New testo 325
M
Your Introduction To Professional Flue Gas Analysis
TÜV approval: O2, CO2, °C
Vol.% O2
Vol.% CO2
Backlit display
ppm CO
°C
h Pa
Two P inputs for draught and
additional pressure measurement
P
P
E ffg
Effn
λ
CO measuring range up to
4000 ppm
Ambient CO measurement to
protect from CO poisoning
Your Gain In Service.
testo 325 M, the successor to testo 325, tried and proven thousands of times over, meets
all the requirements in the adjustment of burners by heating installation technicians and is
also additionally used by gas fitters in service work on account of its P measurement
function.
Summary of analyser functions
• CO measuring range up to 4000 ppm
• Two P pressure measuring ranges
P1: 40 hPa;
P2: 200 hPa
• Ambient CO measurement buffer
• Measuring cell can be easily changed by
customer
• Automatic CO switch-off limits to protect
measuring cells
• User defined display sequence
• Comprehensive accessories for
every measuring task
• Fast and easy printout on
Testo printer
Flue gas input
Ambient air
temperature
T
Flue gas
temperature
2 P inputs
The new Testo system case, particularly robust;
made of ABS plastic with soft foam insert, without
contents
The New testo 325
New practical housing
M
The Extension case system specially for technicians for easy
transport of your work tools.
The robust extension case
system for easy transport
For example, the allpurpose system case,
without contents
...or the tool system case,
without contents
Technical data
Integrated condensate trap
Backlit display
Temperature measurement:
Measuring range:
Accuracy:
Resolution:
Sensor:
Draught/pressure measurement:
P1:
Accuracy:
Resolution:
P2:
Accuracy:
-40 to +600 °C
±0.5 °C (0 to +99.9 °C)
±0.5 % of reading (from +100°C)
0.1 °C
Thermcouple Type K (NiCr-Ni)
acc. to DIN IEC 584, Part 2, Class 1
±40 hPa
±0.03 hPa (±3 hPa)
±1.5 % of reading(±3.1 to ±40 hPa)
0.01 hPa
±200 hPa
±0.5 hPa (±50 hPa)
±1.5 % of reading (±50.1 to ±200 hPa)
Large buttons
Resolution:
Efficiency
Measuring range:
Resolution:
0.1 hPa
0 to 120 %
0.1 %
Battery operation
O2 measurement:
Measuring range:
Accuracy:
Resolution:
Response time t90:
CO2 measurement:
Display range:
Accuracy:
Resolution:
Measurement:
Response time t90:
CO measurement:
Measuring range:
Accuracy:
(H2 level < 10%):
Resolution:
Response time t90:
General technical data:
Weight:
Dimensions:
Transport/storage temperature:
Ambient temperature:
Power supply:
Battery life:
Warranty:
0 to 21 vol.%
±0.2 vol.%
0.1 %
Approx. 40 seconds
0 to CO2max
±0,2 Vol.%
0.01 Vol.%
Dig. calculation based on O2
Approx. 40 seconds
0 to 4000 ppm
± 20 ppm (up to 400 ppm)
± 5 % of reading
1 ppm
Max. 60, typically approx. 40 s
Approx. 500 g
216x68x47
-20 to +50 °C
+4 to +45 °C
Mains unit, batteries or rech. batt.
> 4h
2 years
(excluding wear parts/meas. cells)
CO measuring cell – 1 year
O2 measuring cell – 1,5 years
Ordering by fax
Recommended kit:
Your introduction to professional
measurement technology
Qty.
Part
no.
Simply copy this page, enter quantities and fax to us.
The kit case for your introduction to professional flue gas measurement
The complete kit in the case includes the following:
0563.0332
testo 325 M flue gas analyser incl. battery, calibration protocol and Instruction manual
0632.0322
Compact flue gas probe, 180 mm immersion depth, incl. probe stop, 1.5 m hose.
0600.9544
230 V mains unit, for mains operation and battery recharging
0554.1084
Ambient air probe, 190 mm immersion depth, with probe stop, magnetic clip, Tmax +100 °C
0600.9788
System case for analyser, probe and accessories
0516.0326
SoftCase made of flexible plastic incl. magnetic plate and carrying strap
0516.2572
Printer for cordless printout via infrared, incl. batteries and thermal paper
0554.0545
If you need soot measurement additionally, please order testo’s smoke tester.
Ordering data for testo 325 M
Qty.
Analyser
Part no.
testo 325 M flue gas analyser incl.battery, cal. protocol and Instruction manual
0632.0322
Accessories
Printer for cordless printout via infrared, incl. batteries and thermal paper
0554.0545
Spare thermal paper for printer (6 rolls)
0554.0569
Spare thermal paper for printer (6 rolls), long-term legible measurement data documentation for up to 10 years
0554.0568
230 V mains unit, for mains operation and to recharge batteries
0554.1084
SoftCase made of flexible plastic incl. magnetic plate and carrying strap, protects analyser from dirt and impact
0516.2572
Hose connection set for separate gas pressure measurement, incl. silicone hose and connection adapter
0554.0315
Hose connection set, single, diameter: 4/6 mm, 50 cm long, without connection adapter
0554.0449
Spare particle filter (10 off)
0554.0040
Smoke tester incl. oil, soot sheet, to measure soot in flue gas
0554.0307
Filter paper to determine smoke number, 40 measurement strips for approx. 200 measurements
0554.0308
System case for analyser, probe and accessories, can be easily connected to other cases in seconds
0516.0326
Tool system case with sections for tools, without contents, can be easily connected to system case 0516.0326
0516.0329
All-purpose system case with plastic sections, without contents, can be easily connected to system case 0516.0326
0516.0328
Universal system case without sections, can be easily connected to system case 0516.0326
0516.0331
Plastic transport case
0516.3250
Flue gas probes
Compact flue gas probe, 180 mm immersion depth, incl. probe stop, up to +500°C, NiCr-Ni thermocouple, 1.5m hose
0600.9544
Compact flue gas probe, 300 mm immersion depth, incl. probe stop, up to +500°C, NiCr-Ni thermocouple, 1.5m hose
0600.9542
TÜV approved flue gas probe corresponding to latest instrument test guidelines, 180 mm immersion depth, 2.2 m hose
0600.9543
TÜV approved flue gas probe corresponding to latest instrument test guidelines, 300 mm immersion depth, 2.2 m hose
0600.9546
Flexible flue gas probe, 300 mm immersion depth
0600.9440
High temperature probe up to +1000 °C, 300 mm immersion depth
0600.8540
Temperature probes
Fast-action surface probe with sprung thermocouple strip, also for rough surfaces, measuring range short-term to +500 °C
0602.0392
Ambient air probe, 190 mm immersion depth, with probe stop, magnetic clip, Tmax +100 °C for ambient air temperature measurement in
systems dependent on/independent of ambient air
0600.9788
Mini ambient air probe, 60 mm immersion depth, with probe stop, magnetic clip, Tmax +100 °C, for dual wall clearance temperature measurement in
systems dependent on ambient air
0600.9798
Subject to change without notice.
Sender
Name
Company
Department
Address
Tel/Fax no.
Date, Signature
0981.2954/hd/AC/08.2002
Fax to:
New testo 325
XL
Professional Flue Gas Analysis and Monitoring
TÜV approval: O2, CO2, °C + CO
Vol.% O
Vol.%
Backlit display
ppm CO
°C
h Pa
Two P inputs for draught and
additional pressure measurement
P
P
Effg
Effn
λ
TÜV approved CO measurement
via H2 compensation
Storable O2 additional air
measurement in dependent systems
Your Service Advantage.
The TÜV approved testo 325 XL is used by heating technicians and engineers for
monitoring measurements. It meets all the requirements in the adjustment, maintenance and
servicing of burners.
Summary of analyser functions
• CO measuring range up to 4000 ppm
• CO TÜV approval (H2 compensated)
• Two P pressure measuring ranges
P1: 40 hPa;
P2: 200 hPa
• Battery recharging in instrument, batteries can be used
• Up to 6 fuels can be selected
• Ambient CO measurement buffer
• O2 additional air measurement buffer
• Memory for 20 measurement blocks
• Measuring cells can be easily changed by customer
• Automatic CO switch-off limits to protect
measuring cells
• User defined display sequence
• Comprehensive accessories for
every measuring task
• Fast and easy printout on
Testo printer
Flue gas input
Ambient air
temperature
T
Flue gas
temperature
2 P inputs
The new Testo system case, particularly robust;
made of ABS plastic with soft foam insert, without
contents
The New testo 325
New practical housing
XL
The Extension case system specially for technicians for easy
transport of your work tools.
The robust extension case
system for easy transport
For example, the allpurpose system case,
without contents
...or the tool system case,
without contents
Technical data
Integrated condensate trap
Backlit display
Temperature measurement:
Measuring range:
Accuracy:
Resolution:
Sensor:
Draught / pressure measurement:
P1:
Accuracy:
Resolution:
P2:
Accuracy:
Large buttons
Resolution:
Efficiency
Measuring range:
Resolution:
-40 to +1000 °C
±0.5 °C (0 to +99.9 °C)
±0.5 % of reading (from +100°C)
0.1 °C
Thermcouple Type K (NiCr-Ni)
acc. to DIN IEC 584, Part 2, Class 1
±40 hPa
±0.03 hPa (±3 hPa)
±1.5 % of reading (±3.1 to ±40 hPa)
0.01 hPa
±200hPa
±0.5 hPa (±50 hPa)
±1.5 % of reading (±50.1 to ±200 hPa)
0.1 hPa
0 to 120 %
0.1 %
Memory for 20 memory blocks
Battery recharging + battery operation
O2 measurement:
Measuring range:
Accuracy:
Resolution:
Response time t90:
CO2 measurement:
Display range:
Accuracy:
Resolution:
Measurement:
Response time t90:
CO measurement:
Measuring range:
Accuracy:
(H2 compensated)
Resolution:
Response time t90:
General technical data:
Weight:
Dimensions:
Transport/storage temperature:
Ambient temperature:
Power supply:
Battery life:
Warranty:
0 to 21 vol.%
±0.2 vol.%
0.1 %
Approx. 40 seconds
0 to CO2max
±0.2 vol.%
0.01 vol.%
Dig. calculation based on O2
Approx. 40 seconds
0 to 4000 ppm
±20 ppm (up to 400 ppm)
±5 % of reading (401 to 4000)
1 ppm
Max. 60, typically approx. 40 s
Approx. 500 g
216 x 68 x 47
-20 to +50 °C
+4 to +45 °C
Mains unit, batteries or rech. batt.
> 4h
2 years
(excluding wear parts/meas. cells)
CO measuring cell – 1 year
O2 measuring cell – 1 years
Ordering by fax
Part
no.
Simply copy this page, enter quantities and fax to us.
Recommended kit:
Qty.
Your introduction to reference
measurement technology
The kit case for your introduction to professional flue gas measurement
The complete kit in the case includes the following:
0563.0334
testo 325 XL flue gas analyser incl. rechargeable battery, calibration protocol and Instruction manual
0632.0321
TÜV approved flue gas probe, 180 mm immersion depth, incl. probe stop, 2.2 m hose.
0600.9543
230 V mains unit, for mains operation and battery recharging
0554.1084
Ambient air probe, 190 mm immersion depth, with probe stop, magnetic clip, Tmax +100 °C
0600.9788
System case for analyser, probe and accessories
0516.0326
SoftCase made of flexible plastic incl. magnetic plate and carrying strap
0516.2572
Smoke tester incl. oil, soot sheets, to measure soot in flue gas
0554.0307
If you need printouts additionally, please order the testo printer !
Ordering data for testo 325 XL
Qty.
Analyser
Part no.
testo 325 XL flue gas analyser incl. rechargeable battery, cal. protocol and Instruction manual
0632.0321
Accessories
Printer for cordless printout via infrared, incl. batteries and thermal paper
0554.0545
Spare thermal paper for printer (6 rolls)
0554.0569
Spare thermal paper for printer (6 rolls), long-term legible measurement data documentation for up to 10 years
0554.0568
230 V mains unit, for mains operation and to recharge batteries
0554.1084
SoftCase made of flexible plastic incl. magnetic plate and carrying strap, protects analyser from dirt and impact
0516.2572
Hose connection set for separate gas pressure measurement, incl. silicone hose and connection adapter
0554.0315
Pressure connection hose, single, diameter: 4/6 mm, 50 cm long, without connection adapter
0554.0449
Spare particle filter (10 off)
0554.0040
Smoke tester incl. oil, soot sheet, to measure soot in flue gas
0554.0307
Filter paper to determine smoke number, 40 measurement strips for approx. 200 measurements
0554.0308
System case for analyser, probe and accessories, can be easily connected to other cases in seconds
0516.0326
Tool system case with sections for tools, without contents, can be easily connected to system case 0516.0326
0516.0329
All-purpose system case with plastic sections, without contents, can be easily connected to system case 0516.0326
0516.0328
Universal system case without sections, can be easily connected to system case 0516.0326
0516.0331
Plastic transport case
0516.3250
Flue gas probes
Compact flue gas probe, 180 mm immersion depth, incl. probe stop, up to +500°C, NiCr-Ni thermocouple, 1.5m hose
0600.9544
Compact flue gas probe, 300 mm immersion depth, incl. probe stop, up to +500°C, NiCr-Ni thermocouple, 1.5m hose
0600.9542
TÜV approved flue gas probe corresponding to latest instrument test guidelines, 180 mm immersion depth, 2.2 m hose
0600.9543
TÜV approved flue gas probe corresponding to latest instrument test guidelines, 300 mm immersion depth, 2.2 m hose
0600.9546
Flexible flue gas probe, 300 mm immersion depth
0600.9440
High temperature probe up to +1000 °C, 300 mm immersion depth
0600.8540
Dual wall clearance probe, sickle-shaped O2 additional air probe to measure O2 level in dual wall of systems independent of ambient air
0632.1244
Temperature probes
Fast-action surface probe with sprung thermocouple strip, also for rough surfaces, measuring range short-term to +500 °C
0602.0392
Ambient air probe, 190 mm immersion depth, with probe stop, magnetic clip, Tmax +100 °C for ambient air temperature measurement in
systems dependent on/independent of ambient air
0600.9788
Mini ambient air probe, 60 mm immersion depth, with probe stop, magnetic clip, Tmax +100 °C, for dual wall clearance temperature measurement in
systems dependent on ambient air
0600.9798
Subject to change without notice.
Fax to:
Sender
Name
Company
Department
Address
Tel/Fax no.
Date, Signature
Committing to the future
testo 330-1
testo 330-1 LL
Cutting Edge Technology for Heating
Manufacturers and Fitters!
NEW!
