extension cable without an equipment grounding conductor!

Note:

Route the connection cable so that it does not touch the weighing pan and such that it can not be damaged or hinder your daily operations!

OFF

On

Off

+75 +30

The balance now performs a self-test in which all display segments light up briefly. “OFF” then appears in the display (“OFF” indicates that the balance has been disconnected from the power supply).

Press the «On/Off» key. The display briefly provides information on the installed software version and the normal weight display then appears.

Allow your balance to warm up for 30 minutes to enable it to adapt itself to the ambient conditions.

Putting the balance into operation

12

2.5

Adjusting (calibrating) the balance

An adjustment (i.e. an adjustment to the acceleration due to gravity) is needed when putting into operation for the first time and after every location change. In colloquial language, this operation is frequently also referred to as “calibration”

(to avoid misunderstandings, this term is enclosed in brackets when necessary). You should also adjust (calibrate) your balance at regular intervals in weighing operations to obtain precise results. If you work according to GLP (Good

Laboratory Practice) and SOP (Standard Operating␣ Procedure), please note the stipulated intervals for the adjustment (calibration).

With PG-S balances you have various possibilities for adjusting (calibrating) or checking the balance. You have a choice between

– Adjustment (calibration) or checking the balance,

– internal or external weights,

– automatic or manual initiation of the adjustment operation

The factory setting is fully automatic adjustment (calibration) FACT (Fully Automatic Calibration Technology) with the internal weight. In this setting, you have no need worry about adjusting (calibrating) your balance.

The balance adjusts itself automatically

– after the warm-up phase on connection to the power supply,

– in the case of certified balances, during the warm-up phase (after power failure),

– when a change in the ambient conditions, e.g. the temperature could lead to a noticeable deviation in the measurement.

If your balance is attached to a printer, the adjustment (calibration) is automatically printed out in conformance with GLP. The record opposite is a specimen printed out with the METTLER TOLEDO LC-P45 Printer.

-- BALANCE CALIBRATION -

03.10.97 11:23:34

METTLER TOLEDO

Type: PG5001-S

SNR: 1105238536

SW: 1.75 1.30

Internal Cal. done

Signature:

.........................

--------- END -----------

Weighing made simple

13

3. Weighing made simple

This section shows you how to perform simple weighings, how you can accelerate the weighing process and print out the weighing result and transfer data.

3.1

Switching the balance on and off

Your balance is set in the factory so that it automatically switches to the weighing mode when you load a weight in the standby mode.

On

Off

On

Off

=0 g

OFF

To switch the balance on, press the «On/Off» key briefly. As soon as the normal weight display appears, your balance is ready to weigh.

Note

In section 4.14 you will learn how to perform a display test in which all segments of the display light up briefly when you switch on the balance.

To switch the balance off, press the «On/Off» key and keep it pressed until the message “OFF” appears in the display.

long

After it has been switched off, your balance is in the standby mode. If you wish to perform a weighing, you need now only place the sample on the weighing pan and the balance immediately displays the result. There is no need to switch it on with the «On/Off» key (see also section 4.14).

As your balance needs no warm-up time when in the standby mode and is immediately ready for weighing, we advise you to switch the instrument off only by use of the «On/Off» key and not to disconnect it from the power supply. This also ensures that the balance is always in thermal equilibrium.

Weighing made simple

14

3.2

Zeroing the scale

As a result of environmental influences, your balance may not show exactly «0.0» even though the weighing pan is unloaded. However, you can reset the display of your balance to zero at any time to ensure that the weighing really starts at zero.

The balance does not show exactly zero even though the weighing pan is unloaded.

=2 g

Press the « Ç

Ç

3.3

Taring the balance

The weight of any weighing container can be “tared” at a keystroke to set the display to zero. The taring range covers the entire weighing range of your balance.

If you wish to tare a container, place it on the weighing pan.

Press the « #» key briefly to start the taring operation.

#

------

Net

=0 g

Taring runs automatically. If you tare the balance when it is unstable, the taring procedure will be shown in the display by horizontal segments.

On completion of taring, the zero display and the symbol “Net” for net weight appear. Your balance is again ready for weighing.

When the balance is unloaded, the stored tare value is shown with a negative sign.

The current tare value remains stored until the next taring or zeroing or until the balance is switched off.

Weighing made simple

15

3.4

Performing a simple weighing

Performing a simple weighing is described only for the sake of completeness as this operation comprises only two work steps.

After you have tared the balance, place the weighing sample on the pan.

XKD 04

PG5001-S

METTLER TOLEDO

Wait until the circular symbol of the stability detector fades. Fading of the symbol indicates that the weighing result is stable.

40=6 g

Now read off the weight in the display.

50=5 g

3.5

Weighing with the analog display – DeltaTrac

DeltaTrac is a dynamic graphic indicator which shows the weighing range in use and that still available. You can thus recognize at a glance when the load on the balance approaches the maximum capacity.

0 % in use 15 % in use

55 % in use 95 % in use

Weight within tolerance

Note

You can use the « “» key to switch from the dynamic graphic indicator to a display with two pointers and two tolerance marks (for percent weighing). This enables you to determine the position of the weighing result in regard to the target weight quicker (see section 5.2).

The tolerance range is ±2.5␣ % of the target weight. The setting of this tolerance range is fixed and can be changed only via the interface.

Weighing made simple

16

3.6

DeltaRange

®

balances with movable fine range

METTLER TOLEDO DeltaRange ® balances have a movable fine range with 10 times higher readability. An additional decimal place always appears in the display of this fine range. Thanks to the DeltaRange function, you have the possibility to weigh small amounts of samples into heavy weighing containers.

1 g 1 g

The illustration opposite shows the principle of the movable fine range in which one extra decimal place is shown (in this example, the movable fine range encompasses 2400 grams).

2400 g

0 g 12100 g

After the balance has been switched on, METTLER TOLEDO DeltaRange ® balances operate in the fine range as standard.

=0 g

If the fine range in the display is exceeded, the balance display automatically switches to the lower readability.

4600 g

However, the fine range can always be recalled by taring the balance again.

#

=0 g

3.7

Switching weighing units

Your balance can display the weighing result in two different weighing units. How you preselect the two weighing units is described in sections 4.10 and 4.11.

You can switch between the two weighing units at a keystroke:

The balance shows the result in weighing unit 1.

5200 g

Press the « “» key briefly.

F

%200 kg

Weighing made simple

17

The balance shows the result in weighing unit 2. By pressing the « “» key again, you can return to weighing unit 1.

Notes

If an additional unit (e.g. “%” or “PCS”) is displayed when switching between the two weighing units, you have preselected a function in the menu. You will find further information on the functions in sections 4.6 and 5.1 through 5.4.

You will find a table of the conversion factors between the various weighing units in section 8.2.

3.8

Printing out the weighing result and transferring data

If your balance is connected to a printer via the RS232C interface or via the LocalCAN universal interface, you can transfer current weighing results, identifications and other data to the attached device with a single keystroke.

Menu

56"6 g

Press the « ±» key briefly. As soon as the weighing result is stable, the status indicator of the readability fades and the result is transferred to the attached device.

You will find additional information on the attachment of a printer in sections

6.4 and 6.5 and in the documentation accompanying your printer.

The menu

18

4. The menu

4.1

What is the menu?

The menu allows you to match your balance to your specific weighing needs. In the menu you can change the settings of your balance and activate functions.

SECUrEd The menu contains 20 different options, each of which allows you various selection possibilities.

1. Reset rESEt

2. Adjustment

FACT

3. Automatic adjustm.

call-up

1) 6)

Cal

InFo oFF

4. Function

2)

F nonE

5. Vibration adapter

2

6. Weighing process

adapter

2

7. Repeatability

6ood

14. Peripheral unit

3)

PrinTEr

13. Icons on

15. Data transfer mode

4)

S. off

16. Data transfer

format

4) 5)

S. SICS

12. Power-up mode

1)

Qu. STArT

17. Bau rate

3)

BD 2400

11. Autom. shutdown

ÅoFF -

18. Parity/bit

3)

7B-E

10. Autozero

Å" on

19. Handshake

3)

HS oFF

9. Weighing unit 2

2)

Unit 2 kg

20. Settings

LiST

8. Weighing unit 1

1)

Unit 1 g

1. Reset: Call-up of the factory setting.

2. Adjustment (calibration): Default settings for the type and testing of the adjustment (calibration).

3. Automatic adj. call-up 1) 6) : Switch adjustment call-up to the display on or off.

4. Function 2) :

5. Vibration adapter:

Preselec. of the func. you wish to have available in weighing operation at a keystroke.

