advertisement
Instruction Manual
748213-S
April 2002
Model 755R
Oxygen Analyzer http://www.processanalytic.com
ESSENTIAL INSTRUCTIONS
READ THIS PAGE BEFORE PROCEEDING!
Rosemount Analytical designs, manufactures and tests its products to meet many national and international standards. Because these instruments are sophisticated technical products, you
MUST properly install, use, and maintain them to ensure they continue to operate within their normal specifications. The following instructions MUST be adhered to and integrated into your safety program when installing, using, and maintaining Rosemount Analytical products. Failure to follow the proper instructions may cause any one of the following situations to occur: Loss of life; personal injury; property damage; damage to this instrument; and warranty invalidation.
• Read all instructions prior to installing, operating, and servicing the product.
• If you do not understand any of the instructions, contact your Rosemount Analytical representative for clarification.
• Follow all warnings, cautions, and instructions marked on and supplied with the product.
• Inform and educate your personnel in the proper installation, operation, and maintenance of
the product.
• Install your equipment as specified in the Installation Instructions of the appropriate
Instruction Manual and per applicable local and national codes. Connect all products to the proper electrical and pressure sources.
• To ensure proper performance, use qualified personnel to install, operate, update, program, and maintain the product.
• When replacement parts are required, ensure that qualified people use replacement parts specified by
Rosemount. Unauthorized parts and procedures can affect the product’s performance, place the safe operation of your process at risk, and VOID YOUR WARRANTY. Look-alike substitutions may result in fire, electrical hazards, or improper operation.
• Ensure that all equipment doors are closed and protective covers are in place, except when maintenance is being performed by qualified persons, to prevent electrical shock and personal
injury.
The information contained in this document is subject to change without notice.
Teflon and Viton are registered trademarks of E.I. duPont de Nemours and Co., Inc.
Paliney No.7 is a trademark of J.M. Ney Co., Hartford, CT
SNOOP is a registered trademark of NUPRO Co.
Emerson Process Management
Rosemount Analytical Inc.
Process Analytic Division
1201 N. Main St.
Orrville, OH 44667-0901
T (330) 682-9010
F (330) 684-4434 e-mail: [email protected]
http://www.processanalytic.com
Model 755R
Instruction Manual
748213-S
April 2002
TABLE OF CONTENTS
PREFACE...........................................................................................................................................P-1
Definitions ...........................................................................................................................................P-1
Intended Use Statement.....................................................................................................................P-2
Safety Summary .................................................................................................................................P-2
General Precautions For Handling And Storing High Pressure Gas Cylinders .................................P-4
Documentation....................................................................................................................................P-5
Compliances .......................................................................................................................................P-5
1-0 DESCRIPTION AND SPECIFICATIONS..............................................................................1-1
1-1 Description.............................................................................................................................1-1
1-2
1-3
1-4
1-5
Recorder Output Ranges.......................................................................................................1-1
Mounting ................................................................................................................................1-1
Isolated Current Output Option .............................................................................................1-1
Alarm Option..........................................................................................................................1-2
1-6
1-7
1-8
Electrical Options...................................................................................................................1-2
Remote Range Change Option .............................................................................................1-2
Specifications ........................................................................................................................1-3 a. Performance....................................................................................................................1-3 b. Sample ............................................................................................................................1-3 c. Electrical..........................................................................................................................1-4 d. Physical...........................................................................................................................1-4
2-0 INSTALLATION ....................................................................................................................2-1
2-1 Facility Preparation................................................................................................................2-1 a. Installation Drawings.......................................................................................................2-1 b. Electrical Interconnection Diagram ...............................................................................2-1 c. Flow Diagram ..................................................................................................................2-1
2-2
2-3 d. Location and Mounting....................................................................................................2-1
Calibration Gas Requirements ..............................................................................................2-2 a. Zero Standard Gas..........................................................................................................2-2 b. Span Standard Gas ........................................................................................................2-2
Sample...................................................................................................................................2-2 a. Temperature Requirements ............................................................................................2-2 b. Pressure Requirements - General ..................................................................................2-3 c. Normal Operation at Positive Gauge Pressures.............................................................2-3
2-4
2-5
2-6 d. Operation at Negative Gauge Pressures........................................................................2-4 e. Flow Rate ........................................................................................................................2-4 f.
Materials in Contact with Sample...................................................................................2-4 g. Corrosive Gases .............................................................................................................2-4
Leak Test...............................................................................................................................2-5
Electrical Connections ...........................................................................................................2-6 a. Line Power Connection...................................................................................................2-6 b. Recorder Output Selection and Cable Connections .......................................................2-6 c. Potentiometric Output .....................................................................................................2-7 d. Isolated Current Output (Optional)..................................................................................2-7 e. Output Connections and Initial Setup for Dual Alarm Option .........................................2-8
Remote Range Change Option .............................................................................................2-12
Rosemount Analytical Inc. A Division of Emerson Process Management Contents i
Instruction Manual
748213-S
April 2002
Model 755R
3-0 OPERATION .........................................................................................................................3-1
3-1 Overview................................................................................................................................3-1
3-2
3-3
3-4
Operating Range Selection ...................................................................................................3-1
Startup Procedure .................................................................................................................3-1
Calibration..............................................................................................................................3-1
3-5
3-6 a. Calibration with Zero and Span Standard Gases ...........................................................3-1
Compensation For Composition Of Background Gas ...........................................................3-2 a. Oxygen Equivalent Value of Gases ................................................................................3-4 b. Computing Adjusted Settings for Zero and Span Controls.............................................3-4
Selection Of Setpoints And Deadband On Alarm Option......................................................3-7
3-7
3-8
Current Output Board (Option) ..............................................................................................3-7
Routine Operation .................................................................................................................3-8
3-9 Effect of Barometric Pressure Changes on Instrument Readout ..........................................3-8
3-10 Calibration Frequency ...........................................................................................................3-8
4-0 THEORY................................................................................................................................4-1
4-1 Principles of Operation ..........................................................................................................4-1
4-2
4-3
Variables Influencing Paramagnetic Oxygen Measurements ...............................................4-2 a. Pressure Effects..............................................................................................................4-2
Electronic Circuitry.................................................................................................................4-4 a. Detector/Magnet Assembly.............................................................................................4-4 b. Control Board and Associated Circuitry..........................................................................4-4 c. Power Supply Board Assembly.......................................................................................4-5 d. Isolated Current Output Board (Optional) .......................................................................4-6
5-0 CIRCUIT ANALYSIS.............................................................................................................5-1
5-1
5-2
5-3
Circuit Operation....................................................................................................................5-1
±15 VDC Power Supply.........................................................................................................5-1
Case Heater Control Circuit...................................................................................................5-1
5-4
5-5
5-6
5-7
5-8
5-9
Detector Heater Control Circuit .............................................................................................5-6
Detector Light Source Control Circuit....................................................................................5-7
Detector with First Stage Amplifier ........................................................................................5-8
Buffer Amplifiers U8 and U10 with Associated Anticipation Function ...................................5-10
Digital Output Circuit..............................................................................................................5-10
Analog Output Circuits for Recorder and Alarms ..................................................................5-11 a. First Stage Amplifier........................................................................................................5-11 b. Second Stage Amplifier ..................................................................................................5-11
6-0 MAINTENANCE AND SERVICE ..........................................................................................6-1
6-1 Initial Checkout With Standard Gases...................................................................................6-1 a. Control Board Checkout..................................................................................................6-1
6-2
6-3
6-4
Heating Circuits .....................................................................................................................6-2 a. Case Heater Control Circuit ............................................................................................6-2
Detector/Magnet Heating Circuit ...........................................................................................6-2
Detector Check......................................................................................................................6-4 a. Source Lamp...................................................................................................................6-5 b. Photocell .........................................................................................................................6-5 c. Suspension .....................................................................................................................6-5 ii Contents Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R
Instruction Manual
748213-S
April 2002
6-5
6-6
Replacement Of Detector/Magnet Components ...................................................................6-5 a. Source Lamp...................................................................................................................6-5 b. Photocell .........................................................................................................................6-5 c. Detector...........................................................................................................................6-7
Control Board Setup ..............................................................................................................6-7 a. Power Supply Test..........................................................................................................6-7 b. Detector zero...................................................................................................................6-7 c. U4 Zero ...........................................................................................................................6-8 d. U8 Zero ...........................................................................................................................6-8 e. U10 Zero .........................................................................................................................6-8 f.
Fullscale ..........................................................................................................................6-8 g. Recorder Fullscale ..........................................................................................................6-8
7-0 REPLACEMENT PARTS ......................................................................................................7-1
7-1
7-2
7-3
Circuit Board Replacement Policy.........................................................................................7-1
Matrix – Model 755R Oxygen Analyzer.................................................................................7-2
Selected Replacement Parts.................................................................................................7-3
8-0 RETURN OF MATERIAL ......................................................................................................8-1
8-1
8-2
8-3
Return Of Material .................................................................................................................8-1
Customer Service ..................................................................................................................8-1
Training..................................................................................................................................8-1
LIST OF ILLUSTRATIONS
Figure 1-1.
Model 755R Oxygen Analyzer – Front Panel ........................................................ 1-1
Figure 2-1.
Interconnect of Typical Gas Manifold to Model 755R............................................ 2-3
Figure 2-2.
Model 755R Rear Panel ........................................................................................ 2-5
Figure 2-3.
Connections for Potentiometric Recorder with Non-Standard Span ..................... 2-6
Figure 2-4.
Model 755R Connected to Drive Several Current-Actuated Output Devices........ 2-7
Figure 2-5.
Relay Terminal Connections for Typical Fail-Safe Applications............................ 2-8
Figure 2-6.
Typical Alarm Settings ......................................................................................... 2-10
Figure 2-7.
Alarm Relay Assembly Schematic Diagram ........................................................ 2-11
Figure 3-1.
Control Board - Adjustment Locations................................................................... 3-3
Figure 3-2.
Dial Settings for Alarm Setpoint Adjustments........................................................ 3-7
Figure 4-1.
Functional Diagram of Paramagnetic Oxygen Measurement System................... 4-3
Figure 4-2.
Spherical Body in Non-Uniform Magnetic Field..................................................... 4-4
Figure 4-3.
Detector/Magnet Assembly.................................................................................... 4-7
Figure 5-1.
Two-Comparator OR Circuit .................................................................................. 5-2
Figure 5-2.
Case Heater Control Circuit................................................................................... 5-3
Figure 5-3.
Ramp Generator Circuit......................................................................................... 5-3
Figure 5-4.
Detector Heater Control Circuit.............................................................................. 5-6
Figure 5-5.
Detector Light Source Control Circuit .................................................................... 5-7
Figure 5-6.
Detector with First Stage Amplifier ........................................................................ 5-9
Figure 5-7.
Buffer, Anticipation, and Digital Output Circuits................................................... 5-10
Figure 5-8.
Simplified Analog Recorder Output Circuit .......................................................... 5-12
Figure 6-1.
Detector/Magnet Assembly.................................................................................... 6-3
Figure 6-2.
Pin/Lead Removal ................................................................................................. 6-4
Figure 6-3.
Detector Optical Bench.......................................................................................... 6-4
Figure 6-4.
Lamp Replacement................................................................................................ 6-6
Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii
Instruction Manual
748213-S
April 2002
Model 755R
LIST OF TABLES
Table 2-1.
Remote Range Switching Truth Table................................................................. 2-12
Table 3-1.
Calibration Range for Various Zero-Based Operating Ranges ............................. 3-4
Table 3-2.
Oxygen Equivalent of Common Gases ................................................................. 3-6
617186
620434
646090
652826
654014
654015
656081
DRAWINGS
(LOCATED IN REAR OF MANUAL)
Schematic Diagram, Case Board
Schematic Diagram, Isolated Current Output Board
Schematic Diagram, Remote Range Board
Schematic Diagram, Control Board
Pictorial Wiring Diagram, Model 755R
Installation Drawing, Model 755R
Instructions, Remote Range Selection iv Contents Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R
Instruction Manual
748213-S
April 2002
PREFACE
The purpose of this manual is to provide information concerning the components, functions, installation and maintenance of the 755R.
Some sections may describe equipment not used in your configuration. The user should become thoroughly familiar with the operation of this module before operating it. Read this instruction manual completely.
DEFINITIONS
The following definitions apply to DANGERS, WARNINGS, CAUTIONS and NOTES found throughout this publication.
DANGER .
Highlights the presence of a hazard which will cause severe personal injury, death, or substantial property damage if the warning is ignored.
WARNING .
Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not strictly observed, could result in injury, death, or long-term health hazards of personnel.
CAUTION .
Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not strictly observed, could result in damage to or destruction of equipment, or loss of effectiveness.
NOTE
Highlights an essential operating procedure, condition or statement.
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-1
Instruction Manual
748213-S
April 2002
Model 755R
INTENDED USE STATEMENT
The Model 755R is intended for use as an industrial process measurement device only. It is not intended for use in medical, diagnostic, or life support applications, and no independent agency certifications or approvals are to be implied as covering such application.
SAFETY SUMMARY
If this equipment is used in a manner not specified in these instructions, protective systems may be impaired.
AUTHORIZED PERSONNEL
To avoid explosion, loss of life, personal injury and damage to this equipment and on-site property, all personnel authorized to install, operate and service the this equipment should be thoroughly familiar with and strictly follow the instructions in this manual. SAVE THESE
INSTRUCTIONS.
DANGER .
ELECTRICAL SHOCK HAZARD
Do not operate without doors and covers secure. Servicing requires access to live parts which can cause death or serious injury. Refer servicing to qualified personnel.
For safety and proper performance this instrument must be connected to a properly grounded three-wire source of power.
Optional alarm switching relay contacts wired to separate power sources must be disconnected before servicing.
WARNING
POSSIBLE EXPLOSION HAZARD
This analyzer is of a type capable of analysis of sample gases which may be flammable. If used for analysis of such gases, internal leakage of sample could result in an explosion causing death, personal injury, or property damage. Do not use this analyzer on flammable samples. Use explosionproof version instruments for analysis of flammable samples.
P-2 Preface Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R
Instruction Manual
748213-S
April 2002
WARNING .
PARTS INTEGRITY
Tampering or unauthorized substitution of components may adversely affect safety of this product.
Use only factory documented components for repair.
CAUTION
PRESSURIZED GAS
This module requires periodic use of pressurized gas. See General Precautions for Handling and
Storing High Pressure Gas Cylinders, page P-4
CAUTION
TOPPLING HAZARD
This instrument’s internal pullout chassis is equipped with a safety stop latch located on the left side of the chassis.
When extracting the chassis, verify that the safety latch is in its proper (counter-clockwise) orientation.
If access to the rear of the chassis is required, the safety stop may be overridden by lifting the latch; however, further extraction must be done very carefully to insure the chassis does not fall out of its enclosure.
If the instrument is located on top of a table or bench near the edge, and the chassis is extracted, it must be supported to prevent toppling.
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-3
Instruction Manual
748213-S
April 2002
Model 755R
GENERAL PRECAUTIONS FOR HANDLING AND STORING HIGH
PRESSURE GAS CYLINDERS
Edited from selected paragraphs of the Compressed Gas Association's "Handbook of Compressed
Gases" published in 1981
Compressed Gas Association
1235 Jefferson Davis Highway
Arlington, Virginia 22202
Used by Permission
1. Never drop cylinders or permit them to strike each other violently.
2. Cylinders may be stored in the open, but in such cases, should be protected against extremes of weather and, to prevent rusting, from the dampness of the ground. Cylinders should be stored in the shade when located in areas where extreme temperatures are prevalent.
3. The valve protection cap should be left on each cylinder until it has been secured against a wall or bench, or placed in a cylinder stand, and is ready to be used.
4. Avoid dragging, rolling, or sliding cylinders, even for a short distance; they should be moved by using a suitable hand-truck.
5. Never tamper with safety devices in valves or cylinders.
6. Do not store full and empty cylinders together. Serious suckback can occur when an empty cylinder is attached to a pressurized system.
