Reader
Manual
www.novassay.fr
®
Manual Uranalyze Reader
1. Introduction ............................................................................................. 3
1.1 Technical description ...................................................................... 3
1.2 Technical specifications .................................................................. 4
2. Safety precautions .................................................................................. 5
2.1 Protective clothing .......................................................................... 5
2.2 Power connection ........................................................................... 5
2.3 Housing damage ............................................................................. 5
2.4 Cable break .................................................................................... 5
2.5 Improper use................................................................................... 5
2.6 Transport ........................................................................................ 5
3. Instrument set-up .................................................................................... 6
3.1 Set-up location ................................................................................ 6
3.2 Package content ............................................................................. 7
4. Quickstart ................................................................................................ 8
5. Initial operation ....................................................................................... 8
5.1 Power connection ........................................................................... 9
5.2 Interface ........................................................................................ 10
5.2.1 Serial interface ..................................................................... 10
5.2.2 USB interface ....................................................................... 11
5.2.3 Transmission protocol .......................................................... 11
5.2.4 Barcode scanner/Keyboard.................................................. 11
5.3 Inserting printer paper ................................................................... 11
5.4 Switching on the instrument .......................................................... 12
6. Operation and user guidance ............................................................... 13
6.1 Touch screen ................................................................................ 13
6.2 Operating the user interface ......................................................... 13
6.2.1 Activity buttons ..................................................................... 13
6.2.2 Lists ..................................................................................... 14
6.2.3 Option fields ......................................................................... 14
6.2.4 Input menu ........................................................................... 15
6.3 Stand-by mode ............................................................................. 15
7. Start screen ........................................................................................... 16
8. Measurement preparation .................................................................... 17
9. Measurements ....................................................................................... 18
9.1 Sequence number ........................................................................ 18
9.2 Sample ID ..................................................................................... 18
9.3 Measurement procedure ............................................................... 19
10. Main menu ........................................................................................... 21
1
Manual Uranalyze® Reader EN v1.0 / 06.15
®
Manual Uranalyze Reader
11. Memory ................................................................................................ 22
11.1 Memory list ................................................................................. 22
11.2 Search ........................................................................................ 23
11.2.1 Day .................................................................................... 24
11.2.2 Parameter .......................................................................... 25
11.2.3 Process search results ....................................................... 26
11.3 Delete ......................................................................................... 26
12. Check mode ......................................................................................... 27
13. Settings ................................................................................................ 28
13.1 Test strips ................................................................................... 28
13.1.1 Confidence interval ............................................................ 29
13.1.2 Correction .......................................................................... 30
13.2 Password .................................................................................... 32
13.3 Printer ......................................................................................... 32
13.4 Sound ......................................................................................... 33
13.5 Autostart ..................................................................................... 33
13.6 Battery ........................................................................................ 34
13.7 Language .................................................................................... 35
13.8 Date/Time ................................................................................... 35
13.9 Interface ...................................................................................... 36
13.10 Customization ........................................................................... 37
13.11 Reaction time ............................................................................ 38
13.12 Print settings ............................................................................. 38
14. Connecting external devices .............................................................. 39
14.1 Barcode scanner and keyboard .................................................. 39
14.2 Computer .................................................................................... 39
15. Quality control ..................................................................................... 40
16. Cleaning and maintenance ................................................................. 41
16.1 Daily cleaning ............................................................................. 41
16.2 Weekly cleaning .......................................................................... 41
17. Error messages ................................................................................... 42
18. Service ................................................................................................. 43
18.1 Spare parts and accessories ...................................................... 43
18.2 Disposal ...................................................................................... 43
18.3 Warranty ..................................................................................... 43
18.4 General usage comments ........................................................... 44
18.5 Contact ....................................................................................... 44
2
Manual Uranalyze® Reader EN v1.0 / 06.15
®
Manual Uranalyze Reader
1. Introduction
®
Welcome and thank you very much for choosing our Uranalyze Reader
analysis.
®
®
The Uranalyze Reader is a reflexion photometer for the analysis of Uranalyze
test strips from NOVASSAY. The measurements are carried out under
standardized conditions. Results may be displayed, printed and sent to a
computer. Instrumental evaluation eliminates potential sources of error inherent
to visual evaluation. Results will not vary due to subjective color perception or
different light sources. Additionally, errors resulting from manual result
administration are avoided. Featuring a clear touch screen and intuitive user
®
guidance, the Uranalyze Reader is ideal for quick, efficient and comfortable
work.
1.1 Technical description
The test strips is placed on a measuring slide with inbuilt reference pad, which
is moved into the instrument along a fixed measurement head. While moving in
and out of the instrument, the test strip is measured reflectometrically. A predefined light source illuminates the test pad. A detection unit then measures the
intensity of the reflected light at three different wavelengths. From this
information, remission values are calculated. The timing of the measurements
may vary depending on test strip and parameter. Using the remission values,
the instrument can, in contrast to visual evaluation, show continuous values
within the respective tests’ measuring range. Once the result has been
determined, it is displayed on the screen, printed and sent to the instrument’s
interface.
