advertisement
▼
Scroll to page 2
of 76
LUDLUM MODEL 2241-3 & 2241-3i SURVEY METER November 2014 Serial Number 299171 and Succeeding Serial Numbers LUDLUM MODEL 2241-3 & 2241-3i SURVEY METER November 2014 Serial Number 299171 and Succeeding Serial Numbers Table of Contents Introduction Information on Model 2241-3i Getting Started 1 1-2 2 Unpacking and Repacking 2-1 Battery Installation 2-1 Connecting a Detector 2-2 Operational Check 2-2 Specifications 3 Removable Switchboard Adjustable Parameters 3-1 Internal Detector Specifications for Model 2241-3i 3-4 Identification of Controls and Functions 4 Display 4-1 Display Status Definitions 4-1 Front Panel Controls 4-2 Front Panel Calibration Controls 4-3 Main Board Controls 4-4 Switch Board Controls 4-5 Safety Considerations 5 Environmental Conditions for Normal Use 5-1 Detector Connector 5-1 Warning Markings and Symbols 5-1 Maintenance 6 Operational Check 6-1 Recalibration 6-1 Batteries 6-3 Technical Principle of Operation 7 Main Board 7-1 Switch Board 7-2 Display Board 7-3 Ludlum Measurements, Inc. November 2014 Model 2241-3 & 3i Technical Manual Instrument Setup 8 Entering or Changing Switch Board Parameters 8-1 Loading Default Parameters 8-1 The Function Switch 8-2 Function Switch Position Descriptions and Variables 8-2 Calibration 8-7 General Detector Setup Information 8-8 Counts per minute (C/m) Calibration 8-9 R/hr Calibration 8-11 Determining CC and DTC 8-12 Software License Agreement 8-19 Recycling 9 Parts List 10 Model 2241-3 Survey Meter 10-1 Main Circuit Board, Drawing 408 × 226 10-1 Calibration Board, Drawing 408 × 98 10-3 Display Board, Drawing 408 × 259 10-4 Switch Board, Drawing 408 × 45 10-5 Chassis Wiring Diagram, Drawing 408 × 101 10-5 Drawings and Diagrams 11 RS-232 Output Formats - Appendix A A Binary Output Format A-1 ASCII Output Format A-4 Ludlum Measurements, Inc. November 2014 Model 2241-3 & 3i Technical Manual 1 Section 1 Section Introduction T he Model 2241-3 is a portable microprocessor-based digital Scaler/ Ratemeter designed for use with scintillation, Geiger-Mueller (GM), and proportional-type detectors for measurement of ionizing radiation. Data is presented on a four-digit (six digits in the scaler mode) Liquid Crystal Display (LCD) with moving decimal point. A toggle switch labeled SCA/RATE selects the desired operating mode for the instrument. A five-position switch labeled ˝OFF/DET 1/DET 2/DET 3/DET 4˝ allows for a quick change of operating parameters when changing out detectors. Programmable display units (RATEMETER mode only) can be represented in R/hr, Sv/h, cpm, or cps with multipliers of micro (µ) or milli (m) for R/hr and Sv/h and kilo (k) for cpm or cps. The display units are auto-ranging, enabling the readout to display a broad range of radiation levels. The display also offers lower limit capability. For example, the display can be set to show only values that are greater than or equal to1 µR/hr. This instrument incorporates independent adjustable alarms for RATEMETER and SCALER operating modes. The RATEMETER mode has two alarm indicators. The first-level alarm is indicated by display of the word ˝ALERT˝ on the LCD. The second-level alarm is indicated by display of the word ˝ALARM˝ and by the emitting of a continuous audible tone. The SCALER alarm condition will also display the word ˝ALARM˝ and produce the same audible tone. Both audible alarms may be silenced (acknowledged) by depressing the RESET switch. All alarms are concurrent. Other features include: 1) Dead Time Correction (DTC) to compensate for detector dead time; 2) audible click-per-event with programmable 1, 10, 100, and 1000 divide-by; 3) LCD backlight with programmable ON time; 4) programmable fixed or variable response time; and 5) count overflow visual alarm, indicative of the counting circuitry nearing the maximum counting capability. Ludlum Measurements, Inc. Page 1-1 November 2014 Model 2241-3 & 3i Technical Manual Section 1 All of the features described above may be programmed manually using the internal switch board or by computer through the RS-232 port. Four different detector operating parameters may be stored in non-volatile memory. The switch board can be removed after entering or changing parameters to prevent tampering with setup parameters. A regulated high-voltage power supply, four independent set-point controls adjustable from 400 to 2500 volts, and adjustable discrimination levels, add versatility to the instrument. This supports operation for a broad range of detectors and connecting cable lengths. All of the calibration controls are covered to prevent any inadvertent adjustment to the detector operating parameters. The instrument is powered by two standard ˝D˝ cell batteries. The unit body is made of cast-and-drawn aluminum with beige powder coating, which aids in the decontamination of surfaces. The Model 2241-3i is identical to the Model 2241-3 except for the following: The five-position selector switch located on the front panel has been modified to provide for selection of an internal detector. When selected, the INT position (formerly DET4) switches detector operation to an internal, energy-compensated Geiger-Mueller detector. Appropriate wiring changes to accommodate the internal detector have been made; however, physical dimensions are the same as for a Model 2241-3. A drawing of the front-panel of the 2241-3i is included at the beginning of this manual. A wiring diagram for the instrument is included in the back of the manual with the other drawings. The internal detector specifications can be found at the end of Section 3 of this manual. Ludlum Measurements, Inc. Page 1-2 November 2014 Model 2241-3 & 3i Technical Manual 2 Section 2 Section Getting Started Unpacking and Repacking Remove the calibration certificates and place them in a secure location. Remove the instrument, detectors, and accessories (batteries, cable, etc.), and ensure that all of the items listed on the packing list are in the carton. Check individual item serial numbers and ensure calibration certificates match. The Model 2241-3 serial number is located on the front panel below the battery compartment. Most Ludlum Measurements, Inc. detectors have a label on the base or body of the detector for model and serial number identification. Important! If multiple shipments are received, ensure that the detectors and instruments are not interchanged. Each instrument is calibrated to specific detectors, and are therefore, not interchangeable. To return an instrument for repair or calibration, provide sufficient packing material to prevent damage during shipment. Also provide appropriate warning labels to ensure careful handling. Include detector(s) and related cable(s) for calibration. Every returned instrument must be accompanied by an Instrument Return Form, which can be downloaded from the Ludlum website at www.ludlums.com. Find the form by clicking the “Support” tab and selecting “Repair and Calibration” from the drop-down menu. Then choose the appropriate Repair and Calibration division where you will find a link to the form. Battery Installation Ensure the OFF/DET 1/DET 2/DET 3/DET 4 switch is in the OFF position. Open the battery lid by turning the quarter-turn thumb screw counter- Ludlum Measurements, Inc. Page 2-1 November 2014 Model 2241-3 & 3i Technical Manual Section 2 clockwise. Install two ˝D˝ size batteries in the compartment. Note the (+) and (–) marks inside the battery door. Match the battery polarity to these marks. Close the battery box lid. Note: The center post of “D” size battery is positive. Connecting a Detector Caution! The detector operating voltage (HV) is supplied to the detector by way of the input connector. A mild electric shock may occur if contact is made with the center pin of the input connector. Switch the Model 2241-3 to the OFF position before connecting or disconnecting the cable or detector. Connect a detector to the Model 2241-3 by using the cable provided; firmly pushing the connectors together while twisting clockwise until the connector latches (1/4 turn). The diagram to the left illustrates how this is done. Next, place the detector selector switch in the appropriate position. Operational Check Place the SCA/RATE (scaler/ratemeter) switch in the RATE position. Notice that the display goes through an initialization sequence. The display will show all “8”s with decimal points. Check to make sure all segments display, as illustrated in the diagram to the left. The LCD then displays the firmware number in the format “P-XX YY.” The “XX” is the firmware number, and the “YY” is the firmware version. (The figure to the left is for example only; to illustrate location of display.) The minimum displayable value (for example 00.0 µR/hr) should be shown. When switched to the SCA position, a single “0” will be displayed. Ludlum Measurements, Inc. Page 2-2 November 2014 Model 2241-3 & 3i Technical Manual Section 2 The display will auto-range to the current level (see figure at left). When auto-ranging down, the Model 2241-3 uses multiples of 5. This technique keeps the decimal point from jumping between numbers when viewing values around multiples of 10. Check for a proper background reading: If using a Ludlum Model 44-9 detector, a typical reading would be 25-50 cpm or 8-15 µR/hr. If using a Ludlum Model 44-2 detector, a typical reading would be 1.4-2.6 kcpm or 8-15 µR/hr. If using a Ludlum Model 133-7 detector, a typical reading would be 0 cpm or 8-15 