PC-8423
MEMORY HiLOGGER
efficienc
Highspeed
High
MultiWithst
channe
Voltag
PCefficie
Highspeed
Multichann
Fast 10-ms Sampling
Up to 600 Channels Data Logging
MEMORY HiLOGGER Model 8423 is a data acquisition system capable of
measuring and recording multiple channels at high speed. Acquired data can
be easily analyzed on a personal computer. This model is ideal for acquiring
data for evaluation and testing at development sites. If your evaluation
needs require faster data sampling than was available with former HIOKI
MEMORY HiLOGGERs, or if you just need more measurement channels,
this model has the capabilities you want.
2
Who needs 10 ms high-speed sampling?
- Answer -
To acquire data when converting automobile
electronics for electric or hybrid vehicles
● Fastest measurement interval (sampling interval) is 10 ms
● Acquires up to 600 channels of data with 10 ms sampling interval
● Insulation withstand voltage between the measurement channels in each module is 200 V
(Model 8948)
In the development of electric and hybrid automobiles, the
need to capture sudden swings in various loads requires
a measurement instrument with multi-channel highspeed sampling capability. For this purpose, HIOKI has
developed a very economical logger that can measure with
Sudden-load-change testing of a fuel cell employs dual sampling to
measure with 10-ms (upper trace) and 100-ms sampling (lower trace).
(Timebase: 50 ms/div).
10-ms sampling interval on all channels. Also included is
a dual-sampling function that can measure at two different
sampling rates simultaneously. This new model can follow
waveforms that former 100-ms-sampling instruments could
not.
A 5-Hz pulse waveform is measured using dual sampling: 10-ms
(upper trace) and 100-ms sampling (lower trace)
(Timebase: 50 ms/div).
3
Who needs 120 or 600 channels?
- Answer -
To acquire multi-point temperature distribution data
To measure the voltage of each cell in a stack
● Expandable up to 120 channels with a single instrument
● Up to five instruments can be connected for measuring up to 600 channels
● Isolated to sustain up to 600 V between modules and earth
Temperature distribution is measured to evaluate air
conditioning systems during development. A system to
acquire data on up to 600 channels can be constructed
with merely a LAN or USB connection, providing highly
detailed temperature distribution measurements.
With all channels isolated and a 600V AC/DC maximum
rated voltage to earth, even when the common mode voltage
increases as is common with layered batteries, the voltage
of each individual battery cell can be safely measured.
Radiation through ceiling and roof
Window
Thermocouple
Maximum rated
voltage to earth
between each
module:
600V AC, DC
External wall
Door
Draft
Ventilation
Maximum rated
voltage to earth:
600V AC, DC
Under the floor
Thermocouple
To e v a l u a t e h e a t r a d i a t i o n
characteristics and copy machine
operation, temperat u res at
many points inside the chassis
a nd a na log volt ages f rom t he
control board are simultaneously
measured.
Maximum rated voltage between
each channel: 200V DC
200V in model 8948
120V in model 8949
4
“Simplicity” as a Design Concept
Installation
Easily add input modules: just align and mate the connectors
on the left side of the instrument assembly, and turn
the metal clasp. For added strength, attach the supplied
mounting bracket on the rear, or attach a standard DIN rail
to the rear for tray or rack mounting.
Because the terminal blocks are designed to be removable,
thermocouples can be connected to the terminal block in
hand before connecting the block to a HiLOGGER input
module, with just one touch.
Each measurement module daisy-chains
onto the stacking configuration.
Mounting with a standard DIN rail is supported.
Measurement configuration settings
Logger configuration settings are made from a computer
running the supplied application program. Settings can be
easily made using familiar PC operations. To keep the process
simple, the user is guided sequentially through the setting items.
1
2
3
4
5
6
7
8
5
View your data even while measuring!
Direct connection
with a LAN crossover
or USB cable
Data is recorded on the computer in real time using the
supplied Logger Utility PC application program. View a
trend graph in a window and scroll back to view earlier
waveform data, even while recording.
Post-measurement analysis (New Double-Thumb function*)
The newly developed Double-Thumb function simplifies analysis.
Two windows are displayed side by side, each with a scroll bar at the
bottom containing a thumb (scroll box) that corresponds to the length
and position of that window’s displayed segment within the overall
waveform. The thumbs in the scroll bars of the waveform display
windows show you the position of the segments at a glance, greatly
simplifying scrolling operations.
Entire Recording Length: 1 s/div
Segment A
50 ms/div magnification
Segment A View
Segment B
100 ms/div magnification
Segment B View
6
More Functional Details
Universal isolated inputs for temperature, voltage and pulses
*1 Pt (platinum resistance temperature sensor) and humidity measurements require UNIVERSAL UNIT 8949
*2 Requires optional HUMIDITY SENSOR 9701
With the modular input design, you can select the input modules
appropriate for your measurement application. Select from voltage
and temperature (thermocouple or Pt input*1) and humidity.*1 *2 Also,
Digital Pulse Module 8996 provides 15 input channels for totalization/
rotation counts and Hi/Lo logic measurements. In addition to interchannel input isolation, the PC connection interface is completely
isolated from the measurement terminals, minimizing shock hazards
and interference even when measuring thermocouple and voltage
inputs at the same time.