More assurance and more convenience thanks to an extended sensor life
LL
°C
4 years warranty
on instrument and probe
hPa
O2
CO
NO
∆P
EN 50379-2 and 1. BImSchV
TÜV tested for O2, °C, hPa
TÜV By RgG 250 for testo 330-1 LL
TÜV By RgG 236 for testo 330-1
Highest longevity with even more assurance!
Revolutionary: Testo launches a new generation of measuring cells!
Reduced after-costs
More assurance!
Thanks to the extended sensor
(testo 330-1 LL incl. O2 and CO
Thanks to the longer life of the
life of up to 6 years for O2 and
measurement cells and probe).
complete instrument testo 330-1
CO in the testo 330-1 LL, the
LL, downtime - e.g. due to
after-costs for the user are greatly
exhausted sensors - is drastically
reduced. At least one change of
reduced. The instrument stays
O2 and CO cell can be omitted
ready for use and free from
during the typical phase of usage
malfunctions for longer
of the instrument. In addition to
this, Testo gives a warranty of 4
years on the complete instrument
Exception: Wearing parts such as filter,
thermocouple (12 months), NO/NOlow cell (24
months)
hPa
The right accessories for every application.
The probes
The robust probe handle which
and dual wall clearance probe
which can be attached to the
fits comfortably in your hand
round off our wide range of flue
analyser also has a user-defined
facilitates easy positioning. The
gas probes. The ambient CO
alarm limit. A warning signal is
probe has a quick connection
probe as well as the CO2 probe
triggered if the limit is exceeded.
function ensuring that all gas
are recognised automatically by
The probe is also recognised
paths are connected correctly.
the analyser and the
automatically and the results are
The particle filter accommodated
measurement results immediately
shown in a trend graph in their
in the handle blocks out particles
appear in their own graphically
own menu.
efficiently. Additionally, the probes
supported menu. Furthermore,
require little or no maintenance
the user-defined alarm limit is
and are easy to clean. Different
coupled with an audible signal.
lengths and diameters ensure a
Measurements can be carried out
high flexibility level for all
in their entirety simultaneously
applications. When changing the
with flue gas measurement.
probe, the probe shaft is simply
The gas leak detection probe
attached to the probe handle and
clicks into place. The multi-hole
User-friendly
The new, highly robust fast probe
connection for all gas paths
eliminates any confusion. The
single cable cannot be bent, is
indestructible and saves space.
The menu-driven operational
procedures - in compliance with
standards - help you to avoid
errors and save you time. The
very large display is a perfect
additional aid.
NO
∆P
O2
°C
CO
The heart of the new testo 330-1 LL
With the development of the
protection of the anode material,
new longlife sensors, Testo have
and a lead-free metal alloy. This
once again proven their
makes it not only longer living
innovative spirit.
but also evironmentally more
friendly.
The O2 Longlife sensor
distinguishes itself from the
An additional feature of the new
standard sensor through its even
CO Longlife sensor is also the
more robust design, an
exchangeable NO filter.
improved diffusion barrier for the
At a glance
Testo has put together a special set for heating fitters and technicians. Of course, it
is always possible to add to and supplement the set by selecting from the wide
range of accessories available.
Illustration not
obligatorily identical
The new Longlife set
Our classic
· Exhaust gas analyzer testo 330-1 LL incl. rech. battery and
· testo 330-1 flue gas analyser incl. rech. battery and calibration
calibration protocol
protocol
· Mains unit 100-240 V for mains operation for charging the battery in
the instrument
· 100-240 V mains unit for mains operation or battery recharging in
analyser
· Modular exhaust gas probe, immersion depth 300 mm, Ø 8 mm
· Modular flue gas probe, immersion depth: 300 mm, ø 8 mm
· Combustion air temperature probe, immersion depth 190 mm
· Combustion air temperature probe, immersion depth: 190 mm
· Hose connection set for separate gas pressure measurement
· Hose connection set for separate gas pressure measurement
· Report printer IRDA
· IRDA printer
· Flexible probe shaft length 300 mm, Ø 10 mm, Tmax 500 °C
· Basic system case for instrument, probes and accessories
· Basic system case for instrument, probes and accessories
Part no.
0563 3324 70
Part no.
0563 3314 70
Always a step ahead
testo 330-1 at a glance
Instrument diagnosis
Automatic menu selection
Additional features
• Rechargeable battery status display
• Recognition of connected probe
• TÜV approved in compliance with EN
• Sensor status display
50379 Part 3
• Monitoring of condensate trap level with
“FULL” message
• Display of error status incl. description and
diagnosis
• Display of last maintenance
• Display of instrument temperature
• Operating hour counter
• Interface to Testo Software with analysis
Memory management
• 200 measurement data records incl.
systems number
• Several measurement data records per
system number are possible
• Read in using barcode reader
• IRDA interface for data transfer to
PDA/Notebook
Additional measurement tasks
• USB interface for data transfer to PC
• ∆T measurement, flow/return
• ZIV driver for all standard branch software
• Ambient CO measurement
• Ambient CO2 measurement
• Detection of gas leaks with gas leak
detection probe
• ∆P measurement for gas pressure
• Gas/oil flow rate determination
and graphics functions and online
measurement
• Calculated parameter: Flue gas dewpoint
• Graphics display
• Integrated impact protection with magnet,
recessed display
• Protection class: IP40
• Rechargeable battery lifetime > 6 h with
pump operating
• Battery rechargeable separately and in
instrument
• Adjustable alarm thresholds for ambient
CO/CO2 measurement
Technical data
Temperature
Meas. range
Accuracy
Resolution
Draught measurement
Pressure measurement
O2 measurement
Meas. range
Accuracy
Resolution
Meas. range
Accuracy
Resolution
Meas. range
Resolution
Accuracy
Adjustment time t90
-40 to +1200 °C
±0,5 °C (0.0 to +100.0 °C)
±0,5 % of mv (remaining range)
0.1 °C (-40 to 999,9 °C)
1 °C (remaining range)
-9.99 to +40 hPa
±0.02 hPa or ±5 % of mv (-0.50 to +0.60 hPa)
±0.03 hPa (+0.61 to +3.00 hPa)
±1.5 % of mv (+3.01 to +40.00 hPa)
0.01 hPa
0 to 200 hPa
±0.5 hPa (0.0 to 50.0 hPa)
±1 % of mv (50.1 to 100.0 hPa)
±1.5 % of mv (remaining range)
0.1 hPa
0 to 21 Vol. %
0.1 Vol. %
±0.2 Vol. %
< 20 s
CO measurement
(without H2
compensation)
Meas. range
Resolution
Accuracy
Adjustment time t90
0 to 4000 ppm
1 ppm
±20 ppm (0 to 400 ppm)
±5% of mv (401 to 1000 ppm)
±10% of mv (1001 to 4000 ppm)
< 60 s
Efficiency
(ETA)
Meas. range
Resolution
0 to 120 %
0.1 %
Flue gas loss
Meas. range
Resolution
0 to 99.9 %
0.1 %
CO2 measurement
Display range
Resolution
Accuracy
Measurement
Adjustment time t90
Meas. range
Resolution
Accuracy
0 to CO2 max
0.1 Vol. %
±0.2 Vol. %
Digital calculation from O2
< 40 s
0 to 300 ppm
0.1 ppm
±2 ppm (0.0 to 40.0 ppm)
±5% of mv (remaining range)
< 30 s
Option: NOlow
measurement
Adjustment time t90
Option: NO
measurement
Meas. range
Resolution
Accuracy
Ambient CO
measurement (with CO
probe)
Meas. range
Resolution
Accuracy
Adjustment time t90
Gas leak measurement
for combustible gases
(with gas leak detection
probe)
Adjustment time t90
Range of indication
Signal
Adjustment time t90
0 to 3000 ppm
1 ppm
±5 ppm (0 to 100 ppm)
±5% of mv (101 to 2000 ppm)
±10% of mv (2001 to 3000 ppm)
< 30 s
0 to 500 ppm
1 ppm
±5 ppm (0 to 100 ppm)
±5% of mv (>100 ppm)
Approx. 35 s
0 ... 10,000 ppm CH4 / C3H8
Optical display (LED)
audible display via buzzer
<2s
Ambient CO2
measurement (with
ambient CO2 probe)
Meas. range
General Technical Data
Memory
Weight
Dimensions
Storage temp.
Oper. temp.
Display
Power supply
200 Sites
600 g (without rechargeable battery)
270 x 90 x 65 mm
-20 to +50 °C
-5 to +45 °C
Graphics display: 160 x 240 pixel
Rech. block: 3.7V/2.2Ah
Power: 6V/1.2A
Warranty
testo 330-1
Instrument/probe/cells (O2, CO, NO, NOlow)
Thermocouple and rechargeable battery
Instrument/probe/cells (O2, CO)
NO, NOlow cell
Thermocouple and rechargeable battery
Accuracy
Adjustment time t90
testo 330-1 LL
*
0 to 1 Vol. %
0 to 10000 ppm
±(50 ppm ±2 % of mv) (0 to 5000 ppm)
Approx. 35 s
24 months
12 months
48 months
24 months
12 months
*the larger value applies
Fax Order Form
Qty.
Qty.
Qty.
Part no.
0632 3301
Basic system case for analyser, probes and accessories
0516 3330
0632 3304
Basic system case with two levels for analyser, probes and additional
accessories
0516 3331
Option: Fine draught measurement, Resolution 0.1 Pa, measurement range to
100 Pa (instead of the standard draught measurement)
0440 3921
Product system case with small plastic boxes without contents, can be
interlocked to basic system case
0516 0328
Option: NO meas. cell, meas. range 0 to 3000 ppm, 1 ppm resolution
0440 3922
0440 3931
Tools system case with tools section without content, can be interlocked to
basic system case
0516 0329
Option: NOlow meas. cell, meas. range 0 to 300 ppm, 0.1 ppm resolution
Set testo 330-1 LL: Pro analysis for heating manufacturers and fitters
Part no.
· Exhaust gas analyzer testo 330-1 LL incl. rech. battery and calibration
protocol
· Mains unit 100-240 V for mains operation for charging the battery in the
instrument
· Modular exhaust gas probe, immersion depth 300 mm, Ø 8 mm
· Combustion air temperature probe, immersion depth 190 mm
· Hose connection set for separate gas pressure measurement
· Report printer IRDA
· Flexible probe shaft length 300 mm, Ø 10 mm, Tmax 500 °C
· Basic system case for instrument, probes and accessories
0563 3324 70
Qty.
Part no.
Universal system case w/o pockets, can be interlocked to basic system case
0516 0331
Measurement case (leather) with drawers for instruments and accessories
0516 0303
Probes
Part no.
Modular flue gas probes, available in 2 lengths, incl. positioning cone,
NiCr-Ni thermocouple, 2.2 m hose and particle filter
testo 330-1 Set: Pro analysis for heating technicians and fitters
Part no.
Flue gas probe 180 mm long, Ø 8 mm, Tmax 500 °C
0600 9760
· testo 330-1 flue gas analyser incl. rech. battery and calibration protocol
· 100-240 V mains unit for mains operation or battery recharging in analyser
· Modular flue gas probe, immersion depth: 300 mm, ø 8 mm
· Combustion air temperature probe, immersion depth: 190 mm
· Hose connection set for separate gas pressure measurement
· IRDA printer
· Basic system case for instrument, probes and accessories
0563 3314 70
Flue gas probe 300 mm long, Ø 8 mm, Tmax 500 °C
0600 9761
Flue gas probe 180 mm long, Ø 6 mm, Tmax 500 °C
0600 9762
Flue gas probe 300 mm long, Ø 6 mm, Tmax 500 °C
0600 9763
Probe shaft 180 mm long, Ø 8 mm, Tmax 500 °C
0554 9760
Spare measurement cells
Part no.
Probe shaft 180 mm long, Ø 6 mm, Tmax 500 °C
0554 9762
Spare O2 measurement cell for testo 330-1
0390 0092
Probe shaft 300 mm long, Ø 8 mm, Tmax 500 °C
0554 9761
Spare CO measurement cell (without H2 compensation) for testo 330-1
0390 0095
Probe shaft 300 mm long, Ø 6 mm, Tmax 500 °C
0554 9763
Spare NOlow measurement cell 0 to 300 ppm for testo 330-1/-1 LL
0390 0094
Probe shaft, 335 mm long, with probe stop, Ø 8 mm, Tmax 1000 °C
0554 8764
Spare NO measuring cell
0390 0074
Probe shaft, 700 mm long, with probe stop, Ø 8 mm, Tmax 1000 °C
0554 8765
O2 measurement cell for testo 330-1 LL
0390 0061
Flexible probe shaft 330 mm long, Ø 10 mm, Tmax 500 °C
0554 9764
CO measurement cell for testo 330-1 LL
0554 5762
Probe accessories
0390 0110
Multi-hole probe shaft, 300 mm long, Ø 8 mm, for mean CO calculation
Upgrade: NO meas. cell, meas. range 0 to 3000 ppm, 1 ppm resolution for
testo 330-1/-1 LL
0554 3922
Multi-hole probe shaft, 180 mm long, Ø 8 mm, for mean CO calculation
0554 5763
Upgrade: NOlow meas. cell, meas. range 0 to 300 ppm, 0.1 ppm resolution for
testo 330-1/-1 LL
Hose extension, 2.8 m, extension cable for probe and analyser
0554 1202
0554 3931
6 mm probe stop, PTFE, with spring clamp and handle, Tmax 200 °C
0554 3327
Upgrade testo 330-1 to testo 330-1 LL (only by Testo service)
Part no.
8 mm probe stop, PTFE, with spring clamp and handle, Tmax 200 °C
0554 3328
8 mm probe stop, steel, with spring clamp and handle, Tmax 500 °C
0554 3330
6 mm, probe stop, steel, with spring clamp and handle, Tmax 500 °C
0554 3329
Upgrade testo 330-1 to testo 330-1 LL consisting of:
Qty.
Cases
Option code upgrade testo 330-1 to testo 330-1 LL
Upgrade O2 LL measurement cell
Upgrade CO LL measurement cell
0450 1100
0554 3938
0554 3936
Accessories
Part no.
100-240 V mains unit for mains operation or batt. recharging in analyser
0554 1086
Spare rech. batt. w/ charging station
0554 1087
Additional probes
Dual wall clearance probe
0632 1260
Gas leak probe
0632 3330
Testo printer with wireless IRDA and infrared interface, 1 roll of thermal paper
0554 0547
and 4 AA batteries
Ambient CO probe
0632 1247
Testo IrDA printer with bidirectional IrDA interface and rech. battery, for
communication testo 330 and PDA
0554 0548
Ambient CO2 probe
0632 1240
Spare thermal paper for printer (6 rolls), permanent ink
0554 0568
Connection cable
0430 0143
Barcode reader to read in customer number on site
0554 0461
Barcode labels, self-adhesive (1200 off)
0554 0411
Combustion air temperature probe, immersion depth 300 mm
0600 9791
Adhesive pockets (50 off) for printout, paper barcode labels...