µR/hr. A reference reading (or readings) with a check source should be obtained with the detector(s) in a constant and reproducible manner at the time of calibration or at the time the instrument is received in the field. If at any time the instrument fails to read within 20% of the reference reading when using the same check source, it should be sent to a calibration facility for recalibration and/or repair. If desired, multiple readings may be taken at different distances and/or with different sources so that other ranges or scales are checked. Switch the AUD ON/OFF switch to the ON position and confirm that the external unimorph speaker produces an audible click for each event detected (audio divide-by 1 parameter). The AUD ON/OFF switch will silence the clicks if in the OFF position, however, an audible alarm condition will still be heard. Increase the source activity or lower the alert and alarm points to initiate an and ALARM condition. (Refer to Section 8, Subsection “Entering or Changing Switchboard Parameters.”) Depress the RESET switch to acknowledge the audible alarm. Decrease the radiation activity below the ALERT and ALARM threshold and depress the RESET switch to clear the alarm conditions. If an alarm condition is not present, depressing the RESET switch the first time will reset the alert condition and zero the ratemeter. ALERT Position a check source to produce a ratemeter reading of 100 to 2000 counts per minute or 10-100 µR/hr. While observing the ratemeter fluctuations, select between the fast and slow response time (F/S) positions to observe variations in the display. The S position should respond Ludlum Measurements, Inc. Page 2-3 November 2014 Model 2241-3 & 3i Technical Manual Section 2 approximately five times slower than the F position (for fixed response mode) and three times slower when in variable response mode. The slow response position is normally used when the Model 2241-3 is displaying low numbers, which require a more stable display. The fast response position is used at high count levels. Place the SCA/RATE switch to the SCA position. Depress the COUNT switch located in the end of the carrying handle in order to initiate a count cycle. The word “COUNTING” should be flashing on the LCD during the count cycle and should disappear at the end of the predetermined count time. If a scaler ALARM condition occurs, the RESET switch can be depressed to acknowledge the alarm; however, the COUNT switch must be depressed to clear the visual ALARM and to restart the count cycle. Depress and release the LIGHT switch. The backlight located behind the LCD should illuminate (for pre-programmed ON time). Select the desired F/S, AUD ON/OFF, and RATE/SCA parameters and proceed to use the instrument. Ludlum Measurements, Inc. Page 2-4 November 2014 Model 2241-3 & 3i Technical Manual 3 Section 3 Section Specifications 0 Linearity: Readings are within 10% of true value with a detector connected. Display: a four-digit Liquid Crystal Display (LCD) with digits 1.3 cm (0.5 in.) in height. Two additional 0.5 cm (0.2 in.) high digits are used for the overflow counter (SCALER mode) and exponential powers (parameter setup). Enunciators are provided for display units, ALERT, ALARM, low battery, detector OVERLOAD (not used), counting OVERFLOW, and scaler COUNTING. Warm-up Time: Unit may be used immediately after the LCD initialization sequence is completed (approximately five seconds after power-up). RATEMETER: Depending upon how the instrument was calibrated, the can display in either R/hr, Sv/h, cpm, or cps when the control switch is in the RATEMETER position. RATEMETER SCALER: activated by pushbutton in handle when the three-position switch is in the SCALER position. Count time is adjustable. Calibration Controls: accessible from the front of the instrument (protective cover provided). These controls are preset at the factory or calibration lab and should not be adjusted by field personnel. Discriminator / Input Sensitivity: adjustable from 2 to 100 mV; negative pulse response High Voltage: four independent controls, each adjustable from 400-2500 volts; four separate set points for each detector setupHV 1, HV 2, HV 3, and HV 4, regulated within 0.2% at 1000 Vdc; maximum load of 50 µA. Ludlum Measurements, Inc. Page 3-1 November 2014 Model 2241-3 & 3i Technical Manual Section 3 RESET: a pushbutton for zeroing the display, acknowledging and/or resetting the alarm. Note: The RESET button only silences the alarm in the current mode that the instrument is in. For example, the RESET button will not affect the scaler alarm if the instrument is in the ratemeter mode. Light: display backlight activated by pushbutton Audio: built-in audio speaker (unimorph) with AUD ON/OFF switch; at maximum volume > 60 dB at 2 feet; internal adjustable volume Alert/Alarm: indicated by either an ALERT or ALARM enunciator on the display (RATEMETER mode only) and by an audible tone Power: two each, ˝D˝ cell batteries housed in an externally accessible sealed compartment. Current draw is approximately 35 mA with the backlight OFF. Minimum battery voltage is 2.2 ±0.1 Vdc. Battery Dependence: Meter readings vary by less than 3% from fully charged batteries until the battery symbol appears, indicating the need for recharge or replacement. Battery Life: typically 200 hours with alkaline batteries (display indicates low-battery condition). Instrument will operate for approximately 24 hours after the battery symbol first appears. Size: 16.5 x 8.9 x 21.6 cm (6.5 x 3.5 x 8.5 in.) (H x W x D) Weight: 1.6 kg (3.5 lb), including batteries. Removable Switchboard Adjustable Parameters: Backlight ON Time: 5, 15, 30, 60, 90, 120, 180, or 240 seconds for the backlight to stay on when activated by the pushbutton; factory set at 5. Set Minimum Display: allows lower limit of the auto-ranging display to be fixed. For example, the display can be set to only show values above or equal to 1 µR/hr. Ludlum Measurements, Inc. Page 3-2 November 2014 Model 2241-3 & 3i Technical Manual Section 3 RS-232 Data Dump Mode: enables or disables dump mode to the RS-232 port (˝D˝ type connector). When enabled, the data will be dumped every two seconds. RS-232 Detector Setup Mode: allows for input of detector parameters via the RS-232 port Baud Rate: selects either 150, 300, 600, 1200, 2400, 4800, 9600, or 19200 bps Detector Dead Time Compensation (DTC): adjustable from 0 to 9999 microseconds Calibration Constant: adjustable from 0.001 to 280 × 109 counts/display unit Display Units: can display in R/hr, Sv/h, cpm, or cps Time Base: can display in seconds or minutes Audio Divide: 1, 10, 100, or 1000 events per click Response Time: variable or fixed ratemeter response (All stated times correspond to a range of 10% to 90% of the final reading). Factory default is variable so that the instrument will automatically adjust the response time to the best setting for the current count rate. Variable Response: dependent on the number of counts present, typically 4 to 25 seconds for FAST, and 4 to 60 seconds for SLOW Fixed Response: The parameter is adjustable from 1-25 seconds resulting in a FAST response from approximately 2-50 seconds. The SLOW response is approximately 10-250 seconds. For MDA-type measurements, the fixed response mode is recommended. Ratemeter Alert/Alarm: set at any point corresponding to the pre-selected ratemeter range Scaler Alarm: adjustable from 1 to 999999 counts Scaler Count Time: adjustable from 1 to 9999 seconds Ludlum Measurements, Inc. Page 3-3 November 2014 Model 2241-3 & 3i Technical Manual Section 3 Internal Detector Specifications for Model 2241-3i: Operating Voltage: recommended 550 volts Input Sensitivity: -30 mV ±10 mV Dead Time: typically 80 microseconds Energy Reponse: within ±15% 2 Tube: 30 mg/cm stainless steel (Halogen quench) GM Size: 2.3 x 10.2 (0.9 x 4 in.) (Dia x L) Temperature Range: -20 to 50 °C (-4 to 122 °F) Ludlum Measurements, Inc. Page 3-4 November 2014 Model 2241-3 & 3i Technical Manual 4 Section 4 Section Identification of Controls and Functions Display The Model 2241-3 utilizes a four-digit liquid crystal display (LCD) with a two-digit overflow (SCALER mode) and moving decimal point. The two smaller digits located in the lower right corner of the display indicate counter OVERFLOW when in the scaler counting mode (equivalent to a six-digit scaler) or exponential power when in the parameter setup mode. The upper right corner of the LCD displays units and multiplier(s) - R/hr, mR/hr, or µR/hr; Sv/h, mSv/h or µSv/h; C/m, kC/m, C/s or kC/s. The bottom part of the readout displays the ALARM, ALERT, OFLOW, OVERLOAD (not used) enunciators and the low-battery icon. COUNTING indicates that the scaler mode has been initiated and is in the counting process. Display Status Definitions ALARM: Ratemeter or scaler count has increased above the preset alarm threshold. An audible continuous tone will accompany the latching ALARM condition. Depressing RESET will acknowledge the audible ratemeter and/or scaler alarm. Depressing RESET a second time will reset the ratemeter reading and ratemeter alarm. To reset the scaler ALARM, depress the COUNT switch located in the carrying handle to re-initiate the scaler count cycle. ALERT: Ratemeter count has increased above the preset alert threshold. To reset an ALERT condition, press RESET once if in the non-alarm condition and twice if in an alarm condition. (The first depression in the alarm condition acknowledges the audible