Voltage
100 mV f.s.
1 V f.s.
10 V f.s.
20 V f.s.
100 V f.s.
1-5 V f.s.
Voltage
100 mV f.s.
1 V f.s.
10 V f.s.
20 V f.s.
100 V f.s.
1-5 V f.s.
Thermocouples
100 °C f.s.
500 °C f.s.
2000 °C f.s.
Thermocouples
100 °C f.s.
500 °C f.s.
2000 °C f.s.
Pt/JPt 100
100 °C f.s.
500 °C f.s.
2000 °C f.s.
Humidity
5 to 95% rh
Note: Isolation between channels is possible through the use of semi-conductor
VOLTAGE/TEMP UNIT 8948
relays. Voltage exceeding the product specifications, such as that originating from
lightning surges or other sources, should never be applied between each channel;
otherwise the relays will short and the recorder will be damaged.
UNIVERSAL UNIT 8949
DIGITAL/PULSE UNIT 8996
Real-time saving to CF Card
Each measurement can be saved to a CF Card in real time.
Continuous long-term recording can be performed with
high capacity CF Cards up to 1 GB. Data can be viewed
on a PC using the supplied Logger Utility program.
Recording Times with a 512 MB Card (Voltage, Temperature and Humidity Measurements, but no Pulse Channels)
Recording
intervals
This exclusive technology has been developed to preserve data as reliably
as possible in the event of a power failure, by incorporating memory card
technology with the know-how built into the MEMORY HiLOGGER
8420 -50, 8421-50 and 8422-50 series. The 8423 emphasizes the existing
HiLOGGER functions and maintains internal supply voltage with a
large internal capacitor until all data
has been saved to the ca rd, resulting
in greater reliability when acqui r ing
large amounts of data.
50ms
155 d 08 h 16 min 10 d 08 h 33 min 41 d 10 h 12 min 20ms
100ms
31 d 01 h 39 min 62 d 03 h 18 min "H"
"H"
1s
"H"
10s
1min
10min
1hour
2 d 01 h 42 min 4 d 03 h 25 min 512MB
512MB
(using 30 channels)
1 d 00 h 51 min (using 60 channels)
512MB
(using 120 channels)
12 h 25 min 2 d 01 h 42 min 1 d 00 h 51 min 10 d 08 h 33 min 5 d 04 h 16 min 1 d 07 h 04 min
20 d 17 h 06 min 10 d 08 h 33 min 5 d 04 h 16 min
207 d 03 h 01 min 103 d 13 h 30 min 51 d 18 h 45 min 20 d 17 h 06 min 103 d 13 h 30 min 5 d 04 h 16 min 6 h 12 min
12 h 25 min
2 d 14 h 08 min 310 d 16 h 32 min 500ms
51 d 18 h 45 min 25 d 21 h 22 min 2 d 14 h 08 min
12 d 22 h 41 min
25 d 21 h 22 min
"H"
"H"
"H"
"H"
258 d 21 h 47 min
"H"
"H"
"H"
"H"
"H"
"H"
"H"
"H"
"H"
"H"
"H"
"H"
"H"
"H"
"H"
Note: Actual CF data capacity is less than total CF storage capacity, and waveform file headers are not included
in these calculated values, so we recommend using 90% of these values for estimation purposes.
Note: "H" Periods longer than 1 year is abbreviated.
Recording Times with a 512 MB Card (Pulse Channels use only)
Recording
intervals
512MB
10ms
50ms
20ms
100ms
200ms
A CF Card slot is included as a standard feature,
supporting HIOKI CF Cards up to 1 GB (operation with
non-HIOKI-brand cards is not guaranteed). Using a CF
Card, instrument settings can be easily copied from one
8423 to another.
512MB
(using 15 channels)
10ms
200ms
Enhanced data protection from power failures
512MB
(using 1 channel)
500ms
1s
10s
1min
10min
1hour
512MB
(using 1 channel)
15 d 12 h 49 min 31 d 01 h 39 min 77 d 16 h 08 min (using 15 channels)
1 d 00 h 51 min 2 d 01 h 42 min 5 d 04 h 16 min 155 d 08 h 16 min 10 d 08 h 33 min 51 d 18 h 45 min 310 d 16 h 32 min "H"
"H"
20 d 17 h 06 min 103 d 13 h 30 min 512MB
512MB
(using 30 channels)
12 h 25 min (using 60 channels)
512MB
(using 120 channels)
6 h 12 min 3 h 06 min
2 d 14 h 08 min 1 d 07 h 04 min 15 h 32 min
10 d 08 h 33 min 5 d 04 h 16 min 1 d 00 h 51 min 5 d 04 h 16 min 25 d 21 h 22 min 51 d 18 h 45 min 12 h 25 min 2 d 14 h 08 min 12 d 22 h 41 min 25 d 21 h 22 min 6 h 12 min
1 d 07 h 04 min
2 d 14 h 08 min
6 d 11 h 20 min
12 d 22 h 41 min
258 d 21 h 47 min 129 d 10 h 53 min
"H"
"H"
"H"
"H"
"H"
"H"
"H"
"H"
"H"
"H"
"H"
"H"
"H"
"H"
"H"
"H"
"H"
"H"
Note: Actual CF data capacity is less than total CF storage capacity, and waveform file headers are not included
in these calculated values, so we recommend using 90% of these values for estimation purposes.
