User Manual PTW C-Box

Add to my manuals
39 Pages

advertisement

User Manual PTW C-Box | Manualzz

User Manual

PTW C-Box

Installation and Assembly Manual

Contents

PTW C-Box

Contents

Contents

Illustrations

1 Safety Information

2 General

3 Assembly for application MP2 and MP3

3.1

Control unit in control room

3.2

Control unit and TANDEM in the treatment room

4 Assembly for application MULTIDOS with LA48

4.1

Control unit in the control room

4.2

Control unit and TANDEM in the treatment room

5 Assembly for application MULTIDOS with QC6Plus

6 Assembly for application MULTIDOS or VIVODOS with detector box or DIODE

MATE

7 Assembly for application MULTIDOS or VIVODOS with AL-Box

8 Assembly for application MULTIDOS with M-Chamber-Box

9 Assembly for application MULTIDOS with B-Chamber-Box

10 Order Information

26

29

32

35

38

6

6

11

14

14

18

4

5

23

2

3

2 D418.131.00/01 en

PTW C-Box

Illustrations

Illustrations

Figure 3-1: Wiring MP3 – Tandem (drawing no. 16006.9.005/1)

Figure 3-2: Wiring MP3 / MP3-S (drawing no. 16006.9.001/1)

Figure 3-3: C-Box control room (drawing no. 16006.1.002/2)

Figure 3-4: C-Box treatment room (drawing no. 16006.1.002/1)

Figure 3-5: Connection of control cable L173038

Figure 3-6: Connection of motor cable L173015

Figure 3-7: Wiring MP3 – TANDEM (drawing no. 16006.9.006/1)

Figure 3-8: C-Box control room (drawing no. 16006.1.006/2)

Figure 3-9: C-Box treatment room (drawing no. 16006.1.006/1)

Figure 3-10: Cable (in the wall) RS232 (drawing no. 16006.1.011/1)

Figure 4-1: Wiring RS232 (LA48) / Monitor (drawing no. 16006.9.003/1)

Figure 4-2: C-Box control room (drawing no. 16006.1.004/2)

Figure 4-3: C-Box treatment room (drawing no. 16006.1.004/1)

Figure 4-4: Cable (in the wall) RS232 (drawing no. 16006.1.011/1)

Figure 4-5: Cable (in the wall) Monitor (drawing no. 16006.1.012/1)

Figure 4-6: Wiring MP3 - LA48 (drawing no. 16006.9.007/1)

Figure 4-7: C-Box control room (drawing no. 16006.1.004/2)

Figure 4-8: C-Box Box treatment room (drawing no. 16006.1.004/1)

Figure 4-9: Cable (in the wall) RS232 (drawing no. 16006.1.011/1)

Figure 4-10: Cable (in the wall) Monitor (drawing no. 16006.1.012/1)

Figure 4-11: RS232 connection cable / T25015 (drawing no. 25015.1.001/1)

Figure 5-1: Wiring MULTIDOS – QC6Plus (drawing no. 16006.9.002/1)

Figure 5-2: C-Box control room (drawing no. 16006.1.003/2)

Figure 5-3: C-Box treatment room (drawing no. 16006.1.003/1)

Figure 6-1: Wiring MULTIDOS – detector box (drawing no. 16009.9.001/1)

Figure 6-2: C-Box control room (drawing no. 16006.1.003/2)

Figure 6-3: C-Box treatment room (drawing no. 16006.1.003/1)

Figure 7-1: Wiring MULTIDOS – AL-Box (drawing no. 16008.9.001/1)

Figure 7-2: C-Box control room (drawing no. 16006.1.003/2)

Figure 7-3: C-Box treatment room (drawing no. 16006.1.003/1)

Figure 8-1: Wiring MULTIDOS – M-Chamber-Box (drawing no. 16007.9.001/1)

Figure 8-2: C-Box control room (drawing no. 16006.1.003/2)

Figure 8-3: C-Box treatment room (drawing no. 16006.1.003/1)

Figure 9-1: Wiring MULTIDOS – B-Chamber-Box (drawing no. 16013.9.001/1)

Figure 9-2: C-Box control room (drawing no. 16006.1.003/2)

Figure 9-3: C-Box treatment room (drawing no. 16006.1.003/1)

26

28

28

29

31

31

32

34

20

20

21

21

22

23

25

25

34

35

37

37

13

13

14

16

16

17

17

18

9

10

11

12

6

7

8

9

D418.131.00/01 en 3

1 Safety Information

NOTE

The C-Box is a component of the following systems:

− MP2 / MP3 / MP3-S / MP3-M / MP3-XS and similar moving mechanisms

− MULTIDOS

− VIVODOS

− DIODE MATE

Follow the Safety instructions and General Information given in the manuals to the mentioned systems.