3
Manual Uranalyze® Reader EN v1.0 / 06.15
®
Manual Uranalyze Reader
1.2 Technical specifications
Overview
Mesurement
Operation
Interfaces
Memory
Accessories
(optional)
Dimensions: 7.5 x 16 x 20 cm
Weight: 710 g (without batteries and power pack)
Power supply: mains transformer (100–240 V Input
Output 9 V = 1,5 A alternativ)
Battery operation with 6 mignon batteries 1,5 V (AA)
Principle: reflection photometer
Light source: 1 white LED
Detector: RGB-detector
Touch screen: W x H: 57 x 57 mm
Printer: thermo transfer
1 x USB
1 x RS232 19200 Baud, 8 bit, N unidirectional
1 x PS/2 for keyboard/barcode scanner
Capacity: 200 results
PC keyboard: data entry
Barcode scanner: data entry
Serial cable: PC connection
4
Manual Uranalyze® Reader EN v1.0 / 06.15
®
Manual Uranalyze Reader
2. Safety precautions
2.1 Protective clothing
®
Using the Uranalyze Reader may entail handling chemicals and/or dangerous
substances. In such cases, adhere to the safety measures provided within the
test strip instructions, the in-house regulations of your place-of-work as well as
to good laboratory practice. If in doubt, we recommend to wear at least gloves,
safety goggles and a lab coat.
2.2 Power connection
To ensure safe functioning of the instrument, only use the included power pack.
2.3 Housing damage
If the housing is damaged, the instrument needs to be repaired. Correct
functioning and measurement results cannot be guaranteed anymore.
2.4 Cable break
In case of a broken cable, immediately stop using the instrument and replace
the cable and power pack.
2.5 Improper use
®
®
The Uranalyze Reader instrument is solely designed to read Uranalyze test
strips. Any other usage of the instrument is neither tested nor approved. In the
event of improper use, any claim with regards to warranty and liability are void
and of no effect.
2.6 Transport
To transport the instrument, the cardboard box from the initial shipment
including the protective foamed plastic is well suited. Therefore, we recommend
to keep the box. In case you do not have the original box anymore, use a large
cardboard box and plenty of soft cushioning material, so that the instrument
cannot shift during transport.
5
Manual Uranalyze® Reader EN v1.0 / 06.15
®
Manual Uranalyze Reader
3. Instrument set-up
3.1 Set-up location
Only use the instrument in a suitable location. It should be positioned on a dry,
clean, level, hard and horizontal surface. The operating temperature is between
5–40 °C, while the optimal temperature range for the instrument is between 18–
25 °C. For optimal performance, the humidity should be between 20–80 %.
Make sure to protect the instrument from drastic changes in temperature. Avoid
positioning the instrument close to windows, hot plates, ovens etc. To ensure
correct functioning and reliable results, the instrument should not be exposed to
direct sunlight or other bright, focused light sources.
If the instrument underwent drastic temperature changes (e.g. during transport
of shipment), it should not be operated prior to sufficient acclimatization. Also,
we recommend not to use the instrument close to electrical fields (e.g. micro
wave ovens, radio units etc.), as it may affect measurements.
6
Manual Uranalyze® Reader EN v1.0 / 06.15
®
Manual Uranalyze Reader
3.2 Package content
Carefully take the instrument and all other parts out of the box. Check the
package, instrument and all accessories for visible damage. If a part is
damaged, please contact your distributor or the technical support. The following
picture shows the complete package content (Picture 1).
Check the shipment for completeness. If a part is missing, please contact your
distributor or the technical support as well (contact@novassay.fr).
Note: Keep the original box and the packaging material from the initial shipment
to optimally protect the instrument in case of a return-shipment.
Picture 1 : Package content
®
1- Uranalyze Reader reflexion photometer
2- Power pack 100-240 V, 47/63 Hz, 9V + adapters
3- Printer paper
4- Manual
7
Manual Uranalyze® Reader EN v1.0 / 06.15
®
Manual Uranalyze Reader
4. Quickstart
To start measurements right away, unpack the instrument and place it on a
level, hard surface. Connect the power pack and turn on the instrument using
the on/off switch (Picture 3.8). Once the instrument has performed a self-test,
the start display appears.
Use the test strip as described in the instructions. Then place the strip on the
measurement slide. The instrument recognizes the test strip and starts the
measurement automatically. A progress bar in the display shows the remaining
reaction time. The result is displayed right after the measurement has been
completed.
5. Initial operation
This chapter contains a description of the steps necessary to start-up the
instrument (Picture 2 and Picture 3).
Picture 2 : Instrument front
1- Touch screen
2- Measurement slide
3- Printer
8
Manual Uranalyze® Reader EN v1.0 / 06.15
®
Manual Uranalyze Reader
5.1 Power connection
The shipment includes four adaptors to connect the power pack to different
outlets. Clip the respective adaptor to the power pack. Once the power cord is
®
connected to the "DC IN" port (7) and connected to an outlet, the Uranalyze
Reader is connected to the power supply.