alarm.) The ratemeter will reset to the minimum displayable reading each time the alert is reset. OFLOW (Overflow) [RATEMETER mode]: Indicates that the incoming count exceeds the capability to display stable or reliable readings corresponding to the radiation level being measured. The overflow symbol will appear when the ratemeter exceeds 100 kcps or if the dead time correction is greater than 75%. OFLOW will appear in the SCALER mode when the six-digit display (four digits display and two overflow digits in right corner) reaches “999999” and starts to roll over again. Ludlum Measurements, Inc. Page 4-1 November 2014 Model 2241-3 & 3i Technical Manual Section 4 Low Battery icon: indicates that the batteries have decreased to the minimum operating voltage of 2.2 ±0.1 Vdc. Instrument will continue to operate for approximately 24 hours thereafter. COUNTING: indicates that the scaler COUNT switch has been depressed and that the scaler is accumulating counts for the pre-determined count time Front Panel Controls OFF/DET 1/DET 2/DET 3/DET 4 Switch: a five-position rotary switch that applies power to the instrument and selects the desired set of detector parameters SCA/RATE (Mode) Switch: a protected (locking) toggle switch that toggles between ratemeter or scaler counting modes. The user pulls up and then over to change modes. AUD ON/OFF Switch: The click-per-event audio may be silenced or enabled via this front-panel toggle switch. The audible alarm is independent of the AUD ON/OFF switch and will override the audible click-per-event. An audible alarm can only be silenced by depressing the RESET button. F/S (Fast/Slow) Response Switch: a two-position toggle switch that selects fast or slow counting response time Variable Response: The F position allows the time constant (TC) to vary from 1 to 10 seconds, while the S position varies from 1 to 30 seconds. The response time is automatically adjusted in proportion to the incoming count rate between the F/S TC variables. Fixed Response: The F position corresponds to the selected fixed response time - TC. The S position is five times slower than the selected fast TC. LIGHT (LCD Backlight): A pushbutton switch, when depressed, illuminates the LCD for a pre-programmed time. The backlight ON time can be selected between 5 and 240 seconds during the parameter setup. Ludlum Measurements, Inc. Page 4-2 November 2014 Model 2241-3 & 3i Technical Manual Section 4 RESET Pushbutton: In the non-alarm condition, depressing the RESET switch resets the ratemeter display to the minimum display readout. In an alarm condition (ratemeter or scaler), depressing RESET will silence the audible alarm. Depressing RESET a second time will reset the ratemeter alarm and/or alert condition. The scaler alarm can only be reset by depressing the scaler count switch located in the end of the Model 2241-3 handle. Note: The RESET button only silences the alarm in the current mode that the instrument is in. For example, the RESET button will not affect the scaler alarm if the instrument is in the ratemeter mode. Scaler Count Switch: pushbutton switch located in the end of the Model 2241-3 carrying handle, which when depressed, initializes the start of the scaler count accumulation for the preset scaling time. The SCA/RATE switch must be in the SCA position to initiate the counting cycle. The scaler display uses the two digits in the lower right-hand corner for the two most significant digits of the six-digit readout. Scaling time can be set from 1 to 9999 seconds in the parameter setup by way of the switch board. Depressing the Count switch after a scaler alarm will reset the scaler display to 0, resetting the alarm condition. Front Panel Calibration Controls Note: Remove the front-panel calibration cover to expose the following calibration controls: (Discriminator): a multi-turn potentiometer (approximately 20 revolutions) used to vary the detector pulse-counting threshold from 2 to 100 millivolts. A Ludlum Model 500 Pulser or equivalent should be used in checking or adjusting the pulse discrimination parameter. DISC Ludlum Measurements, Inc. Page 4-3 November 2014 Model 2241-3 & 3i Technical Manual Section 4 Note: When making adjustments to the HV potentiometers, make note of the following precautions: Use a Ludlum Model 500 Pulser or high-impedance voltmeter with a high-voltage probe to measure the high voltage at the detector connector. If a Ludlum Model 500 Pulser is not available, ensure that the impedance of voltmeter used is 1000 megohms or greater. HV1, HV2, HV3, HV4: multi-turn potentiometers (approximately 20 revolutions) that vary the detector voltage from 400 to 2500 volts. The maximum high-voltage output is adjusted by the HV LIMIT potentiometer located on the internal main board. Main Board Controls Note: To access the internal circuit boards, unlatch the latches at each end of the Model 2241-3. Carefully separate the top chassis from the bottom cover (referred to as a can). The can has the audio speaker (unimorph) with a two-conductor cable attached to the main board. The audio plug may be disconnected during the internal control adjustments. HV LIMIT (R027): A multi-turn potentiometer (approximately 20 revolutions) sets the maximum HV limit with the front-panel HV control adjusted to the maximum clockwise position. It is adjustable from 1250 to 2500 Vdc. VOLUME (R002): A multi-turn potentiometer (approximately 20 revolutions) varies the audible click-per-event and alarm audio. Adjust the control to the maximum clockwise position for maximum volume. If the VOLUME control is adjusted to the maximum counterclockwise position, the clicks-per-event, or the audible alarm(s), will not be audible when active. Ludlum Measurements, Inc. Page 4-4 November 2014 Model 2241-3 & 3i Technical Manual Section 4 Switch Board Controls The switch board utilizes a 16-position rotary switch (FUNCTION) to select the 16 setup parameters. (Refer to schematics and component layout drawing near the end of the manual.) All of the setup parameters are stored in the non-volatile EEPROM, which will retain data even after the Model 2241-3 batteries are removed. After the parameters are entered, the switch board can be removed and the Model 2241-3 will continue to operate from the previously programmed information. Changing parameters and information on switchboard controls are covered in detail in Section 8 of this manual. Ludlum Measurements, Inc. Page 4-5 November 2014 Model 2241-3 & 3i Technical Manual 5 Section 5 Section Safety Considerations Environmental Conditions for Normal Use Indoor or outdoor use No maximum altitude Temperature range of –20 to 50 °C (-4 to 122 °F) Maximum relative humidity of less then 95% (non-condensing) Pollution Degree 3 (as defined by IEC 664). (Occurs when conductive pollution or dry nonconductive pollution becomes conductive due to condensation. This is typical of industrial or construction sites.) Detector Connector Caution! The detector operating voltage (HV) is supplied to the detector by way of the input connector. A mild electric shock may occur if contact is made with the center pin of the input connector. Switch the Model 2241-3 to the OFF position before connecting or disconnecting the cable or detector. Warning Markings and Symbols Caution! The operator or responsible body is cautioned that the protection provided by the equipment may be impaired if the equipment is used in a manner not specified by Ludlum Measurements, Inc. Ludlum Measurements, Inc. Page 5-1 November 2014 Model 2241-3 & 3i Technical Manual Section 5 The Model 2241-3 Survey Meter is marked with the following symbols: CAUTION, RISK OF ELECTRIC SHOCK (per ISO 3864, No. B.3.6): designates a terminal (connector) that allows connection to a voltage exceeding 1 kV. Contact with the subject connector while the instrument is on or shortly after turning off may result in electric shock. This symbol appears on the front panel. CAUTION (per ISO 3864, No. B.3.1): designates hazardous live voltage and risk of electric shock. During normal use, internal components are hazardous live. This instrument must be isolated or disconnected from the hazardous live voltage before accessing the internal components. This symbol appears on the front panel. Warning! The operator is strongly cautioned to take the following precautions to avoid contact with internal hazardous live parts that are accessible using a tool: 1. Turn the instrument power OFF and remove the batteries. 2. Allow the instrument to sit for one minute before accessing any internal components. The “crossed-out wheelie bin” symbol notifies the consumer that the product is not to be mixed with unsorted municipal waste when discarding. Each material must be separated. The symbol is placed on the battery compartment. See Section 9, “Recycling,” for further information. The “CE” mark is used to identify this instrument as being acceptable for use within the European Union. Ludlum Measurements, Inc. Page 5-2 November 2014 Model 2241-3 & 3i Technical Manual 6 Section 6 Section Maintenance Instrument maintenance consists of keeping the instrument clean and periodically checking the batteries and the calibration. The Model 2241-3 may be externally cleaned with a damp cloth (using only water as the wetting agent). Do not immerse the instrument in any liquid. Observe the following precautions when cleaning: 1. Turn the instrument OFF and remove the batteries. 