Matching the balance to the ambient conditions.

6. Weighing process adapter: Matching the balance to different types of weighing.

7. Repeatability:

8. Weighing unit 1 1) :

9. Weighing unit 2 2) :

10. Autozero:

Selection of the repeatability of the weighing results.

Specification of the 1st weighing unit in which the balance should show the result.

Specification of the 2nd weighing unit in which the balance should show the result.

Switching the automatic zero correction (Autozero) on or off.

11. Automatic shutdown:

12. Power-up mode

13. Icons:

14. Peripheral unit 3)

1) :

:

Preselection of the time after which the balance should be switched off automatically.

Start without or with display test.

On or off switching of the icons.

Attachment to a printer or host.

15. Send commands 4) : Selection of data transfer mode.

16. Send format 4) 5) :

17. Baud rate

18. Parity/bit

3)

3)

19. Handshake

:

:

3) :

Selection of data transfer format.

Matching the transfer speed.

Matching the character format.

Transfer protocol

20. Settings:

1)

2)

3)

4)

5)

6)

Saving or printing out all menu settings.

With certified balances, these menu options have a fixed setting and can not be changed.

With certified balances, only those weighing units/functions allowed by national weights and measures legislation can be selected.

These menu options are shown only if your balance is equipped with an RS232C interface.

These menu options are shown only if “HoSt” has been selected in menu option 14.

These menu options are shown only if “S.oFF” has not been selected in menu option 15.

This menu option is shown only if “FACT” or “CAL oFF” has not been selected in menu option 2.

Note: You will find a graphical overview of the entire menu with all setting possibi-lities in section 8.1.

The menu

19

4.2

Menu operation

In this section you will learn how to work with the menu. Information regarding the individual menu options and the available settings can be found in the following sections.

How to change from the weighing mode to the menu

The balance is operating in the normal weighing mode.

50"6 g

Press the «Menu» key and keep it pressed until the balance switches to the menu.

Menu

Ÿ≈ENU long

After release of the «Menu» key, the balance shows the first option directly (“Reset”) with the current setting.

rESEt

How to select the menu options

Press the « ±» key briefly.

Menu

The next menu option appears in the display. Each time the « ±» key is pressed, the balance switches to the following menu option.

fACT

After the last menu option (“Settings”), the first menu option (“Reset”) is again shown.

rESEt

The menu

20

Unit 1 g

F

Unit 1 kg x times

Unit 1 kg

Menu

StorEd long

%026 kg

How to select the desired setting in a menu option

Press the « “» key briefly. The display shows the next setting available in the selected menu option. Each time the « “» key is pressed, the balance switches to the next setting. After the last setting, the first is again shown.

How to save your settings and quit the menu

After you have made all settings in the individual menu options, press the «Menu» key and keep it pressed until the balance returns to the weighing mode.

Before the normal weighing result display reappears, the balance briefly confirms saving of the settings.

C

=0 g

How to quit the menu without saving your settings

By pressing the «C» key briefly (your attention is drawn to the termination by a double beep), you can return to the weighing mode at any time without chan-

ging the stored settings.

If you do not press a key for 45 seconds, the balance automatically returns to the weighing mode. Changes you have made in the menu are not saved!

The menu

21

4.3

Reset

In this menu option you have the possibility to reset all menu settings to the factory setting.

Resetting settings to factory setting

If you select this option and then save and quit the menu, all menu settings are reset to the values set in the factory.

rESEt long

Before the return to the weighing mode, the resetting is briefly confirmed in the display.

r donE

4.4

Selection of the adjustment (calibration) and test function

Your balance can be adjusted (calibrated) with internal or external weights. The balance can also be checked by a test with internal or external weights. If you have attached a printer to your balance, the data of the adjustment (calibration) and the results of the test are printed out following GLP recommendations.

fACT

The following settings are available:

Fully automatic internal adjustment (calibration) FACT (Fully

Automatic Calibration Technology)

This is the factory setting. The balance adjusts (calibrates) itself fully automatically

– after the warm-up phase following connection to the power supply,

– when a change in the ambient conditions, e.g. the temperature could lead to a noticeable measurement deviation,

– with certified balances, always independent of the settings in menu option

Adjustment.

CAL int

Internal adjustment (calibration)

Adjustment (calibration) is performed at a keystroke with the built-in weight.

UAr∫CAL

Adjustment (calibration) with external weights (VariCal)

Adjustment (calibration) is performed with a selectable* external weight.

* This function is blocked for certified balances.

The menu

22 tESt int tESt E

CAL oFF

Test of the balance with internal weight

In this setting the accuracy test of the balance is performed with the internal weight.

Test of the balance with external weights

The accuracy of the balance can be checked with any external weight.

You will find information on how to perform the adjustment and test function in sections 2.5, 5.6, 5.7 and 5.8.

Switching off the adjustment and test function

The adjustment or test is switched off with the CAL key.

Note

Irrespective of the settings in the menu option Adjustment (Calibration), the fully automatic adjustment FACT is always active with certified balances.

4.5

Switching automatic adjustment call-up on or off

In this menu option you can switch the call-up of the automatic adjustment (calibration) or test on or off.

Note: If you have set «FACT» in the menu option Adjustment (calibration), the automatic adjustment call-up is always active and will thus be skipped in the menu. It becomes active again as soon as «FACT» is switched off.

Cal

InFo on

The following settings are available:

Automatic adjustment (calibration) or test call-up switched on

This is the factory setting. The balance prompts you with a flashing «Cal» in the display to adjust (calibrate) or test it with the internal weight or external weights.

The call-up is triggered by, e.g. temperature changes of the surroundings.

Cal

InFo oFF

Automatic adjustment (calibration) or test call-up switched off

The automatic adjustment or test call-up is switched off.

Note

With the certified versions of balances, call up of the automatic adjustment or test can not be selected, i.e. FACT is always active.

The menu

23

4.6

Preselecting a function

In this menu option you can preselect a function which you then have available in the weighing mode at a keystroke.

Note: With certified balances, only the functions allowed by the national weights and measures legislation are available.

F nonE

The following functions are available:

No function preselected

There is no function available in the weighing mode (factory setting).

PCS

F count

Stk

F count

Piece counting

Your balance counts the pieces which you place in or remove from the weighing container.

F 100 %

ForŸ≈ulA

F dYn A

F dYn Ÿ≈

Percent weighing

Your balance allows you to weigh in to a preselected value or it determines percent weight differences.

Simple formula weighing

The formula weighing function allows you to weigh in up to 255 individual components, store their weights and totalize these. If your balance is connected to a printer, all individual weights and the total weight of all components are printed out. In addition, up to 99 weighing containers can be tared. Your balance can save and print out the total weight of all weighing containers.

Dynamic weighing with automatic start

Your balance determines an average weighing result over a preset time interval. This setting is suitable for unstable weighing samples (e.g. animals). With this setting, the dynamic weighing starts automatically.

Dynamic weighing with manual start

Analogous to dynamic weighing with automatic start, but the weighing cycle must be started manually.

You will find details on how to work with the functions in section 5.

The menu

24

4.7

Setting the vibration adapter

The vibration adapter is used to match the balance to the ambient conditions (vibrations, drafts at balance location).

2

The following settings are available:

Setting for normal ambient conditions

This is the factory setting matched to normal ambient conditions.

3

Setting for unstable surroundings

The filter setting of the balance is higher than in the factory setting, but the balance is less sensitive to external influences.

1

Setting for virtually disturbance-free, stable surroundings

The filter setting of the balance is lower than in the factory setting, but the balance is more sensitive to external influences.

4.8

Setting the weighing process adapter

The weighing process adapter is used to match your balance to the different types of weighings (absolute weighing, fine addition, etc.).

The following settings are available:

2

3

oFF

Universal setting

This is the factory setting, it is suitable for all types of weighings. The display always corresponds to the current weight.

Absolute weighing

This setting is suitable for checkweighings and for the weight determination of samples.

Special applications

In this setting, the displayed weight value has a fixed relation to the weight change with time.

