Midmark M11, a state-of-the-art sterilizer, offers reliable and efficient sterilization for your dental or medical instruments. With its user-friendly design and advanced features, the M11 ensures optimal performance and safety. It boasts rapid sterilization cycles, reaching 275°F in just 6 minutes, and effectively eliminates microorganisms, ensuring the sterility of your instruments. The M11 also features a convenient digital display for clear visibility of cycle status and error codes, allowing for easy troubleshooting.
advertisement
An ISO 9001 Certified Company
RPI FIELD SERVICE
CALIBRATION KIT
RPI Part #MIK074
For servicing the Midmark • Ritter Ultra Clave
®
For use with these models only:
M9 (S/N Series: CZ, DA, DB, DX, DY, FD & OM)
M9D (S/N Series: FF & FG)
M11 (S/N Series: ES, ET, FP, FR & GB)
M11D (S/N Series: GC & GC)
CALL (800) 221-9723 or (818) 882-8611 • FAX (818) 882-7028 • E-MAIL [email protected] • WEBSITE www.rpiparts.com
T
his guide has been prepared to help troubleshoot, evaluate and repair the Midmark•Ritter M9 and M11 Ultra-Clave
®
.
TABLE OF CONTENTS
1.0 – Leak Detection Guide
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2
Identifies areas in which leaks are most likely to occur.
2.0 – Planned Maintenance (PM) Check List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3
Lists step-by-step instructions for replacing most common PM parts including door gasket, dam gasket, filters and door spring.
3.0 – Error Codes and Other Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6
Provides explanation of error codes as well as other diagnostic tools including a fuse chart, drying cycle dip switch settings information, and service tips focusing on a low water error and inactive display lights.
4.0 – Door Opening Test Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7
Provides a test to verify if the door opening relay on the main PC board is functioning.
5.0 – Calibration Procedure – Pressure/Temperature Potentiometer Adjustments . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-14
Offers step-by-step, detailed calibration procedures as well as an at-a-glance version.
6.0 – RPI Parts to fit Midmark • Ritter M9 & M11 Ultra-Clave . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
15
Listing of RPI parts that fit the Ultra-Clave and includes the OEM Part # for easy reference.
1
1.0 - LEAK DETECTION GUIDE
DOOR and/or DAM GASKET
Check for water or steam leaking around the door.
VENT SOLENOID VALVE
FILL SOLENOID VALVE
BELLOWS VALVE
PRESSURE RELIEF
VALVE
PRESSURE
SENSOR HOSE
Check for water or steam leaking back into the reservoir through the condensing coil.
Check for water leaking back into the reservoir through the fill line by observing an increase in the reservoir water level and/or presence of bubbles.
Check for excessive steam from the bellows line in the reservoir. Note: a slight "hiss" of steam is not uncommon.
Check for water leaking beneath the back of the sterilizer.
Caution: Release pressure in the sterilizer before replacing the pressure relief valve.
Check for steam leaking onto or around the Main PC board.
2
2.0 - PLANNED MAINTENANCE (PM) CHECK LIST
R
E
P
LA
CE
ME
NT P
ARTS INDUS
TRIE
S,
IN
C
.
S
T
E
R
ILIZ
ER
PLAN
NED MAINTENAN
CE
CH
EC
K
LI
S
T
Sterilizer PM KIt (RPI Part #MIK072) for parts to fit the M9 UltraClave ® sterilizer
Sterilizer PM KIt (RPI Part #MIK080) for parts to fit the M11 UltraClave ® sterilizer
DOOR GASKETS: MIG027
fits
M9 and MIG028
fits
M11
Remove the old door gasket. Clean the gasket groove in the door and chamber face. For easy installation of the new gasket, use the “South-
East-West-North” method (i.e. begin installing the gasket at the bottom of the door, then install the left and right sides and finish at the top by squeezing in any of the excess, making sure that the gasket fits flush).
