thermokit

89119-2

89119-3

89119-1

www.minilandbaby.com

English

INDEX

1. INTRODUCTION

2. CONTENTS

3. BODY THERMOMETER

4. PACIFIER THERMOMETER

5. BATH THERMOMETER

6. BATTERY & PRODUCT DISPOSAL INFORMATION

1. INTRODUCTION

Congratulations on the purchase of this kit of thermometers that includes a body thermometer, a bath thermometer and a pacifier thermometer.

Before using this kit, please read this manual carefully.

Please keep this manual as it contains important information.

The features described in this user’s guide are subject to modifications without prior notice.

Only changes and modifications carried by the MINILAND technical service are covered by the product guarantee.

2. CONTENTS

1 Body Thermometer

1 Bath Thermometer

1 Pacifier Thermometer

1 Case for storage and transport

1 Instruction manual and guarantee card

If any of the above is missing, please contact your retailer.

3. BODY THERMOMETER

3.1. SAFETY INSTRUCTIONS

1. Swallowing a battery may be fatal.

2. Do not charge the battery.

3. Do not dispose the battery in fire as it may explode.

4. Do not drop the unit. It is not shock proof.

5. Do not attempt to sterilize with heat or by boiling.

6. Do not store the unit in direct sunlight or at high temperature.

7. Do not bend or bite or pull the unit.

8. Do not attempt to disassemble the unit except to replace the battery.

9. The measurement reading maybe incorrect when the body temperature is lower than ambient temperature.

Battery Warning: Use only the batteries indicated.

English

3.2. PRODUCT FEATURES

3.2.1. PARTS DESCRIPTION

1. Probe

2. Probe cover

3. Display LCD

4. ON / OFF Button

5. Battery cover

3.3. INSTRUCTIONS FOR USE

1. Ensure that the thermometer has been turned off before taking the temperature.

Press and release the power button (4). A short beep will be heard and all the symbols will be turned on for 1 second to show that all parts of the LCD are functioning properly,

188.8

e

M

then the last measurement will be shown on the display for 2 seconds.

32.3

e

M

When the flashing ºC or ºF symbol appears take the body temperature.

32.3

C

Note: You may find it easier to start measuring your child’s body temperature

before turning the thermometer on as it is easy to miss the flashing ºC or ºF symbol. For example: place the probe into your child’s mouth then turn the thermometer on.

2. During the measurement the LCD will display ”Lo °C” or “Lo °F” if the measured temperature is below 32.0°C (89.6°F), or will display ”Hi °C” or “Hi °F” if the measured temperature is above 43.0°C /109.4°F

3. When the probe tip reaches its thermal equilibrium or peak, the ºC (or ºF) symbol stops flashing and the beeper will sound 10 times. The temperature is now ready to be read and will not change when the thermometer is removed.

4. FEVER: 3 fast beeps repeated 10 times. If the measured temperature is higher then 37.8ºC (100ºF), the fever warning function will emit 3 fast beeps 10 times .

5. After measuring, turn off the unit by pressing the power button. If the power is left on, the thermometer will switch off after approximately 9 minutes to save power.

6. Press and release power button to take another reading.

English

3.3.1. METHODS OF USE

When measuring the body temperature, always take the measurement from the same part of the body and at the same time of day.

Whichever method is used, it is recommended to avoid eating or drinking any liquids, exercising, taking showers or baths, or smoking for 15 minutes prior to use.

3.3.1.1. MOUTH (ORAL USE)

1. Cover the thermometer’s tip with the probe cover.

2. Place the probe tip under the tongue as far in as possible and switch the thermometer on. The mouth should remain closed during the measurement to ensure an accurate reading. Be careful not to bite the thermometer.

3. When the peak temperature is reached (15~25 seconds), it will beep 10 times and the “ºC” ( or “ºF” ) sign will stop flashing. The temperature is now ready to read and will not change when the thermometer is removed.

3.3.1.2. ARMPIT (AXILLARY USE)

This method can be used for babies and young children but is less accurate and takes longer. NOTE: This method produces a temperature 1ºC lower than a temperature taken orally. To get a reading temperature closer to an oral temperature, you must take the temperature over a 10-minute period.

1. Wipe the armpit with a towel.

2. Cover the tip of the thermometer with the probe cover. Place the probe tip firmly under the armpit, then bring your arm to your side so your elbow is firmly pressed against the side of your body and switch the thermometer on.

3. When the peak temperature is reached (usually 30 seconds but sometimes longer for this method), it will beep 10 times and the “ºC” (or “ºF”) sign will stop flashing. The temperature can now be read and will not change when the thermometer is removed.

