Autostart AS-2460 Specifications


Add to my manuals
86 Pages

advertisement

Autostart AS-2460 Specifications | Manualzz

Operating Manual

Refrigerated and

Heating Circulators air cooled

F25-HE

F32-HE

F34-HE

FP40-HE

FP50-HE

water cooled

FPW50-HE

Original Operating Manual

1.951.2460-V2 09/14

19512460-V2.doc

English

JULABO GmbH

77960 Seelbach / Germany

Tel. +49 (0) 7823 / 51-0

Fax +49 (0) 7823 / 24 91 [email protected] www.julabo.de

22.09.14

Congratulations!

You have made an excellent choice.

JULABO thanks you for the trust you have placed in us.

This operating manual has been designed to help you gain an understanding of the operation and possible applications of our circulators. For optimal utilization of all functions, we recommend that you thoroughly study this manual prior to beginning operation.

The JULABO Quality Management System

Temperature control devices for research and industry are developed, produced, and distributed according to the requirements of ISO 9001 and

ISO 14001. Certificate Registration No. 01 100044846

Unpacking and inspecting

Unpack the circulator and accessories and inspect them for possible transport damage. Damage should be reported to the responsible carrier, railway, or postal authority, and a damage report should be requested. These instructions must be followed fully for us to guarantee our full support of your claim for protecting against loss from concealed damage. The form required for filing such a claim will be provided by the carrier.

Printed in Germany Changes without prior notification reserved

Important: keep operating manual for future use

2

TABLE OF CONTENTS

Operating manual ................................................................................................................ 5

1. Intended use ................................................................................................................. 5

1.1. Description ............................................................................................................... 5

2. Operator responsibility – Safety recommendations ....................................................... 6

2.1. Disposal ................................................................................................................... 8

2.2. Warranty conditions ................................................................................................. 8

2.3. EC Conformity .......................................................................................................... 9

2.4. Technical specifications ......................................................................................... 15

2.5. Cooling water connection ....................................................................................... 18

Operating instructions ....................................................................................................... 19

3. Safety notes for the user ............................................................................................. 19

3.1. Explanation of safety notes .................................................................................... 19

3.2. Explanation of other notes ...................................................................................... 19

3.3. Safety recommendations ........................................................................................ 20

4. Operating controls and functional elements ................................................................ 22

5. Preparations ............................................................................................................... 26

5.1. Installation .............................................................................................................. 26

5.2. Bath fluids .............................................................................................................. 26

5.3. Temperature application to external systems ......................................................... 28

5.3.1. Tubing ............................................................................................................... 29

5.4. Filling / draining ...................................................................................................... 30

6. Operating procedures ................................................................................................. 31

6.1. Power connection ................................................................................................... 31

6.2. Switching on / Start - Stop ...................................................................................... 31

6.2.1. Switching on the circulator ................................................................................. 31

6.2.2. Switching on the Cooling Machine ..................................................................... 32

7. Setting of temperatures ................................................................................. 33

7.1. Using the pre-settings in the menu ........................................................... 33

7.2. Direct setting of temperatures ................................................................................ 34

8. Safety installations, warning functions ........................................................... 35

8.1. Excess temperature protection ............................................................................... 35

8.1.1. Early warning system, low level protection ......................................................... 36

8.2. Switch-over from warning to shutdown function...................................................... 37

8.3. Over and Sub temperature warning function .......................................................... 38

9. Menu functions.............................................................................................. 39

9.1. MENU PROGRAM – START.................................................................................. 40

HE

3

9.2. MENU PROGRAM – creation, administration ..........................................................43

9.3. MENU PUMP - Setting of pump pressure ...............................................................45

9.4. MENU CONFIG - Configuration of unit ....................................................................46

9.4.1. Remote control via the serial interface ................................................................47

9.4.2. Keypad control or setpoint setting via the analog input .......................................48

9.4.3. AUTOSTART ......................................................................................................48

9.4.4. OFF-MODE ........................................................................................................49

9.4.5. ACTVAR - actuating variable ..............................................................................49

9.4.6. Setting of clock and date ....................................................................................50

9.4.7. RESET – Factory settings...................................................................................50

9.5. MENU CONTROL – Control characteristics and parameters...................................50

9.5.1. CONTROL – Control INTERNAL / EXTERNAL ...................................................52

9.5.2. SELFTUNING .....................................................................................................53

9.5.3. Dynamic internal .................................................................................................53

9.5.4. Control parameters– XP-, TN-, TV- INTERNAL ..................................................54

9.5.5. COSPEED - external ..........................................................................................55

9.5.6. Control parameters – XPU-, XP-, TN-, TV- EXTERNAL ......................................55

9.6. MENU SERIAL - BAUDRATE, HANDSHAKE, PARITY ...........................................56

9.7. MENU ATC - Absolut Temperature Calibration .......................................................57

9.7.1. ATC SENSOR - INTERNAL / EXTERNAL ..........................................................59

9.7.2. ATC STATUS - YES / NO ...................................................................................59

9.7.3. CALIBRATION TYPE: 1 -/ 2 -/ 3 POINT..............................................................60

9.7.4. Example: 3-point calibration for internal control .................................................61

9.8. MENU LIMITS .........................................................................................................63

9.8.1. Limits for internal control .....................................................................................64

9.8.2. Limits for external control ....................................................................................64

9.9. MENU IN/OUT – Analog inputs/outputs (Option) .....................................................66

9.9.1. Outputs of the connector - REG+E-PROG ..........................................................69

9.9.2. Input of the connector - REG+E-PROG ..............................................................70

9.9.3. ALARM-output / Stand-by-input ..........................................................................72

10. Troubleshooting guide / error messages ......................................................................74

11. Electrical connections ..................................................................................................77

12. Remote control ............................................................................................................79

12.1. Setup for remote control ..........................................................................................79

12.2. Communication with a PC or a superordinated data system ...................................79

12.3. List of commands ....................................................................................................80

12.4. Status messages ....................................................................................................83

12.5. Error messages .......................................................................................................83

13. Installation of electronic module ...................................................................................84

14. JULABO Service – Online remote diagnosis ................................................................85

15. Cleaning / repairing the unit .........................................................................................86

4

HE

Operating manual

1. Intended use

JULABO circulators have been designed to control the temperature of specific fluids in a bath tank.

The units feature pump connections for temperature control of external systems (loop circuit).

JULABO circulators are not suitable for direct temperature control of foods, semiluxury foods and tobacco, or pharmaceutical and medical products.

Direct temperature control means unprotected contact of the object with the bath medium (bath fluid).

1.1. Description

1x10

ICC

TCF

ATC

3

RS232

Pt100

SMART

PUMP

The circulators are operated via the splash-proof keypad. The implemented microprocessor technology allows to set and to store different values that can be indicated on the VFD COMFORT-DISPLAY. Three menu keys facilitate adjusting setpoints, warning and safety functions and menu functions.

Temperature- and time-dependent processes can be stored and executed using the integrated programmer.

The control electronics including “ICC - Intelligent Cascade Control” automatically adjust the heat supply to the thermal requirements of the bath.

The TCF - Temperature Control Features give the user access to all important temperature control parameters. This means full control of the control behavior at all times and the ability to manually adjust or adapt the controls to the specific application.

Absolute Temperature Calibration (ATC3) provides high temperature stability at all points in the bath. With the 3-point calibration, an offset is adjusted at three temperatures to ensure an accurate temperature pattern at the selected spot in the bath over the entire temperature range.

Electrical connections:

The serial interface, switchable from RS232 to RS485, allows modern process technology without additional interface.

Connection for Pt100 external sensor for external temperature measurement and control.

Alarm output for external alarm message or control of JULABO refrigerating baths or solenoid valve (cooling water).

The electronic module (option) provides 3 further analog connections (alarm input, standby input, recorder output, programmer input).

The excess temperature protection conforming to IEC 61010-2-010 is a safety installation independent from the control circuit. This protection can be indicated and set on the VFD COMFORT-DISPLAY.

The early warning system for low level signals that bath fluid needs to be refilled before the low level protection conforming to IEC 61010-2-010 causes a complete shutdown of the main functional elements.

Intelligent pump system: The pump capacity (electronically adjustable via the motor speed) enables to adapt to varying conditions for internal and external temperature applications.

5

Operator responsibility – Safety recommendations

2. Operator responsibility – Safety recommendations

The products of JULABO ensure safe operation when installed, operated, and maintained according to common safety regulations. This section explains the potential dangers that may arise when operating the circulator and also specifies the most important safety precautions to preclude these dangers as far as possible.

 The operator is responsible for the qualification of the personnel operating the units.

 The personnel operating the units should be regularly instructed about the dangers involved with their job activities as well as measures to avert these dangers.

 Make sure all persons tasked with operating, installing, and maintaining the unit have read and understand the safety information and operating instructions.

 When using hazardous materials or materials that could become hazardous, the circulator may be operated only by persons who are absolutely familiar with these materials and the circulator.

These persons must be fully aware of possible risks.

If you have any questions concerning the operation of your unit or the information in this manual, please contact us!

Contact JULABO GmbH

Eisenbahnstraße 45

77960 Seelbach / Germany

Tel. +49 (0) 7823 / 51-0

Fax +49 (0) 7823 / 24 91 [email protected] www.julabo.de

Safety instructions for the operator:

 You have received a product designed for industrial use. Nevertheless, avoid strikes to the housing, vibrations, damage to the operating-element panel (keypad, display), and contamination.

 Make sure the product is checked for proper condition regularly (depending on the conditions of use). Regularly check (at least every 2 years) the proper condition of the mandatory, warning, prohibition and safety labels.

 Make sure that the mains power supply has low impedance to avoid any negative effects on instruments being operated on the same mains.

 This unit is designed for operation in a controlled electromagnetic environment. This means that transmitting devices (e.g., cellular phones) should not be used in the immediate vicinity.

Magnetic radiation may affect other devices with components sensitive to magnetic fields

(e.g., monitors). We recommend maintaining a minimum distance of 1 m.

 Permissible ambient temperature: max. 40 °C, min. 5 °C.

 Permissible relative humidity: 50% (40 °C).

 Do not store the unit in an aggressive atmosphere.

 Protect the unit from contamination.

 Do not expose the unit to sunlight.

Appropriate operation

Only qualified personnel is authorized to perform configuration, installation, maintenance and repairs of the circulator.

Routine operation can also be carried out by untrained personnel who should however be instructed by trained personnel.

6

HE

Use:

The bath can be filled with flammable materials. Fire hazard!

There might be chemical dangers depending on the bath medium used.

Observe all warnings for the used materials (bath fluids) and the respective instructions (safety data sheets).

Insufficient ventilation may result in the formation of explosive mixtures. Only use the unit in well ventilated areas.

Only use recommended materials (bath fluids). Only use non-acid and non corroding materials.

When using hazardous materials or materials that could become hazardous, the operator must affix the enclosed safety labels (1 + 2) to the front of the unit so they are highly visible:

1 Warning label W00: Colors: yellow, black

Danger area. Attention! Observe instructions.

(operating manual, safety data sheet)

2 or

Mandatory label M018: Colors: blue, white

Carefully read the user information prior to beginning operation.

Scope: EU

2 Semi S1-0701 Table A1-2 #9

Carefully read the user information prior to beginning operation.

Scope: USA, NAFTA

Particular care and attention is necessary because of the wide operating range.

There are thermal dangers: Burn, scald, hot steam, hot parts and surfaces that can be touched.

Warning label W26: Colors: yellow, black

Hot surface warning.

(The label is put on by JULABO)

Observe the instructions in the manuals for instruments of a different make that you connect to the circulator, particularly the respective safety recommendations. Also observe the pin assignment of plugs and technical specifications of the products.

7

Operator responsibility – Safety recommendations

2.1. Disposal

The circulator contains a back-up battery that supplies voltage to memory chips when the unit is switched off. Do not dispose of the battery with household waste!

Depending on battery regulations in your country, you might be obliged to give back used or defect batteries to gathering places.

The product may be used with oil as bath fluid. These oils fully or partially consist of mineral oil or synthetic oil. For disposal, observe the instructions in the safety data sheets.

These units contains refrigerants– at this time considered not to have any negative effects on the ozone layer. However, during the long operating period of the unit, disposal prescriptions may change.

So only qualified personnel should take care of disposal.

Valid in EU countries

See the current official journal of the European Union – WEEE directive.

Directive of the European Parliament and of the Council on waste electrical and electronic equipment (WEEE).

This directive requires electrical and electronic equipment marked with a crossedout trash can to be disposed of separately in an environmentally friendly manner.

Contact an authorized waste management company in your country.

Disposal with household waste (unsorted waste) or similar collections of municipal waste is not permitted!

2.2. Warranty conditions

JULABO GmbH warrants its products against defects in material or in workmanship, when used under appropriate conditions and in accordance with appropriate operating instructions for a period of ONE YEAR.

Extension of the warranty period – free of charge

With the ‘1PLUS warranty’ the user receives a free of charge extension to the warranty of up to 24 months, limited to a maximum of 10 000 working hours.

To apply for this extended warranty the user must register the unit on the JULABO web site www.julabo.de, indicating the serial no. The extended warranty will apply from the date of JULABO

GmbH’s original invoice.

JULABO GmbH reserves the right to decide the validity of any warranty claim. In case of faults arising either due to faulty materials or workmanship, parts will be repaired or replaced free of charge, or a new replacement unit will be supplied.