WARNING

Installing the C-Box requires technical knowhow and may only be performed by trained technical staff.

The relevant regulations concerning electric safety must especially be observed.

Non-compliance with any of these stipulations voids any liability of PTW for operation and safety of the equipment in question.

PTW-FREIBURG

Physikalisch-Technische Werkstätten

Dr. Pychlau GmbH

Lörracher Str. 7

79115 FREIBURG

GERMANY

Phone: +49 761 49055-0

Fax: +49 761 49055-70 [email protected] www.ptw.de

PTW C-Box

Safety Information

4 D418.131.00/01 en

PTW C-Box

General

2 General

The PTW - C-Box is a universal wall box for the connecting of PTW-Instruments and for various applications.

One C-Box has to be installed in the treatment room and one in the control room. This allows to handle the cables of different measuring systems in a simple and time-saving way.

The advantage of the C-Box is, that you can install only one application at first and integrate other applications later without difficulty.

Following instruments and systems can be connected with the C-Box if it is equipped completely:

− MP2- or MP3-system (as well MP3-S, MP3-M,

MP3-XS) inclusive field detector and reference detector or a similar moving mechanism

− MULTIDOS for LA48 application

(LA = Linear Array)

− MULTIDOS for QC6Plus application

(QC = Quality Control)

− MULTIDOS or VIVODOS for TBI application

(TBI = Total Body Irradiation)

− MULTIDOS or VIVODOS for AL application

(AL = Afterloading)

− MULTIDOS or VIVODOS with DIODE MATE

− MULTIDOS with M-Chamber-Box (Connection box for up to 12 detectors with M-Connector)

− MULTIDOS with B-Chamber-Box (Connection box for up to 12 detectors with BNC-Connector and Banana plug)

NOTE

The C-Box should be installed in connection with a cable channel that the cables are routed through the cable channel from below into the

C-Box. The strain-relief will be guaranteed after fixation of the lower cover.

In the case the cables are pulled through the wall directly from behind necessary wholes have to be drilled. In that case the strain relief is under your own responsibility.

For all applications you have first to unscrew the front panel of the C-Box and the lower cover. Afterwards the C-Box can be fixed to the wall using the enclosed mounting kit (4 screws/4 dowels).

After the connection of the cables you should sign them with the enclosed cable markers adequate to the respective drawings.

D418.131.00/01 en 5

PTW C-Box

Assembly for application MP2 and MP3

3 Assembly for application MP2 and MP3

NOTE

The following instructions apply for the systems

MP2, MP3, MP3-S, MP3-M, MP3-XS and all similar moving mechanisms.

3.1 Control unit in control room

Possible versions:

1. control unit T41013 with separate TANDEM

electrometer, refer to Figure 3-1

2. control unit T43164/U600 resp. T41004, refer to

Figure 3-2

The following cable connections are necessary

(refer to Figure 3-1 and Figure 3-2):

− control cable between water phantom in the treatment room and control unit in the control room consisting of 2 x T43164/K1 and cable (in the wall) L173038

− motor cable between water phantom in the treatment room and control unit in the control room consisting of 2 x T43164/K2 and cable (in the wall) L173015

− cable for detector between water phantom in the treatment room and TANDEM or control unit

(T43164/U600 resp. T41004) in the control room

− cable for reference detector (if fitted) between water phantom in the treatment room and

TANDEM or control unit (T43164/U600 resp.

T41004) in the control room

Figure 3-1: Wiring MP3 – Tandem (drawing no. 16006.9.005/1)

6 D418.131.00/01 en

PTW C-Box

Assembly for application MP2 and MP3

Figure 3-2: Wiring MP3 / MP3-S (drawing no. 16006.9.001/1)

Assembly:

Afterwards pull the cables from below into the

C-Box.