0554 0116
Combustion air temperature probe, immersion depth 190 mm
0600 9787
Instrument cleaner (100 ml)
0554 1207
Combustion air temperature probe, immersion depth 60 mm
0600 9797
Smoke tester with oil, soot sheet, for measuring soot in flue gas
0554 0307
Hose connection set for separate gas pressure measurement
0554 1203
Differential temperature set consisting of 2 pipe clamp probes and adapter
0554 1204
Spare particle filter (10 off)
0554 3385
Full version EasyHeat (PC software)
0554 3332
Upgrade EasyHeat Mobile for PDA
0554 1211
Full version EasyHeat + EasyHeat Mobile (for PC und PDA)
0554 1210
USB connection cable, instrument to PC
0449 0047
ISO calibration certificate/flue gas
0520 0003
Combustion air temperature probe
Additional temperature probes
Mini ambient air probe
0600 3692
Pipe clamp probe
0600 4593
Quick-action surface probe
0604 0194
Connection cable
0430 0143
Sender
Name
Address
Company
City
Department
Date, Signature
Subject to change without notice.
Qty.
Qty.
Instrument/options
testo 330-1 flue gas analyser incl. rechargeable battery and calibration
protocol
testo 330-1 LL exhaust gas anlyzer with Longlife cells, incl. rechargeable
battery and calibration protocol
0981 5634/msp/AC/A/08.2006
Qty.
Committing to the future
testo 330-2
testo 330-2 LL
Cutting Edge Technology for Customer
Service and Maintenance Technicians
NEW!
More robust, longer-living, faster and safer, thanks to an extended sensor life
LL
°C
4 years warranty
on instrument and probe
hPa
O2
CO/H2
NO
ΔP
EN 50379-2 and 1. BImSchV/KÜO
TÜV tested for O2, °C, hPa
TÜV By RgG 251 for testo 330-2 LL
TÜV By RgG 237 for testo 330-2
Highest longevity with even more
assurance!
Revolutionary: Testo launches a new generation of measuring cells!
Reduced after-costs!
Thanks to the extended sensor life
330-2 LL incl. O2 and CO
of up to 6 years for O2 and CO in
measurement cells and probe).
the testo 330-2 LL, the after-costs
for the user are greatly reduced. At
least one change of O2 and CO
cell can be omitted during the
typical phase of usage of the
instrument. In addition to this,
Testo gives a warranty of 4 years
on the complete instrument (testo
More assurance!
Thanks to the longer life of the
complete instrument testo 330-2
LL, downtime - e.g. due to
exhausted sensors - is drastically
reduced. The instrument stays
ready for use and free from
malfunctions for longer.
Exception: Wearing parts such as filter,
thermocouple (12 months), NO/NOlow cell
(24 months)
The heart of the new testo 330-2 LL
With the development of the new
standard sensor through its even
An additional feature of the new
longlife sensors, Testo have once
more robust design, an improved
CO Longlife sensor is also the
again proven their innovative
diffusion barrier for the protection
exchangeable NO filter.
spirit.
of the anode material, and a
lead-free metal alloy. This makes
The O2 Longlife sensor
it not only longer living but also
distinguishes itself from the
evironmentally more friendly.
Even more
convenience
Maintenance and monitoring
work on the most diverse
systems are now even faster and
more effective, since the zeroing
of the gas and draught sensors
can be carried out directly in the
flue.
The following time-consuming
work steps, which up to now
were necessary for zeroing, are
now superfluous:
• Probe removal
• Sensor zeroing
• Re-positioning of the probe in
the flue
Instead, the probe can remain in
the flue for each zeroing, whether
for gas or draught zeroing, and
be conveniently zeroed from the
testo 330.
The right accessories for every application
The probes
The robust probe handle which
and dual wall clearance probe
analyser also has a user-defined
fits comfortably in your hand
round off our wide range of flue
alarm limit. A warning signal is
facilitates easy positioning. The
gas probes. The ambient CO
triggered if the limit is exceeded.
probe has a quick connection
probe as well as the CO2 probe
The probe is also recognised
function ensuring that all gas
are recognised automatically by
automatically and the results are
paths are connected correctly.
the analyser and the
shown in a trend graph in their
The particle filter accommodated
measurement results immediately
own menu.
in the handle blocks out particles
appear in their own graphically
efficiently. Additionally, the probes
supported menu. Furthermore,
require little or no maintenance
the user-defined alarm limit is
and are easy to clean. Different
coupled with an audible signal.
lengths and diameters ensure a
Measurements can be carried out
high flexibility level for all
in their entirety simultaneously
applications. When changing the
with flue gas measurement.
probe, the probe shaft is simply
The gas leak detection probe
attached to the probe handle and
which can be attached to the
clicks into place. The multi-hole
User-friendly!
The new, highly robust fast probe
errors and save you time. The
connection for all gas paths
very large display is a perfect
eliminates any confusion. The
additional aid.
single cable cannot be bent, is
indestructible and saves space.
The menu-driven operational
procedures - in compliance with
standards - help you to avoid
At a glance
The sets for customer service and maintenance technicians
Testo has put together a special set for customer service and maintenance technicians. Of course, it is
always possible to add to and supplement the set by selecting from the wide range of accessories
available.
The new Longlife set
Our classic
· Exhaust gas analyzer testo 330-2 LL incl. rech.
battery and calibration protocol
· testo 330-2 flue gas analyser incl. rechargeable
battery and calibration protocol
· Mains unit 100-240 V for mains operation for
charging the battery in the instrument
· 100-240 V mains unit for mains operation or
battery recharging in instrument
· Modular exhaust gas probe, immersion depth
300 mm, Ø 8 mm
· Modular flue gas probe, immersion depth: 300
mm, ø 8 mm
· Combustion air temperature probe, immersion
depth 190 mm
· Combustion air temperature probe, 190 mm long
· Hose connection set for separate gas pressure
measurement
· Report printer IRDA
Illustration not obligatorily
identical
· Flexible probe shaft length 300 mm, Ø 10 mm,
Tmax 500 °C
· Basic system case for instrument, probes and
accessories
Part no.
0563 3325 70
· Hose connection set for separate gas pressure
measurement
· IRDA printer
· Smoke tester incl. oil, soot sheets and reference
scale
· Basic system case for analyser, probes and
accessories
Part no.
0563 3315 70
Always a step ahead
testo 330-2 / testo 330-2 LL at a glance
Instrument diagnosis
Automatic menu selection
Additional features
• Automatic instrument leak test
• Rechargeable battery status
display
• Sensor status display
• Monitoring of condensate trap
level with “FULL” message
• Display of pump capacity (l/min.)
• Display of error status incl.
description and diagnosis
• Display of last maintenance
• Display of instrument
temperature
• Operating hour counter
• Recognition of connected probe
• Interface to Testo Software with
analysis and graphics functions
and online measurement
• Interface to burners
• Calculated parameter: Flue gas
dewpoint
• Graphics display
• Integrated impact protection with
magnet, recessed display
• Protection class: IP40
• Rechargeable battery lifetime > 6
h with pump operating
• Battery rechargeable separately
and in instrument
• Adjustable alarm thresholds for
ambient CO/CO2 measurement
Initialisation of flue gas sensors
without removal of probes
• The probe can remain in the flue
during initialisation
• Draught zeroing without removal
of the probe (testo 330-2 LL
only). The probe can remain in
the flue during zeroing
Memory management
Additional measurement tasks
• ΔT measurement, flow/return
• Ambient CO measurement
• Ambient CO2 measurement
• Detection of gas leaks with gas
leak detection probe
• ΔP measurement for gas
pressure
• Gas/oil flow rate determination
• 200 meas. data records with
address and systems no.
• Several measurement data
records per system number are
possible
• Read in using barcode reader
• IRDA interface for data transfer
to PDA/Notebook
• USB interface for data transfer to
PC
• ZIV driver for all conventional
branch software
TÜV testing / EN norm
• Accuracy tested for O2, CO2,
CO, NO, NOlow, hPa
• TÜV By RgG 237 for testo 3302, TÜV By RgG 251 for testo
330-2 LL acc.to 1.
BImSchV/KÜO / EN 50379 Part
2 for O2, °C, hPa and CO
Technical Data
Temperature
Meas. range
Accuracy
Resolution
Draught measurement
Pressure measurement
O2 measurement
Meas. range
Accuracy
Resolution
Meas. range
Accuracy
Resolution
Meas. range
Resolution
Accuracy
Adjustment time t90
-40 to +1200 °C
±0,5 °C (0.0 to +100.0 °C)
±0,5 % of mv (remaining range)
0.1 °C (-40 to 999,9 °C)
1 °C (remaining range)
-9.99 to +40 hPa
±0.02 hPa or ±5 % of mv (-0.50 to +0.60 hPa)
±0.03 hPa (+0.61 to +3.00 hPa)
±1.5 % of mv (+3.01 to +40.00 hPa)
0.01 hPa
0 to 200 hPa
±0.5 hPa (0.0 to 50.0 hPa)
±1 % of mv (50.1 to 100.0 hPa)
±1.5 % of mv (remaining range)
0.1 hPa
0 to 21 Vol. %
0.1 Vol. %
±0.2 Vol. %
< 20 s
from 8000 ppm
Adjustment time t90
Display range
Resolution
0 to 8000 ppm
1 ppm
±10 ppm or ±10% of mv (0 to 200 ppm)
±20 ppm or ±5% of mv (201 to 2000 ppm)
±10% of mv (2001 to 8000 ppm)
< 40 s
8000 to 30000 ppm (automatic dilution)
500 ppm
Efficiency
(Eta)
Meas. range
Resolution
0 to 120 %
0.1 %
Flue gas loss
Meas. range
Resolution
0 to 99.9 %
0.1 %
CO2 measurement
Display range
Resolution
Accuracy
Determination
Adjustment time t90
Meas. range
Resolution
Accuracy
0 to CO2 max
0.1 Vol. %
±0.2 Vol. %
Digital calculation from O2
< 40 s
0 to 300 ppm
0.1 ppm
±2 ppm (0.0 to 40.0 ppm)
±5% of mv (remaining range)
< 30 s
CO measurement
(H2 compensated)
Option: NOlow
measurement
Meas. range
Resolution
Accuracy
Adjustment time t90
Option: NO
measurement
Meas. range
Resolution
Accuracy
Ambient CO
measurement (with CO
probe)
Meas. range
Resolution
Accuracy
Adjustment time t90
Gas leak measurement
for combustible gases
(with gas leak detection
probe)
Adjustment time t90
Display range
Signal
Adjustment time t90
0 to 3000 ppm
1 ppm
±5 ppm (0 to 100 ppm)
±5% of mv (101 to 2000 ppm)
±10% of mv (2001 to 3000 ppm)
< 30 s
0 to 500 ppm
1 ppm
±5 ppm (0 to 100 ppm)
±5% of mv (>100 ppm)
Approx. 35 s
0 to 10,000 ppm CH4 / C3H8
Optical display (LED)
audible display via buzzer
<2s
Ambient CO2
measurement (with
ambient CO2 probe)
Meas. range
General Technical Data
Memory
Weight
Dimensions
Storage temp.
Oper. temp.
Display
Power supply
200 Sites
600 g (without rechargeable battery)
270 x 90 x 65 mm
-20 to +50 °C
-5 to +45 °C
Graphics display: 160 x 240 pixel
Rech. block: 3.7V/2.2Ah
Power: 6V/1.2A
Warranty
testo 330-2
Instrument/probe/cells (O2, CO, NO, NOlow)
Thermocouple and rechargeable battery
Instrument/probe/cells (O2, CO)
NO, NOlow cell
Thermocouple and rechargeable battery
Accuracy
Adjustment time t90
testo 330-2 LL
0 to 1 Vol. %
0 to 10000 ppm
±(50 ppm ±2 % of mv) (0 to 5000 ppm)
Approx. 35 s
24 months
12 months
48 months
24 months
12 months
*the larger value applies
Fax Order Form
0632 3302
Basic system case for analyser, probes and accessories
0516 3330
0632 3305
Basic system case with two levels for analyser, probes and additional
accessories
0516 3331
0440 3921
Product system case with small plastic boxes without contents, can be
interlocked to basic system case
0516 0328
Tools system case with tools section without content, can be interlocked to
basic system case
0516 0329
Option: NO meas. cell, meas. range 0 to 3000 ppm, 1 ppm resolution
0440 3922
Option: NOlow meas. cell, meas. range 0 to 300 ppm, 0.1 ppm resolution
0440 3931
Set testo 330-2 LL: Pro analysis: Customer service/technicians
Best.-Nr.
· Exhaust gas analyzer testo 330-2 LL incl. rech. battery and calibration
protocol
· Mains unit 100-240 V for mains operation for charging the battery in the
instrument
· Modular exhaust gas probe, immersion depth 300 mm, Ø 8 mm
· Combustion air temperature probe, immersion depth 190 mm
· Hose connection set for separate gas pressure measurement
· Report printer IRDA
· Flexible probe shaft length 300 mm, Ø 10 mm, Tmax 500 °C
· Basic system case for instrument, probes and accessories
0563 3325 70
testo 330-2 Set: Pro flue gas analysis: Customer service/technicians
Best.-Nr.
Qty.
Qty.
0516 0331
Measurement case (leather) with drawers for instruments and accessories
0516 0303
Probes
Part no.
Flue gas probe 180 mm long, Ø 8 mm, Tmax 500 °C
0600 9760
Flue gas probe 300 mm long, Ø 8 mm, Tmax 500 °C
0600 9761
Flue gas probe 180 mm long, Ø 6 mm, Tmax 500 °C
0600 9762
Flue gas probe 300 mm long, Ø 6 mm, Tmax 500 °C
0600 9763
Probe accessories
Probe shaft 180 mm long, Ø 8 mm, Tmax 500 °C
0554 9760
Probe shaft 180 mm long, Ø 6 mm, Tmax 500 °C
0554 9762
Probe shaft 300 mm long, Ø 8 mm, Tmax 500 °C
0554 9761
Spare measurement cells
Best.-Nr.