7. No part of cylinder should be subjected to a temperature higher than 125
° permitted to come in contact with any part of a compressed gas cylinder.
F (52
°
C). A flame should never be
8. Do not place cylinders where they may become part of an electric circuit. When electric arc welding, precautions must be taken to prevent striking an arc against the cylinder.
P-4 Preface Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R
Instruction Manual
748213-S
April 2002
DOCUMENTATION
The following Model 755R instruction materials are available. Contact Customer Service Center or the local representative to order.
748213 Instruction Manual (this document)
COMPLIANCES
This product satisfies all obligations of all relevant standards of the EMC framework in Australia and New
Zealand.
N96
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-5
Instruction Manual
748213-S
April 2002
Model 755R
P-6 Preface Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R
Instruction Manual
748213-S
April 2002
SECTION 1
DESCRIPTION AND SPECIFICATIONS
1-1 DESCRIPTION
The Model 755R Oxygen Analyzer provides continuous readout of the oxygen content of a flowing gas sample. The determination is based on measurement of the magnetic susceptibility of the sample gas. Oxygen is strongly paramagnetic while most other common gases are weakly diamagnetic.
The instrument provides direct readout of 0 to
100% oxygen concentration on a front panel digital display. In addition, a field-selectable voltage output is provided as standard. An isolated current output of 0 to 20 mA or 4 to
20 mA is obtainable through plug-in of an optional circuit board. Current and voltage outputs may be utilized simultaneously if desired.
An alarm option is also available by way of a relay assembly that mounts at the rear of the case with a cable that plugs into the Control
Board. Customer connections are available on this assembly.
The basic electronic circuitry is incorporated into two master boards designated the Control
Board assembly and the Power Supply Board assembly. The Control Board has receptacles that accept optional plug-in current output board and alarm features.
1-2 RECORDER OUTPUT RANGES
Seven zero-based ranges are available with the Model 755R: 0 to 1%, 0 to 2.5%, 0 to 5%, 0 to 10%, 0 to 25%, 0 to 50%, and 0 to 100%.
Each range is jumper selectable.
1-3 MOUNTING
The Model 755R is a rack-mounted instrument, standard for a 19-inch relay rack (Refer to IEC Standard, Publication 297-1, 1986).
1-4 ISOLATED CURRENT OUTPUT OPTION
An isolated current output is obtainable by using an optional current output board, either during factory assembly or subsequently in the field. The board provides ranges of 0 to 20 or 4 to 20 mA into a maximum resistive load of 1000 ohms.
Digital Display
% O
2
ZERO SPAN
Rosemount Analytical Model 755R
Zero Control Span Control
Figure 1-1. Model 755R Oxygen Analyzer – Front Panel
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-1
Instruction Manual
748213-S
April 2002
1-5 ALARM OPTION
The alarm option contains:
• An alarm circuit incorporating two comparator amplifiers, one each for the
ALARM 1 and ALARM 2 functions. Each amplifier has associated setpoint and deadband adjustments. Setpoint is adjustable from 1% to 100% of fullscale.
Deadband is adjustable from 1% to 20% of fullscale.
• An alarm relay assembly, containing two single-pole, double-throw relays (one each for the ALARM 1 and ALARM 2
Model 755R contacts). These relays may be used to drive external, customer-supplied alarm and/or control devices.
1-6 ELECTRICAL OPTIONS
The analyzer is supplied, as ordered, for operation on either 115 VAC, 50/60 Hz or 230
VAC, 50/60 Hz.
1-7 REMOTE RANGE CHANGE OPTION
This option allows the customer to remotely control the recorder scaling. It disables the internal recorder fullscale range select without affecting the front panel display.
1-2 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R
Instruction Manual
748213-S
April 2002
1-8 SPECIFICATIONS 1 a. Performance
Operating Range (Standard) ......... 0 to 5, 0 to 10, 0 to 25, 0 to 50, and 0 to 100% oxygen
Operating Range (Optional) .......... 0 to 1, 0 to 2.5, 0 to 5, 0 to 10, 0 to 25, 0 to 50, and 0 to 100% oxygen
Response Time ............................. 90% of fullscale, 20 seconds
Reproducibility ............................... 0.01% oxygen or ±1% of fullscale, whichever is greater
Ambient Temperature Limits ......... 32 °F (0°C) to 113°F (45°C)
Zero Drift........................................ ±1% fullscale per 24 hours, provided that ambient temperature does not change by more than 20 °F (11.1°C)
±2.5% of fullscale per 24 hours with ambient temperature change over entire range
Span Drift....................................... ±1% fullscale per 24 hours, provided that ambient temperature does not change by more than 20 °F (11.1°C)
±2.5% of fullscale per 24 hours with ambient temperature change over entire range b. Sample
Dryness ......................................... Sample dewpoint below 110 °F (43°C), sample free of entrained liquids.
Temperature Limits ....................... 50 °F (10°C) to 150°F (65°C)
Operating Pressure ....................... Maximum: 10 psig (68.9 kPa)
....................................................... Minimum: 5 psig vacuum (34.5 kPa vacuum)
Flow Rate ...................................... 50 cc/min. to 500 cc/min.
....................................................... Recommended 250 ±20 cc/min.
Materials in Contact with Sample.. Glass, 316 stainless steel, titanium, Paliney No. 7, epoxy resin,
Viton-A, platinum, nickel, and MgF2
1
Performance specifications are measured at recorder output and are based on constant sample pressure and deviation from set flow held to within 10% or 20 cc/min., whichever is smaller.
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-3
Instruction Manual
748213-S
April 2002
Model 755R c. Electrical
Supply Voltage and Frequency
(selectable when ordered)............ Standard: 115 VAC ±10 VAC, 50/60 Hz
Optional: 230 VAC ±10 VAC, 50/60 Hz
Power Consumption ...................... Maximum: 300 watts
Outputs .......................................... Standard: Field selectable voltage output of 0 to 10mV, 0 to
100mV, 0 to 1V, or 0 to 5VDC
Optional: Isolated current output of 0 to 20mA or 4 to 20mA (with
Current Output Board)
Alarm Option.................................. High-Low Alarm
Contact Ratings ..................... 5 amperes, 240V AC, resistive 3 amperes, 120 VAC inductive
1 amperes, 24V DC, resistive 5 amperes, 30 VDC resistive
5 amperes, 120V AC, resistive 3 amperes, 30 VDC inductive
Setpoint ......................................... Adjustable from 1% to 100% of fullscale
Deadband ...................................... Adjustable from 1% to 20% of fullscale (Factory set at 10% of fullscale) d. Physical
Mounting........................................ 19 inch rack (IEC 297-1, 1986)
Case Classification........................ General Purpose
Weight ........................................... 46 lbs. (21 kg)
Dimensions.................................... 19.0 x 8.7 x 19.2 inches (482.2 x 221 x 487 mm) W x H x D
1-4 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R
Instruction Manual
748213-S
April 2002
SECTION 2
INSTALLATION
2-1 FACILITY PREPARATION
Observe all precautions given in this section when installing the instrument.
a. Installation Drawings
For outline and mounting dimensions, gas connections, and other installation information, refer to Installation Drawing
654015 at the back of this manual.
b. Electrical Interconnection Diagram
Electrical interconnection is also shown in drawing 654015. Refer also to Section 2-
5, page 2-6.
c. Flow Diagram
The flow diagram of Figure 2-1 (page 2-3) shows connection of a typical gas selector manifold to the Model 755R.
d. Location and Mounting
Install the Model 755R only in a non-hazardous, weather-protected area.
Permissible ambient temperature range is
32 °F to 113°F (0°C to 45°C). Avoid mounting where ambient temperature may exceed the allowable maximum.
Magnetic susceptibilities and partial pressures of gases vary with temperature. In the Model 755R, temperature-induced readout error is avoided by control of temperatures in the following areas:
1. Interior of the analyzer is maintained at 140 °F (60°C) by an electrically controlled heater and associated fan.
2. Immediately downstream from the inlet port, prior to entry into the detector, the sample is preheated by passage through a coil maintained at approximately the same temperature as the detector (See Figure 4-3A, page 4-7).
3. The detector is maintained at a controlled temperature of 150 °F (66°C).
Also, avoid excessive vibration. To minimize vibration effects, the detector/magnet assembly is contained in a shock-mounted compartment.
WARNING
POSSIBLE EXPLOSION HAZARD
This analyzer is of a type capable of analysis of sample gases which may be flammable. If used for analysis of such gases, internal leakage of sample could result in an explosion causing death, personal injury, or property damage. Do not use this analyzer on flammable samples. Use explosion-proof version instruments for analysis of flammable samples.
Use reasonable precautions to avoid excessive vibration. In making electrical connections, do not allow any cable to touch the shock-mounted detector assembly or the associated internal sample inlet and outlet tubing. This precaution ensures against possible transmission of mechanical vibration through the cable to the detector, which could cause noisy readout.
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-1
Instruction Manual
748213-S
April 2002
2-2 CALIBRATION GAS REQUIREMENTS
WARNING
HIGH PRESSURE GAS CYLINDERS
Calibration gas cylinders are under pressure. Mishandling of gas cylinders could result in death, injury, or property damage.
Handle and store cylinders with extreme caution and in accordance with the manufacturer’s instructions. Refer to GENERAL
PRECAUTIONS FOR HANDLING & STOR-
ING HIGH PRESSURE CYLINDERS, page
P-4.
Analyzer calibration consists of establishing a zero calibration point and a span calibration point.
Zero calibration is performed on the range that will be used during sample analysis. In some applications, however, it may be desirable to perform span calibration on a range of higher sensitivity (i.e., more narrow span) and then jumper to the desired operating range.
For example, if the operating range is to be 0 to 50% oxygen, span calibration may be performed on the 0 to 25% range to permit use of air as the span standard gas.
Recommendations on calibration gases for various operating ranges are tabulated in Table 3-1 (page 3-4) and are explained in Sections 2-2a (page 2-2) and 2-2b (page 2-2).
Each standard gas should be supplied from a cylinder equipped with dual-stage, metal diaphragm type pressure regulator, with output pressure adjustable from 0 to 50 psig (0 to
345 kPa).
Instrument response to most non-oxygen sample components is comparatively slight, but is not in all cases negligible. During initial installation of an instrument in a given application, effects of the background gas should be calculated to determine if any correction is required (See Section 3-4, page 3-1).
Model 755R a. Zero Standard Gas
In the preferred calibration method, described in Section 3-4a (page 3-1), a suitable zero standard gas is used to establish a calibration point at or near the lower range limit. Composition of the zero standard normally requires an oxygen-free zero gas, typically nitrogen.
b. Span Standard Gas
A suitable span standard gas is required to establish a calibration point at or near the upper range limit. If this range limit is
21% or 25% oxygen, the usual span standard gas is air (20.93% oxygen).
2-3 SAMPLE
Basic requirements for sample are:
1. A 2-micron particulate filter, inserted into the sample line immediately upstream from the analyzer inlet.
2. Provision for pressurizing the sample gas to provide flow through the analyzer.
3. Provision for selecting sample, zero standard, or span standard gas for admission to the analyzer, and for measuring the flow of the selected gas.
a. Temperature Requirements
Sample temperature at the analyzer inlet should be in the range of 50 °F to 150°F
(10 °C to 66°C).
Normally, however, a maximum entry temperature of 110 °F (43°C) is recommended so that the sample temperature will rise during passage of the sample through the analyzer. This precaution prevents cooling of the sample and possible analyzer-damaging condensation.
With a thoroughly dry sample, entry temperature can be as high as 150 °F (66°C) without affecting readout accuracy.
2-2 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R
Needle
Valves
Instruction Manual
748213-S
April 2002
Model 755R
Oxygen Analyzer
Sample In
Zero
Standard
Gas
Two Micron
Filter
Flowmeter
To Vent
Span
Standard
Gas
Figure 2-1. Interconnect of Typical Gas Manifold to Model 755R b. Pressure Requirements - General
Operating pressure limits are as follows: maximum, 10 psig (68.9 kPa); minimum,
5 psig vacuum (34.5 kPa vacuum).
CAUTION
RANGE LIMITATIONS
Operation outside the specified pressure limits may damage the detector, and will void the warranty.
The basic rule for pressure of sample and standard gases supplied to the inlet is to calibrate the analyzer at the same pressure that will be used during subsequent operation, and to maintain this pressure during operation. The arrangement required to obtain appropriate pressure control will depend on the application.
When inputting sample or calibration gases, use the same pressure that will be used during subsequent operation. Refer to Section 2-3c (page 2-3), Normal Operation at Positive Gauge Pressures, or
Section 2-3d (page 2-4) Operation at
Negative Gauge Pressures.
c. Normal Operation at Positive Gauge
Pressures
Normally, the sample is supplied to the analyzer inlet at a positive gauge pressure in the range of 0 to 10 psig (0 to 68.9
kPa).
CAUTION
HIGH PRESSURE GAS CYLINDERS
Pressure surges in excess of 10 psig during admission of sample or standard gases can damage the detector.
Maximum permissible operating pressure is 10 psig (68.9 kPa). To ensure against over-pressurization, insert a pressure relief valve into the sample inlet line. In addition, a check valve should be placed in the vent line if the analyzer is connected to a manifold associated with a flare or other outlet that is not at atmospheric pressure. If the detector is overpressurized, damage will result.
The analyzer exhaust port is commonly vented directly to the atmosphere. Any change in barometric pressure results in a
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-3
Instruction Manual
748213-S
April 2002 directly proportional change in the indicated percentage of oxygen.
Example:
Range, 0% to 5% O
2.
Barometric pressure change after calibration, 1%.
Instrument reading, 5% O
2
.
Readout error = 0.01 x 5% O
2
=
0.05% O
2
.
Fullscale span is 5% O
2.
Therefore, the 0.05% O
2
error is equal to 1% of fullscale.
Thus, if the exhaust is vented to the atmosphere, the pressure effect must be taken into consideration. This may be accomplished in various ways, including manual computation and computer correction of data.
d. Operation at Negative Gauge Pressures
Operation at negative gauge pressures is not normally recommended, but may be used in certain special applications. A suction pump is connected to the analyzer exhaust port to draw sample into the inlet and through the analyzer. Such operation necessitates special precautions to ensure accurate readout. First is the basic consideration of supplying the standard gases to the analyzer at the same pressure that will be used for the sample during subsequent operation. In addition, any leakage in the sample handling system will result in decreased readout accuracy as compared with operation at atmospheric pressure.
The minimum permissible operating pressure is 5 psig vacuum (34.5 kPa vacuum).
Operation of the analyzer below this limit may damage the detector, and will void the warranty.
e. Flow Rate
Operating limits for sample flow rate are as follows: minimum, 50 cc/min; maxi-
Model 755R mum, 500 cc/min. A flow rate of less than
50 cc/min is too weak to sweep out the detector and associated flow system efficiently. Incoming sample may mix with earlier sample, causing an averaging or damping effect. Too rapid a flow will cause back pressure that will affect the readout accuracy. The optimum flow rate is between 200 and 300 cc/min.
Deviation from the set flow should be held to within 10% or 20 cc/min, whichever is smaller. If deviation is held to within these parameters and operating pressure remains constant, zero and span drift will remain within specification limits.
The analyzer should be installed near the sample source to minimize transport time.
Otherwise, time lag may be appreciable.
For example, assume that sample is supplied to the analyzer via a 100-foot
(30.5 m) length of 1/4-inch (6.35 mm) tubing. With a flow rate of 100 cc/min, sample transport time is approximately 6 minutes.
Sample transport time may be reduced by piping a greater flow than is required to the analyzer, and then routing only the appropriate portion of the total flow through the analyzer. The unused portion of the sample may be returned to the stream or discarded.
f. Materials in Contact with Sample
Within the Model 755R, the following materials are exposed to the sample: 316 stainless steel, glass, titanium, Paliney
No.7, epoxy resin, Viton-A, platinum, nickel and MgF
2
coating on mirror.
g. Corrosive Gases
In applications where the sample stream contains corrosive gases, a complete drying of the sample is desirable, as most of these gases are practically inert when totally dry. For corrosive applications consult the factory.