Picture 3 : Instrument back
4- RS232 Interface
7- Power connection
5- USB interface
8- On / Off switch
6- PS/2 port
®
Alternatively, the Uranalyze Reader instrument can run on standard AA
batteries independent from the power network. The battery compartment is on
the bottom side of the instrument (Picture 4). Pay attention to the pre-defined
polarity when inserting the batteries (+/-).
Picture 4 : Instrument bottom
9
Manual Uranalyze® Reader EN v1.0 / 06.15
®
Manual Uranalyze Reader
5.2 Interface
The instrument can be connected to a computer using either a serial or USB
connection.
5.2.1 Serial interface
Protocol RS232 (Picture 3.4), 19 200 baud, 8 bit, no parity, connection plug
arrangement:
Pin No.
1
2
Signal
Nc
3
4
5
6
7
8
9
TxD
Nc
GND
Nc
Nc
Nc
Nc
RxD
Description
Sens
Not wired
Data reception
Input
Senden
Output
Not wired
Signal ground
Not wired
Not wired
Not wired
Not wired
--
10
Manual Uranalyze® Reader EN v1.0 / 06.15
®
Manual Uranalyze Reader
5.2.2 USB interface
USB port type B (Picture 3.5). Prevalent operating systems recognize the
instrument as a serial interface. The driver for the interface component is
available on the NOVASSAY website (www.novassay.fr).
5.2.3 Transmission protocol
Data is sent via the interfaces in clear text. The received data relates to format
of the print out.
5.2.4 Barcode scanner/Keyboard
You can connect a keyboard and/or barcode scanner using the instruments
PS/2 port (Picture 3.6).
5.3 Inserting printer paper
To open the printer paper cover, press the little button above the touch screen.
Move the printer paper cover backwards (Picture 5).
Picture 5 : Open printer paper cover (left) and Opened printer paper cover (right)
11
Manual Uranalyze® Reader EN v1.0 / 06.15
®
Manual Uranalyze Reader
Unroll appr. 10 cm (4ʺ) of printer paper and place the printer paper into the
chamber, so that the paper unrolls downwards and towards the touch screen
(Picture 6).
Picture 6 : Insert printer paper
Close the printer paper cover with an audible click, so that the printer paper
comes out from the chamber (Picture 7).
Picture 7 : Inserted printer paper
Note: Only use special thermal printer paper.
5.4 Switching on the instrument
You can find the On/Off switch (Picture 3.8) on the back of the instrument. Turn
on the instrument. Right after switching on, the instrument performs a self-test.
The touch screen lights up and the start screen appears.
12
Manual Uranalyze® Reader EN v1.0 / 06.15
®
Manual Uranalyze Reader
6. Operation and user guidance
The instrument is designed for easy and intuitive use. Nevertheless, read the
following paragraphs carefully to familiarize yourself with the instrument’s
functions and user guidance.
6.1 Touch screen
All operations are carried out with a touch-sensitive screen. An external keyboard can be connected via the PS/2 interface (see 5.2.4 Barcode
scanner/Keyboard).
Note: Do not touch the touch screen with pointed or hard items.
6.2 Operating the user interface
Please note some basic rules for operating the instrument’s user interface:
• All framed buttons can be activated by touch.
• To confirm changes or entries, you have to press
when leaving the
menu. If you leave menus pressing
, none of your changes or entries will
be adopted.
There are three different categories of buttons.
6.2.1 Activity buttons
Framed activity buttons with a symbol or text initiate a certain activity. Among
others, the following symbols are frequently used:
– Leave the menu / cancel
– Confirm an entry
– Prints display
– Scrolls in a list up or down
13
Manual Uranalyze® Reader EN v1.0 / 06.15
®
Manual Uranalyze Reader
6.2.2 Lists
Within a list, the currently activated entry is highlighted. Only one entry can be
activated within a list at a time (Picture 8). Using the arrow buttons (see 6.2.1
Activity buttons); you can scroll through the list and activate the different entries.
Confirm your selection pressing
. Press
to leave the list.
Strip
Password
Printer
Sound
Autostart
Battery
Language
Date/Time
Interface
Picture 8 : Lists
6.2.3 Option fields
Option fields consist of choice buttons (Picture 9). Choice buttons are circles. If
a choice button is active, it is filled black. Within an option field, only one choice
button can be active.
 Today
 All
 Date
Picture 9 : Option fields
Touch a choice button to activate it. Press
to confirm and save your
changes. If you leave menus that include option fields by pressing
, none
of your choices will be adopted.
14
Manual Uranalyze® Reader EN v1.0 / 06.15
®
Manual Uranalyze Reader
6.2.4 Input menu
You have the option to insert some data into the instrument (e.g. Date/Time).
The input menu appears whenever you want to make such an entry (Picture
10).
1
2
3
4
5
6
7
8
9
ABC
0
Picture 10 : Input menu
Above the input buttons you will find a window which shows your new entry.
You can enter upper case letters or numbers. To switch between the options,
press the button that shows the desired form an entry ( ABC / 123 ).
