Vermeiren Forest 3+ Total comfort Maintenance Manual


Add to my manuals
20 Pages

advertisement

Vermeiren Forest 3+ Total comfort Maintenance Manual | Manualzz

VERMEIREN

Maintenance manual

Beds

Beds

2014-09

Contents

1   Maintenance .......................................................................................................................... 2  

1.1

  Regular Maintenance .................................................................................................. 2  

1.2

  Shipping and storage .................................................................................................. 6  

1.3

  Care ............................................................................................................................. 7  

1.4

  Inspection .................................................................................................................... 8  

1.5

  Troubleshooting ........................................................................................................ 12  

2  

3  

4  

5  

1.6

  Disinfection .............................................................................................................. 12  

Warranty ............................................................................................................................... 14  

Disposal ................................................................................................................................ 14  

Maintenance plan ............................................................................................................... 15  

Disinfection book ............................................................................................................... 15  

Page 1

Beds

2014-09

1 Maintenance

Lifetime of the bed is strongly influenced by its use, storage, regular maintenance, servicing and cleaning. Do not use the bed when something is damaged, therefore read the maintenance instructions carefully. Do not perform repairs on your own.

No persons may be on the bed when performing any maintenance activities.

The manufacturer is not liable for damage caused by lacking or improper service checks.

The maintenance activities to ensure a good condition for your Vermeiren bed are described below:

 Before every use

 Check the castors for visible damage and/or soiling. Remove any dirt as it could impair the motion of the castors. When a castor is damaged, please go see an authorized workshop for repair.

 Check the brakes and replace if necessary.

 See that all the screws, locking pins are tightened properly.

 Check the bed (cleanliness, cracks, damage on structural parts ...) and make it clean. If necessary change the coating.

 Check the functioning of the mechanically adjustable elements for wear and tear

(side railings, lifting pole + triangle, foot-end, head-end).

 Check the condition of the wooden parts.

Only for electrical bed:

 Check the cables for visible signs of damage and whether all plugs are properly inserted as indicated in the instruction manual. In the case of damage immediately unplug the power cable out the wall socket. Plugs must only be pulled out by grasping the plug itself; never pull on the cable. Contact your authorized workshop for repairs.

 Check that the handset function properly. If a function is faulty, kindly disconnect the power cable and contact your specialist dealer immediately.

Only for mechanical bed:

 Check functioning of the gas pressure spring (synchronized, deformation, oiled).

 Check functioning for the mechanical tooth adjustment mechanism (deformation, wear).

 Every 8 weeks: inspection and possibly lubricating or adjusting from (depending on how often the bed is used)

 Operation and condition of the wooden side railings (headrest and foot-end)

 Operation and condition of the metal side railings (if present)

 Placement of the slats in the slatted frame (if available)

 The condition of the wooden parts

 Lifting pole (fit / deformation)

 All screw, bolt and nut fasteners

 Lubricating from moving parts

Page 2

Figures below indicate the greasing points for the different Vermeiren beds:

Alois, Alzeis, Lo-Lo (wooden side rails)

Page 3

Beds

2014-09

Interval

Beds

2014-09

Page 4

Pratic (metal side rails)

Pratic (wooden side rails)

Beds

2014-09

Page 5

Pratic (wooden side rails)

Same for the Pratic HC bed.

 Every 6 months or for every new user

Beds

2014-09

For your convenience, on the back of this manual is a maintenance plan.

Repairing and assembling of spare parts for your bed may only be performed by the specialist dealer.

Only authorized Vermeiren spare parts may be installed.

1.2 Shipping and storage

The shipping and storage of the bed shall be according following instructions:

 Store in a dry place (between +5 °C and +40 °C).

 The relative humidity of the air should be between 30% and 70%.

 Provide sufficient covering or packaging to protect the bed from rust and foreign bodies. (e.g. salt water, sea air, sand, dust).

 Pull the power cable out the wall socket.

 Carefully roll up the power cable with no kinks and place it in the receptacle provided.

 Store all removed parts together in one place (or mark them if necessary) to avoid mixing up with other products when re-assembling (e.g. side rails).

 Components must be stored without being subjected to strains (Do not put too heavy parts on the bed, not clamping between something, ...). Grasp the bed by solid frame parts only.

