Valor™ 2000W Series Instruction Manual

99 Washington Street
Melrose, MA 02176
Phone 781-665-1400
Toll Free 1-800-517-8431
Visit us at www.TestEquipmentDepot.com
Valor™ 2000W Series
Instruction Manual
Valor™ 2000W Series
EN-1
1. INTRODUCTION
This manual contains installation, operation and maintenance instructions for the
Valor™ 2000W Series. Please read the manual completely before using the scale.
1.1 Safety Precautions
Please follow these safety precautions:
 Verify that the AC Adapter input voltage matches the local AC power supply.
 Do not drop loads on the platform.
 Do not place the scale upside down on the platform.
 Disconnect the scale from power supply when cleaning.
 Operate the scale only under ambient conditions specified in these
instructions.
 Service should be performed only by authorized personnel.
 Only use weights within the scale’s capacity as specified in these instructions.
 Do not operate the scale in hostile environments.
 Do not carry the scale by the pan or sub-platform. Use the handholds on the
side of the scale housing.
2. INSTALLATION
2.1 Package Contents
 Scale
 Stainless Steel Pan
 Power Adapter & Plug
 Instruction Manual
 Warranty Card
2.2 Installing Components
Place the stainless steel pan onto the weighing platform before turning the scale on.
Figure 2-1. Install the stainless steel pan onto the weighing platform
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
EN-2
Valor™ 2000W Series
2.3 Selecting the Location
Use the scale on a clean, firm and flat surface. Avoid locations with excessive air
current, vibrations, heat sources, or rapid temperature changes.
2.4 Leveling the Equipment
Adjust the leveling feet so the bubble is centered in the circle. Be sure the
equipment is level each time its location is changed.
Level Bubble
Adjustable Feet
Figure 2-2. Level Indicator
2.5 Power
The AC Adapter is used to power the scale when battery power is not needed.
First connect the AC Adapter plug to the scale input jack located at the bottom of
the scale according to the description below.
Figure 2-3. Slide to plug in
Then connect the AC adapter to the proper AC supply.
Figure 2-4. Connecting the AC adapter to AC supply
Caution: The scale can only be used in a dry environment when
powered by the AC adapter.
Valor™ 2000W Series
EN-3
2.5.1 Battery Power:
The scale can be used on AC power immediately. Allow the battery to charge
for 12 hours before using the scale on battery power. The scale will
automatically switch to battery operation if there is a power failure or the power
cord is removed. With AC power, the scale is constantly charging, so the
battery charge indicator (see item 8 in table 3-2) will remain lit. The scale can
be operated during charging, and the battery is protected against overcharging.
For maximum operating time, the battery should be charged at room
temperature.
During battery operation, the battery symbol indicates the battery charging
status. When charging, the symbol will blink slowly and when fully charged the
symbol will stop blinking.
Symbol
TABLE 2-1
Charge Level
Battery in use:
Symbol displayed
Notes:
When battery symbol blinks fast, approximately 30 minutes working time is left.
When [Lo.bAt] is displayed, the scale will shut off.
Charging the scale must be performed in a dry environment.
CAUTION: Battery is to be replaced only by an authorized
OHAUS service dealer. Risk of explosion can occur if the
rechargeable battery is replaced with the wrong type or if it is
not properly connected. Dispose of the lead acid battery
according to local laws and regulations.
EN-4
Valor™ 2000W Series
3. OPERATION – V22PW
3.1 Controls
Figure 3-1. Front Control Panel Valor 2000W – V22PW
TABLE 3-1
Button
Functions
1
On/Zero
Off
Tare
Cal
Short Press (when on): Sets display to zero
Short Press (when off): Turns scale on
2
Long Press (when on): Turns the scale off
Short Press (in Menu): Selects / accepts displayed setting
Short Press: Enter / clear a Tare value
3
Extended Press : Initiates Calibration procedure
Short Press (in Menu): Toggles through available settings
Notes:
1
Short Press: Press less than 2.5 seconds.
2
Long Press: Press and hold for more than 2.5 seconds.
3
Extended Press: Press and hold for more than 5 seconds.