Any other compensation claims are excluded from this guarantee.

8

2.3. EC Conformity

HE

9

Operator responsibility – Safety recommendations

10

HE

11

Operator responsibility – Safety recommendations

12

HE

13

Operator responsibility – Safety recommendations

14

HE

2.4. Technical specifications

F25-HE

Working temperature range

Temperature stability

Cooling capacity

Medium: ethanol

Refrigerant

Overall dimensions

Bath opening

Bath depth

Filling volume

Weight

Mains power connection

Current draw

Mains power connection

Current draw

Mains power connection

Current draw

Mains power connection

Current draw

°C

°C

-28 ... 200

±0,01

°C kW cm

+20 0 -20

0.26 0.2 0.06

R134a

23x42x64

(WxDxH)

(WxL) cm cm

12x14

14 liters 3 ... 4,5 kg 32

230 V/50 Hz V/ Hz 207-253 / 50

(at 230 V) A 12

230 V/60 Hz V/ Hz -----

(at 230 V) A -----

115 V/60 Hz V/ Hz 103-127 / 60

(at 115 V) A 13

100 V/60 Hz V/ Hz 90-110 / 50-60

(at 100 V) A 13

Working temperature range

Temperature stability

°C

°C

Cooling capacity

Medium: ethanol

Refrigerant

°C kW

Overall dimensions

Bath opening

(WxDxH) cm

(WxL) cm cm

Bath depth

Filling volume liters kg

Weight

Mains power connection

Current draw

Mains power connection

Current draw

Mains power connection

Current draw

230 V/50 Hz

(at 230 V)

V/Hz

A

230 V/60 Hz V/ Hz

(at 230 V) A

115 V/60 Hz V/ Hz

(at 115 V) A

F32-HE

-35 ... 200

±0,01

+20 0 -20 -30

0.45 0.39 0.15 0.06

R134a

31x42x64

18x12

15

5,5 ... 8

38

207-253 / 50

12

207-253 / 60

12

103-127 / 60

14

90-110 / 50-60

14

F34-HE

-30 ... 150

±0.01

+20 0 -20 -30

0.45 0.32 0.14 0.03

R134a

38x58x64

24x30

15

14 ... 20

44

207-253 / 50

12

207-253 / 60

12

103-127 / 60

14

15

Operator responsibility – Safety recommendations

FP40-HE FP50-HE

FPW50-HE

-50 ... 200

Working temperature range

°C -40 ... 200

Temperature stability

°C ±0.01 ±0.01

Cooling capacity

Medium: ethanol

Refrigerant

°C kW

+20 0 -20 -30

0.68 0.5 0.32 0.04

+20 0 -20 -40

0.9 0.8 0.5 0.16

R134a

36x46x71

R404A or R507

42x49x72

Overall dimensions (WxDxH) cm

Bath opening

Bath depth

(WxL) cm cm

23x14

20

18x12

20

Filling volume liters 9 ... 16 5.5 ... 8

Weight

Mains power connection

Current draw

Mains power connection

Current draw

Temperature selection via keypad remote control via PC

Temperature indication

ATC3

Resolution

Temperature control

230 V/50 Hz V/Hz 207-253 / 50

(at 230 V) A 13

230 V/60 Hz V/ Hz 207-253 / 60

(at 230 V) A 13

INT/EXT kg

°C

°C

49

HE digital indication on VFD COMFORT-DISPLAY indication on monitor

57

207-253 / 50

14

207-253 / 60

14

VFD COMFORT-DISPLAY

0.01

±3 / ±9

ICC - Intelligent Cascade Control

Heater wattage (at 230 V)

Heater wattage (at 115 V)

Electronically adj. pump capacity stages kW kW

2,0

1,0

1 ... 4

Flow rate at 0 bar

Max. pressure at 0 liters

Max. suction at 0 liters l/min 22 ... 26 bar 0.7 bar 0.4

Electrical connections:

See page 17

Ambient temperature °C 5 … 40

All measurements have been carried out at: rated voltage and frequency ambient temperature: 20 °C

Technical changes without prior notification reserved.

16

HE

Electrical connections:

External alarm device 24-0 V DC / max. 25 mA

Computer interface RS232 or RS485

External Pt100 sensor

Optional for HE, SE

(Order No. 8900100 Electronic module with analog connections)

Programmer input -100 °C to 400 °C = 0 - 10 V or 0 - 20 mA or 4 - 20 mA

Input for the signal of a flow meter or external manipulated variable

Temperature recorder outputs 0 - 10 V (0 V = -100 °C, 10 V = 400 °C)

Standby input

0 - 20 mA (0 mA = -100 °C, 20 mA = 400 °C)

4 - 20 mA (4 mA = -100 °C, 20 mA = 400 °C) for external emergency switch-off

Alarm output for external alarm signal

Safety installations according to IEC 61010-2-010:

Excess temperature protection

Low liquid level protection

Classification according to DIN 12876-1

Supplementary safety installations

Early warning system for low level

High temperature warning function

Low temperature warning function adjustable from 0 °C ... 320 °C float switch class III float switch optical + audible (in intervals) optical + audible (in intervals) plausibility control Supervision of working sensor

Reciprocal sensor monitoring between working and safety sensors

Alarm message

Warning message difference >35 K optical + audible (permanent) optical + audible (in intervals)

Environmental conditions according to IEC 61 010-1:

Use indoors only.

Altitude up to 2000 m - normal zero.

Ambient temperature: see Technical specifications

Humidity:

Max. relative humidity 80% for temperatures up to +31 °C, linear decrease down to 50% relative humidity at a temperature of +40 °C

Max. mains voltage fluctuations of ±10% are permissible.

Protection class according to IEC 60 529

The unit corresponds to Class I

IP21

Overvoltage category II

Pollution degree

Caution:

2

The unit is not suitable for use in explosive environment

EMC requirements according to EN 61326-1

This unit is an ISM device classified in Group 1 (using high frequency for internal purposes), Class A

(industrial and commercial range).

17

Operator responsibility – Safety recommendations

2.5. Cooling water connection

Cooling water pressure (IN / OUT ) max. 6 bar

Difference pressure (IN - OUT ) 3.5 to 6 bar

Cooling water temperature

Recommended quality of cooling water:

<20 °C pH – value

Sulfate [SO4 2- ]

Hydrocarbonate [HCO3- ] / Sulphate [SO4 2-]

Hardness [Ca2+, Mg2+] / [HCO3-]

Alkalinity

Conductivity

Chloride (CL-)

Phosphate (PO43-)

Ammonia (NH3)

Free Chlorine

Ferri Ions (Fe3+ )

Mangano Ions (Mn2+)

Carbon dioxide (CO2)

Hydrosulfide (H2S)

Content of oxygen

Algae growth

Suspended solids

Notice:

7,5 to 9,0

< 100 ppm

> 1 ppm

> 0,5 dH

60 ppm < [HCO3-] < 300 ppm

< 500 μs / cm

< 50 ppm

< 2 ppm

< 0,5 ppm

< 0,5 ppm

< 0,5 ppm

< 0,05 ppm

< 10 ppm

< 50 ppm

< 0,1 ppm impermissible impermissible

Danger of corrosion of heat exchanger due to unsuitable quality of cooling water.

• Due to its high content of lime hart water is not suitable for cooling and causes calcination of the heat exchanger.

• Ferrous water or water containing ferrous particles will cause formation of rust even in heat exchangers made of stainless steel.

• Chlorous water will cause pitting corrosion in heat exchangers made of stainless steel.

• Due to its corrosive characteristics distilled and deionized water is unsuitable and will cause corrosion of the bath. .

• Due to its corrosive characteristics sea water is not suitable.

• Due to its microbiological (bacteria) components which settle in the heat exchanger untreated and unpurified river water and water from cooling towers is unsuitable.

• Avoid particulate matter in cooling water.

• Avoid putrid water.

Notice: Cooling water circuit

Risk of oil leaking from the cooling circuit (compressor) of the recirculating cooler into the cooling water in case of a fault in the circuit!

Observe the laws and regulations of the water distribution company valid in the location where the unit is operated.

18

HE

Operating instructions

3. Safety notes for the user

3.1. Explanation of safety notes

In addition to the safety warnings listed, warnings are posted throughout the operating manual. These warnings are designated by an exclamation mark inside an equilateral triangle. “Warning of a dangerous situation (Attention! Please follow the documentation).”

The danger is classified using a signal word.

Read and follow these important instructions for averting dangers.

Warning:

Describes a possibly highly dangerous situation. If these instructions are not followed, serious injury and danger to life could result.

Caution:

Describes a possibly dangerous situation. If this is not avoided, slight or minor injuries could result. A warning of possible property damage may also be contained in the text.

Notice:

Describes a possibly harmful situation. If this is not avoided, the product or anything in its surroundings can be damaged.

3.2. Explanation of other notes

Note!

Draws attention to something special.

Important!

Indicates usage tips and other useful information.

This icon is used in the operating instructions to indicate flashing values or parameters which have to be set or confirmed.

19

Safety notes for the user

3.3. Safety recommendations

Follow the safety instructions to avoid personal injury and property damage. Also, the valid safety instructions for workplaces must be followed.

• Only connect the unit to a power socket with an earthing contact (PE – protective earth)!

• The power supply plug serves as a safe disconnecting device from the line and must always be easily accessible.

• Place the unit on an even surface on a base made of nonflammable material.

• Do not stay in the area below the unit.

• Make sure you read and understand all instructions and safety precautions listed in this manual before installing or operating your unit.

• Set the excess temperature safety installation at least 25 °C below the fire point of the bath fluid.

• Observe the limited working temperature range when using plastic bath tanks.

• Never operate the unit without bath fluid in the bath.

• Pay attention to the thermal expansion of bath oil during heating to avoid overflowing of the fluid.

• Prevent water from entering the hot bath oil.

• Do not drain the bath fluid while it is hot!

Check the temperature of the bath fluid prior to draining (e.g., by switching the unit on for a short moment).

• Use suitable connecting tubing.

• Avoid sharp bends in the tubing, and maintain a sufficient distance from surrounding walls.

• Make sure that the tubing is securely attached.

• Regularly check the tubing for material defects (e.g., for cracks).

• Never operate damaged or leaking units.

• Always turn off the unit and disconnect the mains cable from the power source before performing any service or maintenance procedures, or before moving the unit.

• Always turn off the unit and disconnect the mains cable from the power source before cleaning the unit.

• Always empty the bath before moving the unit.

• Transport the unit with care.

• Sudden jolts or drops may cause damage in the interior of the unit.

• Observe all warning labels.

• Never remove warning labels.

• Never operate units with damaged mains power cables.

• Repairs are to be carried out only by qualified service personnel.

 Some parts of the bath tank and the pump connections may become extremely hot during continuous operation. Therefore, exercise particular caution when touching these parts.

20

HE

Caution:

The temperature controlling i.e. of fluids in a reactor constitutes normal circulator practice.

We do not know which substances are contained within these vessels.

Many substances are:

• inflammable, easily ignited or explosive

• hazardous to health

• environmentally unsafe i.e.: dangerous

The user alone is responsible for the handling of these substances!

The following questions shall help to recognize possible dangers and to reduce the risks to a minimum.

• Are all tubes and electrical cables connected and installed?

Note: sharp edges, hot surfaces in operation, moving machine parts, etc.

• Do dangerous steams or gases arise when heating?

Is an exhaust needed when working?

• What to do when a dangerous substance was spilled on or in the unit?

Before starting to work, obtain information concerning the substance and determine the method of decontamination.

Notice: Check the safety installations at least twice a year!

• Excess temperature protection according to IEC 61010-2-010.

With a screwdriver turn back the adjustable excess temperature protection until the shut-down point (actual temperature).

• Low level protection according to IEC 61010-2-010.

To check the function of the float, it can be manually lowered with a screwdriver for example.

21

Operating controls and functional elements

4. Operating controls and functional elements

Front view Rear view

FPW50

22

1a

1b

2

2.2

2.3

4

4.1

2.1

HE

Mains power switch, illuminated for circulator

Mains power switch, illuminated for cooling machine

VFD COMFORT-DISPLAY

Header: Control indicators

Line 1: Actual value internal or external

The display is depending on the selected control mode in the menu > Control < (internal or external).

Line 2: Working temp. setpoint, constantly S xxx.xx

Line 3: Actual value (E = external or I = internal)

Alternating with the display in line 1

Use the keys to indicate further values in line 3

PI Capacity in % - with manipulated variable set to >control<*

PS Capacity in % - with manipulated variable set to >serial<* or

>eprog<*

H Heater capacity in Watts

U Mains voltage Volts

F Flow rate in liters/minute

(providing EPROG input set to >Flowrate<)

*refer to >MENU/CONFIG<  >CONFIG / ACTVAR>

Control indicators in the header:

Heating / Cooling / Alarm /

R

emote control

Control indicators in the header:

Temperature indication Internal or External actual value

Temperature indication in °C (°F not possible on this unit)

Display of set pump pressure stage

Four stages, can be set via the key , under >MENU - PUMP<.

Navigation keys

1. Key: >OK< Start / Stop (pump / heater )

2. >OK< in the menu Menu item / select submenu for setting

Save set value

Save selected parameter

A beep signals the end of setting

After the actions Start, Stop and change from VFD Display to standard display the key is locked for a short time.