2. Allow the instrument to sit for one minute before performing any external cleaning or accessing internal components for maintenance. Operational Check An instrument operational check should be performed prior to each use by exposing the detector to the supplied check source, confirming the proper reading on each scale. See Section 2, Subsection “Operational Check,” for further details. Recalibration Recalibration should be accomplished after any maintenance or adjustment of any kind has been performed on the instrument. Battery and cable replacements (of same length) and exterior cleanings do not normally require the instrument be recalibrated. Note: Ludlum Measurements, Inc. recommends recalibration at intervals no greater than one year. Check the appropriate local state and federal regulations to determine required recalibration intervals. Ludlum Measurements, Inc. Page 6-1 November 2014 Model 2241-3 & 3i Technical Manual Section 6 Ludlum Measurements offers a full-service repair and calibration department. We not only repair and calibrate our own instruments, but most other manufacturers’ instruments as well. See Section 8 “Instrument Setup & Calibration,” for further details on instrument calibration. Batteries The batteries should be removed and the battery contacts cleaned of any corrosion at least every three months. If the instrument has been exposed to a very dusty or corrosive atmosphere, more frequent battery servicing should be used. Use a spanner wrench to unscrew the battery contact insulators, exposing the internal contacts and battery springs. Removing the handle will facilitate access to these contacts. Note: Never store the instrument over 30 days without removing the batteries. Although this instrument will operate at very high ambient temperatures, battery seal failure can occur at temperatures as low as 38 °C (100°F). Ludlum Measurements, Inc. Page 6-2 November 2014 Model 2241-3 & 3i Technical Manual 7 Section 7 Section Refer to the Main Board schematic for the following: Technical Principle of Operation Detector Input/Amplifier Negative-going detector pulses are coupled from the detector through C021 to amplifier U021. R024 and CR021 protect the input of U021 from inadvertent shorts. Self-biased amplifier U021 provides gain in proportion to R022, divided by R025. Transistor pins 4, 5, and 6 of U021, provide amplification. Pins 10-15 of U021 are coupled as a constant current source to pin 6 of U021. The output is self-biased to 2 Vbe (approximately 1.4 volts) at pin 7 of U021. This provides just enough bias current through pin 6 of U021 to conduct all of the current from the constant current source. Positive pulses from pin 7 of U021 are coupled to the discriminator (U011) through R031 and C012. Discriminator Positive pulses from amplifier U021 are coupled to pin 2 of U011 comparator. The discrimination level is set by the DISC control connected to pin 3 of U011. As the positive pulses at pin 2 of U011 increase above DISC reference at pin 3, pin 1 goes low, producing a low pulse. Pin 1 of U011 is normally held high (+5 volts) by R014. The low pulse from pin 1 of U021 is coupled to univibrator U001. U001 shapes and fixes the pulse width to approximately 10 µs. The univibrator is configured in the non-retriggerable mode. Negative pulses from pin 9 of U001 are coupled to the µP for counting. Low Voltage Supply Battery voltage is coupled to DC-DC converter U231. U231 and related components provide +5 V to power the µP, op-amps, and logic circuitry. R135 and R136 provide voltage division for low-battery detection. Pin 6 of U231 provides a low signal when the battery voltage decreases to +2.2 ±0.1 Vdc. U121 provides the +2.5 Vdc reference for the HV and DISC control references. Ludlum Measurements, Inc. Page 7-1 November 2014 Model 2241-3 & 3i Technical Manual Section 7 High Voltage Supply High voltage is developed by blocking oscillator Q241, T141, and C244 and rectified by voltage multiplier CR041-CR043, C041-C043, and C141. High voltage increases as current through R241 increases, with maximum output voltage with Q241 saturated. High voltage is coupled back through R034 to op-amp pin 2 of U131. Resistor network R027, R132 completes the HV division circuit to ground. R027 provides HV limit from 1250-2400 when the HV control on the calibration board is at maximum. The regulated HV output is controlled by the HV1 and HV2 potentiometers located under the CAL cover on the front panel. This control provides the reference for comparator pin 3, U131. During stable operation, the voltage at pin 2 of U131 will equal the voltage at pin 3 of U131. Pin 1 of U131 will cause conduction of Q141 to increase or decrease until the HV finds a level of stability. Microprocessor (µP) U111 controls all of the data, control inputs, and display information. The clock frequency is crystal-controlled by Y221 and related components at 6.144 MHz. The µP incorporates internal memory (ROM), storing the program information. U1 resets the µP at power-up to initiate the start of the program routine. During the program loop, the µP looks at all the input switches for initiation or status changes and responds accordingly. U122 is a 256 x 8 bit EEPROM used to store the setup parameters. The information is transferred serially from the µP. The EEPROM is non-volatile, meaning it retains memory even after power is removed. Audio Click-per-event, divide-by, and alarm audio pulse frequency is generated by the µP and coupled to Q101. Q101 then inverts the pulses and drives the bottom of T101. Bias voltage is provided by the volume control (R002) to the top of T101. Refer to the Switch Board schematic for the following: Ludlum Measurements, Inc. S1 (FUNCTION) S1 is a 16-position binary rotary switch, which selects the programmable parameters for the Model 2241-3. The switch selects the parameters using the hexadecimal numbering system via buss lines SW1-SW4. Page 7-2 November 2014 Model 2241-3 & 3i Technical Manual Section 7 S2-S4 S2-S4 are pushbutton switches which enter/change the variables for each of the 16 parameters. U1 U1 is a +5 V powered RS-232 driver/receiver used to interface the Model 2241-3 to a computer. Refer to Display Board schematic for the following: LCD Drive U111 and U211 are serial input 32-bit LCD drivers. The data is loaded serially into the 32-bit shift registers (internal) via the “D” IN input. The LOAD input instructs the shift register to receive data while the CLOCK input shifts the data through the 32-bit registers. After all the data is loaded, the LOAD line is pulsed by the µP, instructing the registers to transfer the data to the LCD drivers. The backplane (BP) signal from U211 provides the reference signal (approximately 125 Hz at 5 Vdc) to the LCD (DSP1) BP connection. When a segment is illuminated, the signal to that segment will be out-of-phase with the BP signal. If the segment is OFF, the signal will be in-phase with the BP signal. Backlight Drive Depressing the LIGHT button instructs the µP to set the BACKLIGHT line, pin 31 on µP, "low" for the predetermined backlight ON time. (Refer to main board schematic for details.) A "low" condition on pin 31 causes Q212 to conduct sending +3 V to P8-3 on the display board with +3 V at P8-3 (refer to display board schematic). Backlight oscillator Q011, T011, and related components start to oscillate, producing a 2.5 kHz, sine wave signal. The signal is amplified by T011 to 150 volts peak-to-peak to drive the LCD backlight. Ludlum Measurements, Inc. Page 7-3 November 2014 Model 2241-3 & 3i Technical Manual 8 Section 8 Section Instrument Setup & Calibration Entering or Changing Switch Board Parameters On the switch board, select the desired parameter to enter or change by using the corresponding FUNCTION switch position. Depress the ENTER button, and a character on the LCD will start to flash. The flashing character indicates that the program is in the parameter change mode. To change the character, press the UP button until the desired variable is reached. To shift to another character, increment the LEFT pushbutton until the desired character is reached. The LEFT pushbutton switch enables the operator to sequence through all the characters on the LCD associated with a particular parameter. Once the desired data is entered, depress the ENTER button. The LCD characters should stop flashing and the new parameter data should display. Note: The DETECTOR SELECT switch allows the Model 2241-3 to have four sets of operating parameters. Loading Default Parameters To load the default parameters for all detector setups, hold down the UP pushbutton on the switch board until DEF is displayed on the LCD. The table on the following page shows the default values. Ludlum Measurements, Inc. Page 8-1 November 2014 Model 2241-3 & 3i Technical Manual Section 8 Model 2241-3 Dead Time Cal Const Rate Alarm Scaler Alarm Count Time Time Base Units Audio Divide-By Response Check Source Percent CS Rate Alert Cpm Setup 01 0 µs 100e-2 50.0 kcpm 85000 12 s Minutes cpm 1 0 0 0 20.0 kcpm Minimum Display 0.00 cpm Baud Rate LCD Time Off Detector uR/hr Setup 02 0 µs 105e+8 50 µR/hr 85000 12 s Seconds R/hr 1 0 0 0 20 µR/hr 00.0 µR/hr R/Hr Setup 03 0 µs 108e+4 5 R/hr 85000 12 s Seconds R/hr 1 0 0 0 2 R/hr 00.0 µR/hr mR/hr Setup 04 0 µs 198e+6 5 mR/hr 85000 12 s Seconds R/hr 1 0 0 0 2 mR/hr 00.0 µR/hr 9600 5s 0 The Function Switch FUNCTION Switch: a 16-position rotary switch labeled “0-9” and “A-F.” This switch selects a parameter setup mode for the