Initially, the instrument is opting for capital letters. Once you make a change,
the instrument remembers your last choice. If you open the input menu again,
the former choice is still active. For new entries touch a letter/number. To enter
a different letter / number than the first one on the button, press the button
repeatedly.
You can delete an entry by pressing
. Press
to confirm and save your
changes. None of your changes will be adopted, if you leave the input menu by
pressing
.
6.3 Stand-by mode
After three minutes idling, the instrument goes into stand-by mode and the
measurement slide retracts into the housing. Pressing the touch screen reactivates the instrument and you get straight to the start screen. You can also
activate the stand-by mode ahead of time by pressing (P) in the start screen.
Note: If the device is activated from the stand-by mode, it will always go to the
start screen. Any changes in menus or changes in settings that have not been
saved before the instrument went into stand-by mode will be lost.
15
Manual Uranalyze® Reader EN v1.0 / 06.15
®
Manual Uranalyze Reader
7. Start screen
Once the instrument has been switched on and performed the self-test, the start
screen appears (Display 1).
Uranalyze
® Reader
Uranalyze
Test:
SN:
Uranium
0001
ID:
(P)
Display 1: Start screen
The start screen is the starting point for measurements and all other settings. It
shows the currently activated test strip, the sequence number and sample ID.
(P) activates the stand-by mode. To get to the main menu, press
. A new
menu with multiple options appears.
16
Manual Uranalyze® Reader EN v1.0 / 06.15
®
Manual Uranalyze Reader
8. Measurement preparation
The instructions of the test strips describe the necessary steps prior to
measurement. Therefore, please read the package insert carefully to obtain
accurate and reliable results.
Note: as described in the instructions, always shake off excess liquid from the
test strip to get optimal results.
Note: Reaction times are pre-programmed into the instrument for every test.
Therefore, put the test strip on the measurement slide immediately after
immersing the strip into the sample and shaking of the excess liquid.
17
Manual Uranalyze® Reader EN v1.0 / 06.15
®
Manual Uranalyze Reader
9. Measurements
The instrument is designed for the quick and easy measurement of test strips.
The following points outline set-up options with regards to measurements as
well as the actual measurement procedure
9.1 Sequence number
SN: shows a continuous number for every measurement for a given day. The
counter starts at 0001. To change the sequence number, press SN: . The input
menu appears and you can enter the desired number.
For every new day, the sequence number is set back to 0001.
9.2 Sample ID
ID: shows the sample ID for the next measurement. To enter or change the
sample ID, press ID: . The input menu appears and you can enter the desired
sample ID. Once you enter the sample ID, 〔 〕appears. You can lock the
sample ID by pressing〔 〕, 〔x〕appears.
The following measurements are then always run and saved using the same
sample ID. Using the search mode, you can retrieve the results from the
instrument memory (see 11.2 Search). Pressing 〔x〕 twice removes the lock.
You can also enter the sample ID using a barcode scanner or a keyboard (see
5.2.4 Barcode scanner / Keyboard).
Note: You have to enter both sequence number and sample ID before you
begin the measurement. Changing or adding a sequence number and/or
sample ID to a result after the measurement is not possible.
18
Manual Uranalyze® Reader EN v1.0 / 06.15
®
Manual Uranalyze Reader
9.3 Measurement procedure
®
Measuring test strips with Uranalyze Reader is very easy. To start a
measurement, simply place the test strip on the test strip slide with the test pad
facing upwards and forward. The end of the test strip has to touch the end of
the slide’s recess (Picture 11).
A sensor (little gap within the slide) detects the test strips and starts the
measurement automatically.
Picture 11 : Position strip (left) and Push strip forward (right)
Note: It is important to place the test strip at the far end of the slide’s recess.
You can easily move the strip forward on the recess using your finger (make
sure to not touch the test field).
19
Manual Uranalyze® Reader EN v1.0 / 06.15
®
Manual Uranalyze Reader
A progress bar appears showing the remaining reaction time. After a few
seconds, the test strip is drawn into the instrument’s housing. After the
measurement, the slide moves out of the housing again and the result is
displayed in the screen. The measured test strip can be discarded into the
domestic waste.
Measurement’s result appears on the screen (Display 2).
RESULT
Seq. No: 0001
ID:
2015.01.20 9:47
Uranium
64 mg/L
Display 2 : Result
Note: The reaction time is dependent on the test and programmed into the
instrument. Therefore, different tests have different waiting times. In general
though, the time to wait is hardly ever longer than 60 seconds.
Note: If the autostart function is disabled, you have to start a measurement
pressing
(see 13.5 Autostart).
20
Manual Uranalyze® Reader EN v1.0 / 06.15
®
Manual Uranalyze Reader
10. Main menu
Once you press
in the start screen, the main menu appears. Within the
main menu are three options (Display 3).
allows you to display, search and
delete stored data.
enables quality control measurements with special Novassay control
strips.
The button
gets you to the instrument settings.
