advertisement
M211877EN-D
User Guide
Vaisala Indigo™ 201 Analog Output Transmitter
Indigo 201
PUBLISHED BY
Vaisala Oyj
Street address: Vanha Nurmijärventie 21, FI-01670 Vantaa, Finland
Mailing address: P.O. Box 26, FI-00421 Helsinki, Finland
Phone: +358 9 8949 1
Visit our Internet pages at www.vaisala.com
.
© Vaisala Oyj 2017
No part of this manual may be reproduced, published or publicly displayed in any form or by any means, electronic or mechanical
(including photocopying), nor may its contents be modified, translated, adapted, sold or disclosed to a third party without prior written permission of the copyright holder. Translated manuals and translated portions of multilingual documents are based on the original English versions. In ambiguous cases, the English versions are applicable, not the translations.
The contents of this manual are subject to change without prior notice.
Local rules and regulations may vary and they shall take precedence over the information contained in this manual.
Vaisala makes no representations on this manual’s compliance with the local rules and regulations applicable at any given time, and hereby disclaims any and all responsibilities related thereto.
This manual does not create any legally binding obligations for Vaisala towards customers or end users. All legally binding obligations and agreements are included exclusively in the applicable supply contract or the General Conditions of Sale and General Conditions of Service of
Vaisala.
This product contains software developed by Vaisala or third parties. Use of the software is governed by license terms and conditions included in the applicable supply contract or, in the absence of separate license terms and conditions, by the General License Conditions of Vaisala
Group.
This product may contain open source software (OSS) components. In the event this product contains OSS components, then such OSS is governed by the terms and conditions of the applicable OSS licenses, and you are bound by the terms and conditions of such licenses in connection with your use and distribution of the OSS in this product. Applicable OSS licenses are included in the product itself or provided to you on any other applicable media, depending on each individual product and the product items delivered to you.
Table of Contents
Wireless Configuration Interface Overview.......................................................13
Receiving Analog Output Settings from Probe................................................14
Disabling WLAN Functionality with DIP Switch..............................................20
Connecting to Wireless Configuration Interface............................................. 24
Connecting with a Computer.......................................................................24
Connecting with a Mobile Device................................................................25
Logging in to Wireless Configuration Interface...............................................26
Changing Administrator Password..............................................................27
Resetting Administrator Password............................................................. 28
Status View: Analog Outputs.......................................................................32
User Access Settings (Administrator Password)...................................... 34
Factory Reset and Clearing Analog Output Settings............................... 36
Table of Contents
1
Indigo 201 User Guide
Analog Output Configuration Overview.......................................................... 43
Analog Output Configuration Example............................................................44
Clearing Analog Output Settings from Indigo.................................................45
Analog Output Overrange Behavior................................................................. 45
Analog Output Overrange Example........................................................... 45
Setting Relay Activation Limit Without Hysteresis......................................... 48
Setting Relay Activation Limit Using Hysteresis............................................. 49
Starting and Closing Calibration Mode.......................................................52
Restoring Factory Adjustment.................................................................... 53
Adjustment.....................................................59
8. Maintenance and Troubleshooting.............................................................63
Indigo Wireless Connection Troubleshooting..................................................63
Connecting to Indigo WLAN with iPhone........................................................64
Connection Option 1: Use Without Internet...............................................65
Connection Option 2: Disable iPhone Auto-Login....................................66
M211877EN-D
2
List of Figures
Figure 1 Indigo 201 With and Without Display, Wireless Interface
Figure 2 Serial Number on Probe Body (GMP251 Example)....................................8
Figure 5 Indigo 201 Display with Relays, Three Parameters and
Figure 9 Indigo 201 Transmitter Base Main Parts and Screw Positions...............16
Figure 12 WLAN OFF/ON DIP Switch on Indigo Circuit Board...............................20
Figure 13 Wireless Configuration Interface, Desktop Browser View.....................23
Figure 14 Enabling and Accessing Indigo's Wireless Configuration Interface... 24
Figure 16 User Level (User/Admin) in Upper Right Corner of Menu View...........27
Figure 17 Measurements View (Desktop Browser)....................................................29
Probe Example, Desktop Browser)............................30
Figure 19 Calibration Menu (CO
Probe Example, Desktop Browser).................. 33
Figure 21 Example of Analog Output Overrange Behavior ....................................46
Figure 22 Relay Configuration Options (CO
Probe Example)............................... 47
Figure 23 Relay Icons on the Optional Display (Relay A Active, Relay
Figure 26 Measurement Selections, CO
Probe Example.........................................55
Figure 27 Compensation Setpoint and Power-Up Default Selection,
Figure 28 Measurements Tab, CO
Probe Example.................................................... 57
Probe Example..........................................................58
Figure 30 Configuration Tab, CO
Probe Example..................................................... 59
Figure 31 Additional Steps Needed to Connect Notification..................................64
Figure 32 Indigo Transmitter Dimensions in Millimeters (mm)...............................72
List of Figures
3
4
Indigo 201 User Guide
List of Tables
Table 4 Indigo Transmitter Messages Shown on Display.........................................67
Table 8 Wireless Access Point (Module With Internal Chip Antenna).................70
M211877EN-D
Chapter 1 – About This Document
1. About This Document
1.1 Version Information
This document provides detailed instructions for using and maintaining Vaisala Indigo™ 201 analog output transmitter.
Table 1 Document Versions
Document Code Date
M211877EN-D Jun 2017
M211877EN-C
M211877EN-B
Feb 2017
Jan 2017
Description
This manual. User interface updates in place from software version 1.2.0
onwards.
• Calibration procedure and menu interface updated.
• Added information on changing and resetting administrator password.
• Added information on enabling and disabling WLAN functionality with
DIP switch.
• Added SRRC approval (CMIIT ID 2017DJ3476)
Previous version.
• Transmitter software version 1.01 onwards: entered values no longer need to be saved with the Save button. The values are saved automatically when the user leaves the input field.
• Added iPhone-specific WLAN connection instructions.
• Relay configuration information updated.
• Updated calibration information and added calibration instructions.
• Added information on Indigo display messages, probe compatibility, calibration PIN code, and resetting configurations.
• Accuracy specification information updated.
• Wireless configuration interface's menu selections and menu item names updated.
1.2 Related Manuals
Document Code Name
M211876EN
M211799EN
Indigo 201 Analog Output Transmitter Quick Guide
Vaisala CARBOCAPâ Carbon Dioxide Probe GMP251 User Guide
M211897EN
Vaisala CARBOCAPâ Carbon Dioxide Probe GMP252 User Guide
5
6
Indigo 201 User Guide M211877EN-D
1.3 Documentation Conventions
WARNING!
alerts you to a serious hazard. If you do not read and follow instructions carefully at this point, there is a risk of injury or even death.
CAUTION!
warns you of a potential hazard. If you do not read and follow instructions carefully at this point, the product could be damaged or important data could be lost.
Note highlights important information on using the product.
Tip gives information for using the product more efficiently.
1.4 Trademarks
Vaisalaâ is a registered trademark of Vaisala Oyj.
Indigo™ is a trademark of Vaisala Oyj.
Chrome™ is a trademark of Google Inc.
Firefoxâ is a registered trademark of Mozilla Foundation.
Microsoftâ, Windowsâ, Internet Explorerâ, and Edge™ are either registered trademarks or trademarks of Microsoft Corporation in the United States and/or other countries.
Safari™ is a trademark of Apple Inc, registered in the U.S. and other countries.
All other product or company names that may be mentioned in this publication are trade names, trademarks, or registered trademarks of their respective owners.
Chapter 2 – Product Overview
2. Product Overview
2.1 Introduction to Indigo™ Transmitters
1.35
%
| Carbon dioxide
WLAN is activated
Indigo 201
Figure 1 Indigo 201 With and Without Display, Wireless Interface Examples
Vaisala Indigo transmitters are a plug-and-play host device platform for Vaisala Indigocompatible probes. Indigo transmitters extend the feature set of connected probes with a range of additional options for outputs, measurement viewing, status monitoring, and configuration interface access.
Depending on the Indigo transmitter model, a display is available as an optional selection or as a standard feature. In the non-display model, an LED indicator is used for notifications.
Probes can be connected either directly on the transmitter from the probe's connector, or by using a cable between Indigo and the probe.
The configuration interface of Indigo transmitters is a browser-based wireless UI that requires a mobile device or computer that supports wireless connectivity (IEEE 802.11 b/g/n
WLAN).
The Indigo 201 transmitter is designed for analog output applications: the transmitter's output options include 3 analog output channels (current or voltage) and 2 relays.
For more information on Indigo transmitter models, see www.vaisala.com/indigo .
More Information
‣ Indigo 201 Basic Features and Options (page 8)
7
8
Indigo 201 User Guide M211877EN-D
2.2 Indigo 201 Basic Features and
Options
• All Vaisala Indigo-compatible probes can be connected to all Indigo transmitter models
• Plug-and-play probe installation: when unconfigured, Indigo automatically adapts the analog output configuration of the connected Indigo-compatible probe
• Wireless configuration interface: connect to the scalable browser-based UI of Indigo
201 to change probe and transmitter settings, view measurements and review probe and transmitter status
• 3.5” TFT LCD color display or non-display model with LED indicator
• Power supply input 15 ... 30 VDC (20 ... 22 VAC)
• 3 current (mA) or voltage (V) analog outputs
• 2 configurable relays
More Information
‣ Wireless Interface Menus (page 23)
‣ Display and LED Indicator (page 10)
‣ Input and Output Specification (page 19)
‣ Receiving Analog Output Settings from Probe (page 14)
‣ Analog Output Configuration Overview (page 43)
‣ Relay Configuration Overview (page 47)
2.3 Probe Compatibility
All Indigo-compatible probes can be used with all Indigo transmitter models. To verify that your probe is compatible with Indigo, check the serial number on the probe body. All probes intended for use with Indigo manufactured from 2017 onwards (serial numbers starting with the letter N or latter in alphabetical order) have full compatibility.
