sensor 500-D | User manual | DX-D 600 - Agfa HealthCare

Add to my manuals
6 Pages

advertisement

sensor 500-D | User manual | DX-D 600 - Agfa HealthCare | Manualzz

5

DX-D 600

GENERATORS

Generator model SHF 345

(CR only)

Input power 3-phase, 480 v 60 Hz

SHF 545 SHF 645 SHF 845

Max. power (kW) 32 50 64 80

Max. mA 400 640 640 800

Power output

@ 0.1s

Compatible X-ray All

Tubes

640 mA @ 78 kVp

500 mA @ 100 kVp

400 mA @ 125 kVp

320 mA @ 150 kVp

All

640 mA @ 100 kVp

500 mA @ 128 kVp

400 mA @ 150 kVp

E7252X, E7254FX,

E7869X

800 mA @ 100 kVp

640 mA @ 125 kVp

500 mA @ 150 kVp

E7254FX, E7869X

Range of radiographic parameters

mA From 10 mA to 800 mA through the following mA stations: 10, 12.5, 16, 20, 25, 32, 40, 50, 64,

80, 100, 125, 160, 200, 250, 320, 400, 500, 640, 800. (Depending on the Generator model)

mAs

ms

AEC

Product of mA x time values from 0.1 mAs to 500 mAs

From 1 to 10,000 milliseconds through the following time stations: 1, 2, 3, 4, 5, 6, 8, 10, 12, 16,

20, 25, 32, 40, 50, 64, 80, 100, 125, 160, 200, 250, 320, 400, 500, 640, 800, 1,000, 1,250,

1,600, 2,000, 2,500, 3,200, 4,000, 5,000, 6,400, 8,000 and 10,000.

mAs: 0.1 mAs to 500 mAs exposure time: Nominal shortest irradiation time = 1 ms

X-RAY TUBES

Housing Focal spot

Toshiba E7239X 1.0 - 2.0

Toshiba E7876X 0.6 - 1.2

Toshiba E7884X 0.6 - 1.2

Toshiba E7252X 0.6 - 1.2

Toshiba E7254FX 0.6 - 1.2

Toshiba E7869X 0.6 - 1.2

Target angle

16°

12°

12°

12°

12°

12°

Heat capacity (kHU)

140

230

300

300

400

600

Anode speed

Low

Low

Low

High/Low

High/Low

High/Low

DX-D 600

Why Agfa HealthCare?

Agfa HealthCare is a global leader in the fast growing market of integrated IT and imaging systems, offering healthcare facilities a seamless flow of information and a 360 º view of patient care. The company has a unique, holistic approach, enabling it to provide in-depth clinical know-how and fully integrated hospital-wide solutions.

These specialized solutions integrate IT and imaging systems for Radiology, Cardiology, Mammography* and

Orthopaedics. Agfa HealthCare’s enterprise-wide IT platform integrates administrative and clinical data within a healthcare facility and is designed to match the unique needs of healthcare professionals.

www.agfahealthcare.com

Agfa and the Agfa rhombus are trademarks of Agfa-Gevaert N.V., Belgium, or its affiliates. MUSICA and DX are trademarks of Agfa HealthCare NV, Belgium or its affiliates. All rights reserved. All other trademarks are held by their respective owners and are used in an editorial fashion with no intention of infringement. The data in this publication are for illustration purposes only and do not necessarily represent standards or specifications, which must be met by Agfa HealthCare. All information contained herein is intended for guidance purposes only, and characteristics of the products and services described in this publication can be changed at any time without notice. Products and services may not be available for your local area. Please contact your local sales representative at agfa.com for availability information. Agfa HealthCare diligently strives to provide as accurate information as possible, but shall not be responsible for any typographical error.

*NOTE: The DX-D 600 is not indicated for use in mammography.

© 2013 Agfa HealthCare Corp

All rights reserved

Published by

Agfa HealthCare Corp.

10 South Academy Street

Greenville, SC 29601 USA

5SJ9F US 00201303

DX-D 600

D I R E C T R A D I O G R A P H Y S Y S T E M

DX-D 600

HIGH-PRODUCTIVITY, EXCELLENT IMAGE QUALITY,

DIRECT RADIOGRAPHY SYSTEM WITH USER-

FRIENDLY DESIGN IN THREE CONFIGURATION

OPTIONS.

