HT SOLAR I-Ve User manual
Add to my manuals
80 Pages
HT SOLAR I-Ve is a powerful and versatile instrument designed for testing and monitoring photovoltaic systems. It offers a wide range of features and functions to help you quickly and easily assess the performance of your PV system.
With HT SOLAR I-Ve, you can:
- Measure irradiance, temperature, and humidity
- Perform I-V curve measurements
- Test PV systems with single or multiple MPPT inverters
- Generate detailed reports
- Store and manage data on your PC
HT SOLAR I-Ve is the ideal tool for anyone who needs to test and monitor photovoltaic systems. Whether you are a solar installer, a maintenance technician, or a homeowner, HT SOLAR I-Ve can help you ensure that your PV system is operating at peak performance.
advertisement
Copyright HT ITALIA 2017
Release EN 2.00 - 24/03/2017
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
Table of contents:
1.
SAFETY PRECAUTIONS AND PROCEDURES .......................................................... 3
1.1. Preliminary instructions ..................................................................................................... 3
1.2. During use ......................................................................................................................... 4
1.3. After use ............................................................................................................................ 4
1.4. Overvoltage categories - Definitions ................................................................................. 4
2.
GENERAL DESCRIPTION ........................................................................................... 5
2.1. Introduction ........................................................................................................................ 5
2.2. Instrument features ........................................................................................................... 5
3.
PREPARATION FOR USE ........................................................................................... 6
3.1. Initial checks ...................................................................................................................... 6
3.2. Instrument power supply ................................................................................................... 6
3.3. Storage .............................................................................................................................. 6
4.
DESCRIPTION OF PARTS .......................................................................................... 7
4.1. Instrument description ....................................................................................................... 7
4.2. Keyboard description ......................................................................................................... 8
4.3. Display description ............................................................................................................ 8
4.4. Initial screen ...................................................................................................................... 8
5.
MAIN MENU ................................................................................................................. 9
5.1. SET – Instrument settings ................................................................................................. 9
5.1.1.
5.1.2.
5.1.3.
5.1.4.
5.1.5.
5.1.6.
General .................................................................................................................................... 9
Measurement units ................................................................................................................ 10
Date/Time ............................................................................................................................... 10
Remote unit/Solarim. ............................................................................................................. 11
Irradiance ............................................................................................................................... 12
DC Clamp (SOLAR I-Vw or SOLAR I-Ve) ............................................................................. 12
5.2. EFF – SETTINGS FOR TESTING PV SYSTEMS (SOLAR I-Vw or SOLAR I-Ve) .......... 13
5.2.1.
Settings for PV systems with single-MPPT inverter - single-phase AC output ..................... 13
5.2.1.1.
5.2.1.2.
5.2.2.
5.2.2.1.
Instrument settings ........................................................................................................................... 13
PV plant parameters ........................................................................................................................ 14
Settings for PV with single/multi MPPT inverter – single/three phase AC output .......................... 15
Instrument settings ........................................................................................................................... 15
5.2.2.2.
5.2.2.3.
PV plant parameters ........................................................................................................................ 16
MPP300 status ................................................................................................................................ 17
5.2.3.
Selection of the compensation relationship of temperature effects ....................................... 18
5.3. DB – Module database .................................................................................................... 19
5.3.1.
5.3.2.
How to define a new PV module ............................................................................................ 20
5.3.3.
How to modify a PV module ................................................................................................... 21
How to delete a PV module ................................................................................................... 21
6.
HOW TO OPERATE ................................................................................................... 22
6.1. Testing PV systems (SOLAR I-Vw or SOLAR I-Ve) ........................................................ 22
6.1.1.
6.1.2.
Testing of PV systems with single-MPPT inverter - single-phase AC output ........................ 23
Test PV systems with single/multi-MPPT inverter – single/three-phase AC output ................... 27
6.2. I-V curve measurement ................................................................................................... 33
6.2.1.
6.2.2.
I-V curve measurement by use of instrument ........................................................................ 33
I-V curve measurement by use of remote unit SOLAR-02 .................................................... 37
6.2.2.1.
6.2.2.2.
I-V curve test by remote unit SOLAR-02 in RF connection .............................................................. 37
6.2.3.
I-V curve test by remote unit SOLAR-02 in synchro recording ......................................................... 41
Meaning of measurement results .......................................................................................... 46
6.3. Quick Check on PV modules and strings (IVCK) ............................................................ 47
6.3.1.
6.3.2.
6.3.3.
General information ............................................................................................................... 47
6.3.4.
6.3.5.
Preliminary settings ................................................................................................................ 48
Fast check IVCK without irradiance measurement ................................................................ 49
Fast check IVCK with irradiance measurement ..................................................................... 51
Reset Media ........................................................................................................................... 53
6.4. List of display messages ................................................................................................. 54
7.
STORING DATA ......................................................................................................... 55
7.1. Saving test results of PV checks (SOLAR I-Vw or SOLAR I-Ve) .................................... 55
7.2. Saving test results of I-V curve test ................................................................................. 55
EN - 1
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
7.3. Managing the results ....................................................................................................... 56
7.3.1.
7.3.2.
Recall data of PV checks at display (SOLAR I-Vw or SOLAR I-Ve) ..................................... 56
Recall data of I-V curve tests at display ................................................................................. 57
7.3.2.1.
7.3.2.2.
7.3.2.3.
View data – Numerical screens ........................................................................................................ 58
View data – Graphic screens of I-V curve ........................................................................................ 59
View data – Graphic screens of power ............................................................................................. 60
7.3.3.
Delete data ............................................................................................................................. 61
8.
CONNECTING THE INSTRUMENT TO A PC ............................................................ 62
8.1. Connecting by optical/USB cable C2006 ........................................................................ 62
8.2. Connecting by WiFi ......................................................................................................... 63
9.
MAINTENANCE .......................................................................................................... 64
9.1. General ............................................................................................................................ 64
9.2. Battery replacement ........................................................................................................ 64
9.3. Instrument cleaning ......................................................................................................... 64
9.4. End of life ........................................................................................................................ 64
10.
TECHNICAL SPECIFICATIONS ................................................................................. 65
10.1. Yield test technical feratures (SOLAR I-Vw, SOLAR I-Ve) .............................................. 65
10.2. I-V test technical features (I-V and IVCK) ....................................................................... 67
10.3. Safety specifications ........................................................................................................ 68
10.3.1.
General .................................................................................................................................. 68
10.4. General characteristics .................................................................................................... 69
10.5. Environment conditions ................................................................................................... 69
10.6. Accessories ..................................................................................................................... 69
11.
APPENDIX ................................................................................................................. 70
11.1. Testing photovoltaic systems (SOLAR I-Vw, SOLAR I-Ve) ............................................. 70
11.2. Notes on MPPT (Maximum Power Point Tracker) .......................................................... 71
11.3. Theoretical aspects of the I-V curve measurement ......................................................... 72
11.3.1.
Typical errors on I-V curves and troubleshooting .................................................................. 73
12.
SERVICE .................................................................................................................... 75
12.1. Warranty conditions ......................................................................................................... 75
12.2. Service ............................................................................................................................ 75
EN - 2
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
1. SAFETY PRECAUTIONS AND PROCEDURES
The word “instrument” in this manual means generically both the models I-V400w, I-V500w,
SOLAR I-Vw or SOLAR I-Ve e SOLAR I-Ve except notation specifically indicated. This instrument has been designed in compliance with directives IEC/EN61010-1 regarding electronic measuring instruments. Before and while measuring, carefully follow the instructions below:
Do not perform voltage or current measurements in humid environments
Do not perform measurements near explosive gas or material and fuels or in dusty environments
Avoid contact with the circuit tested if no measurement is being performed
Avoid contact with exposed metal parts, test terminals not in use, circuits, etc.
Do not perform any measurement if instrument anomalies are detected, such as deformations, breaks, leakage of substances, no display reading, etc.
Only use original HT accessories
In this manual, following symbols are used:
CAUTION: Follow the instructions given in this manual; improper use may damage the instrument, its components or create dangerous situations for the operator
High voltage: risk of electrical shock
Double insulation
DC voltage or current
AC voltage or current
Ground reference
1.1. PRELIMINARY INSTRUCTIONS
This instrument has been designed for use in the environmental conditions specified in
§ 10.5. Do not use in different environmental conditions.
The instrument may be used for measuring of VOLTAGE and CURRENT in CAT II
1000V DC or CAT III 300V to ground, maximum voltage 1000V between inputs (I-
V400w and SOLAR I-Vw or SOLAR I-Ve) or 1500V (I-V500w and SOLAR I-Ve) Do not use on systems exceeding the limit values specified in § 10.1 and § 10.2
We recommend to follow the ordinary safety rules aimed at: your protection against dangerous currents, the instrument’s protection against improper use.
Only the accessories provided with the instrument guarantee compliance with safety standards. They must be in good conditions and must be replaced, if necessary, with identical models.
Check that batteries are correctly inserted.
Before connecting the test cables to the circuit being tested, check that the desired function has been selected.
EN - 3
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
1.2. DURING USE
We recommend to carefully read the following recommendations and instructions:
CAUTION
Failure to comply with the CAUTIONs and/or instructions may damage the instrument and/or its components or cause dangers to the operator.
The symbol “
” indicates the charge level. When there are five bars, it means that batteries are fully charged; a decrease in the number of bars down to “ ” indicates that the batteries are almost low. In this case, interrupt tests and replace the batteries according to the indications given in § 9.2. The instrument is capable of keeping data stored even
without batteries.
1.3. AFTER USE
When measuring operations are completed, turn off the instrument by pressing and holding the ON/OFF key for a few seconds. Should the instrument remain unused for a long time, remove batteries and follow the indications given in § 3.3.
1.4. OVERVOLTAGE CATEGORIES - DEFINITIONS
Standard IEC/EN61010-1 (Safety requirements for electrical equipment for measurement, control and laboratory use, Part 1: General requirements) defines what a measurement category (usually called “overvoltage category”) is. At § 6.7.4: Measuring circuits it quotes:
Circuits are divided into the following measurement categories:
Measurement Category IV is for measurements performed at the source of the lowvoltage installation.
Examples are electricity meters and measurements on primary overcurrent protection devices and ripple control units.
Measurement Category III is for measurements performed in the building installation.
Examples are measurements on distribution boards, circuit breakers, wiring, including cables, bus-bars, junction boxes, switches, socket-outlets in the fixed installation, and equipment for industrial use and some other equipment, for example, stationary motors with permanent connection to fixed installation.
Measurement Category II is for measurements performed on circuits directly connected to the low voltage installation.
Examples are measurements on household appliances, portable tools and similar equipment.
Measurement Category I is for measurements performed on circuits not directly connected to MAINS.
Examples are measurements on circuits not derived from MAINS, and specially protected (internal) MAINS-derived circuits. In the latter case, transient stresses are variable; for that reason, the norm requires that the transient withstand capability of the equipment is made known to the user.
EN - 4
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
2. GENERAL DESCRIPTION
2.1. INTRODUCTION
The instrument was designed to perform complete yield test on single-phase (and threephase if combined with optional accessory MPP300) PV installations with calculations of efficiencies (SOLAR I-Vw or SOLAR I-Ve) and to perform all I-V curve tests on single modules (panels) or complete strings of photovoltaic (PV) plants in order to verify the reference parameters given by the manufacturers.
2.2. INSTRUMENT FEATURES
The instrument can perform the following tests:
Check Single phase PV installations (EFF - SOLAR I-Vw and SOLAR I-Ve)
DC voltage and current measurements (1000V for SOLAR I-Vw 1500V for SOLAR I-Ve)
AC TRMS voltage and current measurements
DC/AC power measurement
Solar irradiation [W/m
2
] measured by use of reference cell connected to unit SOLAR-02
Module and environ. temp. measurement by means of probe connect to SOLAR-02
Use of compensation relationships on PDC measurements
Immediate evaluation of result OK/NO of a test
Checks with programmable IP from 5s to 60min
Testing single/three-phase PV systems (MPP – SOLAR I-Vw, SOLAR I-Ve + MPP300)
Measurement of 3 DC voltages and currents
Measurement of DC string power and total DC power
Measurement of 3 AC TRMS voltages and currents
Measurement of total AC power
Measurement of irradiation [W/m
2
] by means of a reference cell connected to SOLAR-02
Module and environ. temp. measurement by means of probe connect to SOLAR-02
Application of compensation relationships to DC efficiency.
Immediate evaluation of result OK/NO of a test
Parameter recording of a PV system with 5s to 60min programmable IP
I-V Curve measurement
Voltage, current, power of PV module/string 1000VDC, 15ADC (SOLAR I-Vw, I-V400w) module/string
Activation of manual or automatic measurement
PV module/string temperature measurement
Irradiance measurement [W/m
2
] by using reference cell
Numerical and graphical results of I-V curve with 4-wire method
Comparison of results with standard conditions (STC) and OK/NO final response
Internal customizable database for definition up to 30 PV modules
Annual degradation evaluation of PV modules/strings/fields
Internal memory for saving test and Optical/USB interface for PC connection
Quick Check (IVCK)
Fast test of open voltage and short circuit current up to 1000V DC, 15A (SOLAR I-Vw,
I-V400w) and up to 1500VDC, 15A (SOLAR I-Ve, I-V500w)
Activation of manual or automatic measurement
Immediate evaluation (OK/NO) of test results
The instrument has an innovative electronic function selector to simply set operations with internal parameters, a backlit display, the contrast adjustment and a HELP key in order to give a valid help to the operator during the meter’s connection to the plants. An enable/disable Auto Power OFF feature is available after about 5 minutes of idleness.
EN - 5
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
3. PREPARATION FOR USE
3.1. INITIAL CHECKS
Before shipment, the instrument’s electronics and mechanics have been carefully checked.
All possible precautions have been taken to have the instrument delivered under optimal conditions. However, we recommend to rapidly check the instrument in order to detect possible damage occurred during transport. Should you detect anomalies, please immediately contact the dealer.
It is also recommended to check that the package contains all parts indicated in § 10.6. In case of discrepancies, please contact the dealer. Should it be necessary to return the instrument, please follow the instructions given in § 12.
3.2. INSTRUMENT POWER SUPPLY
The instrument is battery supplied. For battery type and life, see § 10.4.
The symbol “ ” indicates the charge level. When there are five bars next to the battery symbol, it means that batteries are fully charged; a decrease in the number of bars down to “ ” indicates that the batteries are almost low. In this case, interrupt tests and replace the batteries according to the indications given in § 9.2.
The instrument is capable of keeping data stored even without batteries.
The instrument is provided with advanced algorithms to maximize the batteries’ life.
During instrument operation, a further short pressing of the key turns on the display’s backlighting (if battery voltage level is high enough). In order to save battery efficiency, backlighting automatically turns off after approx. 30 seconds.
A frequent use of back lighting reduces the batteries’ life
3.3. STORAGE
In order to guarantee precise measurements, after a long period of the instrument’s storage under extreme environmental conditions, wait until it recovers its normal conditions
(see § 10.5).
EN - 6
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
4. DESCRIPTION OF PARTS
4.1. INSTRUMENT DESCRIPTION
CAPTION:
1. Inputs
2. Display
3. Connector for opto insulated cable
4. Arrows/ENTER key
5. GO/STOP key
6. SAVE key
7. ON/OFF key
8. HELP / key
9. ESC/MENU key
Fig. 1: Description of the front part of the instrument
CAPTION:
1. Input for irradiance probe (I-V) / DC transducer clamp (EFF – SOLAR I-
Vw, SOLAR I-Ve)
2. Input for auxiliary temperature probe
(I-V) / AC transducer clamp (EFF –
SOLAR I-Vw, SOLAR I-Ve) measurement (I-V) / voltage measurement (EFF – SOLAR I-Vw,
SOLAR I-Ve)
P1, P2 inputs for voltage measurement (I-V) / AC voltage
(EFF – SOLAR I-Vw, SOLAR I-Ve)
Fig. 2: Description of the upper part of the instrument
CAPTION:
1. Connector for opto insulated cable connection
Fig. 3: Description of the instrument’s side
EN - 7
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
4.2. KEYBOARD DESCRIPTION
The keyboard includes following keys:
ON/OFF
key to switch on/off the instrument
ESC/MENU
key to exit the selected menu without confirming and to activate menu management
keys to move the cursor through the different screens in order to select the desired programming parameters
ENTER
key to confirm the modifications and the selected programming parameters and to select the function from the menu
GO/STOP
key to start measurements
SAVE
key to save the measured values
HELP
key (long pressure) to display an indicative outline of the connections between the instrument and the system being tested in the function set
key (short pressure) to turn on the display’s backlighting
4.3. DISPLAY DESCRIPTION
The display is a graphic module with a resolution of 128 x 128 dots. The display’s first line indicates the system date/hour and the battery charge indicator.
In the bottom part the functionality of ENTER key and the active mode is displayed
15/05/10 15:34:26
The symbol indicates the presence of an active radio connection with the selected remote unit (SOLAR-02 or
MPP300).
The flashing symbol indicates that the instrument is currently searching for a radio connection with the selected
Selection remote unit (SOLAR-02 or MPP300).
I – V
4.4. INITIAL SCREEN
When turning on the instrument, the instrument displays the initial screen for a few seconds. It displays the following:
The instrument’s model
The manufacturer’s name
Internal presence of modules for RF (Radio Frequency) and WiFi connection
The serial number (SN:) of the instrument
The firmware version (FW:) in the instrument's memory
The date of the last calibration (Calibration date:)
S O L A R
I - V w
H T
R F - W i S N : 1 5 3 4 5 6 7 8
FW: 7.07 H4
Calibration date:
09/02/2016
Then, the instrument switches to the last function selected.
EN - 8
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
5. MAIN MENU
Pressing the MENU/ESC key in any allowable condition of the instrument displays the main menu screen, in which the instrument may be set, the saved measures can be displayed and the desired measuring function may be set (item EFF only appears for instrument SOLAR I-Vw or SOLAR I-Ve).
Select by cursor the desired options and confirm by pressing
ENTER
15/05/10 15:34:26
I - V
I-V Test
E F F Y i e l d T e s t
S E T
S e t t i n g s
DB
Modules
MEM
Data Recall
PC
PC connection
E N T E R t o s e l e c t .
M E N U
5.1. SET – INSTRUMENT SETTINGS
Move the cursor to SET by means of the arrow keys (,) and confirm with ENTER. Subsequently, the displays shows
15/05/10 15:34:26
G e n e r a l the screen which provides access to the various instrument’s settings.
M e a s . U n i t
The settings will remain valid also after switching off the instrument
D a t e a n d T i m e
R e m o t e U n i t / S o l a r i m .
I r r a d i a n c e
D C C l a m p
E N T E R t o s e l e c t .