Note: "H" Periods longer than 1 year is abbreviated.
Trigger function
Level, Window and Logic trigger functions are provided.
You can have one criterion start recording and another stop
recording.
7
Dual Sampling
Two different measurement intervals can be specified
at the same time (one interval setting per input module).
Using dual sampling, the appropriate measurement
interval can be set for each type of object to be measured,
optimizing use of internal memory and CF Card capacity.
ds
ng
ng
pli
pli econ
m
m
a
nd
s
s
sa
w se c o
st 0 0 m
o
l
a
S
F
1
1
Enhanced PC Interface
USB Port Included
A USB 2.0 (mini-B connector) port is included as standard. The
8423 instrument and a PC can be connected by a USB cable (A
to mini-B) for transferring 8423 operating settings and data.
LAN Terminal Included
A 100Base-TX LAN terminal is included as standard. The
8423 instrument and a PC can be connected by a LAN cable
for transferring 8423 operating settings and data.
External Control Inputs Included
Input terminals are provided for external triggering, external start and stop
and external sampling. External signals can be applied as a trigger source
and to start and stop measurements, so data can be acquired by controlled
sampling timing.
Note: External triggering and external sampling share a common terminal, so only one of these control input types can be
used at a time.
8
More Functional Details
All-Channel Synchronous Measurement Capability
When measuring up to 120 channels on combined modules,
all input channels are sampled synchronously. When multiple
8423s are connected via LAN or USB for measuring up to 600
channels, the sampling of each instrument in the system can be
synchronized using optional Connection Cable Model 9683.
As well as PC-based data collection, measurement start and
stop can be controlled by the [START/MARK] and [STOP]
keys on a master 8423.
Note: Any 8423 may be designated as the master. Only the initial acquisition criteria setting needs to
be performed on a PC via USB or LAN.
Enhanced Noise Immunity
A delta-sigma type A/D converter has been incorporated
in the measurement circuitry. The effects of previously
problematic inverter switching noise and 50/60 Hz hum
noise have been greatly reduced by the digital filtering
function using the oversampling principle inherent in this
type of device.
Note: Optimum noise suppression is obtained with recording intervals of two seconds or
longer
System-1
(Max. 120ch)
+
System-2
(Max. 120ch)
+
HUB
(LAN or USB)
Synchronize cable 9683
+
System-5
(Max. 120ch)
LAN cable (100BASE-TX)
or
USB cable (mini B terminal at 8423 side)
9
■Product Specifications
8423 Hardware Specifications (accuracy is specified @23 ±5˚C/73 ±9˚F, 30 to 80 % rh, from 30 minutes after power on, accuracy guaranteed for 1 year, Post-adjustment accuracy guaranteed for 1 year, product guaranteed for 1 year)
Display
Memory capacity
External control
connectors
LCD, 16 characters × 2 lines, 5 × 8 dots / characters
Total 16 M-word (about 16.77 million data points: 32 mega-bytes)
Push-button type terminal block : External trigger/ External sampling input (exclusive OR), External start input, External stop input
External sampling : rise-up, or fall-down (selectable)
Rise-up : Low (0 to 1.0 V) to High (2.5 to 5.0 V)
Fall-down : High (2.5 to 5.0 V) to Low (0 to 1.0 V), or terminal short
Input voltage range : -5 to 10 V DC, Filter ON/OFF possible
Pulse width response : Over 1 ms at "H", over 2 ms at "L" (at filter OFF), Over 2.5 ms at "H", over 4 ms at "L" (at filter ON)
Maximum external sampling period : 10 ms (at digital filter OFF), 20 ms (at digital filter OFF, and synchronous measurement), 5 s (at digital filter ON, and combined with
humidity measurement)
Clock
Accuracy of timebase
Recording intervals
Recording length
Recording mode
Number of data
Durability of battery
No. of connectable units
Environmental conditions
Conforming standards
Power supply
Power consumption
Dimensions & Mass
Accessories
PC Interface
Data storage media
Interface
PC control
Synchronous sampling : Five-units maximum for synchronous connection, Function : Connect via the connection cable model 9683 for synchronous sampling
Auto calendar, leap year auto distinguish, Precision : ±0.2s/ day at power ON, ±3s/ day at power OFF (at 23 ˚C/ 73˚F)
±0.2s/ day on measurement (at 23 ˚C/ 73˚F)
10ms, 20ms, 50ms, 100ms, 200ms, 500ms, 1s, 2s, 5s, 10s, 20s, 30s, 1min, 2min, 5min, 10min, 20min, 30min, 1hr (5s to 1hr when combined with humidity measurement)
Set to arbitrary length or continuous; Data storage : last 16-mega datas in internal memory (for one channel recording. For n channels, 16 M-datas / n data)
