Datasheet for U1252B 50000 Count Handheld DMM

Add to My manuals
14 Pages

advertisement

Datasheet for U1252B 50000 Count Handheld DMM | Manualzz

Agilent U1250 Series Handheld

Digital Multimeters

Data Sheet

Test more – more safely

The Agilent U1250 Series handheld digital multimeters (DMMs) exceed your expectations by delivering powerful features and performance that meet your toughest requirements and applications. The series’ latest multimeters, the U1251B, U1252B and U1253B come in vivid orange cases, offering capabilities and functions equivalent to the U1250A Series.

Key features

Superior contrast from organic LED

(OLED) display 1

50,000-count dual display

Up to 0.025% basic DCV accuracy

True-RMS AC and AC+DC 2 measurements

K-type and J-type 2 temperature measurements

Manual and automated (interval) data logging; internally to DMM and externally to PC

CAT III 1000 V and CAT IV 600 V safety protection

Built-in 20-MHz frequency counter 2

Built-in programmable square-wave generator 2

[1] U1253A/U1253B model

[2] U1252A/U1252B/U1253A/U1253B models

Do more with just one instrument

The basic model, U1251A/U1251B, expands your capabilities beyond typical DMM measurements to include data logging. The

U1252A/U1252B starts with the same foundation, and then adds a 20-MHz frequency counter and programmable square-wave generator — so you’d be able to perform more tests conveniently with one tool. What’s more: both models come bundled with a complete set of accessories to equip you right from the start at no extra cost.

Offering the same functionality as the U1252A/

U1252B, the U1253A/U1253B is the world’s fi rst

OLED handheld DMM. You won’t have to squint to be sure you’re reading it right: On the go or on the bench, you’ll get crystal-clear viewing indoors, even in dark, off-angle situations.

Find problems quickly

Troubleshooting can be tricky, especially when you’re dealing with elusive problems. With the

U1250 Series’data logging capability, you can ensure that every reading gets recorded manually or at intervals you specify. Better yet: you can have virtually unlimited data logging saves when you connect any of the U1250

Series DMM to a PC with the optional IR-to-USB cable.

In addition, the U1250 Series lets you achieve greater confi dence in your measurements with accurate true-RMS AC measurements, low DCV error rate of up to 0.025% and high-resolution display of 50,000 counts.

Uncompromising ruggedness and safety

Figure 1: Automate recording of measurements with bundled

GUI data-logging software.

The U1250 Series DMMs are housed in robust overmold enclosures, rated at CAT III 1000 V and

CAT IV 600 V and operate over a wide temperature range of –20 °C to +55 °C. Each

DMM also include a 30 kA high energy fuse to further protect you against violent fuse failures during high-current measurements. Built tough and certifi ed to stringent industrial standards, the U1250 DMM is what you need to face the demands of everyday tasks.

Take a closer look

OLED display with approximately 160° viewing angle, and high contrast ratio of

2000:1 for crystal-clear viewing 1

50,000 count dual display with true-

RMS reading capability

SHIFT key and one-stop function keys for easy access

Broad range of measurement functions, including temperature and capacitance

Built-in battery charging for optimum capacity

[1] U1253A/U1253B only

[2] U1252A/U1252B/U1253A/U1253B only

Figure 2: U1253B front view

2

20 MHz frequency counter 2

Data logging to internal or external memory

Programmable square-wave generator for stimulation of digital circuits 2

CAT III 1000 V and

CAT IV 600 V overvoltage protection for assurance of your safety while you go about your tasks

DC SPECIFICATIONS

FUNCTION

VOLTAGE (1)

RESISTANCE

CURRENT

(8)

RANGE

50.000 mV

500.00 mV

1000.0 mV

5.0000 V

50.000 V

500.00 V

1000.0 V

500.00 W [3]

5.0000 k W [3]

50.000 k W

500.00 k W

5.0000 M W

50.000 M W [4]

500.00 M W [4]

500.00 nS [5]

500.00 µA

5000.0 µA

50.000 mA

440.00 mA

5.0000 A

10.000 A [7]

-

RESOLUTION

0.001 mV

0.01 mV

0.1 mV

0.0001 V

0.001 V

0.01 V

0.1 V

0.01 W

0.0001 k W

0.001 k W

0.01 k W

0.0001 M W

0.001 M W

0.01 M W

0.01 nS

0.01 µA

0.1 µA

0.001 mA

0.01 mA

0.0001 A

0.001 A

0.1 mV

TEST CURRENT/

BURDEN VOLTAGE

-

-

-

-

-

-

-

1.04 mA

416 µA

41.2 µA

4.12 µA

375 nA

187 nA

187 nA

187 nA

0.06 V (100 W)

0.6 V (100 W)

0.09 V (1 W)

0.9 V (1 W)

0.2 V (0.01 W)

0.4 V (0.01 W)

1.04 mA

ACCURACY ± (% of reading + No. of Least Signifi cant Digit)