1
Made in Finland, Vaisala Oyj
Vanha Nurmijärventie 21, FI-01670 Vantaa
CO2 PROBE GMP251 A2B0A1N1
Output: 0... 20% CO2 RS485 Vaisala protocol
Input: 12 – 30 V d
Serial No. N0310004
c R r
Figure 2 Serial Number on Probe Body (GMP251 Example)
1 Probes with a serial number starting with the letter N have been manufactured in 2017.
2.4 Indigo Transmitter Parts
Chapter 2 – Product Overview
1
6
2
3
4
5
Figure 3 Indigo Transmitter Parts
1 Display (optional)
2 LED indicator (in non-display model)
3 Probe and probe cable connector, locking wheel
4 Wireless (WLAN) configuration interface activation button
5 Rubber lead-through with strain relief for input/output cable
6 Top cover of the transmitter: LED/display, circuit board and connector pins
7 Transmitter base: input and output wiring (screw terminals), mounting base
7
CAUTION!
Only use Vaisala Indigo-compatible probes with the transmitter. Attempting to connect incompatible probes or probe cables can damage the equipment.
More Information
‣ Dimensions (in mm) (page 72)
‣ Indigo Transmitter Base (page 16)
‣ Spare Parts and Accessories (page 71)
9
Indigo 201 User Guide M211877EN-D
2.5 Display and LED Indicator
Indigo 201 can be ordered either with the optional display, or as a non-display model that uses an LED indicator for notifications.
hPa
| Pressure
Figure 4 Indigo 201 Display with One Parameter
1 2
Figure 5 Indigo 201 Display with Relays, Three Parameters and WLAN Notification
1 Relay A and relay B status
2 Wireless configuration interface indicator (SSID text and WLAN symbol)
3 Message field for notifications, warnings and errors
3
10
Chapter 2 – Product Overview
The Indigo display can be configured to show 1-3 measurement parameters. Information about the transmitter and connected probe (for example, notifications and warnings) is shown on a message row at the bottom of the display. Relay and wireless access point activity is also shown on the display.
The parameters shown on the display, display brightness and display mode (numeric or graph) can be configured with the wireless configuration interface in the Settings > General menu.
More Information
2.5.1 Graph Display Mode
Figure 6 Indigo Display in Graph Mode
You can set the Indigo display to show the connected probe's measurement data as a graph.
The display mode can be changed in the wireless configuration interface's Settings >
General menu.
More Information
11
Indigo 201 User Guide M211877EN-D
2.5.2 LED Indicator
In the display model, notifications, warnings and other messages are shown on the display as text rows and indicator graphics. In the LED indicator model, the status of the transmitter and probe is indicated by LED color and whether the LED blinks. The following table lists the
LED notifications:
Table 2 LED Indicator States
LED Color and Flashes
Solid green LED:
Description
A solid green LED means that the transmitter and probe are working normally.
OK
Blinking green LED:
Starting/WLAN Active
Solid orange LED:
No Probe
Solid red LED:
Error
The LED blinks green when the transmitter is starting up, and changes to a steady green when the transmitter is ready for use.
The LED also blinks green when the wireless connection is active, and changes to a steady green when the connection ends.
A solid orange LED is shown when a probe is not connected to the transmitter (or the probe is not properly connected and the transmitter does not recognize it).
A solid red LED means that an alarm or a notification is active, or the probe or the transmitter is in an error state.
To view information about the alarm or notification, connect to Indigo's wireless configuration interface with your web browser and open the Status menu.
More Information
‣ Connecting to Wireless Configuration Interface (page 24)
12
Chapter 2 – Product Overview
2.6 Wireless Configuration Interface
Overview
Indigo transmitters are configured using a wireless browser-based configuration interface
(requires a mobile device or computer with IEE 802.11 b/g/n WLAN wireless connectivity). In addition to probe and transmitter configuration and calibration, you can also use the wireless interface to view measurement data and status information.
Figure 7 Desktop and Mobile Example Views
The wireless configuration interface has two user levels. All users have view-only access (no configuration rights, not password protected), and personnel that carry out configuration tasks can log in with an administrative password that allows changing the transmitter and probe settings.
To use the wireless configuration interface to modify the settings of your Indigo transmitter and the connected probe, you must first enable the transmitter's wireless connection and then connect to Indigo with your mobile device or computer. Most major browsers (for example, Firefox, Chrome, Safari and Internet Explorer) are supported: using the most recent version is recommended.
More Information
‣ Connecting to Wireless Configuration Interface (page 24)
‣ Logging in to Wireless Configuration Interface (page 26)
‣ Wireless Interface Menus (page 23)
‣ Indigo Wireless Connection Troubleshooting (page 63)
13
14
Indigo 201 User Guide M211877EN-D
2.7 Receiving Analog Output Settings from Probe
When you take Indigo 201 in use for the first time and have not entered an analog output configuration, the transmitter automatically adapts the analog output configuration of the first Vaisala Indigo-compatible probe you connect.
If an analog output configuration already exists in Indigo 201 (that is, you have previously connected a probe or configured the settings), you need to set the analog output configuration of the new probe manually. Analog output from the transmitter is halted and does not resume until you have set the new probe's configuration.
Alternatively, you can clear the analog output settings in Indigo 201 configuration. This returns the transmitter to a state where it automatically starts using the analog output configuration of the next connected probe. The analog output settings can be cleared from the Settings > General menu.
Figure 8 Clearing Analog Output Settings
More Information
‣ Attaching Probes and Cables (page 18)
‣ Analog Output Configuration Overview (page 43)
‣ Clearing Analog Output Settings from Indigo (page 45)
Chapter 3 – Installation
3. Installation
3.1 Mounting
1. Open the 4 screws on the transmitter cover and pull the cover off the base.
CAUTION!
To avoid damaging the connector pins of the transmitter, pull the transmitter cover off the base in a straight angle. Do not twist or bend.
2. Place the transmitter base on the installation surface and mount it with 3 screws. See the screw positions in
.
3. Lead the input/output cable inside the transmitter (see
3.3 Wiring Options (page 17) ).
If you are wiring through the lead-through on the bottom of the transmitter, test that the strain relief works with your cable.
4. Connect the input/output cable's wiring to the screw terminals (see 3.2 Indigo
) and reattach the cover when done.
5. When you have finished wiring and reattached the cover, switch on the power supply input.
15
Indigo 201 User Guide
3.2 Indigo Transmitter Base
M211877EN-D
INDIGO 201 ANALOG
OUTPUT TRANSMITTER
Serial No. SX12345678
5
6
1 2 3 4
Figure 9 Indigo 201 Transmitter Base Main Parts and Screw Positions
1 Probe and probe cable connector inside the locking wheel
2 Wireless (WLAN) configuration interface activation button
3 Wiring from the back: cut open the seal
4 Rubber cable lead-through with strain relief
5 Screw terminals for analog outputs 1-3 and relays A and B
6 Screw terminals for 24 V power supply input
CAUTION!
Do not energize the power supply before the wiring has been connected.
16
Chapter 3 – Installation
3.3 Wiring Options
You can wire an input/output cable either through the opening on the back of the transmitter, or through the rubber lead-through on the bottom of the transmitter.
1
2
Ø
7 ... 8 mm
Figure 10 Indigo Wiring Options
1 Wiring from the back: cut the seal open
2 Wiring through the rubber lead-through on the bottom of the transmitter
1. To wire the input/output cable through the back of the transmitter: a. Cut off as much of the seal as is needed to fit your cable through the opening.
b. Lead the cable through the opening and attach a strain relief as needed.
c. If you wire only through the back, plug or seal the lead-through on the bottom.
2. To wire the input/output cable through the rubber lead-through on the bottom: a. Push the input/output cable through the lead-through.
b. The lead-through provides strain relief and holds the cable in place. Tightening is not required.
The recommended cable diameter for wiring through the rubber lead-through is
7 ... 8 mm. If you use a different cable size, test that the strain relief works as intended.
17
Indigo 201 User Guide
3.4 Attaching Probes and Cables
M211877EN-D
3
2
1
Figure 11 Attaching Probes and Cables to Indigo
1 Insert probes into the probe connector with the orientation mark facing out.
2 Probes are locked in place with the locking wheel. Never turn from the probe body.
3 Connect probe cables in the same way as probes: insert the cable in the connector and hold in place while turning the locking wheel.
1. Insert the probe into the probe connector with the orientation mark on the probe body facing out.
2. Hold the probe in the probe connector and lock it in place by turning the locking wheel counterclockwise. Do not turn the probe body when attaching, only the locking wheel on the transmitter.
3. If no previous configuration exists (that is, this is the first time a probe is connected to
Indigo, and no configuration has been set manually), Indigo adapts the analog output settings of the connected Vaisala Indigo-compatible probe automatically.
If a configuration that differs from the attached probe's configuration exists in the transmitter, you must configure the new probe manually to enable analog output.
See 2.7 Receiving Analog Output Settings from Probe (page 14)
.