DAM GASKETS: MIG035
fits
M9 and MIG036
fits
M11
Use CAUTION when removing or installing the dam gasket – the door studs can be sharp. To install the door dam, align and place the dam gasket on and over the door studs. The dam must lie flat against the door (refer to the Installation Instructions included with the
MIG035/MIG036 dam gasket).
DOOR SPRING: MIS076
Remove the old spring. Reuse the screw to install the two new door springs – one on top of the other. This will aid in the door opening.
3
FILTERS: MIF038 or MIF062
Depending on the age of the machine, you will need to replace either the sintered bronze style Chamber Filter (MIF038) or both the mesh style Chamber Filter and Vent Filter (MIF062). See below for instructions.
• Chamber Filter (sintered bronze style filter) – MIF038:
If the machine has only a Chamber Filter – Remove all trays and the rack from the chamber. Clean the chamber, then remove the old chamber filter. Install new filter. Reinstall rack and trays.
• Chamber & Vent Filters (mesh style filter) – MIF062:
If the machine has both a Chamber and a Vent Filter – Remove all trays and the rack from the chamber. Clean the chamber, then remove the chamber filter. Install the new filter. You must also replace the Vent
Filter that serves as a filter for the steam trap/bellows and is located in the vent line to the tee fitting. Reinstall rack and trays.
After replacing the parts, check the reservoir and clean if dirt or debris are present. (Note: Debris is the #1 reason for Steam Trap/Bellows failure.) • Refill reservoir with distilled water only. • Conduct a temperature check using a Max Register Thermometer (RPI Part
#RPT113). • Conduct a pressure check to identify any leaks that may need to be addressed.
3.0 - ERROR CODES AND OTHER DIAGNOSTICS
ERROR CODE 1
Power interruption to control PC board.
1. Overheat thermostat activates (opens).
• Pressure leak during sterilization cycle.
• Water sensor shorted (which will result in bypassing of filling phase).
• Chamber water level too low.
• High input voltage during drying cycle.
• Faulty overheat thermostat.
2. Poor electrical connection or faulty wiring (bad power source to sterilizer).
3. Main PC board may be faulty.
ERROR CODE 2
The Stop button was depressed during the sterilization cycle.
ERROR CODE 3
The On/Standby button was depressed during the sterilization cycle.
4
ERROR CODE 4
Door ajar during sterilization cycle.
• Verify the door latch properly engages the door switch and the switch activates.
• Check door switch continuity and the wiring connections.
ERROR CODE 5
Excessive pressure (Pressure exceeds 35 psi).
• Sterilizer is overloaded.
• Bellows is closing prematurely trapping air in the chamber.
• Temperature probe is dirty or malfunctioning.
• Sterilizer may need to be re-calibrated. Refer to 5.0 - Calibration
Procedure.
ERROR CODE 6
Excessive temperature (Temperature exceeds 277˚F).
• Check for pressure leaks. Refer to 1.0 - Leak Detection Guide.
• Temperature probe is dirty or malfunctioning.
• Sterilizer may need to be re-calibrated. Refer to 5.0 - Calibration
Procedure.
3.0 - ERROR CODES AND OTHER DIAGNOSTICS (continued)
ERROR CODE 7
Pressure display reads less than 24.5 psi.
• Check for pressure leaks. Refer to 1.0 - Leak Detection Guide.
• Temperature probe is dirty or malfunctioning.
• Sterilizer may need to be re-calibrated. Refer to 5.0 - Calibration
Procedure.
ERROR CODE 10
Hardware and/or software problem – Watchdog timer reset error.
1. Poor electrical connection or faulty wiring (bad power source to sterilizer).
2. Main PC board may be faulty.
ERROR CODE 8 or 9
Door malfunction: Error Code 8 – Door switch did not return to normally open condition after door pulse solenoid actuated. Error Code 9 –
Display Pressure exceeded 0.9 PSI during the Dry cycle.
• If pressure reading is not at zero (0), sterilizer needs to be re-calibrated.
Refer to 5.0 - Calibration Procedure.
• If door opens at end of cycle, check door switch continuity (normally open).
• If door does not open at end of cycle, refer to 4.0 - Door Opening Test
Procedure.
