User Manual
PTW C-Box
Installation and Assembly Manual
Contents
PTW C-Box
Contents
Contents
Illustrations
1 Safety Information
2 General
3 Assembly for application MP2 and MP3
3.1
Control unit in control room
3.2
Control unit and TANDEM in the treatment room
4 Assembly for application MULTIDOS with LA48
4.1
Control unit in the control room
4.2
Control unit and TANDEM in the treatment room
5 Assembly for application MULTIDOS with QC6Plus
6 Assembly for application MULTIDOS or VIVODOS with detector box or DIODE
MATE
7 Assembly for application MULTIDOS or VIVODOS with AL-Box
8 Assembly for application MULTIDOS with M-Chamber-Box
9 Assembly for application MULTIDOS with B-Chamber-Box
10 Order Information
2 D418.131.00/01 en
PTW C-Box
Illustrations
Illustrations
Figure 3-1: Wiring MP3 – Tandem (drawing no. 16006.9.005/1)
Figure 3-2: Wiring MP3 / MP3-S (drawing no. 16006.9.001/1)
Figure 3-3: C-Box control room (drawing no. 16006.1.002/2)
Figure 3-4: C-Box treatment room (drawing no. 16006.1.002/1)
Figure 3-5: Connection of control cable L173038
Figure 3-6: Connection of motor cable L173015
Figure 3-7: Wiring MP3 – TANDEM (drawing no. 16006.9.006/1)
Figure 3-8: C-Box control room (drawing no. 16006.1.006/2)
Figure 3-9: C-Box treatment room (drawing no. 16006.1.006/1)
Figure 3-10: Cable (in the wall) RS232 (drawing no. 16006.1.011/1)
Figure 4-1: Wiring RS232 (LA48) / Monitor (drawing no. 16006.9.003/1)
Figure 4-2: C-Box control room (drawing no. 16006.1.004/2)
Figure 4-3: C-Box treatment room (drawing no. 16006.1.004/1)
Figure 4-4: Cable (in the wall) RS232 (drawing no. 16006.1.011/1)
Figure 4-5: Cable (in the wall) Monitor (drawing no. 16006.1.012/1)
Figure 4-6: Wiring MP3 - LA48 (drawing no. 16006.9.007/1)
Figure 4-7: C-Box control room (drawing no. 16006.1.004/2)
Figure 4-8: C-Box Box treatment room (drawing no. 16006.1.004/1)
Figure 4-9: Cable (in the wall) RS232 (drawing no. 16006.1.011/1)
Figure 4-10: Cable (in the wall) Monitor (drawing no. 16006.1.012/1)
Figure 4-11: RS232 connection cable / T25015 (drawing no. 25015.1.001/1)
Figure 5-1: Wiring MULTIDOS – QC6Plus (drawing no. 16006.9.002/1)
Figure 5-2: C-Box control room (drawing no. 16006.1.003/2)
Figure 5-3: C-Box treatment room (drawing no. 16006.1.003/1)
Figure 6-1: Wiring MULTIDOS – detector box (drawing no. 16009.9.001/1)
Figure 6-2: C-Box control room (drawing no. 16006.1.003/2)
Figure 6-3: C-Box treatment room (drawing no. 16006.1.003/1)
Figure 7-1: Wiring MULTIDOS – AL-Box (drawing no. 16008.9.001/1)
Figure 7-2: C-Box control room (drawing no. 16006.1.003/2)
Figure 7-3: C-Box treatment room (drawing no. 16006.1.003/1)
Figure 8-1: Wiring MULTIDOS – M-Chamber-Box (drawing no. 16007.9.001/1)
Figure 8-2: C-Box control room (drawing no. 16006.1.003/2)
Figure 8-3: C-Box treatment room (drawing no. 16006.1.003/1)
Figure 9-1: Wiring MULTIDOS – B-Chamber-Box (drawing no. 16013.9.001/1)
Figure 9-2: C-Box control room (drawing no. 16006.1.003/2)
Figure 9-3: C-Box treatment room (drawing no. 16006.1.003/1)
D418.131.00/01 en 3
1 Safety Information
NOTE
The C-Box is a component of the following systems:
− MP2 / MP3 / MP3-S / MP3-M / MP3-XS and similar moving mechanisms
− MULTIDOS
− VIVODOS
− DIODE MATE
Follow the Safety instructions and General Information given in the manuals to the mentioned systems.
