Rosemount OCX 8800 O2 / Combustibles Transmitter General Purpose Owner's Manual

Add to my manuals
166 Pages

advertisement

Rosemount OCX 8800 O2 / Combustibles Transmitter General Purpose Owner's Manual | Manualzz
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
Oxygen/Combustibles Transmitter
General Purpose OCX 8800 Oxygen/Combustibles Analyzer with Remote Electronics
http://www.raihome.com
HIGHLIGHTS OF CHANGES
Effective December 2007, Rev 1.1
Page/Section
Summary
Throughout
Updated manual number suffixes in page headers to indicate OCX configuration.
Changed references to HART Communicator te read Field Communicator in text and
illustrations.
Page 1-10
Added new illustration Figure 1-6
Page 1-14
Updated Product Matrices.
Page 1-15
Revised/updated Table 1-2 .
Page 2-2
Deleted obsolete CAUTION statement.
Page 2-14
Revised paragraph 4 listing of gases for calibration.
Page 2-18
Revised Figure 2-11 to indicate minimum pressure for actuation of the blowback valve.
Page 2-19
Revised Figure 2-12 to indicate minimum pressure for actuation of the blowback valve.
Page 4-4
Revised 1st paragraph.
Page 4-5
Revised HART menu items in Figure 4-3, Sheet 1 of 4
Page 4-9
Added new step 4 to both D/A Trim Procedures.
Page 5-2
Replaced steps 6 through 10 with a repeat of step 4 instruction.
Page 5-3
Revised steps 2 and 3.
Page 7-3
Revised "Ref Current Err" fault indication heading.
Page 7-7
Revised probable cause statements of "Cal Warning/Cal Failed" fault indication.
Page 8-5
Revised parts list for Sensor Housing Components.
Appendix A
Updated multi-lingual safety pages.
Effective July 2008, Rev 1.2
Page/Section
Summary
General
Updated manual to synchronize revision levels.
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
Table of Contents
Essential Instructions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . i
Preface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ii
Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ii
Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ii
SECTION 1
Description and
Specifications
Component Checklist. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
System Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-12
SECTION 2
Installation
Mechanical
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Electrical Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8
Pneumatic Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-14
Initial Startup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-20
SECTION 3
Configuration and
Startup
Verify Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Initial Power Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Set Test Gas Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
OCX 8800 Reset Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
SECTION 4
Using HART
Communications
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Field Communicator Signal Connections. . . . . . . . . . . . . . . . . . . . . . . 4-1
Field Communicator PC Connections . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
HART Menu Tree. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
D/A Trim Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
SECTION 5
Calibration
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Fully Automatic Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Operator - Initiated Autocalibration . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Manual Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
SECTION 6
Maintenance and Service
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
OCX 8800 Removal
and Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
OCX with
Remote Electronics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
Repair Sensor Housing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-8
Sensor Housing Disassembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-8
Sensor Housing Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-18
Repair Electronics Housing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-28
Electronics Housing Disassembly . . . . . . . . . . . . . . . . . . . . . . . . 6-28
http://www..raihome.com
Instruction Manual
OCX 8800
IM-106-880ASB, Rev 1.2
July 2008
Electronics Housing Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-30
Replace Tube Fittings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-34
Remove Tube Fittings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-34
Install Tube Fittings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-35
SECTION 7
Troubleshooting
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
Grounding. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
Electrical Noise. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
Electrostatic Discharge. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
Total Power Loss . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-2
Diagnostic Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-2
Fault Isolation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-3
Alarm Relay Events . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-11
SECTION 8
Replacement Parts
Sensor Housing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-2
Electronics Housing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-6
O2 Cell and Heater Strut Assembly. . . . . . . . . . . . . . . . . . . . . . . . . . . 8-9
APPENDIX A
Safety Data
Safety Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-2
Safety Data Sheet for Ceramic Fiber Products . . . . . . . . . . . . . . . . . A-24
High Pressure Gas Cylinders . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-30
APPENDIX B
SPA with HART Alarm
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-1
Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-1
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-2
Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-2
APPENDIX C
Return of Materials
Returning Material . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-1
TOC-2
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
OCX 8800
Oxygen/Combustibles Transmitter
ESSENTIAL
INSTRUCTIONS
READ THIS PAGE BEFORE PROCEEDING!
Emerson Process Management designs, manufactures and tests its products
to meet many national and international standards. Because these
instruments are sophisticated technical products, you MUST properly install,
use, and maintain them to ensure they continue to operate within their
normal specifications. The following instructions MUST be adhered to and
integrated into your safety program when installing, using, and maintaining
Emerson’s Rosemount Analytical products. Failure to follow the proper
instructions may cause any one of the following situations to occur: Loss of
life; personal injury; property damage; damage to this instrument; and
warranty invalidation.
• Read all instructions prior to installing, operating, and servicing the
product.
• If you do not understand any of the instructions, contact your
Emerson Process Management representative for clarification.
• Follow all warnings, cautions, and instructions marked on and
supplied with the product.
• Inform and educate your personnel in the proper installation,
operation, and maintenance of the product.
• Install your equipment as specified in the Installation Instructions
of the appropriate Instruction Manual and per applicable local and
national codes. Connect all products to the proper electrical and
pressure sources.
• To ensure proper performance, use qualified personnel to install,
operate, update, program, and maintain the product.
• When replacement parts are required, ensure that qualified people use
replacement parts specified by Emerson Process Management.
Unauthorized parts and procedures can affect the product's
performance, place the safe operation of your process at risk, and
VOID YOUR WARRANTY. Look-alike substitutions may result in fire,
electrical hazards, or improper operation.
• Ensure that all equipment doors are closed and protective covers
are in place, except when maintenance is being performed by
qualified persons, to prevent electrical shock and personal injury.
The information contained in this document is subject to change without
notice.
If a Model 275/375 Universal HART® Communicator is used with this unit, the software
within the Model 275/375 may require modification. If a software modification is required,
please contact your local Emerson Process Management Service Group or National
Response Center at 1-800-654-7768.
http://www..raihome.com
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
PREFACE
The purpose of this manual is to provide a comprehensive understanding of
the OCX 8800 components, functions, installation, and maintenance.
We recommend that you thoroughly familiarize yourself with the Introduction
and Installation sections before installing your transmitter.
The introduction presents the basic principles of the transmitter along with its
performance characteristics and components. The remaining sections contain
detailed procedures and information necessary to install and service the
transmitter.
Before contacting Emerson Process Management concerning any questions,
first consult this manual. It describes most situations encountered in your
equipment's operation and details necessary action.
DEFINITIONS
The following definitions apply to WARNINGS, CAUTIONS, and NOTES
found throughout this publication.
Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not
strictly observed, could result in injury, death, or long-term health hazards of personnel.
Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not
strictly observed, could result in damage to or destruction of equipment, or loss of
effectiveness.
NOTE
Highlights an essential operating procedure, condition, or statement.
SYMBOLS
: EARTH (GROUND) TERMINAL
: PROTECTIVE CONDUCTOR TERMINAL
: RISK OF ELECTRICAL SHOCK
: WARNING: REFER TO INSTRUCTION BULLETIN
NOTE TO USERS
The number in the lower right corner of each illustration in this publication is a
manual illustration number. It is not a part number, and is not related to the
illustration in any technical manner.
ii
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
Section 1
OCX 8800
Description and Specifications
Component Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 1-1
System Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 1-1
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 1-12
COMPONENT
CHECKLIST
A typical OCX 8800 Oxygen/Combustibles Transmitter package should
contain the items shown in Figure 1-1.
Use the product matrix in Table 1-1 at the end of this section to verify your
order number. The first part of the matrix defines the model. The last part
defines the various options and features of the OCX 8800. Check the model
number against the transmitter features and options, making sure options
specified by this number are on or included with the unit. Use this complete
model number for any correspondence with Emerson Process Management.
A list of accessories for use with the OCX 8800 is provided in Table 1-2.
SYSTEM OVERVIEW
Scope
This Instruction Manual supplies details needed to install, startup, operate,
and maintain the OCX 8800. Signal conditioning electronics outputs separate
4-20 mA signals representing oxygen (O2) and combustibles (COe) values.
This information, plus additional details, can be accessed with the HART
Model 275/375 handheld communicator or Rosemount Analytical AMS
software.
System Description
The OCX 8800 is designed to measure oxygen and combustible
concentrations in flue gas temperatures up to 2600°F (1427°C). Electrical
connections, power and communications are made through two 3/4 NPT ports
in the flameproof electronics enclosure using fittings and cables provided by
the customer. Cable installation must meet NEC, IEC and/or other applicable
national or local codes for Class I, Zone 1, Group IIB +H2 T3/T6 permanently
mounted equipment. The transmitter is close coupled to the process and
requires minimal sample conditioning requirements.
http://www..raihome.com
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
Figure 1-1. Typical System Package
1
An
aly
tic
al
MAN 4275A00
English
October 1994
4
Communicator
HART
o
FISHER-ROSEMOUNTTM
2
3
1. Instruction Manual
2. Field Communicator Package (optional)
3. Adapter Plate with Mounting Hardware and Gasket
4. Reference Air and Calibration Set
5. Blowback Hardware (optional)
6. OCX 8800 with Remote Electronics
1-2
5
37390076
6
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
The equipment measures oxygen percentage by reading the voltage
developed across a heated electrochemical cell, which consists of a small
yttria-stabilized, zirconia disc. Both sides of the disc are coated with porous
metal electrodes. When operated at the proper temperature, the millivolt
output of the cell is given by the following Nernst equation:
EMF = KT log10 (P1/P2) + C
Where:
1. P2 is the partial pressure of the oxygen in the measured gas on one side
of the cell.
2. P1 is the partial pressure of the oxygen in the reference air on the
opposite side of the cell.
3. T is the absolute temperature.
4. C is the cell constant.
5. K is an arithmetic constant.
NOTE
For best results, use clean, dry instrument air (20.95% oxygen) as the
reference air.
When the cell is at operating temperature and there are unequal oxygen
concentrations across the cell, oxygen ions will travel from the high oxygen
partial pressure side to the low oxygen partial pressure side of the cell. The
resulting logarithmic output voltage is approximately 50 mV per decade. The
output is proportional to the inverse logarithm of the oxygen concentration.
Therefore, the output signal increases as the oxygen concentration of the
sample gas decreases. This characteristic enables the OCX 8800 to provide
exceptional sensitivity at low oxygen concentrations.
The OCX 8800 measures net oxygen concentration in the presence of all the
products of combustion, including water vapor. Therefore, it may be
considered an analysis on a "wet" basis. In comparison with older methods,
such as the portable apparatus, which provides an analysis on a "dry" gas
basis, the "wet" analysis will, in general, indicate a lower percentage of
oxygen. The difference will be proportional to the water content of the
sampled gas stream.
The OCX 8800 combustibles sensor is a catalytic sensor consisting of two
Resistance Devices (RTD). One RTD is the reference element covered with
an inert coating. The other RTD element is active, coated with a catalyst. As
the sample gases flow by the sensor, the combustible gases oxidize on the
surface of the active element. The oxidation that occurs produces heat and a
temperature rise in the active element. The temperature difference produces
a resistance relationship between the two elements that is directly
proportional to the concentration of combustibles in the sample gases.
1-3
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
The catalyst is specifically designed to detect carbon monoxide (CO), but the
sensor responds to other combustible gases. The sensor is calibrated using
CO, thus the output should be expressed in terms of CO. However, since the
sensor detects other combustible gases, the output cannot just be labeled
CO. The response of the sensor to other combustible gases gives an output
that is equivalent to the sensor detecting CO. The term COe is used in this
manual to describe the sensor output. This term indicates that the sensor is
calibrated in terms of CO, and that the sensor output is equivalent to CO but
not specific to CO.
Dilution air is provided to the COe sensor to ensure there is adequate oxygen
to fully oxidize any combustible gases regardless of the concentration of
oxygen in the process.
System Configuration
Transmitters are available in four lengths, giving the user the flexibility to use
a penetration appropriate to the size of the stack or duct. The length options
are 18 in. (457 mm), 3 ft (0.91 m), 6 ft (1.83 m), or 9 ft (2.7 m). Probes are
available in three material options, 316L stainless steel, inconel 600, and
ceramic to accommodate higher temperatures.
The electronics are contained in a separate housing from the sensors. The
electronics housing may be mounted up to 150 feet away from the sensor
housing.
The electronics control both sensor temperatures and provide individual 4-20
mA isolated outputs that are proportional to the measured oxygen and
combustibles concentrations. The power supply can accept voltages of 100 to
240 VAC and 50 to 60 Hz. The electronics accepts millivolt signals generated
by the sensors and produces the outputs to be used by remotely connected
devices. The outputs are isolated 4-20 mA linearized currents. Refer to
Section 3, Configuration and Startup for specific instructions upon initial
power up.
System Features
1. The O2 cell output voltage and sensitivity increase as the oxygen
concentration decreases.
2. HART communication is standard. To use the HART capability, you
must have either:
a. Model 275/375 Field Communicator.
b. AMS software for the PC.
3. Oxygen cell and heater/thermocouple assembly are field replaceable.
4. Electronics are automatically configured for line voltages from 100 to
240 VAC.
1-4
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
5. An operator can calibrate and diagnostically troubleshoot the OCX 8800
using the HART Interface. Each of the OCX 8800's 4-20 mA output lines
transmit an analog signal proportional to oxygen or combustible levels
detected. The HART output is superimposed on the oxygen 4-20 mA
output line only. This information can be accessed through the following:
•
Model 275/375 Handheld Communicator - The handheld
communicator requires Device Description (DD) software specific
to the OCX 8800. The DD software will be supplied with many
Model 275/375 units, but can also be programmed into existing
units at most Emerson Process Management service offices.
Refer to Section 4, Using HART Communications, for additional
information.
•
Personal Computer (PC) - The use of a personal computer
requires AMS software available from Emerson Process
Management.
•
Selected Distributed Control Systems - The use of distributed
control systems requires input/output (I/O) hardware and AMS
software which permit HART communications.
6. Optional Blowback System. The blowback system periodically blows
instrument air back through the sample line filter and out the sample
tube. This clears out particulate and keeps the sample line filter from
clogging.
System Operation
Figure 1-2 shows the relationship between the components of the OCX 8800.
The sensors and the electronics are contained in separate housings. The
sensor housing and probe mounts to a duct or process wall so that the probe
protrudes into the flue gas stream. An air powered eductor continuously pulls
samples of the process flue gas through the probe to a chamber in front of the
sensor housing where the sample passes the O2 sensor and continues on to
the COe sensor. Dilution air is provided to the COe sensor and reference air
to the O2 sensor. After the gas sample flows past the O2 sensor and through
the COe sensor, it is drawn through the eductor where it mixes with the
eductor air and exits through exhaust back into the system. The electronics
housing contains the CPU and HART boards which convert the sensor inputs
into 4-20 mA analog output signals. The CPU can also initiate and perform
calibrations. Three test gasses and instrument air can be turned on and off by
solenoids. Test gas flow to the sensors is regulated by a flow meter between
the electronics and sensor housings. Instrument air is separated into eductor
air, reference air, and dilution air. The instrument air solenoid does not allow
air flow until the heaters are up to temperature. This minimizes the amount of
sampled process flue gas being pulled into cold sensors causing
condensation.
1-5
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
Figure 1-2. System Operation Diagram
SENSOR
HOUSING
ELECTRONICS
HOUSING
CPU
COe
Combustibles
Sensor
HART
Board
Power
Supply
Probe
Sample
Gas
Optional
Test Gas
Solenoids
O2
Sensor
Eductor
Instrument Air
Solenoid
Low O2
Test Gas
High O2
Test Gas
CO
Test Gas
Instrument
Air
Flow Meter
7 scfh
Eductor Air
Reference Air
Dilution Air
Flow Meter
50 cc/min.
(0.1 scfh)
Handling the OCX 8800
It is important that printed circuit boards and integrated circuits are handled only when
adequate antistatic precautions have been taken to prevent possible equipment damage.
The OCX 8800 is designed for industrial application. Treat each component of the system
with care to avoid physical damage. The probe may contain components made from
ceramics, which are susceptible to shock when mishandled.
System Considerations
Prior to installing your OCX 8800, make sure you have all the components
necessary to make the system installation. Ensure all the components are
properly integrated to make the system functional.
1-6
37390001
Exhaust
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
After verifying that you have all the components, select mounting locations
and determine how each component will be placed in terms of available line
voltage, ambient temperatures, environmental considerations, convenience,
and serviceability. Figure 1-3 shows a typical system wiring. Simplified
installations for the OCX 8800 are shown in Figure 1-4. Figure 1-6 shows the
dimensions for the optional sample tube support.
A source of instrument air is required at the OCX 8800 for reference air,
dilution air, and eductor air. Since the OCX 8800 is equipped with an in-place
calibration feature, provision should be made for connecting test gas tanks to
the OCX 8800 when it is to be calibrated.
NOTE
The electronics module is designed to meet NEMA 4 (IP66) and the electronic
components are rated to temperatures up to 185°F (8 5°C).
Retain packaging in which the unit arrived from the factory in case any
components are to be shipped to another site. This packaging has been
designed to protect the product.
Figure 1-3. OCX 8800 HART
Connections and AMS Application
1-7
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
Figure 1-4. Typical System
Installation
Gases
Duct
OCX 8800 with
REMOTE
ELECTRONICS
Stack
Heater
Power Cable
[up to 150 ft (46 m)]
Signal Cable
[up to 150 ft (46 m)]
Test Gas
Flow Meter
Line Voltage
Instrument Air
Pressure
Supply
Regulator (Reference Gas)
High O2 Test Gas
Low O2 Test Gas
CO Test Gas
Figure 1-5. OCX 8800 HART
1-8
37390064
Dilution
Air
Flow
Meter
4-20 mA Outputs
(2 Twisted Pairs)
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
Connections and AMS Application
1-9
Instruction Manual
OCX 8800
Figure 1-6. Sample Tube Support
1-10
IM-106-880ASB, Rev 1.2
July 2008
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
Figure 1-7. Typical System
Installation
Gases
Duct
OCX 8800 with
REMOTE
ELECTRONICS
Stack
Heater
Power Cable
[up to 150 ft (46 m)]
Signal Cable
[up to 150 ft (46 m)]
Test Gas
Flow Meter
Line Voltage
Instrument Air
Pressure
Supply
Regulator (Reference Gas)
High O2 Test Gas
Low O2 Test Gas
CO Test Gas
37390064
Dilution
Air
Flow
Meter
4-20 mA Outputs
(2 Twisted Pairs)
1-11
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
SPECIFICATIONS
General Purpose OCX
Specifications
Net O2 Range
Combustibles
Accuracy
Oxygen
Combustibles
System Response to
Test Gas
Oxygen
Combustibles
Temperature Limits
Process
Sensors Housing
Electronics Housing
Nominal and Approximate
Shipping Weights
18 in. (457 mm)
probe package
3 ft (0.91 m) probe
package
6 ft (1.83 m) probe
package
9 ft (2.74 m) probe
package
Mounting and Mounting
Positions
Sensors Housing
Electronics Housing
Materials
Probes
Enclosures
Calibration
Calibration Gas Mixtures
Recommended
(Ref. test gas bottles
kit #1A99119G04)
Calibration Gas Flow
Reference Air
Eductor Air
Dilution Air
Blowback Air (optional)
1-12
0-1% to 0-40% O2, fully field selectable
0-1000 ppm to 0-5%, fully field selectable
± 0.75% of reading or 0.05% O2 (whichever is greater)
± 2% range
10 sec T90
25 sec T90
32° to 2600°F (0° to 1427°C)
-40° to 212°F (-40° to 100°C), ambie nt
-40° to 149°F (-40° to 65°C), ambient
-40° to 185°F (-40° to 85°C), internal - operating temperature of
electronics inside instrument housing, as measured by a HART
communicator or AMS software
54 lbs (20 kg)
55 lbs (20.5 kg)
57 lbs (21 kg)
59 lbs (22 kg)
Flange
Wall/Pipe
316L stainless steel - 1300°F (704°C)
Inconel 600 - 1832°F (1000°C)
Ceramic - 2600°F (1427°C)
Low-copper aluminum
Semi-automatic or automatic
0.4% O2, Balance N2
8% O2, Balance N2
1000 ppm CO, Balance Air
7 scfh (3.3 l/m)
2 scfh (1 l/m), clean, dry instrument-quality air (20.95% O2),
regulated to 35 psi (241 kPa)
5 scfh (2.5 l/m), clean, dry, instrument-quality air 20.95% O2),
regulated to 35 psi (241 kPa)
0.1 scfh (0.5 l/m), clean, dry, instrument-quality air (20.95% O2)
regulated to 35 psi (241 kPa)
Clean, dry, instrument-quality air (20.95% O2), regulated to 55 psi
(379 kPa)
Table continued on next page
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
Specifications
Sensors Housing
Electronics Housing
Electrical Noise
Certifications
NEMA 4, IP66 with fitting and pipe on reference exhaust port to
clean, dry atmosphere, two 3/4-14 NPT conduit ports
NEMA 4, IP66 with fitting and pipe on reference exhaust port to
clean, dry atmosphere, two 3/4-14 NPT conduit ports
EN 61326-1, Class A
C
Line Voltage
Pollution Degree
Over Voltage Category
Relative Humidity
Isolated Output
Oxygen
Combustibles
Alarm
Power Consumption
Mounting and Mounting
Positions
Sensor Housing
Electronics Housing
US
APPROVED
Universal 100 to 240 VAC ±10%, 50 to 60 Hz, no switches or
jumpers required, 3/4-14 NPT conduit port
2
II
5 to 95% (non-condensing)
4-20 mAdc, 950 ohm maximum, with HART capability
4-20 mAdc, 950 ohm maximum
Alarm output relay - dry contact, form C, 30mA, 30VDC capacity
750 W maximum
Flange
Wall/Pipe
NOTE
All static performance characteristics are with operating variables constant. Specifications subject to change
without notice.
1-13
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
Table 1-1. Product Matrix - General Purpose OCX 8800
OCX88A
O2/Combustibles Transmitter
Code
00
11
21
31
12
22
32
13
23
14
24
Probe Length and Material
No Probe or Exhaust Tube
18 in. (457 mm) 316 SST
18 in. (457 mm) Inconel 600
18 in. (457 mm) Ceramic
3 ft (0.91 m) 316 SST
3 ft (0.91 m) Inconel 600
3 ft (0.91 m) Ceramic
6 ft (1.83 m) 316 SST
6 ft (1.83 m) Inconel 600
9 ft (2.7 m) 316 SST
9 ft (2.7 m) Inconel 600
Code
10
20
up to 1300°F (704°C)
up to 1832°F (1000°C)
up to 2600°F (1427°C)
up to 1300°F (704°C)
up to 1832°F (1000°C)
up to 2600°F (1427°C)
up to 1300°F (704°C)
up to 1832°F (1000°C)
up to 1300°F (704°C)
up to 1832°F (1000°C)
(3)
Mounting Hardware - Probe Side
No Adapter Plate
Probe Only (ANSI)
Probe Only (DIN)
Code
H1
H2
H3
H4
Electronics Housing - Communications
HART Communications
HART Communications with Local Operator Interface
HART Communications with Calibration Solenoids
HART Communications with Local Operator Interface and Calibration Solenoids
Code
01
02
03
04
05
06
07
08
Electronics Mounting
Integral to Sensor Housing
Remote Electronics and no cable
Remote Electronics and 20 ft (6 m) cable
Remote Electronics and 40 ft (12 m) cable
Remote Electronics and 60 ft (18 m) cable
Remote Electronics and 80 ft (24 m) cable
Remote Electronics and 100 ft (30 m) cable
Remote Electronics and 150 ft (46 m) cable
Code
00
01
02
03
11
12
13
10
(3)
Mounting Hardware - Stack Side
No Adapter Plate (“0” must be chosen under “Mounting Adapter - Probe Side” below)
New Installation - Square weld plate with studs
Model 218/240 Mounting Plate (with Model 218/240 Shield Removed)
Existing Model 218/240 Support Shield
Special Mounting(1)
Model 132 Adapter Plate
Code
0
1
2
11
(3)
Probe Mounting Assembly
(ANSI 2 in. 150 lb) 6" dia. flange, 4.75" BC with 4 x 0.75" dia. holes
(DIN) 185 mm dia. flange, 145 mm BC with 4 x 18 mm dia. holes
Code
0
1
2
3
4
5
OCX88A
(3)
1
1
H3
06
Accessories
None
Flow meters & Ref. Air Set
In-Situ Filter (Stainless Steel only)
In-Situ Filter (SST), Flow meters & Ref. Air Set
Flow meters, and Ref. Air Set with Blowback
In-Situ Filter (SST) with Blowback
In-Situ Filter (SST), Flow meters & Ref. Air Set with Blowback
02
Example
NOTES:
(1)
Provide details of the existing mounting plate as follows:
Plate with studs
Plate without studs
(2)
(3)
1-14
Bolt circle diameter, number, and arrangement of studs, stud thread, stud height above mounting plate.
Bolt circle diameter, number, and arrangement of holes, thread, depth of stud mounting plate with accessories.
For use with stainless steel sample tube only.
For high temperature applications that require a filter, please order 1A99762H03 separately.
(2)
(2)
(2)
(2)
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
Table 1-2. Accessories
OCX 8800
PART NUMBER
DESCRIPTION
1A99119H01
Oxygen test gas bottle; 0.4% O2, balance N2
1A99119H02
Oxygen test gas bottle; 8.0% O2, balance N2
1A 99119H07
CO test gas bottle; 1000 ppm CO, balance air
1A99120H02
Regulator for Oxygen (may need 2)
1A99120H03
Regulator for CO test gas
1A99119G06
Wall mount bracket for test gas bottles
1A99119G05
Test gas regulators kit
1A99119G04
Test gas bottles kit
1A99292H01
Moore Industries SPA for Low O2 Alarm, High COe Alarm,
Calibration Status, and Unit Fail
1A99339H03
Blowback valve, air operated
1A99762H03
Hasteloy In Situ Filter, High Temperature
1A99784H02
375 Field Communicator with 12 Megabyte buffer,
model no. 375HR1EKLU
6A00171G01
Power line filter kit
6A00288G01
Sample Tube Support, 18 in. (457 mm)
6A00288G02
Sample Tube Support, 3 Ft. (0.91 m)
6A00288G02
Sample Tube Support, 6 Ft. (1.83 m)
6A00288G04
Sample Tube Support, 9 Ft. (2.7 m)
6P00162H02
Flange Insulator
1-15
Instruction Manual
OCX 8800
1-16
IM-106-880ASB, Rev 1.2
July 2008
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
Section 2
OCX 8800
Installation
Mechanical Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . page 2-2
Electrical Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 2-8
Pneumatic Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 2-14
Initial Startup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 2-20
Before installing this equipment, read the "Safety instructions for the wiring and installation
of this apparatus" in Appendix A: Safety Data. Failure to follow the safety instructions could
result in serious injury or death.
The OCX88A can be installed in general purpose areas only. Do not install the OCX88A in
hazardous areas.
http://www..raihome.com
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
MECHANICAL
INSTALLATION
Selecting Location
1. The location of the OCX 8800 in the stack or flue is most important for
maximum accuracy in the oxygen analyzing process. The probe must
be positioned so the gas it measures is representative of the process.
Best results are normally obtained if the transmitter is positioned near
the center of the duct (40-60% insertion). Longer ducts may require
several transmitters since the oxygen and combustibles can vary due to
stratification. A point too near the wall of the duct or the inside radius of
a bend, may not provide a representative sample because of the very
low flow conditions. The sensing point should be selected so the
process gas temperature falls within the range of probe material used.
Figure 2-1 through Figure 2-4 provide mechanical installation
references. The ambient temperature inside the electronics housing
must not exceed 185°F (85°C).
2. Check the flue or stack for holes and air leakage. The presence of this
condition will substantially affect the accuracy of the oxygen and
combustibles readings. Therefore, either make the necessary repairs or
install the transmitter up stream of any leakage.
3. Ensure the area is clear of internal and external obstructions that will
interfere with installation and maintenance access to the unit. Allow
adequate clearance for the removal of the OCX 8800.
Do not allow the temperature of the electronics housing to exceed 185°F (85°C) or damage
to the electronics may result.
Whenever a positive stack pressure exists at the installation site, be sure to connect all
pneumatic lines prior to installing the OCX 8800 in the stack or ductwork. Failure to connect
the pneumatic lines can allow the flow of contaminants into the OCX 8800 ports.
Installation
1. Ensure all components are available to install the OCX 8800.
2. The OCX 8800 may be installed intact as it is received.
3. Weld or bolt adapter plate (Figure 2-2) onto the duct.
4. Use the pipe or wall mounting hardware as shown in Figure 2-3 to
mount the electronics housing. Choose a location not to exceed the
length of the electronics cable ordered.
5. Ensure the conduits drop vertically from the OCX 8800 and the conduit
is routed below the level of the conduit ports on the housing to form a
drip loop. Drip loops minimize the possibility that moisture will damage
the electronics (Figure 2-4).
6. Where a positive stack pressure exists at the installation site, connect all
pneumatic lines prior to installing the OCX 8800 in the stack or
ductwork.
2-2
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
NOTE
If process temperatures will exceed 392°F (200°C), use anti-seize compound
on stud threads to ease future removal of the OCX 8800.
7. Insert sample and exhaust tubes through the opening in the mounting
flange and bolt the unit to the flange.
Uninsulated stacks or ducts may cause ambient temperatures in the electronics housing to
exceed 185°F (85°C) and damage the electronics.
8. If insulation is removed to access the duct for OCX 8800 mounting,
make sure to replace insulation afterward.
Enclosures
The OCX 8800 enclosures are designed to meet ingress conditions of IP66.
Each enclosure cover is threaded to its base and sealed with an o-ring that
isolates the threads from external contaminants.
Each cover is secured by a clip attached to the base that engages the cover
between the ribs of the cover sidewall. The clip is held in place by an Allen
head cap screw and lockwasher mounted in a recess. Cover removal and
installation requires an Allen wrench to loosen and tighten the screw.
2-3
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
Figure 2-1. Installation, OCX 8800
NOTE
All dimensions are in inches with millimeters in parentheses.
Insulate if exposed to adverse weather or extreme temperature
changes, install a protective housing and/or insulation
around the unit.
Flange Dia.
B.C. Dia.
Hole Dia.
Table 1. Mounting Flange
DIN
ANSI
5R10244H01 5R10244H02
Flange
7.28
6.00
Dia.
(185)
(152)
0.75
0.71
Hole
Dia.
(19)
(18)
(4) Holes
4.75
5.71
equally
(121)
(145)
spaced on
B.C. dia
See Table 1
Removal Envelope
Dim “B”
Insertion Depth
Dim “A”
See Table 2
SENSOR HOUSING
Optional
In Situ FIilter
8.3
(211)
3 ft
6 ft
9 ft
Dim "A”
Dim "B”
18
(457)
36
(914)
72
(1829)
108
(2743)
34
(864)
52
(1321)
88
(2235)
124
(3150)
Heater Power Cable
18 in.
Signal Cable
Heater Power Cable
Probe
Signal Cable
Table 2. Installation/Removal
0.06 In. Thick Gasket
ANSI 3535B18H02
3535B45H01
DIN
Allow 9 in.
(229 mm) for
Cover Removal
37390009
ELECTRONICS HOUSING
2-4
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
Figure 2-2. Adapter Plate Installation
2-5
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
Figure 2-3. Wall or Pipe Mounting
of Electronics Housing
Note: * Indicates items are included
in mounting kit P/N 4851B40G01.