Page 6

Beds

2014-09

 Check internal cables for squashing and the prevention of kinks.

 When dismantling properly tighten all screws and bolts.

 Make sure all components required for assembling the nursing bed are in place.

 After being re-assembled and passed on to somebody else, all inspections must be performed as described in the "Inspection" chapter.

1.3 Care

The cleaning of your bed is possible with household products. Check before cleaning the bed the power cable is disconnected from the power supply.

Your bed requires regular care if you want it to remain visually attractive so it should be regularly cleaned.

1.3.1 Metal frames, wooden parts

See instructions below for the cleaning of the metal frames and the wooden parts:

 Clean the metal frames and the wooden parts with a cloth moistened in warm water.

Take care that you do not soak the metal frames and the wooden parts.

 Use a mild commercial detergent for removing stubborn dirt.

 Stains can be removed by using a sponge or a soft brush.

 Do not use strong cleaning liquids like solvents, nor use hard brushes.

 Never clean with steam and/or pressure washers.

Damage caused by the use of improper cleaning agents is not the responsibility of

Vermeiren.

1.3.2 Plastic parts

Clean plastic parts of your bed with commercial plastic cleaners. Only use a soft brush or soft sponge. Take note of special product information.

Examples of plastic parts is the handset, etc.

1.3.3 Coating

The high quality of the surface layer guarantees optimal protection against corrosion. If the outer coating is damaged by scratching or in some other way, get your specialist dealer to repair the affected surface.

When cleaning, only use warm water and normal household detergents and soft brushes and cloths. Ensure that no moisture gets into the components (tubes, electronics, ...).

Initially the chrome parts only require rubbing with a dry cloth. Dull places or stubborn dirt can best be removed by using a suitable commercial chrome polish.

Dulling can be avoided by lightly greasing steel parts with Vaseline.

1.3.4 Electronics

L

WARNING:

Check the plug connectors for corrosion or damage before every ride,

since it could affect the efficiency of the electronics.

Wipe the handset, motors with a cloth moistened by a few drops of a commercial domestic cleaner. Do not use any abrasives or sharp-edged polishing equipment like a metal scrubber or brush, since these can scratch the surface of the handset.

Page 7

Beds

2014-09

1.4 Inspection

For your safety and to ensure that your nursing bed remains functional even after prolonged use, it should be serviced regularly by an authorized specialist dealer whose work should be documented in the service plan. The dealer will enter maintenance work in the service plan.

In general we recommend having annual servicing and servicing before each new usage.

The following inspections need to be documented:

 Visually check the frame parts for plastic deformation, cracks and impaired functioning.

(slatted frame, undercarriage, head-end, feet-end, lifting pole, attachment lifting pole, wheels).

 Visually check all plastic parts for cracks and brittle spots.

 Check the solidity and seating of all screws, nut connections.

 Visually check all housings for damage, screws must be securely fixed, seals and gaskets should not exhibit visible damage.

 Visually check for damage to the paintwork (danger of corrosion).

 Check the plug connectors for corrosion or damage, since it could affect the efficiency of the electronics.

 Check the operation of the braking wheels (free running, safety, locking, etc.).

 Verify the amount of grease on the metal joints of movable parts.

 Measure the resistance of the protective earth (Ohm) according to VDE EN62353.

 Measure the residual diversion current (A) according to VDE EN62353.

 Measure the insulation resistance (MOhm) according to VDE EN62353.

 Check the functioning of the side railings (locking, distances according to ISO 60601-2-

52: 2009, also when loaded, deformation and wear through loading).

 Check the condition of the mattress (not when use of the bed is resumed).

 Completeness of the delivery condition, instruction manual available?

Only for electrical bed:

 Condition of the power cable (especially: squashing, abrasion, cuts, visible insulation or inner conductors, visible metallic veins, kinks, bulges, color changes of the outer covering, brittle places).

 Check the internal cabling of the power cable (especially: crushing, abrasion, cuts, visible insulation of the inner conductor, visible metallic veins, kinks, lumpiness, color changes of the outer sleeve, brittle spots).

 Visual inspection of the power cable (EPR or similar).

 Humidity protection of the electric system, at least IPX4 (see Type label).

 Check the device for suspending the power cable during transportation.

 The power cable and other electric cables to be securely placed to avoid shearing, crushing and other mechanical stresses and strains.