WARNING
Installing the C-Box requires technical knowhow and may only be performed by trained technical staff.
The relevant regulations concerning electric safety must especially be observed.
Non-compliance with any of these stipulations voids any liability of PTW for operation and safety of the equipment in question.
PTW-FREIBURG
Physikalisch-Technische Werkstätten
Dr. Pychlau GmbH
Lörracher Str. 7
79115 FREIBURG
GERMANY
Phone: +49 761 49055-0
Fax: +49 761 49055-70 [email protected] www.ptw.de
PTW C-Box
Safety Information
4 D418.131.00/01 en
PTW C-Box
General
2 General
The PTW - C-Box is a universal wall box for the connecting of PTW-Instruments and for various applications.
One C-Box has to be installed in the treatment room and one in the control room. This allows to handle the cables of different measuring systems in a simple and time-saving way.
The advantage of the C-Box is, that you can install only one application at first and integrate other applications later without difficulty.
Following instruments and systems can be connected with the C-Box if it is equipped completely:
− MP2- or MP3-system (as well MP3-S, MP3-M,
MP3-XS) inclusive field detector and reference detector or a similar moving mechanism
− MULTIDOS for LA48 application
(LA = Linear Array)
− MULTIDOS for QC6Plus application
(QC = Quality Control)
− MULTIDOS or VIVODOS for TBI application
(TBI = Total Body Irradiation)
− MULTIDOS or VIVODOS for AL application
(AL = Afterloading)
− MULTIDOS or VIVODOS with DIODE MATE
− MULTIDOS with M-Chamber-Box (Connection box for up to 12 detectors with M-Connector)
− MULTIDOS with B-Chamber-Box (Connection box for up to 12 detectors with BNC-Connector and Banana plug)
NOTE
The C-Box should be installed in connection with a cable channel that the cables are routed through the cable channel from below into the
C-Box. The strain-relief will be guaranteed after fixation of the lower cover.
In the case the cables are pulled through the wall directly from behind necessary wholes have to be drilled. In that case the strain relief is under your own responsibility.
For all applications you have first to unscrew the front panel of the C-Box and the lower cover. Afterwards the C-Box can be fixed to the wall using the enclosed mounting kit (4 screws/4 dowels).
After the connection of the cables you should sign them with the enclosed cable markers adequate to the respective drawings.
D418.131.00/01 en 5
PTW C-Box
Assembly for application MP2 and MP3
3 Assembly for application MP2 and MP3
NOTE
The following instructions apply for the systems
MP2, MP3, MP3-S, MP3-M, MP3-XS and all similar moving mechanisms.
3.1 Control unit in control room
Possible versions:
1. control unit T41013 with separate TANDEM
electrometer, refer to Figure 3-1
2. control unit T43164/U600 resp. T41004, refer to
The following cable connections are necessary
(refer to Figure 3-1 and Figure 3-2):
− control cable between water phantom in the treatment room and control unit in the control room consisting of 2 x T43164/K1 and cable (in the wall) L173038
− motor cable between water phantom in the treatment room and control unit in the control room consisting of 2 x T43164/K2 and cable (in the wall) L173015
− cable for detector between water phantom in the treatment room and TANDEM or control unit
(T43164/U600 resp. T41004) in the control room
− cable for reference detector (if fitted) between water phantom in the treatment room and
TANDEM or control unit (T43164/U600 resp.
T41004) in the control room
Figure 3-1: Wiring MP3 – Tandem (drawing no. 16006.9.005/1)
6 D418.131.00/01 en
PTW C-Box
Assembly for application MP2 and MP3
Figure 3-2: Wiring MP3 / MP3-S (drawing no. 16006.9.001/1)
Assembly:
•
Afterwards pull the cables from below into the
C-Box.
•
You have to rip open the front foil along the perforation for the installation of the 37-pin and 15pin Sub-D-connectors.
•
Remove the cable jacket on both sides on approx. 45 mm.