MENU
Memory
Check mode
Settings
Display 3 : Main menu
21
Manual Uranalyze® Reader EN v1.0 / 06.15
®
Manual Uranalyze Reader
11. Memory
®
The Uranalyze Reader saves up to 200 results in an internal memory. If the
memory is completely full, a new data set will replace the oldest data set in the
memory. Press
in the start screen and
in the main menu to get to the
memory menu (Display 4).
MEMORY MODE
List
Search
Clear
Display 4 : Memory
Within the memory menu, you have three different options. You can display a
complete list with all data sets, search for specific data sets or delete the entire
memory.
11.1 Memory list
Press
to display all data sets in the memory. A new screen appears,
showing the result for the latest measurement. The results are always sorted
according to the measurement date. The heading shows the number of the
currently displayed data sets and the total number of data sets within the
memory
22
Manual Uranalyze® Reader EN v1.0 / 06.15
®
Manual Uranalyze Reader
Memory 1/63
Seq. No: 0001
ID:
2015.01.20 9:47
Uranium
48 mg/L
Display 5: Memory entry
For the respective result, sequence number, sample ID, measurement
date/time, test and the actual result are displayed. On the bottom side of the
screen are 5 activity buttons.
Press
and
to choose between different data sets. Additionally, you may
press
Press
to print the displayed data set or press
to send the data to a PC.
to get to the memory menu.
11.2 Search
Press
to search the memory for specific results. To do so, you can freely
combine two search criteria (date, and parameter). To start a search, all criteria
need to be defined.
23
Manual Uranalyze® Reader EN v1.0 / 06.15
®
Manual Uranalyze Reader
11.2.1 Day
Press DAY to select results from a particular day. Once you press the activity
button, an option field with three choice buttons appears (Display 6).
If you press "Today" only results from the current day will be searched and
selected.
The choice button "All" selects all data sets, irrespective of the date.
The choice button "Date" allows you to select a specific date on which the result
you are looking for was measured. Press "Date" and confirm with
.
A list with all dates on which measurements have been performed appears. Use
the arrow buttons to scroll through the list and press
if you have found
the desired date. Your selection will appear in the "Search" menu. If you leave
the list by pressing
, your selection is not adopted.
DAY
 Today
 All
 Date
Display 6 : Date selection
24
Manual Uranalyze® Reader EN v1.0 / 06.15
®
Manual Uranalyze Reader
11.2.2 Parameter
Press PAR to define additional search criteria for sorting through the memory.
You can search for sample ID, sequence number, not printed and not sent data
sets. As soon as you press PAR , a list with all options appears (Display 7).
SELECT PARAMETER
ID
Seq. No
Not printed
Not sent
All
Display 7 : Parameter selection
Use the arrow buttons to scroll through the list and press
if you have
found the desired criterion. In the case of sample "ID" and "Seq. No", the input
menu appears so that you can enter the respective number. If you chose "All",
no specific criteria are selected.
Press
to confirm your selection. Your selection will appear in the search
menu. If you leave the list by pressing
, your selection is not adopted.
25
Manual Uranalyze® Reader EN v1.0 / 06.15
®
Manual Uranalyze Reader
11.2.3 Process search results
Once you have made a selection for all search criteria, the
appears in
the search menu. To start the search, press
. A new screen with three
different activity buttons appears. The heading shows the number of results
selected given by your search criteria (Display 8). Press
to print,
to
send and
to display the data sets selected.
If you leave the menu by pressing
, you go straight back to the memory
menu.
XX entries found
Print
Send
Display
Display 8 : Select options for search results
11.3 Delete
Within the memory menu, you can also delete the entire memory. To delete the
memory, press
and confirm the following security question with
.
Note: Deleting the memory is irrevocable. Make sure that no relevant data is left
in the memory before you delete all entries.
26
Manual Uranalyze® Reader EN v1.0 / 06.15
®
Manual Uranalyze Reader
12. Check mode
The “Check mode” allows you to check the correct functioning of instrument
using special control strips (Display 9). You can order the control strips from
NOVASSAY.
Once you press
, the instrument asks you to insert the control strip. The
measurement of the control strip starts automatically as soon as the strip is
placed on the slide. Once the measurement has been completed, the measured
values are immediately displayed or printed. The values are so called remission
values for the pre-colored test pads on the control strip. These values have to
be compared to a set of standard values from the control strip instructions. For
additional information on the control strips, please consult the package insert, or
contact NOVASSAY directly
CHECK MODE
Insert control
strip!
Display 9 : Insert control strip
ATTENTION: Do not touch the colored areas of the control strip with your
fingers.
27
Manual Uranalyze® Reader EN v1.0 / 06.15
®
Manual Uranalyze Reader
13. Settings
The "Settings" menu allows you to adjust the instrument to your personal
requirements. Press
in the start screen and then
in the main menu to
get to the settings (Display 10).
Use the arrow buttons to choose an entry and confirm your choice
.
SETTINGS
Strip
Password
Printer
Sound
Autostart
Battery
Language
Date/Time
Interf ace
Display 10 : Settings
13.1 Test strips
Within the "Test strip" menu, there are three more options (Display 11).
