Publication number DS1-010133

May. 2006

UltraScope For DS1000 Series

© Copyright Rigol Technologies, Inc. 2006

All Rights Reserved

Copyright © RIGOL TECHNOLOGIES, INC. 2006 All Rights Reserved.

Rigol products are covered by P.R. China and foreign patents, issued and pending.

Information in this publication supercedes that in all previously published material.

Specification and price change privileges reserved.

NOTE: RIGOL is registered trademark of RIGOL TECHNOLOGIES, INC.

Copyright 2006 Rigol Technologies, Inc.

2

UltraScope For DS1000

Product Description

UltraScope Software introduces a family of Windows 95, 98, 2000, NT, and Windows XP applications that connect your RIGOL DS1000 series oscilloscopes to your PC desktop by

USB or RS-232 .

The UltraScope software provides the following control and analysis features: z Use Data Browser to display captured waveforms, data and measurements. z Use DSO Controller to control the oscilloscope locally or over the network. z Export the waveform by “BMP” format. z Save the data into “TXT” or “EXCEL” file for analysis. z Print the waveforms.

Copyright 2006 Rigol Technologies, Inc.

3

UltraScope For DS1000

System Requirements

To install and run UltraScope software, you must have an IBM-compatible PC with the following installed: z Microsoft Windows 95, 98, NT 4.0, 2000, or Windows XP z 16 MB RAM or greater recommended z CD-ROM drive, 4X or better z Super VGA monitor or better z 10 MB of disk space z Available RS-232 COM port

You must have a DS1000 series oscilloscope and the firmware version must be higher or equal to 01.02.22 . To connect to PC, the oscilloscope must install USB dirver or have an available RS-232 COM port.

Copyright 2006 Rigol Technologies, Inc.

4

UltraScope For DS1000

Contents

This manual is the user guide for the Ultrascope software, and it contains four chapters.

Chapter 1: Installation and Uninstallation

This chapter guides you install or uninstall the software.

Chapter 2: Operating the Ultrascope

This chapter introduces how to use Ultrascope software in detail.

Chapter 3: Prompting messages and Troubleshooting

This chapter gives solutions to frequently encountered problems or failures when testing. And it lists all the system prompting messages and their meanings.

Chapter 4: Support & Service

This chapter contains service information for Ultrascope.

Copyright 2006 Rigol Technologies, Inc.

5

UltraScope For DS1000

List

PRODUCT DESCRIPTION .....................................................................................................3

SYSTEM REQUIREMENTS ...................................................................................................4

CONTENTS..............................................................................................................................5

CHAPTER 1: INSTALLATION AND UNINSTALLATION...................................................7

CHAPTER 2: OPERATING THE ULTRASCOPE ..................................................................8

Description of the user interface............................................................................................9

Description of the software menu........................................................................................10

Description of the Toolbar...................................................................................................12

How to use DSO Controller.................................................................................................14

Shortcut Menu.....................................................................................................................15

Instant Running Control......................................................................................................16

Use DSO Controller to set up CH1& CH2 .........................................................................17

Use DSO Controller to set up Time Base............................................................................20

Use DSO Controller to set up Trigger System ....................................................................22

Use DSO Controller to set up Acquisition System .............................................................34

Use DSO Controller to set up Display System ...................................................................35

How to use the Virtual Keyboard........................................................................................36

How to use Data Browser....................................................................................................37

How to display a waveform ................................................................................................39

How to Measure with Data Browser...................................................................................42

How to get waveform data with Data Browser ...................................................................45

General Options ..................................................................................................................52

CHAPTER 3: PROMPTING MESSAGES AND TROUBLESHOOTING ...........................55

Prompting Messages ...........................................................................................................55

Troubleshooting ..................................................................................................................55

CONTACT RIGOL.................................................................................................................56

Copyright 2006 Rigol Technologies, Inc.

6

UltraScope For DS1000

Chapter 1: Installation and Uninstallation

This section describes how to install Ultrascope software on your computer.

Installing Ultrascope Software

Follow the steps below to install Ultrascope software.

1. Insert the CD-ROM into the CD-ROM drive. Run the Ultrascope for DS1000

Series.EXE file to start the installation wizard.

2. The Welcome to Ultrascope dialog appears. Read the information in the dialog, and click

Next.

3. Select to accept the terms in the License Agreement dialog, and click Next.

4. The Customer Information dialog appears. Read or edit the information in the dialog, and

Click Next.

