User Guide

Add to my manuals
52 Pages

advertisement

User Guide | Manualzz

KODAK 8000C Digital Panoramic and Cephalometric

Extraoral Imaging System

User Guide

Notice

Congratulations on your purchase of the KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging

System. The KODAK 8000C unit is the KODAK 8000 equipped with the cephalostat unit. Thank you for your confidence in our products and we will do all in our power to ensure your complete satisfaction.

The User Guide for the KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System includes information on the cephalometric features. For the panoramic features, see the KODAK 8000 Extraoral Imaging

System (SM722) User Guide. We recommend that you thoroughly familiarize yourself with this Guide in order to make the most effective use of your system.

WARNING: We recommend that you consult the “Safety, Regulatory and the Technical

Specification User Guide” before using the KODAK 8000C Extraoral Imaging Systems.

The information contained in this Guide may be subject to modification without notice, justification or notification to the persons concerned.

No part of this Guide may be reproduced without the express permission of Carestream Health, Inc.

The US Federal law restricts this device to sale by or on the order of a physician.

This document is originally written in English.

Manual Name: KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System User Guide

Part Number: SM735

Revision Number: 02

Print Date: 03/2010

.

The brand names and logos reproduced in this Guide are copyright.

KODAK is a trademark of KODAK used under License.

KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System, complies with Directive

93/42/CEE relating to medical equipment.

0086

Manufacturer

Care stream Hea lth, Inc.

150 Verona Street

Roche ster NY 14 608

Authorized Representative in the European Community

EC REP

TROPHY

4, Rue F. Pelloutier, Croissy-Beaubourg

77435 Marne la Vallée Cedex 2, France

Contents

1—About This Guide

Conventions in this Guide. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–1

2—KODAK 8000C UNIT OVERVIEW

General Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–1

Mobile Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–2

General Functional Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–3

Digital Sensor Locations. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–4

Laser Locations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–5

Control Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–6

X-Ray Remote Control Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–7

Positioning Accessories and Replacement Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–8

3—IMAGING SOFTWARE OVERVIEW

Computer System Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–1

General Software Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–2

KODAK Dental Imaging Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–2

Cephalometric Acquisition Interface Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–2

Cephalometric Acquisition Interface Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–3

Cephalometric Acquisition Window Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–3

Cephalometric Program Pane . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–5

Cephalometric Patient Pane . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–6

Cephalometric Parameter Pane. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–7

4—GETTING STARTED

Switching on the Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4–1

Starting the Imaging Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4–2

Creating a Patient Record. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4–2

Accessing the Cephalometric Acquisition Window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4–3

5—ACQUIRING CEPHALOMETRIC IMAGES

Acquiring a Lateral Image . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–1

Preparing the Unit and Setting the Acquisition Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–1

Acquiring a Frontal AP or PA Image. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–5

Preparing the Unit and Setting the Acquisition Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–5

Preparing and Positioning the Patient . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–6

Acquiring an Oblique Image . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–9

Preparing the Unit and Setting the Acquisition Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–9

Preparing and Positioning the Patient . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–10

Acquiring a Submento-Vertex Image . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–12

Preparing the Unit and Setting the Acquisition Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–12

Preparing and Positioning the Patient . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–13

KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System User Guide (SM735)_Ed 02

iii

Contents

Acquiring a Carpus Image . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–16

Preparing the Unit and Setting the Acquisition Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–16

Preparing and Positioning the Patient . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–17

X-Ray Dose Emission Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–19

6—MAINTENANCE

Daily . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6–1

Monthly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6–1

Annually . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6–1

7—TROUBLESHOOTING

Quick Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–1

iv

Chapter 1

About This Guide

Conventions in this Guide

The following special messages emphasize information or indicate potential risk to personnel or equipment:

WARNING

Warns you to avoid injury to yourself or others by following the safety instructions precisely.

CAUTION

Alerts you to a condition that might cause serious damage.

IMPORTANT

Alerts you to a condition that might cause problems.

NOTE

Emphasizes important information.

TIP

Provides extra information and hints.

KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System User Guide (SM735)_Ed 02

1–1

Conventions in this Guide

1–2 About This Guide

Chapter 2

KODAK 8000C UNIT OVERVIEW

The KODAK 8000C digital panoramic and cephalometric unit is designed to carry out the following radiological examinations:

• Panoramic

• Maxillary Sinus

• Temporomandibular Joints (TMJ)

• Lateral cephalometric

• Frontal (PA or AP) cephalometric

• Oblique cephalometric

• Submento-vertex cephalometric

• Carpus cephalometric

General Overview

The KODAK 8000C digital panoramic and cephalometric unit is composed of the following functional components:

• The unit head that contains all the electronic control

• The rotative arm

• The fixed arm with a control panel

• The panoramic digital sensor

• The x-ray source assembly

• The x-ray remote control

• The chin rest and bite block

• The chin rest base

• The panoramic chin rest and bite block

• The temple supports

• The hand grips

• The cephalostat arm

• The cephalostat head

• The head clamps and ear cones

• The nasion support

• The acquisition software (see “Imaging Software Overview”)

The following figures illustrate the general overview of the KODAK 8000C digital panoramic and cephalometric units.

KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System User Guide (SM735)_Ed 02

2–1

General Overview

IMPORTANT

The Cephalostat can be positioned either on the right or the left side of the KODAK 8000 unit.

Mobile Components

Figure 2-1 illustrates the up and down movement of the KODAK 8000C digital panoramic and cephalometric units mobile component and the rotation of the rotative arm.

Figure 2–1 KODAK 8000 and KODAK 8000C Units Mobile Components

2–2 KODAK 8000C UNIT OVERVIEW

General Functional Components

Figure 2-2 illustrates the general functional components of the KODAK 8000C digital panoramic and cephalometric units.

Figure 2–2 KODAK 8000 and KODAK 8000C Units Functional Components

14

General Overview

15

12

16

16

8

17

13

7

6

4

5

2

10

3

1

11

9

3

4

1

2

ON/OFF button

Chin rest and bite block

Temple supports

Temple supports control knob

7

8

5

6

9

Hand Grips

Control panel

Height adjustment buttons

Panoramic digital sensor

X-Ray source assembly

10

11

12

13

14

15

16

17

Collimator selector

Unit rotative arm

X-Ray remote control

PC hosting the imaging and the acquisition software

Cephalostat arm

Cephalostat head

Head clamps and ear cones

Nasion support

KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System User Guide (SM735)_Ed 02

2–3

General Overview

Digital Sensor Locations

Figure 2-3 illustrates the locations of the digital panoramic and digital cephalometric sensors of the KODAK 8000C digital panoramic and cephalometric units.

Figure 2–3 KODAK 8000 and KODAK 8000C Units Digital Sensor Locations

2–4 KODAK 8000C UNIT OVERVIEW system

Laser Locations

Figure 2-4 illustrates the location of the lasers of the KODAK 8000C digital panoramic and cephalometric units.

Figure 2–4 KODAK 8000 and KODAK 8000C Units Laser Beam Locations

General Overview

4

3

2

1

1

Mid-sagittal plane positioning laser beam

2

Frankfort plane positioning laser beam

3

Canine plane positioning laser beam

4

Cephalometric Frankfort plane positioning laser beam

KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System User Guide (SM735)_Ed 02

2–5

Control Panel

Control Panel

The control panel is an alphanumeric, digital soft touch console. It allows the operator to control certain unit functions. It also displays the operating parameters and error messages.

Figure 2–5 Unit Control Panel

8

Digital Panoramic and Cephalometric System kV mA S

1 2 3 4

1

X-Ray emission LED: Yellow, indicates the x-rays are being emitted.

2

Display Screen: Displays the current acquisition parameters and the error messages.

3

Reset button: Resets the unit arm to the initial position to enable the patient to enter and exit the unit.

4

Laser beam button: Activates the laser positioning beams to correctly position the patient.

2–6 KODAK 8000C UNIT OVERVIEW

X-Ray Remote Control Overview

The x- ray remote control enables you to launch a radiological image acquisition via the exposure button from outside the x-ray room. You must press and hold the exposure button until the end of acquisition. Premature release of the exposure button interrupts the acquisition.

X-Ray Remote Control Overview

Figure 2–6 X-Ray Remote Control

1

1

Exposure button: launches image acquisition.

KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System User Guide (SM735)_Ed 02

2–7

Positioning Accessories and Replacement Parts

Positioning Accessories and Replacement Parts

The following accessories are used when positioning a patient. They are delivered with the KODAK 8000 digital panoramic and KODAK 8000C digital panoramic and cephalometric unit.

Table 2-1 and Table 2-2 list the panoramic and cephalometric positioning accessories.

Table 2–1 Panoramic Positioning Accessories and Replacement Parts

Accessory Description

Panoramic chin rest and TMJ x2

Maxillary sinus chin rest

TMJ x2 and TMJ x4 nose rest

Standard bite block

Bite block for edentulous patients x2

A set of right and left temple supports

Single use sheaths for bite blocks

(500 pcs box)

2–8 KODAK 8000C UNIT OVERVIEW

Positioning Accessories and Replacement Parts

Table 2–2 Cephalometric Positioning Accessories and Replacement Parts

Accessory Description

Head clamps with ear cones x2

Nasion support

KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System User Guide (SM735)_Ed 02

2–9

Positioning Accessories and Replacement Parts

2–10 KODAK 8000C UNIT OVERVIEW

Chapter 3

IMAGING SOFTWARE OVERVIEW

Computer System Requirements

This section specifies the minimum computer system requirements for KODAK 8000C digital panoramic and cephalometric extraoral imaging system software.