2-4 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R
2-4 LEAK TEST
WARNING
TOXIC OR CORROSIVE HAZARD
The sample containment system must be carefully leak checked upon installation and before initial start-up, during routine maintenance and any time the integrity of the sample containment system is broken, to ensure the system is in leak proof condition.
Internal leaks resulting from failure to observe these precautions could result in personal injury or property damage.
Instruction Manual
748213-S
April 2002
For proper operation and safety, system leakage must be corrected, particularly before introduction of toxic or corrosive samples and/or application of electrical power.
To check system for leaks, liberally cover all fittings, seals, and other possible sources of leakage with suitable leak test liquid such as
SNOOP (P/N 837801). Check for leak indicative bubbling or foaming. Leaks that are inaccessible to SNOOP application could evade detection by this method.
A B I C D E
L1/HOT
L2/NEUT
GND
CUR VOLT
OUTPUT OUTPUT
+ - G + -
(Rear terminal cover removed for clarity)
H G H
A. Sample outlet. 1/4” O.D. tube fitting.
B. Sample Inlet. 1/4” O.D. tube fitting.
C. 5/8” diameter hole for optional Dual Alarm Cable. Cable supplied by customer, minimum 24 AWG.
D. 5/8” diameter hole fitted with liquid-tight gland for Recorder Output Cable. Cable supplied by customer,
conductor, minimum 24 AWG.
E. 13/16” diameter hole for Power Cable. Cable supplied by customer, 3 conductor, minimum 18 AWG.
F. TB1: Customer hook-up for Power.
G. TB2: Customer hook-up for Recorder Output.
H. Optional Dual Alarm connections.
I. Connections for Optional Remote Range Change.
Figure 2-2. Model 755R Rear Panel
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-5
Instruction Manual
748213-S
April 2002
Model 755R
2-5 ELECTRICAL CONNECTIONS
WARNING
ELECTRICAL SHOCK HAZARD
For safety and proper performance, this instrument must be connected to a properly grounded three-wire source of supply.
Cable connections for AC power, recorder output, and alarm output are shown in Installation Drawing, 654015, and are explained in the following sections.
a. Line Power Connection
The analyzer is supplied, as ordered, for operation on 115 VAC or 230 VAC, 50/60
Hz. Ensure that the power source conforms to the requirements of the individual instrument, as noted on the name-rating plate.
Electrical power is supplied to the analyzer via a customer-supplied three-conductor cable, type SJT, minimum wire size 18 AWG. Route power cable through conduit and into appropriate opening in the instrument case. Connect power leads to HOT, NEUT, and GND terminals on the I/O board. Connect analyzer to power source via an external fuse or breaker, in accordance with local codes. Do not draw power for associated equipment from the analyzer power cable
(Refer to Figure 2-3 below).
If the analyzer is mounted in a protected rack or cabinet or on a bench, an accessory kit (P/N 654008) is available which provides a 10-foot North American power cord set and a liquid-tight feed through gland for the power cable hole. The kit also contains four enclosure support feet for bench top use.
b. Recorder Output Selection and Cable
Connections
If a recorder, controller, or other output device is used, connect it to the analyzer via a number 22 or number 24 AWG two-conductor shielded cable. Route the cable into the case through the liquid-tight feed through gland in the Recorder Output opening (See Installation Drawing,
654015). Connect the shield only at the recorder end or the analyzer end, not to both at the same time because a ground loop may occur.
NOTE:
Route recorder cable through a separate cable gland (P/N 899329) or conduit not with power cable or alarm output cable.
Cable connections and output selection for potentiometric and current-actuated devices are explained below.
755R
Analyzer
Potentiometric
Recorder
Input
Terminals
(Verify polarity
is correct) Voltage Divider
(Customer Supplied)
Position of Recorder Output
Selector Plug
Minimum Permissible
Resistance for R1 + R2
(ohms)
10 mV 1K
100 mV 10K
1 V 100K
5 V 2K
Figure 2-3. Connections for Potentiometric Recorder with Non-Standard Span
2-6 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R c. Potentiometric Output
1. Insert RECORDER OUTPUT Selector
Plug (See Figure 3-1) in position appropriate to the desired output: 10 mV, 100 mV, 1V or 5V.
2. Connect leads of shielded recorder cable to “REC OUT +” and “-” terminals on the I/O board.
3. Connect the output cable to the appropriate terminals of the recorder or other potentiometric device: a. For device with span of 0 to 10 mV, 0 to 100 mV, 0 to 1V, or 0 to
5V, connect cable directly to input terminals of the device, ensuring correct polarity and range selection.
b. For a device with intermediate span (i.e., between the specified values), connect the cable to the device via a suitable external voltage divider (See Figure 2-3, page 2-6).
Instruction Manual
748213-S
April 2002 d. Isolated Current Output (Optional)
1. Verify that the optional current output board appropriate to desired output is properly in place in its connector. See
Figure 3-1, page 3-3. If originally ordered with the analyzer, the board is factory installed.
2. On I/O board, connect leads of shielded recorder cable to “CUR-
RENT OUT+” and “-” terminals.
3. Connect free end of output cable to input terminals of recorder or other current-actuated device, making sure that polarity is correct. If two or more current-actuated devices are to be used, they must be connected in series (See Figure 2-4 below). Do not exceed the maximum load resistance of 1000 ohms.
Current and voltage outputs may be utilized simultaneously if desired.
mA
+
-
755R
Analyzer
+
Recorder
-
+
Controller
-
+
Remote
Indicator
-
Note: Total series resistance of all devices is not to exceed 1000 ohms.
Figure 2-4. Model 755R Connected to Drive Several Current-Actuated Output Devices
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-7
Instruction Manual
748213-S
April 2002 e. Output Connections and Initial Setup for Dual Alarm Option
If so ordered, the analyzer is factory equipped with alarm output. Alternatively, the alarm feature is obtainable by subsequent installation of the 654019 Alarm Kit.
Model 755R
The alarm output provides two sets of relay contacts for actuation of alarm and/or process-control functions. Leads from the customer-supplied external alarm system connect to terminals on the 654019 Alarm
Assembly (See Figure 2-5 below and Interconnect Drawing 654014).
REQUIREMENT
Low Alarm,
Fail-Safe
High Alarm,
Fail-Safe
Low Control
Limit,
Fail-Safe
No. 1 NO
COM
NC
RESET
RESET
No. 2
NO
COM
NC
No. 1 NO
COM
NC
RESET
RESET
No. 2
NO
COM
NC
TYPICAL CONNECTIONS
No. 1 NO
COM
NC
RESET
RESET
No. 2
NO
COM
NC
Alarm Bell or Lamp
N
H
115 VAC
Alarm Bell or Lamp
N
H
115 VAC
Solenoid
Valve
H
N
115 VAC
REQUIREMENT
Low Control
Limit,
Fail-Safe
Lower
Low Alarm
Indicator,
Fail-Safe
Low Control,
Fail-Safe
High Control,
Fail-Safe
Higher
High Alarm
Indicator,
Fail-Safe
Figure 2-5. Relay Terminal Connections for Typical Fail-Safe Applications
TYPICAL CONNECTIONS
No. 1
Solenoid
Valve NO
COM
NC
H
N
115 VAC
RESET
RESET
No. 2
NO
COM
NC
No. 1 NO
COM
NC
RESET
RESET
No. 2
NO
COM
NC
No. 1 NO
COM
NC
RESET
RESET
No. 2
NO
COM
NC
Alarm Bell or Lamp
Solenoid
Valve
N
H
115 VAC
H
N
115 VAC
Solenoid
Valve
H
N
115 VAC
Alarm Bell or Lamp
N
H
115 VAC
Note the following recommendations:
• A fuse should be inserted into the line between the customer-supplied power supply and the alarm relay terminals on the Alarm Relay Assembly.
• If the alarm contacts are connected to any device that produces radio fre-
• quency interference (RFI), it should be arc suppressed. The 858728 Arc
Suppressor is recommended.
If at all possible, the analyzer should operate on a different AC power source, to avoid RFI.
2-8 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R
Alarm 1 and Alarm 2 output through the
654019 Alarm Relay Assembly is provided by two identical single-pole, double-throw relays. These relay contacts are rated at the following values:
5 amperes 240 VAC resistive
1 ampere 240 VAC inductive
5 amperes 120 VAC resistive
3 amperes 120 VAC inductive
5 amperes 30 VDC resistive
3 amperes 30 VDC inductive
Removal of AC power from the analyzer
(such as power failure) de-energizes both relays, placing them in alarm condition.
Switching characteristics of the Alarm 1 and Alarm 2 relays are as follows:
The Alarm 1 relay coil is de-energized when the display moves downscale through the value that corresponds to setpoint minus deadband. This relay coil is energized when the display moves upscale through the value that corresponds to setpoint plus deadband.
The Alarm 2 relay coil is de-energized when the display moves upscale through the value that corresponds to the setpoint plus deadband. This relay coil is energized when the display moves downscale through the value that corresponds to setpoint minus deadband.
Both the ALARM 1 and ALARM 2 functions generally incorporate automatic rest.
When the display goes beyond the preselected limits, the corresponding relay is de-energized. When the display returns within the acceptable range, the relay is turned on.
The ALARM 1 and/or ALARM 2 alarm functions may be converted to manual reset. The conversion requires the substitution of an external pushbutton or other momentary contact switch for the jumper that connects the RESET terminals on the
Alarm Relay Assembly. If the corresponding relay is now de-energized (i.e., in alarm condition), the relay remains
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-9
Instruction Manual
748213-S
April 2002 de-energized until the operator momentarily closes the switch.
By appropriate connection to the double-throw relay contacts, it is possible to obtain either a contact closure or a contact opening for an energized relay. Also, either a contact closure or a contact opening may be obtained for a de-energized relay. It is important, for fail-safe applications, that the user understands what circuit conditions are desired in event of power failure and the resultant relay de-energization. Relay contacts should then be connected accordingly
(See Figure 2-5, page 2-8).
The ALARM 1 and ALARM 2 circuits have independent setpoint and deadband adjustments (See Figure 3-1, page 3-3). Initially, the ALARM 1 and ALARM 2
Setpoint Adjustments must be calibrated by means of the ALARM 1 and ALARM 2
Calibration Adjustments by the following procedure:
1. Set RANGE Select in a position appropriate to the span standard gas.
2. Inject span standard gas through analyzer at 50 to 500 cc/min.
3. Verify that ALARM 1 and ALARM 2
Deadband Adjustments (See Figure
3-1, page 3-3) are set for minimum value (turned fully counterclockwise).
These potentiometers should be factory-set for minimum deadband. Both potentiometers MUST REMAIN at this setting throughout calibration of the alarm setpoint adjustments.
4. Adjust ALARM 1 control function as follows: a. With ALARM 1 Setpoint Adjustment at 100% (i.e., position 10 on dial), adjust front panel SPAN
Control so that the display or recorder reads exactly fullscale.
b. Set ALARM 1 Calibrate Adjustment (R63) to its clockwise limit.
Instruction Manual
748213-S
April 2002
Carefully rotate R63 counterclockwise the minimum amount required to obtain energization of
ALARM 1 Relay K1 (See Figure
2-6 below and Figure 3-1, page 3-
3). Energization may be verified by connecting an ohmmeter to relay terminals on 654019 Alarm
Relay Assembly.
c. To verify correct adjustment of
R63, adjust front panel SPAN
Control so that the display or recorder reads 99% of fullscale.
Relay K1 should now be
DE-ENERGIZED.
5. Adjust ALARM 2 control function as follows: a. With ALARM 2 Setpoint Adjustment at 100% (i.e., Position 10 on the dial), adjust front panel SPAN
Control so that
Model 755R b. the display or recorder reads exactly fullscale.
c. Set ALARM 2 Calibrate Adjustment (R67) to its clockwise limit.
Carefully rotate R67 counterclockwise the minimum amount required to obtain energization of
ALARM 2 Relay K2 (See Figure
2-5, page 2-8).
d. To verify correct adjustment of
R67, adjust front panel SPAN
Control so that the display or recorder reads 99% of fullscale.
Relay K2 should now be
DE-ENERGIZED.
The ALARM 1 and ALARM 2 Setpoint
Adjustments are now properly calibrated and may be used to select the desired alarm setpoints, as described in Section
3-6 (page 3-7).
A. Typical ALARM 1 Setting
DEADBAND SET FOR
20% OF FULLSCALE
40
INPUT SIGNAL
Percent of Fullscale
30
When input signal moves upscale through this point, the coil of ALARM 1 relay (K1) is energized, providing continuity between the common and normally-closed contacts of the relay.
ALARM 1 Setpoint
B. Typical ALARM 2 Setting
DEADBAND SET FOR
10% OF FULLSCALE
20
When input signal moves downscale through this point, the coil of ALARM 1 relay (K1) is de-energized, providing continuity between the common and normally-open contacts of the relay.
55
INPUT SIGNAL
Percent of Fullscale
50
45
When input signal moves upscale through this point, the coil of ALARM 2 relay (K2) is de-energized, providing continuity between the common and normally-open contacts of the relay.
ALARM 2 Setpoint
When input signal moves upscale through this point, the coil of ALARM 2 relay (K2) is energized, providing continuity between the common and normally-closed contacts of the relay.
Figure 2-6. Typical Alarm Settings
2-10 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R
TB4
NO
1
9
+15V
J5
1
CR1 5
COM
NC
-15V
Alarm
No. 1
Command
2
4
14
K1
13
12
8
Reset
NO
1
9
5
COM
NC CR2
14 13
12 8
K2 Reset
Alarm
No. 2
Command
6
2. CR! AND CR2 ARE ANY 600V, 1 AMP DIODE.
1. RELAYS SHOWN IN ENERGIZED POSITION.
NOTES:
Figure 2-7. Alarm Relay Assembly Schematic Diagram
Alarm
No. 1
Alarm
No. 2
Instruction Manual
748213-S
April 2002
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-11
Instruction Manual
748213-S
April 2002
2-6 REMOTE RANGE CHANGE OPTION
The power supply circuitry on the Remote
Range Board 646004 must be jumpered for the correct line voltage, either 115 VAC or 230
VAC. See Drawings # 656081 (Table 2) and
646090 for correct jumper locations.
On the Remote Range Board, an additional option exists: for using either the on-board 12
V to drive the range select relays or an external 12 V supply.
To use an external supply:
1. Remove the E to F jumper (DWG 646090).
2. Apply the external 12 V to J3-5.
3. Program the remote controller to pull the range bits, J3-1 through J3-4, low. (See
Table 2-1 below.)
Model 755R
To use the internal 12 V supply:
1. Verify the E to F jumper is in place.
2. Connect the controller's common to J3-6 to reference the instrument's common to the controller's common.
NOTE
DO NOT connect anything to J3-5.
3. Connect J3-1 to J3-4, as shown in the truth table below, to switch ranges.
Remember that you are dealing with inverse logic and not normal binary addresses. Also, this process switches the recorder output only, and does not affect the front panel display.
Range 1
Range 2
Range 3
Range 4
J3-4
1
1
1
0
J3-3
1
1
0
1
J3-2
1
0
1
1
J3-1
0
1
1
1
Note: 1 = 12 V, 0 << 1 V.
Table 2-1.
Remote Range Switching Truth Table
Hex
E
D
B
7
2-12 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R
Instruction Manual
748213-S
April 2002
SECTION 3
OPERATION
3-1 OVERVIEW
Preparatory to operation, a familiarization with Figure 3-1 (page 3-3) is recommended. This figure gives locations and summarized descriptions of operating adjustments of the Model 755R Oxygen Analyzer.
3-2 OPERATING RANGE SELECTION
The Model 755R is designed to operate on a single, field-selectable range. A new range may be selected any time the analyzer application changes or any time calibration may require a range change.