5. Click Install to start installation. After all the files have been installed, the Finish button appears.

6. Click Finish to complete the installation process.

Uninstallation

To uninstall Ultrascope for DS1000 Series Oscilloscopes, you can click by the follow steps:

Start-> Program -> Ultrascope for DS1000 Series->Uninstall. Or use the Add/Remove program in the Microsoft Windows Control Panel .

Both uninstall methods activate the Uninstall wizard, which takes you through the uninstall process for Ultrascope.

Copyright 2006 Rigol Technologies, Inc.

7

UltraScope For DS1000

Chapter 2: Operating the Ultrascope

This chapter covers the following topics:

„ Description of the user interface

„ Description of the software menu

„ Description of the Toolbar

„ How to use DSO Controller

„ How to use Data Browser

„ How to use Virtual Keyboard

„ How to set up General Options

Copyright 2006 Rigol Technologies, Inc.

8

UltraScope For DS1000

Description of the user interface

Figure 1: Software interface

Copyright 2006 Rigol Technologies, Inc.

9

UltraScope For DS1000

Description of the software menu

1. File

Waveform, Measure or Wave Data.

Open:

Opens a dialog box that lets you open a Data Sheet or a

DSO Controller file.

Save Data Sheet:

Saves the Data Sheet to its current directory and file name.

Save Controller:

Saves the DSO Controller to its current directory and file name.

Exit:

Exits Ultrascope

2. View

Toolbar:

Turns the toolbar on and off

Show Controller:

Turns the DSO Controller on and off

Show Data Browser:

Turns the Data Browser on and off

3. Tools

Connect to Oscilloscope:

Connects PC to the oscilloscope

Options:

I/O settings and general settings

Register Ultrascope

4. Window

Cascade:

Stacks open windows in a diagonal overlap, starting from the upper left corner of the data browser window.

Tile Horizontal:

Resizes open windows in the data browser window so that all windows are displayed horizontally.

Tile Vertical:

Resizes open windows in the data browser window so that all windows are displayed vertically.

Arrange Icons:

To arrange icons

Copyright 2006 Rigol Technologies, Inc.

10

UltraScope For DS1000

5. Help

About:

Copyright Information

Copyright 2006 Rigol Technologies, Inc.

11

UltraScope For DS1000

Description of the Toolbar

Figure 2: Tool Bar

1( New ): To create a new document of Data Sheet, Controller, Waveform, Measure or Wave

Data.

2( Open ): To open a document of Data Sheet or Controller

3( Save All ): To save all files

4( Connect to Oscilloscope ): Connect PC to the oscilloscope

5( Disconnect ): Disconnect the link between PC and the oscilloscope

6( DSO Controller ): Show the DSO Controller window

7( Data Browser ): Open or close the Data Browser window

8( Add New Waveform ): Add new waveform window

9( Add New Measure ): Add new measure window

10( Add New Data ): Add new data window

11( Add New Record ): Add new record window

12( Add New Memory Waveform ): Add new memory waveform window

13( Options ): I/O settings and general settings

14( Connection Status Lamp ): The lamp will turn blue in connecting status, and it will turn red after disconnect the link between PC and Oscilloscope.

Copyright 2006 Rigol Technologies, Inc.

12

UltraScope For DS1000

Note: After connecting the PC to the oscilloscope, the keys and knobs on the front panel will be locked. To activate the front panel control, you must disconnect the link in Ultrascope software or pressing the FORCE key on the front panel.

Copyright 2006 Rigol Technologies, Inc.

13

UltraScope For DS1000

How to use DSO Controller

Figure 3: The Explorer View and Menu of DSO Controller

The DSO Controller consists of an Explorer View and a Shortcut Menu. The following illustration descripts each component of the DSO Controller.

Explorer View

The Explorer view lists the instrument and its settings, controls, and virtual keyboard in a

Windows Explorer-like tree diagram.

To display the contents of an Explorer view subtree, click on the plus icon next to an item you want to explore. To close a subtree, click on the minus icon next to the subtree. The lowest-level subtrees generally represent the scope control that you can setup or execute.

Download & Upload

You could click the Upload button to transfer parameters to your oscilloscope, and you can get these parameters by pressing the Download button.

Copyright 2006 Rigol Technologies, Inc.

14

UltraScope For DS1000

Shortcut Menu

The Shortcut menu contains commands and settings available in DSO Controller. You could access the menu by right-click on an icon.

New Control File:

Creates a new document for DSO Controller.

Open Control File:

Opens a dialog box that lets you open a DSO Controller file.