18
Chapter 3 – Installation
3.5 Input and Output Specification
Table 3 Input and Output
Property
Analog outputs
Accuracy of analog outputs at 20 °C
Relays
Maximum current
Power consumption
Probe connector
Cable feed throughs
Screw terminal wire size
Specification
3 voltage (V) or current (mA) outputs:
• 0 … 10 V / 0 … 5 V / 0 … 1 V / 1 … 5 V (min. load
1kΩ)
• 0 … 20 mA / 4 … 20 mA (max. load 500 Ω)
± 0.1% full scale
2 configurable relays (VAC/VDC)
Device maximum specification (resistive load):
• Max. switching power 30 W / 37.5 VA
UL-rated maximum specification (resistive load):
• AC: max. 28 V / 0.5 A
• DC: max. 40 V / 0.24 A
• Up to 30 VDC:
• max. switching current 1 A
• max. switching power 30 W
Range 15 ... 30 VDC (20 ... 22 VAC)
Transmitter and connected probe max. 1 A
Transmitter max. 3 W (+ connected probe, varies depending on probe type)
M12/5 connector for probe or probe cable connection
(Vaisala Indigo-compatible probes)
2 options: rubber lead-through on the bottom of the transmitter, and opening with a seal at the back of the transmitter
0.2 mm
2
... 1.5 mm
2
1) Using a power supply with overload protection is recommended for electrical safety.
CAUTION!
Do not modify the unit or use it in ways not described in the documentation.
Improper modification may lead to safety hazards, equipment damage, failure to perform according to specification, or decreased equipment lifetime.
More Information
19
Indigo 201 User Guide M211877EN-D
3.6 Disabling WLAN Functionality with
DIP Switch
If your application requires removing WLAN communication from use, use the DIP switch on the Indigo circuit board.
WLAN
OFF ON
1
Figure 12 WLAN OFF/ON DIP Switch on Indigo Circuit Board
1 WLAN functionality OFF or ON
To switch WLAN communication OFF or ON with the DIP switch:
1. Power off the transmitter.
2. Open the transmitter cover.
CAUTION!
To avoid damaging the connector pins of the transmitter, pull the transmitter cover off the base in a straight angle. Do not twist or bend.
20
Chapter 3 – Installation
3. Use a small screwdriver or a similar tool to change the position of the WLAN DIP switch
(1).
4. When done, close the transmitter cover and power on Indigo.
Indigo does not display a notification about the WLAN functionality being disabled when a user attempts to enable it with the wireless connection activation button.
21
Indigo 201 User Guide M211877EN-D
22
1
2
3
4
Chapter 4 – Wireless Configuration Interface
4. Wireless Configuration
Interface
4.1 Wireless Interface Menus
5
Figure 13 Wireless Configuration Interface, Desktop Browser View
1 Measurements: displays the measurement data of the connected probe
2 Status: contains information about the status of Indigo and the connected probe (for example, notifications and alarms)
3 Calibration: calibrate and adjust probes using references. Available options (for example, adjustment points) vary depending on the probe model.
4 Settings: contains options for configuring the connection and display settings, outputs, relays, probe-specific settings, and general device preferences
• General submenu: device information and general settings, wireless connection and display settings
• Outputs submenu: configuration options for analog outputs 1-3
• Relays submenu: settings for controlling relays A and B
• Probe submenu: probe-specific settings such as environmental compensations and filtering factor
5 Main display area for menus and measurement information (desktop browser example)
More Information
‣ Connecting to Wireless Configuration Interface (page 24)
23
24
Indigo 201 User Guide M211877EN-D
4.2 Connecting to Wireless Configuration
Interface
1
2
Indigo 200
500 ppm
Carbon dioxide concentration
WLAN on
3
Select WLAN to connect to:
1. WLAN XYZ
2. Indigo_IDxx
3. WLAN ABC
Figure 14 Enabling and Accessing Indigo's Wireless Configuration Interface
1 Wireless connection activation button
2 Wireless connection indicator (WLAN symbol) on the Indigo display
3 Choose Indigo (Indigo_ID[xx]) from your wireless device's list of available connections
More Information
‣ Logging in to Wireless Configuration Interface (page 26)
‣ Indigo Wireless Connection Troubleshooting (page 63)
4.2.1 Connecting with a Computer
To connect to the Indigo wireless configuration interface with a computer:
1. Enable wireless connectivity (WLAN) from your computer's settings.
2. Press the wireless connection activation button on the bottom of the Indigo transmitter until the transmitter notifies about a WLAN connection being active.
3. Open your computer's wireless connections menu and choose Indigo_ID[xx]
(transmitter-specific SSID) from the list of available connections.
Chapter 4 – Wireless Configuration Interface
4. When you have established a connection, open your browser. Depending on your browser and system, you are either directed to the wireless configuration interface's login screen, or you may need to: a. open a new browser tab or window, or close and restart your browser, if the browser was already open when you connected to Indigo's access point.
b. acknowledge the connection in a notification prompt before opening your browser or a new browser tab or window.
The default IP address of the Indigo transmitter is http://192.168.1.1/. If you have trouble opening the wireless configuration interface in your browser, try entering the
IP address in the browser's address bar.
5. When the wireless configuration interface opens in your browser, you are prompted to log in. For instructions, see
4.3 Logging in to Wireless Configuration Interface
.
4.2.2 Connecting with a Mobile Device
To connect to the Indigo wireless configuration interface with a mobile device (phone or tablet):
1. Enable wireless connectivity (WLAN) from your mobile device's settings.
2. Press the wireless connection activation button on the bottom of the Indigo transmitter until the transmitter notifies about a WLAN connection being active.
3. Open your device's wireless connections menu and select Indigo_ID[xx] (transmitterspecific SSID) from the list of available connections.
4. Depending on device settings, the Indigo wireless configuration interface either opens automatically in your browser, or you may need to: a. open your browser application manually after connecting to Indigo b. acknowledge the connection in a wireless network prompt (check your device's notifications) before opening your browser.
5. When the wireless configuration interface opens in your browser, you are prompted to log in. For instructions, see
4.3 Logging in to Wireless Configuration Interface
.
More Information
‣ Connecting to Indigo WLAN with iPhone (page 64)
25
Indigo 201 User Guide M211877EN-D
4.3 Logging in to Wireless Configuration
Interface
Figure 15 Indigo Login View
When you open Indigo's wireless configuration interface in your browser, you are prompted to log in. There are 2 available user levels:
• User: view-only access available for all users. Does not require a password.
• Admin: password-protected access. To change settings, you must log in as admin.
To log in:
1. Enter the user name and password: a. To log in as user (view-only access, no configuration rights), select User from the
User name dropdown. Leave the Password field empty.
b. To log in as admin (required for configuration), select Admin in the User name dropdown and type the admin password (default: 12345) in the Password field.
2. Select Log in after entering the login credentials. The wireless configuration interface opens in the Measurements view.
The user level (User or Admin) is shown in the upper right corner of all menu views.
Select the user/admin icon in the upper right corner to change the user level.
More Information
‣ Changing User Level (page 27)
‣ Changing Administrator Password (page 27)
‣ Resetting Administrator Password (page 28)
26
Chapter 4 – Wireless Configuration Interface
4.3.1 Changing User Level
A link that allows logging in with a different user level is included in the upper right corner of all menu views. The text of the link shows the current user level.
1
Figure 16 User Level (User/Admin) in Upper Right Corner of Menu View
1 Current user level (User/Admin): select the text or icon to log in with a different user level
To change between the User and Admin user levels:
1. Select the link that shows the current user level (User or Admin) on the upper right corner of any menu view.
Selecting the link logs you out of the interface, and a new login is required.
2. The login screen opens. Log in as User or Admin.
4.3.2 Changing Administrator Password
To change the administrator password:
1. Connect to the wireless configuration interface and log in as an administrator (default password: 12345). The user level (User or Admin) can be also changed by selecting the user level icon in the upper right corner of any menu view.
2. Open the Settings > General menu.
27
Indigo 201 User Guide
3. In the User Access section, enter the new password (max. 25 characters) in the
Administrator Password field.
M211877EN-D
4. The new password is saved when you tap or click outside the text field, and is in use at the next login.
4.3.3 Resetting Administrator Password
If you have lost or forgotten the admininistrator password and cannot log in, you can reset the password back to default (12345). Note that this requires powering the transmitter off and on.
To reset the password:
1. Power off the transmitter.
2. Keep the wireless connection activation button on the bottom of the transmitter pressed down.
3. While pressing the wireless connection activation button, power on the transmitter.
Keep the button pressed down until the transmitter start-up is complete (a measurement is shown on the display). In the LED indicator model, the LED blinks yellow when the password reset is complete.
4. The administrator password has now been reset to the default 12345.
28
4.4 Measurements View
Chapter 4 – Wireless Configuration Interface
Figure 17 Measurements View (Desktop Browser)
The Measurements view shows the measurement data from the connected probe in numeric and graph format. When you log in to Indigo, the wireless configuration interface opens in this view.
29
Indigo 201 User Guide
4.5 Status View
M211877EN-D
Figure 18 Status View (CO
2
Probe Example, Desktop Browser)
The Status view shows the transmitter status, general information about the Indigo transmitter, and the status of the analog outputs 1-3 and relays A and B.
More Information
‣ Status View: General (page 31)
‣ Status View: Analog Outputs (page 32)
‣ Status View: Relays (page 32)
30
Chapter 4 – Wireless Configuration Interface
4.5.1 Status View: General
The General section of the Status menu shows general information about the transmitter and the connected probe.
• Indigo Status: shows the current status of the transmitter. Errors and notifications are displayed here.
• Indigo Serial Number: the transmitter's serial number.
• WLAN Access Point Name: the network name (SSID) of the transmitter. Can be configured in the Settings menu.
• WLAN MAC Address: unique hardware address of the unit. Cannot be changed.
• Probe: the type and serial number of the probe that is connected to Indigo.
More Information
31
Indigo 201 User Guide M211877EN-D
4.5.2 Status View: Analog Outputs
The Analog Outputs section of the Status menu shows information about the status of
Indigo's analog outputs 1-3.