Display
1
1
2
3
4
5
6
1
1
1
1
1
1
1
12
11
10
9
1
1
1
Figure 3-2. Valor 2000W Display
7
1
8
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
Valor™ 2000W Series
EN-5
TABLE 3-2 Display Symbols
Item Description
Item Description
1 NET symbol
7
Dynamic symbol*
2 Center of Zero symbol
8
Battery charge symbol
3 Negative symbol
9
Pound, Ounce, Pound:Ounce symbol
4 Stable weight symbol
10
Gram, kilogram symbol
5 Preset Tare symbol*
11
Pieces symbol*
6 Pointer symbols*
12
Percent symbol*
Note: * Not Used
Figure 3-3. Valor 2000W – V22PW Rear display.
3.2 Turning Scale On/Off
To turn the scale on, press the On/Zero Off button. The scale performs a display
test, momentarily displays the software version, and then enters the active weighing
mode.
To turn the scale off, press and hold the On/Zero Off button until OFF is displayed.
3.3 Calibration Menu
To enter the calibration menu, extend press Cal button for 5 seconds, [SpaN] will
then be displayed. Press Zero key to perform span calibration or press Cal key to
proceed to GEO settings. Press Zero key to enter GEO settings and then press Cal
key to increase the value. Press Zero key to confirm, [end] will be displayed. Press
Zero key to exit the menu and return to weighing mode.
Cal
Span
GEO
End
Figure 3-4. Menu structure
Span [SpaN] (yes, no)
Initiates a span calibration procedure (zero and span).
GEO [GEO]
Geographical Adjustment Factor (GEO) is used to adjust the calibration based on
the current location. Settings from 0 to 31 are available with 12 being the default.
Refer to table 4-2 to determine the GEO factor that corresponds to your location.
EN-6
Valor™ 2000W Series
3.4 Initial Calibration
When the scale is first installed it should be calibrated to ensure accurate results.
Before performing the calibration, be sure to have the appropriate calibration
weights as listed in table 3-3.
Press and hold Cal for 5 seconds, the display will show [Cal]. When the button is
released, the display will show [SpaN]. Press On/Zero to begin the span calibration.
[--C--] blinks while zero reading is stored. Next, the display shows the calibration
weight value. Place the specified calibration mass on the pan and press On/Zero.
[--C--] blinks while the reading is stored. The scale returns to the previous
application mode and is ready for use.
The message [Err 3.0 Cal] will be displayed if these calibration steps are not
followed or if the wrong weight was used.
The calibration process can be aborted by turning the scale off.
TABLE 3-3
Required Span Calibration Mass (sold separately)
Max
Mass
Max
Mass
1500 g
3000 g
1 kg / 3 lb
2 kg / 6 lb
6000 g
15000 g
5 kg / 15 lb
10 kg / 30 lb
Note: If current unit is lb or oz, the scale will be calibrated by pound weight.
3.5 Unit Menu
Follow below instructions to enter the service menu to set the displayed unit.
With the scale off and LFT off, extend press both the On/Zero and Tare keys at the
same time for more than 8 seconds until [rAmMP] is displayed. Press the Tare key
through the succeeding displays: [rAmMP]*, [LIN]*, [SPAN]*, [CAP]*, [Dius]* until [s.UNIt]
is displayed.

When [s.UNIt] is displayed, press the On/Zero key.

[UNIt kg] will be displayed. Press the Tare key until the active (power on)
unit is displayed.

When the active unit is displayed, press the On/Zero key.

[ON XX] is displayed. Press Tare to switch to [Off XX], then press the
On/Zero key**.

The next unit [UNIt xx] is displayed. Press the Tare key until the desired
unit is displayed.

When the desired unit is displayed, press the On/Zero key.

[Off XX] is displayed. Press the Tare key to switch to [ON XX], then press the
On/Zero key.

Press the Tare key until [End] is displayed, then press the On/Zero key.

Press the Tare key through the succeeding displays: [e.paND]*, [reset]*
until [End] is displayed.
When [End] is displayed, press the On/Zero key to exit to weighing mode.