The above graph “front side” shows an example for standard display.

23

Operating controls and functional elements

4.2

4.3

1. Key: >Return< Stop (pump / heater )

2. >Return< in the menu one menu level down

Correction function for parameters or values (prior to OK) immediately back to standard display

- icon for „keep key pressed down“.

1. Key: >Up / Down <temperature – increase/decrease setpoint

Push key quickly for single steps,

Keep key pressed for fast change.

2. >Up/Down< in the menu selection of menu items / parameters

Menu keys

4.4

4.5

4.6

6

Key: start the menu > warning and safety values<

Key: start the menu >temperature setpoints<

Key: display of MENU structure

Adjustable excess temperature protection according to IEC 61010-2-010

7

8

9 ext Pt100

SERIAL

Socket for external measurement and control sensor or external setpoint programming

Interface RS232: remote control via personal computer

Socket: control cable of JULABO refrigerated circulator or output for alarm messages

Option: Electronic module

10

Order No. 8 900 100

Alarm output (for external alarm signal)

ALARM

11 Standby input (for external emergency switch-off)

STAND-BY

12 Programmer input and temperature recorder output

REG+E-PROG

24

16a

16b

17a

17b

18

19

20

21

22

23

24

Mains fuses for circulator, T16A

Mains fuses for cooling machine, T10A

Mains power cable with plug for circulator

Mains power cable with plug cooling machine

Built-in mains outlet for connection of circulator

Drain screw with drain connection

Venting grid, removable

Pump connectors suction pump pressure pump

Selector dial for cooling machine (only F25, F34)

Position "1" for operation with HE circulator.

Cooling water OUTLET and INLET.

¾"

HE

25

Preparations

5. Preparations

5.1. Installation

• Place the unit on an even surface on a pad made of non-

flammable material.

F34: The circulator fitted with a stainless steel bridge is placed on on the back of the bath tank leaving the bath open on the front side.

• The place of installation should be large enough and provide sufficient air ventilation to ensure the room does not warm up excessively because of the heat the instrument radiates to the environment. (Max. permissible ambient temperature: 40 °C).

With regard to a disturbance in the cooling loop (leakage), the guideline EN 378 prescribes a certain room space to be available for each kg of refrigerant.

The necessary amount of refrigerant is specified on the type plate.

> For 0.25 kg of refrigerant R134a, a room space of 1 m

3 is required.

> For 0.52 kg of refrigerant R404A, a room space of 1 m

3 is required.

> For 0.49 kg of refrigerant R507, a room space of 1 m

3 is required.

• Set selector dial for cooling machine in position "1" for operation with HE circulator.

(only F25, F34)

• Keep at least 20 cm of open space on the front and rear venting grid.

• Do not set up the unit in the immediate vicinity of heat sources and do not expose to sun light

• Before operating the unit after transport, wait about one hour after setting it up. This will allow any oil that has accumulated laterally during transport to flow back down thus ensuring maximum cooling performance of the compressor.

Only water cooled models:

Ensure circulation of cooling water by connecting the tubing to cooling water inlet and outlet on the rear of the refrigerated circulator.

• Cooling water connecter G¾"

• Cooling water see page 18

5.2. Bath fluids

Caution:

Carefully read the safety data sheet of the bath fluid used, particularly with regard to the fire point!

If a bath fluid with a fire point of

≤65 °C is used, only supervised operation is possible.

Water: The quality of water depends on local conditions.

• Due to the high concentration of lime, hard water is not suitable for temperature control because it leads to calcification in the bath.

26

HE

• Ferrous water can cause corrosion - even on stainless steel.

• Chloric water can cause pitting corrosion.

• Distilled and deionized water is unsuitable. Their special properties cause corrosion in the bath, even in stainless steel.

Recommended bath fluids:

Bath fluid soft/decalcified water mixture water/glycol, mixture 1:1

Temperature range

5 °C to 80 °C

-20°C to 50°C

JULABO bath fluids

JULABO

Description

Order Number

Temperature range

Flash point

Fire point

Color

JULABO

Description

Order Number

Temperature range

Flash point

Fire point

Color

10 liters

5 liters

°C

°C

°C

10 liters

5 liters

°C

°C

°C

Thermal

G

8 940 124

8 940 125

-30 ... 80

--

-- light yellow

Thermal

H10

8 940 114

8 940 115

-20 ... 180

190

216 clear

Thermal

HY

8 940 104

8 940 105

-80 ... 55

78

80 clear

Thermal

H20S

8 940 108

8 940 109

0 ... 220

230

274 light brown

Thermal

H5

8 940 106

8 940 107

-50 ...105

124

142 clear

See website for list of recommended bath fluids.

ATTENTION: The maximum permissible viscosity is 70 mm

2

/s.

Caution:

Fire or other dangers when using bath fluids that are not recommended:

Use only nonacidic and noncorrosive bath fluids.

JULABO assumes no liability for damage caused by the selection of an unsuitable bath liquid.

Unsuitable bath fluids are fluids which, e.g.,

• are highly viscous

(much higher than recommended at the respective working temperature)

• have a low viscosity and have creep characteristics

• have corrosive characteristics or

• tend to crack.

No liability for use of other bath fluids!

27

Preparations

5.3. Temperature application to external systems

Caution: Securely attach all tubing to prevent slipping.

If the circulator is operated without external system, close the pump connector (22) with the cap nut.

Temperature application to external, closed systems ext

Pt100

T16A

T16A

SERIAL ALARM STAND-BY REG+

E-PROG

The circulator is used for temperature application to external, closed systems (loop circuit) with simultaneous temperature application in the circulator bath.

• Unscrew the M16x1 collar nuts on the pump connectors with a 19 mm (3/4“) wrench and remove the sealing disks. Using the collar nuts, screw on the tubing connection fittings (for tubing 8 mm or 12 mm in diameter) delivered with the unit and tighten firmly. (Pressure pump: 22, suction pump: 21)

• Push on the tubings, and secure with tube clamps.

21 22

• Attach the tubings to the connectors of the external closed system, e.g., an instrument with a pressure-resistant temperature jacket or a temperature coil, and fasten with tube clamps to prevent slipping.

Temperature application to external, open systems

The circulator is equipped with both a pressure and suction pump for external temperature application in open systems.

Differing flow rates of the pressure and suction pumps should be compensated. To maintain a constant liquid level, the JULABO

„D+S“ Level Adapter is recommended for the external bath tank.

The flow rate of the pressure pump will be then regulated by a built-in float device. The liquid level may be changed by a height adjustment on the „D+S“ Level Adapter.

S = Suction pump connection

D = Pressure pump connection

K = Float

H = Height adjustment

Accessory: „D+S“ Level Adapter Order No. 8 970 410

Important:

The liquid level should be equal in the internal and external baths (absolute height).

If you take out samples (for example Erlenmeyer flasks) from the external bath, turn the circulator off with the Start/Stop key.

28

Return flow safety device

If the liquid levels in the circulator bath and the external system are at different heights, overflowing must be prevented after the power has been turned off.

HE

Flood hazard!

For this reason, shut-off valves can be integrated in the loop circuit.

Order No. Description

8 970 456 Shut-off valve (suitable up to +90 °C)

8 970 457 Shut-off valve (suitable up to +200 °C)

5.3.1. Tubing

Recommended tubing:

Order No. Length

8930008

8930012

1 m

1 m

CR

® tubing

CR

®

tubing

8930108

8930112

8930410

1 m

1 m

1 m

Viton tubing

Viton tubing

8 mm inner dia.

12 mm inner dia.

8 mm inner dia.

12 mm inner dia.

Insulation for tubing 8 mm inner dia.

Temperature range

-20 °C to 120 °C

-20 °C to 120 °C

-50 °C to 200 °C

-50 °C to 200 °C

-50 °C to 100 °C

8930412 1 m Insulation for tubing 12 mm inner dia. -50 °C to 100 °C

8 930 209 0.5 m

8 930 210 1.0 m

8 930 211 1.5 m

8 930 214 3.0 m

Metal tubing, triple insulated, M16x1

8 930 220 0.5 m

8 930 221 1.0 m

8 930 222 1.5 m

8 930 223 3.0 m

Metal tubing, insulated, M16x1

Warning: Tubing:

-100 °C to +350 °C

-50 °C to +200 °C

At high working temperatures the tubing used for temperature application and cooling water supply represents a danger source.

A damaged tubing line may cause hot bath fluid to be pumped out within a short time.

This may result in:

• Burning of skin

• Difficulties in breathing due to hot atmosphere

Safety recommendations

• Employ suitable connecting tubing.

• Make sure that the tubing is securely attached.

• Avoid sharp bends in the tubing, and maintain a sufficient distance from surrounding walls.

• Regularly check the tubing for material defects (e.g. for cracks).

• Preventive maintenance: Replace the tubing from time to time.

29

Preparations

5.4. Filling / draining

Notice:

• Pay attention to the thermal expansion of bath oil during heating to avoid overflowing of the liquid.

• Do not drain the bath fluid while it is hot!

Recommendation: Temperature range 5 °C to 40 °C

Check the temperature of the bath fluid prior to draining (by switching the unit on for a short moment, for example).

• Store and dispose the used bath fluid according to the laws for environmental protection.

Filling

Take care that no liquid enters the interior of the circulator.

• Recommended maximum filling level with water as bath fluid:

30 mm below the tank rim

• Recommended maximum filling level with bath oils:

40 mm below the tank rim

After filling, immerse the samples in the bath or place the lid on the bath, in case the opening is not to be used.

The circulator provides an early warning system for low level that may be triggered when changing samples in the bath.

Draining

• Turn off the circulator and cooling machine.

• Hold the venting grid, pull out and remove.

• Slide a short piece of tube onto the drain port (11) and hold it into a pail.

• Unscrew the drain tap (23) and empty the unit completely.

• Tighten the drain tap.

30

HE

6. Operating procedures

6.1. Power connection

Caution:

• Only connect the unit to a power socket with earthing contact (PE – protective earth)!

• The power supply plug serves as safe disconnecting device from the line and must be always easily accessible.

• Never operate equipment with damaged mains power cables.

• Regularly check the mains power cables for material defects (e.g. for cracks).

• We disclaim all liability for damage caused by incorrect line voltages!

Check to make sure that the line voltage matches the supply voltage specified on the identification plate. Deviations of ±10 % are permissible.

• Connect the circulator with mains power cable (17a) to the mains outlet (18).

• Connect the control cable (9) between the connectors  (9a, 9b).

• Connect the refrigerated circulator with mains power cable (17b) to the mains socket.

6.2. Switching on / Start - Stop

6.2.1. Switching on the circulator

Switching on:

• Turn on the mains power switch (1).

The unit performs a self-test.

Then the software version (example: V 1.xx) appears. The display

OFF“ or „R OFF“ indicates the unit is ready to operate.

The circulator enters the operating mode activated before switching the circulator off:

keypad control mode (manual operation) or

remote control mode (operation via personal computer).

Start:

• Press

key.

The actual bath temperature is displayed on the VFD COMFORT-

DISPLAY. The circulating pump starts with a slight delay.

Stop:

Press key. or

Keep key pressed.

The VFD COMFORT-DISPLAY indicates the message "OFF".

31

Operating procedures

6.2.2. Switching on the Cooling Machine

Switching on:

• Switch on the cooling machine using the switch (1b) .

Control of the cooling machine:

With the mains switch (1b) turned on, the circulator automatically switches the cooling machine off and on.

• It is switched off, if:

- the actual working temperature is increased by >30 °C (cooling is not required).

- the heater operates at full power (>800 W) for longer than 5 minutes.

• It is switched on, if:

- cooling is necessary for maintaining the bath temperature.

After switch-off, the cooling machine automatically switches on only after a delay of 5 minutes for protecting the cooling compressor.

To save energy, turn off the cooling machine with the mains switch

(1b) whenever cooling is not required.

32

HE

7. Setting of temperatures

7.1. Using the pre-settings in the

Werkseinstellungen:

SETPNT 1 25 °C

SETPNT 2 37 °C

SETPNT 3 70 °C

Press the

menu

key to call up the menu for temperature selection.

3 different working temperatures can be adjusted. Their values are freely selectable within the operating temperature range.

The temperatures can be set in start or stop mode.

Press key if a value is to be retained

Setting of working temperature in the menu

1. Press the key . The value flashes

2. Select SETPOINT 1 or 2 or 3 using the key or .

3. Confirm by pressing the key.

The circulator uses the new working temperature value for temperature control.

Example: Adjustment/modification of the pre-setting of "SETPOINT 3"

1. Press the key.

2. Select SETPOINT 3 by pressing the

Example: SETPNT 3 / 70.00 °C

key.

3. Keep the

(example: <70>) key pressed until the integer digits flash .

4. Adjust value by pressing the and confirm by pressing the

key and the

key.

key to 85.00 °C

The decimal digits flash and can be adjusted if desired.

Confirm once more by pressing the

Example on the left: SETPNT 3 / 85.00.

key.

If the active setpoint (SETPNT) is changed, the new value is immediately used for the control of the working temperature.

The heater control indicator flashes.