• Output: The number of the analog output channel (1-3).
• Signal Level: The level of the analog output signal: shows the present measurement as scaled in the analog output settings.
• Mode: The mode and range selected for the analog output (voltage, current or off).
• Parameter: The measured parameter that is sent on the analog output channel.
More Information
‣ Analog Output Configuration Overview (page 43)
4.5.3 Status View: Relays
32
• Relay: relay A or relay B
• State: Shows whether the relay is active (set) or not.
• Mode: Relay activation mode (above or below a trigger level, or off)
• Parameter: The parameter that controls the relay
More Information
‣ Relay Configuration Overview (page 47)
4.6 Calibration Menu
Chapter 4 – Wireless Configuration Interface
Figure 19 Calibration Menu (CO
2
Probe Example, Desktop Browser)
In the Calibration menu, you can calibrate and adjust the measurement of the probe connected to the transmitter with the help of references such as calibration gases that have a known concentration. You can also view the current adjustments and restore the probe's factory adjustments.
The available options vary depending on the type of the connected probe (for example, applicable environmental compensations and the number of adjustment points).
More Information
‣ Calibration Overview (page 51)
4.7 General Settings
The Settings > General menu contains the following configuration options:
•
•
4.7.2 User Access Settings (Administrator Password) (page 34)
•
4.7.3 Display Settings (page 35)
•
4.7.4 Factory Reset and Clearing Analog Output Settings (page 36)
33
Indigo 201 User Guide
•
4.7.5 License Information (page 37)
4.7.1 WLAN Settings
M211877EN-D
• Mode: Select the timeout period for disconnecting the wireless configuration interface, or remove the timeout from use.
• SSID: The network name that is used to identify the transmitter when connecting to it
(default: Indigo_ID[xx]).
4.7.2 User Access Settings (Administrator Password)
Enter a new administrator password and exit the text entry field. The new password is in use in the next login.
If you have forgotten or lost the password, you can reset the password back to the default one (12345). For more information, see the login instructions.
34
4.7.3 Display Settings
Chapter 4 – Wireless Configuration Interface
General display settings:
• Brightness: set the brightness of the display (20%, 50%, 100% or off).
• Mode: Select whether the display shows the measurement information in graph or numeric format.
Graph view settings:
• Parameter: Select which parameter's measurement is shown in the graph view.
• Unit: Select the unit that the measurement is shown in (for example, ppmCO
2
or %CO
2
).
35
Indigo 201 User Guide M211877EN-D
• Rounding: Choose how many decimals of the measurement are shown in the numeric reading shown alongside the graph view.
• Time Scale: Select the time period shown in the measurement graph (1 minute ... 1 hour).
Numeric view settings:
• Parameter (1-3): Select the measurement parameter or parameters shown in the numeric view (up to 3 parameters can be shown simultaneously in the numeric view).
• Unit for Parameter (1-3): Select the unit the measurement is shown in (for example, ppmCO
2
or %CO
2
).
• Rounding for Parameter (1-3): Choose how many decimals of the measurement are shown in the numeric view.
More Information
‣ Display and LED Indicator (page 10)
‣ Graph Display Mode (page 11)
4.7.4 Factory Reset and Clearing Analog Output Settings
• Factory Reset: select this to restore the configuration of the transmitter to factory default settings.
• Clear Analog Output Settings: this option removes the current analog output settings from Indigo's memory. When you clear the analog output settings, Indigo automatically adapts the analog output settings of the next probe you connect.
More Information
‣ Receiving Analog Output Settings from Probe (page 14)
‣ Clearing Analog Output Settings from Indigo (page 45)
36
Chapter 4 – Wireless Configuration Interface
4.7.5 License Information
Select the View software licenses link to view information on Indigo 201 software licenses.
4.8 Output Settings
The Settings > Outputs menu contains the configuration options for analog outputs 1-3.
37
Indigo 201 User Guide M211877EN-D
The analog output settings are automatically imported from the probe if the Indigo transmitter does not have an analog output configuration in place. For more information, see
2.7 Receiving Analog Output Settings from Probe (page 14) .
General output settings:
• Mode: Select whether the analog outputs 1-3 are used for current (mA) or voltage (V) output.
Settings for analog outputs 1-3:
• Output Type: Select the voltage or current range used for the output (for example,
0 ... 10 V)
• Parameter: The measurement parameter that is sent on the analog output channel.
• Unit: The unit used for the output (for example, % if the output is used for %CO
2 measurements).
• Scale Low End: The lower limit of the measurement scale (for example, 0 if the scale is
0 ... 20 %CO
2
)
• Scale High End: The upper limit of the measurement scale (for example, 20 if the scale is 0 ... 20 %CO
2
)
• Clipping Limit: The point (in % above the output range) when the output is clipped due to being out of range. For more information, see
• Error Limit: The point (in % above the output range) when the output switches to error
state due to being out of range. For more information, see 5.4 Analog Output
Overrange Behavior (page 45) .
• Error Output Voltage: The voltage that is output in an error state.
• Error Output Current: The current that is output in an error state.
More Information
‣ Analog Output Configuration Overview (page 43)
38
Chapter 4 – Wireless Configuration Interface
4.9 Relay Settings
The Settings > Relay menu contains the configuration options for controlling relays A and B.
Both relays have the same configuration options (relay A example shown here).
Output Mode
Parameter
Unit
Low Trigger
Level and High
Trigger Level
Error State
Select whether the relay is activated when the measurement exceeds the set trigger, or when the measurement falls below the set trigger. Set to
Off if the relay is not in use.
Select which measurement parameter controls the relay.
Select the unit of the measurement parameter (for example, % if the measurement is in %CO
2
).
If you want to activate the relay above or below a single setpoint without using hysteresis, enter the same value for the low trigger and the high trigger. The Output Mode selection defines whether the relay activates above or below this value. If you want to set a hysteresis, define the limits of the hysteresis with the low and high triggers.
Choose the behavior of the relay when an error state occurs (switched on or off, or remains in its current state).
When you use a single setpoint without hysteresis, set the same value for Low Trigger
Level and High Trigger Level. For instructions on using hysteresis, see
Activation Limit Using Hysteresis (page 49)
.
More Information
‣ Relay Configuration Overview (page 47)
39
Indigo 201 User Guide M211877EN-D
4.10 Probe Settings
The Settings > Probe menu contains probe-specific configuration options (for example, filtering factor and environmental compensations).
This menu also includes the Calibration PIN Code entry field. The PIN code must be in place in order to calibrate probes. The PIN code is in place by default.
The available options depend on the features of the connected probe. The examples here show the probe settings for Vaisala CARBOCAPâ Carbon Dioxide Probe GMP251.
General (All Probes)
The Calibration PIN Code must be in place to enable probe calibration and adjustment in the Calibration menu. The PIN code is in place by default.
If the PIN code has been removed and you need to enter it, check the probe's documentation for information on the code used in the probe model.
40
Measurement (GMP251 Example)
Chapter 4 – Wireless Configuration Interface
Filtering factor
Temperature compensation mode
Pressure/humidity/oxygen compensation on/off
Defines how much past measurements affect the output
(measurement averaging over time). For details, see probe documentation.
Select whether the probe sensor's measurement or a manually entered setpoint is used to set the temperature compensation.
Enable or disable the environmental compensations.
Compensation Setpoints (GMP251 Example)
Enter the setpoint values for the environmental compensations that are in use (enabled in the Measurement selections).
41
Indigo 201 User Guide
Compensation Power-up Defaults (GMP251 Example)
M211877EN-D
The power-up default values are stored to the probe memory and remain in use also after you disconnect or reset the probe.
Note that the environmental compensations you set in the Settings > Probe menu and the compensations you set in the Calibration menu are interconnected: the configuration set in either menu is applied to both.
More Information
‣ Environmental Compensation (page 54)
42
Chapter 5 – Configuring Analog Outputs
5. Configuring Analog Outputs
5.1 Analog Output Configuration
Overview
Indigo 201 has 3 analog outputs that can be configured with the wireless configuration interface's Settings > Outputs menu. Each output has the same set of configuration options
(analog output 1 shown in example):
Figure 20 Analog Output Configuration Options
Note that the Mode selection (under General) defines whether the outputs are used for voltage or current output.
Receiving Analog Output Settings Automatically from Probe
Indigo can receive the analog output settings automatically from the connected probe in the following cases:
43
Indigo 201 User Guide M211877EN-D
• the transmitter is a new unit and has not been configured yet, or
• you have removed the analog output configuration from Indigo manually.
It is recommended to always check the analog output settings from the wireless configuration interface's Settings > Outputs menu after connecting a probe.
More Information
‣ Receiving Analog Output Settings from Probe (page 14)
‣ Clearing Analog Output Settings from Indigo (page 45)
5.2 Analog Output Configuration
Example
This example shows how to configure analog output 1 to use the following settings when a
CO
2
probe is connected to Indigo:
• 4 ... 20 mA current output
• Carbon dioxide measurement scaled to 0 ... 20 %CO
2
• Clipping limit 5%, error limit 10%, and error output 3.6 mA
When you enter a value into a field, the value is saved automatically when you exit the input field (for example, tap on an area outside of the field).
To configure the output:
1. Launch the Indigo 201 wireless configuration interface.
2. Open the Settings > Outputs menu.
3. Select Current outputs from the Mode dropdown.
4. Select 4 ... 20 mA from the Output Type dropdown.
5. Select Carbon dioxide concentration from the Parameter dropdown.
6. Select % from the Unit dropdown.
7. Type 0 in the Scale Low End field.
8. Type 20 in the Scale High End field.
9. Type 5 in the Clipping Limit field and 10 in the Error Limit field.
10. Type 3.6 in the Error Output Current field.
More Information
44
Chapter 5 – Configuring Analog Outputs
5.3 Clearing Analog Output Settings from
Indigo
If you need to restore Indigo 201 to a state where no analog output configuration exists in the transmitter, you can clear the current configuration from the Settings > General menu.
