Supplementary Installation and Operating Instructions UFC 030

© KROHNE 09/2004
Supplementary
Installation and
Operating Instructions
HART Communicator 275
Field Communicator 375
Asset Management Solutions (AMS)
Process Device Manager (PDM)
UFC 030
Ultrasonic Flow Converter
KROHNE Messtechnik GmbH & Co. KG · Ludwig-Krohne-Str. 5 ‚ D-47058 Duisburg
Tel.: 0203-301 0 ‚ Fax: 0203-301 389 · e-mail: krohne@krohne.de
1/20
Supplementary Installation and Operating Instructions UFC 030 HART
HC275, FC375, AMS, PDM
1
General Information
3
2
IDs and Revision numbers
5
3
HART Communicator 275 (HC275), Field Communicator 375 (FC375)
6
4
5
3.1
Installation
6
3.2
Operating
6
Asset Management Solutions (AMS)
6
4.1
Installation
6
4.2
Operating
6
Process Device Manager (PDM)
7
5.1
Installation
7
5.2
Operating
7
KROHNE Messtechnik GmbH & Co. KG · Ludwig-Krohne-Str. 5 ‚ D-47058 Duisburg
Tel.: 0203-301 0 ‚ Fax: 0203-301 389 · e-mail: krohne@krohne.de
2/20
Supplementary Installation and Operating Instructions UFC 030 HART
HC275, FC375, AMS, PDM
1 General Information
The UFC 030 is a “four-wire” transmitter with 4...20mA current output and HART capability. Dependent on device
implementation it is available with active/passive current output (standard) or passive current output (MODIS).
General Characteristics of the UFC 030 HART interface:
• Multidrop Mode is supported
• Burst Mode is not supported
Electrical connection: Refer to section “Electrical connection of the signal inputs and outputs” of the following manual:
• “Installation and Operating Instructions Universal 3-Beam ultrasonic flowmeter UFM 3030 Compact
ultrasonic flowmeter UFC 030 Ultrasonic flow converter UFS 3000 Ultrasonic flow sensor” (KROHNE)
There are two ways of using the HART communication:
a) As a point-to-point connection between the UFC 030 and the HART master equipment. The instrument's current
output may be active or passive.
Point-to-Point Analog/Digital Mode
Primary Master
HART
Modem
Analog
HART
4
mA
Terminal I/C
20
Terminal ⊥
≥ 250 Ω
Power
Supply
4...20 mA
For slaves with
passive current
output (2 wire)
Secondary Master
KROHNE Messtechnik GmbH & Co. KG · Ludwig-Krohne-Str. 5 ‚ D-47058 Duisburg
Tel.: 0203-301 0 ‚ Fax: 0203-301 389 · e-mail: krohne@krohne.de
UFC 030
Addr. 0
3/20
Supplementary Installation and Operating Instructions UFC 030 HART
HC275, FC375, AMS, PDM
b) As a multipoint connection (multidrop) with up to 15 devices (UFC 030 or other HART equipment) in parallel. The
instrument's current outputs must be passive.
Multidrop Mode
Primary Master
HART
Modem
HART
Power
Supply
≥ 250 Ω
Terminal I/C
4 mA
4 mA
Terminal ⊥
...
Secondary Master
Up to 15
Slaves
...
UFC 030
Addr. > 0
UFC 030
Addr. > 0
(passive Current Output)
(passive Current Output)
KROHNE Messtechnik GmbH & Co. KG · Ludwig-Krohne-Str. 5 ‚ D-47058 Duisburg
Tel.: 0203-301 0 ‚ Fax: 0203-301 389 · e-mail: krohne@krohne.de
4/20
Supplementary Installation and Operating Instructions UFC 030 HART
HC275, FC375, AMS, PDM
In case the UFC 030's current output shall work continuously active a 'third wire' is needed to properly connect it
together with two-wire loop powered devices in the same network.
Multidrop Mode (‘three-wire’)
(Connecting two-wire and four-wire devices in the same network)
Primary Master
HART
4 mA
HART
Modem
4 mA
two-wire
loop powered
devices
...
Addr. > 0
Power
Supply
Addr. > 0
HART
Terminal V+
4 mA
≥ 250 Ω
4 mA
Terminal I/C
four-wire
active
source
devices
Secondary Master
UFC 030
Addr. > 0
...
UFC 030
Addr. > 0
2 IDs and Revision numbers
The HART Device Descriptions described in this document have the following IDs and revision numbers:
Manufacturer ID:
69 (0x45)
Device Type:
231 (0xE7)
Device Revision:
2
DD Revision:
1
HART Universal Revision: 5
HC 275 OS Revision:
≥ 4.9
FC 375 System SW Rev.: ≥ 1.6
AMS Version:
≥ 6.0
PDM Version:
≥ 5.2
FDT Version:
≥ 1.2
For information about Transmitter Revisions and related Device Descriptions refer to the KROHNE HART Device
List.