IMPORTANT

It is MANDATORY to check that the computer system configuration is compatible with the computer system requirements for the KODAK

8000C software. If necessary you MUST update your computer system configuration. KODAK 8000C MUST be connected to the computer via a point-to-point Ethernet link and not via a LAN. DO NOT place the computer and the peripheral equipment connected to it in the immediate vicinity of the patient in the unit. Leave at least 1.5 m distance from the unit. The computer and the peripheral equipment must conform to the IEC 60950 standard.

Table 3–1 Minimum Computer System Requirements

Item Viewing

Acquisition

Comments

CPU

RAM

2 GHz Intel Duo Core

2 GB

2 GHz Intel Duo Core

2GB RAM has a major impact on system performance.

Hard disk drive

1.2 GB for software installation

80 GB free space to use the software

1.2 GB for software installation

Graphic board

Monitor

Operating system (32 bits)

Nvidia / ATI based board supporting

Open GL 1.2 with 256 MB of video

RAM on AGP x8 video bus (for example: Nvidia GeForce 6800 GT)

Nvidia / ATI based board supporting Open Glide 1.2 with

256 MB of video RAM on AGP x8 video bus (for example: Nvidia

GeForce 6800 GT)

The video RAM has major impact on system performance.

1 monitor

17" or larger

1024 x 768 minimum screen resolution - 32 bits color mode

1 monitor

17"

1280 x 880 minimum screen resolution

Your monitor is a vital component in displaying quality images. Low-quality screens will prevent you from proper diagnoses and treatment.

Windows 2000 SP4

Windows XP Home / Pro edition SP2

Windows Vista

Windows XP Home / Pro edition SP2

Windows Vista

KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System User Guide (SM735)_Ed 02

3–1

General Software Overview

Table 3–1

CD/DVD drive

Backup Media

Minimum Computer System Requirements (Continued)

Item

Ethernet interface

Viewing

Acquisition

Comments

1 Ethernet interface 1 Ethernet interfaces (100Mbits) If there is a LAN connection a second Ethernet interface is required.

A DVD-BURNER drive is required.

A DVD-BURNER drive is required.

Removable/portable, external hard disk drive

Removable/portable, external hard disk drive.

We strongly recommend a daily backup of x-ray images and patient records.

General Software Overview

The KODAK 8000C digital panoramic and cephalometric extraoral imaging system operates with the following software:

• KODAK dental imaging software

• Cephalometric acquisition interface module

KODAK Dental Imaging Software

The KODAK dental imaging software is a user-friendly working interface that was designed and developed specifically for radiological diagnosis. It is the common imaging platform for all our digital systems for dentistry.

The KODAK dental imaging software has the following features:

• Patient record management using Patient Window features

• Extraoral and intraoral image management using Imaging Window features

NOTE

For a complete information on the KODAK Dental Imaging

Software, see the “KODAK Dental Imaging Software, Quick Start

Guide”.

Cephalometric Acquisition Interface Module

The cephalometric acquisition interface module is a user-friendly working interface that was designed and developed specifically for KODAK 8000C digital panoramic and cepahlometric extraoral imaging system.

3–2 IMAGING SOFTWARE OVERVIEW

Cephalometric Acquisition Interface Module

Cephalometric Acquisition Interface Module

Cephalometric Acquisition Window Overview

The Cephalometric Acquisition Window is the main cephalometric interface with the

KODAK 8000C cephalometric extraoral imaging system that provides you with imaging acquisition functions.

Figure 3–1

Cephalometric Acquisition Window

2

3

4

5

6

1

Information button:

About: Identifies the Software and the Firmware versions

Reset of the Values: Resets to the manufacturing parameter settings

Memorize settings: Memorizes the user preference settings for each patient type (kV, mA and seconds)

Preview Screen: Displays the acquired image in real time.

Selected Parameter Display: Displays the current acquisition parameter settings.

System Status Screen: Displays various alert or warning messages originating from the unit.

X-Rays ON / OFF button: Turns off the x-ray emissions to demonstrate the acquisition process for the patient.

Generator Cooling indicator: Indicates the automatic cooling time (mm:ss) required for the generator to reach 0 for a new acquisition.

KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System User Guide (SM735)_Ed 02

3–3

Cephalometric Acquisition Interface Module

8

9

7

Positioning laser button: Activates the laser positionning beams to correctly position the patient.

Reset button: Resets the rotative arm in the start position.

Ready Indicator LED

Green indicates the unit is ready to start acquisition.

Black indicates the unit is not ready to start acquisition.

10

Exit button: Closes the Acquisition Window.