•
Equip the C-Box in the control room with the
37-pin and the 15-pin Sub-D-plug according to
•
Equip the C-Box in the treatment room with the
37-pin and the 15-pin Sub-D-socket according to
•
Untie and twist the screen and sold it to the sol-
dering ring according to Figure 3-3 and
•
Solder the control cable on both sides to the
Sub-D-connectors according to Figure 3-5.
•
Solder the motor cable on both sides to the Sub-
D-connectors according to Figure 3-6.
•
Pull both raw cables through the wall.
•
Slide at both ends a lead-through bush over both cables.
WARNING
The given wire colors and the pin connections are very important because the wires of the cables are twisted in pairs.
D418.131.00/01 en 7
PTW C-Box
Assembly for application MP2 and MP3
•
Fix the plug holders for the chamber cable
(T2664/U1) on the right and the left hand side of the C-Box with the delivered screws.
•
If the wiring is just for one application you can screw on the front panels now.
•
Afterwards attach the lower covers.
NOTE
Please pay attention that you clamp the leadthrough bushes between cover and housing in the intended drillings.
•
Attach the plugs or the sockets of the detector cables in the holder of the left and right side.
Figure 3-3: C-Box control room (drawing no. 16006.1.002/2)
8 D418.131.00/01 en
PTW C-Box
Assembly for application MP2 and MP3
Figure 3-4: C-Box treatment room (drawing no. 16006.1.002/1)
Figure 3-5: Connection of control cable L173038
D418.131.00/01 en 9
PTW C-Box
Assembly for application MP2 and MP3
Figure 3-6: Connection of motor cable L173015
10 D418.131.00/01 en
PTW C-Box
Assembly for application MP2 and MP3
3.2 Control unit and TANDEM in the treatment room
Figure 3-7: Wiring MP3 – TANDEM (drawing no. 16006.9.006/1)
This assembly is only possible with control unit
T41013 with separate TANDEM electrometer.
The following cable connections are necessary
− RS232 cable between control unit in the treatment room and PC in the control room consisting of 2 x T22373/K67and cable (in the wall)
T16006.1.011
Assembly:
•
You have to rip open the front foil along the perforation for the installation of the 9-pin Sub-Dconnectors.
•
Equip the C-Box in the control room with 9-pin
Sub-D-socket according to Figure 3-8.
•
Equip the C-Box in the treatment room with the
9-pin Sub-D-plug according to Figure 3-9.
•
Pull the raw cable through the wall.
•
Slide at both ends a lead-through bush over both cables.
•
Afterwards pull the cables from below into the
C-Box.
D418.131.00/01 en 11
PTW C-Box
Assembly for application MP2 and MP3
•
Solder the cable for the RS232 interface on both sides to the Sub-D-connectors according to
NOTE
WARNING
The given wire colors and the pin connections are very important because the wires of the cables are twisted in pairs.
Instead of the TANDEM you can also use an
UNIDOS or MULTIDOS as electrometer. To connect the UNIDOS or MULTIDOS to the MP3 control unit please use the special RS232 cable
T25015 (see also section 4.2).
•
If the wiring is just for one application you can screw on the front panels now.
•
Afterwards attach the lower covers.
NOTE
Please pay attention that you clamp the leadthrough bushes between cover and housing in the intended drillings.
Figure 3-8: C-Box control room (drawing no. 16006.1.006/2)
12 D418.131.00/01 en
PTW C-Box
Assembly for application MP2 and MP3
Figure 3-9: C-Box treatment room (drawing no. 16006.1.006/1)
Figure 3-10: Cable (in the wall) RS232 (drawing no. 16006.1.011/1)
D418.131.00/01 en 13
PTW C-Box
Assembly for application MULTIDOS with LA48
4 Assembly for application MULTIDOS with LA48
4.1 Control unit in the control room
Figure 4-1: Wiring RS232 (LA48) / Monitor (drawing no. 16006.9.003/1)
For this application the MULTIDOS is in the treatment room together with the ME48. The wiring has to be done generally in addition to the wiring for the use of the control unit in the control room as de-
The following cable connections are necessary
− RS232 cable between MULTIDOS in the treatment room and PC in the control room consisting of 2 x T22373/K67 and cable (in the wall)
T16006.1.011
− If the monitor signal of the linear accelerator is used as a reference signal it also has to be fed into the ME48.