3.3.1.3. ANUS (RECTAL USE)

Commonly used for babies, young children, or when it is difficult to take an oral

(mouth) or axillary (armpit) temperature.

1. Cover the thermometer’s tip with a probe cover and lubricate with a child-safe water-soluble lubricant e.g. KY for easier insertion. Do not use a petroleum jelly.

2. Gently insert the tip of the thermometer NO MORE THAN 1/2 inch into the rectum. If you detect resistance, STOP. Hold the thermometer in place during the measurement.

3. When the peak temperature is reached (usually 10~15 seconds), the thermometer will beep 4 times and the “ºC” (or “ºF”) sign will stop flashing. The temperature is now ready to read and will not change when the thermometer is removed.

Remember that this method produces a temperature about 0.5~1ºC higher than a temperature taken orally.

English

3.4. MAINTENANCE

3.4.1. CLEANING

1. Clean the thermometer probe before and after use by wiping it with medical alcohol 95% and then rinse with plenty of water.

2. When taking continuous measurements, we recommend using a probe cover, which should not to be reused.

3.4.2. REPLACING THE BATTERY

When the battery flag “ “ appears and flashes, the battery needs to be replaced.

1. Remove the battery cover.

2. Remove the battery with a non-metallic object.

3. Replace with an alkaline LR41.

4. Make sure the positive (+) side is facing up.

5. Close the cover securely.

6. Carefully discard the used battery and keep it out of children’s reach.

3.5. TECHNICAL SPECIFICATION

- Measurement range: 32ºC~42.9ºC / 89.6ºF~109.2ºF

- Measuring time: 10 seconds

Accuracy ºC:

Accuracy Measurement Range

±0.1 ºC

±0.2 ºC from 35.0 ºC to 42.0 ºC remaining range at an ambient of 22ºC

-1 ºC Armpit (axillary use)

- need to add 1 ºC to reading

Accuracy ºF:

Accuracy Measurement Range

±0.2 ºF

±0.3 ºF

-0.5 ºF from 95.0ºF to 107.6ºF remaining range at an ambient of 71.6ºF

Armpit (axillary use) - need to add 0.5 ºF to reading

- Display unit ºC or ºF (factory setting only)

- Power: 1 LR41 button battery, 1.5V

- Battery life approx. 250 hours of continuous use

- Shelf life: 3 years

- Storage temperature: -10ºC to 40ºC (14ºF to 104ºF). Humidity: 30% to 90%

- Operation temperature: 10ºC to 35ºC (50ºF to 95ºF). Humidity 40% to 80%

English

3.5.1. NORMALIZED SYMBOLS

· Attention: Read the operating Instructions for use!

· Applied part of type BF

· Disposal in accordance with Directive 2002/96/EC (WEEE)

0434

· Complies with the European Medical Device Directive (93/42/EEC) and

amended by directive 2007/47/ECrequirements.

Notified body 0434: Det Norske Veritas Certification AS

· Batch code

· Serial number

· The name and the address of the manufacturer.

· Authorized Representative in the European Community

4. PACIFIER THERMOMETER

4.1. SAFETY INSTRUCTIONS

1. Suitable for measuring childrens’ body temperatures.

2. It is designed and manufactured complying with relevant requirements in

EN12470-3: Performance of Compact electrical thermometers (non-predictive and predictive) with maximum device.

3. It has passed EMC emission test, immunity test according to EN60601-1-2 and biocompatibility test according to MDD annex 1.

4. According to the degree of protection against electric shock, it is classified to BF, in accordance with EN60601-1.

5. Avoid dropping the device or subjecting it to heavy knocks.

6. Never put it in boiling water for sterilizing.

7. Never allow chemical solutions to come into contact with the pacifier and keep away from direct sunlight or heat sources as well as strong electromagnetic sources to avoid interference.

8. Do not store the unit in direct sunlight or at high temperature.

9. Do not charge the battery.

10. Do not dispose of the battery in fire as it may explode.

11. Dispose of the product and battery according to relevant national regulations.

Caution! Be sure that both the battery and the thermometer are kept away from children. The pacifier is not a toy! Children MUST NOT be allowed to play with it! Remove the pacifier once you have taken your child’s temperature and place back in its cover.

English

4.2. PRODUCT FEATURES

4.2.1. PARTS DESCRIPTION

1. ON / OFF Button

2. LCD Display

3. Sensor

4. Cover

36.8

c

1

2

3

4.3. INSTRUCTIONS FOR USE

Place the pacifier in your child’s mouth before turning it on.