Favorites allow you to define test strips for the “Favorites” menu. Additionally,
you can activate or disable the confidence interval message in the display. The
option “Correction” allows for the adjustment of the factory-provided parameter
calibration to account for matrix effects and interferences during the analysis.
28
Manual Uranalyze® Reader EN v1.0 / 06.15
®
Manual Uranalyze Reader
STRIP MENU
Favorites
Conf. interval
Correction
Uranium
Display 11 : Strip menu
Use the arrow buttons to choose an entry and confirm your choice with
If you press
, you get back to the settings.
.
13.1.1 Confidence interval
Naturally, test strip results are subject to slight deviations even when they are
generated photometrically. For every test there is a percentage based
confidence interval in which the true value can be found. The instrument can
optionally display the confidence interval (in +/- percent) underneath the result
after every measurement (Display 12). To activate or deactivate the confidence
interval, respectively, choose the desired choice button from the option field and
confirm with
. If you press
, none of your selection is adopted.
CONF. INTERVAL
RESULT
 % On
Seq. No: 0001
ID:
2015.01.20 9:47
 % Off
Uranium
64 mg/L
+/- 15%
Display 12 : Confidence interval (left) and Activated confidence interval (right)
29
Manual Uranalyze® Reader EN v1.0 / 06.15
®
Manual Uranalyze Reader
13.1.2 Correction
The correction mode allows for the adjustment of the NOVASSAY parameter
calibration and affects the result in terms of an additional calculation step. The
menu offers the opportunity of correcting the result with a linear correction
function.
ATTENTION: Faulty measurement results due to manipulated calibration are
the sole responsibility of the operator of the equipment.
By selecting the correction menu four values appear, that can be edited by
choosing each of them (Display 13). Selecting this function always affects the
parameter displayed in the “Test strips” menu.
URANIUM
A = 1.000
B = 1.000
C = 0.000
D= 0.000
Reset
Display 13 : Correction
The meaning of the values A to D is as follows:
A: gradient of the correction function
B: y-axis intercept of the correction function
C: new lower measurement limit
D: new upper measurement limit
Confirming your entry with
stores the settings. If you press
, none
of your selection is adopted. By pressing Reset , the values will be set back to
factory settings.
30
Manual Uranalyze® Reader EN v1.0 / 06.15
®
Manual Uranalyze Reader
Once you select one of these values, the input menu appears and you can
enter a value (Display 14). (Restrictions : The value for A must be > 0; value D
must be bigger than value C; if values C and D are equal to zero, the factoryprovided measuring range will be adjusted depending upon the entered values
for A and B).
Set A value
14_
1
2
3
4
5
6
7
8
9
.
0
Display 14 : Enter correction value
Confirming your entry with
stores the value. If you press
, the
former value will be kept.
The modification of a parameter will be indicated with a star in the parameter list
display as well as on the result display and printout. Also the result in the
memory will contain this indication, which cannot be removed by resetting the
values to factory-settings (Display 15).
RESULT
Seq. No: 0001
ID:
2015.01.28 10:43
*
Uranium
< 5 mg/L
+/- 15%
Display 15 : Result window
31
Manual Uranalyze® Reader EN v1.0 / 06.15
®
Manual Uranalyze Reader
13.2 Password
You can protect access to the instrument settings using a password, thus
restricting the number of user who can access the settings (Display 16). To
install a password, press "On" in the option field and confirm with
. If you
press
, none of your selection is adopted.
PASSWORD
 On
 Off
Display 16 : Password
The input menu appears and you can enter a password. Once you leave the
settings and then go back again, the instrument requires the respective
password.
13.3 Printer
In the printer menu, you can disable or enable the printer (Display 17). Press
the respective choice button in the option field and confirm with
. If you
press
, none of your selection is adopted.
TURN PRINTER
 On
 Off
Display 17 : Printer
32
Manual Uranalyze® Reader EN v1.0 / 06.15
®
Manual Uranalyze Reader
13.4 Sound
In the "Sound" menu, you can disable or enable the instrument sound. Press
the respective choice button in the option field and confirm with
. If you
press
, none of your selection is adopted (Display 18).
SOUND
 On
 Off
Display 18 : Sound
13.5 Autostart
The instrument has an autostart function (see 9. Measurements), which detect a
test strip on the slide and automatically starts the measurement process.
In the autostart menu, you can disable or enable the autostart (Display 19).
Press the respective choice button in the option field and confirm with
. If
you press
, none of your selection is adopted.
AUTOSTART
 On
 Off
Display 19 : Autostart
33
Manual Uranalyze® Reader EN v1.0 / 06.15
®
Manual Uranalyze Reader
If the autostart function is disabled, the start screen shows
(Display 20).
In this case, you have to press the activity button to start a measurement (see
9. Measurement).
Uranalyze
Test:
Uranium
SN:
0001
ID:
(P)
Display 20 : Deactivated autostart
13.6 Battery
You can make some adjustments to conserve energy. Within the menu you will
find to option fields (Display 21). Here you can enable or disable the screen’s
backlight function and the printer. Press the respective choice button(s) and
confirm with
. If you press
, none of your selection is adopted.