S E T
5.1.1. General
1. Move the cursor to “General” by means of the arrow keys
(,) and confirm with ENTER
2. The displays shows the screen which allows:
Setting the instrument language
Setting the enable/disable of auto power off
Adjusting the contrast of display
Enabling the acoustic signal when pressing a key
Setting the enable/disable of WiFi interface inside each section of main menu for using the instrument with the
HTAnalysis APP (in the PC section the WiFi interface is always on). With WiFi enabled the “ ” symbol is shown in the bottom right side of the display. The enable of
WiFi interface means a higher consumption of the batteries
3. By means of the arrow keys (,) and select the desired option by means of the arrow keys (,)
4. Confirm with SAVE and the “Data saved” message is displayed for a while. Press the ESC/MENU key to exit without saving and to go back to the previous screen
15/05/10 15:34:26
L a n g u a g e : E n g l i s h
A U T O P O W E R O F F : N O
C o n t r a s t : 1 0
K e y B e e p : N O
W i F i : N O
S A V E t o s t o r e d a t a
S E T
EN - 9
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
units
This § allows setting the default measurement units of some internal parameters included in the database (DB) for the management of PV modules (§ 5.3)
1. Move the cursor to “Meas. Unit” by means of the arrow keys (,) and confirm with ENTER.
2. The display shows the screen which allows selecting the measurement units of parameters shown by the meter.
3. Press ESC/MENU to exit without saving any setting.
15/05/10 15:34:26
P a r a m e t e r
E N T E R t o s e l e c t
M E N U
4. Move the cursor to “Parameter” by means of the arrow keys (,) and confirm with ENTER.
5. The display shows the screen which allows selecting the measurement units of typical parameters of the modules:
Alpha possible selection: “%/°C” and “mA/°C”
Beta possible selection: “%/°C” and “mV/°C”
Gamma possible selection: “%/°C” and “W/°C”
Tolerance possible selection: “%” and “W”
6. Set the desired units by means of the arrow keys (,).
7. Confirm with SAVE and the “Saved data” message is
15/05/10 15:34:26
A l p h a
B e t a
G a m m a
T o l e r a n c e
: m A / ° C
: m V / ° C
: W / ° C
: %
SAVE to store data
S E T
displayed for a while. Press ESC/MENU key to exit without saving and to go back to the previous screen.
5.1.3. Date/Time
1. Move the cursor to “Date/Time” by means of the arrow
15/05/10 15:34:26 keys (,) and confirm with ENTER
2. The display shows the screen which allows setting the system date/time both in the European (EU) or USA (US) format.
3. Set the values by means of the arrow keys (,).
4. Confirm with SAVE and the “Saved data” message is
Y e a r
M o n t h
D a y
H o u r
M i n u t e
F o r m a t
: 2 0 0 9
: 0 6
: 1 5
: 0 9
: 5 3
: E U displayed for a while. Press ESC/MENU key to exit without saving and to go back to the previous screen.
SAVE to store data
S E T
EN - 10
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
unit/Solarim.
This section allows selecting the type of remote unit to be used (if available) and setting the values of typical parameters (Sensitivity and Alpha) of the reference irradiance cell supplied with the meter. The values of these parameters, which are printed on the
back label of the cell, depend on the type of PV modules on test.
1. Move the cursor to “Remote unit” by means of the arrow keys (,) and confirm with ENTER
2. The display shows the screen which allows:
Select the type of remote unit to be used for testing PV systems (SOLAR I-Vw or SOLAR I-Ve): o
NO: Remote unit disabled o
SOLAR: Use of SOLAR-02 o
MPP300: Use of MPP300 (optional)
Enable/disable the use of the remote unit SOLAR-02 for
I-V measurements (opt. for I-V400w and I-V500w). In case the use of the remote unit has not been enabled, it will be possible to set the sensitivity values (sens.) of the reference cell provided, expressed in “mV/kW*m-2”, and of the Alpha parameter.
3. Set the value by means of the arrow keys (,)
4. Confirm with SAVE and the “Saved data” message is displayed for a while. Press ESC/MENU key to exit without saving and to go back to the previous screen
15/05/10 15:34:26
R e m o t e U E F F
R e m o t e U I - V :
M P P 3 0 0
: N O
S e n s . : 3 1 . 0 m V / k W / m 2
A l p h a : 0 . 0 6 0 % / ° C
S A V E t o s t o r e d a t a
S E T
CAUTION
For EFF measurements (testing of PV systems – SOLAR I-Vw or SOLAR I-
Ve), disabling the remote unit entails:
The impossibility of carrying out Irradiation and Temperature
measurements by means of unit SOLAR-02
The impossibility of using the unit MPP300 (if available)
Consequently, it will be impossible to obtain a result for the test carried out.
EN - 11
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
5.1.5. Irradiance
1. Move the cursor to “Irradiance” by means of the arrow
15/05/10 15:34:26 keys (,) and confirm with ENTER.
2. The display shows the screen with the options: “ Irr min
IV
” which allows to set the minimum threshold of measured irradiance expressed in W/m
2
, used by meter as
I r r m i n I V : 0 W / m 2
I r r m i n E F F : 6 0 0 W / m 2 reference during the I-V curve measurements and “Irr min
EFF
” (SOLAR I-Vw or SOLAR I-Ve which allows to set the minimum threshold of measured irradiance expressed in
W/m
2
, used by meter as reference during the check PV
SAVE to store data
S E T
efficiency measurements
3. Set the value by means of the arrow keys (,). The accuracy indicated in this manual is granted under condition indicated in § 10.1 and § 10.2. In I-V curve measurement the value is set within the 0
800 W/m
2
interval while 400
800 W/m
2
for the PV check operations (SOLAR I-Vw or SOLAR I-Ve)
4. Confirm with SAVE and the “Saved data” message is displayed for a while. Press ESC/MENU key to exit without saving and to go back to the previous screen.
CAUTION
The “0 W/m2” setting for parameter “Min Irr IV” allows carrying out I-V measurements without the following conditions being checked:
Connection of reference cell to input IRR of the instrument
Unsteady irradiation values
Number of modules consistent with the measured open circuit voltage
5.1.6. DC Clamp (SOLAR I-Vw or SOLAR I-Ve)
This option allows setting a possible correction factor K for the DC clamp, in order to improve current measurement. If present, the correction factor is indicated on the rear label of the clamp itself, indicated as:
In case no label is present, set k = 1.000
K= X.xxx
1. Move the cursor to “DC Clamp” by means of the arrow keys (,) and confirm with ENTER
15/05/10 15:34:26
2. The item: “DC clamp k” is shown at display to set the corrective factor in a interval within 0.950 and 1.050. Set the value by means of the arrow keys (, )
3. Confirm with SAVE and the “Saved data” message is
D C c l a m p k : 1 . 0 0 0 displayed for a while. Press ESC/MENU key to exit without saving and to go back to the previous screen
SAVE to store data
S E T
EN - 12
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
5.2. EFF – SETTINGS FOR TESTING PV SYSTEMS (SOLAR I-VW OR SOLAR I-VE)
Further in this manual, the acronym MPPT (Multiple Power Point Tracker) shall indicate the characteristic of the DC/AC converter (inverter), capable of maximizing the DC power which can be taken from the photovoltaic field. See § 11.2 for further details.
5.2.1. Settings for PV systems with single-MPPT inverter - single-phase AC output
Check beforehand the settings made in MENUSETRemote Unit and check that you have selected “SOLAR” as setting for parameter “R. unit”.
1. Press MENU key, move the cursor to “EFF” by means of 15/05/10 15:34:26 the arrow keys (,) and confirm with ENTER. The screen with the values of electrical parameters relative to
P r p
I r r
PV generator is displayed
P n o m
T c
T e
P d c
V d c
I d c n d c
- - -
- - -
W / m 2
0 . 0 0 0 k W
- - -
° C
- - -
° C
0 . 0 k W
0 . 0 0 0
V
0 . 0 A
- - -
G O – S t a r t r e c
Select
E F F
2. Press ENTER key. The options Set PV plant and Set
Instrum.
are displayed
3. By means of the arrow keys (,) to select Set Instrum. and confirm with ENTER. The herewith screen is shown:
15/05/10 15:34:26
P R p
I r r
P n o m
T c
T e
P d c
V d c
I d c n d c
- - -
- - -
W / m 2
0 . 0 0 0 k W
- - -
° C
- - -
° C
0 . 0 k W
0 . 0 0 0
V
0 . 0 A
- - -
Set PV plant
S e t I n s t r u m .
Select
E F F
4. By means of the arrow keys (,) the herewith
15/05/10 15:34:26 parameters are set:
IP (integration period) used by meter during yield test recording. The values I 5s, 10s, 30s, 60s, 120s, 300s,
600s, 900s, 1800s, 3600s
are selectable
I P : 5 s
F S D C c l a m p : 1 0 0 0 A
F S A C c l a m p : 1 0 0 0 A
The full scale (FS) of DC clamp used on DC current measurement with selectable value within 1A
3000A
The full scale (FS) of AC clamp used on AC current measurement with selectable value within 1A
3000A
5. Press the SAVE key to save the settings made; the message “Data saved” will be displayed for a few seconds. Press the ESC/MENU key to exit without saving and go back to the previous screen.
SAVE to store data
E F F
EN - 13
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
5.2.1.2. PV plant parameters
1. Press MENU key, move the cursor onto EFF by using the arrow keys (,) and confirm with ENTER. The display shows the values of the output electrical parameters of the photovoltaic generator (DC side).
15/05/10 15:34:26
P R p
I r r
P n o m
T c
T e
P d c
V d c
I d c n d c
- - -
- - -
W / m 2
0 . 0 0 0 k W
- - -
° C
- - -
° C
0 . 0 k W
0 . 0 0 0
V
0 . 0 A
- - -
G O – s t a r t r e c
Select
E F F
2. Press the ENTER key. The instrument shows the 15/05/10 15:34:26 following options: “Set PV plant“ and “Set Instrum“.
3. Use the arrow keys (,) to select “Set PV plant” and confirm with ENTER. The instrument shows the following screen:
P r p
I r r
P n o m
T c
T e
P d c
V d c
I d c n d c
- - -
W / m 2
- - -
W / m 2
0 . 0 0 0 k W
- - -
° C
- - -
° C
0 . 0 k W
0 . 0 0 0
V
0 . 0 A
- - -
Set PV plant
S e t I n s t r u m e n t
4. By means arrow keys ( , ) the herewith parameters
15/05/10 15:34:26 are set:
Pmax maximum nominal power of the PV installation expressed in kW
Gamma coefficient of variation of power with the temperature, standard parameter of PV module
(range: -0.3
-0.5%/°C typical)
: 3 . 5 0 0 k W
G a m m a
N o c t
: 0 . 4 5 % / ° C
: 4 5 ° C
T e :
4 5 ° C
T . E n v
Noct nominal working temperature of a PV cell, standard parameter of PV module (range: 42
48°C typical)
Select
E F F
SAVE to store data
E F F
Te, Tc set of default values of environmental temperature and PV module temperature. These values are considered by the meter only without the auxiliary probe connected to emote unit SOLAR-02
Corr. Type set the corrective term Ptpv for DC efficiency calculation and for maximum DC efficiency visualization (see § 5.2.3)
EN - 14
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
5.2.2. Settings for PV with single/multi MPPT inverter – single/three phase AC output
See § 11.2 for further details about the meaning of MPPT. This mode requires unsing the remote unit MPP300 (optional). Check beforehand the settings made in
MENU
SETRemote Unit and check that you have selected “MPP300” as setting for parameter “R. unit”.
1. Press the MENU key, position the cursor onto EFF by
15/05/10 15:34:26 using the arrow keys (,) and confirm with ENTER.
The display shows the screen here to the side, which contains the global parameters of the system.
P R p
I r r
P n o m
T c
T e
P d c
P a c n d c n a c
- - -
- - -
W / m 2
1 5 0 . 0 k W
- - -
° C
- - -
° C
- - k W
- - k W
- - -
- - -
G O - s t a r t R e c
Select
M P P
2. Press the ENTER key. The instrument shows the 15/05/10 15:34:26 following options: MPP300 status, System Par. and
Instrument settings
3. Use the arrow keys (,) to select “Instrument
settings
the following screen:
.
” and confirm with ENTER. The instrument shows
P R p
I r r
P n o m
T c
T e
P d c
P a c n d c n a c
- - -
- - -
W / m 2
1 5 0 . 0 k W
- - -
° C
- - -
- - -
- - -
- - -
MPP300 Status
- - -
° C k W k W
Set PV plant
S e t I n s t r u m .
Select
M P P
4. By using the arrow keys ( , ) it is possible to set:
15/05/10 15:34:26
the integration period (IP) which can be used by the instrument when testing the parameters of a PV
system. The following values can be selected: 5s, 10s,
30s, 60s, 120s, 300s, 600s, 900s, 1800s, 3600s
current with value selectable between 1A
The FS of the DC clamp used for measuring DC
3000A
The FS of the AC clamp used for measuring AC current with value selectable between 1A
3000A
I P : 5 s
F S D C c l a m p : 1 0 0 0 A
F S A C c l a m p : 1 0 0 0 A
C l a m p T y p e : S T D
D c i n p u t s
A c s y t e m
1 + 2 + 3
3 p h a s e
The type of AC clamp used: STD (standard) or FLEX
(flexible clamp)
S A V E t o s t o r e d a t a
The number of DC inputs to be used for measuring:
1
, 1+2, 1+2+3
M P P
The type of AC electrical system: SINGLE, 4 wires
5. Press the SAVE key to save the settings made; the message “Data saved” will be displayed for a few seconds. Press the ESC/MENU key to exit without saving and go back to the previous screen.
EN - 15
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
5.2.2.2. PV plant parameters
1. Press MENU key, position the cursor onto EFF by using
15/05/10 15:34:26 the arrow keys (,) and confirm with ENTER. The display shows the screen here to the side, which contains
P R p
I r r the global parameters of the system.
P n o m
T c
T e
P d c
P a c n d c n a c
- - -
- - -
W / m 2
1 5 0 . 0 k W
- - -
° C
- - -
° C
- - k W
- - k W
- - -
- - -
G O – s t a r t r e c .
Select
M P P
2. Press the ENTER key. The instrument shows the 15/05/10 15:34:26 following options: MPP300 status, Set PV plant. and Set
Instrum.
3. Use the arrow keys (,) to select “Set PV plant” and confirm with ENTER. The instrument shows the following screen:
P R p
I r r
P n o m
T c
T e
P d c
V d c
I d c n d c
- - -
- - -
W / m 2
0 . 0 0 0 k W
- - -
° C
- - -
° C
0 . 0 k W
0 . 0 0 0
V
0 . 0 A
- - -
M P P 3 0 0 s t a t u s
Set PV plant
S e t i n s t r u m .
Select
M P P
4. By using the arrow keys ( , ) it is possible to set:
Range coefficient of power variation with temperature, characteristic parameter of PV modules
(typically in range: -0.3
-0.5%/C)
15/05/10 15:34:26
Pmax maximum rated power of PV installation expressed in kW
P m a x
G a m m a
N o c t
: 3 . 5 0 0 k W
: 0 . 4 5 % / ° C
: 4 5 ° C
T e :
4 5 ° C
C o r r . T y p e . :
4 0 ° C
T c :
T . E n v
Noct rated operating temperature of the cell, characteristic parameter of PV modules (typically in range: 42
48°C)
S A V E t o s t o r e d a t a
Te, Tc setting of default values of environmental and PV module temperatures. These values are taken
M P P
into consideration by the instrument only when probe is not connected to unit SOLAR-02
Corr. Type Setting of the compensation relationship on the calculation of Pdc power and the maximization of DC performance (see § 5.2.3)
EN - 16
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
status
If SOLAR I-Vw or SOLAR I-Ve is near MPP300, the general parameters of MPP300 may be displayed.
1. Position the cursor onto EFF by using the arrow keys
15/05/10 15:34:26
(,) and confirm with ENTER. The display shows the screen here to the side, which contains the global parameters of the system.
P R p
I r r
P n o m
T c
T e
P d c
P a c n d c n a c
- - -
- - -
W / m 2
1 5 0 . 0 k W
- - -
° C
- - -
° C
- - k W
- - k W
- - -
- - -
Select
G O – s t a r t r e c .
Select
M P P
2. Press the ENTER key. The instrument shows the 15/05/10 15:34:26 following options: MPP300 status, Set PV plant and Set
P o w e r s u p p l y P . S u p .
Instrum.
3. Use the arrow keys (,) to select “MPP300 status” and confirm with ENTER. The instrument shows the screen here to the side, which indicates the main general parameters of the instrument.
C h a r g e 9 9 %
S O L A R - 0 2
S I d e t e c t e d
V e r s i o n 1 . 0 1
S N 1 1 0 1 0 0 3 0
MPP300 status
Set PV plant
S e t i n s t r u m .
M E N U
EN - 17
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
5.2.3. Selection of the compensation relationship of temperature effects
This option allows selecting the relationship to be used to correct the measurements performed according to the modules’ temperature. The following correction modes are available:
- T.mod.:
Correction factor Rfv2 related to PV module Temp. (Italian guideline CEI-82-25)
- T.env:
- ndc:
Correction factor Rfv2 related to environment Temp. (Italian guideline CEI-82-25)
PRp (Performance ratio for active power) corrected by PV module Temp.
CAUTION
If it’s performed a test according to Italian guideline it is recommended to select
“T.env.” correction
Corr.
Type
Temperature (Tcel) PRp calculation Ref.
T.mod.
Tcel
Tmodule_Me as
T.amb. Tcel
Tamb
NOCT
20
Irr
800
R
fv2
1
1 (Tcel 40)
(if
100
(if
Tcel
Tcel
40
C)
40
C)
-
CEI
82-25
PRp
P ca
Rfv
2
G p
G
STC
P n
dc
Tcel
Tmodule_Me as
PRp
G
STC
G p
1
100
T cel
25
P ca
P n
- - - where:
G p
G
STC
Irradiance on PV module surface
Standard Irradiance = 1000
2
W/m
2
W/m
Nominal Power = sum of all power module (Pmax ) included in the part of PV plant under test
P n
AC Active Power measured
P ca
Rfv
2
NOCT
Thermal Coefficient factor
Absolute value of Pmax thermal coefficient
Normal Operating Cell Temperature (@ 800W/m
2
, 20°C, AM=1.5, vel. Aria =1m/s).
%/
C
%/
C
For further details see § 11.1
EN - 18
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
5.3. DB – MODULE DATABASE
The meter allows defining up to 30 different PV modules, further to a DEFAULT situation
(not editable and not erasable) which can be used as reference case when no piece of information about the module under test is available.
The parameters which can be set, with reference to 1 module, are described below in
Table 1 together with their range, resolution and validity condition.
Symbol Description
Nms Number of modules for string
1
50
Pmax Maximum nominal power of module
50
4800W
Vmpp voltage
15.00
99.99V
100.0
320.0V
Voltage on point of maximum power
15.00
99.99V
100.0
320.0V
Isc Short circuit current
0.5
15.00A
1
1W
P
max
V mpp
I mpp
P
max
0 .
01
0.01V
0.1V
0.01V
0.1V
0.01A
Voc
Voc
Isc
Vmpp
Vmpp
Impp
Impp Current on point of maximum power
Toll -
Toll +
Alpha
Beta
Gamma
Negative tolerance for Pmax provided by the module manufacturer
Positive tolerance for Pmax provided by the module manufacturer
Isc temperature coefficient
Voc temperature coefficient
Pmax temperature coefficient
0.5
15.00A
0%
25.0%
0.01A
0.1%
Isc
Impp
100*Tol
-
/Pnom< 25
0
99W
0
25%
1
0.1%
100*Tol
+
/Pnom< 25
0
99W
1
-0.100
0.100%/°C 0.001%/°C
-15.00
15.00mA/°C 0.01mA/°C
-0.99
-0.01%/°C
0.01%/°C
100*Alpha / Isc
0.1
100*Beta/Voc
0.999
-0.999
-0.001V/°C
0.001V/°C
-0.99
-0.01%/°C
0.01%/°C
NOCT Nominal working temperature of cell
Tech.