Save Control File:

Saves the DSO Controller to its current directory and file name.

Save Control File As: Opens a dialog box that lets you save a DSO Controller file.

Connect To DSO:

Connect PC to the oscilloscope.

Disconnect:

Disconnect the link between PC and the oscilloscope.

Upload:

Upload the settings and parameters from DSO Controller to the oscilloscope.

Download:

Download the settings and parameters from the oscilloscope to DSO

Controller.

Expand All:

Expand the Explorer View to display the lowest-level subtrees

Copyright 2006 Rigol Technologies, Inc.

15

UltraScope For DS1000

Instant Running Control

On the DSO Controller panel, the “Running Control” subtree includes four instant action keys. (See Figure 4)

Figure 4

Double-click “Run|Stop” to pop up a dialogue box, you could select “RUN” or “STOP” to let the instrument run or stop.

Double-click the other three action keys to execute the “Autoset”, “Set Trig to 50%” or

“Force Trigger” actions instantly.

Copyright 2006 Rigol Technologies, Inc.

16

UltraScope For DS1000

Use DSO Controller to set up CH1& CH2

You can set up the parameters of channel 1 or channel 2 at “CH1”, “CH2” subtrees on DSO

Controller panel.

Figure 5

Double-click the lowest subtrees to pop up a list, you could select a value and set the instrument to use that value.

Copyright 2006 Rigol Technologies, Inc.

17

UltraScope For DS1000

Table 1:

Subtree Settings

Display

ON

OFF

Volt/Div

Offset

Coupling

BW

Limit

Probe

2.00mV~

5.00V

-40.00V~+40.00V

(Volt/Div> 100mV)

-2.00V ~ +2.00V

(Volt/Div<200mV)

AC

DC

ON

OFF

1X

10X

100X

1000X

Comments

Select “ON” to display the channel

Select “OFF” to close the channel

Select the Volt/Div value within the range of 2.00mV~ 5.00V

Select the vertical offset value

Set to “AC” coupling

Set to “DC” coupling

Limits the bandwidth to reduce display noise.

To set “OFF” will get full bandwidth.

Set this to match your probe attenuation factor to make the vertical scale readout correct

Copyright 2006 Rigol Technologies, Inc.

18

UltraScope For DS1000

Use DSO Controller to set up LA(Mixed-Signal Oscilloscope)

You can set up the parameters of digital channels at “LA” subtrees on DSO Controller panel..

Figure 6

Double-click the lowest subtrees or the second-subtrees to pop up a list, you could select a value and set the instrument to use that value.

Table 2:

Subtree

Second-subtree Settings

Comments

Dipslay D15-D0

Position D15-D0

ON

OFF

0-15

Turn on or turn off single channel of

D15-D0.

Select the vertical position of the channel D15-D0.

Threshold

TTL

CMOS

ECL

User

Select style of whole digital channel.

The threshold voltage can set by user in user-defined style.

Reset

Reset waveform of the channel

D15-D0.

Copyright 2006 Rigol Technologies, Inc.

19

UltraScope For DS1000

Use DSO Controller to set up Time Base

You could set up the horizontal parameters of the oscilloscope at the “Time Base” subtree on

DSO Controller panel. When the trigger mode is not Alternative, you can set up the value as below figure and table .(See Figure 7)

Figure 7

Table 3:

Subtree Settings Comments

Time/Div 5ns~50s Select the timebase of CH1 from 5ns to 50s

Delay

Format

Y-T

X-Y

Roll

Display the delay value

Select to Y-T format

Select to X-Y format

Select to Roll format

Copyright 2006 Rigol Technologies, Inc.

20

UltraScope For DS1000

When the trigger mode is Alternative, you can set up the value as below figure and table.

Figure 8

Table 4:

Subtree Second-subtree Settings

Time/Div 5ns~50s

CH1

Delay

CH2

Time/Div 5ns~50s

Delay

Comments

Select the timebase of CH1 from 5ns to 50s

Display the delay value of CH1

Select the timebase of CH2 from 5ns to 50s

Display the delay value of CH2

Copyright 2006 Rigol Technologies, Inc.

21

UltraScope For DS1000

Use DSO Controller to set up Trigger System

You could set up the trigger system of the oscilloscope at the “Trigger” subtree on DSO

Controller panel.

Figure 9: Edge Trigger

Copyright 2006 Rigol Technologies, Inc.