*Pipe
Mounting
Clamp
1/4-20 UNC x 2 in. M-F Standoffs.
Recommended for rear piping
clearance when wall mounting.
(4 each by customer)
*Wall or Pipe
Mounting Bracket
and Screws
2-6
Wall Mounting Bolts
or Lag Screws
(4 each by customer)
37020021
Electronics
Housing
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
Figure 2-4. Installation
with Drip Loops
Conduit Drip Loop
Duct Wall
Conduit Drip Loop
Duct Wall
37020004
Conduit Drip Loops
2-7
Instruction Manual
OCX 8800
ELECTRICAL
INSTALLATION
IM-106-880ASB, Rev 1.2
July 2008
All wiring must conform to local and national codes. For reference, factory
wired solenoid power connections are shown in Figure 2-5.
Disconnect and lock out power before connecting the unit to the power supply. Failure to
lock out power could result in serious injury or death.
Install all protective equipment covers and safety ground leads after installation. Failure to
install covers and ground leads could result in serious injury or death.
To meet the Safety Requirements of IEC 1010 (EC requirement), and ensure safe operation
of this equipment, connection to the main electrical power supply must be made through a
circuit breaker (min 10 A) in close proximity and marked for this equipment which will
disconnect all current-carrying conductors during a fault situation. This circuit breaker
should also include a mechanically operated isolating switch. If not, then another external
means of disconnecting the supply from the equipment should be located close by. Circuit
breakers or switches must comply with a recognized standard such as IEC 947.
The OCX88A can be installed in general purpose areas only. Do not install the OCX88A in
hazardous areas.
2-8
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
NOTE
To maintain proper earth grounding, ensure a positive connection exists
between the sensor housing, the electronics housing, and earth. The
connecting ground wire must be 14 AWG minimum. Refer to Figure 2-5.
NOTE
Line voltage, signal, and relay wiring must be rated for at least 105ºC (221ºF).
Electrical Connections
Electrical connections, power and communications are made to the electronic
enclosure. The connections are made through two 3/4 NPT ports in the
enclosure using fittings and cables provided by the customer. Cable
installation must meet NEC, IEC and/or other applicable national or local
codes for Class I, Zone 1, IIB +H2 T3/T6 permanently mounted equipment.
Connect Line Voltage
The OCX 8800 operates on 100 to 240 VAC line voltage at 50 to 60 Hz. The
power supply requires no setup. Connect the line (L wire) to the L terminal,
and the neutral (N wire) to the N terminal on the AC power input terminal
block in the electronics housing. Connect the ground (G wire) to the ground
stud in the electronics housing as shown in Figure 2-5.
Connect 4-20 mA Signals
Connect the 4-20 mA current loop to the 4-20 mA signal output terminals in
the electronics housing as shown in Figure 2-5. Use individual shielded
twisted wire pairs. Terminate the shield at the electronics housing.
O2 4-20 mA Signal
One 4-20 mA signal represents the O2 value. Superimposed on the O2
signal is the HART information accessible through a Model 275/375
Handheld Communicator or AMS software. The O2 signal is at the AOUT 1
terminals.
COe 4-20 mA Signal
Another 4-20 mA signal at the AOUT 2 terminals represents the COe
value. HART information is not available on the COe signal.
Alarm Output Relay
Connect any customer-supplied relay input to the alarm output relay terminal.
Use shielded wire and terminate the shield at the electronics housing. The
alarm output relay terminal is a set of dry, no. 2, form C, contacts with 30 mA,
30 VDC capacity.
2-9
Instruction Manual
OCX 8800
IM-106-880ASB, Rev 1.2
July 2008
Remote Electronics Connections to Sensor Housing
Make the following connections between the electronics and sensor housings
with the electronics cable ordered with the package (Figure 2-6). Braided
cable is available in lengths up to 150 ft. (46 m).
NOTE
Interconnect wiring shown is for Rosemount Analytical supplied cables. For
customer furnished interconnect wiring or cables, refer to Figure 2-7.
Signal Connections
Connect the electronics housing terminals to the corresponding terminals
in the sensor housing. The twisted wire pairs are numbered on the inner
plastic wrapper. Keep twisted pairs together and match the numbers and
wire colors shown in Figure 2-6.
Heater Power Connections
Use the blue, white, orange black, red, and yellow stranded wires in the
heater power cable to connect power to the three heaters in the sensor
housing. Match the wire colors to the corresponding heater power terminal
blocks in the sensor and electronics housings as shown in Figure 2-6.
2-10
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
Figure 2-5. Line Voltage, Earth,
and 4-20 mA Connections
#1
NC
COM
NO
Alarm Output Relay
Terminal Block
4-20 mA Signal Output
Terminal Block
AOUT2+
COe Signal
AOUT2 O2 Signal AOUT1 AOUT1+
{
{
#1
Signal Port
3/4 NPT
G
TOP VIEW
(1/2 SIZE)
Ground Stud
Customer
Wiring
Earth Ground
Typical for Electronics and
Sensor Housing
N
L1
Terminal
Block
G
Ground
Stud
Power Port
3/4 NPT
G
External Tooth
Lockwasher
37390013
EMI Filter
2-11
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
#1
2HTR CO
1HTR CO
2HTR O2
1HTR O2
2HTR SB
1HTR SB
YEL
RED
BLK
ORG
WHT
BLU
SHIELD
GRN
Figure 2-6. Electrical Connections Between Remote Electronics and Sensor Housing
To
ground
screw
Heater Power
Connector (J3)
#1
#1
RED
BLK
T/C CO+
T/C CO-
WHT
BLK
GRN
BLK
Heater Power Cable
#1
T/C SB+
T/C SBT/C O2+
T/C O2-
BLU
BLK
YEL
BRN
BLK
O2 CELL+
O2 CELL-
#1
EXC+
CO ACT+
CO ACT-
COe Sensor
and
CO REF+ Cold Junction
CO REF- Connector (J4)
CJC+
CJCEXC-
RED
WHT
ORG
BLK
BLK
O2 Sensor and
Thermocouple
Connector (J5)
SHLD
To ground
screw
ELECTRONICS HOUSING
#1
2-12
YEL
2
HTR
1 SB
EXC +
BRN
BLK
RED
2
HTR
1 CO
CO
ACT +
-
CO
REF
+
WHT
ORG
+
2
HTR
1 O2
37390014
GRN
BLU
WHT
RED
YEL
ORG
EXC-
-
BLK
BLK
WHT
T/C SB
+
BLK
-
RED
+
BLK
T/C CO
CJC
To ground screw
BLK
BLK
-
GRN
+
BLK
BLU
+
O2
-
T/C O2
-
Signal Cable
SENSOR HOUSING
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
Figure 2-7. Customer-Furnished
Interconnect Wiring or Cables
SIGNAL WIRING OR CABLE
NOTE: For RFI/CE compliance, the connector
must provide 360 degrees of electrical
contact to the cable shield.
ELECTRONICS END
7.25
±0.10
Strip Wire
Ends 3/16”
Typical
Heat Shrink
Tubing
2” Long
1/2” Size
PROBE END
12.5 ±0.10
3/4 NPT Hub Size,
Liquid-tight
Strain Relief
Connector
10.375 ±0.10
Heat Shrink
Tubing
2” Long
1/2” Size
0.5
Typ.
A
See Note
2.0 ±0.25
Typical
7.0” Long Teflon Tubing,
0.042” ID (Cut off drain
wire at probe end of
shield).
Stud Size
#10
7.25 ±0.10
9.25 ±0.10
8.625 ±0.10
6.875 ±0.10
Overall Cable Length
By Customer
150’ Maximum
See Note
8 twisted pairs 24 AWG,
stranded, insulated,
tinned copper
o
conductors, 200 C, 300
volts, with overall braid
of 34 AWG tinned
copper, 90% coverage
and 24 AWG tinned
copper, uninsulated
drain wire.
3.875
±0.10
Ferrule,
Uninsulated
4.625
±0.10
5.375
±0.10
6.125
±0.10
#1
#2
Heat Shrink Tubing
1” Long, 3/16” Size
DETAIL A
(typical on both ends of wiring)
HEATER WIRING OR CABLE
ELECTRONICS END
3/4 NPT Hub Size,
Liquid-tight
Strain Relief
Connectors
4.0 ±0.10
Heat Shrink Tubing
2” Long, 1/2” Size
Heat Shrink Tubing
2” Long, 1/2” Size
0.5
Typ.
Green,
16 AWG
Stud Size
#10
5.5 ±0.10
Ferrule, Uninsulated
2.0 ±0.25 Typical
Green, 16 AWG
4.5 ±0.10
4.25” Long Teflon Tubing,
0.042” ID. (Cut off drain wire
at probe end of shield).
8 Conductors, 16 AWG, Stranded,
O
200 C, 600 volts.
Braided shield - tinned copper, 90%
coverage with 18 AWG 24 tinned copper,
uninsulated, drain wire.
Stud Size #6
37390061
Strip Wire
Ends 3/16”
Typical
PROBE END
2-13
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
PNEUMATIC
INSTALLATION
Pneumatic system connections depend on whether reference air set,
calibration solenoids, and/or blowback equipment options are equipped on
your transmitter. Refer to the following paragraphs and select the option that
applies to your transmitter configuration.
Reference Air Set Option (only)
When no options or only the reference air set option is equipped, use the
following procedure to install the pneumatic system components.
1. Refer to Figure 2-8. Connect the reference air set (regulator/filter and
pressure gage) to the instrument air inlet on the electronics housing and
to the inlet side of the dilution air flow meter.
2. Connect the dilution air flow meter output to the dilution air inlet fitting on
the sensor housing.
.
3. Install an air line between the instrument air outlet fitting on the
electronics housing and the tee fitting on the sensor housing.
Do not use 100% nitrogen as an O2 low gas. It is suggested that O2 low gas be between
0.4% and 2.0% O2. Do not use gases with hydrocarbon concentrations of more than 40
parts per million. Failure to use proper gases will result in erroneous readings.
4. One CO gas and two O2 gases are used to calibrate the OCX 8800:
CO - 1000 ppm or up to 4%, Balance air
O2 low gas - 0.4% , Balance N2
O2 high gas - 8%, Balance N2
Connect the output of the test gas sources to the inlet port of the CAL
GAS flow meter. Install an air line between the flow meter outlet port and
the CAL GAS inlet fitting on the sensor housing.
2-14
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
Figure 2-8. Pneumatic
Installation, OCX with Reference
Air Set (without Autocalibration)
Sensor
Housing
CAL Gas In
Reference Air In
(
CAL Gas
Flow Meter
7 scfh, 20-30 psig
Recommended
(
Dilution Air In
Dilution Air
Flow Meter
0.1 scfh
LO O2
Instrument
Air Out
HI O2
2-Stage
Regulators
CO
Instrument
Air Supply
Pressure Reguator/Filter
35 psig - General Purpose
45 psig - Hazardous Area
Electronics
Housing
37390011
Eductor
Air In
2-15
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
Figure 2-9. Pneumatic
Installation, OCX with Reference
Air Set and Solenoids (with
Autocalibration)
Sensor
Housing
CAL Gas In
(
(
CAL Gas
Flow Meter
7 scfh, 20-30 psig
Recommended
Reference
Air In
Eductor
Air In
CAL Gas Out
Instrument Air Out
Dilution
Air In
Dilution Air
Flow Meter
0.1 scfh
Electronics
Housing
Instrument
Air Supply
Pressure Regulator/Filter
35 psig - General Purpose
45 psig - Hazardous Area
37390012
CO
HI O2
LO O2
2-Stage
Regulators
Reference Air Set and Solenoids Option
When the reference air set and test gas solenoids are included with your
OCX 8800, use the following procedure to install the pneumatic system
components.
1. Install the reference air set according to the instructions in Reference Air
Set Option, steps 1 through 3.
2. Refer to Figure 2-9. Connect the O2 low gas source to the CAL GAS LO
O2 inlet fitting on the electronics housing. Install a shutoff valve and
pressure regulator with gage in the O2 low supply line, as shown.
3. Connect the O2 high gas source to the CAL GAS HI O2 inlet fitting.
Install a shutoff valve and pressure regulator with gage in the O2 high
supply line.
2-16
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
4. Connect the CO high gas to the CAL GAS HI COe inlet fitting. Install a
shutoff valve and pressure regulator with gage in the CO high supply
line.
5. Connect the CAL GAS outlet fitting of the electronics housing to the inlet
port of the CAL GAS flow meter. Install an air line between the flow
meter outlet port and the CAL GAS inlet fitting on the sensor housing.
Reference Air Set, Solenoids, and Blowback Option
The blowback system blows instrument air back through the blowback filter
and out the sample tube of the transmitter. This removes built up dirt and
particulate from the filter and sample line. The blowback option is normally
used in systems that have a dirty process stream.
Installing an OCX 8800 with the blowback option requires the addition of air
operated blowback valve, regulator and gage, and check valve.
Figure 2-10 shows the piping arrangement for the OCX 8800 with the
blowback and autocalibration options. Figure 2-11 shows the piping
arrangement for the OCX 8800 with the blowback option, but without
autocalibration (without test gas solenoids).
When the reference air set, calibration gas solenoids, and blowback options
are included with your transmitter, use the following procedure to install the
pneumatic system components.
1. Connect the calibration gas sources according to the instructions in the
previous paragraph “Reference Air Set and Solenoids Option”, steps 2
through 5.
2. Connect a clean, dry, instrument-quality supply of air (20.95% O2) to the
35/45 psig and 55 psig pressure regulators. The inlet to the 35/45 psig
regulator accepts a 1/8" NPT fitting. The inlet to the 55 psig regulator
accepts a 1/4" NPT fitting.
3. See the upper leg of the instrument air supply. Connect the output of the
35/45 psi regulator/filter to one port of the normally-closed air-operated
solenoid valve, and to the inlet side of the dilution air flow meter.
4. Connect the dilution air flow meter output to the DILUTION AIR inlet
fitting on the sensor housing.
5. Install an instrument air line between the open port of the normally-open
air-operated solenoid valve and the tee fitting on the sensor housing.
6. Connect the output of the 55 psi regulator/filter to one port of the
normally-open air-operated solenoid valve, and to the instrument air
inlet on the back of the electronics housing.
7. Install an air line between the open port of the normally-closed
air-operated solenoid valve and the check valve inlet fitting on the
sensor housing.
2-17
Instruction Manual
OCX 8800
Figure 2-10. Piping Arrangement,
Blowback with Autocalibration
2-18
IM-106-880ASB, Rev 1.2
July 2008
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
Figure 2-11. Piping Arrangement,
Blowback without Autocalibration
2-19
Instruction Manual
OCX 8800
IM-106-880ASB, Rev 1.2
July 2008
8. Install an air line between the instrument air outlet fitting on the
electronics housing and the control air inlet fitting on the air-operated
solenoid valve.
9. There are three settings that need to be specified to set up the blowback
option. These are the blowback interval, duration, and purge time.
Interval - Length of time between blowback events.
(60 minutes recommended.)
Duration - Length of time blowback air is activated.
(5 seconds recommended.)
Purge - Length of time after blowback is complete before
oxygen/combustibles readings are considered valid.
(Set as required by the application.)
These settings are available through HART from the DEVICE SETUP >
DETAILED SETUP > OUTPUT CONDITIONS > BLOWBACK menu.
INITIAL STARTUP
Observe the following Caution and Note. Refer to Section 3: Configuration
and Startup, for OCX 8800 startup information.
Upon completing installation, make sure that the OCX 8800 is turned on and operating prior
to firing up the combustion process. Damage can result from having a cold OCX 8800
exposed to the process gases.
If ducts will be washed down during outages, make sure to power down the OCX 8800 units
and remove them from the wash area.
NOTE
During outages, and whenever possible, leave OCX 8800 units running to
prevent condensation and premature aging from thermal cycling.
2-20
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
Section 3
OCX 8800
Configuration and Startup
Verify Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 3-1
Initial Power Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 3-3
Set Test Gas Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 3-4
OCX 8800 Reset Procedure . . . . . . . . . . . . . . . . . . . . . . . . page 3-4
Install all protective equipment covers and safety ground leads after installation. Failure to
install covers and ground leads could result in serious injury or death.
VERIFY INSTALLATION
Ensure the OCX 8800 is installed correctly. Verify mechanical installation and
all electrical and pneumatic connections. Refer to Section 2, Installation.
Make sure that the OCX 8800 is turned on and operating prior to firing up the combustion
process. Damage can result from having a cold OCX 8800 exposed to the process gases.
NOTE
During outages, and whenever possible, leave all OCX 8800 units running to
prevent condensation and premature aging from thermal cycling.
http://www..raihome.com
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
Verify Configuration
There are three switches on the microprocessor board which are user
configurable for the OCX 8800 (Figure 3-1). SW1 determines if the O2 4-20
mA signal is internally or externally powered. SW2 determines if the COe 4-20
mA signal is internally or externally powered. SW3 sets the rail limits for the
O2 and COe 4-20 mA signals and configures the sample line heater control
circuit. All switches are accessible through holes in the electronics box.
Remove power from the OCX 8800 before changing defaults. If defaults are changed under
power, damage to the electronics may occur.
Verify that the following switch settings are correct for your OCX 8800
installation:
SW1 The two settings are internally or externally powering the O2 4-20
mA signal. The factory setting is for the O2 4-20 mA signal to be internally
powered.
SW2 The two settings are internally or externally powering the COe 4-20
mA signal. The factory setting is for the COe 4-20 mA signal to be
internally powered.
SW3 The factory sets this switch as follows:
• Position 1 determines the O2 4-20 mA signal rail limit. The settings are
high, 21.1 mA, or low, 3.5 mA. The factory setting is low, 3.5 mA.
• Position 2 determines the COe 4-20 mA signal rail limit. The settings
are high, 21.1 mA, or low, 3.5 mA. The factory setting is high, 21.1 mA.
• Positions 3 and 4 must be set as shown for proper software control of
the device heaters.
3-2
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
COe
O2
Figure 3-1. OCX 8800 Defaults
Open
21.1 mA
Switch
Default Postions
Shown
3.5 mA
1
Open
Closed
2
3
4
Closed
O2 21.1 mA/3.5 mA: O2 4-20 mA Signal
Rail Limits:
Open High - 21.1 mA
Closed Low - 3.5 mA
SW3
3 4
1 2
COe 21.1 mA/3.5 mA: COe 4-20 mA Signal
Rail Limits:
Open High- 21.1 mA
Closed Low - 3.5 mA
Internal:
COe 4-20 mA
is Internally
Powered
SW2
INITIAL POWER UP
External:
COe 4-20 mA
Requires an External
Power Supply
(Default)
Internal:
O2 4-20 mA
is Internally
Powered
SW1
External:
O2 4-20 mA
Requires an External
Power Supply
(Default)
37390026
1 2 3 4
Allow adequate time (approximately 60 minutes) for the heaters to begin
operation and for the OCX 8800 to reach normal operating temperature on
power up. Normal operating temperature for the O2 cell is 736°C. Normal
operating temperature for the combustibles cell is 300°C. The normal sample
line temperature is 170°C. During this time the edu ctor air solenoid will remain
closed so no sample is pulled through the analyzer. When the OCX reaches
operating temperature the solenoid will energize, eductor air will begin to flow,
and the unit will begin normal operation.
3-3
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
SET TEST GAS VALUES
Use HART/AMS to set test gas values for calibration. Refer to Section 4,
Using HART Communications for more information.
Setting Test Gas Values with HART
1. Use the 275/375 Field Communicator or AMS software to access the
HART menu.
2. From the DETAILED SETUP menu, select O2 CALIB PARAMS.
3. From O2 CALIB PARAMS, select O2 High Gas. Enter the percent O2
used for the high O2 test gas.
4. From O2 CALIB PARAMS, select Low TG. Enter the percent O2 used
for the low O2 test gas.
5. From the DETAILED SETUP menu, select COe CALIB PARAMS.
6. From COe CALIB PARAMS, select COe Test Gas. Enter the CO
concentration (ppm) used for COe test gas.
OCX 8800 RESET
PROCEDURE
Whenever you correct an equipment alarm or fault condition, the OCX 8800
will either revert to normal operation or continue to indicate an alarm status
condition. If the equipment does not revert to normal operation when a fault
condition is cleared, or if instructed to do so in Section 7, Troubleshooting,
use the following procedure to reset the OCX 8800.
OCX Reset with HART
Remove the OCX 8800 from the process loop and recycle power.
3-4
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
Section 4
OCX 8800
Using HART Communications
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 4-1
Field Communicator Signal Connections . . . . . . . . . . . . page 4-1
Field Communicator PC Connections . . . . . . . . . . . . . . . page 4-4
HART Menu Tree . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 4-5
D/A Trim Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 4-9
OVERVIEW
The Field Communicator is a handheld communications interface device. It
provides a common communications link to all microprocessor-based
instruments that are HART compatible.
To interface with the OCX 8800, the Field Communicator requires a
termination point along the O2 4-20 mA current loop and a minimum load
resistance of 250 ohms between the communicator and the power supply.
The Field Communicator accomplishes its task using a Frequency Shift
Keying (FSK) technique. With the use of FSK, high-frequency digital
communication signals are superimposed on the 4-20 mA oxygen output
signal. The communicator does not disturb the 4-20 mA signal since no net
energy is added to the loop. HART information is not available on the COe
output signal.
The Field Communicator may be interfaced with a personal computer (PC),
providing special software has been installed. To connect the Field
Communicator may to a PC, an interface adapter is required. Refer to the
proper Field Communicator documentation in regard to the PC interface
option.
FIELD COMMUNICATOR
SIGNAL CONNECTIONS
http://www..raihome.com
The Field Communicator can connect to the OCX 8800 oxygen analog output
signal line at any wiring termination in the O2 4-20 mA current loop. There are
two methods of connecting the Field Communicator to the signal line. For
applications in which the signal line has a load resistance of 250 ohms or
more, refer to method 1. For applications in which the signal line load
resistance is less than 250 ohms, refer to method 2.
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
Figure 4-1. Signal Line Connections,
≥ 250 Ohms Load Resistance
Method 1, for Load Resistance ≥ 250 Ohms
Refer to Figure 4-1 and the following steps to connect the Field
Communicator to a signal line 250 ohms or more of load resistance.
Do not make connections to the Field Communicator 's serial port, 4-20 mA signal lines, or
NiCad recharger jack in an explosive atmosphere. Explosions can result in death or serious
injury.
Using the supplied lead set, connect the Field Communicator in parallel to the
OCX 8800. Use any wiring termination points in the oxygen analog output
4-20 mA signal line.
4-2
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
Method 2, for Load Resistance < 250 Ohms
Refer to Figure 4-2 and the following steps to connect the Field
Communicator to a signal line with less than 250 ohms load resistance.
Do not make connections to the Field Communicator's serial port, 4-20 mA signal lines, or
NiCad recharger jack in an explosive atmosphere. Explosions can result in death or serious
injury.
1. At a convenient point, break the oxygen analog output 4-20 mA signal
line and install the optional 250 ohm load resistor.
2. Plug the load resistor into the loop connectors (located on the rear panel
of the Field Communicator).
Figure 4-2. Signal Line Connections,
< 250 Ohms Load Resistance
4-3
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
Field Communicator PC
Connections
There is an option to interface the Field Communicator with a personal
computer. Load the 375 Easy Upgrade Programming Utility software into the
PC. Then link the Field Communicator to the PC through the IRDA Interface.
Refer to the proper Field Communicator documentation in regard to the PC
interface option.
Off-line and On-line Operations
The Field Communicator can be operated both off-line and on-line.
Off-line operations are those in which the communicator is not connected to
the OCX 8800. Off-line operations can include interfacing the Field
Communicator with a PC. (Refer to applicable documentation regarding
HART/PC applications.)
In the on-line mode, the communicator is connected to the oxygen 4-20 mA
analog output signal line. The Field Communicator is connected in parallel to
the OCX 8800 or in parallel to the 250 ohm load resistor.
The opening menu displayed on the LCD is different for on-line and off-line
operations. When powering up a disconnected (off-line) communicator, the
LCD will display the Main Menu. When powering up a connected (on-line)
communicator, the LCD will display the On-line Menu. Refer to the Field
Communicator manual for detailed menu information.
4-4
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
HART MENU TREE
OCX 8800
This section consists of a menu tree for the Field Communicator. This menu is
specific to OCX 8800 applications.
Figure 4-3. HART Menu Tree
(Sheet 1 of 4)
4-5
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
Figure 4-3. HART Menu Tree (Sheet 2 of 4)
STATUS
DIAG/SERVICE
Operate Mode
Status group 1
Status group 2
Status group 3
Status group 4
Status group 5
Status group 6
AO saturated
AO fixed
HW STATUS
Model Number
Snsr Housing
CPLD REGS
Line Freq
Line Volt
PCN Counter
PCD Counter
LOOP TEST
D/A TRIM
BLOWBACK
MAX TEMP
(CONTINUED ON
SHEET 3)
4-6
BCSR0
BCSR1
BCSR2
BCSR3
BCSR4
BCSR5
O2 Loop Test
COe Loop Test
(procedures)
PERFORM
CAL
CAL METHODS
O2OutTracks
COeOutTracks
Cal Step
Cal Time
O2
COe
CAL
STATUS
Cal Step
Cal Time
CAL VALUES
O2 Slope
O2 Const
O2 Snsr R
COe Slope
COe Const
PREV CAL
VALUES
Prev O2 Slope
Prev O2 Const
Prev O2 Snsr R
Prev COe Slope
Prev COe Const
FAILED CAL
VALUES
Bad O2 Slope
Bad O2 Const
Bad COe Slope
Bad COe Const
RESET CAL
CONST
Reset O2 CalConsts
Reset COe CalConsts
Verify Calibration
Status
TimeRemain
O2
COe
Status group 2
O2 Temp Low
O2 Temp Hi
O2 Snsr Open
O2 Snsr R Hi
EEPRM Chksm Fail
O2 Htr Ramp Rate
COe Htr Ramp Rate
SB Htr Ramp Rate
Status group 3
Cal Failed
Cal Warning
SB T/C Open
SB T/C Shorted
SB Temp Low
SB Temp Hi
SB Temp Very Hi
ADC Ref Error
Status group 4
RTD Current Err
COe Temp Low
COe Temp Hi
COe Temp Very Hi
COe T/C Open
COe T/C Shorted
COe T/C Reversed
SB T/C Reversed
Status group 5
COe Htr Failure
SB Htr Failure
Line Voltage Low
Line Voltage Hi
Htr Relay Failed
Output Board Failure
Status group 6
O2 Slope Error
O2 Constant Error
O2 Cal Failed
COe Slope Error
COe Constant Error
COe Cal Failed
O2 Calibration
COe Calibration
O2 and COe Calib
BlBk State
Blowback
O2TempMax
COeTempMax
SBTempMax
BrdTempMax
CJCTempMax
O2 PID
HEATER PID
O2 AO Fixed
COe AO Fixed
O2 D/A Trim
COe D/A Trim
CALIBRATE
VERIFY CALIB
O2 AO Saturated
COe AO Saturated
Status group 1
O2 T/C Open
O2 T/C Shorted
O2 T/C Reversed
ADC Failure
Line Freq Error
O2 Htr Failure
O2 Temp Very Hi
Board Temp Hi
O2 Prop
O2 Int
O2 DutyCyc
O2 SetPt
O2 Temp
COE PID
COe Prop
COe Int
COe DutyCyc
COe SetPt
COe Temp
SB PID
SB Prop
SB Int
SB DutyCyc
SB SetPt
SB Temp
37390039
(CONTINUED FROM
SHEET 1 )
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
Figure 4-3. HART Menu Tree
(Sheet 3 of 4)
BASIC SETUP
DEVICE
INFORMATION
HART
INFORMATION
S/W VERSION
INFO
Date
Descriptor
Message
Final asmbly num
O2 Sensor s/n
COe Sensor s/n
Tag
Poll addr
Dev id
Num req preams
Fld dev rev
Version
Chksum
Bld Num
Bld Date
Restart Cntr
ANALOG
OUTPUT
COe ANALOG
OUTPUT
OUTPUT
CONDITION
DETAILED
SETUP
(CONTINUED ON
SHEET 4)
O2 ANALOG
OUTPUT
O2 AO Type
O2 URV
O2 LRV
O2 Alarm Level
COe AO Type
COe URV
COe LRV
COe Alarm Level
ALARM
RELAY
Trig 1 Event
Trig 2 Event
Trig 3 Event
Alarm State
BLOWBACK
BlBk Enabled
BlBk Intrvl
BlBk Period
BlBk PurgeTm
BlBk State
O2 CALIB
PARAMS
O2OutTracks
O2 High Gas
O2 Low Gas
Gas Time
Purge Time
Solenoids
O2 CalIntv
O2NxtCalTm
COe CALIB
PARAMS
COeOutTracks
COe Test Gas
Gas Time
Purge Time
Solenoids
COe CalIntv
COeNxtCalTm
COe Slope Warn
DEVICE
PARAMS
O2 Slope
O2 Const
O2 T90
COe Slope
COe Const
COe T90
PCNC Enable
PCDC Enable
LOI
PARAMS
User Intface
Luminance
Lockout Time
Revert Time
37390040
(CONTINUED FROM
SHEET 2)
4-7
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
Figure 4-3. HART Menu Tree
(Sheet 4 of 4)
(CONTINUED FROM
SHEET 3)
DEVICE
INFORMATION
Manufacturer
Model
Date
Descriptor
Message
Final asmbly num
O2 Sensor s/n
COe Sensor s/n
Hardware rev
Software rev
HART
INFORMATION
Tag
Poll addr
Dev id
Num req preams
Fld dev rev
Universal rev
O2 CALIB
O2OutTracks
O2 High Gas
O2 Low Gas
Gas Time
Purge Time
Solenoids
O2 CalIntv
O2NxtCalTm
COE CALIB
COeOutTracks
COe Test Gas
Gas Time
Purge Time
Solenoids
COe CalIntv
COeNxtCalTm
COe Slope Warn
O2 DEVICE
CONFIG
O2 Slope
O2 Const
O2 T90
COE DEVICE
CONFIG
COe Slope
COe Const
COe T90
CAL INFO
DEVICE
CONFIG
DEVICE CONFIG
OUTPUTS
CONFIG
4-8
O2 URV
O2 LRV
COe URV
COe LRV
O2 AO Type
COe AO Type
O2 Alarm Level
COe Alarm Level
Trig 1 Event
Trig 2 Event
Trig 3 Event
BlBk Enabled
BlBk Intrvl
BlBk Period
BlBk PurgeTm
PCNC Enable
PCDC Enable
Luminance
Lockout Time
Revert Time
37390041
REVIEW
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
D/A TRIM PROCEDURES
OCX 8800
O2 D/A trim procedure using HART
Use the following procedure to perform the O2 D/A trim procedure using the
Field Communicator. Refer to the HART menu tree, Figure 4-3.
NOTE
To select a menu item, either use the up and down arrow keys to scroll to the
menu item and press the right arrow key or use the number keypad to select
the menu item number.
To return to a preceding menu, press the left arrow key.
1. From the DIAG/SERVICE menu, select D/A TRIM. The Field
Communicator displays O2 D/A Trim.
2. Press the right arrow key to start the procedure. (If you wish to exit D/A
Trim with no changes, select ABORT.)
3. The Field Communicator displays WARNING: Loop should be
removed from automatic control. Remove the OCX 8800 from any
automatic control loops to avoid a potentially dangerous operating
condition and press OK.
4. The Field Communicator displays Connect reference meter to O2
output.