1

The menu

25

Fine addition

This setting is suitable for the weighing in of fine powders, small amounts of liquids, etc.

4.9

Selecting repeatability

In the left lower corner of the display you will find the circular symbol of the stability detector. As soon as the weighing result is within specified limit values over a certain time interval, the weighing result is considered stable and the symbol for the stability detector fades. With the setting for the repeatability (“Repro-Set”), you determine the time interval over which the result must lie within the limit values for it to be considered stable. The better the repeatability, the longer the weighing process lasts.

The following settings are available:

Good

Good repeatability

The weight display is released as stable quickly, this is the factory setting.

bEttEr

Very good repeatability

Slower release until stable weight display.

bESt

Best possible repeatability

Stable weight display is not released until several seconds without change.

Std

Normal repeatability

The weight display is released as stable very quickly, in other words:

The display for the stability detector fades very quickly.

The menu

26

4.10 Selecting weighing unit 1

In this menu option you specify the unit* in which the weighing result should be displayed.

Unit 1 g

The following units * are available:

GN dwt ct mo m lb oz ozt

Display g kg

Designation gram kilogram pound ounce troy ounce grain pennyweight carat momme mesghal

You will find a table with the conversion factors for the different units in section

8.2 of these operating instructions.

* With certified balances, the weighing unit has a fixed setting g (gram) and can not be changed.

The menu

27

4.11 Selecting weighing unit 2

In this menu option you specify the additional unit* in which the weighing result should be displayed.

Unit 2 g

The following units are available: m

H tl

S tl t tl

GN dwt ct mo lb oz ozt

Display g kg

Designation gram kilogram pound ounce troy ounce grain pennyweight carat momme mesghal

Hong Kong taels

Singapore taels

Taiwan taels

You will find a table with the conversion factors for the different units in section

8.2 of these operating instructions.

* With certified balances, only those weighing units allowed by the respective national weights and measures legislation can be selected.

The menu

28

4.12 Switching automatic zero correction (Autozero) on or off

This menu option allows you to switch the automatic zero correction on or off. When it is switched on (factory setting), the zero point is automatically corrected for drift or contamination of the weighing pan.

A" on

The following settings are available:

Autozero switched on

This is the factory setting. The zero point is automatically corrected.

A" oFF

Autozero switched off

The zero point is not automatically corrected. This setting is advantageous for special applications (e.g. evaporation measurements).

Note

With certified balances, this setting is possible only with a resolution of 10 d.

4.13 Preselecting automatic shutdown

When the automatic shutdown is activated, the balance automatically switches itself off after a preselected time (calculated from the last operation) and is switched to the standby mode.

ÅoFF -

The following settings are available:

No automatic shutdown

The automatic shutdown is deactivated (factory setting).

ÅoFF 2`

Automatic shutdown after 2 minutes

As soon as the balance has not been operated for 2 minutes, it switches itself off automatically.

ÅoFF 5`

ÅoFF 10`

The menu

29

Automatic shutdown after 5 minutes

As soon as the balance has not been operated for 5 minutes, it switches itself off automatically.

Automatic shutdown after 10 minutes

As soon as the balance has not been operated for 10 minutes, it switches itself off automatically.

4.14 Selecting the power-up mode

You can set your balance such that it either immediately starts from the standby mode when you load a weight or it must be switched on with the «On/Off» key after which it then performs a display test.

qÙ StArt

The following settings are available:

Quickstart*

This is the factory setting. The balance can be started directly from the standby mode and is immediately ready for weighing. You can load the weight in the standby mode and the balance immediately shows the current weighing result.

* Quickstart is not possible with certified balances.

FÙ StArt

Start with display test

You must switch on the balance with the «On/Off» key. After it has been switched on, it performs a display test in which all display elements light up. On completion of the test, the balance is ready for weighing.

Note

If the balance has been disconnected from the power supply, it always performs a display test after it has been switched on again, even if the “Quickstart” setting has been selected.

The menu

30

4.15 Setting display of the icons

All icons appear in the display.

on

AuTo oFF

If desired, you can also switch off the icons. They disappear after about 10 seconds after you have quit the menu or after about 3 min. after the balance has been switched on.

Note: The menu options 4.16–4.21 are shown only if your balance is equipped with an RS232C interface.

4.16 Selecting peripheral device

In this menu option you can select the desired peripheral unit. The balance stores the corresponding settings (4.17–

4.21) for each peripheral unit separately.

Attachment to a printer (e.g. METTLER TOLEDO LC-P45 Printer).

Factory setting : bd 2400, 7b–E, HS OFF

PrintEr

Attachment to any peripheral unit.

Factory setting: S. oFF, bd 9600, 8b–no, HS Soft

Host

4.17 Selecting data transfer mode

In this menu block you tell the balance how a value should be transferred to a peripheral device (e.g. LC-P45). This menu option appears only if the setting “HoSt” has been selected in the menu option “Selecting peripheral device” (section

4.16).

Data transfer mode switched off

S. oFF

The next possible stable value will be transferred after triggering of the Print/

Transfer command.

S. STb

S. ALL

The menu

31

The current value will be transferred after triggering of the Print/Transfer command.

Only stable values will be automatically transferred after every weight change.

S. AuTo

S. ConT

All values will be automatically transferred.

Note

The menu option “Data transfer mode” appears only if “HoSt” has been selected in the menu option “Selecting peripheral device” (see also section 8.1).

4.18 Selecting data transfer format

S. SICS

With this setting the data transfer formats described in MT-SICS are used. You will find the description in the enclosed description of the operating instructions

(Reference manual MT-SICS).

With this setting the data transfer formats of the PM balances are used.

S. PŸ≈

Note

If you wish to use other data transfer formats of the PM balances with your

PG-S balance, please use the optional R/G-M emulation software cassette which emulates all interface commands of the PM balances (see section 7.3).

The interface is unidirectional. Incoming interface commands in the setting “S.␣ PM” are not processed further.

The menu option “Selecting data transfer format” appears only if the setting

“S. oFF” has not been selected in the menu option “Selecting data transfer mode”

(see also section 8.1).

The menu

32

4.19 Setting baud rate (data transfer rate)

The data transfer rate (baud rate) determines the speed of the transfer via the serial interface. The unit is baud (1 baud

(bd) = 1␣ bit/second).

The following settings are available:

150 bd, 300 bd, 600 bd, 1200 bd, 2400 bd, 4800 bd and 9600 bd bd 2400 bd 1200

4.20 Setting parity/bit

In this menu option you can set the character format for the attached peripheral unit.

The following settings are available:

7 bit/even parity

7B-E

7 bit/no parity

7B-mo

8 bit/no parity

8B-mo

7 bit/odd parity

7B-odd

The menu

33

4.21 Setting handshake

In this menu option the data transfer can be matched to various serial receivers.

The following settings are available:

No handshake

HS oFF

Software handshake (XON/XOFF)

HS SoFt

HS HArd

Hardware handshake

Note

If you have selected this setting, the attached peripheral unit must be switched on. If the peripheral unit is switched off, the balance will be blocked.

4.22 Printing out or saving menu settings

In this menu option you have the possibility to save all menu settings. You can also print out all current settings of the menu assuming your balance is connected to a printer.

As soon as you save your settings and quit the menu, all settings defined in the menu are recorded on the attached printer.

LiSt

With “secure 1” you can protect the menu settings against inadvertent changes.

SECUrE 1

SECUrE 2

With “secure 2” you can protect both the menu settings and also the Ç key, which triggers the adjustment function or lowers the readability of the display, against inadvertent changes.

Note

If the adjustment function “FACT” is set in the menu option, the PG-S balance also automatically performs an internal adjustment in the setting “secure 2”.

The menu

34

-------- LIST ----------

03.10.97 16:49:06

METTLER TOLEDO

Balance

Type: PG5001-S

SNR: 1113170358

SW-Ver.: 1.75 1.30

Cal: FACT

Function: none

Vibr.: 2

Wghpro.: 2

Repro.: good

Unit 1: g

Unit 2: kg

A.Zero: on

A.off: -

Start: Qu. Start

Icons: on

Output: Printer

Baud: 2400

Bit: 7

Parity: even

Handshake: off

--------- END ----------

The record shown opposite is a specimen which can be printed out, e.g. with the METTLER TOLEDO LC-P45 Printer. Depending on the selected settings and the selected connection with LocalCAN or RS232 interface, the representation may differ from the example shown.