22

UltraScope For DS1000

Table 5: Edge Trigger

Subtree

Method

Source

Settings

EDGE

PULSE

VIDEO

SLOPE

ALTERNATIVE

PATTERN

DURATION

CH1

CH2

EXT

EXT5

AC

D15-D0

Slope

Mode

Coupling

Level

RISING

FALLING

BOTH

AUTO

NORMAL

SINGLE

DC

AC

HF

LF

-12.00div to +12.00div

Comments

Select trigger mode as “EDGE”

Select trigger mode as “PULSE”

Select trigger mode as “VIDEO”

Select trigger mode as “SLOPE”

Select trigger mode as “ALTERNATIVE”

Select trigger mode as “PATTERN”

Select trigger mode as “DURATION”

Select the input source as the trigger signal

EXT and EXT5 use the signal applied to the EXT TRIG connector as the source.

Select any digital channel in D15-D0 as trigger source

(only in mixed-signal oscilloscope).

Select to trigger on the rising, falling or both edge of the signal

Select the type of triggering

Selects the components of the trigger signal applied to the trigger circuitry

Set the trigger level from -12.00div to +12.00div

Sensitivit y

0.10div to 1.00div

Set the trigger sensitivity from 0.10div to 1.00div

Holdoff 100ns ~ 1.5s Select the holdoff time from 100ns to 1.5s

Copyright 2006 Rigol Technologies, Inc.

23

UltraScope For DS1000

Figure 10: Video Trigger

Copyright 2006 Rigol Technologies, Inc.

24

UltraScope For DS1000

Table 6: Video Trigger

Subtree

Source

Polarity

Settings

CH1

CH2

EXT

EXT5

POSITIVE

NEGATIVE

Sync

All line

Line num

Odd Field

Even Field

Line number 1 ~ 525/625

Video

Standard

Level

Comments

Select the input source as the trigger signal

EXT and EXT5 use the signal applied to the EXT TRIG connector as the source

Normal triggers on the negative edge of the sync pulse and inverted triggers on the positive edge of the sync pulse

Select to trigger on all lines

Select to trigger on the specified line

Select to trigger on odd field

Select to trigger on even field

Select the trigger line number

NTSC

PAL/SECAM

Select the video standard for sync and the number count

-12.00div to +12.00div

Set the trigger level from -12.00div to +12.00div

Sensitivity

Holdoff

0.10div to 1.00div

Set the trigger sensitivity from 0.10div to 1.00div

100ns ~ 1.5s Select the holdoff time from 100ns to 1.5s

Copyright 2006 Rigol Technologies, Inc.

25

UltraScope For DS1000

Figure 11: Pulse Trigger

Copyright 2006 Rigol Technologies, Inc.

26

UltraScope For DS1000

Table 7: Pulse Trigger

Subtree

Source

Settings

CH1

CH2

EXT

EXT5

D15-D0

Pulse mode

Pulse width

Mode

Coupling

Level

+Less Than

+Greater Than

+Equal

-Less Than

-Greater Than

-Equal

20ns ~ 10s

AUTO

NORMAL

SINGLE

DC

AC

HF

LF

-12.00div to +12.00div

Comments

Select the input source as the trigger signal

EXT and EXT5 use the signal applied to the EXT

TRIG connector as the source.

Select any digital channel in D15-D0 as trigger source (only in mixed-signal oscilloscope).

Select how to compare the trigger pulse.

Select the pulse width from 20ns to 10s

Select the type of triggering

Selects the components of the trigger signal applied to the trigger circuitry

Set the trigger level from -12.00div to +12.00div

Sensitivity

0.10div to 1.00div

Set the trigger sensitivity from 0.10div to 1.00div

Holdoff 100ns ~ 1.5s Select the holdoff time from 100ns to 1.5s

Note: The change of trigger methods will make the trigger subtrees different.

Copyright 2006 Rigol Technologies, Inc.

27

UltraScope For DS1000

Figure 12: Slope Trigger

Copyright 2006 Rigol Technologies, Inc.

28

UltraScope For DS1000

Table 8: Slope Trigger

Subtree

Source

Slope mode

Slope time

Mode

Coupling

LevelA

LevelB

Settings

CH1

CH2

EXT

EXT5

+Less Than

+Greater Than

+Equal

-Less Than

-Greater Than

-Equal

20ns ~ 10s

AUTO

NORMAL

SINGLE

DC

AC

HF

LF

-12.00div to +12.00div

Comments

Select the input source as the trigger signal

EXT and EXT5 use the signal applied to the EXT

TRIG connector as the source

Set the slope condition

Select the slope time from 20ns to 10s

Select the type of triggering

Selects the components of the trigger signal applied to the trigger circuitry

Set the trigger levelA or levelB from -12.00div to

+12.00div

Sensitivity

Holdoff

0.10div to 1.00div

100ns ~ 1.5s

Set the trigger sensitivity from 0.10div to 1.00div

Select the holdoff time from 100ns to 1.5s

Copyright 2006 Rigol Technologies, Inc.