11

X-Ray Emission indicator: Yellow, indicates the x-ray emission status.

12

Selector Button: Selects different acquisition setting options.

Click Program to select examination type options.

Click Patient to select patient type parameters.

Click Parameters to select exposure parameter options.

The Selector button enables you to access the following 3 panes:

Program pane: Examination type options

Patient pane: Patient type parameter options

Parameters pane: Exposure parameter options

3–4 IMAGING SOFTWARE OVERVIEW

Cephalometric Acquisition Interface Module

Cephalometric Program Pane

The cephalometric Program pane enables you to choose different radiological exams as well as different acquisition formats.

Figure 3–2 Cephalometric Program Pane

Radiological exam options:

Click for a lateral exam.

Click for a submento-vertex exam.

Click or exam (AP and PA).

for a frontal

Click for a carpus exam.

Click for an oblique exam.

NOTE

The above list of exam types are only a sample of exam options of the Program pane.

KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System User Guide (SM735)_Ed 02

3–5

Cephalometric Acquisition Interface Module

Cephalometric Patient Pane

The cephalometric Patient pane enables you to choose different patient parameters. The selection of the patient parameters influences the quality of the image. The selected parameters must be based on the patient age and morphology.

Figure 3–3 Cephalometric Patient Pane

1 Patient type parameters:

Click if the patient is a child.

2

Click if the patient is adult.

Patient size parameters:

Click if the patient is small.

Click if the patient is medium.

Click if the patient is large.

3–6 IMAGING SOFTWARE OVERVIEW

Cephalometric Acquisition Interface Module

Cephalometric Parameter Pane

The cephalometric Parameter pane enables you to choose exposure parameters for the radiological image acquisition. If the default parameter setting is not adapted to your patient type, you can manually adapt the parameter settings to the patient type and save this setting as the default setting.

Figure 3–4 Cephalometric Parameter Pane

1 Radiation dose options:

kV: Kilovolt

mA: Milliampere

2

S: Second

Fine-tuning buttons:

3

Click to fine-tune the kV, mA and the second.

Saving parameter button:

Click to save the selected parameters.

KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System User Guide (SM735)_Ed 02

3–7

Cephalometric Acquisition Interface Module

3–8 IMAGING SOFTWARE OVERVIEW

Chapter 4

GETTING STARTED

Switching on the Unit

Before switching on the unit, check that:

• The installation of the unit is complete.

• The PC is switched on.

IMPORTANT

You must switch on the PC and wait for it to be ready for the connection before switching on the unit.

To switch on the unit, follow these steps:

1. On the unit column, press the ON button.

2. You must wait for a minute for the connection between the unit and the PC to be established. If you start the imaging software before the connection is established an error message is displayed. Click OK, close the imaging software and wait for the connection to be established.

3. You can now proceed to start the imaging software.

IMPORTANT

To increase the operating life of the x-ray tube, if the unit has not been used for a month, you must follow the following procedures before use.

1. In the Panoramic Acquisition Window, select the Parameter pane.

2. Select the following series of parameter settings:

• 70 kV - 6.3 mA

• 80 kV - 10 mA

• 85 kV - 10 mA

3. Leave the x-ray room and close the door. For each parameter setting, from the x-ray remote control, press and hold the button to launch the x-ray

The unit is now ready to be used for acquisition.

KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System User Guide (SM735)_Ed 02

4–1

Starting the Imaging Software

Starting the Imaging Software

To start the imaging software, follow these steps:

1. On your desktop, double-click .

OR

From your PC, click Start > All Programs > Kodak > Kodak Dental Software.

4–2 GETTING STARTED

A blank Patient Window is displayed.

2. Create or open an existing patient record.

Creating a Patient Record

To create a patient record, follow these steps:

1. In the Patient Window, from the toolbar, click

OR

From the menu bar, select Patient > New.

2. Enter the required patient information. The Last name, the First name and the Date

of birth fields are required.

3. From the menu bar, select Picture > Insert Picture to add a *.tif or *.bmp picture of the patient to the record. Select the picture from your directory and click Open.

4. Click OK to save. The patient record is automatically assigned a 7-digit number starting with a letter (for example, M0000001).

5. Click to access the Imaging Window.

6. Select an image acquisition.

Accessing the Cephalometric Acquisition Window

Accessing the Cephalometric Acquisition Window

To access the Acquisition Windows, follow these steps:

1. In the Imaging Window, from the toolbar, click to access the Cephalometric

Acquisition Window.

2. Prepare the acquisition parameters and launch an acquisition.

KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System User Guide (SM735)_Ed 02

4–3

Accessing the Cephalometric Acquisition Window

4–4 GETTING STARTED

Chapter 5

ACQUIRING CEPHALOMETRIC IMAGES

Acquiring a Lateral Image

Before acquiring a lateral image, check that you have:

• Reset the unit rotative arm to the start position for patient to enter the unit.