4.23 Canceling secure function

Step 1

Menu

If “secure” is selected in the menu, “secure” appears when it is reentered (initiated by the menu key). If you do not press the « “» key for more than 3 seconds, the balance automatically returns to the weighing mode (menu remains blokked).

long

SECUrEd

After the « “» key has been pressed, “Open” appears. Confirm this within 3␣ seconds by pressing and holding the menu key, entry into the menu is then possible again (menu open).

F

Step 2

Note

The release applies to “SECUrE 1” and “SECUrE 2”.

OPEn

Step 3

Menu long

Special applications and functions

35

5. Special applications and functions

Your balance can do more than just weigh. Built-in applications and functions expand the possibilities of your balance and facilitate your daily work. The following sections will acquaint you with these applications and functions.

5.1

Piece counting

Piece counting presupposes that you have preselected the function “F count” in the menu (see section 4.6).

XKD 05

Load the empty container.

PG5001-S

METTLER TOLEDO

Press the « #» key to tare the balance.

#

=0 g

Your balance now needs the weight of a reference piece number. Press and hold the «F» key until you are prompted to load the reference pieces.

F

SEt 10

Stk long

F

SEt 10

Stk

SEt 20

Stk

Your balance suggest “10” as reference piece number. You can accept this suggestion or select one of the the available reference piece numbers (20, 30,

50, 100 or 5 pieces ) by briefly pressing the « “» key.

Note

We advise you to select a reference piece number as large as possible as the balance determines the average weight per piece and saves it as the reference weight. As it is seldom the case that all pieces have the exactly the same weight, the accuracy of the reference weight increases with increasing reference piece number.

Special applications and functions

36

XKD 06

Now load the selected number of reference pieces.

PG5001-S

METTLER TOL

EDO

Menu

------

Then briefly press the « ±» key. As long as the horizontal dashes are displayed, the balance is calculating the reference weight.

Note

If you do not press a key for 45 seconds, the balance returns to the weighing mode.

After your balance has determined the reference weight, it shows the correct piece number and is now ready for piece counting.

20

Stk

F

55=6 g

You can use the « “» key at any time to switch the display between the piece number display, weighing unit 1 and weighing unit 2.

Note

The current set weight remains stored until it is redetermined or the power supply of the balance is interrupted.

F

=5506 kg

---- PIECE COUNTING ----

APW 0.19990000 g

Out of: 100 Stk

100 Stk

Net 20.0 g

--------- END ----------

If a printer is connected to your balance, the reference weight, the reference piece number, the total piece count as well as the net weight of the total piece count are printed out.

Ç

0

Stk

Note

If a printer is attached, you can start a new piece counting with the « Ç» key.

Special applications and functions

37

5.2

Percent weighing

The “Percent weighing” function allows you to weigh in to a preset value (100 %) and determine the deviations from this target value. The DeltaTrac helps you quickly determine the position of the sample weight regarding the tolerances (see section 3.4).

Percent weighing presupposes that you have preselected the function “F 100 %” in the menu (see section 4.6).

XKD 07 If you are working with a weighing container, place this on the pan and tare the balance with the « #» key.

PG5001-S

METTLER TOL

EDO

F

SEt 100 %

Your balance needs a reference weight that should correspond to 100%. Press and hold the «F» key until you are prompted to load the reference weight. Now load the reference weight.

long

Menu

------

Then briefly press the « ±» key. As long as the horizontal dashes are shown, your balance is calculating the reference weight.

Note

If you do not press a key for 45 seconds, the balance returns to the weighing mode.

On completion of the weighing-in procedure, your balance is ready for percent weighing.

10=0 %

F

121*5 g

You can use the « “» key at any time to switch the display between percent display, weighing unit 1 and weighing unit 2.

Note

The current piece weight remains stored until it is redetermined or the power supply of the balance is interrupted.

F

+2135 kg

The position of the sample weight regarding the tolerances can be quickly determined with the DeltaTrac. The taring range is ±2.5␣ % of the target weight.

The setting of this tolerance range is fixed and can be changed only via the interface.

Weight within tolerance

Special applications and functions

38

5.3

Formula weighing

With the formula weighing function you can weigh and totalize individual weights (components). Your balance processes up to 255 components per formula weighing operation. In addition, you can tare up to 99 weighing containers for each formula weighing operation. If your balance is connected to a printer, the entire formula weighing operation can be recorded.

Formula weighing presupposes that the function “Formula” has been preselected in the menu (see section 4.6).

F

ForŸ≈ulA

Unload the weighing pan.

Press the « “» key briefly and the display confirms that the formula weighing function is active.

After 2 seconds, the normal weight display appears.

=0 g

XKD 07

If you wish to tare a weighing container, place this on the pan.

PG5001-S

#

METTLER TOL

EDO

Net

=0 g

Then press the « #» key briefly.

----- FORMULATION ------

T 1 100.8 g

If your balance is connected to a printer, the tare weight is printed out.

XKD 04

Special applications and functions

39

Add the first container to the weighing container.

PG5001-S

METTLER TOLEDO

F

Then press the « “» key briefly. The display shows “- 1 -” briefly to confirm the weighing in of the first component.

- 1 -

Net

=0 g

----- FORMULATION ------

T 1 100.2 g

1 Comp. 12.0 g

After weighing in of the first component, the display is reset to zero and the balance is now ready for weighing in of the second component.

If a printer is attached, the weight of the components is printed out.

Menu

Net

T

17(6 g

Now weigh in the additional components as described above.

As soon as you have weighed in all components, press the « ±» key briefly to end the formula weighing operation. The total weight of all individual components is displayed briefly.

The balance then returns to the normal weighing mode.

=0 g

The weight memories for tare and net total are now cleared and the balance is ready for the next formula.

Special applications and functions

40

----- FORMULATION ------

T 1 100.2 g

1 Comp. 12.0 g

2 Comp. 2.5 g

3 Comp. 3.3 g

T total 100.2 g

G 118.1 g

N total 17.8 g

--------- END ---------

If a printer is attached to your balance, a record with the net total weight of all components “N total”, the tare weight (weight of the weighing container) “T total” and the total weight (total weight of all components plus tare weight) “G” is printed out.

F

Net

T

1/8 g

During the formula weighing operation you can increase the net total weight to a desired value

Press and hold the «F» key until the net total weight of all components weighed in so far is displayed.

long

XKD 08

Now add the component to the container until the desired net total weight is reached.

PG5001-S

F

METTLER TOL

EDO

Net

T

2%0 g

Press the « “» key briefly and the desired weight is confirmed as an additional component.

F

Net

T

1/8 g

During the formula weighing operation you can always display the totalized total weight and the number of components weighed in so far

Press and hold the «F» key until the weight of all components weighed in so far is displayed.

long

F

Special applications and functions

41

Press and hold the «F» key again until the number “n” of all components weighed in so far is displayed.

n 3

F long

Net

=0 g

Press and hold the «F» key until the balance switches back to the weight display.

You can now weigh in additional components.

long

XKG 09

During the formula weighing operation you can always tare additional weighing containers

Place the additional weighing container on the weighing pan next to weighing containers already tared.

PG5001-S

METTLER T

OL

#

43"1 g

Net

=0 g

Press the « #» key briefly. The balance is now tared with the additional weight of the new weighing container. If your balance is connected to a printer, the tare weight of the new container is printed out. You can now weigh in additional components.

T 2 432.1 g

T total 1434.9 g

G 1613.5 g

N total 178.6 g

--------- END ---------

If you print out the results at the end of the formula weighing operation, all tare weights are totalized and the total weight of all tare containers (“T total”) is recorded.

Special applications and functions

42

5.4

Dynamic weighing of unstable weighing samples

The functions “Dynamic weighing with automatic start” and “Dynamic weighing with manual start” facilitate your weighing of unstable weighing samples (e.g. animals). With this type of weighing, your balance determines the weight over a particular time period and calculates a representative mean value.

Dynamic weighing presupposes that you have preselected the function “F dyn A” or “F dyn M” in the menu (see section␣ 4.6).

XKD 07 If you are working with a weighing container, place it on the weighing pan in the normal weighing mode.

PG5001-S

METTLER TOL

EDO

1ç4 g

Press the « #» key to tare the balance.