To select the operating range, reposition the jumper shown in Figure 3-1 (page 3-3) to the desired location. Each position is labeled as to its fullscale range. Only the analog output (voltage and optional current) is affected by range selection. The digital display always reads 100% oxygen.
3-3 STARTUP PROCEDURE
Inject a suitable on-scale gas (not actual sample) through the analyzer. Turn power
ON. If digital display gives overrange indication, the probable cause is the suspension in the detector is hung up. To correct this condition, turn power OFF, tap detector compartment with fingers, wait 30 seconds, turn power ON.
When on-scale reading is obtained, allow analyzer to warm-up for a minimum of one hour with gas flowing. This warm-up is necessary because a reliable calibration is obtainable only after the analyzer reaches temperature stability. Moreover, the resultant elevated temperature will ensure against condensation within, and possible damage to the detector assembly. After warm-up, the digital display or recorder should give stable, drift-free readout. If so, proceed to Section 3-4 below. Otherwise, refer to Section 6, Maintenance and Service.
3-4 CALIBRATION
Calibration consists of establishing a zero calibration point and a span calibration point (see Table 3-1, page 3-4). Zero and span calibration should be performed on the range that will be used during sample analysis. In some applications, however, it may be desirable to perform span calibration on a range of higher sensitivity (i.e., more narrow span) and then move the jumper to the desired operating range. For example, if the operating range is to be 0 to
50% oxygen, span calibration may be performed on the 0 to 25% range to permit use of air as the span standard gas.
a. Calibration with Zero and Span
Standard Gases
NOTE:
The same flow rate must be maintained for zero, span, and sample to avoid measure error. The exhaust is vented to the atmosphere to avoid back pressure. The following procedure is based on the standards in
Table 3-2 (page 3-6). Performance specifications are based on recorder output.
Set Zero Calibration Point
Inject nitrogen zero standard gas through analyzer at suitable flow rate, preferably 250 cc/min. Allow gas to purge analyzer for a minimum of three minutes.
Adjust ZERO control so that the reading on the digital display or recorder is zero
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-1
Instruction Manual
748213-S
April 2002
Set Span Calibration Point
Inject span standard gas (see Table
3-1, page 3-4) through the analyzer at the same flow rate as was used for zero standard gas. Allow gas to purge analyzer for a minimum of three minutes.
Adjust SPAN control so that reading on display or recorder is appropriate to the span standard gas.
3-5 COMPENSATION FOR COMPOSITION OF
BACKGROUND GAS
Any gas having a composition other than
100% oxygen contains background gas.
The background gas comprises all nonoxygen constituents. Although instrument response to most gases other than oxygen is comparatively slight, it is not in all cases negligible. The contribution of these com-
Model 755R ponents to instrument response is a function of the span and range used, and can be computed for each individual case.
If the zero and span standard gases contain the same background gas as the sample, the routine standardization procedure automatically compensates for the background components. Therefore, the zero and span standard gases would introduce no error.
If the background gas in the sample is different from that in the zero and/or span standard gas(es), background effects must be taken into consideration to ensure correct readout. During adjustment of the front panel ZERO and SPAN controls (see Figure 1-1, page 1-1), the instrument is not set to indicate the true oxygen content of the zero and span standard gases. It is set to indicate a slightly different value, relative to background gas, calculated to provide correct readout during subsequent analysis of the sample gas.
3-2 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R
15
SPAN
E9
E10
100%
E23
E24
50%
8
E11 E13
7
E15 E17 E19 TP5 TP6 TP7 TP8 TP9 TP10 TP11 TP16 TP17 TP18 TP19 TP20
E12
25%
E14 E16
10%
E18
5% 2.5%
E20
1%
3
4
1
2
R8
R9
CR2
I G O
U6
U3
U2
R5
R6
R3
R4
R2 R1
16
1
G
I
O
C5 C4
U4
C2
U1
U5
I G O
C3 CR1 C1
J1
U18
C60
C56
C58
C59
5V 1V .1V .01V
E21 E1 E3 E5 E7
E22 E2 E4 E6 E8
C49
C61
C13
C12
R43
CR5
T1
C55 U12
U15
U17
U14
C68
R42
CR4
C45
+
T1
U21
U16 C57 U13
R57 R58 R59 R60
C14
U4
R21 R2
R50
C53
R20
J2
BL
Instruction Manual
748213-S
April 2002
5 4 3 2 6 10 13
R88 R30 R29 R100
12
9
14
11
R101
C18
U1
1. RECORDER OUTPUT selector plug Provides selectable output of 10 mV, 100 mV, 1 V or 5 V for a voltage recorder.
2. DIGITAL READOUT (R100) Calibration of digital readout.
3. AMPLIFIER U8 ZERO (R29)
4. RESPONSE TIME (R30)
Initial factory zeroing of amplifier U8.
Adjustment of electronic response time.
5. FULLSCALE OUTPUT (R88)
6. DETECTOR COARSE ZERO (R9)
7. CURRENT OUTPUT ZERO (R1)
8. CURRENT OUTPUT SPAN (R2)
9. ALARM 2 CALIBRATION (R67)
10. ALARM 2 SETPOINT (R68)
11. ALARM 2 DEADBAND (R78)
12. ALARM 1 CALIBRATION (R63)
13. ALARM 1 SETPOINT (R64)
14. ALARM 1 DEADBAND (R73)
15. OUTPUT RANGE selector plug
Setting fullscale for 1 V, 0.1 V and 10 mV outputs.
Coarse adjustment of detector zero by shifting the position of the detector within the magnetic field. It is adjusted during factory checkout, and does not require readjustment except if detector is replaced.
Located on Current Output Board, adjustment for zero-level current output, i.e., 4mA or 0mA
Located on Current Output Board, adjustment for fullscale current output: 20mA
Initial calibration of ALARM 2 circuit.
Continuously variable adjustment of setpoint for ALARM 2 circuit, for actuation of external, customer supplied control device(s). Adjustment range is 0 to 100% of fullscale span.
Adjustment of ALARM 2 deadband circuit from 1% to 20% of fullscale. Deadband is essentially symmetrical with respect to setpoint.
Initial calibration of ALARM 1 circuit.
Continuously variable adjustment of setpoint for ALARM 1 circuit, for actuation of external, customer supplied control device(s). Adjustment range is 0 to 100% of fullscale span.
Adjustment of ALARM 1 deadband circuit from 1% to 20% of fullscale. Deadband is essentially symmetrical with respect to setpoint.
Selectable fullscale output range.
16. DETECTOR ISOLATION plug
DIGITAL DISPLAY
ZERO control (R13)
SPAN control (R20)
For servicing and testing of the Control Board.
Display (viewed on front panel) indicates oxygen content of sample.
Accessible on front panel, use to establish zero-calibration point.
Accessible on front panel, use to establish span calibration point.
Figure 3-1. Control Board - Adjustment Locations
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-3
Instruction Manual
748213-S
April 2002
Model 755R
RANGE % OXYGEN
0 to 1
0 to 2.5
0 to 5
0 to 10
0 to 25
0 to 50
0 to 100
RECOMMENDED ZERO
STANDARD GAS
Nitrogen
Nitrogen
Nitrogen
Nitrogen
Nitrogen
Nitrogen
Nitrogen
RECOMMENDED SPAN
STANDARD GAS
0.9% O2, balance N2
2.3% O2, balance N2
4.5% O2, balance N2
9% O2, balance N2
Air (20.93% O2)
45% O2, balance N2
100% O2
Table 3-1.
Calibration Range for Various Zero-Based Operating Ranges a. Oxygen Equivalent Value of Gases
For computation of background corrections, the analyzer response to each component of the sample must be shown.
Table 3-2 (page 3-6) lists the percentage oxygen equivalent values for many common gases.
In equation form:
%O2 Equivalent of Gas =
Analyzer Response to Gas
Analyzer Response to O
2
X 100%
The percentage oxygen equivalent of a gas is the instrument response to the given gas compared to the response to oxygen, assuming that both gases are supplied at the same pressure .
To select a random example from Table
3-2, if analyzer response to oxygen is
+100%, the response to xenon would be
-1.34%.
The oxygen equivalent of a gas mixture is the sum of the contribution of the individual gas components.
Example: Zero Based Range
At lower range limit, i.e., 0% oxygen, composition of sample is 80% CO 2 , 20% N 2 .
From Table 3-1 (page 3-4), the % oxygen equivalents are CO 2 . -0.623 and N 2 , -0.358%.
% oxygen equivalent of mixture =
0.8 x (-0.623) + 0.2 x (-0.358) = (-0.4984) + (-0.0716) = -0.570% Oxygen b. Computing Adjusted Settings for Zero and Span Controls
During instrument calibration, adjusted values may be required in setting the ZERO and SPAN controls to correct for the magnetic susceptibility of the background gas.
The quantities are defined as follows:
BGGst = Oxygen equivalent of background gas in standard gas (see Table 3-2, page 3-6)
BGGs = Oxygen equivalent of background gas sample (see Table 3-2, page 3-6)
OP = operating pressure. Unless special pressure corrections are to be made, the zero standard, span standard and sample gases must all be admitted at the same pressure.
3-4 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R
Instruction Manual
748213-S
April 2002
Use the following equation to compute the adjusted settings for the ZERO and SPAN controls:
Adjusted % O 2 for standard gas =
(A)[100 + (B-C)] - 100 [B-C]
100
Where:
A = true % O 2 of standard gas
B = BGGs
C = BGGst
Example:
Background gas in sample is CO2, oxygen equivalent = -0.623%
Zero gas is 100% N2
Span standard gas is air: 21% O2, 79% N2
Background gas in zero and span standard gases is N2, oxygen equivalent = -0.358%
With N2 zero standard gas flowing, ZERO control is adjusted so digital display reads:
0[100+(-0.623-(-0.358))] - 100[-0.623-(-0.358)]
100
= 0.265% O
2
With air flowing, SPAN control is adjusted so the digital display reads:
21[100 - 0.265) - 100 (-0.265)
100
= 21.209% O
2
≅
21.21
In two limiting cases, the general equation is reduced to simpler forms.
1. If the span standard gas is 100% oxygen, the adjusted oxygen value for setting the SPAN control is the same as the true value (i.e., 100% oxygen).
2. If the zero standard is an oxygen-free zero gas, the adjusted value for setting the ZERO control =
BGGst - BGGs. (If the oxygen-free zero gas is more diamagnetic than the background gas in the sample, this difference is negative. The negative value may be set on the digital display or the recorder if provided with below-zero capability.)
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-5
Instruction Manual
748213-S
April 2002
Model 755R
GAS
Acetylene, C 2 H 2
Allene, C 3 H 4
Ammonia, NH 3
Argon, A
Bromine, Br 2
1,2-Butadiene C 4 H 6
1,3-Butadiene C 4 H 6 n-Butane, C 4 H 10 iso-Butane, C 4 H 10
Butene-1, C 4 H 8 cis Butene-2, C 4 H 8 iso-Butene, C 4 H 8 trans butene-2, C 4 H 8
Carbon Dioxide CO 2
Carbon Monoxide, CO
Ethane, C 2 H 6
Ethylene, C 2 H 4
Helium, H e n-Heptane, C 7 H 16 n-Hexane, C 6 H 12 cyclo-Hexane, C 6 H 12
Hydrogen, H 2
Hydrogen Bromide, Hbr
Hydrogen Chloride, HC1
Hydrogen Fluoride, HF
Hydrogen Iodide, HI
Hydrogen Sulphide, C 2 S
Kryton, Kr
Methane, CH 4
Neon, Ne
Nitric Oxide, NO
Nitrogen, N 2
Nitrogen Dioxide, NO 2 n-Octane, C 8 H 18
Oxygen, O 2 n-Pentane, C 5 H 12 iso-Pentane, C 5 H 12 neo-Pentane, C 5 H 12
Propane, C 3 H 8
Propylene, C 3 H 6
Water, H 2 O
Xenon, Xe
Table 3-2. Oxygen Equivalent of Common Gases
-0.968
-0.651
-0.253
-1.403
-0.751
-0.853
-0.512
-0.205
+44.2
-0.358
-0.354
-0.789
-0.553
-0.059
-2.508
-2.175
-1.915
-0.117
EQUIV. % AS
-0.612
-0.744
-0.479
-0.569
-1.83
-1.047
-1.944
-1.481
-1.485
-1.205
-1.252
-1.201
-1.274
-0.623
+28.7
-2.840
+100.0
-1.810
-1.853
-1.853
-1.135
-0.903
-0.381
-1.340
3-6 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R
Instruction Manual
748213-S
April 2002
3-6 SELECTION OF SETPOINTS AND DEAD-
BAND ON ALARM OPTION
The ALARM 1 and ALARM 2 setpoint adjustments (see Figure 3-1, page 3-3) are adjustable for any desired value from 1% to 100% of the fullscale analyzer span. The adjustment screws are graduated from 0 to 10.
Required dial settings for both setpoint adjustments may be determined from either Figure 3-2 or the appropriate equation that follows:
• Zero-based operating range
• Required control setting =
(desired alarm setpoint)(10) fullscale span
Figure 3-2 example:
Operating range, 0 to 5% oxygen
Desired ALARM 1 setpoint = 4% oxygen
Turn potentiometer R64 to 8
Desired ALARM 2 setpoint = 3% oxygen
Turn potentiometer R68 to 6
The desired deadband may be selected via the appropriate trimming potentiometer, R73, for ALARM 1 deadband adjustment and R78 for ALARM 2 deadband adjustment. For any setpoint, deadband is adjustable from 1% of fullscale (counterclockwise limit) to 20% of fullscale (clockwise limit). Deadband is essentially symmetrical with respect to setpoint.
3-7 CURRENT OUTPUT BOARD (OPTION)
The Current Output is set at the factor for 4 to
20 mA. If a 0 to 20 mA output is required, readjust both the zero and span potentiometers
(R1 and R2) on the Current Output Board.
RANGE
%
OXYGEN
0 to 2.5
0 to 5
0 to 10
0 to 25
0 to 50
0 to 100
PERCENTAGE OXYGEN READOUT versus
ALARM SETPOINT DIAL READING
Percentage Oxygen Readout
0.0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1.0
0 1 2 3 4 5 6 7 8 9 10
Setpoint Dial Reading
Percentage Oxygen Readout
0.0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1.0
0 1 2 3 4 5 6 7 8 9 10
Setpoint Dial Reading
Percentage Oxygen Readout
0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0
0 1 2 3 4 5 6 7 8 9 10
Setpoint Dial Reading
Percentage Oxygen Readout
0 1 2 3 4 5 6 7 8 9 10
0 1 2 3 4 5 6 7 8 9 10
Setpoint Dial Reading
Percentage Oxygen Readout
0 2.5 5 7.5 10 12.5 15 17.5 20 22.5 25
0 1 2 3 4 5 6 7 8 9 10
Setpoint Dial Reading
Percentage Oxygen Readout
0 5 10 15 20 25 30 35 40 45 50
0 1 2 3 4 5 6 7 8 9 10
Setpoint Dial Reading
Percentage Oxygen Readout
0 10 20 30 40 50 60 70 80 90 100
0 1 2 3 4 5 6 7 8 9 10
Setpoint Dial Reading
Figure 3-2. Dial Settings for Alarm Setpoint
Adjustments
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-7
Instruction Manual
748213-S
April 2002
Model 755R
3-8 ROUTINE OPERATION
After the calibration procedure of Section 3-4
(page 3-1), admit sample gas to the analyzer at the same pressure and the same flow rates used for the zero and span gases. The instrument will now continuously indicate the oxygen content of the sample gas.