• Single or Dual-detector, high-productivity, highthroughput general radiography system with three configuration options: manual, auto-tracking and fully automatic “move to” positioning

• Excellent contrast detail provided by MUSICA² processing, producing exam-independent consistent image quality

• DICOM connectivity to PACS and RIS

• Wireless Cesium Iodide DR detector technology with columnar crystal growth for higher sensitivity

(DQE)

• Option to integrate with Agfa HealthCare CR systems for maximum versatility

The DX-D 600 Direct Radiography system combines userfriendly design with Agfa HealthCare’s MUSICA² image quality to create this high-productivity, direct radiography

(DR) solution. The DX-D 600 is family of ceiling mounted

X-ray systems with configurations ranging from a system using manual movement to a fully motorized, autopositioning solution, ideal for facilities with a high patient load that are looking to streamline workflow to increase patient comfort. The fully integrated NX Workstation communicates seamlessly with PACS and RIS, and eliminates the need for a separate generator console. Both APR and

X-ray parameters are captured in the DICOM header after exposure and become part of the patient demographics for future reference.

The DX-D 600 features Cesium Iodide detector technology, which offers excellent image quality and fast image availability. GOS technology (Gadolinium Oxy-Sulphide) is also available. Both detector types are available in fixed or removable formats for versatility, including wireless.

Agfa HealthCare’s proprietary MUSICA² image processing software delivers image consistency for improved workflow and excellent contrast detail to help enhance diagnostic confidence.

2

DX-D 600

Configurations to meet every need

The DX-D 600 is available in three different basic configurations, which can be tailored to meet your workflow and budget. The fully automatic system offers the latest in robotic auto-positioning technology, as well as fully-automated tracking. Position selection is available at tube head or through the exam tree preprogrammed settings for increased speed and convenience. The semi-automatic configuration includes vertical tracking with the wall stand and the table. The fully manual system is easily positioned without the aid of motorization.

Features such as the innovative design with a touch screen control panel at the tube head, generator control and image preview on single NX monitor, grid sensing for both table and wall stand, solid state AEC (Automatic Exposure Control) for high-speed accuracy, and LED lighting are features designed for convenience and quality patient care. simultaneously. Fine bony detail is enhanced in the same image to provide sharpness across the full dynamic range.

The result is few if any manual adjustments are needed, improving both workflow and image consistency.

Detector technology

The cassette-less DX-D 600 direct radiography workflow allows high productivity, and includes the choice of GOS or

Cesium Iodide technology for even higher panel sensitivity

(DQE). The single plate design eliminates tiling artifacts, while the direct availability of images speeds up workflow and reduces patient waiting times. Wireless, fixed or tethered detector options cover a wide range of budget and workflow requirements.

Combine CR flexibility and DR performance

The DX-D 600 can be integrated with Agfa HealthCare’s CR systems, such as the DX-G and DX-M. With Agfa HealthCare’s

CR needle detector technology, the combined systems deliver the high image quality with the flexibility of cassette-based

CR systems.

Excellent ease of operation, in any situation

This solution can efficiently handle demanding X-ray departments. The versatile ceiling suspended tube crane operates with a touch sensor keypad, which controls all the ceiling support movements and displays the X-ray parameters and patient information. The radiographic table allows easy positioning with four way float top movement and a wide vertical range as well with a 660 pound capacity. Manual movement and tracking or move-to positioning options for tube and wall stand allow you to select the level of automation you require for increased department efficiency and patient comfort. Combine room configurations with choice of panels to custom tailor your room your way.

MUSICA²: tuned to optimize image quality

Agfa HealthCare’s “gold standard” MUSICA² image processing software has been specially adapted and tuned to optimize DR image quality improving visualization of fine detail by optimizing both density and contrast

Services & Support

Agfa HealthCare offers service agreement solutions tailored to the individual customer’s situation. The service agreements are available in Basic, Advantage, and PM Only levels, making lifecycle costs predictable.

A worldwide team of some 1,000 service professionals is at your disposal to provide support at all phases of your project.

As an additional service, they can help you customize your examination tree or link RIS protocol codes, for an even higher return on investment. Furthermore, this team carries out tasks that go well beyond maintenance, including value added services such as super user training, staff training and software upgrades.

DX-D 600

DX-D 600 FEATURE PRODUCT MATRIX

Feature

Fully motorized “move to” auto-positioning

Auto-Tracking with Table and Wall Stand

Manual Positioning

TFT Touch Screen at Tube

Tilting Wall Stand

Automatic Collimation

Centering Laser

DAP Meter

FLFS**

DX-D 600

Automatic*

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Optional

* Available in Chest Room Configuration as well.

** FLFS is a Work in Progress.

technical

SPECIFICATIONS

ENVIRONMENTAL REQUIREMENTS

There are no special environmental conditions required for the safe operation of the ceiling suspension. However, it is not designed for the use in the presence of explosive or flammable gases as might be found in operating rooms.