14 D418.131.00/01 en
PTW C-Box
Assembly for application MULTIDOS with LA48
− If the monitor signal is only available in the control room it has to be fed through the C-Box into the treatment room to the ME48 (cable
•
If the wiring is just for one application you can screw on the front panels now.
T43164/K4 and T16006.1.012).
•
Afterwards attach the lower covers.
− For the application of a reference chamber it must be plugged into the chamber socket of the
ME48.
NOTE
Please pay attention that you clamp the leadthrough bushes between cover and housing in the intended drillings. Assembly:
•
You have to rip open the front foil along the perforation for the installation of the 9-pin Sub-Dconnectors as well as the monitor signal connector.
•
Equip the C-Box in the control room with 9-pin
Sub-D-socket and the 7-pin monitor plug accord-
•
Equip the C-Box in the treatment room with the
9-pin Sub-D-plug and the 3-pin monitor socket
•
Pull both raw cables through the wall.
•
Slide at both ends a lead-through bush over both cables.
•
Afterwards pull the cables from below into the
C-Box.
•
Solder the cable for the RS232 interface on both sides to the Sub-D-connectors according to
•
Wire the monitor cable according to Figure 4-5.
WARNING
The given wire colors and the pin connections are very important because the wires of the cables are twisted in pairs.
D418.131.00/01 en 15
PTW C-Box
Assembly for application MULTIDOS with LA48
Figure 4-2: C-Box control room (drawing no. 16006.1.004/2)
Figure 4-3: C-Box treatment room (drawing no. 16006.1.004/1)
16 D418.131.00/01 en
PTW C-Box
Assembly for application MULTIDOS with LA48
Figure 4-4: Cable (in the wall) RS232 (drawing no. 16006.1.011/1)
Figure 4-5: Cable (in the wall) Monitor (drawing no. 16006.1.012/1)
D418.131.00/01 en 17
PTW C-Box
Assembly for application MULTIDOS with LA48
4.2 Control unit and TANDEM in the treatment room
Figure 4-6: Wiring MP3 - LA48 (drawing no. 16006.9.007/1)
For this application the MULTIDOS is in the treatment room together with the ME48. The wiring has to be done generally in addition to the wiring for the use of the control unit T41013 with TANDEM in the
treatment room as described in section 3.2.
The following cable connections are necessary
− RS232 cable between control unit in the treatment room and PC in the control room consisting of 2 x T22373/K67 and cable (in the wall)
T16006.1.011
− special RS232 cable T25015 between
MULTIDOS and control unit or RS232 cable
T22373/K67 between MULTIDOS and PC
NOTE
The cable T22373/K67 could not be routed through the C-Box because the C-Box could only be equipped with one RS232 cable.
− If the monitor signal of the linear accelerator is used as a reference signal it also has to be fed into the ME48.
18 D418.131.00/01 en
PTW C-Box
Assembly for application MULTIDOS with LA48
− If the monitor signal is only available in the control room it has to be fed through the C-Box into the treatment room to the ME48 (cable
•
If the wiring is just for one application you can screw on the front panels now.
T43164/K4 and T16006.1.012, refer to
•
Afterwards attach the lower covers.
NOTE
− For the application of a reference chamber it must be plugged into the chamber socket of the
ME48.
Please pay attention that you clamp the leadthrough bushes between cover and housing in the intended drillings.
Assembly:
•
You have to rip open the front foil along the perforation for the installation of 9-pin
Sub-D-connectors as well as the monitor signal connector.
•
Connect either the special RS232 cable T25015
(refer to Figure 4-11) between MULTIDOS and
control unit or connect the RS232 cable
T22373/K67 between MULTIDOS and PC.
•
Equip the C-Box in the control room with 9-pin
Sub-D-socket and the 7-pin monitor plug accord-
•
Equip the C-Box in the treatment room with the
9-pin Sub-D-plug and the 3-pin monitor socket
•
Pull both raw cables through the wall.
•
Slide at both ends a lead-through bush over both cables.
•
Afterwards pull the cables from below into the
C-Box.
•
Solder the cable for the RS232 interface on both sides to the Sub-D-connectors according to
•
Wire the monitor cable according to Figure 4-10.
