BENCHTOP WIRING ANALYZER
HORIZON II SERIES
Fully updated, low cost, ergonomic benchtop wiring analyzer makes
1500VDC/1067VAC HiPot testing faster and easier than ever!
The Horizon II is the next-generation
Horizon wiring analyzer. It has been
completely repackaged to further
enhance the user experience and improve
overall test performance.
Whether networked, integrated or used
as a standalone tester, Horizon II
provides virtually unlimited data storage
and reporting capabilities with built-in
flash disk and label printer support.
Key features include a fully updated
Windows XP-based user environment,
color touch-screen, real-time temperature
and humidity monitoring, expanded I/O,
flash disk data storage and improved
ergonomics. Operators will appreciate its
oversized, color touch-screen, built-in
‘START’ and ‘STOP’ buttons, and onscreen user help tips which makes it
exceptionally easy to program and test.
This all-in-one tester offers an unrivalled
diversity of test scenarios in a portable
wiring analyzer. All Horizon II models
enable 128-point testing with modular
expansion capability up to 1024 points
capable of switching 1A and HiPot
testing up to 1500 VDC and/or 1067
VAC.
The Horizon II (H2) software will accept
and run any Horizon test program, so
there is no need to re-create test
programs when moving to H2.
Standard test capability includes fault
location, automatic product learning,
component detection and real-time
intermittent open/short detection.
HORIZON II SERIES
PRODUCT DESCRIPTION
[Models]
[Measurement Specifications]
Horizon II multi-point cable and harness test systems are supplied with
a minimum of 128 Low or high-voltage test points and can be
configured with the following sources:
Programmable High Current Source (HCS) Measurement
Model
Horizon II
Voltage (AC)
Voltage (DC)
High Current
H2-LV
-
-
-
H2-HV1
-
1000
1A Fixed
H2-HV2
-
1500
1A Fixed
H2-HV3
1067
-
1A Fixed
H2-HV4
1067
1500
1A Fixed
H2-HV5
1067
1000
1A Fixed
H2-HV6
-
-
1A Fixed
[System Configuration]
•
•
Expansion boxes in
128-point
increments
•
•
Tools for periodic
verification of
system calibration
Label Printer
Support
•
•
•
•
•
The built-in PC and flash drive provides fast and capable computing
power allowing fully independent or networked operation of each
Horizon II tester
The following external computing capabilities can be easily added:
VGA monitor, label or standard printer, keyboard, mouse, network
capability and 3 I/O for PLC control
TCL script language for advanced applications
Supplied with ground probe, instruction manual, calibration
certificate and one-year warranty
Full range of product adapters
Custom adapter kits
Expansion boxes in 128-point increments with cable lengths of 3’,
6’, 9’ and 15’
Tools for periodic verification of system calibration
1A/28VDC programmable high-current source
Handheld digital probe for remote probing
8 I/O
Simplex and Kelvin test adapters
[Test Capabilities]










Auto-start test when product is loaded
Test Speed
500 nets per Sec. - Quick Scan (continuity and isolation)
7000 nets per Sec. - Quick Scan (continuity)
Self-learn known good products
High Voltage Breakdown (HiPot)
Insulation Resistance
Flex test for intermittent failures
Twisted pair verification
Resistors, capacitors, diodes, keys, LEDs
[Physical Data]
•
•
Power:
Dimensions:
•
Weight:

Operation Temp.