You have to rip open the front foil along the perforation for the installation of the 37-pin and 15pin Sub-D-connectors.

Remove the cable jacket on both sides on approx. 45 mm.

Equip the C-Box in the control room with the

37-pin and the 15-pin Sub-D-plug according to

Figure 3-3.

Equip the C-Box in the treatment room with the

37-pin and the 15-pin Sub-D-socket according to

Figure 3-4.

Untie and twist the screen and sold it to the sol-

dering ring according to Figure 3-3 and

Figure 3-4.

Solder the control cable on both sides to the

Sub-D-connectors according to Figure 3-5.

Solder the motor cable on both sides to the Sub-

D-connectors according to Figure 3-6.

Pull both raw cables through the wall.

Slide at both ends a lead-through bush over both cables.

WARNING

The given wire colors and the pin connections are very important because the wires of the cables are twisted in pairs.

D418.131.00/01 en 7

PTW C-Box

Assembly for application MP2 and MP3

Fix the plug holders for the chamber cable

(T2664/U1) on the right and the left hand side of the C-Box with the delivered screws.

If the wiring is just for one application you can screw on the front panels now.

Afterwards attach the lower covers.

NOTE

Please pay attention that you clamp the leadthrough bushes between cover and housing in the intended drillings.

Attach the plugs or the sockets of the detector cables in the holder of the left and right side.

Figure 3-3: C-Box control room (drawing no. 16006.1.002/2)

8 D418.131.00/01 en

PTW C-Box

Assembly for application MP2 and MP3

Figure 3-4: C-Box treatment room (drawing no. 16006.1.002/1)

Figure 3-5: Connection of control cable L173038

D418.131.00/01 en 9

PTW C-Box

Assembly for application MP2 and MP3

Figure 3-6: Connection of motor cable L173015

10 D418.131.00/01 en

PTW C-Box

Assembly for application MP2 and MP3

3.2 Control unit and TANDEM in the treatment room

Figure 3-7: Wiring MP3 – TANDEM (drawing no. 16006.9.006/1)

This assembly is only possible with control unit

T41013 with separate TANDEM electrometer.

The following cable connections are necessary

(refer to Figure 3-7):

− RS232 cable between control unit in the treatment room and PC in the control room consisting of 2 x T22373/K67and cable (in the wall)

T16006.1.011

Assembly:

You have to rip open the front foil along the perforation for the installation of the 9-pin Sub-Dconnectors.

Equip the C-Box in the control room with 9-pin

Sub-D-socket according to Figure 3-8.

Equip the C-Box in the treatment room with the

9-pin Sub-D-plug according to Figure 3-9.

Pull the raw cable through the wall.

Slide at both ends a lead-through bush over both cables.

Afterwards pull the cables from below into the

C-Box.

D418.131.00/01 en 11

Figure 3-10.

PTW C-Box

Assembly for application MP2 and MP3

Solder the cable for the RS232 interface on both sides to the Sub-D-connectors according to

NOTE

WARNING

The given wire colors and the pin connections are very important because the wires of the cables are twisted in pairs.

Instead of the TANDEM you can also use an

UNIDOS or MULTIDOS as electrometer. To connect the UNIDOS or MULTIDOS to the MP3 control unit please use the special RS232 cable

T25015 (see also section 4.2).

If the wiring is just for one application you can screw on the front panels now.

Afterwards attach the lower covers.

NOTE

Please pay attention that you clamp the leadthrough bushes between cover and housing in the intended drillings.

Figure 3-8: C-Box control room (drawing no. 16006.1.006/2)

12 D418.131.00/01 en

PTW C-Box

Assembly for application MP2 and MP3

Figure 3-9: C-Box treatment room (drawing no. 16006.1.006/1)

Figure 3-10: Cable (in the wall) RS232 (drawing no. 16006.1.011/1)

D418.131.00/01 en 13

PTW C-Box

Assembly for application MULTIDOS with LA48

4 Assembly for application MULTIDOS with LA48

4.1 Control unit in the control room

Figure 4-1: Wiring RS232 (LA48) / Monitor (drawing no. 16006.9.003/1)

For this application the MULTIDOS is in the treatment room together with the ME48. The wiring has to be done generally in addition to the wiring for the use of the control unit in the control room as de-

scribed in section 3.1.