Notes:
*Do not change any of these settings (by pressing the On/Zero key). This may affect
the correct set up of the scale.
**The active unit must be turned off first before another unit can be turned on.
Valor™ 2000W Series
EN-7
4. OPERATION – V22XW
4.1 Controls
Figure 4-1. Front Control Panel Valor 2000W – V22XW
TABLE 4-1
Button
Functions
1
Short Press (when on): Sets display to zero
Short Press (when off): Turns scale on
2
Long Press (when on): Turns the scale off
Short Press (in Menu): Selects / accepts displayed setting
Long Press: Toggles through active Units
Short Press (in Menu): Toggles through available settings
Short Press: Initiates an application mode specific response
Long Press: Selects active Mode
Short Press (in Menu): returns to previous settings
Short Press: Enter / clear a Tare value
Long Press: Enters User Menu
Short Press (in Menu): Quickly exit User Menu
Notes:
1
Short Press: Press less than 2.5 seconds.
2
Long Press: Press and hold for more than 2.5 seconds.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
EN-8
Valor™ 2000W Series
LED Indicators
Figure 4-1. Valor 2000W – V22XW Rear display.
The colored LED indicators on the right side of the control panels are used in
Checkweigh mode (section 4.5) and will light up according to the following rules:
(Red) Loads > Upper limit
(Green) Loads ≥ Lower limit and ≤ Upper limit
(Yellow) Loads < Lower limit
Display
See section 3.1 for information about the display and it’s symbols.
4.2 Turning Scale On/Off
To turn the scale on, press the On/Zero Off button. The scale performs a display
test, momentarily displays the software version, and then enters the active weighing
mode.
To turn the scale off, press and hold the On/Zero Off button until OFF is displayed.
4.3 Initial Calibration
When the scale is first installed it should be calibrated to ensure accurate results.
Before performing the calibration, be sure to have the appropriate calibration
weights as listed in table 3-3.
Press and hold Menu until [mMeNU] (Menu) is displayed. When the button is released,
the display will display [C.A.L]. Press Yes to accept, [SpaN] will then be shown. Press
Yes again to begin the span calibration. [--C--] blinks while zero reading is stored.
Next, the display shows the calibration weight value. Place the specified calibration
mass on the pan and press Yes. [--C--] blinks while the reading is stored. The scale
returns to the previous application mode and is ready for use.
The message [Err 3.0 CAl] will be displayed if these calibration steps were not
followed or if the wrong weight was used.
The calibration process can be aborted by turning the scale off.
Valor™ 2000W Series
EN-9
4.4 Weigh Mode
1.
2.
3.
Press and hold Mode until [wJeIGH] (Weigh) is displayed.
If required, place an empty container on the pan and press Tare.
Add sample to the pan or container. The display shows the weight of the
sample.
4.5 Checkweigh Mode
This mode sets low and high weight limits for portion control processes.
1. Press and hold Mode until [CHeCk] (Check) is displayed. [CLr.ref] (clear
references) will then display.
Notes: Press Function to view the low and high reference weight limits.
If there is no reference data stored, [Set.ref] is displayed. Press Yes to
set reference data.
2. Press No to use the stored reference weight limits and proceed to step 5.
3. Press Yes to establish new reference values. The scale will then display [Set.
LO]. Press Yes to view the “Low” limit value. Press Yes to accept or No to edit
the “Low” limit value. The stored value then displays with the first digit
highlighted [000.000 kg]. Repeatedly press No until the desired number
appears. Press Yes to accept and highlight the next digit. Repeat until all the
digits are correct. Press Yes to accept the “low” limit value, [Set. HI] will be
displayed.
4. Repeat the same procedure to accept or edit the “high” value.
5. If required, place an empty container on the pan and press Tare. Place
sample material on the pan or in the container. If the sample weight is under
the target weight range, the yellow LED will light. If the sample is within the
target weight range, the green LED will light. If the sample is over the target
weight range, the red LED will light.
6. To clear the stored reference values press and hold Mode until [CHeCk] is
displayed. Press Yes when [Clr.ref] is displayed.