Operating

• Atmospheric pressure

Minimum

Maximum

• Relative humidity

Minimum

Maximum

• Ambient temperature

Minimum

Maximum

PATIENT TABLE

• Tabletop width

• Tabletop length

• Table height

(motorized adjustment)

700 hPa

1,060 hPa

30 %

75 %

59° F

86° F

86.8 cm (34.17 in)

220 cm (86.61 in)

58 to 92 cm

(22.83 to 36.22 in)

DX-D 600

Semi-Automatic*

-

Yes

Yes

Yes

Optional

Optional

Yes

Optional

-

DX-D 600

Analog Manual

-

-

-

Yes

-

Optional

Optional

Optional

Optional

• X-ray absorption < 1.3 mm Al

(0.05 in)

• Tabletop travel longitudinal 109.5 cm (43.11 in)

(+60 cm, -49.5 cm)

• Tabletop travel transverse

(+23.62 in,

-19.49 in)

25 cm (± 12.5 cm)

9.84 in (± 4.92 in)

• Tabletop material

• Max. patient weight

Carbon fiber

300 kg (660 lbs)

• Bucky travel along table access 61 cm (24.02 in)

• Automatic exposure control 3-field solid state sensors

• Focused Antiscatter included in the table

CEILING MOUNTED X-RAY TUBE SUPPORT

• Minimum source-ceiling distance 726 mm (28.58 in)

• Vertical telescope travel range 1,584 mm (62.36 in)

• Tube rotation range, alpha

• Tube rotation range, beta

-135°

-180°

+135°

+180°

• Software configured rotation stops for both alpha and beta rotations

3

DX-D 600

DETECTOR

• Image size:

43 x 35 cm (16.93 x 13.78 in) 3,072 x 2,560 pixels

43 x 43 cm (16.93 x 16.93 in) 3,072 x 3,072 pixels

• Scintillator: Cesium Iodide/GOS

Photodiode Array

Contiguous a-Si matrix Fast Imaging

Cycle Time

COLLIMATORS

• Inherent filtration

• Full field light localizer

• Rotation

2 mm Aluminum equivalent

>160 lx up to maximum

± 90°

WALLSTAND

• Vertical Movement 1,495 mm (58.9 in)

400 to 1,895 mm above floor (center to center) (15.7 to 74.6 in)

• Tilting bucky -20° to +90°

(horizontal position)

• Table surface-detector plane distance

• Radiation absorption

< 40 mm (1.6 in)

< 0.6 mm Al equivalent (0.02 in)

• Automatic exposure control 3-field solid state

• Optional wall stand grids sensors

150 to 180 cm

(59.1 to 70.9 in)

SYSTEM ACCESSORIES

• Table compression band

• Table hand grips

• Table mattress

• Overhead hands support for wall stand

• Hand grips for wall stand

• Vacudap 2004 (External Dose Area Meter)

INSTALLATION DATA

• Line voltage 3-phase, 480 v 60 Hz

• Automatic line compensation ± 10 %

• Power consumption 105 kVA (Gen 64 kW);

120 kVA (Gen 80 kW)

• Ceiling height 2.60 - 2.80 m

(8.43 - 9.08 feet)

(for normal use)

ELECTRICAL REQUIREMENTS

• Frequency

• Voltage

Minimum

Maximum

• Maximum current

Momentary

Continuous

60 Hz

115 V ~

240 V ~

3.5 A

1.6 A

4

2

DX-D 600

Configurations to meet every need

The DX-D 600 is available in three different basic configurations, which can be tailored to meet your workflow and budget. The fully automatic system offers the latest in robotic auto-positioning technology, as well as fully-automated tracking. Position selection is available at tube head or through the exam tree preprogrammed settings for increased speed and convenience. The semi-automatic configuration includes vertical tracking with the wall stand and the table. The fully manual system is easily positioned without the aid of motorization.

Features such as the innovative design with a touch screen control panel at the tube head, generator control and image preview on single NX monitor, grid sensing for both table and wall stand, solid state AEC (Automatic Exposure Control) for high-speed accuracy, and LED lighting are features designed for convenience and quality patient care. simultaneously. Fine bony detail is enhanced in the same image to provide sharpness across the full dynamic range.

The result is few if any manual adjustments are needed, improving both workflow and image consistency.

Detector technology

The cassette-less DX-D 600 direct radiography workflow allows high productivity, and includes the choice of GOS or

Cesium Iodide technology for even higher panel sensitivity

(DQE). The single plate design eliminates tiling artifacts, while the direct availability of images speeds up workflow and reduces patient waiting times. Wireless, fixed or tethered detector options cover a wide range of budget and workflow requirements.

Combine CR flexibility and DR performance

The DX-D 600 can be integrated with Agfa HealthCare’s CR systems, such as the DX-G and DX-M. With Agfa HealthCare’s

CR needle detector technology, the combined systems deliver the high image quality with the flexibility of cassette-based

CR systems.