BATTERY
Backlight
Printer
 On
 On
 Off
 Off
Display 21 : Battery
34
Manual Uranalyze® Reader EN v1.0 / 06.15
®
Manual Uranalyze Reader
13.7 Language
The instrument can be set to 9 different languages. In the "Language" menu,
you can choose between the different languages (Display 22).
To activate a particular language, select the language using the arrow buttons
and confirm with
. The instrument changes the language automatically
and goes back to the settings menu. If you press
, none of your selection
is adopted.
LANGUAGE
English
Deutsch
Espanol
Français
Italiano
Portugues
Polski
Türkce
Magyar
Display 22 : Language
13.8 Date/Time
In the date "Date/Time" menu, you can adjust the date, the date format, the time
and the time format (Display 23).
SET DATE & TIME
2015
01
09
47
20
YMD
24
Display 23 : Date/Time
35
Manual Uranalyze® Reader EN v1.0 / 06.15
®
Manual Uranalyze Reader
There are three activity buttons for the date (upper row) and two activity buttons
the time (lower row). Once you press one of the activity buttons, the input menu
opens and you can make the necessary adjustments. To change the year, you
only have to insert the last two digits (e.g. 15 pour 2015).
You can set the date to three different formats (Table 1). An activity button
shows the currently active format. To change the format, press the activity
button. The date is adjusted automatically.
Date format
YMD
DMY
MDY
Signification
Year-Month-Day
Day.Month.Year
Month/Day/Year
2015-08-20
20.08.2015
08/20/2015
Table 1 : Date format
You can change the time format between 12 and 24 hours. To make the
respective adjustment, press the activity button at the right end of the row. If you
chose the 12 hour format, an additional activity button appears to set am or pm.
Confirm your entry with
. If you press
, none of your selection is
adopted.
13.9 Interface
In the "Interface" menu, you can disable or enable the instrument’s interface.
The instrument can send data only if the interface is enabled. Press the
respective choice button in the option field and confirm with
.
If you press
, none of your selection is adopted.
36
Manual Uranalyze® Reader EN v1.0 / 06.15
®
Manual Uranalyze Reader
13.10 Customization
In the "Customization" menu, you can personalize the first two lines of a result
print-out with a specific text (Display 24). Once you select the menu item, the
input menu appears and you can enter the next for the first line (24 characters
max.). Confirm your entry with
. The input menu appears again to enter
the text for the second line. Confirm again with
. If you press
,
none of your selection is adopted.
Printer line 1st

ABC
DEF
GHI
JKL
MNO
PQRS
TUV
WXYZ
123
-(#)
Display 24 : Customization
37
Manual Uranalyze® Reader EN v1.0 / 06.15
®
Manual Uranalyze Reader
13.11 Reaction time
In the "Reaction time" menu, you can decide whether the instrument will wait for
the strips reaction time before starting a measurement. By default, the reaction
time is enabled. The reaction time is then shown as a progress bar once the
strip is placed on the strip holder.
If you disable the reaction time, the instrument draws the strip into the housing
immediately and starts the measurement irrespective of the strip’s individual
reaction time. Hence, you need to place the strip onto the strip holder as soon
as the reaction time has run out. In such a way, you can potentially generate a
higher strip throughput.
To activate/deactivate the reaction time, press the respective choice button in
the option field and confirm with
. If you press
, none of your
selection is adopted. The switched off reaction time will be indicated with a
hashtag (#) on the result display and printout.
13.12 Print settings
The last menu item print settings allow you to print an overview over the
adjustments made settings.
Select the entry using the arrow buttons and press
. The overview is
printed automatically.
38
Manual Uranalyze® Reader EN v1.0 / 06.15
®
Manual Uranalyze Reader
14. Connecting external devices
The instrument can be connected to a barcode scanner and/or keyboard as well
as to a computer. If connected to a computer, the instrument can only send, but
not receive data (unidirectional communication).
14.1 Barcode scanner and keyboard
Both barcode scanner and keyboard can be connected via the PS/2 port. Using
a barcode scanner or keyboard enables you to enter sample IDs or to make
entries in open input fields.
14.2 Computer
You can connect the instrument to a computer using its USB or RS/232
interface (see 5.2 Interface). To read data from the instrument, you need a
current terminal program. For further information regarding connection to
computers or laboratory information systems, please contact you distributor or
NOVASSAY directly (see 18.5 Contact).
39
Manual Uranalyze® Reader EN v1.0 / 06.15
®
Manual Uranalyze Reader
15. Quality control
The instrument performs a self-calibration whenever it is turned on. Therefore,
in general, the instrument does not need to be calibrated. To check instrument
performance, we recommend to run multiple standard measurements with
known concentrations within the tests range. If such results are not within the
test’s confidence interval (see 13.1.1 Confidence interval), please contact your
distributor or NOVASSAY directly (see 18.5 Contact).