5. Remove the electronics housing cover.
6. Refer to Figure 2-5. Connect a digital multimeter to read the milliamp
output from the O2 D/A converter circuit. Connect the positive lead to
the AOUT1+ terminal and connect the negative lead to the AOUT1terminal. Then, press OK at the Field Communicator.
7. The Field Communicator displays Setting Fld dev output to 4 mA.
Press OK. Read the O2 millamp output at the digital multimeter. Enter
the reading at the Field Communicator and press ENTER. (Select
ABORT to exit without changes).
8. The Field Communicator displays Setting Fld dev output to 20 mA.
Press OK. Read the O2 millamp output at the digital multimeter. Enter
the reading at the Field Communicator and press ENTER. (Select
ABORT to exit without changes).
9. The Field Communicator displays Setting Fld dev output to 4 mA.
Press OK.
10. The Field Communicator displays Fld dev output 4.00 mA equal to
reference meter? Using the up or down arrow, select 1 Yes or 2 No
and Press ENTER. If No, the process repeats from step 6.
11. The Field Communicator displays Setting Fld dev output to 20 mA.
Press OK.
12. The Field Communicator displays Fld dev output 20.00 mA equal to
reference meter? Using the up or down arrow, select 1 Yes or 2 No
and Press ENTER. If No, the process repeats from step 7.
13. The Field Communicator displays NOTE: Loop may be returned to
automatic control.
4-9
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
COe D/A trim procedure using HART
Use the following procedure to perform the COe D/A trim procedure using the
Field Communicator. Refer to the HART menu tree, Figure 4-3.
NOTE
To select a menu item, either use the up and down arrow keys to scroll to the
menu item and press the right arrow key or use the number keypad to select
the menu item number.
To return to a preceding menu, press the left arrow key.
1. From the DIAG/SERVICE menu, select D/A TRIM. The Field
Communicator displays O2 D/A Trim. Press the up or down arrow to
select COe D/A Trim.
2. Press the right arrow key to start the procedure. (If you wish to exit D/A
Trim with no changes, select ABORT.)
3. The Field Communicator displays WARNING: Loop should be
removed from automatic control. Remove the OCX 8800 from any
automatic control loops to avoid a potentially dangerous operating
condition and press OK.
4. The Field Communicator displays Connect reference meter to
Combustibles output.
5. Remove the electronics housing cover.
6. Refer to Figure 2-5. Connect a digital multimeter to read the milliamp
output from the COe D/A converter circuit. Connect the positive lead to
the AOUT2+ terminal and connect the negative lead to the AOUT2terminal. Then, press OK at the HART communicator.
7. The Field Communicator displays Setting Fld dev output to 4 mA.
Press OK. Read the COe millamp output at the digital multimeter. Enter
the reading at the Field Communicator and press ENTER. (Select
ABORT to exit without changes).
8. The Field Communicator displays Setting Fld dev output to 20 mA.
Press OK. Read the COe millamp output at the digital multimeter. Enter
the reading at the Field Communicator and press ENTER. (Select
ABORT to exit without changes).
9. The Field Communicator displays Setting Fld dev output to 4 mA.
Press OK.
10. The Field Communicator displays Fld dev output 4.00 mA equal to
reference meter? Using the up or down arrow, select 1 Yes or 2 No
and Press ENTER. If No, the process repeats from step 6.
11. The Field Communicator displays Setting Fld dev output to 20 mA.
Press OK.
12. The Field Communicator displays Fld dev output 20.00 mA equal to
reference meter? Using the up or down arrow, select 1 Yes or 2 No
and Press ENTER. If No, the process repeats from step 7.
13. The Field Communicator displays NOTE: Loop may be returned to
automatic control.
4-10
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
Section 5
OCX 8800
Calibration
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 5-1
Fully Automatic Calibration . . . . . . . . . . . . . . . . . . . . . . . . page 5-1
Operator - Initiated Autocalibration . . . . . . . . . . . . . . . . . page 5-3
Manual Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 5-3
OVERVIEW
During a calibration, two calibration gases with known O2 concentrations and
one calibration gas with a known COe concentration are applied to the OCX
8800. Slope and constant values are calculated to determine if the OCX 8800
is correctly measuring net concentrations of O2 and combustibles in the
industrial process.
Before calibrating the OCX 8800, verify that the calibration gas parameters
are correct by setting the test gas values used when calibrating the unit. Refer
to Section 3, Configuration and Startup.
There are three calibration methods available to the OCX 8800; automatic,
operator-initiated automatic, and manual. Calibration commands and menus
can be accessed by HART/AMS.
FULLY AUTOMATIC
CALIBRATION
If the OCX 8800 is equipped with calibration solenoids, the unit can be
programmed to automatically calibrate without any operator action. Refer to
the following paragraphs for using HART/AMS to set up the OCX 8800 for
fully automatic calibration.
Autocalibration Setup using HART
NOTE
Automatic calibration is only available on units equipped with calibration
solenoids.
Use the following procedure to specify a time interval (in hours) at which the
OCX 8800 will automatically calibrate.
1. From the DEVICE SETUP screen, select DETAILED SETUP.
2. From the DETAILED SETUP screen, select O2 CALIB PARAMS or
COE CALIB PARAMS.
3. If the unit is equipped with calibration solenoids and timed automatic
calibrations are desired, select Solenoids, then select Yes. Select No to
disable the calibration solenoids.
http://www..raihome.com
Instruction Manual
OCX 8800
IM-106-880ASB, Rev 1.2
July 2008
4. Select O2 CalIntrvl (O2 calibration interval) and enter the desired time
in hours between automatic O2 calibrations. Select COE Callintrvl and
enter the desired time between automatic COe calibrations. To disable
automatic calibration for O2 and COe, enter 0 for both CalIntrvl
parameters.
5. If desired, the O2NxtCalTm and the COeNxtCalTm (next calibration
time) parameters can be changed to synchronize a calibration at a
specific day or time.
When setting automatic calibration times, CalIntrvl and NxtCalTm should be set so that O2
and COe are NOT calibrated simultaneously.
NOTE
To select a menu item, either use the up and down arrow keys to scroll to the
menu item and press the right arrow key or use the number keypad to select
the menu item number.
To return to a preceding menu, press the left arrow key.
6. Select O2 CalIntrvl (O2 calibration interval) and enter the desired time
in hours between automatic O2 calibrations. Select COE Callintrvl and
enter the desired time between automatic COe calibrations. To disable
automatic calibration for O2 and COe, enter 0 for both CalIntrvl
parameters.
5-2
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OPERATOR - INITIATED
AUTOCALIBRATION
OCX 8800
An operator can initiate an automatic calibration at any time provided that the
unit is equipped with calibration solenoids.
Autocalibration using HART
To initiate an automatic calibration using HART/AMS, perform the following
steps on the HART menu tree. Refer to Section 4, Using HART
Communications, for the HART menu tree.
1. Select DIAG/SERVICE from DEVICE SETUP menu.
2. Select CALIBRATION from the DIAG/SERVICE menu.
3. Select PERFORM CAL from the CALIBRATION menu.
4. Select CAL METHODS from the PERFORM CAL menu.
5. From the CAL METHODS menu, select the type of calibration desired:
O2 Calibration, COe Calibration, or O2 and COe Calibration.
MANUAL CALIBRATION
If a unit is not equipped with calibration solenoids, a calibration must be
performed by an operator following prompts from the unit. Refer to the
following paragraphs for manual calibration.
Manual O2 Calibration using HART
To perform a manual O2 calibration using the Field Communicator or AMS,
use the following procedure. If necessary, refer to Section 4, Using HART
Communications, for the HART menu tree.
NOTE
To select a menu item, either use the up and down arrow keys to scroll to the
menu item and press the right arrow key or use the number keypad to select
the menu item number.
To return to a preceding menu, press the left arrow key.
1. From the DIAG/SERVICE menu, select CALIBRATION. From the
CALIBRATION menu, select PERFORM CAL.
2. From the PERFORM CAL menu, select CAL METHODS. Select O2
CALIBRATION to start O2 calibration.
3. In the first O2 Calibration screen, a Loop should be removed from
automatic control warning appears. Remove the OCX 8800 from any
automatic control loops to avoid a potentially dangerous operating
condition and press OK.
4. The main Calibration screen should look like the following. Press OK to
continue.
OCX: TAG NAME
STATUS: Idle
TIME REMAIN: 0s
O2: 0.4 %, 85.95mV
OK/NEXT to Select
ABORT/CANCEL to Exit
5-3
Instruction Manual
OCX 8800
IM-106-880ASB, Rev 1.2
July 2008
5. From the SELECT ACTION screen, select START/NEXT CALSTEP to
continue calibration, select ABORT CAL to abort calibration or EXIT
CAL to exit calibration. Select one item from the list and press ENTER.
OCX: TAG NAME
SELECT ACTION
1. START/NEXT CALSTEP
2. ABORT CAL
3. EXIT CAL
6. When the Calibration Status is at the AppO2Low step, switch on O2
Low Gas. Verify the O2 concentration measured matches the O2 LOW
GAS parameter in the Setup. Press OK when ready.
7. Select Start/Next Cal Step to start applying the O2 Low Gas. The time
to apply the test gas is specified by the Gas Time.
8. The Calibration Status should be automatically changed to FlowO2Low
and then ReadO2Low for a period of time. During this period, if an
attempt is made to go to the next calibration step by pressing OK and
selecting Start/Next Cal Step, you will be prompted with Operator step
command is not accepted at this time. The Next Cal Step command
is not accepted at this time.
9. When ready, Calibration Status will stop at the AppO2Hi. Switch off the
O2 Low Gas and switch on the O2 High Gas. Verify the O2 concentration
measured matches the O2 HIGH GAS parameter in the Setup. Press
OK when ready.
10. Select Start/Next Cal Step to start applying the O2 High Gas. The time
to apply the test gas is specified by the Gas Time.
11. The Calibration Status should be automatically changed to FlowO2Hi
and then ReadO2Hi for a period of time. During this period, if an attempt
is made to go the next calibration step by pressing OK and selecting
Start/Next Cal Step, you will be prompted with Operator step
command is not accepted at this time. The Next Cal Step command
is not accepted at this time.
12. When ready, Calibration Status will stop at STOP GAS. Switch off the
O2 High Gas. Press OK when ready. Select Start/Next Cal Step to start
purging gas. The time to purge gas is specified by the Purge Time.
13. When the Purge step is complete, the Calibration Status will be at IDLE
if the calibration is successful or CAL RECOMMENDED if the
calibration has failed. A Calibration Failed alarm will be set if the
calibration has failed.
14. When calibration is complete. Select Exit Cal to exit the calibration
method.
5-4
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
Manual COe Calibration using HART
To perform a manual COe calibration using the Field Communicator or AMS,
use the following procedure. If necessary, refer to Section 4, Using HART
Communications, for the HART menu tree.
NOTE
To select a menu item, either use the up and down arrow keys to scroll to the
menu item and press the right arrow key or use the number keypad to select
the menu item number.
To return to a preceding menu, press the left arrow key.
1. From the DIAG/SERVICE menu, select CALIBRATION. From the
CALIBRATION menu, select PERFORM CAL.
2. From the PERFORM CAL menu, select CAL METHODS. Select COe
CALIBRATION to start COe calibration.
3. In the first COe Calibration screen, a Loop should be removed from
automatic control warning appears. Remove the OCX 8800 from any
automatic control loops to avoid a potentially dangerous operating
condition and press OK.
4. The main Calibration screen should look like the following. Press OK to
continue.
OCX: TAG NAME
STATUS: Idle
TIME REMAIN: 0s
COe: 0.20 ppm, 0.00 mV
OK/NEXT to Select
ABORT/CANCEL to Exit
5. From the SELECT ACTION screen, select START/NEXT CALSTEP to
continue calibration, select ABORT CAL to abort calibration or EXIT
CAL to exit calibration. Select one item from the list and press ENTER.
OCX: TAG NAME
SELECT ACTION
1. START/NEXT CALSTEP
2. ABORT CAL
3. EXIT CAL
6. The unit samples reference air as the COe Low Gas. The Calibration
Status should automatically change to ReadCOLow for a period of time.
During this period, if an attempt is made to go to the next calibration
step by pressing OK and selecting Start/Next Cal Step, you will be
prompted with Operator step command is not accepted at this time.
The Next Cal Step command is not accepted at this time.
7. When ready, Calibration Status will stop at the AppCOeHi. Switch on
the COe High Gas. Verify the COe concentration measured matches the
COe HIGH GAS parameter in the Setup. Press OK when ready.
8. Select Start/Next Cal Step to start applying the COe High Gas. The
time to apply the test gas is specified by the Gas Time.
5-5
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
9. The Calibration Status should be automatically changed to FlowCOeHi
and then ReadCOeHi for a period of time. During this period, if an
attempt is made to go the next calibration step by pressing OK and
selecting Start/Next Cal Step, you will be prompted with Operator step
command is not accepted at this time. The Next Cal Step command
is not accepted at this time.
10. When ready, Calibration Status will stop at STOP GAS. Switch off the
COe High Gas. Press OK when ready. Select Start/Next Cal Step to
start purging gas. The time to purge gas is specified by the Purge Time.
11. When the Purge step is complete, the Calibration Status will be at IDLE
if the calibration is successful or CAL RECOMMENDED if the
calibration has failed. A Calibration Failed alarm will be set if the
calibration has failed.
12. When calibration is complete. Select Exit Cal to exit the calibration
method.
Manual O2 and COe Calibration using HART
To perform a manual O2 and COe calibration using the Field Communicator or
AMS, use the following procedure. If necessary, refer to Section 4, Using
HART Communications, for the HART menu tree.
NOTE
To select a menu item, either use the up and down arrow keys to scroll to the
menu item and press the right arrow key or use the number keypad to select
the menu item number.
To return to a preceding menu, press the left arrow key.
1. From the DIAG/SERVICE menu, select CALIBRATION. From the
CALIBRATION menu, select PERFORM CAL.
2. From the PERFORM CAL menu, select CAL METHODS. Select 02 &
COe CALIBRATION to start O2 and COe calibration.
3. In the first Calibration screen, a Loop should be removed from
automatic control warning appears. Remove the OCX 8800 from any
automatic control loops to avoid a potentially dangerous operating
condition and press OK.
4. The main Calibration screen should look like the following. Press OK to
continue.
OCX: TAG NAME
STATUS: Idle
TIME REMAIN: 0s
O2: 0.4 %, 85.95mV
COe: 0.20 ppm, 0.00 mV
OK/NEXT to Select
ABORT/CANCEL to Exit
5-6
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
5. From the SELECT ACTION screen, select START/NEXT CALSTEP to
continue calibration, select ABORT CAL to abort calibration or EXIT
CAL to exit calibration method. Select one from the list and press
ENTER.
OCX: TAG NAME
SELECT ACTION
1. START/NEXT CALSTEP
2. ABORT CAL
3. EXIT CAL
6. When the Calibration Status is at the AppO2Low step, switch on O2
Low Gas. Verify the O2 concentration measured matches the O2 LOW
GAS parameter in Setup. Press OK when ready.
7. Select Start/Next Cal Step to start applying the O2 Low Gas. The time
to apply the test gas is specified by the Gas Time.
8. The Calibration Status should automatically change to FIowO2Low and
then ReadO2Low for a period of time. During this period, if an attempt
is made to go to the next calibration step by pressing OK and selecting
Start/Next Cal Step, you will be prompted with Operator step
command is not accepted at this time. The Next Cal Step command
is not accepted at this time.
9. When ready, Calibration Status will stop at AppO2Hi. Switch off the O2
Low Gas and switch on the O2 High Gas. Verify the O2 concentration
measured matches the O2 HIGH GAS parameter in Setup. Press OK
when ready.
10. Select Start/Next Cal Step to apply the O2 High Gas. The time to apply
the test gas is specified by the Gas Time.
11. The Calibration Status should automatically change to FlowO2Hi, then
ReadO2Hi, and then ReadCOeLo for a period of time. During this
period, if an attempt is made to go the next calibration step by pressing
OK and selecting Start/Next Cal Step, you will be prompted with
Operator step command is not accepted at this time. The Next Cal
Step command is not accepted at this time.
12. When ready, Calibration Status will stop at AppCOeHi. Switch off the O2
High Gas and switch on the COe Gas. Verify the COe concentration
measured matches the COe TEST GAS parameter in the Setup. Press
OK when ready.
13. Select Start/Next Cal Step to start applying the COe Gas. The time to
apply the test gas is specified by the Gas Time.
14. The Calibration Status should automatically change to FlowCOeHi and
then ReadCOeHi for a period of time. During this period, if an attempt is
made to go the next calibration step by pressing OK and selecting
Start/Next Cal Step, you will be prompted with Operator step
command is not accepted at this time. The Next Cal Step command
is not accepted at this time.
15. When ready, Calibration Status will stop at STOP GAS. Switch off the
COe gas. Press OK when ready. Select Start/Next Cal Step to start
purging gas. The time to purge gas is specified by Purge Time.
5-7
Instruction Manual
OCX 8800
5-8
IM-106-880ASB, Rev 1.2
July 2008
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
Section 6
OCX 8800
Maintenance and Service
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 6-1
OCX 8800 Removal and Installation . . . . . . . . . . . . . . . . . page 6-1
Repair Sensor Housing . . . . . . . . . . . . . . . . . . . . . . . . . . . page 6-8
Repair Electronics Housing . . . . . . . . . . . . . . . . . . . . . . . . page 6-28
Replace Tube Fittings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 6-34
OVERVIEW
This section contains the procedures to maintain and service the OCX 8800.
Install all protective equipment covers and safety ground leads after equipment repair or
service. Failure to install covers and ground leads could result in serious injury or death.
It is recommended that the OCX 8800 be removed from the stack for all service activities.
The unit should be allowed to cool and be taken to a clean work area. Failure to comply may
cause severe burns.
Disconnect and lock out power before working on any electrical components. There may be
voltage up to 264 VAC.
OCX 8800 REMOVAL
AND INSTALLATION
http://www..raihome.com
Use the following procedures to remove or install the OCX 8800.
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
OCX with
Remote Electronics
Remove Sensor Housing
1. Turn off power to the system.
2. Shut off the test gasses at the cylinders and close the instrument air
valve.
3. Disconnect the calibration gas, reference air, eductor air, and dilution air
lines from the sensor housing, Figure 6-1.
Figure 6-1. OCX with
Remote Electronics
Duct
Stack
Sensor
Housing
Heater
Power Cable
Te
s
tG
Signal Cable
as
ir
rA
cto
u
Air
Ed
ce
n
r
e
Ai
fer
on
Re
i
t
lu
Di
t
en
um
r
t
Ins
4-20 mA
Outputs
AC Power
Input
Ai
r
Electronics
Housing
Instrument Air
(Reference Gas)
37390044
High O2 Test Gas
Low O2 Test Gas
CO Test Gas
6-2
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
4. Remove the cover from the sensor housing to expose the sensor
housing terminal blocks, Figure 6-2.
5. Disconnect the signal cable from the O2 and T/C terminal blocks, and
from the CO and CJC terminal blocks.
6. Disconnect the heater power cable from the HTR terminal blocks.
7. If moving the sensor housing to another work site, disconnect and
remove the power and signal cables from the sensor housing.
8. Remove insulation to access the mounting bolts. Unbolt the sensor
housing from the stack and take it to a clean work area.
9. Allow the sensor housing to cool to a comfortable working temperature.
Signal Cable
YEL
EXC+
37390028
BLU
To
Ground
Screw
GRN
BRN
WHT
RED
2
HTR SB
1
CO ACT
+
BLK
-
RED
+
YEL
2
HTR CO
1
CO REF
ORG
2
HTR 02
1
BLK
-
+
O2
NOTE: Wire colors shown are for cables
supplied by Emerson Process
Management.
BLK
WHT
ORG
CJC
+
BLK
-
BLK
EXC -
WHT
T/C SB
+
BLK
-
RED
T/C CO
+
BLK
-
T/C O2
+
-
BLK
GRN
SENSOR HOUSING
BLU
Heater Power Cable
Figure 6-2. Sensor
Housing Terminals
6-3
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
Install Sensor Housing
1. Connect the test gas, reference air, eductor air, and dilution air lines to
the sensor housing.
2. Remove the sensor housing cover.
3. If removed, install the power and signal cables and the customer power
and signal conduits and wiring at the sensor housing.
4. Connect the signal cable to O2 and T/C terminal blocks and to the CO
and CJC terminal blocks, Figure 6-2. Connect the heater power cable to
the HTR terminal blocks.
5. Refer to Figure 6-1 and make sure all test gas lines and electrical
connectors are complete.
6. Install the sensor housing cover.
7. Restore power to the system. Allow OCX to reach normal operating
temperature before installing in a hot process stack.
8. Insert and bolt the sensor housing in the stack and install insulation.
9. Turn on the test gases at the cylinders and open the instrument air
supply valve.
Remove Remote Electronics Housing
1. Turn off power to the system.
2. Shut off the test gases at the cylinders and close the instrument air
supply valve, Figure 6-1.
3. Disconnect the test gas and instrument air lines from the remote
electronics housing.
4. Remove the cover from the electronics housing to expose the
electronics housing terminal blocks, Figure 6-3.
5. Disconnect and remove the power leads from the AC power input
terminal block. Remove the ground lead from the ground stud.
6. Disconnect and remove the O2 and COe signal leads from the 4-20 mA
signal output terminal block.
7. If used, disconnect and remove the external relay leads from the alarm
output relay terminal block.
8. Disconnect the signal cable leads from O2 cell and thermocouple
connector (J4), and from COe and CJC connector (J5), Figure 6-4.
9. Disconnect the heater cable leads from the heater power connector
(J3).
10. If moving the electronics housing to another work site, disconnect and
remove the power and signal cables and customer wiring conduits from
the housing.
11. Remove the remote electronics housing from its mounting and move it
to a suitable work area.
6-4
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
Figure 6-3. Electronics Housing
Terminal Blocks
#1
NC
COM
NO
Alarm Output Relay
Terminal Block
4-20 mA Signal Output
Terminal Block
AOUT2+
COe Signal
AOUT2 O2 Signal AOUT1 AOUT1+
{
{
#1
Signal Port
3/4 NPT
G
TOP VIEW
(1/2 SIZE)
Ground Stud
Customer
Wiring
Earth Ground
Typical for Electronics and
Sensor Housing
N
L1
Terminal
Block
G
Ground
Stud
Power Port
3/4 NPT
G
External Tooth
Lockwasher
37390013
EMI Filter
6-5
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
2HTR CO
1HTR CO
2HTR O2
1HTR O2
2HTR SB
1HTR SB
SHIELD
#1
YEL
RED
BLK
ORG
WHT
BLU
GRN
Figure 6-4. Remote Electronics
Housing Cable Connections
To
Ground
Screw
Heater Power
Connector (J3)
#1
#1
RED
BLK
WHT
BLK
GRN
BLK
1
BLU
BLK
#1
#1
COe Sensor
& CJC
Connector (J4)
T/C CO+
T/C COT/C SB+
T/C SBT/C O2+
T/C O2-
YEL
BRN
BLK
O2 CELL+
O2 CELL-
RED
WHT
ORG
BLK
BLK
O2 Cell
& Thermocouple
Connector (J5)
EXC+
CO ACT+
CO ACTCO REF+
CO REFCJC+
CJCEXC-
#1
SHLD
To Ground
Screw
ELECTRONICS HOUSING
Heater
Power Cable
6-6
NOTE: Wire colors shown are for cables supplied
by Emerson Process Management.
37390016
Signal Cable
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
Install Remote Electronics Housing
1. Mount remote electronics housing on wall or pipe within distance of
signal and heater cables in use.
2. Remove the electronics housing cover.
3. If removed, install the power and signal cables and the customer power
and signal conduits and wiring at the electronics housing.
4. Connect the signal cable leads to the O2 cell and thermocouple
connector (J4), and to the COe and CJC connector (J5), Figure 6-4.
5. Connect the heater cable leads to the heater power connector (J3).
6. Connect the line (L1 wire) to the L1 terminal, and the neutral (N wire) to
the N terminal on the AC power input terminal block, Figure 6-4.
7. Connect the ground lead to the ground stud. Secure the connection with
two nuts. Attach a separate ground lead (G wire) from the ground stud
to the G terminal on the power input terminal block.
8. If used, connect external relay leads to the alarm output relay terminal.
9. Connect the test gas and instrument air lines to the remote electronics
housing. Connect the calibration gas line and instrument air line to the
remote electronics housing.
10. Refer to Figure 6-1 and Figure 6-4. Make sure all test gas lines and
electrical connections are complete.
11. Install the cover on the electronics housing.
12. Turn on the test gasses at the cylinders and open the instrument air
valve.
13. Restore power to the system.
6-7
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
REPAIR SENSOR
HOUSING
Use the following procedures to remove damaged components from the OCX
8800 sensor housing and to install new replacement parts. Disassemble the
unit only as needed to replace damaged components. Use the assembly
procedures that apply to install replacement parts and reassemble the unit.
Sensor Housing
Disassembly
Remove Cover and Terminals Insulator
1. Loosen screw (1, Figure 6-5) and slide locking clip (2) away from cover.
Retighten screw (1).
2. With two hands or strap wrench, turn cover (3) counterclockwise to
loosen. Unthread and remove cover.
3. Inspect cover o-ring (4) for wear or damage. Replace cover o-ring if
damaged.
4. Unsnap terminal marking plates (5) and remove terminal insulator (6).
6-8
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
Figure 6-5. Removal of O2 Cell
and Heater Strut Assembly
Screw
Locking Clip
Cover
O-ring
Marking Plate
Terminal Insulator
Reference Air Tube
Sensor Housing
Heater Strut Assembly
Gasket
Screw
Heater Clamp
Heater Rod
3
7
4
9
5
6
11
10
12
8
13
2
1
37390067
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
6-9
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
Figure 6-6. O2 Cell,
Thermocouple, and
Heater Connections
O2 Cell Wires
Thermocouple Wires
+
Return Wire
O2 Heater Wires
+
+
ORG
RED
GRN
YEL
T/C O2
+
Sensor Housing Terminals
T/C CO
+
2
HTR 02
1
2
HTRCO
02
HTR
1
2
HTR SB
1
+
+
-
-
O2 Cell and Heater Strut
Assembly
T/C SB
+
Sample
Block
- EXC Thermocouple
-
Sample Block
Heater Rods
CJC
+
CO REF
+
CO ACT
+
+ EXC
37390068
+
O2
-
Remove O2 Cell and Heater Strut Assembly
1. Remove reference air tube (7, Figure 6-5) from sensor housing (8).
2. See Figure 6-6. Disconnect and tag O2 heater wires, O2 cell and return
wires, and thermocouple wires at the sensor housing terminals.
3. Remove the O2 cell and heater strut assembly (9, Figure 6-5) from
sensor housing (8). Remove and discard gasket (10).
Remove Sample Block Heater Rods
1. Disconnect sample block heater rod wires from terminal block. Refer to
Figure 6-6.
6-10
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
2. Loosen screws (11, Figure 6-5) and rotate heater clamps (12) to release
heater rods (13). One heater clamp secures each heater rod.
3. Slide sample block heater rods (13) out of housing (8).
Remove COe Sensor Assembly
1. Disconnect COe heater, thermocouple, and sensor wires from terminal
blocks. Refer to Figure 6-7.
2. Remove insulator (1, Figure 6-8).
T/C O2
+
RED
-
YEL
T/C CO
+
-
CJC Sensor
T/C SB
+
BLU
- EXC
CJC
+
BLU
2
HTRCO
02
HTR
1
WHT
2
HTR SB
1
RED
WHT
2
HTR 02
1
RED
RED
NOTE: All wires
at these terminals
are in the CJC
current loop.
COe Sensor
Assembly
CO REF
+
CO ACT
+
+ EXC
COe Heater Wires
COe Sensor Wires
COe Thermocouple Wires
37390018
Sensor Housing Terminals
-
-
O2
+
Figure 6-7. COe Sensor,
Thermocouple, and Heater
Connections
6-11
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
NOTE
For easier access, you may remove two screws from base of terminal block
mounting (13) and move terminal block assembly out of the way.
3. Remove tubes (2, 3, and 4) from COe sensor assembly (5), eductor
fittings (12 and 15), and sensor housing (7).
4. Unfasten bayonet connector of COe thermocouple (8) and remove
thermocouple.
5. Loosen clamp screw of COe band heater (9) until heater rotates freely
on sensor holder (11).
Figure 6-8. Removal of
COe Sensor Assembly
4
3
5
8
1
11
9
2
15
10
6
12
13
14
7
6-12
16
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
15.
16.
Insulator
Dilution Air Tube
COe Extractive Tube
Eductor Air Tube
COe Sensor Assembly
Eductor
Sensor Housing
COe Thermocouple
Heater Insulator
COe Band Heater
Sensor Holder
Eductor Elbow
Terminal Block Mounting
Eductor Holder
Tube Fitting
CJC Sensor
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
6. See Figure 6-9. Using straightedge on the sensor holder flat, as shown,
matchmark upper flange of sensor housing to show correct alignment of
sensor holder.
7. With one wrench holding eductor elbow (12, Figure 6-8), and one
wrench on flats of sensor holder (11), unthread and remove COe sensor
assembly (5). Do not allow eductor elbow to turn.
8. Slide band heater (10) and heater insulator (9) from sensor holder (11).
Figure 6-9. Alignment of
COe Sensor Assembly
Matchmark
Straightedge
Sensor
Holder
Flat
37390035
Matchmark
6-13
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
Remove Eductor
The O2 cell and heater strut assembly (9, Figure 6-5) and the COe sensor
assembly (5, Figure 6-8) must be removed before you start this procedure.
1. Use straightedge to matchmark alignment of eductor flat and elbow, as
shown in Figure 6-10.
Figure 6-10. Eductor Alignment
Matchmarks
Eductor Flat
Straightedge
37390036
Matchmark
2. Unscrew terminal block mounting (13, Figure 6-8). Move terminal block
mounting away from eductor (6).
3. Unscrew eductor holder (14) with eductor (6) and fittings (12 and 15)
from sensor housing (7).
4. Clamp flats of eductor (6) in jaws of bench vise.
Use heat resistant gloves when removing mating parts from the eductor. The mating parts
are bonded with a thread sealing compound. The compound softens at 450°F (232°C). The
heated parts can cause severe burns.
5. Use a propane torch to heat the eductor (6) to 450°F (232°C), minimum.
Apply the heat near the screw threads.
6. While heating the eductor (6), use wrench to apply removal torque to
elbow (12), eductor holder (14), or tube fitting (15) until the thread
sealant softens. Remove mating parts (12, 14, and 15).
7. Use MEK or methylene chloride solvent to clean thread sealant residue
from the pipe threads of the mating parts. Refer to applicable MSDS
sheet for solvent handling precautions.
6-14
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
Figure 6-11. Removal of Sample
and Exhaust Tubes
Remove Sample and Exhaust Tubes
Use heat resistant gloves when removing the probe tube or exhaust tube. The tubes are
bonded with a thread sealing compound. The compound softens at 450°F (232°C). The
heated parts can cause severe burns.
1. Secure the sensor housing (1, Figure 6-11) in soft (plastic, wood, or
brass) vice jaws.
2. Use a propane torch to heat the sample tube (2) or exhaust tube (3) to
450°F (232°C), minimum. Apply the heat near the thr eaded end of the
tube.
3. While heating the tube, use a pipe wrench to apply removal torque to
part being removed. Apply torque until the pipe thread sealant softens.