When you remove the analog output configuration from Indigo, Indigo automatically adapts the analog output settings of the next probe that you connect.
To clear the configuration:
1. Launch the Indigo 201 wireless configuration interface.
2. Open the Settings > General menu.
3. Go to the Reset > Clear Analog Output Settings section and select Clear.
5.4 Analog Output Overrange Behavior
The analog output of the probe has a defined behavior when the values measured by the probe are outside the scaled analog output range. At first, the output is clipped when the measurement exceeds a set limit (the measurement continues, but the output does not change from the clipped value).
When the measurement exceeds the second limit (error limit), the analog output switches to the error state defined for the output.
The same clipping and error limits are applied when the measured value drops back to the scaled range: at first the output returns to the clipped value from the error state, and then to normal output.
5.4.1 Analog Output Overrange Example
Consider a probe with 0 ... 5 V output, scaled to 0 ... 200 000 ppm (= 0 ... 20 %) CO
2
.
• When the measured CO
2
rises above 20 %, the output rises above 5 V.
45
Indigo 201 User Guide M211877EN-D
• The output keeps rising until the measurement is 21 %CO
2
, at which point the probe outputs 5.25 V.
• If the CO
2
level rises above 21 %CO
• If the CO
2
level rises above 22 %CO
2
, the output still remains at 5.25 V.
2
, the output enters the error state, which is 0 V for the 0 ... 5 V output.
Output voltage (V)
0.00
5.25
5.00
Error level
Output clipping limit
In error state at
>220000 ppm
(200000 ppm + 10%)
Clipped at
210000 ppm
(200000 ppm + 5%)
Regular measurement
Time
Figure 21 Example of Analog Output Overrange Behavior
This example uses output scaled to 0 ... 5 V and 0 ... 200000 ppm, error level set to 0 V, clipping set to 5 % overrange, and error limit set to 10 % overrange. CO
2
concentrations
(ppm) are indicated for the clipping point and error limit point.
46
Chapter 6 – Configuring Relays
6. Configuring Relays
6.1 Relay Configuration Overview
Indigo transmitters have 2 configurable relays (relay A and relay B). Both relays have configuration options for selecting the parameter that is used to control the relay, activation triggers, hysteresis, and error state behavior.
1
4
4
5
2
3
Figure 22 Relay Configuration Options (CO
2
Probe Example)
1 Output Mode: Select whether the relay activates above or below a set value (or set the relay Off).
2 Parameter: The measurement that is used to control the relay.
3 Unit: Select the unit of the measurement parameter that controls the relay (for example,
% if the measurement is in %CO
2
).
4 Low Trigger Level and High Trigger Level: If you want to activate the relay above or below a single setpoint without using hysteresis, enter the same value for the low trigger and the high trigger. The Output Mode selection defines whether the relay activates above or below this value.
If you want to set a hysteresis, define the limits of the hysteresis with the low and high triggers.
5 Error State: Select which state the relay is set to when an error occurs (on, off, or remains in its current state)
47
Indigo 201 User Guide M211877EN-D
Relay State Information
When one or both relays are enabled, the relay state (active/not active) is shown on the optional display. You can also check the status of the relays in the Status view of the wireless configuration interface.
Figure 23 Relay Icons on the Optional Display (Relay A Active, Relay B Not Active)
Relay Activation Setpoint and Hysteresis
You can define whether the relay is activated when the measurement falls below a set limit
(Active below trigger level), or when the measurement exceeds a set limit (Active above
trigger level). To prevent the relay switching back and forth when the measured value is near to the setpoint value, you can set a hysteresis with the Low Trigger Level and High
Trigger Level settings.
For example, if you want the relay to activate when the measurement exceeds (Active
above trigger level) 2000 units, but do not want the relay to switch off if the measurement momentarily falls between 2000 and 1980 units, set the High Trigger Level to 2000 and
Low Trigger Level to 1980. With this configuration, the relay activates when the measurement exceeds 2000, but does not switch off until the measurement falls below
1980.
6.2 Setting Relay Activation Limit
Without Hysteresis
You can set the relay to activate when the probe measurement exceeds or falls below a set limit. When you configure a single setpoint for relay activation without hysteresis, the relay switches on or off immediately when the measurement moves over or below the setpoint
(depending on the low/high activation mode selection).
When you enter a value into a field, the value is saved automatically when you exit the input field (for example, tap on an area outside of the field).
48
Chapter 6 – Configuring Relays
To define a single setpoint for relay activation:
1. Open the Settings > Relays menu in the wireless configuration interface.
2. In the Relays menu, select the relay activation mode from the Output Mode dropdown menu: a. select Active above trigger level if you want the relay to activate when the probe measurement exceeds the set limit b. select Active below trigger level if you want the relay to activate when the probe measurement falls below the set limit
3. Select the measurement parameter that is used to control the relay with the Parameter dropdown.
4. Set the unit of the measurement parameter with the Unit dropdown.
5. Enter the same measurement limit to both the Low Trigger Level and the High Trigger
Level field.
With this configuration, there is no hysteresis. The relay activates or switches off immediately after passing this point.
6. Select the Error State for the relay.
6.3 Setting Relay Activation Limit Using
Hysteresis
If the measurement you are using to control the relay is likely to move back and forth close to the activation setpoint, you can set a hysteresis that prevents the relay switching on and off too frequently.
When hysteresis is used, the relay activates at the defined limit, but does not switch off immediately when the measurement moves back to the other side of the activation limit.
Instead, with hysteresis, the relay remains active until the measurement reaches the defined tolerated variation limit.
To set a relay activation limit with hysteresis:
1. Open the Settings > Relays menu in the wireless configuration interface.
2. In the Relays menu, select the relay activation mode from the Output Mode dropdown menu: a. select Active above trigger level if you want the relay to activate when the probe measurement exceeds a set limit b. select Active below trigger level if you want the relay to activate when the probe measurement falls below a set limit
49
Indigo 201 User Guide M211877EN-D
3. Select the measurement parameter that is used to control the relay with the Parameter dropdown.
4. Set the unit of the measurement parameter with the Unit dropdown.
5. Enter the relay activation limit either to the Low Trigger Level or the High Trigger
Level field: a. If you are using Active above trigger level (relay activation when the measurement exceeds a set limit), enter the limit to the High Trigger Level field.
b. If you are using Active below trigger level (relay activation when the measurement falls below a set limit), enter the limit to the Low Trigger Level field.
6. To define the hysteresis value: a. If you are using Active above trigger level, enter the limit for tolerated variation below the setpoint to the Low Trigger Level field.
With this option, the relay activates when the measurement exceeds the limit entered in High Trigger Level, and switches off when the measurement falls below the limit entered in the Low Trigger Level field.
b. If you are using Active below trigger level, enter the limit for tolerated variation above the setpoint to the High Trigger Level field.
With this option, the relay activates when the measurement falls below the limit entered in Low Trigger Level, and switches off when the measurement exceeds the limit entered in the High Trigger Level field.
7. Select the Error State for the relay.
6.4 Selecting Relay Error State
You can define whether the relay is switched off or on in an error state, or whether the relay remains in the state it is on the moment an error state occurs.
1. Open the Settings > Relays menu in the wireless configuration interface.
2. In the Relays menu, select the relay error state from the Error State dropdown menu: a. To set the relay to release when an error occurs, select Inactive.
b. To set the relay to activate when an error occurs, select Active.
c. To keep the relay in the same state as it was when the error occurred, select No
change.
50
Chapter 7 – Calibration and Adjustment
7. Calibration and Adjustment
7.1 Calibration Overview
The Calibration menu of the wireless configuration interface contains options for calibrating and adjusting the measurement of the probe you have connected to the transmitter. You can also view the current adjustment and restore the probe's factory adjustments.
The calibration menu interface described in this document is in use from Indigo software version 1.2.0 onwards. See manual revision M211877EN-C or earlier for instructions on using the previous version of the interface.
1
2
3 4
Figure 24 Calibration Menu Main View
1 Calibration tab
2 Configuration tab
3 Diagnostics tab
4 Measurements tab
There are 4 tabs in the Calibration menu:
• Calibration: the main adjustment view with options for making adjustments, viewing adjustments, and restoring factory adjustments.
• Configuration: options for using environmental compensations (probe-specific range of options) that allow compensating for the conditions present in the calibration environment, for example, pressure, temperature, and background gases. Also includes probe-specific configuration options that are not mandatory for use with Indigo.
51
Indigo 201 User Guide M211877EN-D
• Diagnostics: this tab contains information about the status of the measurement and the probe, and shows the current environmental compensation configuration.
• Measurements: this tab shows the current probe measurement in numeric format (use this view, for example, when you need to follow measurement stablization in a reference environment without leaving the Calibration menu).
CAUTION!
Before adjusting a probe's measurement, make sure that you have familiarized yourself with the probe-specific calibration requirements such as possible adjustment limits and environmental compensation interdependencies. See the probe's documentation set for probe-specific information.
The range of available options for certain parameters (for example, environmental compensations and the number of adjustment points) varies depending on the features of the connected probe.
The menu examples presented here show the calibration options for Vaisala CARBOCAPâ
Carbon Dioxide Probe GMP252 (ppmCO
2
measurement).
7.1.1 Starting and Closing Calibration Mode
In order to be able to use the calibration options, you must switch the operation of the probe and Indigo to calibration mode with the Start calibration button.
