User’s
Manual
Models 436101/436102/436103/436104/436106
µR10000 Recorder
IM 04P01B01-01E
Yokogawa Electric Corporation
3rd Edition
Foreword
Thank you for purchasing the YOKOGAWA µR10000 Recorder.
This manual describes the functions (excluding the communication functions), installation
and wiring procedures, operating procedures, and lists the handling precautions of the
µR10000 Recorder. To ensure correct use, please read this manual thoroughly before
beginning operation.
The following three manuals including this manual are available for the µR10000
Recorder.
• Electronic Manuals Provided on the Accompanying CD-ROM
Manual Title
Manual No.
Description
µR10000 Recorder
User’s Manual
IM 04P01B01-01E
This manual.
µR10000/µR20000
IM 04P01B01-17E
Communication Interface
User’s Manual
Explains the communication functions of the
µR10000 Recorder using Ethernet interface and
the RS-422A/485 communication interface.
• Paper Manual
Manual Title
Manual No.
Description
µR10000 Recorder
Operation Guide
IM 04P01B01-02E
Explains concisely the operations of the
µR10000 Recorder.
Notes
• The contents of this manual are subject to change without prior notice as a result of
continuing improvements to the instrument’s performance and functions.
• Every effort has been made in the preparation of this manual to ensure the accuracy
of its contents. However, should you have any questions or find any errors, please
contact your nearest YOKOGAWA dealer as listed on the back cover of this manual.
• Copying or reproducing all or any part of the contents of this manual without the
permission of Yokogawa Electric Corporation is strictly prohibited.
• The TCP/IP software of this product and the document concerning the TCP/IP
software have been developed/created by YOKOGAWA based on the BSD
Networking Software, Release 1 that has been licensed from the University of
California.
Trademarks
• All the brands or names of Yokogawa Electric’s products used in this manual are
either trademarks or registered trademarks of Yokogawa Electric Corporation.
• Microsoft, MS-DOS, Windows, Windows NT, and Windows XP are either registered
trademarks or trademarks of Microsoft Corporation in the United States and/or other
countries.
• Adobe, Acrobat, and PostScript are trademarks of Adobe Systems incorporated.
• For purposes of this manual, the TM and ® symbols do not accompany their
respective trademark names or registered trademark names.
• Company and product names that appear in this manual are trademarks or registered
trademarks of their respective holders.
Revisions
1st Edition
2nd Edition
3rd Edition
December 2004
March 2005
August 2005
3rd Edition: August 2005 (YK)
All Rights Reserved, Copyright © 2004 Yokogawa Electric Corporation
IM 04P01B01-01E
i
Recorder’s Version and Functions Described in This
Manual
The contents of this manual corresponds to the recorder with version 1.11.
µR10000 Versions and Functions
Version
Suffix Code
1.02 or earlier –
1.11
–
–
–
–
/C3
/C7
•
Added or Modified Functions
–
(Added)
The printout/display format of the date can be changed.
(Added)
Key operation to move the printer carriage near the center
position so that the ribbon cassette can be replaced with
the power turned ON (dot model)
(Changed) Selectable range of alarm values during linear scaling
(including 1-5V and SQRT) is –5% to 105% of the scale
span.
(Changed) Specify the date/time for switching DST and standard time
by the month, the nth day of the week of that month, and
time.
(Added)
Modbus slave protocol
Two-wire system
(Changed) Users with the same user name cannot be registered.
Reference
–
Section 7.19
Section 3.4
Section 5.2
Sections 5.5
and 6.13
Communication
manual
Communication
manual
Checking the Version Number
You can check the version number on the System display.
The System display cannot be shown at the factory default condition.
First, register the System display to the display screen.
• Procedure of registering the System display to the display screen: See section 8.2.
• Procedure of displaying the System display: The screen switches each time the DISP key is pressed. Press the DISP
key repeatedly until System display is shown. The displayed contents on the System display switches every 3
seconds. Check the number shown by the “Version:” item.
Software (Sold Separately)
The µR10000 Configuration Software cannot set the new functions (indicated above) available on µR10000
version 1.11. The new functions are planned to be supported on the configuration software to be released after
November 2005.
ii
IM 04P01B01-01E
Safety Precautions
The general safety precautions described here must be observed during all phases of operation.
Safety Standards and EMC Standards
This recorder conforms to IEC safety class I (provided with terminal for protective grounding), Installation Category
II, Measurement Category II (CAT II), and EN61326-1 (EMC standard), class A (use in a commercial, industrial, or
business environment). This recorder is designed for indoor use.
About This Manual
• This manual should be read by the end user.
• Read this manual thoroughly and have a clear understanding of the product before operation.
• This manual explains the functions of the product. YOKOGAWA does not guarantee that the product will suit a
particular purpose of the user.
• Under absolutely no circumstances may the contents of this manual be transcribed or copied, in part or in
whole, without permission.
• The contents of this manual are subject to change without prior notice.
• Every effort has been made in the preparation of this manual to ensure the accuracy of its contents. However,
should you have any questions or find any errors or omissions, please contact your nearest YOKOGAWA
dealer.
Precautions Related to the Protection, Safety, and Alteration of the Product
• The following safety symbols are used on the product and in this manual.
“Handle with care.” To avoid injury and damage to the instrument, the operator must refer to the
explanation in the manual.
Protective ground terminal
AC
“High temperature.” To avoid injury caused by hot surface, do not touch locations where this
symbol appears.
• For the protection and safe use of the product and the system controlled by it, be sure to follow the instructions
and precautions on safety that are stated in this manual whenever you handle the product. Take special note
that if you handle the product in a manner that violate these instructions, the protection functionality of the
product may be damaged or impaired. In such cases, YOKOGAWA does not guarantee the quality,
performance, function, and safety of the product.
• When installing protection and/or safety circuits such as lightning protection devices and equipment for the
product and control system or designing or installing separate protection and/or safety circuits for fool-proof
design and fail-safe design of the processes and lines that use the product and the control system, the user
should implement these using additional devices and equipment.
• If you are replacing parts or consumable items of the product, make sure to use parts specified by
YOKOGAWA.
• This product is not designed or manufactured to be used in critical applications that directly affect or threaten
human lives. Such applications include nuclear power equipment, devices using radioactivity, railway facilities,
aviation equipment, air navigation facilities, aviation facilities, and medical equipment. If so used, it is the user’s
responsibility to include in the system additional equipment and devices that ensure personnel safety.
• Do not modify this product.
IM 04P01B01-01E
iii
Safety Precautions
WARNING
• Power Supply
Ensure that the source voltage matches the voltage of the power supply before turning ON the power.
• Protective Grounding
Make sure to connect the protective grounding to prevent electric shock before turning ON the power.
• Necessity of Protective Grounding
Never cut off the internal or external protective earth wire or disconnect the wiring of the protective earth
terminal. Doing so invalidates the protective functions of the instrument and poses a potential shock
hazard.
• Defect of Protective Grounding
Do not operate the instrument if the protective earth or fuse might be defective. Make sure to check
them before operation.
• Do Not Operate in an Explosive Atmosphere
Do not operate the instrument in the presence of flammable liquids or vapors. Operation in such
environments constitutes a safety hazard.
• Do Not Remove Covers
The cover should be removed by YOKOGAWA’s qualified personnel only. Opening the cover is
dangerous, because some areas inside the instrument have high voltages.
• External Connection
Connect the protective grounding before connecting to the item under measurement or to an external
control unit.
• Damage to the Protective Structure
Operating the recorder in a manner not described in this manual may damage its protective structure.
Exemption from Responsibility
• YOKOGAWA makes no warranties regarding the product except those stated in the WARRANTY that is
provided separately.
• YOKOGAWA assumes no liability to any party for any loss or damage, direct or indirect, caused by the user or
any unpredictable defect of the product.
Handling Precautions of the Software
• YOKOGAWA makes no warranties regarding the software accompanying this product except those stated in
the WARRANTY that is provided separately.
• Use the software on a single PC.
• You must purchase another copy of the software, if you are to use the software on another PC.
• Copying the software for any purposes other than backup is strictly prohibited.
• Please store the original media containing the software in a safe place.
• Reverse engineering, such as decompiling of the software, is strictly prohibited.
• No portion of the software supplied by YOKOGAWA may be transferred, exchanged, sublet, or leased for use
by any third party without prior permission by YOKOGAWA.
iv
IM 04P01B01-01E
Checking the Contents of the Package
Unpack the box and check the contents before operating the recorder. If some of the
contents are not correct or missing or if there is physical damage, contact the dealer
from which you purchased them.
Checking the Model
A name plate is affixed to the recorder. Check that the model name and suffix code
given on the name plate on the rear panel match those on your order.
RECORDER
R
C
US
LR99988
N200
MODEL
SUFFIX
STYLE
SUPPLY
FREQUENCY
NO.
Made in China
MODEL and SUFFIX Code
Model
436101
436102
436103
436104
436106
Suffix Code
Optional Code
-1
-2
/A1
/A2
/A3
/C3
/C7
/F1
/H2
/H3
/M1
/N1
/N2
/N3
/R1
1
2
3
4
5
6
IM 04P01B01-01E
Description
µR10000 1 pen recorder
µR10000 2 pen recorder
µR10000 3 pen recorder
µR10000 4 pen recorder
µR10000 6 dot recorder
Japanese
English & deg F / DST
Alarm output relay 2 points1
Alarm output relay 4 points1
Alarm output relay 6 points1, 2
RS-422A/485 interface3
Ethernet (10BASE-T) interface3
Fail/Chart end detection and output2
Clamped input terminal4
Non-glare door glass
Mathematical function
Cu10, Cu25 RTD input
3 legs isolated RTD4, 5
Expansion inputs6
Remote control 5 points
/A1, /A2, and /A3 cannot be specified simultaneously.
/A3 and /F1 cannot be specified simultaneously.
/C3 and /C7 cannot be specified simultaneously.
/H2 and /N2 cannot be specified simultaneously.
Valid only on the model 436106.
14 types of input including Pt50 RTD, PR40-20, and Platinel TC
v
Checking the Contents of the Package
Standard Accessories
The standard accessories below are supplied with the recorder. Check that all contents
are present and undamaged.
Z-fold chart
paper
Ribbon
cassette
Disposable
felt pen
Manuals for the µR10000/µR20000
(CD-ROM)
Plotter pen
Mounting bracket
µR10000 Recorder Operation Guide
IM 04P01B01-02E
Item
1-Pen
1
Red
1
Green
Disposable felt pen
Blue
Violet
Purple
1
Plotter pen
2
Mounting bracket
Manuals for the µR10000/µR20000 (CD-ROM)
1
µR10000 Recorder Operation Guide
1
IM 04P01B01-02E
Z-fold chart paper
Ribbon cassette
2-Pen
1
1
1
1
2
1
3-Pen
1
1
1
1
1
2
1
1
1
4-Pen Dot Printing
1
1
1
1
1
1
1
1
2
2
1
1
1
1
Optional Accessories (Sold Separately)
The optional accessories below are available for purchase separately. If you make an
order, make sure that all contents are present and undamaged.
For information about ordering accessories, contact the dealer from which you
purchased the recorder.
Item
Z-fold chart paper
Ribbon cassette
Red
Green
Blue
Violet
Purple
Disposable felt pen
Plotter pen
Mounting bracket
Shunt resistor
for the screw terminal (standard)
Shunt resistor
for the clamped input terminal (/H2)
Model
Quantity
(Part Number)
B9565AW
1
1
B9901AX
1
B9902AM
1
B9902AN
1
B9902AP
1
B9902AQ
1
B9902AR
2
B9900BX
1
415920
1
415921
1
415922
1
438920
1
438921
1
438922
Note
10 pcs.
3 pcs.
3 pcs.
3 pcs.
3 pcs.
3 pcs.
250 Ω ± 0.1%
100 Ω ± 0.1%
10 Ω ± 0.1%
250 Ω ± 0.1%
100 Ω ± 0.1%
10 Ω ± 0.1%
Software (Sold Separately)
Item
Note
Model
RXA10-01
µR10000 Configuration software
RXA10-02
With interface unit
* The new functions of the µR10000 version 1.11 are planned to be supported on the
configuration software to be released after November 2005.
vi
IM 04P01B01-01E
Checking the Contents of the Package
Removing the Packing Materials
Open the door, put your finger on the tab at the lower left of the display, and open the
display.
Tab on the display
Open
Display
Open
Remove all packing materials.
• Pen Model
Hinge
• Dot Model
Hinge
CAUTION
To protect the hinges, do not apply vertical force on the display.
IM 04P01B01-01E
vii
How to Use This Manual
This user’s manual consists of the following sections.
For details on communication functions, see the µR10000/µR20000 Communication
Interface User’s Manual (IM 04P01B01-17E) on the CD-ROM.
Chapter
Title and Description
1
Functional Description
Describes the functions of the µR10000 Recorder. Refer to this chapter when you
are unsure of the details of the function that you are operating.
2
Before Using the Recorder
Describes the installation and wiring procedures.
3
Names of Parts and Run Operations
Describes the names of each part of the recorder and the daily operations.
4
Common Operations for Setting Functions and Setup Guide
Describes the execution modes of the recorder, basic setup operations using keys,
and provides a function setup guide.
5
Frequently Used Setup Operations (Setting Mode)
Describes how to change the input range, alarms, chart speed, etc.
6
Setup Operations for Convenient Functions (Setting Mode)
Describes the setup operations for convenient functions such as how to assign tags
to channels and how to set message strings that are to be printed.
7
Setup Operations for Changing/Adding Functions (Basic Setting Mode)
Describes the setup operations for changing or adding functions such as setting the
recorder to detect sensor burnouts and changing the contents that are printed on the
chart paper.
8
Setup Operations for Changing the Displayed Contents
Describes how to select the display type suitable for the application.
9
Operations Related to the Computation Function (/M1 Option)
Describes all operations related to the computation function.
10
Troubleshooting
Describes error message and troubleshooting measures of the µR10000 Recorder.
11
Maintenance
Describes periodic inspection, calibration, pen adjustment/printer carriage
adjustment, and recommended replacement period for worn parts.
12
Specifications
Gives the specifications of the µR10000 Recorder.
Appendix Describes the printout contents.
Index
Note
•
•
viii
This user’s manual covers information regarding the recorders with English as the display/
recording language (suffix code “2”).
For the procedure of setting the display/recording language, see section 7.14, “Changing the
Display/Recording Language.”
IM 04P01B01-01E
How to Use This Manual
1
Conventions Used in This Manual
Unit
K ........ Denotes 1024. Example: 768 KB (file size)
k ........ Denotes 1000.
2
Safety Markings
The following markings are used in this manual.
3
Improper handling or use can lead to injury to the user or
damage to the instrument. This symbol appears on the
instrument to indicate that the user must refer to the user’s
manual for special instructions. The same symbol appears in
the corresponding place in the user’s manual to identify those
instructions. In the manual, the symbol is used in conjunction
with the word “WARNING” or “CAUTION.”
WARNING
4
5
Calls attention to actions or conditions that could cause serious
or fatal injury to the user, and precautions that can be taken to
6
prevent such occurrences.
CAUTION
Calls attentions to actions or conditions that could cause light
injury to the user or damage to the instrument or user’s data,
and precautions that can be taken to prevent such occurrences.
Note
Calls attention to information that is important for proper
7
8
operation of the instrument.
Subheadings
On pages that describe the operating procedures in Chapter 3 through 9, the following
symbols are used to distinguish the procedures from their explanations.
Bold characters denote keys or character strings that are displayed on the screen.
Example: Range, Unit
9
10
Procedure
Follow the numbered steps. All procedures are written with
inexperienced users in mind; depending on the operation, not
all steps need to be taken.
Explanation
11
This subsection describes the setting parameters and the
limitations on the procedures. It does not give a detailed
12
explanation of the function. For details on the function, see
chapter 1.
App
Index
IM 04P01B01-01E
ix
Contents
Foreword ......................................................................................................................................... i
Recorder’s Version and Functions Described in This Manual ......................................................... ii
Safety Precautions ......................................................................................................................... iii
Checking the Contents of the Package .......................................................................................... v
How to Use This Manual .............................................................................................................. viii
Chapter 1 Functional Description
1.1
1.2
1.3
1.4
1.5
1.6
1.7
1.8
1.9
Overview of the Recorder ................................................................................................ 1-1
Measuring Input Section ................................................................................................. 1-2
Alarms .............................................................................................................................. 1-7
Recording ....................................................................................................................... 1-13
Display ........................................................................................................................... 1-22
Computation Function (/M1 Option) ............................................................................... 1-24
FAIL/Chart End Detection and Output Function (/F1 Option) ........................................ 1-27
Remote Control Function (/R1 Option) .......................................................................... 1-28
Other Functions ............................................................................................................. 1-30
Chapter 2 Before Using the Recorder
2.1
2.2
2.3
2.4
2.5
2.6
Handling Precautions ....................................................................................................... 2-1
Installation ........................................................................................................................ 2-2
Input Signal Wiring ........................................................................................................... 2-5
Optional Terminal Wiring .................................................................................................. 2-9
Power Supply Wiring ...................................................................................................... 2-12
Turning ON/OFF the Power Switch ................................................................................ 2-14
Chapter 3 Names of Parts and Run Operations
3.1
3.2
3.3
3.4
3.5
3.6
3.7
3.8
3.9
3.10
3.11
3.12
3.13
Names of Parts ................................................................................................................ 3-1
Installing or Replacing the Chart Paper ........................................................................... 3-4
Installing/Replacing Felt Pens or Plotter Pen (Pen Model) .............................................. 3-7
Installing/Replacing the Ribbon Cassette (Dot Model) .................................................... 3-9
Starting/Stopping the Recording .................................................................................... 3-11
Switching the Display Screen ........................................................................................ 3-12
Printing Measured Values (Manual Printout) ................................................................. 3-13
Printing the Recorder Settings ....................................................................................... 3-14
Clearing the Alarm Printout Buffer ................................................................................. 3-15
Printing Messages ......................................................................................................... 3-16
Resetting the Report Data of the Periodic Printout ........................................................ 3-17
Releasing the Alarm Output (Alarm ACK Operation) ..................................................... 3-18
Activating/Releasing the Key Lock ................................................................................. 3-19
Chapter 4 Common Operations for Setting Functions and Setup Guide
4.1
4.2
4.3
4.4
Run Modes ....................................................................................................................... 4-1
Key Operations ................................................................................................................ 4-2
Menu Structure, Settings, and List of Default Values ....................................................... 4-5
Function Setup Guide .................................................................................................... 4-13
Chapter 5 Frequently Used Setup Operations (Setting Mode)
5.1
5.2
5.3
x
Setting the Input Range ................................................................................................... 5-1
Setting the Alarm .............................................................................................................. 5-9
Setting the Unit on Scaled Channels ............................................................................. 5-12
IM 04P01B01-01E
Contents
5.4
5.5
Changing the Chart Speed ............................................................................................ 5-13
Setting the Date/Time .................................................................................................... 5-14
Chapter 6 Setup Operations for Convenient Functions (Setting Mode)
6.1
6.2
6.3
6.4
6.5
6.6
6.7
6.8
6.9
6.10
6.11
6.12
6.13
Setting the Trend Recording Interval (Dot Model). ........................................................... 6-1
Setting the Filter (Pen Model) .......................................................................................... 6-2
Setting the Moving Average (Dot Model) ......................................................................... 6-3
Setting Recording Zones for Each Channel (Zone Recording) ........................................ 6-4
Setting the Partial Expanded Recording .......................................................................... 6-5
Turning Trend Recording (Dot Model) and Periodic Printout ON/OFF for Each Channel 6-6
Setting Tags on Channels ................................................................................................ 6-7
Setting the Message String .............................................................................................. 6-8
Setting the Secondary Chart Speed (Remote Control Function, /R1) ............................. 6-9
Setting the Alarm Delay Duration ................................................................................... 6-10
Setting the Brightness of the Display and Internal Light ................................................ 6-11
Applying a Bias on the Measuring Input Signal ............................................................. 6-12
Setting the Date/Time When Switching between Standard Time and DST ................... 6-13
Chapter 7 Setup Operations for Changing/Adding Functions (Basic Setting
Mode)
7.1
7.2
7.3
7.4
7.5
7.6
7.7
7.8
7.9
7.10
7.11
7.12
7.13
7.14
7.15
7.16
7.17
7.18
7.19
7.20
Changing the Auxiliary Alarm Function ............................................................................ 7-1
Changing the Integration Time of the A/D Converter ....................................................... 7-4
Setting the Burnout Detection Function of Thermocouples ............................................. 7-5
Setting the RJC Function on Channels Set to TC Input ................................................... 7-6
Changing the Channel Recording Color (Dot Model) ...................................................... 7-8
Recording by Compensating for the Pen Offset along the Time Axis (Pen Model) .......... 7-9
Turning Printouts ON/OFF. ............................................................................................. 7-10
Setting the Periodic Printout Interval and the Type of Measured Values to Be Printed . 7-12
Setting the Bar Graph Display Mode ............................................................................. 7-15
Setting the Key Lock Function ....................................................................................... 7-16
Enabling the Moving Average Function (Dot Model) ...................................................... 7-19
Enabling the Filter Function (Pen Model) ....................................................................... 7-20
Enabling the Partial Expanded Recording Function ...................................................... 7-21
Changing the Display/Recording Language .................................................................. 7-22
Enabling the Bias Function, Low-Cut Function, and Alarm Delay Function ................... 7-23
Changing the Time Printout Format ............................................................................... 7-25
Initializing the Settings ................................................................................................... 7-27
Assigning Functions to the Remote Control Input Terminals (/R1 Option) ..................... 7-28
Changing the Printout/Display Format of the Date ........................................................ 7-30
Changing the Temperature Unit ..................................................................................... 7-31
IM 04P01B01-01E
3
4
5
7
8
9
10
11
12
Key Operations for Changing the Displayed Information ................................................. 8-1
Changing the Displayed Information ................................................................................ 8-3
App
Chapter 9 Operations Related to the Computation Function (/M1 Option)
9.1
9.2
9.3
9.4
9.5
9.6
9.7
2
6
Chapter 8 Setup Operations for Changing the Displayed Contents
8.1
8.2
1
Starting/Stopping/Resetting the Computation .................................................................. 9-1
Setting the Computing Equation ...................................................................................... 9-2
Setting the Unit ................................................................................................................ 9-9
Setting the Constants Used in Equations ...................................................................... 9-10
Setting the Alarm ............................................................................................................ 9-11
Specifying the Timer Used in Statistical Calculations (TLOG) ....................................... 9-13
Setting Recording Zones for Each Channel (Zone Recording) ...................................... 9-15
xi
Index
Contents
9.8
9.9
9.10
9.11
9.12
9.13
9.14
9.15
9.16
9.17
Setting the Partial Expanded Recording ........................................................................ 9-16
Turning Trend Recording (Dot Model) and Periodic Printout ON/OFF
for Each Channel ................................................................................................ 9-17
Setting Tags on Channels .............................................................................................. 9-18
Setting the Alarm Delay Duration ................................................................................... 9-19
Setting the Timer Used in TLOG Computation and Periodic Printout ............................ 9-20
Changing the Channel Recording Color (Dot Model) .................................................... 9-23
Changing the Channel Assignments of Recording Pens (Pen Model) ........................... 9-24
Changing the Type of Report Data Printed in Periodic Printout ..................................... 9-25
Setting the Bar Graph Display Mode ............................................................................. 9-27
Setting the Procedure Taken When the Computed Result Is in Error ............................ 9-28
Chapter 10 Troubleshooting
10.1
10.2
A List of Error Messages ................................................................................................ 10-1
Troubleshooting Flow Charts ......................................................................................... 10-4
Chapter 11 Maintenance
11.1
11.2
11.3
11.4
11.5
11.6
11.7
Periodic Inspection ......................................................................................................... 11-1
Cleaning the Recorder ................................................................................................... 11-2
Replacing the Internal Light LED ................................................................................... 11-3
Calibrating the Recorder ................................................................................................ 11-4
Adjusting the Pen Position (Pen Model) ........................................................................ 11-6
Adjusting the Dot Printing Position (Dot Model) ............................................................. 11-7
Recommended Replacement Periods for Worn Parts ................................................... 11-9
Chapter 12 Specifications
12.1
12.2
12.3
12.4
12.5
12.6
12.7
Input Specifications ........................................................................................................ 12-1
Alarm Function Specifications ........................................................................................ 12-3
Recording Function Specifications ................................................................................. 12-4
Display Function Specifications ..................................................................................... 12-7
Specifications of Optional Functions ............................................................................ 12-13
General Specifications ................................................................................................. 12-17
Dimensional Drawings ................................................................................................. 12-21
Appendix
Appendix 1
Periodic Printout and Printout Using the TLOG Timer (/M1 Option) ................. App-1
Index
xii
IM 04P01B01-01E
Chapter 1 Functional Description
1.1
Overview of the Recorder
1
µR10000 Recorder
Recording example (dot model)
Alarms
For each channel, various alarms such as high limit alarm and low limit alarm can be
assigned to monitor the measured values. Alarm output relays can be used to output
contact signals when alarms occur (/A1, /A2, and /A3 options).
Recording
The measured results are recorded with pens or dots on a chart paper (trend recording).
The chart speed can be selected from 5 to 12000 mm/h on the pen model and 1 to 1500
mm/h on the dot model.
In addition to trend recording, various types of information can be printed or recorded on
the chart paper such as numeric measured values, alarm occurrence/release, and
predefined messages.
Also, the recorder settings can be printed.
Internal Light
A light is provided for easier viewing of the recording area of the chart paper.
Display
Measured values can be displayed numerically or using bar graphs on the large display.
Also, alarm status and chart speed can be displayed.
Communication Function
Using the Ethernet communication interface (/C7 option) or the RS-422A/485
communication interface (/C3 option), the measured values on the recorder can be
output to a PC or a PC can be used to control the recorder.
This manual does not cover the communication functions. For details on communication
functions, see the µR10000/µR20000 Communication Interface User’s Manual (IM
04P01B01-17E) on the CD-ROM.
Other Main Functions
The computation function (/M1 option) can be used to perform various computations from
four arithmetic operations to statistical calculations on 8 and 12 computation channels on
the pen model and dot model, respectively. The computed results can be recorded.
The remote control function (/R1 option) can be used to control the start/stop and other
operations of the recorder by applying contact signals to the dedicated terminals.
The FAIL/chart end detection and output function (/F1 option) can be used to output
contact signals when errors are detected on the recorder or when the chart paper runs
out.
IM 04P01B01-01E
1-1
Functional Description
The µR10000 Recorder (hereafter referred to as the recorder) can be used to assign DC
voltage, 1-5V, thermocouple, RTD, and contact or voltage ON/OFF signal to channels for
measurement. The measured results are recorded with pens or dots on a chart paper
that is fed at a constant speed. The pen model can record up to 4 channels; the dot
model can record up to 6 channels.
1.2
Measuring Input Section
Input Section
Number of Measurement Channels and Scan Interval
The recorder samples the input signals on the measurement channels at the scan
interval to obtain the measured values.
Model
Number of Channels
Scan Interval
1-pen model
2-pen model
3-pen model
4-pen model
Dot model
1
2
3
4
6
125 ms
125 ms
125 ms
125 ms
1s
(However, the scan interval is 2.5 s when the
integration time of the A/D converter is 100 ms.)
Input Type, Measurable Range, and Computation
The recorder can measure the following types of inputs.
Input Type
Measurable Range
DC voltage
DC voltage in the range of ±20 mV to ± 50 V
1-5V
See “1-5V” below.
Thermocouple Temperature range corresponding to each type: R, S, B, K, E, J, T, N, W, L, U,
and WRe
RTD
Temperature range corresponding to each type: Pt100Ω and JPt100Ω
ON/OFF input Contact input: Open contact is OFF (0). Closed contact is ON (1).
Voltage input: Less than 2.4 V is OFF (0). Greater than or equal to 2.4 V is ON (1)
• 1-5V
1-5V is scaled to values in the appropriate unit to be used as measured values. Also,
the low-cut function (input less than 0% is fixed to 0 (scale left value)) can be used.
• Current Input
A shunt resistor is attached to the input terminal. The current signal is converted to a
voltage signal and measured. The measurable range is the range equivalent to the
“DC voltage” range indicated above after converting the current to the voltage signal.
Note
Three types of shunt resistors (250 Ω, 100 Ω, and 10 Ω) are available for current input (see
“Optional Accessories (Sold Separately)” on page v). For example, a 250-Ω shunt resistor is
used to convert the signal to the range of 1 to 5 V for 4 to 20 mA input.
• Range Type, Measurable Range, and Recording Span
Various “range type” are available for the different types of inputs (for example
thermocouple R). Each range type has a preset measurable range (0.0 to 1760.0°C
for thermocouple R). Measurement can be made by specifying an arbitrary range
within the measurable range as the input range. The measured values in the input
range are recorded on the chart paper. The range of measured values that are
recorded is called the recording span.
Measurable range
(Thermocouple R example)
1760.0°C
Input range or
recording span
1500.0°C (rightmost value of span)
300.0°C (leftmost value of span)
0.0°C
<Related Topics>
Setting the input range: Section 5.1
1-2
IM 04P01B01-01E
1.2 Measuring Input Section
Channel set to delta computation
Input
value
Measured value
–
Measured value on the reference channel
DC voltage
Note
A channel whose input type is set to DC voltage, TC, or RTD can be designated as a
reference channel. However, channels set to scaling or square root computation cannot be
designated.
• Scaling
The input values are scaled to values in the appropriate unit to be used as measured
values.
Input value
10 V
Measured value
300.0°C
0V
−100.0°C
• Square Root Computation
When the input type is DC voltage, the square root of the input value is calculated, the
result is scaled to a value in the appropriate unit, and used as the measured value of
the channel. Also, the low-cut function (input less than a given measured value is
fixed to 0 (scale left value)) can be used.
Channel set to square root computation
Input value
√
Scaling
Measured value
Measured value
Result of square
root computation
Low-cut value
Note
The square root computation on the recorder uses the following formula.
Fx = ( Fmax - Fmin )
Vx - Vmin
Vmax - Vmin
+ Fmin
where Vmin (leftmost value of span) < Vmax (rightmost value of span)
Fmin (leftmost value of scale after scaling) < Fmax (rightmost value of scale after
scaling)
Vx is the input voltage and Fx is the scaled value
IM 04P01B01-01E
1-3
1
Functional Description
• Delta Computation
The value obtained by subtracting the measured value of another channel (called the
reference channel) from the input value of the channel set to delta computation is used
as the measured value of that channel. The reference channel must be assigned to a
channel whose channel number is less than that of the channel on which delta
computation is specified. The channel on which delta computation is specified is
automatically set to the same range type as the reference channel.
1.2 Measuring Input Section
Bias
A given value (bias value) is added to the input value and used as the measured value of
that channel.
Biased channel
+
Input value
Measured value
+
Bias value
<Related Topics>
Setting the bias: Section 7.15 and 6.12
Burnout Detection of Thermocouples
This function makes the recording go off the scale to the right or left when the
thermocouple burns out while measuring temperature with a thermocouple. This
function can also be used on 1-5V. The burnout detection function can be set for each
channel.
By default, this function is disabled.
Note
For 1-5V, a burnout occurs when the input value is less than or equal to 0.2 V.
<Related Topics>
Setting the burnout detection function: Section 7.3
Reference Junction Compensation of Thermocouple Input
When measuring the temperature using a thermocouple, the reference junction
compensation on the recorder can be used. When using external reference junction
compensation, you can set the reference voltage. The reference junction compensation
can be set for each channel.
By default, the recorder is configured to use the internal reference junction compensation
function.
Note
When using external reference junction compensation, set an appropriate reference junction
compensation voltage. For example, if the reference junction temperature of the external
reference compensation is T0 °C, set the reference compensation junction voltage to the
thermoelectromotive force of the 0°C reference of T0 °C.
Example when using external reference junction compensation
External reference junction compensation
Recorder
(Hold the contact point of the thermocouple
and copper wire at T0°C)
Copper wire
Thermocouple
<Related Topics>
Setting the reference junction compensation function: Section 7.4
1-4
IM 04P01B01-01E
1.2 Measuring Input Section
1
Noise Elimination from Input Signals
Filter result (output for a step input)
Input
63.2% of the output value
Output response curve
(when using the filter)
2, 5, 10 s (time constant, the time it takes
to reach 63.2% of the output value)
• Moving Average (Dot Model)
The average value of the m most recent values acquired at the scan interval is used
as the measured value of the channel. The number of moving-averaged data points
(m) can be set in the range 2 to 16. The figure below shows an example indicating
the operation of the buffer for the moving average computation when the number of
moving averaged data points is set to 5.
Buffer data for the
nth sampling time
Buffer data for the
n+1th sampling time
Buffer data for the
n+2th sampling time
Most recent data
Most recent data
1
10.0 mV
15.0 mV
10.0 mV
2
5.0 mV
10.0 mV
15.0 mV
3
0.0 mV
5.0 mV
10.0 mV
4
–5.0 mV
0.0 mV
5.0 mV
5
–10.0 mV
–5.0 mV
Deleted
Moving
average
0.0 mV
0.0 mV
Deleted
5.0 mV
8.0 mV
<Related Topics>
Setting the filter: Section 7.12 and 6.2
Setting the moving average: Section 7.11 and 6.3
IM 04P01B01-01E
1-5
Functional Description
Filter and Moving Average
This function used to suppress the effects of noise that is riding on the signal. The pen
model and dot model are equipped with a filter function and a moving average function,
respectively. The function can be set for each measurement channel. However, it does
not operate on channels set to ON/OFF input.
• Filter (Pen Model)
The filter is a low-pass filter. The time constant can be set to 2 s, 5 s, or 10 s.
1.2 Measuring Input Section
Integration Time of the A/D Converter
The recorder uses an A/D converter to convert the sampled analog signal to a digital
signal. By setting the integration time of the A/D converter to match the time period
corresponding to one cycle of the power supply or an integer multiple of one cycle, the
power supply frequency noise can be effectively suppressed.
The integration time of the A/D converter is selected according to the model from the
table below.
Model
Integration Time of the A/D Converter
Pen model
Select 16.7 ms (60 Hz), 20 ms (50 Hz), or Auto
Dot model
Select 16.7 ms (60 Hz), 20 ms (50 Hz), 100 ms or Auto
• If Auto is selected, the recorder detects the power supply frequency and automatically
selects 16.7 ms or 20 ms.
• Because 100 ms is an integer multiple of 16.7 ms and 20 ms, this setting can be used
to suppress the power frequency noise for either frequency, 50 Hz or 60 Hz.
• The scan interval on the dot model is 1 s when the integration time is set to 16.7 ms or
20 ms and 2.5 s when the integration time is set to 100 ms.
<Related Topics>
Setting the A/D integration time: Section 7.2
1-6
IM 04P01B01-01E
1.3
Alarms
1
Alarm Types
Number of Alarm Point Marks
Up to four alarms can be set for each channel.
Alarm Conditions
The eight conditions below are available. The character inside the parentheses is the
symbol used to denote each alarm on the recorder.
• High Limit Alarm (H)
An alarm occurs when the input value exceeds the alarm value.
• Low Limit Alarm (L)
An alarm occurs when the input value falls below the alarm value.
Low limit alarm
High limit alarm
Alarm
value
Alarm occurrence
Alarm release
Measured value
Measured Alarm release
value
Alarm value
Alarm occurrence
• Difference High Limit Alarm (h)*
An alarm occurs when the difference in the input values of two channels is greater
than or equal to the specified value.
• Difference Low Limit Alarm (l)*
An alarm occurs when the difference in the input values of two channels is less than
or equal to the specified value.
*
Can be specified on channels set to delta computation.
• High Limit on Rate-of-Change Alarm (R)
The rate-of-change of the measured values is checked over a certain time (interval).
An alarm occurs if the rate-of-change of the measured value in the rising direction is
greater than or equal to the specified value.
• Low Limit on Rate-of-Change Alarm (r)
The rate-of-change of the measured values is checked over a certain time (interval).
An alarm occurs if the rate-of-change of the measured value in the falling direction is
greater than or equal to the specified value.
