NI 5401 User Manual

NI 5401 User Manual

March 1999 Edition

Part Number 322419A-01

Worldwide Technical Support and Product Information www.natinst.com

National Instruments Corporate Headquarters

11500 North Mopac Expressway Austin, Texas 78759-3504 USA Tel: 512 794 0100

Worldwide Offices

Australia 03 9879 5166, Austria 0662 45 79 90 0, Belgium 02 757 00 20, Brazil 011 284 5011,

Canada (Ontario) 905 785 0085, Canada (Québec) 514 694 8521, China 0755 3904939, Denmark 45 76 26 00,

Finland 09 725 725 11, France 01 48 14 24 24, Germany 089 741 31 30, Hong Kong 2645 3186,

India 91805275406, Israel 03 6120092, Italy 02 413091, Japan 03 5472 2970, Korea 02 596 7456,

Mexico (D.F.) 5 280 7625, Mexico (Monterrey) 8 357 7695, Netherlands 0348 433466, Norway 32 27 73 00,

Singapore 2265886, Spain (Madrid) 91 640 0085, Spain (Barcelona) 93 582 0251, Sweden 08 587 895 00,

Switzerland 056 200 51 51, Taiwan 02 2377 1200, United Kingdom 01635 523545

For further support information, see the Technical Support Resources appendix. To comment on the documentation, send e-mail to [email protected]

.

© Copyright 1999 National Instruments Corporation. All rights reserved.

Important Information

Warranty

The NI 5401 is warranted against defects in materials and workmanship for a period of one year from the date of shipment, as evidenced by receipts or other documentation. National Instruments will, at its option, repair or replace equipment that proves to be defective during the warranty period. This warranty includes parts and labor.

The media on which you receive National Instruments software are warranted not to fail to execute programming instructions, due to defects in materials and workmanship, for a period of 90 days from date of shipment, as evidenced by receipts or other documentation. National Instruments will, at its option, repair or replace software media that do not execute programming instructions if National Instruments receives notice of such defects during the warranty period.

National Instruments does not warrant that the operation of the software shall be uninterrupted or error free.

A Return Material Authorization (RMA) number must be obtained from the factory and clearly marked on the outside of the package before any equipment will be accepted for warranty work. National Instruments will pay the shipping costs of returning to the owner parts which are covered by warranty.

National Instruments believes that the information in this document is accurate. The document has been carefully reviewed for technical accuracy. In the event that technical or typographical errors exist, National Instruments reserves the right to make changes to subsequent editions of this document without prior notice to holders of this edition. The reader should consult National Instruments if errors are suspected. In no event shall National Instruments be liable for any damages arising out of or related to this document or the information contained in it.

E

XCEPT AS SPECIFIED HEREIN

, N

ATIONAL

I

NSTRUMENTS MAKES NO WARRANTIES

,

EXPRESS OR IMPLIED

,

AND SPECIFICALLY DISCLAIMS

ANY WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE

. C

USTOMER

S RIGHT TO RECOVER DAMAGES CAUSED

BY FAULT OR NEGLIGENCE ON THE PART OF

N

ATIONAL

I

NSTRUMENTS SHALL BE LIMITED TO THE AMOUNT THERETOFORE PAID BY THE

CUSTOMER

. N

ATIONAL

I

NSTRUMENTS WILL NOT BE LIABLE FOR DAMAGES RESULTING FROM LOSS OF DATA

,

PROFITS

,

USE OF PRODUCTS

,

OR INCIDENTAL OR CONSEQUENTIAL DAMAGES

,

EVEN IF ADVISED OF THE POSSIBILITY THEREOF

. This limitation of the liability of

National Instruments will apply regardless of the form of action, whether in contract or tort, including negligence.

Any action against National Instruments must be brought within one year after the cause of action accrues. National

Instruments shall not be liable for any delay in performance due to causes beyond its reasonable control. The warranty provided herein does not cover damages, defects, malfunctions, or service failures caused by owner’s failure to follow the National Instruments installation, operation, or maintenance instructions; owner’s modification of the product; owner’s abuse, misuse, or negligent acts; and power failure or surges, fire, flood, accident, actions of third parties, or other events outside reasonable control.

Copyright

Under the copyright laws, this publication may not be reproduced or transmitted in any form, electronic or mechanical, including photocopying, recording, storing in an information retrieval system, or translating, in whole or in part, without the prior written consent of National Instruments Corporation.

Trademarks

ComponentWorks ™ , CVI ™ , LabVIEW ™ , natinst.com

™ , PXI ™ , RTSI ™ , and VirtualBench ™ are trademarks of

National Instruments Corporation.

Product and company names mentioned herein are trademarks or trade names of their respective companies.

WARNING REGARDING MEDICAL AND CLINICAL USE OF NATIONAL INSTRUMENTS PRODUCTS

National Instruments products are not designed with components and testing for a level of reliability suitable for use in or in connection with surgical implants or as critical components in any life support systems whose failure to perform can reasonably be expected to cause significant injury to a human. Applications of National Instruments products involving medical or clinical treatment can create a potential for death or bodily injury caused by product failure, or by errors on the part of the user or application designer. Because each end-user system is customized and differs from

National Instruments testing platforms and because a user or application designer may use National Instruments products in combination with other products in a manner not evaluated or contemplated by National Instruments, the user or application designer is ultimately responsible for verifying and validating the suitability of National Instruments products whenever National Instruments products are incorporated in a system or application, including, without limitation, the appropriate design, process and safety level of such system or application.

Conventions

»

♦ bold italic monospace monospace italic

The following conventions are used in this manual:

The » symbol leads you through nested menu items and dialog box options to a final action. The sequence File»Page Setup»Options directs you to pull down the File menu, select the Page Setup item, and select Options from the last dialog box.

The

symbol indicates that the following text applies only to a specific product, a specific operating system, or a specific software version.

This icon denotes a note, which alerts you to important information.

This icon denotes a caution, which advises you of precautions to take to avoid injury, data loss, or a system crash.

This icon denotes a warning, which advises you of precautions to take to avoid being electrically shocked.

Bold text denotes items that you must select or click on in the software, such as menu items and dialog box options. Bold text also denotes parameter names.

Italic text denotes variables, emphasis, a cross reference, or an introduction to a key concept.

Text in this font denotes text or characters that you should enter from the keyboard. This font is also used for the proper names of functions, variables, and filenames and extensions.

Italic text in this font denotes text that is a placeholder for a word or value that you must supply.

Contents

Chapter 1

Generating Functions with the NI 5401

About Your NI 5401 ......................................................................................................1-1

Connecting Signals ........................................................................................................1-2

ARB Connector ...............................................................................................1-3

SYNC Connector.............................................................................................1-3

PLL Ref Connector..........................................................................................1-4

Pattern Out Connector (PCI Only) ..................................................................1-5

Connector Pin Assignments ..............................................................1-5

Signal Descriptions ...........................................................................1-6

SHC50-68 50-Pin Cable Connector ................................................................1-6

Software Options for Your NI 5401 ..............................................................................1-8

Software Included with Your NI 5401 ............................................................1-8

VirtualBench .....................................................................................1-8

NI-FGEN Instrument Driver .............................................................1-9

Additional National Instruments Development Tools.....................................1-9

LabVIEW ..........................................................................................1-9

LabWindows/CVI .............................................................................1-10

ComponentWorks .............................................................................1-10

Using the Soft Front Panels to Generate Waveforms ....................................................1-11

Generating Standard Functions .......................................................................1-11

Generating Multiple Frequencies in a Sequence .............................................1-13

Waveform Editor .............................................................................................1-15

Power-Up and Reset Conditions....................................................................................1-16

Chapter 2

Function Generator Operation

Generating Waveforms ..................................................................................................2-2

Direct Digital Synthesis (DDS) .....................................................................................2-3

Frequency Hopping and Sweeping..................................................................2-4

Triggering ......................................................................................................................2-4

Trigger Sources ...............................................................................................2-4

Modes of Operation .........................................................................................2-5

Single Trigger Mode .........................................................................2-5

Continuous Trigger Mode .................................................................2-6

Stepped Trigger Mode ......................................................................2-7

Analog Output................................................................................................................2-7

© National Instruments Corporation v

NI 5401 User Manual

Contents

SYNC Output and Duty Cycle........................................................................ 2-9

Output Attenuation.......................................................................................... 2-9

Output Impedance ........................................................................................... 2-10

Output Enable ................................................................................................. 2-10

Pre-Attenuation Offset .................................................................................... 2-11

Phase-Locked Loops and Board Synchronization......................................................... 2-11

Analog Filter Correction................................................................................................ 2-13

RTSI/PXI Trigger Lines ................................................................................................ 2-14

Calibration ..................................................................................................................... 2-15

Appendix A

Specifications

Appendix B

Optional Accessories

Appendix C

Frequency Resolution and Lookup Memory

Appendix D

Technical Support Resources

Glossary

Index

Figures

Figure 1-1.

NI 5401 I/O Connectors........................................................................ 1-2

Figure 1-2.

Output Levels and Load Termination Using a

50

Ω

Output Impedance ........................................................................ 1-3

Figure 1-3.

SYNC Output and Duty Cycle.............................................................. 1-4

Figure 1-4.

NI 5401 50-Pin Digital Connector Pin Assignments ............................ 1-5

Figure 1-5.

SHC50-68 68-Pin Connector Pin Assignments .................................... 1-7

Figure 1-6.

VirtualBench-FG Soft Front Panel for Function Generation................ 1-11

Figure 1-7.

VirtualBench-FG General Settings Dialog Box for the NI 5401.......... 1-12

Figure 1-8.