If the other two setpoints (not activated for control) are changed the

MENU has to be left by pressing the key after the decimal digits have been confirmed

Notice: Refer to SETPOINT MAX / MIN in chapter

9.8. MENU LIMITS

33

Setting of temperatures

7.2. Direct setting of temperatures

The circulator uses the setpoint of SETPNT 1 or 2 or 3 for temperature control

The indicated setpoint temperature can be changed directly any time.

Example: change 25.00 °C to 50.00 °C

1. By pressing the key the circulator switches to the active

SETPOINT< example on the left: >SETPNT / 1 25.00°C<.

The integer digits flash (example: <25>).

2. By pressing the keys

50.00 °C and

and the value is changed to is confirmed by pressing the key.

The decimal digits flash and can be adjusted if desired.

Confirm once more by pressing the key.

The circulator uses the new working temperature value for temperature control.

The temperatures can be set in start or stop mode.

34

HE

8. Safety installations, warning functions

SECVAL

(Security Values)

 SAFETMP

 AL-TYPE

 OVERTMP

 SUBTEMP

Check the safety installations at least twice a year! Refer to ( page 21)

Settings for the excess temperature protection > SAFETMP< and for the warning functions for high > OVERTMP< and low >

SUBTEMP< temperature are made in a menu which is called up by pressing the key .

Menu item > AL-TYPE< allows choosing between a warning and an alarm cut-off for the menu items > OVERTMP< and > SUBTEMP<.

8.1. Excess temperature protection

Warning:

The excess temperature protection must be set at least 25 °C below the fire point of the bath fluid used!

In case of wrong setting there is a fire hazard!

We disclaim all liability for damage caused by wrong settings!

This excess temperature protection is independent of the control circuit.

When activated heater and circulating pump are completely shut down.

The alarm is indicated by optical and audible signals (continuous tone) and the error message "ALARM-CODE 14" appears on the VFD

COMFORT-DISPLAY together with the ticker:

> EXCESS TEMPERATURE PROTECTOR ALARM-CHECK

ADJUSTMENT <

Setting range: 20 °C ... 320 °C

Rough setting can be effected by using the temperature scale.

Exact setting:

1. Press the key to display menu >SAFETMP<.

2. Press the key and the set shutdown value is indicated.

3. Set the new shutdown value within 30 seconds using a screwdriver.

The value is indicated on the VFD COMFORT-DISPLAY

Example: SAFETMP / 100 °C

Recommendation:

Set the excess temperature protection at 5 °C to 10 °C above the working temperature setpoint.

35

Safety installations, warning functions

8.1.1. Early warning system, low level protection

This low level protection is independent of the control circuit and is divided into two sections:

1. Switch in stage 1 recognizes a defined fluid level

.

An audible warning sounds (interval tone) and together with the ticker: > LOW LEVEL WARNING-FILL MEDIUM < a message appears on the VFD COMFORT-DISPLAY:

Refill the bath fluid!

2. Switch in stage 2 recognizes a low fluid level

.

If stage 2 of the low level protection according to IEC 61010-2-

010 is triggered, a complete, all-pole shutdown of heater and circulating pump is effected

A continuous alarm sounds and together with the ticker: > LOW

LEVEL ALARM-FILL MEDIUM < a message appears on the

VFD COMFORT-DISPLAY:

Turn off the unit with the mains switch, refill bath fluid and turn the unit on again!

3. Float

4. Circulating pump

5. Heater

Warning:

When adding bath fluid, always us the type of fluid which is identical with the fluid in the bath.

Bath oils must not contain any water and should be pre-heated approximately to the current bath temperature! Explosion hazard at high temperatures!

36

HE

8.2. Switch-over from warning to shutdown function

If a shutdown of functional elements (e.g. heater, circulating pump) is required when the limit values are exceeded or undercut the circulator can be changed over from warning function >WARNING< to shutdown function

>ALARM<.

Factory setting:

>WARNING<

1. Press the key .

2. Select the menu >SECVAL -AL-TYPE< by pressing the

3. Press the key and the set parameter will flash .

(Example: WARNING)

key.

4. Change the parameter by pressing the pressing the or press the

key.

key and confirm by

key if the parameter is to retained.

Setting >WARNING<

A mere warning function with optical and audible warning signal (interval tone) A message appears on the VFD COMFORT-DISPLAY:

or

OVERTMP SUBTEMP

• Setting >ALARM<

Temperature limit with shutdown of heater and circulating pump.

An audible alarm sounds (continuous tone) and a message appears on the VFD COMFORT-DISPLAY:

OVERTMP

or

SUBTEMP

37

Safety installations, warning functions

8.3. Over and Sub temperature warning function

Over temperature

Sub temperature

If the observance of a working temperature value >SETP< has to be supervised for a sensitive temperature application, then set over and sub temperature warning values. In the example below the SETPOINT 85 °C is surrounded by the values OVERTMP 87 °C and SUBTEMP 83 °C. The electronics immediately register if the actual temperature breaches one of the set limit values. The resulting reaction is defined in a further menu item.

(See chapter 8.2. )

1. Press the key .

2. By pressing the

>SUBTEMP<.

1.

Press the

or key select the menu >OVERTMP< or key. The integer digits flash

2.

Change the values to 87. °C and/or 83. °C by pressing the

key and confirm with the key.

The decimal digits flash and can be adjusted if desired.

Confirm once more by pressing the

See above examples.

key.

and

• The warning functions are only activated if the actual bath temperature remains within the set limit values for 3 seconds after switch-on.

Recommendation:

Set the over temperature warning value >OVERTMP< 5 °C to 10 °C above the working temperature setpoint.

Set the sub temperature warning value >SUBTMP< 5 °C to 10 °C below the working temperature setpoint.

38

9.

Menu level 1

Menu functions

1. Open the menu by pressing the

2. Use the

3. Press the

key.

keys to scroll in menu level 1.

key to change to menu level 2.

Press the key if settings are to be retained.

The term „Menu functions“ refers to settings such as

Start program

Administration and creation of programs

Electronically adjustable pump capacity

Configuration of the unit

REMOTE – on / off (remote control via RS232)

AUTOST – AUTOSTART on / off

OFF-MODE – pump on / off

TIME / DATE – setting time and date

RESET – factory settings

Control characteristics and parameters

C-TYPE – Internal or external control

DYNAMICS - internal

Control parameter - XP-, TN-, TV- INTERNAL

Control parameter - XP-, TN-, TV- XPU-, EXTERNAL

Adjustable interface parameters

BAUD RATE, PARITY, HANDSHAKE

Page 40

Page 43

Page 45

Page 46

Page 50

Page 56

ATC - Absolute Temperature Calibration,

Sensor calibration INTERNAL SENSOR,

Sensor calibration EXTERNAL SENSOR

3-point calibration

Page 57

Limitations of temperature and capacity

SETPOINT MAX / MIN - Maximum and minimum setpoint

Page 63

HEAT MAX – Set maximum heating

COOLING MAX – Set maximum cooling

INTERN MAX / MIN – Limitation of the temperature range

BAND HIGH / LOW – Band limit

HE

39

Menu functions

Menu level 1

Analog inputs/outputs

Recorder output – CHANNEL 1, 2, 3

EPROG – External programmer input

EX-STBY - STAND-BY input

ALARM - output

Page 66

9.1. MENU PROGRAM – START

Start-Menu

Menu level 1

This menu will start a previously set program.

Requirements:

1. Create a program. (refer to next chapter)

2. Return to the Start-MENU and confirm the desired setting of each

MENU item with the key

3. Set a start time (>TIME< >DATE< >YEAR<) if the program is to be started by the internal timer.

> STEP< Program start at section 1 … 10

> RUNS < Number of repetitions 1 ... 99

> END< Status at end of program (STDBY/SETPNT)

Standby or last setpoint

> GO < Time of start (NOW/TIMER)

Press the key if a parameter is to be retained.

Correction function for parameters or values (prior to OK)

Level 2 Parameter level

 or

Set program step with

and example: STEP 1

Set number of runs with

and example: 1 run

Set desired parameters with

and .

(STDBY / SETPNT)

Parameter STanDBY: the circulator switches to –

OFF-.

Parameter SETPoiNT: the circulator constantly keeps the temperature at the value of the last step.

40

HE

Level 2 Parameter level

oer

Submenu TIMER

Parameter level

Confirm >NOW< with the key and the program will start immediately

or start at the set time under parameter (TIMER ).

Set time in the example below:

09. August 2009, 11:15 hrs

 set the time for the start of the program in the submenu >TIMER<.

>TIME< hours/minutes (hh:mm), set both values one after the other and confirm

• hours flash, set by pressing

+ minutes flash, set by pressing

+

>DATE< day/months (TT/MM), set both values one after the other and confirm.

• day flashes, set by pressing

+ month flashes, set by pressing

+

>YEAR< year

Set the year with

and .

The program starts at the set time.

Display of time until start:

In line 3 the notice >TIMER< and the set values for „TIME“ and

„DATE/YEAR“ are alternately indicated

Check the correct setting of the internal real time clock if required

(see MENU CONFIG)

41

Menu functions

A

B

C1

C2

D1

The started program

After the start the program will indicate the currently calculated setpoint in line

2

S XX.XX. The value increases within the time period >TSLICE< until the target temperature >SETPNT< of the section is reached.

If the time period in a section is set to „0“, the next section will not begin until the target temperature has been reached.

Use the edit keys to scroll to line 3. The display changes approximately every 4 seconds between the current section (STEP XX) and the

A remaining time of the section

B remaining time of the program

C current bath temperature

I xxx.xx - internal actual value or

E xxx.xx – external actual value

D RUN – the program has started or

PAUSE – the progress of the program has been interrupted by pressing the key. While the time is stopped the temperature will constantly remain at the last calculated setpoint

Continue with the key.

D2

Termination / Interruption of a program

The program can be terminated any time by pressing the key .

In case of power failure the program is interrupted.

The circulator switches to –OFF- .

If the AUTOSTART-function is activated the programmer starts again at the point in time approx. 5 minutes prior to the interruption. However, an uncontrolled change of the bath temperature has occurred.

42

HE

9.2. MENU PROGRAM – creation, administration

Menu level 1

1 program

10 sections

The integrated programmer permits fast and easy programming of setpoint temperature sequences. This temperature sequence is called program. A program is composed of individual sections (STEP). The sections are defined by duration (TSLICE) and target temperature. The target temperature is the setpoint (SETPNT), which is achieved at the end of a section. The programmer calculates the temperature ramp from the difference in time and temperature.

STEP (Nr.)

SETPNT (°C)

TIME

1

100

2

180

4

180

6

75

8

75

(hh.mm) 00:20 00.10 00:20 00:50 00:20

°C

200 Step 1 2 4 6 8

150

100

75

50

25

Graph1 10 20 30 40 50 60 10 20 30 40 50 60 10 [min] t

Sections without set value and time are skipped. They can be defined retroactively and the integrated into the program.

43

Menu functions

Menu level 1

Level 2

>EDIT< Create, administer program

> STEP<

>SETPNT <

>TSLICE<

Program step (1 ... 10)

Temperature setpoint of step ...

Duration of step ...

> DELETE< delete program step (01 … 10, ALL)

Press key, if a parameter is to be retained.

Correction function for parameters or values (prior to OK)

Level 3 Parameter level

 (STEP 1 … 10)

Set program step with

and

 (Example: EDIT STEP 01)

For STEP 01 the values for SETPOINT 01 and

TSLICE 01 are set one after the other

 (values within working temp. range)

Integer digits flash, set by pressing

+

Decimal digits flash, set by pressing

+

(time slice)

Set duration by pressing and

Delete program

Program steps can be deleted individually or entirely. (STEP 01, 02,… 10, ALL).

Set parameters by pressing and

44

HE

9.3. MENU PUMP - Setting of pump pressure

The pressure of the circulating pump is adjustable in four stages. After setting, the VFD COMFORT-DISPLAY indicates the corresponding value.

Adjustable pump capacity stage 1 ... 4

Examples:

Soll Ist

Factory setting: stage 1

Illuminated display: for pump pressure

Adjusted:Display for the adjusted pump pressure stage in the –OFF- mode.

Effective: Display for the effective pump pressure stage (rotation speed) after start.

For protecting the pump motor, the rotation speed (i.e. the pump pressure stage) changes with the load applied.

Example: The viscosity of the bath fluid (i.e. the load applied to the pump motor) changes with the working temperature in the bath.

Flow rate: 22 ... 26 l/min

Pump capacity stage 1

Pump pressure

Suction pump

[bar]

[bar]

0.4

0.2

2

0.5

0.26

3

0.6

0.33

4

0.7

0.4

Total capacity in a loop circuit

1. Press the

[bar] 0.6

key.

0.76 0.93 1.1

2. Select the menu >PUMP< pressing the key and confirm by pressing the key

The set parameter flashes (example: >LEVEL 2<)

3. Change the parameter by pressing

key. or

Press the

and confirm by pressing the

key if the parameter is to be retained.

45

Menu functions

9.4. MENU CONFIG - Configuration of unit

Menu level 1

Level 2

A RESET can be effected only in the >OFF< mode.

Switch off circulator by pressing the

CONFIGURATION.

key and call up the menu

Parameter level or

 or or

Press the key if a parameter is to be retained. Correction function for parameters and values (prior to OK).

Switch on and off remote control by pressing

and

Control display in the topline

for Remote

For remote control refer to 79

Connect RS232 with PC.