29

UltraScope For DS1000

Figure 13: Alternative Trigger

Copyright 2006 Rigol Technologies, Inc.

30

UltraScope For DS1000

Table 9: Alternative Trigger

Subtree

Second-subtree

CH1

CH2

Trig type

Different

Second-subtrees

Trig type

Different

Second-subtrees

Settings

EDGE

PULSE

VIDEO

SLOPE

EDGE

PULSE

VIDEO

SLOPE

Comments

Select trigger mode as “EDGE”

Select trigger mode as “PULSE”

Select trigger mode as “VIDEO”

Select trigger mode as “SLOPE”

The change of trigger type will make the trigger second-subtrees different

Select trigger mode as “EDGE”

Select trigger mode as “PULSE”

Select trigger mode as “VIDEO”

Select trigger mode as “SLOPE”

The change of trigger type will make the trigger second-subtrees different

Note: The change of trig type will make the alternative trigger second-subtrees different.

Copyright 2006 Rigol Technologies, Inc.

31

UltraScope For DS1000

Figure 14: Pattern Trigger(Mixed-Signal Oscilloscope)

Table 10: Pattern Trigger

Comments Subtree Second-subtree Settings

Code D15-D0

Mode

H

L

X

Rising

Falling

AUTO

NORMAL

SINGLE

Holdoff

100ns

~ 1.5s

Set the pattern on the digital channels

Select the type of triggering

Select the holdoff time from

100ns to 1.5s

Copyright 2006 Rigol Technologies, Inc.

32

UltraScope For DS1000

Figure 15: Duration Trigger(Mixed-Signal Oscilloscope)

Table 11: Duration Trigger

Subtree Second-subtree Settings Comments

Code D15-D0

Qualifier

H

L

X

Less Than

Greater Than

Equal

Set the pattern on the digital channels

Set the duration trigger condition

Duration time

Mode

AUTO

NORMAL

SINGLE

Select the type of triggering

Holdoff 100ns ~ 1.5s

Select the holdoff time from 100ns to 1.5s

Copyright 2006 Rigol Technologies, Inc.

33

UltraScope For DS1000

Use DSO Controller to set up Acquisition System

You could set up the acquisition system of the oscilloscope at the “Acquire” subtree on DSO

Controller panel. (See Figure 16)

Figure 16:

Table 12:

Subtree

Acquisition

Sampling

Mode

Settings

NORMAL

AVERAGE

PEAK

DETECT

REAL-TIME

EQU-TIME

Comments

Set to Normal Acquisition mode

Set to Average Acquisition mode

Set to Peak Detect Acquisition mode

Set to Realtime sampling mode

Set to Equivalent sampling mode

Averages

2 to 256, step

2

Select Number of Averages from 2 to 256 with the step of 2.

The number is only used to the average acquisition mode.

Copyright 2006 Rigol Technologies, Inc.

34

UltraScope For DS1000

Use DSO Controller to set up Display System

You could set up the display system of the oscilloscope at the “Display” subtree on DSO

Controller panel. (See Figure 17)

Figure 17

Table 13:

Subtree

Type

Persist

Clear

Grid

Settings

VECTORS

DOT

ON

OFF

FULL

HALF

NONE

Comments

Vectors fills the space between adjacent sample points in the display

Dots displays only the sample points

Set the sample point remains displayed until turn the persistence “OFF”.

Turn off the persistence

Clear all exiting waveform from screen

Select the graticule display mode

Copyright 2006 Rigol Technologies, Inc.

35

UltraScope For DS1000

How to use the Virtual Keyboard

You could double-click the subtree of “Virtual Keyboard” to pop up a panel shown as Figure

18.

Figure 18: Virtual Keyboard

The virtual keyboard has softkeys in conjunction with the keys and knobs on the front panel of the oscilloscope. The arrangement of the softkeys on the virtual keyboard is the same as those buttons and knobs on the front panel. You can click these softkeys to control the oscilloscope by PC.

Copyright 2006 Rigol Technologies, Inc.