1
Figure 25 Start Calibration Button
1 Start calibration button on the Calibration tab
When you start the calibration mode, the Start calibration button is replaced with the Stop
calibration button. The calibration mode remains active until you close it by selecting Stop
calibration.
You can use other menus while the calibration mode is active, and return to the Calibration menu later to complete your adjustments.
52
Chapter 7 – Calibration and Adjustment
Always close the calibration mode to return the probe and Indigo to normal operating mode. The measurement performance of the probe can be affected when used in calibration mode. You must close the calibration mode with the Stop calibration button also when no changes are made.
More Information
‣ Logging in to Wireless Configuration Interface (page 26)
7.1.2 Restoring Factory Adjustment
Always restore factory adjustment before entering a new adjustment. This prevents any possible earlier adjustments having an effect on the new adjustment you make.
To restore factory adjustment:
1. Connect to the wireless configuration interface and open the Calibration menu.
2. Start the calibration mode with the Start calibration button.
3. On the Calibration tab, scroll down to the parameter you want to adjust (for example,
CO2 adjustment) and select Restore factory adjustment.
4. Restore the factory adjustment with the Restore factory adjustment button for each parameter separately as needed.
5. To verify that the factory adjustment was restored, check the adjustment data information at the bottom of the Calibration tab view.
6. Close the calibration mode with the Stop calibration button.
53
Indigo 201 User Guide
7.1.3 Calibration PIN Code
M211877EN-D
Probe calibration can be locked and unlocked with a calibration PIN code in the Settings >
Probe menu. By default, the calibration PIN code is in place and calibration is enabled. Do not remove the PIN code from the probe settings unless you need to block access to calibration settings.
If the PIN code has been removed and you need to re-enter it, check the probe's documentation for information on the code used in the probe model.
7.2 Environmental Compensation
When making adjustments, you can enter information about the environment in which you perform the adjustment. This allows compensating for environmental factors that have an effect on the measurement (for example, temperature, pressure, or background gases). The selection of environmental compensations available in the Indigo calibration settings depends on the features and configuration of the connected probe.
The environmental compensation selections are available on the Configuration tab of the
Calibration menu. To use environmental compensations, first enable the compensation from the Measurement selections, and then enter the compensation in the Compensation
setpoints fields. You can also set the power-up default compensations that remain in use also after probe reset.
54
Chapter 7 – Calibration and Adjustment
Measurement
Figure 26 Measurement Selections, CO
2
Probe Example
Filtering factor
Temperature compensation mode
Pressure/humidity/oxygen compensation on/off
Defines how much past measurements affect the output
(measurement averaging over time). For details, see probe documentation.
Select whether the probe sensor's measurement or a manually entered setpoint is used to set the temperature compensation.
Enable or disable the environmental compensations.
55
Indigo 201 User Guide
Compensation Setpoints and Power-Up Defaults
M211877EN-D
Figure 27 Compensation Setpoint and Power-Up Default Selection, CO
2
Probe Example
Environmental compensations can have interdependencies: for example, accurate RH measurement requires that also the temperature and pressure configuration match the measurement environment. For more information on the environmental compensation features of the connected probe, refer to the probe's documentation.
Note that the environmental compensations you set in the Settings > Probe menu and the compensations you set in the Calibration menu are interconnected: the configuration set in either menu is applied to both.
More Information
56
Chapter 7 – Calibration and Adjustment
7.3 Measurements Tab
The Measurements tab of the Calibration menu shows the current measurement data in numeric format. The measurement data updates automatically at a 2-second interval.
The information shown in this tab is probe-specific: the example here shows the measurement information for a CO
2
probe.
Figure 28 Measurements Tab, CO
2
Probe Example
7.4 Diagnostics Tab
The Diagnostics tab of the Calibration menu contains status and diagnostics codes. When carrying out diagnostics (for example, contacting Vaisala support), you can identify issues by referring to this information.
In addition to the diagnostics information, this view also shows the environmental compensation values that are currently in use. The status and compensation fields are both read-only.
The information shown on the diagnostics tab is probe-specific: the following figure shows an example of the diagnostics tab information when using a CO
2
probe.
57
Indigo 201 User Guide M211877EN-D
Figure 29 Diagnostics Tab, CO
2
Probe Example
7.5 Configuration Tab
The Configuration tab of the Calibration menu contains probe-specific configuration options and the environmental compensation options.
Note that the probe configuration options available in this view are not always necessary for use with Indigo. For example, the Communication options shown in the following CO
2
probe example apply only to the probe's own communication settings. The communication settings of the Indigo transmitter are configured in a separate menu.
Similarly, the analog output settings shown here apply only for the probe's analog output channels, which are not used when the probe is connected to an Indigo transmitter that has its own output channels and settings.
58
Chapter 7 – Calibration and Adjustment
The information shown in this tab is probe-specific: the example here shows the configuration view for a CO
2
probe.
Figure 30 Configuration Tab, CO
2
Probe Example
7.6 GMP252 Example: 2-point CO
2
Adjustment
To make a 2-point adjustment, you need a low reference and a high reference (select references that are near the low and high ends of your measurement range). Prepare the calibration references (for example, reference gases with known concentrations) before starting the adjustment.
Check the possible probe-specifc adjustment limits and requirements for 2-point calibration (for example, the minimum difference between the low and high reference points) from the probe's documentation.
To make a 2-point adjustment (GMP252 ppmCO
2
example):
1. Connect to the wireless configuration interface and open the Calibration menu.
59
Indigo 201 User Guide M211877EN-D
2. Start the calibration mode with the Start calibration button.
If you cannot enter configurations after selecting Start calibration, check that the calibration PIN code is in place in the Settings > Probe menu.
3. If you need to set environmental compensations, enable and set the required compensations on the Configuration tab.
Note that the environmental compensations you set on the Configuration tab and the compensations you set in the Settings > Probe menu are interconnected: the configuration set in either menu is applied to both.
4. Remove any possible previous adjustments by restoring the factory adjustment: select
Restore factory adjustment for each parameter you are adjusting.
5. Enter the calibration date and calibration information into the corresponding text fields.
60
6. Place the probe in the first reference environment (adjustment point 1) and wait until the measurement has stabilized. You can follow the stabilization from the
Measurements tab.
7. Enter the value of the first reference (for example, 0 if calibrating with a 0 ppmCO
2 reference gas) into the Reference value, point 1 field.
8. After you enter the reference value, the value of the Measured value, point 1 field updates automatically.
Chapter 7 – Calibration and Adjustment
9. Place the probe in the second reference environment (adjustment point 2) and wait until the measurement has stabilized.
10. Enter the value of the second reference (for example, 2000 if calibrating with a 2000 ppmCO
2
reference gas) into the Reference value, point 2 field.
11. When both reference points have been entered, select Store adjustment to save the adjustment.
12. Close the calibration mode with the Stop calibration button.
13. To check that the adjustment was carried out correctly, review the information in the
Adjustment data fields at the bottom of the view.
61
Indigo 201 User Guide M211877EN-D
62
Chapter 8 – Maintenance and Troubleshooting
8. Maintenance and
Troubleshooting
8.1 Cleaning
You can clean the Indigo transmitter body by wiping it with a moist cloth. Standard cleaning agents can be used.
Refer to the probe-specific cleaning instructions when cleaning the probe connected to
Indigo. Do not spray anything directly on the probe connected to Indigo, since that may deposit impurities on the sensor.
Chemical tolerance
The following chemicals can be used to clean Indigo:
• H
2
O
2
(6000 ppm), non-condensing
• Alcohol-based cleaning agents such as ethanol and IPA (70 % Isopropyl Alcohol, 30 % water)
• CaOCI (hypochlorite) max. 0.5 %
• QAC (quaternary ammonium cations) max. 0.5 %)
Avoid exposing the transmitter to chemicals for unnecessarily long periods of time. Do not immerse it in a chemical, and wipe chemicals off the surfaces after cleaning.
8.2 Indigo Wireless Connection
Troubleshooting
Problem
The wireless device has connected to the Indigo access point, but the configuration interface does not launch.
Possible Cause
The device you are using to connect to Indigo does not launch the browser automatically after connecting to the access point.
The wireless connection requires an authentication or acknowledgement before the
Indigo interface opens in your browser.
Remedy
After connecting to Indigo, open your browser application.
Check your device's notifications to see if an authentication or login prompt is present for the Indigo connection. Acknowledge the connection and open your browser application if the interface does not launch automatically.
63
Indigo 201 User Guide M211877EN-D
Problem
The Indigo access point does not show up in your device's list of available WLAN connections.
Indigo shows up in the list of available wireless connections, but connecting to it does not work.
The interface does not open in the browser.
Cannot connect to the Indigo access point with iPhone.
Indigo does not respond when pressing the wireless connection activation button.
Possible Cause Remedy
Indigo access point is not enabled or an error is has occurred.
Your device is too far from the transmitter or obstacles are blocking the signal.
A device is already connected to the Indigo access point.
The browser has issues with loading the landing page.
The iPhone WLAN settings prevent establishing a connection.
WLAN functionality has been disabled with the WLAN ON/OFF
DIP switch on Indigo's circuit board.
Switch off the Indigo WLAN connection, enable the connection again and retry.
Move closer to the transmitter and refresh your device's access point list.
Ensure that your device is the only one that is connecting to the wireless configuration interface.
Enter the default Indigo IP address
http://192.168.1.1 in the browser's address bar.
Connecting to Indigo WLAN with iPhone (page 64) .
Enable WLAN functionality with the WLAN ON/OFF DIP switch. See
the instructions in 3.6 Disabling
.