Switch over setpoint setting by pressing

and

OFF –Setpoint setting with the navigation keys or

Setpoint setting via the analog socket „ext. Pt100 or analog socket >REG+EPROG<

 or

 or

 or or

Switch on and off autostart by pressing

and

AUTOSTART on = on

AUTOSTART off = off

See WARNING page 48

Switch on and off OFFMODE by pressing

and

PUMP ON continuous operation of circulating pump

PUMP OFF circulating pump is linked to Start/Stop

Switch over the input variable by pressing

and

Programming of variables for the parameters >

SERIAL < or > EPROG < is only accepted, if the unit is in Start mode

46

HE

Level 2 Level 3 Parameter level

Hours flash, set by pressing

Minutes flash, set by pressing

+

+

Day flashes, set by pressing

Month flashes, set by pressing

+

+

Year flashes, set by pressing

+

9.4.1. Remote control via the serial interface

Return to factory settings by pressing

RESET returns all set values to the factory setting except for date and time.

A RESET can be effected only in the –OFF- mode.

During the message –RUN- all parameters are reset to factory settings.

Factory setting: OFF

RS232

The control electronics offer two ways of adjusting a setpoint.

1. Adjustment of setpoint using the keypad or the integrated programmer.

2. Adjustment of setpoint via the serial interface RS232 using a PC or a superordinated process control system.

The topline of the VFD-DISPLAY shows a bright „R“ for remote control; - remote control discontinued.

IMPORTANT: additional measures for remote control

Connect the circulator to the PC using an interface cable.

Check the interface parameters of both interfaces (circulator and PC) and make sure they match.

(refer to 12.1. Setup for remote controll page 79)

47

Menu functions

9.4.2. Keypad control or setpoint setting via the analog input

Factory setting:: OFF

The selected mode is indicated on the VFD

COMFORT-DISPLAY

>

OFF

PT100

EPROG

>

9.4.3. AUTOSTART

In addition to the serial interface via remote control the circulator offers the possibility to adjust the setpoint via analog interface >ext. Pt100< or

>REG+E-PROG<.

OFF - Setpoint setting with the navigation keys

or the integrated programmer.

PT100 - Setpoint setting via the analog socket „ext. Pt100“ using an external temperature sensor or an appropriate voltage/current source.

EPROG - Can only be adjusted when an electronic module with analog connections is used (option).

Setpoint setting via the analog interface REG+E-PROG connection with an external voltage or current source or a programmer.

Important:

Connect the external voltage or current source or a programmer to

the circulator via the socket REG+E-PROG (see page 70).

In the menu >MENU IN/OUT< set the parameter >EPROG-

INPUT< and the input variables >EPROG-SIGNAL< (see page 70).

The E-Prog input can only be used either under menu item >SP

EXT < or under menu item >ACTVAR<

(see page 49).

Warning :

For supervised or unsupervised operation with the “AUTOSTART“ function avoid any hazardous situation to persons or property

Take care to fully observe the safety and warning functions of the circulator.

Factory settings: OFF

Notice:

The circulator has been configured and delivered by JULABO in accordance with the NAMUR recommendations. This means for the start mode that the unit must enter a safe operating status after a power failure. This safe operating status is indicated by the message „OFF“ or

R OFF“ on the VFD COMFORT-DISPLAY.

A complete, all-pole shutdown of the main functional elements such as heater and pump motor is effected.

The values set on the circulator remain saved and the unit is restarted by pressing the start/stop key in manual control.

In remote control mode the values need to be resent by the PC via the interface.

If such a safety standard is not required, the NAMUR recommendations can be bypassed with the AUTOSTART function thus allowing a direct start of the circulator by pressing the mains switch or using a timer.

48

HE

9.4.4. OFF-MODE

Factory setting:

PMP OFF

Usually the circulating pump is controlled with the key

or the start/stop command. If the circulating pump is to work in the –OFF- mode, the adjustment can be set in a sub-menu.

The pump motor will be shutdown in case of alarm anyhow.

9.4.5. ACTVAR - actuating variable

Factory setting:

CONTROL

The variable (ACTuating VARiable) corresponds to the extent to which the heater or cooling unit of the circulator is controlled. Heat or cold is applied to the bath according to this variable. If this happens with the control electronics of the circulator, called > CONTROL < in this particular case, the bath temperature is exactly heated and maintained constant at the adjusted setpoint.

Programming of variables for the parameters > SERIAL < or >

EPROG < is only accepted, if the unit is in Start mode.

Possible parameters:

CONTROL – The internal control electronics of the circulator controls the heater und the connected cooling unit. Self-tuning is possible.

SERIAL – The heater or the connected cooling unit receives the control signal via the serial interface. Self-tuning is not possible.

EPROG - The heater or the connected cooling unit receives the control signal via the E-Prog input. Self-tuning is not possible.

Important:

Under set the input variable >EPROG U/I< to

(refer to page 70).

Note:

The E-Prog input can only be used either under menu item

>SP EXT< (refer to page 48) or under menu item > ACTVAR <.

Warning:

The working temperature range of the circulator is determined during configuration. If set to >CONTROL<, this range cannot be exceeded.

If set to > SERIAL < and > EPROG <, heat or cold is applied to the bath without control. The permissible maximum temperature can be exceeded. The user has to take adequate precautions for temperature control.

Materials, such as gaskets or insulations for example, may be damaged or destroyed, if the permissible maximum temperature is exceeded.

The safety and warning functions > their fullest capacity.

< of the instrument must always be used to

49

Menu functions

9.4.6. Setting of clock and date

The internal real time clock allows starting a program any time. The clock is set to the local mean time (MEZ) at the factory.

If the unit is operated in a different time zone, the clock can be adjusted in this menu.

Change summer/winter time in this menu

9.4.7. RESET – Factory settings

A Reset will return all values to factory setting except for date and time.

A RESET can be effected in the >OFF< mode only.

Switch off the circulator by pressing the key menu CONFIGURATION.

and call up the

9.5. MENU CONTROL – Control characteristics and parameters

Menu level 1

The circulator is qualified for internal and external temperature control

The switchover is carried out in the menu >C-TYPE< .(INT or EXT).

For external temperature control and measurement connect a Pt100 external sensor to the socket at the rear of the circulator.

Press the key if a parameter is to be retained. Correction function for parameters or values (prior to OK)

Level 2 Parameter level

 or

 or or

The parameter flashes, switch by pressing

and

The control type can be adjusted in the -OFF- mode only.

Depending on the adjustment only the active parameters are displayed.

The parameter flashes, switch by pressing

and

OFF - no selftuning.

ONCE - single selftuning (factory setting)

ALWAYS - continual selftuning.

50

Level 2

C-TYPE INTERNAL

Parameter level

 or

0.1 … 99.9

3 … 9999

C-TYPE EXTERNAL

0 … 999

0.1 … 99.9

3 … 9999

0 … 999

0.1 … 99.9

HE

The parameter flashes, switch by pressing

and

This parameter affects the temperature sequence in case of internal control.

The parameter flashes, set by pressing

+

The parameter flashes, set by pressing

+

The parameter flashes, set by pressing

+

The parameter flashes, set by pressing

+

The parameter flashes, set by pressing

+

The parameter flashes, set by pressing

+

The parameter flashes, set by pressing

+

51

Menu functions

9.5.1. CONTROL – Control INTERNAL / EXTERNAL

Pt100

Switchover can only be effected if a Pt100 external sensor is connected.

Factory setting: INT

IMPORTANT: Additional measures for external temperature control

Suggested settings for external temperature control:

BAND HIGH / LOW and INTERN MAX / MIN see chapter >MENU LIMITS<.

Sensor calibration of the Pt100 external sensor is carried out in the menu >ADJUST<, submenu >ATC SENOR - EXT<; set ATC

STATUS< to >OFF<

(See page 57).

Attention:

Place the external sensor into the temperature-controlled medium and securely fix the sensor.

Pt100 M+R

Accessory: Pt100 external sensor

Order No. Description Material Cable

8981003

8981005

8981006

8981010

8981015

8981013

8981016

8981014

200x6 mm Ø,

200x6 mm Ø,

20x2 mm Ø,

300x6 mm Ø,

300x6 mm Ø,

600x6 mm Ø,

900x6 mm Ø,

1200x6 mm Ø, stainless steel glass stainless steel stainless steel stainless steel / PTFE coated stainless steel / PTFE coated stainless steel / PTFE coated stainless steel / PTFE coated

8981103 Extension cable for Pt100 sensor

8981020 M+R in-line Pt100 sensor

1.5 m

1.5 m

1.5 m

1.5 m

3 m

3 m

3 m

3 m

3.5 m

The M+R in-line Pt100 sensor is a flow sensor and can be installed loop circuit

52

HE

9.5.2. SELFTUNING

Selftuning:

When performing a selftuning for the controlled system (temperature application system), the control parameters Xp, Tn and Tv are automatically determined and stored.

Possible parameters:

9.5.3. Dynamic internal

OFF - no selftuning

The control parameters ascertained during the last identification are used for control purposes.

ONCE - single selftuning (factory setting)

The instrument performs a single selftuning of the controlled system after each start with the command via the interface.

ALWAYS - continual selftuning

key or after receiving a start

The instrument performs a selftuning of the controlled system whenever a new setpoint is to be reached.

Use this setting only when the temperature application system changes permanently.

°C

°C

NORM temp. stability

APER

Setpoint

NORM temp. stability

APER temperature ramp t

This parameter affects the temperature sequence only in case of internal control.

Factory setting: APER (aperiodic)

Possible parameters:

NORM Allows for reaching the setpoint faster – with setpoint change or ramp function – but overshooting of up to 5 % is possible.

APER Ramp function: the increase of temperature occurs temporally offset and achieves the target temperature without overshooting.

Setpoint change: The temperature increases at the same rate, the target temperature is achieved without overshooting.

With both settings constant temperature is achieved after approximately the same time. t

53

54

Menu functions

9.5.4. Control parameters– XP-, TN-, TV- INTERNAL

Setting range: 0.1 ... 99.9

In most cases the control parameters preset in the factory are adequate for achieving an optimum temperature sequence.

The control parameters allow adjustment to special control processes..

Proportional range >Xp<

The proportional range is the range below the setpoint in which the control circuit reduces the heating capacity from 100% to 0 %

Setting range: 3 ...9999

Reset time >Tn< (Integral component)

Compensation of the remaining control deviation due to proportional regulation. An insufficient reset time may cause instabilities. Excessive reset times will result in unnecessary prolongation of compensation of the control difference.

Lead time >Tv< (Differential component)

Setting range: 0 ... 999

The differential component reduces the transient time. An insufficient lead time will prolong the time required for compensation of disturbance effects and cause high overshooting during run-up. An excessive lead time could cause instabilities (oscillations)

Optimum setting

Optimization instructions for the PID control parameters

Control parameters XP-, TN-, TV- INTERNAL as well as

-EXTERNAL

The heat-up curve reveals possible faulty settings of the control parameter.

Inappropriate settings may produce the following heat-up curves:

Xp too low Tv/Tn too low

Xp too high or Tv too high Tv/Tn too high or Xp too high

HE

9.5.5. COSPEED - external

°C

This parameter affects the temperature pattern only in case of external control.

Possible parameters: 0.0 ... 5.0

During selftuning, the control parameters Xp, Tn and Tv of a controlled system are automatically determined and stored. Depending on the controlled system, time for tuning can be unequally longer. This controller layout allows protection of sensitive objects requiring temperature application.

Int. temperature

Ext. temperature

5

3

0

Setpoint

As soon as a co-speed factor is set, it is considered for calculating the control parameters. As shown in the diagram, tuning times become shorter the higher the co-speed factor is, but overshooting can happen in the internal system. t

9.5.6. Control parameters – XPU-, XP-, TN-, TV- EXTERNAL

Setting range: 0.1 ...99.9

In most cases the control parameters preset in the factors are adequate for achieving an optimum temperature sequence.

The control parameters allow adjustment to special control processes.

Setting range: 3 ...9999

Setting range: 0 ... 999

Setting range: 0.1 ... 99.9

Proportional range >Xpu<

The proportional range Xpu of the cascaded controller is only needed for external control.

55

Menu functions

9.6. MENU SERIAL - BAUDRATE, HANDSHAKE, PARITY

Menu level 1

Level 2

For communication between circulator and a PC or a superordinated process control system the interface parameters of both units must be identical.

For remote control refer to page 79

Factory settings:

4800 Baud even hardware handshake

Parameter level 

Press the retained.

key if a parameter is to be

The parameter flashes, switch by pressing

and

The parameter flashes, switch by pressing

and even: Data bits = 7; Stop bits = 1 odd: Data bits = 7; Stop bits = 1 no: Data bits = 8; Stop bits = 1

The parameter flashes, switch by pressing

and

Xon/Xoff-protocol

Protocol RTS/CTS

(Software handshake)

(Hardware handshake)

56

HE

9.7. MENU ATC - Absolut Temperature Calibration

Example:

1-point calibration

°C

ATC serves to compensate a temperature difference that might occur between circulator and a defined measuring point in the bath tank because of physical properties.

Principle:

For ATC calibration, in steady state the bath temperature at the location of the temperature sensor (CT) is determined at the respective adjusted working temperature. This value is then set on the circulator in the menu

>ATCalibration< under menu item > CTEMP X <.