Continue, Repeat, Timer measurement
For analog "n" channels, ( 16-mega datas / n ) datas
Backup battery for clock and setting conditions: battery life of at least 10 years, For measurement data: none (at 23 ˚C/ 73˚F)
Maximum 8 units (total 120 channels)
Operating temperature and humidity : 0 (32˚F) to 40˚C (104˚F), 30 to 80% rh, Storage temperature and humidity : -10 (14˚F) to 50˚C (122˚F), 80% rh or less, (non-condensating)
Safety : EN61010, EMC : EN61326, EN61000-3-2, EN61000-3-3
(1) Using the AC ADAPTER 9418 -15, 100 to 240 VAC, 50/60 Hz (2) External DC Power: 9.6 V to 15.6 VDC (Please contact HIOKI for connection cord)
Using the AC adapter 9418-15: 55 VA Max. (include AC adapter), 20 VA Max. (main unit only) (when connected with 8 units),
External DC Power: 20 VA Max. (when connected with 8 units)
Approx. 67 mm (2.64 in) W × 133 mm (5.24 in) H × 125 mm (4.92 in) D, 600 g (21.2 oz)
Operating Manual ×1, Quick Start Manual ×1, AC ADAPTER 9418 -15 ×1, USB cable ×1, Connection Plate ×1, CD-R (data collection software
"Logger Utility") ×1, Connector cover ×1, Ferrite clamp ×1
CF card slot × 1 (Up to 1GB), MS-DOS format, Note: Cannot use with the 9830 (2GB) card
LAN : supports 100Base-TX, DHCP, DNS
USB : Ver 2.0, mini-B receptacle
Data acquisition and measurement criteria settings are controlled by the PC data acquisition program; data acquired to internal memory and CF Cards is
downloaded via FTP server function; simple operations (measurement start/stop and data acquisition to internal memory) are available via HTTP server function
Function Specifications
Major Functions
Measurement
parameters
Real time save
Dual sampling
Marking
Trigger function
Control the input units, or output units, Communication to the PC, Data storage to the CF card
Depending on the connected measurement unit: Temperature (thermocouple, Pt), voltage, humidity (used optional sensor), totalized pulses (addition,
instantly), rotation count, digital signal
Measurement data are saved as binary data to the CF Card in real time, and can be saved to separate files at preset times, selectable as full files or
an endless loop with automatic deletion of oldest data.
Two (high-speed and low-speed) recording intervals can be specified for every input module from the following: 10, 20, 50, 100, 200 and 500 ms;
1, 2, 5, 10, 20 and 30 s; 1, 2, 5, 10, 20 and 30 min; and 1 hr (the low-speed setting divided by the high-speed setting must be an integer less than 1,000)
Event mark input : Press [Start / Stop] key at measuremet
Mode : Single / Repeat,
Timing : Start / Stop / Start & Stop, Pre-Trigger : records period before trigger, can be set for real-time saving
Trigger source
Analog input : Maximum 120 channels, depend on number of the input unit.
Pulse totalizer inputs : Maximum 120 channels, depend on number of the input unit.
Logic inputs : Maximum 120 channels, depend on number of the input unit.
External trigger : Rise up or fall down of the external input signal (selectable)
Logical AND or OR for each trigger source, Trigger condition settable for each channels
Trigger type
Level: Triggers when rising or falling through preset level
Window: Triggers when entering or exiting range defined by preset upper and lower limit values
Trigger level resolution : 0.1 % f.s. Logic : 1, 0, × Pattern trigger
External trigger
signal
Rise up : Low level (0 to 1.0 V) to High level (2.5 V to 5.0 V)
Fall down : High level (2.5 V to 5.0 V) to Low level (0 to 1.0 V), or terminal short
Input voltage range : -5 V to 10 V, Filter ON/OFF possible, Pulse width response : more than 1 ms (High period), more than 2 ms (Low period)
at filter OFF, more than 2.5 ms (High period), more than 4 ms (Low period) at filter ON
Alarm output
Alarm Module 8997 can be connected along with various measurement modules (although it cannot be connected alone)
Alarm type
Level: Triggers when rising or falling through preset level
Window: Triggers when entering or exiting range defined by preset upper and lower limit values
Logic pattern : agreement (or disagreement) in the specified pattern
Output latch settings : latch / no latch
Start backup
Possible
10
■ Specification
■ Bundled software specifications
Logger Utility (bundled application software)
Supported units
Model 8423, 8430-20, LR8431-20, LR8432-20, LR8400-20, LR8401-20,
LR8402-20, and LR8410-20
Operating environment
Windows 10/8/7 (32bit/64bit), Vista (32bit/64bit), XP (with SP2 or later)
(32bit)
Real-time data
acquisition
Measurements on multiple loggers connected by LAN or USB can be