Model 2241-3. If the board is not installed, the normal operation mode (counting mode) is selected. If the switch board is installed, the selector switch must be set to the 0 position for normal instrument operation. The following may be changed using the switch board, and are discussed in detail in this section: Detector parameters Durrent detector setup in use RS-232 communication baud rate RS-232 data dump mode RS-232 detector parameters set/read mode Function Switch Position Descriptions and Variables POSITION 0: NORMAL OPERATION places the Model 2241-3 in the normal (counting) operating mode. Unplugging the switch board from the Model 2241-3 main board defaults to the normal operating mode. Ludlum Measurements, Inc. Page 8-2 November 2014 Model 2241-3 & 3i Technical Manual Section 8 POSITION 1: DEAD TIME (µs) allows changing the detector dead time correction for the current detector setup. Setting this parameter to “0” disables dead time correction. The dead time adjusts from 0 to 9999 microseconds (µs). The incoming counts are adjusted for dead time using the following formula: n= m 1 - m Where, n = corrected counts per second m = incoming count per second = system dead time CC = cps x time base rate POSITION 2: calibration constant allows changing the calibration constant for the current detector setup. The calibration constant (CC) adjusts from 0.001 to 280 × 109. The calibration constant converts counts/time base to units/time base. The CC must be set to 1 to readout in cps (counts per second) or cpm (counts per minute). CC CONVERSION TABLE Conversion Rate Multiply by to get CC 3.6 x 109 3.6 x 106 3.6 x 103 6.0 x 107 6.0 x 104 6.0 x 101 3.6 x 107 3.6 x 104 3.6 x 101 6.0 x 105 6.0 x 102 0.6 cps/µR/hr cps/mR/hr cps/R/hr cpm/µR/hr cpm/mR/hr cpm/R/hr cps/µSv/h cps/mSv/h cps/Sv/h cpm/µSv/h cpm/mSv/h cpm/Sv/h Example: The Model 44-9 GM detector produces approximately 3300 cpm/mR/hr for 137Cs: 6.0 x 104 x 3300 = 198 x 106 for CC. Ludlum Measurements, Inc. Page 8-3 November 2014 Model 2241-3 & 3i Technical Manual Section 8 POSITION 3: DISPLAY UNITS selects the display units for the associated detector setup number. The Model 2241-3 and detector may be calibrated in either exposure rate (R/hr or Sv/h) by entering the appropriate Calibration Constant (position 2) and Dead Time correction (position 1). The Model 2241-3 will automatically convert to the correct reading when switching between R and Sv. The time base for count “C” is set independently in position 4. The display units may be set to: R/hr (Roentgens per hour) Sv/h (Sieverts per hour) C/time base (Counts per time) The display is auto-ranging with the appropriate multiplier symbol appearing in front of the “R,” “Sv,” or “C,” indicating the range: µR/hr, mR/hr, R/hr µSv/h, mSv/h, Sv/h C/s, kC/s, C/m, kC/m POSITION 4: (TIME BASE) CPS or CPM selects the display time base for the current detector setup. This time base only applies if the units are set to C/ (Counts/time). The time base for R/hr and Sv/h is fixed in “hr.” For “true” reading (pulser calibration) cpm or cps calibrations, set the Calibration Constant (CC, parameter 2) to read “1.” For geometry calibrations, the detector efficiency can be entered for CC. Example: For alpha scintillation detector with 25% 2π efficiency; enter “250 10-3” in the CC parameter setup.) The display time base may be set to: seconds (s) minutes (m) Ludlum Measurements, Inc. Page 8-4 November 2014 Model 2241-3 & 3i Technical Manual Section 8 POSITION 5: AUDIO DIVIDE-BY selects the audible click-per-event division rate for the current detector setup. If the AUD ON/OFF switch is in the OFF position, no audible click-per-event will be heard. This parameter ranges from: 0 = Divide By 1 1 = Divide By 10 2 = Divide By 100 3 = Divide By 1000 POSITION 6: RESPONSE TIME allows changing the time constant (TC) for the current detector setup. If the response is set to 0, the Model 2241-3 automatically calculates (for variable mode) the time constant based on the incoming cps. If a variable of 1-199 is entered for TC, the response time becomes fixed. Variable Response - Response time is varied in proportion to the incoming count rate. The two-position F/S (Fast/Slow) toggle switch selects the maximum time constant (TC) for the variable mode. The fast position varies the TC from 4-25 seconds, and the slow position varies from 4-60 seconds. Response - The Fast (F) response position is programmable from 2-50 seconds, and the slow response is 5 times slower than the fast TC. For MDA-type measurements, the fixed response time mode is recommended. Fixed POSITION 7: RATEMETER ALARM/ALERT allows changing the ratemeter alarm for the current detector setup. The units of this alarm are the same as the units for the ratemeter display. The fifth push of the left button allows the decimal point to be moved. The ratemeter alarm adjusts from 1 to 999 R/hr (or Sv/h) or 1 to 999 kcpm or 1 to 100 kcps. The units of the alarm are determined by the units for the ratemeter. POSITION 8: SCALER ALARM/COUNT TIME sets the scaler alarm variable from 1-999999, corresponding to the accumulated scaler count. After the scaler alarm variable is entered, the scaler count time is prompted. The scaler count time is adjustable from 1-9999 seconds. POSITION 9: NOT USED Ludlum Measurements, Inc. Page 8-5 November 2014 Model 2241-3 & 3i Technical Manual Section 8 POSITION A: NOT USED POSITION B: LCD Backlight ON TIME is the amount of time that the LCD backlight will stay on after pressing the front panel switch labeled LIGHT. This value is stored in EEPROM. Available values are: 5 seconds 30 seconds 60, 90 seconds 180, 240 seconds. POSITION C: SET MINIMUM DISPLAY sets the ratemeter minimum displayable reading. Depressing the RESET button displays the minimum ratemeter units. The readout will auto-range up to the maximum displayable but will display 0 for ratemeter readings below the user-programmed minimum variable. Minimum displayable values are: 00.0 µ, 000 µ, 0.00 m, 00.0 m, 000 m, 0.00, 00.0, 000 R/hr .000 µ, 000 µ, 00.0 m, 000 m, 0.00, 00.0, 000 Sv/h 0.00, 00.0, 000, 0.00 k, 00.0 k, 000 kcpm, or cps POSITION D: RS-232 DATA DUMP MODE allows the RS-232 port to dump ratemeter data every two seconds. The Model 2241-3 is fully functional during RS-232 data dump with the exception of the audio function. The LCD will alternate between display of the ratemeter and the word “dUP” (representing "dump"). POSITION E: RS-232 DETECTOR PARAMETERS SETUP MODE allows the RS232 port to accept/send a string of parameters corresponding to the current detector setup values. POSITION F: BAUD RATE configures the RS-232 port for the following baud: 150, 300, 600, 1200, 2400, 4800, 9600, and 19200. The data is eight data bits, one stop bit with no parity bit. This value is stored in EEPROM. The baud rate can only be programmed through the switch board. Ludlum Measurements, Inc. Page 8-6 November 2014 Model 2241-3 & 3i Technical Manual Section 8 RS-232 PORT CONNECTOR: This nine-pin “D” type connector is designed as a DCE port. A straight wire cable (extension cable) connects the Model 2241-3 to a computer’s nine-pin RS-232 port. RS-232 CONNECTOR PIN OUT: PIN FUNCTION 1 2 3 4 5 6 7 8 9 NC (No Connection) DATA OUT DATA IN NC NC NC HANDSHAKING IN HANDSHAKING OUT NC Note: Ludlum Measurements, Inc. offers a PC compatible software program, which incorporates the read/write commands necessary to communicate between the PC and the Model 2241-3. The program also incorporates an algorithm to calculate the detector Calibration Constant and Dead Time Constant. The software is offered in a DOS version (part number 1370-025) or a WINDOWS version (part number 1370-024). Read the Software License Agreement at the end of this section before installing any LMI software. Calibration The Model 2241-3 calibration routine consists of entering detector parameters into memory by way of the switch board and adjusting the CAL controls (HV and DISC) for the specific detector operating requirements. The first subsection of calibration will give a general overview of detector setup, including the determination of various detector operating voltages (HV) and the adjustment of counter input sensitivity (DISC). The next subsection deals with pulse generator counts-per-minute calibration. The counts-per minute-parameter setup is used in the initial Ludlum Measurements, Inc. Page 8-7 November 2014 Model 2241-3 & 3i Technical Manual Section 8 instrument checkout procedure and the variables are saved under detector setup number “1” when shipped from Ludlum Measurements, Inc. The following subsection deals with exposure rate calibration. The detector Calibration Constant (CC) and Dead Time Correction (DTC) are the two primary parameters used in the exposure rate calibrations (R/hr and Sv/h). These two constants are alternately varied to achieve linearity at the detector non-linear operating regions. An example of the Ludlum Model 44-9 GM detector calibration is given at the end of this section to illustrate the algorithm used in determining the CC and DTC variables. GENERAL DETECTOR SETUP INFORMATION The operating point for the instrument and probes is established by setting the probe voltage and instrument sensitivity (HV and DISC). The proper selection of this point is the key to instrument performance. Efficiency, background sensitivity, and noise are fixed by the physical makeup of the given detector and rarely vary from unit to unit. However, the