KROHNE Messtechnik GmbH & Co. KG · Ludwig-Krohne-Str. 5 ‚ D-47058 Duisburg
Tel.: 0203-301 0 ‚ Fax: 0203-301 389 · e-mail: krohne@krohne.de
5/20
Supplementary Installation and Operating Instructions UFC 030 HART
HC275, FC375, AMS, PDM
3 HART Communicator 275 (HC275), Field Communicator 375 (FC375)
3.1 Installation
The UFC 030 HART Device Description has to be installed on the HC275 and FC375 respectively. Otherwise the
user will work with the instrument as a generic one thus loosing opportunity for entire instrument control.
For installing DDs on the HC275 a ‘HART Communicator Module Programmer’ is needed (see details in the ‘Module
Programmer User’s Guide’).
For installing DDs on the FC375 the ‘Easy Upgrade Programming Utility’ is needed and the FC375 must have a
System Card with ‘Easy Upgrade’ option (see details in the ‘375 Field Communicator User’s Manual’).
3.2 Operating
Refer to the UFC 030 Menu Tree HC275 / FC375 (Attachment A).
The UFC 030 operation via HC275 / FC375 is made quite close to the manual instrument control via keypad.
Due to limitations of the HC275 / FC375 there are some peculiarities:
• The online help of each parameter is only a short form help. However it contains the function number as a
reference to the device’s local display and the “Installation and Operating Instructions” for a comprehensive
description.
• Some selection lists (e.g. for output functions) may contain items which are actually not valid for the device
concerned. However invalid settings are rejected when trying to send them to the device.
Parameter protection via passwords (Entry Code, Service Code) is the same as on local display. Please refer to the
online help for valid symbols according to device’s keypad.
The set of parameters of the HC275 “standard configuration” is only a partial set which doesn’t contain service
parameters. However the HC275 “full configuration” contains a complete set of parameters. Prior to sending a “full
configuration” to a device the Service Code protection must be disabled in case. Both types of configurations can be
transferred to AMS.
The FC375 always creates a “full” configuration for interaction with AMS. Still the FC375 considers only a partial
parameter set (like the HC275 “standard configuration”) when sending it to a device.
4 Asset Management Solutions (AMS)
4.1 Installation
If the UFC 030 Device Description is not already installed on the AMS System a so called Installation Kit UFC 030
HART AMS is needed (available as download from KROHNE ‘Download Centre’ on the internet or on floppy disk /
CD-ROM from KROHNE).
For installing the DD with the Installation Kit refer to the “AMS Intelligent Device Manager Books Online" section
"Basic AMS Functionality /Device Configurations / Installing Device Types / Procedures /Install device types from
media”. Please read also the “readme.txt”, which is also contained in the Installation Kit.
4.2 Operating
Refer to the UFC 030 Menu Tree AMS (Attachment B).
Due to AMS requirements and conventions the UFC 030 operation differs to some extent from operation with HC275
/ FC375 and via local keypad.
Due to limitations of the HC275 / FC375 which affect also AMS there are some peculiarities:
• The online help of each parameter is only a short form help. However it contains the function number as a
reference to the device’s local display and the “Installation and Operating Instructions” for a comprehensive
description.
• Some selection lists (e.g. for output functions) may contain items which are actually not valid for the device
concerned. However invalid settings are rejected when trying to send them to the device.
Parameter protection via passwords (Entry Code, Service Code) is the same as on local display. Please refer to the
online help for valid symbols according to device’s keypad.
KROHNE Messtechnik GmbH & Co. KG · Ludwig-Krohne-Str. 5 ‚ D-47058 Duisburg
Tel.: 0203-301 0 ‚ Fax: 0203-301 389 · e-mail: krohne@krohne.de
6/20
Supplementary Installation and Operating Instructions UFC 030 HART
HC275, FC375, AMS, PDM
5 Process Device Manager (PDM)
5.1 Installation
If the UFC 030 Device Description is not already installed on the PDM System a so called Device Install is needed
(available as download from KROHNE ‘Download Centre’ on the internet or on floppy disk / CD-ROM from
KROHNE).
For installing the DD with the Device Install refer to the “PDM Manual” section 11.2:”Device Install / Integrating
Devices in SIMATIC PDM with 'Device Install'". Please read also the “readme.txt”, which is also contained in the
Device Install.
5.2 Operating
Refer to the UFC 030 Menu Tree PDM (Attachment C).
Due to PDM requirements and conventions the UFC 030 operation differs to some extent from operation with HC275
/ FC375 and via local keypad.
Due to limitations of the HC275 / FC375 which affect also PDM there are some peculiarities:
• The online help of each parameter is only a short form help. However it contains the function number as a
reference to the device’s local display and the “Installation and Operating Instructions” for a comprehensive
description.
• Some selection lists (e.g. for output functions) may contain items which are actually not valid for the device
concerned. However invalid settings are rejected when trying to send them to the device.
Parameter protection via passwords (Entry Code, Service Code) is the same as on local display. Please refer to the
online help for valid symbols according to device’s keypad.