Remove and discard the used sample tube (2), exhaust tube (3), or
in-situ filter (4).
4. Use MEK or methylene chloride solvent to clean thread sealant residue
from the internal pipe threads in the housing. Refer to applicable MSDS
sheet for solvent handling precautions.
1
3
2
1.
2.
3.
4.
Sensor Housing
Sample Tube
Exhaust Tube
In Situ Filter
37390048
4
6-15
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
Figure 6-12. O2 Cell, Heater,
and Thermocouple, Exploded
View
Do not remove the O2 cell unless you are certain it needs to be replaced. Removal may
damage the cell and platinum pad. Remove the O2 cell for cell replacement only.
Disassemble O2 Cell and Heater Strut Assembly
Do not attempt to replace the O2 cell until all other possibilities for poor
performance have been considered. If cell replacement is needed, order the
O2 cell replacement kit (Refer to Section 8, Replacement Parts).
4
6
The O2 cell replacement kit contains an O2 cell and flange assembly, gaskets,
socket head cap screws, and anti-seize compound. The items are carefully
packaged to preserve precise surface finishes.
7
8
9
Do not remove items from the package until they are ready to be used.
1. Remove the four allen cap screws (1, Figure 6-12) from the O2 cell (2).
Remove the O2 cell. The cell flange has a notch that may be used to
gently pry the flange away from heater tube (3).
12
11
10
NOTE
The pad on the end of contact/thermocouple assembly (4) will sometimes
fuse to the O2 cell (2).
1.
2.
3.
4.
Screw
O2 Cell
Heater Tube
Contact/
Thermocouple
Assembly
5. Gasket
6. Screw
7. Lockwasher
8. Return Wire
9. Heater Strut
Assembly
10. Spring Clip
11. Spring
12. Strut Bracket
2. If the O2 cell is fused to the contact pad, push the O2 cell back into the
heater tube (against spring pressure) and quickly twist the O2 cell. The
cell and contact pad should separate. If the contact pad stays fused to
the cell, a new contact/thermocouple assembly (4) must be installed.
3. Remove and discard gasket (5). Clean the mating surface of heater
tube (3). Remove burrs and raised surfaces with a block of wood and
crocus cloth.
Use care when handling contact and thermocouple assembly. The ceramic rod in this
assembly is fragile.
3
5
4. Remove screws (6), lockwashers (7), return wire (8), and heater strut
assembly (9).
Test Gas
Passage
Holes
1
6-16
37390032
2
5. If replacing contact and thermocouple assembly (4), use a pencil to
mark location of spring clip (10) before removing. Squeeze tabs on
spring clip to remove. Retain spring clip and spring (11); replace if
damaged.
6. While carefully handling new contact and thermocouple assembly (4)
lay old assembly next to new one. Transfer match marks to new
assembly.
7. Carefully guide new contact and thermocouple assembly (4) through
strut bracket (12), spring (11), and spring clip (10) until spring clip
reaches pencil mark.
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
Disassemble COe Sensor Assembly
1. Carefully remove screws (1, Figure 6-13), lockwashers (2), and COe
sensor (3) from sensor holder (4). Remove and discard gasket (5).
2. If damaged, use the following procedure to remove thermocouple
adaptor (6) from sensor holder (4):
a. Use a propane torch to heat the thermocouple adaptor to 450°F
(232°C), minimum.
b. While heating, use a flat-head screwdriver to apply removal torque.
Apply torque until the pipe thread sealant softens. Remove and
discard the thermocouple adaptor.
c. Use MEK or methylene chloride solvent to clean thread sealant
residue from the internal pipe threads in the sensor holder. Refer to
applicable MSDS sheet for solvent handling precautions.
Figure 6-13. COe Sensor,
Exploded View
3
1
2
7
9
6
8
5
1.
2.
3.
4.
5.
6.
7.
8.
9.
Screw
Lockwasher
COe Sensor
Sensor Holder
Gasket
Thermocouple Adapter
Pre-Heater
Plug
Stainless Steel Balls
37390030
4
6-17
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
Always remove the stainless steel balls (approximately 200) from sensor holder before
removing or installing pre-heater. Turning pre-heater in the sensor holder with the stainless
steel balls in place will cause permanent damage to the pre-heater.
3. If pre-heater (7) is to be removed, clamp flats of sensor holder (4) in vise
jaws with plug (8) pointing up. Remove plug. Unclamp sensor holder
and pour stainless steel balls (9) into a container.
NOTE
Pre-heater should only be removed when pre-heater or sensor holder is
damaged. If removal is not required, leave the pre-heater installed in the
sensor holder.
4. Unthread and remove pre-heater (7).
5. Use a cleaning solvent to thoroughly clean stainless steel balls (9) and
pre-heater chamber in sensor housing (4). Refer to applicable MSDS
sheet for solvent handling precautions.
Sensor Housing
Assembly
Assemble COe Sensor Assembly
Always remove the stainless steel balls (approximately 200) from sensor holder before
removing or installing pre-heater. Turning pre-heater in the sensor holder with the stainless
steel balls in place will cause permanent damage to the pre-heater.
1. If pre-heater (7, Figure 6-13) was removed, apply pipe thread sealant
(Loctite #567) to the external pipe threads of pre-heater (7) and plug (8).
Do not apply sealant to the first turn of the pipe threads.
2. Clamp flats of sensor holder (4) in vise jaws with pre-heater port
pointing up.
3. Install and tighten pre-heater (7). Align pre-heater to flat of sensor
holder (4) as shown in Figure 6-14.
4. Invert sensor holder (4, Figure 6-13) in vise and pour stainless steel
balls (9) into plug port. Press down on stainless steel balls and tap
sensor holder with plastic hammer to compact balls in pre-heater
chamber.
5. Install and tighten plug (8).
Use care when installing the combustibles (COe) sensor. The RTD elements are fragile and
correct alignment in sensor holder is required for proper OCX operation.
6. Lubricate and install COe sensor gasket (5). Apply anti-seize compound
to threads of screws (1).
6-18
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
Figure 6-14. COe Sensor and
Pre-Heater Alignment
COe
SENSOR ASSEMBLY
TOP VIEW
THERMOCOUPLE
Sensor
Holder
Sensor
Holder
Flat
Pre-Heater
Flat
Sensor
Holder
Flat
37390034
2 to 2-1/4 in.
(51 to 57 mm)
COe Sensor
Flat
7. Install COe sensor (3), lockwashers (2), and screws (1). Rotate flat of
COe sensor (3) to center of sensor holder (4).
8. Align COe sensor flat parallel to sensor holder flat, as shown in
Figure 6-14. Tighten screws (1, Figure 6-13).
9. If replacing thermocouple adaptor (6), apply anti-seize to the pipe
threads. Install and tighten thermocouple adaptor.
Assemble O2 Sensor and Heater Strut Assembly
1. See Figure 6-12. Assemble O2 cell (2), gasket (5), and heater tube (3).
Make sure the test gas passage holes line up with each other in all
components.
2. Apply a small amount of anti-seize compound to the screw threads and
use screws (1) to secure assembly. Torque to 35 in-lbs (4 N·m).
3. Carefully slide O2 heater strut assembly (9) into heater tube (3).
4. Press down on the back plate of strut bracket (12) to ensure spring (11)
tension is present to hold contact pad against O2 cell (2).
5. Secure strut bracket (12) and return wire (8) with four screws (6) and
lockwashers (7). Make sure return wire (8) is tightly fastened. This is the
ground side connection for the O2 cell.
Install Sample and Exhaust Tubes
1. See Figure 6-11. Apply pipe thread sealant (Loctite #567) to the
replacement sample tube (2) or exhaust tube (3) pipe threads. Do not
apply sealant to the first turn of the pipe threads.
2. Thread the sample tube (2) or exhaust tube (3) into the housing (1). Use
a pipe wrench to tighten the tube.
3. If used, install and tighten in-situ filter (4).
6-19
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
Install Eductor
If installed, the O2 cell and heater strut assembly (9, Figure 6-5) must be
removed from sensor housing (8), before you install the eductor.
1. Apply pipe thread sealant (Loctite #567) to the external pipe threads of
eductor (6, Figure 6-15). Do not apply sealant to the first turn of the pipe
threads.
Figure 6-15. Installation of
Eductor and COe Sensor
4
3
5
8
1
11
9
2
15
10
6
12
13
14
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
15.
16.
Insulator
Dilution Air Tube
COe Extractive Tube
Eductor Air Tube
COe Sensor Assembly
Eductor
Sensor Housing
COe Thermocouple
Heater Insulator
COe Band Heater
Sensor Holder
Eductor Elbow
Terminal Block Mounting
Eductor Holder
Tube Fitting
CJC Sensor
37390047
7
16
6-20
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
2. Install and tighten eductor (6) in eductor holder (14).
3. Install and tighten elbow (12) on eductor (6). Male port of elbow must
point up and be in line with long axis of eductor.
4. Apply anti-seize compound to the external pipe threads of eductor
holder (14).
5. Install and tighten eductor holder (14) in sensor housing (7). Align
eductor with matchmarks, as shown in Figure 6-16.
Figure 6-16. COe Sensor
Parts Alignment
Eductor Flat
Straightedge
37390036
Matchmark
6-21
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
Install COe Sensor Assembly
Figure 6-17. Band Heater Height
COe Sensor
1. Apply pipe thread sealant (Loctite #567) to the exposed pipe threads
of eductor elbow (12, Figure 6-15). Do not apply sealant to the first turn
of the pipe threads.
2. Screw sensor holder (11) onto eductor elbow (12).
0.37 in.
(9,4 mm)
3. With wrenches on eductor elbow (12) and on flats of sensor holder (11),
tighten sensor holder. Do not allow eductor elbow to turn.
4. Tighten sensor holder (11) to align outside flat with matchmark on
sensor housing flange, as shown in Figure 6-18.
Band Heater
Insulator
37390058
The heater insulator prevents current leakage between the band heater and the sensor
holder. Failure to properly install the insulator may cause the device to trip a ground fault
interrupt circuit.
5. Wrap heater insulator (9) around sensor holder (11). Make sure the
insulator joint lines up with the band gap of the COe band heater (10).
6. Slide COe band heater (10, Figure 6-15) up onto sensor holder (11). Do
not tighten the band heater at this time. Heater must rotate freely around
sensor holder.
7. Check for proper height of COe heater thermocouple (Figure 6-14).
Thread bayonet connector up or down to adjust height.
8. Install and fasten thermocouple (8, Figure 6-15).
9. Position band heater as shown in Figure 6-17 and Figure 6-18 and
tighten band heater clamp screw. The heater insulator (9) end joint must
line up with the band gap of the COe band heater (10).
6-22
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
Figure 6-18. COe Sensor
Holder Alignment
Matchmark
Straightedge
Sensor
Holder
Flat
37390035
Matchmark
6-23
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
10. Reconnect the COe sensor, thermocouple, and heater wires at the
sensor housing terminal blocks. Refer to Figure 6-19.
11. Install and fasten the COe insulator (1, Figure 6-15) around COe sensor
assembly (5). All wiring must remain outside of the insulator.
12. If terminal block mounting (13, Figure 6-15) was moved, reinstall with
two base mounting screws.
Figure 6-19. COe Sensor,
Thermocouple, and Heater
Connections
-
-
T/C O2
+
RED
-
YEL
T/C CO
+
-
CJC Sensor
T/C SB
+
BLU
- EXC
CJC
+
BLU
2
HTR
HTRCO
02
1
WHT
2
HTR SB
1
RED
WHT
2
HTR 02
1
RED
RED
Sensor Housing Terminals
+
O2
NOTE: All wires
at these terminals
are in the CJC
current loop.
COe Sensor
Assembly
CO REF
+
CO ACT
+
+ EXC
COe Sensor Wires
COe Thermocouple Wires
6-24
37390018
COe Heater Wires
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
Figure 6-20. Installation of O2 Cell
and Heater Strut Assembly
Screw
Locking Clip
Cover
O-ring
Marking Plate
Terminal Insulator
Reference Air Tube
Sensor Housing
Heater Strut Assembly
Gasket
Screw
Heater Clamp
Heater Rod
3
7
4
9
5
6
11
10
12
8
13
2
1
37390067
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
Install Sample Block Heater Rods
1. Before installing sample block heater rods (13, Figure 6-20), evenly coat
the heater rods with Watlube heater release agent.
6-25
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
Figure 6-21. O2 Cell,
Thermocouple, and
Heater Connections
O2 Cell Wires
Thermocouple Wires
+
Return Wire
O2 Heater Wires
+
+
ORG
RED
GRN
YEL
T/C O2
+
Sensor Housing Terminals
2
HTR
HTRCO
02
1
2
HTR SB
1
+
T/C CO
+
2
HTR 02
1
+
-
O2 Cell and Heater Strut
Assembly
T/C SB
+
Sample
Block
- EXC Thermocouple
-
Sample Block
Heater Rods
CJC
+
CO REF
+
CO ACT
+
+ EXC
37390068
+
O2
-
2. Install the heater rods (13), heater clamps (12), and screws (11).
3. Reconnect the heater rod leads at the sensor housing terminal blocks,
(Figure 6-21).
6-26
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
Install O2 Cell and Heater Strut Assembly
1. Rub a small amount of anti-seize compound on both sides of new
gasket (10, Figure 6-20).
2. Apply anti-seize compound to threads of O2 cell and heater strut
assembly (9) and sensor housing (8).
Stripped threads on the O2 cell and heater strut assembly can allow gas leakage. Gas
leakage can affect the O2 measurements and calibration. Avoid over-tightening the O2 cell
and heater strut assembly.
3. Install O2 cell and heater strut assembly (9) in sensor housing (8). Snug
up, but do not over-tighten the assembly.
4. Reconnect the lead wires from O2 cell, heater, and thermocouple to the
sensor housing terminal blocks. Refer to Figure 6-21.
5. Install reference air tube (7, Figure 6-20) in sensor housing (8). Make
sure that the open end of reference air tube extends into heater tube of
O2 cell and heater strut assembly (9).
Install Terminals Insulator and Cover
1. Install insulator (6, Figure 6-20) over uppermost terminal blocks.
Position one side of insulator against terminal blocks and snap terminal
marking plate (5) to mating stand-off.
2. Position opposite side of insulator (6) and secure with related marking
plate (5).
3. If removed, install cover gasket (4). Screw cover (3) onto sensor
housing (8). Tighten cover firmly.
4. Align locking clip (2) with gap between cover ribs.
5. Loosen screw (1) and slide locking clip (2) fully into gap between cover
ribs. Retighten screw (1).
Sensor Housing Leak Test
1. Install 1/4 NPT cap on dilution air inlet fitting. Install a 1/4 NPT cap on
sample tube (2, Figure 6-11) or plug 1/4 NPT sample inlet port. Capped
or plugged ports must be air tight.
2. If not in place, install exhaust tube (3, Figure 6-11) in exhaust port
according to the instructions provided.
3. Connect a calibrated manometer to the CAL GAS inlet port.
4. Connect and apply clean instrument air at 35 psig (241 kPa gage) to the
instrument air inlet fitting.
5. Observe the manometer reading. The reading should be from 10 to 13
inches, Water Column. Locate and correct leaks if the reading is less
than 10 inches WC.
6-27
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
REPAIR ELECTRONICS
HOUSING
Use the following procedures to remove damaged components from the OCX
8800 electronics housing and to install new replacement parts. Disassemble
the unit only as needed to replace damaged components. Use the assembly
procedures that apply to install replacement parts and reassemble the unit.
Electronics Housing
Disassembly
Remove Cover
1. See Figure 6-22. Loosen screw (1) and slide locking clip (2) away from
cover (3). Retighten screw (1).
2. With two hands or strap wrench, turn cover (3) counterclockwise to
loosen. Unthread and remove cover.
3. Inspect cover gasket (4) for wear or damage. Replace cover gasket if
damaged.
Remove Flash PROM
Electrostatic discharge (ESD) protection is required to avoid damage to the electronic
circuits.
1. Locate Flash PROM access port in electronics stack (5, Figure 6-22).
2. Use suitable IC removal tool to remove Flash PROM (6).
6-28
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
Figure 6-22. Removal/Installation of
Electronics Housing Components
3
13
6
14
5
4
1. Screw
2. Locking Clip
3. Cover
4. O-ring
5. Electronics Stack
6. Flash PROM
7. thru 12. Not used
13. Screw
14. Lockwasher
15. Solenoid Valve (3 way)
16. Solenoid Valve
15
16
2
37390037
1
Remove Electronics Stack
1. Unplug power cable, signal cable, and solenoid lead connectors from
terminals of electronics stack (5, Figure 6-22).
2. Remove two screws (13) and lockwashers (14).
3. Remove electronics stack (5).
6-29
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
Remove Solenoid Valves
1. Disconnect solenoid leads from mating terminal connector.
2. Remove top nut of solenoid valve (15 or 16, Figure 6-22).
3. Remove the solenoid coil assembly and washer.
4. Unthread and remove solenoid valve base.
Remove EMI Filter and Terminal Block
1. Disconnect EMI filter wiring (Figure 6-23) at terminal block (3).
2. Disconnect EMI filter wiring at AC power input terminal block on
electronic stack.
3. Unbolt and remove EMI filter (1) from electronic stack.
4. Remove ground wire (2) from terminal block (3).
5. Unbolt and remove terminal block (3) from electronic stack.
Figure 6-23. Removal/
Installation of EMI Filter
Blue
Brown
Blue
Green
Brown
Green
Green
1
3
Electronics Housing
Assembly
37390083
2
Install EMI Filter and Terminal Block
1. Install replacement EMI filter (1, Figure 6-23) and/or terminal block (3)
on electronic stack.
2. Refer to wiring details in Figure 6-23. Connect EMI filter wiring and
ground wire (2) at terminal block (3).
3. Connect EMI filter wiring at AC power input terminal block on electronic
stack.
6-30
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
Install Solenoid Valves
1. Disassemble replacement solenoid valve (15 or 16, Figure 6-22).
2. Install new solenoid valve base. Be careful not to overtighten.
3. Install new washer and solenoid coil assembly and secure with nut.
4. Connect the solenoid leads to the proper terminations on the solenoid
power terminal block (Figure 6-26).
Install Electronics Stack
1. Install electronics stack (5, Figure 6-22) and secure with lockwashers
(14) and screws (13).
2. See Figure 6-25 and Figure 6-26. Reconnect power cable, signal cable,
and solenoid lead connectors to electronics stack terminals.
Install Flash PROM
Figure 6-24. Flash PROM
Alignment
Corner
Bevel
Electrostatic discharge (ESD) protection is required to avoid damage to the electronic
circuits.
1. Locate Flash PROM access port in electronics stack (5, Figure 6-22).
Flash
PROM
2. See Figure 6-24. Align Flash PROM (6, Figure 6-22) with mating
receptacle in access port of electronics stack as shown. Flash PROM
corner bevel must be in upper left corner of receptacle.
Access
Port
37390060
3. Install Flash PROM (6).
6-31
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
2HTR CO
1HTR CO
2HTR O2
1HTR O2
2HTR SB
1HTR SB
SHIELD
#1
YEL
RED
BLK
ORG
WHT
BLU
GRN
Figure 6-25. Electronics Housing
Cable Connections
To
Ground
Screw
Heater Power
Connector (J3)
#1
#1
RED
BLK
WHT
BLK
GRN
BLK
1
BLU
BLK
#1
#1
COe Sensor
& CJC
Connector (J4)
T/C CO+
T/C COT/C SB+
T/C SBT/C O2+
T/C O2-
YEL
BRN
BLK
O2 CELL+
O2 CELL-
RED
WHT
ORG
BLK
BLK
O2 Cell
& Thermocouple
Connector (J5)
EXC+
CO ACT+
CO ACTCO REF+
CO REFCJC+
CJCEXC-
#1
SHLD
To Ground
Screw
ELECTRONICS HOUSING
Heater
Power Cable
6-32
NOTE: Wire colors shown are for cables supplied
by Emerson Process Management.
37390016
Signal Cable
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
Figure 6-26. Solenoid
Power Terminals
SOL 2
SOL 3
SOL 4
SOL 1
TOP VIEW
SOL 4
CO Test Gas
SOL 3
SOL 2
High O2 Test Gas
Low O2 Test Gas
SOL 1
Instrument Air
#1
37390033
#1
Install Cover
1. If removed, install cover gasket (4, Figure 6-22). Screw cover (3) onto
electronics housing. Tighten cover firmly. Align locking clip (2) with gap
between cover ribs.
2. Loosen screw (1) and slide locking clip (2) fully into gap between cover
ribs. Retighten screw (1).
6-33
Instruction Manual
OCX 8800
REPLACE TUBE
FITTINGS
IM-106-880ASB, Rev 1.2
July 2008
The OCX transmitter housings have special tube fittings that, if clogged or
damaged, must be replaced with the same type of fitting. The special tube
fittings have alpha or numeric codes etched on the fitting. Unetched tube
fittings are standard 1/4 inch, stainless steel fittings.
‘E’ Type Fitting
The ‘E’ type fitting is an eductor drive air fitting for the OCX 8800 general
purpose sensor housing. It is a 1/8 inch tube fitting with a built-in 0.011
inch restrictor orifice. It seats in a threaded base port inside the housing.
‘R’ Type Fitting
The ‘R’ type fitting is a reference air line fitting for the general purpose and
hazardous area sensor housings. This is a 1/4 inch tube fitting with a
built-in 0.007 inch restrictor orifice.
Remove Tube Fittings
The OCX construction includes pipe thread sealant to seal fittings in all ports
that pass through to an outer wall of the instrument housing base. Use the
following instructions to loosen and remove tube fittings that are secured with
pipe thread sealant.
Use heat resistant gloves when removing a damaged tube fitting. The pipe threads are
bonded with a pipe thread sealant. The thread sealant softens at 450°F (232°C). The
heated parts can cause severe burns.
1. Secure sensor housing (1, Figure 6-27) or electronics housing (2) in soft
(plastic, wood, or brass) vice jaws.
2. To soften the pipe thread sealant, use a propane torch to heat the tube
fitting (3, 4, or 5) to 450°F (232°C), minimum.
3. While heating the tube fitting, use a wrench to apply removal torque until
the pipe thread sealant softens. Remove and discard the used fitting.
4. Use MEK or methylene chloride solvent to clean thread sealant residue
from the internal pipe threads in the housing. Refer to applicable MSDS
sheet for solvent handling precautions.
6-34
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
Figure 6-27. Removal
of Tube Fittings
5
3
4
1.
2.
3.
4.
5.
Sensor Housing
Electronics Housing
Tube Fitting
Tube Fitting (Type R)
Tube Fitting (Type E)
1
3
3
Install Tube Fittings
37390070
3
2
1. Verify that the replacement tube fitting is identical to the item removed.
Special fittings are etched with code markings.
2. Apply pipe thread sealant (Loctite #567) to the mating threads of the
tube fitting. Do not apply sealant to the first turn of the external pipe
threads.
3. Install and tighten the tube fitting in the mating port of sensor or
electronics housing.
6-35
Instruction Manual
OCX 8800
6-36
IM-106-880ASB, Rev 1.2
July 2008
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
Section 7
OCX 8800
Troubleshooting
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 7-1
Diagnostic Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 7-2
Fault Isolation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 7-3
Alarm Relay Events . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 7-11
Install all protective equipment covers and safety ground leads after troubleshooting.
Failure to install covers and ground leads could result in serious injury or death.
OVERVIEW
The troubleshooting section describes how to identify and isolate faults that
may develop in the OCX 8800. When troubleshooting the OCX 8800,
reference the following information.
Grounding
It is essential that adequate grounding precautions are taken when installing
the system. Thoroughly check both the probe and electronics to ensure the
grounding quality has not degraded during fault finding. The system provides
facilities for 100% effective grounding and total elimination of ground loops.
Electrical Noise
The OCX 8800 has been designed to operate in the type of environment
normally found in a boiler room or control room. Noise suppression circuits
are employed on all field terminations and main inputs. When fault finding,
evaluate the electrical noise being generated in the immediate circuitry of a
faulty system. Ensure all cable shields are connected to earth.
Electrostatic Discharge
Electrostatic discharge can damage ICs in the electronics. Before removing or
handling the processor board or the ICs, ensure you are at ground potential.
http://www..raihome.com
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
Total Power Loss
In the event that the OCX 8800 will not power up at all, check the incoming
power supply to make sure power is being delivered to the OCX 8800. If the
incoming power supply is good, then check fuses F1 and F6 in the electronics
housing. Refer to Figure 7-1 for fuse locations.
Figure 7-1. Fuse Locations
F6
Neutral (N)
10 Amp, 250 VAC
F1
Line (L1)
10 Amp, 250 VAC
F3
O2 and COe Heater
4 Amp, 250 VAC
37390079
F4
Sample Block Heater
8 Amp, 250 VAC
DIAGNOSTIC ALARMS
Always install a blocking diode on the customers relay coil. Failure to install a blocking diode
may create noise spikes and cause faults in the OCX electronics.
The OCX 8800 is equipped with a set of alarm relay contacts on the
microprocessor board in the electronics housing. This set of dry contacts can
be connected to any customer supplied relay device, 30 VDC, 30 mA
maximum. A blocking diode is required on the customers relay coil.
Any fault condition in the OCX 8800 will trip the alarm relay. The optional SPA
with HART programmable alarm indicates LOW O2, HIGH COe, Calibration
Status, and Unit Failure. For more information refer to Appendix B - SPA with
HART Alarm.
7-2
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
FAULT ISOLATION
OCX 8800
Faults in the OCX 8800 Transmitter are indicated by messages displayed on
the 275/375 Field Commiunicator. Fault indications that can appear are listed
in Table 7-1, Troubleshooting.
If a fault is indicated on the Field Communicator, locate the fault indication in
Table 7-1. For each fault listed, there are related Probable Causes and
Recommended Corrective Actions. The Probable Causes are listed in the
order of most probable to least probable. Starting with the most probable
cause, inspect and test the unit to isolate the actual cause, then use the
Recommended Corrective Action listed to correct the problem.
Table 7-1. Troubleshooting
O2 Sensor R High (Oxygen sensor resistance high, > 5000 Ohms)
O2 Sensor Open (Oxygen sensor disconnected)
Probable Cause
Recommended Corrective Action
Loose or open O2 cell circuit connection
Check O2 cell circuit wires for breaks or loose connections. Repair lead wire break or
loose connections.
Check O2 cell impedance by reading the O2 Snsr R value via HART (see Figure 4-3,
sheet 2). If cell impedance is zero, replace O2 cell with cell replacement kit. If cell
impedance is less than 5000 ohms, check for cell housing ground fault. Repair ground
fault. If cell impedance is greater than 5000 ohms and no ground fault is indicated,
replace O2 cell with cell replacement kit.
O2 cell degraded or failed
Ref Current Err (RTD excitation current error) (reference current should be 5.02 mA)
Probable Cause
Recommended Corrective Action
Loose or open lead or circuit wire connection for
COe sensor or CJC sensor current loop
COe sensor grounded
Check all COe and CJC sensor current loop wiring per Figure 7-3 and Figure 6-7.
Correct wiring faults.
Check resistance of COe sensor leads to ground per Figure 6-7. Replace COe sensor if
resistance is less than 10M Ohms.
Check resistance of both COe sensor elements per Figure 6-7. Replace COe sensor if
resistance of sensor element is not between 100 and 250 Ohms.
Check resistance of CJC sensor to ground per Figure 6-7. Replace CJC sensor if
resistance is less than 10M Ohms.
Check resistance of CJC sensor per Figure 6-7. Replace CJC sensor if resistance of
sensor is not between 100 and 150 Ohms.
COe sensor failed
CJC sensor grounded
CJC sensor failed
O2 Temp Hi (Oxygen sensor heater temperature high, > 750ºC)
Probable Cause
Recommended Corrective Action
High noise in OCX power supply
Check power supply for line noise or voltage fluctuations. Install power line filter kit
(PN 6A00171G01) or high quality line filter for input power.
COe Temp Hi (Combustion sensor heater temperature high, 310ºC)
Probable Cause
Recommended Corrective Action
High noise in OCX power supply
Check power supply for line noise or voltage fluctuations. Install power line filter kit
(PN 6A00171G01) or high quality line filter for input power.
Table continued on next page
7-3
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
SB Temp Hi (Sample block heater temperature high, > 190ºC)
Probable Cause
Recommended Corrective Action
High noise in OCX power supply
Check power supply for line noise or voltage fluctuations. Install power line filter kit
(PN 6A00171G01) or high quality line filter for input power.
O2 Temp Very Hi (Oxygen sensor heater over maximum temperature, > 820ºC)
O2 Htr Rmp Rate (Oxygen sensor heater over maximum temperature ramp rate)
Probable Cause
Recommended Corrective Action
Incorrect O2 heater wiring
Check O2 heater wiring per Figure 7-3 and Figure 6-6. Check the wiring at the heater
and inside the electronics housing. Correct wiring fault. Perform Reset procedure in
Section 3, Configuration and Startup to continue operation.
Check O2 thermocouple wiring per Figure 7-3 and Figure 6-6. Check the wiring at the
thermocouple and inside the electronics housing. Correct wiring fault. Perform Reset
procedure in Section 3, Configuration and Startup to continue operation.
Replace electronics package.
Incorrect O2 thermocouple wiring
Electronics package failure
COe Temp Very Hi (Combustion sensor heater over maximum temperature, > 400ºC)
COe Htr Rmp Rate (Combustibles sensor heater over maximum temperature ramp rate)
Probable Cause
Recommended Corrective Action
Incorrect COe heater wiring
Check COe heater wiring per Figure 7-3 and Figure 6-7. Check the wiring at the heater
and inside the electronics housing. Correct wiring fault. Perform Reset procedure in
Section 3, Configuration and Startup to continue operation.
Check COe thermocouple wiring per Figure 7-3 and Figure 6-7. Check the wiring at the
thermocouple and inside the electronics housing. Correct wiring fault. Perform Reset
procedure in Section 3, Configuration and Startup to continue operation.
Replace electronics package.
Incorrect COe thermocouple wiring
Electronics package failure
SB Temp Very Hi (Sample block heater over maximum temperature, > 260ºC)
SB Htr Rmp Rate (Sample block heater over maximum temperature ramp rate)
Probable Cause
Recommended Corrective Action
Incorrect sample block heater wiring
Check sample block heater wiring per Figure 7-3 and Figure 6-6. Check the wiring at
the heater and inside the electronics housing. Correct wiring fault. Perform Reset
procedure in Section 3, Configuration and Startup to continue operation.
Check sample block thermocouple wiring per Figure 7-3 and Figure 6-6. Check the
wiring at the thermocouple and inside the electronics housing. Correct wiring fault.
Perform Reset procedure in Section 3, Configuration and Startup to continue operation.
Replace electronics package.
Incorrect sample block thermocouple wiring
Electronics package failure
O2 TC Open (Oxygen sensor heater thermocouple open)
Probable Cause
Recommended Corrective Action
O2 thermocouple or thermocouple circuit open
Check O2 thermocouple and circuit wires for breaks or loose connections per Figure 7-3
and Figure 6-6. Repair breaks or loose connections or replace failed thermocouple.
Perform Reset procedure in Section 3, Configuration and Startup to continue operation.
COe TC Open (Combustibles sensor heater thermocouple open)
Probable Cause
Recommended Corrective Action
Combustibles sensor thermocouple open
Check thermocouple resistance (lead to ground) of reference and active thermocouples
per Figure 7-3 and Figure 6-7. If either thermocouple is open or shorted to ground,
replace combustibles sensor.