 Check the operation of the drives (over the full adjustment range of the motors  noises, speed, free running, etc., final switching (e.g. current limiting, main switch), if necessary:

Measure the performance, first with no load, and then with the nominal load ("SWL") to investigate the wear and tear of the motors by comparing the values of the electric current with the values when the bed was delivered. Do not open the motors otherwise the warranty will lapse immediately.

Only for mechanical bed:

 Check functioning of the gas pressure spring (synchronized, deformation, oiled).

 Check functioning for the mechanical tooth adjustment mechanism (deformation, wear).

Page 8

Beds

2014-09

Electrical measurements may only be carried out by personnel trained on the nursing bed and instructed by an electrician about the instruments and procedures to be used.

Only an electrician can approve the nursing bed for use after making the

measurements and performing maintenance.

Servicing must be confirmed in the service plan when at least the above-mentioned aspects have been checked. Should your specialist dealer not perform the services you request, then contact the manufacturer. We shall gladly refer you to authorized specialist dealers in your area.

The manufacturer is not liable for damage caused by the lack of or improper servicing.

Requirement:

1.) Visual check:

1.1) Power cable

EPR cable or comparable quality

1.2) Condition of power cable

In particular look for: Crushing, abrasion, cuts, visible insulation of the inner conductor, visible metallic veins, kinks, lumpiness, color changes of the outer sleeve, brittle spots)

1.3) Status of internal wiring

In particular look for: Crushing, abrasion, cuts, visible insulation of the inner conductor, visible metallic veins, kinks, lumpiness, color changes of the outer sleeve, brittle spots)

1.4) Sufficient ease of pulling and power cable kink protection

1.5) Sufficient ease of pulling and kink protection for internal cables

1.6) The power cable and other electric leads must be wired securely so that shearing, squashing, and other forms of mechanical stress are unlikely.

1.7) Moisture protection IP66

1.8) Inspecting device for suspended power cables during transport.

1.9) Check the frame parts for plastic deformation and/or wear and tear.

- undercarriage

- bed lift

- head-end, seat area, feet-end

- lifting pole

- fastening of lifting pole

- rolling castors

1.10) Housing

Visual inspection of all housings for damage. The screws need to be tightly fastened. The seals must not have any visual damage

Remark: n.a. Yes No Notes n.a. Yes No Notes n.a. Yes No Notes n.a. Yes No Notes n.a. Yes No Notes n.a. Yes No Notes n.a. Yes No Notes n.a. n.a. n.a.

Yes

Yes

Yes

No

No

No

Notes

Notes

Notes n.a. Yes No Notes

Page 9

Distance between side railings taken apart with the lying surface in flat position.

Smallest distance of all accessible openings between the side railing and the lying surface.

Beds

2014-09

Table (A) on Picture (B)

Symbol

A

B

C

D

E

F

G

H

Size

The smallest distance between the elements within the circumference of the side railings in an erect/inserted position, or the area included by the side railings and fixed parts of the bed.

Thickness of the mattress in accordance with regulatory use

Height of the top sides of the side railings over the mattress (see ‘B’) without compression.

Distance between the headrest or food part and the side railings.

Overall length of the side railings or sum total of the lengths of the side railings taken apart on one side of the bed.

Distance between headrest and foot-end without extension of these parts.

Requirement in mm

A

 120

As indicated by manufacturer

C

 220

D

 60 or

D

 235

E

 60 or

E

 235

If D

 235 then F 

60

If D

 60 then

F

 120

G

 1/2 of the length of the lying surface

No requirements

Actual size: mm

Actual size: mm

Actual size: mm

Actual size: mm

Actual size: mm

Actual size: mm

Actual size: mm

Actual size: mm

NA Yes No Notes

Beds whose side railings exceed the indicated dimensions by less than 10 % must not be used with small or underweight patients, otherwise additional measures must be taken to avoid having such patients being clamped in the bed or sliding through.

Side railings that exceed the indicated measurements by 10

% or more may not be used until the deficiencies have been removed and the rails must be exchanged or upgraded.