KROHNE Messtechnik GmbH & Co. KG · Ludwig-Krohne-Str. 5 ‚ D-47058 Duisburg
Tel.: 0203-301 0 ‚ Fax: 0203-301 389 · e-mail: krohne@krohne.de
7/20
Supplementary Installation and Operating Instructions UFC 030 HART
HC275, FC375, AMS, PDM
Attachment A
UFC 030 HART Menu Tree HC275 / FC375
1 Dynamic
Variables
1 View Process
Vars
1 Actual Flow
2 Corrected Flow Opt
3 VOS
4 Analog Input 1 Opt
5 Analog Input 2 Opt
6 F/R Indication
7 Gain
8 Digital Output Opt
9 Totalizer 1 Opt
10 Totalizer 2 Opt
11 Totalizer Sum Opt
1 Function I
2 Value I
3 %Range I
4 Function P Opt
5 Value P Opt
6 %Range P Opt
7 Analog Input 1 Opt
8 Analog Input 2 Opt
2 View In/Outputs
1 Test / Info
2 Configuration /
Test
1 Test Outputs
1 Test Current O.
2 Test Pulse O. Opt
2 Test Inputs
1 Analog Input 1 Opt
2 Analog Input 2 Opt
3 Device Info
1 Manufacturer Rd
2 Model Number Rd
3 Serial Number Rd
4 µP2 HW No. Rd
5 µP2 SW No. Rd
6 FE HW No. Rd
7 DSP HW No. Rd
8 DSP SW No. Rd
4 Sensor Limits
5 Simulation
3 Quit/Reset
4 HART Variables
2 Installation
2, 3
3 Service Menu
5
1 Volume Flow
2 Heat Flow
3 VOS
1 Upper Snsr Limit Rd
2 Lower Snsr Limit Rd
3 Minimum Span Rd
1 Simulation Mode
2 Uncorrected Flow
3 VOS
4 Gain
5 Analog Input 1 Opt
6 Analog Input 2 Opt
7 Digital Input Opt
4
Designations:
Opt Optional, dependent on device
implementation / configuration
Rd Read-only
Loc Local HC275/FC375,
affects only HC275/FC375 views
KROHNE UFC030 HA 45e70201 (1/5)
05/04
KROHNE Messtechnik GmbH & Co. KG · Ludwig-Krohne-Str. 5 ‚ D-47058 Duisburg
Tel.: 0203-301 0 ‚ Fax: 0203-301 389 · e-mail: krohne@krohne.de
8/20
Supplementary Installation and Operating Instructions UFC 030 HART
HC275, FC375, AMS, PDM
UFC 030 HART Menu Tree HC275 / FC375
1 Dynamic
Variables
1
1 Test / Info
2 Configuration /
Test
1
1 Flow Units
2 Full Scale
3 Zero Value
4 Zero Calibration Opt
5 Master TC
6 L.F.Cutoff
7 Meter Size Units Opt
8 Meter Size
9 GK Value
10 Flow Direction
11 VOS Min/Max
Units
12 Min VOS
13 Max VOS
1 Flow
2 Installation
1 Converter Function
2 Temperature
Units Opt
3 Input 1 4mA Opt
4 Input 1 20mA Opt
5 Pressure Units Opt
6 Input 2 4mA Opt
7 Input 2 20mA Opt
8 K0 Opt
9 K1 Opt
10 K2 Opt
11 Density 15 Opt
12 Heat Flow UnitsOpt
13 Full Scale Opt
2 Version
1 Display Flow
2 Function Totalizer
3 Display Totalizer
4 Volume Units Opt
5 Energy Units Opt
6 VOS Units
7 Cyclic Display
8 Error Messages
9 Date
10 Analog Input Opt
11 Signal Gain
12 Display Formats
3 Display
3 Service Menu
3 Quit/Reset
4 HART Variables
4
Designations:
Opt Optional, dependent on device
implementation / configuration
Rd Read-only
Loc Local HC275/FC375,
affects only HC275/FC375 views
5
4 Current Output I
5 Pulse Output P
6 Digital Input C
7 User Data
8 Logging
1 L.F.Cutoff
2 Cutoff on Opt
3 Cutoff off Opt
1 Format Flow Loc
2 Format
Heat Flow Loc, Opt
3 Format Volume Loc
4 Format
Energy Loc, Opt
3
KROHNE UFC030 HA 45e70201 (2/5)
05/04
KROHNE Messtechnik GmbH & Co. KG · Ludwig-Krohne-Str. 5 ‚ D-47058 Duisburg
Tel.: 0203-301 0 ‚ Fax: 0203-301 389 · e-mail: krohne@krohne.de
9/20
Supplementary Installation and Operating Instructions UFC 030 HART
HC275, FC375, AMS, PDM
UFC 030 HART Menu Tree HC275 / FC375
1 Dynamic
Variables
1
1 Test / Info
2 Configuration /
Test
1
2 Installation
1 Flow
2 Version
3 Display
4 Current Output I
5 Pulse Output P
6 User Data
3 Service Menu
3 Quit/Reset
4 HART Variables
1
1 Function I
2 Direction I
3 Range I
4 I 0%
5 I 100%
6 Limit
1 Function P Opt
2 Direction P Opt
3 Digital Output Opt
4 Trip.Point 1 Opt
5 Trip.Point 2 Opt
6 Time Constant Opt
7 Output Opt
8 Pulserate Units Opt
9 Pulsrate Opt
10 Units
Pulse/Unit Opt
11 Pulse/Unit Opt
12 Pulswidth Opt
1 Language
2 Entry Code 1
3 tag
4 Text Volume
5 Text Time
6 Unit Volume
7 Unit Time
8 Enable Reset
9 Error Limit
10 Counter
Decrement
11 Counter Limit
5
4
Designations:
Opt Optional, dependent on device
implementation / configuration
Rd Read-only
Loc Local HC275/FC375,
affects only HC275/FC375 views
KROHNE UFC030 HA 45e70201 (3/5)