U1251A/U1251B

0.05 + 50

0.03 + 5

0.08 +10

0.08 + 5

0.2 + 5

1 + 10

N/A

1 + 20

0.1 + 5

0.2 + 5

0.3 + 10

[2]

[6]

[6]

U1252A/52B/53A/53B

0.05 + 50

0.025 + 5

0.03 + 5

0.05 + 10

0.05 + 5

0.15 +5

1 + 5

3+10 < 200 M W / 8+10 > 200 MW

1 + 10

0.05 + 5

0.15 + 5

0.3 + 10

0.3 + 5

[2]

[6]

[6]

DIODE TEST (8) 0.05 + 5

[1] Input impedance: Refer to Table A on page 9.

[2] The accuracy could be 0.05% + 10 for U1251A and 0.05% + 5 for U1252A/3A. Always use the NULL function to zero out the thermal effect before measuring the signal.

[3] The accuracy of 500 W and 5 kW is specifi ed after NULL function, which is used to subtract the test lead resistance and the thermal effect.

[4] For the range of 50 M W and 500 MW, the R.H. is specifi ed for <60%.

[5] The accuracy is specifi ed for <50 nS and after NULL function with open test lead.

[6] Always use the NULL function to zero out thermal effect with open test lead before measuring the signal. If the NULL function is not used, an additional 20 counts needs to be added to the

DC current accuracy. Thermal effect could occur due to the following:

• Wrong operation to measure the high voltage of 50 V ~ 1000 V for resistance, diode, and mV measurements.

• After battery-charging has completed.

• After measuring a current greater than 440 mA, it is suggested that the meter be left to cool down for twice the measuring time used.

[7] Current can be measured up to 10 A continuously. An additional 0.5% needs to be added to the specifi ed accuracy if the signal measured is in the range of 10 A~20 A for 30 seconds

maximum. After measuring a current of > 10 A, leave the meter to cool down for twice the measuring time used before application of low current measurement.

[8] Maximum open voltage: <+ 4.2V.

U1251A AC VOLTAGE SPECIFICATIONS

FUNCTION

TRUE RMS AC

VOLTAGE [1] [2]

RANGE

50.000 mV

500.00 mV

1000.0 mV

5.0000 V

50.000 V

500.00 V

1000.0 V

RESOLUTION

0.001 mV

0.01 mV

0.1 mV

0.0001 V

0.001 V

0.01 V

0.1 V

30 Hz ~ 45 Hz

ACCURACY ± (% of reading + No. of Least Signifi cant Digit)

FREQUENCY

45 Hz ~ 1 kHz 1 kHz ~ 10 kHz 10 kHz ~ 30 kHz

0.6+40

1.0+40 1.6+60

1.0+60

0.6+25

0.6+40

1.0+25

1.0+40

1.6+40

1.6+40 [3]

N/A

3

U1251B AC VOLTAGE SPECIFICATIONS

FUNCTION RANGE RESOLUTION

TRUE RMS AC

VOLTAGE [1] [2]

50.000 mV

500.00 mV

1000.0 mV

5.0000 V

50.000 V

500.00 V

1000.0 V

0.001 mV

0.01 mV

0.1 mV

0.0001 V

0.001 V

0.01 V

0.1 V

U1251A/U1251B AC CURRENT SPECIFICATIONS

FUNCTION RANGE RESOLUTION

AC CURRENT [2]

500.00 µA [4]

5000.0 µA

50.000 mA

440.00 mA

5.0000 A

10.000 A [5]

U1252A/U1253A AC VOLTAGE SPECIFICATIONS

0.01 µA

0.1 µA

0.001 mA

0.01 mA

0.0001 A

0.001 A

FUNCTION RANGE RESOLUTION

TRUE RMS AC

VOLTAGE [1] [2] [9]

50.000 mV

500.00 mV

1000.0 mV

5.0000 V

50.000 V

500.00 V

1000.0 V

U1252B/U1253B AC VOLTAGE SPECIFICATIONS

0.001 mV

0.01 mV

0.1 mV

0.0001 V

0.001 V

0.01 V

0.1 V

30 Hz ~ 45 Hz

ACCURACY ± (% of reading + No. of Least Signifi cant Digit)

45 Hz ~ 1 kHz

FREQUENCY

1 kHz ~ 5 kHz

0.6+40

5 kHz ~ 30 kHz

1.0+40 1.6+60

1.0+60

0.6+25

0.6+40

1.0+25

1.0+40

3.5+120

N/A

N/A

ACCURACY ± (% of reading + No. of Least Signifi cant Digit)

30 Hz ~ 45 Hz

1.5+50

FREQUENCY

45 Hz ~ 2 kHz 2 kHz ~ 20 kHz

3.0+80

1.5+40 3.0+60

0.8+20

2.0+40 [6]