VirtualBench-FG Signals Settings Dialog Box for the NI 5401........... 1-12

Figure 1-9.

VirtualBench-FG Load Waveform Dialog Box.................................... 1-13

Figure 1-10.

VirtualBench-FG Frequency List Editor Dialog Box ........................... 1-14

Figure 1-11.

Waveform Editor Soft Front Panel ....................................................... 1-15

NI 5401 User Manual

vi www.natinst.com

Contents

Figure 2-1.

NI 5401 Block Diagram ........................................................................2-1

Figure 2-2.

Waveform Data Path Block Diagram....................................................2-2

Figure 2-3.

DDS Building Blocks ............................................................................2-3

Figure 2-4.

Waveform Generation Trigger Sources ................................................2-5

Figure 2-5.

Single Trigger Mode .............................................................................2-6

Figure 2-6.

Continuous Trigger Mode .....................................................................2-6

Figure 2-7.

Stepped Trigger Mode...........................................................................2-7

Figure 2-8.

Analog Output and SYNC Out Block Diagram ....................................2-8

Figure 2-9.

Waveform and Trigger Timings............................................................2-8

Figure 2-10.

Output Attenuation Chain .....................................................................2-9

Figure 2-11.

PLL Architecture for the NI 5401 for PCI ............................................2-12

Figure 2-12.

PLL Architecture for the NI 5401 for PXI ............................................2-12

Figure 2-13.

Analog Filter Correction .......................................................................2-13

Figure 2-14.

RTSI Trigger Lines and Routing for the NI 5401 for PCI ....................2-14

Figure 2-15.

PXI Trigger Lines, 10 MHz Backplane Oscillator, and

Routing for the NI 5401 for PXI ...........................................................2-14

Table

Table 1-1.

Digital Connector Signal Descriptions..................................................1-6

© National Instruments Corporation vii

NI 5401 User Manual

1

Generating Functions with the NI 5401

The NI 5401 User Manual describes the features, functions, and operation of the NI 5401 arbitrary function generator. This device performs comparably to standalone instruments while providing the flexibility of computer-based operation.

About Your NI 5401

Thank you for buying a National Instruments NI 5401 arbitrary function generator. The NI 5401 family consists of two different devices:

• NI 5401 for PCI

• NI 5401 for PXI

Your NI 5401 device has the following features:

• One 12-bit resolution output channel

• Up to 16 MHz sine and transistor-transistor logic (TTL) waveform output

• Software-selectable output impedances of 50

Ω

and 75

Ω

• Output attenuation levels from 0 to 73 dB

• Phase-locked loop (PLL) synchronization to external clocks

• Sampling rate of 40 MS/s

• Digital and analog filter

• 32-bit direct digital synthesis (DDS) for standard function generation

• External trigger input

• Real-Time System Integration (RTSI) and PXI triggers

All NI 5401 devices follow industry-standard Plug and Play specifications on both buses and offer seamless integration with compliant systems.

Detailed specifications for the NI 5401 are in Appendix A, Specifications .

© National Instruments Corporation 1-1

NI 5401 User Manual

Chapter 1 Generating Functions with the NI 5401

Connecting Signals

Figure 1-1 shows the front panels for the NI 5401 for the PXI and PCI

buses. The front panel contains three types of connectors: BNC, SMB, and

50-pin very high-density SCSI (VHDSCSI). The main waveform is generated through the connector labeled ARB.

ARB OUT

LOCK ACCESS

EXT TRIG

SYNC OUT

PLL REF

ARB

SYNC

PLL IN

NI 5401 User Manual

PXI

Figure 1-1. NI 5401 I/O Connectors

PCI

1-2 www.natinst.com

Chapter 1 Generating Functions with the NI 5401

ARB Connector

The ARB connector provides the waveform output. The maximum output levels on this connector depend on the type of load termination. If the output of your NI 5401 terminates into a 50

Ω

load, the output levels are

±5 V, as shown in Figure 1-2. If the output of your NI 5401 terminates into

a high-impedance load (HiZ), the output levels are ±10 V. If the output terminates into any other load, the levels are as follows:

V out

=

±

R

-------------------

R

L

+ R

O

×

10 V where V out

is the maximum output voltage level

R

L

is the load impedance in ohms, and

R

O

is the output impedance on the NI 5401.

By default, R

O

= 50

Ω

, but you can use your software to set it to 75

Ω

.

NI 5401

ARB Load

NI 5401

ARB Load

R

O

=

50

Ω

R

L

=

50

Ω

±5 V

R

O

=

50

Ω

R

L

=

HiZ

±10 V

50

Ω

Load High-Impedance Load

Figure 1-2. Output Levels and Load Termination Using a 50

Ω

Output Impedance

Note Software sets the voltage output levels based on a 50

Ω

load termination.

For more information on waveform generation and analog output

operation, refer to Chapter 2, Function Generator Operation

. For

specifications on the waveform output signal, see Appendix A,

Specifications

.

SYNC Connector

The SYNC connector provides a TTL version of the sine waveform being generated at the output. You can think of the SYNC output as a very high-frequency resolution, software-programmable clock source for many applications. You can also vary the duty cycle of the SYNC output on the

fly by software control, as shown in Figure 1-3. t

p

is the time period of the

© National Instruments Corporation 1-3

NI 5401 User Manual

Chapter 1 Generating Functions with the NI 5401 sine wave being generated and t w

The duty cycle is (t w

/t p

)

×

100%.

is the pulse width of the SYNC output.

ARB Output

SYNC Output

(50% Duty Cycle)

SYNC Output

(33% Duty Cycle) t w t p

Figure 1-3. SYNC Output and Duty Cycle

For your NI 5401 for PCI, you can route the SYNC output to the RTSI lines over the RTSI bus. For your NI 5401 for PXI, you can route the SYNC output to the TTL trigger lines over the TTL trigger bus. The SYNC output is derived from a comparator connected to the analog waveform and provides a meaningful waveform only when you are generating a sine wave

on the ARB output. For more information on SYNC output, see Chapter 2,

Function Generator Operation .

PLL Ref Connector

The PLL Ref connector is a phase-locked loop (PLL) input connector that can accept a reference clock from an external source and frequency lock the

NI 5401 internal clock to this external clock. The reference clock should not deviate more than ±100 ppm from its nominal frequency. The minimum amplitude levels of 1 V pk-pk

are required on this clock. You can lock reference clock frequencies of 1 MHz and 5–20 MHz in 1 MHz steps.

Note You can frequency lock the NI 5401 for PCI to other National Instruments devices over the RTSI bus using the 20 MHz RTSI clock signal. You can frequency lock the

NI 5401 for PXI to other National Instruments devices using the 10 MHz backplane clock.

If no external reference clock is available, the NI 5401 automatically tunes the internal clock to the highest accuracy possible. For more information on

PLL operation, refer to Chapter 2, Function Generator Operation .

NI 5401 User Manual

1-4 www.natinst.com

Chapter 1 Generating Functions with the NI 5401

Pattern Out Connector (PCI Only)

This connector is used on the NI 5401 for PCI to supply the external trigger input to the board.

Connector Pin Assignments

Figure 1-4 shows the NI 5401 50-pin digital connector. Refer to Table 1-1

for a description of the signals.

41

40

39

38

37

36

35

34

33

32

31

30

29

28

27

26

50

49

48

47

46

45

44

43

42

16

15

14

13

12

11

10

9

8

7

6

5

4

3

2

1

25

24

23

22

21

20

19

18

17

RFU

DGND

RFU

DGND

RFU

DGND

RFU

DGND

RFU

DGND

RFU

DGND

RFU

DGND

RFU

DGND

DGND

NC

DGND

NC

DGND

NC

DGND

RFU

DGND

RFU

RFU

RFU

RFU

RFU

RFU

RFU

RFU

RFU

RFU

RFU

RFU

RFU

RFU

RFU

RFU

EXT_TRIG

NC

NC

NC

NC

NC

NC

RFU

RFU

Figure 1-4. NI 5401 50-Pin Digital Connector Pin Assignments

© National Instruments Corporation 1-5

NI 5401 User Manual

Chapter 1 Generating Functions with the NI 5401

Signal Name

DGND

EXT_TRIG

Signal Descriptions

Table 1-1 shows the pin names and signal descriptions used on the NI 5401

digital output connector.

Table 1-1. Digital Connector Signal Descriptions

Type

Input

Digital ground

Description

External trigger—The external trigger input signal is a

TTL-level signal that you can use to start or step through a waveform generation. For more information on trigger sources

and trigger mode, see Chapter 2, Function Generator

Operation

.

Not connected.

Reserved for future use. Do not connect signals to this pin.

NC

RFU

SHC50-68 50-Pin Cable Connector

You can use an optional SHC50-68 50-pin to 68-pin cable for external trigger input. The cable connects to the digital connector on the NI 5401.

Figure 1-5 shows the 68-pin connector pin assignments on the SHC50-68

cable.

Note The SHC50-68 connector uses the same signals as the NI 5401 digital output

connector shown in Table 1-1.