WARNING
The given wire colors and the pin connections are very important because the wires of the cables are twisted in pairs.
D418.131.00/01 en 19
PTW C-Box
Assembly for application MULTIDOS with LA48
Figure 4-7: C-Box control room (drawing no. 16006.1.004/2)
Figure 4-8: C-Box Box treatment room (drawing no. 16006.1.004/1)
20 D418.131.00/01 en
PTW C-Box
Assembly for application MULTIDOS with LA48
Figure 4-9: Cable (in the wall) RS232 (drawing no. 16006.1.011/1)
Figure 4-10: Cable (in the wall) Monitor (drawing no. 16006.1.012/1)
D418.131.00/01 en 21
PTW C-Box
Assembly for application MULTIDOS with LA48
Figure 4-11: RS232 connection cable / T25015 (drawing no. 25015.1.001/1)
22 D418.131.00/01 en
PTW C-Box
Assembly for application MULTIDOS with QC6Plus
5 Assembly for application MULTIDOS with QC6Plus
Figure 5-1: Wiring MULTIDOS – QC6Plus (drawing no. 16006.9.002/1)
With this application the MULTIDOS is in the control room so that you can use the instrument also without a computer.
The following cable connections are necessary
− cable between QC6Plus in the treatment room and MULTIDOS in the control room consisting of
2 x T26011 and cable (in the wall) T16006.1.010
D418.131.00/01 en 23
Assembly:
•
Pull the cable through the wall. On the delivered cable the 25-pin Sub-D-connectors are already installed on both sides.
PTW C-Box
Assembly for application MULTIDOS with QC6Plus
•
You have to rip open the front foil along the perforation for the installation of the 25-pin Sub-Dconnectors as well as the HV connector.
•
•
After this work is finished, fix the front panels.
Afterwards attach the lower covers.
NOTE
Please pay attention that you clamp the leadthrough bushes between cover and housing in the intended drillings.
WARNING
When you pull the cable through the wall you have to be careful that the 25-pin Sub-Dconnector will not become dirty. Please pack it into a plastic bag before pulling through. You have to pay attention that the 25-pin Sub-Dconnector (plug) will be assembled in the control room.
•
Lead the 25-pin Sub-D-connectors through the front panel diagonally from behind and assemble them with the enclosed locking afterwards.
•
Assemble the HV connectors.
•
Solder up the cable core and the screen of the
HV cable.
•
Equip and solder the C-Box in the control room
according to Figure 5-2 and the C-Box in the treatment room according to Figure 5-3.
NOTE
Before soldering the cable core you have to move on the enclosed shrink-sleeve. After soldering it will be pushed as a protection over the soldered contact and shrunk on with hot air.
24 D418.131.00/01 en
PTW C-Box
Assembly for application MULTIDOS with QC6Plus
Figure 5-2: C-Box control room (drawing no. 16006.1.003/2)
Figure 5-3: C-Box treatment room (drawing no. 16006.1.003/1)
D418.131.00/01 en 25
PTW C-Box
Assembly for application MULTIDOS or VIVODOS with detector box or DIODE MATE
6 Assembly for application MULTIDOS or VIVODOS with detector box or DIODE MATE
Figure 6-1: Wiring MULTIDOS – detector box (drawing no. 16009.9.001/1)
If the MULTIDOS or VIVODOS should only be used with detector box or DIODE MATE, there is no need that the HV connector is installed and wired. Nevertheless, we recommend it to prepare the C-Box for a later use with QC6Plus or with a chamber adapter box.
The following cable connections are necessary
− cable between detector box resp. DIODE MATE in the treatment room and MULTIDOS resp.
VIVODOS in the control room consisting of 2 x
T26024 (alternative T26011 in the control room) and cable (in the wall) T16006.1.010
CAUTION
For this application the MULTIDOS resp.
VIVODOS may not be used in the patient environment, it has to be placed in the control room.
26 D418.131.00/01 en
PTW C-Box
Assembly for application MULTIDOS or VIVODOS with detector box or DIODE MATE
Assembly:
•
You have to rip open the front foil along the perforation for the installation of the 25-pin Sub-Dconnectors as well as the HV connector.
•
Pull the cable through the wall. On the delivered cable the 25-pin Sub-D-connectors are already installed on both sides.