8.3 Connecting to Indigo WLAN with iPhone
When you are connecting to Indigo's wireless configuration interface with an iPhone, it may be necessary to change the phone's WLAN settings. The following notification is shown on your phone when additional steps are required:
Figure 31 Additional Steps Needed to Connect Notification
64
Chapter 8 – Maintenance and Troubleshooting
Depending on the current settings in your iPhone, you can connect to Indigo by either disabling the Auto-Login feature or with the Use Without Internet option.
8.3.1 Connection Option 1: Use Without Internet
Activate Indigo's wireless (WLAN) configuration interface with the button on the bottom of the transmitter and select Indigo from your phone's list of available WLAN connections.
To connect to Indigo's wireless configuration interface using the Use Without Internet option:
1. When the notification about additional steps being required is shown, select Cancel.
If selecting Cancel returns you to the list of available WLAN connections, select
Indigo from the list and then select Cancel again in the notification screen.
1
200
200
200
2. In the next dialog, select Use Without Internet.
2
65
Indigo 201 User Guide M211877EN-D
3. Open you browser application (for example, Safari).
4. If the Indigo login screen does not launch automatically when you open your browser, enter Indigo's default IP address 192.168.1.1 in the browser's address bar.
8.3.2 Connection Option 2: Disable iPhone Auto-Login
To connect to Indigo's wireless configuration interface by disabling iPhone's Auto-Login:
1. Press the wireless configuration interface activation button on the bottom of the Indigo transmitter.
2. When Indigo's wireless connection has activated, find Indigo from your phone's list of
WLAN connections. Select the info icon next to Indigo.
200
2
3
3. In the next menu view, disable the Auto-Login feature.
4. Return to the list of WLAN connections and select Indigo.
5. Open you browser application (for example, Safari).
6. If the Indigo login screen does not launch automatically when you open your browser, enter Indigo's default IP address 192.168.1.1 in the browser's address bar.
8.4 Display Messages
The following table lists the display messages that Indigo 201 uses to inform you about the transmitter's state.
In addition to the Indigo transmitter messages, the connected probes have probe-specific messages that are also shown on the display. Messages from the connected probe start with
Probe:. For more information on the probe-specific messages, see the probe's documentation.
66
Chapter 8 – Maintenance and Troubleshooting
Table 4 Indigo Transmitter Messages Shown on Display
Description Error Message
Errors
Low supply voltage
No legal measurement probe
Analog output 1 out of range
Analog output 2 out of range
Analog output 3 out of range
WLAN error
The supply voltage is below the minimum (range: 15 ... 30 VDC or
20 ... 22 VAC).
The connected probe is not compatible with Indigo or a restriction (probe name or serial number) on allowed probes is in place.
The output of analog output 1 has gone out of range.
See above
See above
There is a problem with the wireless connection.
Recommended Action
The error clears when the supply voltage returns to the specified range. Check your power supply and restart Indigo if needed.
Check your probe's Indigo compatibility and possible restrictions on allowed probes.
The error clears when the output returns to range. Check the analog output configuration.
See above
See above
Go through the troubleshooting instructions in
Warnings
Outputs configured for different probe
No measurement probe
An analog output configuration already exists in Indigo, and the settings cannot be received automatically from the probe you are connecting.
The measurement probe is not connected, or the probe is connected poorly and cannot be recognized by Indigo.
Configure the analog output settings for the new probe manually, or reset the analog output settings in Indigo to enable automatic receiving of analog output settings.
Check that you have fastened the probe correctly with the locking wheel.
Notifications
Probe connected:
WLAN on
Indigo displays the name of the connected probe.
The WLAN connection is active and you can connect to the wireless configuration interface.
More Information
‣ Analog Output Configuration Overview (page 43)
‣ Receiving Analog Output Settings from Probe (page 14)
67
Indigo 201 User Guide M211877EN-D
68
Chapter 9 – Technical Data
9. Technical Data
9.1 Specifications
Table 5 Input and Output
Property
Analog outputs
Accuracy of analog outputs at 20 °C
Relays
Maximum current
Power consumption
Probe connector
Cable feed throughs
Screw terminal wire size
Specification
3 voltage (V) or current (mA) outputs:
• 0 … 10 V / 0 … 5 V / 0 … 1 V / 1 … 5 V (min. load
1kΩ)
• 0 … 20 mA / 4 … 20 mA (max. load 500 Ω)
± 0.1% full scale
2 configurable relays (VAC/VDC)
Device maximum specification (resistive load):
• Max. switching power 30 W / 37.5 VA
UL-rated maximum specification (resistive load):
• AC: max. 28 V / 0.5 A
• DC: max. 40 V / 0.24 A
• Up to 30 VDC:
• max. switching current 1 A
• max. switching power 30 W
Range 15 ... 30 VDC (20 ... 22 VAC)
Transmitter and connected probe max. 1 A
Transmitter max. 3 W (+ connected probe, varies depending on probe type)
M12/5 connector for probe or probe cable connection
(Vaisala Indigo-compatible probes)
2 options: rubber lead-through on the bottom of the transmitter, and opening with a seal at the back of the transmitter
0.2 mm
2
... 1.5 mm
2
1) Using a power supply with overload protection is recommended for electrical safety.
Table 6 General
Display (optional) 3.5” TFT LCD color display
69
Indigo 201 User Guide M211877EN-D
Configuration interface
Wireless configuration interface browser support
Browser-based wireless configuration interface (IEEE
802.11 b/g/n WLAN)
• Microsoft Internet Explorer: version 11.0 onward
• Google Chrome: version 57 onward
• Mozilla Firefox: version 50 onward
• Apple Safari: version 10 onward
Table 7 Standards and Compliance
Property
Safety standard
Networking standards (wireless configuration interface WLAN access point)
Electromagnetic compatibility
Specification
IEC/UL/EN 61010-1
IEEE 802.11 b/g/n compliant
Contains
Complies with EMC standard EN61326-1 Generic
Environment
FCC ID QOQ-WGM110
IC 5123A-WGM110
当該機器には電波法に基づく、技術基準適合証明等
を受けた特定無線設備を装着している。
CMIIT ID
R 209-J00197
2017DJ3476
Table 8 Wireless Access Point (Module With Internal Chip Antenna)
Property
Networking standards
Data rates
Frequency band
Modulation
Security
Output power
Receiver sensitivity
Specification
IEEE 802.11 b/g/n
802.11 b: 1, 2, 5.5,11 Mbps : 802.11 g: 6, 9, 12, 18
2402 ~ 2480MHz
802.11 b: DSSS (CCK-11, CCK-5.5, DQPSK-2,DBPSK-1):
802.11g : OFDM Wi-Fi
WEP (128-bit), WPA, WPA2 (Personal)
+16DBm
-85dBm typical
Table 9 Operating and Storage Environment
Property
Operating temperature range
Specification
Model without display: -40 ... + 60 °C (-40 ... + 140 °F)
Model with display: -20 ... + 60 °C (-4 ... + 140 °F)
70
Chapter 9 – Technical Data
Property
Storage temperature range
Operating humidity range
Chemical tolerance
Specification
-40 ... +70 °C (-40 ... 158 °F)
0 ... 100 %RH (non-condensing)
Temporary exposure during cleaning:
• H
2
O
2
(6000 ppm, non-condensing)
• Alcohol-based cleaning agents such as ethanol and
IPA (max. 70 % concentrate)
• CaOCI (hypochlorite) max. 0.5 %
• QAC (quaternary ammonium cations) max. 0.5 %)
Table 10 Mechanics
Property
Housing classification
Housing color
Housing material
Display window material
Connection screw terminals
Plastic material flammability (UL rating)
Weight
Dimensions (H×W×D)
Input/output cable recommended diameter (with cable gland strain relief)
Specification
IP65
White (RAL9003)
PC/ABS plastic
PMMA plastic
26 AWG ... 20 AWG
UL94 HB
402 g
149×135×43 mm (5.87×5.31×1.7 inch)
7 ... 8 mm (0.25 ... 0.31 inch)
9.2 Spare Parts and Accessories
Information on spare parts, accessories, and calibration products is available online at www.vaisala.com
and store.vaisala.com
.
Table 11 Spare Parts and Accessories
Description
Probe connection cable, 1 m
Probe connection cable, 3 m
Probe connection cable, 5 m
Probe connection cable, 10 m
Order Code
INDIGOCABLE1M
INDIGOCABLE3M
INDIGOCABLE5M
INDIGOCABLE10M
71
Indigo 201 User Guide
9.3 Dimensions (in mm)
135 mm 43 mm
M211877EN-D
50 mm
Ø
4
.0
mm
Secondary mounting option with 4 screws: designed for Würth 4.0×1.8 mm Plastofast
(self-tapping screws for plastics).
Hole depth 12 mm.