Degr
0
100°C
1°C
Effects due to PV technology
STD (standard)
CAP(capacitive eff.)
HIT (hybrid techn.)
Percentage of annual performance degradation
0.0
25.0%/yr
0.1%/yr
Table 1: Typical parameters of PV modules
CAUTION
The “Tech” item is referred to the choose of the technology of the module on test. Select the “STD” option for test on “STANDARD” PV modules,
“CAP” option for test on higher capacitive effects PV modules or HIT
(modules with HIT/HIP hybrid technology
A wrong choose of the type of technology can lead to a negative outcome of the final test
The “Degr” item is referred to the power performance degradation of a PV module/strings/field expressed in annual percentage (maximum limit set =
25%)
EN - 19
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
5.3.1. How to define a new PV module
1. Move the cursor to “DB” by means of the arrow keys
(,) and confirm with ENTER. The display shows the screen with:
The type of selected module
The parameter associated to the module (see Table 1)
15/05/10 15:34:26
T y p e : D E F A U L T
P m a x = 1 8 5
V o c = 4 4 . 5 V
V m p p = 3 7 . 5
2. Select the “DEFAULT” module by means of the arrow keys (,) and confirm with ENTER.
I m p p = 4 . 9 5
T o l l - =
Select
D B
3. Press ENTER, select the “New” command and confirm again with ENTER. Use the arrow keys (,) to scroll all internal parameters.
15/05/10 15:34:26
T y p e : D E F A U L T
P m a x = 1 8 5
V o c = 4 4 . 5 V
V m p p = 3 7 . 5
I m p p = 4 . 9 5
T o l l - =
N e w
Select
D B
4. By using the internal virtual keyboard it is possible to define the name of the module (ex: SUNPOWER 210) by
15/05/10 15:34:26
T y p e :
means the arrow keys (,,,). Press ENTER to digit the characters of the desired name.
5. Press the SAVE key to save the inserted name of the module as defined or ESC/MENU key to exit without saving.
P m a x = 1 8 5
V o c = 4 4 . 5
K E Y B O A R D
V
S U N P O W E R 2 1 0
A B C D E F G H I J K L M N O P
Q R S T U V W X Y Z - + 0 1 2 3
4 5 6 7 8 9 S P A C E D E L
6. Digit the value of each parameter of the defined module
(see Table 1) based on the manufacturer’s data sheet.
Move the cursor to the row of the parameter by means of the arrow keys (,) and set the value by means of the arrow keys (,). Press and hold (,) for a quick
15/05/10 15:34:26
T y p e : S U N P O W E R 2 1 0
P m a x =
0 W
V o c = 0 . 0 V
V m p p = V
A setting of values.
7. Confirm with SAVE and the “Saved data” message is displayed for a while. Press ESC/MENU key to exit without saving and to go back to the previous screen.
I m p p = 0 . 0 0
T o l l - =
D B
CAUTION
If the value of any parameter is unknown press and hold the HELP key for some seconds to set the default values
After pressing the SAVE key, the meter checks all the conditions shown in
Table 1 and, if one or more of these conditions do not occur, some error messages are shown by the display (§ 6.5) and the meter does not save the configuration before any error is solved. Solve the error condition before saving.
EN - 20
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
5.3.2. How to modify a PV module
1. Select the PV module to be modified from the internal database by means of arrow keys (,).
2. Press the ENTER key and select the “Modify” command by means the arrow key ().
3. Confirm the selection with ENTER.
15/05/10 15:34:26
T y p e :
S U N P O W E R 2 1 0
P m a x = 2 1 0 W
V o c = 4 7 . 7 0 V
V
A
N e w
M o d i f y
D e l e t e
D e l . A l l
S e l e c t
D B
4. By using the internal virtual keyboard it is possible to define a different name of the module by means of arrow keys (,,,). Press ENTER to digit any character of
15/05/10 15:34:26
T y p e :
S U N P O W E R 2 1 0
P m a x = 1 8 5 the desired name.
5. Press the SAVE key to save the new name of the module as defined or to access the new setting of parameters.
V o c = 4 4 . 5
K E Y B O A R D
V
S U N P O W E R 2 1 0
A B C D E F G H I J K L M N O P
Q R S T U V W X Y Z - + 0 1 2 3
4 5 6 7 8 9 S P A C E D E L
SAVE/ESC
6. Select the desired parameters to be modified by means of arrow keys (,) and change values by means of arrow displayed for a while. Press ESC/MENU key to exit without saving and to go back to the previous screen.
15/05/10 15:34:26
T y p e : S U N P O W E R 2 1 0
keys (,). Press and hold the (,) for a quick setting of values. If the value of any parameter is unknown, press and hold the HELP key for some seconds to set the default values.
7. Confirm with SAVE and the “Saved data” message is
P m a x =
2 1 0
W
V o c = 4 7 . 7 0
V m p p = 4 0 . 0 0
V
V
I s c = 5 . 7 5
A
T o l l - =
D B
5.3.3. How to delete a PV module
1. Select the PV module to be deleted from the internal database by means of arrow keys (,).
15/05/10 15:34:26
T y p e :
S U N P O W E R 2 1 0
2. Press the ENTER key and select “Delete” command by means of arrow key () to delete the selected module.
P m a x = 2 1 0
V o c = 4 7 . 7 0
W
V
3. Press the ENTER key and select “Del. All” command by
V
A means of arrow key () to delete all modules in the
N e w database (except for “DEFAULT”).
M o d i f y
4. Confirm the selection with ENTER or press ESC/MENU to exit
D e l e t e
D e l . A l l
S e l e c t
D B
CAUTION
It is not possible to modify or delete the “DEFAULT” PV module which is the standard reference factory module.
EN - 21
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
6. HOW TO OPERATE
6.1. TESTING PV SYSTEMS (SOLAR I-VW OR SOLAR I-VE)
For the sake of simplicity, further in this manual, the word “string” will be used, although often the term “photovoltaic field” would be more correct. From the point of view of the instrument, the management of a single string or of more parallel strings (photovoltaic field) is identical. Furthermore, the acronym MPPT (Multiple Power Point Tracker) shall indicate the characteristic of the DC/AC converter (inverter), capable of maximizing the DC power which can be taken from the photovoltaic field (see the § 11.2 for further details), the acronym PRp shall indicated the Performance ratio (evaluated on active powers).
CAUTION
For the evaluation of the PRP only, the measurement of the DC (voltage
and current) is not strictly necessary
Conversely it is necessary if you want to evaluate the performance of the
photovoltaic section (ndc) and DC / AC conversion (nac)
Simbol Description
PRp Performance Ratio (calculated through Active power)
Irr Irradiance
Pnom Total Nominal power of the PV section under test.
Tc Module
Te Environment
Pdc, Pdcx Total DC power measured, DC power of PV field x (x=1,2,3)
Pac, Pacx Total AC power measured, AC power of phase x (x=1,2,3,) ndc DC nac AC efficiency
Vdc, Vdcx DC Voltage, DC voltage of PV field x (x=1,2,3)
Idc, Idcx DC current, DC current of PV field x (x=1,2,3)
Vac, Vacx AC Voltage, AC voltage of PV field x (x=1,2,3)
Iac, Iacx AC current, AC current of PV field x (x=1,2,3)
Symbols description
Unit
W/m
2 kW
°C
°C kW
A
V
A kW
V
EN - 22
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
6.1.1. Testing of PV systems with single-MPPT inverter - single-phase AC output
The SOLAR I-Vw or SOLAR I-Ve meter (master) can perform check tests of efficiencies on single phase PV installations by using the irradiance and temperature probes which are connected to the remote unit SOLAR-02. This remote unit communicates with the master unit (for the synchronization and download data operations) through a wireless radiofrequency (RF) which is active with a maximum distance of about 1m between units.
CAUTION
The maximum voltage between C1, and C2 inputs is 1000VDC (for SOLAR
I-Vw) or 1500VDC (for SOLAR I-Ve) and between P1 and P2 inputs is 265V
AC rms. Do not measure voltages exceeding the limits prescribed by this manual. Should you exceed the voltage limits you could damage the instrument and/or its components or endanger your safety
In order to guarantee the operator’s safety, while making the connections, disable the system being measured by means of the switches/breakers upstream and downstream of the DC/AC converter (inverter).
Fig. 4: Connection of meter for check on single phase PV installation
1. Check and, if necessary, set the sensitivity of the reference cell on SOLAR-02 consistently with the type of PV modules which will be measured (please refer to the User Manual of
SOLAR-02)
2. We recommend making a preliminary evaluation of the value of irradiation on the surface of the PV modules being tested by means of unit SOLAR-02 (operating independently) and the reference cell
3. Switch on SOLAR I-Vw or SOLAR I-Ve. Check and if necessary modify the basic settings of meter relative to the minimum irradiance threshold, the FS of DC and AC transducer clamps, the IP and the parameters of the PV installation on test (see § 5.1.1,
§5.1.5 and §5.2)
EN - 23
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
4. In order to guarantee the operator’s safety, disable the system being measured by means of the switches/breakers upstream and downstream of the DC/AC converter (inverter)
5. In case of inverters provided with more than one power tracker (MPPT), only leave the
string corresponding to the first MPPT connected
, as shown in Fig. 4. Then, it will be necessary to repeat the operations described below by only leaving the string corresponding to the second MPPT connected, then to the third, etc.
6. Bring SOLAR I-Vw or SOLAR I-Ve and SOLAR-02 nearer (maximum distance of 1m between them). All instruments must be switched on (see the User Manual of SOLAR-2 for further details)
7. On SOLAR I-Vw or SOLAR I-Ve, press the MENU key, select the function EFF and press
ENTER
; wait for the two units to start communicating with each other. This condition is highlighted by the contemporary presence of the following indicators:
Symbol steady (not flashing) on the display of SOLAR I-Vw or SOLAR I-Ve
Symbol steady (not flashing) on the display of SOLAR-02
8.
Connect the C2 and C1 inputs respectively to the positive and negative polarity of the output string. Connect the P1 and P2 inputs to the phase and neutral conductors respecting the colors as shown in Fig. 4
9. Connect the output connector of the DC clamp to the IDC1 input.
CAUTION
BEFORE CONNECTING THE DC CLAMPS TO THE CONDUCTORS
Switch on the clamp, check the LED indicating the status of the clamp’s internal batteries (if present), select the correct range, press the ZERO key on the DC clamp and check on the display of SOLAR I-Vw or SOLAR I-Ve, the actual zeroing of the corresponding Idc value (values up to 0.02A are acceptable).
10. Connect the DC current clamp to the positive string output conductor by respecting
the direction of the arrow
on the clamp itself as indicated in Fig. 4. Position the DC clamp in point so far from the inverter and avoid that the jaws stay close to the negative conductor
11. Connect the AC current clamp to the Phase L1 conductor by respecting the direction
of the arrow
on the clamp itself as indicated in Fig. 4. Position the AC clamp as far as possible from the inverter and avoid that the jaws stay close to the neutral conductor.
Connect the output of the clamp to the IAC input of the instrument
12. Power on the system on test
EN - 24
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
13. The display shows the first screen containing the values of the electrical parameters on the DC side of the inverter.
15/05/10 15:34:26
P R p
I r r
- - -
- - -
P n o m 3 . 5 0 0
W / m 2
T c 4 5
T e 3 0
I d c 8 . 0 1 n d c
- - -
° C
P d c 3 . 1 2 5
V d c 3 8 9
A
G O – s t a r t r e c .
Select
E F F
14. By means of arrow key to access to the second screen with the values of electrical parameters of the AC side of means the search by the meter of RF connection with the remote unit SOLAR-02
15/05/10 15:34:26
k W inverter. Before to start the recording:
Verify the symbol “ ” blinking at display which
Verify that the Pac active power should be a positive
I d c 8 . 0 1 n d c - - -
V
° C
P a c 3 . 0 1 2
V
A value. In the opposite case open the AC clamp and rotate of 180 degree on the phase conductor
Verify that the AC efficiency
ac = Pac / Pdc should be a consistent value (e.g.: a situation of
ac > 1 is not physically possible)
Select
G O – s t a r t r e c
E F F
15. With remote unit always close to the main unit press
15/05/10 15:34:26
GO/STOP
key to start the recording. The “Rec. Start
waiting…
” message is shown at display of main unit and the
“HOLD” message is shown at display of SOLAR-02 more than the indication of residual time for “00” instance
P d c 3 . 1 2 5
V d c 3 8 9
I d c 8 . 0 1
- - - ° C n d c
P a c 3 . 0 1 2
V
A
R e c . S t a r t w a i t i n g …
Select
E F F
16. When the “00” instant after the press of GO/STOP key is
15/05/10 15:35:00 reached, the recording starts and the units are correctly
k W synchronized. In these conditions the message “Rec
running…
” is displayed on the main unit and the
“Recording…” message is displayed on the SOLAR-02
I d c 8 . 0 1 n d c - - -
V
° C
P a c 3 . 0 1 2
V
A
R e c . R u n n i n g …
Select
E F F
EN - 25
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
15/05/10 15:35:00 recording status by pressing the MENU key. The following information will be shown:
starting date and time of recording
the value set for the integration period
the number of periods elapsed from the beginning of the recording
the remaining memory capacity for recording
S t a r t
1 4 / 0 2 / 0 0 1 7 : 1 8 : 0 0
P e r i o d : 5 s
I P N u m b e r
R e c . T i m e
6 1
0 d 1 h
R e c . r u n n i n g
Press the ESC key to exit the screen
Select
M P P
18. Now it is possible to bring the unit SOLAR-02 near the PV strings to measure irradiation and temperature by means of the relevant probes. When the distance between unit SOLAR-02 and SOLAR I-Vw or SOLAR I-Ve does not allow the RF connection, on the display of
SOLAR-02, the symbol “ ” flashes for approx. 30s and then disappears, while SOLAR I-
Vw or SOLAR I-Ve continues searching for the connection for approx. 1 minute.
19. Position the reference cell onto the surface of the PV modules. Please refer to the relevant
User Manual for a correct assembly.
20. Put the temperature sensor in contact with the rear side of the module and fasten it with some tape; prevent touching it with your fingers (as this could alter the measure).
21. Wait some seconds in order to permit a stable measure by probes and then connect the irradiation and temperature probes respectively to the inputs PYRA/CELL and TEMP of the SOLAR-02 unit.
22. Wait for “READY” message displayed on SOLAR-02. This event indicates that the instrument has detected some data with solar irradiation > minimum limit threshold (see
§ 5.1.5)
23. Wait at least 1 minute in order to recording more valid samples
24. Disconnect the irradiation and temperature probes from unit SOLAR-02 and bring the unit near SOLAR I-Vw or SOLAR I-Ve (max distance 1m).
25. The main unit SOLAR I-Vw or SOLAR I-Ve must be in EFF mode. If there is no flashing symbol “ ”, press key to activate the RF connection search again.
26. Press the key on SOLAR-02 to active the RF connection. On the main unit the message
“ RF connection active” is displayed
27. To stop testing, press the GO/STOP key on instrument
SOLAR I-Vw or SOLAR I-Ve and confirm with ENTER that you want to stop recording.
28. The display of SOLAR I-Vw or SOLAR I-Ve will show the message “DOWNLOAD” to indicate that the data will be
15/05/10 15:35:00
9 7 1
4 5 . 1
3 0 . 5
° C
° C
0 . 9 5
W / m 2
I r r
P n o m 3 . 5 0 0
T c
T e
P d c 3 . 1 2 5
P a c 2 . 9 6 0 n a c k W n d c 0 . 8 6 transferred to the main unit during its various phases.
29. After the automatic data transfer phase, the instrument:
Do not show any results if do not exist on the PV installation a “stable irradiance” condition more than the minimum irradiance threshold.
Select
A n a l y s i s R e s u l t
E F F
Display the best performance values if during the recording, the Irradiance values reached the “stable” condition and its values were higher than the minimum irradiance threshold.
30. Press SAVE to save the results (see § 7.1) or ESC to exit the screen of the results and go back to the initial screen
EN - 26
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
6.1.2. Test PV systems with single/multi-MPPT inverter – single/three-phase AC output
The instrument SOLAR I-Vw or SOLAR I-Ve, used together with remote units SOLAR-02 and MPP300 (optional), allows testing PV systems characterized by 1 or more strings (with the same direction and inclination) and single-phase or three-phase output.
The remote unit MPP300 is capable of communicating with SOLAR I-Vw or SOLAR I-Ve
(to manage synchronization and data download) and with the remote unit SOLAR-02 (for recording irradiation and temperature values) via a wireless radiofrequency (RF) connection, which is active up to a maximum distance of 1m between the units.
Fig. 5: Connection of MPP300 for testing a single-phase PV system
EN - 27
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
Fig. 6: Connection of MPP300 for testing a three-phase PV system
CAUTION
When SOLAR I-Vw or SOLAR I-Ve is set in order to use MPP300 as a remote unit, ALL connections relevant to electrical quantities (voltages and currents) must be carried out on unit MPP300. SOLAR I-Vw or SOLAR I-
Ve must have no voltage nor current connected to its inputs.
The maximum voltage for the inputs of MPP300 is 1000VDC between inputs VDC1, VDC2, VDC3 and 600VAC between inputs VAC1, VAC2,
VAC3. Do not measure voltages exceeding the limits given in this manual.
Exceeding these limits could result in electrical shocks to the user and damage to the instrument.
In order to guarantee the operator’s safety, while making the connections, disable the system being measured by means of the switches/breakers upstream and downstream of the DC/AC converter (inverter).
1. Check and, if necessary, set the sensitivity of the reference cell on SOLAR-02 consistently with the type of PV modules which will be measured (please refer to the User Manual of
SOLAR-02).
2. We recommend making a preliminary evaluation of the value of irradiation on the surface of the PV modules being tested by means of unit SOLAR-02 (operating independently) and the reference cell
3. Switch on SOLAR I-Vw or SOLAR I-Ve, check and, if necessary, change the settings relevant to the type of remote unit, to the minimum irradiation threshold, to the full scale of the AC and DC clamps, to the integration period and to the parameters of the system being measured (see § 5.1.4, § 5.1.5, § 5.1.6, § 5.2.2).
4. In order to guarantee the operator’s safety, disable the system being measured by means of the switches/breakers upstream and downstream of the DC/AC converter (inverter).
5. Bring SOLAR I-Vw or SOLAR I-Ve, SOLAR-02 and unit MPP300 nearer (maximum distance of 1m between them). All instruments must be switched on (see the User
Manuals of SOLAR-2 and MPP300 for further details).
6. On SOLAR I-Vw or SOLAR I-Ve, press the MENU key, select the function EFF and press
ENTER
; wait for the three units to start communicating with each other. This condition is highlighted by the contemporary presence of the following indicators:
Symbol steady (not flashing) on the display of SOLAR I-Vw or SOLAR I-Ve
Symbol steady (not flashing) on the display of SOLAR-02
MASTER and REMOTE LEDs flashing green on unit MPP300
7. Connect the VDC1(+) and VDC1(-) inputs of unit MPP300 to the output terminals of the string, respecting the polarities and the colours indicated in Fig. 5 or Fig. 6
8. Repeat the operation described in the step above for other possible DC power trackers to be monitored by using the VDC2 and VDC3 inputs according to the number of DC inputs set (see § 5.2.1.1).