Probe shaft 300 mm long, Ø 6 mm, Tmax 500 °C
0554 9763
Spare O2 measurement cell for testo 330-2
0390 0092
Probe shaft, 335 mm long, with probe stop, Ø 8 mm, Tmax 1000 °C
0554 8764
Spare CO measurement cell (H2 compensated) for testo 330-2
0390 0109
Probe shaft, 700 mm long, with probe stop, Ø 8 mm, Tmax 1000 °C
0554 8765
Spare NOlow measurement cell 0 to 300 ppm for testo 330-2/-2 LL
0390 0094
Flexible probe shaft 330 mm long, Ø 10 mm, Tmax 500 °C
0554 9764
Spare NO measuring cell 0-3000 ppm für testo 330-2/-2 LL
0390 0074
Multi-hole probe shaft, 300 mm long, Ø 8 mm, for mean CO calculation
0554 5762
O2 measurement cell für testo 330-2 LL
0390 0061
Multi-hole probe shaft, 180 mm long, Ø 8 mm, for mean CO calculation
0554 5763
CO measurement cell für testo 330-2 LL
0390 0090
Hose extension, 2.8 m, extension cable for probe and analyser
0554 1202
Upgrade: NO meas. cell, meas. range 0 to 3000 ppm, 1 ppm resolution für
testo 330-2/-2 LL
0554 3922
6 mm probe stop, PTFE, with spring clamp and handle, Tmax 200 °C
0554 3327
8 mm probe stop, PTFE, with spring clamp and handle, Tmax 200 °C
0554 3328
Upgrade: NOlow meas. cell, meas. range 0 to 300 ppm, 0.1 ppm resolution für
testo 330-2/-2 LL
0554 3931
8 mm probe stop, steel, with spring clamp and handle, Tmax 500 °C
0554 3330
Upgrade testo 330-2 to testo 330-2 LL (only possible by Testo service!)
Part no.
6 mm, probe stop, steel, with spring clamp and handle, Tmax 500 °C
0554 3329
Additional probes
Upgrade testo 330-2 to testo 330-2 LL consisting of
Qty.
Universal system case w/o pockets, can be interlocked to basic system case
Modular flue gas probes, available in 2 lengths, incl. positioning cone,
NiCr-Ni thermocouple, 2.2 m hose and particle filter
· testo 330-2 flue gas analyser incl. rechargeable battery and calibration
0563 3315 70
protocol
· 100-240 V mains unit for mains operation or battery recharging in instrument
· Modular flue gas probe, immersion depth: 300 mm, ø 8 mm
· Combustion air temperature probe, 190 mm long
· Hose connection set for separate gas pressure measurement
· IRDA printer
· Smoke tester incl. oil, soot sheets and reference scale
· Basic system case for analyser, probes and accessories
Qty.
Cases
Option code upgrade testo 330-2 to testo 330-2 LL
Upgrade O2 LL measurement cell
Upgrade CO LL measurement cell
0450 1100
0554 3938
0554 3937
Dual wall clearance probe
0632 1260
Accessories
Best.-Nr.
Gas leak probe
0632 3330
100-240 V mains unit for mains operation or batt. recharging in analyser
0554 1086
Ambient CO probe
0632 1247
Spare rech. batt. w/ charging station
0554 1087
Ambient CO2 probe
0632 1240
Connection cable
0430 0143
Testo printer with wireless IRDA and infrared interface, 1 roll of thermal paper
0554 0547
and 4 AA batteries
Testo IrDA printer with bidirectional IrDA interface and rech. battery, for
communication testo 330 and PDA
0554 0548
Spare thermal paper for printer (6 rolls), permanent ink
0554 0568
Combustion air temperature probe, immersion depth 300 mm
0600 9791
Barcode reader to read in customer number on site
0554 0461
Combustion air temperature probe, immersion depth 190 mm
0600 9787
Barcode labels, self-adhesive (1200 off)
0554 0411
Combustion air temperature probe, immersion depth 60 mm
0600 9797
Adhesive pockets (50 off) for printout, paper barcode labels...
0554 0116
Instrument cleaner (100 ml)
0554 1207
Smoke tester with oil, soot sheet, for measuring soot in flue gas
0554 0307
Hose connection set for separate gas pressure measurement
0554 1203
Differential temperature set consisting of 2 pipe clamp probes and adapter
0554 1204
Spare particle filter (10 off)
0554 3385
Full version EasyHeat (PC software)
0554 3332
Upgrade EasyHeat Mobile for PDA
0554 1211
Full version EasyHeat + EasyHeat Mobile (for PC und PDA)
0554 1210
USB connection cable, instrument to PC
0449 0047
ISO calibration certificate/flue gas
0520 0003
Combustion air temperature probe
Additional temperature probes
Mini ambient air probe
0600 3692
Pipe clamp probe
0600 4593
Quick-action surface probe
0604 0194
Connection cable
0430 0143
Sender
Name
Address
Company
City
Department
Date, Signature
Subject to change without notice.
Qty.
Best.-Nr.
Best.-Nr.
testo 330-2 flue gas analyser incl. rechargeable battery and calibration
protocol
testo 330-2 LL flue gas analyser with Longlife cells with integrated draught
and gas zeroing, incl. rech. batt. and calibration protocol
Option: Fine draught measurement, Resolution 0.1 Pa, measurement range to
100 Pa (instead of the standard draught measurement)
Qty.
Qty.
Instrument/options
0981 5644/msp/AC/A/08.2006
Qty.
200
Analyser for industrial flue gas
testo 325-I provides affordable flue gas
measurement for SO2. It combines precision
with user-friendly operation and low costs. It
is the ideal partner for checking emissions
and monitoring thermal processes.
The reading is constantly displayed and
updated during the measurement. The data
can be printed with date and time on the
cordless Testo printer at the touch of a
button.
Prints
HOLD
Adjusts flue gas feedback for NOx reduction
User-friendly operation and
handling - large display
ppm
SO2
Measuring cell can be easily
changed by user
ppm
NO
Magnetic SoftCase protects from
dirt and impact
Prints readings with date and time
ppm
CO
testo 325-I
testo 325-I
testo 325-I
testo 325-I
SO2 Set
NO Set
COlow Set
COhigh Set
testo 325-I, SO2 Set includes analyser and
sampling probe (with Tygon® hose), incl.
batteries and calibration protocol
testo 325-I, NO Set includes analyser and
sampling probe, incl. batteries and calibration
protocol
testo 325-I, COlow Set includes analyser and
sampling probe, incl. batteries and calibraton
protocol
testo 325-I, COhigh Set includes analyser and
sampling probe, incl. batteries and calibration
protocol
Part no.
Part no.
Part no.
Part no.
0563 3260
testo 325-I SO2
0563 3261
testo 325-I NO
0563 3262
testo 325-I COlow
0563 3263
testo 325-I COhigh
Checks emissions in coal and
heavy oil burners
Checks emissions in motors and
burners
Emission control and adjustment
of gas burners
Checks the atmosphere of thermal
processes in the production sector
(tunnel furnaces, hardening
furnaces, smelting and annealing)
Monitors flue gas desulphurisation
systems
Monitors nitrogen removal
systems/catalytic converters
Localisation of ambient air leaks
in long flue gas paths
Adjustment of process burners
Process monitoring in the glass
and ceramics industry
Process monitoring in the glass
and ceramics industry
Suitable probes at a glance
Description
Sampling probe, 700 mm immersion depth, incl.
cone for positioning, Tmax +1000 °C, 3 m hose
Illustration
Part no.
700 mm
Ø 8 mm
0699 3451/3
201
Practical accessories and recommended sets
testo 325-I, Set for emission checks
Instrument protection thanks to integrated
condensate trap
Testo printer
Prints readings on location
Part no.
ppm
SO2
0554 0545
ppm
NO
ppm
CO
SoftCase made of plastic with magnetic plate,
carrier strap
Protects instrument from dirt and impact
Part no.
0516 2570
TopSafe (indestructible protection case) IP 42, incl.
magnetic plate, carrier strap
Part no.
testo 325-I, SO2 Set includes analyser and
sampling probe (with Tygon® hose), incl.
batteries and calibration protocol
Testo printer with 1 roll of thermal paper and 4 AA
size batteries
SoftCase made of plastic with magnetic plate,
carrier strap
Transport case (plastic) for instrument, probes
and accessories
0516 0444
Ordering data for Accessories
Part no.
Sealing cone with knurled screw for sampling probe
0554 9050
Mains unit 230 V, for instrument (European plug)
For mains operation and battery recharging
0554 1084
Testo printer with 1 roll of thermal paper and 4 AA size batteries
Prints readings on location
0554 0545
Recharger for printer (with 4 standard rech. batt.)
Rechargeable batteries are recharged externally
0554 0110
Spare thermal paper for printer (6 rolls)
0554 0569
Enhanced thermal paper for printer (6 rolls)
Measurement data documentation legible for up to 10 years
0554 0568
SoftCase made of plastic with magnetic plate, carrier strap
0516 2570
Protects instrument from dirt and impact
TopSafe (indestructible protection case) IP 42, incl. magnetic plate, carrier strap 0516 0444
Spare particle filter (10 off)
For CO flue gas probe
0554 0040
Transport case (plastic) for instrument, probes and accessories
For safe and orderly storage
0516 3250
Technical data
0563 3260
0563 3261
0563 3262
0563 3263
Meas. range
0... +3000 ppm
SO2
0... +1000 ppm NO 0... +2000 ppm CO 0... +40000 ppm
CO
Accuracy
±5% of mv
(+400... +3000
ppm SO2)
±20 ppm SO2 (0...
+400 ppm SO2)
±5% of mv
(+400... +1000
ppm NO)
±20 ppm NO (0...
+400 ppm NO)
±5% of mv
(+400... +2000
ppm CO)
±20 ppm CO (0...
+400 ppm CO)
±5% of mv
(+800... +2000
ppm CO)
±10% of mv
(+2000... +40000
ppm CO)
±40 ppm CO (0...
+800 ppm CO)
1 ppm SO2 (0...
+3000 ppm SO2)
1 ppm NO (0...
+1000 ppm NO)
1 ppm CO (0...
+2000 ppm CO)
5 ppm CO (0...
+9999 ppm CO)
± 1 digit
Resolution
0.001 Vol. % CO
(+1... +4 Vol. % CO)
Reaction time
80 s
60 s
Dimensions
Weight
Oper. temp.
Storage temp.
Battery type
Battery life
Material/Housing
Power supply
Voltage
Display
216x68x47 mm
500 g
+4... +45 °C
-20... +50 °C
4 AA batteries
4h
ABS
Mains unit
115/230 V / 50/60 Hz
LCD, 2 lines
60 s
70 s
Common data
Warranty:
Analyser: 2 years (excluding working parts e.g.
measuring cells)
Measuring cells: 1/2 year
Power supply:
Via battery or mains unit
testo 330-1
New Measurement Engineering for
Heating Fitters and Technicians
NEW!
Increased safety and more convenience
°C
hPa
O2
CO
NO
∆P
Increased safety and convenience in flue gas analysis
The new flue gas analyser generation offers the heating fitter and technician everything that he needs to carry
his measurement tasks competently.
1
2
3
Increased safety
More convenience
User-friendly
The new testo 330-1 opens up
With testo 330 you have the
The new, highly robust fast probe
new opportunities for planning
practical benefit of Testo’s
connection for all gas paths
applications. The times in which
rechargeable batteries which can
eliminates any confusion. The
an analyser suddenly lets the
be recharged outside and inside
single cable cannot be bent, is
user down belong to the past.
the analyser. The lithium ion
indestructible and saves space.
testo 330’s instrument diagnosis
rechargeable battery is very
The menu-driven operational
provides you with information
compact and has a lifetime of
procedures - in compliance with
about the qualitative status of
more than six hours. The
standards - help you to avoid
functions and wearing parts.
rechargeable battery can be
errors and save you time. The
Using testo 330, it is therefore
changed quickly and easily by the
very large display is a perfect
easier to plan your daily work.
user. The cell can be changed
additional aid.
with the same speed and ease as
the rechargeable battery.
hPa
4
NO
O2
∆Π
°C
CO
5
6
Long life of wearing Faster and more
parts
effective
Totally robust
analyser concept
The typical service life of the
testo 330 is calibrated in 30
The new testo 330 not only
wearing parts for Testo’s own
seconds. Fuel, for example, can
stands out on account of its
measurement cells is now three
be selected at the same time and
distinctive and ergonomic design
years. The lithium ion
the analyser is then ready for use.
but also because of its unique
rechargeable batteries - free
When measuring differential
robust housing.
from the memory effect and total
pressure, differential temperature
The material used functions as a
discharge - are currently
or gas flow rate just by quickly
built-in protection from impact
unbeatable in terms of service
and simply attaching the probe,
and shock. The large backlit
life.
the analyser can automatically
display is slightly recessed in the
recognise the job and the
housing for better protection.
corresponding measurement
menu appears immediately.
The right accessories for every application.
The probes
The robust probe handle which
and dual wall clearance probe
which can be attached to the
fits nicely in your hand facilitates
round off our wide range of flue
analyser also has a user-defined
easy positioning. The probe has a
gas probes. The ambient CO
alarm limit. A warning signal is
quick connection function
probe as well as the CO2 probe
triggered if the limit is exceeded.
ensuring that all gas paths are
are recognised automatically by
The probe is also recognised
connected correctly. The particle
the analyser and the
automatically and the results are
filter accommodated in the
measurement results immediately
shown in a trend graph in their
handle blocks out particles
appear in their own graphically
own menu.
efficiently. Additionally, the probes
supported menu. Furthermore,
require little or no maintenance
the user-defined alarm limit is
and are easy to clean. Different
coupled with an audible signal.
lengths and diameters ensure a
Measurements can be carried out
high flexibility level for all
in their entirety simultaneously
applications. When changing the
with flue gas measurement.
probe, the probe shaft is simply
The gas leak detection probe
attached to the probe handle and
clicks into place. The multi-hole
The new versatile
infrared printer
New upload and
configuration software
The new testo printer – cordless
When developing the software,
with infared interface – saves
user-friendly operation and
data for printing. This saves time
handling had top priority. A
since the analyser transmits the
modern USB interface is available
data in two seconds and is ready
for this purpose.
for operation immediately after
Data cannot only be uploaded, it
the data is transmitted. The
can also be conveniently filed and
printer can be used anywhere.
managed. It is also possible to
control the analyser via software
when measuring online.
At a glance
The set for the heating fitter and technician
Testo has put together a special set for heating fitters and technicians. Of course,
it is always possible to add to and supplement the set by selecting from the wide
range of accessories available.
· testo 330-1 flue gas analyser incl. rech. battery and calibration protocol
· 100-240 V mains unit for mains operation or battery recharging in analyser
· Modular flue gas probe, immersion depth: 300 mm, ø 8 mm
· Combustion air temperature probe, immersion depth: 190 mm
· Hose connection set for separate gas pressure measurement
· IRDA printer
· Basic system case for instrument, probes and accessories
Part no.
0563 3314 70
Always a step ahead
testo 330-1 at a glance
Instrument diagnosis
Automatic menu selection
Additional features
• Rechargeable battery status display
• Recognition of connected probe
• Interface to Testo Software with analysis
and graphics functions and online
• Sensor status display
measurement
• Monitoring of condensate trap level with
„FULL“ message
• Display of pump capacity (l/min.)