In addition to the following points, the full functionality (i.e. locking, fixing elements) of the side railings must be inspected. Any defects must be eliminated. Keeping the sizes for the side railing components in compliance with the following sketch. n.a. Yes No Notes n.a. Yes No Notes n.a. Yes No Notes

Page 10

Picture (B) on Table (A)

H

D

G

C

A

D

Beds

2014-09

Remark: When the bar for the side railings are external (dark gray marking), they need to be kept at distance A to the lying surface.

A F

B

2.) Measurement tests

2.1) Protective earth resistance

Actual size: Ohm

2.2) Residual diversion current

Actual size: A

2.3) Insulation resistance

Actual size: MOhm

3.) Function tests

3.1) Functional check of side railings:

- locking, e.g. play in the locking device

- distances, including under load

- deformation

- wear and tear owing to load

3.2) Functional check of brakes:

- safety

- braking neutral

3.3) Functional check of drives:

- move through complete adjustment scope for the motors

(sounds, speed, smooth operations, etc.) - Final switch off (e.g. current limitation, main switch)

Measure performance, first with no load, and then with the nominal load ("SWL") to inspect for possible wear and tear on the motors by comparing electrical current input.

Compare with the values given at the time of delivery. The manufacturer will provide you with the limits.

Other: n.a. Yes No Notes n.a. Yes No Notes n.a. Yes No Notes n.a. Yes No Notes n.a. Yes No Notes n.a. Yes No Notes n.a. Yes No Notes n.a. Yes No Notes

Page 11

Beds

2014-09

1.5 Troubleshooting

The following list will assist you with the troubleshooting of your bed. If you cannot found any fault please contact your specialist dealer.

Problem

Power light indicator of the control box is not working.

Power light indicator of the control box does not lit, motor is not working, relay in control box make a clicking sound.

Power light indicator of the control box does not lit, motor is not working, relay in control box make no clicking sound.

Power light indicator is lit, motor is working, but spindle is not working.

Power light indicator is lit, motor and spindle are not working.

Controle box is working, motor works only in one direction.

Cause of problem

Not connected to the power supply.

Fuse is damaged.

Control box is damaged.

Motor plug is not connected very well in the control box.

Motor is damaged.

Control box is damaged.

Control box is damaged.

Handset is damaged.

Worm gear is loose or worn.

Supply voltage is too low.

Bed is overloaded.

Cables are not good connected.

Cables are broken.

Control box is damaged.

Handset is damaged.

Motor is damaged.

Handset is damaged.

Solution

Connect the control box to the power supply.

Send the motor or control box for reparation.

Connect the motor plug in the control box.

Send the motor or control box for reparation.

Send the control box or handset for reparation.

Send the motor for reparation.

Supply voltage must be 220V AC.

Provide less weight on the bed, and let the motors rest.

Connect the cables.

Change the cable.

Send control box, handset or motor for reparation.

Send handset for reparation.

1.6 Disinfection

L

WARNING:

Dangerous Products - The use of disinfectants is restricted to

L

WARNING:

Dangerous Products, risk of skin irritation - Wear suitable protective clothing. Take note of the product information of the solutions concerned.

All parts of the bed can be treated by scrubbing with a disinfectant.

All steps taken to disinfect rehabilitation equipment, their components or other accessory parts are to be recorded in a disinfection report containing a minimum of the following information (with product documentation appended):

Date of the disinfection concentration

Table 1: Example of a disinfection book

Abbreviations used in column 2 (reason):

V = Suspected infection IF = Infection case W = Repetition I = Inspection

For a blank sheet of a disinfection book see § 5.

The recommended disinfectants for scrubbing (based on the list provided by the Robert Koch

Institute, RKI) are standing in the table below. The current state of the disinfectants included in the RKI list can be obtained from the Robert Koch Institute (RKI) (homepage: www.rki.de).

Page 12

Beds

2014-09

Active substance

Product name Laundry disinfection

Surface disinfection

(scrubbing-

/wiping disinfection)

Disinfection of excretions

1 part sputum or stools + 2 parts diluted solution or 1 part urine + 1 part diluted solution

Sputum Stools Urine

Area of effectiveness

Manufacturer or

Supplier

Phenol or phenol derivative

Chlorine, organic or inorganic substances with active chlorine

Amocid

Gevisol

Helipur m-cresylic soap solution (DAB

6)

Phenol

Chloramin-T

DAB 9

Clorina

Trichlorol

Apesin AP100

2

%

1

0,5

1

1

1,5

1, 5

2

Hr.