3+60,

<3 A/5 kHz

20 Hz ~ 45 Hz

ACCURACY ± (% of reading + No. of Least Signifi cant Digit)

45 Hz ~ 1 kHz

0.4+40

FREQUENCY

1 kHz ~ 10 kHz 10 kHz ~ 20 kHz 20 kHz~100 kHz [7]

0.7+40

0.4+25 0.75+40

3.5+120

1.5+60

0.4+25 0.6+25

1.5+40

0.4+25

0.4+40 0.4+40 1.5+40 [3]

3.5+120 [3]

N/A

FUNCTION

TRUE RMS AC

VOLTAGE [1] [2] [9]

RANGE

50.000 mV

500.00 mV

1000.0 mV

5.0000 V

50.000 V

500.00 V

1000.0 V

RESOLUTION

0.001 mV

0.01 mV

0.1 mV

0.0001 V

0.001 V

0.01 V

0.1 V

20 Hz ~ 45 Hz

ACCURACY ± (% of reading + No. of Least Signifi cant Digit)

45 Hz ~ 1 kHz

0.4+40

FREQUENCY

1 kHz ~ 5 kHz

0.7+40

5 kHz ~ 15 kHz 15 kHz~100 kHz [8]

0.75+40

1.5+60

0.4+25

0.4+40

0.4+25

0.6+25

0.4+25

0.4+40

1.5+40

N/A

N/A

3.5+120

N/A

N/A

4

U1252A/U1252B/U1253A/U1253B AC CURRENT SPECIFICATIONS

FUNCTION RANGE RESOLUTION

20 Hz ~ 45 Hz

ACCURACY ± (% of reading + No. of Least Signifi cant Digit)

45 Hz ~ 1 kHz

FREQUENCY

1 kHz ~ 20 kHz 20 kHz~100 kHz [7]

TRUE RMS AC

CURRENT [2] [9]

500.00 µA [4]

5000.0 µA

50.000 mA

0.01 µA

0.1 µA

0.001 mA

1.0+20

0.7+20

0.75+20

1.5+20

5.0+80

440.00 mA

5.0000 A

10.000 A [5]

0.01 mA

0.0001 A

0.001 A

1.5+20 [6]

3+60,

<3 A/5 kHz

N/A

[1] Input impedance: Refer to Table A on page 9.

[2] AC mV/V and AC µA/mA/A specifi cations are true RMS AC coupled, valid from 5% to 100% of range. The crest factor may be up to 3 at full scale, up to 5 at half- scale, except for

1000 mV and 1000 V ranges, where the crest factor is 1.5 at full scale and 3 at half scale.

[3] For U1251A / U1252A / U1253A, the input signal is lower than 20,000,000 V-Hz (the product of voltage and frequency).

[4] Input current >35 µArms.

[5] Current can be measured from 2.5 A up to 10 A continuously. An additional 0.5% needs to be added to the specifi ed accuracy if the signal measured is in the range of 10 A ~ 20 A for 30

seconds maximum. After measuring a current of >10 A, leave the meter to cool down for twice the measuring time used before application of low current measurement.

[6] Input current < 3 Arms.

[7] The additional error to be added as frequency >20 kHz and signal input <10% of range: 3 counts of LSD per kHz.

[8] The additional error to be added as frequency >15 kHz and signal input <10% of range: 3 counts of LSD per kHz.

[9] Crest factor ≤ 3.0 at full scale, 5.0 at half scale except the 1000 mV and 1000 V ranges where it is 1.5 at full scale, 3.0 at half scale. For non-sinusoidal waveform, add 0.1% of reading

± 0.3% of range.

U1252A/U1253A AC+DC VOLTAGE SPECIFICATIONS

FUNCTION RANGE RESOLUTION

TRUE RMS AC

VOLTAGE [1] [2] [10]

50.000 mV

500.00 mV

1000.0 mV

5.0000 V

50.000 V

500.00 V

1000.00 V

0.001 mV

0.01 mV

0.1 mV

0.0001 V

0.001 V

0.01 V

0.1 V

U1252B/U1253B AC+DC VOLTAGE SPECIFICATIONS

FUNCTION

TRUE RMS AC

VOLTAGE [1] [2] [10]

RANGE

50.000 mV

500.00 mV

1000.0 mV

5.0000 V

50.000 V

500.00 V

1000.00 V

30 Hz ~ 45 Hz

1.5+80

ACCURACY ± (% of reading + No. of Least Signifi cant Digit)

FREQUENCY

45 Hz ~ 1 kHz 1 kHz ~ 10 kHz 10 kHz ~ 20 kHz 20 kHz~100 kHz [3]

0.4+60 0.7+60 0.8+60 3.5+220

0.4+30 0.8+45

3.5+125

0.4+30

0.6+30

1.5+65

1.5+45

0.4+30

0.4+45 0.4+45 1.5+45 [5]