#

=0 g

F

=0 g

Press the « “» key briefly. The symbol of the weighing process adapter in the display confirms that dynamic weighing has been activated.

Your balance is set in the factory so that the weight is determined over a period of 3 seconds. You need perform the following 3 steps only if you wish to change this time interval.

Press and hold the «F» key until the time display appears.

F

t ≠ 3” long

F

Menu

t ≠ 3”

t ≠ 5”

=0 g

Special applications and functions

43

By pressing the « “» key briefly, you can select one of the available time intervals (1, 2, 3, 5, 10 or 20 seconds).

Notes

The more unstable the weighing sample, the longer the time interval which should be selected.

If you do not press a key for 45 seconds, the balance quits the display without changing the inputted value.

Then press the « ±» key briefly to confirm the selected time interval.

Your balance is now ready for dynamic weighing.

Load the weighing sample.

9%9 g

If you have selected the function “Dynamic weighing with automatic start” in the menu, the weighing starts automatically on relative stability. However, the

weighing sample must weigh at least 5 grams.

If you have selected the function “Dynamic weighing with manual start” in the menu, press the « ±» key briefly to start the weighing.

The remaining weighing time (in seconds) is displayed continuously.

Menu

-- 5 --

-- 1 --

‹ 9%9 g

On elapse of the weighing time, read off the result. The asterisk symbol “ * ” lights up in the lower left corner of the display. This symbol indicates that the value is the mean value of the performed weighings, in other words a calculated result.

The result remains in the display until the weighing sample is removed. If you wish to weigh the same weighing sample again, press the « ±» key briefly.

Special applications and functions

44

The set weighing time (time interval) remains stored until it is changed or the power supply of the balance is interrupted.

By briefly pressing the « “» key, you can switch between the normal weighing mode and dynamic weighing at any time.

By pressing and holding the «F» key when in the dynamic weighing mode, you can recall the preselected time interval to the display and change it.

5.5

Below-the-balance weighings

Your PG-S balance is equipped with a hanger for weighings below the weighing pan.

Carefully turn the balance over and detach the cover of the hanger (use a screwdriver if necessary). Store the cover in a safe place to ensure it is not lost.

Fasten a suitable suspension device to the hook of the hanger to accommodate the weighing sample. You can use the hanger to perform weighings up to the nominal load of your balance model.

Note

On completion of the below-the-balance weighing, re-attach the cover of the hanger.

Special applications and functions

45

5.6

Adjustment (calibration) with internal weight

Depending on the setting selected in the menu (see section 4.4), the adjustment (calibration) can be performed with the built-in, internal weight fully automatically (FACT) or semi-automatically.

Ç

Cal long

CAL int

Fully automatic internal adjustment (calibration) FACT

Your balance is set in the factory for the fully automatic adjustment with the internal adjustment weight. You are already familiar with this setting from sections 2.5 and 4.4.

Semi-automatic adjustment (calibration)

If your balance is outside the adjustment tolerance and depending on whether you have set the automatic adjustment call-up in the menu (see section 4.5), the balance uses a flashing «Cal» in the display to prompt you to adjust

(calibrate) with the internal weight at a keystroke. With certified balances, the adjustment (calibration) with the internal weight is performed automatically in accordance with the national weights and measures legislation. An adjustment

(calibration) with external weights is not allowed by the weights and measures legislation.

If you wish to adjust your balance with an internal weight, proceed as follows:

Make sure that “FACT” or the “Adjustment (calibration) with internal weight

(Cal int)” is selected in the menu (see section 4.4).

Ensure that the weighing pan is unloaded. There is no need to tare the balance before the adjustment (calibration).

Start the adjustment operation by pressing and holding the «Cal» key. The balance briefly shows that adjustment (calibration) is being performed with the internal weight.

Note

If “SECUrEd 2” is switched on in the menu, the Ç

Cal key is blocked.

Special applications and functions

46

The following displays appear during the adjustment (calibration):

The internal adjustment weight is being loaded.

Cal

------

Cal

=0

Cal

------

The internal adjustment weight is being raised.

The balance is processing the adjustment results.

The balance reports successful completion of the adjustment (calibration).

CAL donE

The balance automatically returns to the weighing mode.

=0 g

C

Abort

-- BALANCE CALIBRATION -

03.10.97 11:23:34

METTLER TOLEDO

Type: PG3001-S

SNR: 1105238536

SW: 1.75 1.30

Internal Cal. done

Signature:

.........................

--------- END -----------

You can always abort an ongoing adjustment (calibration) by briefly pressing the «C» key (double beep).

If the adjustment (calibration) can not be performed properly (e.g. as a result of vibrations), the balance aborts the adjustment operation and “Abort” appears in the display. Press the «C» key to clear this message and restart the adjustment operation.

If your balance is connected to a printer, the adjustment (calibration) is recorded automatically in conformance with GLP. The record shown opposite is a specimen printed with the METTLER TOLEDO LC-P45 Printer. Depending on the attached printer, the printout may differ somewhat from the example shown.

Special applications and functions

47

5.7

Adjustment (calibration) with external weights (VariCal)

Depending on the setting selected in the menu (see section 4.4), the adjustment (calibration) can be performed with the built-in weight or with external weights. In the factory setting, the balance is set to adjustment with the internal weight, which you are already familiar with from section 2.5.

Ç

Cal

UAr∫ CAL

If you wish to adjust your balance with external weights, proceed as follows:*

* The adjustment (calibration) with external weights is not possible with certified balances.

Make sure that “Adjustment (calibration) with external weights (VariCal)”

is selected in the menu (see section 4.4).

Ensure that the weighing pan is unloaded. There is no need to tare the balance before the adjustment (calibration).

Start the adjustment operation by pressing and holding the «Cal» key. The balance shows briefly that adjustment is being performed with external weights.

long

The balance now prompts you to select the desired weight. If you do not press a key for 3 seconds, the balance automatically initiates the adjustment process.

CAL 1000 g

F

CAL 2000 g

If you do not wish to adjust with the suggested weight, you can select a different weight by pressing the « “» key briefly. The available weights depend on the balance model.

Menu

Confirm the selected weight with the « ±» key. This initiates the adjustment process. The balance determines the zero point.

Cal

------

Cal

200=0 g

You are then prompted to load the weight.

Special applications and functions

48

XKD 10 Place the requested weight in the middle of the weighing pan.

PG5001-S

METTLER TOLEDO

Cal

------

During the adjustment the horizontal segments are displayed.

Note

You can abort the ongoing adjustment at any time by pressing the «C» key briefly.

On completion of the adjustment operation, you are prompted to remove the weight. Lift the weight off the weighing pan.

=0 g

After removal of the weight, the balance shows the end of the adjustment operation and then returns to the weighing mode.

CAL donE

C

Abort

-- BALANCE CALIBRATION -

03.10.97 11:34:23

METTLER TOLED

Type: PG5001-S

SNR: 1105238536

SW: 1.75 1.30

Weight ID: ..............

Weight: 1000.0 g

External Cal. done

Signature:

.........................

--------- END -----------

Note

If the adjustment (calibration) can not be performed properly (e.g. as a result of vibrations), the balance aborts the adjustment operation and “Abort” appears in the display. Press the «C» key to clear this message and restart the adjustment operation.

If your balance is connected to a printer, the adjustment is recorded automatically in conformance with GLP. The record shown opposite is a specimen printed with the METTLER TOLEDO LC-P45 Printer. Depending on the attached printer, the printout may differ somewhat from the example shown.

Special applications and functions

49

5.8

Test of the balance with internal weight or external weights

You can always test the accuracy of your balance. This test is performed either with the built-in weight or with external weights, depending on your setting in the menu (see section 4.4).

Ç

Cal tESt int

Test of the balance with the internal weight

Make sure that the “Test of the balance with the internal weight” (test int) is

selected in the menu (see section 4.4).

Ensure that the weighing pan is unloaded. There is no need to tare the balance before the test.

Initiate the test procedure by pressing and holding the «Cal» key. The balance briefly confirms that the test is being performed with the internal weight.

long

The following displays appear during the test:

------

The balance determines the zero point.

=0

The balance is processing the results of the test.

------

The balance confirms that the test has been performed.

tSt donE

d =2

Over a period of 10 seconds, the balance now shows the difference (deviation) between the adjustment (calibration) and the current test weighing.