selection of the operating point makes a marked difference in the apparent contribution of these three sources of count. In setting the operating point, the final result of the adjustment is to establish the system gain so that the desirable signal pulses (including background radiation) are above the discrimination level and the unwanted pulses from noise are below the discrimination level and are therefore not counted. The total system gain can be controlled by adjusting either the instrument sensitivity or the high voltage. HV controls the gain of the detector; and DISC (Discriminator) controls the instrument counting threshold (sensitivity). In the special case of GM detectors, a minimum voltage must be applied to establish the Geiger-Mueller characteristic. Further changes in HV will have little effect on this type of detector. GM Detectors: The output pulse height of the GM detector is not proportional to the energy of the detected radiation. Adjusting DISC will have minimal effect on observed count rate unless the DISC setting is so low that the instrument will double-pulse. Ludlum Measurements, Inc. Page 8-8 November 2014 Model 2241-3 & 3i Technical Manual Section 8 For most GM detectors, set DISC for 30-40 millivolts and adjust HV to the GM detector recommended high voltage. Most GM detectors operate at 900 volts, although some miniature detectors operate at 450-550 volts. If a recommended setting is unavailable, plot count rate versus HV to produce a plateau graph. Adjust the HV for 25-50 volts above the knee or start of the plateau. For mixed detector use, both sensitivity and high voltage may be tailored for other detectors as long as the GM detector is operated within the recommended voltage range. Caution must be observed in lowering the input sensitivity to ensure that the counter does not double or multi-pulse. Alpha Air-Proportional Detectors: For air proportional alpha detectors, set the DISC for 2 millivolt discrimination. Adjust HV until the detector just breaks down (shown by a rapid increase of count rate without a source present). Measure the HV output, then decrease the HV setting to operate 100 volts below breakdown. Proportional Detectors: For proportional detectors, set the DISC control for 2-millivolt discrimination (near maximum clockwise). Expose the detector to a check source and plot count rate versus HV, similar to the one in the figure below. Refine the HV adjustment for optimum source efficiency with a minimum acceptable background count. Scintillators: Set the DISC for 10 millivolts. Plot background and source counts versus HV to produce a plateau graph similar to the one in the figure. Adjust the HV to 25-50 volts above the knee or start of the plateau. This provides the most stable operating point for the detector. COUNTS PER MINUTE (C/M) CALIBRATION This procedure will set up the Model 2241-3 for the Counts/minute (C/m) mode of operation. Refer to Section 8, (Page 8-2 and following) for more information on setup parameter variables. A Ludlum Model 500 Pulser or equivalent is required. If the pulser does not have a high-voltage display, use a high-impedance voltmeter with at least 1000 megohms input resistance to measure the detector high voltage. Ludlum Measurements, Inc. Page 8-9 November 2014 Model 2241-3 & 3i Technical Manual Section 8 Switch the SCA/RATE switch to the RATE position. Select position DET1 on the detector selector switch located on the front panel. Select FUNCTION switch positions 1-6 and adjust for the following parameters: Switch Pos. 1 2 3 4 5 6 Parameter Function 0000 s-6 0100 -2 c/ m 1 000 s Dead Time Calibration Constant Display Units Timebase Audio Divide-by Response Time Position 7 selects the desired ratemeter ALERT and ALARM trip points. If the parameters are undetermined, arbitrarily choose “0050 kC/m” for the alarm and “0045 kC/m” for the alert to confirm operation of the alert/alarm function. Position 8 selects the scaler alarm parameter and the scaler count time. If the values are unknown, set the scaler alarm to “4500ALARM00” and the count time to “0060” (60 second count time). Position 9 is not used, and position A is not used. Switch to position B and enter “15” for a 15-second backlight ON time. Switch to position C and enter “00.0 C/m” for the minimum displayable value. Select position 0 to return to normal operation. Connect the Model 500 Pulser to detector input and adjust HV and DISC to the specific detector operating parameters. Ludlum Measurements, Inc. Page 8-10 November 2014 Model 2241-3 & 3i Technical Manual Section 8 Adjust the pulser amplitude to 1.5 times the Model 2241-3 discrimination level. Adjust the pulser output to 800 cpm and confirm that the Model 2241-3 reads 800 cpm ±10% on the ratemeter setting. Adjust the pulser output to 200 cpm and confirm that the Model 2241-3 reads 200 cpm ±10% on the ratemeter setting. Adjust the pulser output to 800 cpm, take a one minute count and confirm that the digital scaler readout displays 800 cpm ±2%. Adjust the pulser output to 200 cpm, take a one minute count and confirm that the digital scaler readout displays 200 cpm ±2%. Confirm that the 20% and 80% readings for the upper decades are within the pulser input by decading the pulser count output. Confirm that the scaler readout is within 2% of the pulser input rate. Ensure that the ALERT and ALARMs function by inputting the preset alarm levels as to initiate the alert and alarm conditions. R/hr CALIBRATION The following calibration procedure assumes that detector Calibration Constant (CC) and Dead Time Constant (DTC) are already known. If these constants must be determined, reference the following subsection, “Determining CC and DTC.” Switch the toggle switch to DET2. Detector setup number “1” is usually reserved for the Counts/minute parameter calibration. Rotate the FUNCTION switch counterclockwise to position 1 and enter the detector Dead Time in µs. Rotate to position 2 and enter the Calibration Constant. Enter the desired parameters for positions 3-F. Switch to position 0 for normal operation. Expose the detector to calibrated radiation fields extending from the lower to the upper operating range of the detector. Confirm that the linearity is within 10% of each respective reading. If the readings are off on the lower detector operating region, vary CC. If the readings are off at the upper end of the detector operating region, adjust DTC. Ludlum Measurements, Inc. Page 8-11 November 2014 Model 2241-3 & 3i Technical Manual Section 8 DETERMINING CC AND DTC This procedure contains the algorithm (hi-lo method) for determining the CC (Calibration Constant) and the DTC (Dead Time Correction). An example of the Ludlum Model 44-9 GM detector calibration is used in conjunction with the algorithm calculations to aid in solving the equations. Note: Ludlum Measurements, Inc. offers a PC-compatible software program, which incorporates the read/write commands necessary to communicate between a PC and the Model 2241-3. The program also incorporates the algorithm to calculate the detector CC and DTC. The software is offered in a DOS version (part number 1370-025) or a WINDOWS version (part number 1370-024). Read the Software License Agreement at the close of this section prior to installing any LMI software. Hi-Lo Method: The hi-lo method refers to the placement of the detector in a radiation field using a two-point (CC and DT) calibration to make linear the detector response, even in the non-linear operating regions of the detector. The low-radiation field (CC) should be a field that yields from 2 to 5 % count loss. The high radiation field (DT) should be a field that yields from 30 to 60 % count loss. The algorithm ignores background counts, and therefore, the low field must be at least ten times the background count. The following summary lists the calibration constraints. Calibration and Dead Time Calibration Constraints FIELD CONSTRAINT BACKGROUND *10 times less than low field LOW FIELD Yields from 2 to 5 % count loss HIGH FIELD Yields from 30 to 60 % count loss * This constraint only applies when using two sources (two fields) or a radiation range calibrated without background consideration. Ludlum Measurements, Inc. Page 8-12 November 2014 Model 2241-3 & 3i Technical Manual Section 8 Preliminary CPS Setup Refer to Section 8, Subsection “Function Switch Position Descriptions and Variables,” for cps readout variables. Select position DET1 on the detector selector switch located on the front panel. Starting with FUNCTION switch position 1, enter the following variables: SWITCH POS. Equation 1 CPS LO 2% = 1 49 x DT PARAMETER 1 2 3 4 5 6 7 8 9 A B C D-F 0000s-6 0100 -2 C/ m N/A N/A N/A 0060 s Not Used Not Used N/A 000 C/s N/A FUNCTION Dead Time Calibration Constant Display Units Timebase Audio Divide-By Response Time Ratemeter Alm./Alert ScalerAlm./Count Time LCD Backlight Set Minimum Display RS-232 Parameters Equation 2 CPS LO 5% = 1 19 x DT Equation 3 CPS HI 30% = 1 2.3333 x DT Equation 4 CPS HI 60% = The equations to the left (Equations 1-4) determine the hi and lo radiation fields used to acquire counts for the CC and DTC algorithm. These calculations require an unknown variable, DT (Dead Time). Typical dead times for some of the standard LMI detectors are referenced in the table at the end of this section. The lo count field should be a field that yields between 2 and 5% count loss. The hi count field (CPSHI) should be a field that yields between 30 and 