NI 5401 User Manual

1-6 www.natinst.com

Chapter 1 Generating Functions with the NI 5401

52

53

54

55

56

57

58

49

50

51

46

47

48

43

44

45

40

41

42

35

36

37

38

39

65

66

67

68

62

63

64

59

60

61

19

20

21

22

23

16

17

18

13

14

15

10

11

12

7

8

9

1

2

3

4

5

6

30

31

32

33

34

27

28

29

24

25

26

NC

NC

NC

NC

EXT_TRIG

RFU

NC

NC

NC

NC

NC

RFU

RFU

RFU

RFU

RFU

RFU

RFU

RFU

RFU

RFU

RFU

RFU

RFU

RFU

RFU

RFU

RFU

RFU

RFU

RFU

RFU

RFU

RFU

DGND

DGND

DGND

DGND

DGND

DGND

DGND

DGND

DGND

DGND

DGND

DGND

RFU

DGND

DGND

DGND

DGND

DGND

DGND

DGND

DGND

DGND

DGND

DGND

DGND

DGND

DGND

DGND

DGND

DGND

DGND

DGND

DGND

DGND

Figure 1-5. SHC50-68 68-Pin Connector Pin Assignments

© National Instruments Corporation 1-7

NI 5401 User Manual

Chapter 1 Generating Functions with the NI 5401

Software Options for Your NI 5401

This section describes the NI-FGEN driver software and development tools that you can use to create application software for your NI 5401.

Software Included with Your NI 5401

Your NI 5401 kit includes several VirtualBench soft front panels to help you get up and running quickly with your waveform generator. These soft front panels are an onscreen interface similar to standalone instruments. An

NI-FGEN instrument driver is also included, which you can use with a wide variety of development tools to build applications for your NI 5401.

These software tools are discussed in the following sections.

VirtualBench

Similar to standalone instruments, VirtualBench acquires, controls, analyzes, and presents data. However, since VirtualBench operates on your

PC, it provides additional processing, storage, and display capabilities.

VirtualBench loads and saves waveform data in a form that popular spreadsheet programs and word processors can use. It can also generate reports—a complement to the raw data storage—by adding timestamps, measurements, user names, and comments. You can print the waveforms and the settings of VirtualBench to a printer connected to the PC.

VirtualBench has two components—VirtualBench-FG and Waveform

Editor—that you can use with your NI 5401. These components are described in the following sections.

VirtualBench-FG

VirtualBench-FG transforms your PC into a fully featured function generator that rivals desktop models by using the DDS capabilities of your

NI 5401. VirtualBench-FG emulates benchtop function generators, so you can quickly learn to use computer-based instruments.

With VirtualBench-FG, you can generate a variety of waveforms, including five standard waveforms: sine, square, triangle, rising exponential, and falling exponential. Using VirtualBench-FG, you load waveforms from an

ASCII text file and generate them repeatedly. You can generate these waveforms with a resolution of approximately 10 mHz and perform frequency sweeps and shift-keying. As with all VirtualBench instruments, you can load and save instrument settings.

NI 5401 User Manual

1-8 www.natinst.com

Chapter 1 Generating Functions with the NI 5401

Waveform Editor

You use the Waveform Editor to create, sketch, and edit complex waveforms that the VirtualBench-FG player can then generate. A library of standard waveforms for creating complex waveforms is included, and you can also write equations to create arbitrary waveforms and view the waveforms in a time or frequency domain.

NI-FGEN Instrument Driver

To create your application, you need an industry-standard software driver such as NI-FGEN to control your instrument. The NI-FGEN driver includes a set of standard functions for configuring, creating, starting, and stopping waveform generation. The instrument driver reduces your program development time and simplifies instrument control by eliminating the need to learn a complex programming protocol for your instrument.

NI-FGEN is in a standard instrument driver format that works with

LabVIEW, LabWindows/CVI, and conventional programming languages such as C, C++, and Visual Basic.

Refer to the NI-FGEN readme.txt

file for more details on the NI-FGEN instrument driver. This file can be launched from the

Start»Programs»National Instruments FGEN menu.

Note An NI-FGEN Instrument Driver Quick Reference Guide is included in your NI 5401 kit. This reference guide helps you program your NI 5401.

Additional National Instruments Development Tools

The following sections describe several additional tools that you can use to develop complex applications for your NI 5401. The NI-FGEN instrument driver exposes the Application Programming Interfaces (APIs) to these development environments.

LabVIEW

LabVIEW is a graphical programming language for building instrumentation systems. With LabVIEW, you quickly create front panel user interfaces, giving you interactive control of your software system. To specify the functionality, you assemble block diagrams—a natural design notation for engineers and scientists. LabVIEW has all of the same development tools and language capabilities of a standard language such

© National Instruments Corporation 1-9

NI 5401 User Manual

Chapter 1 Generating Functions with the NI 5401 as C, including looping and case structures, configuration management tools, and compiled performance.

Note Use the NI-FGEN instrument driver to program and control your NI 5401 using

LabVIEW.

LabWindows/CVI

LabWindows/CVI is an interactive, ANSI C programming environment designed for automated test applications.

LabWindows/CVI has an interactive drag-and-drop editor for building your user interface and a complete ANSI C development environment for building your test program logic. The LabWindows/CVI environment has a wide collection of automatic code-generation tools and utilities that accelerate your development process, without sacrificing any of the power and flexibility of a language such as C. In addition, the LabWindows/CVI run-time libraries are compatible with standard C/C++ compilers, including Visual C++ and Borland C++ under Windows.

Note Use the NI-FGEN instrument driver to program and control your NI 5401 using

LabWindows/CVI.

ComponentWorks

ComponentWorks is a collection of 32-bit ActiveX controls for building virtual instrumentation systems. ComponentWorks gives you the power and flexibility of standard development tools, such as Microsoft Visual

Basic or Visual C++, with the instrumentation expertise of National

Instruments. Based on ActiveX technology, ComponentWorks controls are easy to configure using property sheets and are easy to control from your programs using high-level properties and methods. ComponentWorks features instrumentation-based graphical user interface (GUI) tools, including graphs, meters, gauges, knobs, dials, and switches.

Note Use the NI-FGEN instrument driver to program and control your NI 5401 using

ComponentWorks.

NI 5401 User Manual

1-10 www.natinst.com

Chapter 1 Generating Functions with the NI 5401

Using the Soft Front Panels to Generate Waveforms

You use the VirtualBench soft front panels to interactively control your

NI 5401 as you would a desktop function generator.

Generating Standard Functions

If you need to generate standard waveforms such as a sine, square, ramp, or DC signal, you can use the VirtualBench-FG soft front panel shown in

Figure 1-6. Launch the front panel by selecting Start»Programs»National

Instruments FGEN»VirtualBench FG. You use this front panel to control the frequency, amplitude, offset, and type of waveform generated.

The maximum sine frequency you can generate is 16 MHz. The maximum amplitude is 5 V pk

into a 50

Ω

load. If the load is a high-impedance load, the actual levels will be twice that shown on the front panel.

Figure 1-6. VirtualBench-FG Soft Front Panel for Function Generation

© National Instruments Corporation 1-11

NI 5401 User Manual

Chapter 1 Generating Functions with the NI 5401

To control additional instrument parameters, select Edit»54xx Settings to

bring up the dialog box shown in Figures 1-7 and 1-8.

Figure 1-7. VirtualBench-FG General Settings Dialog Box for the NI 5401

Figure 1-8. VirtualBench-FG Signals Settings Dialog Box for the NI 5401

Note Refer to the online help for further information about the 54xx Settings dialog box.

NI 5401 User Manual

1-12 www.natinst.com

Chapter 1 Generating Functions with the NI 5401

You can also load a custom waveform pattern with VirtualBench-FG. This waveform should be a text file and should contain exactly 16,384 samples.

If the defined waveform does not contain exactly 16,384 samples, you may see undesired effects in your waveform output. Follow these steps to load a custom waveform:

1.

Select File»Load Waveform to bring up the dialog box shown in

Figure 1-9.

Figure 1-9. VirtualBench-FG Load Waveform Dialog Box

2.

Specify the delimiter used in the text file, the number of columns, the start line, and the number of samples.

3.

Click OK to return to the main VirtualBench-FG screen shown in

Figure 1-6.

4.

Click the User button to use the information in the text file as the source for the waveform.

5.

Click the On button to generate the waveform.

Generating Multiple Frequencies in a Sequence

If desired, you can generate multiple frequencies in a sequence, which can include frequency sweeping, hopping, and so on. You can list up to 512 different frequencies and specify the duration of generation for each of

© National Instruments Corporation 1-13

NI 5401 User Manual

Chapter 1 Generating Functions with the NI 5401 them using the VirtualBench-FG Frequency List Editor. Follow these steps to create a list of multiple frequencies:

1.

Select Window»Frequency List Editor from the VirtualBench-FG

soft front panel to bring up the dialog box shown in Figure 1-10.

NI 5401 User Manual

Figure 1-10. VirtualBench-FG Frequency List Editor Dialog Box

2.

Specify the frequency and duration of each function in the sequence.

3.

Save the sequence by selecting File»Save.

4.

To return to the main VirtualBench-FG screen shown in Figure 1-6,

select File»Close.

5.

Select File»Load Frequency List to load the frequency list.

You can combine the frequency list generation with different trigger modes to get the desired frequency generation.

1-14 www.natinst.com

Chapter 1 Generating Functions with the NI 5401

Waveform Editor

You can use the Waveform Editor shown in Figure 1-11 to create a custom

waveform. To launch the Waveform Editor, select Start»Programs»

National Instruments FGEN»Waveform Editor. You can select waveforms from the function library, write equations, or draw them manually. Each segment can have more than one waveform component in it, and you can perform a variety of math functions on each component.

Figure 1-11. Waveform Editor Soft Front Panel

This soft front panel is resizable so you can view the waveforms you create with as much precision as you wish. You can save the waveforms in the following formats:

• Voltage ( .wfm

)

• Text ( .txt

)

• Binary ( .bin

)

Text waveforms are the only format you can use with the NI 5401, and they must contain exactly 16,384 samples to function properly.