• Inspect door mechanism for smooth operation. Lubricate if necessary.
• Check door switch and wire connections.
• Check door pulse solenoid.
ERROR CODE 11
Software interruption error.
• Run additional cycles. If problem continues, replace the Main PC board.
ERROR CODE 12
Ram Test Error.
• Run additional cycles. If problem continues, replace the Main PC board.
5
3.0 - ERROR CODES AND OTHER DIAGNOSTICS
(continued)
FUSE CHART (Fuses located on Main PC Board)
Printer Fuse
Fast Acting; 1/4" dia. x 1-1/4" Lg.; 3 Amp, 250 Volt
Line Fuse(s)
Fast Acting; 1/4" dia. x 1-1/4" Lg.; 15 Amp, 250 Volt
LOW WATER ERROR
Ground Circuit continuity failed the (5) minute time limit.
If water is continuously filling the chamber:
• Check for an open circuit or corrosion on the Water Lever Sensor.
• Check temperature probe by disconnecting it from Main PC board.
• The fill valve may be stuck in the Open position.
If water flow is restricted or not present:
• Check for debris (dirt) on the filter and in the fill lines.
• Check fill valve for proper operation or debris.
• Water level sensor malfunction.
DRYING CYCLE DIP SWITCH SETTING
If sterilizer is overheating in Dry mode, check the settings.
Caution: Disconnect the power cord before changing these settings.
Off = Position switch to the left. On = Position switch to the right.
Coolest: Setting #1
#1 OFF
#2 OFF
#3 ON
#4 ON
Cool: Setting #2
#1 OFF
#2 OFF
#3 OFF
#4 ON
Warm: Setting #3
#1 OFF
#2 OFF
#3 OFF
#4 OFF
Warmest: Setting #4
#1 OFF
#2 OFF
#3 ON
#4 OFF
If sterilizer needs to dry even cooler -
Setting #5
#1 OFF
#2 ON
#3 ON
#4 ON
DISPLAY LIGHTS INACTIVE
Sterilizer displays temperature and pressure, but the LED status lights do not light up.
• Disconnect the power cord and wait 30 seconds (this will reset the internal CPU). Run additional cycles.
6
4.0 - DOOR OPENING TEST PROCEDURE
Note: This test will only verify that the door opening relay on the main PC board is functioning. A complete cycle is necessary to verify function of the sensors and the latching mechanism.
1. To execute test, start any sterilization cycle. After chamber has filled with water, the "sterilizing" LED will light up.
2. Disconnect the power cord.
3. Open the door.
4. Depress STOP and POUCHES button simultaneously while re-connecting the power cord.
5. Sequentially depress each button from left to right across the panel.
6. Close the door. In sequence, this will make active all relays on the Main PC board.
7. To repeat this procedure, simply repeat Steps #2 through #6.
7
5.0 - CALIBRATION PROCEDURE - PRESSURE/TEMPERATURE POTENTIOMETER ADJUSTMENTS
IMPORTANT
The Calibration Procedure that begins on page 9 requires the sterilizer to be plugged in and powered up while the adjustments are being performed. Do not touch any components except for the adjusting screws of the potentiometers. Use the Trim
Pot Adjustment Tool (RPI Part# RPT460) to set the potentiometers. DANGER: Failure to comply with
these instructions may result in an electrical shock, which may result in severe personal injury or death.
8
5.0 - CALIBRATION PROCEDURE - PRESSURE/TEMPERATURE POTENTIOMETER ADJUSTMENTS
(continued)
The pressure zero, pressure range, and temperature potentiometers must not be adjusted exclusively. If one potentiometer needs to be adjusted, then the entire potentiometer adjustment procedure must be performed, and it must be performed in the proper sequence as described in the steps below.
INSTRUCTIONS
1) Remove the power cord from the A.C. outlet. Remove right hand side panel.
2) Open the sterilizer door.
3) To begin diagnostics, go to page 10, see Figure 1.
AT-A-GLANCE
CALIBRATION PROCEDURE
1) Open door.