This can be a 1-point, 2-point or 3-point calibration.

M B (INT) T

°C

3-point calibration

°C

°C

B (EXT)

TT

T

T

T

1 = Original curve

°C

Pt100 ext

°C

TT

M

M = Temperature measuring instrument with temperature sensor

B = Bath tank (INTernal or EXTernal)

T = circulator

CT = Temperature on measuring point

TT = Temperature on circulator

57

Menu functions

Menu level 1

Level 2

Level 2

Parameter level or

Press the key if parameter is to be retained.

Correction function for parameters or values (prior to OK).

The parameter flashes, switch by pressing

and

On level 2 a (I) is indicated for internal or an (E) for external.

 or

Example:

The parameter flashes, switch by pressing

and

>NO< Carry out an ATC calibration

>YES< return to standard operation after

calibration.

The parameter flashes, switch by pressing

and

A >1-point<, >2-point< or >3-point< calibration can be carried out.

The selected calibration is indicated on level 2 by 1 or

2 or 3.

The value >TMPVAL< is only indicated

In addition the measured temperature value

>CALVAL X< is saved during the next step.

Integer digits flash, set by pressing

+

Decimal digits flash, set by pressing

+

If only a 1-point calibration is carried out, the following menu items are not indicated anymore

Parameter level

The value is only indicated

58

HE

Level 2

Parameter level

Integer digits flash, set by pressing

+

Decimal digits flash, set by pressing

+

If only a 2-point calibration is carried out, the following menu items are not indicated anymore

The value is only indicated

Integer digits flash, set by pressing

+

Decimal digits flash, set by pressing

+

9.7.1. ATC SENSOR - INTERNAL / EXTERNAL

In the first submenu the ATC function is set for the >INTERN< internal or the

>EXTERN< external temperature sensor.

Calibration can be carried out for the internal temperature sensor and for the external temperature sensor connected to the socket „ext. Pt100“.

The circulator is able to save both parameter sets. However only the one which has been set under menu item >ATC SENSOR < is displayed.

9.7.2. ATC STATUS - YES / NO

In the second submenu the ATC function for the temperature sensor selected above is activated >YES< or deactivated >NO<.

>YES< (factory setting) The controller of the circulator uses the original curve of the temperature sensor or the new curve measured during the

ATC calibration.

Important: Set to >NO< during the calibration process

>NO< An ATC calibration is to be carried out.

Important: Set to >YES< after calibration.

In the > ATC STATUS < >YES< the ATC calibration always affects the current working temperature; also the one set via interface.

59

Menu functions

9.7.3. CALIBRATION TYPE: 1 -/ 2 -/ 3 POINT

A >1-point<, >2-point< or >3-point< calibration can be carried out.

First geometrically define the location for calibration (measuring point CT), then determine the temperature values of the calibration points.

The type of calibrations also determines the number of the following pairs of values indicated on the LCD DIALOG-DISPLAY.

Pairs of values:

TMPVAL X: Circulator temperature 1 or 2 or 3 (actual value TT )

The actual temperature of the bath is simultaneously saved with the

“calibration value“ >CALVAL< and can be indicated for control purposes

(value does not flash).

CALVAL X: Calibration temperature 1 or 2 or 3 (actual value CT )

The „calibration value“ is determined with a temperature measuring device and saved under menu item >CALVAL<.

(value flashes )

60

9.7.4. Example: 3-point calibration for internal control

Menu level 1

In the temperature range from 80 °C to 160 °C the calibration curve of the temperature sensor (TT) is to be adjusted to the actual temperatures at measuring point (CT).

1. Set circulator to internal control:

MENU CONTROL page 50

The type of control can be set only in the –OFF- mode.

HE

80.00 °C

120.00 °C

160.00 °C

Menu level 1

2. Set working temperature setpoint – SETPNT:

Refer to „Direct temperature setting “ page 34

• By pressing the key

the circulator switches to the active

>SETPOINT< see example on the left: >SETPNT / 1 25.00°C<.

The integer digits flash (Example: <25>).

• Change the value to 80.00 °C by pressing the keys

and confirm by pressing the key

The decimal digits flash.

.

Confirm once more by pressing the key .

• The bath is heated up.

Wait for approx. 5 minutes until the temperature is constant.

3. Reading of temperature measuring device

and

Read the value of measuring point CT on the device and enter under menu item >CALVAL X< by using the keypad.

>CALVAL 1< (79.70 °C)

>CALVAL 2< (119.50 °C)

>CALVAL 3< (159.30 °C)

4. Calibration

Press the key if parameter is to be retained. Correction function for parameters or values (prior to OK).

Setting is required only for the first calibration point.

Level 2 Parameter level

Set SENSOR INTERN:

• The parameter flashes, switch by pressing

and .

An ATC calibration is to be carried out.

Set to >NO<

• The parameter flashes, switch by pressing

and .

61

Menu functions

• The parameter flashes, switch by pressing

and .

A >3-point< calibration is carried out.

The value >TMPVAL< is only indicated

In addition the measured value >CALVAL X< is saved during the following step

• Integer digits flash, set by pressing

(79) +

• Decimal digits flash, set by pressing

(70) +

The first of 3 points is calibrated.

Return to 2. Set working temperature value SETPNT: 120.00 °C

The value is only indicated

• Integer digits flash, set by pressing

(119) +

• Decimal digits flash, set by pressing

(50) +

The second of 3 points is calibrated.

Return to 2. set working temperature value SETPNT: 160.00 °C

The value is only indicated.

• Integer digits flash, set by pressing

(159) +

• Decimal digits flash, set by pressing

(30) +

The 3-point calibration is completed

5. Return to standard operation

• Set >YES< after calibration.

(Standard operation)

62

HE

9.8. MENU LIMITS

Menu level 1

Level 2

Parameter level

0 … 100 %

Press the key if parameter is to be retained.

Correction function for parameters or values (prior to

OK).

• Integer digits flash, set by pressing

+

• Decimal digits flash, set by pressing

+

• Integer digits flash, set by pressing

+

• Decimal digits flash, set by pressing

+

• The value flashes, set by pressing

+

0 … 100 %

• The value flashes, set by pressing mit

+

In case of external control these menu items are additionally indicated.

• Integer digits flash, set by pressing

+

• Decimal digits flash, set by pressing

+

• Integer digits flash, set by pressing

+

• Decimal digits flash, set by pressing

+

• The value flashes, set by pressing

+

• The value flashes, set by pressing

+

63

Menu functions

9.8.1. Limits for internal control

SETPOINT MAX / MIN – Maximum and minimum setpoint

Restriction of the adjustable temperature range.

The limitation of the operating temperature range effects the temperature setting in the menu with the key .

Only setting of working temperatures which lie within the determined limits is possible

Existing settings for SETPNT 1, -2, -3, as well as those for >OVERTMP<

and > SUBTMP < (refer to page 35), are automatically deferred into the

limit range.

Setting range: -94,90 °C ... +300,0 °C

SET MAX > SET MIN

Interchange of values is not possible.

Set maximum heating / cooling

The heating and cooling capacity of the unit are adjustable. 100 % corresponds to the technical specification of the equipment.

Setting range:

HEAT MAX – 0 to 100 % in 1 % steps

COOLING MAX – 0 to 100 % in 1 % steps

9.8.2. Limits for external control

> SAFETMP <

INTERN MAX / MIN

Restriction for the temperature range of the internal bath.

Setting range: -94,9 °C ... +300,0 °C

The limits INT MAX and INT MIN are only active in external control. INT

MAX and INT MIN determine fixed limits for the temperature within the internal bath. The temperature controller cannot exceed these limits even if it would be necessary for achieving the temperature in an external system. Therefore it is possible that the external setpoint cannot be achieved.

Sense of limit setting:

Protects the bath fluid from overheating.

Prevents an undesired alarm shutdown by the excess temperature protection - >ALARM CODE 14<.

Set the value of > INT MAX at least 5 °C below the value of

>SAFETMP<.

Protects the pump motor from high viscosity of the bath fluid at low temperatures.

For refrigerated circulators. Freezing protection when using water as bath fluid.

64

°C

200

150

100

75

50

25

10 20 30 40 50 60 10 20 30 40 50 60 10 [min] t

HE

BAND HIGH

BAND HIGH / LOW – Band limitation

The band limitation is active during external control. Varied, practiceoriented setting are feasible for heat-up and cool-down phases.

Setting range: 0 °C ... 200 °C

BAND HIGH and BAND LOW allow for the limitation of the difference between the temperatures in the internal bath and the external system to any maximum value for the heat-up and cool-down phase.

During the heat-up phase this difference value is always added to the actual external temperature. During the cool-down phase the difference value is subtracted.

Internal bath external system

Sense of a band limitation:

Protection of objects and samples by gentle temperature control

Protection of e.g. glass reactors from thermal shock.

BAND LOW

65

Menu functions

9.9. MENU IN/OUT – Analog inputs/outputs (Option)

In order to use the analog inputs and outputs, the circulator must be equipped with the electronic module available as option.

Order No. 8 900 100 Electronic module

ALARM STAND-BY REG+

E-PROG

This submenu enables setting of the input and output values for the programmer input and the temperature recorder outputs of socket REG+E-

PROG.

The >STAND-BY< input and the >ALARM< output are configurable.

Menu level 1

Press the key if parameter is to be retained. Correction function for parameters or values (prior to OK).

Level 2 Level 3 Parameter level

 

The parameter flashes, switch by pressing

and

ACTINT [°C] internal actual temperature value

(bath temperature)

ACTEXT [°C] external actual temperature value (external sensor)

S-POINT [°C] active setpoint temperature

POWER [%]

• Integer digits flash, set by pressing

+

• Decimal digits flash, set by pressing

+

 

The parameter flashes, switch by pressing

and

66

Level 2

Level 3

Parameter level

HE

• Integer digits flash, set by pressing

+

• Decimal digits flash, set by pressing

+

The parameter flashes, switch by pressing

and

• Integer digits flash, set by pressing

+

• Decimal digits flash, set by pressing

+

The parameter flashes, switch by pressing

and

The parameter flashes, switch by pressing

and

 or

The parameter flashes, switch by pressing

and

Define the indication of

or

VOLTAGE CURRENT

67

Menu functions

Level 2

 or

Level 3 Parameter level

Lower VALue

Upper VALue

• Integer digits flash, set by pressing

+

• Decimal digits flash, set by pressing

+

The parameter flashes, switch by pressing

and

The parameter flashes, switch by pressing

and

The parameter flashes, switch by pressing

and

68

HE

9.9.1. Outputs of the connector - REG+E-PROG

CHANNEL 3

EPROG

CHANNEL 2

REG+E-PROG :

1.

Select CHANNEL 1, 2 or 3

2.

First define the desired output value for CHANNELs 1 to 3:

CHANNEL 1

internal actual temperature value (bath temperature))

REG+E-PROG

external actual temperature value (external sensor)

1.

periodic or intermittent heating or cooling

2.

3.

active setpoint temperature

(SETPoint 1, 2, 3,/ integr. programmer /external programmer)

3.

Then select the display size for CHANNELs 1 to 3:

Voltage outputs CHANNELs 1 and 2

Assign the voltage values of 0 V to the lowest and 10 V to the highest necessary temperature (°C) or power rating (%) required as an output value.

CHANNEL 3: Output for temperature value (°C) / power (%)

The current output (channel 3) offers 2 ranges for selection:

0 mA to 20 mA or

4 mA to 20 mA

Assign the current values 0 mA or 4 mA to the lowest and 20 mA to the highest temperature (°C) or power rating (%) required as an output value.

or

0 V

10 °C

197 °C

0 mA

4 mA

∆ Τ = 200 K

210 °C

202 °C

∆ Τ = 5 K

20 mA

Example 1:

lowest temperature value: 10 °C

highest temperature value 210 °C

Fig. shows 200

°C scaled to paper width rise: 50 mV/°C

Example 2: lowest temperature value: 197 °C highest temperature value: 202 °C

Fig. shows 5

°C scaled to paper width rise: 2000 mV/°C

69

Menu functions

9.9.2. Input of the connector - REG+E-PROG

CHANNEL 3

EPROG

CHANNEL 2

CHANNEL 1

Setting needs to be carried out, if

1.

setpoint programming is to be made via an external voltage or current source or programmer.

For this, in the menu > MENU / CONFIG < first set the menu item > SETP < to >EPROG<.

REG+E-PROG

> >

2.

the heater variable should be controlled via an external control pulse.

For this, in the menu > MENU / CONFIG < set the menu item >ACTVAR< to >EPROG<.

>FLOWRAT<

Indication on display

> >

(refer to page 49).

3.

the signal of an external flow meter should be registered.

>FLOWRAT< see below.

FLOWRAT

The E-Prog input can only be used either under menu item >CONFIG

-SP EXT< < or under menu item >CONFIG ACTVAR<

If the input is neither occupied by > SP EXT < or > ACTVAR <, the signal of a flow meter can be connected.

First set >MENU CONFIG<

> then >MENU IN/OUT<

>

Beispiel:

>

Settings on input - E-PROG

>

1.

Connect the external voltage or current source or programmer to socket REG+E-PROG of the circulator.

2.

Selecting the signal:

The programmer (E-PROG) input of the circulator can be matched to the output signal of the external voltage or current source. voltage input = U  current input = I

3.