© National Instruments Corporation 1-15

NI 5401 User Manual

Chapter 1 Generating Functions with the NI 5401

Power-Up and Reset Conditions

When you power up your computer, the NI 5401 is in the following state:

• The output is disabled and set to 0 V.

• The trigger mode is set to continuous.

• The trigger source is set to automatic (the software provides the triggers).

• The digital filter is enabled.

• Output attenuation remains unchanged from its previous setting.

• The analog filter remains unchanged from its previous setting.

• Output impedance remains unchanged from its previous setting.

When you reset the board using NI-FGEN or any other application software, your NI 5401 is in the same state as shown at power up, previously listed, with the following differences:

• Output attenuation is set to 0 dB.

• The analog filter is enabled.

• Output impedance is set to 50

Ω.

• The PLL reference source is set to internal tuning.

• The SYNC duty cycle is set to 50%.

NI 5401 User Manual

1-16 www.natinst.com

Function Generator Operation

This chapter describes how to use your NI 5401.

Figure 2-1 shows the NI 5401 block diagram.

RTSI/PXI Trigger Bus

Instruction

FIFOs

DDS +

Lookup

Memory

IFIFO

Control

Clock

Controls

Filter

Controls

RTSI

Control

Waveform

Sequencer

Trigger

Control

DDS

Control

Analog

Control

Data Path

Bus

Interface

PXI/PCI Channel

Digital

Filter

DAC

Attenuators,

Filter, and

Amplifier

ARB

SYNC

Level

Crossing

Detector

PLL and

Clocking

PLL Ref

Figure 2-1. NI 5401 Block Diagram

2

NI 5401 User Manual

© National Instruments Corporation 2-1

Chapter 2 Function Generator Operation

The NI 5401 has several main components:

• A PXI or PCI bus interface that handles Plug and Play protocols for assigning resources to the device and providing drivers for the data and address bus that are local to the device

• A waveform sequencer that performs multiple functions such as arbitrating the data buses and controlling the triggers, filters, attenuators, clocks, PLL, RTSI switch, instruction FIFO, and DDS

• The data from the memory is fed to a digital-to-analog converter

(DAC) through a half-band interpolating digital filter. The output from the DAC goes through the filter to the amplifiers, attenuators, and, finally, the I/O connector.

Generating Waveforms

The NI 5401 generates waveforms using DDS, which is used for generating standard waveforms that are repetitive in nature, such as sine, TTL, square, and triangular waveforms. DDS mode limits you to one buffer, and the buffer size must be exactly equal to 16,384 samples.

Figure 2-2 shows a block diagram of the data path for waveform

generation. The data for waveform generation comes from DDS lookup memory. This data is interpolated by a half-band digital filter and then fed to a high-speed DAC. The data has a pipeline delay of 26 update clocks through this digital filter. Although the digital filter can be disabled through software, there will still be a 26 update clock delay.

NI 5401 User Manual

A

12 Bits

Filter

MUX

12 12 DAC

B

DDS Lookup

Memory

Digital Filter

Enable

DDS

Div/2 80 MHz Oscillator

Figure 2-2. Waveform Data Path Block Diagram

On the NI 5401, the high-speed DAC is always updated at 80 MHz, but the update clock for memory is 40 MHz.

2-2 www.natinst.com

Chapter 2 Function Generator Operation

Direct Digital Synthesis (DDS)

Direct digital synthesis (DDS) is a technique for deriving, under digital control, an analog frequency source from a single reference clock frequency. This technique produces high-frequency accuracy and resolution, temperature stability, wideband tuning, and rapid and phase-continuous frequency switching.

The NI 5401 uses a 32-bit, high-speed accumulator with a lookup memory

and a 12-bit DAC for DDS-based waveform generation. Figure 2-3 shows

the building blocks for DDS-based waveform generation.

Frequency DDS

(14)

Lookup

Memory

Data Out (16)

Time Sequencer

Frequency

Time

Instruction FIFO

16-Bit

Counter

Div/2 80 MHz Oscillator

Figure 2-3. DDS Building Blocks

The lookup memory is dedicated to the DDS. You can store one cycle of a repetitive waveform—a sine, triangular, square, or arbitrary wave—in the lookup memory. Then, you can change the frequency of that waveform by sending just one instruction. You can use DDS mode for very fine frequency resolution function generation. You can generate sine waves of up to 16 MHz with the NI 5401. Waveform generation always loops back to the beginning of the lookup memory after passing through the end of the lookup memory.

The NI 5401 uses a lookup waveform memory for storing the waveform buffer and FIFO memory for storing the staging list, which contains multiple frequency list information. This FIFO is referred to as an

instruction FIFO.

© National Instruments Corporation 2-3

NI 5401 User Manual

Chapter 2 Function Generator Operation

Each stage is made up of two instructions: the frequency, which specifies the frequency of the waveform to be generated, and the time, which specifies the time for which the frequency is to be generated.

Note You cannot specify the number of iterations for a waveform to be generated.

Frequency Hopping and Sweeping

You can define a staging list for performing frequency hops and sweeps.

The entire staging list uses the same buffer loaded into the lookup memory.

All stages differ in the frequency to be generated.

Note The minimum time that a frequency should be generated is 2 µs. Therefore, the maximum hop rate from frequency to frequency is 500 kHz.

The maximum number of stages that can be stored in the instruction FIFO for DDS mode is 512. For more information on the waveform generation process, refer to your software documentation.

Triggering

You use triggering to start and step through a waveform generation. The trigger sources and modes of operation are explained in the following sections.

Trigger Sources

Trigger sources are software selectable. By default, the software produces the trigger sources. You can also use an external trigger from a pin on the digital I/O connector, the RTSI trigger lines on the RTSI bus, or the TTL

trigger lines on the PXI trigger bus on the backplane. Figure 2-4 shows the

trigger sources for the NI 5401.

NI 5401 User Manual

2-4 www.natinst.com

Chapter 2 Function Generator Operation

RTSI/PXI Trigger

Lines <0..6>

7

RTSI/PXI Trigger

External Trigger

Software Trigger

Digital

MUX

Start Trigger

Trigger Select

Figure 2-4. Waveform Generation Trigger Sources

If you need to automatically trigger the waveform generation, use software to generate the triggers. A rising TTL edge is required for external triggering. For more information on triggering over RTSI lines, see the

RTSI/PXI Trigger Lines section later in this chapter.

Modes of Operation

The NI 5401 has three triggering modes—single, continuous, and stepped—described in the following sections.

Single Trigger Mode

The waveform you define in the staging list is generated only once by going through the entire staging list. Only one trigger is required to start the waveform generation.

In single trigger mode, after the NI 5401 receives a trigger, the waveform generation starts at the first stage and continues through the last stage. The last stage is generated repeatedly until you stop the waveform generation.

Figure 2-5 illustrates a single trigger mode of operation.

© National Instruments Corporation 2-5

NI 5401 User Manual

Chapter 2 Function Generator Operation

Start Trigger

End of All Stages f1,

∆

T1 f2,

∆

T2 f3,

∆

T3 f4

Last Stage Generated

Continuously Until Stopped

Figure 2-5. Single Trigger Mode

For example, assume that one cycle of a sine wave is stored in the DDS lookup memory. For stage 1, f1 specifies the sine frequency to be generated for time

∆

T1, f2 and

∆

T2 for stage 2, and so on. If there are four stages in the staging list, f4 will be generated continuously until the waveform generation is stopped.

Continuous Trigger Mode

The waveform you define in the staging list is generated infinitely by continually cycling through the staging list. After a trigger is received, the waveform generation starts at the first stage, continues through the last stage, and loops back to the start of the first stage, continuing until you stop the waveform generation. Only one trigger is required to start the waveform generation.

Figure 2-6 illustrates a continuous trigger mode of operation.

Start Trigger

End of All Stages

Repeat

Until Stopped

(f1,

∆

T1

)

(f2,

∆

T2

)

(f3,

∆

T3

)

(f4,

∆

T4

)

(f1,

∆

T1

)

(f2,

∆

T2

)

Figure 2-6. Continuous Trigger Mode

NI 5401 User Manual

2-6 www.natinst.com

Chapter 2 Function Generator Operation

Stepped Trigger Mode

After a start trigger is received, the waveform defined by the first stage is generated. Then, the device waits for the next trigger signal. On the next trigger, the waveform described by the second stage is generated, and so on.

Once the staging list is exhausted, the waveform generation returns to the first stage and continues in a cyclic fashion.

Figure 2-7 illustrates a stepped trigger mode of operation. Switching from

stage to stage is phase continuous. In this mode, the time instruction is not used. The trigger paces the waveform generation from one frequency to the other.

Start Trigger Start Trigger Start Trigger Start Trigger Start Trigger f1 f2 f3 f4

End of All Stages f1

Figure 2-7. Stepped Trigger Mode

Analog Output

Analog waveforms are generated as follows:

1. The 12-bit digital waveform data is fed to a high-speed DAC.

2. A lowpass filter filters the DAC output.

3. This filtered signal is amplified before it goes to a 10 dB attenuator.

Note The DAC output can be fine-tuned for gain and offset. Since the offset is adjusted before the main attenuators and amplifier, it is referred to as pre-attenuation offset. This fine-tuning of gain and offset is performed by separate DACs.

4. The output from the 10 dB attenuator then goes to the main amplifier, which can provide up to ±5 V levels into 50

Ω

. An output relay can switch between ground level and the main amplifier. Refer to the

Output Enable section of this document for additional information

about this relay.