The following cable connections are necessary

(refer to Figure 4-1):

− RS232 cable between MULTIDOS in the treatment room and PC in the control room consisting of 2 x T22373/K67 and cable (in the wall)

T16006.1.011

− If the monitor signal of the linear accelerator is used as a reference signal it also has to be fed into the ME48.

14 D418.131.00/01 en

PTW C-Box

Assembly for application MULTIDOS with LA48

− If the monitor signal is only available in the control room it has to be fed through the C-Box into the treatment room to the ME48 (cable

If the wiring is just for one application you can screw on the front panels now.

T43164/K4 and T16006.1.012).

Afterwards attach the lower covers.

− For the application of a reference chamber it must be plugged into the chamber socket of the

ME48.

NOTE

Please pay attention that you clamp the leadthrough bushes between cover and housing in the intended drillings. Assembly:

You have to rip open the front foil along the perforation for the installation of the 9-pin Sub-Dconnectors as well as the monitor signal connector.

Equip the C-Box in the control room with 9-pin

Sub-D-socket and the 7-pin monitor plug accord-

ing to Figure 4-2.

Equip the C-Box in the treatment room with the

9-pin Sub-D-plug and the 3-pin monitor socket

according to Figure 4-3.

Pull both raw cables through the wall.

Slide at both ends a lead-through bush over both cables.

Afterwards pull the cables from below into the

C-Box.

Solder the cable for the RS232 interface on both sides to the Sub-D-connectors according to

Figure 4-4.

Wire the monitor cable according to Figure 4-5.

WARNING

The given wire colors and the pin connections are very important because the wires of the cables are twisted in pairs.

D418.131.00/01 en 15

PTW C-Box

Assembly for application MULTIDOS with LA48

Figure 4-2: C-Box control room (drawing no. 16006.1.004/2)

Figure 4-3: C-Box treatment room (drawing no. 16006.1.004/1)

16 D418.131.00/01 en

PTW C-Box

Assembly for application MULTIDOS with LA48

Figure 4-4: Cable (in the wall) RS232 (drawing no. 16006.1.011/1)

Figure 4-5: Cable (in the wall) Monitor (drawing no. 16006.1.012/1)

D418.131.00/01 en 17

PTW C-Box

Assembly for application MULTIDOS with LA48

4.2 Control unit and TANDEM in the treatment room

Figure 4-6: Wiring MP3 - LA48 (drawing no. 16006.9.007/1)

For this application the MULTIDOS is in the treatment room together with the ME48. The wiring has to be done generally in addition to the wiring for the use of the control unit T41013 with TANDEM in the

treatment room as described in section 3.2.

The following cable connections are necessary

(refer to Figure 4-6):

− RS232 cable between control unit in the treatment room and PC in the control room consisting of 2 x T22373/K67 and cable (in the wall)

T16006.1.011

− special RS232 cable T25015 between

MULTIDOS and control unit or RS232 cable

T22373/K67 between MULTIDOS and PC

NOTE

The cable T22373/K67 could not be routed through the C-Box because the C-Box could only be equipped with one RS232 cable.

− If the monitor signal of the linear accelerator is used as a reference signal it also has to be fed into the ME48.

18 D418.131.00/01 en

PTW C-Box

Assembly for application MULTIDOS with LA48

− If the monitor signal is only available in the control room it has to be fed through the C-Box into the treatment room to the ME48 (cable

If the wiring is just for one application you can screw on the front panels now.

T43164/K4 and T16006.1.012, refer to

Figure 4-1).

Afterwards attach the lower covers.

NOTE

− For the application of a reference chamber it must be plugged into the chamber socket of the

ME48.

Please pay attention that you clamp the leadthrough bushes between cover and housing in the intended drillings.

Assembly:

You have to rip open the front foil along the perforation for the installation of 9-pin

Sub-D-connectors as well as the monitor signal connector.

Connect either the special RS232 cable T25015

(refer to Figure 4-11) between MULTIDOS and

control unit or connect the RS232 cable

T22373/K67 between MULTIDOS and PC.

Equip the C-Box in the control room with 9-pin

Sub-D-socket and the 7-pin monitor plug accord-

ing to Figure 4-7.

Equip the C-Box in the treatment room with the

9-pin Sub-D-plug and the 3-pin monitor socket

according to Figure 4-8

Pull both raw cables through the wall.