High limit on rate-of-change alarm
Low limit on rate-of-change alarm
Change in the
measured value
Measured
T2
value
Measured T1
value
Amount of change
in the setting
| T2−T1 |
T1
Amount of change
in the setting
| T2−T1 |
T2
Change in the
measured value
t1
t2−t1
Interval
t2
Time
t1
t2−t1
t2
Time
Interval
The alarm value of the rate-of-change alarm is set using an absolute value. The
interval is derived using the following equation and set using the number of samples.
Interval = the scan interval × the number of samples
IM 04P01B01-01E
1-7
Functional Description
This function generates an alarm when the measured data meets a certain condition.
The alarm occurrence/release can be recorded on the chart paper. The alarm status can
be displayed on the screen.
Also, alarm output relays can be used to output contact signals when alarms occur (/A1,
/A2, and /A3 options).
1.3 Alarms
• Delay High Limit Alarm (T)
An alarm occurs when the measured value remains above the alarm value for a
specified time period (alarm delay period).
• Delay Low Limit Alarm (t)
An alarm occurs when the measured value remains below the alarm value for a
specified time period (alarm delay period).
Delay High Limit Alarm Example (T is the specified delay)
Measured value
X1
X2
X3
X4
Alarm value
T1
T
Alarm occurrence
Alarm release
• Alarm does not occur at T1, because the time period is shorter than the specified
alarm delay period (T).
• The measured value exceeds the alarm value at time X2, and the alarm occurs at
time X3 at which the specified alarm delay period elapses (the time when the alarm
occurs is the time at X3).
• The measured value falls below the alarm value at time X4, and the alarm is
released.
Note
•
•
The alarm detection operation is reset when a power failure occurs. The operation
restarts after the power recovers.
If the alarm value is changed while a delay alarm is occurring, the alarm is released if the
new alarm value does not meet the alarm condition.
Alarm Hysteresis
Hysteresis can be specified to the values for activating and releasing the alarm. The
hysteresis applies only to high limit alarm (H) and low limit alarm (L). The hysteresis
width can be set in the range of 0.0% (Off) to 1.0% of the recording span in 0.1 steps.
The setting applies to all high limit alarms and low limit alarms. By default, the hysteresis
width is set to 0.5%.
Low limit alarm
High limit alarm
Alarm occurrence
Alarm
value
Measured
value
Alarm release
Hysteresis
(1% or less)
Measured value
Alarm release
Alarm value
Alarm occurrence
<Related Topics>
Setting alarms: Section 5.2
Setting the alarm delay function: Section 7.15 and 6.10
Setting the alarm hysteresis: Section 7.1
1-8
IM 04P01B01-01E
1.3 Alarms
1
Alarm Indication
Non-hold
Hold
Alarm ACK
Alarm occurrence
Alarm
Alarm ACK
or
Alarm release
Blinking
Blinking
ON
ON
Alarm
indication
OFF
or
OFF
OFF
<Related Topics>
Setting the non-hold/hold operation of the alarm indicator: Section 7.1
Alarm Recording
The alarm occurrence/release can be recorded on the chart paper. See section 1.4.
Alarm Output Relay (/A1, /A2, and /A3 Options)
Contact signals can be generated from alarm output relays when alarms occur. The
number of output relays is 2 (/A1), 4 (/A2), or 6 (/A3). The alarm output relays are
denoted as I01 to I06 on the recorder.
The following functions can be assigned to the alarm output relay.
Diagnosis Output
The diagnosis output can be assigned to alarm output relay I01.
The relay is activated when there is an error in the plotter operation on the pen model,
when a burnout is detected, or when there is an error in the A/D converter. Output relay
I01 is normally energized and de-energizes when an error is detected (de-energized
operation and non-hold operation).
NO
C
Normal
NC
NO
C
NC
Malfunction
NO
C
NC
Power-OFF
NO: Normally Opened, C: Common, NC: Normally Closed
Note
If diagnosis output is enabled, I01 becomes a relay dedicated to diagnosis output.
<Related Topics>
Setting the diagnosis output: Section 7.1
IM 04P01B01-01E
1-9
Functional Description
The alarm status can be displayed on the screen. For details on the display, see section
1.5.
Non-Hold/Hold Operation of the Alarm Indication
The alarm indication can be set to operate in the following fashion when the alarm
condition is no longer met.
• Clear the alarm indication (non-hold).
• Hold the alarm indication until the alarm ACK operation is executed (hold).
The default setting is non-hold.
1.3 Alarms
Reflash Alarm
When multiple alarms are assigned to one alarm output relay, this function notifies the
occurrence of subsequent alarms after the relay is activated by the first alarm. When
subsequent alarms occur, the output relay is released temporarily (approximately 500
ms).
The reflash alarm function is set to three output relays (I01, I02, and I03 (I01 and I02 for
the /A1 option)).
By default, the reflash alarm is disabled.
Alarm
Channel 1
Channel 2
Channel 3
Alarm output relay
(Reflash alarm ON)
Approx. 500 ms Approx. 500 ms
Alarm output relay
(Reflash alarm OFF)
Note
•
•
If the reflash alarm is enabled, I01 to I03 are set to reflash alarm operation. In this case, I01
to I03 are set to OR operation and non-hold operation regardless of the settings specified in
“AND/OR Operation of Alarm Output Relays” and “Non-Hold/Hold Operation of Alarm Output
Relays” described below.
If diagnosis output is enabled, I01 is set to diagnosis output.
<Related Topics>
Setting the reflash alarm: Section 7.1
1-10
IM 04P01B01-01E
1.3 Alarms
Channel 01
Alarm
Channel 02
AND
Alarm output relay
OR
The alarm output relays assigned to AND operation are specified as follows: “I01 (first
relay) to Ixx (where xx is the relay number).”
The default setting is “no AND relay.”
Note
•
•
If the reflash alarm is enabled, I01 to I03 are fixed to OR operation. Specifying AND
produces no effect.
If diagnosis output is enabled, I01 is set to diagnosis output. Specifying AND produces no
effect.
<Related Topics>
Setting the AND operation: Section 7.1
Energized/De-energized Operation of Alarm Output Relays
You can select whether the alarm output relay is energized or de-energized when an
alarm occurs. If de-energized is selected, the status of the alarm output relay when an
alarm occurs is the same as the status that results when the power is shut down. The
setting applies to all alarm output relays.
The default setting is energized.
Energize
NO
C
NC
NO
C
NC
De-energize NO
C
NC
NO
C
NC
When power is
shut down
When alarm is
not occurring
NO
NO
C
NC
C
NC
When alarm is
occurring
NO : Normally Opened, C : Common, NC : Normally Closed
Note
If diagnosis output is enabled, I01 is fixed to de-energized operation.
<Related Topics>
Setting the energized/de-energized operation of alarm output relays: Section 7.1
IM 04P01B01-01E
1-11
1
Functional Description
AND/OR Operation of Alarm Output Relays
When multiple alarms are assigned to one alarm output relay, the condition for activating
the alarm output relay can be selected from the following:
• AND: Activated when all assigned alarms are occurring simultaneously.
• OR: Activated when any of the specified alarms is occurring.
1.3 Alarms
Non-Hold/Hold Operation of Alarm Output Relays
The alarm output relay can be set to operate in the following fashion when the alarm
condition is no longer met.
• Turn off the relay output (non-hold).
• Hold the relay output until the alarm ACK operation is executed (hold).
The setting applies to all alarm output relays.
The default setting is non-hold.
Non-hold
Alarm occurrence
Alarm
Hold
Alarm
ACK
Alarm Alarm
ACK ACK
or
Alarm release
Activated
Alarm output
relay
Released
or
Note
•
•
If the reflash alarm is enabled, I01 to I03 are fixed to non-hold operation. Specifying Hold
produces no effect.
If diagnosis output is enabled, I01 is fixed to non-hold operation. Specifying Hold produces
no effect.
<Related Topics>
Setting the non-hold/hold operation of alarm output relays: Section 7.1
Alarm ACK Operation
The alarm acknowledge (alarm ACK) operation releases all alarm indications and relay
outputs (/A1, /A2, and /A3 options) that are activated when the alarm indication or alarm
output relay is set to hold operation. This operation can be executed from the front panel
key.
<Related Topics>
Alarm ACK operation: Section 3.12
1-12
IM 04P01B01-01E
1.4
Recording
1
Functional Description
The recorder is capable of recording the measured values with pens or dots (trend
recording) as well as various other types of information.
Recording Example on the Pen Model
Recording Example on the Dot Model
The recording examples may appear differently from the actual recording as a result of functional
improvements made on the recorder after this manual was written.
<Related Topics>
Starting/Stopping recording: Section 3.5
IM 04P01B01-01E
1-13
1.4 Recording
Trend Recording
The measured values are printed within a width of 100 mm.
Recording Method (Pen Model)
• The measured value is updated every scan interval and continuously recorded.
• The recording colors in order from channel 01 are red, green, blue, and violet.
Recording Method (Dot Model)
• The most recent measured value is recorded with a dot every dot printing interval.
The dot printing interval is in the range of 10 s to 90 s. There are two recording
methods from which you can select. One method automatically adjusts the dot
printing interval according to the chart speed so that the dots do not overlap. The
other method records at the fastest dot printing interval at all times.
• The recording colors in order from channel 01 are purple, red, green, blue, brown, and
black. The recording color of each channel can be changed among these six colors.
• For each channel, trend recording can be enabled or disabled.
<Related Topics>
Setting the dot printing interval: Section 6.1
Changing the recording color: Section 7.5
Enabling/Disabling trend recording for each channel: Section 6.6
Chart Speed
On the pen model, the chart speed can be selected from 82 settings in the range of 5 to
12000 mm/h.
On the dot model, the chart speed can be set in the range of 1 to 1500 mm/h in 1-mm
steps.
The default setting is 20 mm/h.
<Related Topics>
Setting the chart speed: Section 5.4
Zone Recording
A recording zone is assigned to each channel.
Zone 1
Zone 2
Zone 3
<Related Topics>
Setting the zone recording: Section 6.4
1-14
IM 04P01B01-01E
1.4 Recording
1
Compressed
Functional Description
Partial Expanded Recording
This function expands a section of the recording range. By default, partial expanded
recording is disabled.
Expanded
<Related Topics>
Setting the partial expanded recording: Section 7.13 and 6.5
Pen Offset Compensation (Pen Model)
This function compensates for the pen offset (phase difference) along the time axis.
On 2-pen, 3-pen, and 4-pen recorders, there are offsets along the time axis (phase
difference) between the pens. This offset is corrected when pen offset compensation is
used.
Same time
Below is an explanation for the 2-pen model.
The recording of these two pens are offset by an amount of phase P. If pen offset
compensation is enabled, the measured values of pen 1 are stored in the memory, and
recorded when the chart paper is fed by an amount corresponding to P.
Reference pen (pen 2)
Pen 1
Recorder front panel
Chart paper
P
Chart feeding direction
By default, this function is disabled.
<Related Topics>
Setting the pen offset compensation: Section 7.6
IM 04P01B01-01E
1-15
1.4 Recording
Printout
The figure below is used to explain the printout contents. The actual printout and font
are different from those illustrated in the figure. The printout positions are also slightly
different.
Printout Example on the Pen Model
Manual printout
Nov.09.04 15:00
1
223.5mg/cm3
3 H 591.6˚C
2
4
437.2µS/cm
−0.222V
New chart speed printout
50mm/h*14:55
Periodic printout
Time tick cancel mark
Nov.09.04!
13:50*
Offset compensation mark
1
218.7mg/cm3
2
390.6µS/cm
3
H 598.4˚C
4
d −0.222V
0.0
1CH
Alarm
RED
Delta computation
50mm/h_
Scale
500.0
mg/cm3
Time tick
Recording color
Buffer overflow mark
Alarm printout
1H3*10:09
1H3 10:05
Message printout
09:52*START#205 ABCDEF
Recording start printout
08:00*25mm/h
Time tick
The time ticks are marks that indicate the positions of the date/time on the chart paper.
Time tick cancel mark
An exclamation point (!) is printed when the periodic printout time tick was not printed at the
correct position.
<Related Topics>
Setting the channel printout (dot model)/pen color printout (pen model): Section 7.7
Setting the alarm printout, new chart speed printout, and recording start printout:
Section 7.7 and 7.16
Clearing the alarm printout buffer: Section 3.9
Setting the periodic printout: Section 6.6, 7.7, and 7.8
Executing manual print: Section 3.7
Printing messages: Section 6.8 and 3.10
Printing settings: Section 3.8
1-16
IM 04P01B01-01E
1.4 Recording
1
Printout Example on the Dot Model
Nov.09.04 16:00
1
223.5mg/cm3
3 H 591.6˚C
5
−0.665V
Functional Description
Manual printout
2
437.2µS/cm
4
−0.222V
6 L −0.448V
New chart speed printout
_50mm/h*14:55
Periodic printout
Nov.09.04
Time tick
13:50
1
218.7mg/cm3
2
390.6µS/cm
3
H 598.4˚C
4
d −0.222V
5
−0.995V
6
L −0.448V
0.0
1CH
Delta computation
50mm/h_
Alarm
Scale
500.0
mg/cm3
Buffer overflow mark
Time tick
Alarm printout
1H3*10:09
1H3 10:05
Message printout
09:52*START#205 ABCDEF
Recording start printout
_08:00*25mm/h
Channel printout
Time tick
Channel Printout (Dot Model Only)
Prints the channel No. or tag by the trend recording. The channel No. or tag is printed
every approximately 25 mm on the chart paper. The channel printout can be enabled or
disabled. By default, the channel printout is enabled.
IM 04P01B01-01E
1-17
1.4 Recording
Alarm Printout
Alarm information is printed when an alarm occurs or releases.
Time of alarm occurrence/release
Indicates that there are alarms that are not
printed because the alarm printout buffer is full.
Level number
Alarm type
Channel No. or tag
: Alarm occurrence,
: Alarm release
• The print condition can be set to (1) print when alarms occur and release, (2) print
only when alarms occur, or (3) do not print.
• Alarms that occur while an alarm printout is in progress are temporarily saved to the
buffer memory in a printout-wait condition. Alarms are cleared from the buffer
memory when they are printed.
• The number alarms that can be stored in the buffer is 8 and 12 on the pen model and
dot model, respectively. Alarms that occur while the buffer is full are not printed. A
buffer overflow mark is printed when there are alarms that cannot be printed because
the buffer is full.
• The time printout format can be selected.
Type
Hour:Minute
Printout format
10 : 00
Nov. 09
Month:Day:
Hour:Minute
10 : 00
Nov. 09. 2004
Year:Month:Day:
Hour:Minute:Second 10 : 00 : 00
Type
Hour:Minute:Second
Month:Day:
Hour:Minute:Second
Printout format
10 : 00 : 00
Nov. 09
00 : 00 : 00
* The format of year, month, and day varies depending on the setting (see the next page).
Periodic Printout
Measured values and other items are printed at the preset interval.
• Printout Contents (for details, see appendix 1)
• Date/time, time ticks (marks that indicate the positions of the date/time on the chart
paper); measured values, alarm status, scale (leftmost and rightmost values of
span) and recording color (pen model) for each channel; and chart speed can be
printed. When pen offset is being executed on the pen model, the pen offset marks
are printed. On the pen model, if a time tick is not printed at the correct position, a
time tick cancel mark (!) is printed.
• Printout of measured values and alarm status can be enabled or disabled for each
channel.
• Printout of the scale and recording colors (pen model) can be enabled or disabled.
The scale can be printed when the recording zone is greater than or equal to 40 mm.
• The measured values for each channel can be selected from the following data types.
Instantaneous value (measured value at the time of periodic printout)
Average value (average of the measured values over an interval)
Minimum value (minimum value of the measured values over an interval)
Maximum value (maximum value of the measured values over an interval)
Minimum value, maximum value, and average value
Sum value (summed value of the measured values over an interval)
• Interval
The printout interval can be set by specifying the value or set automatically in sync
with the chart speed. Periodic printout is disabled at the following chart speeds.
Pen model: Less than or equal to 9 mm/h or greater than or equal to 1600 mm/h
Dot model: Less than or equal to 9 mm/h or greater than or equal to 101 mm/h
1-18
IM 04P01B01-01E
1.4 Recording
Manual Printout
Measured values and alarm status can be printed manually using the keys. When
manual printout is executed, trend recording stops and restarts when manual printout is
complete.
Message Printout
Preset messages can be printed on the chart paper using the keys. Five messages,
each within 16 characters, can be registered in advance.
• If message printout is executed while another message is being printed, the most
recent message is temporarily stored to the buffer memory in a printout-wait condition.
Messages are cleared from the buffer memory when they are printed.
• The number of messages that can be stored in the buffer is 5. If message printout is
executed when the buffer is full, the message is not printed. A buffer overflow mark is
printed when there are messages that cannot be printed because the buffer is full.
• The time printout format can be selected.
New Chart Speed Printout
• When the chart speed is changed, the time tick (dot model), the date/time of change,
and the new chart speed are printed. An asterisk (*) shows there are messages that
cannot be printed.
• The time printout format can be selected.
Recording Start Printout
When recording is started, the time tick (dot model), the time, and the chart speed can
be printed. An asterisk (*) shows there are messages that cannot be printed.
• The recording start printout can be enabled or disabled. By default, the recording
start printout is disabled.
• The time printout format can be selected.
Printout/Display Format of the Date
The printout/display format of the date can be selected from the list below. The setting
applies to all dates printed and shown on the display.
Type
Printout/Display Format
Y/M/D
2005/08/31
M/D/Y
08/31/2005
D/M/Y
31/08/2005
D.M.Y
31.08.2005
M.D.Y
Aug.31.2005
<Related Topics>
Setting the printout/display format of the date: Section 7.19
IM 04P01B01-01E
1-19
1
Functional Description
• Turning ON/OFF the Periodic Printout
Periodic printout can be turn ON/OFF. By default, periodic printout is enabled with the
interval synchronized to the chart speed. Also, the measured values that are printed
are the instantaneous values.
1.4 Recording
Setting Printout
List or Setup List can be printed. When setting printout is executed, trend recording
stops and restarts when the printout is complete.
List contains settings such as the input range and alarm for each channel.
Setup List contains settings of basic specifications such as the alarm output relay
operation and printout method.
• Printout Example of List on the Pen Model
The printout examples may appear differently from the actual printout as a result of functional
improvements made on the recorder after this manual was written.
1-20
IM 04P01B01-01E
1.4 Recording
1
• Printout Example of List on the Dot Model
Functional Description
The printout examples may appear differently from the actual printout as a result of functional
improvements made on the recorder after this manual was written.
IM 04P01B01-01E
1-21
1.5
Display
Displayed Information
Main Display
The recorder is capable of displaying measured values, alarm information, and so on the
VFD (Vacuum Fluorescent Display). Using normal fonts, 17 characters × 2 lines can be
displayed. Using large fonts, 11 characters can be displayed.
Status display
Main display
RECORD KEY LOCK
MATH
CHART END
ALARM
1 2 3 4 5 6
There are 22 display types available. Also, different displays can be assigned to the top
and bottom sections of the main display. These display types can be registered to
screens 1 to 15 and switched during operation using the keys.
Status Display
A status display is available at the bottom of the main display. The indicators below turn
ON/OFF.
• RECORD
This indicator turns ON when recording is started and turns OFF when stopped.
• ALARM 1 2 3 4 5 6
The numbers correspond to measurement channel numbers. The indicator turns ON
when an alarm occurs and turns OFF when the alarm is released.
If the alarm indication is set to hold operation, the indicator blinks when an alarm
occurs. When alarm ACK operation is performed, the indicator turns ON or OFF.
• MATH
This indicator turns ON when computation on the computation function (/M1 option) is
started and turns OFF when stopped.
• KEYLOCK
This indicator turns ON when key lock is enabled and turns OFF when disabled.
• CHART END
This indicator turns ON when the remaining amount of chart paper falls to
approximately 2 cm on models with the FAIL/chart end function (/F1 option).
Display Types
The following display types are available.
• Displays that show measured values
Measured values can be displayed numerically or using bar graphs. The channels
that are displayed can be switched automatically at constant intervals.
The bar graph displays 0 to 100% using 101 points.
0%
•
•
•
•
•
101 dots
100%
Displays that show alarms
Displays that show the date/time and chart speed
Displays that show the remote control input and alarm output relay statuses
Displays that show the status
System display
Different display types can be assigned to the top and bottom sections of the main display.
1-22
IM 04P01B01-01E
1.5 Display
1
Functional Description
Several display examples are shown below. For details on the display types, see
“Display Function Specifications” in section 12.4.
1 Channel Digital + 1 Bar Graph Display
Channel No.
Alarm status
Measured value
Unit
Bar graph
Scale
Alarm point mark
Flag Display
Flag (the number indicates the channel No.)
Scale
Alarm Status Display
Symbol indicating
an alarm
Computation
channel
Measurement channel
Channel on which an alarm is occurring
Channel on which an alarm is not occurring
Channel on which an alarm is not set
Date/Time and Chart Speed
The date format can be specified.
Date
Aug.31.2005 12000
11:26:37
mm/h
Time
Chart speed
Status Display
Alarm status
Remaining amount of chart paper is less than approx. 2 cm (/F1 option)
<Related Topics>
Assigning display types to screens: Chapter 8
Switching displays: Section 3.6
Setting the bar graph display mode: Section 7.9
Changing the date format: Section 7.19
Setting the display brightness: Section 6.11
Displaying measured values: Section 12.4.
IM 04P01B01-01E
1-23
1.6
Computation Function (/M1 Option)
Computation Function
Computing equations can be set up using measured values and other computed values
as variables in computation-dedicated channels, and the computation can be executed.
The computed result can be recorded. Computation is performed every scan interval.
Channels Dedicated to Computations
The computation-dedicated channels below can be used.
Model
Number of Channels
Channel Names
Pen model
8
0A, 0B, 0C, 0D, 0E, 0F, 0G, and 0J
Dot model
12
0A, 0B, 0C, 0D, 0E, 0F, 0G, 0J, 0K, 0M, 0N, and 0P
Types of Computations
In the table below, y represents the computed result. X and n represent the measured
value, a constant, etc.
Type
Syntax in the Equation
Description
Four arithmetic operation +, –, *, /
Determines addition, subtraction, multiplication, and division.
Power
**
Determines the power. y = Xn
Square root
SQR()
Determines the square root.
Absolute value
ABS()
Determines the absolute value.
Common logarithm
LOG()
Determines the common logarithm. y=log10x
Exponent
EXP()
Determines the exponent. y = ex
Relational computation
.LT., .LE., .GT., .GE.,
.EQ., .NE.
Determines <, ≤, >, or ≥ of two elements and outputs “0” or “1.”
Logical computation
AND, OR, XOR, NOT
Determines the AND (logical product), OR (logical sum),
XOR (exclusive logical sum) of two elements,
NOT (negation) of an element and outputs “0” or “1.”
TLOG computation
TLOG.SUM(), TLOG.
MAX(), TLOG.MIN(),
TLOG.AVE(),
TLOG.P-P(),
Determines the sum (SUM), the maximum (MAX),
the minimum (MIN), the average (AVE),
and the maximum – minimum (P-P) at specified time
intervals. There are three timers for setting the interval.
Data That Can Be Used in Equations
In TLOG computation, the measured values of measurement channels or computed
values of computation channels can only be used. For all other computations, all of the
data types below can be used.
• Measured Values on Measurement Channels
Specified by the channel No. For channels that have scaling enabled, the scaled
values are used in the computation.
• Computed Values on Computation Channels
Specified by the channel No.
• Constants (K01 to K30)
The values assigned to K01 to K30 can be used as constants.
Range of values (maximum significant digits is 5):
–9.9999E+29 to –1.0000E–30, 0, 1.0000E–30 to 9.9999E+29
1-24
IM 04P01B01-01E
1.6 Computation Function (/M1 Option)
<Related Topics>
Setting the computing equation, constant, and unit: Section 9.2 to 9.4
Setting the TLOG timer: Section 9.6 and 9.12
Handing of the Unit in Computations
In computations, measured/computed values are handled as values without units. For
example, if the measured value of channel 01 is 20 mV and the measured value of
channel 02 is 20 V, the computed result of 01 + 02 is 40.
Recording Computation Channels
Trend Recording
Trend recording is possible by setting the recording span for each computation channel.
The recording span of computation channels can be set in the range of –9999999 to
99999999 excluding the decimal.
Zone recording and partial expanded recording are also possible.
Assignment of Computation Channels to the Pens (Pen Model)
Trend recording is possible by assigning computation channels to arbitrary pens. Only
display or printout is possible for measurement channels or computation channels that
are not assigned to a pen. Assignment of pens is possible only on models with the
computation function.
Pen
1
2
3
4
Channel 01 02 03 04 0A 0B 0C 0D 0E 0F 0G 0J
Measurement channels
Computation channels
<Related Topics>
Changing the channel assignment of recording pens: Section 9.14
Changing the Channel Recording Color (Dot Model)
The recording colors of computation channels can be changed. The default recording
colors of channels are indicated below.
IM 04P01B01-01E
Recording color
Purple
Red
Green
Blue
Brown
Black
Computation channels
0A
0G
0B
0J
0C
0K
0D
0M
0E
0N
0F
0P
1-25
1
Functional Description
• Communication Input Data (Pen model: C01 to C08, dot model: C01 to C12)
This data type is set using the communication interface (/C7 or /C3 option).
Range of values (maximum significant digits is 5):
–9.9999E+29 to –1.0000E–30, 0, 1.0000E–30 to 9.9999E+29
For the procedure of setting the communication input data, see the µR10000/
µR20000 Communication Interface User’s Manual (IM04P01B01-17E).
• Status of the Remote Control Input Terminal (D01 to D05, /R1 option)
The status of the remote control input signal (1 or 0) can be used in the computation.
Use D01 to D05 (the number following the letter D is the remote control input terminal
number) to specify the status in the equations.
1.6 Computation Function (/M1 Option)
• The Number of Channels Being Trend Recorded and the Fastest Recording
Interval
The fastest recording interval varies depending on the number of channels being
trend recorded. Only display or printout is possible on measurement channels or
computation channels that are not trend recorded.
Number of Channels Being Trend Recorded
Fastest Recording Interval
6 channels or less
7 to 12 channels
13 to 18 channels
10 s
15 s
20 s
<Related Topics>
Changing the channel recording color: Section 9.13
Enabling/Disabling trend recording for each channel: Section 9.9
Printout
Various printouts can be executed in the same fashion as measurement channels.
The computed values of computation channels can be printed at constant intervals by
using the periodic printout timer or the two TLOG timers. The TLOG timers can be used
only on the computation channels. For the printout contents when printing using the
TLOG timer, see appendix 1.
<Related Topics>
Setting the periodic printout: Section 7.8, 9.6, and 9.15
Setting the printout using the TLOG timer: Section 9.6 and 9.12
Alarms
Up to 4 alarms can be assigned to each computation channel.
• The alarm types are high limit alarm (H), low limit alarm (L), delay high limit alarm (T),
and delay low limit alarm (t).
• Hysteresis can be specified to the values for activating and releasing the alarm. The
hysteresis applies only to high limit alarm (H) and low limit alarm (L). The hysteresis
width can be set in the range of 0.0% (Off) to 1.0% of the recording span in 0.1 steps.
The setting applies to all high limit alarms and low limit alarms. By default, the
hysteresis width is set to 0% (Off).
<Related Topics>
Setting alarms: Section 9.5
Setting the alarm delay function: Section 7.15 and 9.11
Setting the alarm hysteresis: Section 7.1
Display
Computation channels can be displayed on the VFD as with the measurement channels.
<Related Topics>
Setting the bar graph display mode: Section 9.16
Starting/Stopping Computation
Computation can be started or stopped using keys.
<Related Topics>
Starting/Stopping computation: Section 9.1
1-26
IM 04P01B01-01E
1.7
FAIL Output
When a failure occurs in the CPU of the recorder, a relay contact signal (1 signal) is
output. The relay is energized when the CPU is normal and de-energizes when a CPU
failure occurs. Therefore, relay output is carried out also when the power is turned OFF
(including a power failure). This relay behavior cannot be changed.
• FAIL Relay Output
NO
C
NC
NO
Normal
C
NC
Failure
NO
C
NC
Power-OFF
NO: Normally Opened, C: Common, NC: Normally Closed
Chart End Output
When the remaining amount of chart paper falls to approximately 2 cm, recording is
stopped, and the relay contact signal (1 relay) is output. The relay is energized in this
case. This relay behavior cannot be changed.
NO
C
Normal
NC
NO
C
NC
Chart end
NO
C
NC
Power-OFF
NO: Normally Opened, C: Common, NC: Normally Closed
IM 04P01B01-01E
1-27
1
Functional Description
FAIL/Chart End Detection and Output Function
(/F1 Option)
1.8
Remote Control Function (/R1 Option)
Specified operations can be carried out by applying remote signals (contact or open
collector signals) to the remote control input terminals.
There are five remote control input terminals. An action can be assigned to each
terminal.
µR10000 Recorder
Contact
Open collector
Assignable Functions
• Recording start/stop
• Remote input signal: Rising edge signifies start; falling edge signifies stop
• Starts/stops recording.
• Applying a rising edge signal when recording is already in progress produces no
effect. Applying a falling edge signal when recording is stopped produces no
effect.
• Chart Speed Switch
• Remote input signal: Level
• The chart paper is fed at the secondary chart speed while a level signal is applied
to the terminal. The secondary chart speed is set in advance.
• Message 1 Printout to Message 5 Printout
• Remote input signal: Trigger
• Prints message 1 to 5 on the chart paper.
• Manual Printout
• Remote input signal: Trigger
• Executes a manual printout.
• Alarm ACK
• Remote input signal: Trigger
• Alarm ACK is executed on all alarms that are occurring. This is valid only when the
alarm indication or output relay operation is set to hold.
• Internal Clock Adjustment
• Remote input signal: Trigger
• The internal clock of the recorder is adjusted to the nearest hour depending on the
time when the remote signal is applied.
1-28
Time When Signal Is Input
Adjustment
00 min 00 s to 01 min 59 s
Truncates the minutes and seconds.
Example: 10 hours 01 min 50 s becomes 10 hours 00 min 00 s.
02 min 00 s to 57 min 59 s
The time is not changed.
58 min 00 s to 59 min 59 s
Rounds up the minutes and seconds.
Example: 10 hours 59 min 50 s becomes 11 hours 00 min 00 s.
IM 04P01B01-01E
1.8 Remote Control Function (/R1 Option)
• Computation Reset
• Remote input signal: Trigger
• The computed results of computation channels are reset. This is valid only on
models with the computation function (/M1 option) and while the computation is
stopped. For all other cases, applying the remote signal produces no effect.
Remote Signal (Edge, Trigger, and Level)
The above actions are carried out on the rising or falling edge of the remote signal
(edge), the ON signal lasting at least 250 ms (trigger), or the ON/OFF signal (level).
Rising/Falling edge
Rising
Trigger
Level
Falling
250 ms or more
250 ms or more
Operates at the
secondary chart speed
For contact inputs, the remote signal rises when the contact switches from open to
closed and falls when the contact switches from closed to open. For open collector
signals, the remote signal rises when the collector signal (voltage level of the remote
control terminal) goes from high to low and falls when the collector signal goes low to
high.
<Related Topics>
Assigning functions to the remote control input terminals: Section 7.18
Setting the secondary chart speed: Section 6.9
IM 04P01B01-01E
1-29
1
Functional Description
• Computation Start/Stop
• Remote input signal: Rising edge signifies start; falling edge signifies stop
• Starts/stops the computation. This is valid only on models with the computation
function (/M1 option).
• If the computation is started, applying a rising edge signal produces no effect. If
the computation is stopped, applying a falling edge signal produces no effect.
1.9
Other Functions
Key Lock
Key lock is a function that prohibits key operations. When key lock is enabled, pressing
keys produces no effect. To release the key lock, a password is entered.
Key Lock Items
Each of the following keys can be included or excluded from the key lock function.
CHARACTER
RCD
MENU
DISP
FUNC
ESC/?
SHIFT
1 MENU
FEED
CH UP
Keys that can be locked
In the case of the FUNC key, each function of the FUNC key can be included or
excluded from the key lock function.
FUNC key functions: Alarm ACK execution, computation start/stop, manual printout and
setting printout, message printout, printout buffer clear, resetting of
the periodic printout report data (the average, etc.), and pen
replacement (pen model), ribbon cassette replacement (dot model)
<Related Topics>
Setting the key lock function: Section 7.10
Using the key lock function: Section 3.13
Language
The display and recording language can be set to English or Japanese.
<Related Topics>
Changing the language: Section 7.14
DST
If the recorder is used in a region that has DST, time can be switched automatically
between DST and standard time by setting the date/time when switching from the
standard time to DST and the date/time when switching back from DST to standard time.
When switching from standard time to DST, the clock is set ahead by 1 hour. When
switching back from DST to standard time, the clock is set back by 1 hour.
<Related Topics>
Using the DST: Section 6.13
Temperature Unit
The temperature unit can be set to Celsius or Fahrenheit. The setting applies to all
channels.
<Related Topics>
Changing the temperature unit: Section 7.20
1-30
IM 04P01B01-01E
Chapter 2 Before Using the Recorder
2.1
Handling Precautions
This section describes the precautions to be taken when using the recorder. Read this
section before use.
• Use care when cleaning the recorder, especially any plastic parts. When cleaning,
wipe using a dry soft cloth. Do not use chemicals such as benzene or thinner, since
these may cause discoloring and deformation.
• Keep electrically charged objects away from the recorder as this may cause
malfunction.
• Do not apply volatile chemicals to the door glass, display, panel keys, etc. Do not
allow rubber and vinyl products to remain in contact with the recorder for long periods
of time. This may damage the recorder.
• When not in use, make sure to turn OFF the power switch.
• If there are any symptoms of trouble such as strange odors or smoke coming from the
recorder, immediately turn OFF the power switch and the power supply source. Then,
contact your nearest YOKOGAWA dealer.
IM 04P01B01-01E
2-1
Before Using the Recorder
Handling Precautions
2
2.2
Installation
Installation Location
Install the recorder indoors in a location that meets the following conditions. See also
the normal operating conditions described in section 12.6, “General Specifications.”
• Instrument Panel
The recorder is designed for panel mounting.
• Well-Ventilated Location
To prevent overheating, install the recorder in a well-ventilated location.
For the panel cut dimensions when arranging multiple recorders, see page 2-4.
Follow the panel cut dimensions providing adequate space between instruments
when other instruments are arranged on the panel.
• Minimum Mechanical Vibrations
Choose an installation location with the minimum mechanical vibration.
Installing the recorder in a location with large mechanical vibration not only causes
adverse effects on the mechanism but also may hinder normal recording.
• Horizontal
Install the recorder horizontally (However, the recorder can be inclined up to 30
degrees backwards for panel mounting).
Note
•
•
Condensation may occur if the recorder is moved to another place where both the ambient
temperature and humidity are higher, or if the temperature changes rapidly. In addition,
measurement errors will result when using thermocouples. In this case, let the recorder
adjust to the new environment for at least one hour before using it.
The chart paper may be adversely affected by a rapid change in the ambient temperature and
humidity.
Do not install the instrument in the following places.
• Outdoors
• In direct sunlight or near heat sources
Install the recorder in a place with small temperature fluctuations near room
temperature (23°C). Placing the recorder in direct sunlight or near heat appliances
can cause adverse effects on the internal circuitry.
• Where an excessive amount of soot, steam, moisture, dust, or corrosive gases
are present
Soot, steam, moisture, dust, and corrosive gases will adversely affect the recorder.
Avoid such locations.
• Near Strong Magnetic Field Sources
Do not bring magnets or instruments that produce electromagnetic fields close to the
recorder. Operating the recorder in strong magnetic fields can cause errors in the
measurements.
2-2
IM 04P01B01-01E
2.2 Installation
Installation Procedure
Fix in place
Screw temporarily
Panel
Panel Case
Mounting bracket
Mounting bracket
Attachment screw
In contact
with each other
Front
Torque driver
(flat blade)
Attachment screw
(The figure shows the case when the mounting
brackets are used on the top and bottom of the case.)
CAUTION
Tightening the screws too much can deform the case or damage the bracket.
IM 04P01B01-01E
2-3
2
Before Using the Recorder
The recorder should be mounted on a steel panel of thickness 2 mm to 26 mm.
1.
Insert the recorder from the front side of the panel.
2.
As shown in the figure below, mount the recorder to the panel using the
mounting brackets that come with the package.