3-9 EFFECT OF BAROMETRIC PRESSURE
CHANGES ON INSTRUMENT READOUT
If the analyzer exhaust port is vented through a suitable absolute backpressure regulator, barometric pressure changes do not affect the percent oxygen readout. However, if the analyzer exhaust port is vented directly to the atmosphere, any change in barometric pressure after instrument standardization will result in a directly proportional change in the indicated percentage of oxygen. This effect may be compensated in various ways. If desired, correction may be made by the following equation:
True % Oxygen = (Pst/Pan)(Indicated % Oxygen)
Where:
Pst = Operating pressure during standardization
Pan = Operating pressure sample analysis
Example: U.S. Units
Pst = 760 mm Hg
Pan = 740 mm Hg
Indicated % O2
True % O2 = (760/740)(40%) = 41.1% O2
Example: S.I. Units
Pst = 101 kPa
Pan = 98.2 kPa
Indicated % O2 = 40%
True % O2 = (101/98.2)(40%) = 41.1%
O2
3-10 CALIBRATION FREQUENCY
The appropriate calibration interval will depend on the accuracy required in the particular application, and is best determined by keeping a calibration log. If the analyzer exhaust port is vented directly to the atmosphere, the greatest source of error is normally the variation in barometric pressure. If desired, effects of barometric pressure variation can be minimized by calibrating immediately before taking readings, for example, at the beginning of each shift.
3-8 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R
Instruction Manual
748213-S
April 2002
SECTION 4
THEORY
4-1 PRINCIPLES OF OPERATION
Oxygen is strongly paramagnetic while most other common gases are weakly diamagnetic.
The paramagnetism of oxygen may be regarded as the capability of an oxygen molecule to become a temporary magnet when placed in a magnetic field. This is analogous to the magnetization of a piece of soft iron.
Diamagnetic gases are analogous to non-magnetic substances.
With the Model 755R, the volume magnetic susceptibility of the flowing gas sample is sensed in the detector/magnet assembly. As shown in the functional diagram of Figure 5-1, a dumbbell-shaped, nitrogen-filled, hollow glass test body is suspended on a platinum/nickel alloy ribbon in a non-uniform magnetic field.
Because of the “magnetic buoyancy” effect, the spheres of the test body are subjected to displacement forces, resulting in a displacement torque that is proportional to the volume magnetic susceptibility of the gas surrounding the test body.
Measurement is accomplished by a null-balance system, where the displacement torque is opposed by an equal, but opposite, restorative torque. The restorative torque is due to electromagnetic forces on the spheres, resulting from a feedback current routed through a titanium wire conductor wound lengthwise around the dumbbell.
In effect, each sphere is wound with a one-turn circular loop. The current required to restore the test body to null position is directly proportional to the original displacement torque, and is a linear function of the volume magnetic susceptibility of the sample gas.
The restoring current is automatically maintained at the correct level by an electro-optical feedback system. A beam of light from the source lamp is reflected off the square mirror attached to the test body, and onto the dual photocell.
The output current from the dual photocell is equal to the difference between the signals developed by the two halves of the photocell.
This difference, which constitutes the error signal, is applied to the input of an amplifier circuit that provides the restoring current.
When the test body is in null position, both halves of the photocell are equally illuminated, the error signal is zero, and the amplifier is unequal. This condition results in application of an error signal to the input of the amplifier circuit. The resultant amplifier output signal is routed through the current loop, thus creating the electromagnetic forces required to restore the test body to null position.
Additionally, the output from the amplifier is conditioned as required to drive the digital display, and recorder if used. The electronic circuitry involved is described briefly in Section 4-3 (page 4-4) and in greater detail in
Section 5.
Rosemount Analytical Inc. A Division of Emerson Process Management Theory 4-1
Instruction Manual
748213-S
April 2002
4-2 VARIABLES INFLUENCING PARAMAG-
NETIC OXYGEN MEASUREMENTS
Variables that influence paramagnetic oxygen measurements include: operating pressure
(See Section 4-3a, page 4-4), sample temperature, interfering sample components, and vibration (See Section 2-1d, page 2-1).
a. Pressure Effects
Although normally calibrated for readout in percent oxygen, the Model 755R actually responds to oxygen partial pressure.
The partial pressure of the oxygen component in a gas mixture is proportional to the total pressure of the mixture. Thus readout is affected by pressure variations.
For instance, assume that an instrument is calibrated for correct readout with a standard gas containing 5% oxygen, admitted at the normal sea level atmospheric pressure of 14.7 psia (101.3 kPa).
If the operating pressure now drops to one-half the original value (i.e., to 7.35
psia {50.65 kPa}) and the calibration controls are left at the previously established settings, the display reading for the standard gas will drop to 2.5%.
Model 755R
It is therefore necessary to calibrate the instrument at the same pressure that will be used during subsequent operation, and to maintain this pressure during operation.
Typically, the sample gas is supplied to the analyzer inlet at slightly above ambient pressure, and is discharged to ambient pressure from the analyzer outlet.
However, in some applications, it is necessary to insert an absolute back pressure regulator into the exhaust line to prevent the readout error that would otherwise result from fluctuations in exhaust pressure. The regulator must be mounted in a temperature-controlled housing (See
Section 2-3c, page 2-3).
Operation at negative gauge pressure is not normally recommended, but is used in certain special applications (See Section
2-3d, page 2-4).
CAUTION
PRESSURE MINIMUM
Never subject the sensing unit to an absolute pressure of less than 500 mm Hg
(66.7 kPa).
4-2 Theory Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R
Instruction Manual
748213-S
April 2002
Displacement
Torque
Restoring
Torque
Electromagnetic
Axis
Balancing
Weight
Electromagnetic
Axis
Restoring
Current
Balancing Weight
Nitrogen-Filled Hollow Glass
Test Body
Mirror
Titanium Wire Conductor
Restoring
Current
Platinum/Nickel Alloy
Suspension Ribbon
TEST BODY DETAIL
Displacement
Torque
Restoring
Torque
Magnet
Shaded Pole Pieces (4)
Dual Photocell
BT1, BT2
Restoring
Current
Test Body
Source Lamp
DS1
DETECTOR/MAGNET
ASSEMBLY
CONTROL
ASSEMBLY
Zero
Span
% Oxygen
Readout
Figure 4-1. Functional Diagram of Paramagnetic Oxygen Measurement System
Rosemount Analytical Inc. A Division of Emerson Process Management Theory 4-3
Instruction Manual
748213-S
April 2002
Model 755R
Shaded
Pole
Piece
Sphere
(Magnetic Susceptibility = k o
)
F k
Sample Gas
(Magnetic Susceptibility = k )
Note:
As percentage of oxygen in sample gas increases, displacement force (F k
) increases.
Figure 4-2. Spherical Body in Non-Uniform Magnetic Field
4-3 ELECTRONIC CIRCUITRY
Electronic circuitry is shown in the Control
Board schematic diagram, Drawing 652826, and is described briefly in the following sections. For detailed circuit analysis, refer to
Section 5 Circuit Analysis.
a. Detector/Magnet Assembly
A cross-sectional view of the optical bench and detector assemblies is shown in Figure 4-3B, page 4-7. Source lamp
DS1, powered by a supply section within the Power Supply Board assembly (See
Section 4-3c, page 4-5) directs a light beam onto the mirror attached to the test body. The mirror reflects the beam onto dual photocell BT1, BT2.
The difference between the signals developed by the two halves of the photocell constitutes the error signal supplied to the input of amplifier U1 on the Control Board assembly. Amplifier U1 drives U2 which,
4-4 Theory in turn, supplies the restoring current to the titanium wire loop on the test body
(See Section 4-1, page 4-1).
Detector temperature is sensed by thermistor RT1, an integral part of the detector assembly (See Figure 4-3B, page 4-7).
The thermistor provides the input signal to the detector temperature control section of the Power Supply Board assembly:
HR1, mounted on the top of the magnet, and HR2, mounted permanently on the rear of the detector assembly.
b. Control Board and Associated Circuitry
The Control Board consists of signal conditioning and control circuitry.
This circuitry includes the following:
Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R
Input Amplifier U1
This amplifier receives the error signal from the dual photocell of the detector assembly and drives amplifier U2.
Amplifier U2 and Associated Zero Adjustment
Amplifier U2 supplies the restoring current to the titanium wire loop of the test body within the detector assembly. Front panel
ZERO Control R13 applies an adjustable zero biasing signal to the input of U2 to permit establishing a zero calibration point on the display or recorder. With zero standard gas flowing through the analyzer, the ZERO control is adjusted for the appropriate reading.
Amplifier U4 and Associated Span
Adjustment
Amplifier U4 and associated feed back resistors provide a signal amplification of
X4. Front panel SPAN adjustment R20 modifies the value of the input resistance and hence the signal amplification factor.
Adjustment range is approximately ±30%.
Amplifier U8
This unity gain amplifier provides zeroing capability and a buffered output for the anticipation circuit feeding U10.
Amplifier U10
U10 is an inverting buffer amplifier that incorporates an anticipation arrangement in its input network, thus providing slightly faster response on the readout device(s).
Potentiometer R30 provides a continuously variable adjustment of 5 to 25 seconds for the electronic anticipation time and is factory-set for 20 seconds.
Since the anticipation network attenuates the signal, a gain of 10 is provided in U10 to restore the signal to the desired fullscale range of 0 to 10 VDC.
Instruction Manual
748213-S
April 2002
The output signal from U10 is routed to two output circuits: a digital and an analog.
In the Digital Output Circuit, the signal from U10 passes to an integrating analog-to-digital converter. The resulting digital signal drives the liquid crystal display.
In the Analog Output Circuit, the output from U10 is provided as an input to the recorder output amplifier. This circuitry provides scale expansion, and amplification preparatory to use for potentiometric recorder, voltage-to-current conversion for current recorder, and/or alarm functions.
Potentiometric output is strap-selectable for 0 to 10 mV, 0 to 100 mV, 0 to 1 V, or 0 to 5VDC. Potentiometer R88 permits adjustment of recorder span on 0 to 1 V, 100 mV and 10 mV outputs.
c. Power Supply Board Assembly
The Power Supply Board assembly contains power supply and temperature control circuitry. The assembly is mounted within the analyzer case.
As shown in DWG 617186, the various circuits operate on main power transformer T1. During instrument assembly, the two primary windings of T1 are factory-connected for operation on either 115
VAC or 230 VAC, as noted on the name rating plate.
The same circuit board contains the following:
Source Lamp Power Supply Section
This circuit provides a regulated output of
2.20 VDC to operate incandescent source lamp DS1 within the optical bench assembly. One secondary of main power transformer T1 drives a fullwave rectifier consisting of CR7 and CR8. The output of
DS1 is held constant by a voltage regulator circuit utilizing U7, Q4 and Q5.
Rosemount Analytical Inc. A Division of Emerson Process Management Theory 4-5
Instruction Manual
748213-S
April 2002
±15 V Power Supply Section
This section provides DC voltage required for various amplifiers and other circuits.
Fullwave rectifier bridge CR5 provides both positive and negative outputs. Each is routed through an associated series type integrated circuit, voltage regulator, providing regulated outputs of +15 V and
-15 V.
Detector Temperature Control Section
This section maintains the detector at a controlled temperature of 150 °F (66°C).
Temperature is sensed by RT1, a resistance element permanently attached to the detector assembly. The signal from the sensor is applied to amplifier AR6, which drives transistors Q2 and Q3, thus controlling application of DC power from full wave rectifier bridge CR6 to two heaters within the detector/magnet assembly:
HR1, mounted on the top of the magnet and HR2, permanently mounted on the rear of the detector assembly.
Model 755R
Detector Compartment Temperature
Control Section
This section maintains the interior of the detector compartment at a controlled temperature of 140 °F (60°C). Temperature is sensed by a thermistor located in the detector compartment and plugged into the Control Board assembly.
The circuit provides an on-off control of heater element HR3 via TRIAC element
Q7. Heater HR3 is a part of the heater/fan assembly.
a. Isolated Current Output Board (Optional)
An isolated current output is obtainable by insertion of an optional plug-in circuit board into receptacle J1 on the Control
Board (see Figure 3-1, page 3-3). The current outputs available by this board are
0 to 20 mA or 4 to 20 mA.
4-6 Theory Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R
Instruction Manual
748213-S
April 2002
Sample Pre-Heating Coil
Sample Inlet Tube
Sample Outlet Tube
Detector Assembly
Optical Bench Assembly
Mounting Screws (2)
A. Exploded View of Detector/Magnet Assembly
Magnet Assembly
Integral Heater (HR2)
Sensor (RT1)
Integral 5-Micron
Diffusion Screen
Test Body
Mirror
Photocell
Lock Screws (2)
Dual Photocell
Optical Bench Assembly
Detector Assembly
Source Lamp
B. Sectional Top View of Optical Bench and
Detector Assemblies
Lamp Retaining
Set Screw
Dual Photocell
Lamp Viewing Hole
Source Lamp
Assembly
C. Exploded View of Optical Bench Assembly
Figure 4-3. Detector/Magnet Assembly
Rosemount Analytical Inc. A Division of Emerson Process Management Theory 4-7
Instruction Manual
748213-S
April 2002
Model 755R
4-8 Theory Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R
Instruction Manual
748213-S
April 2002
5-1 CIRCUIT OPERATION
The electronic circuitry of the Model 755R
Oxygen Analyzer consists of the following:
• A detector compartment heater circuit.
• A detector heater circuit.
• A ±15 VDC power supply.
• A voltage regulating circuit for a stable light source.
• A detector circuit with a first-stage amplifier to provide a feedback current for mechanical feedback to the detector and a scaling amplifier circuit to give an output change of 0 to +2.5 V for a 0 to
100% change of the operating span.
• A digital output circuit for the digital read-out.
• An analog output circuit for recorder, optional alarms and current output.
5-2 ±15 VDC POWER SUPPLY
Refer to Drawing 617186. The components of the ±VDC power supply circuit are located in the lower left-hand corner of the Power
Supply Board. 19 VAC should be measured with respect to ground at CR5 (WO4). +15
VDC should be measured at the C27 (+) lead and -15 VDC at the C28 (-) lead. If the specified voltage measurements are obtained, the power supply is working correctly.
5-3 CASE HEATER CONTROL CIRCUIT
The case heater control circuit utilizes four voltage-comparators (LM339 quad comparator). An understanding of how one of these comparators functions is necessary before any circuit analysis can be attempted.
SECTION 5
CIRCUIT ANALYSIS
In Figure 5-1 (page 5-2), comparators 1 and
2 are depicted having a comparator within an overall comparator symbol. Also within this symbol, the base of the NPN transistor is connected to the output of the comparator. A -15 VDC is supplied to the emitter.
The collector is illustrated as the overall output for the comparator package.
When the non-inverting terminal of comparator 2 is more positive than the inverting terminal, the transistor does not conduct and the collector of the transistor or comparator output is at whatever potential is then present on the collector.
When the non-inverting terminal of comparator 2 is less positive (more negative) than the inverting terminal, the transistor conducts and the output of the comparator is
-15 V. This value is the output of the OR circuit.
Comparator 2 is biased at 0 volts on the inverting terminal. Comparator 1 is biased at about 159 mV on the non-inverting terminal.
Positive feedback or hysteresis is built into each comparator circuit for stability or positive action. This is achieved by the 20 M resistances, R70 and R73.
An approximate 8 V peak-to-peak AC signal is applied to comparators 1 and 2. As the signal starts going positive, comparator 2 transistor ceases conducting and comparator 1 transistor is off.
When the signal exceeds the +159 mV on the non-inverting terminal, it turns on comparator 1 and the output is -15 V. Comparator 1 stays on until the signal drops below +159 mV, at which time the output will be the value of the OR bus.
As the AC signal goes negative with respect to ground, the transistor of comparator 2 conducts and the output is again -15 V. The
Rosemount Analytical Inc. A Division of Emerson Process Management Circuit Analysis 5-1
Instruction Manual
748213-S
April 2002
Model 755R output remains at -15 VDC until the incoming signal crosses zero value and the positive signal causes the comparator 2 transistor to cease to conduct.
Summing the effects of the two comparators in the OR circuit results in no output from the comparators for about 4 ° of the sine wave,
2 ° after the signal goes positive (0 to 2°) and
2 ° before the positive signal reaches 180°
(178 ° to 180°).