3.5+125 [5]

N/A

RESOLUTION

0.001 mV

0.01 mV

0.1 mV

0.0001 V

0.001 V

0.01 V

0.1 V

30 Hz ~ 45 Hz

1.5+80

ACCURACY ± (% of reading + No. of Least Signifi cant Digit)

FREQUENCY

45 Hz ~ 1 kHz

0.4+60

1 kHz ~ 5 kHz

0.7+60

5 kHz ~ 15 kHz 15 kHz~100 kHz [4]

0.8+60 3.5+220

0.4+30 0.8+45

3.5+125

0.4+30

0.6+30

1.5+65

0.4+45

0.4+30

0.4+45

1.5+45

N/A

N/A

N/A

N/A

5

U1252A/U1252B/U1253A/U1253B AC+DC CURRENT SPECIFICATIONS

FUNCTION RANGE RESOLUTION

ACCURACY ± (% of reading + No. of Least Signifi cant Digit)

FREQUENCY

30 Hz ~ 45 Hz 45 Hz ~ 1 kHz 1 kHz ~ 20 kHz

TRUE RMS AC

CURRENT [10]

500.00 µA [6]

5000.0 µA

50.000 mA

440.00 mA

5.0000 A

10.000 A [7]

0.01 µA

0.1 µA

0.001 mA

0.01 mA

0.0001 A

0.001 A

1.1+25

1.2+25

1.8+30 [8]

0.8+25

0.9+25

0.9+30

0.9+25

0.8+25

0.9+25

3.3+70,

<3 A/5 kHz

U1251A/U1251B/U1252A/U1252B TEMPERATURE SPECIFICATIONS

THERMOCOUPLE

TYPE

ºC

–200 ~ 1372 ºC

RANGE

ºF

–328 ~ 2502 ºF

RESOLUTION

0.1 ºC/0.1 ºF K

J

(for U1252A/U1252B)

–210 ~ 1200 ºC –346 ~ 2192 ºF 0.1 ºC/0.1 ºF

ºC

0.3% + 3 ºC

0.3% + 3 ºC

ACCURACY

ºF

0.3% + 6 ºF

0.3% + 6 ºF

U1253A/U1253B TEMPERATURE SPECIFICATIONS

THERMOCOUPLE

TYPE

RANGE

K

J

ºC

–200 ~ –40 °C

–40 ~1372 °C

–210 ~ –40 °C

–40 ~ 1200 °C

ºF

–104 ~ –40 °F

–40 ~ 2502 °F

–346 ~ –40 °F

–40 ~ 2192 °F

CAPACITANCE SPECIFICATIONS

RANGE

10.000 nF

100.00 nF

1000.0 nF

10.000 µF

100.00 µF

1000.0 µF

10.000 mF

100.00 mF

FREQUENCY SPECIFICATIONS [5]

RANGE

RESOLUTION

0.001 nF

0.01 nF

0.1 nF

0.001 µF

0.01 µF

0.1 µF

0.001 mF

0.01 mF

RESOLUTION

RESOLUTION

0.1 ºC/0.1 ºF

0.1 °C /0.1 °F

0.1 °C /0.1 °F

0.1 °C /0.1 °F

ACCURACY

1% + 8

1% + 5

3% + 10

ºC

1% + 3 °C

1% + 1 °C

1% + 3 °C

1% + 1 °C

4 times/sec.

1 time/sec.

0.1 times/sec.

0.01 times/sec

ACCURACY

MEASURING RATE AT FULL

SCALE

U1251A/2A

0.02%+3 [9]

ACCURACY

U1253A

0.02%+3 [9]

ºF

1% + 5.4 °F

1% + 1.8 °F

1% + 5.4 °F

1% + 1.8 °F

MAX. DISPLAY

11000 counts

MIN. INPUT

FREQUENCY

99.999 Hz

999.99 Hz

9.9999 kHz

99.999 kHz

999.99 kHz

0.001 Hz

0.01 Hz

0.0001 kHz

0.001 kHz

0.01 kHz

0.02%+3, <600 kHz 0.02%+3, <600 kHz

1 Hz

6

[1] Input impedance: Refer to Table A on page 9.

[2] Crest factor ≤ 3.0 at full scale, 5.0 at half scale except the 1000 mV and 1000 V ranges where it is 1.5 at full scale, 3.0 at half scale.

For non-sinusoidal waveform, add 0.1% of reading ± 0.3% of range.

[3] The additional error to be added as frequency >20 kHz and signal input <10% of range: 3 counts of LSD per kHz.

[4] The additional error to be added as frequency >15 kHz and signal input <10% of range: 3 counts of LSD per kHz.

[5] For U1252A / U1253A, the input signal is lower than 20,000,000 V-Hz (the product of voltage and frequency).

[6] Input current >35 µArms.