•
•
After this work is finished, fix the front panels.
Afterwards attach the lower covers.
NOTE
Please pay attention that you clamp the leadthrough bushes between cover and housing in the intended drillings.
WARNING
When you pull the cable through the wall you have to be careful that the 25-pin Sub-Dconnector will not become dirty. Please pack it into a plastic bag before pulling through. You have to pay attention that the 25-pin Sub-Dconnector (plug) will be assembled in the control room.
•
Lead the 25-pin Sub-D-connectors through the front panel diagonally from behind and assemble them with the enclosed locking afterwards.
•
Assemble the HV connectors.
•
Solder up the cable core and the screen of the
HV cable.
•
Equip and solder the C-Box in the control room
according to Figure 6-2 and the C-Box in the treatment room according to Figure 6-3.
CAUTION
For safety purposes please use only the cable
T26024 without HV connector to connect the detector box or the DIODE MATE system in the treatment room.
NOTE
Before soldering the cable core you have to move on the enclosed shrink-sleeve. After soldering it will be pushed as a protection over the soldered contact and shrunk on with hot air.
D418.131.00/01 en 27
PTW C-Box
Assembly for application MULTIDOS or VIVODOS with detector box or DIODE MATE
Figure 6-2: C-Box control room (drawing no. 16006.1.003/2)
Figure 6-3: C-Box treatment room (drawing no. 16006.1.003/1)
28 D418.131.00/01 en
PTW C-Box
Assembly for application MULTIDOS or VIVODOS with AL-Box
7 Assembly for application MULTIDOS or VIVODOS with AL-
Box
Figure 7-1: Wiring MULTIDOS – AL-Box (drawing no. 16008.9.001/1)
If the MULTIDOS or VIVODOS should only be used with AL-Box, there is no need that the HV connector is installed and wired. Nevertheless, we recommend it to prepare the C-Box for a later use with
QC6Plus or with a chamber adapter box.
The following cable connections are necessary
− cable between AL-Box in the treatment room and MULTIDOS resp. VIVODOS in the control room consisting of 2 x T26024 (alternative
T26011 in the control room) and cable (in the wall) T16006.1.010
CAUTION
For this application the MULTIDOS resp.
VIVODOS may not be used in the patient environment, it has to be placed in the control room.
D418.131.00/01 en 29
Assembly:
•
Pull the cable through the wall. On the delivered cable the 25-pin Sub-D-connectors are already installed on both sides.
PTW C-Box
Assembly for application MULTIDOS or VIVODOS with AL-Box
•
You have to rip open the front foil along the perforation for the installation of the 25-pin Sub-Dconnectors as well as the HV connector.
•
•
After this work is finished, fix the front panels.
Afterwards attach the lower covers.
NOTE
Please pay attention that you clamp the leadthrough bushes between cover and housing in the intended drillings.
WARNING
When you pull the cable through the wall you have to be careful that the 25-pin Sub-Dconnector will not become dirty. Please pack it into a plastic bag before pulling through. You have to pay attention that the 25-pin Sub-Dconnector (plug) will be assembled in the control room.
•
Lead the 25-pin Sub-D-connectors through the front panel diagonally from behind and assemble them with the enclosed locking afterwards.
•
Assemble the HV connectors.
•
Solder up the cable core and the screen of the
HV cable.
•
Equip and solder the C-Box in the control room
according to Figure 7-2 and the C-Box in the treatment room according to Figure 7-3.
CAUTION
For safety purposes please use only the cable
T26024 without HV connector to connect the
AL-Box in the treatment room.
NOTE
Before soldering the cable core you have to move on the enclosed shrink-sleeve. After soldering it will be pushed as a protection over the soldered contact and shrunk on with hot air.
30 D418.131.00/01 en
PTW C-Box
Assembly for application MULTIDOS or VIVODOS with AL-Box
Figure 7-2: C-Box control room (drawing no. 16006.1.003/2)
Figure 7-3: C-Box treatment room (drawing no. 16006.1.003/1)
D418.131.00/01 en 31
PTW C-Box
Assembly for application MULTIDOS with M-Chamber-Box
8 Assembly for application MULTIDOS with M-Chamber-Box
Figure 8-1: Wiring MULTIDOS – M-Chamber-Box (drawing no. 16007.9.001/1)
The M-Chamber-Box offers the possibility to connect up to twelve detectors with M-Connector to the
MULTIDOS.