During the period that neither comparator is conducting, the value on the OR bus is the potential from the temperature-sensing bridge plus the effect of the ramp generator, probably -1.88 ±0.03 V.
The on-off effect of the comparators to the
OR circuit results in application of a positive-going pulse (from -15 V to -1.89 V) to
5-2 Circuit Analysis the temperature bridge at the rate of 120 pulses per second.
Capacitor C36 is added to the input circuit to delay the incoming AC signal so that the pulses will occur at or just after the line frequency crossover point.
Circuits for a ramp generator and a temperature-sensing bridge are part of the case heater control circuit (See Figure 5-2, page
5-3 and Figure 5-3, page 5-3).
On initial application of power to comparator of Figure 5-2 (page 5-3), no potential exists on the inverting terminal because no charge exists on capacitor, C37. If the transistor of comparator 3 does not conduct, +15V is at the output terminal. With +15V at the output, the potential on the non-inverting terminals will be about ±2.3 V because of the resistance divider, R75, R76.
100µ
-1.7V
-15V
INPUT
159mV
COMP 1
COMP 2
R69
2M
+15V
R71
21.5K
0°
ON
OFF
180°
OFF
ON
360°
0°
ON
OFF
180°
-
+
1
+15V
-15V
R70
20M
R72
4.75K
R68
3.3K
C38
0.18uF
-
+
2
+15V
-15V
R73
20M
-1.88 VDC
Source
OUTPUT
Figure 5-1. Two-Comparator OR Circuit
Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R
Instruction Manual
748213-S
April 2002
120 V
RMS
T1
19 VAC
TO POWER
SUPPLY
19 VAC
12
CR9
R67
10K
R72
4.75K
C36
.18uF
CR10
R69
2 M
R71
21.5K
R68
3.3K
-15V
+15V
-
+
1
R70
-
20M
2
+
R73
20M
R82
9.07K
C39
.01uF
RT1
R74
590K
C37
1.0uF
R85
R83
63.4K
11.0K
R84
169K
-
+
3
R78
249K
R75
210K
R76
37.4K
C40
2200uF
R80
10K
R77
10K
R79
10K
Q6
-
+
4
R86
20M
R87
10K
CR11
R81
56.2
C38
.18uF
T2
-15V
Figure 5-2. Case Heater Control Circuit
-15V
OFF OFF OFF
+2.3V
-2.3V
INPUT FROM
MULTIVIBRATOR
R74
590K
R82
9.09K
RT1
C37
1.0uF
-15V to 1.88V ±0.3V
R83
63.4K
R76
TO
37.4K
COMPARATOR
R84
169K
-15V
R78
-
+3
249K
+15V
-15V
R75
210K
C40
2200uF
6 Hz
R77
10K
+15V
R79
10K Q6
R80
10K
R81
56.2
C38
.18uF
R87
10K
-15V
Figure 5-3. Ramp Generator Circuit
T2
Rosemount Analytical Inc. A Division of Emerson Process Management Circuit Analysis 5-3
Instruction Manual
748213-S
April 2002
Model 755R
Capacitor C37 will now start to charge positively through R78. When the positive potential across C37 and at the inverting terminal of comparator 3 exceeds the potential on the non-inverting terminals, the transistor conducts. The output is -15 V. A full 30 V drop appears across R77.
The potential on the non-inverting terminal will now be about -2.3 V. C37 will not discharge through R78 until its potential exceeds that on the non-inverting terminal. At that time, comparator 3 will switch polarity and start charging
C37 again. The result is that the potential across C37 will vary almost linearly with time and form a ramp signal of about 6 Hz.
As the potential across C37 increases and decreases linearly, it affects the potential at the top of the bridge circuit between R82 and
R83 through R74. Because of the ramp action charging and discharging C37, the potential between R82 and R83 varies approximately from -1.85 V to -1.92 VDC.
The temperature sensing device, RT1, in the bridge circuit is a thermistor. The bridge is designed to control the temperature in the case at 135 °F (57°C). When the temperature is
135 °F (57°C), the resistance of the thermistor
RT1 will be at its lowest and the potential at the junction of RT1 and R84 should be the same as the junction of R82 and R83. Comparator 4 (See Figure 5-4, page 5-6) does not allow pulses from the OR circuit (comparators
1 and 2) to operate Q6 or Triac Q7 in the case heater (See Figure 5-5, page 5-7).
Theoretically, at 135 °F (57°C) the potential at the junction of RTR1 and R84 is -1.85VDC.
This is equivalent to a resistance of 21.2 K. By substituting a decade box for the thermistor and placing 20.2 K into the bridge, the heater should be off. With 22.7 K, the heater should be full on.
Since the potential at the junction of R82 and
R83 can vary between 1.85V and 1.92V according to the 6 Hz ramp, and the potential at the junction of RT1 and R84 may vary around or within these limits, depending on temperature, the error signal to comparator 4 may
5-4 Circuit Analysis vary from 0mV to some absolute value. The polarity of the error signal will depend on the deviation from the desired temperature and the ramp value at the function of R82 and
R83.
The input from the OR circuit comparator (See
Figure 5-1, page 5-2) is either -15 VDC or the ramp effect on the bridge. When -15V, the junction of R82 and R83 is also this value.
The error signal into comparator 4 is negatively large to the inverting terminal. Comparator 4 output transistor does not conduct.
The base of Q6 is positive; therefore, Q6 does not conduct and a charge builds up on capacitor C38.
The input from the OR comparators 1 and 2 form multivibrator circuit, pulses 120 times a second. For about 100 microseconds the junction of R82 and R83 is some value between -1.85 V and -1.92 V, depending on the ramp generator. For this brief period of time
(one pulse), comparator 4 compares the potential of junction R82, R83 with junction RT1,
R84 of the bridge circuit. If the temperature at
RT1 is low, the potential at the non-inverting terminal of comparator 4 is more negative and the output is -15 V.
The base of Q6 is zero, because of the voltage drops across R79 and R80. Therefore,
Q6 conducts. Energy, stored in C38, flows through Q6 as current and capacitor C38 discharges to zero potential. No current flows through the primary winding of transformer
T2.
At the end of the 100 microsecond pulse, the
NPN transistor in the output of comparator 4 ceases to conduct, so the signal on the base of Q6 is +15V. Q6 ceases to conduct. C38 starts to charge, driving electrons (current) through the primary of T2. This induces a pulse into the secondary of T2 and to the gate of Triac Q7 turning it on.
At the beginning of the next 100 microsecond pulse, comparator 4 output is again -15V, with zero volts on the base of Q6. Q6 again conducts, discharging C38. At the end of the 100 microsecond pulse, Q6 ceases to conduct.
Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R
C38 charges and a pulse appears at the gate of Triac Q7, turning it on again.
The charging time for C38 is about one-half a time constant (C38, R87) and ten time constants (R81, C38) are available for discharging C38.
The above action is repeated as long as the temperature is low, causing an error between
R82, R83 junction and RT1, R84 junction. As the temperature approaches the desired case temperature of 135 °F (57°C), differences between these two junctions will exist for only part of each ramp and the number of pulses operating Q7 will be proportional to the amount of error sensed by the 6 Hz ramp.
The pulses arrive at Q7 just as the supply AC line voltage is passing the zero-volt crossover
Instruction Manual
748213-S
April 2002 point. The purpose of C36 is to delay the timing pulse, relative to line frequency, so that a pulse arrives at the gate of Triac Q7 as the line potential just passes the zero-volt crossover point (0 ° and 180° of line phase).
Varistor, RV1 is a temperature sensitive resistance device. When case temperature is low, such as ambient, the value of RV1 is low.
Applying power at that temperature might cause a current surge to damage Triac Q7.
RV1 with its low initial value of resistance acts as a bypass and most of the current is shunted through it. As the temperature increases and approaches the desired case temperature, the resistance of RV1 increases to a large value. This limits the current through it and gives fine control of the heater to Triac Q7 and the temperature-sensing circuit.
Rosemount Analytical Inc. A Division of Emerson Process Management Circuit Analysis 5-5
Instruction Manual
748213-S
April 2002
5-4 DETECTOR HEATER CONTROL CIRCUIT
Figure 5-4 below is a simplified heater control circuit drawing for the detector. Heaters 1 and
2 are actually connected in parallel and have a combined resistance of about 17 ohms.
The thermistor resistance (RT1) in the resistance bridge varies inversely with temperature. The bridge is designed to maintain the temperature of the detector at 150 °F (65.5°C).
The junction point between R55 and R56 is maintained at a specific voltage since these resistances maintain a definite ratio. The thermistor resistance is 149 K at 150F
(65.5
°C) and increases rapidly as the temperature decreases. R59 in this bridge circuit represents the setpoint value for temperature.
Suppose that, at temperature, resistance of the bridge (R55, R56, R59 and RT1) equals
149 K.
Model 755R
If the temperature goes down, RT1 increases in resistance and causes the junction of RT1 and R59 to go positive in voltage value. Since
R55 and R56 are of equal resistance, their junction is at zero volts. Therefore, terminal 3 of AR6 is more positive than terminal 2 and the base of Q2 is positive. Q2 conducts, allowing alternating current to flow through heaters 1 and 2. The voltage drop across the heaters, when completely cold, would be about 20 VAC and, when controlling, would be
AC of very low amplitude.
As the temperature increases, the resistance of RT1 decreases and the junction point between RT1 and R59 becomes less positive.
Terminal 3 of AR6 becomes less positive with respect to terminal 2. The output of AR causes Q2 and Q3 to conduct less. When terminal 3 equals terminal 2, or is less than terminal 2, the output of AR6 is zero or less.
Q2 and Q3 do not conduct and the heater would not be supplying heat energy to the detector.
120 V
RMS
HR1 +2
F1
CR6
WO4
25 VAC
+15V
R55
700K
R59
700K
R56
149K
-15V
RT1
C31
.01uF
R58
5M
2
-
3 +
U6
6
R62
1K
R88
5M
CR12
R60
100
Q2
Q3
R61
2.0
Figure 5-4. Detector Heater Control Circuit
5-6 Circuit Analysis Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R
5-5 DETECTOR LIGHT SOURCE CONTROL
CIRCUIT
Refer to Figure 5-5 below. The detector light source control circuit maintains the light output from the bulb (DS1) as uniform as possible, regardless of voltage fluctuations or aging of the bulb.
The power source for the light bulb is a center-tapped secondary of transformer T1. This
AC voltage is rectified by CR7 and CR8 and filtered (C32), presenting an approximate +8.5
V bus to the current-limiting Darlington configuration of Q4.
Q4 controls the basic amount of current through DS1.
Amplifier AR7 has a fixed value, approximately +2.2 VDC on terminal 3. The output of
AR7 is positive, causing Q4 to conduct. As Q4 conducts, electrons flow from the center-tap of
T1 to ground and from ground through DS1 for an input voltage to terminal 2 of AR7, through R66 to develop a bias on the base of
Q5, through Q4 to the +8.5 V bus, and back to the secondary. As Q5 conducts, some of the
Instruction Manual
748213-S
April 2002 current going through DS1 is shunted from the main current path, and goes through Q5, which acts as a variable feedback resistance, goes to the positive output potential of AR7.
As DS1 ages, its light emission decreases and its resistance increases. The current through DS1 tends to decrease, causing a decrease in the voltage drop across DS1 and the input potential to terminal 2 of AR7. Now the output AR7 will increase, causing Q4 to conduct more current through R66. As the potential across R66 increases, Q5 will conduct more current, causing a further increase in current flow through DS1. The net result is that the voltage across DS1 will remain uniform and the operation of Q4 and Q5 will adjust the gain of AR7 to maintain the light emission from DS1 uniform for a long period of time.
Voltage fluctuations in the 115 VAC supply could cause some variation in the amount of current flowing through the bulb DS1. However, the voltage drop across DS1 would cause AR7 to adjust Q4 and the voltage drop across R66 to adjust Q5. The net result would still be uniform current flow through DS1 and uniform light emission.
120 V
RMS
T1
6.1 VAC
CR7
6.1 VAC
CR8
C31
2000uF
+
VR3
9.0V
α+8.5V BUS
+15V
R63
7.5K
R64
14K
R65
α2.2V
4530
3
2
-
+
AR7
C34
.01uF
C35
.01uF
Q5
Q4
R66
1.0
DS1
Figure 5-5. Detector Light Source Control Circuit
Rosemount Analytical Inc. A Division of Emerson Process Management Circuit Analysis 5-7
Instruction Manual
748213-S
April 2002
Model 755R
5-6 DETECTOR WITH FIRST STAGE AMPLIFIER
Refer to Figure 5-6, page 5-9. The detector assembly consists of a test body suspended on a platinum wire and located in a non-uniform magnetic field.
The test body is constructed of two hollow glass spheres forming a dumbbell shape.
They are filled and sealed with pure, dry nitrogen. Around the test body, a titanium wire is chemically etched in order to form a feedback loop that can create a counteracting magnetic force to the test body displacement caused by oxygen concentration in the test assembly magnetic field.
Attached to the center arm of the test body dumbbell is a diamond-shaped mirror. Attached to the mirror are two separate platinum wires in tension with the supports for the test body. The supports are isolated from ground and are electrically connected to the feedback loop and the electronics for that loop. The platinum wires form a fulcrum around which the test body pivots.
The detector operates in the following fashion.
If the sample gas contains oxygen, it collects in the non-uniform magnetic field around the test body. Oxygen, because of its paramagnetic qualities, gathers along the magnetic lines of flux and forces the dumbbell of the test body out of the magnetic field.
A light source is focused on the test body mirror. As the test body moves out of the magnetic field, the mirror distributes light unevenly on two photocells (BT1 and BT2). The photocells create a current proportional to light. This current is converted to a ± voltage by U1 and
U2 located on the connector board in the detector housing. This voltage is then presented to comparator U1 on the controller board. The output of U1 goes to U2. The output of U2 causes current to flow through the feedback loop attached to the dumbbell.
This feedback current creates an electro-magnetic field that attracts the dumbbell and mirror into the test assembly magnetic field until the mirror reflects light almost uni-
5-8 Circuit Analysis formly on each photocell. A current proportional to the oxygen concentration in the magnetic field of the test assembly has to be flowing through the feedback loop in order to maintain balance and provide a reading of the oxygen content of a sample.
Resistances R7, R8 and the resistance of the wire in the feedback loop determine the gain of amplifier U2. The mirror on the dumbbell is positioned by the amount of current in the feedback loop. The mirror reflects light from the source (DS1) to the photocells (BT1,
BT2). This repositioning of the mirror is a form of mechanical feedback to the input of the amplifier U1.
The net result is that the output of U1 could vary from 0 to -70 mV, or 0 to -7.0 V, depending on the range of the instrument. R4,
C3 and R5, C7 form damping circuits for the input amplifier U1 and to smooth out noise that might be introduced by the measurement source.
Diode CR2 is a low-leakage device. Its purpose in the circuit is to ensure that the dumbbell and mirror are positioned correctly with respect to the photocells on initial application of power.
If the dumbbell was out of position on start-up, the mirror might reflect light from the source onto one of the photocells. If the photocell output was positive, the current in the feedback loop would be in the wrong direction and its electromagnetic field would cause the dumbbell to be further repelled from the permanent magnetic field. The result would be error, not balance.
On application of AC power, capacitor C1 has no charge. The current will have to flow through R2. Initially the full 30 V drop (the difference between the +15 VDC and -15
VDC power) will appear cross R2. The cathode of CR2 will be initially at -15 VDC. The anode of CR2 will be some value more positive than -15 VDC. CR2 will conduct. The input terminal of U1 will be negative and the current through the feedback loop around U2
Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R will cause the dumbbell and mirror to be positioned correctly in the test body.
As the charge on C1 increases, the cathode of CR2 becomes more positive. When it exceeds that on the anode, CR2 ceases to conduct and isolates the +15 VDC and -15 VDC power supply from the input circuit.