Slide at both ends a lead-through bush over both cables.

Afterwards pull the cables from below into the

C-Box.

Solder the cable for the RS232 interface on both sides to the Sub-D-connectors according to

Figure 4-9.

Wire the monitor cable according to Figure 4-10.

WARNING

The given wire colors and the pin connections are very important because the wires of the cables are twisted in pairs.

D418.131.00/01 en 19

PTW C-Box

Assembly for application MULTIDOS with LA48

Figure 4-7: C-Box control room (drawing no. 16006.1.004/2)

Figure 4-8: C-Box Box treatment room (drawing no. 16006.1.004/1)

20 D418.131.00/01 en

PTW C-Box

Assembly for application MULTIDOS with LA48

Figure 4-9: Cable (in the wall) RS232 (drawing no. 16006.1.011/1)

Figure 4-10: Cable (in the wall) Monitor (drawing no. 16006.1.012/1)

D418.131.00/01 en 21

PTW C-Box

Assembly for application MULTIDOS with LA48

Figure 4-11: RS232 connection cable / T25015 (drawing no. 25015.1.001/1)

22 D418.131.00/01 en

PTW C-Box

Assembly for application MULTIDOS with QC6Plus

5 Assembly for application MULTIDOS with QC6Plus

Figure 5-1: Wiring MULTIDOS – QC6Plus (drawing no. 16006.9.002/1)

With this application the MULTIDOS is in the control room so that you can use the instrument also without a computer.

The following cable connections are necessary

(refer to Figure 5-1):

− cable between QC6Plus in the treatment room and MULTIDOS in the control room consisting of

2 x T26011 and cable (in the wall) T16006.1.010

D418.131.00/01 en 23

Assembly:

Pull the cable through the wall. On the delivered cable the 25-pin Sub-D-connectors are already installed on both sides.

PTW C-Box

Assembly for application MULTIDOS with QC6Plus

You have to rip open the front foil along the perforation for the installation of the 25-pin Sub-Dconnectors as well as the HV connector.

After this work is finished, fix the front panels.

Afterwards attach the lower covers.

NOTE

Please pay attention that you clamp the leadthrough bushes between cover and housing in the intended drillings.

WARNING

When you pull the cable through the wall you have to be careful that the 25-pin Sub-Dconnector will not become dirty. Please pack it into a plastic bag before pulling through. You have to pay attention that the 25-pin Sub-Dconnector (plug) will be assembled in the control room.

Lead the 25-pin Sub-D-connectors through the front panel diagonally from behind and assemble them with the enclosed locking afterwards.

Assemble the HV connectors.

Solder up the cable core and the screen of the

HV cable.

Equip and solder the C-Box in the control room

according to Figure 5-2 and the C-Box in the treatment room according to Figure 5-3.

NOTE

Before soldering the cable core you have to move on the enclosed shrink-sleeve. After soldering it will be pushed as a protection over the soldered contact and shrunk on with hot air.

24 D418.131.00/01 en

PTW C-Box

Assembly for application MULTIDOS with QC6Plus

Figure 5-2: C-Box control room (drawing no. 16006.1.003/2)

Figure 5-3: C-Box treatment room (drawing no. 16006.1.003/1)

D418.131.00/01 en 25

PTW C-Box

Assembly for application MULTIDOS or VIVODOS with detector box or DIODE MATE

6 Assembly for application MULTIDOS or VIVODOS with detector box or DIODE MATE

Figure 6-1: Wiring MULTIDOS – detector box (drawing no. 16009.9.001/1)

If the MULTIDOS or VIVODOS should only be used with detector box or DIODE MATE, there is no need that the HV connector is installed and wired. Nevertheless, we recommend it to prepare the C-Box for a later use with QC6Plus or with a chamber adapter box.

The following cable connections are necessary

(refer to Figure 6-1):

− cable between detector box resp. DIODE MATE in the treatment room and MULTIDOS resp.

VIVODOS in the control room consisting of 2 x

T26024 (alternative T26011 in the control room) and cable (in the wall) T16006.1.010

CAUTION

For this application the MULTIDOS resp.

VIVODOS may not be used in the patient environment, it has to be placed in the control room.