[7] Current can be measured from 2.5 A up to 10 A continuously. An additional 0.5% needs to be added to the specifi ed accuracy if the signal measured is in the range of 10 A ~ 20 A for 30

seconds maximum. After measuring a current of >10 A, leave the meter to cool down for twice the measuring time used before application of low current measurement.

[8] Input current < 3 Arms.

[9] For non-square wave signals, add 5 counts.

[10] AC voltage and AC current specifi cations are true RMS AC coupled, valid from 5% to 100% of range.

U1251A/U1251B FREQUENCY SENSITIVITY DURING VOLTAGE MEASUREMENT

INPUT RANGE

FREQUENCY SENSITIVITY AND TRIGGER LEVEL

MINIMUM SENSITIVITY

(R.M.S. Sine Wave)

20 Hz - 100 kHz >100 kHz ~ 200 kHz (Maximum input for specifi ed accuracy = 10 x Range or 1000 V)

50.000 mV

500.00 mV

1000.0 mV

5.0000 V

50.000 V

500.00 V

1000.0 V

10 mV

25 mV

40 mV

0.25 V

2.5 V

25 V

50 V

15 mV

35 mV

50 mV

0.5 V

5 V

N/A

N/A

TRIGGER LEVEL FOR DC COUPLING

< 100 kHz

10 mV

60 mV

100 mV

0.5 V / 1.25 V (< 100 Hz)

5 V

50 V

300 V

>100 kHz ~ 200 kHz

15 mV

70 mV

150 mV

0.6 V

6 V

N/A

N/A

U1252A/U1252B/U1253A/U1253B FREQUENCY SENSITIVITY DURING VOLTAGE MEASUREMENT

INPUT RANGE

FREQUENCY SENSITIVITY AND TRIGGER LEVEL

MINIMUM SENSITIVITY

(R.M.S. Sine Wave)

20 Hz ~ 200 kHz >200 kHz ~ 500 kHz

TRIGGER LEVEL FOR DC COUPLING

< 100 kHz >100 kHz ~ 500 kHz (Maximum input for specifi ed accuracy = 10 x Range or 1000 V)

50.000 mV

500.00 mV

1000.0 mV

5.0000 V

50.000 V

500.00 V

1000.0 V

10 mV

70 mV

120 mV

0.3 V

3 V

30 V < 100 kHz

50 V < 100 kHz

25 mV

150 mV

300 mV

1.2 V

5 V

N/A

N/A

10 mV

70 mV

120 mV

0.6 V

6 V

60 V

120 V

25 mV

150 mV

300 mV

1.5 V

15 V

N/A

N/A

FREQUENCY SENSITIVITY DURING CURRENT MEASUREMENT

INPUT RANGE

500.00 µA

5000.0 µA

50.000 mA

440.00 mA

5.0000 A

10.000 A

MINIMUM SENSITIVITY (R.M.S. Sine Wave)

20 Hz ~ 20 kHz

100 µA

250 µA

10 mA

25 mA

1 A

2.5 A

7

PEAK HOLD

SIGNAL WIDTH

Single event > 1 ms

Repetitive > 250 µs

ACCURACY FOR DC mV/VOLTAGE/CURRENT

2% + 400 for all ranges

2% + 1000 for all ranges

DUTY CYCLE AND PULSE WIDTH [1]

FUNCTION

DUTY CYCLE

PULSE WIDTH

MODE

DC Coupling

500 ms

2000 ms

RANGE

0.01% ~ 99.99%

0.01 ms

0.1 ms

ACCURACY AT FULL SCALE

0.3% per kHz + 0.3%

0.2% + 3

0.2% + 3

[1] The positive or negative pulse width must be greater than 10 µs, and the duty cycle range should be considered. The pulse width range is determined by the frequency of the signal.

U1252A/U1252B/U1253A/U1253B FREQUENCY COUNTER SPECIFICATIONS

DIVISION

1

(secondary display “-1-“)

RANGE

99.999 Hz

RESOLUTION

0.001 Hz

ACCURACY ± (% of reading +

No. of Least Signifi cant Digit)

U1252A

0.02% + 3 [1]

U1253A

0.02% + 3 [1]

999.99 Hz

9.9999 kHz

0.01 Hz

0.0001 kHz

99.999 kHz

999.99 kHz

0.001 kHz

0.01 kHz

0.002% + 5,

<985 kHz

0.002% + 5,

< 985 kHz

SENSITIVITY

100 mV R.M.S.

200 mV R.M.S.

100

(secondary display “-100-“)

9.9999 MHz

9.9999 MHz

99.999 MHz

0.0001 MHz

0.0001 MHz

0.001 MHz

0.002% + 5,

< 1 MHz

0.002% + 5,

< 20 MHz

0.002% + 5,

< 20 MHz

400 mV R.M.S.

600 mV R.M.S.