EN-10
Valor™ 2000W Series
4.6 Menu Settings
The User Menu allows the customizing of scale settings.
Note: Valor 2000W – V22PW does not have these Menus.
4.6.1 Menu Navigation
User Menu:
Menu:
C.A.L
Span
Lin
Menu
GEO
Items:
End
M.O.d.E
Reset
Check
End
U.n.i.t
kg
g
oz
lb
lb:oz
End
E.n.d
To Enter the Menu Mode
Press and hold Menu until [mMeNU] (Menu) is displayed. When released the first submenu [C.a.l] (Cal) will be shown.
Press Yes to enter the displayed sub-menu or press No to advance to the next.
Selecting a sub-menu will display the first menu item. Press Yes to view the menu
item setting or press No to move to the next menu item. When viewing the setting,
press Yes to accept the setting, or press No to change the setting. When [End] is
displayed, press Yes to return to the sub-menu selections or No to return to the first
item in the current menu. Bold indicates factory default setting.
Note: The Calibration / Menu Mode indicator is displayed when in the Menu Mode.
4.6.2 Cal Menu
Enter this menu to perform calibrations.
 Span [SpaN] (yes, no)
Initiates a span calibration procedure (zero and span).
 Lin [LIN] (yes, no)
Initiates a linearity calibration procedure (zero, mid-point and span).
 GEO [GEO]
Geographical Adjustment Factor (GEO) is used to adjust the calibration
based on the current location. Settings from 0 to 31 are available with 12
being the default. Refer to table 4-2 to determine the GEO factor that
corresponds to your location.

End Cal [End]
Advance to the next menu or return to the top of the current menu.
4.6.3 Mode Menu
Enter this menu to activate modes so they will be available for use with the Mode
button. Weigh mode is always active by default.
 Reset [reset] (no, yes)
Reset the Mode menu to factory defaults.
 Check [CHECk] (off, on)
Set the sub-mode.
 End Mode [End]
Advance to the next menu or return to the top of the current menu.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
Valor™ 2000W Series
EN-11
4.6.4 Unit Menu
Enter this menu to activate units so they will be accessible with the Units button.
The units in the menu must be turned “on” to be active.
Note: Available units vary by model and local regulations.
4.6.5 End Menu
Press ‘Yes’ to advance to the Calibration menu. Press ‘No’ to exit the menu and
return to the current application mode.
TABLE 4-2. GEO CODES
Latitude
0°00'
5°46'
5°46'
9°52'
9°52'
12°44'
12°44'
15°06'
15°06'
17°10'
17°10'
19°02'
19°02'
20°45'
20°45'
22°22'
22°22'
23°54'
23°54'
25°21'
25°21'
26°45'
26°45'
28°06'
28°06'
29°25'
29°25'
30°41'
30°41'
31°56'
31°56'
33°09'
33°09'
34°21'
34°21'
35°31'
35°31'
36°41'
36°41'
37°50'
37°50'
38°58'
38°58'
40°05'
40°05'
41°12'
41°12'
42°19'
42°19'
43°26'
43°26'
44°32'
44°32'
45°38'
45°38'
46°45'
46°45'
47°51'
47°51'
48°58'
48°58'
50°06'
50°06'
51°13'
51°13'
52°22'
52°22'
53°31'
53°31'
54°41'
54°41'
55°52'
55°52'
57°04'
57°04'
58°17'
58°17'
59°32'
59°32'
60°49'
60°49'
62°90'
62°90'
63°30'
63°30'
64°55'
64°55'
66°24'
66°24'
67°57'
67°57'
69°35'
69°35'
71°21'
71°21'
73°16'
73°16'
75°24'
75°24'
77°52'
77°52'