• Selected the patient record.

• Accessed the Imaging Window.

• Accessed the Cephalometric Acquisition Window.

Preparing the Unit and Setting the Acquisition Parameters

To set the acquisition parameters, follow these steps:

1. Raise and block the sensor manually.

2. On the x-ray source assembly, set the collimator selector to LA.

3. Position the head clamps manually for the lateral exam.

IMPORTANT

You must position the head clamps manually because they are not positioned automatically from the Program pane exam type selection. In this case, the relevant exam type selection icon becomes active.

4. In the Cephalometric Acquisition Window, click the Program button to access the

Program pane. In the Program pane the for a lateral exam is active.

5. Click the Patient button to access the Patient pane. Select the patient:

• Type

• Size

6. If the default parameter setting is not adapted to your patient type, click the

Parameter button and in the Parameter pane select the appropriate parameters. To

fand select save the new parameter settings as the default settings, click

Memorize settings.

7.

Preparing and Positioning the Patient

To prepare and position the patient, follow these steps:

KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System User Guide (SM735)_Ed 02

5–1

Acquiring a Lateral Image

1. Ask the patient to remove all metal objects.

2. Ask the patient to wear a lead apron. Ensure that the apron lays flat across the patient’s shoulders.

3. Press and hold the height adjustment buttons to raise the cephalostat head.

4. Open the head clamps and ask the patient to stand up straight, in front of the cephalometric unit, in the appropriate position.

5. Press and hold the height adjustment buttons to level the ear cones to the patient’s auditory canals.

6. Insert gently one cone in the auditory canal of the patient. Turn gently the button to close the arms. Insert gently the second cone in the auditory canal of the patient.

7.

On the control panel, click to turn ON the Frankfort laser positioning beam.

Align the patient with the Frankfort laser beam.

5–2 ACQUIRING CEPHALOMETRIC IMAGES

8. Lower the nasion support to a vertical position.

Acquiring a Lateral Image

KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System User Guide (SM735)_Ed 02

5–3

Acquiring a Lateral Image

Launching the X-ray

To launch the x-ray, follow these steps:

1. Leave the x-ray room and close the door. You must keep visual contact with the patient during acquisition.

IMPORTANT

To stop the acquisition, if any problem, release the exposure button of the remote control or press the red emergency stop button.

2. Launch the x-ray with the remote control. Press and hold the exposure button until the end of acquisition. The turns yellow, indicating x-ray emission. The image appears on the Preview Screen of the Cephalometric Acquisition Window. When the acquisition ends, the Cephalometric Acquisition Window disappears and the acquired image is transferred automatically to the Imaging Window.

3. Check the image quality, if satisfactory, remove the ear cones and the nasion support and release the patient.

5–4 ACQUIRING CEPHALOMETRIC IMAGES

Acquiring a Frontal AP or PA Image

Before acquiring a frontal AP or PA image, check that you have:

• Reset the unit rotative arm to the start position for patient to enter the unit.

• Selected the patient record.

• Accessed the Imaging Window.

• Accessed the Cephalometric Acquisition Window.

Acquiring a Frontal AP or PA Image

Preparing the Unit and Setting the Acquisition Parameters

To acquire a frontal AP or PA image, follow these steps:

1. Raise and block the sensor manually.

2. On the x-ray source assembly, set the collimator selector to AP/PA.

3. Position the head clamps manually for the frontal AP or PA exam.

IMPORTANT

You must position the head clamps manually because they are not positioned automatically from the Program pane exam type selection. In this case, the relevant exam type selection icon becomes active.

4. In the Cephalometric Acquisition Window, click the Program button to access the

Program pane. In the Program pane, click for a frontal PA exam.

5. Click the Patient button to access the Patient pane. Select the patient:

• Type

• Size

6. If the default parameter setting is not adapted to your patient type, click the

Parameter button and in the Parameter pane select the appropriate parameters. To save the new parameter settings as the default settings, click

Memorize settings.

fand select

KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System User Guide (SM735)_Ed 02

5–5

Acquiring a Frontal AP or PA Image

Preparing and Positioning the Patient

To prepare and position the patient, follow these steps:

1. Ask the patient to remove all metal objects.

2. Ask the patient to wear a lead apron. Ensure that the apron lays flat across the patient’s shoulders.

3. Press and hold the height adjustment buttons to raise the cephalostat head.

4. Open and position the head clamps parallel to the cephalometric sensor. Ask the patient to stand up straight in front of the cephalometric unit in the following positions:

• For a frontal AP, facing the generator

• For a frontal PA, facing the cephalometric sensor

5. Press and hold the height adjustment buttons to level the ear cones to the patient’s auditory canals.

6. Insert gently one cone in the auditory canal of the patient. Turn gently the button to close the arms. Insert gently the second cone in the auditory canal of the patient.