• Display of error status incl. description and
diagnosis
• Display of last maintenance
• Display of instrument temperature
• Operating hour counter
Memory management
• 200 measurement data records incl.
systems number
• Several measurement data records per
system number are possible
• Read in using barcode reader
• IRDA interface for data transfer to
PDA/Notebook
• USB interface for data transfer to PC
Additional measurement tasks
• ∆T measurement, flow/return
• Ambient CO measurement
• Ambient CO2 measurement
• Detection of gas leaks with gas leak
detection probe
• ∆P measurement for gas pressure
• Gas/oil flow rate determination
• ZIV driver for all standard branch software
• Calculated parameter: Flue gas dewpoint
• Graphics display
• Integrated impact protection with magnet,
recessed display
• Protection class: IP40
• Rechargeable battery lifetime > 6 h with
pump operating
• Rechargeable battery separate and
rechargeable in instrument
• Adjustable alarm thresholds for ambient
CO/CO2 measurement
Technical data
Temperature
Meas. range
Accuracy
Resolution
Draught measurement
Pressure measurement
O2 measurement
CO measurement
(without H2
compensation)
Meas. range
Accuracy
Resolution
Meas. range
Accuracy
Resolution
Meas. range
Resolution
Accuracy
Adjustment time t90
-40 to +1200 °C
±0,5 °C (0.0 to +100.0 °C)
±0,5 % of mv (remaining range)
0.1 °C (-40 to 999,9 °C)
1 °C (remaining range)
-40 to 40 hPa
±0.02 hPa or ±0.5 % of mv (-0.50 to +0,60 hPa)
±0.03 hPa or ±0.5 % of mv (+0.61 to +3.00 hPa)
±1.5 % of mv (+3.01 to +40.00 hPa)
0.01 hPa
0 to 200 hPa
±0.5 hPa (0.0 to 50.0 hPa)
±1 % of mv (50.1 to 100.0 hPa)
±1.5 % of mv (remaining range)
0.1 hPa
0 to 21 Vol. %
0.1 Vol. %
±0.2 Vol. %
< 20 s
Adjustment time t90
0 to 4000 ppm
1 ppm
±20 ppm (0 to 400 ppm)
±5% of mv (401 to 1000 ppm)
±10% of mv (1001 to 4000 ppm)
< 60 s
Efficiency
(ETA)
Meas. range
Resolution
0 to 120 %
0.1 %
Flue gas loss
Meas. range
Resolution
0 to 99.9 %
0.1 %
CO2 measurement
Display range
Resolution
Accuracy
Measurement
Adjustment time t90
Meas. range
Resolution
Accuracy
0 to CO2 max
0.1 Vol. %
±0.2 Vol. %
Digital calculation from O2
< 40 s
0 to 300 ppm
0.1 ppm
±2 ppm (0.0 to 40.0 ppm)
±5% of mv (remaining range)
< 30 s
Option: NOlow
measurement
Meas. range
Resolution
Accuracy
Adjustment time t90
Option: NO
measurement
Meas. range
Resolution
Accuracy
Ambient CO
measurement (with CO
probe)
Meas. range
Resolution
Accuracy
Adjustment time t90
Gas leak measurement
for combustible gases
(with gas leak detection
probe)
Adjustment time t90
Range of indication
Signal
Adjustment time t90
Ambient CO2
measurement (with
ambient CO2 probe)
Meas. range
General Technical Data
Memory
Weight
Dimensions
Storage temp.
Oper. temp.
Display
Power supply
Accuracy
Adjustment time t90
Warranty
IRDA printer
Printer type
Reception radius
Dimensions
Weight
Oper. temp.
Storage temp.
Power supply
*
*
0 to 3000 ppm
1 ppm
±5 ppm (0 to 100 ppm)
±5% of mv (101 to 2000 ppm)
±10% of mv (2001 to 3000 ppm)
< 30 s
0 to 500 ppm
1 ppm
±5 ppm (0 to 100 ppm)
±5% of mv (>100 ppm)
Approx. 35 s
0 ... 10,000 ppm CH4 / C3H8
Optical display (LED)
audible display via buzzer
<2s
0 to 1 Vol. %
0 to 10000 ppm
±(50 ppm ±2 % of mv) (0 to 5000 ppm)
Approx. 35 s
200 Sites
600 g (without rechargeable battery)
270 x 90 x 65 mm
-20 to +50 °C
-5 to +45 °C
Graphics display: 160 x 240 pixel
Rechargeable block: 3.7 V / 2.2 Ah
Mains unit: 6 V / 1.2 A
Analyser/cells/probe 2 years, rechargeable battery 1 year
Infrared-controlled thermal printer, adjustable contrast,
graphics can be printed
Max. 2 m
186 x 91 x 61 mm
430 g incl. batteries
0 to +50 °C
-40 to +60 °C
4 round cell batteries, 1.5 V or NC rechargeable batteries
*the larger value applies
testo 330-1
New Measurement Engineering for
Heating Fitters and Technicians
NEW!
Increased safety and more convenience
°C
hPa
O2
CO
NO
∆P
Increased safety and convenience in flue gas analysis
The new flue gas analyser generation offers the heating fitter and technician everything that he needs to carry
his measurement tasks competently.
1
2
3
Increased safety
More convenience
User-friendly
The new testo 330-1 opens up
With testo 330 you have the
The new, highly robust fast probe
new opportunities for planning
practical benefit of Testo’s
connection for all gas paths
applications. The times in which
rechargeable batteries which can
eliminates any confusion. The
an analyser suddenly lets the
be recharged outside and inside
single cable cannot be bent, is
user down belong to the past.
the analyser. The lithium ion
indestructible and saves space.
testo 330’s instrument diagnosis
rechargeable battery is very
The menu-driven operational
provides you with information
compact and has a lifetime of
procedures - in compliance with
about the qualitative status of
more than six hours. The
standards - help you to avoid
functions and wearing parts.
rechargeable battery can be
errors and save you time. The
Using testo 330, it is therefore
changed quickly and easily by the
very large display is a perfect
easier to plan your daily work.
user. The cell can be changed
additional aid.
with the same speed and ease as
the rechargeable battery.
hPa
4
NO
O2
∆Π
°C
CO
5
6
Long life of wearing Faster and more
parts
effective
Totally robust
analyser concept
The typical service life of the
testo 330 is calibrated in 30
The new testo 330 not only
wearing parts for Testo’s own
seconds. Fuel, for example, can
stands out on account of its
measurement cells is now three
be selected at the same time and
distinctive and ergonomic design
years. The lithium ion
the analyser is then ready for use.
but also because of its unique
rechargeable batteries - free
When measuring differential
robust housing.
from the memory effect and total
pressure, differential temperature
The material used functions as a
discharge - are currently
or gas flow rate just by quickly
built-in protection from impact
unbeatable in terms of service
and simply attaching the probe,
and shock. The large backlit
life.
the analyser can automatically
display is slightly recessed in the
recognise the job and the
housing for better protection.
corresponding measurement
menu appears immediately.
The right accessories for every application.
The probes
The robust probe handle which
and dual wall clearance probe
which can be attached to the
fits nicely in your hand facilitates
round off our wide range of flue
analyser also has a user-defined
easy positioning. The probe has a
gas probes. The ambient CO
alarm limit. A warning signal is
quick connection function
probe as well as the CO2 probe
triggered if the limit is exceeded.
ensuring that all gas paths are
are recognised automatically by
The probe is also recognised
connected correctly. The particle
the analyser and the
automatically and the results are
filter accommodated in the
measurement results immediately
shown in a trend graph in their
handle blocks out particles
appear in their own graphically
own menu.
efficiently. Additionally, the probes
supported menu. Furthermore,
require little or no maintenance
the user-defined alarm limit is
and are easy to clean. Different
coupled with an audible signal.
lengths and diameters ensure a
Measurements can be carried out
high flexibility level for all
in their entirety simultaneously
applications. When changing the
with flue gas measurement.
probe, the probe shaft is simply
The gas leak detection probe
attached to the probe handle and
clicks into place. The multi-hole
The new versatile
infrared printer
New upload and
configuration software
The new testo printer – cordless
When developing the software,
with infared interface – saves
user-friendly operation and
data for printing. This saves time
handling had top priority. A
since the analyser transmits the
modern USB interface is available
data in two seconds and is ready
for this purpose.
for operation immediately after
Data cannot only be uploaded, it
the data is transmitted. The
can also be conveniently filed and
printer can be used anywhere.
managed. It is also possible to
control the analyser via software
when measuring online.
At a glance
The set for the heating fitter and technician
Testo has put together a special set for heating fitters and technicians. Of course,
it is always possible to add to and supplement the set by selecting from the wide
range of accessories available.
· testo 330-1 flue gas analyser incl. rech. battery and calibration protocol
· 100-240 V mains unit for mains operation or battery recharging in analyser
· Modular flue gas probe, immersion depth: 300 mm, ø 8 mm
· Combustion air temperature probe, immersion depth: 190 mm
· Hose connection set for separate gas pressure measurement
· IRDA printer
· Basic system case for instrument, probes and accessories
Part no.
0563 3314 70
Always a step ahead
testo 330-1 at a glance
Instrument diagnosis
Automatic menu selection
Additional features
• Rechargeable battery status display
• Recognition of connected probe
• Interface to Testo Software with analysis
and graphics functions and online
• Sensor status display
measurement
• Monitoring of condensate trap level with
„FULL“ message
• Display of pump capacity (l/min.)
• Display of error status incl. description and
diagnosis
• Display of last maintenance
• Display of instrument temperature
• Operating hour counter
Memory management
• 200 measurement data records incl.
systems number
• Several measurement data records per
system number are possible
• Read in using barcode reader
• IRDA interface for data transfer to
PDA/Notebook
• USB interface for data transfer to PC
Additional measurement tasks
• ∆T measurement, flow/return
• Ambient CO measurement
• Ambient CO2 measurement
• Detection of gas leaks with gas leak
detection probe
• ∆P measurement for gas pressure
• Gas/oil flow rate determination
• ZIV driver for all standard branch software
• Calculated parameter: Flue gas dewpoint
• Graphics display
• Integrated impact protection with magnet,
recessed display
• Protection class: IP40
• Rechargeable battery lifetime > 6 h with
pump operating
• Rechargeable battery separate and
rechargeable in instrument
• Adjustable alarm thresholds for ambient
CO/CO2 measurement
Technical data
Temperature
Meas. range
Accuracy
Resolution
Draught measurement
Pressure measurement
O2 measurement
CO measurement
(without H2
compensation)
Meas. range
Accuracy
Resolution
Meas. range
Accuracy
Resolution
Meas. range
Resolution
Accuracy
Adjustment time t90
-40 to +1200 °C
±0,5 °C (0.0 to +100.0 °C)
±0,5 % of mv (remaining range)
0.1 °C (-40 to 999,9 °C)
1 °C (remaining range)
-40 to 40 hPa
±0.02 hPa or ±0.5 % of mv (-0.50 to +0,60 hPa)
±0.03 hPa or ±0.5 % of mv (+0.61 to +3.00 hPa)
±1.5 % of mv (+3.01 to +40.00 hPa)
0.01 hPa
0 to 200 hPa
±0.5 hPa (0.0 to 50.0 hPa)
±1 % of mv (50.1 to 100.0 hPa)
±1.5 % of mv (remaining range)
0.1 hPa
0 to 21 Vol. %
0.1 Vol. %
±0.2 Vol. %
< 20 s
Adjustment time t90
0 to 4000 ppm
1 ppm
±20 ppm (0 to 400 ppm)
±5% of mv (401 to 1000 ppm)
±10% of mv (1001 to 4000 ppm)
< 60 s
Efficiency
(ETA)
Meas. range
Resolution
0 to 120 %
0.1 %
Flue gas loss
Meas. range
Resolution
0 to 99.9 %
0.1 %
CO2 measurement
Display range
Resolution
Accuracy
Measurement
Adjustment time t90
Meas. range
Resolution
Accuracy
0 to CO2 max
0.1 Vol. %
±0.2 Vol. %
Digital calculation from O2
< 40 s
0 to 300 ppm
0.1 ppm
±2 ppm (0.0 to 40.0 ppm)
±5% of mv (remaining range)
< 30 s
Option: NOlow
measurement
Meas. range
Resolution
Accuracy
Adjustment time t90
Option: NO
measurement
Meas. range
Resolution
Accuracy
Ambient CO
measurement (with CO
probe)
Meas. range
Resolution
Accuracy
Adjustment time t90
Gas leak measurement
for combustible gases
(with gas leak detection
probe)
Adjustment time t90
Range of indication
Signal
Adjustment time t90
Ambient CO2
measurement (with
ambient CO2 probe)
Meas. range
General Technical Data
Memory
Weight
Dimensions
Storage temp.
Oper. temp.
Display
Power supply
Accuracy
Adjustment time t90
Warranty
IRDA printer
Printer type
Reception radius
Dimensions
Weight
Oper. temp.
Storage temp.
Power supply
*
*
0 to 3000 ppm
1 ppm
±5 ppm (0 to 100 ppm)
±5% of mv (101 to 2000 ppm)
±10% of mv (2001 to 3000 ppm)
< 30 s
0 to 500 ppm
1 ppm
±5 ppm (0 to 100 ppm)
±5% of mv (>100 ppm)
Approx. 35 s
0 ... 10,000 ppm CH4 / C3H8
Optical display (LED)
audible display via buzzer
<2s
0 to 1 Vol. %
0 to 10000 ppm
±(50 ppm ±2 % of mv) (0 to 5000 ppm)
Approx. 35 s
200 Sites
600 g (without rechargeable battery)
270 x 90 x 65 mm
-20 to +50 °C
-5 to +45 °C
Graphics display: 160 x 240 pixel
Rechargeable block: 3.7 V / 2.2 Ah
Mains unit: 6 V / 1.2 A
Analyser/cells/probe 2 years, rechargeable battery 1 year
Infrared-controlled thermal printer, adjustable contrast,
graphics can be printed
Max. 2 m
186 x 91 x 61 mm
430 g incl. batteries
0 to +50 °C
-40 to +60 °C
4 round cell batteries, 1.5 V or NC rechargeable batteries
*the larger value applies
Fax Order Form
Qty.
Qty.
testo 330-1 Set: Pro analysis for heating technicians and fitters
Part no.
Qty.