Define the input variable:

Setpoint programmed by external voltage or current

Manipulated (ACTuated) variable for the heater with an external control pulse

Signal of an external flow sensor

70

HE

Examples :

4.

– Setting the LOW value: (See below

)

First adjust and set the lowest voltage or current on the external voltage or current source (e.g. 0 V or 0 mA).

Then after approx. 30 secs enter the corresponding temperature value (e.g. 20.00 °C).on the circulator by pressing

and

5.

– Setting the HIGH value: (See below

)

First adjust and set the highest voltage or current on the external voltage or current source (e.g. 10 V or 20 mA).

Then after approx. 30 secs enter the corresponding temperature value (e.g. 300 °C).on the circulator by pressing

and

Example:

Set the external voltage or current source output for the equivalent of

50 °C temperature setpoint.

Press - to standard display.

The temperature value adjusted and set on the external voltage or current source is displayed in line 2 of the VFD COMFORT-DISPLAY for control purposes. (Example: SP 50.00 °C).

200

150

100

80

60

40

20

°C

300

250

B

A

This EPROG input enables the use of different voltage and current values as program parameters.

– Setting the LOW value: (See below

)

1.

Adjust and set the lowest desired value on the voltage or current source resp. (Example A: 1 V).

Wait appr. 30 seconds.

2.

Assign a lower temperature threshold value to this adjusted voltage/current value by pressing the appropriate buttons on the keypad of the instrument (Example A: 20

°C ) and set by pressing

.

1 2 3 4 5 6 7 8 9 10 V

2 4 6 8 10 14 18 20 mA

– Setting the Upper value: (See below

)

1. Adjust and set the highest desired value on the voltage or current source resp. (Example A: 10 V).

Wait appr. 30 seconds.

2. Assign an upper temperature threshold value to this adjusted voltage/current value by pressing the appropriate buttons on the keypad of the instrument (Example A: 200

°C) and set by pressing .

Example B in the diagram serves to illustrate that the end point values are freely selectable (Ex: 8 mA and 16 mA).

71

Menu functions

Example out of diagram A:

• Adjusting the voltage source for an output of 7.6 V!

The instrument calculates this value from the rise angle of the two predecided end points (in example A: 7.6 V correspond to an external setpoint temperature of 152.0

°C ).

After returning the VFD display to standard display, this value is displayed in line 2 (Example: SP 152.00

°C).

Notice:

If this adjustment is not correctly performed at two different points, the setpoint setting will be incorrect.

 Important:

The usable temperature range between > LVAL < and > UVAL < is limited to the configured working temperature range of the circulator resp. of the unit combination

(working temperature range see technical specifications)

9.9.3. ALARM-output / Stand-by-input

Alarm output (10)

(for external alarm signal)

Possible parameters:

ALARM

(STANDBY or ALARM or AL-STBY)

This socket is a potential-free change-over contact. With the adjustments in the menu item > AL-OUT Function< all operating conditions can be signaled without having to change the pin assignments.

Signification of the terms under menu item >Function< :

The circulator is in condition >STANDBY< or >ALARM< or >AL-STBY< - both conditions are signaled.

Alarm-TYPE

Possible parameters:

( NORMAL or INVERS )

For >TYPE- NORMAL< pins 2 and 3 are connected in any case according to the selected > AL-OUT Function<

For >TYPE- INVERS < pins 2 and 1 are connected in any case according to the selected > AL-OUT Function<

Switching capacity max. 30 W / 40 VA

Switching voltage max. 125 V

∼/−

Switching current max. 1 A

72

HE

EX-STBY: External Stand-by input

(for external switch-off)

AK

1

AK

2

1

STAND-BY

3

2

3

STAND-BY

Possible parameters:

INACTIV - standby input is ignored

ACTIV - standby input is active

Activate the standby input:

1. Under menu item > EXT-STBY <, set the parameter to >ACTIV<.

2. Connect an external contact ‚AK‘ (e.g. for switch-off) or an alarm contact of the superordinated system.

In case the connection between Pin 2 and Pin 3 is interrupted by opening the contact ‚AK‘, a complete shutdown of the circulating pump and heater is effected, and the unit enters the condition „E OFF“.

If the contact is reclosed, the instrument returns to the standby state and

E OFF“ is displayed. Press to start.

Additional tips for using the STANDBY input:

The standby function can be used in conjunction with the AUTOSTART feature

1. If the Autostart function is NOT turned ON, the standby input is used as described above.

2. If the Autostart function is enabled, the instrument will revert back to the original method of entering the setpoint (i.e. keypad, RS232,

Analog input, etc.).

Entering the setpoint with the keypad. e.g.

As described above, a bipolar shutdown is accompanied by displaying the „E OFF“ state. The circulator starts again when the contact is reclosed. The temperature of the bath fluid changed during the „E OFF“ state.

Entering the setpoint with the programmer.

The display E OFF appears. The setpoint value and the time are both held at the current value. The temperature of the bath fluid will be held constant at this temperature. The programmer continues once the contact is reclosed.

Notice: This is not an actual shutoff feature.

73

Troubleshooting guide / error messages

10. Troubleshooting guide / error messages

Alarm with complete shutdown:

If one of the following failures occur a complete, all-pole shutdown of the heater and circulating pump is effected.

+

„ “ lights up and a continuous signal sounds.

The code for the cause of alarm is indicated on the VFD COMFORT-

DISPLAY.

Alarm without shutdown:

The code for the cause of alarm is indicated on the VFD COMFORT-

DISPLAY. The warning signal sounds in regular intervals. The messages appear every 10 seconds.

Press the key to stop the signal

Error message with ticker: >LOW LEVEL ALARM-FILL MEDIUM <

Low level alarm

The circulator is operated without or insufficient bath fluid.

Switch the unit off with the mains switch, refill bath fluid and switch on!

Tube breakage has occurred (insufficient filling level of bath fluid caused by pumping-out)

Replace the tubing and refill bath liquid.

The float is defect (e.g. transport damage).

Repair by authorized JULABO service personnel.

Error message with ticker:

> REFRIGERATOR ALARM-CHECK CONNECTION <

During the self-test after switch-on a short –circuit is detected between pin 2 and pin 4 of the control line or the control line was disconnected during operation.

Reconnect the control line or repair short-circuit.

Error message with ticker:

> EXCESS TEMPERATURE WARNING-CHECK LIMITS <

Excess temperature warning or

Excess temperature alarm

Type of warning: set to >warning< or >alarm< (refer to page 37)

Error message with ticker:

> LOW TEMPERATURE WARNING-CHECK LIMITS <

Low temperature warning or

Low temperature alarm.

Typ of warning: set to >warning< or >alarm< (refer to page 37)

Error message with ticker:

> WORKING SENSOR ALARM-CALL SERVICE <

Cable of working temperature sensor is disconnected or short-circuited.

74

HE

Error message with ticker:

>SENSOR DIFFERENCE ALARM-CHECK VISCOSITY AND PUMP STAGE<

Defect of working or excess temperature protector.

Working temperature sensor and excess temperature protector report a temperature difference of more than 35 K.

Error message with ticker:

> INTERNAL HARDWARE ERROR-CALL SERVICE <

Other errors

Error in A/D converter

Error message with ticker:

> EXCESS TEMPERATURE PROTECTOR ALARM-CHECK ADJUSTMENT <

Excess temperature protector defect.

The protection temperature is below the set working temperature setpoint.

Set the protection temperature to a higher value.

Error message with ticker:

> EXTERNAL SENSOR ALARM-CHECK EXTERNAL SENSOR <

External control was set but the Pt100 external senor was not connected or is defect.

Error message with ticker:

> CLEAN CONDENSER OR CHECK COOLING WATER <

Insufficient cooling of condenser. Clean the air-cooled condenser. Check the flow and the temperature of the cooling water of a water-cooled condenser.

Error message with ticker:

> COMPRESSOR FAILURE-CHECK REFRIGERATOR <

Stage 1 of the compressors does not work.

Automatic restart after short cool-down, message E 21 goes off.

Stage 2 of the compressor does not work.

Cooling machine – overload protection

The driving motor of the cooling compressor is equipped with an overload protection which is triggered by increased internal temperatures or excessive current consumption.

Shutdown can be caused by

- insufficient ventilation,

- insufficient wall distance,

- soiled condenser,

- high room temperature

- switching off and on in short sequence

Excess temperature in stage 1 of the compressor.

Excess temperature in stage 2 of the compressor.

Short circuit of control line to cooling machine during self-test.

75

Troubleshooting guide / error messages

Error message with ticker:

> SAFETY SENSOR ALARM-CALL SERVICE <

The cable of the excess temperature protector has been disconnected or short-circuited

Ext. Pt100 sensor input without signal, but setpoint programming set to external Pt100.

45

60

4 s

30

15

Error message with ticker:

> LOW LEVEL WARNING-FILL MEDIUM <

The early warning system for low level reports a critical fluid level. Refill bath fluid.

By quickly switching off and restarting the unit the alarm is cancelled.

If the error occurs once more after the restart, a remote diagnosis is required.

Error message with ticker:

> CONFIGURATION ERROR-PRESS OK<

The configuration of the circulator does not correspond with its current application.

Press the key for a non-recurring, automatic change of the configuration.

In this case please call the JULABO Technical Service or an authorized dealer.

Disturbances that are not indicated.

The electronic pump motor is overload-protected by an electronic current limiter. If viscosity of the bath fluid is or becomes too high, the motor stops running.

Warning:

Before exchanging the fuses, turn off the mains power switch and disconnect the power plug from the mains socket!

Only use fine fuses with a nominal value as specified.

Fuses:

The mains fuses on the rear of the unit may easily be exchanged as shown on the left.

Circulator: Fine fuses T 16 A , dia.5 x 20 mm

Cooling machine: Fuse T 10.0 A, dia.5 x 20 mm

Example:

Manufacturer

Schurter

Wickmann

Supplier

Schurter

Wickmann

Type

G-fuse insert SPT

T16A 5x20mm

G- fuse insert

T10,0A 5x20 mm

Order No.

No. 0001.2516

No. 19195

76

HE

11. Electrical connections

Notice:

Use shielded cables only.

The shield of the connecting cable is electrically connected to the plug housing.

The unit ensures safe operation if connecting cables with a maximum length of 3 m are used. The use of longer cables does not affect proper performance of the unit, however external interferences may have a negative impact on safe operation (e.g. cellular phones).

2 1

Pt100

3 4 ext. Pt100

Shield Plug

2 1

Socket for external Pt100 sensor

Pin assignment:

Pin Signal

1

2

I+

U+

3

4

U-

I-

The shield of the connecting cable is electrically connected to the plug housing and the sensor tube.

3 4

Look on soldering side.

5 1

9 6

RS232C

RS232 serial interface

This port can be used to connect a computer with an RS232 cable for remote control of the circulator.

Pin assignments RS232:

Pin 2

Pin 3

Pin 5

Pin 7

Pin 8

RxD

TxD

0 V

RTS

CTS

Pin 1; 4; 6, 9 Reserved - do not use!

RS232 interface cable

Circulator (9-pol)

Pin 2 RxD

Pin 3 TxD

Pin 5 GND

Pin 7 RTS

Pin 8 CTS

Receive Data

Transmit Data

Signal GND

Request to send

Clear to send

PC (9-pol)

Pin 3 TxD

Pin 2 RxD

Pin 5 GND

Pin 8 CTS

Pin 7 RTS

Accessories: Order No.

8 980 073

8 900 110

Description

RS232 interface cable 9-pol./9-pol. , 2,5 m

USB interface adapter cable

77

Electrical connections

ALARM

2

3

4

5

/ Control output

The connector may only be used for control of a JULABO refrigerated circulator or JULABO MVS Solenoid valve controller for cooling water.

Pin assignment:

Pin Signal (only with attached JULABO equipment)

1 +24 V (I max. current 25 mA)

0 V

Alarm relay

Reserved - do not use!

Cooling pulse

Alarm output

(for external alarm signal)

This potential-free change-over contact is activated in case of an alarm when pins 2 and 3 are connected.

Switching capacity max. 30 W / 40 VA

Switching voltage max. 125 V

∼/−

Switching current max. 1 A

Functional description see page 72

1

AK

2

3

STAND-BY

2

1

3

5

4

6

REG+E-PROG

STAND-BY input

(for external emergency switch-off)

Pin assignment: Pin

1

2

3

Functional description see page 6

Signal not connected

5 V / DC

0 V

Programmer input / temperature recorder output

Pin

1 Voltage output

2 Voltage output

Channel 1

Channel 2

3 GND for outputs

4 Programmer input EPROG

5 Current output

6 GND for Progammer

Channel 3

Functional description see page 69

Signal

0 ... 10 V

0 ... 10 V

0 V

0 to 10 V / 0 to 20 mA

0 to 20 mA / 4 to 20 mA

0 V

78

HE

12. Remote control

12.1. Setup for remote control

5 1

9 6

RS232C

1. Check the interface parameters for both interfaces (on circulator and PC) and make sure they match.

(Serial interface refer to page 56)

2. In the menu > MENU CONFIG < set the menu item

to

.

3. Connect both units with an interface cable..

Like all parameters which can be entered through the keypad, interface parameters are stored in memory even after the circulator is turned off.