controlled to sequentially acquire, display and save waveform data (for
recording up to 10 million samples)
Number of controllable instruments: up to 5 units
(This software is compatible only with the LR8410-20, LR8400 -20series,
LR8431-20, 8423, and 8430-20)
Display: Waveforms (time-axis divided display possible), numerical values
(logging), and alarm status can be displayed at the same time
Numerical value display: Can be monitored in a separate window
Scroll: Waveform scroll while measuring
Data saving destination: Real-time data transfer to Excel, or Real-time data
acquisition file (LUW format)
Event marks: Can be set while measuring
Data acquisition
settings
Data acquisition settings for the logger or logging station
Saving: The setting for multiple loggers or logging stations can be saved
together in one file (LUS format); Instrument configuration settings can
be sent and received
Waveform display
Processed data file: Real-time data acquisition file (LUW format), Record
to internal memory data (MEM format)
Display format: Simultaneously display waveform and numerical value,
(time-axis divided display possible)
Maximum number of channels: 675 channerls (measurement data) + 60
channels (waveform processing data)
Others: Display each channel’s waveform on 10 sheets, scroll, record event
mark, cursor, screen hard copy, numerical value display
VOLTAGE/TEMP UNIT 8948
Input
Voltage
Processing items: Four arithmetic operations
Number of processing channels: 60 channerls
Target data: Real-time data acquisition file (LUW format), record to internal memory data (MEM format), data acquired in real time, waveform
processing data
Parameter calcuCalculation items: Average, peak, maximum values, time to maximum
lations
values, minimum values, time to minimum values, ON time, OFF time,
count the number of ON time and OFF time, standard deviation, integration, area values, totalization
Target data: Real-time data acquisition file (LUW format), record to internal memory data (MEM format)
Search functions Search mode: Event mark, time and date, maximum position, minimum
position, maximum pole, minimum pole, alarm position, level, window,
amount of change
Print functions
Supported printer: Printer compatible with the OS
Target data: Real-time data acquisition file (LUW format), record to internal memory data (MEM format)
Print format: Waveform image, report format, list print (channel settings,
event, cursor value)
Print area: The entire area, area between cursors A and B
Print preview: Supported
(accuracy specified @23 ±5˚C/73 ±9˚F, 30 to 80% rh., from 30 minutes after power on and after zero point adjustment, accuracy and Post-adjustment accuracy and product guaranteed for 1 year)
100mV f.s.
1V f.s.
10V f.s.
20V f.s.
100V f.s.
1–5V f.s.
Setting Range
A/D conversion
Filter function
Max. allowable
input
Waveform processing
Measurement parameters : Voltage, Thermocouples (K, E, J, T, N, W, R, S, B)
Terminal : M3 (mm) screw terminals (2 terminals/1ch), terminal block removable, supplied terminal block cover
Number of channels : 15 channels isolated from each other and chassis, (voltage or thermocouple selectable for each channels)
Input impedance: 1MΩ (850kΩ when open-circuit polling is enabled)
Setting Range
Measurement
parameters
Target data: Real-time data acquisition file (LUW format), record to internal memory data (MEM format)
Data conversion Converted sections: All data, designation section
Format: CSV format (separate by comma, space, tab), transfer to Excel
spreadsheet, arbitrary data thinning
K 100˚C f.s.
K 500˚C f.s.
K 2000˚C f.s.
E 100˚C f.s.
E 500˚C f.s.
E 2000˚C f.s.
Thermocouples J 100˚C f.s.
Excluding standard
J 500˚C f.s.
reference contact
J 2000˚C f.s.
accuracy
T 100˚C f.s.
T 500˚C f.s.
T 2000˚C f.s.
N 100˚C f.s.
N 500˚C f.s.
N 2000˚C f.s.
Measurement range
-150mV to +150mV
-1.5V to +1.5V
-15V to +15V
-30V to +30V
-100V to +100V
1V to 5V
Measurement range
-100˚C to 100˚C
-200˚C to 500˚C
-200˚C to 1350˚C
-100˚C to 100˚C
-200˚C to 500˚C
-200˚C to 1000˚C
-100˚C to 100˚C
-200˚C to 500˚C
-200˚C to 1200˚C
-100˚C to 100˚C
-200˚C to 400˚C
-200˚C to 400˚C
-100˚C to 100˚C
-200˚C to 500˚C
-200˚C to 1300˚C
Resolution
5mV
50mV
500mV
1mV
5mV
500mV
Resolution
0.01˚C
0.05˚C
0.1˚C
0.01˚C
0.05˚C
0.1˚C
0.01˚C
0.05˚C
0.1˚C
0.01˚C
0.05˚C
0.1˚C
0.01˚C
0.05˚C
0.1˚C
Accuracy
±0.1% f.s.
Note: at 1-5 V
range, f.s.=10 V
Accuracy
Setting Range
Resolution
0˚C to 100˚C
0˚C to 500˚C
0˚C to 1700˚C
0˚C to 100˚C
0˚C to 500˚C
0˚C to 1700˚C
0˚C to 1800˚C
0.01˚C
0.05˚C
0.1˚C
0.01˚C
0.05˚C
0.1˚C
0.1˚C
W 100˚C f.s.
W 500˚C f.s.
W 2000˚C f.s.