05/04
KROHNE Messtechnik GmbH & Co. KG · Ludwig-Krohne-Str. 5 ‚ D-47058 Duisburg
Tel.: 0203-301 0 ‚ Fax: 0203-301 389 · e-mail: krohne@krohne.de
10/20
Supplementary Installation and Operating Instructions UFC 030 HART
HC275, FC375, AMS, PDM
UFC 030 HART Menu Tree HC275 / FC375
1 Dynamic
Variables
1
2 Configuration /
Test
1 Test / Info
1
2 Installation
2, 3
3 Service Menu
5
1 Quit Errors
2 Reset Totalizers Opt
3 Master Reset
4 Reset Conf. Flag
3 Quit/Reset
1 Communication
1 Poll addr
2 Num req preamsRd
3 Num resp preams
2 Device inform.
1 Manufacturer Rd
2 Model Rd
3 Dev id Rd
4 Tag
5 Date
6 Dev des
7 Message
8 Write protect Rd
9 Final asmbly num
10 Snsr s/n
11 Revision #'s
4 HART Variables
3 Dyn. Vars Settings
Designations:
Opt Optional, dependent on device
implementation / configuration
Rd Read-only
Loc Local HC275/FC375,
affects only HC275/FC375 views
1 Universal rev Rd
2 Fld dev rev Rd
3 Software rev Rd
4 Hardware rev Rd
1 PV is Rd
2 SV is
3 TV is
4 4V is
KROHNE UFC030 HA 45e70201 (4/5)
05/04
KROHNE Messtechnik GmbH & Co. KG · Ludwig-Krohne-Str. 5 ‚ D-47058 Duisburg
Tel.: 0203-301 0 ‚ Fax: 0203-301 389 · e-mail: krohne@krohne.de
11/20
Supplementary Installation and Operating Instructions UFC 030 HART
HC275, FC375, AMS, PDM
UFC 030 HART Menu Tree HC275 / FC375
1 Dynamic
Variables
1
1 Test / Info
2 Installation
2 Configuration /
Test
3 Service Menu
1
2, 3
1 Service Code
2 System Opt
3 Current Output I
Opt
4 Sensor Opt
5 Measuring Path Opt
6 An. Inp. Calib. Opt
7 Device
Information Opt
3 Quit/Reset
4 HART Variables
4
Designations:
Opt Optional, dependent on device
implementation / configuration
Rd Read-only
Loc Local HC275/FC375,
affects only HC275/FC375 views
8 Raw Values Opt
1 Reset Menu Par.
2 Reset Service Par.
3 Reset Calib. Par.
4 Reset All Par.
5 Reset Factory Par.
6 Meter Size Table
7 Version
1 Calibration 4 mA
2 Calibration 20 mA
3 Calibrate Current
Output I
1 Dead Time
2 Trigger Select
3 Trigger Level 1
4 Trigger Margin
5 Sensor Frequency
6 Zero Value Path 1
....
1 Select Path
2 Weight 1
...
5 Calibrate Path
6 Length 1
...
9 T. Window Start
10 T. Window End
1 Calibrate Analog
Input 1
2 Calibrate Analog
Input 2
3 Input 1 0mA
4 Input 1 20mA
5 Input 2 0mA
6 Input 2 20mA
1 Manufacturer
2 Model No.
3 Serial Number Rd
4 µP2 HW No.
5 Front end HW No.
6 DSP HW No.
7 Available I/O
1 Path 1 active
...
1 SI units
2 Pulse Output
3 Current Output
4 Digital Input
5 Analog IO module
6 HART Interface Rd
7 FF Interface Rd
8 PA Interface Rd
9 MODIS Version Rd
1 Rel. Flow Speed 1
...
4 SOS 1
...
7 Gain 1
...
KROHNE UFC030 HA 45e70201 (5/5)
05/04
KROHNE Messtechnik GmbH & Co. KG · Ludwig-Krohne-Str. 5 ‚ D-47058 Duisburg
Tel.: 0203-301 0 ‚ Fax: 0203-301 389 · e-mail: krohne@krohne.de
12/20
Supplementary Installation and Operating Instructions UFC 030 HART
HC275, FC375, AMS, PDM
Attachment B
UFC 030 HART Menu Tree AMS
Basic Setup
Sensor
• Zero Value
• Flow Direction
Process Input
• Flow Units
• Full Scale
• Master TC
• L.F. Cutoff
• Cutoff on Opt
• Cutoff off Opt
Correction
Current / Pulse Output
• Function I
• Range I
• Function P Opt
• Output Opt
Display
• Function Totalizer
• Display Totalizer
Correction
• Converter Function
Configure
Version
• Converter Function
• Temperature Units Opt
• Input 1 4mA Opt
• Input 1 20mA Opt
• Pressure Units Opt
• Input 2 4mA Opt
• Input 2 20mA Opt
• K0 Opt
• K1 Opt
• K2 Opt
• Density 15 Opt
• Heat Flow Units Opt
• Full Scale Opt
Compare
Clear Offline
----------------------------------Status/Conditions
Process Variables
3
Scan Device
Volume Flow Limits
• Upper Sensor Limit Rd
• Minimum Span Rd
----------------------------------Calibration Management
Diagnostics and Test
Calibrate
2
Reset
Paramter Protection
----------------------------------Rename
Unassign
Assign / Replace
----------------------------------Audit Trail
Record Manual Event
Drawings / Notes ...