• Use two brackets to support the top and bottom or the left and right sides of
the case. (Remove the seal that is covering the holes for the mounting
brackets beforehand.)
• The proper torque for tightening the mounting screws is 0.7 to 0.9 Nm.
• Mount the recorder to the panel according to the procedure below.
• First, attach the two mounting brackets and temporarily fasten the
attachment screws.
• Next, fix the recorder in place by tightening the attachment screws with the
appropriate torque. When the recorder is approximately perpendicular to
the panel as you fasten the screws, press the mounting bracket against the
case so that they are in contact with each other.
2.2 Installation
External Dimensions of the Recorder
Unit: mm (approx. inch)
Unless otherwise specified, tolerance is ±3%
(however, tolerance is ±0.3 mm when below 10 mm).
220 (8.66)
(1.08) 27.5
151.5 (5.96)
178 (7.01)
Mounting panel
thickness
144 (5.67)
2 to 26
9.4 (Dimensions before attaching
(0.37) the mounting bracket)
144
(5.67)
136.5
(5.37)
151.5
(5.96)
+2
0
7.5 (Dimensions after attaching
(0.30) the mounting bracket)
Panel Cutout
Single-Unit Mounting
Side-by-Side Mounting
(vertically, max. 3 units)
Side-by-Side Mounting (horizontally)
+2
137 0
(5.39)
+2
137 0
(5.39)
+2
137 0
(5.39)
175 min.
(6.89)
+2
137 0
(5.39)
L
L
Units
2-4
+2
0
L (mm)
2
282
3
426
4
570
5
714
6
858
7
1002
8
1146
9
1290
10
1434
n
+2
0
(144xn)-6
175 min.
(6.89)
Attach the mounting brackets to the top and
bottom when mounting the recorders side-by-side
horizontally or right and left when mounting the
recorders side-by-side vertically.
Unit: mm (approx. inch)
Unless otherwise specified, tolerance is ±3%
(however, tolerance is ±0.3 mm when below 10 mm).
IM 04P01B01-01E
2.3
Input Signal Wiring
General Precautions to Be Taken While Wiring
2
Before Using the Recorder
WARNING
To prevent the possibility of electric shock when wiring, make sure that the
power supply source is turned OFF.
CAUTION
• If a strong tension is applied to the cable wired to the recorder, the terminals of
the recorder and/or the cable can be damaged. In order to prevent tension from
being applied directly on the terminals, fasten all wiring cables to the rear of the
mounting panel.
• Do not apply a voltage exceeding the following value to the input terminals as
this may damage the recorder.
• Maximum input voltage
Voltage range less than or equal to 200 mVDC, TC, RTD, and DI: ±10 VDC
Ranges other than those listed above: ±60 VDC
• Maximum common-mode voltage
±60 VDC (under measurement category II conditions)
• The recorder is an INSTALLATION CATEGORY II product.
Precautions to Be Taken While Wiring
Take the following precautions when wring the input signal cables.
It is recommended that crimp-on lug with insulation sleeves (designed for 4-mm
screws) be used when connecting the input/output signal wires to the terminals.
However, this does not apply clamped terminals (/H2).
Crimp-on lug with insulation
sleeves (for 4 mm screws)
For clamped terminals (/H2), the following wire is recommended.
• Conductive cross-sectional area for single wire: 0.14 mm2 to 1.5 mm2, stranded wire:
0.14 mm2 to 1.0 mm2
• Length of the stripped section of the wire: Approx. 5 mm
Take measures to prevent noise from entering the measurement circuit.
• Move the measurement circuit away from the power cable (power circuit) and ground
circuit.
• It is desirable that the object being measured does not generate noise. However, if
this is unavoidable, isolate the measurement circuit from the object. Also, ground the
object being measured.
• Shielded wires should be used to minimize noise caused by electrostatic induction.
Connect the shield to the ground terminal of the recorder as necessary (make sure
you are not grounding at two points).
• To minimize noise caused by electromagnetic induction, twist the measurement circuit
wires at short, equal intervals.
• Make sure to earth ground the protective ground terminal through minimum resistance
(less than 100 Ω).
IM 04P01B01-01E
2-5
2.3 Input Signal Wiring
When using internal reference junction compensation on the thermocouple input,
take measures to stabilize the temperature at the input terminal.
• Always use the terminal cover.
• Do not use thick wires which may cause large heat dissipation (cross sectional area of
0.5 mm2 or less recommended).
• Make sure that the ambient temperature remains reasonably stable. Large
temperature fluctuations can occur if a nearby fan turns ON or OFF.
Connecting the input wires in parallel with other devices can cause signal
degradation, affecting all connected devices.
If you need to make a parallel connection, then
• Turn the burnout detection function OFF.
• Ground the instruments to the same point.
• Do not turn ON or OFF another instrument during operation. This can have adverse
effects on the other instruments.
• RTDs cannot be wired in parallel.
Arrangement of the Measuring Input Terminals
A terminal cover is screwed in place on the measuring input terminal block on the rear
panel. A label indicating the terminal arrangement is affixed to the cover.
Measuring input terminal block
Terminal cover
attachment screws
Pen Model
b
–/B
+/A
Channel 1
Channel 2
Channel 3
Channel 4
Screw input terminal
2-6
+/A
Channel 2 –/B
b
+/A
–/B
b
Channel 1
Channel 3
Channel 4
Clamped input terminal
(/H2 option)
IM 04P01B01-01E
2.3 Input Signal Wiring
Dot Model
b
–/B
+/A
Channel 1
Channel 4
Channel 5
Channel 6
b
2
Channel 1
Channel 4
Channel 3
Channel 6
Channel 5
Screw input terminal
1.
2.
+/A
–/B
b
Before Using the Recorder
Channel 2
Channel 3
+A
Channel 2 –/B
Clamped input terminal
(/H2 option)
Turn OFF the recorder and remove the terminal cover.
Connect the signal wires to the terminals.
Screw input terminal
Clamped input terminal
Signal wire
Crimp-on lug with
insulation sleeves
Signal wire
Flat-blade
screwdriver
Philips
screwdriver
Note
Input signal wires of diameter less than or equal to 0.3 mm may not be secured firmly for
clamped terminals (/H2). Fold over the conducting section of the wire, for example, to make
sure that the wire is securely connected to the clamped terminal.
3.
IM 04P01B01-01E
Replace the terminal cover and fasten it with screws.
The proper torque for tightening the screws is 0.6 N-m.
2-7
2.3 Input Signal Wiring
Measuring Input Wiring
Thermocouple input
b
Resistance temperature detector input
–/B +/A
b
–/B +/A
Leadwire resistance:
10 Ω max./wire. The resistance
of the three wires should be
equal.
Extension leadwire
b
B
A
DC voltage input, 1-5V input,
and ON/OFF input
b –/B +/A
DC current input
b
–/B +/A
Shunt resistor
+
–
DC voltage
1-5V
ON/OFF
Example: For a 4 to 20 mA
input, a shunt resistor of
250 Ω ± 0.1% can be used to
convert to 1-5V input.
+
–
DC current
Note
RTD input terminals A and B on the dot model are isolated on each channel. Terminal b is
shorted internally across all channels. However, for 3 legs isolated RTDs (/N2 option), input b
is also isolated for each channel.
2-8
IM 04P01B01-01E
2.4
Optional Terminal Wiring
General Precautions to be Taken While Wiring the Input/Output Signal Wires
2
Before Using the Recorder
WARNING
• To prevent electric shock while wiring, ensure that the power supply source is
turned OFF.
• If a voltage of more than 30 VAC or 60 VDC is to be applied to the output
terminals, use ring-tongue crimp-on lugs with insulation sleeves on all terminals
to prevent the wires from slipping out when the screws become loose.
Furthermore, use double-insulated wires (dielectric strength of 2300 VAC or
more) for the signal wires on which a voltage of more than 30 VAC or 60 VDC is
to be applied. For all other wires, use basic insulated wires (dielectric strength
of 1390 VAC). To prevent electric shock, attach the terminal cover after wiring
and make sure not to touch the terminals.
Crimp-on lug with insulation sleeves
(for 4 mm screws)
CAUTION
• To prevent fire, use signal wires having a temperature rating of 70°C or more.
• If a strong tension is applied to the cable wired to the recorder, the terminals of
the recorder and/or the cable can be damaged. In order to prevent tension from
being applied directly on the terminals, fasten all wiring cables to the rear of the
mounting panel.
Arrangement of the Optional Terminals
As shown in the figure below, the optional terminal block is located on the rear panel.
The optional terminal block is provided on the recorder when an option that requires
input/output is installed such as the alarm output relay (/A1, /A2, or /A3 option), FAIL/
chart end output (/F1 option), and remote control function (/R1 option). A terminal cover
is screwed in place on the optional terminal block. A label indicating the terminal
arrangement is affixed to the terminal block.
Optional terminal block
Terminal cover
attachment screws
IM 04P01B01-01E
2-9
2.4 Optional Terminal Wiring
/A1
/A1/F1
/A1/R1
/A1/F1/R1
NO
C
NC I01
NO
C
NC I01
NO
C
NC I01
NO
C
NC I01
NO
C
NC I02
NO
C
NC I02
NO
C
NC I02
NO
C
NC I02
NO
C
NC CE
NO
C
NC CE
NO
C
NC FAIL
NO
C
1
2
NC FAIL
3
4
5
/A2
(/F1)
/A2/F1
1
2
3
4
5
C
(/R1)
/A2/R1
C
(/F1)
(/R1)
/A2/F1/R1
NO
C
NC I01
NO
C
NC I01
NO
C
NC I01
NO
C
NC I01
NO
C
NC I02
NO
C
NC I02
NO
C
NC I02
NO
C
NC I02
NO
C
NC I03
NO
C
NC I03
NO
C
NC I03
NO
C
NC I03
NO
C
NC I04
NO
C
NC I04
NO
C
NC I04
NO
C
NC I04
NO
C
NC CE
NO
C
NC CE
NO
C
NC FAIL
NO
C
1
2
NC FAIL
3
4
5
/A3
(/F1)
/A3/R1
1
2
3
4
5
C
/F1
C
NC I01
NO
C
NC I01
NO
C
NC I02
NO
C
NC I02
NO
C
NC I03
NO
C
NC I03
NO
C
NC I04
NO
C
NC I04
NO
C
NC I05
NO
C
NC I05
NO
C
NO
C
NC I06
NO
C
NO
C
1
2
NC I06
3
4
5
C
C
(/R1)
/R1
NO
/F1/R1
(/R1)
(/F1)
NC CE
(/F1)
NC FAIL
(/R1)
1
2
3
4
5
C
(/R1)
CE: Chart end
NC CE
(/F1)
NC FAIL
NO
C
NO
C
1
2
3
4
5
C
(/R1)
Alarm Output Relay Terminals and FAIL/Chart End Output Relay Terminals
NC (Normally Closed), C (Common), NO (Normally Opened)
Remote Control Input Terminals
1 to 5 (remote control input terminals), C (Common)
Alarm output terminals correspond to I01 to I06 in the alarm output relay settings.
Remote control input terminals 1 to 5 correspond to numbers 1 to 5 in the remote control
input settings.
2-10
IM 04P01B01-01E
2.4 Optional Terminal Wiring
Wiring Procedure
1.
2.
Turn OFF the recorder and remove the terminal cover.
Connect the input signal wires to the terminals.
2
Before Using the Recorder
Philips
screwdriver
Crimp-on lug with
insulation sleeves
3.
Replace the terminal cover and fasten it with screws.
The proper torque for tightening the screws is 0.6 N-m.
Alarm output
FAIL/chart end output
Relay contact output
NO
Remote control input
Relay contact input
(Voltage-free contact)
1
2
3
Transistor input
(Open collector)
1
2
3
C NC
4
5
C
4
5
C
Note
To reduce noise, use a shielded cable for the wiring of the remote control input terminals.
Connect the shield to the ground terminal of the recorder.
• Relay Contact Output Specifications
Output type:
Relay
Contact rating:
250 VAC (50/60 Hz)/3 A, 250 VDC/0.1 A (resistive load)
Dielectric strength:
1500 VAC at 50/60 Hz for one minute (between output
terminals and the ground terminal)
• Relay Contact Input/Transistor Input Specifications
Input signal:
• Voltage-free contact: Contact closed at 200 Ω or less and
contact open at 100 kΩ or greater
• Open collector: 0.5 V or less (30 mADC) when turned ON,
leakage current of 0.25 mA or less when turned OFF
Input format:
Photocoupler isolation (shared common)
Dielectric strength:
500 VDC for one minute between input terminals and the
ground terminal
IM 04P01B01-01E
2-11
2.5
Power Supply Wiring
Precautions to Be Taken While Wiring the Power Supply
Make sure to follow the warnings below when wiring the power supply. To prevent
electric shock and damage to the recorder, observe the following warnings.
WARNING
• To prevent electric shock when wiring, ensure the main power supply is turned
OFF.
• To prevent the possibility of fire, use 600 V PVC insulated wire (JISC3307) or an
equivalent wire for power wiring.
• Make sure to earth ground the protective earth terminal through a grounding
resistance less than 100 Ω before turning ON the power.
• Use crimp-on lugs (designed for 4 mm screws) for power and ground wiring
termination.
• To prevent electric shock, make sure to close the transparent cover for the
power supply wires.
• Make sure to provide a power switch (double-pole type) on the power supply
line in order to separate the recorder from the main power supply. Put an
indication on this switch as the breaker on the power supply line for the recorder
and indications of ON and OFF.
Switch specifications
Rated power current: 1 A or more
Rated rush current: 60 A or more
Use a switch complied with IEC 60947-1, 3.
• Connect a fuse between 2 A and 15 A in the power supply line. Use a fuse
approved by CSA (for the use in North America) or VDE (for the use in Europe).
• Do not add a switch or fuse to the ground line.
Use a power supply that meets the following conditions:
Item
Power Supply Specifications
Rated supply voltage
100 to 240 VAC
Allowable power supply voltage range
90 to 132/180 to 264 VAC
Rated supply voltage frequency
50/60 Hz
Allowable power supply frequency range
50/60 Hz±2%
Maximum power consumption
Pen model: 40 VA, dot model: 40 VA
Note
Do not use a supply voltage in the range 132 to 180 VAC, as this may have adverse effects
on the measurement accuracy.
2-12
IM 04P01B01-01E
2.5 Power Supply Wiring
Wiring Procedure
The power supply terminals and a protective ground terminal are located on the rear
panel.
1.
Turn OFF the recorder and open the transparent power terminal cover.
2
Before Using the Recorder
Power terminal block
Screw for fixing the power
terminal cover in place
Power terminal cover
Open
2.
Wire the power cord and the protective ground cord to the power supply
terminals.
Use ring-tongue crimp-on lugs (designed for 4 mm screws).
L
N
Philips screwdriver
3.
IM 04P01B01-01E
Crimp-on lug with insulation sleeves
Power cord
Protective ground cord
Replace the power terminal cover, and fasten it with screws.
The proper torque for tightening the screws is 0.6 N-m.
2-13
2.6
Turning ON/OFF the Power Switch
CAUTION
Check the following points before turning ON the power switch.
• The power cord/wires are connected correctly to the recorder.
• The recorder is connected to the correct power supply (see section 2.5).
The power switch is located inside the door at the lower right.
The power switch is a push button. Press once to turn it ON and press again to turn it
OFF.
When the power switch is turned ON, a self-diagnosis program runs for a few seconds,
and the recorder is ready for operation.
Note
•
•
•
•
If the input wires are connected in parallel with other devices, do not turn ON/OFF the power
switch of the recorder or another device during operation. This can have adverse effects on
the measured values.
If nothing is displayed when the power switch is turned ON, turn OFF the power switch and
check the points listed above one more time. After checking the points, turn ON the power
switch again. If the unit still does not work, it is probably malfunctioning. Contact your
nearest YOKOGAWA dealer for repairs.
If an error message is displayed on the screen, take measures according to the description in
chapter 10, “Troubleshooting.”
Turn ON the power switch, let the recorder warm up for at least 30 minutes, and then start the
measurements.
Checking the Date/Time
The date/time is shown on the display when the DISP key is pressed several times. To
correct the date/time, see the reference indicated below.
<Related Topics>
Setting the date/time: See section 5.5
2-14
IM 04P01B01-01E
Chapter 3 Names of Parts and Run Operations
3.1
Names of Parts
Front
Name plate
The model name is written
on the name plate.
3
Door
Tag plate
Used to write
channel names.
Chart cassette
Holds the chart paper.
Pen model
Mounting hole
There is one hole on each of the
top, bottom, left, and right panels.
The hole is covered with a seal.
Power switch
Turns ON/OFF the power each
time the switch is pressed.
Dot model
Recording pen
Records the measured value.
Plotter pen
Prints various types
of information.
Ink ribbon
Six-color ink.
Printer carriage
Records measured values
and prints various types of
information.
Display and key panel (see the next page)
There is an internal light on the bottom section
of the display and key panel. It lights up the
recording area of the chart paper.
IM 04P01B01-01E
3-1
Names of Parts and Run Operations
Display and key panel
Hold the tab at the lower
left and pull to open.
3.1 Names of Parts
Display and Key Panel
Status display
Displays the following information.
RECORD
Illuminates while recording measured values.
KEY LOCK
Illuminates when key lock is enabled.
MATH
Illuminates when computation on the computation function (/M1 option)
is in progress.
CHART END
Illuminates when the chart paper is out (/F1 option).
ALARM 1 to 6 Illuminates when an alarm is occurring on channels 1 to 6.
Main display
Displays the measured values. Also, displays the setup screen when
setting functions.
RECORD KEY LOCK
CHARACTER
RCD
MENU
MATH
DISP
CHART END
FUNC
1 2 3 4 5 6
ALARM
ESC/?
SHIFT
1 MENU
FEED
CH UP
Seven keys are available.
For all keys except RCD, functions marked above the keys are enabled when
setting functions or when the FUNC key or the DISP MENU key is pressed.
<While setting functions, when the FUNC key/DISP MENU key is pressed>
CHARACTER Key: Changes the character type when entering a character. Press
this key while holding down the SHIFT key to switch the character type in reverse
order.
UP/DOWN Key: Switches the setup item or the value.
Press this key while holding down the SHIFT key to switch the setup item
or the value in reverse order.
LEFT/RIGHT Key: Moves the cursor to the right when entering a value
or character. Press this key while holding down the SHIFT key to move
the cursor to the left..
ESC Key: Cancels the operation.
When pressed with the SHIFT key, the display of the comment
on the setting turns ON/OFF.
SHIFT Key: Used with the
CHARACTER key.
key,
key, or the
ENTER Key: Confirms the setup item or value.
CHARACTER
RCD
MENU
DISP
FUNC
ESC/?
SHIFT
1 MENU
FEED
CH UP
<During normal operation>
CH UP key
Switches the displayed channel.
(when manual switching is specified)
FEED key
Feeds the chart paper.
DISP MENU key
Hold this key down for 3 seconds to switch to the data display
setup screen. Hold this key down for 3 seconds also to exit
from the data display setup screen.
FUNC key
Used when executing manual printout, message printout, etc.
DISP key
Switches the screen in the main display.
MENU key
Hold this key down for 3 seconds to enter Setting mode. Hold this key down for
3 seconds also to exit from Setting mode.
RCD key
Starts/stops recording.
Note
This manual denotes the operation of pressing a key while holding down the SHIFT key as
SHIFT + the other key (for example: SHIFT +
key).
3-2
IM 04P01B01-01E
3.1 Names of Parts
Rear Panel
Heatsink
Dissipates the internal heat.
Power terminal block
The power terminal and protective ground terminal.
Optional terminal block
This is where terminals or ports used by options such
as alarm output relays and communication interface
are installed.
3
Names of Parts and Run Operations
Measuring input terminal block
Measuring input terminals
Ethernet port (/C7 option)
IM 04P01B01-01E
3-3
3.2
Installing or Replacing the Chart Paper
CAUTION
• Do not install or remove the chart cassette with the chart paper guide open.
This may damage the stopper.
• Continuing to record or print without the chart paper on the dot model can cause
damage to the chart cassette platen (the cylindrical section that holds the paper
during the recording operation). Be sure to replace the chart paper ahead of
time.
Loading the Chart Paper
1.
2.
Open the door.
If recording is in progress, press the
RCD
key to stop the recording.
Remove the chart cassette.
Gently pressing the left and right stoppers inward. The bottom section of the
chart cassette comes out. Gently lift the chart cassette and pull it out from the
recorder case.
Stopper
3-4
IM 04P01B01-01E
3.2 Installing or Replacing the Chart Paper
3.
Open the chart holder and the chart paper guide.
Chart holder
This sheet is provided on models
with the chart end detection
function (/F1 option).
Do not remove or bend this sheet.
3
4.
Load the chart paper.
Riffle the chart thoroughly before loading.
Make sure that the sprocket teeth of the chart drives are properly engaged in the
chart paper perforations. Make sure not to load the chart paper backwards.
Z-fold chart paper
Sprocket teeth
5.
Close the chart holder and close the chart paper guide.
Chart holder
Chart paper guide
IM 04P01B01-01E
The side with the
long rectangular
holes is the right side.
3-5
Names of Parts and Run Operations
Chart paper guide
3.2 Installing or Replacing the Chart Paper
6.
Replace the chart cassette back into the recorder case.
Align the left and right projections of the sprocket section with the guide grooves
of the recorder and press the entire chart cassette into the recorder case. The
chart cassette is fixed in place with the stoppers.
Stopper
Feeding the Chart Paper
7.
Press the FEED key to assure that the chart moves two or more folds smoothly
into the chart receiver.
If it moves unsteadily, do the installing procedure again.
3-6
IM 04P01B01-01E
3.3
Installing/Replacing Felt Pens or Plotter Pen
(Pen Model)
CAUTION
• Do not press or pinch the felt tip to prevent deformation.
• Do not move the penholder left or right by force to protect the driving
mechanism.
• Make sure to remove the pen cap before installation.
• Use pen caps of the same ink color. If a pen cap of a different ink color is used
on the pen, the remaining ink in the cap may be absorbed through the pen tip,
and the ink may change its color.
3
1.
2.
3.
4.
Open the door.
If recording is in progress, press the RCD key to stop the recording.
Open the display and key panel section.
You can open the display and key panel section by holding the tab at the lower
left and pulling it toward you.
Hold the felt pen cartridge and pull it out from the pen holder.
If the pen (pen holder) is at a position that is not easily accessible, see “When
the Pen (Pen Holder) Is at a Position That Is Not Easily Accessible” on the next
page.
Remove the cap from the new felt pen and insert the pen firmly into the pen
holder.
From the top: pen 1 (red),
pen 2 (green), pen 3 (blue),
and pen 4 (violet).
Display and key panel
IM 04P01B01-01E
3-7
Names of Parts and Run Operations
Replacing Felt Pens
3.3 Installing/Replacing Felt Pens or Plotter Pen (Pen Model)
When the Pen (Pen Holder) Is at a Position That Is Not Easily Accessible
If the pen (pen holder) is at a position that is not easily accessible, carry out the
procedure below to move it near the center position.
1.
Turn ON the power switch and press the FUNC key.
( DISP ) key several times to display the Pen exchange screen.
2.
Press the
Func=Pen
3.
exchange
Press the
( CH UP ) key.
The pen (pen holder) moves near the center position, and the Pen exchange
= End appears.
Note
When the pen moves, a line is drawn on the chart paper.
4.
5.
Replace the pen.
Press the
( CH UP ) key.
The screen returns to the data display screen.
Replacing the Plotter Pen
1.
2.
3.
4.
3-8
Open the door.
If recording is in progress, press the RCD key to stop the recording.
Open the display and key panel section.
Hold the plotter pen cartridge and pull it out from the pen holder.
Remove the cap from the new plotter pen and insert the pen firmly into the pen
holder.
IM 04P01B01-01E
3.4
Installing/Replacing the Ribbon Cassette (Dot
Model)
CAUTION
• Improper cassette insertion may cause the color to change or damage the ribbon.
• Do not apply upward force to the printer carriage. If you do, the carriage position
may be offset, and the recorder may not print correctly.
2.
3.
Open the door.
If recording is in progress, press the RCD key to stop the recording.
Press the FUNC key.
( DISP ) key several times to display R. exchange.
Press the
Func=R.
4.
5.
exchange
Press the
( CH UP ) key.
The printer carriage moves near the center position, and R. exchange = End is
displayed.
Open the display and key panel section.
You can open the display and key panel section by holding the tab at the lower
left and pulling it toward you.
Note
If the recorder is OFF, pinch the printer carriage and move it near the center position.
Printer carriage
6.
Remove the ribbon cassette.
Pull the left-hand part of the ribbon cassette so that the cassette holder tab
disengages, and pull the ribbon cassette out from the recorder case.
Holder tab
IM 04P01B01-01E
3-9
Names of Parts and Run Operations
1.
3
3.4 Installing/Replacing the Ribbon Cassette (Dot Model)
7.
Install a new ribbon cassette.
First, insert the right-hand part and then the left-hand part into the cassette holder.
Check that the cassette is properly engaged with the cassette holder tab.
If inserting the ribbon cassette is difficult, turn the ribbon feeding knob in the
direction of the arrow to align the ribbon feeding shaft of the cassette with the
ribbon feeding shaft hole of the holder.
Cassette holder
Holder tab
Hole for the ribbon
feed shaft
Ribbon feeding knob
8.
9.
3-10
Turn the ribbon feeding knob in the direction of the arrow a half turn or more to
check that the ribbon is feeding properly. If the ribbon is loose, turn the knob in
the direction of the arrow to tighten it.
Return the display and key panel section to its original position, and press the
( CH UP ) key.
The screen returns to the data display screen.
IM 04P01B01-01E
3.5
Starting/Stopping the Recording
This section describes the procedures for starting/stopping the recording and checking
the recorded result.
Procedure
3
Stopping the Recording
While recording is in progress, press the RCD key to stop recording.
The word “RECORD” on the status display clears.
Feeding the Chart Paper
The chart paper is fed while the
FEED
key is held down.
Viewing the Recorded Results
Pull the front cover tab of the chart cassette to open the front cover. The recorded chart
paper can be pulled out for viewing.
Recorded chart paper can
be pulled out.
Front cover tab
<Related Topics>
Changing the chart speed: Section 5.4
IM 04P01B01-01E
3-11
Names of Parts and Run Operations
Starting the Recording
Press the RCD key to start recording.
The status display shows the word “RECORD.”
3.6
Switching the Display Screen
This section describes the procedure for switching the display screen.
To change the displayed content, see chapter 8.
Procedure
Switching the Display Screen
The display screen switches each time the DISP key is pressed.
Screen 01 through 15 are switched in order. Screens that are set to “Skip” are skipped.
Channel Auto Switching
On screens that show the measured values and computed values, the displayed channel
is automatically switched in ascending order. The switching interval can be set to 1 s, 2
s, 3 s, 4 s, or 5 s.
Measurement channels set to Skip and computation channels set to OFF are not
displayed.
Switching the Displayed Channel Using Keys
If auto switching is not specified, the channel switches each time the
in ascending order. All channels are displayed in order.
CH UP
key is pressed
Note
The operation is different between auto channel switching and manual switching using keys
on the 2-channel digital display.
3-12
During auto switching
01
02
03
04
05
06
When switching using keys
01
02
02
03
03
04
IM 04P01B01-01E
3.7
Printing Measured Values (Manual Printout)
The measured values of all channels are printed.
Procedure
Starting the Manual Printout
1.
Press the FUNC key.
2.
Press the
key to select Print out and then press the
3.
out
Press the
key with ManualStart shown on the screen.
Manual printout starts. The screen returns to the data display screen.
Print=ManualStart
Note
•
•
•
When manual printout is executed, trend recording is suspended. However, the recorder
continues the measurement and alarm detection (in the background).
When manual printout is complete, trend recording resumes.
If an alarm occurs during the manual printout, the alarm is printed after the trend recording
resumes.
Aborting the Manual Printout
The following procedure is for aborting the manual printout that is in progress.
1.
Press the FUNC key.
2.
Press the
key to select Print out and then press the
key.
3.
Press the
key with Manual Stop shown on the screen.
Manual printout stops. The screen returns to the data display screen.
Print=Manual
Stop
* When the
key or
key is pressed while holding down the SHIFT key, the operation is
reversed as when the respective key is pressed by itself.
IM 04P01B01-01E
3-13
Names of Parts and Run Operations
Func=Print
3
key.
3.8
Printing the Recorder Settings
This section explains the procedure for printing the recorder settings. There are two sets
of settings that can be printed: List and Setup.
List: Prints the settings of Setting mode (input range for each channel, etc.)
Setup: Prints the settings of Basic Setting mode
Note
•
•
•
•
The printout takes several minutes to tens of minutes to complete.
When this printout is executed, trend recording is suspended. However, the recorder
continues the measurement and alarm detection (in the background).
When the printout is complete, trend recording resumes.
If an alarm occurs during the printout, the alarm is printed after the trend recording resumes.
Procedure
Starting the List Printout
1.
Press the FUNC key.
2.
Press the
key to select Print out and then press the
key.
3.
Press the
key to display List Start and then press the
key.
The List printout starts. The screen returns to the data display screen.
Print=List
Start
Aborting the List Printout
1.
Press the FUNC key.
2.
Press the
key to select Print out and then press the
key.
3.
Press the
key to display List Stop and then press the
key.
The List printout stops. The screen returns to the data display screen.
Print=List
Stop
Starting the Setup Printout
1.
Press the FUNC key.
2.
Press the
key to select Print out and then press the
key.
3.
Press the
key to display Setup Start and then press the
key.
The Setup printout starts. The screen returns to the data display screen.
Print=Setup
Start
Aborting the Setup Printout
1.
Press the FUNC key.
2.
Press the
key to select Print out and then press the
key.
3.
Press the
key to display Setup Stop and then press the
key.
The Setup printout stops. The screen returns to the data display screen.
Print=Setup
Stop
* When the
key or
key is pressed while holding down the SHIFT key, the operation is
reversed as when the respective key is pressed by itself.
3-14
IM 04P01B01-01E
3.9
Clearing the Alarm Printout Buffer
Alarm information waiting to be printed is temporarily stored in the buffer memory. This
operation clears all of the alarm information in the buffer. This function can be used to
prevent unneeded alarm printouts from being executed.
3
Procedure
Press the
Press the
FUNC
key.
key to select Buffer clear and then press the
Func=Buffer
3.
key.
clear
Press the
key to select Alarm and then press the
key.
The data in the alarm printout buffer is cleared. The screen returns to the data
display screen.
Buf.clear=Alarm
* When the
key or
key is pressed while holding down the SHIFT key, the operation is
reversed as when the respective key is pressed by itself.
IM 04P01B01-01E
3-15
Names of Parts and Run Operations
1.
2.
3.10 Printing Messages
This section explains the procedure for printing the preset character strings. For the
procedure of setting the character strings, see section 6.8.
Note
•
•
Messages can be printed only during trend recording. However, regardless of whether trend
recording is ON or OFF, messages waiting to be printed are temporarily stored in the buffer
memory.
Message printouts are not performed when the chart speed is greater than or equal to 1600
mm/h and 101 mm/h on the pen model and dot model, respectively.
Procedure
Printing a Message
1.
Press the FUNC key.
2.
Press the
key to select Message and then press the
key.
Func=Message
3.
Press the
key to select message number and then press the
key.
The message printout starts. The screen returns to the data display screen.
Message=Message1
Displays the preset message.
Clearing the Message Printout Buffer
Messages waiting to be printed are temporarily stored in the buffer memory. This
operation clears the messages in the buffer.
1.
Press the FUNC key.
2.
Press the
key to select Buffer clear and then press the
key.
3.
Press the
key to select Message and then press the
key.
The messages in the printout buffer are cleared. The screen returns to the data
display screen.
Buf.clear=Message
* When the
key or
key is pressed while holding down the SHIFT key, the operation is
reversed as when the respective key is pressed by itself.
3-16
IM 04P01B01-01E
3.11 Resetting the Report Data of the Periodic
Printout
This operation resets the past report data when the recorder is configured to print the
report data (the average, the minimum, the maximum, or the sum) of the measured
values in periodic printout.
3
Procedure
key.
Func=Periodic
3.
Press the
key to select Reset and then press the
key.
The report data is reset, and the calculation of the report data starts again from
that point. The screen returns to the data display screen.
Log
data=Reset
* When the
key or
key is pressed while holding down the SHIFT key, the operation is
reversed as when the respective key is pressed by itself.
IM 04P01B01-01E
3-17
Names of Parts and Run Operations
Resetting the Report Data
1.
Press the FUNC key.
key to select Periodic and then press the
2.
Press the
3.12 Releasing the Alarm Output (Alarm ACK
Operation)
This operation releases the alarm indication or relay output (/A1, /A2, or /A3 option)
when the alarm indication or output relay is set to hold operation.
Procedure
Releasing the Alarm Output
1.
Press the FUNC key.
2.
Press the
key with Alarm ACK shown on the screen.
The alarm indication or relay output is released. The screen returns to the data
display screen.
Func=Alarm
ACK
Explanation
Alarm Output Relay Operation
• When the alarm is released
The alarm output is released when the alarm ACK operation is carried out.
• When an alarm is occurring
The alarm output is held even after the alarm ACK operation. If the alarm ACK
operation is carried out again after the alarm is released, the alarm output is released.
Indication (Status Display)
Blinks when an alarm occurs.
• When the alarm is released
The alarm indication turns OFF when the alarm ACK operation is carried out.
• When an alarm is occurring
The alarm indication turns ON after the alarm ACK operation. The alarm indication
turns OFF when the alarm is released.
Alarm ACK
Alarm ACK
Alarm
or
Alarm output
relay
or
Blinking
Alarm indication
OFF
3-18
Blinking
or
Alarm ACK
ON
OFF
IM 04P01B01-01E
3.13 Activating/Releasing the Key Lock
When the recorder is configured to use the key lock function, this operation activates or
releases the key lock.
Note
Key lock does not apply to the DISP and CH UP keys.
3
Names of Parts and Run Operations
Procedure
Activating the Key Lock
1.
Press the FUNC key.
2.
Press the
key to select Keylock and then press the
key.
The key lock is activated. The screen returns to the data display screen.
Func=Keylock
Releasing the Key Lock
Note
A password is required to release the key lock.
1.
2.
3.
Press the FUNC key.
Press the
key to select Keylock and then press the
key.
Enter the password for releasing the key lock. The password values are shown
with asterisks.
Press the
key to increment the value.
Press the SHIFT +
key to decrement the value.
Press the
key to move the cursor to the right.
Press SHIFT +
key to move the cursor to the left.
Password=
4.
Press the
key.
The key lock is released. The screen returns to the data display screen.
<Related Topics>
Changing the password: Section 7.10
* When the
key or
key is pressed while holding down the SHIFT key, the operation is
reversed as when the respective key is pressed by itself.
IM 04P01B01-01E
3-19
Chapter 4 Common Operations for Setting Functions and Setup Guide
4.1
Run Modes
The recorder has three run modes.
Operation Mode
This mode is used for normal recording operation. The recorder enters this mode when
the power is turned ON. The operations that can be carried out in this mode are
described in chapter 3. The details on the data display setup screen are explained in
chapter 8.
Basic Setting Mode
This mode is used to set the basic specifications of the recorder such as the
thermocouple burnout detection function and the alarm output relay operation. This
mode cannot be entered while the recorder is recording or while computation is in
progress on the computation function (/M1 option). Measurement, recording, and alarm
detection cannot be carried out in this mode. The operations that can be carried out in
this mode are described in chapter 7.
Power ON
Display “End=Store” and press
Setting mode
Operation mode
Hold down
1 MENU
Basic Setting mode
for 3 s
Data display setup screen
Hold down
MENU for 3 s
IM 04P01B01-01E
Hold down
+
for 3 s
4-1
4
Common Operations for Setting Functions and Setup Guide
Setting Mode
This mode is used to set the input range, alarms, chart speed, and other parameters.
These settings can be changed while recording is in progress. However, the input range
of measurement channels and the computing equation, unit, constant, and TLOG setting
of computation channels cannot be changed while computation (/M1 option) is in
progress. The operations that can be carried out in this mode are described in chapter 5
and 6.
4.2
Key Operations
This section describes basic operations on the panel keys to change various settings.
Functions are set in Setting mode or Basic Setting mode.