0˚C to 100˚C
0˚C to 500˚C
0˚C to 2000˚C
0.01˚C
0.05˚C
0.1˚C
W : Wre5-26
Accuracy
±0.05% f.s. ±3.5˚C
(0˚C to less than 400˚C)
(Temperatures less than
400˚C measured by B
thermocouples are not
guaranteed for accuracy)
±0.05% f.s. ±2˚C
(400˚C and above)
Standard reference contact
±0.05% f.s. ±1˚C
Accuracy with internal
compensation, add to measurement accuracy
Switching
±0.5˚C (K, E, J, T)
±1.0˚C (N, R, S, B, W)
Switchable between internal and external
Resolution : 16 bit, Maximum sampling speed : 10 ms
Digital filter : OFF, 50 Hz, 60 Hz (With 50 and 60 Hz settings, the digital filter is automatically set according to recording interval)
Max. allowable input : 100 V DC (maximum voltage between input terminals that does not cause damage), Max. rated voltage between channels : 200 V DC
Max. rated voltage to earth : 600 V DC, AC (Upper limit voltage that does not cause damage when applied between input channel and chassis, and between each input channels)
Conforming standards Safety : EN61010, EMC : EN61326
Dimensions & Mass Approx. 38.5 mm (1.52 in) W × 133 mm (5.24 in) H × 141.2 mm (5.56 in) D mm, 550 g (19.4 oz)
Accessories
Measurement range
R 100˚C f.s.
R 500˚C f.s.
R 2000˚C f.s.
S 100˚C f.s.
Thermocouples S 500˚C f.s.
Excluding standard
S 2000˚C f.s.
reference contact
B 2000˚C f.s.
accuracy
Connection Plate ×1, Operating Manual ×1
11
■ Specification
UNIVERSAL UNIT 8949 Input
(accuracy specified @23 ±5˚C/73 ±9˚F, 30 to 80% rh., from 30 minutes after power on and after zero point adjustment, accuracy and Post-adjustment accuracy and product guaranteed for 1 year)
Measurement parameters : Voltage, Thermocouples (K, E, J, T, N, W, R, S, B), Resistance temperature sensor (Pt 100, JPt 100), Humidity (only use with the Model 9701 sensor)
Terminal : Screw-type terminals (4 terminals/1ch), terminal block removable, supplied terminal block cover Number of channels : 15 channels (input type selectable for each
channels), Iisolated from each other and chassis (at voltage or thermocouples), Not isolated from each other and common GND (at resistance temperature sensor or humidity)
Input impedance: 1MΩ (850kΩ when open-circuit polling is enabled at thermocouples), 2MΩ (when resistance temperature sensor)
Setting Range
Voltage
100mV f.s.
1V f.s.
10V f.s.
20V f.s.
100V f.s.
1–5V f.s.
Setting Range
Measurement
parameters
A/D conversion
Filter function
Max. allowable
input
K 100˚C f.s.
K 500˚C f.s.
K 2000˚C f.s.
E 100˚C f.s.
E 500˚C f.s.
E 2000˚C f.s.
Thermocouples J 100˚C f.s.
Exclude the standard
J 500˚C f.s.
reference contact
J 2000˚C f.s.
accuracy
T 100˚C f.s.
T 500˚C f.s.
T 2000˚C f.s.
N 100˚C f.s.
N 500˚C f.s.
N 2000˚C f.s.
Measurement range
-150mV to +150mV
-1.5V to +1.5V
-15V to +15V
-30V to +30V
-60V to +60V
1V to 5V
Measurement range
-100˚C to 100˚C
-200˚C to 500˚C
-200˚C to 1350˚C
-100˚C to 100˚C
-200˚C to 500˚C
-200˚C to 1000˚C
-100˚C to 100˚C
-200˚C to 500˚C
-200˚C to 1200˚C
-100˚C to 100˚C
-200˚C to 400˚C
-200˚C to 400˚C
-100˚C to 100˚C
-200˚C to 500˚C
-200˚C to 1300˚C
Resolution
5mV
50mV
500mV
1mV
5mV
500mV
Accuracy
±0.1% f.s.
Note: at 1-5 V
range, f.s.=10 V
Resolution
0.01˚C
0.05˚C
0.1˚C
0.01˚C
0.05˚C
0.1˚C
0.01˚C
0.05˚C
0.1˚C
0.01˚C
0.05˚C
0.1˚C
0.01˚C
0.05˚C
0.1˚C
Setting Range
Accuracy
Resolution
0˚C to 100˚C
0˚C to 500˚C
0˚C to 1700˚C
0˚C to 100˚C
0˚C to 500˚C
0˚C to 1700˚C
0˚C to 1800˚C
0.01˚C
0.05˚C
0.1˚C
0.01˚C
0.05˚C
0.1˚C
0.1˚C
W 100˚C f.s.
W 500˚C f.s.
W 2000˚C f.s.
0˚C to 100˚C
0˚C to 500˚C
0˚C to 2000˚C
0.01˚C
0.05˚C
0.1˚C
W : Wre5-26
Accuracy
±0.05% f.s. ±3.5˚C
(0˚C to less than 400˚C)
(Temperatures less than
400˚C measured by B
thermocouples are not
guaranteed for accuracy)
±0.05% f.s. ±2˚C
(400˚C and above)
Standard reference contact accuracy
±0.5˚C (K, E, J, T) ±1.0˚C (N, R, S, B, W)
Switching
Switchable between internal and external
with internal compensation, add to measurement accuracy
±0.05% f.s. ±1˚C
Measurement range
R 100˚C f.s.