5. The output of this relay goes to a series of passive attenuators.

6. The output of the attenuators goes through a selectable output impedance of 50 or 75

Ω

to the I/O connector.

© National Instruments Corporation 2-7

NI 5401 User Manual

Chapter 2 Function Generator Operation

12 DAC

Figure 2-8 shows the essential block diagram of analog waveform

generation.

Lowpass

Filter

Pre Amp

10 dB

Attenuator

Main Amp

Output

Enable

Attenuators

(63 dB in 1 dB steps)

50

Ω 25

Ω

Arb

50

Ω

/75

Ω

Selector

Gain

DAC

Offset

DAC

Level

DAC

Comparator

-

+

50

Ω

SYNC

Figure 2-8. Analog Output and SYNC Out Block Diagram

Figure 2-9 shows the timing relationships of the trigger input and

waveform output. T d1

is the pulse width on the trigger signal. T d2

is the time

delay from trigger to output on Arb output. Refer to Appendix A,

Specifications , for more information on these timing parameters.

T d1

Trigger Input Signal

(Slope: Positive, TTL)

Waveform Output

(±5 V pp into 50

Ω

)

T d2

Figure 2-9. Waveform and Trigger Timings

Note You can switch off the analog lowpass filter at any time during waveform generation.

When you change this setting, the bouncing of electromechanical relays on the NI 5401 distorts the output signal for about 10 ms.

NI 5401 User Manual

2-8 www.natinst.com

Chapter 2 Function Generator Operation

SYNC Output and Duty Cycle

The SYNC output is a TTL version of the sine waveform generated at the output. The signal from the pre-amplifier is sent to a comparator, where it is compared against a level set by the level DAC. The output of this comparator is sent to the SYNC connector through a hysteresis buffer and a 50

Ω

series resistor to reverse terminate reflected pulses.

You can use the SYNC output as a very high-frequency resolution, software-programmable clock source for many applications. You also can vary the duty cycle of SYNC output, on the fly, by changing the output of the level DAC. The SYNC output might not carry meaning for other types of generated waveforms.

Note You can change the duty cycle of the SYNC output at any time during waveform generation.

Output Attenuation

Figure 2-10 shows the NI 5401 output attenuator chain. The output

attenuators are made of resistor networks and may be switched in any combination. The maximum attenuation possible on the NI 5401 is 73 dB.

1 dB 2 dB 4 dB 8 dB 16 dB 32 dB

In Out

Figure 2-10. Output Attenuation Chain

By attenuating the output signal, you keep the dynamic range of the DAC; that is, you do not lose any bits from the digital representation of the signal because the attenuation is done after the DAC and not before it.

attenuation (in decibels) = –20 log

10

(V o

/V i

)

where V o

= desired voltage level for the output signal

V i

= input voltage level.

Note For the NI 5401, V i

= ±5 V for a terminated load and ±10 V for an unterminated load.

© National Instruments Corporation 2-9

NI 5401 User Manual

Chapter 2 Function Generator Operation

NI-FGEN calculates the value of the output attenuation chain, which you can control by changing the peak-to-peak amplitude parameter. 0 dB attenuation corresponds to an amplitude of 10 V pk-pk

. The maximum attenuation of 73 dB corresponds to an amplitude of 2.24 mV pk-pk

. Any amplitude less then this is coerced to this value.

Note You can change the output attenuation at any time during waveform generation.

When you change this setting, the bouncing of electromechanical relays on the NI 5401 distorts the output signal for about 10 ms.

Output Impedance

As shown in Figure 2-10, before the signal reaches the output connector,

you can select an output impedance of 50

Ω

or 75

Ω

. If the load impedance is 50

Ω

and all the attenuators are off (an output attenuation of 0 dB), the output levels are ±5 V.

Most applications use a load impedance of 50

Ω

, but applications such as video device testers require 75

Ω

. If the load is a very high-input impedance load (~1 M

Ω

), you will see output levels up to ±10 V.

Note You can change the output impedance at any time during waveform generation.

When you change this setting, the bouncing of electromechanical relays on the NI 5401 distorts the output signal for about 10 ms.

Output Enable

You can switch off the waveform generation at the output connector by

controlling the output enable relay, as shown in Figure 2-8. When the

output enable relay is off, the output signal level goes to ground level.

Note Even though the output enable relay is in the off position, the waveform generation process continues internally on the NI 5401.

You can use this feature to disconnect and connect different devices to the

NI 5401 on the fly.

Note You can change the output enable state at any time during waveform generation.

When you change this setting, the bouncing of electromechanical relays on the NI 5401 distorts the output signal for about 10 ms.

NI 5401 User Manual

2-10 www.natinst.com

Chapter 2 Function Generator Operation

Pre-Attenuation Offset

The NI 5401 hardware supports a DC offset of up to ±2.5 V before the attenuation chain. Unless the 10 dB attenuator is switched in, which occurs when the amplitude is less then 3.16 V pk-pk

, the waveform maximum plus the offset must not exceed ±5 V into 50

Ω

. If it does, the waveform is

clipped. Refer to Figure 2-8 for a diagram showing the location of the

10 dB attenuator.

NI-FGEN automatically calculates the pre-attenuation offset value based on the DC offset and amplitude values, so the allowable DC offset range is dependent on the amplitude. For example, if you have an amplitude of

1 V pk-pk

, the maximum DC offset you can apply is 0.25 V, which corresponds to a pre-attenuation offset of 2.5 V.

Note You can change the DC Offset at any time during waveform generation. Refer to your software documentation for additional information.

Phase-Locked Loops and Board Synchronization

Figure 2-11 illustrates the block diagram for the NI 5401 for PCI PLL circuit. Figure 2-12 illustrates the block diagram for the NI 5401 for PXI

PLL circuit. The PLL consists of a voltage-controlled crystal oscillator

(VCXO) with a tuning range of ±100 ppm. This VCXO generates the main clock of 80 MHz.

The PLL can lock to a reference clock source from the external connector, from a RTSI Osc line on the RTSI bus (NI 5401 for PCI), or from a

10 MHz Osc line on the PXI backplane bus (for NI 5401 for PXI). The PLL can also be tuned internally using a calibration DAC (CalDAC). National

Instruments accurately performs this tuning during manufacturing. Refer to the

RTSI/PXI Trigger Lines

section later in this manual for additional information on using the RTSI and 10 MHz Osc lines.

The reference and VCXO clock are compared by a phase comparator running at 1 MHz. The loop filters the error signal and sends it to the control pin of the VCXO to complete the loop.

© National Instruments Corporation 2-11

NI 5401 User Manual

Chapter 2 Function Generator Operation

Board Clock (Master)

Master/Slave

RTSI Clock (Slave)

RTSI

Switch

RTSI Osc

(20 MHz)

Source

Loop

Filter

Tune

DAC

14

Phase

Comp

PLL Ref

(1 V pk-pk min)

80 MHz

Div/4

20 MHz

Board Clock

VCXO

Figure 2-11. PLL Architecture for the NI 5401 for PCI

10 MHz Osc

Source

Loop

Filter

CAL

DAC

Phase

Comp

PLL Ref

(1 V pk-pk min)

80 MHz

Div/8

10 MHz

VCXO

Figure 2-12. PLL Architecture for the NI 5401 for PXI

You can frequency lock to an external reference clock source of 1 MHz and from 5–20 MHz in 1 MHz increments. The PLL can lock to a signal level of at least 1 V pk-pk

.

Caution Do not increase the voltage level of the clock signal at the PLL reference input connector by more than the specified limit, 5 V pk-pk

.

Note If two or more NI 5401 devices are locked to each other using the same reference clock, they are frequency locked, but the phase relationship is indeterminate.

NI 5401 User Manual

2-12 www.natinst.com

Chapter 2 Function Generator Operation

Analog Filter Correction

The NI 5401 can correct for slight deviations in the flatness of the frequency characteristic of the analog lowpass filter in its passband, as

shown in Figure 2-13. Curve A shows a typical lowpass filter curve. The

response of the filter is stored in an onboard EEPROM in 1 MHz increments up to 16 MHz. Curve C is the correction applied to the frequency response. The resulting Curve B is a flat response over the entire passband. If you want to generate a sine wave at a particular frequency with filter correction applied, you have to specify that frequency through your software.

A

B

C

Frequency (MHz)

A. Typical Analog Filter Characteristics

B. Corrected Filter Characteristics

C. Correction Applied

Figure 2-13. Analog Filter Correction

Note You can change the filter frequency correction at any time during waveform generation.

© National Instruments Corporation 2-13

NI 5401 User Manual

Chapter 2 Function Generator Operation

RTSI/PXI Trigger Lines

The NI 5401 for PCI contains seven trigger lines and one RTSI clock line available over the RTSI bus to send and receive NI 5401-specific

information to other boards that have RTSI connectors. Figure 2-14 shows

the RTSI trigger lines and routing of NI 5401 for PCI signals to the RTSI switch.

SYNC

Start Trigger

RTSI Trigger

Board Clock

Master/Slave

RTSI Clock

RTSI

Switch

RTSI 0

RTSI 1

RTSI 2

RTSI 3

RTSI 4

RTSI 5

RTSI 6

RTSI Osc

Figure 2-14. RTSI Trigger Lines and Routing for the NI 5401 for PCI

Figure 2-15 shows the PXI trigger lines and routing of NI 5401 for PXI

signals to the RTSI switch.