CAUTION
The detectors may not be used with patient contact.
The following cable connections are necessary
− cable between M-Chamber-Box in the treatment room and MULTIDOS in the control room consisting of 2 x T26011 and cable (in the wall)
T16006.1.010
32 D418.131.00/01 en
•
Pull the cable through the wall. On the delivered cable the 25-pin Sub-D-connectors are already installed on both sides.
PTW C-Box
Assembly for application MULTIDOS with M-Chamber-Box
Assembly:
•
You have to rip open the front foil along the perforation for the installation of the 25-pin Sub-Dconnectors as well as the HV connector.
•
•
After this work is finished, fix the front panels.
Afterwards attach the lower covers.
NOTE
Please pay attention that you clamp the leadthrough bushes between cover and housing in the intended drillings.
WARNING
When you pull the cable through the wall you have to be careful that the 25-pin Sub-Dconnector will not become dirty. Please pack it into a plastic bag before pulling through. You have to pay attention that the 25-pin Sub-Dconnector (plug) will be assembled in the control room.
•
Lead the 25-pin Sub-D-connectors through the front panel diagonally from behind and assemble them with the enclosed locking afterwards.
•
Assemble the HV connectors.
•
Solder up the cable core and the screen of the
HV cable.
•
Equip and solder the C-Box in the control room
according to Figure 8-2 and the C-Box in the treatment room according to Figure 8-3.
NOTE
Before soldering the cable core you have to move on the enclosed shrink-sleeve. After soldering it will be pushed as a protection over the soldered contact and shrunk on with hot air.
D418.131.00/01 en 33
PTW C-Box
Assembly for application MULTIDOS with M-Chamber-Box
Figure 8-2: C-Box control room (drawing no. 16006.1.003/2)
Figure 8-3: C-Box treatment room (drawing no. 16006.1.003/1)
34 D418.131.00/01 en
PTW C-Box
Assembly for application MULTIDOS with B-Chamber-Box
9 Assembly for application MULTIDOS with B-Chamber-Box
Figure 9-1: Wiring MULTIDOS – B-Chamber-Box (drawing no. 16013.9.001/1)
The M-Chamber-Box offers the possibility to connect up to twelve detectors with B-Connector (BNC
+ Banana plug) to the MULTIDOS.
CAUTION
The detectors may not be used with patient contact.
The following cable connections are necessary
− cable between B-Chamber-Box in the treatment room and MULTIDOS in the control room consisting of 2 x T26011 and cable (in the wall)
T16006.1.010
D418.131.00/01 en 35
Assembly:
•
Pull the cable through the wall. On the delivered cable the 25-pin Sub-D-connectors are already installed on both sides.
PTW C-Box
Assembly for application MULTIDOS with B-Chamber-Box
•
You have to rip open the front foil along the perforation for the installation of the 25-pin Sub-Dconnectors as well as the HV connector.
•
•
After this work is finished, fix the front panels.
Afterwards attach the lower covers.
NOTE
Please pay attention that you clamp the leadthrough bushes between cover and housing in the intended drillings.
WARNING
When you pull the cable through the wall you have to be careful that the 25-pin Sub-Dconnector will not become dirty. Please pack it into a plastic bag before pulling through. You have to pay attention that the 25-pin Sub-Dconnector (plug) will be assembled in the control room.
•
Lead the 25-pin Sub-D-connectors through the front panel diagonally from behind and assemble them with the enclosed locking afterwards.
•
Assemble the HV connectors.
•
Solder up the cable core and the screen of the
HV cable.
•
Equip and solder the C-Box in the control room
according to Figure 9-2 and the C-Box in the treatment room according to Figure 9-3.
NOTE
Before soldering the cable core you have to move on the enclosed shrink-sleeve. After soldering it will be pushed as a protection over the soldered contact and shrunk on with hot air.