60% count loss. 1.5 DT Ludlum Measurements, Inc. Page 8-13 November 2014 Model 2241-3 & 3i Technical Manual Section 8 Reference the table at the end of this section to determine the cps/exposure rate (cps/ER). The conversion can be determined by placing the detector in a radiation field, which produces from 50 to 200 cps. Calculate the count/exposure = cps / ER rate using the equation to the left. cps radiation field in exposure rate units For example, exposing an LMI Model 449 to a 2 mR/hr 137Cs field yields approximately 110 cps so that: 110 cps = 55 cps/mR/hr 2 mR/hr The typical dead time for a Model 44-9 is approximately 85 µs. Therefore, using 85 µs for “DT” in equations 1-4, the lo field should be between 240 and 619 cps, and the hi field is between 5040-17,650 cps. Dividing the cps values by the 55 cps/mR/hr conversion equates to between 4-11 mR/hr for the lo field and 91-320 mR/hr for the hi field. Select a calibrated field between the lo and hi data points determined above: lo (CALlo) = 8 mR/hr hi (CALhi) = 200 mR/hr The following procedure outlines the hi-lo method Abbreviations used: units = Sv, R, counts. CALlo = lo field calibration point. CALhi = hi field calibration point. CORRlo = recorded field at low calibration point. CORRhi = recorded field at high calibration point. DT = dead time constant entered into Model 2241. CC = calibration constant entered into Model 2241. fd and ad are intermediate steps in calculating DT. fcal is an intermediate step in calculating CC. Ludlum Measurements, Inc. Page 8-14 November 2014 Model 2241-3 & 3i Technical Manual Section 8 CC and DTC Algorithm Equations (5) and (6) convert units per time (R/hr Display Units) to units per second: units units second time Insert the cps lo data point (8 mR/hr for the Model 44-9 example) determined from equations (1) and (2): Equation 5 CALlo = (0.008 1h 1m R ) x( ) x( ) = 2.22 x 10 6 s h 60 m 60 s Insert the cps hi data point (200 mR/hr for the Model 44-9 example) determined from equations (3) and (4): Equation 6 CALhi = (0.200 R 1h 1m ) = 55.6 x 10 6 s )x ( )x ( h 60 m 60 s Place the detector in the low field and enter the counts per second: Equation 7 CORRlo = SAMPLlo = count time counts s Note: The low field count sample should be 3000 counts. Use the scaler and adjust the count time to accumulate count 3000. Ludlum Measurements, Inc. Page 8-15 November 2014 Model 2241-3 & 3i Technical Manual Section 8 As an example, assume a 60-second count sample in a low field of 8 mR/hr: Example CORR lo 26,427 60 = = 440 C/s Place detector in the high field and enter the counts per second: Equation 8 CORR hi SAMPL hi = count time = counts s Counts/second sample in high field of 200 mR/hr: Example CORR hi = 5830 = 1 5830 C/s Insert the values calculated in equations (5), (6), (7), and (8) and solve for fd: Equation 9 f d = CAL hi - CORR hi x CAL CORR lo lo units s = Example f d = 55.6 x 10 6 - 5830 x 2.22 x 10 440 6 = 26.2 x 10 6 s Solve for ad: Equation 10 a d = ( CAL hi x CORR hi ) - ( CAL lo x CORR hi )= units count s 2 Example 6 6 a d = (55.6 x 10 x 5830) - (2.22 x 10 x 5830) = Ludlum Measurements, Inc. Page 8-16 31.1 x 10 6 November 2014 Model 2241-3 & 3i Technical Manual Section 8 Enter the results of equations (9) and (10) into equation (11) to solve for DT: Equation 11 DT = f d s count = ad Example DT = 26 . 2 x10 6 = 31 . 1 x10 2 8 . 4 x10 6 s Solve for fcal: Equation 12 f cal = CAL lo - ( CAL lo x CORR lo x DT) = units s Example f cal = 2.22 x 10 6 - (2.22 x 10 6 x 440 x 84 x 10 6 ) = 2.14 x 10 6 s Enter the result of equation (12) into: Equation 13 CC = CORR f cal lo count units = and solve for CC: Example CC = 440 2.14 x 10 6 = 206 x 10 6 Enter the CC and DT values (positions 1 and 2 of the FUNCTION switch), derived from the equations above. Perform an “R/hr calibration” as described in the previous subsection in order to ensure that the instrument and detector have been correctly calibrated. Ludlum Measurements, Inc. Page 8-17 November 2014 Model 2241-3 & 3i Technical Manual Section 8 Model 44-9 Detector Parameter Setup FUNCTION PARAMETER 1 2 4 5-8 B-C D-F 0084 s-6 0206 06 N/A as desired as desired if applicable Typical Count Rate and Dead Time for LMI Detectors MODEL & TYPE COUNT RATE 44-6, GM 44-9, GM 44-7, GM 133-2, GM 133-4, GM 133-6, GM 44-2, Gamma Scint. 44-10, Gamma Scint. 44-3, Low-Energy Gamma Scint. 44-21, Beta/Gamma Scint. 43-5, Alpha Scint. DEAD TIME in µs (microseconds) 20 cps/mR/hr 55 cps/mR/hr 35 cps/mR/hr 17.5 cps/mR/hr 2 cps/mR/hr 0.3 cps/mR/hr 2800 cps/mR/hr 15,000 cps/mR/hr N/A, operated in Counts/units mode N/A, operated in Counts/units mode N/A, operated in Counts/units mode 90-110 µs 80-90 µs 240-290 µs 40-55 µs 40-55 µs 40-55 µs 8-12 µs* 18-20 µs 8-12 µs* 8-12 µs* 20-28 µs Note: The data represented in the table above is typical. Actual values may vary among detector and instrument combinations. This table represents some of the common detectors operated with the Model 2241-3. Consult the LMI sales department for information concerning detectors not listed in the table above. *The dead time values for these scintillation detectors are due to the dead time of the Model 2241-3 electronics. Ludlum Measurements, Inc. Page 8-18 November 2014 Model 2241-3 & 3i Technical Manual Ludlum Measurements, Inc. Section 8 Page 8-19 November 2014 Model 2241-3 & 3i Technical Manual Ludlum Measurements, Inc. Section 8 Page 8-20 November 2014 Model 2241-3 & 3i Technical Manual Ludlum Measurements, Inc. Section 8 Page 8-21 November 2014 Model 2241-3 & 3i Technical Manual 9 Section 9 Section Recycling L udlum Measurements, Inc. supports the recycling of the electronic products it produces for the purpose of protecting the environment and to comply with all regional, national, and international agencies that promote economically and environmentally sustainable recycling systems. To this end, Ludlum Measurements, Inc. strives to supply the consumer of its goods with information regarding reuse and recycling of the many different types of materials used in its products. With many different agencies – public and private – involved in this pursuit, it becomes evident that a myriad of methods can be used in the process of recycling. Therefore, Ludlum Measurements, Inc. does not suggest one particular method over another, but simply desires to inform its consumers of the range of recyclable materials present in its products, so that the user will have flexibility in following all local and federal laws. The following types of recyclable materials are present in Ludlum Measurements, Inc. electronic products, and should be recycled separately. The list is not all-inclusive, nor does it suggest that all materials are present in each piece of equipment: Batteries Glass Aluminum and Stainless Steel Circuit Boards Plastics Liquid Crystal Display (LCD) Ludlum Measurements, Inc. products, which have been placed on the market after August 13, 2005, have been labeled with a symbol recognized internationally as the “crossed-out wheelie bin.” This notifies the consumer that the product is not to be mixed with unsorted municipal waste when discarding. Each material must be separated. The symbol will be placed near the AC receptacle, except for portable equipment where it will be placed on the battery lid. The symbol appears as such: Ludlum Measurements, Inc. Page 9-1 November 2014 Model 2241-3 & 3i Technical Manual Section 10 Section 10 Model 2241-3 Survey Meter Main Circuit Board, Drawing 408 × 285 CAPACITORS Ludlum Measurements, Inc. Parts List Reference Description Part Number UNIT Completely Assembled Model 2241-3 Survey Meter 48-2864 Completely Assembled Main Circuit Board 5408-285 0.1µF, 50V 0.1µF, 50V 47pF, 100V 0.001µF, 100V 0.1µF, 50V 100pF, 3KV 0.0047µF, 3KV 100pF, 3KV 0.0047µF, 3KV 0.0047µF, 3KV 47uF, 10V 47uF, 10V 27pF, 100V 0.0047µF, 3KV 0.1µF, 50V 0.01µF, 50V 47µF, 10V 0.01µF, 50V 47µF, 10V 100pF, 100V 0.001µF, 100V 0.0047F, 3KV 04-5663 04-5663 04-5660 04-5659 04-5663 04-5532 04-5547 04-5532 04-5547 04-5547 04-5666 04-5666 04-5658 04-5547 04-5663 04-5664 04-5666 04-5664 04-5666 04-5661 04-5659 04-5547 BOARD C1 C3 C001-C002 C011 C012 C021 C031 C032 C033 C041-C043 C101 C121 C122-C123 C131 C132-C133 C134 C135 C136 C137 C138 C139 C141 Page 10-1 November 2014 Model 2241-3 & 3i Technical Manual Section 10 Reference Description Part Number C241 C242 C243 C251 1µF, 35V 68µF, 10V 0.1µF, 50V 68µF, 10V 04-5656 04-5654 04-5663 04-5654 TRANSISTORS Q101 Q141 Q211 Q212 Q241 2N7002L MMBT3904LT1 2N7002L MMBT4403LT1 MJD210 RL 05-5840 05-5841 05-5840 05-5842 05-5843 INTEGRATED CIRCUITS U1 U001 U011 U021 U111 U121 U122 U131 U231 MAX810LEUR CD74HC4538M TLC372ID CA3096M; 16=GND N87C51FC LM285MX-2.5 X24C02S8T5 LM358D LT1073CS8-5 06-6424 06-6297 06-6290 06-6288 06-6303 06-6291 06-6299 06-6312 05-5852 DIODES CR021 CR031 CR041-CR044 CR231 CR241 CR242 MMBD7000LT1 GI250-2 GI250-2 CXSH-4 EB33 MMBD914LT1 CXSH-4 EB33 07-6355 07-6266 07-6266 07-6358 07-6353 07-6358 R002 R027 10K; 3269X1-103, VOLUME 1M; 3269X1-105, HV LIMIT 09-6921 09-6906 R001 R011-R012 R013 R014 R015 R021 R022 R023 R024-R025 100K, 1/4W, 1% 10K, 1/4 W, 1% 1K, 1/4W, 1% 10K, 1/4W, 1% 100K, 1/4W, 1% 1M, 1/4W, 5% 392K, 1/8W, 1% 10K, 1/4W, 1% 4.75K, 1/4W, 1% 12-7834 12-7839 12-7832 12-7839 12-7834 10-7028 12-7841 12-7839 12-7858 POTENTIOMETERS / TRIMMERS RESISTORS Ludlum Measurements, Inc. Page 10-2 November 2014 Model 2241-3 & 3i Technical Manual Section 10 Reference Description Part Number R026 R031 R032 R033-R034 