On completion of the test, the balance automatically returns to the weighing mode.

Special applications and functions

50

C

Abort

Notes

You can always abort an ongoing test by pressing the «C» key briefly.

If the test can not be performed properly (e.g. as a result of vibrations), the balance aborts the operation and “Abort” appears in the display. Press the «C» key (double beep) to clear this message and restart the test.

----- BALANCE TEST -----

03.10.97 11:34:23

METTLER TOLEDO

Type: PG12001-S

SNR: 1105238536

SW: 1.75 1.30

Target: 800.0

Actual: 800.2

Diff: 0.2

Internal test done

Signature:

........................

--------- END ----------

If your balance is connected to a printer, the measured difference is recorded automatically. The record shown opposite is a specimen printed with the

METTLER TOLEDO LC-P45 Printer. Depending on the attached printer, the printout may differ somewhat from the example shown.

Test of the balance with external weights

Make sure that the “Test of the balance with external weights” (test E) is

selected in the menu (see section 4.4).

Ensure that the weighing pan is unloaded. There is no need to tare the balance before the test.

Initiate the test procedure by pressing and holding the «Cal» key. The balance briefly confirms that the test is being performed with an external weight.

Ç tESt E long

The balance prompts you to load the external weight. Place your weight on the pan.

LoAd

Special applications and functions

51

During the test, the horizontal segments are displayed.

------

The balance now prompts you to remove your weight. Lift off the weight.

=0 g

After removal of the weight, the balance processes the results of the test.

------

The balance confirms that the test has been performed and then automatically returns to the weighing mode.

tSt donE

C

Abort

----- BALANCE TEST -----

03.10.97 15:21:17

METTLER TOLEDO

Type: PG5001-S

SNR: 1105238536

SW: 1.75 1.30

Weight ID: .............

Target: .............

Actual: 2000.0 g

Diff: .............

External test done

Signature:

........................

--------- END ----------

Notes

You can always abort an ongoing test by pressing the «C» key briefly.

If the test can not be performed properly (e.g. as a result of vibrations), the balance aborts the operation and “Abort” appears in the display. Press the «C» key (double beep) to clear this message and restart the test.

If your balance is connected to a printer, the measured weight of the external test weight is recorded automatically. You can now enter the target weight (“Target”) and the difference (“Diff”) in the record by hand. The record shown opposite is a specimen printed out with the METTLER TOLEDO LC-P45 Printer. Depending on the attached printer, the printout may differ somewhat from the example shown.

Further important information regarding your PG-S balance

52

6 Further important information

6.1

Error messages

Error messages in the display draw your attention to incorrect operation or that the balance could not execute a procedure properly.

Error message Cause

Overload

Rectification

Unload weighing pan

íååååì

ñ----ó

Underload Check that the weighing pan is positioned correctly.

Preselect desired function in the menu.

No function preselected

F nonE

Error 1

No stability

– On taring or adjustment (calibration)

– On loading the reference weight for the functions “Piece counting” or “Percent weighing”.

Ensure more stable ambient conditions. If not possible, check settings of the repeatability and vibration adapter (see sections

4.9 and 4.7).

No or wrong adjustment weight Load requested weight.

Error 2

Error 3

Error 4

Faulty reference

(Reference weight or reference piece number too low)

Increase reference weight or reference piece number.

Internal fault

Error Nos 4, 5, 6 and 7 signal internal balance problems

Inform METTLER TOLEDO customer service

Error 7

Error message

Error 8

PrESS 0

=0

Abort

Balance blocked

Further important information regarding your PG-S balance

53

Cause Rectification

Error in zero setting (certifiable balances only)

Zero setting outside zero setting range

Ensure that zero setting is performed in the admissible range of ±2% of the full load and not on underload or overload.

Negative taring range

The balance has been tared, the tare removed (negative tare weight) and an attempt has been made to retare the balance

Reset balance to zero with the « Ç» key and repeat taring operation

Wrong or missing weighing pan Mount correct weighing pan.

Unload weighing pan.

Adjustment (calibration) or test could not be performed properly.

The balance aborts the procedure. This error message is caused by external disturbing influences (e.g. vibrations or powerful drafts).

Press «C» key (a double beep sounds as confirmation) to clear the error message.

Close all draft shield doors.

Possibly select a more suitable location for the balance.

In the menu option Handshake, setting “HS

Hard” selected.

Switch on attached peripheral device.

6.2

Preventive maintenance and care

Servicing

Regular servicing of your balance by an authorized service engineer ensures constant accuracy for years to come and prolongs the lifetime of the instrument. Ask your METTLER TOLEDO dealer for details of the available service options.

Cleaning

The balance housing and the weighing pan are made of high-grade, resistant materials. All commercially available cleaning agents may thus be used for cleaning.

PG-S balances can best be cleaned with a damp cloth.

PG-S 13

Further important information regarding your PG-S balance

54

6.3

Changing the protective cover

If you operate your balance in an environment liable to cause contamination, we recommend you cover it with the supplied transparent protective cover for the keypad and display.

Contaminated protective covers of all balance models can be changed, see

Optional equipment in section 7.3.

Cleaning beneath the weighing pan

– Remove the weighing pan and the pan mount.

– You can now clean the dirty parts under running water.

PG5001-S

Max

Min 5100g

0.5g

FACT

METTLER TOLEDO

XKD 03

6.4

RS232C interface

SGND

5

9

1)

2)

Ready to send

Ready to receive

DATA

RxD

TxD

«HAND-SHAKE»

IN

1)

OUT

2)

6

1

PG-S 16

Each PG-S balance is equipped with an RS232C interface as standard for the attachment of a peripheral device (e.g. printer or computer).

– 9-pin socket

– Matching to the other device (transfer parameters) is possible via the menu settings (see sections 4.16–4.21)

The versatile features of the PG-S balances regarding documentation of the results can not be utilized to the full until a printer, e.g. the LC-P45 from METTLER

TOLEDO is attached. The printed results make a decisive contribution to a simple working procedure following GLP/GMP.

Notes

If you wish to attach several peripheral devices at the same time, it is simple to retrofit your PG-S balance with the LocalCAN universal interface (see section

6.5).

You will find a detailed description of the available interface commands in the enclosed brochure “Reference manual MT-SICS”.

Further important information regarding your PG-S balance

55

6.5

LocalCAN universal interface

As an option, your PG-S balance can be retrofitted with a LocalCAN universal interface. As you can attach up to five peripheral units at the same time, it offers you a high degree of flexibility in data interchange.

Peripheral units (see section 7.3) from METTLER TOLEDO which have the connection cable as part of their standard equipment can be attached in a simple manner to the balance.

The communication is particularly well supported by the commands of the standard and extended command set. The reference manual (705184) that you received with the LC-RS or LC-CL cable describes the functioning of these commands in an easily surveyed manner.

XKD 11

The features and benefits of the LocalCAN universal interface can be summarized as follows:

– Attachment of up to five peripheral units to a balance at the same time.

– Support of standard interfaces such as RS232C or CL.

– Rugged, 4-pin connector with reversed voltage and pullout protection.

– Dependable data transfer thanks to built-in CAN controller.

– Open cabling system, i.e. each peripheral unit except auxiliary displays have an additional connection.

– Simple configuration of the parameters without recourse to the operating instructions of the PG-S balance.

The versatile features of the PG-S balances regarding documentation of the results can not be utilized to the full until a printer, e.g. the LC-P45 from METTLER

TOLEDO is attached. The printed results make a decisive contribution to a simple working procedure following GLP/GMP.

2

3

1

4

Technical data of the LocalCAN universal interface

Cable length between two devices, maximum 10 m.

Total cable lengths of all attached devices, maximum 15 m.

Pin assignment (balance end)

Pin No.