1. Press and release the power button (1). A short beep will be heard and all the symbols will be turned on for 1 second to show that all parts of the LCD are functioning properly,

188.8

e

M

then the last measurement will be shown on the display for 2 seconds.

32.3

e

M

After this, the screen will very briefly show 37.0ºC (98.6ºF) if operating correctly.

37.0

e

M

When the flashing ºC or ºF symbol appears take the body temperature.

32.3

C

2. On reaching its thermal equilibrium or peak, the ºC (or ºF) symbol stops flashing and the beeper will sound 10 times. The temperature is now ready to be read and will not change once the pacifier is removed.

3. If the measured temperature is higher then 37.8ºC (100ºF), the fever warning function will emit 3 fast beeps 10 times .

4. LCD will display “HiºC” or “HiºF” if the measured temperature is above 43.0ºC

(109.9ºF).

5. LCD will display “LoºC” or “LoºF” if the measured temperature is below

32.0ºC (90ºF).

6. After measuring, turn off the unit by pressing the power button. If the power is left on, the thermometer will switch off after approximately 9 minutes to save power.

7. Press and release power button to take another reading.

8. When it is used to take a child’s temperature, close supervision must be provided by an adult.

4.3.1. METHOD OF USE

You are advised to take the temperature at a specific time of the day, if it is earlier or later, the temperature could be lower or higher.

Your child’s mouth should be closed and they should breath slowly through their nose for 1 minute for a stable temperature. In taken air could affect the accuracy of the result.

The measurement time is between three and five minutes and the normal temperature is about 37.0ºC (98.6ºF).

English

Notes:

- A normal temperature may vary slightly from one person to another and vary

according to the time of day.

- This method produces approximately the same temperature as a temperature

taken orally with the body thermometer.

- Cleaning is required before every measurement.

- Measurement is inaccurate when the battery is low.

4.4. MAINTENANCE:

4.4.1. CLEANING

The thermometer is waterproof and can be cleaned with water or by wiping it with an antiseptic liquid or antiseptic wipe and then rinse with plenty of water.

Notes:

- Other lotions or cleaning methods may damage the thermometer.

- The product doesn’t need any calibration and is recommended to use for a

period of 3 years.

- Never put in boiling water for sterilizing.

4.4.2. REPLACE UNIT

The battery mark will flash if the battery is low. The digital pacifier thermometer is a sealed unit and must be replaced when the battery becomes low.

To maintain the life of the battery, this unit features an automatic shut-off function that activates approximately within 8~10 minutes of operation.

4.5. TECHNICAL SPECIFICATION

- Measurement range: 32.0ºC~42.9ºC / 90ºF~109.9ºF

- Measuring time: about 2~4 minutes

Accuracy ºC:

Accuracy Measurement Range

±0.1 ºC

±0.2 ºC from 35.0 ºC to 42.0 ºC remaining range at an ambient of 22ºC

Accuracy ºF:

Accuracy

±0.2 ºF

±0.4 ºF

Measurement Range

from 95.0ºF to 107.6ºF remaining range at an ambient of 71.6ºF

- Resolution: 0.1ºC/ºF

- Power: 1 x LR41 button battery, 1.5V

- Battery life: About 250 hours or 1000 times.

- Storage temperature: -10ºC to 40ºC (14 ºF to 104ºF). Humidity: 30% to 90%

- Operation temperature: 10 ºC to 35ºC (50 ºF to 95ºF). Humidity 40% to 80%

- Weight: 16g

- Display: 31/2 LCD digits

English

4.5.1. NORMALIZED SYMBOLS

· Attention: Read the operating Instructions for use!

· Applied part of type BF

· Disposal in accordance with Directive 2002/96/EC (WEEE)

· Complies with the European Medical Device Directive (93/42/EEC) and

amended by directive 2007/47/ECrequirements.

0434

Notified body 0434: Det Norske Veritas Certification AS

· Batch code

· Serial number

· The name and the address of the manufacturer.