R111-R113 R121 R122 R131 R132 R133 R134 R135 R136 R141 R211 R231 R241 R242 8.25K, 1/8W, 1% 4.7M, 1/4W, 5% 1M, 1/4W, 5% 1G, FHV-1, 2% 22.1K, 1/4W, 1% 100Ohm, 1/4W, 1% 6.81K, 1/4W, 1% 1M, 1/4W, 1% 511K, 1/8W, 1% 750K, 1/4W, 1% 1M, 1/4W, 1% 82.5K, 1/8W, 1% 10K, 1/4W, 1% 22.1K, 1/4W, 1% 2.21K, 1/4W, 1% 100Ohm, 1/4W, 1% 2.21K, 1/4W, 1% 200Ohm, 1/8W, 1% 12-7838 10-7030 10-7028 12-7686 12-7843 12-7840 12-7857 12-7844 12-7896 12-7882 12-7844 12-7849 12-7839 12-7843 12-7835 12-7840 12-7835 12-7846 CRYSTALS Y221 6.144 MHZ, 2=GND, 3=GND 01-5262 INDUCTOR L231 100uH, CTX100-2 21-9740 TRANSFORMERS T101 T141 4275-083, AUDIO L8050 4275-083 40-0902 MISCELLANEOUS P1-P2 P3 P4 P5 1-640456-3, MTA100X13 640456-6, MTA100X6 640456-2, MTA100X2 1-640456-2, MTA100X12 13-8100 13-8095 13-8073 13-8061 BOARD Completely Assembled Calibration Board 5408-113 1M, HV1 SET 100K, DISC. 1M, HV2 SET 1M, HV3 SET 1M, HV4 SET 09-6814 09-6813 09-6814 09-6814 09-6814 Calibration Board, Drawing 408 × 98 POTENTIOMETERS Ludlum Measurements, Inc. R115 R116 R117 R118 R119 Page 10-3 November 2014 Model 2241-3 & 3i Technical Manual Section 10 Reference Description Part Number RESISTORS R110 R111 R112 R113 R114 R120 1M, 1/3W 10K, 1/3W 1M, 1/3W 1K, 1/3W 1M, 1/3W 1M, 1/3W 12-7751 12-7748 12-7751 12-7750 12-7751 12-7751 CONNECTOR P7 CONN-640456-7, MTA100 13-8115 BOARD Completely Assembled Display Board 5408-259 Display Board, Drawing 408 × 259 CAPACITORS C1 27PF, 100V 04-5658 INTEGRATED CIRCUITS U1 U2 AY0438-I/L AY0438-I/L 06-6358 06-6358 RESISTORS R001-R004 R005 10.0K, 1%, 125mW 392 Ohm, 1%, 1/8 W 12-7839 12-7054 MISCELLANEOUS J1 DS1 DSP1 CONN-640456-8, MTA100 EL-BACKLIGHT-LED MAIN DISPLAY; LCD-8246-365-4E1-A/W-REV1 13-8039 07-6527 Ludlum Measurements, Inc. Page 10-4 07-6383 November 2014 Model 2241-3 & 3i Technical Manual Switch Board, Drawing 408 × 45 Section 10 Reference Description Part Number BOARD Completely Assembled Switch Board 5408-052 CAPACITORS C1-C2 C3-C4 C5 C6 4.7µF, 10V 10µF, 20V 4.7µF, 10V 100µF, 10V 04-5578 04-5592 04-5578 04-5576 INTEGRATED CIRCUITS U1 MAX220EPE 06-6359 SWITCHES S1 S2 S3 S4 350134GSK; FUNCTION; 16 POS LEFT UP ENTER 08-6721 08-6716 08-6716 08-6716 RESISTORS R1-R2 22K 10-7070 MISCELLANEOUS P6 P10 CONN-1-640456-3, MTA100 CONN-208006-2 13-8100 13-8451 S1 S2 S3-S4 S5 S6 S7 * 30-1-PB GRAYHILL PA-1012 7101-SYZ-QE C&K 30-1-PB GRAYHILL 7201KZQE Toggle MPS-103F Switch Cap 08-6517 08-6540 08-6511 08-6517 08-6749 08-6699 08-6871 RESISTOR R1 10 M, 1/4W, 5% 10-7031 CONNECTORS J1 J2 J3 J4 J5 J6 J7 J8 J9 J10 P10 CONN-1-640442-3, MTA100 CONN-640442-6, MTA100 CONN-640442-2, MTA100 CONN-1-640442-2, MTA100 CONN-1-640442-3, MTA100 CONN-640442-7, MTA100 CONN-640442-8, MTA100 Series "C" -UG706/U JACK-09-9011-1-4193 Handle Pin 13-8138 13-8171 13-8178 13-8407 13-8138 13-8172 13-8184 13-7777 18-9080 7408-055 Chassis Wiring Diagram, Drawing 408 × 101 SWITCHES Ludlum Measurements, Inc. Page 10-5 November 2014 Model 2241-3 & 3i Technical Manual Section 10 Reference Description Part Number AUDIO DS1 UNIMORPH 21-9251 BATTERY B1-B2 ˝D˝ Duracell Battery 21-9313 MISCELLANEOUS * Model 2241 Switch Board Add On Model 2241 Digital Bezel Assembly Portable Battery Contact Set Model 2241-3 Main Harness Portable Harness Can Wires Switch Board Harness Portable Can Assembly (MTA) Can Gasket Portable Knob Assembled Battery Lid Portable Battery Gasket Portable Calibration Cover with Screws Model 2241-3 Flat Handle Assembly without Clip Cable-C 1 meter (39 in.) only Model 2241-3 Casting * * * * * * * * * * * * * * Ludlum Measurements, Inc. Page 10-6 4408-053 4408-051 2001-042 8408-121 8363-462 8408-027 4363-441 22-9773 08-6613 2363-191 7363-183 9363-200 4408-179 40-1004 7408-185 November 2014 Model 2241-3 & 3i Technical Manual Section 11 Section 11 Drawings Main Circuit Board, Drawings 408 × 285 (3 sheets) Main Circuit Board Component Layout, Drawing 408 × 286 Calibration Board, Drawing 408 × 98 Calibration Board Component Layout, Drawings 408 × 99 (2 sheets) Display Board, Drawing 408 × 259 Display Board Component Layout, Drawings 408 × 260 (2 sheets) Switch Board, Drawing 408 × 45 Switch Board Component Layout, Drawing 408 × 46 Wiring Diagram, Drawing 408 ×101 Wiring Diagram for Model 2241-3i, Drawing 408 × 263 Ludlum Measurements, Inc. Page 11-1 November 2014 Model 2241-3 & 3i Technical Manual Appendix A Appendix A RS-232 Output Formats The Ludlum Model 2241 series of instruments has an RS-232 serial communications port that can be used to log readings and read or set instrument parameters. There are two formats available. Most Model 2241 instruments have the binary format outlined below, but some newer Model 2241-2 units have an ASCII output, which is also outlined below following the binary format explanation. The RS-232 port is configured at 9600 baud, 8 data bits, no parity, and 1 stop bit (9600,8,N,1). Ludlum Measurements can supply a Windows-based software that can be used to help calibrate the instruments, but note that it will not communicate with the newer ASCII output Model 2241-2 units. Binary Output Format (15 Bytes) BYTE01 BYTE02 BYTE03 BYTE04 BYTE05 BYTE06 BYTE07 BYTE08 BYTE09 BYTE10 BYTE11 BYTE12 BYTE13 BYTE14 BYTE15 RatemeterCPS+0 RatemeterCPS+1 RatemeterCPS+2 RatemeterCPE+3 Scaler+0 Sclaer+1 Scaler+2 Scaler+3 Scaler+4 CountTime+0 CountTime+1 CountTimeLeft+0 CountTimeLeft+1 Carriage Return (0DH) Line Feed (0AH) MSB LSB MSB LSB MSB LSB MSB LSB The ratemeter value is in cps and is scaled by a factor of 256. To get the ratemeter reading in cpm, take the value and divide by 256, and then multiply by 60. Ludlum Measurements, Inc. Page A-1 November 2014 Model 2241-3 & 3i Technical Manual Appendix A RS-232 Commands E – auto dump off A – auto dump on C – start scaler F – set scaler count time R – send parameters from instrument to computer S – read parameters from computer to instrument O – output once All commands must be sent in upper case laters. Output of “R” Command – Read Parameters BYTE01 Detector+0 BYTE02 DeadConstant+0 MSB BYTE03 DeadConstant+1 LSB BYTE04 CaConstant+0 MSB BYTE05 CalConstant+1 BYTE06 CalConstant+2 BYTE07 CalConstant+3 BYTE08 CalConstant+4 BYTE09 CalConstant+5 LSB BYTE10 RateAlarm+0 MSB BYTE11 RateAlarm+1 BYTE12 RateAlarm+2 BYTE13 RateAlarm+3 BYTE14 RateAlarm+4 LSB Ludlum Measurements, Inc. BYTE15 BYTE16 BYTE17 BYTE18 BYTE19 BYTE20 BYTE21 BYTE22 BYTE23 BYTE24 BYTE25 ScalerAlarm+0 ScalerAlarm+1 ScalerAlarm+2 ScalerAlarm+3 ScalerAlarm+4 CountTime+0 CountTime+1 Units+0 TimeBase+0 AudioDivideBy+0 Response+0 BYTE26 BYTE27 BYTE28 BYTE29 BYTE30 BYTE31 BYTE32 BYTE33 RateAlert+0 RateAlert+1 RateAlert+2 RateAlert+3 RateAlert+4 CheckSource+0 CheckSource+1 CheckSource+2 Page A-2 MSB LSB MSB LSB MSB LSB MSB November 2014 Model 2241-3 & 3i Technical Manual BYTE34 BYTE35 BYTE36 BYTE37 BYTE38 BYTE39 CheckSource+3 CheckSource+4 PercentCS+0 MinDisplay+0 Carriage Return (0DH) Line Feed (0AH) Appendix A LSB Input of “S” Command – Send Parameters BYTE1 DeadCosntant+0 MSB BYTE2 DeadConstant+1 LSB BYTE3 CalConstant+0 MSB BYTE4 CalConstant+1 BYTE5 CalConstant+2 BYTE6 CalCosntant+3 BYTE7 CalConstant+4 BYTE8 CalConstant+5 LSB BYTE9 RateAlarm+0 MSB BYTE10 RateAlarm+1 BYTE11 RateAlarm+2 BYTE12 RateAlarm+3 BYTE13 RateAlarm+4 LSB Ludlum Measurements, Inc. BYTE14 BYTE15 BYTE16 BYTE17 BYTE18 BYTE19 BYTE20 BYTE21 BYTE22 BYTE23 BYTE24 ScalerAlarm+0 ScalerAlarm+1 ScalerAlarm+2 ScalerAlarm+3 ScalerAlarm+4 CountTime+0 CountTime+1 Units+0 TimeBase+0 AudioDivdeBy+0 Response+0 BYTE25 BYTE26 BYTE27 BYTE28 BYTE29 BYTE30 BYTE31 BYTE32 BYTE33 BYTE34 BYTE35 BYTE36 RateAlert+0 RateAlert+1 RateAlert+2 RateAlert+3 RateAlert+4 CheckSource+0 CheckSource+1 CheckSource+2 CheckSource+3 CheckSource+4 PercentCS+0 MinDisplay+0 Page A-3 MSB LSB MSB LSB MSB LSB MSB LSB November 2014 Model 2241-3 & 3i Technical Manual Appendix A Input of “F” Command – Set Count Time BYTE1 CountTime+0 MSB BYTE2 CountTime+1 LSB Units Timebase 1 = Sv 2 = cpm 0=R 0 = min 1 = seconds AudioDivide 0 = Auto 1 = Manual ASCII Output Format This special firmware changes the format of the auto-dump from binary to ASCII. The output interval remains the same (2 seconds). 2241 2241-2 2241-3 Special Firmware 40804n10 Special Firmware 40806n13 Special Firmware 40806n16 08/06/2003 04/04/2003 07/07/2004 The old format was (binary): Byte 1 Byte 2 Byte 3 Byte 4 Byte 5 Byte 6 Byte 7 Byte 8 Byte 9 Byte 10 Byte 11 Byte 12 Byte 13 Byte 14 Byte 15 Ratemeter CPS + 0 Ratemeter CPS + 1 Ratemeter CPS + 2 Ratemeter CPS + 3 Scaler + 0 Scaler + 1 Scaler + 2 Scaler + 3 Scaler + 4 Count Time + 0 Count Time + 1 Count Time Left + 0 Count Time Left + 1 Carriage Return (0DH) Line Feed (0AH) and is now replaced with (ASCII): byte 1 byte 2 byte 3 byte 4 byte 5 byte 6 byte 7 Ludlum Measurements, Inc. x Ratemeter x Raetemter x Ratemeter x Ratemeter x Ratemeter y Display units Carriage Return (0DH) Page A-4 November 2014 Model 2241-3 & 3i Technical Manual byte 8 Appendix A Line Feed (0AH) The ratemeter is displayed as 5 ASCII digits with a decimal, if necessary, and matches the LCD display on the 2241-2. The display mode is a value from 0 to 9 representing the display units. CPS KCPS CPM KCPM µR mR R µSv mSv Sv 0 1 2 3 4 5 6 7 8 9 Note: The LMI Model 2241 calibration software is not compatible with this firmware version. Example output: 02.991 01.801 01.221 00.831 00.581 004160 Ludlum Measurements, Inc. (=2.99 kcps) (=1.80 kcps) (=1.22 kcps) (=0.83 kcps) (=0.58 kcps) (=416 cps) Page A-5 November 2014
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Related manuals
advertisement