80°56'
80°56'
85°45'
85°45'
90°00'
0
325
325
650
650
975
975
1300
0
1060
1060
2130
2130
3200
3200
4260
5
5
6
6
7
7
8
8
9
9
10
10
11
11
12
12
13
13
14
14
15
15
16
16
17
17
18
18
19
19
20
20
21
21
22
22
23
23
24
24
25
25
26
26
27
27
28
28
29
29
30
30
31
4
5
5
6
6
7
7
8
8
9
9
10
10
11
11
12
12
13
13
14
14
15
15
16
16
17
17
18
18
19
19
20
20
21
21
22
22
23
23
24
24
25
25
26
26
27
27
28
28
29
29
30
30
4
4
5
5
6
6
7
7
8
8
9
9
10
10
11
11
12
12
13
13
14
14
15
15
16
16
17
17
18
18
19
19
20
20
21
21
22
22
23
23
24
24
25
25
26
26
27
27
28
28
29
29
30
3
4
4
5
5
6
6
7
7
8
8
9
9
10
10
11
11
12
12
13
13
14
14
15
15
16
16
17
17
18
18
19
19
20
20
21
21
22
22
23
23
24
24
25
25
26
26
27
27
28
28
29
29
Elevation in meters
1300
1625
1950
1625
1950
2275
Elevation in feet
4260
5330
6400
5330
6400
7460
GEO value
3
2
2
3
3
2
4
3
3
4
4
3
5
4
4
5
5
4
6
5
5
6
6
5
7
6
6
7
7
6
8
7
7
8
8
7
9
8
8
9
9
8
10
9
9
10
10
9
11
10
10
11
11
10
12
11
11
12
12
11
13
12
12
13
13
12
14
13
13
14
14
13
15
14
14
15
15
14
16
15
15
16
16
15
17
16
16
17
17
16
18
17
17
18
18
17
19
18
18
19
19
18
20
19
19
20
20
19
21
20
20
21
21
20
22
21
21
22
22
21
23
22
22
23
23
22
24
23
23
24
24
23
25
24
24
25
25
24
26
25
25
26
26
25
27
26
26
27
27
26
28
27
27
28
28
27
29
28
28
2275
2600
2600
2925
2925
3250
3250
3575
7460
8530
8530
9600
9600
10660
10660
11730
1
2
2
3
3
4
4
5
5
6
6
7
7
8
8
9
9
10
10
11
11
12
12
13
13
14
14
15
15
16
16
17
17
18
18
19
19
20
20
21
21
22
22
23
23
24
24
25
25
26
26
27
27
1
1
2
2
3
3
4
4
5
5
6
6
7
7
8
8
9
9
10
10
11
11
12
12
13
13
14
14
15
15
16
16
17
17
18
18
19
19
20
20
21
21
22
22
23
23
24
24
25
25
26
26
27
0
1
1
2
2
3
3
4
4
5
5
6
6
7
7
8
8
9
9
10
10
11
11
12
12
13
13
14
14
15
15
16
16
17
17
18
18
19
19
20
20
21
21
22
22
23
23
24
24
25
25
26
26
0
0
1
1
2
2
3
3
4
4
5
5
6
6
7
7
8
8
9
9
10
10
11
11
12
12
13
13
14
14
15
15
16
16
17
17
18
18
19
19
20
20
21
21
22
22
23
23
24
24
25
25
26
EN-12
Valor™ 2000W Series
5. MAINTENANCE
5.1 Cleaning
The housing may be cleaned with a cloth dampened with a mild detergent if
necessary. Do not use solvents, chemicals, alcohol, ammonia or abrasives to clean
the housing or control panels.
5.2 Cleaning the Plastic Pan
5.2.1 Uninstalling and cleaning the plastic pan
Please follow the steps below to clean the plastic pan:
1. Remove the stainless steel pan
2. Remove the four thumb screws holding the battery cover.
3. Unplug the two battery clips and remove the battery.
4. Using a Phillips screwdriver, remove the two screws located at the bottom of the
battery compartment.
5. Remove the plastic pan.
6. Clean the plastic pan.
3
5
2
4
1
6
Figure 5-1. Pan cleaning
5.2.2 Reinstalling the pan after cleaning
Please follow the steps below to reinstall the plastic pan (see also the steps in figure
5-1, in reversed order):
1. Using a Phillips screwdriver, attach the plastic pan to the housing with the
two screws.
2. Place the battery in the battery compartment and attach the two battery clips.
Attach the red wire to the positive (red) battery terminal and the black wire to
the negative (black) battery terminal.