26 D418.131.00/01 en

PTW C-Box

Assembly for application MULTIDOS or VIVODOS with detector box or DIODE MATE

Assembly:

You have to rip open the front foil along the perforation for the installation of the 25-pin Sub-Dconnectors as well as the HV connector.

Pull the cable through the wall. On the delivered cable the 25-pin Sub-D-connectors are already installed on both sides.

After this work is finished, fix the front panels.

Afterwards attach the lower covers.

NOTE

Please pay attention that you clamp the leadthrough bushes between cover and housing in the intended drillings.

WARNING

When you pull the cable through the wall you have to be careful that the 25-pin Sub-Dconnector will not become dirty. Please pack it into a plastic bag before pulling through. You have to pay attention that the 25-pin Sub-Dconnector (plug) will be assembled in the control room.

Lead the 25-pin Sub-D-connectors through the front panel diagonally from behind and assemble them with the enclosed locking afterwards.

Assemble the HV connectors.

Solder up the cable core and the screen of the

HV cable.

Equip and solder the C-Box in the control room

according to Figure 6-2 and the C-Box in the treatment room according to Figure 6-3.

CAUTION

For safety purposes please use only the cable

T26024 without HV connector to connect the detector box or the DIODE MATE system in the treatment room.

NOTE

Before soldering the cable core you have to move on the enclosed shrink-sleeve. After soldering it will be pushed as a protection over the soldered contact and shrunk on with hot air.

D418.131.00/01 en 27

PTW C-Box

Assembly for application MULTIDOS or VIVODOS with detector box or DIODE MATE

Figure 6-2: C-Box control room (drawing no. 16006.1.003/2)

Figure 6-3: C-Box treatment room (drawing no. 16006.1.003/1)

28 D418.131.00/01 en

PTW C-Box

Assembly for application MULTIDOS or VIVODOS with AL-Box

7 Assembly for application MULTIDOS or VIVODOS with AL-

Box

Figure 7-1: Wiring MULTIDOS – AL-Box (drawing no. 16008.9.001/1)

If the MULTIDOS or VIVODOS should only be used with AL-Box, there is no need that the HV connector is installed and wired. Nevertheless, we recommend it to prepare the C-Box for a later use with

QC6Plus or with a chamber adapter box.

The following cable connections are necessary

(refer to Figure 7-1):

− cable between AL-Box in the treatment room and MULTIDOS resp. VIVODOS in the control room consisting of 2 x T26024 (alternative

T26011 in the control room) and cable (in the wall) T16006.1.010

CAUTION

For this application the MULTIDOS resp.

VIVODOS may not be used in the patient environment, it has to be placed in the control room.

D418.131.00/01 en 29

Assembly:

Pull the cable through the wall. On the delivered cable the 25-pin Sub-D-connectors are already installed on both sides.

PTW C-Box

Assembly for application MULTIDOS or VIVODOS with AL-Box

You have to rip open the front foil along the perforation for the installation of the 25-pin Sub-Dconnectors as well as the HV connector.

After this work is finished, fix the front panels.

Afterwards attach the lower covers.

NOTE

Please pay attention that you clamp the leadthrough bushes between cover and housing in the intended drillings.

WARNING

When you pull the cable through the wall you have to be careful that the 25-pin Sub-Dconnector will not become dirty. Please pack it into a plastic bag before pulling through. You have to pay attention that the 25-pin Sub-Dconnector (plug) will be assembled in the control room.

Lead the 25-pin Sub-D-connectors through the front panel diagonally from behind and assemble them with the enclosed locking afterwards.

Assemble the HV connectors.

Solder up the cable core and the screen of the

HV cable.

Equip and solder the C-Box in the control room

according to Figure 7-2 and the C-Box in the treatment room according to Figure 7-3.

CAUTION

For safety purposes please use only the cable

T26024 without HV connector to connect the

AL-Box in the treatment room.

NOTE

Before soldering the cable core you have to move on the enclosed shrink-sleeve. After soldering it will be pushed as a protection over the soldered contact and shrunk on with hot air.

30 D418.131.00/01 en

PTW C-Box

Assembly for application MULTIDOS or VIVODOS with AL-Box

Figure 7-2: C-Box control room (drawing no. 16006.1.003/2)

Figure 7-3: C-Box treatment room (drawing no. 16006.1.003/1)

D418.131.00/01 en 31

PTW C-Box

Assembly for application MULTIDOS with M-Chamber-Box

8 Assembly for application MULTIDOS with M-Chamber-Box

Figure 8-1: Wiring MULTIDOS – M-Chamber-Box (drawing no. 16007.9.001/1)

The M-Chamber-Box offers the possibility to connect up to twelve detectors with M-Connector to the

MULTIDOS.