MIN. INPUT

FREQUENCY

MAXIMUM

MEASUREMENT

LEVEL

0.5 Hz

1 MHz

< 30 Vpp

U1252A/U1252B/U1253A/U1253B SQUARE WAVE OUTPUT

FREQUENCY

OUTPUT [2]

DUTY CYCLE [3]

PULSE WIDTH [4]

AMPLITUDE

RANGE

0.5, 1, 2, 5, 6 [6] , 10, 15, 20, 25, 30, 40, 50, 60,

75, 80, 100, 120, 150, 200, 240, 300, 400, 480,

600, 800, 1200, 1600, 2400, 48000 Hz

0.39% ~ 99.60%

1/Frequency

Fixed 0 ~ +2.8 V

RESOLUTION

0.01 Hz

0.390625%

Range/256

0.1 V

ACCURACY

0.005% + 2

0.4% of full scale [4]

0.2 ms + Range/256

0.2 V

8

MEASURING RATE

FUNCTION

ACV

ACV + dB

DCV

AC + DC V

W /nS

Diode

Capacitance

DCI

ACI

AC + DC I

Temperature

7

7

2

TIMES/SECOND

7

14

2

6

7

7

14

4 (< 100 µF)

Frequency 1 (>10 Hz)

Duty cycle/Pulse width 0.5 (>10 Hz)

MANUAL AND INTERVAL DATA LOGGING

LOGGING

TYPE

MAXIMUM DATA POINTS [5]

U1251A U1252A

MANUAL

INTERVAL

100

200

DECIBEL (dB) CALCULATION dB BASE

1 m W (dBm)

1 V (dBV)

REFERENCE

1-9999 W

1 V

100

200

U1253A

100

1000

DEFAULT REFERENCE

50 W

1 V

[1] All frequency counters are susceptible to errors. Shielding inputs from external noise pickup is critical to minimize measurement errors. For non-square wave signals, add 5 counts.

[2] Output impedance: 3.5 k W maximum.

[3] The positive or negative pulse width must be greater than 50 µs for adjustment of the duty cycle or pulse width under different frequencies. Otherwise, the accuracy and range will be

different from the specifi cations defi ned.

[4] For signal frequencies greater than 1 kHz, an addition of 0.1% per kHz is added to the accuracy.

[5] For data logging to PC, maximum number of data points is dependent on available hard disk space.

[6] For the U1253A/U1253B model.

INPUT IMPEDANCE

Table A

FUNCTION

DC Voltage

RANGE

50 mV to 1000 mV

5 V to 1000 V

AC Voltage

AC + DC Voltage

50 mV to 1000 mV

5 V to 1000 V

50 mV to 1000 mV

5 V to 1000 V

U1251A

> 1 G W

10 M W (nominal)

> 1 G W

1.1 M W (nominal) in parallel with < 100pF

N/A

U1252A/U1253A

> 1 G W

10 M W (nominal), with

1.1 M W in parallel at dual display

> 1 G W

1.1 M W (nominal) in parallel with < 100pF

> 1 G W

1.1 M W (nominal) in parallel with < 100 pF

U1251B U1252B/U1253B

10 M W 10

10 M W (nominal),

with 10 M W in parallel at dual display

10 M W (nominal), with 10 M W in parallel at dual display

10 M W in parallel with

< 100 pF

10 M W in parallel with

< 100 pF

N/A 10 M W

10 M W (nominal) in parallel with 10 M W,

< 100 pF

9

GENERAL SPECIFICATIONS

DISPLAY

Both primary and secondary displays are 5-digit on the LCD display.

Both the primary and secondary displays offer a maximum resolution

of 50,000 counts. Automatic polarity indication.

CONNECTIVITY

IR to USB

POWER CONSUMPTION

105 mVA / 420 mVA (with backlight) maximum (U1251A/U1251B)

165 mVA / 480 mVA (with backlight) maximum (U1252A/U1252B)

420 mVA maximum (U1253A/U1253B)

BATTERY TYPE

9 V Alkaline battery (ANSI/NEDA 1604A or IEC 6LR61)

9 V Carbon-zinc battery (ANSI/NEDA 1604D or IEC6F22)

7.2 V or 8.4 V Ni-MH Rechargeable battery

BATTERY LIFE

U1251A/U1251B: 72 hours typical

U1252A/U1252B: 36 hours typical

U1253A/U1253B: 8 hours typical

OPERATING ENVIRONMENT

• Full accuracy at –20 o C to 55 o C ; and to 80% RH for temperatures

up to 35 °C, decreasing linearly to 50% RH at 55 °C

• 0 to 2000 m altitude per IEC 61010-1 2nd Edition CAT III, 1000 V

/ CAT IV, 600 V

STORAGE COMPLIANCE

–40 o C to 70 o C with battery removed

SAFETY COMPLIANCE

• IEC 61010-1:2001 / EN610101-1:2001

• Canada: CSA C22.2 No. 61010-1:2004

MEASUREMENT CATEGORY

CAT III 1000 V / CAT IV 600 V

EMC COMPLIANCE

Certifi ed to IEC/EN 61326: 2002, CISPR 11, and equivalents for

Group 1, Class A

COMMON MODE REJECTION RATIO (CMRR)