3. Attach the battery cover with the four thumb screws.
4. Place the scale upright and install the stainless steel pan.
Caution: Risk of explosion can occur if the battery is not properly
connected.
Valor™ 2000W Series
EN-13
5.3 Troubleshooting
The following table lists common problems and possible causes and remedies.
If the problem persists, contact OHAUS or your authorized dealer.
TABLE 5-1
Symptom
Possible Cause
Remedy
Cannot turn on
No power to scale
Battery discharged
Verify connections and power
source
Improper calibration
Unstable environment
Unstable environment
Incorrect calibration weight
Perform calibration
Move scale to suitable location
Move the scale to suitable location
Use correct calibration weight
Err 3.0 Cal
Incorrect calibration weight
See table 3-4 for correct weights
Err 8.1
Err 8.2 _LOad_
Power on zero range
exceeded
Power on zero under range
Err 8.3
Overload
Poor accuracy
Cannot calibrate
ˉLOadˉ
ˉLOadˉ
Err 8.4 _LOad_
Under load
Err 8.5 ˉtareˉ
Err 9 Data
Tare out of range
Err13 mMEmM
Fail to write EEPROM.
Err 53 C.SUmM
Invalid checksum data
Lo.Bat
Battery is discharged.
Internal data error.
Battery fails to charge fully Battery is defective.
Clear pan
Install pan
Load exceeds scale maximum
capacity
Reading below min. range - Reinstall pan.
Tare value exceeds maximum.
Contact an authorized service agent
Contact an authorized service
agent
Contact an authorized service
agent.
Connect the power and charge the
battery
Have battery replaced by OHAUS
authorized service dealer.
5.4 Service Information
If the troubleshooting section does not resolve or describe your problem, contact
your authorized OHAUS service agent. Please visit our web site, www.ohaus.com
to locate the OHAUS office nearest you.
6. TECHNICAL DATA
The technical data is valid under the following ambient conditions:
Indoor use only
Operating temperature: -10ºC (14ºF) to 40ºC (104ºF)
Relative humidity: 10% to 90% relative humidity, non-condensing
Altitude: Up to 2000 m
Power: AC Adapter (supplied) - 12 VDC 0.84 A output, internal rechargeable sealed
lead-acid battery
Mains supply voltage fluctuations: up to ±10% of the nominal voltage
Installation Category: II
Pollution Degree: 2
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
EN-14
Valor™ 2000W Series
6.1 Specifications
TABLE 6-1
V22PWE1501T
V22XWE1501T
1.5 kg x 0.0002 kg
1500 g x 0.2 g
3 lb x 0.0005 lb
48 oz x 0.01 oz
7500
0.2 g
± 0.2 g
V22PWE3T
V22PWE6T
V22PWE15T
V22XWE3T
V22XWE6T
V22XWE15T
3 kg x 0.0005 kg 6 kg x 0.001 kg 15 kg x 0.002 kg
3000 g x 0.5 g
6000 g x 1 g
15000 g x 2 g
Capacity x Readability (Max x d)
6 lb x 0.001 lb
15 lb x 0.002 lb
30 lb x 0.005 lb
96 oz x 0.02 oz 240 oz x 0.05 oz 480 oz x 0.1 oz
Maximum Displayed Resolution
6000
6000
7500
Repeatability (at 20° C)
0.5 g
1g
2g
Linearity (at 20° C)
± 0.5 g
± 1g
± 2g
Weighing Units
g, kg, lb, oz, lb:oz *
Tare Range
To capacity by subtraction
Stabilization Time
≤ 0.5 seconds
Safe Overload Protection
150 % of scale capacity
2 x Red LED (front and rear)
Weight Display
6-digit 7-segment, 20.5 mm / 0.8 in characters
Keyboard
V22PW: 2 buttons, V22XW: 4 buttons
Application Modes
V22PW: Weighing, V22XW: Weighing, Checkweighing
Battery Operating Time (at 20ºC)
Typically 50 hours with 12-hour full charge
V22PW: ABS housing with 304 stainless steel platform
Construction
V22XW: ABS bottom housing with 304 stainless steel top housing and platform
Ingress Protection
IPX8
Pan Dimensions
190 x 242 mm / 7.5 x 9.5 in
V22PW: 3.0 kg / 6.6 lb
Net Weight
V22XW: 3.9 kg / 8.6 lb
V22PW: 4.0 kg / 8.8 lb
Shipping Weight
V22XW: 4.9 kg / 10.8 lb
Shipping Dimensions
410 x 370 x 220 mm / 16.1 x 14.6 x 8.7 in
MODEL
Note: * For V22PW, see section 3.5.