36

UltraScope For DS1000

How to use Data Browser

Figure 19: The Data Browser and Shortcut Menu

The Data Browser consists of a Branch View and a Shortcut Menu. The following illustration descripts each component of the Data Browser.

Branch View

The branch view lists the data sheet of waveforms, measures, and data. To display the contents in the data sheet, double click the lowest-level item on the list, and this selected item will be displayed on the right of the main window.

Shortcut Menu

The Shortcut menu contains commands and settings available in Data Browser. You could access the menu by right-click on an icon.

New Data Sheet:

Creates a new document for Data Browser.

Open Data Sheet:

Opens a dialog box that lets you open a Data Sheet file.

Save Data Sheet:

Saves the Data Sheet to its current directory and file name.

Copyright 2006 Rigol Technologies, Inc.

37

UltraScope For DS1000

Save Data Sheet As: Opens a dialog box that lets you save a Data Sheet file.

Add New:

Adds a new waveform, measure, data , record or memory waveform window

Copyright 2006 Rigol Technologies, Inc.

38

UltraScope For DS1000

How to display a waveform

Add a waveform window:

You could right-click the Data Browser panel to pop-up a shortcut menu, select Add New->

Waveform to display a new window.

Remove the waveform window:

Select Remove item on the shortcut menu to close the waveform window.

Figure 20: The Waveform Panel

The waveform panel provides waveform display, zooming, storage, and printing for analysis.

The following description illuminates how to operate these functions.

Copyright 2006 Rigol Technologies, Inc.

39

UltraScope For DS1000

Refresh

After you click the Refresh button on this panel or press the F5 key on your PC keyboard, the signal currently on the oscilloscope’s screen will be displayed on the waveform window.

You could select CH1, CH2, Math, LA(Mixed-Signal Oscilloscope) or all of them to display the signal connected to channel 1, channel 2, Math, LA(Mixed-Signal Oscilloscope) or all channels. You must open the channel that you want to display in the DSO.

Controller. Otherwise, the selection item in conjunction with this channel will be disabled as gray color.

Display

The panel toolbar contains controls for zooming the waveform, selecting into the Best Fit mode, setting the display in Color or B&W mode, setting the Grid ON or OFF.

The percent zoom field displays the current data sheet magnification setting, where 100% is the system-default zoom value. Clicking on the drop-down menu arrow displays a list of magnification values from which you can select. You can also enter numeric values into this field by moving the cursor into the field, and typing in a new value after the Best Fit mode disabled.

The Best Fit mode fits the graticule to the current window size, allowing you to view the entire graticule.

Output

You could output the waveform by exporting currently displayed window into “BMP” format file to its directory and file name.

Previewing and Printing

To preview and set up printing page of waveform, click the Preview button. To print the page directly, just click the Print button.

First Page

Go to Page Last Page

Zoom Level Close Preview Select Page

Previous Page

Next Page

Print

Select Default Printer

Figure21: Toolbar on the preview page

Print All Pages

Copyright 2006 Rigol Technologies, Inc.

40

UltraScope For DS1000

Sizing, Tiling and Cascading Windows

To scroll data or size the data sheet, use normal windows procedures.

To auto arrange or otherwise manipulate windows or icons, select the appropriate menu items in the Ultrascope Window menu.

Copyright 2006 Rigol Technologies, Inc.

41

UltraScope For DS1000

How to Measure with Data Browser

Add a measure window:

You could right-click the Data Browser panel to pop-up a shortcut menu, select Add New->

Measure to display a new window.

Remove the measure window:

Right-click the measurement branch of the Data Browser to display the shortcut menu and then select Remove to close the measure window.

Figure 22: The Measurement Panel

This panel provides 18 parameters for measuring, including Vpp, Vmax, Vmin, Vavg, Vamp,

Vtop, Vbase, Vrms, Vover, Vpre, Frequency, Rise Time, Fall Time, Period, +Pulse Width,

-Pulse Width,+Duty or -Duty.

Copyright 2006 Rigol Technologies, Inc.

42

UltraScope For DS1000

The description of the functional buttons on the measurement panel is shown as follows:

Figure 23: The Control bar on the measurement panel

Refresh

You could get the auto measurement results from the oscilloscope by clicking the Refresh button on this panel or pressing the F5 key on your PC keyboard. The panel will display a list of measurement values for currently selected channel.

CH1, CH2

You could select CH1 or CH2 for measuring the signal connected to channel 1 or channel 2.