· Authorized Representative in the European Community

5. BATH THERMOMETER

5.1. SAFETY INSTRUCTIONS

1. The device is safe for children to play with. It complies with EU toys safety and electromagnetic standards.

2. Swallowing a battery may be fatal.

3. Do not charge the battery.

4. Do not dispose of the battery in fire as it may explode.

5. Do not store the unit in direct sunlight or at high temperature.

6. Do not attempt to disassemble the unit except to replace the battery.

7. Use CR2032 batteries only.

8. Ensure batteries are inserted with the correct polarity.

9. Do not mix different battery types.

10. Do not mix new and used batteries together.

11. Consumed batteries should be removed from the device and disposed of in accordance with the makers recommendation.

12. Remove the batteries for long periods of non-use.

13. The supply terminals are not to be short-circuited.

14. Remove battery compartment lid using tweezers or similar object.

15. Toy is to be operated in water only when fully assembled in accordance with the instructions

5.2. PRODUCT FEATURES

5.2.1. PARTS DESCRIPTION

1. Display LCD

2. Alarm indicator

3. Battery compartment

4. Sensor

Front Rear

English

26

2

1

CLOSE

OPEN

3

4

5.3. INSTRUCTIONS FOR USE

The main function of this penguin is to measure the bath water temperature. It floats and is waterproof.

1. The thermometer is switched on by tapping the base of the penguin against the palm of your hand.

2. If the water temperature is less than 33ºC, the penguin displays “COLD”.

3. If the water temperature is between 33ºC – 38ºC it displays only the temperature, because the water is at the ideal temperature.

4. And if the water temperature is higher than 38ºC the penguin will alert you by flashing and displaying “HOT”.

5. To maintain the life of the battery, this unit features an automatic shut-off function that activates approximately within 60 minutes of operation.

6. Children must be accompanied and supervised by an adult when using this for bathing.

5.4. MAINTENANCE

5.4.1. HOW TO CLEAN

The thermometer can be cleaned using a soft damp cloth.

NOTE! Never clean the bath thermometer by wiping it with alcohol.

5.4.2. HOW TO REPLACE THE BATTERY

When the display is dim this indicates that the battery needs changing.

1. Open the battery cover by using tweezers or similar object.

2. Replace the CR2032 battery. Please make sure the (+) side is facing upwards and the waterproof lining is in place.

3. Close the battery cover securely.

English

A B C

+

CR2032

D

CLOSE

OPEN

CLOSE

OPEN

5.5. TECHNICAL SPECIFICATION

- Measurement range: 0ºC~50ºC / 32ºF~122ºF

- Measuring time: 2 seconds

- Accuracy: +/-1 ºC (+/-2ºF)

- Resolution: 1ºC / 1.8ºF

- Storage temperature: -10ºC~60ºC / 14ºF~140ºF

- Power: 1xCR2032

- Battery life: about 1 year (based on 1 hour using per day)

- Display unit ºC or ºF (factory setting only)

- High temperature alarm ≧39ºC

6. BATTERY & PRODUCT DISPOSAL INFORMATION

Remember to dispose of the product responsibly.

Do not mix products bearing the crossed out bin symbol with your general household waste.

For the correct collection and treatment of these products take them to the collection points designated by your local authority. Alternatively contact the retailer who sold you the product.

Miniland S.A. P.Ind. La Marjal I C/ La Patronal 8-10

03430 ONIL (Alicante) SPAIN

Technical support: +34 966 557 775

Fax +34 965 565 454 www.minilandbaby.com · [email protected]

©

Miniland, S.A. 2012

89119 thermokit

Cotronic Technology Ltd.

Floor 4~6, Block 7, West of ZhouShi Road, XiXiang Street,

BaoAn Zone, Shenzhen, Guangdong, China

EC REP

Locus Electronic Limited

28 Eustace Crescent, Rochester, Kent, ME2 2GL, U.K.

Ref. 89119-1 Ref. 89119-2 Ref. 89119-3

0434 0434

260514

advertisement

Key Features

  • High-precision sensor for accurate temperature readings
  • Versatile use with oral, axillary, and rectal modes
  • Fast and reliable temperature readings within seconds
  • Fever warning function with audible alerts
  • Auto power-off feature for extended battery life
  • Large and easy-to-read LCD display
  • Compact and portable design for convenient use
  • Replaceable battery for long-lasting operation
  • Waterproof design for easy cleaning and disinfection

Related manuals

Frequently Answers and Questions

How do I take a temperature reading with the thermometer?
Press the power button to turn on the thermometer. Wait for the display to show the last measurement. When the flashing ºC or ºF symbol appears, place the probe in the desired measurement location (mouth, armpit, or rectum). The thermometer will beep when the peak temperature is reached, indicating that the reading is ready.
What does the fever warning function do?
The fever warning function alerts you with audible beeps when the measured temperature is higher than 37.8ºC (100ºF), indicating a potential fever.
How do I replace the battery in the thermometer?
When the low battery indicator appears on the display, replace the battery by opening the battery cover and inserting a new CR2032 battery with the positive (+) side facing up.

advertisement