Table continued on next page
7-4
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
SB TC Open (Sample block heater thermocouple open)
Probable Cause
Recommended Corrective Action
Sample block thermocouple or thermocouple
circuit open
Check sample block thermocouple and circuit wires for breaks or loose connections per
Figure 7-3 and Figure 6-6. Repair breaks or loose connections or replace failed
thermocouple. Perform Reset procedure in Section 3, Configuration and Startup to
continue operation.
O2 TC Shorted (Oxygen sensor heater thermocouple shorted)
Probable Cause
Recommended Corrective Action
O2 thermocouple or thermocouple circuit shorted
Check O2 thermocouple and circuit wires for short circuit condition per Figure 7-3 and
Figure 6-6. Repair shorted wiring or replace failed thermocouple. Perform Reset
procedure in Section 3, Configuration and Startup to continue operation.
Perform Reset procedure in Section 3, Configuration and Startup to continue operation.
If alarm persists, refer to the O2 Htr Failure alarm procedure.
Slow heatup during cold start
COe TC Shorted (Combustibles sensor heater thermocouple shorted)
Probable Cause
Recommended Corrective Action
Combustibles sensor thermocouple shorted
Check thermocouple resistance (lead to ground) of reference and active thermocouples
per Figure 7-3 and Figure 6-7. If either thermocouple is open or shorted to ground,
replace combustibles sensor.
Perform Reset procedure in Section 3, Configuration and Startup to continue operation.
If alarm persists, refer to the COe Htr Failure alarm procedure.
Slow heatup during cold start
SB TC Shorted (Sample block heater thermocouple shorted)
Probable Cause
Recommended Corrective Action
Sample block thermocouple or thermocouple
circuit shorted
Check sample block thermocouple and circuit wires for short circuit condition per
Figure 7-3 and Figure 6-6. Repair shorted wiring or replace failed thermocouple.
Perform Reset procedure in Section 3, Configuration and Startup to continue operation.
Perform Reset procedure in Section 3, Configuration and Startup to continue operation.
If alarm persists, refer to the SB Htr Failure alarm procedure.
Slow heatup during cold start
O2 TC Reversed (Oxygen sensor heater thermocouple reversed)
Probable Cause
Recommended Corrective Action
O2 thermocouple wires reversed
Check O2 thermocouple wiring per Figure 7-3 and Figure 6-6. Check the wiring at the
sensor and inside the electronics housing. Correct reversed-wires fault. Perform Reset
procedure in Section 3, Configuration and Startup to continue operation.
COe TC Reversed (Combustibles sensor block heater thermocouple reversed)
Probable Cause
Recommended Corrective Action
Combustibles thermocouple wires reversed
Check combustibles thermocouple wiring per Figure 7-3 and Figure 6-7. Check the
wiring at the sensor and inside the electronics housing. Correct reversed-wires fault.
Perform Reset procedure in Section 3, Configuration and Startup to continue operation.
SB TC Reversed (Sample block heater thermocouple reversed)
Probable Cause
Recommended Corrective Action
Sample block thermocouple wires reversed
Check sample block thermocouple wiring per Figure 7-3 and Figure 6-6. Check the
wiring at the sensor and inside the electronics housing. Correct reversed-wires fault.
Perform Reset procedure in Section 3, Configuration and Startup to continue operation.
Table continued on next page
7-5
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
ADC Failure (Voltage to digital conversion could not complete)
ADC Ref Error (Voltage to digital conversion not accurate)
Probable Cause
Recommended Corrective Action
Incorrect wiring between electronics and sensor
housings
Electronics package failure
Check all wiring between the electronics and sensor housings per Figure 7-3. Correct
wiring faults.
Replace electronics package.
O2 Htr Failure (Oxygen sensor heater could not reach final temperature)
Probable Cause
Recommended Corrective Action
O2 heater circuit wiring open
Check O2 cell heater circuit for broken wire or loose connection per Figure 7-3 and
Figure 6-6. Repair broken wire or loose connection.
Check resistance of O2 heater per Figure 6-6. Normal O2 heater resistance is 62.5
Ohms. Replace O2 heater if heater is open or has a large resistance.
Check heater fuse F3 in electronics housing per Figure 7-1. If open, locate and correct
cause of overload. If F3 is not open, or if cause of overload cannot be found, replace
electronics package.
O2 heater open
Heater electronics failure
COe Htr Failure (Combustibles sensor heater could not reach final temperature)
Probable Cause
Recommended Corrective Action
COe heater circuit wiring open
Check COe heater circuit for broken wire or loose connection per Figure 7-3 and
Figure 6-7. Repair broken wire or loose connection.
Check resistance of COe heater per Figure 6-7. Normal COe heater resistance is 97.7
Ohms. Replace COe heater if heater is open or has a large resistance.
Check heater fuse F3 in electronics housing per Figure 7-1. If open, locate and correct
cause of overload. If F3 is not open, or if cause of overload cannot be found, replace
electronics package.
COe heater open
Heater electronics failure
SB Htr Failure (Sample block heater could not reach final temperature)
Probable Cause
Recommended Corrective Action
Sample block heater circuit wiring open
Check sample block heater circuit for broken wire or loose connection per Figure 7-3
and Figure 6-6. Repair broken wire or loose connection.
Check resistance of sample block heater per Figure 6-6. Normal sample block heater
resistance is 36.4 Ohms each (18.2 Ohms with both heaters in parallel). Replace
sample block heater if heater is open or has a large resistance.
Check heater fuse F4 in electronics housing per Figure 7-1. If open, locate and correct
cause of overload. If F4 is not open, or if cause of overload cannot be found, replace
electronics package.
If above probable causes are not causing the SB heater failure, install flange insulator
(PN 6P00162H01).
Sample block heater open
Heater electronics failure
Sensor housing exposed to high wind and/or
extreme cold temperatures
Cal Warning (Calibration warning)
Cal Failed (Calibration failed)
Probable Cause
Recommended Corrective Action
Calibration gas supply low or gas connection
leaking
Check calibration gas supplies and connections. Adjust gas pressure and flow.
Replenish low calibration gas supplies and tighten or repair loose or leaking
connections. When calibration gas supplies are adequate, recalibrate.
Check O2 cell impedance by reading the O2 Snsr R value via HART menu tree (see
Figure 4-3, sheet 2). If cell impedance is zero, replace O2 cell. If cell impedance is less
than 5000 ohms, check for cell housing ground fault. Repair ground fault. If cell
impedance is greater than 5000 ohms and no ground fault is indicated, replace O2 cell.
O2 cell degraded or failed:
O2 Slope Error (Slope <34.5 mV/Dec or >57.5
mV/Dec)
O2 Constant Error (Constant not between -20
mV to +20 mV)
7-6
Flow calibration gas to the O2 cell. Read the cell millivolt output. Plot the cell millivolt
output and the calibration gas O2 concentration on the chart shown in Figure 7-2. If the
plotted values do not fall on the slope line shown in Figure 7-2, replace the O2 cell.
Table continued on next page
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
Cal Warning (Calibration warning)
Cal Failed (Calibration failed)
Probable Cause
Recommended Corrective Action
COe sensor degraded
Replace COe sensor or increase warning level (SYSTEM, CALIB SETUP, COe Slope
Warn via HART.)
COe Slope Error (Slope not between 200
ppm/Ohm to 4,500 ppm/Ohm)
COe Constant Error (Constant not between
-100 Ohms to +100 Ohms)
Low sample gas flow in sensor housing due to
flow path plugging
Check the following portions of the flow path for plugging:
• blowback filter
• in-situ filter
• eductor outlet path
Table continued on next page
Figure 7-2. Oxygen and Cell
Output
20
9
8
6
5
4
3
2
0.9
0.7
0.5
(73)
1
0.8
0.6
0.4
0.3
0.2
0.1
-20
0
20
40
60
80
100
ANALYZER OUTPUT (MILLIVOLT)
120
37390051
OXYGEN CONCENTRATION (%)
7
(23)
10
7-7
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
Board Temp Hi (Electronics temperature maximum exceeded, > 85ºC)
Probable Cause
Recommended Corrective Action
Electronics housing exposed to high ambient
temperature
Insulate housing from source of high temperature and/or install cooling fan to remove
heat from housing. Perform Reset procedure in Section 3, Configuration and Startup to
continue operation.
EEPRM Chksm Fail (Non-volatile parameter storage corrupted)
Probable Cause
Recommended Corrective Action
Unit powered down during calibration parameter
storage
Flash PROM failure
Perform Reset procedure in Section 3, Configuration and Startup. Recalibrate the OCX
and check/trim analog outputs.
Replace electronics package.
O2 Temp Low (Oxygen sensor heater temperature low, < 710ºC)
Probable Cause
Recommended Corrective Action
Sensor housing exposed to high wind and/or
extreme cold temperatures
High noise or voltage fluctuations in power
supply
Install sensor housing flange insulator (PN 6P00162H01).
Check power supply for line noise or voltage fluctuations. Install filter power line kit
(PN 6A00171G01) or high quality line filter for input power.
COe Temp Low (Combustion sensor heater temperature low, < 290ºC)
Probable Cause
Recommended Corrective Action
Sensor housing exposed to high wind and/or
extreme cold temperatures
High noise or voltage fluctuations in power
supply
Install sensor housing flange insulator (PN 6P00162H01).
Check power supply for line noise or voltage fluctuations. Install power line filter kit
(PN 6A00171G01) or high quality line filter for input power.
SB Temp Low (Sample block heater temperature low, < 150ºC)
Probable Cause
Recommended Corrective Action
Sensor housing exposed to high wind and/or
extreme cold temperatures
High noise or voltage fluctuations in power
supply
Install sensor housing flange insulator (PN 6P00162H01).
Check power supply for line noise or voltage fluctuations. Install power line filter kit
(PN 6A00171G01) or high quality line filter for input power.
Line Freq Error (AC power line frequency out of usable range, < 45 Hz or > 66 Hz)
Probable Cause
Recommended Corrective Action
High noise or voltage fluctuations in power
supply
AC power line frequency is outside the usable
range of the OCX universal power supply
Electronics package failure
Check power supply for line noise or voltage fluctuations. Install power line filter kit
(PN 6A00171G01) or high quality line filter for input power.
Correct power supply frequency. AC power line frequency must be between 50 and
60 Hz.
Check power supply frequency with a calibrated oscilloscope or frequency meter and
compare with line frequency. Replace electronics package if they do not agree within
1 Hz.
Line Voltage Low (AC power line voltage below minimum, < 85 VAC)
Probable Cause
Recommended Corrective Action
High noise or voltage fluctuations in power
supply
Electronics package failure
Check power supply for line noise or voltage fluctuations. Install power line filter kit
(PN 6A00171G01) or high quality line filter for input power.
Check power supply voltage and compare with line voltage. Replace electronics
package if they do not agree within 5%.
Table continued on next page
7-8
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
Line Voltage High (AC power line voltage above maximum, > 278 VAC)
Probable Cause
Recommended Corrective Action
High noise or voltage fluctuations in power
supply
Electronics package failure
Check power supply for line noise or voltage fluctuations. Install power line filter kit
(PN 6A00171G01) or high quality line filter for input power.
Check power supply voltage and compare with line voltage. Replace electronics
package if they do not agree within 5%.
Htr Relay Failed (Heater relay failure)
Probable Cause
Recommended Corrective Action
High noise or voltage fluctuations in power
supply
Electronics package failure
Check power supply for line noise or voltage fluctuations. Install power line filter kit
(PN 6A00171G01) or high quality line filter for input power.
Replace electronics package.
Out Brd Failure (Output board failure)
Probable Cause
Recommended Corrective Action
Electronics package failure
Replace electronics package.
7-9
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
GRN
#1
2HTR CO
1HTR CO
2HTR O2
1HTR O2
2HTR SB
1HTR SB
YEL
RED
BLK
ORG
WHT
BLU
SHIELD
Figure 7-3. Electrical Connections Between Electronics and Sensor Housing
To
ground
screw
Heater Power
Connector (J3)
#1
#1
RED
BLK
T/C CO+
T/C CO-
WHT
BLK
GRN
BLK
Heater Power Cable
#1
T/C SB+
T/C SBT/C O2+
T/C O2-
BLU
BLK
YEL
BRN
BLK
O2 CELL+
O2 CELL-
#1
EXC+
CO ACT+
CO ACT-
COe Sensor
and
CO REF+ Cold Junction
CO REF- Connector (J4)
CJC+
CJCEXC-
RED
WHT
ORG
BLK
BLK
O2 Sensor and
Thermocouple
Connector (J5)
SHLD
To ground
screw
ELECTRONICS HOUSING
#1
7-10
YEL
HTR
1 SB
EXC +
BRN
CO
ACT +
HTR
1 CO
2
2
RED
BLK
-
-
CO
REF
+
WHT
ORG
+
2
HTR
1 O2
37390014
GRN
BLU
WHT
RED
YEL
ORG
EXC-
-
BLK
BLK
WHT
+
BLK
-
T/C SB
RED
+
BLK
T/C CO
CJC
To ground screw
BLK
BLK
-
GRN
+
BLK
BLU
+
O2
-
T/C O2
-
Signal Cable
SENSOR HOUSING
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
ALARM RELAY EVENTS
OCX 8800
The OCX 8800 contains an alarm relay that can be configured to activate on
one of twelve different groups of events. These event groups, and the
conditions that trigger them, are listed in Table 7-2, Alarm Relay Event
Groups.
Table 7-2. Alarm Relay
Event Groups
Alarm Relay Event
In Calibration
O2 Temperature Error
Heater Failure
O2 Sensor Error
Calibration Failure
Calibration Warning
Board Temperature High
Unit Failure
Alarms/Conditions
Calibration in progress
Oxygen sensor heater temperature low (O2 Temp Low)
Oxygen sensor heater temperature high (O2 Temp Hi,
O2 Temp Very Hi)
RTD excitation current error (Ref Curr Err)
Oxygen sensor heater could not reach final temperature
(O2 HTR Failure)
Combustibles sensor heater could not reach final temperature
(COe Htr Failure)
Sample Block sensor heater could not reach final temperature
(SB Htr Failure)
Oxygen sensor resistance high (O2 Sensor R High)
Oxygen sensor disconnected (O2 Sensor Open)
Calibration failed (Cal Failed)
Calibration warning (Cal Warning)
Electronics temperature maximum exceeded (Board Temp Hi)
Any non-recoverable or heater relay off alarm (O2 Temp Hi,
O2 Temp Very Hi, COe Temp Hi, COE Temp Very Hi,
SB Temp Hi, SB Temp Very Hi, O2 Htr Ramp Rate,
COe Htr Rmp Rate, SB Htr Ramp Rate, O2 TC Shorted,
O2 TC Reversed, COe TC Shorted, COe TC Reversed,
SB TC Shorted, SB TC Reversed, ADC Failure,
ADC Ref Error, Board Temp Hi, EEPRM Chksm Fail,
Line Freq Error, Line Voltage Low, Line Voltage Hi,
Htr Relay Failed, Out Brd Fail)
Sample Block
Temperature Error
Sample block heater temperature low (SB Temp Low)
Sample block heater temperature high (SB Temp Hi,
SB Temp Very Hi)
RTD excitation current error (Ref Curr Err)
COe Sensor
Temperature Error
Combustibles sensor heater temperature low (COe Temp Low)
Combustibles sensor heater temperature high (COe Temp Hi,
COe Temp Very Hi)
RTD excitation current error (Ref Curr Err)
Power Input Error
AC power line frequency out of usable range (Line Freq Err)
AC power line voltage below minimum (Line Voltage Low)
AC power line voltage above maximum (Line Voltage Hi)
All Alarms
Any alarm
7-11
Instruction Manual
OCX 8800
7-12
IM-106-880ASB, Rev 1.2
July 2008
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
Section 8
OCX 8800
Replacement Parts
Sensor Housing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 8-2
Electronics Housing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 8-6
O2 Cell and Heater Strut Assembly . . . . . . . . . . . . . . . . . page 8-9
http://www..raihome.com
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
SENSOR HOUSING
Figure 8-1. Sensor Housing
Components
3
2
4
1
5
11
6
7
8
9
10
37390072
12
(Requires
item 13 for
assembly)
8-2
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
Index No.
Part Number
1
2
3
4
5
6
7
8
9
10
11
12
13
5R10190G02
6P00177H01
4851B46G03
1A99786H01
6P00163H01
1A99746H02
1A99749H01
1A99747H01
5R10200H01
6A00123G01
6P00155H02
1A98765H02
1A99520H01
OCX 8800
Description
O2 Cell and Heater Assembly
Heater Insulator, Mica
Combustibles Sensor Replacement Kit
Heater Leads Insulator, 11 inches long
COe Insulator
Band Heater
Thermocouple
Elbow
Eductor
CJC Sensor (RTD assembly, ring type)
Insulator
Heater Rod
Watlube Heater Release Agent
8-3
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
Figure 8-2. Sensor Housing
Components
1
3
2
7
4
5
6
8
9
10
17
14
16
16
15
16
12
8-4
11, 11A
37390074
13
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
Index No.
Part Number
Description
1
2
3
1A99089H02
5R10246H02
1A98448H02
Cover Gasket (O-ring)
Blowback Filter, 20 Micron, Stainless Steel
O-ring
4
5R10247H01
Adaptor
5
5R10183H02
Tube, Sample, 18 in. (457 mm) 316 Stainless steel
5R10183H06
Tube, Sample, 18 in. (457 mm) Inconel 600
5R10227G01
Tube, Sample, 18 in. (457 mm) Ceramic
5R10183H03
Tube, Sample, 3 ft. (0.91 m) 316 Stainless steel
5R10183H07
Tube, Sample, 3 ft. (0.91 m) Inconel 600
5R10227G02
Tube, Sample, 3 ft. (0.91 m) Ceramic
5R10183H04
Tube, Sample, 6 ft. (1.83 m) 316 Stainless steel
5R10183H08
Tube, Sample, 6 ft. (1.83 m) Inconel 600
5R10183H05
Tube, Sample, 9 ft. (2.7 m) 316 Stainless steel
5R10183H09
Tube, Sample, 9 ft. (2.7 m) Inconel 600
6
5R10183H01
Tube, Exhaust
7
1A68017H02
COe Extractive
8
5R10185H08
Dilution Air Tube
9
5R10185H03
Eductor Drive Air Tube
10
5R10185H04
Reference Air Tube
11
6A00146G01
Heater Cable Assembly, Remote Electronics, 20 ft (6 m)
11A
12
12
13
14
6A00146G02
Heater Cable Assembly, Remote Electronics, 40 ft (12 m)
6A00146G03
Heater Cable Assembly, Remote Electronics, 60 ft (18 m)
6A00146G04
Heater Cable Assembly, Remote Electronics, 80 ft (24 m)
6A00146G05
Heater Cable Assembly, Remote Electronics, 100 ft (30 m)
6A00146G06
Heater Cable Assembly, Remote Electronics, 150 ft (45 m)
6A00147G01
Signal Cable Assembly, Remote Electronics, 20 ft (6 m)
6A00147G02
Signal Cable Assembly, Remote Electronics, 40 ft (12 m)
6A00147G03
Signal Cable Assembly, Remote Electronics, 60 ft (18 m)
6A00147G04
Signal Cable Assembly, Remote Electronics, 80 ft (24 m)
6A00147G05
Signal Cable Assembly, Remote Electronics, 100 ft (30 m)
6A00147G06
Signal Cable Assembly, Remote Electronics, 150 ft (45 m)
1A99762H02
In Situ Filter, 10 Micron (Stainless Steel Sample Tube only)
1A99762H03
6P00162H01
Hasteloy In Situ Filter, 10 Micron, High Temperature (Stainless Steel and Inconel Sample
Tubes only)
Flange Insulator (optional)
3535B18H02
Flange Gasket, ANSI
3535B45H01
Flange Gasket, DIN
15
5R10279G01
Tube Fitting, Type "R"
16
771B870H04
Tube Fitting, Standard
17
5R10279G02
Tube Fitting, Type "E"
8-5
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
ELECTRONICS
HOUSING
Figure 8-3. Electronics Housing
Components
1
6
5
4
2
3
8
37390080
7
8-6
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
Index No.
Part Number
1
2
3
4
5
6
1A97902H01
1A97905H02
1A97905H01
6A00132G01
Special Order
1A97913H06
1A99766H01
1A99766H02
1A99089H02
5R10219G01
5R10199G01
7
8
8A
OCX 8800
Description
Hose
Solenoid Valve, 3-Way
Solenoid Valve, Test Gas
Electronics Stack, HART
Flash PROM, Programmed
Fuse (F1 and F6), 10 Amp, 250 VAC
Fuse (F3), 4 Amp, 250 VAC
Fuse (F4), 8 Amp, 250 VAC
Cover Gasket (O-ring)
Cover, Blind
Cover, Window
8-7
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
Figure 8-4. EMI Filter
and Terminal Block
Blue
Brown
Blue
Green
Brown
Green
Green
1
37390083
2
3
8-8
Index No.
Part Number
1
2
3
1A98467H01
5R10238G01
1A99714H01
Description
Filter, EMI
Ground Wire
Terminal Block
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
O2 CELL AND HEATER
STRUT ASSEMBLY
Figure 8-5. O2 Cell and
Heater Strut Assembly
1
2
3
4
37390055
7
5
6
Index No.
1
2
3
4
5
6
7
Part Number
4851B44G01
5R10211G02
5R10188G01
Ref
Ref
Ref
4851B45G01
Description
Contact and Thermocouple Assembly
Heater Strut Assembly
Heater Tube
Gasket (part of O2 Cell Replacement Kit, item 7)
O2 Cell (part of O2 Cell Replacement Kit, item 7)
Screw (part of O2 Cell Replacement Kit, item 7)
O2 Cell Replacement Kit
8-9
Instruction Manual
OCX 8800
8-10
IM-106-880ASB, Rev 1.2
July 2008
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
Appendix A
OCX 8800
Safety Data
Safety Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page A-2
Safety Data Sheet for Ceramic Fiber Products . . . . . . . . page A-24
High Pressure Gas Cylinders . . . . . . . . . . . . . . . . . . . . . . page A-30
http://www..raihome.com
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
SAFETY INSTRUCTIONS
IMPORTANT
SAFETY INSTRUCTIONS FOR THE WIRING
AND INSTALLATION OF THIS APPARATUS
The following safety instructions apply specifically to all EU member
states. They should be strictly adhered to in order to assure compliance
with the Low Voltage Directive. Non-EU states should also comply with
the following unless superseded by local or National Standards.
1. Adequate earth connections should be made to all earthing points,
internal and external, where provided.
2. After installation or troubleshooting, all safety covers and safety grounds
must be replaced. The integrity of all earth terminals must be maintained
at all times.
3. Mains supply cords should comply with the requirements of IEC227 or
IEC245.
4. All wiring shall be suitable for use in an ambient temperature of greater
than 75°C.
5. All cable glands used should be of such internal dimensions as to
provide adequate cable anchorage.
6. To ensure safe operation of this equipment, connection to the mains
supply should only be made through a circuit breaker which will
disconnect all circuits carrying conductors during a fault situation. The
circuit breaker may also include a mechanically operated isolating
switch. If not, then another means of disconnecting the equipment from
the supply must be provided and clearly marked as such. Circuit
breakers or switches must comply with a recognized standard such as
IEC947. All wiring must conform with any local standards.
7. Where equipment or covers are marked with the symbol
to the right, hazardous voltages are likely to be present
beneath. These covers should only be removed when
power is removed from the equipment - and then only
by trained service personnel.
8. Where equipment or covers are marked with the symbol
to the right, there is a danger from hot surfaces beneath.
These covers should only be removed by trained
service personnel when power is removed from the
equipment. Certain surfaces may remain hot to the
touch.
9. Where equipment or covers are marked with the symbol
to the right, refer to the Operator Manual for instructions.
10. All graphical symbols used in this product are from one
or more of the following standards: EN61010-1, IEC417,
and ISO3864.
11. Where equipment or labels are marked "Do Not Open While Energized"
or similar, there is a danger of ignition in areas where an explosive
atmosphere is present. This equipment should only be opened when
power is removed and adequate time as specified on the label or in the
instruction manual has been allowed for the equipment to cool down and then only by trained service personnel.
A-2
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
DŮLEŽITÉ
Bezpečnostní pokyny pro zapojení a instalaci zařízení
Následující bezpečnostní pokyny se speciálně vztahují na všechny
členské státy EU. Pokyny by měly být přísně dodržovány, aby se
zajistilo splnění Směrnice o nízkém napětí. Pokud nejsou pokyny
nahrazeny místními či národními normami, měly by je dodržovat i
nečlenské státy EU.
1. U všech zemnicích bodů, interních a externích, by mělo být vytvořeno
odpovídající uzemnění.
2. Po instalaci nebo odstranění problémů musí být vyměněny všechny
bezpečnostní kryty a uzemnění. Vždy musí být zajištěna integrita všech
zemnicích svorek.
3. Sí˙ové kabely by měly odpovídat požadavkům normy IEC227 nebo
IEC245.
4. Všechna zapojení by měla být vhodná pro použití při vnějších teplotách
nad 75 °C.
5. Všechna použitá kabelová hrdla by měla mít takové vnitřní rozměry, aby
zajistila odpovídající zakotvení kabelu.
6. Správnou činnost zařízení zajistíte, vytvoříte-li připojení k napájecímu
zdroji pouze přes jistič, který v případě poruchy odpojí všechny obvody
s konduktory. Jistič může také obsahovat mechanický odpojovač.
Pokud ho neobsahuje, musí být zajištěn a jasně označen jiný způsob
odpojení zařízení od zdroje. Jističe nebo přepínače musí odpovídat
uznávaným normám, např. IEC947. Všechna zapojení musí odpovídat
místním normám.
7. Je-li zařízení nebo kryt označen symbolem na pravé
straně, pravděpodobně se uvnitř nachází nebezpečné
napětí. Tyto kryty by měly být sejmuty pouze po
odpojení zařízení od zdroje - a to pouze kvalifikovaným
zaměstnancem.
8. Je-li zařízení nebo kryt označen symbolem na pravé
straně, povrch zařízení může být velmi horký. Tyto kryty
by měly být sejmuty pouze kvalifikovaným
zaměstnancem po odpojení zařízení od zdroje. Některé
povrchy mohou být stále horké.
9. Je-li zařízení nebo kryt označen symbolem na pravé
straně, přečtěte si nejprve instrukce v návodu k obsluze.
10. Všechny grafické symboly používané u výrobku
pocházejí z následujících norem: EN61010-1, IEC417 a
ISO3864.
11. Pokud je zařízení nebo štítky označeno varováním „Je-li zařízení pod
napětím, neotvírejte jej“ či podobným, může dojít ve výbušném prostředí
ke vznícení. Zařízení lze otevřít pouze po jeho odpojení od zdroje a
ponechání dostatečného času na vychladnutí, jak je uvedeno na štítku
nebo v návodu k obsluze - a to pouze kvalifikovaným zaměstnancem.
A-3
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
VIGTIGT
Sikkerhedsinstruktion for tilslutning og installering af dette udstyr.
Følgende sikkerhedsinstruktioner gælder specifikt i alle
EU-medlemslande. Instruktionerne skal nøje følges for overholdelse af
Lavsspændingsdirektivet og bør også følges i ikke EU-lande medmindre
andet er specificeret af lokale eller nationale standarder.
1. Passende jordforbindelser skal tilsluttes alle jordklemmer, interne og
eksterne, hvor disse forefindes.
2. Efter installation eller fejlfinding skal alle sikkerhedsdæksler og
jordforbindelser reetableres.
3. Forsyningskabler skal opfylde krav specificeret i IEC227 eller IEC245.
4. Alle ledningstilslutninger skal være konstrueret til omgivelsestemperatur
højere end 75°C.
5. Alle benyttede kabelforskruninger skal have en intern dimension, så
passende kabelaflastning kan etableres.
6. For opnåelse af sikker drift og betjening skal der skabes beskyttelse
mod indirekte berøring gennem afbryder (min. 10A), som vil afbryde alle
kredsløb med elektriske ledere i fejlsitua-tion. Afbryderen skal indholde
en mekanisk betjent kontakt. Hvis ikke skal anden form for afbryder
mellem forsyning og udstyr benyttes og mærkes som sådan. Afbrydere
eller kontakter skal overholde en kendt standard som IEC947.
7. Hvor udstyr eller dæksler er mærket med dette symbol,
er farlige spændinger normalt forekom-mende bagved.
Disse dæksler bør kun afmonteres, når
forsyningsspændingen er frakoblet - og da kun af
instrueret servicepersonale.
8. Hvor udstyr eller dæksler er mærket med dette symbol,
forefindes meget varme overflader bagved. Disse
dæksler bør kun afmonteres af instrueret
servicepersonale, når forsyningsspænding er frakoblet.
Visse overflader vil stadig være for varme at berøre i op
til 45 minutter efter frakobling.
9. Hvor udstyr eller dæksler er mærket med dette symbol,
se da i betjeningsmanual for instruktion.
10. Alle benyttede grafiske symboler i dette udstyr findes i
én eller flere af følgende standarder:- EN61010-1,
IEC417 & ISO3864.
11. Når udstyr eller etiketter er mærket "Må ikke åbnes, mens udstyret
tilføres strøm” eller lignende, er der fare for antændelse i områder, hvor
der er en eksplosiv atmosfære. Dette udstyr må kun åbnes, når
strømkilden er fjernet, og der er gået tilstrækkelig tid til, at udstyret er
kølet ned. Den nødvendige tid hertil er angivet på etiketten eller i
brugervejledningen. Udstyret må kun åbnes af en faglært person.
A-4
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
BELANGRIJK
Veiligheidsvoorschriften voor de aansluiting en installatie van dit toestel.
De hierna volgende veiligheidsvoorschriften zijn vooral bedoeld voor de
EU lidstaten. Hier moet aan gehouden worden om de onderworpenheid
aan de Laag Spannings Richtlijn (Low Voltage Directive) te verzekeren.
Niet EU staten zouden deze richtlijnen moeten volgen tenzij zij reeds
achterhaald zouden zijn door plaatselijke of nationale voorschriften.
1. Degelijke aardingsaansluitingen moeten gemaakt worden naar alle
voorziene aardpunten, intern en extern.
2. Na installatie of controle moeten alle veiligheidsdeksels en -aardingen
terug geplaatst worden. Ten alle tijde moet de betrouwbaarheid van de
aarding behouden blijven.
3. Voedingskabels moeten onderworpen zijn aan de IEC227 of de IEC245
voorschriften.
4. Alle bekabeling moet geschikt zijn voor het gebruik in
omgevingstemperaturen, hoger dan 75°C.
5. Alle wartels moeten zo gedimensioneerd zijn dat een degelijke kabel
bevestiging verzekerd is.
6. Om de veilige werking van dit toestel te verzekeren, moet de voeding
door een stroomonderbreker gevoerd worden (min 10A) welke alle
draden van de voeding moet onderbreken. De stroomonderbreker mag
een mechanische schakelaar bevatten. Zoniet moet een andere
mogelijkheid bestaan om de voedingsspanning van het toestel te halen
en ook duidelijk zo zijn aangegeven. Stroomonderbrekers of
schakelaars moeten onderworpen zijn aan een erkende standaard
zoals IEC947.
7. Waar toestellen of deksels aangegeven staan met het
symbool is er meestal hoogspanning aanwezig. Deze
deksels mogen enkel verwijderd worden nadat de
voedingsspanning werd afgelegd en enkel door getraind
onderhoudspersoneel.