NI 5401 User Manual

SYNC

Start Trigger

RTSI Trigger

BOARD_SYNC

RTSI

Switch

TRIG 0

TRIG 1

TRIG 2

TRIG 3

TRIG 4

TRIG 5

PXI STAR

PXI 10MHz Osc

Figure 2-15. PXI Trigger Lines, 10 MHz Backplane Oscillator, and Routing for the

NI 5401 for PXI

2-14 www.natinst.com

Chapter 2 Function Generator Operation

The NI 5401 can receive a hardware trigger from another board as a RTSI trigger signal on any of the RTSI/PXI trigger lines.

You can also route signals as follows:

• Route the Start Trigger signal generated on the NI 5401 to other boards through any of the RTSI/PXI bus trigger lines.

• Route the SYNC output generated on the NI 5401 to other boards through any of the RTSI/PXI bus trigger lines.You can use this signal to give other boards an accurate and fine frequency resolution clock.

NI 5401 for PCI

For frequency locking to other boards as a master, the NI 5401 sends an onboard 20 MHz signal to the RTSI Osc line as a board clock signal. For locking to other devices as a slave, the NI 5401 receives the RTSI Osc line as a RTSI clock signal.

NI 5401 for PXI

For frequency locking to other boards, the NI 5401 for PXI receives the

PXI backplane 10 MHz Osc as a reference clock signal. All the NI 5401s for PXI use this common signal as the reference clock for frequency locking.

Note Refer to your software documentation for selecting and routing signals to the

RTSI/PXI trigger bus.

Calibration

Calibration is the process of minimizing measurement errors by making small circuit adjustments. On the NI 5401, NI-FGEN automatically makes these adjustments by retrieving predetermined constants from the onboard

EEPROM, calculating correction values, and writing those values to the

CalDACs.

National Instruments calibrates all NI 5401 devices to the levels indicated

in Appendix A, Specifications . Factory calibration involves procedures

such as nulling the offset and gain errors. However, since offset and gain errors may drift with time and temperature, you may need to recalibrate your device. Contact National Instruments to recalibrate your NI 5401.

© National Instruments Corporation 2-15

NI 5401 User Manual

A

Specifications

This appendix lists the specifications for the NI 5401. These specifications are typical at 25 °C unless otherwise stated. The operating temperature range is 0 to 50 °C.

Analog Output

Number of channels ............................... 1

Resolution .............................................. 12 bits

Maximum update rate ............................ 40 MHz

DDS accumulator ................................... 32 bits

Frequency range

Sine ................................................. 16 MHz, max

SYNC (TTL)................................... 16 MHz, max

Square ............................................. 1 MHz, max

Ramp ............................................... 1 MHz, max

Triangle ........................................... 1 MHz, max

Frequency resolution.............................. 9.31 mHz

Voltage Output

Ranges

..................................................... ±

5 V into a 50

Ω

load

±

10 V into a high-impedance load

Accuracy

................................................ ±

0.1 dB

Output attenuation.................................. 0 to 73 dB

Resolution ....................................... 0.001 dB steps

NI 5401 User Manual

© National Instruments Corporation A-1

Appendix A Specifications

Pre-attenuation offset

Range...............................................±2.5 V into 50

Ω

1

Accuracy..........................................±5 mV

Output coupling .....................................DC

Output impedance ..................................50

Ω or

75 Ω

software selectable

Load impedance .....................................50

Ω

or greater

Output enable..........................................Software switchable

Protection................................................Short-circuit protected

Sine Spectral Purity

Harmonic products and spurs

Up to 1 MHz....................................–60 dBc

Up to 16 MHz..................................–35 dBc

Phase noise .............................................–105 dBc/Hz at 10 kHz from carrier

Filter Characteristics

Digital

Type .................................................Half-band interpolating

Selection ..........................................Software switchable (enable or disable)

Taps ................................................67

Filter coefficients ............................Fixed 20-bit

Data interpolating frequency ..........80 MS/s

Pipeline signal delay .......................26 sampling periods

Analog

Type .................................................7th-order L-C lowpass filter

Passband ripple ................................±2 dB

1 With less than 10 dB of attenuation, signal maximum plus offset (before attenuation) must not exceed ±5 V (into 50

Ω)

NI 5401 User Manual

A-2 www.natinst.com

Appendix A Specifications

Waveform Specifications

Memory .................................................. 16,384 16-bit samples

Segment length....................................... 16,384 samples, exact

Segment linking (instruction FIFO) ....... 512 links

Timing I/O

Update clock .......................................... Internal, 40 MHz only

Frequency locking

External reference sources .............. Input connector, RTSI clock line, or internal

Reference clock frequencies ........... 1 MHz, 5–20 MHz in 1 MHz steps

Frequency locking range................. ±100 ppm

Triggers

Digital Trigger

Compatibility ........................................ TTL

Response ............................................... Rising edge

Pulse width (T d1

) .................................... 20 ns min

Trigger to waveform output delay (T d2

) ................................... 28 sample clocks + 150 ns max

RTSI

Trigger lines .......................................... 7

Clock lines.............................................. 1

Bus Interface

Type ....................................................... Slave

Operational Modes

Type ....................................................... Single, continuous, stepped

© National Instruments Corporation A-3

NI 5401 User Manual

Appendix A Specifications

SYNC Out

External Clock Reference Input

Frequency ...............................................1 MHz or 5–20 MHz in 1 MHz steps

Amplitude ...............................................1 V pk-pk

level

5 V pk-pk

Internal Clock

Frequency ...............................................40 MHz

Initial accuracy .......................................±5 ppm

Temperature stability (0 to 5 °C) ............±25 ppm

Aging (1 year).........................................±5 ppm

Mechanical

Level ......................................................TTL

Duty cycle...............................................20% to 80%, software controllable

Connectors

ARB (output) ...................................SMB/BNC

SYNC (output).................................SMB/BNC

PLL reference (input) ......................SMB

External trigger in............................50-pin digital (PCI),

SMB (PXI)

Size .........................................................1 slot

Power requirements ...............................5 V, 3.5 A max

12 V, 125 mA

NI 5401 User Manual

A-4 www.natinst.com

B

Optional Accessories

National Instruments offers a variety of products to use with your NI 5401, including probes, cables, and other accessories:

• Shielded and unshielded I/O connector blocks (SCB-68, TBX-68,

CB-68)

• RTSI bus cables

For more specific information about these products, refer to your National

Instruments catalogue or Web site, or call the office nearest you.

Cabling

The following list gives recommended part numbers for cables that you can use with your NI 5401 device:

• BNC male to BNC male, 50

Ω

cable from ITT Pomona Electronics

(part number BNC-C-xx)

• BNC male to BNC male, 75

Ω

cable from ITT Pomona Electronics

(part number 2249-E-xx)

• BNC female to RCA phono plug adapter, from ITT Pomona

Electronics (part number 5319)

• BNC 50

Ω

feed-through terminator adapter from ITT Pomona

Electronics (part number 4119-50)

• BNC female-female adapter from ITT Pomona Electronics

(part number 3283)

© National Instruments Corporation B-1

NI 5401 User Manual

C

Frequency Resolution and

Lookup Memory

For DDS-based waveform generation, you must first load one cycle of the desired waveform into the lookup memory. The size of the DDS lookup memory is 16,384 samples. Each sample is 16 bits wide.

Note One cycle of the waveform buffer loaded into the memory should be exactly equal to the size of the DDS lookup memory.

F c

= update clock for the accumulator.

For the NI 5401, F c

= 40 MHz.

F a

= desired frequency of the output signal

N = accumulator size in bits

For the NI 5401, N = 32.

FCW = frequency control word to be loaded into the accumulator to generate F a

.

The frequency control word is calculated using the formula:

FCW = (2 N * F a

) / F c

The frequency resolution is then given by: frequency resolution = F c

/ 2 N = (40

×

10 6 ) / 2 32 = 9.31322 mHz

For example, if you need to generate a frequency of 10 MHz, then the FCW is (2 32 * 10E6)/40E6, which equals 1,073,741,824. If you need to generate a frequency of 1 Hz, then the FCW is (2 32 * 1)/40E6, which equals 107.

Note On the NI 5401, the maximum frequency of a sine wave you can generate reliably is limited to 16 MHz. Other waveforms such as square or triangular waves are limited to

1 MHz.

© National Instruments Corporation C-1

NI 5401 User Manual

Appendix C Frequency Resolution and Lookup Memory

You can also synthesize arbitrary waveforms using DDS. Generating arbitrary waveforms this way will be very limited; you are restricted to a single buffer, and this buffer should be exactly equal to the size of the lookup memory (16,384 samples).

To update every sample of an arbitrary waveform in lookup memory at the maximum clock rate of 40 MHz, the software writes an FCW value of

2 (N–L) , where N is the size of the accumulator and L is the number of address bits of lookup memory (L = 14 bits). Thus, the FCW value for the

NI 5401 equals 262,144. Since FCW = (2 N * F a so you would write a frequency value of (2

) / F

(32–14) c

, F a

= (2 (N–L) * F c

) / 2 N ,

× (

40

×

10 6 )) / 2 32 , which equals 2.441 kHz

If you want to update every sample in lookup memory at an integral subdivision, D, of the maximum clock rate, then you want an FCW value of 2 (N–L–D+1) . In other words, for an effective update rate of every sample at half the maximum clock rate, write a frequency value of

(2 (32–14–2+1)

× (

40

×

10 6 )) / 2 32 , which equals 1.221 kHz.

NI 5401 User Manual

C-2 www.natinst.com

D

Technical Support Resources

This appendix describes the comprehensive resources available to you in the Technical Support section of the National Instruments Web site and provides technical support telephone numbers for you to use if you have trouble connecting to our Web site or if you do not have internet access.