The oscilloscope must display the channel that you want to measure; you could change this setting in the DSO Controller. If the channel is close, the selection item associated with this channel will be disabled as gray color.

Export

You could save the measurement results as “TXT” or “EXCEL” file in your computer.

Info Bar

This bar displays channel, vertical scale, horizontal scale, sampling rate and date info in conjunction with the waveform for measuring.

Copyright 2006 Rigol Technologies, Inc.

43

UltraScope For DS1000

Pass & Fail

Figure 24: Pass & Fail Setting window

The Ultrascope software enables users to set pass/fail test parameters. These parameters include Vpp, Vmax, Vmin, Vavg, Vamp, Vtop, Vbase, Vrms, Vover, Vpre, Frequency, Rise

Time, Fall Time, Period, +Pulse Width, -Pulse Width,+Duty or -Duty.

You could set pass/fail factors range. If the measurement value is out of range, the measurement panel will display test result as “fail”. If the measurement value is under limit, it will show “pass” on the Pass&Fail list in the measurement panel.

To set the range of these pass/fail factors according to the following steps:

1. Select parameters for pass/fail factors from the list, and click to fill “√” in the square block to the right of the list.

2. After enable the parameter, input number and select unit for this parameter to set pass

/fail range.

Copyright 2006 Rigol Technologies, Inc.

44

UltraScope For DS1000

How to get waveform data with Data Browser

Add a data window:

You could right-click the Data Browser panel to pop-up a shortcut menu, select Add New->

Data to display a new window.

Remove the data window:

Right-click the data branch of the Data Browser to display the shortcut menu and then select

Remove to close the data window.

Figure 25: Data window

The data window displays the time and voltage data of the current waveform window on the oscilloscope. These digits are horizontal time values and vertical values of the sampling points.

Copyright 2006 Rigol Technologies, Inc.

45

UltraScope For DS1000

Data Acquisition Control Bar

Figure 26: Data Acquisition Control Bar

Refresh

Click the Refresh button on the data acquisition control bar or press the F5 key on your PC keyboard to get the waveform data from the oscilloscope.

CH1, CH2

You could select CH1 or CH2 for downloading the waveform data from channel 1 or channel

2. Notice that you must set the channel for display in the DSO Controller before selecting channels on this data acquisition bar.

Export

The waveform data can be saved as “TXT” or “EXCEL” file in your hard disk for further analysis.

Info Bar

The channels, vertical scale, horizontal scale, sampling rate, record size and date info could be displayed on this info bar.

Copyright 2006 Rigol Technologies, Inc.

46

UltraScope For DS1000

How to display a record file

Add a record window:

You could right-click the Data Browser panel to pop-up a shortcut menu, select Add New->

Record to display a new window.

Remove the record window:

Select Remove item on the shortcut menu to close the record window.

Figure 27: The Record Panel

The record panel provides waveform display, storage, and printing for analysis. The following description illuminates how to operate these functions.

Copyright 2006 Rigol Technologies, Inc.

47

UltraScope For DS1000

Figure 28: The Record Tool Bar

Figure 29: The Record Control Bar

Open

Press to open the record file you have saved. Then the waveform will be play back.

Color or B&W mode

Setting the display in Color or B&W mode.

Save

You could output the waveform by exporting currently displayed window into “BMP” format file to its directory and file name.

Previewing and Printing

To preview and set up printing page of waveform, click the Preview button. To print the page directly, just click the Print button.

The open record file path

This message disaply the record file path which you have open.

Copyright 2006 Rigol Technologies, Inc.

48

UltraScope For DS1000

Replay

The record waveform will be played back in continues.

Pause

Pause the waveform to play back.

Stop

Stop the waveform to play back.

Go to start frame

Press to go to the start frame.

Go to end frame

Press to go to the end frame.

The last frame

Press to go to the last frame.

The next frame

Press to go to the next frame.

Circulation and single replaying

Choosing to circulation or single replaying the record waveform.

Default time

Press to set the frame interval time is 50.00ms.

The frame interval time

This message display the frame interval time.

Current and total frame

This message display the current and total frame.

Sub time

Press to set the frame interval time sub.

Add time

Press to set the frame interval time add.

Copyright 2006 Rigol Technologies, Inc.

49

UltraScope For DS1000

How to display a memory waveform

Add a memory waveform window:

You could right-click the Data Browser panel to pop-up a shortcut menu, select Add New->

Memory Waveform to display a new window.