36 D418.131.00/01 en
PTW C-Box
Assembly for application MULTIDOS with B-Chamber-Box
Figure 9-2: C-Box control room (drawing no. 16006.1.003/2)
Figure 9-3: C-Box treatment room (drawing no. 16006.1.003/1)
D418.131.00/01 en 37
10 Order Information
PTW C-Box
Order Information
Cabling of MP2 or MP3 / Control unit in the control room according to Section 3.1
C-Box for TBA/MULTIDOS hard-cabling for wall mounting in the treatment room or in the control room.
For installation of cable sets for MP3/MP3-S and/or VIVODOS/MULTIDOS.
Width x height x depth: (131 x 186 x 113) mm
(2 pieces necessary)
TBA cable installation set, with reference probe
Replaces connection cables for MP3 and field probe included in basic equipment.
6 m cable set between MP3 and wall, 20 m for firm installation in the wall, 6 m between wall and control unit. Different lengths available. or
TBA cable installation set, without reference probe
Replaces connection cables for MP3 and field probe included in basic equipment.
6 m cable set between MP3 and wall, 20 m for firm installation in the wall, 6 m between wall and control unit. Different lengths available.
T16006.1.001
L981062
L981063
Cabling of MP2 or MP3 / Control unit and TANDEM in the treatment room according to Section 3.2
C-Box for TBA/MULTIDOS hard-cabling for wall mounting in the treatment room or in the control room.
For installation of cable sets for MP3/MP3-S and/or VIVODOS/MULTIDOS.
Width x height x depth: (131 x 186 x 113) mm
(2 pieces necessary)
Monitor/RS232 cable installation set cable set 20 m for firm installation in the wall. Different lengths available.
RS232 interface cable,
9-pin male / female
Length X = 20 m or 30 m
T16006.1.001
L981061
T22373/K67-X
Cabling of MULTIDOS with LA48 / Control unit in the treatment room according to Section 4.1
C-Box for TBA/MULTIDOS hard-cabling for wall mounting in the treatment room or in the control room.
For installation of cable sets for MP3/MP3-S and/or VIVODOS/MULTIDOS.
Width x height x depth: (131 x 186 x 113) mm
(2 pieces necessary)
Monitor/RS232 cable installation set cable set 20 m for firm installation in the wall. Different lengths available.
Monitor cable for MP3 monitor interface connection cable between monitor/scan interface T431641 and MP3 control unit.
Length 3 m.
RS232 interface cable,
9-pin male / female
Length X = 20 m or 30 m
T16006.1.001
L981061
T43164/K4-3
T22373/K67-X
38 D418.131.00/01 en
Cabling of MULTIDOS with LA48 / Control unit and TANDEM in the treatment room according to Sec-
PTW C-Box
Order Information
C-Box for TBA/MULTIDOS hard-cabling for wall mounting in the treatment room or in the control room.
For installation of cable sets for MP3/MP3-S and/or VIVODOS/MULTIDOS.
Width x height x depth: (131 x 186 x 113) mm
(2 pieces necessary)
Monitor/RS232 cable installation set cable set 20 m for firm installation in the wall. Different lengths available.
Monitor cable for MP3 monitor interface connection cable between monitor/scan interface T431641 and MP3 control unit.
Length 3 m.
RS232 interface cable
25-pin male / 9-pin male.
Connects a MULTIDOS/UNIDOS electrometer to a MP3 control unit T41013,
Length 2 m
RS232 interface cable,
9-pin male / female
Length X = 20 m or 30 m
Cabling of
– MULTIDOS with QC6Plus according to Section 5
– MULTIDOS or VIVODOS with detector box or DIODE MATE according to Section 6
– MULTIDOS or VIVODOS with AL-Box according to Section 7
– MULTIDOS with M-Box according to Section 8
– MULTIDOS with B-Box according to Section 9
T16006.1.001
L981061
T43164/K4-3
T25015-2
T22373/K67-X
C-Box for TBA/MULTIDOS hard-cabling for wall mounting in the treatment room or in the control room.
For installation of cable sets for MP3/MP3-S and/or VIVODOS/MULTIDOS.
Width x height x depth: (131 x 186 x 113) mm
(2 pieces necessary)
VIVODOS/MULTIDOS cable installation set cable set 20 m for firm installation in the wall, 6 m between wall and MULTIDOS. Different lengths available.
T16006.1.001
L981064
D418.131.00/01 en 39