EN - 28
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
9. Connect the output connector of the DC clamp to the IDC1 input of unit MPP300.
CAUTION
BEFORE CONNECTING THE DC CLAMPS TO THE CONDUCTORS
Switch on the clamp, check the LED indicating the status of the clamp’s internal batteries (if present), select the correct range, press the ZERO key on the DC clamp and check on the display of SOLAR I-Vw or SOLAR I-Ve the actual zeroing of the corresponding Idc value (values up to 0.02A are acceptable).
10. Insert the DC current clamp onto the positive output conductor of the string, respecting the
direction of the arrow
found on the clamp itself as indicated in Fig. 5 or Fig. 6. Position the clamp toroid as far as possible from the inverter and from the negative output conductor of the string itself.
11. Repeat the operations described in the two steps above for other possible DC power trackers to be monitored by using the IDC2 and IDC3 inputs according to the number of DC inputs set (see § 5.2.1.1).
12. Connect the VAC1 and N inputs of unit MPP300 to the Phase and Neutral conductors respectively, respecting the polarities and the colors indicated in Fig. 5 or Fig. 6. In case of three-phase systems in which no Neutral conductor is available, connect input N to earth.
13. In case of inverter with three-phase output (see settings in § 5.2.1.1), repeat the operation described in the step above for the remaining phases by using the VAC2 and
VAC3
inputs of MPP300.
14. Connect the AC clamp to the Phase L1 conductor, respecting the direction of the arrow found on the clamp itself as indicated in Fig. 5 or Fig. 6. Position the clamp toroid as far as possible from the inverter and from the Neutral conductor. Connect the clamp output to the
IAC1
input of MPP300.
15. In case of inverter with three-phase output (see settings in § 5.2.1.1), repeat the operation described in the step above for the remaining phases by using the IAC2 and IAC3 inputs of
MPP300.
16. Restore the operation of the electrical system being measured.
17. The display of SOLAR I-Vw or SOLAR I-Ve will show the values of the general electrical parameters of the system being measured. In particular, in this screen:
Pdc = General DC power (sum of the string powers)
Pac = AC power (if single-phase) or sum of the ac powers (if three-phase)
We recommend checking that the values of the electrical
15/05/10 15:34:26
P R p
I r r
- - -
- - -
W / m 2
k W
T c
T e k W
P a c 2 . 9 6 0 n d c n a c
- - -
- - -
° C
° C
- - -
0 . 9 5 k W
G O – s t a r t r e c parameters (Pnom, Pdc, Pac) and of the ac performance
(
ac) are consistent with the system being measured (e.g.:
ac > 1 is not physically acceptable)
Select
M P P
EN - 29
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
18. On SOLAR I-Vw or SOLAR I-Ve, press key () to access the second screen which contains the values of the output DC parameters of the strings according to the number of DC inputs set (see § 5.2.1.1). In particular, in this screen:
Vdcx = DC voltage of string x
Idcx = DC current of string x
Pdx = DC power of string x.
15/05/10 15:34:26
V d c 1 4 6 0 . 1 k W
V d c 2 4 6 1 . 4
V d c 3 4 6 2 . 5
V
A
I d c 1 2 . 2 5
I d c 2 2 . 3 1 A
I d c 3 2 . 2 1
P d c 1 1 . 0 3 5
P d c 2 1 . 0 6 6
P d c 3 1 . 0 2 4 k W k W
G O – s t a r t r e c
We recommend checking that the values of the electrical
Select
M P P
parameters (Vdc, Idc, Pdc) are consistent with the system being measured.
19. On SOLAR I-Vw or SOLAR I-Ve, press key () to access the third screen which contains the values of the electrical parameters on the AC side of the inverter, consistently with the settings made in § 5.2.2 (single-phase, threephase 4 wires). In particular, in this screen:
Vacxy = AC voltage between Phase and Neutral (if single-phase) or between Phases x and y (if threephase)
Iacx = AC current of phase x
Pacx = AC power of phase x
We recommend checking that the values of the electrical parameters (Vac, Iac, Pac) are consistent with the system being measured.
15/05/10 15:34:26
V
V a c 2 3 4 0 1 . 1
V a c 3 1 4 0 0 . 1
I a c 1 4 . 2 6
A
I a c 3 4 . 2 7
P a c 1 9 8 7
W
P a c 2 9 8 6
W
G O – s t a r t r e c
Select
M P P
E x a m p l e o f a s c r e e n f o r
P V s y s t e m s w i t h t h r e e p h a s e o u t p u t
15/05/10 15:34:26 distance approx. 1m), press the GO/STOP key on SOLAR I-
Vw or SOLAR I-Ve to start testing. Consequently:
The display of SOLAR I-Vw or SOLAR I-Ve shows the message “rec. start waiting”
The display of SOLAR-02 shows the message “HOLD” and the time, expressed in seconds, remaining before the recording is started
P R p
I r r
T c
T e k W
P a c 2 . 9 6 0 n d c n a c
- - -
- - -
- - -
- - -
W / m 2
k W
° C
° C
- - -
0 . 9 5 k W
On MPP300, the STATUS LED turns on green (not
Select flashing)
R e c . r u n n i n g
M P P
Upon reaching the instant “00” after pressing the
GO/STOP
synchronized with each other. In these conditions:
key, the test is started and the three units are
The display of SOLAR I-Vw or SOLAR I-Ve shows the message “rec. running”
“Recording…”
P R p
I r r
T c
T e k W
P a c 2 . 9 6 0
The display of SOLAR-02 shows the message n a c
- - -
- - -
- - -
- - -
W / m 2
k W
° C
° C
- - -
0 . 9 5 k W
On MPP300, the STATUS LED flashes green
R e c . r u n n i n g
Select
M P P
EN - 30
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
15/05/10 15:35:00 recording status by pressing the MENU key. The following information will be shown:
Starting date and time of recording
The value set for the integration period
The number of periods elapsed from the beginning of the recording
The remaining memory capacity for recording
S t a r t
1 4 / 0 2 / 0 0 1 7 : 1 8 : 0 0
P e r i o d : 5 s
I P N u m b e r
R e c . T i m e
6 1
0 d 1 h
Select
R e c . R u n n i n g
M P P
Press the ESC key to exit the screen
23. Now it is possible to bring the unit SOLAR-02 near the PV strings to measure irradiation and temperature by means of the relevant probes. When the distance between unit SOLAR-
02 and MPP300 does not allow the RF connection, on the display of SOLAR-02 the symbol “ ” flashes for approx. 30s and then disappears. Unit MPP300 steadily searches for the RF connection with unit SOLAR-02.
24. Position the reference cell onto the surface of the PV modules. Please refer to the relevant
User Manual for a correct assembly.
25. Put the temperature sensor in contact with the rear side of the module and fasten it with some tape; prevent touching it with your fingers (as this could alter the measure).
26. Wait for a few seconds to allow the probes to reach a steady measure and then connect the irradiation probe to input PYRA/CELL and the temperature probe to input TEMP of unit
SOLAR-02.
27. Wait for the message “READY” to appear on the display of SOLAR-02 to indicate that the unit has detected the data with solar irradiation > minimum threshold set (see § 5.1.5).
28. With the message “READY” shown on the display, wait for approximately 1 minute in
order to take a certain number of samples
.
29. Disconnect the irradiation and temperature probes from unit SOLAR-02 and bring the unit near unit MPP300. Bring the main unit SOLAR I-Vw or SOLAR I-Ve near MPP300 too. The three units must be near each other (max distance 1m).
30. The main unit SOLAR I-Vw or SOLAR I-Ve must be in EFF mode; if no flashing symbol
“ ” appears, press key to activate the RF connection search again.
31. Press key on SOLAR-02 to activate the RF connection again. Consequently, the main unit will show the message “active radio connection”.
EN - 31
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
32. To stop testing, press the GO/STOP key on instrument
15/05/10 15:35:00
SOLAR I-Vw or SOLAR I-Ve and confirm pressing ENTER that you want to stop recording.
33. The display of SOLAR I-Vw or SOLAR I-Ve will show the message “DATA DOWNLOAD” to indicate that the data will be transferred to the main unit during its various phases.
34. After the automatic data transfer phase, the instrument:
Do not show any results if do not exist on the PV installation a “stable irradiance” condition more than the minimum irradiance threshold.
P R p
I r r
0 . 8 1 5
T c
T e
9 7 1
4 5 . 1
3 0 . 5
k W
° C
° C
0 . 9 4 7
W / m 2 k W
P a c 2 . 9 6 0 n a c k W n d c 0 . 8 0 6
Select
A n a l y s i s R e s u l t
E F F
Display the best performance values if during the recording, the Irradiance values reached the “stable” condition and its values were higher than the minimum irradiance threshold.
35. Press SAVE to save the results (see § 7.1) or ESC to exit the screen of the results and go back to the initial screen.
EN - 32
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
6.2. I-V CURVE MEASUREMENT
The meter can perform I-V curve tests in automatic or manual activation mode by using one of the herewith modes:
I-V curve test with Irr/Temp measurement directly performed by the instrument
I-V curve test with Irr/Temp measurement performed by use of remote unit SOLAR-02
For theoretical information on measurement see the § 11.1
6.2.1. I-V curve measurement by use of instrument
CAUTION
The maximum voltage among P1, P2, C1, and C2 inputs is 1000V DC (for I-
V400w and SOLAR I-Vw) or 1500VDC (for I-V500w and SOLAR I-Ve). Do not measure voltages exceeding the limits prescribed by this manual
The maximum current measured by the meter is 15A. Do not perform test on strings of PV modules in parallel with them
Do not perform test on PV modules or strings connected to DC/AC converter
1. Turn on the meter by pressing ON/OFF key
2. Check that option NO relative to the selection of remote unit SOLAR-02 should be set on the instrument (see § 5.1.4)
3. Check that the values set on the “Remote unit” section (see § 5.1.4) should be consistent with the reference cell depending on the kind of module/string on test
4. Check that the value set on “Irradiance” section (see § 5.1.5) should be consistent with the measurement to perform. It is suggested to perform the measurement with a threshold
700 W/m
2
in compliance to the IEC/EN60891 guideline
5. Back to the main menu and select the “I-V” option
6. Press ENTER, by means of arrow keys (,) select
15/05/10 15:34:26 item “Meas. Type ”
7. Using the key, select item “I-V Test” and confirm by pressing ENTER
V d c
I r r
= 0 . 0 V
= 0 W / m 2
T c
= - - - ° C
Module: SUNPOWER 210
I - V T e s t
S e t I V F a s t C h e c k
M e a
Select I - V
8. The instrument displays a screen similar to the one reported here to the side, where:
15/05/10 15:34:26
V d c
= 0 . 0 V
Vdc = DC output voltage from module measured between C1 and C2 inputs of meter
Irr = irradiance measured by supplied reference cell
I r r
= - - - W / m 2
T c
= A u t o
Tc = temperature of module cell. The herewith options are available for this field:
AUTO automatic mode of temperature measurement
Number MAN or AUX mode of temp. measurement
“ - - - “ AUX mode with no connected probe
Module = type of selected module
Temp = measurement mode of module temperature
Start = measurement activation mode
M o d u l e : P A N E L 0 1
Temp: Auto
Start: Manual
Select I - V
EN - 33
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
9. Press ENTER key, select the “Settings” item and confirm again with ENTER to access the next screen which allows setting the type of PV module and the number of modules relative to the PV string under test.
15/05/10 15:34:26
V d c
= 0 . 0 V
I r r
= - - - W / m 2
T c
= - - - ° C
Module: PANEL01
S e t t i n g s
M e a s . T y p e
Select I - V
10. By means of arrow keys (,) select the type of PV
15/05/10 15:34:26 module included in the database of meter (§ 5.3.1).
11. By means of arrow keys (,) select the item “N. of
Type : SUNPOWER 210
N . o f M o d . : 1 5
O p e r . y r s : 4 . 5
Mod.” and by means of arrow keys (,) set the number of modules of the PV string under test. The maximum number of modules is 50
T e m p : A u t o
S t a r t : M a n u a l
P m a x = 2 1 0
V o c =
12. By means of arrow keys (,) select the “Oper. yrs” item and by means of arrow keys (,) set the service operating years of the PV module/string/field from start installation
V m p p = 4 0 . 0 0
I s c = 5 . 7 5
I m p p = 5 . 2 5
(see § 6.2.3). The maximum set value is 25.0 (0.5 = 6
S E T months)
13. By means of arrow keys (,) select the type of PV module included in the database of meter (§ 5.3.1). By means of arrow keys (,) select the “Temp” item and by means of arrow keys (,) choose the type of module temperature among the modes listed:
Auto” automatic measurement performed by the meter depending on the measured value of open voltage of the modules (recommended method)
“Manual” setting of a module temperature value by the operator on the corresponding “Value” field
“Aux” setting of temperature of module performed by auxiliary probe
14. By means of arrow keys (,) select the “Start” item and by means of arrow keys
(,) choose the type of measurement activation between the modes listed:
Auto” test automatically activated by the instrument with stable values of voltage among inputs for at least 1s and with values within the measurement range (see §
10.2)
“Manual” test activated from the user by press of GO/STOP key
15. Confirm with SAVE key or press ESC/MENU key to exit without saving
16. Mount the M304 (inclinometer) supplied accessory, place and hold it on the plane of module.
Verify that the sun shadow falls on the disc within the “limit internal circle” on the disc itself (see Fig. 7b). If this is not the case, the incidence angle between sun rays and the module surface is too high and not complying with the test conditions declared by the module manufacturer. As a consequence the measurements performed by the meter are not ok and all measurements steps shall be repeated at another daytime
17. Fix the support of the cell to the module by using the supplied set of screws and mount the reference cell on it possibly with output terminals downwards with respect to the
module
. Rotate the cell upwards to lean it on the small wing of the support so as the cell is perfectly parallel to the plane of module, then fix it with the relevant screws supplied
18. Connect the output of reference cell (depending on the type of module under test) of reference cell at the IRR. input of meter by using the cable supplied with the cell
EN - 34
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
19. Connect, if used, the auxiliary temperature probe (optional accessory) to the AUX input of meter and on the rear of module, fixing it with adhesive tape
20. Connect the meter to the module/string under test as shown in Fig. 7a. In particular, connect the negative output pole of module/string to P1, C1 input terminals and the positive output pole of module/string to P2, C2 input terminals. In Auto activation mode is recommended the use of KITKELVIN optional accessory
Fig. 7a: Instrument connection to module/string FV Fig. 7b: Positioning of inclinometer M304
CAUTION
The method used by the instrument meter for output VDC and IDC measurements of PV module/string is the “4-wire”. Therefore, it is possible to use also test cables connected to P1, C1, P2, C2 inputs of different length without performing any calibration of cable resistance. It is recommended to
use cables with a §
4mm
2
21. After connection to the plant, the real time values of
15/05/10 15:34:26 parameters are shown by meter:
Vdc = DC output voltage of PV module/string
Irr = Irradiance value measured by sensor
Tc = Value of module temperature
V d c
I r r
= 3 6 7 V
= 1 0 4 5 W / m 2
T c
= 4 5 ° C
Module: SUNPOWER 210
Select
I - V
CAUTION
At the press of GO/STOP key some different error messages can be displayed by the meter (see § 6.4) and, due of this, the test cannot start.
Check and remove, if possible, this error message before start the test
22. Press GO/STOP key (in Manual activation mode) to start power during test.
15/05/10 15:34:26 the test. If none of the previous error messages are
V d c
= 3 6 7 V detected, the message “Measuring…” is shown on the display for some seconds, depending on the level of
I r r
= 1 0 4 5 W / m 2
Tc
= 45 °C
Module: SUNPOWER 210
Select
Measuring…
I - V
EN - 35
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
23. At the end of test, the values of typical parameters (all referred to a single module in standard conditions STC)
15/05/10 15:34:26 are shown by the meter together with the final response (§
6.2.3) based on the automatic calculation performed and relevant to:
I m p p = 4 . 7 A
Translation of I-V curve to STC condition
Verification of tolerance % of the maximum power declared by the manufacturer.
P m a x = 2 0 0
F F =
D P m a x = 2 . 1
Results @ STC – Response: OK
Select I - V
CAUTION
The instrument refer all the values of parameters to one only module at the STC condition
The total voltage of the string, obtained at OPC, is divided for the number of module of it. Considering this “average” value, together with the measured current, the instrument perform the I-V curve @ OPC calculation which is the translated at the STC condition
24. Press ENTER for the display of measurement results both in numeric and graphical mode referred to STC or OPC conditions (see § 6.2.3).
25. Use the arrow key for the selection of table or graph
15/05/10 15:34:26
I m p p = 4 . 7 A visualization.
26. Press SAVE key to store the result of test (see § 7.2) or
P m a x = 2 0 0
F F =
ESC/MENU
key to exit without saving and to go back to the main menu
D P m a x = 2 . 1
T a b l e
27. The “Auto” activation mode require to disconnect and
G r a p h
– Response: OK
Select reconnect the test leads in order to start automatically a
I - V new measurement
EN - 36
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
6.2.2. I-V curve measurement by use of remote unit SOLAR-02
The I-V curve test with Irr/Temp measurement by use of remote unit SOLAR-02 can be performed in one of the herewith mode:
Remote unit SOLAR-02 in RF connection with main unit
Remote unit SOLAR-02 in synchro recording (no RF connection with main unit)
6.2.2.1. I-V curve test by remote unit SOLAR-02 in RF connection
This mode is possible ONLY if the distance between the main unit and the remote unit
SOLAR-02 can establish always a stable RF connection. This distance limit is influenced by obstacles, air humidity, etc., so it is considered valid about some meters.
CAUTION
The maximum voltage among P1, P2, C1, and C2 inputs is 1000V DC (for I-
V400w and SOLAR I-Vw) or 1500VDC (for I-V500w and SOLAR I-Ve). Do not measure voltages exceeding the limits prescribed by this manual
The maximum current measured by the meter is 15A. Do not perform test on strings of PV modules in parallel with them
Do not perform test on PV modules or strings connected to DC/AC converter
1. Turn on the meter by pressing ON/OFF key
2. Check that option YES relative to the selection of remote unit SOLAR-02 should be set on the instrument (see § 5.1.4)
3. Check that the values set on the “Remote unit” section (see § 5.1.4) should be consistent with the reference cell depending on the kind of module/string on test
4. Check that the value set on “Irradiance” section (see § 5.1.5) should be consistent with the measurement to perform. It is suggested to perform the measurement with a threshold
700 W/m
2
in compliance to the IEC/EN60891 guideline
5. Back to the main menu and select the “I-V” option
6. Press ENTER, by means of arrow keys (,) select item “Meas. Type ”
7. Using the key, select item “I-V Test” and confirm by
15/05/10 15:34:26
V d c
= 0 . 0 V pressing ENTER
I r r
= 0 W / m 2
T c
= - - - ° C
Modulo: SUNPOWER 210
I - V T e s t
S e t I V F a s t C h e c k
M e a
Select I - V
8. Switch on the remote unit SOLAR-02 and wait that “Radio connection active” message is displayed
EN - 37
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
9. The instrument displays :
Vdc = DC output voltage from module/string
Irr = irradiance measured by supplied reference cell
Tc = temperature of module. The herewith options are available for this field:
AUTO automatic mode
Number MAN or AUX mode
“ - - - “ AUX mode with no connected probe
Module = type of selected module
Temp = measurement mode of module temperature
Start = measurement activation mode
The symbol “
” fixed at display means that a stable
RF connection is established with the remote unit
10. Press ENTER key, select the “Settings” item and confirm again with ENTER to access the next screen which allows setting the type of PV module and the number of modules relative to the PV string under test.