Entering Setting Mode
Hold down the MENU key for 3 seconds.
The Setting mode display appears. The top and bottom lines are the setup item and
comment, respectively.
The section that is blinking in the setup item that you change. In this manual, the section
that you change appears shaded.
Setup item
Comment
Set=Range
Input range
The item to be controlled blinks.
and
The panel keys are set to the functions marked above the keys.
CHARACTER
RCD
MENU
DISP
FUNC
ESC/?
SHIFT
1 MENU
FEED
CH UP
Exiting from Setting Mode (Returning to Operation Mode)
Hold down the MENU key for 3 seconds.
The screen returns to operation mode.
Entering Basic Setting Mode
Basic Setting mode is entered from Setting mode.
Hold down the MENU key for 3 seconds to enter Setting mode. Next, hold down both the
( DISP ) key and the
(
FUNC )
key for 3 seconds.
The Basic Setting mode display appears. The top and bottom lines are the setup item
and comment, respectively.
The section that is blinking in the setup item that you change. In this manual, the section
that you change appears shaded.
Setup item
Comment
Basic=Alarm
Auxiliary alarm
The item to be controlled blinks.
Exiting from Basic Setting Mode (Returning to Operation Mode)
This operation is used to return to Operation mode after changing the settings in Basic
Setting mode.
Press the ESC key several times to return to the Basic= screen.
Press the
key or SHIFT +
key to select End and then press the
key. The
setup save screen appears.
Basic=End
Save Settings
Press the
key to select Store and then press the
key. The setting is applied,
and the screen returns to Operation mode. If you select Abort and press the
key,
the setting is discarded, and the screen returns to Operation mode.
End=Store
Save settings
4-2
and
IM 04P01B01-01E
4.2 Key Operations
Changing the Settings
Note
The comment line shows useful information such as a description of the setup item and the
range of selectable values. Read the comment and change the items as necessary.
The selected item change each time you press the
key. The selected item changes
in reverse order if you press the
key while holding down the SHIFT key.
SHIFT key +
key
Skip
Volt
Mode=TC
4
key
This manual denotes the operation of pressing a key while holding down the SHIFT key
as SHIFT + the other key (for example: SHIFT +
key).
After you make a selection, press the
key. The next screen appears.
When the Setting complete screen appears, the changed item is applied.
01-01 Channel
Setting complete
Using the ESC Key
If you press the ESC ( MENU ) key, the operation is cancelled, and the display returns to a
higher level menu. In other words, if you do not show the Setting Complete screen, the
changes you made up to that point are discarded.
Press the ESC ( 1 MENU ) key while holding down the SHIFT ( FEED ) key to show or hide the
comment that is displayed at the bottom half of the screen.
Entering Values
Use the
key or SHIFT +
key to move the cursor. Use the
key or SHIFT +
key to change a digit value. You repeat these steps to enter the value.
SHIFT key +
key
key
Span_L=
-2.000/
-2.000
2.000V
Cursor
Span_L=
-2.000/
8
7
6
-1.500
4
3
2
When you press the
IM 04P01B01-01E
key
SHIFT key +
key
key, the change is applied and the next screen is displayed.
4-3
Common Operations for Setting Functions and Setup Guide
RTD
1-5V
4.2 Key Operations
Entering Characters
Use the
key or SHIFT +
key to move the cursor.
SHIFT key +
key
key
Unit=ppm
CHR:a-z
Cursor
Use the CHARACTER key or SHIFT + CHARACTER key to select the character type.
key or SHIFT +
key to select a character. You repeat these steps to
Use the
set the character string.
Unit=ppm
CHR:a-z
Character type
The character type changes in the following order: uppercase alphabet, lowercase
alphabet, numbers, and symbols.
p
o
n
Unit=ppm
CHR:a-z
l
k
j
key
SHIFT key +
When you press the
key
key, the change is applied and the next screen is displayed.
Inserting Characters
Press the
key or SHIFT +
key to move the cursor to the position where the
character is to be inserted.
Press the CHARACTER key or SHIFT + CHARACTER key to show Ins DISP and then
key. A space for one character is inserted. Enter the character.
press the
Position to insert the character
Unit=ppm
CHR:Ins
DISP
Display Ins DISP
Deleting a Character
Use the
key or SHIFT +
key to move the cursor to the character to be deleted.
Press the CHARACTER key or SHIFT + CHARACTER key to show Del DISP and then
key. The character is deleted.
press the
Deleting an Entire Character String
Press the CHARACTER key or SHIFT + CHARACTER key to show Clear DISP and
then press the
key. The entire character string is deleted.
Copying & Pasting a Character String
Show the copy source character string.
Press the CHARACTER key or SHIFT + CHARACTER key to show Copy DISP and
then press the
key. The character string is saved to the memory.
Show the copy destination.
Press the CHARACTER key or SHIFT + CHARACTER key to show Paste DISP and
then press the
key. The character string is pasted.
4-4
IM 04P01B01-01E
4.3
Menu Structure, Settings, and List of Default
Values
Operation Menus Using the FUNC Key (Operation Mode)
Below are the items that are operated using the
the parentheses is references.
FUNC
key. Information indicated inside
FUNC key
: Use the
Alarm ACK
(section 3.12)
Math
(section 9.1)
: Use the
key.
key.
Start/Stop
4
Reset
Common Operations for Setting Functions and Setup Guide
Print out
(sections 3.7, 3.8)
ManualStart/Manual Stop
List Start/List Stop
Setup Start/Setup Stop
Message
(section 3.10)
Message 1
Message 2
Message 3
Message 4
Message 5
Buffer clear
(sections 3.9, 3.10)
Alarm
Message
Keylock
(section 3.13)
Password
Periodic
Reset
(section 3.11)
Pen exchange
(section 3.3)
Pen exchange = End
R. exchange
(section 3.4)
R. exchange = End
Alarm ACK
Alarm output release operation. This operation is valid only when the alarm indication or
relay output is set to hold operation.
Math (Models with the Computation Function (/M1 Option))
Starts/stops the computation.
Print out
Starts/Stops the manual printout or setting printout (List or Setup).
Message
Prints messages.
Buffer clear
Clears the data waiting to be printed in the alarm printout buffer or the message printout
buffer.
Keylock
Activates or releases the key lock. This operation is possible only when the use of the
key lock function is enabled.
Periodic
Resets the computed value when printing of the report data (average, etc.) is specified in
periodic printout.
IM 04P01B01-01E
4-5
4.3 Menu Structure, Settings, and List of Default Values
Pen exchange (Pen Model)
Moves the recording pen to a position that is easily accessible for replacement on the
pen model.
R. exchange (Dot Model)
Moves the printer carriage near the center position when replacing the ribbon cassette
on the dot model.
4-6
IM 04P01B01-01E
4.3 Menu Structure, Settings, and List of Default Values
Menu Structure of Setting Mode
Below are the setup items in Setting mode. Information indicated inside the parentheses
is references.
Range
CH
Volt
Range
Span_L
Span_R
TC
Range
Span_L
Span_R
RTD
Range
Span_L
Span_R
1-5V
Span_L
Scale
Type
Delta
Ref.CH
DI
Range
Span_L
Span_R
SQRT
Range
Span_L
Span_R
(section 5.1)
Span_R
Range
Scale_L
Span_L
Span_L
Scale_R
Span_R
Low-cut
Scale_L
Span_R
4
Scale_R
Low-cut
CH
Bias
CH
Level
CH
Unit
Low-cut point
Bias
(section 6.12)
Alarm
Alarm
Type
Value
Relay
Relay_No.
(section 5.2)
Unit
(section 5.3)
Chart
Key operation
Hold down the MENU key for 3 seconds
in Operation mode to enter Setting mode.
mm/h
(section 5.4)
Clock
(Date/Time)
(section 5.5)
Aux
Trend
: Use the
Mode
(section 6.1)
Zone
: Use the
CH
Left
Right
CH
Partial
Expand
CH
Trend
Periodic
CH
Tag
Boundary
(section 6.5)
Print out
Hold down both the
and
keys
for 3 seconds in Setting mode to enter
Basic Setting mode.
(section 6.6)
Tag
key.
Hold down the MENU key for 3 seconds
in Setting mode to return to Operation mode.
(section 6.4)
Partial
key.
(section 6.7)
Message
Message No.
(Message)
(section 6.8)
Chart 2
mm/h
(section 6.9)
Moving_AVE
CH
No. of samples
CH
Resp. Time
CH
Duration
(section 6.3)
Filter
(section 6.2)
Alarm delay T
(section 6.10)
Brightness
Display
Light
(section 6.11)
DST
DST
Start month
Start day
Start time
End month
End day
End time
(section 6.13)
Math
Formula
CH
Mode
CH
Unit
f.
Span_L
Span_R
(section 9.2)
Unit
(section 9.3)
Constant
Number
K##
(section 9.4)
Alarm
CH
Level
Alarm
CH
Timer_No.
Type
Value
Relay
Relay_No.
(section 9.5)
TLOG
SUM scale
(section 9.6)
Aux
Zone
CH
Left
Right
CH
Partial
Expand
CH
Trend
Periodic
CH
Tag
CH
Duration
(section 9.7)
Partial
are not displayed in the default
condition. To display these items,
settings must be changed in Basic
Setting mode.
Bound
(section 9.8)
Print out
(section 9.9)
Tag
(section 9.10)
Alarm delay T
(section 9.11)
IM 04P01B01-01E
4-7
Common Operations for Setting Functions and Setup Guide
Scale_L
Skip
Bias
Scale_R
4.3 Menu Structure, Settings, and List of Default Values
Menu Structure of Basic Setting Mode
Below are the setup items in Basic Setting mode. Information indicated inside the
parentheses is references.
Alarm
Diagnosis
Reflash
AND
Act
Behavior
(section 7.1)
Indicator
A/D
Increase
Decrease
(section 7.2)
Burnout
CH
Burnout
CH
RJC
Volt
(section 7.4)
Color
: Use the
Channel
Color
keys for 3 seconds in
key.
: Use the
key.
Chart speed
Scale
(section 7.5)
POC
M_hysteresis
Key operation
Hold down both the
and
Setting mode to enter this mode.
(section 7.3)
RJC
Hysteresis
Integrate
POC
(section 7.6)
Print
CH/Tag
Channel
Alarm
Record On
Periodic
Ref.Time
Interval
Pen color
(section 7.7)
Print 1
Mode
(section 7.8)
Print 2
CH
Mode
CH
Graph
SUM scale
(section 7.8)
Bar graph
(section 7.9)
Keylock
Keylock
Password
(Free/Lock)
(section 7.10)
Moving_AVE
Moving_AVE
(section 7.11)
Filter
Filter
(section 7.12)
Partial
Partial
(section 7.13)
Language
Lang
(section 7.14)
Date format
Type
(section 7.19)
Temperature
Temp
(section 7.20)
Add function
Personalize
Bias
SQRT low-cut
Alarm
Message
1-5V low-cut
Alarm delay
(section 7.15)
(section 7.14)
Time print
RCD On
C.Speed
(section 7.16)
Initialize
Mode
Are you sure?
(section 7.17)
Remote
Remote No.
(No.)
(section 7.18)
Timer (TLOG)
Math
Timer No.
Mode
Interval
Ref.Time
Reset
Print
(section 9.12)
Color
Channel
Color
(section 9.13)
Output pen
Pen
CH
(section 9.14)
Print 2
CH
Mode
CH
Graph
SUM scale
(section 9.15)
Bar graph
(section 9.16)
Error data
Error
Over
(section 9.17)
RS422/485
Address
Baud rate
Data length
Parity
Protocol
(Communication)
Ethernet
Host
Host
Domain
(Communication)
Local IP
A
M
IP address
DNS
G
Subnet mask
DNS
P
S
Gateway
Suffix_P
Primary server Secondary server Suffix primary
P_Adj
Login
Login
Login set
Level
Timeout
Timeout
K.Alive
Keep alive
P_Adj
Pen No.
Register
User
Suffix_S
Suffix secondary
Password
Duration
(Value)
(sections 11.5, 11.6)
End
End
(section 4.2)
4-8
IM 04P01B01-01E
4.3 Menu Structure, Settings, and List of Default Values
Setup Items in Setting Mode and Their Default Values
Items with an asterisk on the left are not displayed in the default condition. To display
these items, settings must be changed in Basic Setting mode.
Setup Item
Pen/Dot
Selectable Range or Selections
Default Value
Range > CH
Range > Mode
-
01
Volt
Range > Mode > Range
-
01 to 06
Volt/TC/RTD/1-5V/Scale
/Delta/DI/SQRT/Skip
20mV/60mV/200mV/2V/6V/20V/50V/1-5V
/R/S/B/K/E/J/T/N/W/L/U/WRe/Pt/JPt/Level/Cont
/(Slections for /N1 and /N3 options)
*Bias > CH
*Bias > Bias
*Bias > Bias
-
01 to 06
On/Off
±10 % of the span of the measurable range
01
Off
-
Alarm > CH
Alarm > Level
Alarm > Alarm
Alarm > Type
Alarm > Value
Alarm > Relay
Alarm > Relay No.
-
01 to 06
1/2/3/4
On/Off
H/L/h/l/R/r (*/T/t)
Depends on the alarm type.
On/Off
I01 to I06
01
1
Off
H
Off
I01
Unit > CH
Unit > Unit
-
01 to 06
6 characters or less
01
Blank
Chart
Pen Model 82 types (pen model)
Dot Model 1 to 1500 mm/h (dot model)
2V
4
Clock
-
Date/Time
-
Aux > Trend
Aux > Zone > CH
Aux > Zone > Left, Right
*Aux > Partial > CH
*Aux > Partial > Partial
*Aux > Partial > Expand
*Aux > Partial > Boundary
Aux > Print out > CH
Aux > Print out > Trend
Aux > Print out > Periodic
Aux > Tag > CH
Aux > Tag > Tag
Aux > Message > Message No.
Aux > Message > (Message)
Aux > Chart2
Dot Model
Dot Model
Pen Model
Dot Model
Dot Model
Dot Model
Pen Model
Pen Model
-
Auto/Fix
01 to 06
Within the recording span range (mm)
01 to 06
On/Off
1 to 99%
Within the recording span range
01 to 06
On/Off
On/Off
01 to 06
7 characters or less
1 to 5
16 characters or less
82 types (pen model)
1 to 1500 mm/h (dot model)
01 to 06
Off, 2 to 16
01 to 06
2s/5s/10s
01 to 06
0 to 3600s
1/2/3/4/5/6/7/8
1/2/3/4
Not/Use
Apr/May/Jun/Jul/Aug/Sep/Oct/Nov/Dec/Jan/Feb/Mar
1st-Sun/.../Last-Mon
0:00 to 23:00
Apr/May/Jun/Jul/Aug/Sep/Oct/Nov/Dec/Jan/Feb/Mar
1st-Sun/.../Last-Mon
0:00 to 23:00
Auto
01
Left: 0, Right: 100
01
Off
50
01
On
On
01
1 to 6
1
Blank
20 mm/h
20 mm/h
01
Off
01
Off
01
10s
4
2
Not
Apr
1st-Sun
0:00
Apr
Last-Mon
0:00
*Aux > Moving_AVE > CH
*Aux > Moving_AVE > No. of samples
*Aux > Filter > CH
*Aux > Filter > Resp. Time
*Aux > Alm delay T > CH
*Aux > Alm delay T > Duration
Aux > Brightness > Display
Aux > Brightness > Light
Aux > DST
Aux > DST > Start month
Aux > DST > Strt day
Aux > DST > Start time
Aux > DST > End month
Aux > DST > End day
Aux > DST > End time
Computation function (/M1 option)
Setup Item
Pen/Dot
Selectable Range or Selections
Default Value
Math > Formula > CH
Math > Formula > Mode
Math > Formula > f.
Math > Unit > CH
-
0A/0B/0C/0D/0E/0F/0G/0J/0K/0M/0N/0P
On/Off
120 characters or less
0A/0B/0C/0D/0E/0F/0G/0J/0K/0M/0N/0P
0A
Off
01
0A
IM 04P01B01-01E
4-9
Common Operations for Setting Functions and Setup Guide
20 mm/h
20 mm/h
4.3 Menu Structure, Settings, and List of Default Values
Setup Item
Pen/Dot
Selectable Range or Selections
Default Value
Math > Unit > Unit
Math > Constant > No.
Math > Alarm > CH
Math > Alarm > Level
Math > Alarm > Alarm
Math > Alarm > Type
Math > Alarm > Value
Math > Alarm > Relay
Math > Alarm > Relay No.
Math > TLOG > CH
Math > TLOG > Timer No.
Math > TLOG > SUM scale
Math > Aux > Zone > CH
Math > Aux > Zone > Left, Right
*Math > Aux > Partial > CH
*Math > Aux > Partial > Partial
*Math > Aux > Partial > Expand
*Math > Aux > Partial > Bound
Math > Aux > Print out > CH
Math > Aux > Print out > Trend
Math > Aux > Print out > Periodic
Math > Aux > Tag > CH
Math > Aux > Tag > Tag
*Math > Aux > Alm delay T > CH
*Math > Aux > Alm delay T > Duration
Dot Model
-
6 characters or less
K01 to K30
0A/0B/0C/0D/0E/0F/0G/0J/0K/0M/0N/0P
1/2/3/4
On/Off
H/L (*/T/t)
–9999999 to 99999999 excluding the decimal
On/Off
I01 to I06
0A/0B/0C/0D/0E/0F/0G/0J/0K/0M/0N/0P
Periodic/1/2
Off, /s, /min, /h, /day
0A/0B/0C/0D/0E/0F/0G/0J/0K/0M/0N/0P
Within the recording span range (mm)
0A/0B/0C/0D/0E/0F/0G/0J/0K/0M/0N/0P
On/Off
1 to 99%
Within the recording span range
0A/0B/0C/0D/0E/0F/0G/0J/0K/0M/0N/0P
On/Off
On/Off
0A/0B/0C/0D/0E/0F/0G/0J/0K/0M/0N/0P
7 characters or less
0A/0B/0C/0D/0E/0F/0G/0J/0K/0M/0N/0P
0 to 3600s
Blank
K01
0A
1
Off
H
Off
101
0A
Periodic
Off
0A
Left: 0, Right: 100
0A
Off
50
0A
On
On
0A
A to P
0A
10s
Setup Items in Basic Setting Mode and Their Default Values
Setup Item
Pen/Dot
Selectable Range or Selections
Default Value
Alarm > Diagnosis
Alarm > Reflash
Alarm > AND
-
Off
Off
None
Alarm > Act
Alarm > Behavior
Alarm > Indicator
Alarm > Increase
Alarm > Decrease
Alarm > Hysteresis
Alarm > M_Hysteresis
-
On/Off
On/Off
None/I01/I01-I02/I01-I03/I01-I04/I01-I05
/I01-I06
Energize/De_energize
Nonhold/Hold
Nonhold/Hold
01 to 15
01 to 15
Off/0.1% to 1.0%
Off/0.1% to 1.0%
Energize
Nonhold
Nonhold
01
01
0.5%
Off
A/D > Integrate
-
Auto/50Hz/60Hz/100ms
Auto
Burnout > CH
Burnout > Burnout
-
01 to 06
Off/Up/Down
01
Off
RJC > CH
RJC > RJC
RJC > RJC > Volt
-
01 to 06
Internal/External
–20000 to 20000 µV
01
Internal
0 µV
Color > Channel
Color > Color
Dot Model
Dot Model
01 to 06
Purple/Red/Green/Blue/Brown/Black
01: Purple, 02: Red,
03: Green, 04: Blue,
05: Brown, 06: Black
POC > POC
Pen Model On/Off
Off
Print > CH/Tag
Print > Channel
Print > Alarm
Print > Record On
Print > Chart speed
Print > Scale
Print > Pen color
Dot Model
Pen Model
CH/Tag
On/Off
On1/On2/Off
On/Off
On/Off
On/Off
On/Off
CH
On
On1
Off
On
On
On
Print1 > Periodic
Print1 > Ref. Time
Print1 > Interval
-
Auto
00:00
1h
Print1 > Mode
-
Auto/Manual
Hour 0 to 23 (1 hour steps)
10min/15min/20min/30min/1h/2h/3h/4h
/6h/8h/12h/24h
Inst/Report/Off
4-10
Inst
IM 04P01B01-01E
4.3 Menu Structure, Settings, and List of Default Values
Pen/Dot
Selectable Range or Selections
Default Value
Print2 > CH
Print2 > Mode
Print2 > SUM scale
-
01 to 06
AVE/MIX/SUM/MIN/MAX/INST
Off, /s, /min, /h, /day
01
AVE
Off
Bar graph > CH
Bar graph > Graph
-
01 to 06
Normal/Center
01
Normal
Keylock > Keylock
Keylock > Password
Keylock > RCD
Keylock > Feed
Keylock > Menu
Keylock > Disp Menu
Keylock > Alarm ACK
Keylock > Math
Keylock > Print out
Keylock > Message
Keylock > Buffer clear
Keylock > Periodic
Keylock > Pen exchange
Keylock > R. exchange
Pen Model
Dot Model
Not/Use
Numbers and spaces within 4 digits
Free/Lock
Free/Lock
Free/Lock
Free/Lock
Free/Lock
Free/Lock
Free/Lock
Free/Lock
Free/Lock
Free/Lock
Free/Lock
Free/Lock
Not
Blank
Free
Free
Free
Free
Free
Free
Free
Free
Free
Free
Free
Free
Moving_AVE > Moving_AVE
Dot Model
Not/Use
Not
Filter > Filter
Pen Model Not/Use
Not
Partial > Partial
-
Not/Use
Not
Language > Lang
-
English/Japanese
English
Date format > Type
-
Y/M/D M/D/Y D/M/Y D.M.Y M.D.Y
M.D.Y
Temperature > Temp
-
C/F
C
Personalize > Add function > Bias
Personalize > Add function > SQRT low-cut Personalize > Add function > 1-5V low-cut Personalize > Add function > Alarm delay -
Not/Use
Not/Use
Not/Use
Not/Use
Not
Use
Not
Not
Personalize > Time print > Alarm
-
HH:MM
Personalize > Time print > Message
-
Personalize > Time print > RCD On
-
Personalize > Time print > C.Speed
-
HH:MM, HH:MM:SS, M/D H:M, M/D H:M:S,
YMD H:M:S
HH:MM, HH:MM:SS, M/D H:M, M/D H:M:S,
YMD H:M:S, None
HH:MM, HH:MM:SS, M/D H:M, M/D H:M:S,
YMD H:M:S
HH:MM, HH:MM:SS, M/D H:M, M/D H:M:S,
YMD H:M:S
4
Common Operations for Setting Functions and Setup Guide
Setup Item
HH:MM
HH:MM
HH:MM
Setting initialization
Initialize > Mode
Initialize > Mode > Are you sure?
-
Setup+Set/Set
No/Yes
Setup+Set
No
-
1/2/3/4/5
Record On Off /Chart speed/Time adjust/
Math start stop/Math reset/Manual print/
Alarm ACK/Message1/Message2/Message3/
Message4/Message5/None
1
Record On/Off
Remote control function (/R1 option)
Remote > Remote No.
Remote > Remote No. > No.
Computation function (/M1 option)
Setup Item
Pen/Dot
Selectable Range or Selections
Default Value
Math > Timer (TLOG) > Timer No.
Math > Timer (TLOG) > Mode
Math > Timer (TLOG) > Interval (Relative)
Math > Timer (TLOG) > Interval (Absolute)
-
1
Absolute
01:00
1h
Math > Timer (TLOG) > Ref. Time
Math > Timer (TLOG) > Reset
Math > Timer (TLOG) > Print
Math > Color > Channel
Dot Model
1/2
Off/Relative/Absolute
10 min to 24 h (1 min steps)
10min/12min/15min/20min/30min/1h/2h/3h
/4h/6h/8h/12h/24h
Hour 0 to 23 (1 hour steps)
On/Off
On/Off
0A/0B/0C/0D/0E/0F/0G/0J/0K/0M/0N/0P
IM 04P01B01-01E
00:00
On
On
0A
4-11
4.3 Menu Structure, Settings, and List of Default Values
Setup Item
Pen/Dot
Selectable Range or Selections
Default Value
Math > Color > Color
Dot Model
Purple/Red/Green/Blue/Brown/Black
Math > Output pen > Pen ← CH
Pen Model Pen: 1 to 4, Channel: 01 to 04/0A to 0J
Math > Print2 > CH
Math > Print2 > Mode
Math > Print2 > SUM scale
Math > Bar graph > CH
Math > Bar graph > Graph
Math > Error data > Error
Math > Error data > Over
-
Purple: 0A/0G
Red: 0B/0J
Green: 0C/0K
Blue: 0D/0M
Brown: 0E/0N
Black: 0F/0P
1pen: 01, 2pen: 02,
3pen: 03, 4pen: 04,
0A
AVE
Off
0A
Normal
+Over
Skip
0A/0B/0C/0D/0E/0F/0G/0J/0K/0M/0N/0P
AVE/MIX/SUM/MIN/MAX/INST
Off, /s, /min, /h, /day
0A/0B/0C/0D/0E/0F/0G/0J/0K/0M/0N/0P
Normal/Center
+Over/–Over
Skip/Limit
RS-422A/485 communication interface function (/C3 option).
Setup Item
Pen/Dot
Selectable Range or Selections
Default Value
RS-422/485 > Address
RS-422/485 > Baud rate
RS-422/485 > Data length
RS-422/485 > parity
RS-422/485 > Protocol
-
1 to 32
1200/2400/4800/9600/19200/38400
7/8
Odd/Even/None
NORMAL/MODBUS
1
9600
8
Even
NORMAL
Ethernet communication interface function (/C7 option)
Setup Item
Pen/Dot
Selectable Range or Selections
Default Value
Ethernet > Host > Host
Ethernet > Host > Domain
Ethernet > Local IP > A
Ethernet > Local IP > M
Ethernet > Local IP > G
Ethernet > DNS > DNS
Ethernet > DNS > P
Ethernet > DNS > S
Ethernet > DNS > Suffix_P
Ethernet > DNS > Suffix_S
Ethernet > Login > Login
Ethernet > LoginSet > Level
Ethernet > LoginSet > Register
-
64 characters or less
64 characters or less
IP address
IP address (Subnet mask)
IP address (Default gateway)
On/Off
IP address (Primary DNS sever)
IP address (Secondary DNS sever)
Primary domain suffix
Secondary domain suffix
Use/Not
Admin/User1 to User6
On/Off
Ethernet > LoginSet > User
-
16 characters or less
Ethernet > LoginSet > Password
-
4 characters or less
Ethernet > Timeout > Timeout
Ethernet > Timeout > Duration
Ethernet > K. Alive > Keep alive
-
On/Off
1 to 120 min
On/Off
Blank
Blank
0.0.0.0
0.0.0.0
0.0.0.0
Off
0.0.0.0
0.0.0.0
Blank
Blank
Not
Admin
Admin and User1
are On
Admin: admin
User1 to 6: user1 to
user6
Administrator: 0
User1 to 6: 1 to 6
Off
1 min
On
Setup Item
Pen/Dot
Selectable Range or Selections
Default Value
P_Adj > P_Adj
Dot Model Hysteresis/Zero/Full
Pen Model Zero/Full
Pen Model 1/2/3/4
Hysteresis
Zero
1
Setup Item
Pen/Dot
Selectable Range or Selections
Default Value
End > End
-
Store/Abort
Store
Adjustment
P_Adj > Pen No.
Store
4-12
IM 04P01B01-01E
4.4
Function Setup Guide
This section explains the settings necessary to use various functions of the recorder.
Read the section corresponding to the function you wish to use.
Note
This section contains all the settings related to each item. If the desired setting is the same
as the default value, you do not have to set it.
Description
Reference
Section
Date/Time setting
Use Clock in Setting mode
5.5
DST
Sets the date/time for switching between DST and standard time using
Aux > DST in Setting mode.
6.13
Setting initialization
Use Initialize in Basic Setting mode to initialize the settings of Setting mode
and Basic Setting mode to their default values.
7.17
4
Common Operations for Setting Functions and Setup Guide
Item
Measuring input functions
Item
Description
Input range of each channel
Set Range in Setting mode.
Reference
Section
5.1
1-5V
• Unit
5.3
Use Unit in setting mode to set the unit after scaling.
• Low-cut
7.15
Use Personalize > Add function > 1-5V low-cut in Basic Setting mode and
select Use or Not.
5.1
If Use is selected, turn On/Off the low-cut function using Range in Setting mode.
If Not is selected, the Low-cut item does not appear in the Range setting.
Scaling
• Unit
Use Unit in setting mode to set the unit after scaling.
Square Root Computation • Unit
Use Unit in setting mode to set the unit after scaling.
• Low-cut
Use Personalize > Add function > SQRT low-cut in Basic Setting mode
and select Use or Not.
If Use is selected, set the low-cut value using Range in Setting mode.
If Not is selected, the Low-cut item does not appear in the Range setting.
5.3
5.3
7.15
5.1
Bias
Use Personalize > Add function > Bias in Basic Setting mode and select Use
or Not.
If Use is selected, set the bias value that is added to the input using Bias in
Setting mode. If Not is selected, the Bias item does not appear.
7.15
6.12
Filter (pen model)
Use Filter in Basic Setting mode and select Use or Not.
If Use is selected, set the filter time constant using Aux > Filter in Setting mode.
If Not is selected, the Aux > Filter item does not appear.
7.12
6.2
Moving average (dot model)
Use Moving_AVE in Basic Setting mode and select Use or Not.
If Use is selected, set the number of samples of moving average using Aux >
Moving_AVE in Setting mode.
If Not is selected, the Aux > Moving_AVE item does not appear.
RJC of TC input
Use RJC in Basic Setting mode to select whether to use the recorder’s
RJC function or an external RJC function.
7.11
6.3
7.4
Burnout detection function of TC input and 1-5V input
Use Burnout in Basic Setting mode to set the burnout detection function for
each channel.
7.3
Integration time of the A/D converter
Use A/D in Basic Setting mode to set the integration time of the A/D converter.
7.2
Temperature Unit
7.20
IM 04P01B01-01E
Select the temperature unit using Temperature in Basic Setting mode.
4-13
4.4 Function Setup Guide
Alarm functions
Item
Description
Reference
Section
Alarms for each channel
Use Alarm in Setting mode.
5.2
Delay high limit alarm/delay low limit alarm
• Alarm type
Use Personalize > Add function > Alarm delay in Basic Mode and select Use or Not.
If Use is selected, delay high limit alarm and delay low limit alarm become
selectable under Alarm > Type in Setting mode.
If Not is selected, alarm delay cannot be selected.
• Alarm delay time
Use Aux > Alm delay T in Setting mode to set the alarm delay time.
High limit on rate-of-change alarm/low limit on rate-of-change alarm
Use Alarm in Setting mode and set the channel to rate-of-change alarm.
Use Alarm > Increase in Basic Setting mode to set the interval of the high limit on
rate-of-change alarm; use Alarm > Decrease in Basic Setting mode to set the
interval of the low limit on rate-of-change alarm.
Use the alarm ACK operation
• Relay operation and display operation
Set Alarm > Behavior in Basic Setting mode to Hold and set Alarm > Indicator
to Hold.
• Executing the alarm ACK operation
Use FUNC > Alarm ACK in Operation mode to execute the release of the
alarm output.
Set a hysteresis on the alarm occurrence/release value of high limit alarm and low limit alarm
Use Alarm > Hysteresis in Basic Setting mode to set the hysteresis to be applied
to the high limit alarm and low limit alarm of measurement channels.
Likewise, use Alarm > M_Hysteresis to set the hysteresis to be applied to the high
limit alarm and low limit alarm of computation channels (/M1 option).
7.15
5.2
6.10
5.2
7.1
7.1
3.12
7.1
Change the alarm output relay operation
If Alarm > Act in Basic Setting mode set to De_energize, alarm output relay is
energized during normal operation and de-energized when an alarm occurs.
7.1
Diagnosis output
Use Alarm > Diagnosis in Basic Setting mode to set the function.
7.1
Item
Description
Reference
Section
Chart speed
Use Chart in Setting mode to set the chart speed.
5.4
Recording functions
Turn trend recording ON/OFF (dot model)
Use Aux > Print out in Setting mode to set whether to trend record on each
measurement channel.
6.6
Change the recording color of measurement channels (dot model).
Use Color in Basic Setting mode to set the recording color of measurement channels. 7.5
Record by compensating for the pen offset along the time axis (pen model)
Use Aux > POC in Setting mode to turn On/Off offset compensation.
7.6
Record by setting recording zone for each channel
Use Aux > Zone in Setting mode to set the recording zone for each channel.
6.4
Partial expanded recording Use Partial in Basic Setting mode and select Use or Not.
If Use is selected, set the display mode using Aux > Partial in Setting mode.
If Not is selected, the Aux > Partial item does not appear.
4-14
7.13
6.5
IM 04P01B01-01E
4.4 Function Setup Guide
Item
Description
Reference
Section
Periodic printout
Mar.31.2004!
15:50*
1
1.000V
2
-1.000V
0.000
1
RED
50mm/h_
Channel number
or tag
Measured value
2.000
V
Scale
Recording color
Channel/tag selection
Use Print > CH/Tag in Basic Setting mode to select whether to use channel
numbers or tags in printouts.
Alarm occurrence/release printout
• Turn printout On/Off
Use Print > Alarm in Basic Setting mode to select whether to print the alarm
occurrence and release, print only the alarm occurrence, or not print.
• Time printout format
Use Personalize > Time print > Alarm in Basic Setting mode to set the time
printout format when printing alarm occurrence/release.
Recording Start Printout
7.8
4
7.8
Common Operations for Setting Functions and Setup Guide
• Printout interval
Use Print 1 in Basic Setting mode to set the periodic printout interval.
• Type of measured values to be printed/disable periodic printout
Use Print1 in Basic Setting mode to set the type of measured values to be
printed.
You can select Inst, Report (statistical calculation value), or Off.
If Report is selected, use Print2 in Basic Setting mode to select the report
data type.
• Scale printout and recording color printout (pen model)
Use Print > Scale in Basic Setting mode to turn scale printout On/Off.
Use Print > Pen color in Basic Setting mode to turn pen color printout On/Off
(pen model).
• Turn On/Off periodic printout for each channel
Use Aux > Print out in Setting mode to enable/disable the periodic printout on
each measurement channel.
7.7
6.6
7.7
7.7
7.16
• Turn printout On/Off
Use Print > Record On in Basic Setting mode to enable/disable the recording
start printout.
• Time printout format
Use Personalize > Time print > RCD On in Basic Setting mode to set the time
printout format.
7.7
• Turn printout On/Off
Use Print > Chart speed in Basic Setting mode to set whether to print the new
chart speed when the chart speed is changed.
• Time printout format
Use Personalize > Time print > C. Speed in Basic Setting mode to set the time
printout format.
7.7
7.16
New chart speed printout
7.16
Message printout
• Set the message string
6.8
Use Aux > Message to set the message string to be printed.
• Time printout format
7.16
Use Personalize > Time print > Message in Basic Setting mode to set the time
printout format.
• Execute the message printout
3.10
Use FUNC key > Message in Operation mode to execute the message printout.
Tag
• Channel/tag selection
Use Print > CH/Tag in Basic Setting mode to select whether to use channel
numbers or tags in printouts.
• Tag name
Use Aux > Tag in Setting mode to set the tag name.
7.7
Use Date format > Type in Basic Setting mode to set the printout/display format
of the date.
7.19
Date format
IM 04P01B01-01E
6.7
4-15
4.4 Function Setup Guide
Display functions
Item
Description
Reference
Section
Display brightness
Use Aux > Brightness > Display in Setting mode to set the display brightness.
6.11
Bar graph display mode
Use Bar graph in Basic Setting mode to set the bar graph display mode.
7.9
Display type
Assigns the display type on the data display setup screen.
Chapter 8
Date format
Same as the Date format in “Recording functions.”
7.19
Item
Description
Reference
Section
Key lock
• Target keys and password
Use Keylock in Basic Setting mode to set the keys to be key-locked and the
password.
• Enable the key lock
Use FUNC key > Keylock in Operation mode to turn key lock ON/OFF.
7.10
Language
Use Language in Basic Setting mode to set the display and recording language.
7.14
Internal light brightness
Use Aux > Brightness > Light in Setting mode to set the internal light brightness.