NI Web Support

To provide you with immediate answers and solutions 24 hours a day,

365 days a year, National Instruments maintains extensive online technical support resources. They are available to you at no cost, are updated daily, and can be found in the Technical Support section of our Web site at www.natinst.com/support .

On-Line Problem-Solving and Diagnostic Resources

KnowledgeBase—A searchable database containing thousands of frequently asked questions (FAQs) and their corresponding answers or solutions, including special sections devoted to our newest products.

The database is updated daily in response to new customer experiences and feedback.

Troubleshooting Wizards—Step-by-step guides lead you through common problems and answer questions about our entire product line.

Wizards include screen shots that illustrate the steps being described and provide detailed information ranging from simple getting started instructions to advanced topics.

Product Manuals—A comprehensive, searchable library of the latest editions of National Instruments hardware and software product manuals.

Hardware Reference Database—A searchable database containing brief hardware descriptions, mechanical drawings, and helpful images of jumper settings and connector pinouts.

Application Notes—A library with more than 100 short papers addressing specific topics such as creating and calling DLLs, developing your own instrument driver software, and porting applications between platforms and operating systems.

© National Instruments Corporation D-1

NI 5401 User Manual

Appendix D Technical Support Resources

Software-Related Resources

Instrument Driver Network—A library with hundreds of instrument drivers for control of standalone instruments via GPIB, VXI, or serial interfaces. You also can submit a request for a particular instrument driver if it does not already appear in the library.

Example Programs Database—A database with numerous, non-shipping example programs for National Instruments programming environments. You can use them to complement the example programs that are already included with National Instruments products.

Software Library—A library with updates and patches to application software, links to the latest versions of driver software for National

Instruments hardware products, and utility routines.

Worldwide Support

National Instruments has offices located around the globe. Many branch offices maintain a Web site to provide information on local services. You can access these Web sites from www.natinst.com/worldwide .

If you have trouble connecting to our Web site, please contact your local

National Instruments office or the source from which you purchased your

National Instruments product(s) to obtain support.

For telephone support in the United States, dial 512 795 8248. For telephone support outside the United States, contact your local branch office:

Australia 03 9879 5166, Austria 0662 45 79 90 0, Belgium 02 757 00 20,

Brazil 011 284 5011, Canada (Ontario) 905 785 0085,

Canada (Québec) 514 694 8521, China 0755 3904939,

Denmark 45 76 26 00, Finland 09 725 725 11, France 01 48 14 24 24,

Germany 089 741 31 30, Hong Kong 2645 3186, India 91805275406,

Israel 03 6120092, Italy 02 413091, Japan 03 5472 2970,

Korea 02 596 7456, Mexico (D.F.) 5 280 7625,

Mexico (Monterrey) 8 357 7695, Netherlands 0348 433466,

Norway 32 27 73 00, Singapore 2265886, Spain (Madrid) 91 640 0085,

Spain (Barcelona) 93 582 0251, Sweden 08 587 895 00,

Switzerland 056 200 51 51, Taiwan 02 2377 1200,

United Kingdom 01635 523545

NI 5401 User Manual

D-2 www.natinst.com

Glossary

Prefix

µmk-

M-

Meaning micromillikilomega-

Value

10 – 6

10 –3

10 3

10 6

Numbers/Symbols

±

%

+

/

°

Ω

+5 V percent positive of, or plus negative of, or minus plus or minus per degree ohm

+5 V output signal

A

A amplification

ARB attenuation amperes method of scaling the signal level to a higher level normal waveform output signal decreasing the amplitude of a signal

© National Instruments Corporation G-1

NI 5401 User Manual

Glossary

B

b

B

BNC buffer bus

D

DAC dB dBc bit—one binary digit, either 0 or 1 byte—eight related bits of data, an eight-bit binary number. Also used to denote the amount of memory required to store one byte of data.

a type of coaxial signal connector temporary storage for acquired or generated data the group of conductors that interconnect individual circuitry in a computer.

Typically, a bus is the expansion vehicle to which I/O or other devices are connected. Examples of PC buses are the AT bus (also known as the ISA bus) and the PCI bus.

C

C

CalDAC clock

Celsius calibration DAC hardware component that controls timing for reading from or writing to groups

CMOS complementary metal-oxide semiconductor continuous trigger mode repeats a staging list until waveform generation is stopped counter coupling a circuit that counts external pulses or clock pulses (timing) the manner in which a signal is connected from one location to another digital-to-analog converter—an electronic device, often an integrated circuit, that converts a digital number into a corresponding analog voltage or current decibel—the unit for expressing a logarithmic measure of the ratio of two signal levels: dB = 20 log

10

V1/V2, for signals in volts decibel referred to carrier level

NI 5401 User Manual

G-2 www.natinst.com

DC

DC coupled

DDS

DDS mode

DGND digital word driver dynamic range

Glossary direct current allowing the transmission of both AC and DC signals direct digital synthesis—a digital technique of frequency generation using a numerically controlled oscillator (NCO), a dedicated lookup memory, and a DAC a method of waveform generation that uses built-in DDS functionality to generate very high frequency resolution standard waveforms digital ground signal

See word.

software that controls a specific hardware device the ratio of the largest signal level a circuit can handle to the smallest signal level it can handle (usually taken to be the noise level), normally expressed in dB

E

EEPROM external trigger

EXT_TRIG electrically erasable programmable read-only memory—ROM that can be erased with an electrical signal and reprogrammed a voltage pulse from an external source that triggers an event such as A/D conversion external trigger input signal

© National Instruments Corporation G-3

NI 5401 User Manual

Glossary

F

FIFO filters frequency hop frequency resolution frequency sweep

G

gain

GUI

H

hardware

HiZ

Hz first-in first-out memory buffer—the first data stored is the first data sent to the acceptor. FIFOs are often used on DAQ devices to temporarily store incoming or outgoing data until that data can be retrieved or output. For example, an analog input FIFO stores the results of A/D conversions until the data can be retrieved into system memory, a process that requires the servicing of interrupts and often the programming of the DMA controller.

This process can take several milliseconds in some cases. During this time, data accumulates in the FIFO for future retrieval. With a larger FIFO, longer latencies can be tolerated. In the case of analog output, a FIFO permits faster update rates, because the waveform data can be stored on the

FIFO ahead of time. This again reduces the effect of latencies associated with getting the data from system memory to the DAQ device.

digital or analog circuits that change the frequency characteristics of a waveform change from one frequency to another the smallest frequency change that can be generated by a NI 5411/5431 change the frequency of a waveform in a controlled manner the factor by which a signal is amplified, sometimes expressed in decibels graphical user interface the physical components of a computer system, such as the circuit boards, plug-in boards, chassis, enclosures, peripherals, cables, and so on high impedance hertz—the number of cycles or repetitions per second

NI 5401 User Manual

G-4 www.natinst.com

Glossary

I

I/O k

K

IFIFO instruction FIFO

K kS

Kword

L

latch level DAC linking looping lowpass filter input/output—the transfer of data to/from a computer system involving communications channels, operator interface devices, and/or data acquisition and control interfaces instruction FIFO the FIFO that stores the waveform generation staging list kilo—the standard metric prefix for 1,000, or 10 3 , used with units of measure such as volts, hertz, and meters kilo—the prefix for 1,024, or 2 10 , used with B in quantifying data or computer memory

1,000 samples

1,024 words of memory a digital device that stores digital data based on a control signal the calibration DAC used to change the voltage levels to another device linking different buffers stored in the waveform memory repeating the same buffer in the waveform memory. This method of waveform generation decreases memory requirements.

a circuit used to smooth the waveform output and removed unwanted high frequency contents form the signal

© National Instruments Corporation G-5

NI 5401 User Manual

Glossary

M

m

M master/slave

MB

N

noise meters

(1) Mega, the standard metric prefix for 1 million or 10 6 , when used with units of measure such as volts and hertz; (2) mega, the prefix for 1,048,576, or 2 20 , when used with B to quantify data or computer memory locking the NI 5401 clock in frequency to an external phase locking reference clock source megabytes of memory an undesirable electrical signal—Noise comes from external sources such as the AC power line, motors, generators, transformers, fluorescent lights, soldering irons, CRT displays, computers, electrical storms, welders, radio transmitters, and internal sources such as semiconductors, resistors, and capacitors. Noise corrupts signals you are trying to send or receive.

O

output enable relay a relay switch at the output of the NI 5401 that can enable the waveform generation at any time or that can connect the output to ground

P

passband

PCI

PCLK peak-to-peak the range of frequencies which a device can properly propagate or measure

Peripheral Component Interconnect—a high-performance expansion bus architecture originally developed by Intel to replace ISA and EISA. It is achieving widespread acceptance as a standard for PCs and work-stations; it offers a theoretical maximum transfer rate of 132 Mbytes/s.

digital pattern clock output a measure of signal amplitude; the difference between the highest and lowest excursions of the signal

NI 5401 User Manual

G-6 www.natinst.com

Glossary pipeline

PLL

PLL Ref

Plug and Play devices ppm pre-attenuation offset protocol

PXI a high-performance processor structure in which the completion of an instruction is broken into its elements so that several elements can be processed simultaneously from different instructions phase-locked loop—a circuit that synthesizes a signal whose frequency is exactly proportional to the frequency of a reference signal a PLL input that accepts an external reference clock signal and phase locks to it the NI 5401 internal clock devices that do not require dip switches or jumpers to configure resources on the devices—also called switchless devices parts per million an offset provided to the signal before it reaches the attenuators the exact sequence of bits, characters, and control codes used to transfer data between computers and peripherals through a communications channel, such as the GPIB bus

PCI eXtensions for Instrumentation

R

resolution

RTSI bus the smallest signal increment that can be detected by a measurement system. Resolution can be expressed in bits, in proportions, or in percent of full scale. For example, a system has 12-bit resolution, one part in 4,096 resolution, and 0.0244 percent of full scale.