The front panel zero potentiometer R13 and detector coarse zero potentiometer add or subtract current to the input of U2 to offset any currents that may occur because of any
Instruction Manual
748213-S
April 2002 imbalance in the detector and the photocells
BT1 and BT2.
The output current that U2 must provide to restore the dumbbell is a measure of the displacing force and thus is a function of both (a) the % oxygen concentration of the sample and (b) the sample pressure.
The output from the U1 and U2 loop is further amplified by U4 to provide a 0 to 10 VDC output that constitutes signal V.
BT1
BT2
DS1
CURRENT
FEEDBACK
LOOP
R3
1K
-
+
U1
R1
1K
-
+
U2
DETECTOR
HOUSING
R2
1K
R2
1K
CONTROL
BOARD
+15V
-15V
TP6
C1
3.3uF
R3
110
R1
10
R2
249K
TP7
C2
1000pf
CR2
C3
.47uF
R4
1.13K
R6
118K
C4
.01uF
-
+
U1
R5
2M
TP8
E21 E22
R20
20K
TP20
R21
49.9K
R22
49.9K
R12
200K
R7
1.77K
C7
.47uF
R23
150K
TP10
-
+
U4
SIG. Vx 0 -+ 10V
R10
3.01K
C8
-
.0022uF
U2
+
+15V
R13
20K
FRONT
PANEL
ZERO
-15V
+15V
R9
20K
DETECTOR
COARSE
ZERO
-15V
R8
1.77K
Figure 5-6. Detector with First Stage Amplifier
Rosemount Analytical Inc. A Division of Emerson Process Management Circuit Analysis 5-9
Instruction Manual
748213-S
April 2002
Model 755R
5-7 BUFFER AMPLIFIERS U8 AND U10 WITH
ASSOCIATED ANTICIPATION FUNCTION
Refer to Figure 5-8, page 5-12. U8 is a unity gain amplifier that provides zeroing capability and a buffered output for the anticipation circuit feeding U10.
U10 is an inverting buffer amplifier that incorporates an anticipation arrangement in its input network, thus providing slightly faster response on the readout device(s).
Potentiometer R30 provides a continuously variable adjustment of 5 to 25 seconds for the electronic response time (90% of fullscale) and is factory-set for 20 seconds.
Since the anticipation network attenuates the signal, a gain of 10 is provided by the feedback network associated with U10 to restore the signal to the desired fullscale range of 0 to
10 VDC.
The output signal from U10 is routed to two output circuits:
Digital output circuit (See Section 5-8, page 5-10).
Analog output circuits for recorder, V/I and alarms (See Section 5-9, page 5-11).
5-8 DIGITAL OUTPUT CIRCUIT
Refer to Figure 5-7 below. The output signal from buffer amplifier U10 is routed through an attenuator and filter network to an integrating analog-to-digital converter. It converts the signal into an equivalent digital value in the range of 0.00% to 99.99%. Any value above
99.99% will be preceded by an over-range bit, for example, 1.1123.
The output of the ADC consists of binary-coded decimal characters that are input to the liquid crystal controller and display chip characters sequentially in time. The BCD characters are converted into seven-line codes to drive the bar segments of the liquid crystal display.
A separate regulator circuit, which operates from the +15 VDC supply, provides a regulated 5 VDC for the digital functions associated with the display.
R30
20K
TP11
C31
1.0uF
R36
1M
R31
2K
R61
1M
C36
.47uF
-
+
U10
R37
2M
C38
.22uF
R38
100K
R39
11K
TP16 REF 8052A
DIGITAL
DISPLAY
R49
20K
C40
1.0uF
ADC
71C03
DISPLAY
DRIVER AND
CONTROL
TP16
5V
U18
5V
REGULATOR
+15V
To Analog Output Circuit
(Figure 6-8)
5-10 Circuit Analysis
Figure 5-7. Buffer, Anticipation, and Digital Output Circuits
Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R
5-9 ANALOG OUTPUT CIRCUITS FOR RE-
CORDER AND ALARMS
Refer to Figure 5-8, page 5-12. The analog output circuits utilize two amplifiers, first-stage amplifier and second-stage amplifier.
a. First Stage Amplifier
Permits selection of the desired fullscale oxygen range for the recorder via jumper-selectable signal amplification for scale expansion. This amplifier permits selecting the desired fullscale oxygen range for the recorder by an appropriate jumper selection of one of seven recorder spans. The following recorder spans are available: 1, 2.5, 5, 10, 25, 50, and 100%.
b. Second Stage Amplifier
Provides (a) a jumper-selectable output for a potentiometric recorder and (b) an output to drive the voltage-to-current and/or alarm option(s), if used. This amplifier is an inverting configuration that provides a signal attenuation of 2X, thus reducing the 10-volt fullscale input signal to obtain a 5-volt fullscale output. This output is routed to:
1. Recorder Output Resistor Network. It provides a jumper-selectable output of 0 to 10 mV, 0 to 100 mV, 0 to 1 V, or 0 to 5 VDC for a potentiometric recorder.
Instruction Manual
748213-S
April 2002
2. Current Output Receptacle J1. This connector accepts the optional plug-in current-output board.
3. Dual Alarm Amplifier Circuit. This circuit drives the optional 654019 Alarm
Relay Assembly.
Oxygen is strongly paramagnetic while most other common gases are weakly diamagnetic. The paramagnetism of oxygen may be regarded as the capability of an oxygen molecule to become a temporary magnet when placed in a magnetic field. This is analogous to the magnetization of a piece of soft iron. Diamagnetic gases are analogous to non-magnetic substances.
With the Model 755R, the volume magnetic susceptibility of the flowing gas sample is sensed in the detector/magnet assembly. As shown in the functional diagram of Figure 5-1 (page 5-2), a dumbbell-shaped, nitrogen-filled, hollow glass test body is suspended on a platinum/nickel alloy ribbon in a non-uniform magnetic field.
Because of the “magnetic buoyancy” effect, the spheres of the test body are subjected to displacement forces, resulting in a displacement torque that is proportional to the volume magnetic susceptibility of the gas surrounding the test body.
Rosemount Analytical Inc. A Division of Emerson Process Management Circuit Analysis 5-11
Instruction Manual
748213-S
April 2002
Model 755R
FROM
U10
R84
20K
R85
2M
R102
40K
R86
80K
R52
200K
R87
400K
R53
800K
R82
2M
C46
100
50
25
10
5
2.5
1
-
+
U13
.1uF
Recorder Span
(Jumper Selectable)
TP18
R50A
20K
R50B
20K
R50C
20K
C55
.1uF
-
+
U16
R88
500
E1
5V
E2
To Alarm and V/I
To Recorder
R57
3.83K
R58
909
1V
E3 E4 Recorder Output
(Jumper Selectable)
E5
100mV
E6
R59
90.9
E7
10mV
E8
R60
10
5-12 Circuit Analysis
Figure 5-8. Simplified Analog Recorder Output Circuit
Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R
Instruction Manual
748213-S
April 2002
SECTION 6
MAINTENANCE AND SERVICE
The information provided in this section will aid in isolation of a malfunction to a particular assembly or circuit board. A few detailed checks are included, to aide in locating the defective assembly.
It is recommended that those familiar with circuit analysis, refer to Section 6 Circuit Analysis of this manual.
WARNING
ELECTRICAL SHOCK HAZARD
Do not operate without doors and covers secure. Servicing requires access to live parts which can cause death or serious injury. Refer servicing to qualified personnel.
For safety and proper performance this instrument must be connected to a properly grounded three-wire source of power.
Optional alarm switching relay contacts wired to separate power sources must be disconnected before servicing.
WARNING
PARTS INTEGRITY
Tampering or unauthorized substitution of components may adversely affect safety of this product. Use only factory documented components for repair
6-1 INITIAL CHECKOUT WITH STANDARD
GASES
If instrument readings do not meet specifications, the first step in troubleshooting is to isolate the analyzer from the sample stream and the sample handling system.
Admit zero and span standard gases to the analyzer. Observe readout on digital display, and on recorder, if used.
Digital display gives correct reading with standard gases, but not with sample gas
The sample and the sample handling system are suspect. Check these areas.
Digital display gives correct readings with standard gases, but the alarm or output devices do not
Check these devices individually.
Digital display gives overrange readings with standard gases, as well as sample gas
The problem is likely with detector or the electronic circuitry. Turn power OFF. Tap detector compartment with fingers, wait 30 seconds, reapply power. If the suspension within the detector assembly is hung up, this may correct the problem. If not, proceed with checks of the detector and electronic circuitry.
Digital display gives erratic readings with standard gases, as well as sample gas
If zero and span standard gases give noisy or drifting readings, the problem is probably in the detector or the temperature control circuits. Proceed with checks of the detector and electronics. In general, before concluding that the detector is defective and must be replaced, verify correct operation of all circuits that could cause erratic readings.
a. Control Board Checkout
The Detector Isolation Plug located on the
Control Board (Figure 3-1, page 3-3), removes the detector signal, allowing the input voltage to go to zero. The display should register near zero or on scale, and
TP20 should read zero voltage. To test the remainder of the measuring circuit, do the following:
Voltage
1. Set RANGE Select to lowest range.
Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 6-1
Instruction Manual
748213-S
April 2002
Model 755R
2. Adjust R29 clockwise and counterclockwise. The display should follow accordingly and remain steady within the adjustment limits of R29. If this condition is met, refer to Section 6-6a
(page 6-7) for Control Board setup.
Before replacing the Control Board, test for -15V at the junction of C1/J4-
7. Use the junction of CR1/R2 for
+15V, or any source of ±15V on the board for the respective voltages.
3. If adjustment of R29 is not possible, replace the Control Board.
Alarms
Set RANGE Select to lowest range or use zero and span gases.
Current Output
Set RANGE Select to lowest range or use zero and span gases.
When checkout complete, re-install Detector Isolation Plug. Configure Control
Board to original setup.
If the Control Board functions correctly, the problem is either located in the Detector/Magnet Assembly or related to temperature control.
6-2 HEATING CIRCUITS
To ensure against damage from overheating in the event of malfunction, the heating circuits receive power via thermal fuses F2 and
F3. If temperature of a heated area exceeds the permissible maximum, the associated fuse melts, opening the circuit.
NOTE:
The thermal fuses should be plugged in,
NOT SOLDERED, as the fuse element might melt and open the circuit.
a. Case Heater Control Circuit
The case heater control circuit receives power via thermal fuse F2 (setpoint
6-2 Maintenance and Service
75 °C). This fuse, accessible on the
Power Supply Board, may be checked for continuity.
Detector compartment heater element
HR3, mounted on the heater/fan assembly, has a normal resistance of 20 ohms.
To verify heater operation, carefully place a hand on top of detector compartment.
Heat should be felt. If not, check the case heating circuit.
Temperature sensor RT1 has a cold resistance of 22.7K ohms and a normal operating resistance of 20.2K ohms, indicating normal operating temperature.
As a further check, disconnect plug P6 on the Control Board, thus disconnecting temperature sensor RT1. Substitute a decade resistor box to simulate the resistance of RT1. Also, connect an AC voltmeter from the hot side of the line to the neutral side of F2, located inside the detector compartment.
Set the decade box for 20.2K ohms to simulate RT1 at controlling temperature.
The voltmeter should show pulses of 1
VAC.
CAUTION
OVERHEATING
Avoid prolonged operation with the decade box set at 22.2K ohms, overheating may result.
Set the decade box for 22.2K ohms to simulate RT1 resistance at ambient temperature. The voltmeter should show pulses of 120 VAC.
6-3 DETECTOR/MAGNET HEATING CIRCUIT
Heater HR1 is attached to the magnet.
Heater HR2 is attached to the rear of the detector. Combined resistance of these two parallel-connected heaters, as measured at
Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R pins 15 and 16 of the detector connector J12, should be approximately 89 ohms.
If resistance is correct, and the combined resistance is incorrect, heater HR1 may be open.
To reach the leads of HR1, remove the circuit board on the heater assembly. Resistance of
HR1 should be approximately 21 ohms.
To check operation of the heater circuit, connect a voltmeter across R61 on the Power
Supply Board. Normally, the voltage will be 4
Instruction Manual
748213-S
April 2002
VDC when cold and will drop to approximately
0.4 VDC at control temperature. Temperature sensor RT1 is mounted in the detector, with leads accessible at pins 10 and 11 of detector connector J12. The sensor resistance should be 1M ohms at 25 °C and approximately 149K ohms at operating temperature of 65 °C.
A. Detector/Magnet Assembly - Exploded View B. Optical Bench - Exploded View
Sample Pre-Heating
Coil
Connector
J12
Photocell
Lock Screws (2)
Connector
Board
Sample Inlet
Tube
Sample Outlet Tube
Dual
Photocell
Lamp Retaining
Set Screw
Detector Assembly
Magnet
Assembly
Optical Bench Assembly
Mounting Screws (2)
Lamp Viewing
Hole
Source Lamp
Assembly
Figure 6-1. Detector/Magnet Assembly
Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 6-3
Instruction Manual
748213-S
April 2002
6-4 DETECTOR CHECK
To isolate the detector as the problem, it is necessary to check the source lamp, photocells, and suspension (see Figure Figure
6-1B, page 6-3). These components are connected via J12 on the optical bench assembly.
Pin/leads may be removed from connector
J12 by use of an improvised pin removal tool, such as a paper clip (see Figure 6-2 below).
Model 755R
Connect J12 has slots at top and bottom. To remove a connector pin/lead, insert the tool into the upper or lower slot and push down on the end to release the keeper on the pin.
When inserting a pin/lead, its keeper must face toward the slot opening in the connector in order to lock in. If inserted otherwise, the pin/lead will be forced out when the two connectors are joined.
Upper Slot
Keeper
Side View of Connector
Lower Slot
Connector Pin/
Leads in Place
Connector Pin Removed
Improvised Pin Removal Tool, Such as a Paper Clip
Figure 6-2. Pin/Lead Removal
WHT
WHT
BLK
BLK
PUR
GRN
RT1
HR2
Suspension
Heater
Suspension
Terminals
10
J12
18
1
When dual photocell is installed, the gap between the two photocells should be in position indicated by this line.
Optical Bench
Figure 6-3. Detector Optical Bench
Hole for Source Lamp
6-4 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R a. Source Lamp
The simplest check of the source lamp is to verify that it is lit. Another check is done by removing the housing cover and viewing the lamp through the photocell alignment hole (see Figure 6-3, page 6-4).
If the photocell is not illuminated, test the voltage across TP2 to TP5 (ground). This voltage should be 2.2 V ±0.2 VDC. If reading is correct, the lamp may be burned out; also inspect the cable for continuity. If voltage reading is not 2.2 V
±0.2 VDC, the Power Supply Board must be replaced.
b. Photocell
If the photocells are on, observe through the photocell alignment hole. The image should be steady. Disconnect the line power and observe the image when you reconnect the power. It should come up from the side and seek a position that equally illuminates the photocells.
c. Suspension
Turn electrical power to instrument OFF.
Remove optical bench assembly (see
Figure 6-1A, page 6-3). With 100% nitrogen flowing through the analyzer, note position of the suspension. Then admit air and note response of the suspension.
It should rotate clockwise as viewed from the top, and to the right as viewed though the window. Failure to rotate indicates that the suspension has been damaged and detector assembly must be replaced.
See Section 6-5c, page 6-7.
If the suspension has been changed, the cause may be improper operating conditions.
Instruction Manual
748213-S
April 2002
6-5 REPLACEMENT OF DETECTOR/MAGNET
COMPONENTS a. Source Lamp
Removal/Installation
The source lamp is held in the optical bench assembly by a set screw (see Figure 6-1B, page 6-3). The two lamp leads are connected to J12.
The red line on the lamp base must align with the set screw (see Figure 6-4A). The base of the lamp should extend from the hole approximately 1/4 inch. Tighten set screw when lamp is aligned.