CAUTION

The detectors may not be used with patient contact.

The following cable connections are necessary

(refer to Figure 8-1):

− cable between M-Chamber-Box in the treatment room and MULTIDOS in the control room consisting of 2 x T26011 and cable (in the wall)

T16006.1.010

32 D418.131.00/01 en

Pull the cable through the wall. On the delivered cable the 25-pin Sub-D-connectors are already installed on both sides.

PTW C-Box

Assembly for application MULTIDOS with M-Chamber-Box

Assembly:

You have to rip open the front foil along the perforation for the installation of the 25-pin Sub-Dconnectors as well as the HV connector.

After this work is finished, fix the front panels.

Afterwards attach the lower covers.

NOTE

Please pay attention that you clamp the leadthrough bushes between cover and housing in the intended drillings.

WARNING

When you pull the cable through the wall you have to be careful that the 25-pin Sub-Dconnector will not become dirty. Please pack it into a plastic bag before pulling through. You have to pay attention that the 25-pin Sub-Dconnector (plug) will be assembled in the control room.

Lead the 25-pin Sub-D-connectors through the front panel diagonally from behind and assemble them with the enclosed locking afterwards.

Assemble the HV connectors.

Solder up the cable core and the screen of the

HV cable.

Equip and solder the C-Box in the control room

according to Figure 8-2 and the C-Box in the treatment room according to Figure 8-3.

NOTE

Before soldering the cable core you have to move on the enclosed shrink-sleeve. After soldering it will be pushed as a protection over the soldered contact and shrunk on with hot air.

D418.131.00/01 en 33

PTW C-Box

Assembly for application MULTIDOS with M-Chamber-Box

Figure 8-2: C-Box control room (drawing no. 16006.1.003/2)

Figure 8-3: C-Box treatment room (drawing no. 16006.1.003/1)

34 D418.131.00/01 en

PTW C-Box

Assembly for application MULTIDOS with B-Chamber-Box

9 Assembly for application MULTIDOS with B-Chamber-Box

Figure 9-1: Wiring MULTIDOS – B-Chamber-Box (drawing no. 16013.9.001/1)

The M-Chamber-Box offers the possibility to connect up to twelve detectors with B-Connector (BNC

+ Banana plug) to the MULTIDOS.

CAUTION

The detectors may not be used with patient contact.

The following cable connections are necessary

(refer to Figure 9-1):

− cable between B-Chamber-Box in the treatment room and MULTIDOS in the control room consisting of 2 x T26011 and cable (in the wall)

T16006.1.010

D418.131.00/01 en 35

Assembly:

Pull the cable through the wall. On the delivered cable the 25-pin Sub-D-connectors are already installed on both sides.

PTW C-Box

Assembly for application MULTIDOS with B-Chamber-Box

You have to rip open the front foil along the perforation for the installation of the 25-pin Sub-Dconnectors as well as the HV connector.

After this work is finished, fix the front panels.

Afterwards attach the lower covers.

NOTE

Please pay attention that you clamp the leadthrough bushes between cover and housing in the intended drillings.

WARNING

When you pull the cable through the wall you have to be careful that the 25-pin Sub-Dconnector will not become dirty. Please pack it into a plastic bag before pulling through. You have to pay attention that the 25-pin Sub-Dconnector (plug) will be assembled in the control room.

Lead the 25-pin Sub-D-connectors through the front panel diagonally from behind and assemble them with the enclosed locking afterwards.

Assemble the HV connectors.

Solder up the cable core and the screen of the

HV cable.

Equip and solder the C-Box in the control room

according to Figure 9-2 and the C-Box in the treatment room according to Figure 9-3.

NOTE

Before soldering the cable core you have to move on the enclosed shrink-sleeve. After soldering it will be pushed as a protection over the soldered contact and shrunk on with hot air.