Signal

1

2 negative signal line (–CAN) positive signal line (+CAN)

3

4 plus pin of power supply (V CAN) for peripherals minus pin of power supply (0 V) for peripherals

Technical data and optional equipment

56

7. Technical data and optional equipment

7.1 Technical data of the PG-S balances

Power supply

100…240 V, 50/60 Hz, 9…20VA

Ambient conditions for PG-S balances

Height above sea leve up to 4000 m

Temperature 5–40º C

Atmospheric humidity max. 85% RH @ +30° C

Overvoltage category II

Pollution degree 2

Standard equipment

• RS232C interface

• Protective cover

• Feedthrough for below-the-balance weighing

• Device for theft protection

• Device for stand fastening

• Operating instructions

• Description of interface commands (Reference manual MT-SICS)

Technical data

Readability

Maximum capacity

Taring range

Repeatability (s)

Linearity 1)

Stabilization time (typical)

Adjustment

• with internal weight

• with external weights

Sensitivity

• Temperature drift 1) 2)

• Long-term drift 1) 3)

Weighing pan

Dimensions (w/h/d)

Net weight

PG3001-S PG5001-S

0.1 g

3100 g

0 ... 3100 g

0.1 g

5100 g

0 ... 5100 g

0.08 g

±0.1 g

0.08 g

±0.1 g

1…2 s 1…2 s internal, fully automatic motorized initiation (FACT) test possibility for checking the sensitivity

2 / 3 kg 2 / 3 / 4 / 5 kg

±10 ppm / °C

±0.007 %

240x240 mm

240x100x366 mm

6 kg

240x240 mm

240x100x366 mm

6 kg

Technical data and optional equipment

57

PG8001-S

0.1 g

8100 g

0 ... 8100 g

0.08 g

±0.1 g

1…2 s

4 / 5 / 6 / 7 / 8 kg

240x240 mm

240x100x366 mm

6 kg

Technical data PG8000-S PG12001-S

Readability

Maximum capacity

Fine range

Taring range

Repeatability (s)

Linearity 1)

Stabilization time (typical)

Adjustment

• with internal weight

• with external weights

Sensitivity

• Temperature drift 1) 2)

• Long-term drift 1) 3)

Weighing pan

Dimensions (w/h/d)

Net weight

1 g

8100 g

0 ... 8100 g

0.3 g

±0.5 g

0.1 g

12100 g

0 ... 12100 g

0.08 g

±0.1 g

1…2 s 1…2 s internal, fully automatic motorized initiation (FACT) test possibility for checking the sensitivity

4 / 5 / 6 / 7 / 8 kg 4 / 5 / 6 / 7 / 8 / 9 / 10 / 11 / 12 kg

±10 ppm / °C

±0.007 %

240x240 mm

240x100x366 mm

6 kg

240x240 mm

240x100x366 mm

6 kg

1)

2)

3)

In the temperature range 15 ... 30 °C

1 ppm = 1/1 000 000 (referred to the current weight display)

Sensitivity deviation/year after first-time startup with self-calibration FACT switched on

PG12001-S

DeltaRange ®

0.1 g/1 g

12100 g

2400 g

0 ... 12100 g

0.08 g/0.3 g

±0.1g/±0.5 g

1…2 s

4 / 5 / 6 / 7 / 8 / 9 / 10 / 11 / 12 kg

240x240 mm

240x100x366 mm

6 kg

Technical data and optional equipment

58

7.2

Dimensions

240

366

345

14

191

280

225

131

42

10

Technical data and optional equipment

59

7.3 Optional equipment

With optional equipment from the METTLER TOLEDO product range, you can enhance the functionality of your PG-S balance. You have the following options available:

Normal paper printers

LC-P45 Printer: Printer with built-in applications (calibration and adjustment records conforming to GLP, statistical evaluations, totalization functions, etc.)

LC-P43 Printer (only with LC option): Printer for recording the results

GA42 Printer (only with RS232C): Printer for recording the results

Cables and cabling accessories

RS9–RS25: (m/f), length 2 m

RS9–RS9: (m/f), length 1 m

RS9–RS9: (m/m), length1 m

Antitheft device

Steel cable with lock

LC-P45

LC-P43

GA-42

11101052

11101051

21250066

590101

11101055 LC option LocalCAN universal interface (LC-G)

Protective covers

Protective cover for the entire PG-S balance

R/G-M emulation

Software cassette emulates interface commands and menu options of the PM balances.

For the use of PG-S balances in systems together with PM balances.

Weights

Available as OIML weights (E2 and F1, with certificate) or as adjustment (calibration) weights

(not OIML): 20 g, 50 g, 100 g and 200 g

11101643

21201308 on request

Technical data and optional equipment

60

This optional equipment can be used only with the LocalCAN universal interface!

Auxiliary displays

LC-AD: Auxiliary display, active, with bench stand

LC-ADS: Auxiliary display, active, with balance stand for PG-S

LC-PD: LCD auxiliary display, passive, with bench stand

LC-PDS: LCD auxiliary display, passive, with balance stand for PG-S

Foot switch

LC-FS: Foot switch with adjustable function

Cables and cabling accessories

LC-RS25: Cable for the attachment of a printer or computer with RS-232C, 25-pin

(m/f), such as IBM XT or compatibles

LC-RS9: Cable for the attachment of a computer with RS-232C, 9-pin such as IBM AT or compatibles

LC-CL: Cable for the attachment of a device with METTLER TOLEDO CL interface (5-pin)

LC-LC03: Extension cable for LocalCAN, 0.3 m

LC-LC2: Extension cable for LocalCAN, 2 m

LC-LC5: Extension cable for LocalCAN, 5 m

LC-LCT: T-piece for LocalCAN

229140

229150

229100

229070

229060

229050

229065

229130

239270

229115

229116

229118

Bar-code reader: LC-BCR usable for operation of the application software

Differential Weighing 238494

Differential weighing

Application software for differential weighing with bar-code reader LC-BCR

Application software for differential weighing

229145

238495

238494

Operating or installation instructions are enclosed with many of the options. For further information and for details on how to order the optional equipment, please contact your nearest METTLER TOLEDO dealer.

Appendix

61

8. Appendix

8.1

Overview of menu

Weighing mode

SECUrEd

1 Reset rESEt

Menu

“ OPEn

2 Adjustment fA[T tESt E [AL oFF

3 Autom. adjustm.

call-up

1) 6)

Cal lnFo oFF

Cal

InFo on

[ALimT

UAri CAL

4 Function 2)

F nonE

PCS

F count

5 Vibration adapter

2

1 3

F dYn A

Stk

F count

ForŸ≈UlA F 100

6 Weighing process

adapter

2

1 3

OFF

7 Repeatability

6ood

Std bESt bEttEr

8 Weighing unit 1 1)

Unit 1 g

9 Weighing unit 2 2)

Unit 2 g m

Unit 1 kg

Unit 1 mo

Unit 1 lb

Unit 1 tl

Unit2 t kg

Unit 2 tl

Unit2 S lb

Unit 2 ct

Unit 1 dwt

Unit 1 oz

Unit 1 ozt

Unit 1

GN

Unit 1 tl

Unit2 H oz

Unit 2 m

Unit 2 ozt

Unit 2 mo

Unit 2

GN

Unit 2 ct

Unit 2 dwt

Unit 2

10 Autozero

A" on

A" oFF

11 Autom. shutdown

ÅoFF -

ÅoFF 2'

12 Power-up mode 1) qÙ StArt

13 Icons on

AuToOFF

ÅoFF 5'

14 Peripheral unit 3)

PòinTEò

HoST 15 Send commands 4)

S. oFF

S. ConT S. STb

16 Send format 5)

S. SICS

S. PŸ≈ S. AuTo S. ALL see key

4)

,

5)

17 Baud rate 3)

BD 2400

18 Parity/bit 3)

7B-E

19 Handshake 3)

HS oFF

20 Settings

LiST

BD 1200 BD 4800 7B-odd 7B-no H S HArd H S SoFt SECUòE2 SECUòE1

BD 600 BD 9600 8B-no

BD 300 BD 150

4)

5)

6)

1)

2)

3)

With certified balances, these menu options have a fixed setting and can not be changed.

With certified balances, only the weighing units/functions allowed by the respective national weights and measures legislation may be selected.

These menu options are shown only if your balance is equipped with an RS232C interface.

These menu options are shown only if “HoSt” has been selected in menu option 14.

These menu options are shown only if “S.oFF” has not been selected in menu option 15.

This menu option is shown only if “FACT” or “CAL oFF” has not been selected in menu option 2

Appendix

62

8.2

Conversion table for weight units

Unit Gram g

1 g

1 mg

1 oz

1 ozt

1 GN

1 dwt

1 ct/C.M.