12

12

12

12

12

12

12

%

5

5

6

5

3

2,5

2,5

3

Hr.

6

4

4

4

2

2

2

2

%

5

5

6

5

5

6

Hr.

4

4

4

4

4

4

% Hr.

5 6

5

6

6

6

%

5

5

6

Hr.

2

2

2

A

A

A

A

A

A

1

B

A

1

B

A

1

B

Lysoform

Schülke & Mayr

B. Braun

Lysoform

Lysoform

Per combinations

Formaldehyde and/or other aldehydes or derivatives

Dismozon pur

2

Perform

2

Wofesteril

2

Aldasan 2000

Antifect FD 10

Antiseptica surface disinfection 7

Apesin AP30

4

4

3

2

4

3

3

4

1

4

4

4

4

6

AB

AB

AB

AB

AB

AB

AB

Tana

PROFESSIONAL

Bode Chemie

Schülke & Mayr

Kesla Pharma

Lysoform

Schülke & Mayr

Antiseptica

Bacillocid special

Buraton 10F

Desomed A

2000

Hospital disinfectant cleaner

Desomed

Perfekt

Formaldehydesolution (DAB

10), (formaline)

Incidin Perfekt

Incidin Plus

Kohrsolin

Lysoform

Lysoformin

Lysoformin

2000

Melsept

Melsitt

Minutil

Multidor

Nûscosept

1,5

1

2

4

3

12

12

12

12

12

5

6

3

3

8

7

3

3

8

3

5

5

4

4

4

4

6

6

4

4

4

6

4

6

6

6

A

AB

AB

AB

AB

AB*

AB

AB

A

AB

AB

AB

AB

Tana

PROFESSIONAL

Bode Chemie

Schülke & Mayr

Desomed

Dreiturm

Desomed

Ecolab

Ecolab

Bode Chemie

Lysoform

Lysoform

Lysoform

Optisept

Pursept-FD

Ultrasol F

Tensodur 103

2

4

2

3

2

12

12

12

12

12

4

10

6

3

5

7

7

5

6

4

4

6

4

4

4

4

AB

AB

AB

AB

AB

AB*

AB*

AB

Amphoteric surfactants

(amfotensiden)

Lye Lime-milk

3

20 6

A

A

3

B

1 Not effective against myco-bacteria when service disinfecting, especially in the presence of blood.

2 Not suitable for disinfecting blood-contaminated or porous surfaces (e.g. raw wood).

3 Useless for tuberculosis; preparation of Lime-milk: 1 part dissolved lime (calcium hydroxide) + 3 parts water.

* Checked for effectiveness on viruses in accordance with checking methods of the RKI (Federal Health Reporting 38 (1995) 242).

A: Suitable for killing vegetative bacterial germs including myco-bacteria as well as fungi, including fungal spores.

B: Suitable for deactivating viruses.

Table 2: Disinfectants

B. Braun

B. Braun

Ecolab

Ecolab

Dr. Nüsken

Chemie

Dr. Schumacher

Merz

Fresenius Kabi

MFH Marienfelde

Kindly consult your specialist dealer if you have queries on matters related to disinfection; he will gladly assist you.

Page 13

Beds

2014-09

2 Warranty

Excerpt from the "General Business Conditions":

(...)

5. The terms of the guarantee may differ from country to country. Consult your specialist dealer for the guarantee period for warranty claims.

(…)

Claims are invalid when:

- the agreed condition is only minimally deviated from

- usability is only minimally impaired

- wear and tear is normal

- faults are due to improper assembly or neglected maintenance

- faults are due to improper use

- Faults are due to improper handling by customers or their contractual partners or

workshops

- for damage resulting from faulty or negligent treatment, excessive wear, inappropriate

equipment, faulty installation or due to special influences not included in the contract

after the transfer of goods.

(…)

The terms of the guarantee may different from country to country. Please contact your dealer for further information.

3 Disposal

A local waste disposal company will tell you where you can take the nursing bed to be recycled free of charge. It may not be disposed of with domestic waste.

The manufacturer is responsible for taking back and recycling the bed and shall meet the requirements of European Directive 2002/96/EC on Waste Electrical and

Electronic Equipment.