15/05/10 15:34:26
V d c
= 0 . 0 V
I r r
= - - - W / m 2
T c
= A u t o ° C
Module: PANEL01
Temp: Auto
Start: Manual
Select
I – V
15/05/10 15:34:26
V d c
= 0 . 0 V
I r r
= - - - W / m 2
T c
= - - - ° C
Module: PANEL01
S e t t i n g s
M e a s . T y p e
Select
I - V
11. By means of arrow keys (,) select the type of PV
15/05/10 15:34:26 module included in the database of meter (§ 5.3.1)
Type : SUNPOWER 210
12. By means of arrow keys (,) select the item “N. of
Mod.” and by means of arrow keys (,) set the number
N . o f M o d . : 1 5
O p e r . y r s 4 . 5 of modules of the PV string under test. The maximum
T e m p : A u t o
S t a r t : M a n u a l number of modules is 50
P m a x = 2 1 0
V o c =
13. By means of arrow keys (,) select the “Oper. yrs” item and by means of arrow keys (,) set the service operating years of the PV module/string/field from start installation
V m p p = 4 0 . 0 0
I s c = 5 . 7 5
I m p p = 5 . 2 5
(see § 6.2.3). Maximum set value is 25.0 (0.5 = 6 months)
S E T
14. By means of arrow keys (,) select the type of PV module included in the database of meter (§ 5.3.1). By means of arrow keys (,) select the “Temp” item and by means of arrow keys (,) choose the type of module temperature among the modes:
Auto” automatic measurement performed by the meter depending on the measured value of open voltage of modules (recommended method)
“Manual” setting of a module temperature value by the operator on the corresponding “Value” field
“Aux” setting of temperature of module performed by auxiliary probe
15. By means of arrow keys (,) select the “Start” item and by means of arrow keys
(,) choose the type of measurement activation between the modes listed:
Auto” test automatically activated by the instrument with stable values of voltage among inputs for at least 1s and with values within the measurement range (see §
10.2)
“Manual” test activated from the user by press of GO/STOP key
16. Confirm with SAVE key or press ESC/MENU key to exit without saving
EN - 38
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
17. Mount the M304 (inclinometer) supplied accessory, place and hold it on the plane of module.
Verify that the sun shadow falls on the disc within the “limit internal circle” on the disc itself (see Fig. 8). If this is not the case, the incidence angle between sun rays and the module surface is too high and not complying with the test conditions declared by the module manufacturer. As a consequence the measurements performed by the meter are not ok and all measurements steps shall be repeated at another daytime
18. Fix the support of the cell to the module by using the supplied set of screws and mount the reference cell on it possibly with output terminals downwards with respect to the
module
. Rotate the cell upwards to lean it on the small wing of the support so as the cell is perfectly parallel to the plane of module, then fix it with the relevant screws supplied
19. Connect the output of reference cell (depending on the type of module under test) at the PYRA/CELL input of remote unit SOLAR-02 by using the supplied cable
20. Connect, if used, the auxiliary temperature probe to the TEMP input of remote unit
SOLAR-02 and on the rear of module, fixing it with adhesive tape
21. Connect the meter to the module/string under test as shown in Fig.9. In particular, connect the negative output pole of module/string to P1, C1 input terminals and the positive output pole of module/string to P2, C2 input terminals. In Auto activation mode is recommended the use of KITKELVIN optional accessory
Fig. 8: Positioning the M304 inclinometer
Fig. 9: Connection of units to the PV module/string
EN - 39
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
22. The instrument displays:
Vdc = DC output voltage from module/string.
Irr = irradiance measured by supplied reference cell
Tc = temperature of module. The herewith options are available for this field:
AUTO automatic mode of temperature measurement
Number MAN or AUX mode of temp. measurement
“ - - - “ AUX mode with no connected probe
Module = type of selected module
Temp = measurement mode of module temperature
Start = measurement activation mode
The symbol “
” fixed at display means that a stable
RF connection is established with the SOLAR-02
15/05/10 15:34:26
V d c
= 3 6 7 V
I r r
= 1 0 4 5 W / m 2
T c
= 4 5 ° C
Module: SUNPOWER 210
Temp: Auto
Start: Manual
Select
I – V
CAUTION
At the press of GO/STOP key some different error messages can be displayed by the meter (see § 6.3) and, due of this, the test cannot start.
Check and remove, if possible, this error message before start the test
23. Press GO/STOP key (in Manual activation mode) to start
15/05/10 15:34:26 the test. If none of the previous error messages are
V d c
= 3 6 7 V detected, the message “Measuring…” is shown on the display for some seconds, depending on the level of
I r r
= 1 0 4 5 W / m 2 power during test.
Tc
= 45 °C
Module: SUNPOWER 210
Measuring…
Select I - V
24. At the end of test, the values of typical parameters (all referred to a single module in standard conditions STC) are shown by the meter together with the final response (§
6.2.3) based on the automatic calculation performed and
15/05/10 15:34:26
I m p p = 4 . 7 A relevant to:
Translation of I-V curve to STC condition
P m a x = 2 0 0
F F =
Verification of tolerance % of the maximum power
D P m a x = 2 . 1 declared by the manufacturer.
Results @ STC – Response: OK
Select
I - V
25. Press ENTER for the visualization of measurement results referred at STC or OPC conditions both in numeric and graphical mode (§ 6.2.3)
26. Use the arrow key for the selection of table or graph
15/05/10 15:34:26
I m p p = 4 . 7 A visualization
27. Press SAVE key to store the result of test (see § 7.2) or
ESC/MENU
key to exit without saving and to go back to
P m a x = 2 0 0
F F =
D P m a x = 2 . 1
T a b l e the main menu
28. The “Auto” activation mode require to disconnect and
G r a p h
– Response: OK
Select I - V reconnect the test leads in order to start automatically a new measurement
EN - 40
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
6.2.2.2. I-V curve test by remote unit SOLAR-02 in synchro recording
This mode, which consists in an independent recording of the irradiance and temperature parameters by the remote unit SOLAR-02, permits to perform I-V curve test also with wide distance between the instrument and SOLAR-02 without any connection between them.
On the other hand, due to the impossibility for the meter to have immediately the values of irradiance/temperature, it is necessary wait the transfer of data from remote unit SOLAR-
02 to main unit in order to have all the available results.
CAUTION
The maximum voltage among P1, P2, C1, and C2 inputs is 1000V DC (for I-
V400w and SOLAR I-Vw) or 1500VDC (for I-V500w and SOLAR I-Ve). Do not measure voltages exceeding the limits prescribed by this manual
The maximum current measured by the meter is 15A. Do not perform test on strings of PV modules in parallel with them
Do not perform test on PV modules or strings connected to DC/AC converter
1. Turn on the meter by pressing ON/OFF key
2. Check that option YES relative to the selection of remote unit SOLAR-02 should be set on the instrument (see § 5.1.4)
3. Check that the value set on “Irradiance” section (see § 5.1.5) should be consistent with the measurement to perform. It is suggested to perform the measurement with a threshold
700 W/m
2
in compliance to the IEC/EN60891 guideline
4. Back to the main menu and select the “I-V” measurement mode
5. Press ENTER, by means of arrow keys (,) select item “Meas. Type ”
15/05/10 15:34:26
V d c
= 0 . 0 V
6. Using the key, select item “I-V Test” and confirm by pressing ENTER
I r r
= 0 W / m 2
T c
= - - - ° C
Module: SUNPOWER 210
I - V T e s t
S e t I V F a s t C h e c k
M e a
Select
I - V
7. Switch on the remote unit SOLAR-02 and wait that “Radio connection active” is displayed
8. Check the status of internal batteries of SOLAR-02 (the “ displayed)
” symbol should be not
9. Check that the values of sensitivity and temperature coefficient set on the SOLAR-02 section should be consistent with the reference cell depending on the kind of module/string on test (see user manual of SOLAR-02)
EN - 41
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
10. The instrument displays:
Vdc = DC output voltage from module/string
Irr = irradiance measured by supplied reference cell
Tc = temperature of module. The herewith options are available for this field:
AUTO automatic mode of temperature measurement
Number MAN or AUX mode of temp. measurement
“ - - - “ AUX mode with no connected probe
Module = type of selected module
Temp = measurement mode of module temperature
Start = measurement activation mode
The symbol “
” fixed at display means that a stable
RF connection is established with the SOLAR-02
11. Press ENTER key, select the “Settings” item and confirm again with ENTER to access the next screen which allows setting the type of PV module and the number of modules relative to the PV string under test.
15/05/10 15:34:26
V d c
= 0 . 0 V
I r r
= - - - W / m 2
T c
= A u t o ° C
Module: PANEL01
Temp: Auto
Start: Manual
Select
I – V
15/05/10 15:34:26
V d c
= 0 . 0 V
I r r
= - - - W / m 2
T c
= - - - ° C
Module: PANEL01
S e t t i n g s
M e a s . T y p e
Select
I - V
12. By means of arrow keys (,) select the type of PV
15/05/10 15:34:26 module included in the database of meter (§ 5.3.1)
Type : SUNPOWER 210
13. By means of arrow keys (,) select the item “N. of
Mod.” and by means of arrow keys (,) set the number
N . o f M o d . : 1 5
O p e r . y r s 4 . 5
T e m p : A u t o of modules of the PV string under test. The maximum
S t a r t : : M a n u a l
P m a x = 2 1 0 number of modules is 50
14. By means of arrow keys (,) select the “Oper. yrs” item and by means of arrow keys (,) set the service operating years of the PV module/string/field from start installation
4 7 . 7 0
V m p p = 4 0 . 0 0
I s c = 5 . 7 5
I m p p = 5 . 2 5
(see § 6.2.3). Maximum set value is 25.0 (0.5 = 6 months)
S E T
15. By means of arrow keys (,) select the type of PV module included in the database of meter (§ 5.3.1). By means of arrow keys (,) select the “Temp” item and by means of arrow keys (,) choose the type of module temperature among the modes:
Auto” automatic measurement performed by the meter depending on the measured value of open voltage of modules (recommended method)
“Manual” setting of a module temperature value by the operator on the corresponding “Value” field
“Aux” setting of temperature of module performed by auxiliary probe
16. By means of arrow keys (,) select the “Start” item and by means of arrow keys
(,) choose the type of measurement activation between the modes listed:
Auto” test automatically activated by the instrument with stable values of voltage among inputs for at least 1s and with values within the measurement range (see §
10.2)
“Manual” test activated from the user by press of GO/STOP key
17. Confirm with SAVE key or press ESC/MENU key to exit without saving
EN - 42
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
18. Mount the M304 (inclinometer) supplied accessory, place and hold it on the plane of module.
Verify that the sun shadow falls on the disc within the “limit internal circle” on the disc itself (see Fig. 10). If this is not the case, the incidence angle between sun rays and the module surface is too high and not complying with the test conditions declared by the module manufacturer. As a consequence the measurements performed by the meter are not ok and all measurements steps shall be repeated at another daytime
19. Fix the support of the cell to the module by using the supplied set of screws and mount the reference cell on it possibly with output terminals downwards with respect to the
module
. Rotate the cell upwards to lean it on the small wing of the support so as the cell is perfectly parallel to the plane of module, then fix it with the relevant screws supplied
20. Connect the output of reference cell (depending on the type of module under test) at the PYRA/CELL input of remote unit SOLAR-02 by using the supplied cable
21. Connect, if used, the auxiliary temperature probe to the TEMP input of remote unit
SOLAR-02 and on the rear of module, fixing it with adhesive tape
22. Connect the meter to the module/string under test as shown in Fig.11. In particular, connect the negative output pole of module/string to P1, C1 input terminals and the positive output pole of module/string to P2, C2 input terminals. In Auto activation mode is recommended the use of KITKELVIN optional accessory
Fig. 10: Positioning the M304 inclinometer
Fig.11: Connection of units to the PV module/string
EN - 43
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
23. The instrument displays:
Vdc = DC output voltage from module/string
Irr = irradiance measured by supplied reference cell
Tc = temperature of module. The herewith options are available for this field:
15/05/10 15:34:26
V d c
I r r
= 3 6 7 V
= 1 0 4 5 W / m 2
T c
= 4 5 ° C
AUTO automatic mode of temperature measurement
Number MAN or AUX mode of temp. measurement
“ - - - “ AUX mode with no connected probe
Module = type of selected module
Temp = measurement mode of module temperature
Start = measurement activation mode
The symbol “
” fixed at display means that a stable
RF connection is established with the remote unit
SOLAR-02
Module: SUNPOWER 210
Temp: Auto
Start: Manual
Select
I – V
24. Press ENTER key, select the “Activate Rec.” item and 15/05/10 15:34:26 confirm with ENTER. The message “Remote unit in
V d c
= 0 . 0 V recording…” is displayed on the meter. In this condition is possible to start the I-V curve tests always considering
I r r
= - - - W / m 2 that the complete results will be available only after the stop of recording of SOLAR-02 and transfer them from the same remote unit
T c
= - - - ° C
A c t i v a t e R e c .
S e t t i n g s
M e a s . T y p e
Select
I – V
CAUTION
At the press of GO/STOP key some different error messages can be displayed by the meter (see § 6.4) and, due of this, the test cannot start.
Check and remove, if possible, this error message before start the test
25. Press GO/STOP key (in Manual activation mode) to start power during test.
15/05/10 15:34:26 the test. If none of the previous error messages are
V d c
= 3 6 7 V detected, the message “Measuring…” is shown on the display for some seconds, depending on the level of
I r r
= - - - W / m 2
Tc
= Auto °C
Module: SUNPOWER 210
Select
Measuring…
I - V
EN - 44
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
26. At the end of test, based on the descriptions of previous
15/05/10 15:34:26 points, the message “STC Data available after Stop rec” is displayed and then the results at OPC conditions are displayed
1 5 . 2
V
I m p p = 4 . 7
I s c =
F F =
I r r =
T c =
A
P m a x = 2 0 0
7 7 . 1
- - - W / m
2
Select
Results @ OPC
I - V
27. Press ENTER for the visualization of measurement results both in numeric and graphical mode (§ 6.2.3)
28. Use the arrow key for the selection of table or graph
15/05/10 15:34:26 visualization
29. Press SAVE key to store the result of test or ESC/MENU key to exit without saving and to go back to the main screen
30. The “Auto” activation mode require to disconnect and reconnect the test leads in order to start automatically a new measurement
I m p p = 4 . 7
I s c =
P m a x = 2 0 0
F F =
I r r
=
T a b l e
- - -
%
W / m
2
G r a p h
Select I - V
31. At the end of I-V test, press the ENTER key, select the
15/05/10 15:34:26
“Stop recording” item and confirm with ENTER. The messages “Wait…” and then “STC Data available in MEM” are displayed
V d c
= 0 . 0 V
I r r
= - - - W / m 2
32. In this conditions, for the I-V tests performed with irradiance values over the limit threshold and stable, the correspondent STC results will be available inside memory
T c
= A u t o ° C
S t o p R e g .
S e t t i n g s
M e a s . T y p e
Select I - V
33. The message “Download data” is displayed on the instrument and, at the same time, the message “Send” is displayed on SOLAR-02 which means the transfer of irradiance values and, if necessary, the module temperature values on the main unit
34. At the end of data transfer the instrument automatically join the average value of irradiance (detected with IP = 5s) to each I-V test performed and, correspondingly, will calculate the results translated to STC conditions and the Yes/No response for each test. The complete results are recalled at display as described in § 7.3.2. See the §
6.2.3 for the meaning of measurement results
EN - 45
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
6.2.3. Meaning of measurement results
The parameter measured and calculated by meter have the following meaning:
Parameter Description
Pmax Maximum power of module measured by the meter
DPmax
FF
Difference % of maximum measured power from nominal power (@ STC)
Fill Factor %
Vmpp Voltage on point of maximum power
Isc Short circuit current
Impp Current on point of maximum power
Table 2: List of parameter measured by the meter
Where:
DP
MAX
%
100
P
MAX
Nom
P
Age
Nom
P
Age
check parameter which defines the final test response
P
Nom
Age
P
Nom
1
Oper
.
Yrs
YearDegr
100
%
P
Nom
= nominal power of module
FF = 100 x [(Vmpp x Impp ) / (Voc x Isc)] = Fill Factor represents a kind of “efficiency” of module/string as it is a ratio between the maximum power measured and the open power
The meter gives the following final response:
Response Condition Note
OK
- Tol
(-)
+
Meter
Meas
Tol
(+)
-
Meter
(1)
OK*
NO OK*
The previous relation (1) is not verified but the following is valid:
- Tol
(-)
Meas
Tol
(+)
The relations (1) and (2) are not verified but the following is valid:
- Tol
(-)
-
Meter
Meas
Tol
(+)
+
Meter
(2)
(3)
NO OK
None of the relations (1), (2) and (3) are verified (4) where:
Tol
(-)
= Tol
(-)
(%)*Pnom
Negative tolerance in absolute value, declared by manufacturer
Tol
(+)
= Tol
(+)
(%)*Pnom
Positive tolerance in absolute value, declared by manufacturer
Meas
= Pmax – Pnom DPmax which defines the difference between measured values and declared values
Meter
Absolute error of the measurement chain (meter + transducers in the point of measure) considering the error % and the declared dgt
(1) OK Positive response of test also considering the error of measurement chain
(2) OK* Response of test positive less than the error of measurement chain
(3) NO OK* Response of test negative less than the error of measurement chain
(4) NO OK Negative response of test also considering the error of measurement chain
EN - 46
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
6.3. QUICK CHECK ON PV MODULES AND STRINGS (IVCK) information
This function permits to carry out a quick Test of a panel/string just measuring open circuit voltage and short circuit current in compliance with IEC/EN62446. Consequently both irradiance and temperature module values can be measured (using corresponding probes).
Irradiation measurement can be executed only adopting one of the following modes:
Irradiance sensor directly connected to I-V400w, I-V500w , SOLAR I-Vw or SOLAR I-Ve
Irradiance sensor connected to SOLAR-02 through RF to I-V400w, I-V500w , SOLAR I-
Vw or SOLAR I-Ve
Irradiance measurements are always effected in real time, it is impossible to start a “remote” recording of irradiance values through SOLAR-02
If minimum IV irradiance threshold is set (see § 5.1.5):
=
the instrument does not control the presence of cell, irradiance changes, number of modules and does not display error messages should it be impossible to calculate Voc and Isc values transposed to Voc and Isc under STC. This mode is suitable to effect a test session on a considerable number of strings as quickly as possible.
> 0 (suggested > 700) the instrument carries out all tests provided for I-V, manages all conditions and error messages for I-V test (wrong num. Mod., Temp. out of range, cell presence, Min Irr, etc..) and calculates values of Voc and Isc under STC. This mode is recommended whenever thorough tests are to be effected on modules/strings under test.
In general the result page will include:
Description of the module
Irradiance and temperature values (if available)
Average values of Voc and Isc calculated as average of the corresponding values under
OPC on the last 10 tests saved into memory. If the number of tests is < 10 the average is calculated on the number of available tests. The first test will display dashes in the field
“average values” as there are no previous tests on which average can be calculated.