Excellent ease of operation, in any situation

This solution can efficiently handle demanding X-ray departments. The versatile ceiling suspended tube crane operates with a touch sensor keypad, which controls all the ceiling support movements and displays the X-ray parameters and patient information. The radiographic table allows easy positioning with four way float top movement and a wide vertical range as well with a 660 pound capacity. Manual movement and tracking or move-to positioning options for tube and wall stand allow you to select the level of automation you require for increased department efficiency and patient comfort. Combine room configurations with choice of panels to custom tailor your room your way.

MUSICA²: tuned to optimize image quality

Agfa HealthCare’s “gold standard” MUSICA² image processing software has been specially adapted and tuned to optimize DR image quality improving visualization of fine detail by optimizing both density and contrast

Services & Support

Agfa HealthCare offers service agreement solutions tailored to the individual customer’s situation. The service agreements are available in Basic, Advantage, and PM Only levels, making lifecycle costs predictable.

A worldwide team of some 1,000 service professionals is at your disposal to provide support at all phases of your project.

As an additional service, they can help you customize your examination tree or link RIS protocol codes, for an even higher return on investment. Furthermore, this team carries out tasks that go well beyond maintenance, including value added services such as super user training, staff training and software upgrades.

DX-D 600

DX-D 600 FEATURE PRODUCT MATRIX

Feature

Fully motorized “move to” auto-positioning

Auto-Tracking with Table and Wall Stand

Manual Positioning

TFT Touch Screen at Tube

Tilting Wall Stand

Automatic Collimation

Centering Laser

DAP Meter

FLFS**

DX-D 600

Automatic*

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Optional

* Available in Chest Room Configuration as well.

** FLFS is a Work in Progress.

technical

SPECIFICATIONS

ENVIRONMENTAL REQUIREMENTS

There are no special environmental conditions required for the safe operation of the ceiling suspension. However, it is not designed for the use in the presence of explosive or flammable gases as might be found in operating rooms.

Operating

• Atmospheric pressure

Minimum

Maximum

• Relative humidity

Minimum

Maximum

• Ambient temperature

Minimum

Maximum

PATIENT TABLE

• Tabletop width

• Tabletop length

• Table height

(motorized adjustment)

700 hPa

1,060 hPa

30 %

75 %

59° F

86° F

86.8 cm (34.17 in)

220 cm (86.61 in)

58 to 92 cm

(22.83 to 36.22 in)

DX-D 600

Semi-Automatic*

-

Yes

Yes

Yes

Optional

Optional

Yes

Optional

-

DX-D 600

Analog Manual

-

-

-

Yes

-

Optional

Optional

Optional

Optional

• X-ray absorption < 1.3 mm Al

(0.05 in)

• Tabletop travel longitudinal 109.5 cm (43.11 in)

(+60 cm, -49.5 cm)

• Tabletop travel transverse

(+23.62 in,

-19.49 in)

25 cm (± 12.5 cm)

9.84 in (± 4.92 in)

• Tabletop material

• Max. patient weight

Carbon fiber

300 kg (660 lbs)

• Bucky travel along table access 61 cm (24.02 in)

• Automatic exposure control 3-field solid state sensors

• Focused Antiscatter included in the table

CEILING MOUNTED X-RAY TUBE SUPPORT

• Minimum source-ceiling distance 726 mm (28.58 in)

• Vertical telescope travel range 1,584 mm (62.36 in)

• Tube rotation range, alpha

• Tube rotation range, beta

-135°

-180°

+135°

+180°

• Software configured rotation stops for both alpha and beta rotations

3

DX-D 600

DETECTOR

• Image size:

43 x 35 cm (16.93 x 13.78 in) 3,072 x 2,560 pixels

43 x 43 cm (16.93 x 16.93 in) 3,072 x 3,072 pixels

• Scintillator: Cesium Iodide/GOS

Photodiode Array

Contiguous a-Si matrix Fast Imaging

Cycle Time

COLLIMATORS

• Inherent filtration

• Full field light localizer

• Rotation

2 mm Aluminum equivalent

>160 lx up to maximum

± 90°

WALLSTAND

• Vertical Movement 1,495 mm (58.9 in)

400 to 1,895 mm above floor (center to center) (15.7 to 74.6 in)

• Tilting bucky -20° to +90°

(horizontal position)

• Table surface-detector plane distance

• Radiation absorption

< 40 mm (1.6 in)

< 0.6 mm Al equivalent (0.02 in)

• Automatic exposure control 3-field solid state

• Optional wall stand grids sensors

150 to 180 cm

(59.1 to 70.9 in)