8. Waar toestellen of deksels aangegeven staan met het
symbool is er gevaar voor hete oppervlakken. Deze
deksels mogen enkel verwijderd worden door getraind
onderhoudspersoneel nadat de voedingsspanning
verwijderd werd. Sommige oppper-vlakken kunnen 45
minuten later nog steeds heet aanvoelen.
9. Waar toestellen of deksels aangegeven staan met het
symbool gelieve het handboek te raadplegen.
10. Alle grafische symbolen gebruikt in dit produkt, zijn
afkomstig uit een of meer van devolgende standaards:
EN61010-1, IEC417 en ISO3864.
11. Op plaatsen waar uitrusting of etiketten zijn voorzien van een melding
als "Niet openen bij aanwezigheid van spanning" bestaat er
brandgevaar in omgevingen waar een explosieve atmosfeer aanwezig
is. Deze uitrusting mag uitsluitend worden geopend wanneer het niet
meer onder spanning staat en de uitrusting gedurende de
voorgeschreven tijd op het etiket of in de handleiding is afgekoeld - en
dan uitsluitend door voldoende opgeleid onderhoudspersoneel.
A-5
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
BELANGRIJK
Veiligheidsinstructies voor de bedrading en installatie van dit apparaat.
Voor alle EU lidstaten zijn de volgende veiligheidsinstructies van
toepassing. Om aan de geldende richtlijnen voor laagspanning te
voldoen dient men zich hieraan strikt te houden. Ook niet EU lidstaten
dienen zich aan het volgende te houden, tenzij de lokale wetgeving
anders voorschrijft.
1. Alle voorziene interne- en externe aardaansluitingen dienen op
adequate wijze aangesloten te worden.
2. Na installatie, onderhouds- of reparatie werkzaamheden dienen alle
beschermdeksels /kappen en aardingen om reden van veiligheid weer
aangebracht te worden.
3. Voedingskabels dienen te voldoen aan de vereisten van de normen
IEC 227 of IEC 245.
4. Alle bedrading dient geschikt te zijn voor gebruik bij een omgevings
temperatuur boven 75°C.
5. Alle gebruikte kabelwartels dienen dusdanige inwendige afmetingen te
hebben dat een adequate verankering van de kabel wordt verkregen.
6. Om een veilige werking van de apparatuur te waarborgen dient de
voeding uitsluitend plaats te vinden via een meerpolige automatische
zekering (min.10A) die alle spanningvoerende geleiders verbreekt
indien een foutconditie optreedt. Deze automatische zekering mag ook
voorzien zijn van een mechanisch bediende schakelaar. Bij het
ontbreken van deze voorziening dient een andere als zodanig duidelijk
aangegeven mogelijkheid aanwezig te zijn om de spanning van de
apparatuur af te schakelen. Zekeringen en schakelaars dienen te
voldoen aan een erkende standaard zoals IEC 947.
7. Waar de apparatuur of de beschermdeksels/kappen
gemarkeerd zijn met het volgende symbool, kunnen
zich hieronder spanning voerende delen bevinden die
gevaar op kunnen leveren. Deze beschermdeksels/
kappen mogen uitsluitend verwijderd worden door
getraind personeel als de spanning is afgeschakeld.
8. Waar de apparatuur of de beschermdeksels/kappen
gemarkeerd zijn met het volgende symbool, kunnen
zich hieronder hete oppervlakken of onderdelen
bevinden. Bepaalde delen kunnen mogelijk na 45 min.
nog te heet zijn om aan te raken.
9. Waar de apparatuur of de beschermdeksels/kappen
gemarkeerd zijn met het volgende symbool, dient men
de bedieningshandleiding te raadplegen.
10. Alle grafische symbolen gebruikt bij dit produkt zijn
volgens een of meer van de volgende standaarden:
EN 61010-1, IEC 417 & ISO 3864.
11. Op plaatsen waar uitrusting of etiketten zijn voorzien van een melding als
"Niet openen bij aanwezigheid van spanning" bestaat er brandgevaar in
omgevingen waar een explosieve atmosfeer aanwezig is. Deze uitrusting
mag uitsluitend worden geopend wanneer het niet meer onder spanning
staat en de uitrusting gedurende de voorgeschreven tijd op het etiket of in
de handleiding is afgekoeld - en dan uitsluitend door voldoende opgeleid
onderhoudspersoneel.
A-6
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
WICHTIG
Sicherheitshinweise für den Anschluß und die Installation dieser Geräte.
Die folgenden Sicherheitshinweise sind in allen Mitgliederstaaten der
europäischen Gemeinschaft gültig. Sie müssen strickt eingehalten
werden, um der Niederspannungsrichtlinie zu genügen.
Nichtmitgliedsstaaten der europäischen Gemeinschaft sollten die
national gültigen Normen und Richtlinien einhalten.
1. Alle intern und extern vorgesehenen Erdungen der Geräte müssen
ausgeführt werden.
2. Nach Installation, Reparatur oder sonstigen Eingriffen in das Gerät müssen
alle Sicherheitsabdeckungen und Erdungen wieder installiert werden. Die
Funktion aller Erdverbindungen darf zu keinem Zeitpunkt gestört sein.
3. Die Netzspannungsversorgung muß den Anforderungen der IEC227
oder IEC245 genügen.
4. Alle Verdrahtungen sollten mindestens bis 75°C ih re Funktion dauerhaft
erfüllen.
5. Alle Kabeldurchführungen und Kabelverschraubungen sollten in Ihrer
Dimensionierung so gewählt werden, daß diese eine sichere
Verkabelung des Gerätes ermöglichen.
6. Um eine sichere Funktion des Gerätes zu gewährleisten, muß die
Spannungsversorgung über mindestens 10 A abgesichert sein. Im
Fehlerfall muß dadurch gewährleistet sein, daß die
Spannungsversorgung zum Gerät bzw. zu den Geräten unterbrochen
wird. Ein mechanischer Schutzschalter kann in dieses System integriert
werden. Falls eine derartige Vorrichtung nicht vorhanden ist, muß eine
andere Möglichkeit zur Unterbrechung der Spannungszufuhr
gewährleistet werden mit Hinweisen deutlich gekennzeichnet werden.
Ein solcher Mechanismus zur Spannungsunterbrechung muß mit den
Normen und Richtlinien für die allgemeine Installation von
Elektrogeräten, wie zum Beispiel der IEC947, übereinstimmen.
7. Mit dem Symbol sind Geräte oder Abdeckungen gekennzeichnet, die eine
gefährliche (Netzspannung) Spannung führen. Die Abdeckungen dürfen
nur entfernt werden, wenn die Versorgungsspannung unterbrochen wurde.
Nur geschultes Personal darf an diesen Geräten Arbeiten ausführen.
8. Mit dem Symbol sind Geräte oder Abdeckungen gekennzeichnet, in
bzw. unter denen heiße Teile vorhanden sind. Die Abdeckungen dürfen
nur entfernt werden, wenn die Versorgungsspannung unterbrochen
wurde. Nur geschultes Personal darf an diesen Geräten Arbeiten
ausführen. Bis 45 Minuten nach dem Unterbrechen der Netzzufuhr
können derartig Teile noch über eine erhöhte Temperatur verfügen.
9. Mit dem Symbol sind Geräte oder Abdeckungen gekennzeichnet, bei
denen vor dem Eingriff die entsprechenden Kapitel im Handbuch
sorgfältig durchgelesen werden müssen.
10. Alle in diesem Gerät verwendeten graphischen Symbole entspringen
einem oder mehreren der nachfolgend aufgeführten Standards:
EN61010-1, IEC417 & ISO3864.
11. Wenn Geräte oder Etiketten mit dem Hinweis "Nicht unter Spannung
öffnen" oder ähnlichen Hinweisen versehen sind, besteht in
explosionsgefährdeten Umgebungen Entzündungsgefahr. Das Gerät
darf nur geöffnet werden, wenn es nicht ans Stromnetz angeschlossen
und entsprechend der Zeitangaben auf dem Etikett bzw. in der
Betriebsanleitung ausreichend abgekühlt ist. Das Gerät darf nur von
geschultem Service-Personal geöffnet werden.
A-7
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
ΣΗΜΑΝΤΙΚΟ
Οδηγιεσ ασφαλειασ για την καλωδιωση και εγκατασταση τησ συσκευησ
Οι ακόλουθες οδηγίες ασφαλείας εφαρµόζονται ειδικά για όλες τις χώρες
µέλη της Ευρωπαϊκής Κοινότητας. Θα πρέπει να ακολουθούνται αυστηρά
ώστε να εξασφαλιστεί η συµβατότητα µε τις οδηγίες για τη Χαµηλή Τάση.
Χώρες που δεν είναι µέλη της Ευρωπαϊκής Κοινότητας θα πρέπει επίσης
να ακολουθούν τις οδηγίες, εκτός εάν αυτές αντικαθίστανται από τα
Τοπικά ή Εθνικά πρότυπα.
1. Επαρκείς συνδέσεις γείωσης θα πρέπει να γίνονται σε όλα τα σηµεία
γείωσης, εσωτερικά και εξωτερικά, όπου υπάρχουν.
2. Μετά την εγκατάσταση ή την αντιµετώπιση σφαλµάτων, όλα τα καλύµµατα
ασφαλείας και οι γειώσεις ασφαλείας πρέπει να επανεγκαθίστανται. Η καλή
κατάσταση όλων των ακροδεκτών γείωσης πρέπει να συντηρείται διαρκώς.
3. Τα καλώδια τροφοδοσίας πρέπει να πληρούν τις απαιτήσεις των IEC227
ή IEC245.
4. Όλες οι καλωδιώσεις θα πρέπει να είναι κατάλληλες για χρήση σε
θερµοκρασία χώρου υψηλότερη από 75°C.
5. Όλοι οι στυπιοθλίπτες θα πρέπει να είναι τέτοιων εσωτερικών
διαστάσεων, ώστε να παρέχουν επαρκή στερέωση των καλωδίων.
6. Για τη διασφάλιση ασφαλούς λειτουργίας αυτής της συσκευής, η σύνδεση
τροφοδοσίας θα πρέπει να γίνεται µόνο µέσω ασφαλειοδιακόπτη, ο
οποίος θα αποσυνδέει όλους τους ηλεκτροφόρους αγωγούς των
κυκλωµάτων, στη διάρκεια κατάστασης σφάλµατος. Ο ασφαλειοδιακόπτης
µπορεί επίσης να περιλαµβάνει µηχανικό διακόπτη αποµόνωσης. Εάν δεν
περιλαµβάνει, τότε άλλα µέσα αποσύνδεσης της συσκευής από την
τροφοδοσία πρέπει να παροχηθούν και να σηµανθούν σαφώς ως τέτοια.
Οι ασφαλειοδιακόπτες ή διακόπτες πρέπει να συµµορφώνονται µε
αναγνωρισµένα πρότυπα όπως το IEC947. Όλες οι καλωδιώσεις πρέπει
να συµµορφώνονται µε τα τοπικά πρότυπα.
7. Όπου συσκευές ή καλύµµατα είναι σηµασµένα µε το
σύµβολο που εικονίζεται δεξιά, επικίνδυνες τάσεις
ενυπάρχουν κάτω από αυτά. Αυτά τα καλύµµατα θα
πρέπει να αφαιρούνται µόνο όταν έχει αφαιρεθεί η
τροφοδοσία από τη συσκευή - και στην περίπτωση αυτή,
µόνο από ειδικευµένο τεχνικό προσωπικό.
8. Όπου συσκευές ή καλύµµατα είναι σηµασµένα µε το
σύµβολο που εικονίζεται δεξιά, υπάρχει κίνδυνος από
καυτές επιφάνειες κάτω από αυτά. Τέτοια καλύµµατα θα
πρέπει να αφαιρούνται µόνο από ειδικευµένο τεχνικό
προσωπικό, όταν έχει αφαιρεθεί η τροφοδοσία από τη
συσκευή. Κάποιες επιφάνειες µπορούν να παραµένουν ζεστές στην αφή.
9. Όπου συσκευές ή καλύµµατα είναι σηµασµένα µε το
σύµβολο που εικονίζεται δεξιά, ανατρέξτε στις οδηγίες
χρήσης της συσκευής.
10. Όλα τα γραφικά σύµβολα που χρησιµοποιούνται σε αυτό
το προϊόν είναι από ένα ή περισσότερα από τα εξής
πρότυπα: EN61010-1, IEC417 και ISO3864.
11. Όπου συσκευή ή ετικέτα είναι σηµασµένη µε την ένδειξη "Μην ανοίγετε ενώ
βρίσκεται σε λειτουργία" ή άλλη παρόµοια, υπάρχει κίνδυνος ανάφλεξης σε
περιοχές µε εκρηκτική ατµόσφαιρα. Ο παρών εξοπλισµός πρέπει να
ανοίγεται µόνο όταν είναι εκτός ρεύµατος και αφού παρέλθει ο κατάλληλος
χρόνος που αναγράφεται στην ετικέτα ή στο εγχειρίδιο οδηγιών ώστε να
ψυχθεί και µόνο από εκπαιδευµένο προσωπικό συντήρησης.
A-8
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
OLULINE TEAVE
Juhtmestiku ja seadme paigaldamisega seotud ohutusjuhised
Alljärgnevad ohutusjuhised rakenduvad eriti kõigi Euroopa Liidu
liikmesriikide suhtes. Antud juhiseid tuleb täpselt järgida, et kindlustada
vastavus madalpinge direktiiviga. Euroopa Liitu mittekuuluvad riigid
peavad samuti alljärgnevaid juhiseid järgima, va juhul, kui on olemas
vastavad kohalikud riiklikud standardid.
1. Ettenähtud maanduspunktide, nii sisemiste kui väliste jaoks tuleb
tagada nõuetekohased maaühendused.
2. Pärast paigaldamist või rikketuvastust tuleb kõik turvaümbrised ja
turvamaandused uuesti oma kohale seada. Kõigis olukordades tuleb
säilitada kõigi maandusklemmide terviklikkus.
3. Toitejuhtmed peavad vastama IEC227 või IEC245 nõuetele.
4. Kogu juhtmestik peab sobima kasutamiseks üle 75°C õhutemperatuuri
juures.
5. Kõik juhtmetihendid peavad sisemõõtmete poolest tagama
nõuetekohased kaabliühendused.
6. Seadme ohutu töötamise tagamiseks peab ühendus toiteallikaga
toimuma vaid läbi automaatkorgi, mis veaolukorras lülitab välja kõik
voolukandjad. Automaatkorgil võib olla ka mehhaaniliselt reguleeritav
lahklüliti. Vastasel juhul peab seadme toiteallikast lahtiühendamiseks
olema teine ja selgelt osutatud moodus. Automaatkorgid või -lülitid
peavad vastama tunnustatud standarditele nagu nt IEC947. Kogu
juhtmestik peab vastama kohalikele standarditele.
7. Seadmel või ümbristel asuv paremale osutav sümbol
tähistab selle all leiduvat ohtlikku pinget. Selliste
sümbolitega ümbriseid võib eemaldada vaid juhul, kui
seade on toiteallikast lahti ühendatud ning ka siis ainult
vastavate oskustega spetsialisti poolt.
8. Seadmele või ümbristele märgitud paremale osutava
sümboli all valitseb kuumadest pindadest tulenev oht.
Nimetatud sümbolitega ümbriseid võib eemaldada vaid
vastavate oskustega spetsialist, kui seade on
toiteallikast lahti ühendatud. Teatud pinnad võivad
puudutamise jaoks liiga kuumad olla.
9. Seadmel või ümbristel leiduva paremale osutava
sümboli korral vt juhiste jaoks Toimimisjuhendit.
10. Kõik selle toote juures kasutatavad graafilised sümbolid
lähtuvad ühest või enamast järgmistest standarditest:
EN61010-1, IEC417 ja ISO3864.
11. Kui seadmele või siltidele on kirjutatud "Ärge avage voolutarbimine
korral" vms, valitseb plahvatusohtlikus keskkonnas süttimise oht.
Seadet võib avada ainult siis, kui toide on lahti ühendatud ning seadmel
on võimaldatud sildil või kasutusjuhendis osutatud aja jooksul maha
jahtuda -- ning ka sellisel juhul ainult vastavate oskustega spetsialisti
poolt.
A-9
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
TÄRKEÄÄ
Turvallisuusohje, jota on noudatettava tämän laitteen asentamisessa ja
kaapeloinnissa.
Seuraavat ohjeet pätevät erityisesti EU:n jäsenvaltioissa. Niitä täytyy
ehdottomasti noudattaa jotta täytettäisiin EU:n matalajännitedirektiivin
(Low Voltage Directive) yhteensopivuus. Myös EU:hun kuulumattomien
valtioiden tulee nou-dattaa tätä ohjetta, elleivät kansalliset standardit
estä sitä.
1. Riittävät maadoituskytkennät on tehtävä kaikkiin maadoituspisteisiin,
sisäisiin ja ulkoisiin.
2. Asennuksen ja vianetsinnän jälkeen on kaikki suojat ja suojamaat
asennettava takaisin pai-koilleen. Maadoitusliittimen kunnollinen
toiminta täytyy aina ylläpitää.
3. Jännitesyöttöjohtimien täytyy täyttää IEC227 ja IEC245 vaatimukset.
4. Kaikkien johdotuksien tulee toimia >75°C lämpötil oissa.
5. Kaikkien läpivientiholkkien sisähalkaisijan täytyy olla sellainen että
kaapeli lukkiutuu kun-nolla kiinni.
6. Turvallisen toiminnan varmistamiseksi täytyy jännitesyöttö varustaa
turvakytkimellä (min 10A), joka kytkee irti kaikki jännitesyöttöjohtimet
vikatilanteessa. Suojaan täytyy myös sisältyä mekaaninen erotuskytkin.
Jos ei, niin jännitesyöttö on pystyttävä katkaisemaan muilla keinoilla ja
merkittävä siten että se tunnistetaan sellaiseksi. Turvakytkimien tai
kat-kaisimien täytyy täyttää IEC947 standardin vaatimukset
näkyvyydestä.
7. Mikäli laite tai kosketussuoja on merkitty tällä merkillä
on merkinnän takana tai alla hengenvaarallisen
suuruinen jännite. Suojaa ei saa poistaa jänniteen
ollessa kytkettynä laitteeseen ja poistamisen saa
suorittaa vain alan asian-tuntija.
8. Mikäli laite tai kosketussuoja on merkitty tällä merkillä
on merkinnän takana tai alla kuuma pinta. Suojan saa
poistaa vain alan asiantuntija kun jännite-syöttö on
katkaistu. Tällainen pinta voi säilyä kosketuskuumana
jopa 45 mi-nuuttia.
9. Mikäli laite tai kosketussuoja on merkitty tällä merkillä
katso lisäohjeita käyt-töohjekirjasta.
10. Kaikki tässä tuotteessa käytetyt graafiset symbolit ovat
yhdestä tai useammasta seuraavis-ta standardeista:
EN61010-1, IEC417 & ISO3864.
11. Jos laitteessa tai tarrassa on merkintä "Älä avaa, kun virta on kytketty"
tai vastaava, räjähdysvaarallisissa tiloissa on syttymisen vaara. Nämä
laitteet voidaan avata vain silloin, kun virta ei ole kytkettynä ja laitteen
on annettu jäähtyä tarrassa tai oppaassa määritetyn ajan. Tällöinkin
laitteet saa avata vain koulutettu huoltohenkilökunta.
A-10
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
IMPORTANT
Consignes de sécurité concernant le raccordement et l'installation de
cet appareil.
Les consignes de sécurité ci-dessous s'adressent particulièrement à
tous les états membres de la communauté européenne. Elles doivent
être strictement appliquées afin de satisfaire aux directives concernant
la basse tension. Les états non membres de la communauté européenne
doivent également appliquer ces consignes sauf si elles sont en
contradiction avec les standards locaux ou nationaux.
1. Un raccordement adéquat à la terre doit être effectuée à chaque borne
de mise à la terre, interne et externe.
2. Après installation ou dépannage, tous les capots de protection et toutes
les prises de terre doivent être remis en place, toutes les prises de terre
doivent être respectées en permanence.
3. Les câbles d'alimentation électrique doivent être conformes aux normes
IEC227 ou IEC245.
4. Tous les raccordements doivent pouvoir supporter une température
ambiante supérieure à 75°C.
5. Tous les presse-étoupes utilisés doivent avoir un diamètre interne en
rapport avec les câbles afin d'assurer un serrage correct sur ces derniers.
6. Afin de garantir la sécurité du fonctionnement de cet appareil, le
raccordement à l'alimentation électrique doit être réalisé exclusivement
au travers d'un disjoncteur (minimum 10A.) isolant tous les conducteurs
en cas d'anomalie. Ce disjoncteur doit également pouvoir être actionné
manuellement, de façon mécanique. Dans le cas contraire, un autre
système doit être mis en place afin de pouvoir isoler l'appareil et doit
être signalisé comme tel. Disjoncteurs et interrupteurs doivent être
conformes à une norme reconnue telle IEC947.
7. Lorsque les équipements ou les capots affichent le
symbole suivant, cela signifie que des tensions
dangereuses sont présentes. Ces capots ne doivent
être démontés que lorsque l'alimentation est coupée, et
uniquement par un personnel compétent.
8. Lorsque les équipements ou les capots affichent le
symbole suivant, cela signifie que des surfaces
dangereusement chaudes sont présentes. Ces capots ne
doivent être démontés que lorsque l'alimentation est
coupée, et uniquement par un personnel compétent.
Certaines surfaces peuvent rester chaudes jusqu'à
45 mn.
9. Lorsque les équipements ou les capots affichent le
symbole suivant, se reporter au manuel d'instructions.
10. Tous les symboles graphiques utilisés dans ce produit
sont conformes à un ou plusieurs des standards
suivants: EN61010-1, IEC417 & ISO3864.
11. Les équipements comportant une étiquette avec la mention " Ne pas
ouvrir sous tension " ou toute autre mention similaire peuvent créer un
risque d'incendie dans les environnements explosifs. Ces équipements
ne doivent être ouverts que lorsqu'ils sont hors tension et que la durée de
refroidissement requise indiquée sur l'étiquette ou dans le manuel
d'instructions s'est écoulée. En outre ils ne doivent être ouverts que par
un personnel qualifié.
A-11
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
FONTOS
Biztonsági elıírások a készülék vezetékeléséhez és üzembeállításához
A következı biztonsági elıírások kifejezetten vonatkoznak az összes
EU-tagállamra. Ezeket szigorúan be kell tartani a Kisfeszültségő
irányelvnek való megfelelés biztosításához. A nem EU-tagállamok
szintén tartsák be a következıket, kivéve ha a helyi és nemzeti
szabványok azt másként nem írják elı.
1. A megfelelı földelést biztosítani kell az összes rendelkezésre álló
földelési ponton, legyen az belsı vagy külsı.
2. Az üzembeállítás vagy hibaelhárítás után az összes biztonsági
burkolatot és biztonsági földvezetéket ki kell cserélni. A földelıkapcsok
sértetlenségét mindig biztosítani kell.
3. A tápvezetékeknek eleget kell tenniük az IEC227 vagy IEC245
szabványokban megfogalmazott követelményeknek.
4. Az összes vezetéknek alkalmasnak kell lennie a 75 °C-nál magasabb
környezeti hımérséklet melletti használatra.
5. Az összes használt kábelvezetı tömszelencének olyan belsı
méretőnek kell lennie, hogy biztosítsák a kábelek megfelelı lekötését.
6. A berendezés biztonságos mőködésének biztosításához az elektromos
hálózathoz való csatlakozást csak megszakítón keresztül szabad
megvalósítani, amely az összes áramot szállító vezetéket bontja
hibahelyzet esetén. A megszakító magában foglalhat egy mechanikusan
mőködtethetı áramtalanító kapcsolót is. Ellenkezı esetben biztosítani kell
a berendezés elektromos hálózatról történı lekapcsolásának más módját,
és ezt világosan jelezni kell. A megszakítóknak vagy kapcsolóknak meg
kell felelniük egy elismert szabványnak, például az IEC947 szabványnak.
Az összes vezetéknek meg kell felelnie az összes helyi szabványnak.
7. Ha a berendezés vagy a burkolata a jobb oldalon
látható szimbólummal jelzett, alatta valószínőleg
veszélyes feszültség van jelen. Az ilyen burkolat csak a
berendezés áramtalanítása után távolítható el - és csak
képzett szervizszakember végezheti el.
8. Ha a berendezés vagy a burkolata a jobb oldalon
látható szimbólummal jelzett, fenn áll a veszélye, hogy
alatta forró felületek találhatóak. Az ilyen burkolatot
csak képzett szervizszakember távolíthatja el a
berendezés áramtalanítása után. Bizonyos felületek
érintésre forróak maradhatnak.
9. Ha a berendezés vagy a burkolata a jobb oldalon
látható szimbólummal jelzett, tekintse meg az
Üzemeltetési útmutató arra vonatkozó utasításait.
10. A terméken használt grafikus szimbólumok a következı
szabványok legalább egyikébıl származnak:
EN61010-1, IEC417 és ISO3864.
11. Ha a berendezésen vagy a címkéken a „Ne nyissa ki bekapcsolt
állapotban” vagy hasonló felhívás szerepel, robbanásveszélyes
környezetben fennáll a gyulladás veszélye. Ez a berendezés csak
áramtalanítás után nyitható ki, a címkén vagy a kezelési útmutatóban
szereplı, a berendezés lehőlését biztosító megfelelı idıi ráhagyás után és csak képzett szervizszakember végezheti el.
A-12
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
IMPORTANTE
Norme di sicurezza per il cablaggio e l'installazione dello strumento.
Le seguenti norme di sicurezza si applicano specificatamente agli stati
membri dell'Unione Europea, la cui stretta osservanza è richiesta per
garantire conformità alla Direttiva del Basso Voltaggio. Esse si
applicano anche agli stati non appartenenti all'Unione Europea, salvo
quanto disposto dalle vigenti normative locali o nazionali.
1. Collegamenti di terra idonei devono essere eseguiti per tutti i punti di
messa a terra interni ed esterni, dove previsti.
2. Dopo l'installazione o la localizzazione dei guasti, assicurarsi che tutti i
coperchi di protezione siano stati collocati e le messa a terra siano
collegate. L'integrità di ciscun morsetto di terra deve essere
costantemente garantita.
3. I cavi di alimentazione della rete devono essere secondo disposizioni
IEC227 o IEC245.
4. L'intero impianto elettrico deve essere adatto per uso in ambiente con
temperature superiore a 75°C.
5. Le dimensioni di tutti i connettori dei cavi utilizzati devono essere tali da
consentire un adeguato ancoraggio al cavo.
6. Per garantire un sicuro funzionamento dello strumento il collegamento alla
rete di alimentazione principale dovrà essere eseguita tramite interruttore
automatico (min.10A), in grado di disattivare tutti i conduttori di circuito in
caso di guasto. Tale interruttore dovrà inoltre prevedere un sezionatore
manuale o altro dispositivo di interruzione dell'alimentazione, chiaramente
identificabile. Gli interruttori dovranno essere conformi agli standard
riconosciuti, quali IEC947.
7. Il simbolo riportato sullo strumento o sui coperchi di
protezione indica probabile presenza di elevati voltaggi.
Tali coperchi di protezione devono essere rimossi
esclusivamente da personale qualificato, dopo aver
tolto alimentazione allo strumento.
8. Il simbolo riportato sullo strumento o sui coperchi di
protezione indica rischio di contatto con superfici ad alta
temperatura. Tali coperchi di protezione devono essere
rimossi esclusivamente da personale qualificato, dopo
aver tolto alimentazione allo strumento. Alcune superfici
possono mantenere temperature elevate per oltre 45 minuti.
9. Se lo strumento o il coperchio di protezione riportano il
simbolo, fare riferimento alle istruzioni del manuale
Operatore.
10. Tutti i simboli grafici utilizzati in questo prodotto sono
previsti da uno o più dei seguenti standard: EN61010-1,
IEC417 e ISO3864.
11. L'indicazione "Non aprire sotto tensione" o simili sull'apparecchiatura o
sulle etichette segnala il pericolo di accensione nelle aree in cui è
presente un'atmosfera esplosiva. L'apparecchiatura può essere aperta
solo quando l'alimentazione è scollegata ed è trascorso il tempo
indicato sull'etichetta o nel manuale delle istruzioni per consentirne il
raffreddamento. L'operazione può essere effettuata esclusivamente da
personale dell'assistenza qualificato.
A-13
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
SVARBU
šio prietaiso laidų prijungimo ir instaliacijos saugos instrukcijos
Toliau išvardinti saugumo reikalavimai taikomi konkrečiai visoms ES
šalims nar÷ms. Jų turi būti griežtai paisoma, kad būtų užtikrintai
laikomasi Žemos įtampos direktyvos. Ne ES nar÷s taip pat turi laikytis
toliau pateikiamų reikalavimų nebent juos pakeičia vietiniai ar
Nacionaliniai standartai.
1. Turi būti atliktas tinkamas įžeminimas visuose įžeminimo taškuose,
vidiniuose ir išoriniuose, kur numatyta.
2. Visos apsaugin÷s dangos ir įžemikliai po instaliacijos ar remonto turi
būti pakeisti. Visų įžeminimo terminalų vientisumo priežiūra turi būti
atliekama nuolat.
3. Matinimo tinklo laidai turi atitikti IEC227 ar IEC245 reikalavimus.
4. Visi laidai turi būti tinkami naudojimui aplinkos temperatūtoje,
aukštesn÷je nei 75°C.
5. Visi naudojamų kabelių riebokšliai turi būti tokių vidinių matmenų, kad
būtų galimas tinkamas kabelio pritvirtinimas.
6. Saugaus šio prietaiso veikimo užtikrinimui, prijungimas prie maitinimo
tinklo turi būti atliekamas tik per automatinį pertraukiklį, kuris atjungs
visas grandines nešančius konduktorius linijos gedimo metu.
Automatinis pertraukiklis taip pat gali tur÷ti mechaniškai veikiantį
įzoliavimo jungiklį. Jeigu ne, tuomet turi būti nurodytos kitos įrenginio
atjungimo priemon÷s, ir aiškai pažym÷tos, kad jos tokios yra.
Automatiniai perjungikliai ar jungikliai turi atitikti pripažintus standartus,
tokius kaip IEC947. Visi laidai turi atitikti visus vietinius standartus.
7. Kur įrenginys ar dangos yra pažym÷ti simboliu dešin÷je,
žemiau turi būti pavojinga įtampa. Šios dangos turi būti
nuimamos tik tada, kai srov÷ yra pašalinta iš įrenginio ir tik tuomet tai turi atlikti apmokytas personalas.
8. Ten kur įrenginys ar dangos yra pažym÷ti simboliu
dešin÷je, ten yra pavojus nuo karštų paviršių apačioje.
Šios dangos gali būti nuimamos tik apmokyto
personalo, kai srov÷ yra pašalinta iš įrenginio. Tam tikri
paviršiai gali išlikti karšti liečiant.
9. Ten kur įrenginys ar dangos yra pažym÷ti simboliu
d÷šin÷je, žr. nurodymus Valdymo instrukcijose.
10. Visi grafiniai simboliai naudojami šiam produktui yra iš
vieno ar daugiau toliau išvardintų standartų:
EN61010-1, IEC417, ir ISO3864.