Realign the photocell per Section 6-5b, page 6-5.
b. Photocell
Removal/Installation
Refer to Figure 6-1B, page 6-3. Note location of photocell leads in connector J12.
Remove leads. Remove photocell lock screws (2), slide photocell out.
Reverse the removal procedure for installation. Align photocell (see below).
Alignment
The adjustments in this procedure are made on the Control Board. With zero gas flowing:
1. Place a digital voltmeter between the wiper of zero potentiometer (R13) and
TP7 (ground). Adjust for 0 VDC.
2. Remove the voltmeter from R13 and place on R10 (see Figure 6-4B, page
6-6). Adjust R9 for 0 VDC.
3. Remove the voltmeter from R10 and place on TP8. Move the photocell to obtain a DC voltage as close to 0 mV as possible, but no more than ±750 mV.
Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 6-5
Instruction Manual
748213-S
April 2002
4. Apply power to instrument and allow to warm-up approximately one hour.
5. Set front panel ZERO at mid-range
(i.e., five turns from either end).
6. Connect digital voltmeter from slider of R9 to chassis ground. With a steady flow of 50 to 500 cc/min of nitrogen zero gas going through instrument, adjust R9 for 0 V.
Model 755R
7. Connect the voltmeter between TP10 and circuit ground (TP7). Adjust front panel ZERO for reading of exactly zero on voltmeter.
All internal adjustments are now properly set. The instrument may be calibrated per Section 3-4, page 3-1.
DETAIL A
Set Screw
1/4"
DETAIL B
Voltmeter
Lead
R10
C8 C1 R8 R7
Figure 6-4. Lamp Replacement
Red Mark for
Alignment
6-6 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R c. Detector
Removal
Prior to removal of the detector, remove power from instrument and stop flow of sample gas.
1. Remove the four screws securing the detector cover plate.
2. Disconnect cable from J12 on the detector assembly.
NOTE:
Note how the rubber sample lines are looped into a "long coil". When reinstalling the sample lines they must be configured in the same way. This precaution isolates the detector from the effects of mechanical vibration. Otherwise vibration waves could travel upward along the tubing walls, resulting in noisy readout.
3. Refer to Figure 6-1, page 6-3. Using needle-nose pliers, squeeze the hose clamps to disconnect the rubber sample lines from the metal inlet and outlet tubes of the detector assembly.
4. Remove the two screws at the bottom of the detector assembly, slide detector out.
Installation
1. Install replacement detector assembly and connect cable to J12.
2. Seat the detector assembly firmly against the magnet pole pieces and tighten attaching screws.
3. Reconnect rubber sample lines to metal inlet and outlet tubes on detector assembly.
4. Apply power to instrument and allow to warm up approximately one hour.
Instruction Manual
748213-S
April 2002
Calibration
1. On the Control Board, set the front panel ZERO control (R13) at midrange (i.e., five turns from either end).
2. Connect a digital voltmeter from the slider of R9 to chassis ground. With a steady flow of 50 to 500 cc/min. of nitrogen gas passing through the instrument, adjust R9 for zero volts.
3. Connect the voltmeter between TP10 and circuit ground (TP7). Adjust front panel ZERO control (R13) for reading of exactly zero on voltmeter.
4. With all internal adjustments now properly set, the instrument may be calibrated per Section 3-4, page 3-1.
6-6 CONTROL BOARD SETUP a. Power Supply Test
1. TP7 (circuit ground) is ground point for all voltage tests.
2. Counterclockwise end of front panel
ZERO potentiometer (R13 on Control
Board): -15 VDC ±0.5 VDC.
3. Clockwise end of ZERO potentiometer: +15 VDC ±5 VDC.
4. Set ZERO potentiometer to obtain a reading of .0 VDC ±10 mV at slider.
5. Measure TP19: +5 VDC ±0.25 VDC.
b. Detector zero
1. Flow 250 cc/min nitrogen.
2. Monitor TP8, adjust R9 for 0 VDC
±2mV.
Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 6-7
Instruction Manual
748213-S
April 2002 c. U4 Zero
1. Monitor TP5, adjust R100 for 1 VDC
±2mV.
2. Monitor TP10, adjust R13 (ZERO) for
0.0 VDC ±5mV.
d. U8 Zero
Monitor TP11, adjust R29 for 0.0 VDC
±5mV.
e. U10 Zero
1. Monitor TP16, adjust R29 for 0.0 VDC
±5mV.
NOTE:
This adjustment requires a "long time" constant. Allow adequate time.
Model 755R
2. Adjust R29 to obtain a reading of
00.00 ±5 counts on the display.
f. Fullscale
1. Monitor TP11. Flow 100% oxygen, adjust SPAN potentiometer R20 for
10.00 V ±5mV.
2. Monitor TP16, adjust R20 for 10.00 V
±5mV. The display must read 100.00
±5 counts.
g. Recorder Fullscale
1. Flow nitrogen at 250 cc/min, monitor
TP16, and adjust front panel ZERO potentiometer for .000 VDC.
2. Flow 100% oxygen for span gas. Recorder output for 1 V, 100 mV, or 10 mV should read 100% of span gas.
Adjust R88 if necessary.
6-8 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R
Instruction Manual
748213-S
April 2002
SECTION 7
REPLACEMENT PARTS
The following parts are recommended for routine maintenance and troubleshooting of the
Model 755R Oxygen Analyzer. If the troubleshooting procedures do not resolve the problem, contact Rosemount Analytical Customer
Service Center.
WARNING
PARTS INTEGRITY
Tampering or unauthorized substitution of components may adversely affect safety of this product. Use only factory-documented components for repair.
7-1 CIRCUIT BOARD REPLACEMENT POLICY
In most situations involving a malfunction of a circuit board, it is more practical to replace the board than to attempt isolation and replacement of the individual component. The cost of test and replacement will exceed the cost of a rebuilt assembly. As standard policy, rebuilt boards are available on an exchange basis.
Because of the exchange policy covering circuit boards the following list does not include individual electronic components. If circumstances necessitate replacement of an individual component, which can be identified by inspection or from the schematic diagrams, obtain the replacement component from a local source of supply.
Rosemount Analytical Inc. A Division of Emerson Process Management Replacement Parts 7-1
Instruction Manual
748213-S
April 2002
7-2 MATRIX – MODEL 755R OXYGEN ANALYZER
755R Model 755R Oxygen Analyzer
755R
Code Ranges
01 0-5, 10, 25, 50, and 100% O
2
(Standard)
02 0-1, 2.5, 5, 10, 25, 50, and 100% O
2
(Extended)
99 Special
Code Output
01 0-10 mV, 0-100 mV, 0-1 V or 0-5 VDC (Standard)
02 0, 4-20 mA (Current)
99 Special
Code Alarm Relays
00 No Alarm
01 Dual Alarm
99 Special
Code Case
01 Standard
02 Standard with Tropicalization
03 EMC Kit
04 EMC Kit with Tropicalization
99 Special
Code Operation
01 115 VAC, 50/60 Hz (Standard)
02 230 VAC, 50/60 Hz
99 Special
01 01 01 01 01
Code Remote Range
00 None
01 Standard ( 0-5, 10, 25, 50, 100%)
02 Extended ( 0-1, 2.5, 5, 10, 25%) (1)
03 0-1, 2.5, 5, 10, 50% (1)
04 0-1, 2.5, 5, 25, 50% (1)
05 0-1, 2.5, 5, 25, 100% (1)
06 0-1, 2.5, 10, 25, 100% (1)
99 Special
00
Code Feature
00 Features as selected
99 Special
00 Example
Model 755R
7-2 Replacement Parts Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R
7-3 SELECTED REPLACEMENT PARTS
092114 Fuse, 1/2A (240VAC) (Package of 5)
777362 Fuse, Heater 3A (120VAC) (Package of 15)
777361 Fuse, Heater 1.5A (240VAC) (Package of 15)
861649 Thermal Fuse (F2,F3)
656189 Detector/Optical Bench Assembly (0 to 1%)
616418 Source Lamp Kit
622356 Photocell
621023 Current Output Board (0 to 20mA, 4 to 20mA)
622351 Connector Board
631773 Power Supply Board
652830 Control Board Kit
654004 Thermistor - Case Heater
654022 Display Assembly
654078 Viton Tubing (Sample In)
654079 Viton Tubing (Sample Out)
654080 Fan Assembly
654081 Case Heater
809374 Fuse, 3/4A (Power Transformer, 115VAC)
860371 Alarm Relay
645407 Shock Mount (Package of 4)
Instruction Manual
748213-S
April 2002
Rosemount Analytical Inc. A Division of Emerson Process Management Replacement Parts 7-3
Instruction Manual
748213-S
April 2002
Model 755R
7-4 Replacement Parts Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R
Instruction Manual
748213-S
April 2002
SECTION 8
RETURN OF MATERIAL
8-1 RETURN OF MATERIAL
If factory repair of defective equipment is required, proceed as follows:
1. Secure a return authorization from a
Rosemount Analytical Inc. Sales Office or
Representative before returning the equipment. Equipment must be returned with complete identification in accordance with Rosemount instructions or it will not be accepted.
Rosemount CSC will provide the shipping address for your instrument.
In no event will Rosemount be responsible for equipment returned without proper authorization and identification.
2. Carefully pack the defective unit in a sturdy box with sufficient shock absorbing material to ensure no additional damage occurs during shipping.
3. In a cover letter, describe completely:
• The symptoms that determined the equipment is faulty.
• The environment in which the equipment was operating (housing, weather, vibration, dust, etc.).
• Site from where the equipment was removed.
• Whether warranty or non-warranty service is expected.
• Complete shipping instructions for the return of the equipment.
4. Enclose a cover letter and purchase order and ship the defective equipment according to instructions provided in the
Rosemount Return Authorization, prepaid, to the address provided by Rosemount
CSC.
Rosemount Analytical Inc.
Process Analytical Division
Customer Service Center
1-800-433-6076
If warranty service is expected, the defective unit will be carefully inspected and tested at the factory. If the failure was due to the conditions listed in the standard Rosemount warranty, the defective unit will be repaired or replaced at Rosemount’s option, and an operating unit will be returned to the customer in accordance with the shipping instructions furnished in the cover letter.
For equipment no longer under warranty, the equipment will be repaired at the factory and returned as directed by the purchase order and shipping instructions.
8-2 CUSTOMER SERVICE
For order administration, replacement Parts, application assistance, on-site or factory repair, service or maintenance contract information, contact:
Rosemount Analytical Inc.
Process Analytical Division
Customer Service Center
1-800-433-6076
8-3 TRAINING
A comprehensive Factory Training Program of operator and service classes is available. For a copy of the Current Operator and Service
Training Schedule contact the Technical
Services Department at:
Rosemount Analytical Inc.
Customer Service Center
1-800-433-6076
Rosemount Analytical Inc. A Division of Emerson Process Management Return of Material 8-1
Instruction Manual
748213-S
April 2002
Model 755R
8-2 Return of Material Rosemount Analytical Inc. A Division of Emerson Process Management
WARRANTY
Goods and part(s) (excluding consumables) manufactured by Seller are warranted to be free from defects in workmanship and material under normal use and service for a period of twelve (12) months from the date of shipment by Seller. Consumables, glass electrodes, membranes, liquid junctions, electrolyte, o-rings, etc., are warranted to be free from defects in workmanship and material under normal use and service for a period of ninety (90) days from date of shipment by
Seller. Goods, part(s) and consumables proven by Seller to be defective in workmanship and/or material shall be replaced or repaired, free of charge, F.O.B. Seller's factory provided that the goods, part(s) or consumables are returned to Seller's designated factory, transportation charges prepaid, within the twelve (12) month period of warranty in the case of goods and part(s), and in the case of consumables, within the ninety (90) day period of warranty. This warranty shall be in effect for replacement or repaired goods, part(s) and the remaining portion of the ninety (90) day warranty in the case of consumables. A defect in goods, part(s) and consumables of the commercial unit shall not operate to condemn such commercial unit when such goods, part(s) and consumables are capable of being renewed, repaired or replaced.
The Seller shall not be liable to the Buyer, or to any other person, for the loss or damage directly or indirectly, arising from the use of the equipment or goods, from breach of any warranty, or from any other cause. All other warranties, expressed or implied are hereby excluded.
IN CONSIDERATION OF THE HEREIN STATED PURCHASE PRICE OF THE GOODS,
SELLER GRANTS ONLY THE ABOVE STATED EXPRESS WARRANTY. NO OTHER
WARRANTIES ARE GRANTED INCLUDING, BUT NOT LIMITED TO, EXPRESS AND IMPLIED
WARRANTIES OR MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
Limitations of Remedy. SELLER SHALL NOT BE LIABLE FOR DAMAGES CAUSED BY
DELAY IN PERFORMANCE. THE SOLE AND EXCLUSIVE REMEDY FOR BREACH OF
WARRANTY SHALL BE LIMITED TO REPAIR OR REPLACEMENT UNDER THE STANDARD
WARRANTY CLAUSE. IN NO CASE, REGARDLESS OF THE FORM OF THE CAUSE OF
ACTION, SHALL SELLER'S LIABILITY EXCEED THE PRICE TO BUYER OF THE SPECIFIC
GOODS MANUFACTURED BY SELLER GIVING RISE TO THE CAUSE OF ACTION. BUYER
AGREES THAT IN NO EVENT SHALL SELLER'S LIABILITY EXTEND TO INCLUDE
INCIDENTAL OR CONSEQUENTIAL DAMAGES. CONSEQUENTIAL DAMAGES SHALL
INCLUDE, BUT ARE NOT LIMITED TO, LOSS OF ANTICIPATED PROFITS, LOSS OF USE,
LOSS OF REVENUE, COST OF CAPITAL AND DAMAGE OR LOSS OF OTHER PROPERTY
OR EQUIPMENT. IN NO EVENT SHALL SELLER BE OBLIGATED TO INDEMNIFY BUYER IN
ANY MANNER NOR SHALL SELLER BE LIABLE FOR PROPERTY DAMAGE AND/OR THIRD
PARTY CLAIMS COVERED BY UMBRELLA INSURANCE AND/OR INDEMNITY COVERAGE
PROVIDED TO BUYER, ITS ASSIGNS, AND EACH SUCCESSOR INTEREST TO THE GOODS
PROVIDED HEREUNDER.
Force Majeure. Seller shall not be liable for failure to perform due to labor strikes or acts beyond
Seller's direct control.
Instruction Manual
748213-S
April 2002
Model 755R
Emerson Process Management
Rosemount Analytical Inc.
Process Analytic Division
1201 N. Main St.
Orrville, OH 44667-0901
T (330) 682-9010
F (330) 684-4434
ASIA - PACIFIC
Fisher-Rosemount
Singapore Private Ltd.
1 Pandan Crescent
Singapore 128461
Republic of Singapore
T 65-777-8211
F 65-777-0947 http://www.processanalytic.com
Fisher-Rosemount GmbH & Co.
Industriestrasse 1
63594 Hasselroth
Germany
T 49-6055-884 0
F 49-6055-884209
EUROPE, MIDDLE EAST, AFRICA
Fisher-Rosemount Ltd.
Heath Place
Bognor Regis
West Sussex PO22 9SH
England
T 44-1243-863121
F 44-1243-845354
LATIN AMERICA
Fisher - Rosemount
Av. das Americas
3333 sala 1004
Rio de Janeiro, RJ
Brazil 22631-003
T 55-21-2431-1882
© Rosemount Analytical Inc. 2001
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Related manuals
advertisement
Table of contents
- 29 Set Zero Calibration Point
- 30 Set Span Calibration Point
- 41 Input Amplifier U1
- 41 Amplifier U2 and Associated Zero Adjustment
- 41 Amplifier U4 and Associated Span Adjustment
- 41 Amplfier U8
- 41 Amplifier U10
- 41 Source Lamp Power Supply Section
- 42 +/-15V Power Supply Section
- 42 Detector Temperature Control Section
- 42 Detector Compartment Temperature Control Section
- 57 Voltage
- 58 Alarms
- 58 Current Output