SYSTEM ACCESSORIES

• Table compression band

• Table hand grips

• Table mattress

• Overhead hands support for wall stand

• Hand grips for wall stand

• Vacudap 2004 (External Dose Area Meter)

INSTALLATION DATA

• Line voltage 3-phase, 480 v 60 Hz

• Automatic line compensation ± 10 %

• Power consumption 105 kVA (Gen 64 kW);

120 kVA (Gen 80 kW)

• Ceiling height 2.60 - 2.80 m

(8.43 - 9.08 feet)

(for normal use)

ELECTRICAL REQUIREMENTS

• Frequency

• Voltage

Minimum

Maximum

• Maximum current

Momentary

Continuous

60 Hz

115 V ~

240 V ~

3.5 A

1.6 A

4

2

DX-D 600

Configurations to meet every need

The DX-D 600 is available in three different basic configurations, which can be tailored to meet your workflow and budget. The fully automatic system offers the latest in robotic auto-positioning technology, as well as fully-automated tracking. Position selection is available at tube head or through the exam tree preprogrammed settings for increased speed and convenience. The semi-automatic configuration includes vertical tracking with the wall stand and the table. The fully manual system is easily positioned without the aid of motorization.

Features such as the innovative design with a touch screen control panel at the tube head, generator control and image preview on single NX monitor, grid sensing for both table and wall stand, solid state AEC (Automatic Exposure Control) for high-speed accuracy, and LED lighting are features designed for convenience and quality patient care. simultaneously. Fine bony detail is enhanced in the same image to provide sharpness across the full dynamic range.

The result is few if any manual adjustments are needed, improving both workflow and image consistency.

Detector technology

The cassette-less DX-D 600 direct radiography workflow allows high productivity, and includes the choice of GOS or

Cesium Iodide technology for even higher panel sensitivity

(DQE). The single plate design eliminates tiling artifacts, while the direct availability of images speeds up workflow and reduces patient waiting times. Wireless, fixed or tethered detector options cover a wide range of budget and workflow requirements.

Combine CR flexibility and DR performance

The DX-D 600 can be integrated with Agfa HealthCare’s CR systems, such as the DX-G and DX-M. With Agfa HealthCare’s

CR needle detector technology, the combined systems deliver the high image quality with the flexibility of cassette-based

CR systems.

Excellent ease of operation, in any situation

This solution can efficiently handle demanding X-ray departments. The versatile ceiling suspended tube crane operates with a touch sensor keypad, which controls all the ceiling support movements and displays the X-ray parameters and patient information. The radiographic table allows easy positioning with four way float top movement and a wide vertical range as well with a 660 pound capacity. Manual movement and tracking or move-to positioning options for tube and wall stand allow you to select the level of automation you require for increased department efficiency and patient comfort. Combine room configurations with choice of panels to custom tailor your room your way.

MUSICA²: tuned to optimize image quality

Agfa HealthCare’s “gold standard” MUSICA² image processing software has been specially adapted and tuned to optimize DR image quality improving visualization of fine detail by optimizing both density and contrast

Services & Support

Agfa HealthCare offers service agreement solutions tailored to the individual customer’s situation. The service agreements are available in Basic, Advantage, and PM Only levels, making lifecycle costs predictable.

A worldwide team of some 1,000 service professionals is at your disposal to provide support at all phases of your project.

As an additional service, they can help you customize your examination tree or link RIS protocol codes, for an even higher return on investment. Furthermore, this team carries out tasks that go well beyond maintenance, including value added services such as super user training, staff training and software upgrades.

DX-D 600

DX-D 600 FEATURE PRODUCT MATRIX

Feature

Fully motorized “move to” auto-positioning

Auto-Tracking with Table and Wall Stand

Manual Positioning

TFT Touch Screen at Tube

Tilting Wall Stand

Automatic Collimation

Centering Laser

DAP Meter

FLFS**

DX-D 600

Automatic*

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Optional

* Available in Chest Room Configuration as well.

** FLFS is a Work in Progress.

technical

SPECIFICATIONS

ENVIRONMENTAL REQUIREMENTS

There are no special environmental conditions required for the safe operation of the ceiling suspension. However, it is not designed for the use in the presence of explosive or flammable gases as might be found in operating rooms.