R 500˚C f.s.
R 2000˚C f.s.
S 100˚C f.s.
Thermocouples S 500˚C f.s.
Exclude the standard
S 2000˚C f.s.
reference contact
B 2000˚C f.s.
accuracy
Setting Range
100˚C f.s.
Resistance
temperature sensor 500˚C f.s.
2000˚C f.s.
100˚C f.s.
Resistance
temperature sensor 500˚C f.s.
JPt 100, JIS C 1604-1989
2000˚C f.s.
Humidity
100% rh
Pt 100, JIS C 1604-1997
Measurement range
-100˚C to 100˚C
-200˚C to 500˚C
-200˚C to 800˚C
-100˚C to 100˚C
-200˚C to 500˚C
-200˚C to 500˚C
5.0 to 95.0% rh
Resolution
0.01˚C
0.05˚C
0.1˚C
0.01˚C
0.05˚C
0.1˚C
0.1% rh
Accuracy
±0.05% f.s. ±0.5˚C
±0.05% f.s. ±0.5˚C
Refer to the accuracy
table
Resolution : 16 bit, Maximum sampling speed : 10 ms (5 s when combined with humidity measurement)
Digital filter : OFF, 50 Hz, 60 Hz (With 50 and 60 Hz settings, the digital filter is automatically set according to recording interval)
Max. allowable input : 60 V DC (maximum voltage between input terminals that does not cause damage), Max. rated voltage between channels : 120 V DC
Max. rated voltage to earth : 600 V DC, AC (Upper limit voltage that does not cause damage when applied between input channel and chassis, and between each input channels)
Conforming standards Safety : EN61010, EMC : EN61326
Dimensions & Mass Approx. 38.5 mm (1.52 in) W × 133 mm (5.24 in) H × 141.2 mm (5.56 in) D mm, 530 g (18.7 oz)
Accessories
DIGITAL/PULSE UNIT 8996 Input
Pulse input
Digital input
n Humidity sensor 9701 accuracy
Flat-blade Screwdriver ×1 (for terminal block), Connection Plate ×1, Operating Manual ×1
(product guaranteed for one year)
Input signal condition : No-voltage 'a' contact (normally open contact), open collector or voltage input, Digital / Pulse input selectable for each channels
Measurement parameters : Voltage, Totalized pulses (integrated or instantaneous), Rotation count, ON/OFF digital signal
Terminal : M3 (mm) screw terminals (2 terminals/1ch), terminal block removable, supplied terminal block cover
Number of channels : 15 channels (digital / pulse selectable for each channels) (common ground for CH-1 to CH-5, common ground for CH-6 to CH-10, common ground
for CH-11 to CH-15) Input impedance : 1.1MΩ
Totalized pulses
Rotation count
Logic detection
level
Setting Range
1,000M pulse f.s.
5,000/n (r/s) f.s.
Measurement range
0 to 1,000M pulse
0 to 5,000/n (r/s)
Resolution
1 pulse
1/n (r/s)
Note: n = pulses per rotation (1 to 1,000)
HIGH = at least 1.0 V, LOW = 0 to 0.5 V
HIGH = at least 4.0 V, LOW = 0 to 1.5 V
Pulse input period
with filter OFF
with filter ON
Filter
Slope
Detection level
200 μs or more (both H and L periods must be at least 100 μs)
100 ms or more (both H and L periods must be at least 50 ms)
Chatter-prevention filter : can be set ON/OFF for each channels
Rising or falling edge can be set for each channel
HIGH = at least 1.0 V, LOW = 0 to 0.5 V
HIGH = at least 4.0 V, LOW = 0 to 1.5 V
Max. allowable input
50 V DC (maximum voltage between input terminals that does not cause damage)
Max. rated voltage to each
channels
33 V AC rms, 70 V DC (Upper limit voltage that does not cause damage when applied between CH-1 to CH-5 each channel and CH-6 to CH-10 each channel, CH-6 to CH-10
Max. rated voltage to earth 600 V DC, AC
(Upper limit voltage that does not cause damage when applied between CH-1 to CH-5 each channel and chassis, CH-6 to CH-10 each channel and chassis, CH-11 to CH-15 each channel and chassis, and between each UNITs)
each channel and CH-11 to CH-15 each channel, CH-1 to CH-5 each channel and CH-11 to CH-15 each channel)
Conforming standards Safety : EN61010, EMC : EN61326
Dimensions & Mass Approx. 38.5 mm (1.52 in) W × 133 mm (5.24 in) H × 141.2 mm (5.56 in) D mm, 500 g (17.6 oz)
Accessories
Connection Plate ×1, Operating Manual ×1
ALARM UNIT 8997
Output
(product guaranteed for one year)
Output type : open collector (active low)
Alarm parameters : Use up to 15 channels in response to analog input, pulse input, rotation count, or ON/OFF digital signal
Terminal : M3 (mm) screw terminals (2 terminals/1ch)
Number of channels : 15 channels isolated from each other and chassis
Output sink current Maximum switching capability : 5 to 60 V DC @10 mA (open collector drive)
Output latch settings : Latch / No latch at every recording interval
Output refresh
Max. rated voltage to earth
Max. rated voltage to each
channels
Conforming standards
Dimensions & Mass
Accessories
600 V DC, AC (Upper limit voltage that does not cause damage when applied between each output channel and chassis, and between each units)
33 V AC rms, 70 V DC (Upper limit voltage that does not cause damage when applied between each output channels)
Safety : EN61010, EMC : EN61326
Approx. 38.5 mm (1.52 in) W × 133 mm (5.24 in) H × 141.2 mm (5.56 in) D mm, 500 g (17.6 oz)
Connection Plate ×1, Operating Manual ×1
8423 Options in Detail
Model : MEMORY HiLOGGER 8423
Model No. (Order Code) (Note)
(main unit only)
8423
Note: 8423 cannot operate alone. You must install one or more optional input modules in the unit.