6.11
Other functions
3.13
Computation function (/M1 option)
Item
Description
Reference
Section
Computing equation and recording span
Use Math > Formula in Setting mode to set the computing equation and
recording span for each channel.
9.2
TLOG computation
9.12
If TLOG computation is selected, use Math > Timer in Basic Setting mode to
set the timer operation
Use Math > TLOG > Timer No. in Setting mode to set the timer used on each
computation channel.
9.6
Unit
Use Math > Unit in Setting mode to set the unit for each channel.
9.3
Constants
Use Math > Constant in Setting mode to define constants to be used in the
computing equations.
9.4
Alarms
Use Math > Alarm in Setting mode to set the alarm for each channel.
• Delay high limit alarm/delay low limit alarm
• Alarm type
Use Personalize > Add function > Alarm delay in Basic Mode and select
Use or Not (common with measurement channels).
If Use is selected, delay high limit alarm and delay low limit alarm become
selectable under Math > Alarm > Type in Setting mode. If Not is selected,
alarm delay cannot be selected.
• Alarm delay time
Use Math > Aux > Alm delay T in Setting mode to set the alarm delay time.
9.5
7.15
9.11
Record by setting recording zone for each channel
Use Math > Aux > Zone in Setting mode to set the recording zone for each channel. 9.7
Partial expanded recording Use Partial in Basic Setting mode and select Use or Not.
7.13
If Use is selected, set the display mode using Math > Aux > Partial in Setting mode. 9.8
If Not is selected, the Math > Aux > Partial item does not appear.
Change the recording color (dot model)
Use Color in Basic Setting mode to set the recording color of each computation
channel.
Recording pen assignment (pen model)
Assigns measurement/computation channels to the recording pens.
9.13
9.14
Turn trend recording ON/OFF (dot model)
Use Math > Aux > Print out > Trend in Setting mode to set whether to trend record 9.9
each computation channel.
4-16
IM 04P01B01-01E
4.4 Function Setup Guide
Item
Description
Reference
Section
Periodic printout
Mar.31.2004!
15:50*
1
1.000V
A
-100.00UNIT
-200.00
A
RED
50mm/h_
Channel number
or tag
Computed value
200.00
UNIT
Scale
Recording color
7.8
4
9.15
9.6
Common Operations for Setting Functions and Setup Guide
• Type of computed values to be printed/disable periodic printout
Common with measurement channels.
• If Report is selected, use Math > Print2 in Basic Setting mode to select
the report data type.
• Printout interval
Set Math > TLOG > Timer No. in Setting mode to Periodic.
The interval is common with measurement channels.
• Turn On/Off periodic printout for each channel
Use Math > Aux > Print out > Periodic in Setting mode to enable/disable the
periodic printout for each computation channel.
• Scale printout and recording color printout (pen model)
Common with measurement channels.
9.9
7.7
Periodic printout using TLOG timer (dedicated to computation channels)
Mar.31.2004!
15:50*
TAG-00A
100.00UNITA
B
1.0
-200.00
TAG-00A
RED
TLOG1
Channel number
or tag
Computed value
200.00
UNITA
Scale
Recording color
Timer number
• Use Math > Timer (TLOG) in Basic Setting mode to set the printout interval.
• Use Math > Timer (TLOG) > Print in Basic Setting mode to turn the printout On.
• Use Math > Aux > Print out > Periodic in Setting mode to enable/disable the
periodic printout for each computation channel.
• Set Math > TLOG > Timer No. in Setting mode to set the timer used on each
computation channel.
• Scale printout and recording color (pen model)
Common with periodic printout.
9.12
9.12
9.9
9.6
Alarm occurrence/release printout
Common with measurement channels.
7.7
Tag
• Channel/tag selection
Common with measurement channels.
• Tag name
Use Math > Aux > Tag in Setting mode to set the tag name.
7.7
Use Math > Bar graph in Basic Setting mode to set the bar graph display mode.
9.16
Bar graph display mode
Procedure when computed results are in error
Use Math > Error data in Basic Setting mode to set the procedure taken when the
computed results are in error.
9.10
9.17
Remote control function (/R1 option)
Item
Description
Remote control function (/R1 option)
• Assign functions to the remote control input terminals
Use Remote in Basic Setting mode to set the function to be assigned to the
remote control input terminal.
• Secondary chart speed
If “chart speed switching” is assigned, use Aux > Chart 2 in Setting mode to
set the secondary chart speed.
IM 04P01B01-01E
Reference
Section
7.18
6.9
4-17
Chapter 5 Frequently Used Setup Operations (Setting Mode)
5.1
Setting the Input Range
Input range is set for each measurement channel. Set unused channels to Skip.
• The input range cannot be changed on models with the computation function (/M1
option) when computation is in progress.
• If you change the input range, set the bias, alarm, and partial expanded recording
again.
Procedure
TC, RTD, and DC Voltage
1.
2.
3.
5.
Set=Range
Set the channel range.
CH=01-01
First channel Last channel
ESC/?
IM 04P01B01-01E
Mode=TC
Select TC, RTD, or Volt.
Range=R
Select the range type.
Span_L=
0.0
Span_R=
1760.0
01-01 Channel
Setting complete
Select the left span value.
Select the right span value.
The new setting takes effect.
5-1
5
Frequently Used Setup Operations (Setting Mode)
4.
Hold down the MENU key for 3 seconds to enter Setting mode.
Press the
key with Range shown on the screen.
key.
Set each item and press the
Use the
key or SHIFT +
key to select values.
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up
to then are cancelled, and the display returns to a higher level menu.
When the Setting complete screen appears, do either of the following:
Press the
key to set other channels.
If you are done, press the ESC key.
Hold down the MENU key for 3 seconds to return to Operation mode.
5.1 Setting the Input range
Description
• Selectable Range of Input Range, Span Left, and Span Right
The input range, span left, and span right can be set in the range shown below. Span
left and span right cannot be set to the same value.
DC voltage (Volt)
Range Type
Selectable Span Range
20 mV
–20.00 to 20.00 mV
60 mV
–60.00 to 60.00 mV
200 mV
–200.0 to 200.0 mV
2V
–2.000 to 2.000 V
6V
–6.000 to 6.000 V
20 V
–20.00 to 20.00 V
50 V
–50.00 to 50.00 V
Thermocouple (TC)
Range Type
Selectable Span Range (°C)
Selectable Span Range (°F)
R
0.0 to 1760.0 °C
32 to 3200 °F
S
0.0 to 1760.0 °C
32 to 3200 °F
B
0.0 to 1820.0 °C
32 to 3308 °F
K
−200.0 to 1370.0 °C
−328 to 2498 °F
E
−200.0 to 800.0 °C
−328.0 to 1472.0 °F
J
−200.0 to 1100.0 °C
−328.0 to 2012.0 °F
T
−200.0 to 400.0 °C
−328.0 to 752.0 °F
N
0.0 to 1300.0 °C
32 to 2372 °F
W
0.0 to 2315.0 °C
32 to 4199 °F
L
−200.0 to 900.0 °C
−328.0 to 1652.0 °F
U
−200.0 to 400.0 °C
−328.0 to 752.0 °F
WRe
0.0 to 2400.0 °C
32 to 4352 °F
RTD
Range Type
Selectable Span Range (°C)
Selectable Span Range (°F)
PT (Pt100)
−200.0 to 600.0 °C
−328.0 to 1112.0 °F
JPT (JPt100)
−200.0 to 550.0 °C
−328.0 to 1022.0 °F
RTD
Range Type
Selectable Span Range (°F)
PT (Pt100)
−200.0 to 600.0 °C
−328.0 to 1112.0 °F
JPT (JPt100)
−200.0 to 550.0 °C
−328.0 to 1022.0 °F
/N1 Option
Selections
Cu1
Cu2
Cu3
Cu4
Cu5
Cu6
Cu25
5-2
Selectable Span Range (°C)
Input Type
Cu10 (GE)
Cu10 (L&N)
Cu10 (WEED)
Cu10 (BAILEY)
Cu10: a = 0.00392 at 20°C
Cu10: a = 0.00393 at 20°C
Cu25*: a = 0.00425 at 0°C
/N3 Option
Selections
PR
PLATI
NiMo
W/WRe
N2
Kp
Pt4
Pt3
Ni1
Ni2
Ni3
J263
Cu8
Cu10
Input Type
Note
PR40-20
Thermo
PLATINEL
couple
NiNiMo
W/WRe26
Type N (AWG14)
Kp vs Au7Fe
Pt25
RTD
Pt50
Ni100 (SAMA)
Ni100 (DIN)
Ni120
J263*B
Cu53
Cu100: a = 0.00425 at 0°C
IM 04P01B01-01E
5.1 Setting the Input range
Linear Scaling
Set=Range
Set the channel range.
CH=01-01
First channel Last channel
Select Scale.
Type=Volt
Select Volt, TC, RTD, or DI.
Range=2V
Select the range type.
Span_L=
-2.000
Set the left span value.
Span_R=
2.000
Scale_L=
0.00
Scale_R=
200.00
01-01 Channel
Setting complete
5
Frequently Used Setup Operations (Setting Mode)
ESC/?
Mode=Scale
Set the right span value.
Set the left scale value.
Set the right scale value.
The new setting takes effect.
Description
• Scale Left and Scale Right
Selectable range (mantissa): –20000 to 30000
Decimal position: XXXXX, XXXX.X, XXX.XX, XX.XXX, X.XXXX
The decimal position is the position specified by scale left.
Note
The displayable/printable range of scale values is –19999 to 30000 excluding the decimal.
IM 04P01B01-01E
5-3
5.1 Setting the Input range
Delta Computation
Set=Range
Set the channel range.
CH=02-02
First channel
ESC/?
Last channel
Mode=Delta
Select Delta.
Ref.CH=01
Set the reference channel.
Span_L=
-2.000
Set the left span value.
Span_R=
2.000
02-02 Channel
Setting complete
Set the right span value.
The new setting takes effect.
Description
• Reference Channel
The reference channel must be a channel that is smaller in channel number than the
channel being set.
• Span Left and Span Right
The range type is set to the same type as the reference channel. The span left and
span right can be set in the range shown below. Span left and span right cannot be
set to the same value.
5-4
Input Type
Range Type
Selectable Span Range (°C)
DC voltage
(Volt)
20 mV
60 mV
200 mV
2V
6V
20 V
50 V
–20.00 to 20.00 mV
–60.00 to 60.00 mV
–200.0 to 200.0 mV
–2.000 to 2.000 V
–6.000 to 6.000 V
–20.00 to 20.00 V
–50.00 to 50.00 V
Selectable Span Range (°F)
Thermocouple R
(TC)
S
B
K
E
J
T
N
W
L
U
WRe
−1760.0 to 1760.0 °C
–1760.0 to 1760.0 °C
–1820.0 to 1820.0 °C
–1570.0 to 1570.0 °C
–1000.0 to 1000.0 °C
–1300.0 to 1300.0 °C
–600.0 to 600.0 °C
–1300.0 to 1300.0 °C
–1999.9 to 2315.0 °C
–1100.0 to 1100.0 °C
–600.0 to 600.0 °C
–1999.9 to 2400.0 °C
–3168 to 3168 °F
–3168 to 3168 °F
–3276 to 3276 °F
–2826 to 2826 °F
–1800 to 1800 °F
–1999.9 to 2340.0 °F
–1080.0 to 1080.0 °F
–2340 to 2340 °F
–4167 to 4167 °F
–1980 to 1980 °F
–1080.0 to 1080.0 °F
–4320 to 4320 °F
RTD
–800.0 to 800.0 °C
–750.0 to 750.0 °C
–1440.0 to 1440.0 °F
–1350.0 to 1350.0 °F
PT (Pt100)
JPT (JPt100)
IM 04P01B01-01E
5.1 Setting the Input range
ON/OFF Input
Set=Range
Set the channel range.
CH=01-01
First channel
MOde=DI
Select DI.
Range=Level
Select Cont or Level.
Span_L=0
Set the left span value.
Span_R=1
Set the right span value.
01-01 Channel
Setting complete
The new setting takes effect.
5
Description
• Selectable Range Type, Span Left, and Span Right
The range type, span left, and span right can be set in the range shown below. Span
left and span right cannot be set to the same value.
Range Type
IM 04P01B01-01E
Selectable Span Values
Level
0: Less than 2.4 V, 1: 2.4 V or higher, within ±6 V
Contact
0: Open, 1: Closed
5-5
Frequently Used Setup Operations (Setting Mode)
ESC/?
Last channel
5.1 Setting the Input range
1-5V Input
Set=Range
Set the channel range.
CH=01-01
First channel Last channel
Select 1-5V.
Mode=1-5V
Span_L=
1,000
Set the left span value.
Span_R=
5,000
Set the right span value.
Scale_L=
0.00
Scale_R=
200.00
(Select the right scale value.)
Low-cut=Off
ESC/?
01-01 Channel
Setting complete
Set the left scale value.
Set the right scale value.
Turn On or Off the low-cut function.
The new setting takes effect.
Description
• Span Left and Span Right
Selectable span left range: 0.8 to 1.2 V
Selectable span right range: 4.8 to 5.2 V
• Scale Left and Scale Right
Selectable range (mantissa): –20000 to 30000
Decimal position: XXXXX, XXXX.X, XXX.XX, XX.XXX, X.XXXX
The decimal position is the position specified by scale left.
Make sure that scale left is less than scale right.
• Low-cut
On: Cuts values below 0% of the recording span (set to 0% (scale left value)).
Off: Does not apply the low-cut function.
Note
•
•
5-6
The displayable/printable range of scale values is –19999 to 30000 excluding the decimal.
If the leftmost value of the scale is set to –20000 (excluding the decimal), the value when the
low-cut function is applied is displayed as “–Over.”
IM 04P01B01-01E
5.1 Setting the Input range
Square Root Computation
Set=Range
Set the channel range.
CH=01-01
First channel Last channel
Select SQRT.
Range=2V
Select the range type of DC voltage.
20 mV, 60 mV, 200 mV, 2 V, 6 V, 20 V,
or 50 V
Span_L=
-2.000
Set the left span value.
Span_R=
2.000
Scale_L=
0.00
Scale_R=
200.00
Set the right span value.
Set the left scale value.
Set the right scale value.
Low-cut=Off
Turn On or Off the low-cut function.
Low-cut
Set the low-cut point.
point=0.0
5
Frequently Used Setup Operations (Setting Mode)
Mode=SQRT
0.0 to 5.0% of the recording span, 0.1 steps.
ESC/?
01-01 Channel
Setting complete
The new setting takes effect.
Description
• Input Type in Which Square Root Computation Can Be Performed
Square root computation can be performed when set to DC voltage input.
• Span Left and Span Right
Same as the selectable span range of the DC voltage input.
Make sure that span left is less than span right.
• Scale Left and Scale Right
Selectable range (mantissa): –20000 to 30000
Decimal position: XXXXX, XXXX.X, XXX.XX, XX.XXX, X.XXXX
The decimal position is the position specified by scale left.
Make sure that scale left is less than scale right.
• Low-cut and Low-cut Point
If low-cut is set to On, set the low-cut point.
Selectable range of the low-cut point: 0.0 to 5.0% of the recording span
Values below the low-cut point is set to 0% (scale left value) of the recording span.
IM 04P01B01-01E
5-7
5.1 Setting the Input range
Note
•
•
The displayable/printable range of scale values is –19999 to 30000 excluding the decimal.
If the leftmost value of the scale is set to –20000 (excluding the decimal) and the low-cut
value is set to –20000, the value when the low-cut function is applied is displayed as “–Over.”
Skip (Unused Channels)
Set=Range
CH=01-01
Set the channel range.
First channel Last channel
ESC/?
Mode=Skip
Select Skip.
01-01 Channel
Setting complete
The new setting takes effect.
Note
Do not use channels set to skip as a reference channel of delta computations or in computing
equations of the computation function (/M1 option).
5-8
IM 04P01B01-01E
5.2
Setting the Alarm
The alarm is set for each channel.
If you change the input range, set the alarm again.
Procedure
1.
2.
3.
4.
Set=Alarm
Set the channel range.
CH=01-01
First channel Last channel
Level=1
Select the level number between 1 and 4.
Alarm=Off
Select On to set the alarm.
When set to On, the settings below can
be entered.
Type=H
Select the alarm type.
Value=
ESC/?
IM 04P01B01-01E
1.000
Select the value used to turn On the alarm.
Relay=Off
Turn relay On when outputting alarms on
the relay. This setting is valid only on
models with the /A1, /A2, or /A3 option.
Relay_No.=I01
Specify the alarm relay No.
01-01 CH/level 1
Setting complete
The new setting takes effect.
5-9
5
Frequently Used Setup Operations (Setting Mode)
5.
Hold down the MENU key for 3 seconds to enter Setting mode.
Press the
key or SHIFT +
key to select Alarm and then press the
key.
Set each item and press the
key.
Use the
key or SHIFT +
key to select values.
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up
to then are cancelled, and the display returns to a higher level menu.
When the Setting complete screen appears, do either of the following:
Press the
key to set other channels.
If you are done, press the ESC key.
Hold down the MENU key for 3 seconds to return to Operation mode.
5.2 Setting the Alarm
Explanation
Channel Range
The only case when an alarm can be set on multiple channels simultaneously is when
the channels are set to the same range type (for example channel 01 and 02 are set to 2
V range). For channels on which scaling is set, the channels must be set to the same
range type, same span values, and same scaling values.
Alarm Type
Symbol
Name
H
High limit alarm
L
Low limit alarm
h
Difference high limit alarm
Can be specified on channels set to
delta computation.
l
Difference low limit alarm
Can be specified on channels set to
delta computation.
R
High limit on rate-of-change alarm
r
Low limit on rate-of-change alarm
T
Delay high limit alarm*
t
Delay low limit alarm*
*
Note
Selectable only when the alarm delay function is enabled. See section 7.15.
Note
On channels set to delta computation, alarms can be detected on the values illustrated in the
figure below.
Alarm on a channel set to delta computation
Alarm
H, L, R, r, T, and t
Input value
Alarm
h and l
Measured value
–
Measured value on the reference channel
If high limit on rate-of-change alarm or low limit on rate-of-change alarm is specified, set
the interval over which the rate-of-change is calculated.
If delay high limit alarm or delay low limit alarm is specified, set the delay time.
<Related Topics>
Setting the interval of the high limit on rate-of-change alarm or low limit on rate-ofchange alarm: Section 7.1
Setting the delay time of delay high limit alarm or delay low limit alarm: Section 6.10
Enabling the alarm delay function: Section 7.15
Setting the alarm hysteresis: Section 7.1
Alarm Value
• High Limit Alarm/Low Limit Alarm and Delay High Limit Alarm/Delay Low Limit
Alarm
The following values can be specified.
• For DC voltage, TC, and RTD input: Values within the measurable range in the
specified range (example: –2.000 to 2.000 V for the 2 V range).
• For ON/OFF input (DI): 0 or 1.
• For linear scaling (1-5V, scaling, and square root): A value within –5 to 105% of the
scaling span except within the range of –20000 to 30000 (excluding the decimal
point).
5-10
IM 04P01B01-01E
5.2 Setting the Alarm
Note
For channels whose the leftmost value of the scale is set to –20000 (excluding the decimal),
the alarm value of –20000 is invalid. Set a value that is greater than or equal –19999.
• Difference High Limit Alarm/Difference Low Limit Alarm
Values in the measurable range can be specified. Measurable range refers to
“Selectable Span Range” in the table on page 5-4.
• High Limit on Rate-of-Change Alarm/Low Limit on Rate-of-Change Alarm
A value greater than or equal to 1 digit can be specified. For example, 1 digit
corresponds 0.001 for the 2 V range (measurable range: –2.000 to 2.000 V). The
maximum value that can be specified is the width of the measurable range (4.000 V
for 2 V range).
For ON/OFF input, only “1” can be specified.
Relay No.
The selectable relay numbers are listed below.
I01 and I02 on the /A1 option
I01, I02, I03, and I04 on the /A2 option
I01, I02, I03, I04, I05, and I06 on the /A3 option
Frequently Used Setup Operations (Setting Mode)
IM 04P01B01-01E
5
5-11
5.3
Setting the Unit on Scaled Channels
Units are set on channels whose input range is set to Scaling, 1-5V, or SQRT.
Procedure
1.
2.
3.
4.
5.
Hold down the MENU key for 3 seconds to enter Setting mode.
Press the
key or SHIFT +
key to select Unit and then press the
key.
Set each item and press the
key.
Use the
key or SHIFT +
key to select values.
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up
to then are cancelled, and the display returns to a higher level menu.
When the Setting complete screen appears, do either of the following:
Press the
key to set other channels.
If you are done, press the ESC key.
Hold down the MENU key for 3 seconds to return to Operation mode.
Set=Unit
CH=01-01
First channel
ESC/?
Set the channel range.
Last channel
Unit=
Set the unit.
01-01 Channel
Setting complete
The new setting takes effect.
Explanation
Characters That Can Be Used for Units
A unit is set using up to six characters.
The available characters are as follows:
Alphabet, numbers, symbols (%, #, °, @, +, –, *, /, (, ), µ, Ω, 2, 3, .), and space
5-12
IM 04P01B01-01E
5.4
Changing the Chart Speed
This section explains the details of changing the chart speed.
Procedure
1.
2.
3.
4.
Set=Chart
mm/h=
ESC/?
Set the chart speed.
20
Chart speed
Setting complete
The new setting takes effect.
Explanation
Chart Speed
• Pen Model
The chart speed can be selected from 82 settings shown below.
Chart speed on the pen model (unit: mm/h)
5
24
60
135
320
720
1500
3600
10800
6
25
64
150
360
750
1600
4000
12000
8
30
72
160
375
800
1800
4320
9
32
75
180
400
900
2000
4500
10
36
80
200
450
960
2160
4800
12
40
90
225
480
1000
2250
5400
15
45
96
240
500
1080
2400
6000
16
48
100
250
540
1200
2700
7200
18
50
120
270
600
1350
2880
8000
20
54
125
300
675
1440
3000
9000
• Dot Model
The chart speed can be set in the range of 1 to 1500 mm/h in 1 mm steps.
IM 04P01B01-01E
5-13
5
Frequently Used Setup Operations (Setting Mode)
5.
Hold down the MENU key for 3 seconds to enter Setting mode.
Press the
key or SHIFT +
key to select Chart and then press the
key.
Set the chart speed and press the
key.
Use the
key or SHIFT +
key to select values.
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up
to then are cancelled, and the display returns to a higher level menu.
When the Setting complete screen appears, do either of the following:
To correct the setting, press the
key.
If you are done, press the ESC key.
Hold down the MENU key for 3 seconds to return to Operation mode.
5.5
Setting the Date/Time
This section explains the details of setting the date/time.
Procedure
1.
2.
3.
4.
5.
Hold down the MENU key for 3 seconds to enter Setting mode.
Press the
key or SHIFT +
key to select Clock and then press the
key.
Set the date and time and press the
key.
Use the
key or SHIFT +
key to select values.
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up
to then are cancelled, and the display returns to a higher level menu.
When the Setting complete screen appears, do either of the following:
To correct the setting, press the
key.
If you are done, press the ESC key.
Hold down the MENU key for 3 seconds to return to Operation mode.
Set=Clock
04/02/17
year
ESC/?
month
Clock
Setting
10:39:47
Set the time.
day
The new setting takes effect.
complete
Explanation
Date/Time
The year is set with the lower two digits.
Values in the range 80 and 99 correspond to 1980 to 1999.
Recorders with Version 1.02 or Earlier
The following is the setting menu.
5-14
04/02/17 10:39:47
Set the time.
DST=Summer
Select Summer or Winter, whichever the set time belongs to.
IM 04P01B01-01E
Chapter 6 Setup Operations for Convenient Functions (Setting Mode)
6.1 Setting the Trend Recording Interval (Dot
Model).
This section explains the details of setting the trend recording interval (dot printing
interval) on the dot model.
Procedure
1.
2.
3.
4.
5.
6.
Hold down the MENU key for 3 seconds to enter Setting mode.
Press the
key or SHIFT +
key to select Aux and then press the
key.
Press the
key or SHIFT +
key to select Trend and then press the
key.
Set the dot printing interval and press the
key.
Use the
key or SHIFT +
key to select values.
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up
to then are cancelled, and the display returns to a higher level menu.
When the Setting complete screen appears, do either of the following:
To correct the setting, press the
key.
If you are done, press the ESC key.
Hold down the MENU key for 3 seconds to return to Operation mode.
Aux=Trend
ESC/?
Mode=Auto
Select the trend recording interval
mode.
Trend Interval
Setting complete
The new setting takes effect.
Explanation
Mode
Auto: The trend recording interval is set according to the chart speed in the range of 10
s to 90 s so that the dots do not overlap. The interval is displayed on the System
display (see page 12-9).
Fix:
The trend recording interval is fixed to 10 s/6 dots. On models with the
computation function (/M1 option), the dot printing interval varies depending on
the number of measurement and computation channels to be trend recorded.
Number of channels
to be recorded
6 or less
7 to 12
13 to 18
Trend recording
interval
10 s
15 s
20 s
Dot
When set to Auto
IM 04P01B01-01E
When set to Fix
6-1
Setup Operations for Convenient Functions (Setting Mode)
Set=Aux
6
6.2
Setting the Filter (Pen Model)
This section explains the details of setting a filter on the measurement channels on the
pen model.
Procedure
1.
2.
3.
4.
5.
6.
Hold down the MENU key for 3 seconds to enter Setting mode.
Press the
key or SHIFT +
key to select Aux and then press the
key.
Press the
key or SHIFT +
key to select Filter and then press the
key.
Set each item and press the
key.
Use the
key or SHIFT +
key to select values.
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up
to then are cancelled, and the display returns to a higher level menu.
When the Setting complete screen appears, do either of the following:
Press the
key to set other channels.
If you are done, press the ESC key.
Hold down the MENU key for 3 seconds to return to Operation mode.
Set=Aux
Aux=Filter
CH=01-01
First channel
ESC/?
Set the channel range.
Last channel
Resp.Time=Off
Set the filter time constant.
01-01 Channel
Setting complete
The new setting takes effect.
Explanation
Filter Time Constant
The time constant is selectable from 2 s, 5 s, and 10 s. Select Off when not using the
filter.
<Related Topics>
Enabling the filter function: Section 7.12
6-2
IM 04P01B01-01E
6.3
Setting the Moving Average (Dot Model)
This section explains the details of setting the moving average function on measurement
channels on the dot model.
Procedure
1.
2.
3.
4.
5.
Set=Aux
Aux=Moving_AVE
CH=01-01
Set the channel range.
First channel Last channel
No.of
ESC/?
samples=Off
01-01 Channel
Setting complete
Set the number of samples of the
moving average. Moving average is
not performed when Off is selected.
The new setting takes effect.
Explanation
The Number of Samples of Moving Average
The number of samples can be set to a integer between 2 and 16. Select Off when not
using the moving average function.
<Related Topics>
Enabling the moving average function: Section 7.11
IM 04P01B01-01E
6-3
6
Setup Operations for Convenient Functions (Setting Mode)
6.
Hold down the MENU key for 3 seconds to enter Setting mode.
Press the
key or SHIFT +
key to select Aux and then press the
key.
Press the
key or SHIFT +
key to select Moving_AVE and then press
the
key.
Set each item and press the
key.
Use the
key or SHIFT +
key to select values.
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up
to then are cancelled, and the display returns to a higher level menu.
When the Setting complete screen appears, do either of the following:
Press the
key to set other channels.
If you are done, press the ESC key.
Hold down the MENU key for 3 seconds to return to Operation mode.
6.4
Setting Recording Zones for Each Channel
(Zone Recording)
This section explains the details of setting the recording zone for each measurement
channel.
Procedure
Hold down the MENU key for 3 seconds to enter Setting mode.
Press the
key or SHIFT +
key to select Aux and then press the
key.
key or SHIFT +
key to select Zone and then press the
Press the
key.
Set each item and press the
key.
key or SHIFT +
key to select values.
Use the
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up
to then are cancelled, and the display returns to a higher level menu.
When the Setting complete screen appears, do either of the following:
Press the
key to set other channels.
If you are done, press the ESC key.
Hold down the MENU key for 3 seconds to return to Operation mode.
1.
2.
3.
4.
5.
6.
Set=Aux
Aux=Zone
CH=01-01
Set the channel range.
First channel Last channel
Left=
ESC/?
0
Set the leftmost value of the
recording zone.
Right=100
Set the rightmost value of the
recording zone.
01-01 Channel
Setting complete
The new setting takes effect.
Explanation
Zone
The recording zone can be set in the following range.
Leftmost position: 0 mm to 95 mm
Rightmost position: 5 mm to 100 mm
Set a larger value for the rightmost position than the leftmost position.
Set at least 5 mm for the zone width.
Note
If the zone is set to a value less than 40 mm, scale printout of periodic printout cannot be
executed.
6-4
IM 04P01B01-01E
6.5
Setting the Partial Expanded Recording
This section explains the details of expanding a section of the recording range.
If you change the input range, set the partial expanded recording again.
Procedure
1.
2.
3.
4.
5.
Set=Aux
Aux=Partial
Set the channel range.
CH=01-01
First channel Last channel
Partial=Off
Select On to use the partial expanded
recording function.
Expand=50
Set the boundary position as a
percentage where the recording span
is assumed to be 100%.
Boundary=
ESC/?
0.000
01-01 Channel
Setting complete
Set the boundary value within the
recording span.
The new setting takes effect.
Explanation
Expand
Set a value between 1 and 99.
<Related Topics>
Enabling the partial expanded recording function: Section 7.13
IM 04P01B01-01E
6-5
6
Setup Operations for Convenient Functions (Setting Mode)
6.
Hold down the MENU key for 3 seconds to enter Setting mode.
Press the
key or SHIFT +
key to select Aux and then press the
key.
Press the
key or SHIFT +
key to select Partial and then press the
key.
Set each item and press the
key.
Use the
key or SHIFT +
key to select values.
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up
to then are cancelled, and the display returns to a higher level menu.
When the Setting complete screen appears, do either of the following:
Press the
key to set other channels.
If you are done, press the ESC key.
Hold down the MENU key for 3 seconds to return to Operation mode.
6.6
Turning Trend Recording (Dot Model) and
Periodic Printout ON/OFF for Each Channel
This section explains the details of setting the trend recording and periodic printout for
each measurement channel.
The following settings can be entered for each channel on the dot model.
• Turn trend recording ON/OFF.
• Turn the printing of measured values ON/OFF during periodic printout.
The following setting can be entered for each channel on the pen model.
• Turn the printing of measured values ON/OFF during periodic printout.
Procedure
1.
2.
3.
4.
5.
6.
Hold down the MENU key for 3 seconds to enter Setting mode.
Press the
key or SHIFT +
key to select Aux and then press the
key.
Press the
key or SHIFT +
key to select Print out and then press the
key.
key.
Set each item and press the
Use the
key or SHIFT +
key to select values.
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up
to then are cancelled, and the display returns to a higher level menu.
When the Setting complete screen appears, do either of the following:
Press the
key to set other channels.
If you are done, press the ESC key.
Hold down the MENU key for 3 seconds to return to Operation mode.
Set=Aux
Aux=Print
out
CH=01-01
Set the channel range.
First channel Last channel
ESC/?
Trend=On
Turn trend recording On/Off.
Periodic=On
Select whether to print the measured
values during periodic printout (On/Off).
01-01 Channel
Setting complete
The new setting takes effect.
Explanation
Trend Recording (Dot Model)
On: Turns On the trend recording of the specified channel.
Periodic Printout
On: Turns On the printing of the measured values of the specified channel during
periodic printout.
6-6
IM 04P01B01-01E
6.7
Setting Tags on Channels
This section explains the details of setting tags to measurement channels.
Procedure
1.
2.
3.
4.
5.
6.
Hold down the MENU key for 3 seconds to enter Setting mode.
Press the
key or SHIFT +
key to select Aux and then press the
key.
Press the
key or SHIFT +
key to select Tag and then press the
key.
Set each item and press the
key.
Use the
key or SHIFT +
key to select values.
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up
to then are cancelled, and the display returns to a higher level menu.
When the Setting complete screen appears, do either of the following:
Press the
key to set other channels.
If you are done, press the ESC key.
Hold down the MENU key for 3 seconds to return to Operation mode.
Setup Operations for Convenient Functions (Setting Mode)
Set=Aux
Aux=Tag
CH=01-01
Set the channel range.
First channel Last channel
ESC/?
Tag=
Set the tag.
01-01 Channel
Setting complete
The new setting takes effect.
Explanation
Characters That Can Be Used for Tags
Set the tag using up to 7 characters.
The available characters are as follows:
Alphabet, numbers, symbols (%, #, °, @, +, –, *, /, (, ), µ, Ω, 2, 3, .), and space
<Related Topics>
Printing tags in place of channel numbers: Section 7.7
IM 04P01B01-01E
6
6-7
6.8
Setting the Message String
This section explains the details of setting the message strings to be printed on the chart
paper. Up to five message strings can be registered.
Procedure
1.
2.
3.
4.
5.
6.
Hold down the MENU key for 3 seconds to enter Setting mode.
Press the
key or SHIFT +
key to select Aux and then press the
key.
Press the
key or SHIFT +
key to select Message and then press the
key.
Set each item and press the
key.
Use the
key or SHIFT +
key to select values.
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up
to then are cancelled, and the display returns to a higher level menu.
When the Setting complete screen appears, do either of the following:
To correct the setting, press the
key.
If you are done, press the ESC key.
Hold down the MENU key for 3 seconds to return to Operation mode.
Set=Aux
Aux=Message
Message
No.=1
Select the message No.
Set message string.
CHR:A-Z
ESC/?
Message
Setting
1
complete
The new setting takes effect.
Explanation
Characters That Can Be Used for Messages
Set the message using up to 16 characters.
The available characters are as follows:
Alphabet, numbers, symbols (%, #, °, @, +, –, *, /, (, ), µ, Ω, 2, 3, .), and space
<Related Topics>
Printing messages: Section 3.10
6-8
IM 04P01B01-01E
6.9
Setting the Secondary Chart Speed (Remote
Control Function, /R1)
This section explains the details of setting the secondary chart speed when the chart
speed is to be switched using the remote control function (/R1 option).
Procedure
1.
2.
3.
4.
5.
Set=Aux
Aux=Chart2
mm/h=
ESC/?
20
Chart speed2
Setting complete
Set the chart speed.
The new setting takes effect.
Explanation
Chart Speed
See section 5.4.
<Related Topics>
Setting the remote control function: Section 7.18
IM 04P01B01-01E
6-9
6
Setup Operations for Convenient Functions (Setting Mode)
6.
Hold down the MENU key for 3 seconds to enter Setting mode.
Press the
key or SHIFT +
key to select Aux and then press the
key.
key or SHIFT +
key to select Chart2 and then press the
Press the
key.
Set each item and press the
key.
key or SHIFT +
key to select values.
Use the
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up
to then are cancelled, and the display returns to a higher level menu.
When the Setting complete screen appears, do either of the following:
To correct the setting, press the
key.
If you are done, press the ESC key.
Hold down the MENU key for 3 seconds to return to Operation mode.
6.10 Setting the Alarm Delay Duration
This section explains the details of setting the duration of the delay alarm.
Procedure
1.
2.
3.
4.
5.
6.
Hold down the MENU key for 3 seconds to enter Setting mode.
Press the
key or SHIFT +
key to select Aux and then press the
key.
Press the
key or SHIFT +
key to select Alarm delay and then press
the
key.
Set each item and press the
key.
Use the
key or SHIFT +
key to select values.
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up
to then are cancelled, and the display returns to a higher level menu.
When the Setting complete screen appears, do either of the following:
Press the
key to set other channels.
If you are done, press the ESC key.
Hold down the MENU key for 3 seconds to return to Operation mode.
Set=Aux
Aux=Alm
delay
T
Set the channel range.
CH=01-01
First channel Last channel
Duration=
ESC/?
10
01-01 Channel
Setting complete
Set the alarm delay time.
The new setting takes effect.
Explanation
Duration
The duration can be set to a value in the range of 1 to 3600 s.
However, if the scan interval on the dot model is 2.5 s, the actual duration is an integer
multiple of 2.5 s. If the set value is not equal to an integer multiple of 2.5 s, the duration
is the smallest integer multiple of 2.5 s greater than the set value.