Real-Time System Integration bus—the National Instruments timing bus that connects DAQ boards directly, by means of connectors on top of the boards, for precise synchronization of functions

S

s

S sampling rate seconds samples the rate, in samples per second (S/s), at which each sample in the waveform buffer is updated

© National Instruments Corporation G-7

NI 5401 User Manual

Glossary

SCSI sequence list shift-keying single trigger mode

SMB

S/s stage staging list stepped trigger mode

SYNC

Small Computer System Interface (bus)

See staging list.

frequency shift keying (FSK) when the arbitrary waveform generator goes through the staging list only once

Sub Miniature Type B connector that features a snap coupling for fast connection samples per second—used to express the rate at which a DAQ board samples an analog signal in Arb mode, specifies the buffer to be generated, the number of loops on that buffer, the marker position for that buffer, and the sample count for the buffer; for DDS mode, specifies the frequency to be generated of the waveform in the lookup memory and the time for which that frequency has to be generated a buffer that contains linking and looping information for multiple waveforms; also known as a sequence list or waveform sequence a mode of waveform generation used when you want a trigger to advance the waveforms specified by the stages in the staging list

TTL version of the sine waveform output signal generated by the NI 5401

T

trigger

TTL

U

update rate any event that causes or starts some form of data capture transistor-transistor logic the rate at which a DAC is updated

NI 5401 User Manual

G-8 www.natinst.com

Glossary

V

V

VCXO

VHDSCSI volts voltage controlled crystal oscillator very high-density SCSI

W

waveform waveform buffer waveform linking and looping waveform memory waveform segment waveform sequence waveform staging word multiple voltage readings taken at a specific sampling rate the collection of 16-bit data samples stored in the waveform memory that represent a desired waveform. Also known as a waveform segment.

See linking, looping.

physical data storage on the NI 5401 for storing the waveform data samples

See waveform buffer.

See staging list.

See linking, looping.

The standard number of bits that a processor or memory manipulates at one time. Microprocessors typically use 8-, 16-, or 32-bit words.

© National Instruments Corporation G-9

NI 5401 User Manual

Index

A

analog filter correction, 2-13 analog output, 2-7 to 2-11 analog output and SYNC out block diagram, 2-8 output attenuation, 2-9 to 2-11 output enable, 2-10 output impedance, 2-10 pre-attenuation offset, 2-11 specifications, A-1

SYNC output and duty cycle, 2-9 waveform and trigger timings (figure), 2-8

ARB connector, 1-3 attenuation of output, 2-9 to 2-11

B

block diagram for NI 5401, 2-1 bus interface specifications, A-3

C

cables, optional, B-1 calibration, 2-15 clocks external clock reference input, A-4 internal clock, A-4

ComponentWorks software, 1-10 connectors. See I/O connectors.

continuous trigger mode, 2-6 conventions used in manual, iv

D

DDS. See direct digital synthesis (DDS).

DGND signal (table), 1-6 diagnostic resources, online, D-1 digital trigger specifications, A-3 direct digital synthesis (DDS) building blocks for DDS (figure), 2-3 description, 2-3 to 2-4 frequency hopping and sweeping, 2-4 frequency resolution and lookup memory,

C-1 to C-2 duty cycle of SYNC output, 2-9

E

external clock reference input, A-4

EXT_TRIG signal (table), 1-6

F

filters analog filter correction, 2-13 characteristics, A-2 frequency direct digital synthesis, 2-4 frequency hops and sweeps, 2-4 frequency resolution and lookup memory,

C-1 to C-2 generating multiple frequencies in sequence, 1-13 to 1-14 function generator operation analog filter correction, 2-13 analog output, 2-7 to 2-11 output attenuation, 2-9 to 2-11 output enable, 2-10 output impedance, 2-10 pre-attenuation offset, 2-11

SYNC output and duty cycle, 2-9 block diagram, 2-1 calibration, 2-15 direct digital synthesis (DDS), 2-3 to 2-4 frequency hopping and sweeping, 2-4

© National Instruments Corporation I-1

NI 5401 User Manual

Index phase-locked loops and board synchronization, 2-11 to 2-12

RTSI/PXI trigger lines, 2-14 to 2-15 triggering, 2-4 to 2-7 continuous trigger mode, 2-6 single trigger mode, 2-5 to 2-6 stepped trigger mode, 2-7 trigger sources, 2-4 to 2-5 waveform generation, 2-2

I

impedance, output, 2-10 instruction FIFO, 2-3 internal clock specifications, A-4

I/O connectors, 1-2 to 1-7

ARB connector, 1-3

I/O connectors on front panel (figure), 1-2

Pattern Out connector, 1-5 to 1-6 pin assignments (figure), 1-5 signal descriptions (table), 1-6

PLL Ref connector, 1-4

SHC50-68 50-pin cable connector,

1-6 to 1-7 pin assignments (figure), 1-7 signals, 1-6

SYNC connector, 1-3 to 1-4

I/O connectors (figure), 1-2

L

LabVIEW and LabWindows/CVI software,

1-9 to 1-10 lookup memory description, 2-3 frequency resolution and lookup memory,

C-1 to C-2 lookup waveform memory, 2-3

M

mechanical specifications, A-4 multiple frequencies, generating in sequence,

1-13 to 1-14

N

National Instruments Web support, D-1 to D-2

NI 5401. See also function generator operation.

block diagram, 2-1 components, 2-2 connecting signals, 1-2 to 1-7

ARB connector, 1-3

I/O connectors (figure), 1-2

Pattern Out connector, 1-5 to 1-6

PLL Ref connector, 1-4

SHC50-68 50-pin cable connector,

1-6 to 1-7

SYNC connector, 1-3 to 1-4 features, 1-1

PLL architecture (figures), 2-12 power-up and reset conditions, 1-16 software options, 1-8 to 1-10

NI-FGEN instrument driver, 1-9

O

online problem-solving and diagnostic resources, D-1 operational mode specifications, A-3 optional accessories, B-1 output attenuation, 2-9 to 2-11

P

Pattern Out connector, 1-5 to 1-6 pin assignments (figure), 1-5 signal descriptions (table), 1-6 phase-locked loop (PLL) Ref connector, 1-4

NI 5401 User Manual

I-2 www.natinst.com

phase-locked loops and board synchronization, 2-11 to 2-12 pin assignments (figure)

Pattern Out connector, 1-5

SHC50-68 50-pin cable connector, 1-7

PLL Ref connector, 1-4 power-up and reset conditions, 1-16 pre-attenuation offset, 2-11 problem-solving and diagnostic resources, online, D-1

R

reset conditions, 1-16

RTSI trigger specifications, A-3

RTSI/PXI trigger lines, 2-14 to 2-15

S

SHC50-68 50-pin cable connector, 1-6 to 1-7 pin assignments (figure), 1-7 signals, 1-6 sine spectral purity, A-2 single trigger mode, 2-5 to 2-6 soft front panels. See VirtualBench software.

software options, 1-8 to 1-10

ComponentWorks, 1-10

LabVIEW, 1-9 to 1-10

LabWindows/CVI, 1-10

NI-FGEN instrument driver, 1-9

VirtualBench, 1-8

VirtualBench-FG, 1-8

Waveform Editor, 1-9 software-related resources, D-2 specifications, A-1 to A-4 analog output, A-1 bus interface, A-3 external clock reference input, A-4 filter characteristics, A-2 internal clock, A-4 mechanical, A-4

© National Instruments Corporation I-3

Index operational modes, A-3 sine spectral purity, A-2

SYNC out, A-4 timing I/O, A-3 triggers, A-3 voltage output, A-1 to A-2 waveform, A-3 staging list, 2-3 stepped trigger mode, 2-7

SYNC connector, 1-3 to 1-4

SYNC output description, 2-9 duty cycle, 2-9 specifications, A-4

T

technical support resources, D-1 to D-2 time, in direct digital synthesis, 2-4 timing I/O specifications, A-3 trigger specifications, A-3 digital trigger, A-3

RTSI, A-3 triggering, 2-4 to 2-7 continuous trigger mode, 2-6 single trigger mode, 2-5 to 2-6 stepped trigger mode, 2-7 trigger sources, 2-4 to 2-5

V

VirtualBench software description, 1-8 generating multiple frequencies in sequence, 1-13 to 1-14 generating standard functions, 1-11 to

1-13

VirtualBench-FG software description, 1-8

Frequency List Editor, 1-14

General Settings dialog box (figure), 1-12

NI 5401 User Manual

Index

Load Waveform dialog box (figure), 1-13

Signal Settings dialog box (figure), 1-12 voltage output specifications, A-1 to A-2

W

Waveform Editor creating custom waveforms, 1-15 description, 1-9 waveform generation. See also function generator operation.

analog waveforms, 2-7 to 2-11 data path block diagram (figure), 2-2 overview, 2-2 using soft front panels generating multiple frequencies in sequence, 1-13 to 1-14 generating standard functions,

1-11 to 1-13 using Waveform Editor, 1-15 waveform specifications, A-3

Web support from National Instruments,

D-1 to D-2 online problem-solving and diagnostic resources, D-1 software-related resources, D-2

Worldwide technical support, D-2

NI 5401 User Manual

I-4 www.natinst.com