The values of Voc and Isc measured under OPC as well as any partial outcome (the latter available only if STC values are not available) obtained through comparison with average values.
The values of Voc and Isc calculated under STC (if available) and any partial outcome obtained comparing the values calculated under STC with the nominal ones (inserted in
DB modules).
The overall test outcome (OK(NO). The overall result will be calculated based on the partial results.
Based on partial outcome under STC (if these are available)
Based on partial outcome under OPC (if STC values are not available)
The instrument will not display any overall outcome if no partial result is available.
In case of negative outcome it is advisable to effect measurement of I-V feature (see § 6.2) in order to thoroughly analysis of the module/string under test.
EN - 47
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
settings
1. Turn the instrument on by pressing ON/OFF key.
2. Press ESC/MENU to display the main menu.
3. Select “I-V” and press ENTER to enter the FV module performance measurement section.
15/05/10 15:34:26 4. In case the current mode is not IVCK, press ENTER, select “Measurement ” by means of arrow keys
(,).Use the arrow key to enter the internal submenu, select “IV Check” and confirm by ENTER to start up the initial screen of rapid test IVCK.
V d c
I r r
T c
= 0 . 0 V
= 0 W / m 2
= - - - ° C
Module: SUNPOWER 210
S e t R s T e s t
M e a I - V F a s t C h e c k
Select I - V
5. Press ENTER, select “Settings” and confirm by pressing ENTER again to enter the following screen where it is possible to set module type and number of modules making the string under test.
A v g R e s e t
S e t t i n g s
M e a s T y p e
I r r 9 8 0 W / m 2
T c ( A U T O ) 4 9 ° C
V o c M e d @ O P C
I s c M e d @ O P C
V o c @ O P C
I s c @ O P C
6 4 7
V
5 . 4 3
A
6 4 6
V
- - -
A
V
A
M o d u l e : S U N P W R 2 1 0
Select
I V C K
6. Use the arrow keys ( , ) to select the module type among the ones stored inside the instrument’s database (see § 5.3.1)
7. Use the arrow keys (,), select “Mod. per Str” and use the arrow keys ( , ) to enter the number of
15/05/10 15:34:26
Tipo : SUNPOWER 210
M o d . x S t r : 1 5
T e m p :
V a l u e
T o l V o c
T o l I s c
: 5 1 ° C
:
:
3 % ( + 4 % )
3 % ( + 4 % ) modules for the string under test. The maximum number of modules to be set is 50
8. Use the arrow keys (,), select “Temp” and use the arrow keys ( , ) to select the measurement type of
V o c
I s c
=
=
6 4 . 7 %
6 . 2 0 %
S E T T I N G
the module temperature among the modes:
“Auto” automatic measurement carried out according to the measured open circuit voltage value of modules (recommended method)
Manual entering the value of module temperature in the corresponding field “Value”
Aux temperature measurement with probe
9. Use the arrow keys (,), select “Tol Voc” and “Tol Isc” then use the arrow keys ( , ) to set the tolerance values for open circuit voltage (Voc) and short circuit current (Isc) provided by the module manufacturer (values allowed: +0% .. +25%). Please note that the instrument’s uncertainty is indicated in brackets beside the above mentioned values and it will be added to the value entered to get the outcome (OK/NO).
10. By means of arrow keys (,) select the “Start” item and by means of arrow keys
(,) choose the type of measurement activation between the modes listed:
“Auto” test automatically activated by the instrument with stable values of voltage among inputs for at least 1s and with values within range (see § 10.2)
“Manual” test activated from the user by press of GO/STOP key
11. Press SAVE to store the measurements effected or ESC/MENU to quit without saving.
EN - 48
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
6.3.3. Fast check IVCK without irradiance measurement
CAUTION
The maximum voltage among inputs P1, P2, C1 and C2 is 1000V DC (for I-
V400w and SOLAR I-Vw) or 1500V (for I-V500w and SOLAR I-Ve). Do not measure voltages exceeding the limits expressed in § 10.2
The maximum current allowable for the instrument is 15A. Do not effect tests in parallel on FV module strings.
Never effect tests on modules or FV strings connected to DC/AC converter.
1. Turn on the instrument pressing ON/OFF
2. Check that remote unit SOLAR-02 is not selected (see § 5.1.4– setting NO)
3. Check that the value of minimum irradiance set in the section “Irradiance” (see §5.1.5) is equal to 0
4. Go back to main MENU and select “I-V”
5. Select “I-V” and press ENTER to enter the section of FV module performance measurement. Generally when selecting I-V on general menu the instrument automatically shows the last mode under use (I-V or IVCK).
6. In case the current mode is not IVCK, press ENTER,
ENTER
to start up the initial screen of fast check IVCK.
15/05/10 15:34:26 select “Measurement ” by means of arrow keys
(,). Use the arrow key to enter the internal
V d c
= 0 . 0 V submenu, select “IV Check” and confirm pressing
I r r
= 0 W / m 2
T c
= - - - ° C
Module: SUNPOWER 210
I m p I V C h e c k
M i s
Selection
7. Check the preliminary settings based on descriptions of § 6.3.2
I - V
8. Connect the instrument to the module/string under test as shown in the following picture. In particular connect the negative pole of the module/string to terminals P1, C1 and it positive pole to terminals P2, C2. In Auto activation mode is recommended the use of KITKELVIN optional accessory
CAPTION:
P1: Black cable
P2: Blue cable
C1: Green cable
C2: Red cable
1: Module or FV string
Fig. 10: Connection to module/FV string for IVCK tests without irradiance measurement
EN - 49
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
9. The initial screen under IVCK mode displays the values of:
Module under use
The average values of Voc and Isc under OPC conditions.
The value of Voc measured under OPC conditions
15/05/10 15:34:26
M o d u l e :
S U N P W R 2 1 0
I r r . - - - W / m 2
T c ( A U T O ) - - - ° C
V o c M e d @ O P C 6 4 7 V
I s c M e d @ O P C 5 . 4 3 A
V o c @ O P C 6 4 6 V
I s c @ O P C - - - A
V o c @ S T C - - - V
I s c @ S T C - - - A
Selection
I V C K
CAUTION
When pressing the GO/STOP key the instrument can provide different error messages (see §6.3) and, as direct consequence, not carry out the test. Check and eliminate, if possible, causes of the problems before continuing the test.
10. Press GO/STOP (in Manual activation mode) to start the test. In case of no error conditions, the instrument displays the screen with results as shown sideways. It displays as follows:
The module under use
The average values of Voc and Isc under OPC conditions
15/05/10 15:34:26
M o d u l e :
I r r - - - W / m 2
T c ( A U T O )
V o c M e d @ O P C
I s c M e d @ O P C
V o c @ O P C
- - ° C
6 4 7
5 . 4 3
6 4 7
5 . 3 5
V
A
V O K
A O K
I s c @ S T C - - A
S U N P W R 2 1 0
I s c @ O P C
V o c @ S T C - - - V
The values of Voc and Isc measured under OPC and relative partial outcome resulting after comparing the average values. In general:
Selection
OUTCOME: OK
I V C K
O
utcome
Voc
@
OPC
OK
if 100
VocAvg
@
OPC
Voc
@
OPC
VocAvg
@
OPC
Tol Voc
4%
O
utcome
Isc
@
OPC
OK
if 100
IsAvg
@
OPC
Isc
@
OPC
Tol Isc
4%
IscAvg
@
OPC
The overall value of outcome: o
OK: if all results under OPC are OK, o
NO if one of the results under OPC is NO
11. Press SAVE to store the test result into the instrument’s memory (see §7.2) or press
ESC/MENU
to quit without saving and go back to main measurement screen
12. The “Auto” activation mode require to disconnect and reconnect the test leads in order to start automatically a new measurement
NOTES on displayed AVERAGE VALUES
The average values of Voc and Isc are displayed on the outcome page. Such values include the average values of Voc and Isc under OPC conditions calculated as mean of the last 10 tests previously saved. If the user effected and saved a number of tests <10 or he reset the average values (see 6.3.5) the average value displayed during test N+1 will be calculated on N available values.
EN - 50
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
6.3.4. Fast check IVCK with irradiance measurement
CAUTION
The maximum voltage among inputs P1, P2, C1 and C2 is 1000V DC (for I-
V400w and SOLAR I-Vw) or 1500V (for I-V500w and SOLAR I-Ve). Do not measure voltages exceeding the limits expressed in § 10.2
The maximum current allowable for the instrument is 15A. Do not effect tests in parallel on FV module strings.
Never effect tests on modules or FV strings connected to DC/AC converter.
1. Turn on the instrument pressing ON/OFF
2. Irradiance measurement can be carried out just through the following modes:
Measurement through reference cell directly connected to I I-V400w, I-V500w, SOLAR
I-Vw or SOLAR I-Ve
Measurement through reference cell connected to SOLAR-02 under RF to I-V400w, I-
V500w, SOLAR I-Vw or SOLAR I-Ve
Check that the remote unit SOLAR-02 setting is consistent with the type of measurement to be effected (see §5.1.4).
3. Check the minimum irradiance value set in the “irradiance” section (see § 5.1.5).
4. Go back to main MENU and select “I-V”
5. Select “I-V” and press ENTER to enter the FV module performance measurement section.
Generally when selecting I-V on general menu the instrument automatically shows the last mode under use (I-V or IVCK).
6. In case the current mode is not IVCK, press ENTER,
ENTER
to start up the initial screen of fast check IVCK
15/05/10 15:34:26 select “Measurement ” by means of arrow keys
(,). Use the arrow key to enter the internal
V d c
= 0 . 0 V submenu, select “IV Check” and confirm pressing
I r r
= 0 W / m 2
T c
= - - - ° C
Module: SUNPOWER 210
I m p I V C h e c k
M i s
Selezione
7. Check the preliminary settings based on descriptions of § 6.3.2
I - V
8. Connect the instrument to the module/string under test as shown in the following picture. In particular connect the negative pole of the module/string to terminals P1, C1 and it positive pole to terminals P2, C2. In Auto activation mode is recommended the use of KITKELVIN optional accessory
Caption:
P1: Black
P2: Blue cable cable
C1: Green cable
Fig. 14: Instrument’s connection for
IVCK tests with direct measurement of Irrad/Temp
Fig. 15: Instrument’s connection for
IVCK tests with measurement of
Irrad/Temp through SOLAR-02
C2: Red cable
1: Module or FV string
2: Reference cell
3: Temp probe (if required)
4 Remote unit SOLAR-02
EN - 51
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
9. Connect irradiance probes and temperature probe (if necessary) according to previous pictures and to effected settings (see § 6.3.2).
10. In the initial screen of IVCK mode the following values are
displayed:
Module under use
Irradiance (through direct measurement or SOLAR_02 under
RF connection)
The temperature value (if mode MAN or AUX) and corresponding measurement mode. If mode AUTO ”- - -“
The average values of Voc and Isc under OPC conditions
15/05/10
15:34:26
M o d u l e :
S U N P W R 2 1 0
I r r . 9 8 0 W / m 2
T c ( A U T O )
V o c M e d @ O P C
I s c M e d @ O P C
V o c @ O P C
I s c @ O P C
V o c @ S T C
I s c @ S T C
- - ° C
6 4 7
5 . 4 3
6 4 6
- - -
- - -
- - -
V
A
V
A
V
A
The following values acquired in real time are also displayed:
Open circuit voltage
Selection I V C K
Any symbol of RF connection with SOLAR-02
CAUTION
When pressing the GO/STOP key the instrument can provide different error messages (see §6.3) and, as direct consequence, not carry out the test. Check and eliminate, if possible, causes of the problems before continuing the test.
11. Press GO/STOP (in Manual activation mode) to start up the
test. In case of no error conditions, the instrument displays the screen with the following results:
The module under use
The irradiance value
The cell temperature value
The average values of Voc and Isc under OPC
The values of Voc and Isc measured under OPC
15/05/10 15:34:26
M o d u l e :
I r r 9 3 2 W / m 2
T c ( A U T O )
V o c M e d @ O P C
I s c M e d @ O P C
V o c @ O P C
I s c @ O P C
V o c @ S T C
I s c @ S T C
S U N P W R 2 1 0
5 7 ° C
6 4 7
V
5 . 4 3
A
6 4 7
V
5 . 3 5
7 8 7
5 . 7 2
A
V O K
A O K
OUTCOME: OK
The values of Voc and Isc measured under STC and relative
Selection I V C K partial outcome resulting after comparing the rated values.
In general:
Outcome Voc
@
STC
OK
if 100
VocNom
@
STC
Voc
@
STC
VocNom
@
STC
Tol Voc
4%
Outcome Isc
@
STC
OK
if 100
IscNom
@
STC
Isc
@
STC
Tol Isc
4%
IscNom
@
STC
The rated values of Voc and Isc are the values of the DB built-in modules (see §5.3)
The overall value of outcome: o
OK: if all results under STC are OK, o
NO if one of the results under STC is NO
12. Press SAVE to store the test result into the instrument’s memory (see §7.2) or press
ESC/MENU
to quit without saving and go back to main measurement screen
13. The “Auto” activation mode require to disconnect and reconnect the test leads in order to start automatically a new measurement
EN - 52
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
Media
If irradiance values are measured the instrument provides an outcome comparing the measured values with the average values calculated according to previously saved measurements.
Therefore in this case the average values calculated by the instrument play an important role.
In case you start a new set of measurements with significant variations in irradiance and temperature it is advisable to set the reference average values to zero in order to recalculate based on new measurements.
To reset the average values keep to the following steps:
1. Under mode IVCK, press ENTER key, select “Reset
Media
” and confirm by pressing ENTER to zero the average values calculated till then.
15/05/10 15:34:26
M o d u l e : S U N P W R 2 1 0
I r r 9 8 0 W / m 2
T c ( A U T O ) 4 9 ° C
V o c M e d @ O P C - - - V
I s c M e d @ O P C - - - A
V o c @ O P C 6 4 6 V
I s c @ O P C - - - A
R e s e t M e d i a
S e t t i n g s
M e a s u r e m e n t
V
A
Selection I V C K
The average values are automatically reset even modifying and saving one of the following parameters:
FV module type
Number of modules per string
The average values are not reset if the operator changes working mode (for example he goes to the overall detection of the I-V curve to effect string analysis in deep) and go back to this mode.
EN - 53
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
6.4. LIST OF DISPLAY MESSAGES
MESSAGE DESCRIPTION
Voltage too low
Vin > 1000
Irradiance too low
NTC Error
Please wait for cooling…
Memory full
Pulse width too long
Current too low
Vdc wrong connection
Negative voltage
Data base full
Data @ STC unavailable
Irradiance too high
Data unavailable
Isc too high
Wrong date
Error 1/2/3/4/5: contact Assistance
Error EEPROM : contact Assistance
Error FLASH : contact Assistance
Error RTC : contact Assistance
Battery low
Error: Vmpp >= Voc
Error: Impp >= Isc
Error: Vmpp * Impp >= Pmax
Error: alpha too high
Error: beta too high
Error: Toll too high
Error: gamma too high
Module already present
Delta-Irrad. too high. Retry
Voltage not steady
Unsteady Current
Firmware mismatch
Ref. Cell temp over range
PV module temp over range
Voc out of range. Continue (ENTER/ESC)?
Ref. Cell temp not detected (ENTER/ESC)
Error during memory writing
Error Radio: contact Assistance
Error Radio transmission
Error during download
Recording downloaded
Radio connection activated
Wait data analysis
Unable execute analysis
I < Lim
CAUTION: internal short
Remote unit not detected Enter/Esc
Rem. Unit in rec
STC Data available after Stop rec.
STC Data available in MEM
Verify connect. P1
Error in data download from SOLAR-02
SOLAR-02 not detected. Stop Rec.?
MPP300 not detected
SOLAR-02 not detected.
MPP300: negative AC power
MPP300: inverted AC and DC voltages
No power supply to MPP. Continue? Enter/Esc
C1 C2 not connected
C1 C2 not connected or fuse blown
Fuse blown
Incorrect language version
C1/C2: neg. voltage.
Check the voltage between C1 and C2 input terminals
DC output voltage from module/string > 1000V (I-V400w, SOLAR I-Vw)
Irradiance value lower than threshold limit
Internal NTC damaged. Contact service dpt
Instrument overheated. Wait before starting the tests again
Internal memory full. Download data to PC
Anomalous condition. Repeat test with more modules
Measured current lower than the minimum detectable
Check the voltage between C1 and C2 input terminals
Check the polarity of input terminals
The number of modules defined is > 30
The meter does not calculate the date at STC conditions
Irradiance value higher than the maximum range
Generic error. Repeat the test
Output current higher than maximum range
Set a correct date/hour on meter
Contact service dpt
Contact service dpt
Contact service dpt
Contact service dpt
Low battery indication. Fit new batteries inside meter
Check the settings of module inside DB
Check the settings of module inside DB
Check the settings of module inside DB
Check the settings of module inside DB
Check the settings of module inside DB
Check the settings of module inside DB
Check the settings of module inside DB
Name of module just used inside DB
Unstable condition on irradiance. Repeat the test
Anomalous condition. Repeat test with more modules
Difference between 2 consec. Instant. values of current is > 0.13A
Problem with internal FW. Contact service dpt
Temperature measured by reference cell is too high
Temperature of module over maximum range
Voc measured not compliance with the set number of modules
(check also the Voc and Beta parameters values in the module’s data sheet)
Measure on the cell of module not performed
Problem with the access to internal memory
Contact service dpt
Contact service dpt
Contact service dpt
Recording corrected downloaded on the instrument
Corrected RF connection with remote unit SOLAR-02
Download data from SOLAR-02 and wait for response
Problem on downloaded data from SOLAR-02. Verify settings
Output current lower that the minimum measurable
Contact service dpt
Unit SOLAR-02 do not RF connected to the instrument
Unit SOLAR-02 in recording parameters Irr/Temp
Stop the recording to obtain STC data
Data STC condition save on memory
Check the proper insertion of test cable in the input P1
SOLAR I-Vw/SOLAR I-Ve used with MPP300. Download error from SOLAR-02.
SOLAR I-Vw or SOLAR I-Ve used with MPP300. SOLAR-02 not detected when stop Rec.
SOLAR I-Vw or SOLAR I-Ve set for use together with MPP300. MPP300 not detected
SOLAR I-Vw or SOLAR I-Ve set for use together with MPP300. SOLAR-02 not detected.
MPP300 has detected negative AC powers.
MPP300 has detected inverted AC and DC voltages
MPP300 has not detected the presence of the external power supply.
Check the connections of the instrument to module/string on test
Check the connections of the instrument and repeat the test. If there is problem again contact service dpt
Update the language file of the instrument
Check the polarity of the connections
EN - 54
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
7. STORING DATA
The meter allows storing 99 results of PV checks (only SOLAR I-Vw or SOLAR I-Ve) and
249 tests of I-V curve in its internal memory. In the PV check operation it’s also possible to save the instantaneous values at display. The saved data can be recalled at display and deleted in each moment, and can be associated (for I-V curve test) to reference numerical markers relevant to the installation name, the PV string and the PV module (max 255 markers).