11. Ten, kur įrenginys ar etiket÷s yra pažym÷ti "Neatidaryti esant srov÷s
tiekimui" ar panašiai, yra užsidegimo pavojus tose vietose, kur yra
sprogstamoji atmosfera. Šis įrenginys gali būti atidarytas tuomet, kai yra
pašalinta srov÷, ir pra÷jęs atitinkamas laikas, nurodytas etiket÷je ar
valdymo instrukcijoje, pakankamas įrenginio ataušimui - ir tai tik
apmokyto personalo.
A-14
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
SVARĪGI
Droš ības norādījumi š īs iekārtas pievienoš anai un uzstādīš anai
Turpmākie drošības norādījumi attiecas uz visām ES dalībvalstīm. Tie ir
stingri jāievēro, lai nodrošinātu atbilstību Zemsprieguma direktīvai.
Turpmāk norādītais jāievēro arī valstīs, kas nav ES dalībvalstis, ja vien
šos norādījumus neaizstāj vietējie vai valsts standarti.
1. Visi pieejamie iekšējie un ārējie zemējuma punkti ir atbilstoši jāiezemē.
2. Pēc uzstādīšanas vai problēmu risināšanas visi drošības pārsegi un
drošības zemējuma savienojumi ir jāpievieno atpakaĜ. Visiem zemējuma
savienojumiem vienmēr jābūt iezemētiem.
3. Elektropadeves vadiem jāatbilst IEC227 vai IEC245 prasībām.
4. Visai elektroinstalācijai jābūt piemērotai lietošanai apkārtējā
temperatūrā, kas pārsniedz 75°C.
5. Visu izmantoto kabeĜu blīvju iekšējiem izmēriem jābūt tādiem, lai
atbilstoši nostiprinātu kabeli.
6. Lai nodrošinātu šīs iekārtas drošu darbību, savienojums ar
elektropadeves tīklu jāizveido, izmantojot slēdzi, kas kĜūmes gadījumā
atvienos visas ėēdes, kurās ir vadītāji. Slēdzī var būt iestrādāts arī
mehānisks pārtraucējslēdzis. Ja tāda nav, tad ir jāuzstāda cita veida
ierīce iekārtas atvienošanai no strāvas padeves un tā atbilstoši un skaidri
jāmarėē. Slēdžiem jāatbilst kādam vispāratzītam standartam, piemēram,
IEC947. Visai elektroinstalācijai jāatbilst vietējiem standartiem.
7. Vietās, kur iekārta vai tās pārsegi ir marėēti ar labajā
pusē norādīto simbolu, visticamāk, zem tiem ir bīstams
spriegums. Šos pārsegus drīkst noĦemt tikai tad, ja
iekārta ir atvienota no strāvas padeves, – un šos darbus
drīkst veikt tikai atbilstoši apmācīti remontdarbu
darbinieki.
8. Vietās, kur iekārta vai tās pārsegi ir marėēti ar labajā
pusē norādīto simbolu, apdraudējumu izraisa zem tiem
esošās karstās virsmas. Šos pārsegus drīkst noĦemt tikai
atbilstoši apmācīti remontdarbu darbinieki, kad iekārta ir
atvienota no strāvas padeves. Iespējams, dažas virsmas
arī pēc iekārtas atvienošanas paliks karstas.
9. Ja iekārta vai pārsegi ir marėēti ar labajā pusē esošo
simbolu, skatiet operatora rokasgrāmatā ietvertos
norādījumus.
10. Visi šajā izstrādājumā izmantotie grafiskie simboli atbilst
vienam vai vairākiem no šiem standartiem: EN61010-1,
IEC417 un ISO3864.
11. Ja iekārtai vai uzlīmēm ir marėējums "Neatvērt, kamēr pieslēgta strāvai"
vai tamlīdzīga norāde, tas nozīmē, ka sprādzienbīstamā vidē ir
uzliesmošanas bīstamība. Šo iekārtu drīkst atvērt tikai tad, ja ir atvienota
strāva un ir nogaidīts iekārtas atdzišanai nepieciešamais laiks, kas
norādīts uzlīmē vai ekspluatācijas rokasgrāmatā, – un šos darbus drīkst
veikt tikai atbilstoši apmācīti remontdarbu darbinieki.
A-15
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
IMPORTANTI
STRUZZJONIJIET TAS-SIGURTÀ GĦALL-WIRING U
L-INSTALLAZZJONI TAT-TAGĦMIR
L-istruzzjonijiet tas-sigurtà japplikaw speċifikament għall-Istati Membri ta'
l-UE. Dawn għandhom jiġu osservati b'mod strett biex tkun żgurata lkonformità mad-Direttiva dwar il-Vultaġġ Baxx. Stati li mhumiex membri ta'
l-UE għandhom ukoll ikunu konformi ma' dan li ġej ħlief jekk dawn ikunu
sostituti mill-Istandards lokali jew Nazzjonali.
A-16
1.
Konnessjonijiet adegwati ta' l-ert għandhom isiru għall-punti kollha ta'
l-ert, interni u esterni, fejn ikun ipprovdut.
2.
Wara l-installazzjoni jew meta tipprova ssolvi xi problema, l-għatjien
kollha tas-sigurtà u l-erts tas-sigurtà għandhom jitpoġġew lura f'posthom.
L-integrità tat-terminali kollha ta' l-ert għandha tinżamm f'kull ħin.
3.
Il-wajers tal-provvista tad-dawl għandhom ikunu konformi ml-ħtiġijiet ta'
IEC227 jew IEC245.
4.
Il-wiring kollu għandu jkun adattat għall-użu f'temperatura ta' l-ambjent ta'
iktar minn 75°C.
5.
Il-glands tal-kejbils kollha li jintużw iridu jkunu ta' daqs intern tali li
jipprovdu ankoraġġ adegwat lill-kejbil.
6.
Biex tiżgura t-tħaddim sigur ta' dan it-tagħmir, il-konnessjoni mal-provvista
tad-dawl għandha ssir biss permezz ta' circuit breaker li jiskonnetta
l-kondutturi kollha li jkunu jġorru ċ-ċirkuwiti f'sitwazzjoni meta jkun hemm
il-ħsara. Is-circuit breaker jista wkoll jinkludi swiċċ li jiżola li jaħdem b'mod
mekkaniku. Jekk dan ma jkunx il-każ, mezz ieħor ta' kif it-tagħmir jiġi
skonnettjat minn mal-provvista tad-dawl għandu jkun ipprovdut, u jkun
immrkat b'mod ċar li hu hekk. Is-circuit breakers jew swiċċijiet iridu jkunu
konformi ma' standard rikonoxxut bħal IEC947. Il-wiring kollu jrid ikun
konformi ma' l-istandards lokali, jekk ikun hemm.
7.
Meta t-tagħmir jew l-għatjien ikunu mmarkati bis-simbolu fuq il-lemin,
x'aktarx li jkun hemm vultaġġi perikolużi taħthom. Dawn l-għatjien
għandhom jitneħħew biss meta titneħħa l-provvista tad-dawl
mit-tagħmir - u minn ħaddiema tal-manutenzjoni mħarrġa biss.
8.
Meta t-tagħmir jew l-għatjien ikunu mmarkati bis-simbolu fuq il-lemin,
ikun hemm periklu mill-uċuħ jaħarqu li jkun hemm taħthom. Dawn
l-għatjien għandhom jitneħħew biss minn ħaddiema tal-manutenzjoni
mħarrġa meta titneħħa l-provvista tad-dawl mit-tagħmir. Ċerti wċuħ
jistgħu jibqgħu jaħarqu meta tmisshom.
9.
Meta t-tagħmir jew l-għatjien ikunu mmarkati bis-simbolu fuq il-lemin,
irreferi għall-Manwal ta' l-Operatur għall-istruzzjonijiet.
10.
Is-simboli grafiċi kollha użati f'dan il-prodott huma minn wieħed jew
iktar mill-istandards li ġejjin: EN61010-1, IEC417, u ISO3864.
11.
Fejn it-tagħmir u t-tikketti huma mmarkati bil-kliem “Tiftaħx Meta
Jkun Enerġizzat” jew kliem simili, hemm periklu ta' nar f'żoni fejn
atmosfera esplossiva hi preżenti. It-tagħmir għandu jinfetaħ biss meta
l-provvista tad-dawl tkun mitfija u jkun għadda ħin biżżejjed, kif speċifikat
fuq it-tikketta jew fil-manwal ta' l-istruzzjonijiet, biex it-tagħmir ikun kesaħ
– u t-tagħmir għandu jinfetaħ biss minn staff li jkun imħarreġ.
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
VIKTIG
Sikkerhetsinstruks for tilkobling og installasjon av dette utstyret.
Følgende sikkerhetsinstruksjoner gjelder spesifikt alle EU medlemsland
og land med i EØS-avtalen. Instruksjonene skal følges nøye slik at
installasjonen blir i henhold til lavspenningsdirektivet. Den bør også
følges i andre land, med mindre annet er spesifisert av lokale- eller
nasjonale standarder.
1. Passende jordforbindelser må tilkobles alle jordingspunkter, interne og
eksterne hvor disse forefinnes.
2. Etter installasjon eller feilsøking skal alle sikkerhetsdeksler og
jordforbindelser reetableres. Jordingsforbindelsene må alltid holdes i
god stand.
3. Kabler fra spenningsforsyning skal oppfylle kravene spesifisert i IEC227
eller IEC245.
4. Alle ledningsforbindelser skal være konstruert for en
omgivelsestemperatur høyere en 750°C.
5. Alle kabelforskruvninger som benyttes skal ha en indre dimensjon slik at
tilstrekkelig avlastning oppnåes.
6. For å oppnå sikker drift og betjening skal forbindelsen til
spenningsforsyningen bare skje gjennom en strømbryter (minimum 10A)
som vil bryte spenningsforsyningen til alle elektriske kretser ved en
feilsituasjon. Strømbryteren kan også inneholde en mekanisk operert
bryter for å isolere instrumentet fra spenningsforsyningen. Dersom det
ikke er en mekanisk operert bryter installert, må det være en annen måte
å isolere utstyret fra spenningsforsyningen, og denne måten må være
tydelig merket. Kretsbrytere eller kontakter skal oppfylle kravene i en
annerkjent standard av typen IEC947 eller tilsvarende.
7. Der hvor utstyr eller deksler er merket med symbol for
farlig spenning, er det sannsynlig at disse er tilstede bak
dekslet. Disse dekslene må bare fjærnes når
spenningsforsyning er frakoblet utstyret, og da bare av
trenet servicepersonell.
8. Der hvor utstyr eller deksler er merket med symbol for
meget varm overflate, er det sannsynlig at disse er
tilstede bak dekslet. Disse dekslene må bare fjærnes
når spenningsforsyning er frakoblet utstyret, og da bare
av trenet servicepersonell. Noen overflater kan være for
varme til å berøres i opp til 45 minutter etter
spenningsforsyning frakoblet.
9. Der hvor utstyret eller deksler er merket med symbol,
vennligst referer til instruksjonsmanualen for instrukser.
10. Alle grafiske symboler brukt i dette produktet er fra en
eller flere av følgende standarder: EN61010-1, IEC417
& ISO3864.
11. Når utstyr eller merkelapper bærer advarselen "Må ikke åpnes under
spenning" eller lignende, innbærer det fare for eksplosjon i områder
med en eksplosiv atmosfære. Utstyret skal bare åpnes når det ikke er
noen strømtilførsel, og etter at det har hatt tilstrekkelig tid til å kjøle ned,
som spesifisert på merkelappen eller i håndboken. Selv da skal utstyret
bare åpnes av erfarne serviceteknikere.
A-17
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
WAśNE!
Zalecenia dotyczące bezpieczeństwa w zakresie podłączania i instalacji
tego urządzenia
Następujące zalecenia dotyczą zwłaszcza stosowania urządzenia we
wszystkich krajach Unii Europejskiej. NaleŜy się ściśle do nich
stosować w celu zapewnienia zgodności z dyrektywą niskonapięciową.
W przypadku instalacji urządzenia w krajach nienaleŜących do Unii
Europejskiej naleŜy równieŜ przestrzegać poniŜszych zaleceń, chyba Ŝe
są one zastąpione lokalnymi lub ogólnokrajowymi standardami.
1. Urządzenie naleŜy podłączyć kablem uziemiającym do wszystkich
punktów uziemienia (wewnętrznych i zewnętrznych).
2. Po instalacji lub czynnościach serwisowych naleŜy zamknąć wszystkie
pokrywy zabezpieczające i ponownie podłączyć uziemienie. NaleŜy
pilnować, by nie doszło do przerwania uziemienia.
3. Przewody zasilające powinny być zgodne z wymaganiami normy
IEC227 lub IEC245.
4. Wszystkie przewody powinny być odpowiednie do uŜytku w środowisku
o temperaturze wyŜszej niŜ 75°C.
5. Wszystkie dławnice powinny mieć wymiary wewnętrzne zapewniające
pewne umocowanie przewodów.
6. W celu zapewnienia bezpiecznej pracy urządzenie naleŜy podłączyć do
sieci tylko za pośrednictwem wyłącznika automatycznego, który w razie
awarii odłączy wszystkie obwody, w których przepływa prąd. Wyłącznik
automatyczny moŜe być równieŜ wyposaŜony w mechaniczny odłącznik
napięcia. W przeciwnym razie naleŜy zapewnić i jasno oznaczyć inną
moŜliwość odłączenia urządzenia od zasilania. Wyłączniki automatyczne
oraz odłączniki powinny być zgodne z uznawanymi standardami, takimi
jak norma IEC947. Wszystkie przewody muszą być zgodne z lokalnymi
przepisami.
7. Pod pokrywami lub elementami urządzenia oznaczonymi
symbolem pokazanym na rysunku po prawej stronie
moŜe występować niebezpieczne napięcie elektryczne.
Te pokrywy mogą być zdejmowane tylko po odłączeniu
zasilania, wyłącznie przez odpowiednio przeszkolonych
pracowników serwisu.
8. Pod pokrywami lub elementami urządzenia
oznaczonymi symbolem pokazanym na rysunku po
prawej stronie znajdują się gorące powierzchnie. Te
pokrywy mogą być zdejmowane tylko po odłączeniu
zasilania, wyłącznie przez odpowiednio przeszkolonych
pracowników serwisu. Niektóre powierzchnie mogą
pozostać nagrzane przez pewien czas po odłączeniu zasilania.
9. W przypadku sprzętu oraz pokryw oznaczonych
symbolem pokazanym na rysunku po prawej stronie
naleŜy zapoznać się ze wskazówkami w Instrukcji
operatora i stosować się do nich.
10. Wszystkie symbole graficzne zastosowane do
oznaczenia produktu pochodzą z następujących norm: EN61010-1,
IEC417 lub ISO3864.
11. Oznaczenie „Nie otwierać, gdy urządzenie jest pod napięciem” lub
podobne oznaczenia informują o ryzyku zapłonu w miejscach, gdzie
występuje zagroŜenie wybuchem. Urządzenie naleŜy otwierać tylko po
odłączeniu zasilania i po upływie czasu na ostygnięcie urządzenia
oznaczonego na etykiecie lub w instrukcji obsługi. Urządzenie mogą
otwierać wyłącznie odpowiednio przeszkoleni pracownicy serwisu.
A-18
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
IMPORTANTE
Instruções de segurança para ligação e instalação deste aparelho.
As seguintes instruções de segurança aplicam-se especificamente a
todos os estados membros da UE. Devem ser observadas rigidamente
por forma a garantir o cumprimento da Directiva sobre Baixa Tensão.
Relativamente aos estados que não pertençam à UE, deverão cumprir
igualmente a referida directiva, exceptuando os casos em que a
legislação local a tiver substituído.
1. Devem ser feitas ligações de terra apropriadas a todos os pontos de
terra, internos ou externos.
2. Após a instalação ou eventual reparação, devem ser recolocadas todas
as tampas de segurança e terras de protecção. Deve manter-se sempre
a integridade de todos os terminais de terra.
3. Os cabos de alimentação eléctrica devem obedecer às exigências das
normas IEC227 ou IEC245.
4. Os cabos e fios utilizados nas ligações eléctricas devem ser adequados
para utilização a uma temperatura ambiente até 75ºC.
5. As dimensões internas dos bucins dos cabos devem ser adequadas a
uma boa fixação dos cabos.
6. Para assegurar um funcionamento seguro deste equipamento, a
ligação ao cabo de alimentação eléctrica deve ser feita através de um
disjuntor (min. 10A) que desligará todos os condutores de circuitos
durante uma avaria. O disjuntor poderá também conter um interruptor
de isolamento accionado manualmente. Caso contrário, deverá ser
instalado qualquer outro meio para desligar o equipamento da energia
eléctrica, devendo ser assinalado convenientemente. Os disjuntores ou
interruptores devem obedecer a uma norma reconhecida, tipo IEC947.
7. Sempre que o equipamento ou as tampas contiverem o
símbolo, é provável a existência de tensões perigosas.
Estas tampas só devem ser retiradas quando a energia
eléctrica tiver sido desligada e por Pessoal da
Assistência devidamente treinado.
8. Sempre que o equipamento ou as tampas contiverem o
símbolo, há perigo de existência de superfícies quentes.
Estas tampas só devem ser retiradas por Pessoal da
Assistência devidamente treinado e depois de a energia
eléctrica ter sido desligada. Algumas superfícies
permanecem quentes até 45 minutos depois.
9. Sempre que o equipamento ou as tampas contiverem o
símbolo, o Manual de Funcionamento deve ser
consultado para obtenção das necessárias instruções.
10. Todos os símbolos gráficos utilizados neste produto
baseiam-se em uma ou mais das seguintes normas:
EN61010-1, IEC417 e ISO3864.
11. Sempre que o equipamento ou as etiquetas apresentarem o aviso "Não
abrir quando ligado à corrente" ou semelhante, existe um risco de ignição
em atmosferas explosivas. Este equipamento só deve ser aberto depois
de desligado da corrente eléctrica e o tempo de arrefecimento adequado
especificado na etiqueta ou no manual de instruções ter decorrido.
O equipamento só pode ser aberto por técnicos qualificados.
A-19
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
DÔLEŽITÉ
Bezpečnostné pokyny pre zapojenie káblov a inš taláciu tohto prístroja
Nasledovné bezpečnostné pokyny sa vzt’ahujú konkrétne na všetky
členské štáty EÚ. Musia byt’ striktne dodržané, aby sa zaistila zhoda so
Smernicou o nízkom napätí. Štáty, ktoré nie sú členskými štátmi EÚ by
mali nasledovné pokyny taktiež dodržiavat’, pokiaľ nie sú nahradené
miestnymi alebo národnými normami.
1. Adekvátne uzemnenia musia byt’ vykonané na všetkých bodoch
uzemnenia, interných aj externých, tam, kde sú poskytnuté.
2. Po inštalácii alebo riešení problémov musia byt’ všetky bezpečnostné
kryty a bezpečnostné uzemnenia vymenené. Integrita všetkých
uzemňovacích terminálov musí byt’ vždy zachovaná.
3. Káble siet’ového napájania musia byt’ v zhode s požiadavkami IEC227
alebo IEC245.
4. Všetky káblové pripojenia by mali byt’ vhodné pre používanie v teplote
okolia vyššej, ako 75°C.
5. Všetky použité káblové priechodky musia mat’ také vnútorné rozmery,
aby poskytovali adekvátne uchopenie kábla.
6. Pre zaistenie bezpečnej prevádzky tohto zariadenia musí byt’ pripojenie
k siet’ovému napájaniu zapojené len cez prerušovač obvodu, ktorý
počas poruchovej situácie odpojí všetky obvody elektrických vodičov.
Prerušovač obvodu by mal obsahovat’ aj mechanicky ovládaný úsekový
vypínač. Ak nie, musí byt’ poskytnutý iný spôsob odpojenia zariadenia
od siet’ového napájania a tento spôsob musí byt’ zreteľne označený.
Prerušovače obvodu alebo spínače musia byt’ v zhode s uznanou
normou, ako napr. IEC947. Všetky káblové pripojenia musia vyhovovat’
akýmkoľvek miestnym normám.
7. Tam, kde je zariadenie alebo kryty označené symbolom
na pravej strane, sa pravdepodobne nachádza
nebezpečné napätie. Tieto kryty by sa mali odoberat’
len vtedy, keď je zariadenie odpojené od elektrickej
energie a len vyškoleným servisným personálom.
8. Tam, kde je zariadenie alebo kryty označené symbolom
na pravej strane, existuje nebezpečenstvo horúcich
povrchov. Tieto kryty by mali byt’ odstraňované len
vyškoleným servisným personálom, pričom je
zariadenie odpojené od elektrickej energie. Určité
povrchy môžu ostat’ horúce na dotyk.
9. V miestach, kde je zariadenie alebo kryty označené
symbolom na pravej strane, si kvôli pokynom pozrite
Operátorskú príručku.
10. Všetky obrázkové symboly použité pri tomto produkte
zodpovedajú jednej alebo viacerým nasledujúcim
normám: EN61010-1, IEC417 a ISO3864.
11. V miestach, kde je zariadenie alebo značky označené nápisom
"Neotvárat’ pod elektrickým prúdom" alebo podobné, existuje
nebezpečenstvo vznietenia v oblastiach s prítomnost’ou výbušného
ovzdušia. Toto zariadenie sa smie otvárat’ len v prípade odpojenia od
elektrického napájania a ponechania zariadenia vychladnút’ po dobu
uplynutia dostatočného času tak, ako je to uvedené na štítku alebo
v návode na použitie - a len vyškoleným servisným personálom.
A-20
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
POMEMBNO
Varnostna navodila za povezavo in vgradnjo naprave
Naslednja varnostna navodila veljajo za vse države članice EU. Zaradi
zagotovitve skladnosti z nizkonapetostno direktivo morate navodila
strogo upoštevati. V državah, ki niso članice EU, je treba upoštevati tudi
naslednje smernice, razen če jih ne zamenjujejo lokalni ali nacionalnimi
standardi.
1. Do vseh ozemljitvenih točk, notranjih in zunanjih, ki so na voljo, morajo
biti speljane ustrezne ozemljitvene povezave.
2. Po vgradnji ali odpravljanju težav je treba namestiti vse varnostne
pokrove in zaščitne ozemljitve. Brezhibnost vseh ozemljitvenih
priključkov je treba nenehno preverjati.
3. Omrežni napajalni kabli morajo biti skladni z zahtevami standarda
IEC227 ali IEC245.
4. Vsa napeljava mora biti primerna za uporabi pri temperaturi okolja, višji
od 75 °C.
5. Notranje dimenzije kabelskih tesnilk morajo zagotavljati ustrezno
pritrditev kablov.
6. Za zagotovitev varnega delovanja opreme mora biti povezava z omrežnim
napajanjem vzpostavljena prek odklopnega stikala, ki v primeru napake
izklopi vse tokokroge s prevodniki. Odklopno stikalo lahko vključuje tudi
mehansko izolacijsko stikalo. V nasprotnem primeru morajo biti
zagotovljeni in jasno označeni drugi načini za izklop opreme iz napajanja.
Odklopna in druga stikala morajo biti skladna z uveljavljenimi standardi,
kot je IEC947. Vsa napeljava mora biti skladna z lokalnimi standardi.
7. V opremi ali pod pokrovi, ki so označeni s simbolom na
desni, je prisotna nevarna napetost. Te pokrove je
dovoljeno odstraniti samo, če je napajanje opreme
izklopljeno. To lahko izvaja samo usposobljeno servisno
osebje.
8. Pri opremi ali pod pokrovi, ki so označeni s simbolom na
desni, so prisotne nevarne vroče površine. Te pokrove
lahko odstranjuje samo usposobljeno servisno osebje.
Napajanje opreme mora biti izklopljeno. Določene
površine so lahko vroče.
9. Pri opremi ali pokrovih, ki so označeni s simbolom na
desni, si za navodila oglejte priročnik za upravljanje.
10. Vsi uporabljeni grafični simboli so iz enega ali več
naslednjih standardov: EN61010-1, IEC417 in ISO3864.
11. Če je na opremi ali oznakah navedeno "Ne odpirajte, če je pod
napetostjo" ali podobno opozorilo, je na območjih z eksplozivnim
ozračjem prisotna nevarnost vžiga. To opremo je dovoljeno odpirati
samo, če je napajanje izklopljeno in je poteklo dovolj časa, da se
oprema ohladi, kot je navedeno na oznaki ali v priročniku z navodili.
Opremo lahko odpira samo usposobljeno servisno osebje.
A-21
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
IMPORTANTE
Instrucciones de seguridad para el montaje y cableado de este aparato.
Las siguientes instrucciones de seguridad, son de aplicacion especifica
a todos los miembros de la UE y se adjuntaran para cumplir la normativa
europea de baja tension.
1. Se deben preveer conexiones a tierra del equipo, tanto externa como
internamente, en aquellos terminales previstos al efecto.
2. Una vez finalizada las operaciones de mantenimiento del equipo, se
deben volver a colocar las cubiertas de seguridad aasi como los
terminales de tierra. Se debe comprobar la integridad de cada terminal.
3. Los cables de alimentacion electrica cumpliran con las normas IEC 227
o IEC 245.
4. Todo el cableado sera adecuado para una temperatura ambiental de
75ºC.
5. Todos los prensaestopas seran adecuados para una fijacion adecuada
de los cables.
6. Para un manejo seguro del equipo, la alimentacion electrica se realizara
a traves de un interruptor magnetotermico ( min 10 A ), el cual
desconectara la alimentacion electrica al equipo en todas sus fases
durante un fallo. Los interruptores estaran de acuerdo a la norma IEC
947 u otra de reconocido prestigio.
7. Cuando las tapas o el equipo lleve impreso el simbolo
de tension electrica peligrosa, dicho alojamiento
solamente se abrira una vez que se haya interrumpido
la alimentacion electrica al equipo asimismo la
intervencion sera llevada a cabo por personal
entrenado para estas labores.
8. Cuando las tapas o el equipo lleve impreso el simbolo,
hay superficies con alta temperatura, por tanto se abrira
una vez que se haya interrumpido la alimentacion
electrica al equipo por personal entrenado para estas
labores, y al menos se esperara unos 45 minutos para
enfriar las superficies calientes.
9. Cuando el equipo o la tapa lleve impreso el simbolo, se
consultara el manual de instrucciones.
10. Todos los simbolos graficos usados en esta hoja, estan
de acuerdo a las siguientes normas EN61010-1,
IEC417 & ISO 3864.
11. Cuando el equipo o las etiquetas tienen la indicación " No abrir mientras
reciba energía" u otra similar, existe el peligro de ignición en zonas
donde haya un ambiente explosivo. Este equipo sólo debe ser abierto
por personal de servicio cualificado después de apagarlo y dejar pasar
el intervalo de tiempo correspondiente indicado en la etiqueta o el
manual de instrucciones para que el equipo se enfríe.
A-22
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
VIKTIGT
Säkerhetsföreskrifter för kablage och installation av denna apparat.
Följande säkerhetsföreskrifter är tillämpliga för samtliga
EU-medlemsländer. De skall följas i varje avseende för att
överensstämma med Lågspännings direktivet. Icke EU medlemsländer
skall också följa nedanstående punkter, såvida de inte övergrips av
lokala eller nationella föreskrifter.
1. Tillämplig jordkontakt skall utföras till alla jordade punkter, såväl internt
som externt där så erfordras.
2. Efter installation eller felsökning skall samtliga säkerhetshöljen och
säkerhetsjord återplaceras. Samtliga jordterminaler måste hållas
obrutna hela tiden.
3. Matningsspänningens kabel måste överensstämma med föreskrifterna i
IEC227 eller IEC245.
4. Allt kablage skall vara lämpligt för användning i en
omgivningstemperatur högre än 75ºC.
5. Alla kabelförskruvningar som används skall ha inre dimensioner som
motsvarar adekvat kabelförankring.
6. För att säkerställa säker drift av denna utrustning skall anslutning till
huvudströmmen endast göras genom en säkring (min 10A) som skall
frånkoppla alla strömförande kretsar när något fel uppstår. Säkringen
kan även ha en mekanisk frånskiljare. Om så inte är fallet, måste ett
annat förfarande för att frånskilja utrustningen från strömförsörjning
tillhandahållas och klart framgå genom markering. Säkring eller
omkopplare måste överensstämma med en gällande standard såsom t
ex IEC947.
7. Där utrustning eller hölje är markerad med vidstående
symbol föreliggerisk för livsfarlig spänning i närheten.
Dessa höljen får endast avlägsnas när strömmen ej är
ansluten till utrustningen - och då endast av utbildad
servicepersonal.
8. När utrustning eller hölje är markerad med vidstående
symbol föreligger risk för brännskada vid kontakt med
uppvärmd yta. Dessa höljen får endast avlägsnas av
utbildad servicepersonal, när strömmen kopplats från
utrustningen. Vissa ytor kan vara mycket varma att
vidröra även upp till 45 minuter efter avstängning av
strömmen.
9. När utrustning eller hölje markerats med vidstående
symbol bör instruktionsmanualen studeras för
information.
10. Samtliga grafiska symboler som förekommer i denna
produkt finns angivna i en eller flera av följande
föreskrifter:- EN61010-1, IEC417 & ISO3864.
11. För utrustning som markerats med föreskrifter som "Öppna inte när
strömmen är på", eller liknande, råder explosionsrisk när det
förekommer explosiva ångor. Utrustningen får endast öppnas efter att
strömmen stängts av och efter att utrustningen fått svalna under så lång
tid som anges i instruktionsboken. Öppnandet får endast utföras av
utbildad servicepersonal.
A-23
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
SAFETY DATA SHEET
FOR CERAMIC FIBER
PRODUCTS
JULY 1, 1996
SECTION I. IDENTIFICATION
PRODUCT NAME
Ceramic Fiber Heaters, Molded Insulation Modules and Ceramic Fiber
Radiant Heater Panels.
CHEMICAL FAMILY
Vitreous Aluminosilicate Fibers with Silicon Dioxide.
CHEMICAL NAME
N.A.
CHEMICAL FORMULA
N.A.
MANUFACTURER'S NAME AND ADDRESS
Watlow Columbia
2101 Pennsylvania Drive
Columbia, MO 65202
573-814-1300, ext. 5170
573-474-9402
HEALTH HAZARD SUMMARY WARNING
•
Possible cancer hazard based on tests with laboratory animals.
•
May be irritating to skin, eyes and respiratory tract.
•
May be harmful if inhaled.
•
Cristobalite (crystalline silica) formed at high temperatures (above
1800ºF) can cause severe respiratory disease.
SECTION II. PHYSICAL DATA
APPEARANCE AND ODOR
Cream to white colored fiber shapes. With or without optional white to
gray granular surface coating and/or optional black surface coating.
SPECIFIC WEIGHT: 12-25 LB./CUBIC FOOT
BOILING POINT: N.A.
VOLATILES (% BY WT.): N.A.
WATER SOLUBILITY: N.A.
SECTION III. HAZARDOUS INGREDIENTS
MATERIAL, QUANTITY, AND THRESHOLD/EXPOSURE LIMIT VALUES
Aluminosilicate (vitreous) 99+ %
1 fiber/cc TWA
CAS. No. 142844-00-06
10 fibers/cc CL
Zirconium Silicate
0-10% 5 mg/cubic meter (TLV)
Black Surface Coating**
0 - 1% 5 mg/cubic meter (TLV)
Armorphous Silica/Silicon Dioxide
0-10% 20 mppcf (6 mg/cubic meter)
PEL (OSHA 1978) 3 gm cubic meter
(Respirable dust): 10 mg/cubic meter,
Intended TLV (ACGIH 1984-85)
**Composition is a trade secret.
A-24
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
SECTION IV. FIRE AND EXPLOSION DATA
FLASH POINT: None
FLAMMABILITY LIMITS: N.A.