Operating

• Atmospheric pressure

Minimum

Maximum

• Relative humidity

Minimum

Maximum

• Ambient temperature

Minimum

Maximum

PATIENT TABLE

• Tabletop width

• Tabletop length

• Table height

(motorized adjustment)

700 hPa

1,060 hPa

30 %

75 %

59° F

86° F

86.8 cm (34.17 in)

220 cm (86.61 in)

58 to 92 cm

(22.83 to 36.22 in)

DX-D 600

Semi-Automatic*

-

Yes

Yes

Yes

Optional

Optional

Yes

Optional

-

DX-D 600

Analog Manual

-

-

-

Yes

-

Optional

Optional

Optional

Optional

• X-ray absorption < 1.3 mm Al

(0.05 in)

• Tabletop travel longitudinal 109.5 cm (43.11 in)

(+60 cm, -49.5 cm)

• Tabletop travel transverse

(+23.62 in,

-19.49 in)

25 cm (± 12.5 cm)

9.84 in (± 4.92 in)

• Tabletop material

• Max. patient weight

Carbon fiber

300 kg (660 lbs)

• Bucky travel along table access 61 cm (24.02 in)

• Automatic exposure control 3-field solid state sensors

• Focused Antiscatter included in the table

CEILING MOUNTED X-RAY TUBE SUPPORT

• Minimum source-ceiling distance 726 mm (28.58 in)

• Vertical telescope travel range 1,584 mm (62.36 in)

• Tube rotation range, alpha

• Tube rotation range, beta

-135°

-180°

+135°

+180°

• Software configured rotation stops for both alpha and beta rotations

3

DX-D 600

DETECTOR

• Image size:

43 x 35 cm (16.93 x 13.78 in) 3,072 x 2,560 pixels

43 x 43 cm (16.93 x 16.93 in) 3,072 x 3,072 pixels

• Scintillator: Cesium Iodide/GOS

Photodiode Array

Contiguous a-Si matrix Fast Imaging

Cycle Time

COLLIMATORS

• Inherent filtration

• Full field light localizer

• Rotation

2 mm Aluminum equivalent

>160 lx up to maximum

± 90°

WALLSTAND

• Vertical Movement 1,495 mm (58.9 in)

400 to 1,895 mm above floor (center to center) (15.7 to 74.6 in)

• Tilting bucky -20° to +90°

(horizontal position)

• Table surface-detector plane distance

• Radiation absorption

< 40 mm (1.6 in)

< 0.6 mm Al equivalent (0.02 in)

• Automatic exposure control 3-field solid state

• Optional wall stand grids sensors

150 to 180 cm

(59.1 to 70.9 in)

SYSTEM ACCESSORIES

• Table compression band

• Table hand grips

• Table mattress

• Overhead hands support for wall stand

• Hand grips for wall stand

• Vacudap 2004 (External Dose Area Meter)

INSTALLATION DATA

• Line voltage 3-phase, 480 v 60 Hz

• Automatic line compensation ± 10 %

• Power consumption 105 kVA (Gen 64 kW);

120 kVA (Gen 80 kW)

• Ceiling height 2.60 - 2.80 m

(8.43 - 9.08 feet)

(for normal use)

ELECTRICAL REQUIREMENTS

• Frequency

• Voltage

Minimum

Maximum

• Maximum current

Momentary

Continuous

60 Hz

115 V ~

240 V ~

3.5 A

1.6 A

4

5

DX-D 600

GENERATORS

Generator model SHF 345

(CR only)

Input power 3-phase, 480 v 60 Hz

SHF 545 SHF 645 SHF 845

Max. power (kW) 32 50 64 80

Max. mA 400 640 640 800

Power output

@ 0.1s

Compatible X-ray All

Tubes

640 mA @ 78 kVp

500 mA @ 100 kVp

400 mA @ 125 kVp

320 mA @ 150 kVp

All

640 mA @ 100 kVp

500 mA @ 128 kVp

400 mA @ 150 kVp

E7252X, E7254FX,

E7869X

800 mA @ 100 kVp

640 mA @ 125 kVp

500 mA @ 150 kVp

E7254FX, E7869X

Range of radiographic parameters

mA From 10 mA to 800 mA through the following mA stations: 10, 12.5, 16, 20, 25, 32, 40, 50, 64,

80, 100, 125, 160, 200, 250, 320, 400, 500, 640, 800. (Depending on the Generator model)

mAs

ms

AEC

Product of mA x time values from 0.1 mAs to 500 mAs

From 1 to 10,000 milliseconds through the following time stations: 1, 2, 3, 4, 5, 6, 8, 10, 12, 16,

20, 25, 32, 40, 50, 64, 80, 100, 125, 160, 200, 250, 320, 400, 500, 640, 800, 1,000, 1,250,

1,600, 2,000, 2,500, 3,200, 4,000, 5,000, 6,400, 8,000 and 10,000.

mAs: 0.1 mAs to 500 mAs exposure time: Nominal shortest irradiation time = 1 ms

X-RAY TUBES

Housing Focal spot

Toshiba E7239X 1.0 - 2.0

Toshiba E7876X 0.6 - 1.2

Toshiba E7884X 0.6 - 1.2

Toshiba E7252X 0.6 - 1.2

Toshiba E7254FX 0.6 - 1.2

Toshiba E7869X 0.6 - 1.2

Target angle

16°

12°

12°

12°

12°

12°

Heat capacity (kHU)

140

230

300

300

400

600

Anode speed

Low

Low

Low

High/Low

High/Low

High/Low

DX-D 600

Why Agfa HealthCare?