2) Depress LIQUIDS and PACKS buttons at the same time as plugging in the power cord.
3.) Starting with UNWRAP button, depress and release all buttons from left to right across panel.
4) Turn pressure zero pot #1 counter-clockwise to 0.0 PSI. Then turn clockwise until the display pressure reads 0.1 PSI.
5) Depending on if the board is old or new, connect multi-meter leads as shown in Figure 2 on page 13 (old), or Figure 3 on page 14 (new).
6) Adjust temperature pot #3 to 2.56 VDC reading on multi-meter.
7) Connect the test pressure gauge, close the door and press the START button.
8) Check for pressure leaks at this time. Refer to 1.0 Leak Detection Guide.
9) After display reaches 272˚F, adjust the pressure range pot #2 until the display pressure matches test pressure gauge to within +/- 0.5 PSI
10) Depress the STOP button to end the test cycle.
11) After the sterilizer vents and the door opens, unplug the power cord
(main PC board to reset itself).
9
Figure 1 – Diagnostics Procedure (M9 & M11 and M9D & M11D Style Key Pad Displays)
Step #3 continued from page 9.
Read the IMPORTANT note on page 8, before you begin the following steps.
a) Press and hold the LIQUIDS and PACKS buttons simultaneously while connecting the power cord into an A.C. outlet. b) Release the LIQUIDS and PACKS buttons simultaneously to cause the display PC Board LED's to individually illuminate, and then extinguish one at a time in a left to right sequence. c) Press and release the buttons in the following order and one at a time: UNWRAPPED, POUCHES, LIQ-
UIDS, PACKS, START, STOP and ON/STANDBY. When each button is pressed, the corresponding LED will illuminate and stay illuminated. When the sequence is completed, the PRESSURE (PSI) display will display the pressure that the control PC board is reading and the TEMP (˚F)/TIME (MIN: SECONDS) display will display the temperature that the control PC board is reading. When the displays register proper values, continue with calibration – proceed to Step #4, page 11.
M9 & M11 Style Key Pad Display
M9D & M11D Style Key Pad Display
10
5.0 - CALIBRATION PROCEDURE - PRESSURE/TEMPERATURE POTENTIOMETER ADJUSTMENTS
(continued)
4) PRESSURE ZERO ADJUSTMENT (Potentiometer #1): Turn the adjusting screw of the pressure zero potentiometer, #1 (see Figures 2 and 3, pages
13-14), in the counter-clockwise direction until the Pressure (psi) display reads 0.0 psi (0.0 kPa). (See Notes 1 and 2 below.) Then turn clockwise until pressure display reads 0.1 psi (kPa).
5) TEMPERATURE ADJUSTMENT (Potentiometer #3): Continue by determining the style of PCB and confirming meter lead attachment points (see
Figures 2 and 3, pages 13-14). Set the multi-meter to read VDC, at least the 5 Volts range. Connect the red lead of the multi-meter to Test Point
(A) and the black lead of the multi-meter to Test Point (B). (See Figures 2 and 3, pages 13-14.)
CAUTION
Test Points for the multi-meter leads must be connected exactly as shown in the illustration - Figures 2 and 3, pages 13-14. Connecting a lead to the wrong side of the resistor or wrong pin of the chip will result in an incorrect adjustment.
6) The multi-meter should read 2.550 VDC +/- 0.001 VDC. If the multi-meter reading is not 2.550 VDC +/- 0.001 VDC, turn the adjusting screw on the temperature potentiometer #3 in a clockwise direction to lower the voltage setting or a counter-clockwise direction to raise the voltage setting until the multi-meter reading is 2.550 VDC, +/- 0.001 VDC. (See Notes 1 and 2 below.)
NOTE
In this step, if the digital multi-meter being used has only three digits, turn the adjusting screw on the potentiometer #3 counter-clockwise
(see Figures 2 and 3, pages 13-14) then adjust to a setting of 2.54 VDC and then adjust counter-clockwise to 2.55 VDC.
7) For temperature verification in a later step, place a Max-Register Thermometer on the tray and close the sterilizer door.