· testo 330-1 flue gas analyser incl. rech. battery and calibration protocol
· 100-240 V mains unit for mains operation or battery recharging in analyser
· Modular flue gas probe, immersion depth: 300 mm, ø 8 mm
· Combustion air temperature probe, immersion depth: 190 mm
· Hose connection set for separate gas pressure measurement
· IRDA printer
· Basic system case for instrument, probes and accessories
0563 3314 70
Modular flue gas probes, available in 2 lengths, incl. positioning cone,
NiCr-Ni thermocouple, 2.2 m hose and particle filter
Instrument / Options / Upgrades
testo 330-1 flue gas analyser incl. rechargeable battery and calibration
protocol
Option: Fine draught measurement, Resolution 0.1 Pa, measurement range to
100 Pa (instead of the standard draught measurement)
Qty.
Part no.
Probes
Part no.
Flue gas probe 180 mm long, Ø 8 mm, Tmax 500 °C
0600 9760
Flue gas probe 300 mm long, Ø 8 mm, Tmax 500 °C
0600 9761
Flue gas probe 180 mm long, Ø 6 mm, Tmax 500 °C
0600 9762
Flue gas probe 300 mm long, Ø 6 mm, Tmax 500 °C
0600 9763
Probe accessories
0632 3301
Probe shaft 180 mm long, Ø 8 mm, Tmax 500 °C
0554 9760
0440 3921
Probe shaft 180 mm long, Ø 6 mm, Tmax 500 °C
0554 9762
Probe shaft 300 mm long, Ø 8 mm, Tmax 500 °C
0554 9761
Option: NO meas. cell, meas. range 0 to 3000 ppm, 1 ppm resolution
On request
Upgrade: NO meas. cell, meas. range 0 to 3000 ppm, 1 ppm resolution
On request
Probe shaft 300 mm long, Ø 6 mm, Tmax 500 °C
0554 9763
Probe shaft 300 mm, Ø 8 mm, Tmax 1000 °C
0554 8764
Option: NOlow meas. cell, meas. range 0 to 300 ppm, 0.1 ppm resolution
0440 3931
Upgrade: NOlow meas. cell, meas. range 0 to 300 ppm, 0.1 ppm resolution
0554 3931
Multi-hole probe shaft, 300 mm long, Ø 8 mm, for mean CO calculation
0554 5762
Accessories
Part no.
Hose extension, 2.80 m, extension cable for probe and analyser
0554 1202
100-240 V mains unit for mains operation or batt. recharging in analyser
0554 1086
Cone 8 mm, steel, with grip clamp and grip, Tmax 500 °C
0554 3330
Spare rechargeable battery with recharging station
0554 1087
Cone 6 mm, steel, with spring clamp and grip, Tmax 500 °C
0554 3329
IRDA printer
0554 0547
Spare thermal paper for printer (6 rolls), permanent ink
0554 0568
Barcode reader to read in customer number on site
0554 0461
Probe shaft 700 mm long, Ø 8 mm, Tmax 1000 °C
0554 8765
Flexible probe shaft 330 mm long, Ø 10 mm, Tmax 500 °C
0554 9764
Additional probes
Dual wall clearance probe
0632 1260
Gas leak probe, available from December 2004
0632 3330
Ambient CO probe
0632 1247
Barcode labels, self-adhesive (1200 off)
0554 0411
Adhesive pockets (50 off) for printout, paper barcode labels...
0554 0116
Smoke tester with oil, soot sheet, for measuring soot in flue gas
0554 0307
Hose connection set for separate gas pressure measurement
0554 1203
Ambient CO2 probe
0632 1240
Differential temperature set consisting of 2 pipe clamp probes and adapter
0554 1204
Connection cable
0440 0143
Spare particle filter (10 off)
0554 3385
Combustion air temperature probe
testo 330 software with analysis and graphics functions, online measurement 0554 3332
Combustion air temperature probe, immersion depth 300 mm
0600 9791
USB connection cable, instrument to PC
0449 0047
Combustion air temperature probe, immersion depth 190 mm
0600 9787
ISO calibration certificate/Flue gas
0520 0003
Combustion air temperature probe, immersion depth 60 mm
0600 9797
Qty.
Cases
Part no.
Mini ambient air probe
0600 3692
Pipe clamp probe
0600 4593
Basic system case for analyser, probes and accessories
0516 3330
Basic system case with reinforced base for analyser, probes and additional
accessories
0516 3331
Product system case with small plastic boxes without contents, can be
interlocked to basic system case
0516 0328
Quick-action surface probe
0604 0194
Tools system case with tools section without content, can be interlocked to
basic system case
0516 0329
Connection cable
0440 0143
Universal system case w/o pockets, can be interlocked to basic system case
0516 0331
Measurement case (leather) with drawers for instruments and accessories
0516 0303
Spare measurement cells
Part no.
Spare O2 measurement cell
0390 0092
Spare CO measurement cell (without H2 compensation)
0390 0095
Spare NOlow measurement cell 0 to 300 ppm
0390 0094
Spare NO measurement cell 0 to 3000 ppm
On request
Sender
Name
Company
Department
Address
City
Signature
0981 5634/hd/AC/A/07.2004
Qty.
Subject to change without notice.
Additional temperature probes
Fax Order Form
Qty.
Qty.
testo 330-1 Set: Pro analysis for heating technicians and fitters
Part no.
Qty.
· testo 330-1 flue gas analyser incl. rech. battery and calibration protocol
· 100-240 V mains unit for mains operation or battery recharging in analyser
· Modular flue gas probe, immersion depth: 300 mm, ø 8 mm
· Combustion air temperature probe, immersion depth: 190 mm
· Hose connection set for separate gas pressure measurement
· IRDA printer
· Basic system case for instrument, probes and accessories
0563 3314 70
Modular flue gas probes, available in 2 lengths, incl. positioning cone,
NiCr-Ni thermocouple, 2.2 m hose and particle filter
Instrument / Options / Upgrades
testo 330-1 flue gas analyser incl. rechargeable battery and calibration
protocol
Option: Fine draught measurement, Resolution 0.1 Pa, measurement range to
100 Pa (instead of the standard draught measurement)
Qty.
Part no.
Probes
Part no.
Flue gas probe 180 mm long, Ø 8 mm, Tmax 500 °C
0600 9760
Flue gas probe 300 mm long, Ø 8 mm, Tmax 500 °C
0600 9761
Flue gas probe 180 mm long, Ø 6 mm, Tmax 500 °C
0600 9762
Flue gas probe 300 mm long, Ø 6 mm, Tmax 500 °C
0600 9763
Probe accessories
0632 3301
Probe shaft 180 mm long, Ø 8 mm, Tmax 500 °C
0554 9760
0440 3921
Probe shaft 180 mm long, Ø 6 mm, Tmax 500 °C
0554 9762
Probe shaft 300 mm long, Ø 8 mm, Tmax 500 °C
0554 9761
Option: NO meas. cell, meas. range 0 to 3000 ppm, 1 ppm resolution
On request
Upgrade: NO meas. cell, meas. range 0 to 3000 ppm, 1 ppm resolution
On request
Probe shaft 300 mm long, Ø 6 mm, Tmax 500 °C
0554 9763
Probe shaft 300 mm, Ø 8 mm, Tmax 1000 °C
0554 8764
Option: NOlow meas. cell, meas. range 0 to 300 ppm, 0.1 ppm resolution
0440 3931
Upgrade: NOlow meas. cell, meas. range 0 to 300 ppm, 0.1 ppm resolution
0554 3931
Multi-hole probe shaft, 300 mm long, Ø 8 mm, for mean CO calculation
0554 5762
Accessories
Part no.
Hose extension, 2.80 m, extension cable for probe and analyser
0554 1202
100-240 V mains unit for mains operation or batt. recharging in analyser
0554 1086
Cone 8 mm, steel, with grip clamp and grip, Tmax 500 °C
0554 3330
Spare rechargeable battery with recharging station
0554 1087
Cone 6 mm, steel, with spring clamp and grip, Tmax 500 °C
0554 3329
IRDA printer
0554 0547
Spare thermal paper for printer (6 rolls), permanent ink
0554 0568
Barcode reader to read in customer number on site
0554 0461
Probe shaft 700 mm long, Ø 8 mm, Tmax 1000 °C
0554 8765
Flexible probe shaft 330 mm long, Ø 10 mm, Tmax 500 °C
0554 9764
Additional probes
Dual wall clearance probe
0632 1260
Gas leak probe, available from December 2004
0632 3330
Ambient CO probe
0632 1247
Barcode labels, self-adhesive (1200 off)
0554 0411
Adhesive pockets (50 off) for printout, paper barcode labels...
0554 0116
Smoke tester with oil, soot sheet, for measuring soot in flue gas
0554 0307
Hose connection set for separate gas pressure measurement
0554 1203
Ambient CO2 probe
0632 1240
Differential temperature set consisting of 2 pipe clamp probes and adapter
0554 1204
Connection cable
0440 0143
Spare particle filter (10 off)
0554 3385
Combustion air temperature probe
testo 330 software with analysis and graphics functions, online measurement 0554 3332
Combustion air temperature probe, immersion depth 300 mm
0600 9791
USB connection cable, instrument to PC
0449 0047
Combustion air temperature probe, immersion depth 190 mm
0600 9787
ISO calibration certificate/Flue gas
0520 0003
Combustion air temperature probe, immersion depth 60 mm
0600 9797
Qty.
Cases
Part no.
Mini ambient air probe
0600 3692
Pipe clamp probe
0600 4593
Basic system case for analyser, probes and accessories
0516 3330
Basic system case with reinforced base for analyser, probes and additional
accessories
0516 3331
Product system case with small plastic boxes without contents, can be
interlocked to basic system case
0516 0328
Quick-action surface probe
0604 0194
Tools system case with tools section without content, can be interlocked to
basic system case
0516 0329
Connection cable
0440 0143
Universal system case w/o pockets, can be interlocked to basic system case
0516 0331
Measurement case (leather) with drawers for instruments and accessories
0516 0303
Spare measurement cells
Part no.
Spare O2 measurement cell
0390 0092
Spare CO measurement cell (without H2 compensation)
0390 0095
Spare NOlow measurement cell 0 to 300 ppm
0390 0094
Spare NO measurement cell 0 to 3000 ppm
On request
Sender
Name
Company
Department
Address
City
Signature
0981 5634/hd/AC/A/07.2004
Qty.
Subject to change without notice.
Additional temperature probes
testo 330-2
New Measurement Engineering for
Customer Service and Service Technicians
NEW!
Robust, long-lasting and faster
°C
hPa
O2
CO/H2
The latest measurement engineering:
robuster, longer-lasting and faster
The new flue gas analyser generation offers Customer Service and service technicians
all the features required to compently carry out the measurement tasks at hand.
Regardless of whether in the context of breakdowns or emergencies, during the
monitoring of legal limit values or during daily maintenance assignments, testo 330-2
is a reliable partner for the complete spectrum of measurement tasks. With the option
of carrying out NOlow measurements on lowNOx burner systems, this check –
although not compulsory but relevant for the environment - can be completed to
highest efficiency levels.
1 Completely robust instrument concept
1
For applications in extreme
blockage caused by condensate.
stands out on account of its
conditions, such as highly
It is easy to gain access to the
distinctive and ergonomic design
contaminated burners, Testo’s
condensate trap for emptying
but also on account of its unique
patented technology guarantees
purposes. It is located just in front
robust housing design. The
efficient display of the CO level of
of the measurement cell in the
material used acts as a built-in
up to 30,000 ppm. The
housing.
protection from impact and
measurement cell is not
The particle filter absorbs
shock. The large backlit display is
subjected to any more strain than
impurities at the point where they
positioned slightly
during normal measurements.
accumulate.That is why the filter
recessed in the
The design of the gas paths and
cartridge is located in the probe
housing for
Testo measurement cells in
handle, for easy access.
better
connection with the positioning of
the condensate trap prevents
protection.
The new testo 330 not only
2
2 Long life of wearing parts
The typical service life of Testo’s
own measurement cells now lies
at more than three years.
The lithium ion rechargeable
batteries – free from the memory
effect and total discharge - are
currently unbeatable in terms of
service life.
The design layout of the gas
paths and an additional filter
mean that the gas pump is
protected thus increasing service
life.
3 Faster and more effective
testo 330-2 supports the fast
flow rate by quickly and simply
easily document all results on a
switch between measurement
attaching the probe, the analyser
data carrier such as a PDA or
and maintenance. The probe can
can automatically recognise the
laptop.
remain positioned in the flue
job and the corresponding
during the zero phase of the
measurement menu appears
measurement cells. testo 330 is
immediately. The infrared
calibrated in less than 30
interface for the control
seconds. Fuel, for example, can
electronics of the burner systems
be selected at the same time and
facilitates fast error diagnosis on
the analyser is then ready for use.
the heating unit. The built-in IRDA
When measuring ∆p, ∆t or gas
interface makes it possible to
3
4 Increased safety
The new testo 330-2 opens up
testo 330’s instrument diagnosis
also shown. Using testo 330
new opportunities for planning
provides you with information
makes it easier to plan your daily
applications. The times in which
about the qualitative status of
work.
an analyser suddenly lets the
functions and wearing parts. The
user down belong to the past.
level on the condensate trap is
4
5 More convenience
and user-friendliness
With testo 330 you have the
practical benefit of Testo’s
rechargeable batteries which can
be recharged outside and inside
the analyser. The lithium ion
rechargeable battery is very
compact and has a lifetime of
more than six hours. The
rechargeable battery can be
changed quickly and easily by the
user.
The new, highly robust fast probe
connection for all gas paths
eliminates any confusion. The
single cable cannot be bent, is
indestructible and saves space.
The menu-driven operational
procedures – in compliance with
standards – help you to avoid
errors and save you time. The
very large display is a perfect
additional aid.
5
The right accessories for every application
The probes
The robust probe handle which
and dual wall clearance probe
analyser also has a user-defined
fits nicely in your hand facilitates
round off our wide range of flue
alarm limit. A warning signal is
easy positioning. The probe has a
gas probes. The ambient CO
triggered if the limit is exceeded.
quick connection function
probe as well as the CO2 probe
The probe is also recognised
ensuring that all gas paths are
are recognised automatically by
automatically and the results are
connected correctly. The particle
the analyser and the
shown in a trend graph in their
filter accommodated in the
measurement results immediately
own menu.
handle blocks out particles
appear in their own graphically
efficiently. Additionally, the probes
supported menu. Furthermore,
require little or no maintenance
the user-defined alarm limit is
and are easy to clean. Different
coupled with an audible signal.
lengths and diameters ensure a
Measurements can be carried out
high flexibility level for all
in their entirety simultaneously
applications. When changing the
with flue gas measurement.
probe, the probe shaft is simply
The gas leak detection probe
attached to the probe handle and
which can be attached to the
clicks into place. The multi-hole
The new versatile infrared printer
The new testo printer – cordless
with infrared interface – saves
data for printing. This saves time
since the analyser is ready for
operation immediately after the
data is transmitted. The printer
can be used anywhere.