36 D418.131.00/01 en

PTW C-Box

Assembly for application MULTIDOS with B-Chamber-Box

Figure 9-2: C-Box control room (drawing no. 16006.1.003/2)

Figure 9-3: C-Box treatment room (drawing no. 16006.1.003/1)

D418.131.00/01 en 37

10 Order Information

PTW C-Box

Order Information

Cabling of MP2 or MP3 / Control unit in the control room according to Section 3.1

C-Box for TBA/MULTIDOS hard-cabling for wall mounting in the treatment room or in the control room.

For installation of cable sets for MP3/MP3-S and/or VIVODOS/MULTIDOS.

Width x height x depth: (131 x 186 x 113) mm

(2 pieces necessary)

TBA cable installation set, with reference probe

Replaces connection cables for MP3 and field probe included in basic equipment.

6 m cable set between MP3 and wall, 20 m for firm installation in the wall, 6 m between wall and control unit. Different lengths available. or

TBA cable installation set, without reference probe

Replaces connection cables for MP3 and field probe included in basic equipment.

6 m cable set between MP3 and wall, 20 m for firm installation in the wall, 6 m between wall and control unit. Different lengths available.

T16006.1.001

L981062

L981063

Cabling of MP2 or MP3 / Control unit and TANDEM in the treatment room according to Section 3.2

C-Box for TBA/MULTIDOS hard-cabling for wall mounting in the treatment room or in the control room.

For installation of cable sets for MP3/MP3-S and/or VIVODOS/MULTIDOS.

Width x height x depth: (131 x 186 x 113) mm

(2 pieces necessary)

Monitor/RS232 cable installation set cable set 20 m for firm installation in the wall. Different lengths available.

RS232 interface cable,

9-pin male / female

Length X = 20 m or 30 m

T16006.1.001

L981061

T22373/K67-X

Cabling of MULTIDOS with LA48 / Control unit in the treatment room according to Section 4.1

C-Box for TBA/MULTIDOS hard-cabling for wall mounting in the treatment room or in the control room.

For installation of cable sets for MP3/MP3-S and/or VIVODOS/MULTIDOS.

Width x height x depth: (131 x 186 x 113) mm

(2 pieces necessary)

Monitor/RS232 cable installation set cable set 20 m for firm installation in the wall. Different lengths available.

Monitor cable for MP3 monitor interface connection cable between monitor/scan interface T431641 and MP3 control unit.

Length 3 m.

RS232 interface cable,

9-pin male / female

Length X = 20 m or 30 m

T16006.1.001

L981061

T43164/K4-3

T22373/K67-X

38 D418.131.00/01 en

Cabling of MULTIDOS with LA48 / Control unit and TANDEM in the treatment room according to Sec-

tion 4.2

PTW C-Box

Order Information

C-Box for TBA/MULTIDOS hard-cabling for wall mounting in the treatment room or in the control room.

For installation of cable sets for MP3/MP3-S and/or VIVODOS/MULTIDOS.

Width x height x depth: (131 x 186 x 113) mm

(2 pieces necessary)

Monitor/RS232 cable installation set cable set 20 m for firm installation in the wall. Different lengths available.

Monitor cable for MP3 monitor interface connection cable between monitor/scan interface T431641 and MP3 control unit.

Length 3 m.

RS232 interface cable

25-pin male / 9-pin male.

Connects a MULTIDOS/UNIDOS electrometer to a MP3 control unit T41013,

Length 2 m

RS232 interface cable,

9-pin male / female

Length X = 20 m or 30 m

Cabling of

– MULTIDOS with QC6Plus according to Section 5

– MULTIDOS or VIVODOS with detector box or DIODE MATE according to Section 6

– MULTIDOS or VIVODOS with AL-Box according to Section 7

– MULTIDOS with M-Box according to Section 8

– MULTIDOS with B-Box according to Section 9

T16006.1.001

L981061

T43164/K4-3

T25015-2

T22373/K67-X

C-Box for TBA/MULTIDOS hard-cabling for wall mounting in the treatment room or in the control room.

For installation of cable sets for MP3/MP3-S and/or VIVODOS/MULTIDOS.

Width x height x depth: (131 x 186 x 113) mm

(2 pieces necessary)

VIVODOS/MULTIDOS cable installation set cable set 20 m for firm installation in the wall, 6 m between wall and MULTIDOS. Different lengths available.

T16006.1.001

L981064

D418.131.00/01 en 39