NOTES:
1) Turning the adjusting screw of the pressure range potentiometer in the clockwise direction raises the sterilizer pressure reading while turning the adjusting screw in the counter-clockwise direction lowers the sterilizer pressure reading. 2) Turning adjusting screw 1/4 turn raises or lowers the temperature of the sterilizer approximately 3O˚F (1.67˚C). On old style PC boards, turning the adjusting screw of the temperature potentiometer in the clockwise direction raises the sterilizer temperature while turning the adjusting screw in the counter clockwise direction lowers the sterilizer temperature. On new style PC boards, it is the opposite.
11
5.0 - CALIBRATION PROCEDURE - PRESSURE/TEMPERATURE POTENTIOMETER ADJUSTMENTS
(continued)
8) CONNECTING THE TEST PRESSURE GAUGE TO THE STERILIZER: Disconnect the black tubing from Pressure Sensor and from the barb fitting.
Keep black tubing in a safe place for re-installation upon completion of calibration procedure. Then connect the RPI Test Pressure Gauge to pressure sensor and the barb fitting – secure using clamps. (Important: See Notes 1 and 2 below. See illustration on page 8.)
9) Press the START button to initiate a test calibration cycle and then wait until the temperature on the TEMP (˚F) / TIME (MIN: SECONDS) display reaches 272 - 273˚F (133 -134˚C). From a cold start, the wait should be about 9-10 minutes.
10) PRESSURE RANGE ADJUSTMENT (Potentiometer #2): Turn the adjusting screw of the pressure range potentiometer, #2 (see Figures 2 and 3, pages 13-14) in the clockwise or counter-clockwise direction until the reading on the PRESSURE (psi) display matches the reading of the Test
Pressure Gauge within a tolerance of +/- O.5 psi (3.5 kPa).
11) Check the multi-meter reading, it must still read 2.550 VDC, +/- 0.001 VDC.
12) Press the STOP button to end the test calibration cycle and let the sterilizer vent.
13) Disconnect the power cord from the A.C. outlet, then remove the multi-meter leads from test points A and B (see Figures 2 and 3, page 13-14).
14) Open the sterilizer door and remove the Max-Register Thermometer (RPI Part #RPT113) from the sterilizer tray. Verify that the chamber temperature has reached 272-273˚F (133-134˚C).
15) Repeat Steps #3 and #4 of this procedure (see page 10-11). The display must read 0.1 psi (1 kPa) with the sterilizer door open. If it does not, repeat Calibration Procedure starting with Step #4, page 11.
16) Disconnect the power cord from the A.C. outlet. Remove the Test Pressure Gauge from the sterilizer. Using new cable ties, re-install the original black tubing removed in Step #8. Re-install right hand side panel.
17) Run a final test cycle and use Max-Register Thermometer (RPI Part #RPT113) to confirm chamber temperature.
NOTES:
1) CAUTION Place and tighten the hose Kwik Clamps™ on the fittings in order to secure the hose (see illustration, page 8). 2) After disconnecting Test Pressure
Gauge from the sterilizer, reconnect the black tubing to Pressure Sensor and tube fitting using new hi-temp cable ties (see illustration, page 8).
12
Figure 2 – Old Style PC Board
NOTE:
The pressure zero, pressure range, and temperature potentiometers must not be adjusted exclusively. If one potentiometer needs to be adjusted, then the entire potentiometer adjustment procedure must be performed, and it must be performed in the proper sequence.
13
Figure 3 – New Style PC Board
NOTE:
The pressure zero, pressure range, and temperature potentiometers must not be adjusted exclusively. If one potentiometer needs to be adjusted, then the entire potentiometer adjustment procedure must be performed, and it must be performed in the proper sequence.
14
M11
M9
Parts mentioned in this booklet are manufactured by Replacement Parts Industries, Inc. to fit Midmark
¥
Ritter M9 and M11 Ultra-Clave ® .
All product names used in this document are trademarks or registered trademarks of their respective holders. All rights reserved © RPI 2005.
15
MIK074INS REV A 11/05