Agfa HealthCare is a global leader in the fast growing market of integrated IT and imaging systems, offering healthcare facilities a seamless flow of information and a 360 º view of patient care. The company has a unique, holistic approach, enabling it to provide in-depth clinical know-how and fully integrated hospital-wide solutions.

These specialized solutions integrate IT and imaging systems for Radiology, Cardiology, Mammography* and

Orthopaedics. Agfa HealthCare’s enterprise-wide IT platform integrates administrative and clinical data within a healthcare facility and is designed to match the unique needs of healthcare professionals.

www.agfahealthcare.com

Agfa and the Agfa rhombus are trademarks of Agfa-Gevaert N.V., Belgium, or its affiliates. MUSICA and DX are trademarks of Agfa HealthCare NV, Belgium or its affiliates. All rights reserved. All other trademarks are held by their respective owners and are used in an editorial fashion with no intention of infringement. The data in this publication are for illustration purposes only and do not necessarily represent standards or specifications, which must be met by Agfa HealthCare. All information contained herein is intended for guidance purposes only, and characteristics of the products and services described in this publication can be changed at any time without notice. Products and services may not be available for your local area. Please contact your local sales representative at agfa.com for availability information. Agfa HealthCare diligently strives to provide as accurate information as possible, but shall not be responsible for any typographical error.

*NOTE: The DX-D 600 is not indicated for use in mammography.

© 2013 Agfa HealthCare Corp

All rights reserved

Published by

Agfa HealthCare Corp.

10 South Academy Street

Greenville, SC 29601 USA

5SJ9F US 00201303

DX-D 600

D I R E C T R A D I O G R A P H Y S Y S T E M

DX-D 600

HIGH-PRODUCTIVITY, EXCELLENT IMAGE QUALITY,

DIRECT RADIOGRAPHY SYSTEM WITH USER-

FRIENDLY DESIGN IN THREE CONFIGURATION

OPTIONS.

• Single or Dual-detector, high-productivity, highthroughput general radiography system with three configuration options: manual, auto-tracking and fully automatic “move to” positioning

• Excellent contrast detail provided by MUSICA² processing, producing exam-independent consistent image quality

• DICOM connectivity to PACS and RIS

• Wireless Cesium Iodide DR detector technology with columnar crystal growth for higher sensitivity

(DQE)

• Option to integrate with Agfa HealthCare CR systems for maximum versatility

The DX-D 600 Direct Radiography system combines userfriendly design with Agfa HealthCare’s MUSICA² image quality to create this high-productivity, direct radiography

(DR) solution. The DX-D 600 is family of ceiling mounted

X-ray systems with configurations ranging from a system using manual movement to a fully motorized, autopositioning solution, ideal for facilities with a high patient load that are looking to streamline workflow to increase patient comfort. The fully integrated NX Workstation communicates seamlessly with PACS and RIS, and eliminates the need for a separate generator console. Both APR and

X-ray parameters are captured in the DICOM header after exposure and become part of the patient demographics for future reference.

The DX-D 600 features Cesium Iodide detector technology, which offers excellent image quality and fast image availability. GOS technology (Gadolinium Oxy-Sulphide) is also available. Both detector types are available in fixed or removable formats for versatility, including wireless.

Agfa HealthCare’s proprietary MUSICA² image processing software delivers image consistency for improved workflow and excellent contrast detail to help enhance diagnostic confidence.

5

DX-D 600

GENERATORS

Generator model SHF 345

(CR only)

Input power 3-phase, 480 v 60 Hz

SHF 545 SHF 645 SHF 845

Max. power (kW) 32 50 64 80

Max. mA 400 640 640 800

Power output

@ 0.1s

Compatible X-ray All

Tubes

640 mA @ 78 kVp

500 mA @ 100 kVp

400 mA @ 125 kVp

320 mA @ 150 kVp

All

640 mA @ 100 kVp

500 mA @ 128 kVp

400 mA @ 150 kVp

E7252X, E7254FX,

E7869X

800 mA @ 100 kVp

640 mA @ 125 kVp

500 mA @ 150 kVp

E7254FX, E7869X

Range of radiographic parameters

mA From 10 mA to 800 mA through the following mA stations: 10, 12.5, 16, 20, 25, 32, 40, 50, 64,