Remove the memory waveform window:

Select Remove item on the shortcut menu to close the memory waveform window.

Figure 30: The Memory Waveform Panel

The memory waveform panel provides the saved waveform display, zooming, storage, and printing for analysis. The following description illuminates how to operate these functions.

Copyright 2006 Rigol Technologies, Inc.

50

UltraScope For DS1000

Open

You can press this button to select the waveform file which you have saved.

Display

The panel toolbar contains controls for zooming the waveform, setting the display in Color or B&W mode, setting the Grid ON or OFF.

The percent zoom field displays the current data sheet magnification setting, where 100% is the system-default zoom value. Clicking on the drop-down menu arrow displays a list of magnification values from which you can select. You can also enter numeric values into this field by moving the cursor into the field.

Output

You could output the waveform by exporting currently displayed window into “BMP” format file or the Memory data into CSV format file to its directory and file name.

Previewing and Printing

To preview and set up printing page of waveform, click the Preview button. To print the page directly, just click the Print button.

Volt/Div

You can change the volt/div value to zoom the waveform in vertical direction.

Time/Div

You can change the time/div value to zoom the waveform in horizontal direction.

Copyright 2006 Rigol Technologies, Inc.

51

UltraScope For DS1000

General Options

Figure 31: Page 1 in General Options

The general options include 3 pages of settings for I/O selection, startup option, refresh and communication.

I/O Selection

You could select USB or RS-232.

Startup Option

In this section, you could set three operations while the Ultrascope software is starting up.

1. Startup Hardware Check: It will check the connection between the oscilloscope and the computer. If the communication is available, it will connect the oscilloscope with PC automatically while the software starting up.

Copyright 2006 Rigol Technologies, Inc.

52

UltraScope For DS1000

2. Recent Data Sheet File Load: It will open the data sheet file that you use last time after the software starts up.

3. Recent Control File Load: It will open the control file that you use last time after the software starts up.

Figure 32: Page 2 in General Options

Refresh

You could select to refresh the waveform or measurement data in single or auto mode.

Time Setting

You could select the refreshing time interval from a list of 1s, 2s, 5s, 10s and 20s. The default value is 1s.

Copyright 2006 Rigol Technologies, Inc.

53

UltraScope For DS1000

Figure 33: Page 3 in General Options

The communication settings include setups for RS-232 and USB.

RS-232 Setting:

1. RS232 Port: You could select the RS-232 port at COM1 or COM2 according to which port you use on your computer.

2. Baud Rate: You could select the baud rate from a list of 300, 2400, 4800, 9600, 19200 and 38400. The default value is 9600.

Note: The baud rate setting in Ultrascope must be same as the setting at the oscilloscope.

Otherwise, the Ultrascope will pop up an error message as “ Device not found”.

Copyright 2006 Rigol Technologies, Inc.

54

UltraScope For DS1000

Chapter 3: Prompting messages and Troubleshooting

Prompting Messages

Device not found: Prompting that the software does not detect the connection between the oscilloscope and the computer.

File not found: Prompting that Ultrascope could not find the control file or data sheet file when starting up.

This Item Query Only: Prompting that this item is only for display the setting or status of the oscilloscope. You could not double-click this item to setup the oscilloscope.

Save the changes to the Control File: Prompting users to save the changes of the control file before quit Ultrascope software.

Save the changes to the Data Sheet File: Prompting users to save the changes of the data sheet file before quit Ultrascope software.

Troubleshooting

1. After running the Ultrascope, the software could not detect the connection between the instrument and the computer. Please inspect the settings of the software and the oscilloscope according to the following steps:

(1) Make sure that the oscilloscope is powered on.

(2) Check the cable connection between the oscilloscope and PC.

(3) Check if the settings for the baud rate of the oscilloscope are same as those of the

Ultrascope software. (The default baud rate is 9600.)

(4) After the inspections above, restart the Ultrascope software.

(5) If the problem still remains, please contact RIGOL for help.

Copyright 2006 Rigol Technologies,Inc.

55

UltraScope For DS1000

Contact RIGOL

Product Support: For questions about using RIGOL measurement products, call in China:

Tel: 86-10-82899325

9:00 a. m.–5: 00 p. m. Beijing time

Or contact us by e- mail:

[email protected]

For product support outside of China, contact your local RIGOL distributor or sales office.

For a listing of worldwide service centers, visit our web site.

Web site:www.rigol.com

Copyright 2006 Rigol Technologies,Inc.

56

UltraScope For DS1000

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Related manuals

advertisement