Ø
4
.5
mm
25 mm
114 mm
2.5 mm
Figure 32 Indigo Transmitter Dimensions in Millimeters (mm)
72
Chapter 9 – Technical Data
73
Indigo 201 User Guide M211877EN-D
74
EU Declaration of Conformity
BG: С настоящото Vaisala Oyj декларира, че този тип радиосъоръжение Indigo 201 е в съответствие с Директива 2014/53/ЕС. Цялостният текст на ЕС декларацията за съответствие може да се намери на следния интернет адрес: www.vaisala.com/indigo
CS: Tímto Vaisala Oyj prohlašuje, že typ rádiového zařízení Indigo 201 je v souladu se směrnicí 2014/53/EU. Úplné znění EU prohlášení o shodě je k dispozici na této internetové adrese: www.vaisala.com/indigo
DA: Hermed erklærer Vaisala Oyj , at radioudstyrstypen Indigo 201 er i overensstemmelse med direktiv 2014/53/EU. EU-overensstemmelseserklæringens fulde tekst kan findes på følgende internetadresse: www.vaisala.com/indigo
DE: Hiermit erklärt Vaisala Oyj , dass der Funkanlagentyp Indigo 201 der Richtlinie
2014/53/EU entspricht. Der vollständige Text der EU-Konformitätserklärung ist unter der folgenden Internetadresse verfügbar: www.vaisala.com/indigo
EL:Με την παρούσα ο/η Vaisala Oyj , δηλώνει ότι ο ραδιοεξοπλισμός Indigo 201 πληροί την
οδηγία 2014/53/ΕΕ. Το πλήρες κείμενο της δήλωσης συμμόρφωσης ΕΕ διατίθεται στην
ακόλουθη ιστοσελίδα στο διαδίκτυο: www.vaisala.com/indigo
EN: Hereby, Vaisala Oyj declares that the radio equipment type Indigo 201 is in compliance with Directive 2014/53/EU. The full text of the EU declaration of conformity is available at the following internet address: www.vaisala.com/indigo
ES: Por la presente, Vaisala Oyj declara que el tipo de equipo radioeléctrico Indigo 201 es conforme con la Directiva 2014/53/UE. El texto completo de la declaración UE de conformidad está disponible en la dirección Internet siguiente: www.vaisala.com/indigo
ET: Käesolevaga deklareerib Vaisala Oyj , et käesolev raadioseadme tüüp Indigo 201 vastab direktiivi 2014/53/EL nõuetele. ELi vastavusdeklaratsiooni täielik tekst on kättesaadav järgmisel internetiaadressil: www.vaisala.com/indigo
FI: Vaisala Oyj vakuuttaa, että radiolaitetyyppi Indigo 201 on direktiivin 2014/53/EU mukainen. EU-vaatimustenmukaisuusvakuutuksen täysimittainen teksti on saatavilla seuraavassa internetosoitteessa: www.vaisala.com/indigo
FR: Le soussigné, Vaisala Oyj , déclare que l'équipement radioélectrique du type Indigo 201 est conforme à la directive 2014/53/UE. Le texte complet de la déclaration UE de conformité est disponible à l'adresse internet suivante: www.vaisala.com/indigo
HR: Vaisala Oyj ovime izjavljuje da je radijska oprema tipa Indigo 201 u skladu s Direktivom
2014/53/EU. Cjeloviti tekst EU izjave o sukladnosti dostupan je na sljedećoj internetskoj adresi: www.vaisala.com/indigo
HU: Vaisala Oyj igazolja, hogy a Indigo 201 típusú rádióberendezés megfelel a 2014/53/EU irányelvnek. Az EU-megfelelőségi nyilatkozat teljes szövege elérhető a következő internetes címen: www.vaisala.com/indigo
IT: Il fabbricante, Vaisala Oyj , dichiara che il tipo di apparecchiatura radio Indigo 201 è conforme alla direttiva 2014/53/UE. Il testo completo della dichiarazione di conformità UE è disponibile al seguente indirizzo Internet: www.vaisala.com/indigo
75
Indigo 201 User Guide M211877EN-D
LT: Aš, Vaisala Oyj , patvirtinu, kad radijo įrenginių tipas Indigo 201 atitinka Direktyvą
2014/53/ES. Visas ES atitikties deklaracijos tekstas prieinamas šiuo interneto adresu: www.
www.vaisala.com/indigo
LV: Ar šo Vaisala Oyj deklarē, ka radioiekārta Indigo 201 atbilst Direktīvai 2014/53/ES. Pilns
ES atbilstības deklarācijas teksts ir pieejams šādā interneta vietnē: www.vaisala.com/indigo
MT: B'dan, Vaisala Oyj , niddikjara li dan it-tip ta' tagħmir tar-radju Indigo 201 huwa konformi mad-Direttiva 2014/53/UE. It-test kollu tad-dikjarazzjoni ta' konformità tal-UE huwa disponibbli f'dan l-indirizz tal-Internet li ġej: www.vaisala.com/indigo
NL: Hierbij verklaar ik, Vaisala Oyj , dat het type radioapparatuur Indigo 201 conform is met
Richtlijn 2014/53/EU. De volledige tekst van de EU-conformiteitsverklaring kan worden geraadpleegd op het volgende internetadres: www.vaisala.com/indigo
PL: Vaisala Oyj niniejszym oświadcza, że typ urządzenia radiowego Indigo 201 jest zgodny z dyrektywą 2014/53/UE. Pełny tekst deklaracji zgodności UE jest dostępny pod następującym adresem internetowym: www.vaisala.com/indigo
PT: O(a) abaixo assinado(a) Vaisala Oyj declara que o presente tipo de equipamento de rádio Indigo 201 está em conformidade com a Diretiva 2014/53/UE. O texto integral da declaração de conformidade está disponível no seguinte endereço de Internet: www.vaisala.com/indigo
RO: Prin prezenta, Vaisala Oyj declară că tipul de echipamente radio Indigo 201 este în conformitate cu Directiva 2014/53/UE. Textul integral al declarației UE de conformitate este disponibil la următoarea adresă internet: www.vaisala.com/indigo
SK: Vaisala Oyj týmto vyhlasuje, že rádiové zariadenie typu Indigo 201 je v súlade so smernicou 2014/53/EÚ. Úplné EÚ vyhlásenie o zhode je k dispozícii na tejto internetovej adrese: www.vaisala.com/indigo
SL: Vaisala Oyj potrjuje, da je tip radijske opreme Indigo 201 skladen z Direktivo 2014/53/EU.
Celotno besedilo izjave EU o skladnosti je na voljo na naslednjem spletnem naslovu: www.vaisala.com/indigo
SV: Härmed försäkrar Vaisala Oyj att denna typ av radioutrustning Indigo 201
överensstämmer med direktiv 2014/53/EU. Den fullständiga texten till EU-försäkran om
överensstämmelse finns på följande webbadress: www.vaisala.com/indigo
Warranty
For standard warranty terms and conditions, see www.vaisala.com/warranty .
Please observe that any such warranty may not be valid in case of damage due to normal wear and tear, exceptional operating conditions, negligent handling or installation, or unauthorized modifications. Please see the applicable supply contract or Conditions of Sale for details of the warranty for each product.
76
Technical Support
Contact Vaisala technical support at [email protected]
. Provide at least the following supporting information:
• Product name, model, and serial number
• Name and location of the installation site
• Name and contact information of a technical person who can provide further information on the problem
For more information, see www.vaisala.com/support .
Recycling
Recycle all applicable material.
Follow the statutory regulations for disposing of the product and packaging.
77
Indigo 201 User Guide M211877EN-D
78
www.vaisala.com
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Related manuals
advertisement
Table of contents
- 7 About This Document
- 7 Version Information
- 7 Related Manuals
- 8 Documentation Conventions
- 8 Trademarks
- 9 2. Product Overview
- 9 Introduction to Indigo™ Transmitters
- 10 Indigo 201 Basic Features and Options
- 10 Probe Compatibility
- 11 Indigo Transmitter Parts
- 12 Display and LED Indicator
- 13 Graph Display Mode
- 14 LED Indicator
- 15 Wireless Configuration Interface Overview
- 16 Receiving Analog Output Settings from Probe
- 17 3. Installation
- 17 Mounting
- 18 Indigo Transmitter Base
- 19 Wiring Options
- 20 Attaching Probes and Cables
- 21 Input and Output Specification
- 22 Disabling WLAN Functionality with DIP Switch
- 25 4. Wireless Configuration Interface
- 25 Wireless Interface Menus
- 26 Connecting to Wireless Configuration Interface
- 26 Connecting with a Computer
- 27 Connecting with a Mobile Device
- 28 Logging in to Wireless Configuration Interface
- 29 Changing User Level
- 29 Changing Administrator Password
- 30 Resetting Administrator Password
- 31 Measurements View
- 32 Status View
- 33 Status View: General
- 34 Status View: Analog Outputs
- 34 Status View: Relays
- 35 Calibration Menu
- 35 General Settings
- 36 WLAN Settings
- 36 User Access Settings (Administrator Password)
- 37 Display Settings
- 38 Factory Reset and Clearing Analog Output Settings
- 39 License Information
- 39 Output Settings
- 41 Relay Settings
- 42 4.10 Probe Settings
- 45 5. Configuring Analog Outputs
- 45 Analog Output Configuration Overview
- 46 Analog Output Configuration Example
- 47 Clearing Analog Output Settings from Indigo
- 47 Analog Output Overrange Behavior
- 47 Analog Output Overrange Example
- 49 6. Configuring Relays
- 49 Relay Configuration Overview
- 50 Setting Relay Activation Limit Without Hysteresis
- 51 Setting Relay Activation Limit Using Hysteresis
- 52 Selecting Relay Error State
- 53 Calibration and Adjustment
- 53 Calibration Overview
- 54 Starting and Closing Calibration Mode
- 55 Restoring Factory Adjustment
- 56 Calibration PIN Code
- 56 Environmental Compensation
- 59 Measurements Tab
- 59 Diagnostics Tab
- 60 Configuration Tab
- 61 Adjustment
- 65 8. Maintenance and Troubleshooting
- 65 Cleaning
- 65 Indigo Wireless Connection Troubleshooting
- 66 Connecting to Indigo WLAN with iPhone
- 67 Connection Option 1: Use Without Internet
- 68 Connection Option 2: Disable iPhone Auto-Login
- 68 Display Messages
- 71 9. Technical Data
- 71 Specifications
- 73 Spare Parts and Accessories
- 74 Dimensions (in mm)
- 77 EU Declaration of Conformity
- 78 Warranty
- 79 Technical Support
- 79 Recycling