Packaging materials can be taken to disposal or recycling centers.

Page 14

Beds

2014-09

Date Maintenance

1/1/2014 Greasing and general servicing non

Remarks

Date of the disinfection

Reason Specification

Abbreviations used in column 2 (reason):

V = Suspected infection IF = Infection case W = Repetition I = Inspection

Substance and concentration

Signature

Paraph

Page 15

Beds

2014-09

SERVICE

The bed was serviced:

Dealer´s stamp:

Date:

Dealer´s stamp:

Date:

Dealer´s stamp:

Date:

Dealer´s stamp:

Date:

Dealer´s stamp:

Date:

Dealer´s stamp:

Date:

Dealer´s stamp:

Date:

Dealer´s stamp:

Date:

Dealer´s stamp:

Date:

Dealer´s stamp:

Date:

 For service checklists an additional technical information, please see our specialist dealers nearest to you.

More information on our website at: www.vermeiren.com.

Belgium

N.V. Vermeiren N.V.

Vermeirenplein 1 / 15

B-2920 Kalmthout

Tel: +32(0)3 620 20 20

Fax: +32(0)3 666 48 94 website: www.vermeiren.be e-mail: [email protected]

France

Germany

Vermeiren Deutschland GmbH

Wahlerstraße 12 a

D-40472 Düsseldorf

Tel: +49(0)211 94 27 90

Fax: +49(0)211 65 36 00 website: www.vermeiren.de e-mail: [email protected]

Austria

Vermeiren France S.A.

Z. I., 5, Rue d´Ennevelin

F-59710 Avelin

Tel: +33(0)3 28 55 07 98

Fax: +33(0)3 20 90 28 89 website: www.vermeiren.fr

L. Vermeiren Ges. mbH

Winetzhammerstraße 10

A-4030 Linz

Tel: +43(0)732 37 13 66

Fax: +43(0)732 37 13 69 website: www.vermeiren.at e-mail: [email protected] e-mail: [email protected]

Italy

Vermeiren Italia

Viale delle Industrie 5

I-20020 Arese MI

Tel: +39 02 99 77 07

Fax: +39 02 93 58 56 17 website: www.reatime.it e-mail: [email protected]

Vermeiren Suisse S.A.

Hühnerhubelstraße 59

CH-3123 Belp

Tel: +41(0)31 818 40 95

Fax: +41(0)31 818 40 98 website: www.vermeiren.ch e-mail: [email protected]

Poland

Vermeiren Polska Sp. z o.o ul. Łączna 1

PL-55-100 Trzebnica

Tel: +48(0)71 387 42 00

Fax: +48(0)71 387 05 74 website: www.vermeiren.pl e-mail: [email protected]

Vermeiren Iberica, S.L.

Carratera de Cartellà, Km 0,5

Sant Gregori Parc Industrial Edifici A

17150 Sant Gregori (Girona)

Tel: +34 972 42 84 33

Fax: +34 972 40 50 54 website: www.vermeiren.es e-mail: [email protected]

Czech Republic

Vermeiren ČR S.R.O.

Nadrazni 132

702 00 Ostrava 1

Tel: +420 596 133 923

Fax: +420 596 133 277 website: www.vermeiren.cz e-mail: [email protected]

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Related manuals

Frequently Answers and Questions

What is the recommended interval for lubricating or adjusting moving parts on the Vermeiren Interval XXL bed?
Every 8 weeks, or more frequently depending on the frequency of use.
What are the main points to check during regular maintenance of the Vermeiren Interval XXL bed?
Check for visible damage, ensure brakes and castors are functioning properly, tighten screws and locking pins, inspect for cleanliness and cracks, and check the operation of mechanically adjustable elements.
How often should a general review and inspection of the Vermeiren Interval XXL bed be performed?
Every 6 months or for every new user.
What are the recommended storage conditions for the Vermeiren Interval XXL bed?
Store in a dry place between +5 °C and +40 °C, with a relative humidity between 30% and 70%, and provide sufficient covering to protect from rust and foreign bodies.
How can I clean the metal frames and wooden parts of the Vermeiren Interval XXL bed?
Use a cloth moistened with warm water and mild commercial detergent, avoiding strong cleaning liquids or harsh brushes. Never clean with steam or pressure washers.
Download PDF