New upload and configuration software
When developing the software,
conveniently filed and managed.
user-friendly operation and
It is also possible to control the
handling had top priority. A
analyser via software when
modern USB interface is available
measuring online.
for this purpose. Data cannot
only be uploaded, it can also be
At a glance
The set for Customer Service and service technicians
Testo has put together a special
set for customer service and
service technicians. Of course, it
is always possible to add to
and supplement the set
· testo 330-2 flue gas analyser incl. rechargeable battery and
calibration protocol
· 100-240 V mains unit for mains operation or battery recharging in
instrument
· Modular flue gas probe, immersion depth: 300 mm, ø 8 mm
by selecting from the
· Combustion air temperature probe, 190 mm long
comprehensive range
· Hose connection set for separate gas pressure measurement
of accessories
· IRDA printer
available.
· Smoke tester incl. oil, soot sheets and reference scale
· Basic system case for analyser, probes and accessories
Part no.
0563 3315 70
Always a step ahead
testo 330-2 at a glance
Instrument diagnosis
Automatic menu selection
Additional features
• Automatic instrument leak test
• Recognition of connected
• Interface to Testo Software
• Rechargeable battery status
probe
• Sensor status display
• Monitoring of condensate trap
level with „FULL“ message
• Display of pump capacity
(l/min.)
Initialisation of flue gas sensors
without removal of probes
• The probe can remain in the
flue during initialisation
temperature
• Operating hour counter
Memory management
• 200 meas. data records with
address and systems no.
• Several measurement data
records per system number are
Additional measurement tasks
• ∆T measurement, flow/return
• Ambient CO measurement
• Ambient CO2 measurement
• Detection of gas leaks with gas
leak detection probe
• ∆P measurement for gas
pressure
• Gas/oil flow rate determination
• Calculated parameter: Flue gas
dewpoint
• Graphics display
with magnet, recessed display
description and diagnosis
• Display of instrument
measurement
• Interface to burners
• Integrated impact protection
• Display of error status incl.
• Display of last maintenance
with analysis and graphics
functions and online
display
possible
• Read in using barcode reader
• Protection class: IP40
• Rechargeable battery lifetime >
6 h with pump operating
• Rechargeable battery separate
and rechargeable in instrument
• Adjustable alarm thresholds for
ambient CO/CO2 measurement
• IRDA interface for data transfer
to PDA/Notebook
• USB interface for data transfer
to PC
• ZIV driver for all conventional
branch software
TÜV approval / EN standard
• Accuracy approved for O2,
CO2, CO, NO, NOlow, °C, hPa
testo 330-3
New Measurement Engineering for
Heating Technicians and Inspectors
NEW!
A new dimension in speed and efficiency
°C
hPa
O2
CO/H2
testo 330-3 – A new dimension in speed and efficiency
The new flue gas analyser generation offers heating technicians all the features required to competently carry
out the inspection tasks at hand: the analyser has a special menu and automatic mean calculation. The course
of the temperature during hot spot detection is shown in graphs on the display. Draught measurement is also
1
Highest availability and
long life of wearing parts
2
Faster and more effective
testo 330 has the added
as the structuring of customer
The typical service life of Testo’s
advantage that testo
and measurement data can be
own measurement cells is now
rechargeable batteries can be
carried out according to
more than three years. The Testo
recharged outside the instrument
postcode, town, street, name or
design eliminates the formation of
as well in the instrument. The
system number. The new testo
condensate in the actual
lithium ion rechargeable battery is
330 also has a USB interface as
measurement cell. This and the
very compact and can be quickly
well as built-in IRDA interface.
special design of the gas paths in
and easily changed.
This makes it possible to transmit
the instrument combined with the
When changing probes - by
the results to the corresponding
position of the condensate trap
quickly and simply attaching the
branch software.
prevent blockages and increases
probe or probe shaft - the
service life considerably. The
analyser can automatically
design layout of the gas paths
recognise the job and the
and an additional filter mean that
corresponding measurement
the gas pump is protected thus
menu appears immediately. The
increasing service life. The lithium
probe can remain positioned in
ion rechargeable batteries – free
the flue during the zero phase of
from the memory effect and total
the measurement cells. testo 330
discharge - are currently
is calibrated in less than 30
unbeatable in terms of service
seconds. Fuel, for example, can
life. The service life of the
be selected at the same time and
rechargeable battery is more than
the analyser is then ready for use.
six hours when the pump is
The memory management
operating.
function in the instrument
facilitates easy and user-friendly
structured representations of the
data. The search function as well
shown in the same menu; the probe can remain positioned in the flue during the zero phase. This improves
measurement accuracy when checking the systems and helps to save you time. CO switch-off can be carried
out manually if necessary.
3
Increased safety
The new testo 330-3 opens up
new opportunities for planning
applications. The times in which
an analyser suddenly lets the
user down belong to the past.
testo 330’s instrument diagnosis
provides you with information
about the qualitative status of
functions and wearing parts.
Included is the quality check on
measurement cells, filter status,
level display in the condensate
trap, leak-proofness of
instrument, pump capacity,
rechargeable battery status and
the date of the last maintenance.
Your daily work is therefore easier
to plan.
New upload and configuration
software
When developing the software, userfriendly operation and handling had top
priority. Additionally, not only can data
be uploaded, it can also be
conveniently filed and managed. It is
also possible to control the analyser via
software when measuring online.
Long life of wearing parts.
The typical service life of the
wearing parts for Testo’s own
measurement cells is now three
years. The lithium ion rechargeable
batteries – free from the memory
effect and total discharge – are
currently unbeatable in terms of
service life.
4
5
More convenience and
user-friendliness
Totally robust instrument
concept
The service cover, which can be
The particle filter absorbs
easily removed without any tools,
contaminants at the points where
makes for easy care and
they occur. For this reason, the
maintenance of the instrument.
filter cartridge is located in the
Changing the lithium ion
probe handle for easy access.
rechargeable battery and
For applications in extreme
measurement cell could not be
conditions, such as highly
easier. Since the calibration data
contaminated burners, Testo’s
is saved in the measurement cell,
patented technology guarantees
it can be changed in compliance
efficient measurements up to a
with TÜV without the need for
CO measurement range of
sample gas calibration. The
30,000 ppm. The measurement
condensate trap is easy to
cell is not subjected to any more
access and to empty. It is located
strain than during normal
just in front of the measurement
measurements.
cell in the housing.
The new testo 330 not only
The new, very robust quick probe
stands out on account of its
connection for all gas paths
distinctive and ergonomic design
eliminates any confusion. The
but also on account of its unique
single cable cannot be bent, is
robust housing design.
indestructible and saves space.
The material used acts as a built-
The menu-driven operational
in protection from impact and
procedures – in accordance with
shock. The large backlit display is
standards – help you to avoid
positioned slightly recessed in the
errors and save you time. The
housing for better protection.
very large, backlit display is a
perfect additional aid. The size of
the display means you can switch
from four to eight parameters.
The right accessories for every application.
The probes
The robust probe handle which
and dual wall clearance probe
gas leak protection probe which
fits nicely in your hand facilitates
round off our wide range of flue
can be attached to the analyser
easy positioning. The probe has a
gas probes. The ambient CO
also has a user-defined alarm
quick connection function
probe as well as the CO2 probe
limit. A warning signal is triggered
ensuring that all gas paths are
are recognised automatically by
if the limit is exceeded. The probe
connected while at the same time
the analyser and the
is also recognised automatically
eliminating any confusion. The
measurement results immediately
and the results are shown in a
particle filter accommodated in
appear in their own graphically
trend graph in their own menu.
the handle blocks out particles
supported menu. Furthermore,
efficiently. Additionally, the probes
the user-defined alarm limit is
require little or no maintenance
coupled with an audible signal.
and are easy to clean. Different
Measurements can be carried out
lengths and diameters ensure a
in their entirety simultaneously
high flexibility level for all
with flue gas measurement. The
applications. When changing the
probe, the probe shaft is simply
attached to the probe handle and
clicks into place. The multi-hole
The new versatile infrared
printer
New upload and
configuration software
The new testo printer – cordless
When developing the software,
with infrared interface – saves
user-friendly operation and
data for printing. This saves time
handling had top priority.
since the analyser is ready for
Additionally, not only can data be
operation immediately after the
uploaded, it can also be
data is transmitted. The printer
conveniently filed and managed.
can be used anywhere.
It is also possible to control the
analyser via software when
measuring online.
At a glance
The set for the heating technician
Testo has put together a set for heating technicians in order to make selection
easier for you. Of course, it is always possible to add to and supplement the set
by selecting from the comprehensive range of accessories available.
· testo 330-3 flue gas analyser incl. rechargeable battery and calibration protocol
· 100-240 V mains unit for mains operation or recharging in analyser
· Modular flue gas probe, immersion depth: 300 mm, ø 8 mm
· CO multi-hole probe shaft suitable for the modular flue gas probe
· Combustion air temperature probe, 190 mm long
· Smoke tester incl. oil, soot sheets and reference scale
· Basic system case for analyser, probes and accessories
Part no.
0563 3316 70
testo 330-3 at a glance
testo 330-3 at a glance
Instrument diagnosis
Automatic menu selection
Additional features
• Automatic instrument leak test
• Recognition of connected probe
• BImSchV menu with manual CO switch-off
and qA mean value
• Rechargeable battery status display
• Sensor status display
• Monitoring of condensate trap level with
“FULL” message
• Display of pump power (l/min.)
• Display of error status incl. description and
Initialisation of pressure sensor for draught
measurement without removal of probe
• The probe can remain in the flue during
initialisation
Memory management
• Display of instrument temperature
• 400 measurement data records incl. object
address and system number
• Several measurement data records per
Additional measurement tasks
• ∆T measurement, flow/return
• Ambient CO measurement
• Ambient CO2 measurement
• Detection of gas leaks with gas leak
detection probe
• ∆P measurement for gas pressure
• Gas/oil flow rate measurement
• Graphics display
• Graphic hot spot detection
• Built-in impact protection with magnet,
recessed display
• Protection class IP40
diagnosis
• Display of last maintenance
• Operating hour counter
• Calculated parameter: Flue gas dew point
system number are possible
• Read in using barcode reader
• IRDA interface for data transfer to
PDA/Notebook
• USB interface for data transfer to PC
• ZIV driver for all conventional branch
software
• Rechargeable battery lifetime > 6 h with
pump operating
• Rechargeable battery separate and
rechargeable in instrument
• Adjustable alarm thresholds for ambient
CO/CO2 measurement
TÜV approval / EN standard
• Accuracy approved for O2, CO2, CO, NO,
NOlow, °C, hPa
• Approved measurement cell interchange
CO (adjustment without testing gas)
Technical Data
Temperature
Meas. range
Accuracy
Resolution
Draught measurement
Pressure measurement
O2 measurement
CO measurement
(H2 compensated)
Meas. range
Accuracy
Resolution
Meas. range
Accuracy
Resolution
Meas. range
Resolution
Accuracy
Adjustment time t90
-40 to 40 hPa
±0.02 hPa or ±0.5 % of mv (-0.50 to +0.60 hPa)
±0.03 hPa or ±0.5 % of mv (+0.61 to +3.00 hPa)
±1.5 % of mv (+3.01 to +40.00 hPa)
0.01 hPa
0 to 200 hPa
±0.5 hPa (0.0 to 50.0 hPa)
±1 % of mv (50.1 to 100.0 hPa)
±1.5 % of mv (remaining range)
0.1 hPa
0 to 21 Vol. %
0.1 Vol. %
±0.2 Vol. %
< 20 s
Adjustment time t90
Display range
Resolution
0 to 8000 ppm
1 ppm
±10 ppm or ±10% of mv (0 to 200 ppm)
±20 ppm or ±5% of mv (201 to 2000 ppm)
±10% of mv (2001 to 8000 ppm)
< 40 s
8000 to 30000 ppm (automatic dilution)
500 ppm
Efficiency (Eta)
Meas. range
Resolution
0 to 120 %
0.1 %
Flue gas loss
Meas. range
Resolution
0 to 99.9 %
0.1 %
CO2 measurement
Display range
Resolution
Accuracy
Measurement
Adjustment time t90
Meas. range
Resolution
Accuracy
0 to CO2 max
0.1 Vol. %
±0.2 Vol. %
Digital calculation from O2
< 40 s
0 to 300 ppm
0.1 ppm
±2 ppm (0.0 to 40.0 ppm)
±5% of mv (remaining range)
< 30 s
from 8000 ppm
Option: NOlow
measurement
Meas. range
Resolution
Accuracy
-40 to +1200 °C
±0,5 °C (0.0 to +100.0 °C)
±0,5 % of mv (remaining range)
0.1 °C (-40 to 999,9 °C)
1 °C (remaining range)
Adjustment time t90
Option: NO
measurement
Meas. range
Resolution
Accuracy
Ambient CO
measurement (with CO
probe)
Meas. range
Resolution
Accuracy
Adjustment time t90
Gas leak measurement
for combustible gases
(with gas leak detection
probe)
Adjustment time t90
Display range
Signal
Adjustment time t90
Ambient CO2
measurement (with
ambient CO2 probe)
Meas. range
General Technical Data
Memory
Weight
Dimensions
Storage temp.
Oper. temp.
Display
Power supply
Accuracy
Adjustment time t90
Warranty
IRDA printer
Printer type
Reception radius
Dimensions
Weight
Oper. temp.
Storage temp.
Power supply
*
*
*
*
0 to 3000 ppm
1 ppm
±5 ppm (0 to 100 ppm)
±5% of mv (101 to 2000 ppm)
±10% of mv (2001 to 3000 ppm)
< 30 s
0 to 500 ppm
1 ppm
±5 ppm (0 to 100 ppm)
±5% of mv (>100 ppm)
Approx. 35 s
0 to 10,000 ppm CH4 / C3H8
Optical display (LED)
audible display via buzzer
<2s
0 to 1 Vol. %
0 to 10000 ppm
±(50 ppm ±2 % of mv) (0 to 5000 ppm)
Approx. 35 s
400 Sites
600 g (without rechargeable battery)
270 x 90 x 65 mm
-20 to +50 °C
-5 to +45 °C
Graphics display: 160 x 240 pixel
Rechargeable block: 3.7 V / 2.2 Ah
Mains unit: 6 V / 1.2 A
Analyser/cells/probe 2 years, rechargeable battery 1 year
Infrared-controlled thermal printer, adjustable contrast,
graphics can be printed
Max. 2 m
186 x 91 x 61 mm
430 g incl. batteries
0 to +50 °C
-40 to +60 °C
4 round cell batteries, 1.5 V or NC rechargeable batteries
*the larger value applies
Fax Order Form
Part no.
Qty.