1 mo

1 m

1 tl (HK)

1 tl (SGP/Mal)

1 tl (Taiwan)

1

0.001

28.34952

31.10347

0.06479891

1.555174

0.2

3.75

4.608316

37.429

37.79937

37.5

Milligram mg

1000

1

28349.52

31103.47

64.79891

1555.174

200

3750

4608.316

37429

37799.37

37500

Ounce oz

(avdp)

0.03527396

0.0000352740

1

1.097143

0.002285714

0.05485714

0.007054792

0.1322774

0.1625536

1.320269

1.333333

1.322773

Troy ounce ozt

Grain

GN

0.03215075

0.0000321508

0.9114585

1

0.002083333

0.05

0.006430150

0.1205653

0.1481608

1.203370

1.215278

1.205653

15.43236

0.01543236

437.500

480

1

24

3.086472

57.87134

71.11718

577.6178

583.3334

578.7134

Pennyweight dwt

0.6430149

0.000643015

18.22917

20

0.04166667

1

0.1286030

2.411306

2.963216

24.06741

24.30556

24.11306

Unit

1 g

1 mg

1 oz

1 ozt

1 GN

1 dwt

1 ct/C.M.

1 mo

1 m

1 tl (HK)

1 tl (SGP/Mal)

1 tl (Taiwan)

Carat ct/C.M.

(metr.) koil

5

0.005

141.7476

155.5174

0.3239946

7.775869

1

18.75

23.04158

187.1450

188.9968

187.5

Momme mo

Mesghal m

0.2666667

0.000266667

7.559873

8.294260

0.01727971

0.4147130

0.05333333

1

1.228884

9.981068

10.07983

10

0.216999

0.000216999

6.151819

6.749423

0.01406130

0.3374712

0.04339980

0.8137461

1

8.122056

8.202425

8.137461

Tael tl

(Hong Kong)

0.02671725

0.0000267173

0.7574213

0.8309993

0.001731249

0.04154997

0.005343450

0.1001897

0.1231215

1

1.009895

1.001897

Tael tl

(Singapore)

(Malaysia)

0.02645547

0.0000264555

0.75

0.8228570

0.001714286

0.04114285

0.005291094

0.09920800

0.1219152

0.9902018

1

0.9920800

Tael tl

(Taiwan)

0.02666667

0.0000266667

0.7559874

0.8294261

0.001727971

0.04147131

0.005333333

0.1

0.1228884

0.9981068

1.007983

1

Appendix

63

8.3

SOP (Standard Operating Procedure)

In the documentation of a GLP test, the SOPs represent a relatively small but none the less very important part.

Practical experience has confirmed that SOPs produced in-house can be followed much better than those produced by an external, anonymous source.

In what follows , you will find a brief overview of the responsibilities in regard to SOPs, as well as a check list for the production of an SOP.

Responsibilities in regard to SOPs

Head of the inspection and test equipment

Test director

Personnel

GLP quality assurance arranges the production of SOPs approves SOPs with date and signature ensures that SOPs are available approves SOPs on behalf of the management follow the SOPs and other directives checks that valid SOPs are available checks that SOPs are followed checks whether and how modifications are documented

Appendix

64

Check list for the production of SOPs

Administrative matters

1. Use of SOP blank forms

2. Name of the inspection and test equipment

3. Date (production date of the SOP)

4. Filing identifier (code plan) for SOPs

5. Number of pages (1 of n)

6. Title

7. Date of putting into force

8. Modification information

9. Names of people (departments) responsible for the implementation

10. Date and signatures: a) author(s) b) checker c) person authorized for approval

11. Distribution list

Contents of the SOP

1.

Introduction and aim

2.

Required material

3.

Description of the work steps

4.

Description of the documentation

5.

Data processing and evaluation

6.

Documentation, samples, etc. to be stored

7.

Archiving directions

Yes

Yes

No

No

8.4

Index

A

Abort 46, 48, 50, 51, 53

Absolute weighing 24

Accuracy test 22

Adjustment operation 12

Adjustment to acceleration due to gravity 12

Adjustment tolerance 45

Adjustment weight 52

Adjustment 12, 21, 57

Ambient conditions 11, 12, 18, 21, 24, 52, 56

Animals 42

Antitheft device 59

Asterisk 43

Atmospheric humidity 56

Automatic adjustment call-up 22

Automatic shutdown 28

Automatic zero correction 28

Autozero 28

Auxiliary display 60

B

Backweighing 60

Bar-code reader 60

Baud rate 32

Below-the-balance weighings 44

Brief keystroke 7

C

Cables 59, 60

Care 53

CE declaration of conformity 6

Checkweighing 24

Cleaning 53

Components 38, 40, 41

Computers 54

Conversion table for weight unit 62

D

Data transfer rate 32

Data 17

DeltaRange ® 16

DeltaTrac 15, 37

Deviation 51

Dimensions 57

Display test 13, 18, 29

Double beep 20

Drift 28

Dynamic weighing 23, 42

E

Equipment features 6

Error messages 52

Evaporation measurements 28

F

F count 35

FACT 12, 21, 45

Factory setting 21

Fine addition 24, 25

Fine range 16

Foot switch 60

Formula weighing 23, 38

Formula 38

Function 23, 35

Appendix

65

Appendix

66

G

GLP 6, 12, 21, 54, 55, 59, 63

Good Laboratory Practice 6, 12

Graphic indicator 16

H

Handshake 18, 33

Hazardous environment 8

I

Icons 30

Individual components 39

Interface 60

Internal adjustment 21

ISO 9001 6

K

Key designations 7

Keypad 8

L

Leveling the balance 11

Linearity 57

List 33

LocalCAN universal interface 55, 59

Location change 10

Location selection 10

M

Maximum capacity 57

Menu options 19

Menu overview 61

Menu settings 33

Menu 18, 61

N

N total 40

Net total weight 40

Net weight 57

O

On and off switching 13

Open 34

Optional equipment 56, 59

Overload 52

P

Packaging 9

Percent weighing 23, 37

Peripheral device 30, 32, 33, 54, 55

Piece counting 23, 35

Pin assignment 55

Power cable 8

Power supply 11, 36, 37, 44

Power-up mode 29

Printer 17, 33, 34, 46, 48, 50, 51, 54, 59

Printing out settings 33

Protective cover 54, 59

Putting into operation 9

Q

Quickstart 29

R

Readability 16, 57

Record 12, 34, 40, 46, 48, 50, 51

Reference piece number 35, 52

Reference weight 35, 36, 37, 52

Reference 52

Repeatability 25, 52, 57

Repro-Set 25

Reset 21

RS232C interface 54

S

Safety 7, 8

Saving settings 20

Secure 33

Self-test 11

Serial interface 32

Service 53

Settings 19

Short-form operating instructions 6

Software version 11

SOP 6, 12, 63

Stability detector 15, 25

Stability 43

Stabilization time 57

Standard equipment 9

Standard Operating Procedure 6, 12, 63

Standby mode 14

Standby 28, 29

T

T total 40

Target weight 51

Taring range 14, 57

Taring 14

Technical data 56

Temperature 12, 21, 56

Test of the balance 22, 49, 50

Testing 12, 49

Thermal equilibrium 13

Tolerance marks 15

Tolerances 37

Total weight 39

U

Underload 52

Unit 26, 27

Unstable weighing samples 23, 42

V

VariCal 21, 45, 47

Vibration adapter 24, 52

W

Warm-up phase 12, 21

Warm-up time 13

Weighing container 14, 41

Weighing mode 19

Weighing pan 53, 57

Weighing process adapter 24, 42

Weighing result 17

Weighing unit 16, 17, 26, 27

Weights 12, 21

Z

Zero point 28

Appendix

67

To protect your METTLER TOLEDO product's future:

METTLER TOLEDO Service assures the quality, measuring accuracy and preservation of value of all METTLER TOLEDO products for years to come.

Please send for full details of our attractive terms of service.

Thank you.

*P21203050*

Subject to technical changes and to the availability of the accessories supplied with the instrument.

Printed on recycled paper. Because we care.

© Mettler-Toledo GmbH

2002 21203050B Printed in Switzerland 0206/2.12

Mettler-Toledo GmbH, Laboratory & Weighing Technologies, CH-8606 Greifensee, Switzerland

Phone +41-1-944 22 11, Fax +41-1-944 30 60, Internet: http://www.mt.com

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Related manuals

Download PDF

advertisement

Table of contents