EXTINGUISHING MEDIA
Use extinguishing agent suitable for type of surrounding fire.
UNUSUAL FIRE AND EXPLOSION HAZARDS / SPECIAL FIRE
FIGHTING PROCEDURES
N.A.
SECTION V. HEALTH HAZARD DATA
THRESHOLD LIMIT VALUE
(See Section III)
EFFECTS OF OVER EXPOSURE
•
EYE - Avoid contact with eyes. Slightly to moderately irritating.
Abrasive action may cause damage to outer surface of eye.
•
INHALATION - May cause respiratory tract irritation. Repeated or
prolonged breathing of particles of respirable size may cause
inflammation of the lung leading to chest pain, difficult breathing,
coughing and possible fibrotic change in the lung (Pneumoconiosis).
Pre-existing medical conditions may be aggravated by exposure:
specifically, bronchial hyper-reactivity and chronic bronchial or lung
disease.
•
INGESTION - May cause gastrointestinal disturbances. Symptoms
may include irritation and nausea, vomiting and diarrhea.
•
SKIN - Slightly to moderate irritating. May cause irritation and
inflammation due to mechanical reaction to sharp, broken ends of
fibers.
EXPOSURE TO USED CERAMIC FIBER PRODUCT
Product which has been in service at elevated temperatures (greater
than 1800ºF/982ºC) may undergo partial conversion to cristobalite, a
form of crystalline silica which can cause severe respiratory disease
(Pneumoconiosis). The amount of cristobalite present will depend on
the temperature and length of time in service. (See Section IX for
permissible exposure levels).
SPECIAL TOXIC EFFECTS
The existing toxicology and epidemiology data bases for RCF's are still
preliminary. Information will be updated as studies are completed and
reviewed. The following is a review of the results to date:
EPIDEMIOLOGY
At this time there are no known published reports demonstrating
negative health outcomes of workers exposed to refractory ceramic
fiber (RCF). Epidemiologic investigations of RCF production workers
are ongoing.
1. There is no evidence of any fibrotic lung disease (interstitial fibrosis)
whatsoever on x-ray.
2. There is no evidence of any lung disease among those employees
exposed to RCF that had never smoked.
A-25
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
3. A statistical "trend" was observed in the exposed population between
the duration of exposure to RCF and a decrease in some measures
of pulmonary function. These observations are clinically insignificant.
In other words, if these observations were made on an individual
employee, the results would be interpreted as being within the
normal range.
4. Pleural plaques (thickening along the chest wall) have been
observed in a small number of employees who had a long duration of
employment. There are several occupational and non-occupational
causes for pleural plaque. It should be noted that plaques are not
"pre-cancer" nor are they associated with any measurable effect on
lung function.
TOXICOLOGY
A number of studies on the health effects of inhalation exposure of rats
and hamsters are available. Rats were exposed to RCF in a series of
life-time nose-only inhalation studies. The animals were exposed to 30,
16, 9, and 3 mg/m3, which corresponds with approximately 200, 150,
75, and 25 fibers/cc.
Animals exposed to 30 and 16 mg/m3 were observed to have
developed a pleural and parenchymal fibroses; animals exposed to 9
mg/m3 had developed a mild parenchymal fibrosis; animals exposed to
the lowest dose were found to have the response typically observed any
time a material is inhaled into the deep lung. While a statistically
significant increase in lung tumors was observed following exposure to
the highest dose, there was no excess lung cancers at the other doses.
Two rats exposed to 30 mg/m3 and one rat exposed to 9 mg/m3
developed masotheliomas.
The International Agency for Research on Cancer (IARC) reviewed the
carcinogenicity data on man-made vitreous fibers (including ceramic
fiber, glasswool, rockwool, and slagwool) in 1987. IARC classified
ceramic fiber, fibrous glasswool and mineral wool (rockwool and
slagwool) as possible human carcinogens (Group 2B).
EMERGENCY FIRST AID PROCEDURES
A-26
•
EYE CONTACT - Flush eyes immediately with large amounts of
water for approximately 15 minutes. Eye lids should be held away
from the eyeball to insure thorough rinsing. Do not rub eyes. Get
medical attention if irritation persists.
•
INHALATION - Remove person from source of exposure and move
to fresh air. Some people may be sensitive to fiber induced irritation
of the respiratory tract. If symptoms such as shortness of breath,
coughing, wheezing or chest pain develop, seek medical attention. If
person experiences continued breathing difficulties, administer
oxygen until medical assistance can be rendered.
•
INGESTION - Do not induce vomiting. Get medical attention if
irritation persists.
•
SKIN CONTACT - Do not rub or scratch exposed skin. Wash area of
contact thoroughly with soap and water. Using a skin cream or lotion
after washing may be helpful. Get medical attention if irritation
persists.
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
SECTION VI. REACTIVITY DATA
STABILITY/CONDITIONS TO AVOID
Stable under normal conditions of use.
HAZARDOUS POLYMERIZATION/CONDITIONS TO AVOID
N.A.
INCOMPATIBILITY/MATERIALS TO AVOID
Incompatible with hydrofluoric acid and concentrated alkali.
HAZARDOUS DECOMPOSITION PRODUCTS
N.A.
SECTION VII. SPILL OR LEAK PROCEDURES
STEPS TO BE TAKEN IF MATERIAL IS RELEASED OR SPILLED
Where possible, use vacuum suction with HEPA filters to clean up
spilled material. Use dust suppressant where sweeping if necessary.
Avoid clean up procedure which may result in water pollution. (Observe
Special Protection Information Section VIII.)
WASTE DISPOSAL METHODS
The transportation, treatment, and disposal of this waste material must
be conducted in compliance with all applicable Federal, State, and Local
regulations.
SECTION VIII. SPECIAL PROTECTION INFORMATION
RESPIRATORY PROTECTION
Use NIOSH or MSHA approved equipment when airborne exposure
limits may be exceeded. NIOSH/MSHA approved breathing equipment
may be required for non-routine and emergency use. (See Section IX
for suitable equipment).
Pending the results of long term health effects studies, engineering
control of airborne fibers to the lowest levels attainable is advised.
VENTILATION
Ventilation should be used whenever possible to control or reduce
airborne concentrations of fiber and dust. Carbon monoxide, carbon
dioxide, oxides of nitrogen, reactive hydrocarbons and a small amount
of formaldehyde may accompany binder burn off during first heat. Use
adequate ventilation or other precautions to eliminate vapors resulting
from binder burn off. Exposure to burn off fumes may cause respiratory
tract irritation, bronchial hyper-reactivity and asthmatic response.
SKIN PROTECTION
Wear gloves, hats and full body clothing to prevent skin contact. Use
separate lockers for work clothes to prevent fiber transfer to street
clothes. Wash work clothes separately from other clothing and rinse
washing machine thoroughly after use.
EYE PROTECTION
Wear safety glasses or chemical worker's goggles to prevent eye
contact. Do not wear contact lenses when working with this substance.
Have eye baths readily available where eye contact can occur.
A-27
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
SECTION IX. SPECIAL PRECAUTIONS
PRECAUTIONS TO BE TAKEN IN HANDLING AND STORING
General cleanliness should be followed.
The Toxicology data indicate that ceramic fiber should be handled with
caution. The handling practices described in this MSDS must be strictly
followed. In particular, when handling refractory ceramic fiber in any
application, special caution should be taken to avoid unnecessary
cutting and tearing of the material to minimize generation of airborne
dust.
It is recommended that full body clothing be worn to reduce the potential
for skin irritation. Washable or disposable clothing may be used. Do not
take unwashed work clothing home. Work clothes should be washed
separately from other clothing. Rinse washing machine thoroughly after
use. If clothing is to be laundered by someone else, inform launderer of
proper procedure. Work clothes and street clothes should be kept
separate to prevent contamination.
Product which has been in service at elevated temperatures (greater
than 1800°F/982°C) may undergo partial conversion t o cristobalite, a
form of crystalline silica. This reaction occurs at the furnace lining hot
face. As a consequence, this material becomes more friable; special
caution must be taken to minimize generation of air-borne dust. The
amount of cristobalite present will depend on the temperature and
length in service.
IARC has recently reviewed the animal, human, and other relevant
experimental data on silica in order to critically evaluate and classify the
cancer causing potential. Based on its review, IARC classified
crystalline silica as a group 2A carcinogen (probable human
carcinogen).
The OSHA permissible exposure limit (PEL for cristobalite is 0.05
mg/m3 (respirable dust). The ACGIH threshold limit value (TLV) for
cristobalite is 0.05 mg/m3 (respirable dust) (ACGIH 1991-92). Use
NIOSH or MSHA approved equipment when airborne exposure limits
may be exceeded. The minimum respiratory protection recommended
for given airborne fiber or cristobalite concentrations are:
CONCENTRATION
Concentration
0-1 fiber/cc or 0-0.05 mg/m3
cristobalite (the OSHA PEL)
Up to 5 fibers/cc or up to 10 times the
OSHA PEL for cristobalite
Up to 25 fibers/cc or 50 times the
OSHA PEL for cristobalite (2.5
mg/m3)
Greater than 25 fibers/cc or 50 times
the OSHA PEL for cristobalite (2.5
mg/m3)
A-28
Personal Protective Equipment
Optional disposable dust respirator (e.g.
3M 9970 or equivalent).
Half face, air purifying respirator
equipped with high efficiency particulate
air (HEPA) filter cartridges (e.g. 3M 6000
series with 2040 filter or equivalent).
Full face, air purifying respirator with
high efficiency particulate air (HEPA)
filter cartridges (e.g. 3M 7800S with
7255 filters or equivalent) or powered air
purifying respirator (PARR) equipped
with HEPA filter cartridges (e.g. 3M
W3265S with W3267 filters or
equivalent).
Full face, positive pressure supplied air
respirator (e.g. 3M 7800S with W9435
hose & W3196 low pressure regulator kit
connected to clean air supply or
equivalent).
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
If airborne fiber or cristobalite concentrations are not known, as
minimum protection, use NIOSH/MSHA approved half face, air purifying
respirator with HEPA filter cartridges.
Insulation surface should be lightly sprayed with water before removal to
suppress airborne dust. As water evaporates during removal, additional
water should be sprayed on surfaces as needed. Only enough water
should be sprayed to suppress dust so that water does not run onto the
floor of the work area. To aid the wetting process, a surfactant can be
used.
After RCF removal is completed, dust suppressing cleaning methods,
such as wet sweeping or vacuuming, should be used to clean the work
area. If dry vacuuming is used, the vacuum must be equipped with
HEPA filter. Air blowing or dry sweeping should not be used. Dust
suppressing components can be used to clean up light dust.
Product packaging may contain product residue. Do not reuse except to
reship or return Ceramic Fiber products to the factory.
A-29
Instruction Manual
OCX 8800
HIGH PRESSURE GAS
CYLINDERS
IM-106-880ASB, Rev 1.2
July 2008
GENERAL PRECAUTIONS FOR HANDLING AND STORING HIGH
PRESSURE GAS CYLINDERS
• Edited from selected paragraphs of the Compressed
Gas Association's "Handbook of Compressed Gases"
published in 1981
Compressed Gas Association
1235 Jefferson Davis Highway
Arlington, Virginia 22202
Used by Permission
1. Never drop cylinders or permit them to strike each other violently.
2. Cylinders may be stored in the open, but in such cases, should be
protected against extremes of weather and, to prevent rusting, from
the dampness of the ground. Cylinders should be stored in the
shade when located in areas where extreme temperatures are
prevalent.
3. The valve protection cap should be left on each cylinder until it has
been secured against a wall or bench, or placed in a cylinder stand,
and is ready to be used.
4. Avoid dragging, rolling, or sliding cylinders, even for short distance;
they should be moved by using a suitable handtruck.
5. Never tamper with safety devices in valves or cylinders.
6. Do not store full and empty cylinders together. Serious suckback can
occur when an empty cylinder is attached to a pressurized system.
7. No part of cylinder should be subjected to a temperature higher than
52°C (125°F). A flame should never be permitted to come in contact
with any part of a compressed gas cylinder.
8. Do not place cylinders where they may become part of an electric
circuit. When electric arc welding, precautions must be taken to
prevent striking an arc against the cylinder.
A-30
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
Appendix B
OCX 8800
SPA with HART Alarm
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page B-1
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page B-1
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page B-2
Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page B-2
OVERVIEW
This section describes the SPA with HART Alarm option for the OCX 8800.
DESCRIPTION
The Moore Industries SPA with HART Alarm, Figure B-1, is a 4-wire (line or
mains powered), site-programmable, digital process alarm. It connects to a
standard HART field device, and provides up to four, fully configurable,
contact-closure outputs based on readings of the HART digital data. The four
OCX 8800 alarm outputs recognized by the SPA are Low O2, High COe,
Calibration Status, and OCX Unit Failure.
37020018
Figure B-1. SPA with
HART Alarm
http://www..raihome.com
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
Figure B-2. OCX 8800 and
SPA Interface Connections
INSTALLATION
Refer to Figure B-2 for the typical interface connections for the OCX 8800 and
the SPA with HART alarm. Refer to the Moore Industries SPA user's manual
for additional information concerning SPA installation, setup, and operation.
SETUP
Setup of the SPA for communication with the OCX 8800 includes setting
internal jumpers and dip switches and configuring the SPA operating
parameters via a menu-driven selection and calibration procedure.
Jumper and Switch Settings
SPA jumper and switch settings are shown in Figure B-3. If the SPA with
HART was factory-configured by Emerson Process Management for
operation with your OCX 8800, jumper and switch setting adjustments are not
required. However, you may use the following procedure to verify that the
jumper and switch settings are correct. Adjust or verify jumper and switch
settings as follows:
Electrostatic discharge (ESD) protection is required to avoid damage to the SPA electronic
circuits.
1. Refer to Figure B-3. Turn the SPA over and slide the access cover out.
Before changing any jumper or switch position, take adequate
precautions to avoid an electrostatic discharge.
B-2
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
Figure B-3. SPA Jumper
and Dip Switch Settings
NOTICE
REMOVE THIS PANEL
FOR SERVICE ACCESS.
STATIC SENSITIVE.
USE CAUTION WHEN
CHANGING JUMPERS
Access Cover
SPA - Bottom View
Source Current Dip Switches
(shown in correct position)
Figure B-4. SPA Setup for
Calibration
Password
OFF
Failsafe Dip Switches
(shown in correct position)
2. Verify that the Password Jumper is set to the OFF position. If the jumper
is in the ON position, reposition the jumper.
3. Check the position of the Failsafe Dip Switches. Position the dip
switches as shown in Figure B-3.
Fluke Model 87
Multimeter
or equivalent
OCX
8800
Password
ON
37020022
Password
Jumper Pins
4. Check the position of the Source Current Dip Switches. Position the dip
switches as shown in Figure B-3.
mA
Optional
5. Reinstall the SPA access cover.
Configuration/Calibration
> 250Ω
< 950Ω
IN
Prior to operation, the SPA operating parameters must be configured via a
menu-driven setup procedure. At the end of the configuration procedure, the
SPA analog output signal is calibrated to insure valid communications.
IN
1. See Figure B-4. Connect a calibrated ammeter (Fluke Model 87 or
equivalent, accurate to ±0.025%) to the SPA analog output terminals.
Observe polarity.
AC
AC or DC
Power Input ACC
GND
37390024
2. Connect a 90 to 260 VAC or 22 to 300 VDC power source to the SPA
power terminals. When connecting an AC power source, use the AC
and ACC (AC Common) terminals. For a DC source, use the AC and
Ground terminals.
3. If desired, you can connect the 4 to 20 mA O2 signal wires from the
OCX 8800 analog output terminal block to the SPA Input terminals. (The
OCX must be operational to transmit the O2 signal. Observe polarity.)
B-3
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
NOTE
The O2 signal connection is not required for SPA configuration or calibration.
The OCX interface will allow you to observe the O2 signal level when the SPA
configuration procedure is completed.
4. Observe the front panel of the SPA, Figure B-5:
a. A process value display in the SPA display window indicates that the
SPA is operational. Four pushbuttons are located below the display
window.
b. Pressing a left-hand pushbutton scrolls up () or down () through
the SPA command menu, a submenu, or parameter values list.
c. Pressing the VIEW pushbutton displays rail limits and alarm relay
configurations. There are five sequential displays in the VIEW mode.
While in the VIEW mode, the up (), down (), and SELECT
pushbuttons are disabled.
NOTE
In the VIEW mode, you can scroll through and display the output zero and full
scale settings and the alarm relay trip points and configurations.
d. Pressing the SELECT pushbutton selects the displayed menu or
submenu command or selects a displayed parameter variable.
Figure B-5. SPA Front Panel
SPA
Display
Window
Right-Hand
Pushbuttons
37020017
Left-Hand
Pushbuttons
B-4
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
5. Figure B-6 shows the SPA menu, submenus, and parameter values that
must be selected to configure the SPA for use with the OCX 8800. Use
the following instructions and selections shown to properly configure the
SPA.
a. Press the SELECT pushbutton. Observe the display window on the
SPA front panel. The display window should read SET HART.
b. See the SET HART command in Figure B-6. To the right of the
command window is the SET HART submenu and related parameter
values that must be selected via the front panel pushbuttons on the
SPA.
c. In the submenu views shown:
means press the down pushbutton.
means press the up pushbutton.
means press the SELECT pushbutton one time.
means press the or pushbutton until the desired parameter
value is shown in the SPA window.
d. Proceed through the SPA menu, selecting the menu commands and
parameter values indicated. After completing the sequence in the
first column, go to the top of the second column and continue.
e. To exit the menu, repeatedly press SELECT to display any main
menu command. Then, press or until CONF EXIT is displayed.
Select CONF EXIT.
f. Detailed instructions concerning the configuration menu and the
submenu structure for each main command are provided in the SPA
user's manual.
B-5
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
Figure B-6. SPA Configuration Menu
for OCX 8800 Communication
SET
HART
 SET MSTR   PRIM
 SET FUCT   NRML
 NUM VARS   02
 NUM TRYS   03
 EXIT HART  SGNL SRCE
SCLE INPT
SCLE
DSPL
 AOUT SRCE   SV
SGNL  AL2 SRCE   PV
SRCE  AL3 SRCE   PV
 AL4 SRCE   PV
 EXIT SRCE  FLT SRCE
FLT
SRCE
 PV DSPL
 SET ZERO   0000 PCT
 SET FULL   25 PCT
 SV DSPL
 SET ZERO   0000 PPM
 SET FULL   1000 PPM
 EXIT DSPL  SCLE OUT
SCLE
OUT
 SET ZERO   4.000 MA
 SET FULL   20.00 MA
 EXIT Z/FS
 TRIM OUT
TRIM
OUT
 TRIM ZERO   4.000 MA
 TRIM FULL   20.00 MA
 EXIT TRIM  CONF ALRM
 CONF OPTS
 SET LINR   LINR OFF
CONF  PV SCLE   AUTO
OPTS  DSPL EGU   CSTM   PPM
SET
EGU
SCLE
INPT
 PV EGU   PCT
 SV EGU   CSTM   PPM
(select P, select P, select M)
 EXIT EGU  SCLE INPT
 PV SCLE  PV ZERO  0000 PCT
 PV FULL  25 PCT
 SCLE SV  SET ZERO  0000 PPM
 SET FULL  1000 PPM
 EXIT SCLE  SCLE DSPL
SCLE DSPL
B-6
Refer to
CONF SPA manual
ALRM for settings or
 PASS WORD
PASS
WORD
CONF
EXIT
 CONF EXIT
 SPA setup complete;
observe process value display.
37390003
(select P, select P, select M)
 DSPL SRCE   SV
 AL2 SEL  AL2 FLT
 SET FAIL  FAIL LOW
 EXIT OPTS  SET EGU
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
Appendix C
Return of Materials
RETURNING MATERIAL
If factory repair of defective equipment is required, proceed as follows:
1. Secure a return authorization number from an Emerson Process
Management sales office or representative before returning the
equipment. Equipment must be returned with complete identification in
accordance with Emerson Process Management instructions or it will
not be accepted.
In no event will Emerson Process Management be responsible for
equipment returned without proper authorization and identification.
2. Carefully pack defective unit in a sturdy box with sufficient shock
absorbing material to ensure that no additional damage will occur during
shipping.
3. In a cover letter, describe completely:
a. The symptoms from which it was determined that the equipment is
faulty.
b. The environment in which the equipment has been operating
(housing, weather, vibration, dust, etc.).
c. Site from which equipment was removed.
d. Whether warranty or nonwarranty service is requested.
e. Complete shipping instructions for return of equipment.
f. Reference the return authorization number.
4. Enclose a cover letter and purchase order and ship the defective
equipment according to instructions provided in Emerson Process
Management Return Authorization, prepaid, to:
Emerson Process Management
RMR Department
Daniel Headquarters
11100 Britmore Park Drive
Houston, TX 77041
If warranty service is requested, the defective unit will be carefully inspected
and tested at the factory. If failure was due to conditions listed in the standard
Rosemount Analytical warranty, the defective unit will be repaired or replaced
at Emerson Process Management's option, and an operating unit will be
returned to the customer in accordance with shipping instructions furnished in
the cover letter.
For equipment no longer under warranty, the equipment will be repaired at the
factory and returned as directed by the purchase order and shipping
instructions.
http://www..raihome.com
Instruction Manual
OCX 8800
C-2
IM-106-880ASB, Rev 1.2
July 2008
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
Index
A
Accessories . . . . . . . . . . . . . . 1-15
Alarm Relay Events . . . . . . . . 7-11
Assemble O2 Sensor . . . . . . . 6-19
Autocalibration Setup, HART . . 5-1
Autocalibration, HART . . . . . . . 5-3
B
Band Heater Height . . . . . . . . 6-22
Blowback System . . . . . . 1-5, 2-17
Blowback with
Autocalibration . . . . . . . . . . . 2-18
Blowback without
Autocalibration . . . . . . . . . . . 2-19
C
Cable Connections . . . . . . . . . . 6-6
Cell Output . . . . . . . . . . . . . . . . 7-7
COe Band Heater . . . . . . . . . . 6-14
COe Sensor
Assembly . 6-11, 6-17, 6-18, 6-22
COe Sensor Holder
Alignment . . . . . . . . . . . . . . . 6-23
COe Sensor Parts . . . . . . . . . . 6-21
COe Sensor, Thermocouple, and
Heater Connections . . . . . . . 6-24
Combustibles Sensor . . . . . . . . 1-3
Component Checklist . . . . . . . . 1-1
Configuration . . . . . . . . . . . . . . 3-2
D
D/A Trim Procedure . . . . . . . . . 5-8
Defaults . . . . . . . . . . . . . . . . . . . 3-3
Diagnostic Alarms . . . . . . . . . . . 7-2
E
Eductor . . . . . . . . . . . . . 6-14, 6-20
Eductor Holder . . . . . . . . . . . . 6-14
Electrical Noise . . . . . . . . . . . . . 7-1
Electronics Housing
Components . . . . . . . . . . . . . . 8-6
Electronics Housing
Disassembly . . . . . . . . . . . . . 6-28
http://www.raihome.com
Electronics Housing
Terminal Blocks . . . . . . . . . . . 6-5
Electrostatic Discharge . . . . . . . 7-1
Essential Instructions . . . . . . . . . i-i
F
Factory Repair . . . . . . . . . . . . . .C-1
Fault Indications . . . . . . . . . . . . 7-3
Fault Isolation . . . . . . . . . . . . . . 7-3
Fitting, ‘E’ Type . . . . . . . . . . . . 6-34
Fitting, ‘R’ Type . . . . . . . . . . . . 6-34
Fuse Locations . . . . . . . . . . . . . 7-2
G
Grounding . . . . . . . . . . . . . . . . . 7-1
H
Handling . . . . . . . . . . . . . . . . . . 1-6
HART Communicator . . . . . . . . 4-1
HART Connections . . . . . . 1-7, 1-8
HART PC Connections . . . . . . . 4-4
HART Signal Connections . . . . . 4-1
HART Signal Line
Connections . . . . . . . . . . 4-2, 4-3
Heater Strut Assembly . . . . . . 6-19
I
Install Eductor . . . . . . . . . . . . . 6-20
Install Electronics Stack . . . . . . 6-31
Install Flash PROM . . . . . . . . . 6-31
Install Remote Electronics . . . . . 6-7
Install Sensor Housing . . . . . . . 6-4
Install Solenoid Valves . . . . . . 6-31
Install Tube Fittings . . . . . . . . . 6-35
Instrument Air . . . . . . . . . . . . . . 1-7
M
Manual Calibration,
HART . . . . . . . . . . . .5-3, 5-5, 5-6
Material Safety Data Sheet . . .A-24
N
Nernst Equation . . . . . . . . . . . . .1-3
O
O2 Cell and
Heater Strut . . . . . . . . . 6-25, 6-27
O2 Cell and Heater Strut
Assembly . . . 6-9, 6-10, 6-16, 8-9
O2 Cell Output Voltage . . . . . . .1-4
O2 Cell Replacement Kit . . . . .6-16
O2 Cell, Heater, and
Thermocouple . . . . . . . . . . . .6-16
O2 Cell, Thermocouple, and
Heater Connections . . 6-10, 6-26
OCX Specifications . . . . . . . . .1-12
OCX with Remote
Electronics . . . . . . . . . . . . . . .6-2
Operation Diagram . . . . . . . . . .1-6
Oxygen and Cell Output . . . . . .7-7
P
Personal Computer (PC) . . . . . .1-5
Power Up . . . . . . . . . . . . . . . . . .3-3
Pre-Heater . . . . . . . . . . . . . . . .6-18
Pre-Heater Alignment . . . . . . .6-19
Product Matrix . . . . . . . . . . . . .1-14
R
Reference Air Tube . . . . . . . . .6-10
Removal and Installation . . . . . .6-1
Remove Eductor . . . . . . . . . . .6-14
Remove EEprom . . . . . . . . . . .6-28
Remove Electronics Stack . . . .6-29
Remove Flash PROM . . . . . . .6-28
Remove Remote Electronics . . .6-4
Remove Sensor Housing . . . . . .6-2
Remove Solenoid Valves . . . . .6-30
Remove Tube Fittings . . . . . . .6-34
Repair Sensor Housing . . . . . . .6-8
Replace Tube Fittings . . . . . . .6-34
Reset Procedure . . . . . . . . . . . .3-4
Resistance Devices (RTD) . . . .1-3
Returning Material . . . . . . . . . . C-1
RTD . . . . . . . . . . . . . . . . . . . . . .1-3
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
S
Sample and Exhaust
Tubes . . . . . . . . . . . . . 6-15, 6-19
Sample Block Heater
Rods . . . . . . . . . . . . . . 6-10, 6-25
Selected Distributed
Control Systems . . . . . . . . . . . 1-5
Sensor Housing
Components . . . . . . . . . . 8-2, 8-4
Sensor Housing
Disassembly . . . . . . . . . . . . . . 6-8
Sensor Housing Leak Test . . . 6-27
Sensor Housing Terminals . . . . 6-3
Solenoid Power Terminals . . . 6-33
SPA Configuration Menu . . . . . . B-6
Index-2
SPA Front Panel . . . . . . . . . . . .B-4
SPA Interface Connections . . . .B-2
SPA Setup for Calibration . . . . .B-3
SPA with HART Alarm . . . . . . . .B-1
Specifications . . . . . . . . . . . . . 1-12
SW1 . . . . . . . . . . . . . . . . . . . . . 3-2
SW2 . . . . . . . . . . . . . . . . . . . . . 3-2
SW3 . . . . . . . . . . . . . . . . . . . . . 3-2
System Configuration . . . . . . . . 1-4
System Description . . . . . . . . . . 1-1
System Features . . . . . . . . . . . . 1-4
System Operation . . . . . . . . . . . 1-5
System Overview . . . . . . . . . . . 1-1
System Package . . . . . . . . . . . . 1-2
T
Terminals Insulator . . . . . . . . .6-27
Test Gas Values . . . . . . . . . . . . .3-4
Total Power Loss . . . . . . . . . . . .7-2
Transmitter Package . . . . . . . . .1-1
Typical System
Installation . . . . . . .1-8, 1-10, 1-11
Typical System Package . . . . . .1-2
W
Warranty Service . . . . . . . . . . . C-1
WARRANTY
Rosemount Analytical warrants that the equipment manufactured and sold by it
will, upon shipment, be free of defects in workmanship or material. Should any
failure to conform to this warranty become apparent during a period of one year
after the date of shipment, Rosemount Analytical shall, upon prompt written notice
from the purchaser, correct such nonconformity by repair or replacement, F.O.B.
factory of the defective part or parts. Correction in the manner provided above
shall constitute a fulfillment of all liabilities of Rosemount Analytical with respect to
the quality of the equipment.
THE FOREGOING WARRANTY IS EXCLUSIVE AND IN LIEU OF ALL OTHER
WARRANTIES OF QUALITY WHETHER WRITTEN, ORAL, OR IMPLIED
(INCLUDING ANY WARRANTY OF MERCHANTABILITY OF FITNESS FOR
PURPOSE).
The remedy(ies) provided above shall be purchaser's sole remedy(ies) for any
failure of Rosemount Analytical to comply with the warranty provisions, whether
claims by the purchaser are based in contract or in tort (including negligence).
Rosemount Analytical does not warrant equipment against normal deterioration
due to environment. Factors such as corrosive gases and solid particulates can
be detrimental and can create the need for repair or replacement as part of normal
wear and tear during the warranty period.
Equipment supplied by Rosemount Analytical Inc. but not manufactured by it will
be subject to the same warranty as is extended to Rosemount Analytical by the
original manufacturer.
At the time of installation it is important that the required services are supplied to
the system and that the electronic controller is set up at least to the point where it
is controlling the sensor heater. This will ensure, that should there be a delay
between installation and full commissioning that the sensor being supplied with ac
power and reference air will not be subjected to component deterioration.
3951
7/08
Instruction Manual
IM-106-880ASB, Rev 1.2
July 2008
OCX 8800
Rosemount Analytical and the Rosemount Analytical logotype are registered trademarks of Rosemount Analytical Inc.
HART is a registered trademark of the HART Communications Foundation.
All other marks are the property of their respective owners.
WORLD HEADQUARTERS
Emerson Process Management
Rosemount Analytical Inc.
6565P Davis Industrial Parkway
Solon, OH 44139
T 440 914 1261
T 800 433 6076
F 440 914 1262
E [email protected]
EUROPE, MIDDLE EAST, AFRICA
Emerson Process Management
Shared Services Limited
Heath Place
Bognor Regis
West Sussex PO22 9SH
England
T 44 1243 863121
F 44 1243 845354
ROSEMOUNT ANALYTICAL EUROPE
Emerson Process Management
Gmbh & co. OHG
Industriestrasse 1
63594 Hasselroth
Germany
T 49 6055 884 0
F 49 6055 884 209
GAS CHROMATOGRAPHY
CENTER AND LATIN AMERICA
Emerson Process Management
Rosemount Analytical Inc.
11100 Brittmoore Park Drive
Houston, TX 77041
T 713 467 6000
F 713 827 3329
http://www.raihome.com
© 2008 Emerson Process Management. All rights reserved.
ASIA-PACIFIC
Emerson Process Management
Asia Pacific Private Limited
1 Pandan Crescent
Singapore 128461
Republic of Singapore
T 65 6 777 8211
F 65 6 777 0947
E [email protected]

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!