7.1. SAVING TEST RESULTS OF PV CHECKS (SOLAR I-VW OR SOLAR I-VE)
1. Press SAVE key with a measured result shown at display or for saving the instantaneous values. The right screen with a virtual keyboard is displayed by the meter
2. Use the arrow keys (,) and ( , ) to fit a short description (max 12 chars) relative to the PV check
3. Press SAVE key to store the result of test or ESC/MENU key to exit without saving
15/05/10 15:34:26
P R p
I r r
P n o m
T c
0 . 8 5 1
8 2 7
W / m 2
3 . 5 0 0 k W
4 5
° C
T e
3 0
° C
K E Y B O A R D
A B C D E F G H I
A B C D E F G H I J K L M N O P
Q R S T U V W X Y Z - + 0 1 2 3
4 5 6 7 8 9 S P A C E D E L
SAVE for store data
7.2. SAVING TEST RESULTS OF I-V CURVE TEST
1. Press SAVE key with a measured result shown at display.
The following screen is displayed where the herewith items are shown:
The first memory location available (“Measurement”)
The numerical marker “Installation”
The numerical marker “String”
The numerical marker “Module”
The field “Comment” where the operator can include a short description (max 14 char) by using virtual keyboard
15/05/10 15:34:26
M e a s u r e m e n t : 0 0 7
I n s t a l l a t i o n :
0 1 0
S t r i n g : 0 0 9
M o d u l e : 0 0 4
C o m m e n t :
K E Y B O A R D
A B C D E F G H I
A B C D E F G H I J K L M N O P
Q R S T U V W X Y Z - + 0 1 2 3
4 5 6 7 8 9 S P A C E D E L
SAVE for store data
C o m m e n t
MEM – IV
2. Use the arrow keys (,) to select the items and (,) to set the numeric value of the markers and to use the virtual keyboard. “Comments” field modifications are possible only by changing the number of the “Installation” marker inserting one among the available markers
3. Press ENTER to digit any character of the desired name
4. Press SAVE key to store the result of test or ESC/MENU key to exit without saving.
EN - 55
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
7.3. MANAGING THE RESULTS
7.3.1. Recall data of PV checks at display (SOLAR I-Vw or SOLAR I-Ve)
1. Press ESC/MENU key to back to the main menu, select the “MEM” item and confirm with ENTER to access inside the section od saved data
2. Using arrows keys (,) and the arrow key select the
15/05/10 15:34:26
MEM TYPE
001 IST 08/04/2010 item “Recall”, the item “Yield Test” and confirm with
ENTER
for the visualization of PV checks results only
3. Using arrow key the herewith labels can be displayed:
TYPE means the type of saved data: “REC” for a complete PV check, “*REC” when the values of
Y i e l d T e s t
V i e
I - V T e s t
R e c a l l
D e l e t e
irradiance/temperature measured by SOLAR-02 are not available for the meter and “IST” for the saving of
Select
MEM - EFF instantaneous values at display
DATA means the date/hour od saved data on meter
Comments means the description gives by the user during the data saved operation
4. Select the “IST” data, the item “View” and confirm with
ENTER
. The herewith screen is displayed:
5. Press the arrow keys (,) to access to the two available screens
6. Press ESC/MENU to back to previous screen
7. Select the “REC” data, the item “View” and confirm with
ENTER
. The herewith screen is displayed:
15/05/10 15:35:00
P d c 3 . 1 2 5
V d c 3 8 9 V
I d c 8 . 0 1 n d c 0 . 8 8
P a c 3 . 0 1 2
V
A
Select
A n a l y s i s r e s u l t s
E F F
8. Press the arrow keys (,) to access to the two available screens
9. The final values of PV check relative to the maximum efficiency condition detected are displayed
10. Press ESC/MENU to back to previous screen
11. Selecting the “*REC” data, the item “View” and confirm with ENTER the message “Unable to execute analysis” is displayed due to the missed of irradiance/temperature
15/05/10 15:35:00
k W
V
A n d c 0 . 8 8
P a c 3 . 0 1 2
V a c 2 3 1
A n a c 0 . 9 6 values recorded by SOLAR-02. The partial values of
Select
this measurement are only visible after downloading the data onto PC
(see § 8)
E F F
EN - 56
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
7.3.2. Recall data of I-V curve tests at display
1. Press ESC/MENU key to back to the main menu, select the “MEM” item and confirm with ENTER to access inside the section od saved data
2. Using arrows keys (,) and the arrow key select the
15/05/10 15:34:26
MEM DATA
001
002
08/04/2010 10:38
13/04/2010 12:15 item “Recall”, the item “I-V Test” and confirm with ENTER for the visualization of I-C curve results only
3. The “DATA” label means the date/hour od saved data on meter
T e s t
V i e
I - V T e s t
R e c a l l
D e l e t e
Select MEM I - V
4. Use arrow key to select the “Comments” label 15/05/10
15:34:26
MODULE # 1
MODULE # 2
5. The description gives by the user during the data saved operation (see § 7.2) is displayed
MEM Comments
6. The symbol “*” near the number of measure means that
001 the meter have performed the I-V curve test with recording of irradiance/temperature values by use of remote unit
002*
SOLAR-02, but this values did not transferred or are not available. For this measurements the translated STC
values won’t be available
Select
7. By means of the arrow key , select the “parameters” label.
8. The markers relevant to Installation, String and Module selected by the user during the saving procedure (§ 7.2) are shown on the display.
9. Press ESC/MENU to exit the screen and to go back to the
15/05/10
15:34:26
001
002
0 0 1 0 0 1 0 0 1
0 0 1 0 0 1 0 0 2 main menu.
MEM I - V
Select
MEM – IV
EN - 57
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
7.3.2.1. View data – Numerical screens
1. Select a row corresponding to a saved result and press 15/05/10 15:34:26
ENTER
to access the § for the display of stored data. The herewith screen shows a list of saved tests.
001 001 001 001
2. Select the “View” item and press ENTER to open the section of data visualization in the herewith modes:
002 001
Numeric screens of parameters measured at the STC conditions and at the operative conditions (OPC)
Graphic screens relative to the I-V curves saved at STC and OPC conditions
V i e w
D e l e t e
Select
MEM – IV
3. The first screen shows the values of measured parameters, referred to 1 module, translated to the
V o c 4 8 . 0 V standard conditions (STC) according to § 6.2.3
4. Press arrow key , select by means of arrow keys (,)
V m p p 3 9 . 8 the option “OPC – Avg” and press ENTER.
A
A
P m a x W
%
D P m a x S T C
O P C – A v g
D a t a O P C
I - V G r a p h
P W R G r a p h
- O K
Select
I – V
5. The meter shows the values measured on a string at the real operating conditions (OPC). The values are averaged
on a single module
(corresponding to the total values should the string be made of a single module).
6. Press arrow key in the first screen, select the option
“OPC” by means of arrow keys (,) and press ENTER.
15/05/10 15:34:26
V o c
4 . 8 5
I s c
P m a x 1 8 9
F F
9 2 7
T c
R e s u l t s @ O P C – A v g
Select
I – V
7. The meter shows the total values measured at the real operating conditions (OPC) relative to the string under
test.
8. Press ESC/MENU key to exit and go back to the previous screen.
15/05/10 15:34:26
V o c
5 . 2 2
W
0 . 7 7
I r r
T c
R e s u l t s @ O P C
Select
I – V
EN - 58
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
7.3.2.2. View data – Graphic screens of I-V curve
1. With screen of numerical values relative to standard
15/05/10 15:34:26 conditions (STC) select the item “I-V Graph” by means of arrow key and press ENTER or the arrow key .
2. Select the “STC” option and press ENTER.
3. The herewith screen is shown by the meter.
V o c 4 8 . 0 V
V
I m p p 5 . 2 4
A
I s c
A
P m a x 2 0 8
W
F F
D P m a x
D a t a
S T C
0 . 7 8
O P C – A v g
O P C
I - V G r a p h
P W R G r a p h - O K
Select
I – V
%
4. The graph is the I-V curve relative to the object under test translated to the standard conditions (STC) and referred to 1 module.
5. Press ESC/MENU key to exit the screen and to go back to the memory section.
15/05/10 15:34:26
I - V G r a p h @ S T C – O K
Select
I – V
6. With screen of numerical values measured by meter
15/05/10 15:34:26 select the item “I-V Graph” by means of arrow key and press ENTER or the arrow key .
7. Select the option “OPC – Avg” and press ENTER. The I-
V curve graph of a string measured at the real operative conditions (OPC) and displayed by the meter is averaged
on a single module
.
8. Press ESC/MENU key to exit the screen and to go back to the memory section
I - V G r a p h @ O P C - A v g
Select
I – V
9. With screen of numerical values measured by meter
15/05/10 15:34:26 select the item “IV Graph” by means of arrow key and press ENTER or the arrow key .
10. Select the option “OPC” and press ENTER. The I-V curve graph measured at the real operative conditions (OPC) and displayed by the meter corresponds to the total
value relative to the string under test.
11. Press ESC/MENU key to exit from screen and back to the memory §
I - V G r a p h @ O P C
Select
I – V
EN - 59
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
7.3.2.3. View data – Graphic screens of power
1. With screen of numerical values measured by the meter select the item “PWR Graph” by means of arrow key and press ENTER or the arrow key .
2. Select the option “STC” and press ENTER.
3. The herewith screen is shown by the meter.
15/05/10 15:34:26
V o c 4 8 . 0 V
V
I m p p 5 . 2 4
A
I s c
A
P m a x 2 0 8
W
F F
D P m a x
D a t a S T C
0 . 7 8
%
0 . 5
%
I - V G O P V – A v g
P W R O P C
Select
4. The graph shows values of output power from module/string translated to the standard conditions (STC).
15/05/10 15:34:26
5. Press ESC/MENU key to exit the screen and to go back to the memory §.
I – V
P W R G r a p h @ S T C – O K
Select
I – V
6. With screen of numerical values measured by meter,
15/05/10 15:34:26 select the item “PWR Graph” by means of arrow key and press ENTER or the arrow key .
7. Select the option “OPC – Avg” and press ENTER. The meter shows the graph of output power from a single
module of a string
measured at the real operative conditions (OPC).
8. Press ESC/MENU key to exit the screen and to go back to the memory §.
P W R G r a p h @ O P C - A v g
Select
I – V
9. With screen of numerical values measured by the meter select the item “PWR Graph ” by means of arrow key and press ENTER or the arrow key .
15/05/10 15:34:26
10. Select the option “OPC” and press ENTER. The meter shows the graph of total output power relevant to the
string under test
measured at the real operative conditions (OPC).
11. Press ESC/MENU key to exit the screen and to go back to the memory §.
P W R G r a p h @ O P C
Select
I – V
EN - 60
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
1. Press ENTER key in the memory § to show the submenu
2. Select the “Delete” item and press the arrow key . The options below are shown by the meter:
Del. Last delete the last measurement saved
Del. All delete all data in the memory
3. Select the desired option with arrow keys (,) and
15/05/10 15:34:26
MEM DATA
001
1 5 / 0 5 / 1 0
1 2 : 3 2 confirm with ENTER
4. Press ESC/MENU to exit the screen and go back to main menu
V i e w D e l . L a s t
D e l e t e
D e l . A l l
Select
MEM – IV
EN - 61
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
8. CONNECTING THE INSTRUMENT TO A PC
8.1. CONNECTING BY OPTICAL/USB CABLE C2006
CAUTION
The connection between instrument and PC is realized by means C2006 cable.
In order to transfer the data onto a PC it is necessary to install the management software Topview in advance and the drivers of cable C2006 on the PC itself.
Before connecting, it is necessary to select the port to be used and the right baud rate (57600 bps) on the PC. To set these parameters, start the
TopView
software and refer to the program's on-line help.
The selected port must not be engaged by other devices or applications, e.g. a mouse, a modem, etc.
Optical port emits invisible LED radiations. Do not bring the beam at eye level. Class 1M LED apparatus according to IEC/EN 60825-1
To transfer the saved data to the PC, keep to the following procedure:
1. Switch on the instrument by pressing ON/OFF key
2. Connect the instrument to the PC via the provided optical/USB cable C2006.
3. Press ESC/MENU key to open the main menu.
4. By means of arrow keys (,) select the item “PC Connection” to access the PC mode section and confirm with ENTER
15/05/10 15:34:26
I - V
I-V Test
EFF Yield Test
SET
Settings
DB
Modules
MEM
Data Recall
PC
PC Connection
E N T E R t o s e l e c t .
M E N U
5. The below screen is displayed by the meter:
15/05/10 15:34:26
PC – RS232
WiFI ON
M E N U
6. Use data management software TopView to transfer the content of the instrument memory to the PC (see on-line help of this software for all details)
EN - 62
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
8.2. CONNECTING BY WIFI
CAUTION
The connection between instrument and PC is realized by means of WiFi interface which is normally active on the instrument. The WiFi function is not available if in the initial screen of the instrument (see § 4.4) the “WiFi-OFF” message is displayed
In order to transfer the data onto a PC it is necessary to install the management software Topview in advance, and WiFi interface shall be available on the PC itself
Before connecting it is necessary to enable WiFi connection on the instrument as below mentioned, and select/connect the “wireless network”
(WiFi) which the instrument makes available
Before connecting, it is necessary to select the “WiFi” port on the “PC
Instrument Connection” of TopView software and refer to the program online help.
To transfer the saved data to the PC, keep to the following procedure:
1. Switch on the instrument by pressing ON/OFF key
2. Press ESC/MENU key to open the main menu.
3. By means of arrow keys (,) select “PC Connection” to access the PC mode section and confirm by ENTER
15/05/10 15:34:26
I - V
I-V Test
EFF Yield Test
SET
Settings
DB
Modules
MEM
Data Recall
PC
PC Connection
E N T E R t o s e l e c t .
M E N U
4. The below screen is displayed by the meter:
15/05/10 15:34:26
PC – RS232
WiFI ON
M E N U
5. Enable the WiFi connection on the PC (if necessary, it is possible to use a WiFi adaptor). Select and connect the WiFi network which the instrument makes available in
“Network connection Wireless networks” section
6. Use data management software TopView to transfer the content of the instrument memory to the PC (see software on-line help for all details)
EN - 63
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
9. MAINTENANCE
9.1. GENERAL
The instrument you purchased is a precision instrument. When using and storing it, please observe the recommendations listed in this manual in order to prevent any possible damage or danger. Do not use the instrument in environments with high humidity levels or high temperatures. Do not directly expose it to sunlight. Always switch off the instrument after using it. Should the instrument remain unused for a long time, remove batteries in order to prevent liquids from leaking out of them, and the instrument internal circuits from being damaged.
9.2. BATTERY REPLACEMENT
When the low battery symbol “ ” appears on the LCD display, or if during a test the meter gives the message “low battery”, it is necessary to replace the batteries.
CAUTION
This operation must be carried out by skilled technicians only. Before carrying out this operation, make sure that all cables have been removed from the input leads.
1. Swicth off the instrument by pressing and holding the ON/OFF key.
2. Remove the cables from the input leads.
3. Unscrew the cover fastening screw from the battery compartment and remove it.
4. Remove all the batteries from the battery compartment and replace with new batteries of the same type only (§ 10.4) making sure to respect the indicated polarities.
5. Restore the battery compartment cover into place and fasten it by means of the relevant screw.
6. Do not dispose of the used batteries into the environment. Use relevant containers for disposal.
9.3. INSTRUMENT CLEANING
Use a dry and soft cloth to clean the instrument. Never use wet cloths, solvents, water, etc.
9.4. END OF LIFE
CAUTION
: this symbol indicates that the equipment, the batteries and its accessories must be collected separately and disposed of in the right way.
EN - 64
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
10. TECHNICAL SPECIFICATIONS
10.1. YIELD TEST TECHNICAL FERATURES (SOLAR I-VW, SOLAR I-VE)
Accuracy is calculated as [%reading+(number of dgt) * resolution] at 23°C ± 5°C, <80%HR
VDC Voltage (SOLAR I-Vw)
Range [V]
15.0
999.9
VDC Voltage (SOLAR I-Ve)
Resolution [V]
0.1
Accuracy
(0.5rdg + 2dgt)
Range [V]
15.0
99.9
100.0
1499.0
AC TRMS Voltage
Resolution [V]
0.1
0.3
Range [V]
50.0
265.0
Max crest factor: 1.5
Resolution [V]
DC Current (by external transducer clamp)
0.1
Accuracy
(0.5rdg + 2dgt)
Accuracy
(0.5rdg + 2dgt)
Range [mV]
-1100
-5
5
1100
Resolution [mV]
0.1
Accuracy
(0.5rdg + 0.6mV)
The value of current is ALWAYS displayed with positive sign ; The value of current transduced in voltage less then 5mV is zeroed
AC TRMS Current (by external transducer clamp)
Range [mV]
1
1200
Resolution [mV]
0.1
Max crest factor: 2.0 ; The value of current transduced in voltage less then 5mV is zeroed
Accuracy
(0.5rdg + 0.6mV)
FS DC and AC clamp [A] Resolution [A]
Minimum read value [A]
DC AC
1< FS
10
10< FS
100
100< FS
1000
DC Power (Vmis > 150V) (SOLAR I-Vw)
Accuracy FS clamp [A]
1< FS
10
10< FS
100
Range [W]
0.000k
9.999k
10.00k
99.99k
0.000k
9.999k
10.00k
99.99k
100< FS
1000
0.00k
99.99k
100.0k
999.9k
Vmis = voltage at which power is measured ; Imis = measured current
Resolution [W]
0.001k
0.01k
0.001k
0.01k
0.01k
0.1k
DC Power (Vmis > 150V) (SOLAR I-Ve)
FS clamp [A]
1< FS
10
10< FS
100< FS
100
1000
Range [W]
0.000k
9.999k
10.00k
99.99k
0.000k
9.999k
10.00k
99.99k
100.0k
999.9k
0.00k
99.99k
100.0k
999.9k
1000k
9999k
Resolution [W]
Vmis = voltage at which power is measured ; Imis = measured current
0.001k
0.01k
0.001k
0.01k
0.1k
0.01k
0.1k
1k
(0.7rdg + 3dgt)
(Imis < 10%FS)
(0.7rdg)
(Imis
10%FS)
Accuracy
(0.7rdg + 3dgt)
(Imis < 10%FS)
(0.7rdg)
(Imis
10%FS)
EN - 65
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
AC Power (Vmis > 200V, PF=1) (SOLAR I-Vw)
FS clamp [A]
1< FS
10
10< FS
100
Range [W]
0.000k
9.999k
10.00k
99.99k
0.000k
9.999k
10.00k
99.99k
Resolution [W]
100< FS
1000
0.00k
99.99k
100.0k
999.9k
Vmis = voltage at which power is measured ; Imis = measured current
0.001k
0.01k
0.001k
0.01k
0.01k
0.1k
AC Power (Vmis > 200V, PF=1) (SOLAR I-Ve)
FS clamp [A]
1< FS
10
10< FS
100
Range [W]
0.000k
9.999k
10.00k
99.99k
0.000k
9.999k
10.00k
99.99k
100.0k
999.9k
Resolution [W]
100< FS
1000
0.00k
99.99k
100.0k
999.9k
1000k
9999k
Vmis = voltage at which power is measured ; Imis = measured current
0.001k
0.01k
0.001k
0.01k
0.1k
0.01k
0.1k
1k
Frequency
Range [Hz]
47.5
63.0
Resolution [Hz]
0.1
Irradiance (with reference reference cell)
Range [mV]
1.0
65.0
Temperature (with auxiliary probe)
Range [°C]
-20.0
100.0
Resolution [mV]
0.1
Resolution [°C]
0.1
Accuracy
(0.7rdg + 3dgt)
(Imis < 10%FS)
(0.7rdg)
(Imis
10%FS)
Accuracy
(0.7rdg + 3dgt)
(Imis < 10%FS)
(0.7rdg)
(Imis
10%FS)
Accuracy
(0.2rdg + 1dgt)
Accuracy
(1.0rdg + 5dgt)
Accuracy
(1.0rdg + 1°C)
EN - 66
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
10.2. I-V TEST TECHNICAL FEATURES (I-V AND IVCK)
I-V, IVCK: DC Voltage @ OPC (SOLAR I-Vw, I-V400w)
Range [V] (*)
5.0
999.9
Resolution [V]
0.1
(*) The I-V curve measurement start for VDC > 15V and the accuracy is defined for VDC > 20V
I-V, IVCK: DC Voltage @ OPC (SOLAR I-Ve, I-V500w)
Range [V] (*)
15.0
99.9
Resolution [V]
0.1
100.0
1499.9
0.3
(*) The I-V curve measurement start for VDC > 15V and the accuracy is defined for VDC > 20V
I-V: DC Current @ OPC
Uncertainty (*)
(1.0rdg + 2dgt)
Uncertainty (*)
(0.5rdg + 2dgt)
Range [A]
0.10
15.00
Resolution [A]
0.01
Uncertainty
(1.0rdg + 2dgt)
(*) Maximum current = 15A for Voc ≤ 1000V, Maximum current = 10A for Voc > 1000V (I-V500w and SOLAR I-Ve)
I-V: DC Power @ OPC (Vmpp > 30V, Impp > 2A)
Range [W] (*)
50
99999
Resolution [W]
1
Uncertainty
(1.0rdg + 6dgt)
Vmpp = Maximum power voltage, Impp = Maximum Power Current
(*) The value of max measured power is relative to a max FF value of approx. 0.7 Pmax= 1000V x 15A x 0.7 = 10500W
Pmax= 1500V x 10A x 0.7 = 10500W
I-V, IVCK: DC Voltage @ STC
Uncertainty (*,**)
4.0rdg + 2dgt)
Range [V]
5.0
999.9
I-V: DC Current @ STC
Resolution [V]
0.1
Range [A]
0.10
99.00
Resolution [A]
0.01
I-V: DC Power @ STC (Vmpp > 30V, Impp > 2A)
Uncertainty (**)
(4.0 rdg + 2 dgt)
Range [W]
50
99999
Resolution [W]
1
Global Uncertainty (**)
(5.0 rdg + 1 dgt)
Vmpp = Maximum power voltage, Impp = Maximum Power Current
(*) Measurements of I-V curve start for VDC > 15V and the accuracy is defined for VDC > 20V
(**) Test conditions:
Test cond.: Steady Irrad.
