Firmware Update Description inEos X5
Firmware Update Description inEos X5
The following is a detailed description of the new firmware update procedure for the
inEos X5.
For all inEos X5 scanners with a serial number lower than 1125 it is recommended to first
test if the scanner starts the firmware update mode before installation of the inLab SW 4.2.5.
Please therefore carefully read points 5 and 6 of this description.
1. If the firmware update is necessary the following message will be automatically
shown when scanning the first case after installation of inLab SW 4.2.5.
Click on the >>OK<< button.
2. Proceed to the system menu and click on >>Configuration<<.
3. Please select >>Devices<< and >>inEos X5<<.
SIRONA Dental Systems
Page 1 of 6
CAD/CAM inLab V4.2.5
Software-Beschreibung
4. In the next window select >>Update Firmware<< from the step menu.
5. Switch off the inEos X5 with the main switch now. Confirm with >>OK<<.
SIRONA Dental Systems
Seite 2 von 6
CAD/CAM inLab V4.2.5
Software-Beschreibung
6. The inEos X5 must now be set in the firmware update mode. To activate this mode,
please follow the instructions below:
a. Press and hold the blue button on the front side of the inEos X5
b. Switch on the device with main switch and keep the blue button pressed
c. Press the blue button until the status light turns yellow
d. Release the blue button as soon as the light turns yellow
e. Confirm with >>OK<<
The update process will start automatically.
Note:
For inEos X5 scanners with a serial number lower than 1100 the status light at the
front will not turn yellow, rather it will turn off. As soon as the light turns off, release
the button. The firmware update can then be performed as well.
SIRONA Dental Systems
Seite 3 von 6
CAD/CAM inLab V4.2.5
Software-Beschreibung
7. Do not disconnect the scanner from the power supply during the update process.
Confirm with >>OK<<.
8. The update process will now start. The update status can be monitored in the step
menu.
SIRONA Dental Systems
Seite 4 von 6
CAD/CAM inLab V4.2.5
Software-Beschreibung
9. As soon as the update process is finished a message will be shown in the step menu.
Switch the scanner off and on.
Confirm with >>OK<<.
10. A message will be shown in the step menu that the firmware update was successful.
Confirm with >>Continue<<.
SIRONA Dental Systems
Seite 5 von 6
CAD/CAM inLab V4.2.5
Software-Beschreibung
11. Return back to scanning with >>End configuration<<.
SIRONA Dental Systems
Seite 6 von 6