7.

On the control panel, click to turn on the Frankfort laser positioning beam. Align the patient with the Frankfort laser beam for the frontal AP only.

Figure 5–1 Frontal AP Figure 5–2 Frontal PA

5–6 ACQUIRING CEPHALOMETRIC IMAGES

Figure 5–3 Frontal AP Figure 5–4 Frontal PA

Acquiring a Frontal AP or PA Image

Launching the X-ray

To launch the x-ray, follow these steps:

1. Leave the x-ray room and close the door. You must keep visual contact with the patient during acquisition.

IMPORTANT

To stop the acquisition, if any problem, release the exposure button of the remote control or press the red emergency stop button.

KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System User Guide (SM735)_Ed 02

5–7

Acquiring a Frontal AP or PA Image

Figure 5–5 Frontal AP

Figure 5–6 Frontal PA

2. Launch the x-ray with the remote control. Press and hold the exposure button until the end of acquisition. The turns yellow indicating x-ray emission. The image appears on the Preview Screen of the Cephalometric Acquisition Window. When the acquisition ends, the Cephalometric Acquisition Window disappears and the acquired image is automatically transferred to the Imaging Window.

3. Check the image quality. If satisfactory, remove the ear cones and release the patient.

5–8 ACQUIRING CEPHALOMETRIC IMAGES

Acquiring an Oblique Image

Before acquiring an oblique image, check that you have:

• Reset the unit rotative arm to the start position for patient to enter the unit.

• Selected the patient record.

• Accessed the Imaging Window.

• Accessed the Cephalometric Acquisition Window.

Preparing the Unit and Setting the Acquisition Parameters

To acquire an oblique image, follow these steps:

1. Raise and block the sensor manually.

2. On the x-ray source assembly, set the collimator selector to AP/PA.

3. Position the head clamps manually for the oblique exam with the desired angle.

IMPORTANT

You must position the head clamps manually because they are not positioned automatically from the Program pane exam type selection. In this case, the relevant exam type selection icon becomes active.

4. In the Cephalometric Acquisition Window, click the Program button to access the

Program pane. In the Program pane:

• The

for an oblique exam is active.

• Click

to select the desired angle.

5. Click the Patient button to access the Patient pane. Select the patient type.

6. Click the Patient button to access the Patient pane. Select the patient:

• Type

• Size

7.

If the default parameter setting is not adapted to your patient type, click the

Parameter button and in the Parameter pane select the appropriate parameters. To save the new parameter settings as the default settings, click

Memorize settings.

fand select

Acquiring an Oblique Image

KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System User Guide (SM735)_Ed 02

5–9

Acquiring an Oblique Image

Preparing and Positioning the Patient

To prepare and position the patient, follow these steps:

1. Ask the patient to remove all metal objects.

2. Ask the patient to wear a lead apron. Ensure that the apron lays flat across the patient’s shoulders.

3. Press and hold the height adjustment buttons to raise the cephalostat head.

4. Open the head clamps and ask the patient to stand up straight in front of the cephalometric unit, in the appropriate position.

5. Press and hold the height adjustment buttons to level the ear cones to the patient’s auditory canals.

6. Insert gently one cone in the auditory canal of the patient. Turn gently the button to close the arms. Insert gently the second cone in the auditory canal of the patient.

7.

Lower the nasion support to a vertical position.

5–10 ACQUIRING CEPHALOMETRIC IMAGES

Launching the X-ray

To launch the x-ray, follow these steps:

1. Leave the x-ray room and close the door. You must keep visual contact with the patient during acquisition.

IMPORTANT

To stop the acquisition, if any problem, release the exposure button of the remote control or press the red emergency stop button.

Acquiring an Oblique Image

2. Launch the x-ray with the remote control. Press and hold the exposure button until the end of acquisition. The turns yellow indicating x-ray emission. The image appears on the Preview Screen of the Cephalometric Acquisition Window. When the acquisition ends, the Cephalometric Acquisition Window disappears and the acquired image is automatically transferred to the Imaging Window.

3. Check the image quality. If satisfactory, remove the ear cones and the nasion support.

Release the patient.

KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System User Guide (SM735)_Ed 02

5–11

Acquiring a Submento-Vertex Image

Acquiring a Submento-Vertex Image

Before acquiring a submento-vertex image, check that you have:

• Reset the unit rotative arm to the start position for patient to enter the unit.

• Selected the patient record.

• Accessed the Imaging Window.

• Accessed the Cephalometric Acquisition Window.