Help ...
Sensor
Meter
• Meter Size Table Opt
• Meter Size Units Opt
• Meter Size
• GK Value
• Zero Value
• Flow Direction
Analog Input Calibration Opt
• Input 1 0mA
• Input 1 20mA
• Input 2 0mA
• Input 2 20mA
Sensor Opt
• Dead Time
• Trigger Select
• Trigger Level 1
• Trigger Margin
• Sensor Frequency
Heat Flow Limits Opt
• Upper Sensor Limit Rd
• Minimum Span Rd
Process Input
Current Output
Pulse Output
• Flow Units
• Full Scale
• Master TC
Current Output
• Function I
• Direction I
• Range I
• I 0%
• I 100%
• Limit
• Calibration 4mA Opt
• Calibration 20mA Opt
•
•
•
•
•
•
•
•
•
•
•
•
Local Display
• Language
• Cyclic Display
• Error Messages
• Date
• Text Volume
• Text Time
• Unit Volume
• Unit Time
Display Formats
• Format Flow Loc
• Format Heat Flow Opt, Loc
• Format Volume Loc
• Format Energy Opt, Loc
L.F. Cutoff
• L.F. Cutoff
• Cutoff on Opt
• Cutoff off Opt
• VOS Min/Max Units
• Min VOS
• Max VOS
Function P Opt
Direction P Opt
Digital Output Opt
Trip. Point 1 Opt
Trip. Point 2 Opt
Time Constant Opt
Output Opt
Pulserate Units Opt
Pulserate Opt
Units Pulse/Unit Opt
Pulse/Unit Opt
Pulsewidth Opt
Display
Designations:
Opt Optional, dependent on device
implementation / configuration
Rd Read-only
Loc Local AMS, affects only AMS views
•
•
•
•
•
•
•
•
Display Flow
Function Totalizer
Display Totalizer
Volume Units Opt
Energy Units Opt
VOS Units
Analog Input Opt
Signal Gain
KROHNE UFC 030 HA 45e70201 (1/3)
09/04
KROHNE Messtechnik GmbH & Co. KG · Ludwig-Krohne-Str. 5 ‚ D-47058 Duisburg
Tel.: 0203-301 0 ‚ Fax: 0203-301 389 · e-mail: krohne@krohne.de
13/20
Supplementary Installation and Operating Instructions UFC 030 HART
HC275, FC375, AMS, PDM
UFC 030 HART Menu Tree AMS
Measuring Path
• Path 1 active
...
• Weight 1
...
• Length 1
...
• Zero Value Path 1
Simulation
Plausibility Filter
• Error Limit
• Counter Decrement
• Counter Limit
•
•
•
•
•
•
Simulation Mode
Uncorrected Flow
VOS
Gain
Analog Input 1 Opt
Analog Input 2 Opt
• Version
• Enable Reset
Available I/O Opt
• SI Units Opt
• Pulse Output Opt
• Current Output Opt
• Digital Input Opt
• Analog IO module Opt
• HART Interface Opt, Rd
• FF Interface Opt, Rd
• PA Interface Opt, Rd
• MODIS Version Opt, Rd
• Time Window Start
• Time Window End
Device
Configure
Compare
Clear Offline
----------------------------------Status/Conditions
Process Variables
Scan Device
3
Device Info
• Manufacturer Rd
• Model Number Rd
• Serial Number Rd
• µP1 HW No. Rd
• µP2 SW No. Rd
• FE HW No. Rd
• DSP HW No. Rd
• DSP SW No. Rd
----------------------------------Calibration Management
Diagnostics and Test
Calibrate
Reset
Paramter Protection
-----------------------------------
HART
Revision Numbers
• Universal rev Rd
• Fld dev rev Rd
• Software rev Rd
• Hardware rev Rd
Preambles
• Num req preams Rd
• Num resp preams Rd
Rename
Unassign
Assign / Replace
Identification
• Manufacturer Rd
• Model Rd
• Dev ID Rd
• Poll Addr
• Tag
• Date
• Dev des
• Message
• Write Protect Rd
Final asmbly num
Snsr s/n
Dynamic Variables Settings
• PV is Rd
• SV is
• TV is
• 4V is
----------------------------------Audit Trail
Record Manual Event
Drawings / Notes ...
Test Current Output I
Test Pulse Output P Opt
Test Analog Inputs Opt
Help ...