700W/m
2
, spectrum AM 1.5,solar incidence vs perpendicular.
25°, Cells Temp. [15..65°C]
Global accuracy include contribute of solar sensor and its measuring circuit
Irradiance (with reference cell)
Range [mV]
1.0
100.0
Temperature (with auxiliary probe)
Resolution [mV]
0.1
Accuracy
(1.0rdg + 5dgt)
Range [°C]
-20.0
100.0
Resolution [°C]
0.1
Accuracy
(1.0rdg + 1°C)
EN - 67
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
10.3. SAFETY SPECIFICATIONS
10.3.1. General
Instrument safety: IEC/EN61010-1
EMC: IEC/EN61326-1
Technical literature: IEC/EN61187
Accessory safety:
Measurements:
IEC / EN61010-031
IEC/EN60891 (I-V curve test)
IEC/EN 60904-5 (Temperature measurement) insulation
Pollution level:
Category of measure:
2
CAT II 1000V DC, CAT III 300V to ground
Max 1000V among inputs P1, P2, C1, C2 (SOLAR I-Vw,
I-V400w)
Max 1500V among inputs P1, P2, C1, C2 (SOLAR I-Ve,
I-V500w)
EN - 68
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
10.4. GENERAL CHARACTERISTICS
Display and memory
Type of display:
Memory capacity:
Saved data:
PC interface:
LCD custom, 128x128 pxl, backlighted
256kbytes
99 yield tests (SOLARI-Vw/I-Ve); 249 I-V, 999IVCK optical/USB and WiFi
RF module specifications
Frequency range:
R&TTE category:
Max transmission power:
Max distance of RF connection
2.400
2.4835GHz
Class 1
30
W
1m
PV Yield Test (SOLAR I-Vw or SOLAR I-Ve)
Integration period (IP): 5,10,30,60,120,300,600,900,1800,3600s
Memory authonomy SOLAR-02: approx 1.5 hours (@ PI = 5s)
Max distance RF connection: approx 8 days (@ PI = 600s)
1m (3 ft)
Memory of SOLAR-02 - MPP300: approx. 1.5 hours (@ PI = 5s) approx. 8 days (@ PI = 600s)
Power supply
Battery type: 6 x1.5 V alkaline type AA LR06
1W
Low battery indication:
Battery life:
Auto power off: symbol “ “ is displayed approx 120 hours (PV Yield test)
>249 I-V curve test, 999 IVCK test after 5 minutes of idleness
Mechanical data
Dimensions (L x W xH):
Weight (batteries included):
235 x 165 x 75mm ; 9 x 6 x 3in
1.2kg (42 ounces)
10.5. ENVIRONMENT CONDITIONS
Reference temperature:
Working temperature:
Relative humidity allowed:
Storage temperature:
Storage humidity:
Max height of use:
23° ± 5°C (73°F ± 41°F)
0 ÷ 40°C (32°F ÷ 104°F)
<80%RH
-10 ÷ 60°C (14°F ÷ 140°F)
<80%RH
2000m (6562 ft) (*)
CAUTION
(*) Information about the use of meter at altitude from 2000 to 5000m
As for voltage inputs P1, P2, C1, C2 the instrument is to be considered downgraded to overvoltage category CAT I 1000V DC and CAT II 300V to ground, max 1000V among inputs (for SOLAR I-Vw and I-V400w) and max 1500V (for SOLAR I-Ve and I-V500w).
Markings and symbols indicated on the instrument are to be considered valid when using it at altitude lower than 2000m
This instrument complies with the prescriptions of the European directive on low voltage 2014/35/EU (LVD) and EMC 2014/30/EU
This instrument satisfies the requirements of 2011/65/EU (RoHS) directive and
2012/19/EU (WEEE) directive
10.6. ACCESSORIES
See only the standard and optional accessory indicated in the enclosed packing list
EN - 69
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
11. APPENDIX
11.1. TESTING PHOTOVOLTAIC SYSTEMS (SOLAR I-VW, SOLAR I-VE)
According to the requirements of the laws in force, the result of the test depends on settings about Temperature effects compensation and PRp calculations:
Corr.Typ
e
Tcel value PRp calculation
Guidelin e
Tmod
Tcel = PV Module Temp. measured
Tcel = PV module Temp. calculated
Tamb or
Tenv
Tcel
Tamb
NOCT
20
G p
800
PRp
Rfv
2
P ca
G p
G
STC
P n
CEI
82-25
(Italian
Guideline)
nDC
Tcel = PV Module Temp. measured
PRp
G
STC
G p
1
100
T cel
25
P ca
P n
- - - where:
G p
Irradiance on PV module surface
W/m
2
G
STC
Standard Irradiance = 1000
2
W/m
P n
Nominal Power = sum of all power module (Pmax ) included in the part of PV plant under test
AC Active Power measured
P ca
Rfv
2
1
1 (Tcel 40)
100
(if
(if
Tcel
Tcel
40
C)
40
C)
Thermal Corrective factor
Absolute value of Pmax thermal coefficient
%/
C
NOCT Normal Operating Cell Temperature (@ 800W/m
2
, 20°C,
AM=1.5, vel. Aria =1m/s).
%/
C
Previous relationship are valid if Irradiance > Min Irraddiance value and the Irradiance values are “steady”: if IP
1min (Irr max – Irr min) < 20W/m
2
The final OUTCOME can be:
Non-displayable: if the obtained values are inconsistent (e.g. PRp >1.15) or if irradiation has never reached a steady value > minimum threshold set.
The maximum performance point (PRp) of the system
The highest performance (maximum value of PRp) is detected according with previous relationships.
EN - 70
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
11.2. NOTES ON MPPT (MAXIMUM POWER POINT TRACKER)
Solar irradiation on a surface such as the surface of a photovoltaic system has extremely variable characteristics, since it depends on the position of the sun with respect to the surface and on atmospheric conditions (typically, on the presence of clouds). A photovoltaic module presents, for different solar irradiation values, and for different temperature values, a range of characteristic curves of the type shown in the following figure. In particular, the figure shows three I-V curves (in bold) which correspond to three values (1000, 800, 600W/m
2
) of solar irradiation.
On each characteristic curve there is one single point in which the power transfer towards a hypothetical charge supplied by the photovoltaic module is maximized. The maximum power point corresponds to the voltage-current pair for which the product V*I is maximum, where V is the value of voltage at the module’s terminals and I is the current which runs in the circuit obtained by closing the module on a hypothetical charge.
With reference to the figure above, the product V*I is represented, for the three solar irradiation values mentioned above, through the three curves in thinner lines.
The figure shows that, as stated above, these curves only have one single maximum point.
For example, for 1000W/m
2
, the maximum power point corresponds to a voltage value of approx. 36V and to a current value of approx. 5.5A.
Obviously, if the power provided by the system is maximized, it is possible to make the most of the system, both in case the system is connected to mains, and in case it is standalone.
MPPT is an inbuilt device in the inverters. It typically reads the voltage and current values at any instant, calculates their product (i.e. the power in Watts) and, by causing small variations in the conversion parameters (duty cycle), it is capable of determining, by comparison, if the photovoltaic module is working in maximum power conditions or not.
According to the result, it operates again on the circuit in order to bring the system to an optimal condition.
EN - 71
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
The reason why MPPTs are used is simple: a photovoltaic system without MPPTs may operate anyway; however, with the same solar irradiation, it provides less energy.
There are inverters with 1, 2 or also 3 inbuilt MPPTs available on the market. Typically, the inverters with more than one MPPT are used in systems where:
- the different photovoltaic systems it consists of “forcibly” have different inclinations or directions. In this way, each single MPPT manages its own photovoltaic field, maximizing its performance for the corresponding irradiation and temperature characteristics (without being influenced by the other photovoltaic fields).
- a greater service continuity is sought. With more MPPTs it is possible to put one single photovoltaic field out of service, while the others continue producing energy towards the remaining MPPTs.
11.3. THEORETICAL ASPECTS OF THE I-V CURVE MEASUREMENT
The I-V curve test is performed as described below:
The meter performs the I-V curve measurement of the PV module connected to it, further to the real time measurement of irradiance and module temperature
The results of measurements are automatically “translated” to the standard conditions
(STC) of irradiance at 1000 W/m
2
and module temperature at 25°C. The accuracy indicated in this manual is granted under condition indicated in § 10.2.
The meter performs a comparison between the maximum nominal power, with the margin of percentage tolerance declared by the module’s manufacturer, which is included in the kind of module selected on the meter from the database (§ 5.3.1), and the measured value
If the comparison of power is within the declared tolerance, the final response of the meter will be “OK” or will be “NO OK” in the opposite case, and the module will not be complying with the specifications declared by the manufacturer (§ 6.2.3)
EN - 72
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
11.3.1. Typical errors on I-V curves and troubleshooting
Measurement
Good measurement
Event Troubleshooting
The measured curve extrapolated No error, save data and to STC (blue) and the curve as declared by the manufacturer
(black) of the modules match. move to next string to test
Low short circuit current
Low open circuit voltage
High short circuit current
Orient the irradiance sensor differently than the string under test
sensor properly
Set the instrument properly
Wrong module selection at configuration of PV system
Clean the modules
Remove obstruction
Environmental contaminants on modules (soiling, dirt, snow, debris)
Far distance obstruction (shading)
Aging blindness, humidity, etc.
Replace affected modules
Irradiance sensor oriented differently than the string under test
Shades on the irradiance sensor
Orient the irradiance sensor properly
Clean the irradiance sensor
Dirt on the irradiance sensor
configuration of PV system
Irradiance sensor defect
Remove obstruction
Check the irradiance sensor for blindness, humidity, etc.
Set the instrument properly
Replace affected sensor
Low temperature measurement
Wrong numbers of module selection at configuration of PV system
Shorted bypass diodes
Attach the thermocouple properly to a typical temperature location
Select the AUTO mode
Set the instrument properly
Replace affected modules
Low current ratio Impp/Isc
Soiling across modules
Tapered shades
Small current mismatch
Degradation of cell shunt resistance
Clean the modules
Un-shade the modules
Check the current match
Replace affected modules
Low voltage ratio Vmpp/Voc
Voltage drop on cables to PV module string
Check cables, connectors, contacts, and string wirings
Check cable length and cross section
Check modules for bad connections or oxidation
Replace affected modules
EN - 73
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
Steps in the I-V curve
Small or partial shading of one Remove obstructions module of the string under test
Reflections
Random environmental contaminants on modules (soiling, dirt, snow, debris, etc.)
Strong current mismatch
Re-test when un-shaded
Remove reflections
Clean the modules
Check the current match
Replace affected modules
Characteristic curve not linear
Cracks of cells or glass
Burn marks
measurement
Leopard spot clouds shading the
Re-test when clouds are gone
Repeat measurement at an modules
Low irradiation measurement during irradiation of minimum
700W/m
2
(IEC/EN61829)
EN - 74
I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
12. SERVICE
12.1. WARRANTY CONDITIONS
This instrument is guaranteed against any defect in material and manufacturing in compliance with the general sales terms and conditions. Throughout the period of guarantee all defective parts may be replaced and the manufacturer reserves the right to repair or replace the product.
If the instrument is to be returned to the after-sales service or to a dealer transportation costs are on the customer’s behalf. Shipment shall be however agreed upon. A report must always be enclosed to a rejected product stating the reasons of its return. To ship the instrument use only the original packaging material; any damage that may be due to nooriginal packing shall be charged to the customer. The manufacturer declines any responsibility for damages caused to persons and/or objects.
Warranty is not applied in the following cases:
Repair and/or replacement of accessories and battery (not covered by warranty)
Any repair that might be necessary as a consequence of a misuse of the instrument or of its use with no compatible devices.
Any repair that might be necessary as a consequence of improper packaging.
Any repair that might be necessary as a consequence of service actions carried out by unauthorized personnel.
Any modification of the instrument carried out without the authorization of the manufacturer.
Use not provided for in the instrument’s specifications or in the instruction manual.
The contents of this manual may not be reproduced in any form whatsoever without the manufacturer’s authorization.
All our products are patented and their trade marks registered. The manufacturer reserves the right to modify the product specifications and prices if this is aimed at technological improvements
.
12.2. SERVICE
If the instrument does not operate properly, before contacting the after-sales service check cables as well as test leads and replace them if necessary. Should the instrument still operate improperly check that the operation procedure is correct and conforms with the instructions given in this manual. If the instrument is to be returned to the after-sales service or to a dealer, transportation costs are on the customer’s behalf. Shipment shall be however agreed upon. A report must always be enclosed to a rejected product stating the reasons of its return. To ship the instrument use only the original packaging material; any damage that may be due to non-original packing shall be charged to the customer.
EN - 75
HT INSTRUMENTS SA
C/ Legalitat, 89
08024 Barcelona - ESP
Tel.: +34 93 408 17 77, Fax: +34 93 408 36 30 eMail: [email protected] eMail: [email protected]
Web: www.htinstruments.es
HT INSTRUMENTS USA LLC
3145 Bordentown Avenue W3
08859 Parlin - NJ - USA
Tel: +1 719 421 9323 eMail: [email protected]
Web: www.ht-instruments.com
HT ITALIA SRL
Via della Boaria, 40
48018 Faenza (RA) - ITA
Tel: +39 0546 621002
Fax: +39 0546 621144 eMail: [email protected]
Web: www.ht-instruments.com
HT INSTRUMENTS GMBH
Am Waldfriedhof 1b
D-41352 Korschenbroich - GER
Tel: +49 (0) 2161 564 581
Fax: + 49 (0) 2161 564 583 eMail: [email protected]
Web: www.ht-instruments.de
HT INSTRUMENTS BRASIL
Rua Aguaçu, 171, bl. Ipê, sala 108
13098321 Campinas SP - BRA
Tel: +55 19 3367.8775
Fax: +55 19 9979.11325 eMail: [email protected]
Web: www.ht-instruments.com.br
HT ITALIA CHINA OFFICE
意大利
HT 中国办事处
Room 3208, 490# Tianhe road, Guangzhou - CHN
地址:广州市天河路
490 号壬丰大厦 3208 室
Tel.: +86 400-882-1983, Fax: +86 (0) 20-38023992 eMail: [email protected]
Web: www.guangzhouht.com
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Key Features
- Measures irradiance, temperature, and humidity
- Performs I-V curve measurements
- Tests PV systems with single or multiple MPPT inverters
- Generates detailed reports
- Stores and manages data on your PC
Related manuals
Frequently Answers and Questions
What is the HT SOLAR I-Ve?
What can I do with the HT SOLAR I-Ve?
Who is the HT SOLAR I-Ve for?
advertisement
Table of contents
- 5 SAFETY PRECAUTIONS AND PROCEDURES
- 7 GENERAL DESCRIPTION
- 8 PREPARATION FOR USE
- 9 DESCRIPTION OF PARTS
- 11 MAIN MENU
- 11 General
- 12 Measurement units
- 12 Date/Time
- 13 Remote unit/Solarim
- 14 Irradiance
- 14 DC Clamp (SOLAR I-Vw or SOLAR I-Ve)
- 15 Settings for PV systems with single-MPPT inverter - single-phase AC output
- 15 Instrument settings
- 16 PV plant parameters
- 17 Settings for PV with single/multi MPPT inverter – single/three phase AC output
- 17 Instrument settings
- 18 PV plant parameters
- 19 MPP300 status
- 20 Selection of the compensation relationship of temperature effects
- 22 How to define a new PV module
- 23 How to modify a PV module
- 23 How to delete a PV module
- 24 HOW TO OPERATE
- 25 Testing of PV systems with single-MPPT inverter - single-phase AC output
- 29 Test PV systems with single/multi-MPPT inverter – single/three-phase AC output
- 35 I-V curve measurement by use of instrument
- 39 I-V curve measurement by use of remote unit SOLAR
- 39 I-V curve test by remote unit SOLAR-02 in RF connection
- 43 I-V curve test by remote unit SOLAR-02 in synchro recording
- 48 Meaning of measurement results
- 49 General information
- 50 Preliminary settings
- 51 Fast check IVCK without irradiance measurement
- 53 Fast check IVCK with irradiance measurement
- 55 Reset Media
- 57 STORING DATA
- 58 Recall data of PV checks at display (SOLAR I-Vw or SOLAR I-Ve)
- 59 Recall data of I-V curve tests at display
- 60 View data – Numerical screens
- 61 View data – Graphic screens of I-V curve
- 62 View data – Graphic screens of power
- 63 Delete data
- 64 CONNECTING THE INSTRUMENT TO A PC
- 66 MAINTENANCE
- 67 TECHNICAL SPECIFICATIONS
- 70 General
- 72 APPENDIX
- 75 Typical errors on I-V curves and troubleshooting
- 77 SERVICE