Accuracy
0Ω – 2.0Ω
±(5% + 100mΩ), with tare
4-Wire Resistance Measurement
Range
Accuracy
0Ω – 2.0Ω
±(2.5% + 5mΩ);
Functional Test (Voltage and Current)
Feature
Range
Accuracy
Resolution
Programmed Voltage
2V – 28V
±2.5%
0.1V
Programmed Current
1mA – 1000mA
±2.5%
1mA
4-Wire Voltage Drop Measurement
[Options and Accessories]
•
•
•
2-Wire Resistance Measurement
Range
Range
Accuracy
0V – 2.0V
±(2.5% + 2mV)
Fixed High Current Source Resistance Measurement
4-Wire Resistance Measurement
Range
Accuracy
Comments
0Ω – 2.0Ω
±(2% + 2mΩ)
Typical ±(2%+1mΩ)
4-Wire Voltage Drop Measurement
Range
Accuracy
0V – 2.0V
±(2.5% + 2mV)
DC HiPot Generator
Settings and Detections
Feature
Range
Accuracy
Output Voltage
50V – 1500V
±(1% + 2V)
1V
Dwell Time
0.05 – 600 sec
±2% -0/+ 50ms
10ms
Range
Accuracy
Resolution
1uA – 2.5mA
±(5% + 0.5µA)
0.1µA
Leakage Current Measurement
Insulation Resistance (IR) Measurements
Range
Accuracy
Lowest Voltage for max. accuracy over entire IR range
50KΩ - 100MΩ
±5%
100
100MΩ - 200MΩ
±5%
200
200MΩ - 500MΩ
±5%
500
500MΩ - 1GΩ
±5%
1000
1GΩ - 1.5GΩ
±5%
1500
AC HiPot Generator
110/220VAC, 50/60 Hz, 1A
532mm W x 273mm D x 175mm H
20.94” W x 10.75” D x 6.87” H
15kg (33 lbs) - main unit
5kg (11 lbs) - expansion box
0 – 40 °C
Settings and Detections
Feature
Range
Accuracy
Output Voltage
50V – 1067V
±(10% + 5V)
1V
Output Frequency
50Hz/60Hz
±5%
N/A
Dwell Time
0.4 – 600 sec.
±2% -0/+ 250ms
20ms @ 50Hz
16.66ms @ 60Hz
Arc Detection
Automatic
Aperture 1-10uSec
Voltage 100V - 1500V
Current 1% to 100%
Sensitivity
10mA/uSec
Neutralization Mode
On/Off
N/A
N/A
Arcs Only Mode
On/Off
N/A
N/A
[Measurement Specifications]
HA106 Standard
Adapter
MSR Resistance Measurements
2-Wire Measurements
Range
Accuracy
Comments
0Ω – 100KΩ
± (2.5% + 0.1Ω)
Typical ±(2%+0.02Ω)
± (2.5% + 1Ω)
Typical ±(2%+ 0.5Ω)
Range
Accuracy
Resolution
±7%
Typical ±5%
0.1mA – 8mA
±(10% + 5µA)
5µA
10.1MΩ – 25MΩ
±15%
25.1MΩ– 50MΩ
±30%
Use DC HiPot for better accuracy
in these ranges
Range
Accuracy
Comments
0Ω – 400Ω
± (2.5% + 2mΩ)
Typical reading : Typical ±(2%+1mΩ)
MSR Capacitance Measurements
Capacitance Measurement
Range
2. CableTest Systems reserves the right to update
specifications at any time, please visit
www.cabletest.com for latest updates.
Leakage Current Measurement
0Ω – 100KΩ
Notes:
1. Certain features in this datasheet are optional.
Some of the technical parameters may under
certain conditions, for further information please
refer to full version datasheet or contact CableTest
Systems.
Resolution
101KΩ – 10MΩ
4-Wire Measurements
HA304 Custom
Adapter Kit
Resolution
50pF -1uF
50pF -1uF
1uF – 10mF
Accuracy
± (5% + 10pF)
± (5% + 20pF)
Comments
With tare
Systems with more than 256 test points
± (5% + 2nF)
Without tare
± (5% + 5nF)
Systems with more than 256 test points
± 10%
With or without tare
Insulation Resistance (IR) Measurements
Range
Accuracy
Lowest Voltage for max. accuracy over entire IR range
12kΩ – 1MΩ
±20%
100 V
25kΩ – 2MΩ
±20%
200 V
62kΩ – 5MΩ
±20%
500 V
93kΩ – 7MΩ
±20%
750 V
125kΩ – 10MΩ
±20%
1000 V
Open as PDF