Example: The actual duration for 1 s and 2 s is 2.5 s.
<Related Topics>
Enabling the alarm delay function: Section 7.15
6-10
IM 04P01B01-01E
6.11 Setting the Brightness of the Display and
Internal Light
This section explains the details of setting the brightness of the display and the internal
light.
Procedure
1.
2.
3.
4.
5.
Set=Aux
Aux=Brightness
ESC/?
Display=4
Select the brightness of the display.
Light=Off
Select the brightness of the internal
light.
Brightness
Setting complete
The new setting takes effect.
Explanation
Display Brightness
The brightness can be set to an integer between 1 and 8. The darkest setting is 1; the
brightest setting is 8.
Internal Light Brightness
The brightness of the internal light can be set to an integer between 1 and 4. The
darkest setting is 1; the brightest setting is 4. Select Off to turn Off the internal light.
IM 04P01B01-01E
6-11
6
Setup Operations for Convenient Functions (Setting Mode)
6.
Hold down the MENU key for 3 seconds to enter Setting mode.
Press the
key or SHIFT +
key to select Aux and then press the
key.
key or SHIFT +
key to select Brightness and then press the
Press the
key.
Set each item and press the
key.
key or SHIFT +
key to select values.
Use the
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up
to then are cancelled, and the display returns to a higher level menu.
When the Setting complete screen appears, do either of the following:
To correct the setting, press the
key.
If you are done, press the ESC key.
Hold down the MENU key for 3 seconds to return to Operation mode.
6.12 Applying a Bias on the Measuring Input Signal
This section explains the details of setting a bias on the measuring input signal.
If you change the input range, set the bias again.
Procedure
1.
2.
3.
4.
5.
Hold down the MENU key for 3 seconds to enter Setting mode.
Press the
key or SHIFT +
key to select Bias and then press the
key.
Set each item and press the
key.
key or SHIFT +
key to select values.
Use the
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up
to then are cancelled, and the display returns to a higher level menu.
When the Setting complete screen appears, do either of the following:
Press the
key to set other channels.
If you are done, press the ESC key.
Hold down the MENU key for 3 seconds to return to Operation mode.
Set=Bias
CH=01-01
First channel
ESC/?
Set the channel range.
Last channel
Bias=Off
Select whether to use the bias.
(When bias is On)
Bias=
0.000
Set the bias value.
01-01 Channel
Setting complete
The new setting takes effect.
Explanation
Channel Range
The only case when a bias can be set on multiple channels simultaneously is when the
channels are set to the same range type (for example channel 01 and 02 are set to 2 V
range). For channels on which scaling is set, the channels must be set to the same
range type, same span values, and same scaling values.
Bias
The range of bias that can be specified is ±10% of the span of the measurable range at the
specified range type. For example, the range is –0.4 V to 0.4 V for the range type 2 V.
For channels on which scaling is set, the range is ±10% of the scaling span.
Note
A bias cannot be set on channels set to ON/OFF input (DI).
<Related Topics>
Enabling the bias function: Section 7.15
6-12
IM 04P01B01-01E
6.13 Setting the Date/Time When Switching between
Standard Time and DST
This section explains the details of setting the date/time when switching from the
standard time to DST and the date/time when switching back from DST to standard time
if the recorder is used in a region that has DST. When the preset date/time arrives, the
recorder internal clock automatically switches.
Procedure
1.
2.
3.
4.
5.
IM 04P01B01-01E
6-13
6
Setup Operations for Convenient Functions (Setting Mode)
6.
Hold down the MENU key for 3 seconds to enter Setting mode.
Press the
key or SHIFT +
key to select Aux and then press the
key.
Press the
key or SHIFT +
key to select DST and then press the
key.
Set each item and press the
key.
key or SHIFT +
key to select values.
Press the
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up
to then are cancelled, and the display returns to a higher level menu.
When the Setting complete screen appears, do either of the following:
Press the
key to set other channels.
If you are done, press the ESC key.
Hold down the MENU key for 3 seconds to return to Operation mode.
6.13 Setting the Date/Time When Switching between Standard Time and DST
Set=Aux
Aux=DST
DST=Use
Select Use.
Start
Month, day of the week, and time
when the DST starts.
Month
Strt
month=Apr
day=1st-Sun
The n th day of the week
of the month
Strt
day=1st-Sun
Start
ESC/?
time=2:00
End
month=Oct
End
day=Last-Mon
End
day=Last-Mon
End
time=1:00
DST
Setting
Time
Month, day of the week, and time
when the DST ends.
Month
The n th day of the week
of the month
Time
The new setting takes effect.
complete
Explanation
DST
Select Use to use DST.
Start month, Strt day, and Start time
Set the date/time for switching from standard time to DST. Specify the day as the n th
day of the week of the month.
End month, End day, and End time
Set the date/time for switching from DST to standard time. Specify the day as the n th
day of the week of the month.
6-14
IM 04P01B01-01E
6.13 Setting the Date/Time When Switching between Standard Time and DST
Recorders with Version 1.02 or Earlier
The following is the setting menu.
Time=Summer
Select standard time (Winter) or DST (Summer).
Summer=Off
Select On to set the time to switch between standard time and DST.
Time=04/07/05 00
Set the time to switch.
DST
Setting complete
The new setting takes effect.
6
Setup Operations for Convenient Functions (Setting Mode)
IM 04P01B01-01E
6-15
Chapter 7 Setup Operations for Changing/Adding Functions (Basic Setting Mode)
7.1
Changing the Auxiliary Alarm Function
This section explains the details of setting the alarm system items listed below.
• Diagnosis output using the alarm output relay (I01)
• Reflash alarm operation of the alarm output relay (I01 to I03)
• AND/OR operation of alarm output relays
• Energized/De-energized operation of alarm output relays when alarms occur
• Hold/Non-hold operation of alarm output relays
• Hold/Non-hold operation of alarm indications
• The interval of the rate-of-change alarm
• Alarm hysteresis
Basic Setting mode cannot be entered when recording is in progress or when
computation is in progress on models with the computation function (/M1 option).
Procedure
IM 04P01B01-01E
7-1
7
Setup Operations for Changing/Adding Functions (Basic Setting Mode)
Changing the Settings
1.
Hold down the MENU key for 3 seconds to enter Setting mode.
2.
Hold down the
and
keys simultaneously for 3 seconds to enter Basic
Setting mode.
3.
Press the
key with Alarm shown on the screen.
4.
Set each item and press the
key.
Use the
key or SHIFT +
key to select values.
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up
to then are cancelled, and the display returns to a higher level menu.
5.
When the Setting complete screen appears, do either of the following:
To correct the setting, press the
key.
If you are done, press the ESC key.
7.1 Changing the Auxiliary Alarm Function
Basic=Alarm
Diagnosis=Off
Select On to set relay I01 to diagnosis
output.
Reflash=Off
Select On to set relays I01, I02, and I03
to reflash alarm operation.
AND=None
Select the range of relays that are to
operate using AND logic.
Act=Energize
Select the relay’s energized or
de-energized operation.
Behavior=Nonhold
Select the relay’s hold or non-hold
operation.
Indicator=Nonhold
Select the relay’s hold or non-hold
operation.
Increase=01
Set the interval of the high limit on
rate-of-change alarm.
Decrease=01
Set the interval of the low limit on
rate-of-change alarm.
Hysteresis=0.5%
Set the hysteresis of the measurement
channels.
(models with /M1)
ESC/?
M_Hysteresis=0.0%
Set the hysteresis of the computation
channels.
Alarm
Setting
The new setting takes effect.
complete
Applying the Changes and Returning to Operation Mode
Press the
key or SHIFT +
key to select End and then press the
key. Press
the
key to select Store and then press the
key. The changes are applied, and
the screen returns to Operation mode. If you select Abort and press the
key, the
changes are discarded, and the screen returns to Operation mode. Press the ESC key
to return to the Basic= screen.
Explanation
Diagnosis
When set to On, alarm output relay I01 becomes a dedicated diagnosis output relay, and
the operation is fixed to OR, de-energized, and non-hold. Diagnosis output is a function
which outputs a relay signal when an error is detected in the recording section, burnout
detection function, or the A/D converter.
7-2
IM 04P01B01-01E
7.1 Changing the Auxiliary Alarm Function
Reflash
Turns On/Off the reflash alarm operation of alarm output relays I01, I02, and I03.
When set to On, alarm output relays I01, I02, and I03 set to reflash alarm operation, and
the operation is fixed to OR, de-energized, and non-hold.
AND
Sets the AND/OR operation of alarm output relays.
None: No relays are set to AND operation. All relays are set to OR operation.
I01: Only I01 is set to AND operation.
I01-I02: I01 and I02 are set to AND operation.
I01-I03: I01 to I03 are set to AND operation.
I01-I04: I01 to I04 are set to AND operation.
I01-I05: I01 to I05 are set to AND operation.
I01-I06: I01 to I06 are set to AND operation.
Act
Sets whether the output relay is energized or de-energized when an alarm occurs.
Energize:
Energizes the relay when an alarm occurs. The relay is de-energized
during normal operation.
De_energize: De-energizes the relay when an alarm occurs. The relay is energized
during normal operation.
Indicator
Specifies whether to set the alarm indication to hold operation or non-hold operation.
Nonhold:
Releases the alarm indication at the same time the alarm is released.
Hold:
Holds the alarm indication until the alarm ACK operation is executed.
Increase
Set the interval of the high limit on rate-of-change alarm to an integer between 1 and 15.
The interval is set to scan interval × (1 to 15).
Pen model: The scan interval is 125 ms. The maximum interval is 1.875 s.
Dot model: The maximum interval when the scan interval set to 1 s is 15 s. The
maximum interval when the scan interval set to 2.5 s is 37.5 s.
Decrease
Set the interval of the low limit on rate-of-change alarm to an integer between 1 and 15.
The setup procedure is the same as Increase.
Hysteresis
Sets the alarm hysteresis of the measurement channels.
The hysteresis width can be set in the range of 0.0% (Off) to 1.0% of the recording span
in 0.1 steps.
The hysteresis applies to all high limit and low limit alarms of measurement channels.
M_Hysteresis (Models with the Computation Function (/M1 Option))
Sets the alarm hysteresis of the computation channels.
The hysteresis width can be set in the range of 0.0% (Off) to 1.0% of the recording span
in 0.1 steps.
The hysteresis applies to all high limit and low limit alarms of computation channels.
IM 04P01B01-01E
7-3
7
Setup Operations for Changing/Adding Functions (Basic Setting Mode)
Behavior
Specifies whether to set the alarm output relay to hold operation or non-hold operation.
Nonhold:
Releases the relay output at the same time the alarm is released.
Hold:
Holds the relay output until the alarm ACK operation is executed.
7.2
Changing the Integration Time of the A/D
Converter
This section explains the details of setting the integration time of the A/D converter.
Basic Setting mode cannot be entered when recording is in progress or when
computation is in progress on models with the computation function (/M1 option).
Procedure
Changing the Settings
1.
Hold down the MENU key for 3 seconds to enter Setting mode.
and
keys simultaneously for 3 seconds to enter Basic
2.
Hold down the
Setting mode.
3.
Press the
key or SHIFT +
key to select A/D and then press the
key.
4.
Set each item and press the
key.
Use the
key or SHIFT +
key to select values.
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up
to then are cancelled, and the display returns to a higher level menu.
5.
When the Setting complete screen appears, do either of the following:
To correct the setting, press the
key.
If you are done, press the ESC key.
Basic=A/D
ESC/?
Integrate=Auto
Select the integration time of the
A/D converter.
A/D Integrate
Setting complete
The new setting takes effect.
Applying the Changes and Returning to Operation Mode
Press the
key or SHIFT +
key to select End and then press the
key. Press
the
key to select Store and then press the
key. The changes are applied, and
the screen returns to Operation mode. If you select Abort and press the
key, the
changes are discarded, and the screen returns to Operation mode. Press the ESC key
to return to the Basic= screen.
Explanation
Integrate
Selects the integration time of the A/D converter.
50Hz: Sets the integration time to 20 ms.*
60Hz: Sets the integration time to 16.7 ms.*
Auto: Sets the integration time (20 ms or 16.7 ms) to a value synchronized with the
power supply frequency.*
100ms (dot model): Sets the integration time to 100 ms. The scan interval is set to
2.5 s.
*
7-4
The scan interval is 125 ms (pen model) and 1 s (dot model) when set to 50Hz, 60Hz, or
Auto.
IM 04P01B01-01E
7.3
Setting the Burnout Detection Function of
Thermocouples
This section explains the details of setting the burnout detection function of
thermocouples to channels set to 1-5V or TC input.
Basic Setting mode cannot be entered when recording is in progress or when
computation is in progress on models with the computation function (/M1 option).
Procedure
Changing the Settings
1.
Hold down the MENU key for 3 seconds to enter Setting mode.
2.
Hold down the
and
keys simultaneously for 3 seconds to enter Basic
Setting mode.
3.
Press the
key or SHIFT +
key to select Burnout and then press the
key.
4.
Set each item and press the
key.
key or SHIFT +
key to select values.
Use the
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up
to then are cancelled, and the display returns to a higher level menu.
5.
When the Setting complete screen appears, do either of the following:
Press the
key to set other channels.
If you are done, press the ESC key.
CH=01-01
Set the channel range.
First channel Last channel
ESC/?
Burnout=Off
Select the recording operation that is
carried out when a burnout is detected.
01-01 Channel
Setting complete
The new setting takes effect.
Applying the Changes and Returning to Operation Mode
Press the
key or SHIFT +
key to select End and then press the
key. Press
the
key to select Store and then press the
key. The changes are applied, and
the screen returns to Operation mode. If you select Abort and press the
key, the
changes are discarded, and the screen returns to Operation mode. Press the ESC key
to return to the Basic= screen.
Explanation
Burnout
Selects the action taken when a burnout is detected at the 1-5V input or the
thermocouple input.
Up:
Records off the scale on the 100% side.
Down: Records off the scale on the 0% side.
Select Off when not using the burnout detection function.
IM 04P01B01-01E
7-5
Setup Operations for Changing/Adding Functions (Basic Setting Mode)
Basic=Burnout
7
7.4
Setting the RJC Function on Channels Set to
TC Input
This section explains the details of setting RJC on channels set to TC input.
Basic Setting mode cannot be entered when recording is in progress or when
computation is in progress on models with the computation function (/M1 option).
Procedure
Changing the Settings
1.
Hold down the MENU key for 3 seconds to enter Setting mode.
and
keys simultaneously for 3 seconds to enter Basic
2.
Hold down the
Setting mode.
3.
Press the
key or SHIFT +
key to select RJC and then press the
key.
4.
Set each item and press the
key.
Use the
key or SHIFT +
key to select values.
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up
to then are cancelled, and the display returns to a higher level menu.
5.
When the Setting complete screen appears, do either of the following:
Press the
key to set other channels.
If you are done, press the ESC key.
Basic=RJC
Set the channel range.
CH=01-01
First channel Last channel
RJC=Internal
Select the RJC mode of the
thermocouple input.
(When RJC is set to External)
Volt=
ESC/?
0
01-01 Channel
Setting complete
Set the compensation voltage when
using the external RJC function.
The new setting takes effect.
Applying the Changes and Returning to Operation Mode
Press the
key or SHIFT +
key to select End and then press the
key. Press
the
key to select Store and then press the
key. The changes are applied, and
the screen returns to Operation mode. If you select Abort and press the
key, the
changes are discarded, and the screen returns to Operation mode. Press the ESC key
to return to the Basic= screen.
7-6
IM 04P01B01-01E
7.4 Setting the RJC Function on Channels Set to TC Input
Explanation
RJC
Sets the RJC mode.
Internal: Uses the RJC function on the recorder.
External: Uses an external RJC function. If external is selected, set the compensation
voltage.
Volt
Sets the compensation voltage when using an external RJC function.
The compensation voltage can be set in the range of –20000 µV to 20000 µV.
7
Setup Operations for Changing/Adding Functions (Basic Setting Mode)
IM 04P01B01-01E
7-7
7.5
Changing the Channel Recording Color (Dot
Model)
This section explains the details of changing the trend recording color on the dot model.
Basic Setting mode cannot be entered when recording is in progress or when
computation is in progress on models with the computation function (/M1 option).
Procedure
Changing the Settings
1.
Hold down the MENU key for 3 seconds to enter Setting mode.
and
keys simultaneously for 3 seconds to enter Basic
2.
Hold down the
Setting mode.
3.
Press the
key or SHIFT +
key to select Color and then press the
key.
4.
Set each item and press the
key.
Use the
key or SHIFT +
key to select values.
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up
to then are cancelled, and the display returns to a higher level menu.
5.
When the Setting complete screen appears, do either of the following:
Press the
key to set other channels.
If you are done, press the ESC key.
Basic=Color
ESC/?
Channel=01
Select the channel.
Color=Purple
Select the recording color.
01 Channel Color
Setting complete
The new setting takes effect.
Applying the Changes and Returning to Operation Mode
Press the
key or SHIFT +
key to select End and then press the
key. Press
the
key to select Store and then press the
key. The changes are applied, and
the screen returns to Operation mode. If you select Abort and press the
key, the
changes are discarded, and the screen returns to Operation mode. Press the ESC key
to return to the Basic= screen.
Explanation
Color
Sets the recording color of measurement channels.
By default, the recording colors are set as follows:
Channel01: Purple
Channel 02: Red
Channel 03: Green
Channel 04: Blue
Channel 05: Brown
Channel 06: Black
7-8
IM 04P01B01-01E
7.6
Recording by Compensating for the Pen Offset
along the Time Axis (Pen Model)
This section explains the details of setting the compensation for the pen offset along the
time axis (pen model).
Basic Setting mode cannot be entered when recording is in progress or when
computation is in progress on models with the computation function (/M1 option).
Procedure
Changing the Settings
1.
Hold down the MENU key for 3 seconds to enter Setting mode.
2.
Hold down the
and
keys simultaneously for 3 seconds to enter Basic
Setting mode.
3.
Press the
key or SHIFT +
key to select POC and then press the
key.
4.
Set each item and press the
key.
key or SHIFT +
key to select values.
Use the
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up
to then are cancelled, and the display returns to a higher level menu.
5.
When the Setting complete screen appears, do either of the following:
To correct the setting, press the
key.
If you are done, press the ESC key.
Select On to record by compensating
for the pen offset along the time axis.
POC=Off
ESC/?
POC
Setting
The new setting takes effect.
complete
Applying the Changes and Returning to Operation Mode
Press the
key or SHIFT +
key to select End and then press the
key. Press
the
key to select Store and then press the
key. The changes are applied, and
the screen returns to Operation mode. If you select Abort and press the
key, the
changes are discarded, and the screen returns to Operation mode. Press the ESC key
to return to the Basic= screen.
Explanation
POC (Pen Offset Compensation)
On: Enables pen offset compensation.
Off: Disables pen offset compensation.
Note
When pen offset compensation is enabled, all pens are aligned with the reference pen (pen
that is furthest to the back: pen 2 on the 2-pen model, pen 3 on the 3-pen model, and pen 4
on the 4-pen model) for recording. When recording is started, only the reference pen moves
until the offset is compensated. This is not a malfunction.
When pen offset compensation is enabled, an asterisk is printed by the time on the periodic
printout.
IM 04P01B01-01E
7-9
Setup Operations for Changing/Adding Functions (Basic Setting Mode)
Basic=POC
7
7.7
Turning Printouts ON/OFF.
This section explains the details of turning ON/OFF the printouts of various items.
Basic Setting mode cannot be entered when recording is in progress or when
computation is in progress on models with the computation function (/M1 option).
Procedure
Changing the Settings
1.
Hold down the MENU key for 3 seconds to enter Setting mode.
2.
Hold down the
and
keys simultaneously for 3 seconds to enter Basic
Setting mode.
3.
Press the
key or SHIFT +
key to select Print and then press the
key.
4.
Set each item and press the
key.
Use the
key or SHIFT +
key to select values.
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up
to then are cancelled, and the display returns to a higher level menu.
5.
When the Setting complete screen appears, do either of the following:
To correct the setting, press the
key.
If you are done, press the ESC key.
Basic=Print
CH/Tag=Channel
(Dot model)
Channel=On
Select whether to print the alarm
occurrence/release.
Record
Select On to enable recording start
printout.
On=Off
speed=On
Scale=On
(Pen model)
Pen color=On
7-10
Select whether to print the channel
number beside the trend recording.
Alarm=On1
Chart
ESC/?
Select whether to use the channel
number or tags for the printout.
Printout
Setting complete
Prints when the chart speed is changed.
Select On to print the channel scale in
periodic printout.
Select On to print the pen color in
periodic printout.
The new setting takes effect.
IM 04P01B01-01E
7.7 Turning Printouts ON/OFF.
Applying the Changes and Returning to Operation Mode
key or SHIFT +
key to select End and then press the
key. Press
Press the
the
key to select Store and then press the
key. The changes are applied, and
key, the
the screen returns to Operation mode. If you select Abort and press the
changes are discarded, and the screen returns to Operation mode. Press the ESC key
to return to the Basic= screen.
Explanation
CH/Tag
Channel: Prints the channel number.
Tag: Prints the tag.
Channel (Dot Model)
When On is selected channel No. is printed by the trend recording.
Alarm
On1: Prints the alarm information when an alarm occurs or releases.
On2: Prints the alarm information only when an alarm occurs.
Off:
Does not print alarm information.
Pen Color Printout
When On is selected, the recording color of the channel whose scale is printed is printed.
This is valid on channels whose scale printout is turned On.
7
Setup Operations for Changing/Adding Functions (Basic Setting Mode)
IM 04P01B01-01E
7-11
7.8
Setting the Periodic Printout Interval and the
Type of Measured Values to Be Printed
This section explains the details of setting the periodic printout interval and the type of
measured values to be printed.
Basic Setting mode cannot be entered when recording is in progress or when
computation is in progress on models with the computation function (/M1 option).
Procedure
Periodic Printout Interval
Changing the Settings
1.
Hold down the MENU key for 3 seconds to enter Setting mode.
2.
Hold down the
and
keys simultaneously for 3 seconds to enter Basic
Setting mode.
3.
Press the
key or SHIFT +
key to select Print1 and then press the
key.
key.
4.
Set each item and press the
Use the
key or SHIFT +
key to select values.
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up
to then are cancelled, and the display returns to a higher level menu.
5.
When the Setting complete screen appears, do either of the following:
To correct the setting, press the
key.
If you are done, press the ESC key.
Basic=Print1
Periodic=Auto
Select the periodic printout interval mode.
Ref.Time=00:00
Set the reference time.
(When Period is set to Manual)
ESC/?
Interval=1h
Select the periodic printout interval.
Mode=Inst
Set the type of measured values to print.
If report is selected, set the report data
type (next page). Select OFF to disable
periodic printout.
The new setting takes effect.
Printout1
Setting complete
Applying the Changes and Returning to Operation Mode
Press the
key or SHIFT +
key to select End and then press the
key. Press
the
key to select Store and then press the
key. The changes are applied, and
the screen returns to Operation mode. If you select Abort and press the
key, the
changes are discarded, and the screen returns to Operation mode. Press the ESC key
to return to the Basic= screen.
7-12
IM 04P01B01-01E
7.8 Setting the Periodic Printout Interval and the Type of Measured Values to Be Printed
Types of Report Data to Be Printed
Changing the Settings
1.
Hold down the MENU key for 3 seconds to enter Setting mode.
and
keys simultaneously for 3 seconds to enter Basic
2.
Hold down the
Setting mode.
3.
Press the
key or SHIFT +
key to select Print2 and then press the
key.
4.
Set each item and press the
key.
Use the
key or SHIFT +
key to select values.
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up
to then are cancelled, and the display returns to a higher level menu.
5.
When the Setting complete screen appears, do either of the following:
Press the
key to set other channels.
If you are done, press the ESC key.
Basic=Print2
CH=01-01
Set the channel range.
First channel Last channel
Select the type of report data to print.
SUM
Select the sum scale.
scale=Off
Printout2
Setting complete
7
The new setting takes effect.
Applying the Changes and Returning to Operation Mode
Press the
key or SHIFT +
key to select End and then press the
key. Press
the
key to select Store and then press the
key. The changes are applied, and
the screen returns to Operation mode. If you select Abort and press the
key, the
changes are discarded, and the screen returns to Operation mode. Press the ESC key
to return to the Basic= screen.
IM 04P01B01-01E
7-13
Setup Operations for Changing/Adding Functions (Basic Setting Mode)
ESC/?
Mode=AVE
7.8 Setting the Periodic Printout Interval and the Type of Measured Values to Be Printed
Explanation
Periodic Printout Interval
Periodic
Select the periodic printout interval mode.
Auto:
Automatically sets the printout interval in sync with the chart speed.
Manual: Set the printout interval manually.
Ref. Time
Sets the reference time for determining the times for executing the periodic printout.
The reference time is set in the range of 00 to 23 in 1 hour steps. Minutes cannot be
specified.
Interval
Select the interval from 10, 12, 15, 20, 30 minutes, 1, 2, 3, 4, 6, 8, 12, and 24 hours.
Mode
Sets the type of measured values to be printed.
Inst:
Prints the measured value at that point.
Report: Prints the report data over the interval. If Report is selected, set the type of
report data. Proceed to “Type of Report Data to Be Printed” below.
Off:
Disables periodic printout.
Types of Report Data to Be Printed
Mode
Sets the type of report data to be printed.
AVE: Prints the average value over the interval.
MIX: Prints the minimum, maximum, and average values over the interval.
SUM: Prints the sum value over the interval. If SUM is selected, set the sum scale.
MIN: Prints the minimum value over the interval.
MAX: Prints the maximum value over the interval.
INST: Prints the instantaneous value.
SUM scale
Set this item only when SUM is selected.
SUM sums the data every computation interval. For flow values that have units /s, /min,
or /h, a simple summation results in the actual value not matching the computed result,
because the scan interval and the unit of the input values are different. In such cases,
set the sum scale to match the unit of the input value. In effect, the sum value with the
same unit as that of the input value is calculated.
For example, if the scan interval is 1 s, and the input value is 100 m3/min, a simple
summation would add 100 every 1 s resulting in 6000 after one minute. However, if the
sum scale is set to /min, then 1 s/60 s is multiplied every scan interval before the value is
added giving a result with an m3/min unit.
Off: Simply sums the measured values.
/s:
Sums by converting the measured values to a value over 1 second.
/min: Sums by converting the measured values to a value over 1 minute.
/h:
Sums by converting the measured values to a value over 1 hour.
/day: Sums by converting the measured values to a value over 1 day.
7-14
IM 04P01B01-01E
7.9
Setting the Bar Graph Display Mode
This section explains the details of setting the bar graph display mode.
Basic Setting mode cannot be entered when recording is in progress or when
computation is in progress on models with the computation function (/M1 option).
Procedure
Changing the Settings
1.
Hold down the MENU key for 3 seconds to enter Setting mode.
2.
Hold down the
and
keys simultaneously for 3 seconds to enter Basic
Setting mode.
3.
Press the
key or SHIFT +
key to select Bar graph and then press the
key.
4.
Set each item and press the
key.
Use the
key or SHIFT +
key to select values.
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up
to then are cancelled, and the display returns to a higher level menu.
5.
When the Setting complete screen appears, do either of the following:
Press the
key to set other channels.
If you are done, press the ESC key.
Basic=Bar
7
graph
First channel
ESC/?
Last channel
Graph=Normal
Select the display mode of the bar graph.
01-01 Channel
Setting complete
The new setting takes effect.
Applying the Changes and Returning to Operation Mode
Press the
key or SHIFT +
key to select End and then press the
key. Press
the
key to select Store and then press the
key. The changes are applied, and
the screen returns to Operation mode. If you select Abort and press the
key, the
changes are discarded, and the screen returns to Operation mode. Press the ESC key
to return to the Basic= screen.
Explanation
Graph
Normal: Sets the base point of the bar graph to the left or right of the recording span
whichever is smaller in value.
Center: Sets the base point of the bar graph to the 50% position of the recording span.
Starting point of the bar graph
Standard
Starting point of the bar graph
Center-zero
IM 04P01B01-01E
7-15
Setup Operations for Changing/Adding Functions (Basic Setting Mode)
Set the channel range.
CH=01-01
7.10 Setting the Key Lock Function
This section explains the details of setting the keys that can be locked and the password
for releasing the key lock.
Basic Setting mode cannot be entered when recording is in progress or when
computation is in progress on models with the computation function (/M1 option).
Procedure
Changing the Settings
1.
Hold down the MENU key for 3 seconds to enter Setting mode.
2.
Hold down the
and
keys simultaneously for 3 seconds to enter Basic
Setting mode.
3.
Press the
key or SHIFT +
key to select Keylock and then press the
key.
4.
Set each item and press the
key.
Use the
key or SHIFT +
key to select values.
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up
to then are cancelled, and the display returns to a higher level menu.
5.
When the Setting complete screen appears, do either of the following:
To correct the setting, press the
key.
If you are done, press the ESC key.
7-16
IM 04P01B01-01E
7.10 Setting the Key Lock Function
Basic=Key
Lock
Keylock=Use
Select whether to use the key lock
function.
Password=
Set the password for releasing the key
lock.
RCD=Free
Select Lock to apply the key lock function
to the RCD key.
Feed=Free
Select Lock to apply the key lock function
to the FEED key.
Menu=Free
Select Lock to apply the key lock function
to the MENU key.
Disp
Select Lock to apply the key lock function
to the DISP MENU.
Menu=Free
Alarm
ACK=Free
Math=Free
Select Lock to apply the key lock function
to the computation start/stop operation.
Printout=Free
Select Lock to apply the key lock function
to the various printout operations.
Message=Free
Select Lock to apply the key lock function
to the message printout operation.
Buffer
Select Lock to apply the key lock function
to the operation of clearing the printout
buffer.
clear=Free
Periodic=Free
Select Lock to apply the key lock function
to the operation of resetting the report
data in periodic printout.
(Pen model)
Pen
exchange=Free
(Dot model)
R.
ESC/?
IM 04P01B01-01E
exchange=Free
Keylock
Setting
Select Lock to apply the key lock function
to the operation of moving the pen to a
position where the pen can be easily
replaced.
Select lock to apply the key lock to the
operation of replacing the ribbon cassette
with the power turned ON.
The new setting takes effect.
complete
7-17
7
Setup Operations for Changing/Adding Functions (Basic Setting Mode)
Select Lock to apply the key lock function
to the alarm ACK operation.
7.10 Setting the Key Lock Function
Applying the Changes and Returning to Operation Mode
key or SHIFT +
key to select End and then press the
key. Press
Press the
the
key to select Store and then press the
key. The changes are applied, and
key, the
the screen returns to Operation mode. If you select Abort and press the
changes are discarded, and the screen returns to Operation mode. Press the ESC key
to return to the Basic= screen.
Explanation
Keylock
Sets whether to use or not use the key lock function.
If Not is selected, the setup ends here.
Password
Sets the password for releasing the key lock. Set the password using numbers and
spaces within 4 digits.
Operation of Keys to Be Key-Locked
Sets the operation of the keys to be key-locked. Select Lock or Free for each key
operation.
RCD:
RCD key
Feed:
FEED key
Menu:
MENU key
Disp Menu:
1 MENU key
Alarm ACK:
Alarm ACK operation under the FUNC key
Math:
Computation start/stop/reset operation under the FUNC key
Print out
Printout start/stop/reset operation under the FUNC key
Message
Message printout operation under the FUNC key
Buffer clear:
Clear operation of the printout buffer memory under the FUNC key
Periodic:
Clear operation of the report data of periodic printout under the FUNC key
Pen exchange: Operation of moving the pen to an easily accessible position of
replacement under the FUNC key (pen model)
R. exchange: Operation carried out when replacing the ribbon cassette with the power
turned ON under the FUNC key (dot model)
<Related Topics>
Activating/Releasing the key lock: Section 3.13
7-18
IM 04P01B01-01E
7.11 Enabling the Moving Average Function (Dot
Model)
This section explains the details of enabling/disabling the moving average function on
the dot model.
Basic Setting mode cannot be entered when recording is in progress or when
computation is in progress on models with the computation function (/M1 option).
Procedure
Changing the Settings
1.
Hold down the MENU key for 3 seconds to enter Setting mode.
2.
Hold down the
and
keys simultaneously for 3 seconds to enter Basic
Setting mode.
3.
Press the
key or SHIFT +
key to select Moving_AVE and then press
the
key.
4.
Set each item and press the
key.
key or SHIFT +
key to select values.
Use the
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up
to then are cancelled, and the display returns to a higher level menu.
5.
When the Setting complete screen appears, do either of the following:
To correct the setting, press the
key.
If you are done, press the ESC key.
ESC/?
Moving_AVE=Not
Enable/Disable moving average.
Move AVE
Setting complete
The new setting takes effect.
Applying the Changes and Returning to Operation Mode
Press the
key or SHIFT +
key to select End and then press the
key. Press
the
key to select Store and then press the
key. The changes are applied, and
the screen returns to Operation mode. If you select Abort and press the
key, the
changes are discarded, and the screen returns to Operation mode. Press the ESC key
to return to the Basic= screen.
Explanation
Moving_AVE
Use: Enables the setting of the number of samples of the moving average in Setting
mode.
Not: The Moving_AVE item does not appear in Setting mode.
<Related Topics>
Setting the moving average function: Section 6.3
IM 04P01B01-01E
7-19
Setup Operations for Changing/Adding Functions (Basic Setting Mode)
Basic=Moving_AVE
7
7.12 Enabling the Filter Function (Pen Model)
This section explains details of enabling/disabling the input filter function on the pen
model.
Basic Setting mode cannot be entered when recording is in progress or when
computation is in progress on models with the computation function (/M1 option).
Procedure
Changing the Settings
1.
Hold down the MENU key for 3 seconds to enter Setting mode.
2.
Hold down the
and
keys simultaneously for 3 seconds to enter Basic
Setting mode.
3.
Press the
key or SHIFT +
key to select Filter and then press the
key.
4.
Set each item and press the
key.
Use the
key or SHIFT +
key to select values.
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up
to then are cancelled, and the display returns to a higher level menu.
5.
When the Setting complete screen appears, do either of the following:
To correct the setting, press the
key.
If you are done, press the ESC key.
Basic=Filter
ESC/?
Filter=Not
Enable/Disable the filter function.
Filter
Setting
The new setting takes effect.
complete
Applying the Changes and Returning to Operation Mode
Press the
key or SHIFT +
key to select End and then press the
key. Press
the
key to select Store and then press the
key. The changes are applied, and
the screen returns to Operation mode. If you select Abort and press the
key, the
changes are discarded, and the screen returns to Operation mode. Press the ESC key
to return to the Basic= screen.
Explanation
Filter
Use: Enables the setting of the filter time constant in Setting mode.
Not: The Filter item does not appear in Setting mode.
<Related Topics>
Setting the filter: Section 6.2
7-20
IM 04P01B01-01E
7.13 Enabling the Partial Expanded Recording
Function
This section explains the details of enabling/disabling the partial expanded recording
function.
Basic Setting mode cannot be entered when recording is in progress or when
computation is in progress on models with the computation function (/M1 option).
Procedure
Changing the Settings
1.
Hold down the MENU key for 3 seconds to enter Setting mode.
2.
Hold down the
and
keys simultaneously for 3 seconds to enter Basic
Setting mode.
3.
Press the
key or SHIFT +
key to select Partial and then press the
key.
4.
Set each item and press the
key.
key or SHIFT +
key to select values.
Use the
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up
to then are cancelled, and the display returns to a higher level menu.
5.
When the Setting complete screen appears, do either of the following:
To correct the setting, press the
key.
If you are done, press the ESC key.
ESC/?
Partial=Not
Enable/Disable partial expanded recording.
Partial
Setting
The new setting takes effect.
complete
Applying the Changes and Returning to Operation Mode
Press the
key or SHIFT +
key to select End and then press the
key. Press
the
key to select Store and then press the
key. The changes are applied, and
the screen returns to Operation mode. If you select Abort and press the
key, the
changes are discarded, and the screen returns to Operation mode. Press the ESC key
to return to the Basic= screen.
Explanation
Partial
Use: Enables the setting of the boundary position and boundary value of the partial
expanded recording function in Setting mode.