· testo 330-3 flue gas analyser incl. rechargeable battery and calibration
protocol
· 100-240 V mains unit for mains operation or recharging in analyser
· Modular flue gas probe, immersion depth: 300 mm, ø 8 mm
· CO multi-hole probe shaft suitable for the modular flue gas probe
· Combustion air temperature probe, 190 mm long
· Smoke tester incl. oil, soot sheets and reference scale
· Basic system case for analyser, probes and accessories
0563 3316 70
Modular flue gas probes, available in 2 lengths, incl. positioning cone,
NiCr-Ni thermocouple, 2.2 m hose and particle filter
Instrument / Options / Upgrades
testo 330-3 flue gas analyser incl. rechargeable battery and calibration
protocol
Option: Fine draught measurement, Resolution 0.1 Pa, measurement range to
100 Pa (instead of the standard draught measurement)
Qty.
Qty.
Qty.
Part no.
Probes
Part no.
Flue gas probe 180 mm long, Ø 8 mm, Tmax 500 °C
0600 9760
Flue gas probe 300 mm long, Ø 8 mm, Tmax 500 °C
0600 9761
Flue gas probe 180 mm long, Ø 6 mm, Tmax 500 °C
0600 9762
Flue gas probe 300 mm long, Ø 6 mm, Tmax 500 °C
0600 9763
Probe accessories
0632 3303
Probe shaft 180 mm long, Ø 8 mm, Tmax 500 °C
0554 9760
0440 3921
Probe shaft 180 mm long, Ø 6 mm, Tmax 500 °C
0554 9762
Probe shaft 300 mm long, Ø 8 mm, Tmax 500 °C
0554 9761
Option: Gas zeroing, Probe can remain in flue during initialisation
0440 3331
Option: NO meas. cell, meas. range 0 to 3000 ppm, 1 ppm resolution
On request
Upgrade: NO meas. cell, meas. range 0 to 3000 ppm, 1 ppm resolution
On request
Option: NOlow meas. cell, meas. range 0 to 300 ppm, 0.1 ppm resolution
0440 3931
Upgrade: NOlow meas. cell, meas. range 0 to 300 ppm, 0.1 ppm resolution
0554 3931
Probe shaft 300 mm long, Ø 6 mm, Tmax 500 °C
0554 9763
Probe shaft 300 mm, Ø 8 mm, Tmax 1000 °C
0554 8764
Probe shaft 700 mm long, Ø 8 mm, Tmax 1000 °C
0554 8765
Flexible probe shaft 330 mm long, Ø 10 mm, Tmax 500 °C
0554 9764
Multi-hole probe shaft, 300 mm long, Ø 8 mm, for mean CO calculation
0554 5762
Hose extension, 2.80 m, extension cable for probe and analyser
0554 1202
Accessories
Part no.
Cone 8 mm, steel, with grip clamp and grip, Tmax 500 °C
0554 3330
100-240 V mains unit for mains operation or batt. recharging in analyser
0554 1086
Cone 6 mm, steel, with spring clamp and grip, Tmax 500 °C
0554 3329
Spare rechargeable battery with recharging station
0554 1087
IRDA printer
0554 0547
Spare thermal paper for printer (6 rolls), permanent ink
0554 0568
Barcode reader to read in customer number on site
0554 0461
Barcode labels, self-adhesive (1200 off)
0554 0411
Adhesive pockets (50 off) for printout, paper barcode labels...
0554 0116
Additional probes
Dual wall clearance probe
0632 1260
Gas leak probe, available from December 2004
0632 3330
Ambient CO probe
0632 1247
Smoke tester with oil, soot sheet, for measuring soot in flue gas
0554 0307
Ambient CO2 probe
0632 1240
Hose connection set for separate gas pressure measurement
0554 1203
Connection cable
0440 0143
Differential temperature set consisting of 2 pipe clamp probes and adapter
0554 1204
Spare particle filter (10 off)
Combustion air temperature probe
0554 3385
Combustion air temperature probe, immersion depth 300 mm
0600 9791
testo 330 software with analysis and graphics functions, online measurement 0554 3332
Combustion air temperature probe, immersion depth 190 mm
0600 9787
USB connection cable, instrument to PC
0449 0047
Combustion air temperature probe, immersion depth 60 mm
0600 9797
ISO calibration certificate/Flue gas
0520 0003
Cases
Part no.
Basic system case for analyser, probes and accessories
0516 3330
Basic system case with two levels for analyser, probes and additional
accessories
0516 3331
Product system case with small plastic boxes without contents, can be
interlocked to basic system case
0516 0328
Tools system case with tools section without content, can be interlocked to
basic system case
0516 0329
Universal system case w/o pockets, can be interlocked to basic system case
0516 0331
Measurement case (leather) with drawers for instruments and accessories
0516 0303
Spare measurement cells
Part no.
Spare O2 measurement cell
0390 0092
Spare CO measurement cell (H2 compensated)
0390 0109
Spare NOlow measurement cell 0 to 300 ppm
0390 0094
Spare NO measurement cell 0 to 3000 ppm
On request
Additional temperature probes
Mini ambient air probe
0600 3692
Pipe clamp probe
0600 4593
Quick-action surface probe
0604 0194
Connection cable
0440 0143
Sender
Name
Subject to change without notice.
Qty.
testo 330-3 Set: Pro analysis for heating technicians
Company
Department
Address
City
Signature
0981 5654/hd/AC/A/07.2004
Qty.
Technical Data
Temperature
Meas. range
Accuracy
Resolution
Draught measurement
Pressure measurement
O2 measurement
Meas. range
Accuracy
Resolution
Meas. range
Accuracy
Resolution
Meas. range
Resolution
Accuracy
Adjustment time t90
-40 to +1200 °C
±0,5 °C (0.0 to +100.0 °C)
±0,5 % of mv (remaining range)
0.1 °C (-40 to 999,9 °C)
1 °C (remaining range)
-40 to 40 hPa
±0.02 hPa or ±0.5 % of mv (-0.50 to +0,60 hPa)
±0.03 hPa or ±0.5 % of mv (+0.61 to +3.00 hPa)
±1.5 % of mv (+3.01 to +40.00 hPa)
0.01 hPa
0 to 200 hPa
±0.5 hPa (0.0 to 50.0 hPa)
±1 % of mv (50.1 to 100.0 hPa)
±1.5 % of mv (remaining range)
0.1 hPa
0 to 21 Vol. %
0.1 Vol. %
±0.2 Vol. %
< 20 s
from 8000 ppm
Adjustment time t90
Display range
Resolution
0 to 8000 ppm
1 ppm
±10 ppm or ±10% of mv (0 to 200 ppm)
±20 ppm or ±5% of mv (201 to 2000 ppm)
±10% of mv (2001 to 8000 ppm)
< 40 s
8000 to 30000 ppm (automatic dilution)
500 ppm
Efficiency
(ETA)
Meas. range
Resolution
0 to 120 %
0.1 %
Flue gas loss
Meas. range
Resolution
0 to 99.9 %
0.1 %
CO2 measurement
Display range
Resolution
Accuracy
Measurement
Adjustment time t90
Meas. range
Resolution
Accuracy
0 to CO2 max
0.1 Vol. %
±0.2 Vol. %
Digital calculation from O2
< 40 s
0 to 300 ppm
0.1 ppm
±2 ppm (0.0 to 40.0 ppm)
±5% of mv (remaining range)
< 30 s
CO measurement
(H2 compensated)
Option: NOlow
measurement
Meas. range
Resolution
Accuracy
Adjustment time t90
Option: NO
measurement
Meas. range
Resolution
Accuracy
Ambient CO
measurement (with CO
probe)
Meas. range
Resolution
Accuracy
Adjustment time t90
Gas leak measurement
for combustible gases
(with gas leak detection
probe)
Adjustment time t90
Range of indication
Signal
Adjustment time t90
Ambient CO2
measurement (with
ambient CO2 probe)
Meas. range
General Technical Data
Memory
Weight
Dimensions
Storage temp.
Oper. temp.
Display
Power supply
Accuracy
Adjustment time t90
Warranty
IRDA printer
Printer type
Reception radius
Dimensions
Weight
Oper. temp.
Storage temp.
Power supply
*
*
*
*
0 to 3000 ppm
1 ppm
±5 ppm (0 to 100 ppm)
±5% of mv (101 to 2000 ppm)
±10% of mv (2001 to 3000 ppm)
< 30 s
0 to 500 ppm
1 ppm
±5 ppm (0 to 100 ppm)
±5% of mv (>100 ppm)
Approx. 35 s
0 ... 10,000 ppm CH4 / C3H8
Optical display (LED)
audible display via buzzer
<2s
0 to 1 Vol. %
0 to 10000 ppm
±(50 ppm ±2 % of mv) (0 to 5000 ppm)
Approx. 35 s
200 Sites
600 g (without rechargeable battery)
270 x 90 x 65 mm
-20 to +50 °C
-5 to +45 °C
Graphics display: 160 x 240 pixel
Rechargeable block: 3.7 V / 2.2 Ah
Mains unit: 6 V / 1.2 A
Analyser/cells/probe 2 years, rechargeable battery 1 year
Infrared-controlled thermal printer, adjustable contrast,
graphics can be printed
Max. 2 m
186 x 91 x 61 mm
430 g incl. batteries
0 to +50 °C
-40 to +60 °C
4 round cell batteries, 1.5 V or NC rechargeable batteries
*the larger value applies
Fax Order Form
Part no.
· testo 330-2 flue gas analyser incl. rechargeable battery and calibration pro- 0563 3315 70
tocol
· 100-240 V mains unit for mains operation or battery recharging in instrument
· Modular flue gas probe, immersion depth: 300 mm, ø 8 mm
· Combustion air temperature probe, 190 mm long
· Hose connection set for separate gas pressure measurement
· IRDA printer
· Smoke tester incl. oil, soot sheets and reference scale
· Basic system case for analyser, probes and accessories
Qty.
Instrument / Options / Upgrades
Part no.
Qty.
Qty.
Qty.
Qty.
0440 3330
Option: NO meas. cell, meas. range 0 to 3000 ppm, 1 ppm resolution
On request
Upgrade: NO meas. cell, meas. range 0 to 3000 ppm, 1 ppm resolution
On request
Option: NOlow meas. cell, meas. range 0 to 300 ppm, 0.1 ppm resolution
0440 3931
Upgrade: NOlow meas. cell, meas. range 0 to 300 ppm, 0.1 ppm resolution
0554 3931
Part no.
Flue gas probe 180 mm long, Ø 8 mm, Tmax 500 °C
0600 9760
Flue gas probe 300 mm long, Ø 8 mm, Tmax 500 °C
0600 9761
Flue gas probe 180 mm long, Ø 6 mm, Tmax 500 °C
0600 9762
Flue gas probe 300 mm long, Ø 6 mm, Tmax 500 °C
0600 9763
Probe accessories
testo 330-2 flue gas analyser incl. rechargeable battery and calibration proto- 0632 3302
col
Option: Fine draught measurement, Resolution 0.1 Pa, measurement range to 0440 3921
100 Pa (instead of the standard draught measurement)
Option: Zero initialisation, Probe can remain in flue during initialisation
Probes
Modular flue gas probes, available in 2 lengths, incl. positioning cone,
NiCr-Ni thermocouple, 2.2 m hose and particle filter
Probe shaft 180 mm long, Ø 8 mm, Tmax 500 °C
0554 9760
Probe shaft 180 mm long, Ø 6 mm, Tmax 500 °C
0554 9762
Probe shaft 300 mm long, Ø 8 mm, Tmax 500 °C
0554 9761
Probe shaft 300 mm long, Ø 6 mm, Tmax 500 °C
0554 9763
Probe shaft 300 mm, Ø 8 mm, Tmax 1000 °C
0554 8764
Probe shaft 700 mm long, Ø 8 mm, Tmax 1000 °C
0554 8765
Flexible probe shaft 330 mm long, Ø 10 mm, Tmax 500 °C
0554 9764
Multi-hole probe shaft, 300 mm long, Ø 8 mm, for mean CO calculation
0554 5762
Hose extension, 2.80 m, extension cable for probe and analyser
0554 1202
Accessories
Part no.
Cone 8 mm, steel, with grip clamp and grip, Tmax 500 °C
0554 3330
100-240 V mains unit for mains operation or batt. recharging in analyser
0554 1086
Cone 6 mm, steel, with spring clamp and grip, Tmax 500 °C
0554 3329
Spare rechargeable battery with recharging station
0554 1087
IRDA printer
0554 0547
Spare thermal paper for printer (6 rolls), permanent ink
0554 0568
Barcode reader to read in customer number on site
0554 0461
Barcode labels, self-adhesive (1200 off)
0554 0411
Adhesive pockets (50 off) for printout, paper barcode labels...
0554 0116
Additional probes
Dual wall clearance probe
0632 1260
Gas leak probe, available from December 2004
0632 3330
Ambient CO probe
0632 1247
Smoke tester with oil, soot sheet, for measuring soot in flue gas
0554 0307
Ambient CO2 probe
0632 1240
Hose connection set for separate gas pressure measurement
0554 1203
Connection cable
0440 0143
Differential temperature set consisting of 2 pipe clamp probes and adapter
0554 1204
Spare particle filter (10 off)
Combustion air temperature probe
0554 3385
Combustion air temperature probe, immersion depth 300 mm
0600 9791
testo 330 software with analysis and graphics functions, online measurement 0554 3332
Combustion air temperature probe, immersion depth 190 mm
0600 9787
USB connection cable, instrument to PC
0449 0047
Combustion air temperature probe, immersion depth 60 mm
0600 9797
ISO calibration certificate/Flue gas
0520 0003
Cases
Part no.
Basic system case for analyser, probes and accessories
0516 3330
Basic system case with reinforced base for analyser, probes and additional
accessories
0516 3331
Product system case with small plastic boxes without contents, can be interlocked to basic system case
0516 0328
Tools system case with tools section without content, can be interlocked to
basic system case
0516 0329
Universal system case w/o pockets, can be interlocked to basic system case
0516 0331
Measurement case (leather) with drawers for instruments and accessories
0516 0303
Spare measurement cells
Part no.
Spare O2 measurement cell
0390 0092
Spare CO measurement cell (H2 compensated)
0390 0109
Spare NOlow measurement cell 0 to 300 ppm
0390 0094
Spare NO measurement cell 0 to 3000 ppm
On request
Additional temperature probes
Mini ambient air probe
0600 3692
Pipe clamp probe
0600 4593
Quick-action surface probe
0604 0194
Connection cable
0440 0143
Sender
Name
Subject to change without notice.
testo 330-2 Set: Pro flue gas analysis: Customer service/technicians
Company
Department
Address
City
Signature
0981 5644/hd/AC/A/07.2004
Qty.