12.2. Communication with a PC or a superordinated data system

If the circulator is put into remote control mode via the configuration level, the

VFD COMFORT-DISPLAY will read “R -OFF-” = REMOTE STOP. The circulator is now operated via the computer.

In general, the computer (master) sends commands to the circulator (slave).

The circulator sends data (including error messages) only when the computer sends a query.

In remote control mode, the start command and all values to be set must be resent by the PC via the interface in case of a power interruption.

AUTOSTART is not possible.

A transfer sequence consists of:

• command

• space ( ⇔; Hex: 20)

• parameter (decimal separation with a period)

• end of file (

↵; Hex: 0D)

The commands are divided into in and out commands.

in commands: retrieve parameters

out commands: set parameters

Important times for a command transmission:

To ensure a safe data transfer, the time gap between two commands should be at least 250 ms.

The circulator automatically responds to an in command with a data string followed by a LF (Line Feed). The next command should only be sent after

10 ms.

The out commands are valid only in remote control mode.

Command to set the working temperature > SETPNT 1< to 55.5 °C out_sp_00

⇔ 55.5↵

Command to retrieve the working temperature > SETPNT 1< in_sp_00

Response from the circulator: 55.5

79

Remote control

12.3. List of commands

out commands: Setting temperature values or parameters.

Command out_mode_01 out_mode_01 out_mode_01 out_mode_02 out_mode_02 out_mode_02 out_mode_03 out_mode_03 out_mode_04 out_mode_04 out_mode_05 out_mode_05 out_mode_08 out_mode_08 out_sp_00 out_sp_01 out_sp_02 out_sp_03 out_sp_04 out_sp_06 out_sp_07 out_par_04 out_par_06 out_par_07 out_par_08 out_par_09 out_par_10 out_par_11

80

Parameter

0

1

2

0

1

2

0

1

0

1

0

1

0

1 xxx.xx xxx.xx xxx.xx xxx.xx xxx.xx xxx.xx x x.x xxx xxx xxx xxx xxx xxx

Response of circulator

Use working temperature > SETPNT 1<

Use working temperature > SETPNT 2<

Use working temperature > SETPNT 3<

Selftuning „off“.

Temperature control by using the stored parameters.

Selftuning „once“

Single selftuning of controlled system after the next start.

Selftuning „always“

Continual selftuning of controlled system whenever a new setpoint is to be reached.

Set external programmer input to voltage.

Voltage 0 V ... 10 V

Set external programmer input to current.

Current 0 mA ... 20 mA

Temperature control of internal bath.

External control with Pt100 sensor.

Stop the unit = R –OFF-.

Start the unit.

Set the control dynamics - aperiodic

Set the control dynamics - standard

Set working temperature. „SETPNT 1“

Set working temperature. „SETPNT 2“

Set working temperature. „SETPNT 3“

Set high temperature warning limit „OverTemp“

Set low temperature warning limit „SubTemp“

Set manipulated variable for the heater via serial interface

-99.99 ... +100 [%]

Set the pump pressure stage. (1 ... 4)

CoSpeed for external control 0 ... 5.0

Xp control parameter of the internal controller. 0.1 ... 99.9

Tn control parameter of the internal controller. 3 ... 9999

Tv control parameter of the internal controller. 0 ... 999

Xp control parameter of the cascade controller. 0.1 ... 99.9

Proportional portion of the cascade controller. 1 ... 99.9

Tn control parameter of the cascade controller. 3 ... 9999

Command version status in_pv_00 in_pv_01 in_pv_02 in_pv_03 in_pv_04 in_sp_00 in_sp_01 in_sp_02 in_sp_03 in_sp_04 in_sp_05

Command out_par_12 out_par_13 out_par_14 out_par_15 out_par_16 out_hil_00

Parameter xxx xxx xxx xxx xxx

-xxx

HE

Response of circulator

Tv control parameter of the cascade controller. 0 ... 999

Maximum internal temperature of the cascade controller.

Minimum internal temperature of the cascade controller.

Band limit (upper) 0 ... 200 °C

Band limit (lower) 0 ... 200 °C

Set the desired maximum cooling power (0 % to 100 %).

Note: Enter the value with a preceding negative sign!

This adjustment only makes sense if FP refrigerating units are used.

Set the desired maximum heating power (10 % to 100 %). out_hil_01 xxx

in commands: Asking for parameters or temperature values to be displayed. in_sp_07 in_sp_08 in_par_00 in_par_01 in_par_02 in_par_03

Parameter none none none none none none none none none none none none none none none none none none none

Response of circulator

Number of software version (V X.xx)

Status message, error message (see page 83)

Actual bath temperature.

Heating power being used (%).

Temperature value registered by the external Pt100 sensor.

Temperature value registered by the safety sensor.

Setpoint temperature of the excess temperature protection

Working temperature „SETP 1“

Working temperature „SETP 2“

Working temperature „SETP 3“

High temperature warning limit „OverTemp“

Low temperature warning limit „SubTemp“

Setpoint temperature of the external programmer

(REG+E-PROG) .

1. Adjusted pump stage in the -OFF- condition.

2. Pump stage corresponding to the effective rotation speed

after start.

Value of a flowrate sensor connected to the E-Prog input

Temperature difference between working sensor and safety sensor

Te - Time constant of the external bath.

Si - Internal slope

Ti - Time constant of the internal bath.

81

Remote control

Command in_par_04 in_par_05 in_par_06 in_par_07 in_par_08 in_par_09 in_par_10 in_par_11 in_par_12 in_par_13 in_par_14 in_par_15 in_par_16 in_mode_01 in_mode_02 in_mode_03 in_mode_04 in_mode_05 in_mode_08 in_hil_00 in_hil_01

Parameter none none none none none none none none none none none none none none none none none none none none none

Response of circulator

CoSpeed - Band limit (max. difference between the temperatures in the internal bath and external system).

Factor pk/ph0:

Ratio of max. cooling capacity versus max. heating capacity

Xp control parameter of the internal controller.

Tn control parameter of the internal controller.

Tv control parameter of the internal controller.

Xp control parameter of the cascade controller.

Proportional portion of the cascade controller.

Tn control parameter of the cascade controller.

Tv control parameter of the cascade controller.

Adjusted maximum internal temperature of the cascade controller.

Adjusted minimum internal temperature of the cascade controller.

Band limit (upper)

Band limit (lower)

Selected setpoint:

0 = SETP 1

1 = SETP 2

2 = SETP 3

3 = Last setpoint setting was carried out

through an external programmer

Selftuning type:

0 = Selftuning „off“

1 = Selftuning „once“

2 = Selftuning „alwayS“

Type of the external programmer input:

0 = Voltage 0 V to 10 V

1 = Current 0 mA to 20 mA

Internal/external temperature control:

0 = Temperature control with internal sensor.

1 = Temperature control with external Pt100 sensor.

Circulator in Stop/Start condition:

0 = Stop

1 = Start

Adjusted control dynamics

0 = aperiodic

1 = standard

Max. cooling power (%).

Max. heating power (%).

82

HE

12.4. Status messages

Status messages

00 MANUAL STOP

01 MANUAL START

02 REMOTE STOP

03 REMOTE START

Description

Circulator in „OFF“ state.

Circulator in keypad control mode.

Circulator in „r OFF“ state.

Circulator in remote control mode.

12.5. Error messages

Error messages

-01 LOW LEVEL ALARM

-02 REFRIGERATOR ALARM

Description

Low liquid level alarm.

Control cable of the refrigerated circulator or MVS solenoid valve controller short-circuited or interrupted.

High temperature warning. -03 EXCESS TEMPERATURE

WARNING

-04 LOW TEMPERATURE WARNING

-05 WORKING SENSOR ALARM

-06 SENSOR DIFFERENCE ALARM

-07 I

2

C-BUS ERROR

-08 INVALID COMMAND

-09 COMMAND NOT ALLOWED IN

CURRENT OPERATING MODE

-10 VALUE TOO SMALL

-11 VALUE TOO LARGE

-12 TEMPERATURE MEASUREMENT

ALARM

-13 WARNING : VALUE EXCEEDS

TEMPERATURE LIMITS

-14 EXCESS TEMPERATURE

PROTECTOR ALARM

Low temperature warning.

Working temperature sensor short-circuited or interrupted.

Sensor difference alarm.

Working temperature and safety sensors report a temperature difference of more than 35 K.

Internal error when reading or writing the I

2

C bus.

Invalid command.

Invalid command in current operating mode.

Entered value too small.

Entered value too large.

Error in A/D converter.

Value lies outside the adjusted range for the high and low temperature warning limits. But value is stored.

Excess temperature protector alarm

-15 EXTERNAL SENSOR ALARM

-20 WARNING: CLEAN CONDENSOR

OR CHECK COOLING WATER

CIRCUIT OF REFRIGERATOR

-21 WARNING: COMPRESSOR STAGE

1 DOES NOT WORK

-22 WARNING: COMPRESSOR STAGE

2 DOES NOT WORK

External control selected, but external Pt100 sensor not connected.

Cooling of the condenser is affected.

Clean air-cooled condenser.

Check the flow rate and cooling water temperature on water-cooled condenser.

Compressor stage 1 does not work.

Compressor stage 2 does not work.

83

Installation of electronic module

Error messages Description

-23 WARNING: HIGH TEMPERATURE

ON COMPRESSOR STAGE 1

-24 WARNING: HIGH TEMPERATURE

ON COMPRESSOR STAGE 2

-25 REFRIGERATOR WARNING

Excess temperature on compressor stage 1.

Excess temperature on compressor stage 2.

Error in the cooling machine.

-26 WARNING: STAND-BY PLUG IS

MISSING

External standby contact is open.

(see page 72 and 78)

-30 CONFIGURATION ERROR:

CONFIRM BY PRESSING <OK> ON

CIRCULATOR

The configuration of the circulator does not conform to its present use.

Press >OK< to automatically perform a single modification of the configuration.

-33 SAFETY SENSOR ALARM

-38 EXTERNAL SENSOR SETPOINT

PROGRAMMING ALARM

-40 NIVEAU LEVEL WARNUNG

Excess temperature sensor short-circuited or interrupted.

Ext. Pt100 sensor input without signal and setpoint programming set to external Pt100.

Low liquid level warning in the internal reservoir.

13. Installation of electronic module

Caution:

Always turn off the unit and disconnect the mains cable from the power source before performing any service or maintenance procedures, or before moving the unit.

Configuration, installation, maintenance and repairs on the circulator may only be carried out by qualified personnel.

B

ALARM STAND-BY REG+

E-PROG

A

• Turn off the unit with the mains switch and disconnect the power plug.

• Open the screws (A) and remove the dummy plate.

• Carefully insert the upper edge (B) into the rear panel from below, align the electronic module at top and bottom and applying light pressure connect the 15-pole connector.

• Attach the electronic module using the screws (A).

• The circulator is ready for use. It automatically recognizes the installed electronic module.

84

HE

14. JULABO Service – Online remote diagnosis

JULABO circulators of the HighTech series are equipped with a black box. This box is implemented in the controller and records all significant data for the last 30 minutes.

In case of a failure, this data can be read out from the unit by using special software. This software is available as a free download from www.julabo.de \ EasyBlackBox.

• Installation is easy and is performed step by step.

Please observe the instructions.

• Data read-out is possible in the conditions

“OFF”, “R OFF” or “ALARM”.

• Connect the circulator to the computer using an interface cable.

• Start the EasyBlackBox program.

The program asks for the port used (COM1, ......) and the baud rate of the unit.

You do not have this information on hand? Simply try it out!

The program continues to send the request until the correct settings are made.

• Data is read out and shown on the monitor divided into the sections

>Einstellungen/Settings<,

>Alarmspeicher/Alarms stored<,

>Blackbox<

 see example

• After pressing >Speichern/Save<, a text file is created. The program suggests a filename -

>C:\model description and barcode no.<.

Modifications are possible.

• E-mail this file to [email protected], JULABO's service department. JULABO is thus able to provide rapid support.

85

86

Cleaning / repairing the unit

15. Cleaning / repairing the unit

Caution:

Always turn off the unit and disconnect the mains cable from the power source before cleaning the unit.

Prevent humidity from entering into the circulator.

Electrical connections and any other work must be performed by qualified personnel only.

To maintain the full cooling performance, clean the condenser from time to time.

• Switch off the unit, disconnect mains power cable.

• Hold the venting grid, pull out and remove.

• Clean the ribbed condenser with a vacuum cleaner.

• Replace the venting grid.

• Switch on the unit.

Cleaning:

For cleaning the bath tank and the immersed parts of the circulator, use low surface tension water (e.g., soap suds).

Clean the outside of the unit using a wet cloth and low surface tension water.

The circulator is designed for continuous operation under normal conditions. Periodic maintenance is not required.

The tank should be filled only with a bath fluid recommended by JULABO.

To avoid contamination, it is essential to change the bath fluid from time to time.

Repairs

Before asking for a service technician or returning a JULABO instrument for repair, please contact an authorized JULABO service station.

When returning the unit:

• Clean the unit in order to avoid any harm to the service personnel.

• Attach a short fault description.

• During transport the unit has to stand upright. Mark the packing correspondingly.

• When returning a unit, take care of careful and adequate packing.

• JULABO is not responsible for damages that might occur from insufficient packing.

JULABO reserves the right to carry out technical modifications with repairs for providing improved performance of a unit.

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Related manuals

advertisement