Not: The Partial item does not appear in Setting mode.
<Related Topics>
Setting the partial expanded recording function: Section 6.5
IM 04P01B01-01E
7-21
Setup Operations for Changing/Adding Functions (Basic Setting Mode)
Basic=Partial
7
7.14 Changing the Display/Recording Language
This section explains the details of changing the display/recording language.
Basic Setting mode cannot be entered when recording is in progress or when
computation is in progress on models with the computation function (/M1 option).
Procedure
Changing the Settings
1.
Hold down the MENU key for 3 seconds to enter Setting mode.
2.
Hold down the
and
keys simultaneously for 3 seconds to enter Basic
Setting mode.
3.
Press the
key or SHIFT +
key to select Language and then press the
key.
4.
Set each item and press the
key.
Use the
key or SHIFT +
key to select values.
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up
to then are cancelled, and the display returns to a higher level menu.
5.
When the Setting complete screen appears, do either of the following:
To correct the setting, press the
key.
If you are done, press the ESC key.
Basic=Language
ESC/?
Lang=English
Select the display/recording language.
Language
Setting complete
The new setting takes effect.
Applying the Changes and Returning to Operation Mode
Press the
key or SHIFT +
key to select End and then press the
key. Press
the
key to select Store and then press the
key. The changes are applied, and
the screen returns to Operation mode. If you select Abort and press the
key, the
changes are discarded, and the screen returns to Operation mode. Press the ESC key
to return to the Basic= screen.
Explanation
Lang (Language)
English:
Uses English for the display and recording.
Japanese: Uses Japanese for the display and recording.
7-22
IM 04P01B01-01E
7.15 Enabling the Bias Function, Low-Cut Function,
and Alarm Delay Function
This section explains the details of enabling/disabling the bias function, low-cut function,
and the alarm delay function.
Basic Setting mode cannot be entered when recording is in progress or when
computation is in progress on models with the computation function (/M1 option).
Procedure
Changing the Settings
1.
Hold down the MENU key for 3 seconds to enter Setting mode.
and
keys simultaneously for 3 seconds to enter Basic
2.
Hold down the
Setting mode.
3.
Press the
key or SHIFT +
key to select Personalize and then press
key.
the
4.
Press the
key with Add function shown on the screen.
5.
Set each item and press the
key.
key or SHIFT +
key to select values.
Use the
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up
to then are cancelled, and the display returns to a higher level menu.
6.
When the Setting complete screen appears, do either of the following:
To correct the setting, press the
key.
If you are done, press the ESC key.
Mode=Add
function
Bias=Not
Enable/Disable the bias function.
SQRT
low-cut=Use
Enable/Disable the low-cut function on
channels set to square root computation.
1-5V
low-cut=Not
Enable/Disable the low-cut function on
channels set to 1-5V.
Alarm
ESC/?
delay=Not
Personalize
Setting complete
Enable/Disable the alarm delay function.
The new setting takes effect.
Applying the Changes and Returning to Operation Mode
Press the
key or SHIFT +
key to select End and then press the
key. Press
the
key to select Store and then press the
key. The changes are applied, and
the screen returns to Operation mode. If you select Abort and press the
key, the
changes are discarded, and the screen returns to Operation mode. Press the ESC key
to return to the Basic= screen.
IM 04P01B01-01E
7-23
Setup Operations for Changing/Adding Functions (Basic Setting Mode)
Basic=Personalize
7
7.15 Enabling the Bias Function, Low-Cut Function, and Alarm Delay Function
Explanation
Bias
Enables/Disables the bias function.
Use: Enables the setting of the bias value in Setting mode.
Not: The Bias item does not appear in Setting mode.
SQRT low-cut
Enables/Disables the square root low-cut function.
Use: Enables the setting of the low-cut function when a channel is set to square root
computation in Setting mode.
Not: The SQRT low-cut item does not appear in Setting mode.
1-5V low-cut
Enables/Disables the 1-5V low-cut function.
Use: Enables the setting of the low-cut function when a channel is set to 1-5V in
Setting mode.
Not: The 1-5V low-cut item does not appear in Setting mode.
Alarm delay
Enables/Disables the alarm delay function.
Use: Enables the setting of delay high limit alarm or delay low limit alarm for the alarm
type.
Not: The delay high limit alarm and delay low limit alarm items do not appear as
possible selections for the alarm type.
7-24
IM 04P01B01-01E
7.16 Changing the Time Printout Format
This section explains the details of changing the time printout format of alarm printout,
message printout, recording start printout, and new chart speed printout.
Basic Setting mode cannot be entered when recording is in progress or when
computation is in progress on models with the computation function (/M1 option).
Procedure
Basic=Personalize
Mode=Time
Alarm=HH:MM
Select the time printout format of the
alarm printout.
Message=HH:MM
Select the time printout format of the
message printout.
RCD
Select the time printout format of the
recording start printout.
C.
ESC/?
IM 04P01B01-01E
print
On=HH:MM
Speed=HH:MM
Personalize
Setting complete
Select the time printout format of the
new chart speed printout.
The new setting takes effect.
Applying the Changes and Returning to Operation Mode
Press the
key or SHIFT +
key to select End and then press the
key. Press
the
key to select Store and then press the
key. The changes are applied, and
the screen returns to Operation mode. If you select Abort and press the
key, the
changes are discarded, and the screen returns to Operation mode. Press the ESC key
to return to the Basic= screen.
7-25
7
Setup Operations for Changing/Adding Functions (Basic Setting Mode)
Changing the Settings
1.
Hold down the MENU key for 3 seconds to enter Setting mode.
2.
Hold down the
and
keys simultaneously for 3 seconds to enter Basic
Setting mode.
3.
Press the
key or SHIFT +
key to select Personalize and then press
the
key.
4.
Press the
key or SHIFT +
key to select Time print and then press the
key.
5.
Set each item and press the
key.
Use the
key or SHIFT +
key to select values.
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up
to then are cancelled, and the display returns to a higher level menu.
6.
When the Setting complete screen appears, do either of the following:
To correct the setting, press the
key.
If you are done, press the ESC key.
7.16 Changing the Time Printout Format
Explanation
Alarms
Sets the time printout format of the alarm printout.
HH:MM:
Hour:Minute
HH:MM:SS: Hour:Minute:Second
M/D H:M:
Month Day Hour:Minute
M/D H:M:S: Month Day Hour:Minute:Second
YMD H:M:S: Month Day Year Hour:Minute:Second
Message
Sets the time printout format of the message printout.
HH:MM:
Hour:Minute
HH:MM:SS: Hour:Minute:Second
M/D H:M:
Month Day Hour:Minute
M/D H:M:S: Month Day Hour:Minute:Second
YMD H:M:S: Month Day Year Hour:Minute:Second
None:
Does not print the date/time.
RCD On
Sets the time printout format of the record start time printout
The format is the same as that of the alarm printout.
C. Speed
Selects the time printout format of the new chart speed printout.
The format is the same as that of the alarm printout.
Note
The printout format of the year, month, and day varies depending on the printout/display
format of the date (see section 7.19).
7-26
IM 04P01B01-01E
7.17 Initializing the Settings
This section explains the details of initializing the recorder settings to their factory
default.
Be careful, because all settings except the date/time setting in the recorder will be
initialized.
Basic Setting mode cannot be entered when recording is in progress or when
computation is in progress on models with the computation function (/M1 option).
Procedure
1.
2.
3.
4.
5.
Basic=Initialize
Mode=Setup+Set
Are
ESC/?
you
sure?=No
Select the range to be initialized.
Select Yes and press
to execute
initialization and return to Operation
mode.
Explanation
Items That Are Not Initialized
The following items are not initialized.
• Adjustment value of the pen position and the adjustment value of the printer carriage
position (Basic Setting mode)
• Date/Time (Setting mode)
IM 04P01B01-01E
7-27
7
Setup Operations for Changing/Adding Functions (Basic Setting Mode)
Hold down the MENU key for 3 seconds to enter Setting mode.
Hold down the
and
keys simultaneously for 3 seconds to enter Basic
Setting mode.
Press the
key or SHIFT +
key to select Initialize and then press the
key.
Press the
key to select the settings to be initialized and then press the
key.
Set:
Initializes the settings in Setting mode.
Setup+Set: Initializes the settings in Basic Setting mode and Setting mode.
To execute the initialization, press the
key to select Yes and then press the
key.
The settings are initialized, and the recorder returns to Operation mode.
To cancel the initialization, press the ESC key The screen returns to the
Basic=Initialize screen.
If you press the
key to select No and press then the
key, nothing
happens.
7.18 Assigning Functions to the Remote Control
Input Terminals (/R1 Option)
This section explains the details of assigning functions to the input terminals of the
remote control function.
Basic Setting mode cannot be entered when recording is in progress or when
computation is in progress on models with the computation function (/M1 option).
Procedure
Changing the Settings
1.
Hold down the MENU key for 3 seconds to enter Setting mode.
2.
Hold down the
and
keys simultaneously for 3 seconds to enter Basic
Setting mode.
3.
Press the
key or SHIFT +
key to select Remote and then press the
key.
4.
Set each item and press the
key.
key or SHIFT +
key to select values.
Use the
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up
to then are cancelled, and the display returns to a higher level menu.
5.
When the Setting complete screen appears, do either of the following:
To correct the setting, press the
key.
If you are done, press the ESC key.
Basic=Remote
Remote
No.=1
1=Record
ESC/?
On/Off
Remote No.1
Setting complete
Select the remote control input terminal
number.
Select the function to be assigned.
The new setting takes effect.
Applying the Changes and Returning to Operation Mode
Press the
key or SHIFT +
key to select End and then press the
key. Press
the
key to select Store and then press the
key. The changes are applied, and
the screen returns to Operation mode. If you select Abort and press the
key, the
changes are discarded, and the screen returns to Operation mode. Press the ESC key
to return to the Basic= screen.
7-28
IM 04P01B01-01E
7.18 Assigning Functions to the Remote Control Input Terminals (/R1 Option)
Explanation
Remote No.
The remote control terminal numbers are from 1 to 5.
Function to Be Assigned
Record On/Off: Starts/stops recording.
Alarm ACK:
Executes alarm output release.
Time adjust:
Adjusts the internal clock to the nearest hour.
Manual print:
Executes manual printout.
Message #:
Prints message # (where # is a value between 1 and 5).
Chart speed:
Changes the chart speed.
Math start/stop: Starts/stops the computation on the computation function (/M1 option).
Math reset:
Resets the computed result of the computation function (/M1 option).
None:
No function is assigned.
<Related Topics>
Setting the secondary chart speed: Section 6.9
7
Setup Operations for Changing/Adding Functions (Basic Setting Mode)
IM 04P01B01-01E
7-29
7.19 Changing the Printout/Display Format of the
Date
This section explains how to change the printout/display format of the year, month, and
day.
Basic Setting mode cannot be entered when recording is in progress or when
computation is in progress on models with the computation function (/M1 option).
Procedure
Changing the Settings
1.
Hold down the MENU key for 3 seconds to enter Setting mode.
2.
Hold down the
and
keys simultaneously for 3 seconds to enter Basic
Setting mode.
3.
Press the
key or SHIFT +
key to select Date format and then press
the
key.
4.
Set each item and press the
key.
key or SHIFT +
key to select values.
Use the
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up
to then are cancelled, and the display returns to a higher level menu.
5.
When the Setting complete screen appears, do either of the following:
To correct the setting, press the
key.
If you are done, press the ESC key.
Basic=Date
ESC/?
format
Type=M.D.Y
Select the date format.
Date format
Setting complete
The new setting takes effect.
Applying the Changes and Returning to Operation Mode
Press the
key or SHIFT +
key to select End and then press the
key. Press
the
key to select Store and then press the
key. The changes are applied, and
the screen returns to Operation mode. If you select Abort and press the
key, the
changes are discarded, and the screen returns to Operation mode. Press the ESC key
to return to the Basic= screen.
Explanation
Date Format
7-30
Setting
Format Example
Y/M/D
2005/8/31
M/D/Y
8/31/2005
D/M/Y
31/8/2005
D.M.Y
31.8.2005
M.D.Y
Aug.31.2005
Note
Default value
IM 04P01B01-01E
7.20 Changing the Temperature Unit
This section explains the details of changing the temperature unit on the temperature
measurement channels.
Basic Setting mode cannot be entered when recording is in progress or when
computation is in progress on models with the computation function (/M1 option).
Procedure
Changing the Settings
1.
Hold down the MENU key for 3 seconds to enter Setting mode.
2.
Hold down the
and
keys simultaneously for 3 seconds to enter Basic
Setting mode.
3.
Press the
key or SHIFT +
key to select Temperature and then press
the
key.
4.
Set each item and press the
key.
Use the
key or SHIFT +
key to select values.
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up
to then are cancelled, and the display returns to a higher level menu.
5.
When the Setting complete screen appears, do either of the following:
To correct the setting, press the
key.
If you are done, press the ESC key.
ESC/?
Temp=C
Select the temperature unit.
Temperature
Setting complete
The new setting takes effect.
Applying the Changes and Returning to Operation Mode
Press the
key or SHIFT +
key to select End and then press the
key. Press
the
key to select Store and then press the
key. The changes are applied, and
the screen returns to Operation mode. If you select Abort and press the
key, the
changes are discarded, and the screen returns to Operation mode. Press the ESC key
to return to the Basic= screen.
Explanation
Temp (Temperature)
The setting is applied to all temperature measurement channels.
C: Use Celsius.
F: Use Fahrenheit.
IM 04P01B01-01E
7-31
Setup Operations for Changing/Adding Functions (Basic Setting Mode)
Basic=Temperature
7
Chapter 8 Setup Operations for Changing the Displayed Contents
8.1
Key Operations for Changing the Displayed
Information
Displaying the Data Display Setup Screen
Hold the 1 MENU key for 3 seconds to show the data display setup screen.
Power ON
Operation mode
Hold down 1 MENU or 3 s to switch.
Data display setup mode
The panel keys are set to the functions marked above the keys as shown below.
CHARACTER
RCD
MENU
DISP
FUNC
ESC/?
SHIFT
1 MENU
FEED
CH UP
Returning from the Data Display Setup Screen to Data Display Screen
Hold the 1 MENU key for 3 seconds to return from the data display setup screen to data
display screen.
Note
When the data display setup screen is shown, operations such as measurement, alarm, and
recording continue.
8
Setup Operations for Changing the Displayed Contents
IM 04P01B01-01E
8-1
8.1 Key Operations for Changing the Displayed Information
Data Display Setup Menu
This menu is used to assign display types to screen 01 to 15. The
indicates a
display type. The items to set for each type are shown to the right of the display type.
Screen
Screen 01
Screen 15
Display type
1-channel digital display
Display update rate
End
2-channel digital display
Display update rate
End
4-channel digital display
Display update rate
End
6-channel digital display (dot model)
Display update rate
1-channel digital display + 1 channel bar graph display
End
Display update rate
1-channel digital display + 4-channel bar graph display (pen model)
2-channel digital display + 2-channel bar graph display
Flag display
On/Off for each channel
Channel alarm status display
Display update rate
Display update rate
4-channel bar graph (pen model), 6-channel bar graph (dot model)
End
End
End
End
End
End
Date/Time/Chart speed display
End
DI/DO display (when the /R1, or /A
option is installed)
Multiple display
(Display in which different
screens can be assigned
to the top and bottom sections)
Top section
End
Channel No.
Bottom section
Display update rate
Channel No.
1-channel digital display: tag display
Display update rate
End
2-channel digital display: tag display
Display update rate
End
1-channel digital display + 1-channel bar graph display: tag display
Display update rate
Display update rate
End
End
1-channel digital display + 4-channel bar graph display: tag display (pen model)
Status display
End
System display
End
Lights off
End
Skip
End
Display update rate
End
The display update rate corresponds to “Interval” in the setup menu. It is the time interval at
which the displayed channel is switched.
8-2
IM 04P01B01-01E
8.2
Changing the Displayed Information
This section explains the details of registering display types to screens 01 to 15. As an
example, the procedure of setting 1-channel digital display is explained below.
Procedure
1-Channel Digital Display
1.
Hold the 1 MENU key for 3 seconds to show the data display setup screen.
2.
Press the
key or SHIFT +
key to select the screen number and then
press the
key.
A sample screen of the display type appears. A section of the display blinks to
indicate that this is a data display setup screen.
3.
Press the
or SHIFT +
key to switch the display. Select the 1-channel
digital display and then press the
key.
4.
Press the
key or SHIFT +
key to select the channel switching interval
and then press the
key.
5.
When the Setting complete screen appears, press the
key.
1
MENU
Hold the
key down for 3 seconds to exit from the data display setup
screen.
Screen
No.=01
Select the screen No.
Current display type name.
1 200.0 mV
Select the display type.
8
Select the channel switching interval.
Screen No.01
Setting complete
The new setting takes effect.
Explanation
Interval
Sets the channel switching interval.
Select the interval from 1 s, 2 s, 3 s, 4 s, 5 s, and manual.
1s, 2s, 3s, 4s, 5s: Switches the channels at the specified interval.
Manual:
Switch the channel manually.
Assigning Other Display Types (See the menu on the previous page)
• When assigning the flag display, specify whether to display the flag for each channel.
• If the channel switching interval is set to manual when assigning the display in which
different screens can be assigned to the top and bottom sections, set the channel
number.
<Related Topics>
Display types: Section 12.4
IM 04P01B01-01E
8-3
Setup Operations for Changing the Displayed Contents
Interval=Auto2s
Chapter 9 Operations Related to the Computation Function (/M1 Option)
9.1
Starting/Stopping/Resetting the Computation
This section explains the details of starting/stopping the computation and resetting the
computed values of computation channels.
Procedure
Starting the Computation
1.
Press the FUNC key.
2.
Press the
key to show Math and then press the
key.
3.
Press the
key with Start shown on the screen.
The computation starts, and the screen switches to the data display screen.
The word MATH in the status display comes on.
Math=Start
Stopping the Computation
1.
Press the FUNC key.
2.
Press the
key to show Math and then press the
key.
key with Stop shown on the screen.
3.
Press the
The computation stops, and the screen switches to the data display screen.
The word MATH in the status display clears.
Math=Stop
Resetting the Computed Values of All Computation Channels to 0
This operation can be executed only when the computation is stopped.
1.
Press the FUNC key.
2.
Press the
key to show Math and then press the
key.
3.
Press the
key to show Reset and then press the
key.
The computed values of all computation channels are reset, and the screen
switches to the data display screen.
9
Operations Related to the Computation Function (/M1 Option)
Math=Reset
* When the
key or
key is pressed while holding down the SHIFT key, the operation is
reversed as when the respective key is pressed by itself.
IM 04P01B01-01E
9-1
9.2
Setting the Computing Equation
The computing equation cannot be entered when the computation is in progress. If you
change the computing equation or the recording span, set the alarm and the partial
expanded recording again.
Procedure
1.
2.
3
4.
5.
6.
Hold down the MENU key for 3 seconds to enter Setting mode.
Press the
key or SHIFT +
key to select Math and then press the
key.
Press the
key with Formula shown on the screen.
Set each item and press the
key.
Use the
key or SHIFT +
key to select values.
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC/? key in the middle of the operation, the settings entered up
to then are cancelled, and the display returns to a higher level menu.
When the Setting complete screen appears, do either of the following:
Press the
key to set other channels.
If you are done, press the ESC/? key.
Hold down the MENU key for 3 seconds to return to Operation mode.
Set=Math
Math=Formula
Set the channel range.
CH=0A-0A
First channel
ESC/?
Last channel
Mode=Off
Select On to use the computing equation.
f.=01
Set the equation.
Span_L=
-200.00
Span_R=
200.00
0A-0A Channel
Setting complete
Set the left span value.
Set the right span value.
The new setting takes effect.
Setting the Computing Equation
The CHARACTER,
, and
keys are used to set the computing equation.
Use the
key or SHIFT +
key to select the position for entering a character.
Use the CHARACTER key or SHIFT + CHARACTER key to select the character type.
key or SHIFT +
key to select the character.
Use the
9-2
IM 04P01B01-01E
9.2 Setting the Computing Equation
Each time you press the CHARACTER key the display changes as shown below.
key or SHIFT +
key switches the character within the selected
Pressing the
character type and displays the character at the cursor position.
Display (CHARACTER key) Description
Character (
01...
Measurement channel No.
Pen model: 01 to 04
Dot model: 01 to 06
0A...
Computation channel No.
Pen model: 0A, 0B, 0C, 0D, 0E, 0F,
0G, 0J
Dot model: 0A, 0B, 0C, 0D, 0E, 0F,
0G, 0J, 0K, 0M, 0N, 0P
+, –, *, /...
Symbol and space
+, –, *, /, (, ), space
Function
Function
SQR(, ABS(, LOG(, EXP(, .EQ.,
.NE., .GT., .LT., .GE., .LE., AND,
NOT, XOR, OR, TLOG.AVE(,
TLOG.MAX(, TLOG.MIN(,
TLOG.SUM(, TLOG.P-P(, space
key)
K01-K30
Constant
K01 to K30
D01-D05
Status of the remote
D01 to D05
control terminals
C01-C12
Communication input data
Pen model: C01 to C08
Dot model: C01 to C12
Ins DISP
Insert a character
Del DISP
Delete a character
Clear DISP
Delete the entire string
Copy DISP
Store the entire string in the
memory
Past DISP
Paste the string in the memory
Inserting a Character
Press the
key or SHIFT +
key to move the cursor to the position where the
character is to be inserted.
Press the CHARACTER key or SHIFT + CHARACTER key to show Ins DISP and then
press the
key. A space for one character is inserted. Enter the character.
Deleting an Entire Character String
Press the CHARACTER key or SHIFT + CHARACTER key to show Clear DISP and
then press the
key. The entire character string is deleted.
Copying & Pasting a Character String
Show the copy source character string.
Press the CHARACTER key or SHIFT + CHARACTER key to show Copy DISP and
then press the
key. The character string is saved to the memory.
Show the copy destination.
Press the CHARACTER key or SHIFT + CHARACTER key to show Paste DISP and
then press the
key. The character string is pasted.
IM 04P01B01-01E
9-3
Operations Related to the Computation Function (/M1 Option)
Deleting a Character
Use the
key or SHIFT +
key to move the cursor to the character to be deleted.
Press the CHARACTER key or SHIFT + CHARACTER key to show Del DISP and then
press the
key. The character is deleted.
9
9.2 Setting the Computing Equation
Explanation
Computing Equation
Follow the rules below when writing computing equations.
• Use up to 120 characters to write equations.
• The precedence of computing terms can be specified using parentheses.
• Specify the channels in the equation using 2-digit channel numbers.
Example: 01, 0A
• Specify the constants (K), communication input data (C), and remote control input
terminal status (D) in the computing equation using 2-digit numbers.
Example: K01, C01, D01
• The data value for its own channel and channels that are greater in number than its
own channel are substituted with the data from the previous scan.
Order of Precedence in Computations
The order of precedence of computation in equations is as follows:
Type
Function
Power
Logical negation
Multiplication and division
Addition and subtraction
Greater than and less than
Equal and not equal
Logical product
Logical sum and exclusive logical sum
Operator
(high order of precedence)
ABS(), SQR(), LOG(), EXP(), TLOG.MAX(),
TLOG.MIN(), TLOG.P-P(), TLOG.SUM(),
TLOG.AVE()
**
NOT
*, /
+, –
.GT., .LT., GE., LE.
.EQ., .NE.
AND
OR, XOR
(low order of precedence)
Data That Can Be Used in Equations
In TLOG computation, the measured values of measurement channels or computed
values of computation channels can only be used. For all other computations, all of the
data types below can be used.
• Measured Values on Measurement Channels
Specified by the channel No. For channels that have linear scaling enabled, the
scaled values are used in the computation.
Pen model: 01 to 04, dot model: 01 to 06
• Computed Values on Computation Channels
Specified by the channel No.
Pen model: 0A, 0B, 0C, 0D, 0E, 0F, 0G, 0J
Dot model: 0A, 0B, 0C, 0D, 0E, 0F, 0G, 0J, 0K, 0M, 0N, 0P
• Constants (K01 to K30)
Enter the values as K01 to K30 in the equations.
Range of values (maximum significant digits is 5):
–9.9999E+29 to –1.0000E–30, 0, 1.0000E–30 to 9.9999E+29
• Communication Input Data (Pen model: C01 to C08, dot model: C01 to C12)
Enter the data as C01 to C12 in the equations.
Range of values (maximum significant digits is 5):
–9.9999E+29 to –1.0000E–30, 0, 1.0000E–30 to 9.9999E+29
For the procedure of setting the communication input data, see the µR10000/µR20000
Communication Interface User’s Manual (IM04P01B01-17E).
9-4
IM 04P01B01-01E
9.2 Setting the Computing Equation
• Status of the Remote Control Input Terminal (D01 to D05, /R1 option)
The status of the remote control input signal (1 or 0) can be used in the computation.
Use D01 to D05 (the number following the letter D is the remote control input terminal
number) to specify the status in the equations.
Four Arithmetic Operation
The data that can be used in equations are measured values, computed values,
constants, communication input data, and remote control input terminal status.
Equation Examples
• Addition
01+02
(Determines the sum of the measured values of channel 1 and channel
2.)
• Subtraction 01–02
(Determines the difference of the measured values of channel 1 and
channel 2.)
• Multiplication 01*K03
(Multiplies constant K03 to the measured value of channel 1.)
• Division
01/K02
(Divides the measured value of channel 1 by constant K02.)
Note
By setting an equation similar to the one shown below, you can determine the accumulation
of a specified channel (in this case channel 01) and make the result the computed value of
computation channel 0A.
Computing equation of computation channel 0A: 0A+01
Power and Other Computations
Note
You can determine the natural logarithm by setting an equation like the following.
From logbX = logaX/loga, we obtain logeX = log10X/log10e
Hence, to determine the natural logarithm of channel 1, we set
Constant
K01 = 0.43429(log10e)
Computing equation and set the equation to LOG(01)/K01.
IM 04P01B01-01E
9-5
9
Operations Related to the Computation Function (/M1 Option)
The data that can be used in equations are measured values, computed values,
constants, communication input data, and remote control input terminal status. You can
nest a computing element inside the parentheses of another computing element.
Equation Examples
• Power
01**02
(Determines the measured value of channel 1 to the power of
the measured value of channel 2.)
• Square root
SQR(02)
(Determines the square root of the measured value of channel
2.)
• Absolute value
ABS(02)
(Determines the absolute value of the measured value of
channel 2.)
• Common logarithm
LOG(01)
(Determines the common logarithm (log10) of the measured
value of channel 1.)
• Exponent
EXP(01)
(Determines e to the power of the measured value of channel
1.)
9.2 Setting the Computing Equation
Relational Computation
The data that can be used in equations are measured values, computed values,
constants, communication input data, and remote control input terminal status. You can
specify a computing equation that performs relational computation on a computing
element. (Example: 01.LT.ABS(02))
Equation Examples
02.LT.03
If the measured value of channel 2 is less than the measured value of channel 3, the
computed result is “1.” Otherwise, the result is “0.”
02.GT.03
If the measured value of channel 2 is greater than the measured value of channel 3, the
computed result is “1.” Otherwise, the result is “0.”
02.EQ.03
If the measured value of channel 2 is equal to the measured value of channel 3, the
computed result is “1.” Otherwise, the result is “0.”
02.NE.03
If the measured value of channel 2 is not equal to the measured value of channel 3, the
computed result is “1.” Otherwise, the result is “0.”
02.GE.03
If the measured value of channel 2 is greater than or equal to the measured value of
channel 3, the computed result is “1.” Otherwise, the result is “0.”
02.LE.03
If the measured value of channel 2 is less than or equal to the measured value of
channel 3, the computed result is “1.” Otherwise, the result is “0.”
Logical Computation
Checks whether the two data values, e1 and e2 (e1 only for NOT), are zeroes or nonzeroes, and computes according to the conditions. The data that can be used in
equations are measured values, computed values, constants, communication input data,
and remote control input terminal status. You can specify a computing equation that
performs logical computation on a computing element.
AND
Logical Product
(Syntax)
e1ANDe2
(Condition)
If the two data values e1 and e2 are both non-zeroes, the computed
result is “1.” Otherwise, it is “0.”
(Explanation)
e1 = 0
→
e1ANDe2 = 0
e2 = 0
e1 ≠ 0
→
e1ANDe2 = 0
→
e1ANDe2 = 0
→
e1ANDe2 = 1
e2 = 0
e1 = 0
e2 ≠ 0
e1 ≠ 0
e2 ≠ 0
9-6
IM 04P01B01-01E
9.2 Setting the Computing Equation
OR
Logical sum
(Syntax)
(Condition)
(Explanation)
e1ORe2
If the two data values e1 and e2 are both zeroes, the computed result is
“0.” Otherwise, it is “1.”
e1 = 0
→
e1ORe2 = 0
e2 = 0
e1 ≠ 0
e2 = 0
e1 = 0
e2 ≠ 0
e1 ≠ 0
e2 ≠ 0
XOR
Exclusive OR
(Syntax)
(Condition)
(Explanation)
→
e1ORe2 = 1
→
e1ORe2 = 1
→
e1ORe2 = 1
e1XORe2
If the two data values e1 and e2 are zero and non-zero or non-zero and
zero, the computed result is “1.” Otherwise, it is “0.”
e1 = 0
→
e1XORe2 = 0
e2 = 0
e1 ≠ 0
e2 = 0
e1 = 0
e2 ≠ 0
e2 ≠ 0
e1XORe2 = 1
→
e1XORe2 = 1
→
e1XORe2 = 0
9
Operations Related to the Computation Function (/M1 Option)
e1 ≠ 0
→
NOT
Logical negation
(Syntax)
NOTe1
(Condition)
The result is the inverse of the status of data e1 (zero or non-zero).
(Explanation)
e1 = 0 →
NOTe1 = 1
e1 ≠ 0 →
NOTe1 = 0
Equation Example
01-02OR03.GT.04
Determines the OR of the computed results of “01-02” and “03.GT.04”.
IM 04P01B01-01E
9-7
9.2 Setting the Computing Equation
TLOG Computation
Only measured values and computed values can be used in TLOG computation. In the
explanation below, e1 is used to represent a measurement or computation channel. You
cannot specify an equation that contains a computing element inside e1. In addition,
only one TLOG computation can be specified in a single computing equation.
TLOG.MAX( )
Maximum value
(Syntax)
TLOG.MAX(e1)
(Condition)
Determines the maximum value of channel e1.
TLOG.MIN( )
Minimum value
(Syntax)
TLOG.MIN(e1)
(Condition)
Determines the minimum value of channel e1.
TLOG.AVE( )
Average value
(Syntax)
(Condition)
TLOG.AVE(e1)
Determines the average value of channel e1.
TLOG.SUM( )
Sum value
(Syntax)
(Condition)
TLOG.SUM(e1)
Determines the sum of channel e1.
TLOG.P-P( )
Maximum − minimum value
(Syntax)
TLOG.P-P(e1)
(Condition)
Determines the maximum − minimum value of channel e1.
Equation Example
TLOG.MAX(01) + K01*SQR(02)
Examples of Equations That Are Not Allowed
TLOG.AVE(01) + TLOG.AVE(02)
Reason: TLOG appears twice in one equation.
TLOG.AVE(ABS(01))
Reason: A computing element is used inside the parentheses.
9-8
IM 04P01B01-01E
9.3
Setting the Unit
This section explains the details of setting the unit on computation channels.
Procedure
1.
2.
3.
4.
5.
6.
Hold down the MENU key for 3 seconds to enter Setting mode.
Press the
key or SHIFT +
key to select Math and then press the
key.
Press the
key or SHIFT +
key to select Unit and then press the
key.
Set each item and press the
key.
Use the
key or SHIFT +
key to select values.
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up
to then are cancelled, and the display returns to a higher level menu.
When the Setting complete screen appears, do either of the following:
Press the
key to set other channels.
If you are done, press the ESC key.
Hold down the MENU key for 3 seconds to return to Operation mode.
Set=Math
Math=Unit
CH=0A-0A
First channel
Last channel
9
Unit=
Set the unit.
0A-0A Channel
Setting complete
The new setting takes effect.
Operations Related to the Computation Function (/M1 Option)
ESC/?
Set the channel range.
Explanation
Characters That Can Be Used for Units
Set the unit using up to 6 characters.
The available characters are as follows:
Alphabet, numbers, symbols (%, #, °, @, +, –, *, /, (, ), µ, Ω, 2, 3, .), and space
IM 04P01B01-01E
9-9
9.4
Setting the Constants Used in Equations
This section explains the details of setting constants to be used in computing equations.
Procedure
1.
2.
3.
4.
5.
6.
Hold down the MENU key for 3 seconds to enter Setting mode.
Press the
key or SHIFT +
key to select Math and then press the
key.
Press the
key or SHIFT +
key to select Constant and then press the
key.
Set each item and press the
key.
Use the
key or SHIFT +
key to select values.
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up
to then are cancelled, and the display returns to a higher level menu.
When the Setting complete screen appears, do either of the following:
Press the
key to set other channels.
If you are done, press the ESC key.
Hold down the MENU key for 3 seconds to return to Operation mode.
Set=Math
Math=Constant
ESC/?
No.=K01
Select the constant number.
K01=1
Set the value.
K01
Setting
The new setting takes effect.
complete
Explanation
Constant
Up to 30 constants can be registered. Enter the values as K01 to K30 in the equations.
Set the constants in the range shown below. The maximum number of significant digits
is 5.
–9.9999E+29 to –1.0000E–30, 0, 1.0000E–30 to 9.9999E+29
9-10
IM 04P01B01-01E
9.5
Setting the Alarm
This section explains the details of setting alarms for each computation channel.
If you change the computing equation or the recording span, set the alarm again.
Procedure
1.
2.
3.
4.
5.
6.
Hold down the MENU key for 3 seconds to enter Setting mode.
Press the
key or SHIFT +
key to select Math and then press the
key.
Press the
key or SHIFT +
key to select Alarm and then press the
key.
Set each item and press the
key.
Use the
key or SHIFT +
key to select values.
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up
to then are cancelled, and the display returns to a higher level menu.
When the Setting complete screen appears, do either of the following:
Press the
key to set other channels.
If you are done, press the ESC key.
Hold down the MENU key for 3 seconds to return to Operation mode.
Set=Math
Math=Alarm
Set the channel range.
CH=0A-0A
First channel
Last channel
Select the level number between 1 and 4.
Alarm=Off
Select On to set the alarm. When set
to On, the settings below can be entered.
Type=H
Select the alarm type.
Value=
ESC/?
IM 04P01B01-01E
0.00
Select the value used to turn On the alarm.
Relay=Off
Turn relay On when outputting alarms on
the relay. This setting is valid only on
models with the /A1, /A2, or /A3 option.
Relay
Specify the alarm output relay No.
No.=I01
0A-0A CH/level 1
Setting complete
The new setting takes effect.
9-11
9
Operations Related to the Computation Function (/M1 Option)
Level=1
9.5 Setting the Alarm
Explanation
Alarm Type
Symbol
Name
H
High limit alarm
L
Low limit alarm
T
Delay high limit alarm*
t
*
Delay low limit alarm*
Selectable only when the alarm delay function is enabled. See section 7.15.
If delay high limit alarm or delay low limit alarm is specified, set the alarm delay time.
<Related Topics>
Setting the delay time of the delay high limit alarm or delay low limit alarm: Section 9.11
Enabling the alarm delay function: Section 7.15
Setting the alarm hysteresis: Section 7.1
Alarm Value
Can be set to a value in the range of –9999999 to 99999999 excluding the decimal (–
9999999 to 99999999 is the displayable range).
Relay No.
The selectable relay numbers are listed below.
I01 and I02 on the /A1 option
I01, I02, I03, and I04 on the /A2 option
I01, I02, I03, I04, I05, and I06 on the /A3 option
9-12
IM 04P01B01-01E
9.6
Specifying the Timer Used in Statistical
Calculations (TLOG)
On channels that perform statistical calculation (TLOG), the timer for determining the
interval is specified. The sum scale is also specified when using statistical calculation.
The timer is also used for the periodic printout of computed values.
Procedure
1.
2.
3.
4.
5.
6.
Hold down the MENU key for 3 seconds to enter Setting mode.
Press the
key or SHIFT +
key to select Math and then press the
key.