For additional quality control assessments, you can also use the instrument’s
check mode and special control strips (see 12. Check mode).
40
Manual Uranalyze® Reader EN v1.0 / 06.15
®
Manual Uranalyze Reader
16. Cleaning and maintenance
To ensure correct functioning of the instrument for a long time, the instrument
needs to be cleaned and maintained carefully. Generally, the instrument has to
be turned off and disconnected from the power supply, before you start cleaning
it.
16.1 Daily cleaning
The test strip slide has to be cleaned every day as to ensure the correct
functioning of the instrument.
Carefully remove the transport slide. To do so, pull on the upper black plastic
cover. Rinse the slide with clear water, if necessary, you can use some mild
cleaning agent as well. To disinfect the slide, you can use alcohol based
disinfectants (70-85 % ethanol or isopropanol). After disinfection, rinse the slide
with water again. Also clean the metal part of the transport slide. To fit the
plastic cover on the metal part again, simply push the plastic cover onto the
metal part. You can push the entire slide completely into the housing to ensure
a proper fir of metal part and plastic cover.
Note: Do not clean the transport tray in a dish-washer.
16.2 Weekly cleaning
Housing and touch screen should be cleaned at least once a week. Wipe the
housing with a damp cloth. If necessary, you can use a mild cleaning agent.
Then, dry the housing with a soft cloth. Make sure that no water penetrates the
housing or gets into the printer. To clean the touch screen, use a soft, lint-free
cloth. If necessary you can use special LCD cleaning fluid, which is
commercially available.
Note: Never use aggressive, organic solvents such as acetone for cleaning the
instrument. We recommend isopropanol for cleaning and disinfection.
41
Manual Uranalyze® Reader EN v1.0 / 06.15
®
Manual Uranalyze Reader
17. Error messages
If you cannot solve a problem with the instructions in the table, please contact
your distributor or NOVASSAY directly (see 18.5 Contact, p. 95). Error
messages are displayed as cleartext and are self explanatory (Table 2).
Error Message
Source
Solution
"Remove strip"
Test strip has not been removed Remove test strip and
from slide after measurement in start new measurement
the autostart mode
"Wrong position"
Strip has not been pushed to Start a new measurement
the far end of the slide
and push the strip to the
end of the slide
"No paper"
There is no paper left or the Insert new paper and close
printer cover is still open
printer cover
"Battery empty"
Empty batteries
"Instrument doesn’t start"
No power or power connection Check if all connections
broken
are correctly set and if the
outlet is working
Change
batteries
connect to an outlet
or
Table 2 : Error messages
42
Manual Uranalyze® Reader EN v1.0 / 06.15
®
Manual Uranalyze Reader
18. Service
18.1 Spare parts and accessories
Description
Reflectometer Uranalyze® Reader suitable for evaluation of Uranalyze® test
strips incl. Power supply, adapter, manual and 1 roll of printer paper
Printer paper for Uranalyze® Reader
18.2 Disposal
Disposal according to EU-directive 2002/96/EC. In accordance with EUdirective 2002/96/EC, NOVASSAY takes back old instruments and disposes off
them free of charge.
Note: Disposal via public collection systems is prohibited.
18.3 Warranty
The guarantee period is 24 months. The original commercial invoice serves as
the guarantee card and must be presented in case of a claim during the given
period. The guarantee is not valid in cases of improper handling and the
guarantee does not cover defects caused by the external power supply. The
guarantee is limited to repairs of defective parts or, in our sole discretion,
replacement with a brand new and defect-free instrument. Any guarantee
claims made or alleged do not influence the total guarantee period of 24
months. The right of withdrawal does not exist. Further claims are excluded and
these especially include claims for compensation caused by direct or indirect
damages. Additionally our current Terms and Conditions for the Sale of Goods
as printed on price lists shall apply.
43
Manual Uranalyze® Reader EN v1.0 / 06.15
®
Manual Uranalyze Reader
18.4 General usage comments
®
Only use the instrument as instructed to analyze Uranalyze test strips. Please
do not open the instrument or make any other unauthorized modifications. The
instrument is a highly sensitive measuring device with high accuracy for the
®
reflectometric evaluation of Uranalyze test strips. During the production
process, first all optical components are adjusted with especially designed tools.
Subsequently, technical experts calibrate the gauge head. Unauthorized
modifications or opening the instrument in an inappropriate way can result in
maladjustments of the gauge head, or similar damage, which in turn can lead to
wrong results.
Note: Due to the reasons stated above, NOVASSAY declines any responsibility
for the correct functioning of the instrument if it has been opened or used
improperly and objects to any warranty claims made in such a case.
NOVASSAY’s warranty and liability obligations as to product(s) or to any party
on account thereof shall be void and of no effect if the product(s) or any part
thereof have been subject to unauthorized change, abuse, misuse, accident or
alteration, unauthorized application or installation not in compliance with the
product(s) specifications as set forth in the product leaflets/instructions hereof
(“unauthorized action”) and such unauthorized action resulted in the nonconformance or defect of the product(s).