Preparing the Unit and Setting the Acquisition Parameters

To acquire a submento-vertex image, follow these steps:

1. Raise and block the sensor manually.

2. On the x-ray source assembly, set the collimator selector to AP/PA.

3. Position the head clamps manually for a frontal AP exam..

IMPORTANT

You must position the head clamps manually because they are not positioned automatically from the Program pane exam type selection. In this case, the relevant exam type selection icon becomes active.

4. In the Cephalometric Acquisition Window, click the Program button to access the

Program pane. In the Program pane, click for a submento-vertex exam.

5. Click the Patient button to access the Patient pane. Select the patient:

• Type

• Size

6. If the default parameter setting is not adapted to your patient type, click the

Parameter button and in the Parameter pane select the appropriate parameters. To save the new parameter settings as the default settings, click

Memorize settings.

fand select

5–12 ACQUIRING CEPHALOMETRIC IMAGES

Acquiring a Submento-Vertex Image

Preparing and Positioning the Patient

To prepare and position the patient, follow these steps:

1. Ask the patient to remove all metal objects.

2. Ask the patient to wear a lead apron. Ensure that the apron lays flat across the patient’s shoulders.

3. Press and hold the height adjustment buttons to raise the cephalostat head.

4. Open the head clamps and ask the patient to stand up straight in front of the cephalometric unit, in the appropriate position.

5. Press and hold the height adjustment buttons to level the ear cones to the patient’s auditory canals.

6. Insert gently one cone in the auditory canal of the patient. Turn gently the button to close the arms. Insert gently the second cone in the auditory canal of the patient.

KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System User Guide (SM735)_Ed 02

5–13

Acquiring a Submento-Vertex Image

Launching the X-ray

To launch the x-ray, follow these steps:

1. Leave the x-ray room and close the door. You must keep visual contact with the patient during acquisition.

IMPORTANT

To stop the acquisition, if any problem, release the exposure button of the remote control or press the red emergency stop button.

5–14 ACQUIRING CEPHALOMETRIC IMAGES

Acquiring a Submento-Vertex Image

2. Launch the x-ray with the remote control. Press and hold the exposure button until the end of acquisition. The turns yellow indicating x-ray emission. The image appears on the Preview Screen of the Cephalometric Acquisition Window. When the acquisition ends, the Cephalometric Acquisition Window disappears and the acquired image is automatically transferred to the Imaging Window.

3. Check the image quality. If satisfactory, remove the ear cones and release the patient.

KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System User Guide (SM735)_Ed 02

5–15

Acquiring a Carpus Image

Acquiring a Carpus Image

Before acquiring a carpus image, check that you have:

• Reset the unit rotative arm to the start position for patient to enter the unit.

• Selected the patient record.

• Accessed the Imaging Window.

• Accessed the Cephalometric Acquisition Window.

Preparing the Unit and Setting the Acquisition Parameters

To acquire a carpus image, follow these steps:

1. Raise and block the sensor manually.

2. On the x-ray source assembly, set the collimator selector to AP/PA.

3. Position the head clamps manually for a frontal AP exam.

IMPORTANT

You must position the head clamps manually because they are not positioned automatically from the Program pane exam type selection. In this case, the relevant exam type selection icon becomes active.

4. In the Cephalometric Acquisition Window, click the Program button to access the

Program pane. In the Program pane, click for a carpus exam.

5. Click the Patient button to access the Patient pane. Select the patient:

• Type

• Size

6. If the default parameter setting is not adapted to your patient type, click the

Parameter button and in the Parameter pane select the appropriate parameters. To save the new parameter settings as the default settings, click

Memorize settings.

fand select

5–16 ACQUIRING CEPHALOMETRIC IMAGES

Preparing and Positioning the Patient

To prepare and position the patient, follow these steps:

1. Ask the patient to remove all metal objects.

2. Ask the patient to wear a lead apron. Ensure that the apron lays flat across the patient’s shoulders.

3. Press and hold the height adjustment buttons to raise the cephalostat head.

4. Ask the patient to stand on the side of the cephalometric unit, open the hand fully and place it on the sensor.

Acquiring a Carpus Image

WARNING

Make sure that the patient is not positioned in front of the x-ray emission.

KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System User Guide (SM735)_Ed 02

5–17

Acquiring a Carpus Image

Launching the X-ray

To launch the x-ray, follow these steps:

1. Leave the x-ray room and close the door. You must keep visual contact with the patient during acquisition.

IMPORTANT

To stop the acquisition, if any problem, release the exposure button of the remote control or press the red emergency stop button.

2. Launch the x-ray with the remote control. Press and hold the exposure button until the end of acquisition. The turns yellow indicating x-ray emission. The image appears on the Preview Screen of the Cephalometric Acquisition Window. When the acquisition ends, the Cephalometric Acquisition Window disappears and the acquired image is automatically transferred to the Imaging Window.