80, 100, 125, 160, 200, 250, 320, 400, 500, 640, 800. (Depending on the Generator model)

mAs

ms

AEC

Product of mA x time values from 0.1 mAs to 500 mAs

From 1 to 10,000 milliseconds through the following time stations: 1, 2, 3, 4, 5, 6, 8, 10, 12, 16,

20, 25, 32, 40, 50, 64, 80, 100, 125, 160, 200, 250, 320, 400, 500, 640, 800, 1,000, 1,250,

1,600, 2,000, 2,500, 3,200, 4,000, 5,000, 6,400, 8,000 and 10,000.

mAs: 0.1 mAs to 500 mAs exposure time: Nominal shortest irradiation time = 1 ms

X-RAY TUBES

Housing Focal spot

Toshiba E7239X 1.0 - 2.0

Toshiba E7876X 0.6 - 1.2

Toshiba E7884X 0.6 - 1.2

Toshiba E7252X 0.6 - 1.2

Toshiba E7254FX 0.6 - 1.2

Toshiba E7869X 0.6 - 1.2

Target angle

16°

12°

12°

12°

12°

12°

Heat capacity (kHU)

140

230

300

300

400

600

Anode speed

Low

Low

Low

High/Low

High/Low

High/Low

DX-D 600

Why Agfa HealthCare?

Agfa HealthCare is a global leader in the fast growing market of integrated IT and imaging systems, offering healthcare facilities a seamless flow of information and a 360 º view of patient care. The company has a unique, holistic approach, enabling it to provide in-depth clinical know-how and fully integrated hospital-wide solutions.

These specialized solutions integrate IT and imaging systems for Radiology, Cardiology, Mammography* and

Orthopaedics. Agfa HealthCare’s enterprise-wide IT platform integrates administrative and clinical data within a healthcare facility and is designed to match the unique needs of healthcare professionals.

www.agfahealthcare.com

Agfa and the Agfa rhombus are trademarks of Agfa-Gevaert N.V., Belgium, or its affiliates. MUSICA and DX are trademarks of Agfa HealthCare NV, Belgium or its affiliates. All rights reserved. All other trademarks are held by their respective owners and are used in an editorial fashion with no intention of infringement. The data in this publication are for illustration purposes only and do not necessarily represent standards or specifications, which must be met by Agfa HealthCare. All information contained herein is intended for guidance purposes only, and characteristics of the products and services described in this publication can be changed at any time without notice. Products and services may not be available for your local area. Please contact your local sales representative at agfa.com for availability information. Agfa HealthCare diligently strives to provide as accurate information as possible, but shall not be responsible for any typographical error.

*NOTE: The DX-D 600 is not indicated for use in mammography.

© 2013 Agfa HealthCare Corp

All rights reserved

Published by

Agfa HealthCare Corp.

10 South Academy Street

Greenville, SC 29601 USA

5SJ9F US 00201303

DX-D 600

D I R E C T R A D I O G R A P H Y S Y S T E M

DX-D 600

HIGH-PRODUCTIVITY, EXCELLENT IMAGE QUALITY,

DIRECT RADIOGRAPHY SYSTEM WITH USER-

FRIENDLY DESIGN IN THREE CONFIGURATION

OPTIONS.

• Single or Dual-detector, high-productivity, highthroughput general radiography system with three configuration options: manual, auto-tracking and fully automatic “move to” positioning

• Excellent contrast detail provided by MUSICA² processing, producing exam-independent consistent image quality

• DICOM connectivity to PACS and RIS

• Wireless Cesium Iodide DR detector technology with columnar crystal growth for higher sensitivity

(DQE)

• Option to integrate with Agfa HealthCare CR systems for maximum versatility

The DX-D 600 Direct Radiography system combines userfriendly design with Agfa HealthCare’s MUSICA² image quality to create this high-productivity, direct radiography

(DR) solution. The DX-D 600 is family of ceiling mounted

X-ray systems with configurations ranging from a system using manual movement to a fully motorized, autopositioning solution, ideal for facilities with a high patient load that are looking to streamline workflow to increase patient comfort. The fully integrated NX Workstation communicates seamlessly with PACS and RIS, and eliminates the need for a separate generator console. Both APR and

X-ray parameters are captured in the DICOM header after exposure and become part of the patient demographics for future reference.

The DX-D 600 features Cesium Iodide detector technology, which offers excellent image quality and fast image availability. GOS technology (Gadolinium Oxy-Sulphide) is also available. Both detector types are available in fixed or removable formats for versatility, including wireless.

Agfa HealthCare’s proprietary MUSICA² image processing software delivers image consistency for improved workflow and excellent contrast detail to help enhance diagnostic confidence.

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project