Designations:
Opt Optional, dependent on device
implementation / configuration
Rd Read-only
Loc Local AMS, affects only AMS views
Quit Errors
Reset Totalizers Opt
Master Reset
Reset Conf. Changed Flag
Reset Menu Parameters Opt
Reset Service Param. Opt
Reset Calibration Par. Opt
Reset All Parameters Opt
Reset Factory Par. Opt
Zero Calibration Opt
Calibrate Current O. I Opt
Calibrate Analog Inp. 1 Opt
Calibrate Analog Inp. 2 Opt
Calibrate Path Opt
Entry Code 1
Service Code
KROHNE UFC 030 HA 45e70201 (2/3)
09/04
KROHNE Messtechnik GmbH & Co. KG · Ludwig-Krohne-Str. 5 ‚ D-47058 Duisburg
Tel.: 0203-301 0 ‚ Fax: 0203-301 389 · e-mail: krohne@krohne.de
14/20
Supplementary Installation and Operating Instructions UFC 030 HART
HC275, FC375, AMS, PDM
UFC 030 HART Menu Tree AMS
Overview
• Primary variable out of
limits
• Non-primary variable out
of limits
• Primary variable analog
output saturated
• Primary variable analog
output fixed
• Cold start
• Configuration changed
• Field device malfunction
1
•
•
•
•
•
•
•
Path 1 Signal Failure
…
Path 3 Signal Failure
Sensor open
Sensor short circuit
Unreliable
Empty Pipe Error
•
•
•
•
•
•
•
•
•
Signal Proc. Contr. Error
Peripheral Contr. Error
HW error frontend
Communication error
EEPROM R/W Error
Checks. err. service par.
Checks. err. menu par.
Simulation mode active
Restart
Configure
Compare
IO / Totalizers
Clear Offline
----------------------------------Status/Conditions
Process Variables
Scan Device
----------------------------------Calibration Management
Diagnostics and Test
2
Sensor / Device
Calibrate
Reset
•
•
•
•
•
•
•
•
•
•
Raw Values
Underrun analog input 1
Overrun analog input 1
Underrun analog input 2
Overrun analog input 2
20mA Analog Input HW
Error
Calibration Data Lost
Current Output Overrange
Current Output in fixed
mode
Pulse Output Overrange
Pulse Output in fixed
mode
• Totalizer Overrun Error
• Totalizer Storage Error
• PRimary flow Overrun
Error
• SECondary flow Overrun
Error
• Relative
Flow Speed 1 ... 3 Opt
Totalizers
• Totalizer 1 Opt
• Totalizer 2 Opt
• Totalizer Sum Opt
Current Output
• Value
• Percent of Range
• SOS 1 ... 3 Opt
• Gain 1 ... 3 Opt
Paramter Protection
----------------------------------Rename
Unassign
Assign / Replace
----------------------------------Audit Trail
Record Manual Event
Process Values
• Actual Flow
• Corrected Flow
• VOS
Opt
Correction
• Analog Input 1 Opt
• Analog Input 2 Opt
Pulse Output P
• Value Opt
• Percent of Range Opt
Drawings / Notes ...
Additional Values
• F/R Indication
• Gain
Help ...
Device
• Tag
• Descriptor Opt
HART
• Polling Address
• Device ID
Designations:
Opt Optional, dependent on device
implementation / configuration
Rd Read-only
Loc Local AMS, affects only AMS views
KROHNE UFC 030 HA 45e70201 (3/3)
09/04
KROHNE Messtechnik GmbH & Co. KG · Ludwig-Krohne-Str. 5 ‚ D-47058 Duisburg
Tel.: 0203-301 0 ‚ Fax: 0203-301 389 · e-mail: krohne@krohne.de
15/20
Supplementary Installation and Operating Instructions UFC 030 HART
HC275, FC375, AMS, PDM
Attachment C
UFC 030 HART Menu Tree PDM
Menu Bar
File
Communication Path
----------------------------------Load To Device
Load To PG/PC
----------------------------------Set Address
----------------------------------Simulation
Device
View
Help
2,3
Simulation
• Simulation Mode
• Uncorrected Flow
• VOS
• Gain
• Analog Input 1 Opt
• Analog Input 2 Opt
Test
Inputs Opt
• Analog Input 1
• Analog Input 2
Test
Options
Outputs
• Test Current Output I
• Test Pulse Output P Opt
----------------------------------Reset
• Reset
• <Quit Errors>
• <Reset Totalizers> Opt
• <Master Reset>
• <Reset Conf. Changed Fl.>
Reset
Reset Device Parameters Opt
• <Reset Menu Par.>
• <Reset Service Par.>
• <Reset Calibration Par.>
• <Reset All Par.>
• <Reset Factory Par.>
----------------------------------Calibration
Zero Calibration
• Zero Value 1 ... 3 Opt, Rd
• <Zero Calibration> Opt
Calibration
Current Output Calibr. Opt
• Calibration 4 mA Opt
• Calibration 20 mA Opt
• <Calibrate Current O.> Opt
Analog Input Calibr. Opt
• Input 1 0mA Opt
• Input 1 20mA Opt
• Input 2 0mA Opt
• Input 2 20mA Opt
• <Calibr. An. Inp. 1> Opt
• <Calibr. An. Inp. 2> Opt