Thermocouples are not provided by HIOKI, and must be purchased from a separate vendor.
Input/output modules
VOLTAGE/TEMP UNIT UNIVERSAL UNIT 8949
15-channels, Voltage, Thermocouple,
8948
Input options
15-channles, Voltage,
Thermocouple input
Resistance temperature sensor,
Humidity measurement
HUMIDITY SENSOR
9701
Thermocouple
1-channel, for Universal unit
8949
DIGITAL/PULSE UNIT 8996
15-channels, ON/OFF logic signal,
Totalized pulses (integrated or
instantaneous), Rotation count
15-channels, Opencollector output
*For reference only.
Please purchase
locally.
PC CARD 1G 9729
1 GB capacity
PC CARD 512M 9728
512 MB capacity
AC ADAPTER
9418-15
100 to 240V AC
*LOGGER UTILITY is a bundled software
For the PC
Use only PC Cards sold by HIOKI.
Compatibility and performance are not
guaranteed for PC cards made by other
manufacturers. You may be unable to
read from or save data to such cards.
*9418-15 is supplied as standard
Other options
Storage media
* PC CARD 2G 9830 : cannot use
with the 8423 * Supplied with PC
Card adapter
PC Card Precaution
Example: Connect up 8 measurement modules for a
120-channel system
ALARM UNIT 8997
CONNECTION CABLE
9683
LOGGER UTILITY SF1000
The control of the measurement of
loggers, real-time data collectiont
For synchronization, cable
length 1.5 m (4.92 ft)
LAN CABLE 9642
Straight Ethernet cable, supplied
with straight to cross conversion
adapter, 5 m (16.41 ft) length
38.5 mm
(1.52 in)
67 mm
(2.64 in)
380.3 mm (14.97 in)
141.2 mm (5.56 in)
133 mm (5.24 in)
133 mm (5.24 in)
125 mm (4.92 in)
141.2 mm (5.56 in)
133 mm (5.24 in)
133 mm (5.24 in)
141.2 mm (5.56 in)
■ Appearance/Dimension Illustration
38.5 mm
(1.52 in)
Model 8948
Model 8996
Model 8997
38.5 mm
(1.52 in)
Model 8949
Model 8423
■ Configuration Examples
Input unit × 1
Input unit × 2
Input unit × 4
15-channels Isolated 30-channels Isolated 60-channels Isolated
Model 8423 × 1
Model 8948 × 1
Model 8423 × 1
Model 8948 × 2
Model 8423 × 1
Model 8948 × 4
Input unit × 8
120-channels Isolated
Model 8423 × 1
Model 8948 × 8
(Input unit × 8) system × 2
240-channels Isolated
(Input unit × 8) system × 4
480-channels Isolated
(Input unit × 8) system × 5
600-channels Isolated
Model 8423 × 2
Model 8948 × 16
Synchronization cable 9683 × 2
Model 8423 × 4
Model 8948 × 32
Synchronization cable 9683 × 4
Model 8423 × 5
Model 8948 × 40
Synchronization cable 9683 × 5
Note: Company names and Product names appearing in this catalog are trademarks or registered trademarks of various companies.
HIOKI (Shanghai) SALES & TRADING CO., LTD.
TEL +86-21-63910090 FAX +86-21-63910360
http://www.hioki.cn / E-mail: info@hioki.com.cn
DISTRIBUTED BY
HIOKI SINGAPORE PTE. LTD.
TEL +65-6634-7677 FAX +65-6634-7477
HEADQUARTERS
E-mail: info-sg@hioki.com.sg
81 Koizumi, Ueda, Nagano, 386-1192, Japan
TEL +81-268-28-0562 FAX +81-268-28-0568
HIOKI KOREA CO., LTD.
http://www.hioki.com / E-mail: os-com@hioki.co.jp TEL +82-2-2183-8847 FAX +82-2-2183-3360
E-mail: info-kr@hioki.co.jp
HIOKI USA CORPORATION
TEL +1-609-409-9109 FAX +1-609-409-9108
http://www.hiokiusa.com / E-mail: hioki@hiokiusa.com
All information correct as of June 23, 2016. All specifications are subject to change without notice.
8423E12-66E
Printed in Japan