U1251A/U1251B/U1252A/U1252B: >90 dB at DC, 50/60 Hz ± 0.1%

(1 k W unbalanced)

U1253A/U1253B: >100 dB at DC, 50/60 Hz ± 0.1% (1 k W unbalanced)

NORMAL MODE REJECTION RATIO (NMRR)

U1251A/U1251B/U1252A/U1252B: >60 dB at DC, 50/60 Hz ± 0.1%

U1253A/U1253B: >90 dB at DC, 50/60 Hz ± 0.1%

TEMPERATURE COEFFICIENT

0.15 * (specifi ed accuracy)/ o C (from 20 o C to 18 o C or 28 o C to 55 o C)

SHOCK and VIBRATION

Tested to IEC/EN 60068-2

DIMENSIONS (HxWxD)

203.5 mm x 94.4 mm x 59.0 mm

WEIGHT

U1251A/U1251B: 504±5 g with battery

U1252A/U1252B/U1253A/U1253B: 527±5 g with battery

CHARGING TIME (only U1252A/3A)

< 220 minutes approx. at the environment of 10 o C to 30 o C

WARRANTY

3 years for main unit

3 months for standard accessories unless otherwise specifi ed.

10

ORDERING INFORMATION

U1251A U1251B U1252A U1252B U1253A

STANDARD SHIPPED ACCESSORIES

DESCRIPTION

Alligator clips

SMT grabbers

Fine-tip test probes

Test probe leads (4-mm)

Test probe leads (19-mm)

Mini grabber

Alkaline 9 V battery

Rechargeable Ni-MH battery with power adapter

Soft carrying case

Certifi cate of Calibration

Test Report

Quick Start Guide

U1251A

U1253A OPTION

Option 001 Bundled SMT grabbers, fi ne-tip test probes and soft carrying case

U1252A

U1253B

APPLICABLE MODELS

U1253A U1251B

OPTIONAL ACCESSORIES

MEASURING ACCESSORIES (NON-TEMPERATURE)

U1252B

U1253B

U1160A Standard test lead kit

Includes two test leads (red and black), alligator clips, fi ne-tip test probes, SMT grabbers and mini grabber (black).

• Test leads: CAT III 1000 V, 15 A

• Alligator clips: CAT III 1000 V, 10 A

• Fine-tip test probes: CAT II 300 V, 3 A

• SMT grabbers: CAT II 300 V, 3 A

• Mini grabber: CAT II 300 V, 3 A

U1161A Extended test lead kit

Includes two test leads (red and black), two test probes, medium-sized alligator clips and 4-mm banana plugs.

• Test leads: CAT III 1000 V, CAT IV 600 V, 15 A

• Test probes: CAT III 1000 V, 15 A

• Medium-sized alligator clips: CAT III 600 V, 10 A

• 4-mm banana plugs: CAT II 600 V, 10 A

U1162A Alligator clips

• One pair of insulated alligator clips (red and black).

Recommended for use with Agilent standard test

leads.

• Rated CAT III 1000 V, 10 A.

11

U1163A SMT grabbers

• One pair of SMT grabbers (red and black).

Recommended for use with Agilent standard test leads.

• Rated CAT II 300 V, 3 A.

U1164A Fine-tip test probes

• One pair of fi ne-tip test probes (red and black).

Recommended for use with Agilent standard test leads.

• Rated CAT II 300 V, 3 A.

U1165A Test probe leads

• Rated CAT III 1000 V, 15 A

MEASURING ACCESSORIES (NON-TEMPERATURE)

U1168A Standard test lead kit

Includes two test leads (red and black), 19-mm and 4-mm test probes, alligator clips, fi ne-tip test probes, SMT grabbers and mini grabber (black).

• Test leads: CAT III 1000 V, CAT IV 600 V, 15 A

• Test probes (19-mm tip): CAT III 1000 V, CAT IV 600 V, 15 A

• Test probes (4-mm tip): CAT III 1000 V, CAT IV 600 V, 15 A

(highly recommended for CAT IV environment)

• Alligator clips: CAT III 1000 V, 10 A

• Fine-tip test probes: CAT II 300 V, 3 A

• SMT grabber: CAT II 300 V, 3 A

• Mini grabber: CAT II 300 V, 3 A

U1169A Test probe leads

Includes two test leads (red and black), and a pair each of

19-mm and 4-mm test probes.