6.2 Drawings and Dimensions
A
C
TABLE 6-2
A
B
B
Figure 6-1. Dimensions
C
V22PW
256 mm
10.1 in
280 mm
11 in
121 mm
4.8 in
V22XW
256 mm
10.1 in
288 mm
11.3 in
124 mm
4.9 in
Valor™ 2000W Series
EN-15
6.3 Compliance
Compliance to the following standards is indicated by the corresponding mark on
the product.
Mark
Standard
This product conforms to the EMC Directive 2004/108/EC, the Low
Voltage Directive 2006/95/EC. The Declaration of Conformity is
available online at
europe.ohaus.com/europe/en/home/support/compliance.aspx.
AS/NZS CISPR 11
CAN/CSA-C22.2 No. 61010-1-12
UL Std. No. 61010-1 (3rd edition)
NSF/ANSI 169–2009
NSF/ANSI/3-A 14159-1-2010
Disposal In conformance with the European Directive 2002/96/EC on Waste
Electrical and Electronic Equipment (WEEE) this device may not be
disposed of in domestic waste. This also applies to countries outside
the EU, per their specific requirements.
The Batteries Directive 2006/66/EC introduces new requirements
from September 2008 on removability of batteries from waste
equipment in EU Member States. To comply with this Directive, this
device has been designed for safe removal of the batteries at end-oflife by a waste treatment facility.
Please dispose of this product in accordance with local regulations at the collecting
point specified for electrical and electronic equipment. If you have any questions,
please contact the responsible authority or the distributor from which you
purchased this device.
Should this device be passed on to other parties (for private or professional use),
the content of this regulation must also be related.
Disposal instructions in Europe are available online at
europe.ohaus.com/europe/en/home/support/weee.aspx.
Thank you for your contribution to environmental protection.
EN-16
Valor™ 2000W Series
FCC Note
This equipment has been tested and found to comply with the limits for a Class B
digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to
provide reasonable protection against harmful interference when the equipment is
operated in a commercial environment. This equipment generates, uses, and can
radiate radio frequency energy and, if not installed and used in accordance with the
instruction manual, may cause harmful interference to radio communications.
Operation of this equipment in a residential area is likely to cause harmful
interference in which case the user will be required to correct the interference at his
own expense.
Industry Canada Note
This Class B digital apparatus complies with Canadian ICES-003.
ISO 9001 Registration
In 1994, OHAUS Corporation, USA, was awarded a certificate of registration to ISO
9001 by Bureau Veritus Quality International (BVQI), confirming that the OHAUS
quality management system is compliant with the ISO 9001 standard’s
requirements. On June 21, 2012, OHAUS Corporation, USA, was re-registered to
the ISO 9001:2008 standard.
Product Registration
Protect your investment. Register your product with your local OHAUS dealer or
online at www.ohaus.com.
Limited Warranty
OHAUS products are warranted against defects in materials and workmanship from
the date of delivery through the duration of the warranty period. During the warranty
period OHAUS will repair, or, at its option, replace any component(s) that proves to
be defective at no charge, provided that the product is returned, freight prepaid, to
OHAUS. This warranty does not apply if the product has been damaged by accident
or misuse, exposed to radioactive or corrosive materials, has foreign material
penetrating to the inside of the product, or as a result of service or modification by
other than OHAUS. In lieu of a properly returned warranty registration card, the
warranty period shall begin on the date of shipment to the authorized dealer. No
other express or implied warranty is given by OHAUS Corporation. OHAUS
Corporation shall not be liable for any consequential damages.