3. Check the image quality, if satisfactory, then release the patient.

5–18 ACQUIRING CEPHALOMETRIC IMAGES

X-Ray Dose Emission Information

X-Ray Dose Emission Information

Compliance with

EURATOM 97/43 Directive

You can right-click on each image to display the estimated emitted dose received by the patient. You can use this information to calculate the effective dose received by the patient for the image.

The radiation emission dose is expressed in mGy.cm2. This dose is measured at the primary collimator outlet. The dose is accurate to +/-30%. The primary slot is 0.5 mm wide and 18.2 mm high.

KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System User Guide (SM735)_Ed 02

5–19

X-Ray Dose Emission Information

5–20 ACQUIRING CEPHALOMETRIC IMAGES

Chapter 6

MAINTENANCE

This section describes the maintenance tasks that you need to perform regularly for your

KODAK 8000C unit and the accessories.

WARNING

Switch off the unit, then, clean all accessible parts of the machine with an alcohol-based non-corrosive product. Avoid using liquids inside the unit. Follow the alcohol-based product manufacturer recommendations for safety precautions.

CAUTION

You can use the usual disinfectant products, but we recommend that you protect the unit from contamination by using barriers available from dental distributors. Follow the disinfectant product manufacturer recommendations for safety precautions.

Daily

Carry out the following maintenance tasks:

Table 6–1 Daily Maintenance Tasks

Accessories

Head clamps and ear cones

Nasion support

All components that come into contact with the patient and the

operator

Outer covers of the unit

Maintenance Tasks

Clean with medical-grade 76% alcohol disinfectant before the next patient is x-rayed.

Clean all components with medical-grade 76% alcohol disinfectant before the next patient is x-rayed.

Wipe the unit with a dry cloth at the end of each day's operation.

WARNING

Do not use detergents or solvents to clean the outer covers of the unit.

Monthly

Wipe the outer covers of the unit with a soft, dry cloth.

Annually

We recommend a general inspection of the unit carried out by an authorized service technician.

KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System User Guide (SM735)_Ed 02

6–1

Annually

6–2 MAINTENANCE

Chapter 7

TROUBLESHOOTING

Quick Troubleshooting

Occasionally, malfunctions can occur during use in the event of an incorrect action. An information (I) error code is displayed on the Display Screen of the unit Control Panel and the message is displayed on the popup on the

Acquisition Window System Status Screen. In some cases, an audible warning is also issued.

The following table lists the information messages, their description and the action to take:

IMPORTANT

If an “E” message is displayed, the malfunction persists or more serious conditions occur, contact a qualified technician.

When you call the qualified technician have the following information ready:

Model Number: KODAK 8000C

Error Code Number: E xx

Message displayed on the popup on the Acquisition

Window.

Table 7–1 Information Messages

Information Error Code Information Message

I 1

X-Ray tube cooling

I 2

I 3

I 4

I 5

I 6

Thermal security

Release handswitch

Ceph sensor position

Head clamps position

Wrong rotative arm position

Description Action

Cooling in progress.

Cooling in progress.

The acquisition has ended.

The cephalometric sensor position does not match the selected radiological program.

Wait until the Generator Cooling Indicator on the Acquisition Window reaches zero.

Wait until the Generator Cooling Indicator on the Acquisition Window reaches zero.

Release the exposure button of the x-ray remote control.

Position correctly the sensor for:

LA, for lateral position

AP/PA, for all the other radiologica programs

The head clamps position does not match the selected exam program.

Position manually the head clamps for the selected exam.

The exposure button of the x-ray remote control is inactive because the rotative arm is not in the start position.

Press position.

to reset the rotative arm in start

KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System User Guide (SM735)_Ed 02

7–1

Quick Troubleshooting

Table 7–1 Information Messages (Continued)

Information Error Code Information Message

I 7

Raise sensor

I 8

I 9

I 10

I 13

I 14

I 15

Lower sensor

Cooling Ceph sensor

Collimator position

Mains too low

Mains too high

Interface inactive

Description Action

The panoramic sensor is incorrectly positioned.

The panoramic sensor is incorrectly positioned.

Check that the panoramic sensor is fully retracted and locked in the upper position.

Check that the panoramic sensor is fully retracted and locked in the lower position.

The Cephalometric sensor is in the cooling process.

Wait until the Ready Indicator LED is green.

The collimator is not correctly positioned or does not match the radiological program selection.

Reposition the collimator.

There is a problem with the mains current input.

There is a problem with the mains current input.

Call a qualified technician.

Call a qualified technician.

The Acquisition Window cannot be accessed.

1. Check that the unit is switched on.

2. Wait for the connection between the unit and the PC.

3. Check that the Acquisition Window is not masked by another application, in this case close the masking application.

7–2 TROUBLESHOOTING