• Test leads: CAT III 1000 V, CAT IV 600 V, 15 A

• Test probes (19-mm tip): CAT III 1000 V, CAT IV 600 V, 15 A

• Test probes (4-mm tip): CAT III 1000 V, CAT IV 600 V, 15 A

(highly recommended for CAT IV environment)

U1583B AC current clamp

• Dual range: 40 A and 400 A

• Rated CAT III 600 V

• BNC-to-banana-plug adapter provided for use with DMMs

MEASURING ACCESSORIES (TEMPERATURE)

U1180A Thermocouple adapter+lead kit, J and K types

Includes thermocouple adapter, thermocouple bead J-type and thermocouple bead K-type.

• T/C adapter J/K-type

• T/C bead J-type: –20 °C to 200 °C

• T/C bead K-type: –20 °C to 200 °C

U1181A Immersion temperature probe

• Type-K T/C for use in oil and other liquids

• Measurement range: –50 °C to 700 ºC

• Includes adapter U1184A for connection to DMM

U1182A Industrial surface temperature probe

• Type-K T/C for use on still surfaces

• Measurement range: –50 ºC to 400 ºC

• Includes adapter U1184A for connection to DMM

U1183A Air temperature probe

• Type-K T/C for use in air and non-caustic gas

• Measurement range: –50 ºC to 800 ºC

• Includes adapter U1184A for connection to DMM

U1184A Temperature probe adapter

• Mini-connector-to-banana-plug adapter for use with DMM

U1185A J-type thermocouple and adapter

• T/C adapter J/K-type

• T/C bead J-type: –20 °C to 200 °C

12

MEASURING ACCESSORIES (TEMPERATURE)

U1186A K-type thermocouple and adapter

• T/C adapter J/K-type

• T/C bead J-type: –20 °C to 200 °C

CABLE CARRYING CASE

U1173A IR-to-USB cable

• For remote control and data logging to PC

• Max. baud rate: 19,200 bits per second

HANGING KIT

U1172A Transit case (aluminium-clad)

The robust casing to transport your DMM and accessories

• Aluminum-clad, black panel construction

• Dimension: 18” (H) x 13” (W) x 6“ (D)

• Weight: 4 kg

U1174A Soft carrying case

The convenient way to carry your DMM and essential accessories

• Dimension: 9” (H) x 5” (W) x 3” (D)

AC ADAPTOR

U1171A Magnetic hanging kit

For fastening of DMM to a steel surface so both hands are free.

U1170A AC adaptor

Includes AC power cord based on country

13

Agilent Email Updates www.agilent.com/find/emailupdates

Get the latest information on the products and applications you select.

www.lxistandard.org

LXI is the LAN-based successor to

GPIB, providing faster, more efficient connectivity. Agilent is a founding member of the LXI consortium.

Agilent Channel Partners

www.agilent.com/find/channelpartners

Get the best of both worlds: Agilent’s measurement expertise and product breadth, combined with channel partner convenience.

www.agilent.com

www.agilent.com/find/handhelddmm

Remove

all

doubt

Our repair and calibration services will get your equipment back to you, performing like new, when promised. You will get full value out of your Agilent equipment throughout its lifetime. Your equipment will be serviced by Agilent-trained technicians using the latest factory calibration procedures, automated repair diagnostics and genuine parts. You will always have the utmost confidence in your measurements. For information regarding self maintenance of this product, please contact your Agilent office.

Agilent offers a wide range of additional expert test and measurement services for your equipment, including initial start-up assistance, onsite education and training, as well as design, system integration, and project management.

For more information on repair and calibration services, go to:

For more information on Agilent

Technologies’ products, applications or services, please contact your local Agilent office. The complete list is available at: www.agilent.com/find/contactus

Americas

Canada (877) 894 4414

Latin America 305 269 7500

United States (800) 829 4444

Asia Pacific

Australia

China

Hong Kong

India

Japan

Korea

Malaysia

Singapore

Taiwan

Thailand

Europe & Middle East

Austria

Belgium

Denmark

Finland

France

1 800 629 485

800 810 0189

800 938 693

1 800 112 929

0120 (421) 345

080 769 0800

1 800 888 848

1 800 375 8100

0800 047 866

1 800 226 008

43 (0) 1 360 277 1571

32 (0) 2 404 93 40

45 70 13 15 15

358 (0) 10 855 2100

0825 010 700* www.agilent.com/find/removealldoubt

Germany

Ireland

Israel

49 (0) 7031 464 6333

1890 924 204

972-3-9288-504/544

Italy 39 02 92 60 8484

Netherlands 31 (0) 20 547 2111

Spain

Sweden

34 (91) 631 3300

0200-88 22 55

Switzerland 0800 80 53 53

United Kingdom 44 (0) 118 9276201

Other European Countries: www.agilent.com/find/contactus

Revised: October 1, 2009

Product specifications and descriptions in this document subject to change without notice.

© Agilent Technologies, Inc. 2010

Printed in USA, May 19, 2010

5989-5509EN