Measuring Path Calibr. Opt
• <Calibrate Path> Opt
----------------------------------Parameter Protection
----------------------------------HART Communication
Parameter Protection
• Entry Code Lock Rd
• <Activcte/Deactivate Entry Code Lock>
• Service Code Lock Rd
• <Deactivate Service Code Lock>
HART Communication
Preambles
• Request Preambles Rd
• Response Preambles
Dynamic Variables Settings
• PV is Rd
• SV is
• TV is
• 4V is
Designations:
Opt Optional, dependent on device
implementation / configuration
Rd Read-only
Loc Local PDM, affects only PDM views
KROHNE UFC030 HA 45e80101 (1/5)
09.04
KROHNE Messtechnik GmbH & Co. KG · Ludwig-Krohne-Str. 5 ‚ D-47058 Duisburg
Tel.: 0203-301 0 ‚ Fax: 0203-301 389 · e-mail: krohne@krohne.de
16/20
Supplementary Installation and Operating Instructions UFC 030 HART
HC275, FC375, AMS, PDM
UFC 030 HART Menu Tree PDM
Menu Bar
1
File
Device
View
Measured Values
• Actual Flow
• Corrected Flow Opt
• VOS
• F/R Indication
• Gain
• Digital Output Opt
Display
Options
Help
Totalizer Opt
• Totalizer 1 Opt
• Totalizer 2 Opt
• Totalizer Sum Opt
Analog InputsOpt
• Analog Input 1
• Analog Input 2
Raw Values Opt
• Relative
Flow Speed 1 ... 3 Opt
• SOS 1 ... 3 Opt
• Gain 1 ... 3 Opt
• Device Status
Y t Diagram
• Actual Flow
• Corrected Flow Opt
• VOS
Yt Diagram
Current Outputs
• Function I
• Digital Value I
• Analog Value I
• Percent of Range
• Full Scale Rd
Outputs/Inputs
Device Status
Pulse Output Opt
• Function P
• Digital Value P Opt
• Analog Value P Opt
• Percent of Range Opt
• Full Scale Rd
Analog Inputs Opt
• Analog Input 1 Opt
• Analog Input 2 Opt
3
----------------------------------Toolbar
Status Bar
----------------------------------Update
Designations:
Opt Optional, dependent on device
implementation / configuration
Rd Read-only
Loc Local PDM, affects only PDM views
KROHNE UFC030 HA 45e80101 (2/5)
09.04
KROHNE Messtechnik GmbH & Co. KG · Ludwig-Krohne-Str. 5 ‚ D-47058 Duisburg
Tel.: 0203-301 0 ‚ Fax: 0203-301 389 · e-mail: krohne@krohne.de
17/20
Supplementary Installation and Operating Instructions UFC 030 HART
HC275, FC375, AMS, PDM
UFC 030 HART Menu Tree PDM
Menu Bar
1
File
Device
Display
Yt Diagram
Outputs/Inputs
2
View
Device
Device Info
• Manufacturer Rd
• Model Number Rd
• Serial Number Rd
• µP1 HW No. Rd
• µP2 SW No. Rd
• FE HW No. Rd
• DSP HW No. Rd
• DSP SW No. Rd
Options
Help
HART IDs
• Tag
• Manufacturer Rd
• Device Type Rd
• HART Device ID Rd
• Universal Revision Rd
• Device Revision Rd
• Software Revision Rd
• Hardware Revision Rd
• Date Rd
• Final Assembly No. Rd
• Sensor Serial No. Rd
Status
Device Status
Device Status
• Unreliable
• Empty Pipe Error
• Signal Proc. Contr. Error
• Peripheral Contr. Error
• HW error frontend
• Communication error
• EEPROM R/W Error
• Checks. err. service par.
• Checks. err. menu par.
• Simulation mode active
• Restart
Overview
• Primary variable out of
limits
• Non-primary variable out
of limits
• Analog Output outside the
operating range limits
• Analog Output in fixed
mode
• Cold start
• Configuration changed
• Field device malfunction
Sensor
• Sensor open
• Sensor short circuit
• Path 1 Signal Failure
• …
• Path 3 Signal Failure
I/O
• Calibration Data Lost
• 20mA Analog Input HW
Error
• Underrun analog input 1
• Overrun analog input 1
• Underrun analog input 2
• Overrun analog input 2
• Current Output Overrange
• Pulse Output Overrange
• Current Output in fixed
mode
• Pulse Output in fixed
mode
Totalizers
• Totalizer Overrun Error
• Totalizer Storage Error
• PRimary flow Overrun
Error
• SECondary flow Overrun
Error
----------------------------------Toolbar
Status Bar
----------------------------------Update
Designations:
Opt Optional, dependent on device
implementation / configuration
Rd Read-only
Loc Local PDM, affects only PDM views
KROHNE UFC030 HA 45e80101 (3/5)
09.04
KROHNE Messtechnik GmbH & Co. KG · Ludwig-Krohne-Str. 5 ‚ D-47058 Duisburg
Tel.: 0203-301 0 ‚ Fax: 0203-301 389 · e-mail: krohne@krohne.de
18/20
Supplementary Installation and Operating Instructions UFC 030 HART
HC275, FC375, AMS, PDM
UFC 030 HART Menu Tree PDM
Parameter Table
Identification
Identification
Operation Unit
Input
...
Process Connection
• Meter Size Table Opt
• Meter Size Units Opt
• Meter Size
Operation Unit
• Tag
• Descriptor
• Message
Device
----------------------------------Input
Device
Measuring Limits
•
•
•
•
•
•
•
•
•
•
•
Operation Parameter
Sensor
Version
Analog Input Opt
Measuring Path Opt
Plausibility Filter
----------------------------------Output
Manufacturer
Model Number Rd
Serial Number Rd
µP1 HW No. Rd
µP2 SW No. Rd
FE HW No. Rd
DSP HW No. Rd
DSP SW No. Rd
Version Rd
Meter Size Table Opt
Enable reset
Rd
Available I/O
• SI Units
• Pulse Output
• Current Output
• Digital Input
• Analog IO module
• HART Interface Rd
• FF Interface Rd
• PA Interface Rd
• MODIS Version Rd
• Time/log Module Rd
Opt
Current Output I
5
Pulse Output P Opt
----------------------------------Human Interface
Local Display
5
Display Formats
HART
• Manufacturer Rd
• Device Type Rd
• HART Device ID Rd
• Universal Revision Rd
• Field Device Revision Rd
• Software Revision Rd
• Hardware Revision Rd
• Date
• Final Assembly No.
• Sensor Serial No.
Input
Designations:
Opt Optional, dependent on device
implementation / configuration
Rd Read-only
Loc Local PDM, affects only PDM views
•
•
•
•
•
•
•
•
•
•
•
•
Flow Units
Full Scale
Zero Value
Master TC
GK Value
Flow Direction
VOS Min/Max Units
Min VOS
Max VOS
L.F. Cutoff
Cutoff on Opt
Cutoff off Opt
Measuring Limits
Volume Flow
• Upper Sensor Limit Rd
• Minimum Span Rd
Heat Flow
• Upper Sensor Limit Rd
• Minimum Span Rd
VOS
• Upper Sensor Limit Rd
• Lower Sensor Limit Rd
• Minimum Span Rd
Operation Parameter
Sensor Opt
• Dead Time
• Trigger Select
• Trigger Level 1
• Trigger Margin
• Sensor Frequency
Version Opt
• Converter Function
• Temperature Units Opt
• Input 1 4mA Opt
• Input 1 20mA Opt
• Pressure Units Opt
• Input 2 4mA Opt
• Input 2 20mA Opt
• K0 Opt
• K1 Opt
• K2 Opt
• Density 15 Opt
• Heat Flow Units Opt
• Full Scale Opt
Analog Input Calibration Opt
• Input 1 0mA
• Input 1 20mA
• Input 2 0mA Opt
• Input 2 20mA Opt
Measuring Path Opt
• Path 1 active ...
• Weight 1 ...
• Length 1 ...
• Zero Value 1 ...
• Time Window Start
• Time Window End
Plausibility Filter
• Error Limit
• Counter Decrement
• Counter Limit
KROHNE UFC030 HA 45e80101 (4/5)
09.04
KROHNE Messtechnik GmbH & Co. KG · Ludwig-Krohne-Str. 5 ‚ D-47058 Duisburg
Tel.: 0203-301 0 ‚ Fax: 0203-301 389 · e-mail: krohne@krohne.de
19/20
Supplementary Installation and Operating Instructions UFC 030 HART
HC275, FC375, AMS, PDM
UFC 030 HART Menu Tree PDM
Parameter Table
Output
• IO Fitted Rd
Identification
Operation Unit
Device
4
----------------------------------Input
Measuring Limits
Operation Parameter
Sensor
Version
Analog Input Opt
4
Measuring Path Opt
Plausibility Filter
----------------------------------Output
Current Output I
Pulse Output P Opt
----------------------------------Human Interface
Local Display
Display Formats
Current Output I
• Function I
• Direction I
• Range I
• I 0%
• I 100%
• Limit
• Calibration 4mA Opt
• Calibration 20mA Opt
Pulse Output P Opt
• Function P
• Direction P
• Digital Output Opt
• Trip. Point 1 Opt
• Trip. Point 2 Opt
• Time Constant
• Output Opt
• Pulsrate Units Opt
• Pulserate Opt
• Units Pulse/Unit Opt
• Pulse/Unit Opt
• Pulsewidth Opt
Human Interface
• Display Flow
• Function Totalizer
• Display Totalizer Opt
• Volume Units Opt
• Energy Units Opt
• VOS Units
• Analog Input Opt
• Signal Gain
Local Display
• Language
• Cyclic Display
• Error Messages
• Date Opt
• Text Volume
• Text Time
• Unit Volume
• Unit Time
Display Formats
• Format Flow Loc
• Format Heat Flow Opt, Loc
• Format Volume Loc
• Format Energy Opt, Loc
Designations:
Opt Optional, dependent on device
implementation / configuration
Rd Read-only
Loc Local PDM, affects only PDM views
KROHNE UFC030 HA 45e80101 (5/5)
09.04
KROHNE Messtechnik GmbH & Co. KG · Ludwig-Krohne-Str. 5 ‚ D-47058 Duisburg
Tel.: 0203-301 0 ‚ Fax: 0203-301 389 · e-mail: krohne@krohne.de
20/20