Press the
key or SHIFT +
key to select TLOG and then press the
key.
key.
Set each item and press the
Use the
key or SHIFT +
key to select values.
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up
to then are cancelled, and the display returns to a higher level menu.
When the Setting complete screen appears, do either of the following:
Press the
key to set other channels.
If you are done, press the ESC key.
Hold down the MENU key for 3 seconds to return to Operation mode.
Set=Math
Math=TLOG
CH=0A-0A
Timer
SUM
ESC/?
IM 04P01B01-01E
9
Last channel
No=Periodic Select the timer number (Periodic, 1, or 2).
scale=Off
0A-0A Channel
Setting complete
Select the sum scale.
The new setting takes effect.
9-13
Operations Related to the Computation Function (/M1 Option)
First channel
Set the channel range.
9.6 Specifying the Timer Used in Statistical Calculations (TLOG)
Explanation
Timer No.
There are three timers.
Periodic: The timer for determining the periodic printout interval. The computed result
is printed every interval, and the computed TLOG value is reset.
1:
The timer of timer No. 1. Specify whether the computed value is printed every
interval and whether to reset the computed value.
2:
The timer of timer No. 2. Specify whether the computed value is printed every
interval and whether to reset the computed value.
• Timer operation
The table below shows how each timer operates.
Timer #
Item
Periodic
Application
Target channel
Periodic
Measurement and
computation channels
Timer mode**
Absolute time mode
Reset computed TLOG value Yes
Print when computation
No
stops
1 and 2*
TLOG computation
Computation channels
Absolute or relative time mode
Yes/No selectable
Yes
* For the timer 1 and 2 settings, see section 9.12.
** Absolute time mode: Keeps time from the reference time.
Relative time mode: Starts the time in sync with the starting of the computation.
SUM scale
Set the sum scale when using statistical calculation (TLOG.SUM).
For the definition of the term “sum scale,” see page 7-14.
Off: Simply sums the computed values.
/s:
Sums by converting the computed values to a value over 1 second.
/min: Sums by converting the computed values to a value over 1 minute.
/h:
Sums by converting the computed values to a value over 1 hour.
/day: Sums by converting the computed values to a value over 1 day.
<Related Topics>
Setting the periodic printout timer: Section 7.8
Setting Timer 1 and Timer 2: Section 9.12
9-14
IM 04P01B01-01E
9.7
Setting Recording Zones for Each Channel
(Zone Recording)
This section explains the details of setting the recording zone for each computation
channel.
Procedure
Hold down the MENU key for 3 seconds to enter Setting mode.
Press the
key or SHIFT +
key to select Math and then press the
key.
Press the
key or SHIFT +
key to select Aux and then press the
key.
key or SHIFT +
key to select Zone and then press the
key.
Press the
Set each item and press the
key.
Use the
key or SHIFT +
key to select values.
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up
to then are cancelled, and the display returns to a higher level menu.
When the Setting complete screen appears, do either of the following:
Press the
key to set other channels.
If you are done, press the ESC key.
Hold down the MENU key for 3 seconds to return to Operation mode.
1.
2.
3.
4.
5.
6.
7.
Set=Math
Math=Aux
Aux=Zone
CH=0A-0A
Left=
ESC/?
Last channel
0
Set the leftmost value of the recording
zone.
Right=100
Set the rightmost value of the recording
zone.
0A-0A Channel
Setting complete
The new setting takes effect.
Explanation
Zone
The recording zone can be set in the following range.
Leftmost position: 0 mm to 95 mm
Rightmost position: 5 mm to 100 mm
Set a larger value for the rightmost position than the leftmost position.
Set at least 5 mm for the zone width.
Note
If the zone is set to a value less than 40 mm, scale printout of periodic printout cannot be
executed.
IM 04P01B01-01E
9-15
Operations Related to the Computation Function (/M1 Option)
First channel
9
Set the channel range.
9.8
Setting the Partial Expanded Recording
This section explains the details of expanding a section of the recording range.
If you change the computing equation or the recording span, set the partial expanded
recording again.
Procedure
1.
2.
3.
4.
5.
6.
7.
Hold down the MENU key for 3 seconds to enter Setting mode.
Press the
key or SHIFT +
key to select Math and then press the
key.
Press the
key or SHIFT +
key to select Aux and then press the
key.
Press the
key or SHIFT +
key to select Partial and then press the
key.
Set each item and press the
key.
Use the
key or SHIFT +
key to select values.
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up
to then are cancelled, and the display returns to a higher level menu.
When the Setting complete screen appears, do either of the following:
Press the
key to set other channels.
If you are done, press the ESC key.
Hold down the MENU key for 3 seconds to return to Operation mode.
Set=Math
Math=Aux
Aux=Partial
Set the channel range.
CH=0A-0A
First channel
Partial=Off
Select On to use the partial expanded
recording function.
Expand=50
Set the boundary position as
a percentage where the recording span
is assumed to be 100%.
Bound=
ESC/?
Last channel
0.00
0A-0A Channel
Setting complete
Set the boundary value within the
recording span.
The new setting takes effect.
<Related Topics>
Enabling the partial expanded recording function: Section 7.13
9-16
IM 04P01B01-01E
9.9
Turning Trend Recording (Dot Model) and
Periodic Printout ON/OFF for Each Channel
This section explains the details of setting the trend recording and periodic printout for
each computation channel.
The following settings can be entered for each channel on the dot model.
• Turn trend recording ON/OFF.
• Turn the printing of computed values ON/OFF during periodic printout.
The following setting can be entered for each channel on the pen model.
• Turn the printing of computed values ON/OFF during periodic printout.
Procedure
1.
2.
3.
4.
5.
6.
7.
Hold down the MENU key for 3 seconds to enter Setting mode.
key or SHIFT +
key to select Math and then press the
Press the
key.
Press the
key or SHIFT +
key to select Aux and then press the
key.
Press the
key or SHIFT +
key to select Print out and then press the
key.
Set each item and press the
key.
Use the
key or SHIFT +
key to select values.
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up
to then are cancelled, and the display returns to a higher level menu.
When the Setting complete screen appears, do either of the following:
Press the
key to set other channels.
If you are done, press the ESC key.
Hold down the MENU key for 3 seconds to return to Operation mode.
Set=Math
9
Aux=Print
out
CH=0A-0A
Set the channel range.
First channel Last channel
ESC/?
IM 04P01B01-01E
Trend=On
Turn trend recording On/Off
(dot model).
Periodic=On
Select whether to print the computed
values during periodic printout and
printout using the TLOG timer (On/Off).
0A-0A Channel
Setting complete
The new setting takes effect.
9-17
Operations Related to the Computation Function (/M1 Option)
Math=Aux
9.10 Setting Tags on Channels
This section explains the details of setting the tag on computation channels.
Procedure
1.
2.
3.
4.
5.
6.
7.
Hold down the MENU key for 3 seconds to enter Setting mode.
Press the
key or SHIFT +
key to select Math and then press the
key.
Press the
key or SHIFT +
key to select Aux and then press the
key.
Press the
key or SHIFT +
key to select Tag and then press the
key.
Set each item and press the
key.
Use the
key or SHIFT +
key to select values.
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up
to then are cancelled, and the display returns to a higher level menu.
When the Setting complete screen appears, do either of the following:
Press the
key to set other channels.
If you are done, press the ESC key.
Hold down the MENU key for 3 seconds to return to Operation mode.
Set=Math
Math=Aux
Aux=Tag
CH=0A-0A
First channel
ESC/?
Set the channel range.
Last channel
Tag=
Set the tag.
0A-0A Channel
Setting complete
The new setting takes effect.
Explanation
Characters That Can Be Used for Tags
Set the tag using up to 7 characters.
The available characters are as follows:
Alphabet, numbers, symbols (%, #, °, @, +, –, *, /, (, ), µ, Ω, 2, 3, .), and space
<Related Topics>
Printing tags in place of channel numbers: Section 7.7
9-18
IM 04P01B01-01E
9.11 Setting the Alarm Delay Duration
This section explains the details of setting the duration of the delay high limit and delay
low limit alarms.
Procedure
1.
2.
3.
4.
5.
6.
7.
Hold down the MENU key for 3 seconds to enter Setting mode.
Press the
key or SHIFT +
key to select Math and then press the
key.
Press the
key or SHIFT +
key to select Aux and then press the
key.
Press the
key or SHIFT +
key to select Alarm delay and then press
the
key.
Set each item and press the
key.
Use the
key or SHIFT +
key to select values.
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up
to then are cancelled, and the display returns to a higher level menu.
When the Setting complete screen appears, do either of the following:
Press the
key to set other channels.
If you are done, press the ESC key.
Hold down the MENU key for 3 seconds to return to Operation mode.
Set=Math
Math=Aux
Aux=Alm
delay
Last channel
Duration=
ESC/?
10
0A-0A Channel
Setting complete
Set the alarm delay time.
The new setting takes effect.
Explanation
Duration
The duration can be set to a value in the range of 1 to 3600 s.
However, if the scan interval on the dot model is 2.5 s, the actual duration is an integer
multiple of 2.5 s. If the set value is not equal to an integer multiple of 2.5 s, the duration
is the smallest integer multiple of 2.5 s greater than the set value.
Example: The actual duration for 1 s and 2 s is 2.5 s.
<Related Topics>
Enabling the alarm delay function: Section 7.15
IM 04P01B01-01E
9-19
Operations Related to the Computation Function (/M1 Option)
Set the channel range.
CH=0A-0A
First channel
9
T
9.12 Setting the Timer Used in TLOG Computation
and Periodic Printout
This section explains the details of setting the timer used in TLOG computation and
periodic printout.
Basic Setting mode cannot be entered when recording is in progress or when
computation is in progress on models with the computation function (/M1 option).
Procedure
Changing the Settings
1.
Hold down the MENU key for 3 seconds to enter Setting mode.
2.
Hold down the
and
keys simultaneously for 3 seconds to enter Basic
Setting mode.
3.
Press the
key or SHIFT +
key to select Math and then press the
key.
4.
Press the
key with Timer (TLOG) shown on the screen.
key.
5.
Set each item and press the
Use the
key or SHIFT +
key to select values.
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up
to then are cancelled, and the display returns to a higher level menu.
6.
When the Setting complete screen appears, do either of the following:
To correct the setting, press the
key.
If you are done, press the ESC key.
Basic=Math
Math=Timer(TLOG)
Timer
No.=1
Mode=Absolute
Select the timer No.
Select the timer type.
(When the mode is set to Absolute)
Interval=1h
Select the timer interval.
(When the mode is set to Absolute)
ESC/?
9-20
Ref.Time=00:00
Set the reference time.
Reset=Off
Select On to reset the computed
values every interval.
Print=Off
Select On to print the computed
values every interval.
Math Timer
Setting complete
The new setting takes effect.
IM 04P01B01-01E
9.12 Setting the Timer Used in TLOG Computation and Periodic Printout
Applying the Changes and Returning to Operation Mode
key or SHIFT +
key to select End and then press the
key. Press
Press the
the
key to select Store and then press the
key. The changes are applied, and
key, the
the screen returns to Operation mode. If you select Abort and press the
changes are discarded, and the screen returns to Operation mode. Press the ESC key
to return to the Basic= screen.
Explanation
Timers
Timer features
• The timer determines the interval used to calculate the report data (the average,
minimum, maximum, sum value), etc.
• The timer is also used to set the interval at which the computed values are printed on
the chart paper.
Number of Timers
There are two timers: Timer No. 1 and 2.
Timer Type
There are two timer modes, absolute time mode and relative time mode.
• Absolute Time Mode
After the computation is started, the timer times out at the times determined by the
reference time (on the hour) and the interval. The reference time is set on the hour
(00 to 23).
Example 1: Reference time: 14:00
Interval:
12h
The timeout times are hour 2 and hour 14.
Example 2: Reference time: 00:00
Interval:
10min
The timeout times are 0 hour, 0 hour 10 min, 0 hour 20 min, ... 23 hour
40 min, and 23 hour 50 min. For example, if computation is started at 9
hour 36 min, the timeout times are 09 hour 40 min, 09 hour 50 min, 10
hour, and so on.
Setup Item and Corresponding Display
Display
Description
Absolute
Absolute time mode
Relative
Relative time mode
Ref. Time
Reference time
Interval
Interval
For absolute time mode timers, set the reference time and interval; for relative time mode
timers, set the interval.
Timers That Are Used
The timer that is used can be set for each channel.
<Related Topics>
Setting the timer to be used for each channel: Section 9.6
IM 04P01B01-01E
9-21
Operations Related to the Computation Function (/M1 Option)
• Relative Time Mode
The clock is started in sync with the starting of the computation, and the timer times
out every specified interval. In this mode, the timer and the clock stop when a power
failure occurs.
Example:
Interval:
00:15
The timer times out every 15 minutes after the computation is started.
9
9.12 Setting the Timer Used in TLOG Computation and Periodic Printout
Reset
Selects whether to reset the TLOG computed value at each interval. The figure below
illustrates the reset operation for sum computation (TLOG.SUM).
Example: TLOG.SUM computation
Timer timeout
1
Reset
2
Reset
3
Reset
Reset: On
Reset: Off
When reset is ON, the sum value is calculated over each interval. When set to Off, the
sum value from computation start is calculated.
Printout
On: Prints the computed value every interval specified by the timer. Computation
channels set to Off do not apply.
Off: Does not print.
9-22
IM 04P01B01-01E
9.13 Changing the Channel Recording Color (Dot
Model)
This section explains the details of changing the recording colors of computation
channels.
Basic Setting mode cannot be entered when recording is in progress or when
computation is in progress on models with the computation function (/M1 option).
Procedure
Changing the Settings
1.
Hold down the MENU key for 3 seconds to enter Setting mode.
and
keys simultaneously for 3 seconds to enter Basic
2.
Hold down the
Setting mode.
3.
Press the
key or SHIFT +
key to select Math and then press the
key.
key or SHIFT +
key to select Color and then press the
key.
4.
Press the
5.
Set each item and press the
key.
Use the
key or SHIFT +
key to select values.
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up
to then are cancelled, and the display returns to a higher level menu.
6.
When the Setting complete screen appears, do either of the following:
To correct the setting, press the
key.
If you are done, press the ESC key.
Basic=Math
Math=Color
Select the channel.
Color=Purple
Select the recording color.
0A Channel color
Setting complete
The new setting takes effect.
9
Applying the Changes and Returning to Operation Mode
Press the
key or SHIFT +
key to select End and then press the
key. Press
the
key to select Store and then press the
key. The changes are applied, and
the screen returns to Operation mode. If you select Abort and press the
key, the
changes are discarded, and the screen returns to Operation mode. Press the ESC key
to return to the Basic= screen.
Explanation
Color
This section explains the details of changing the recording colors of computation
channels.
The default values are as follows:
Purple: channel 0A, 0G, Red: channel 0B, 0J, Green: channel 0C, 0K,
Blue: channel 0D, 0M, Brown: channel 0E, 0N, Black: channel 0F, 0P
IM 04P01B01-01E
9-23
Operations Related to the Computation Function (/M1 Option)
ESC/?
Channel=0A
9.14 Changing the Channel Assignments of
Recording Pens (Pen Model)
Measurement channels and computation channels can be arbitrarily assigned to
recording pens.
Basic Setting mode cannot be entered when recording is in progress or when
computation is in progress on models with the computation function (/M1 option).
Procedure
Changing the Settings
1.
Hold down the MENU key for 3 seconds to enter Setting mode.
and
keys simultaneously for 3 seconds to enter Basic
2.
Hold down the
Setting mode.
3.
Press the
key or SHIFT +
key to select Math and then press the
key.
4.
Press the
key or SHIFT +
key to select Output pen and then press the
key.
key.
5.
Set each item and press the
Use the
key or SHIFT +
key to select values.
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up
to then are cancelled, and the display returns to a higher level menu.
6.
When the Setting complete screen appears, do either of the following:
To correct the setting, press the
key.
If you are done, press the ESC key.
Basic=Math
Math=Output
Pen=1
ESC/?
Pen 1
Setting
pen
CH=01
Select the pen number and channel
number.
The new setting takes effect.
complete
Applying the Changes and Returning to Operation Mode
Press the
key or SHIFT +
key to select End and then press the
key. Press
the
key to select Store and then press the
key. The changes are applied, and
the screen returns to Operation mode. If you select Abort and press the
key, the
changes are discarded, and the screen returns to Operation mode. Press the ESC key
to return to the Basic= screen.
Explanation
Output pen
Measurement channels and computation channels can be arbitrarily assigned to
recording pens.
The default values are as follows:
Pen No. 1 (red): Channel 01
Pen No. 2 (green): Channel 02
Pen No. 3 (Blue): Channel 03
Pen No. 4 (violet): channel 04
9-24
IM 04P01B01-01E
9.15 Changing the Type of Report Data Printed in
Periodic Printout
This section explains the details of selecting the type of computed value (the average,
minimum, maximum, sum, or instantaneous) printed in periodic printout
Basic Setting mode cannot be entered when recording is in progress or when
computation is in progress on models with the computation function (/M1 option).
Procedure
Changing the Settings
1.
Hold down the MENU key for 3 seconds to enter Setting mode.
2.
Hold down the
and
keys simultaneously for 3 seconds to enter Basic
Setting mode.
3.
Press the
key or SHIFT +
key to select Math and then press the
key.
4.
Press the
key or SHIFT +
key to select Print2 and then press the
key.
5.
Set each item and press the
key.
Use the
key or SHIFT +
key to select values.
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up
to then are cancelled, and the display returns to a higher level menu.
6.
When the Setting complete screen appears, do either of the following:
To correct the setting, press the
key.
If you are done, press the ESC key.
Basic=Math
Math=Print2
First channel
ESC/?
Set the channel range.
Last channel
Mode=AVE
Select the type of report data to print.
SUM
Select the sum scale.
scale=Off
Periodic
Setting complete
The new setting takes effect.
Applying the Changes and Returning to Operation Mode
Press the
key or SHIFT +
key to select End and then press the
key. Press
the
key to select Store and then press the
key. The changes are applied, and
the screen returns to Operation mode. If you select Abort and press the
key, the
changes are discarded, and the screen returns to Operation mode. Press the ESC key
to return to the Basic= screen.
IM 04P01B01-01E
9-25
Operations Related to the Computation Function (/M1 Option)
CH=0A-0A
9
9.15 Changing the Type of Reported Data Printed in Periodic Printout
Explanation
Mode
Sets the type of report data to be printed.
AVE: Prints the average value over the interval.
MIX: Prints the minimum, maximum, and average values over the interval.
SUM: Prints the sum value over the interval. If SUM is selected, set the sum scale.
MIN: Prints the minimum value over the interval.
MAX: Prints the maximum value over the interval.
INST: Prints the instantaneous value.
SUM scale
Set this item only when SUM is selected.
For the definition of the term “sum scale,” see page 7-14.
Off:
Simply sums the computed values.
/s:
Sums by converting the computed values to a value over 1 second.
/min: Sums by converting the computed values to a value over 1 minute.
/h:
Sums by converting the computed values to a value over 1 hour.
/day: Sums by converting the computed values to a value over 1 day.
<Related Topics>
Setting the periodic printout interval: Section 7.8
9-26
IM 04P01B01-01E
9.16 Setting the Bar Graph Display Mode
This section explains the details of setting the bar graph display mode.
Basic Setting mode cannot be entered when recording is in progress or when
computation is in progress on models with the computation function (/M1 option).
Procedure
Changing the Settings
1.
Hold down the MENU key for 3 seconds to enter Setting mode.
2.
Hold down the
and
keys simultaneously for 3 seconds to enter Basic
¥U‘ting mode.
3.
Press the
key or SHIFT +
key to select Math and then press the
key.
4.
Press the
key or SHIFT +
key to select Bar graph and then press the
key.
5.
Set each item and press the
key.
Use the
key or SHIFT +
key to select values.
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up
to then are cancelled, and the display returns to a higher level menu.
6.
When the Setting complete screen appears, do either of the following:
Press the
key to set other channels.
If you are done, press the ESC key.
Basic=Math
Math=Bar
graph
9
First channel
ESC/?
Set the channel range.
Last channel
Graph=Normal
Select the display mode of the bar
graph.
0A-0A Channel
Setting complete
The new setting takes effect.
Applying the Changes and Returning to Operation Mode
Press the
key or SHIFT +
key to select End and then press the
key. Press
the
key to select Store and then press the
key. The changes are applied, and
the screen returns to Operation mode. If you select Abort and press the
key, the
changes are discarded, and the screen returns to Operation mode. Press the ESC key
to return to the Basic= screen.
Explanation
Graph
Normal: Sets the base point of the bar graph to the left or right of the recording span
whichever is smaller in value.
Center: Sets the base point of the bar graph to the 50% position of the recording span.
IM 04P01B01-01E
9-27
Operations Related to the Computation Function (/M1 Option)
CH=0A-0A
9.17 Setting the Procedure Taken When the
Computed Result Is in Error
This section explains the details of setting the procedure taken when the computed result
is in error.
Basic Setting mode cannot be entered when recording is in progress or when
computation is in progress on models with the computation function (/M1 option).
Procedure
Changing the Settings
1.
Hold down the MENU key for 3 seconds to enter Setting mode.
2.
Hold down the
and
keys simultaneously for 3 seconds to enter Basic
Setting mode.
3.
Press the
key or SHIFT +
key to select Math and then press the
key.
4.
Press the
key or SHIFT +
key to select Error data and then press the
key.
5.
Set each item and press the
key.
Use the
key or SHIFT +
key to select values.
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up
to then are cancelled, and the display returns to a higher level menu.
6.
When the Setting complete screen appears, do either of the following:
Press the
key to set other channels.
If you are done, press the ESC key.
Basic=Math
Math=Error
ESC/?
data
Error=+Over
Select the computation error procedure.
Over=Skip
Select the input overflow procedure.
Math exception
Setting complete
The new setting takes effect.
Applying the Changes and Returning to Operation Mode
Press the
key or SHIFT +
key to select End and then press the
key. Press
the
key to select Store and then press the
key. The changes are applied, and
the screen returns to Operation mode. If you select Abort and press the
key, the
changes are discarded, and the screen returns to Operation mode. Press the ESC key
to return to the Basic= screen.
9-28
IM 04P01B01-01E
9.17 Setting the Procedure Taken When the Computed Result Is in Error
Explanation
Error
Specifies how to display or print the computed result when computation errors occur.
+Over: Set to +over. Displayed/Printed as “+Over.”
–Over: Set to –over. Displayed/Printed as “–Over.”
Computation that result in errors
• x/0
• SQR(-x)
• LOG(-x)
• When a channel set to Skip is included in the equation
Over
Selects the procedure when an “over” value* is input for TLOG.SUM or TLOG.AVE
computation.
*
An “over” value on a measurement channel refers to +Over or –Over. See section 12.4. An
“over” value on a computation channel refers to a value that is less than or equal to
–100000000 or greater than or equal to 100000000 excluding the decimal.
Skip: The “over” value is not used in the computation.
Limit: The following limit value is used for the computation.
Channel
Limit Value
Measurement channel
• For scaling (1-5V, scaling, and square root)
–Over: –315000, +Over: 31500 (excluding the decimal)
• For input types other than scaling
–Over: The minimum measurable value at the specified range type
(Example: –2.000 at 2V range)
+Over: The maximum measurable value at the specified range type
(Example: 2.000 at 2V range)
Computation channel
–Over: –100000000, +Over: 100000000 (excluding the decimal)
9
Operations Related to the Computation Function (/M1 Option)
IM 04P01B01-01E
9-29
Chapter 10 Troubleshooting
10.1 A List of Error Messages
There are cases in which error codes and messages are displayed on the screen during
operation. The error messages and their description are listed below.
Setting Errors
Code
Message
1
System error.
Contact your nearest YOKOGAWA dealer.
2
Incorrect date or time setting.
Check the setting.
3
A disabled channel is selected.
The channel does not exist.
4
Incorrect function parameter.
Incorrect communication parameter.
5
The input numerical value exceeds the set range.
–
6
Incorrect input character string.
The entered character cannot be used.
7
Too many characters.
–
8
Incorrect input mode.
Incorrect range mode (Volt, TC, Scale, etc.) setting.
9
Incorrect input range code.
Incorrect range type (2 V, R, PT100, etc.) setting.
10
Format error.
Incorrect character string format.
11
Range settings are not same within the selected
channels.
Channels whose range differs cannot be set simultaneously.
12
An invalid characters.
Contains an invalid character.
13
Ref. CH error.
Specify a channel whose range is set to voltage, TC, or RTD for
the reference channel.
21
Cannot set an alarm for a SKIPPED channel.
–
22
The upper and lower span limits are equal.
This is not allowed.
23
The upper and lower scale limits are equal.
This is not allowed.
24
The lower span limit is greater than the upper span
limit.
–
25
The lower scale limit is greater than the upper scale –
limit.
26
Bias cannot be set to the SKIPPED channel.
–
27
Bias cannot be set to the DI channel.
–
30
The partial boundary value exceeds the range of the –
span.
31
Partial is invalid on the SKIPPED channel.
–
35
The upper and lower limits of the printing zone
are equal.
Set the rightmost value of the zone – the leftmost value ≥ 5
mm.
36
The lower limit of the printing zone is greater
than the upper limit.
Set the rightmost value of the zone – the leftmost value ≥ 5
mm.
37
The printing zone is narrower than the minimum
width (5 mm).
Set the rightmost value of the zone – the leftmost value ≥ 5
mm.
38
Partial is invalid on the DI channel.
–
47
All items in DISP menu parameters are set to SKIP. –
48
Start = Finish.
The DST start time and end time cannot be set to the same time.
49
Invalid or missing DST time settings.
Since the time gains one hour when the DST starts, the set-up
time does not exist.
61
There is no channel specified by the MATH
expression.
Set a computation channel.
62
MATH expression grammar is incorrect.
Enter the equation correctly.
63
MATH expression sequence is incorrect.
Enter the equation correctly.
64
MATH upper and lower span values are equal.
This is not allowed.
70
MATH constant description is incorrect.
Incorrect computation constant syntax.
71
The range of the MATH constant is exceeded.
–
10
Troubleshooting
IM 04P01B01-01E
Explanation/Countermeasures
10-1
10.1 A List of Error Messages
Code
Message
72
MATH channel is turned off
Explanation/Countermeasures
–
81
All space or ‘quit’ string cannot be specified.
–
86
The key-lock release password is incorrect.
Enter the correct password.
87
This key is locked.
–
91
Password is incorrect.
Enter the correct password.
100
IP address doesn’t belong to class A, B, or C.
–
101
The result of the masked IP address is all 0s or 1s.
–
102
SUBNET mask is incorrect.
Set a correct subnet mask.
103
The net part of default gateway is not equal to
that of IP address.
Set the correct default gateway.
151
This action is invalid during calculation.
–
160
This action is invalid during chart end.
–
161
This action is invalid during pen hold.
–
162
Cannot set an number for a skipped data.
–
163
This action is invalid during record.
–
164
This action is invalid during manual printing.
–
165
This action is invalid during list printing.
–
166
This action is invalid during setup list printing.
–
167
This action is invalid during chart feed.
–
169
This action is invalid during ribbon hold.
–
Operation Errors
Code
Message
Explanation/Countermeasures
232
There is no available data.
There is no data for periodic printout or data for calculating TLOG
when the timer expired.
Communication Errors
Code
Message
Explanation/Countermeasures
300
Command is too long.
–
301
Too many number of commands delimited with ’;’.
Keep the number of commands separated by sub delimiters
under 10.
302
This command has not been defined.
–
303
Data request command can not be enumerated with –
sub-delimiter.
350
Command is not permitted to the current user level.
–
351
This command cannot be specified in the current
mode.
–
352
The option is not installed.
–
353
This command cannot be specified in the current
setting.
–
354
This command is not available during calculation.
–
390
Command error.
–
391
Delimiter error.
–
392
Parameter error.
–
393
No permission.
–
394
No such connection.
–
395
Use “quit” to close this connection.
Attempted to disconnect its own connection.
396
Failed to disconnect.
–
397
No TCP control block.
The control block of the specified connection cannot be found.
10-2
IM 04P01B01-01E
10.1 A List of Error Messages
Code
Message
Explanation/Countermeasures
400
Input username.
–
401
Input password.
–
402
Select username from ‘admin’ or ‘user’.
If the recorder is configured not to use the user name and
password, use user names ‘admin’ or ‘user’.
403
Login incorrect, try again!
–
404
No more login at the specified level is acceptable.
–
420
Connection has been lost.
–
421
The number of simultaneous connection has been
exceeded.
–
422
Communication has timed-out.
–
Warning Messages
Code
Message
Explanation/Countermeasures
600
Initialized.
Settings and measured data have been initialized.
System Errors
Message
Explanation/Countermeasures
902
RAM failure.
Contact your nearest YOKOGAWA dealer.
910
A/D error.
Contact your nearest YOKOGAWA dealer.
921
A/D calibration value error.
Contact your nearest YOKOGAWA dealer.
922
A/D calibration is in the wrong order.
Contact your nearest YOKOGAWA dealer.
930
Memory acquisition failure.
Contact your nearest YOKOGAWA dealer.
940
The ethernet module is down.
Contact your nearest YOKOGAWA dealer.
950
A/D number error.
Contact your nearest YOKOGAWA dealer.
951
EEPROM write error.
Contact your nearest YOKOGAWA dealer.
960
Ribbon error
Contact your nearest YOKOGAWA dealer.
961
Printer error
Contact your nearest YOKOGAWA dealer.
962
Plotter error
Contact your nearest YOKOGAWA dealer.
963
Pen 1 error
Contact your nearest YOKOGAWA dealer.
964
Pen 2 error
Contact your nearest YOKOGAWA dealer.
965
Pen 3 error
Contact your nearest YOKOGAWA dealer.
966
Pen 4 error
Contact your nearest YOKOGAWA dealer.
IM 04P01B01-01E
10
Troubleshooting
Code
10-3
10.2 Troubleshooting Flow Charts
Follow the flow charts to correct the problem.
Nothing works (pen and chart paper do not move).
Is the power switch
turned ON? Does the switch
work properly?
NO
Turn ON the power properly.
Section 2.6
Connect properly.
Section 2.5
YES
Is the power supply
connected properly?
NO
YES
Is the power supplied
properly?
NO
Supply the power at the specified
voltage and frequency.
YES
Contact your nearest
YOKOGAWA dealer.
• The reading error is large.
• The indication is unstable.
• The pen is off the scale on either the 0% or 100% side.
Does the input meet
the specifications?
NO
Change the input to meet
the specifications.
YES
Are the span and range
settings correct?
NO
Adjust the span and range.
YES
To the next page
10-4
IM 04P01B01-01E
10.2 Troubleshooting Flow Charts
From the previous page
YES
Is the input free of noise?
NO
YES
Is the environmental
temperature kept constant?
NO
•
•
•
•
•
Separate the input wiring from the noise source.
Ground the recorder properly.
Ground the measured object properly.
Insulate the TC from the measured object.
Use shielded cables for input.
Select the integration time of the A/D converter
(See section 7.2).
• Use input filter (pen model: see sections 7.12 and
6.2), moving average (dot model: see sections
7.11 and section 6.3).
• Attach the input terminal cover properly.
• Protect the input terminal from wind or fans.
• Keep the temperature fluctuation near the input
terminal to a minimum.
YES
• Wire correctly (see section 2.3).
Is the input wiring correct?
NO
YES
Is the input wiring parallel
with other equipment?
NO
YES
Is the RJC setting
correct (TC input)?
NO
•
•
•
•
Install the terminal block correctly.
Fasten the terminal screws correctly.
Insulate the RTD from ground.
If burnt, replace the TC.
• Stop the TC burnout upscale/
downscale function of other equipment.
• Ground the recorder and other equipment at the
same grounding pole.
• Disconnect parallel wiring.
(Use a double-element TC, for instance.)
Set correctly.
Section 7.4
10
YES
NO
Adjust the pen position or dot printing position
Section 11.5
Section 11.6
YES
Contact your nearest
YOKOGAWA dealer.
IM 04P01B01-01E
10-5
Troubleshooting
Is the pen
position and dot
printing position adjustment
correct?
10.2 Troubleshooting Flow Charts
Display and other functions do not work properly.
Is the input free of noise?
NO
YES
• Separate the input and
communication wiring from the
noise source.
• Ground the recorder properly.
• Use shielded cables for input and
communication wiring.
• Insert an isolation transformer
into the power line.
• Insert a power line filter into the
power line.
• If an inductive load is connected
to an alarm contact output, use a
surge suppresser on that line.
Contact your nearest
YOKOGAWA dealer.
Keys do not work.
Is the key lock released?
NO
Release the key lock using the
FUNC key (see section 3.13).
YES
Contact your nearest
YOKOGAWA dealer.
10-6
IM 04P01B01-01E
Chapter 11 Maintenance
11.1 Periodic Inspection
Check the operation periodically to keep the recorder in good working order.
Perform the following checks and replace worn parts as needed.
• Is the indication and recording functioning properly? If not, see chapter 10.
• Are recorded lines or printed characters clear (not blurred)?
For the replacement procedure of felt pens and plotter pens, see section 3.3 (pen
model).
For the replacement procedure of the ribbon cassette, see section 3.4 (dot model).
• Is the chart paper feeding properly (no paper jams)? If not, see chapter 10.
• Is there enough chart paper remaining?
Remaining chart length is printed on the left margin of the chart at intervals of 20 cm.
For the chart paper replacement procedure, see section 3.2.
Remaining amount of chart paper
11
Maintenance
IM 04P01B01-01E
11-1
11.2 Cleaning the Recorder
CAUTION
• When cleaning, be sure not to scratch the flexible printed circuit board of the
plotter carriage.
• Do not apply lubricating oil to the shaft.
Pen Model
To maintain smooth operation, it is recommended that the plotter carriage shaft be
cleaned once a year.
Cleaning Procedure
Wipe the dust off the plotter carriage shaft using a soft lint-free cloth or paper.
If the dirt does not come off easily, apply ethyl alcohol to the cloth or paper.
Dot Model
To maintain smooth operation, wipe the dust off the two shafts of the printer carriage
using a soft lint-free cloth or paper. It is recommended that these shafts be cleaned
once a year.
11-2
IM 04P01B01-01E
11.3 Replacing the Internal Light LED
If the brightness of the internal light degrades with time, replace the LED.
For the replacement parts, see section 11.7.
CAUTION
• The LED is connected to the internal circuitry through the flexible printed circuit
board. Do not pull the LED with force.
• Be sure not to break the flexible printed circuit board.
Replacement Procedure
1.
2.
3.
Open the door and turn OFF the power switch.
Open the display and key panel section.
Pull out the internal light LED assembly that is attached to the end of the display
and key panel section.
Display and key panel
4.
Remove the LED assembly from the connector of the flexible printed circuit
board.
Flexible printed circuit board
LED assembly
Connector
11
Connect the new LED assembly.
Push the LED assembly back into the display and key panel section.
Be sure not to bend or scratch the flexible printed circuit board.
Maintenance
5.
6.
Stored condition of the flexible printed circuit board
Flexible printed circuit board
7.
IM 04P01B01-01E
Turn the power switch ON and check that the internal light comes on a few
seconds later.
11-3
11.4 Calibrating the Recorder
Calibrate the measured value against the input.
It is recommended that the recorder be calibrated once a year to assure its
measurement accuracy.
For details regarding calibration, contact your nearest YOKOGAWA dealer.
Required Instruments
A calibration instrument with appropriate resolution is required for calibrating the
recorder.
Recommended Instrument
• DC voltage standard: Model 5520A by FLUKE or equivalent
Main Specifications
Output accuracy: ±(0.005% + 1 µV)
• Decade resistance box: Yokogawa M&C Model 2793-01 or equivalent
Main Specifications
Accuracy of output range 0.1 to 500 Ω: ±(0.01% + 2 mΩ)
Resolution: 0.001 Ω
• 0°C standard temperature device: ZC-114/ZA-10 by Coper Electronics or equivalent
Main Specifications
Standard temperature stability accuracy: ±0.05°C
For information on purchasing the calibration instruments, contact your nearest
YOKOGAWA dealer.
Calibration Procedure
1.
2.
3.
Wire the recorder and the calibration instrument as shown in the following figure,
and adequately warm up the instruments (the warm-up time of the recorder is at
least 30 minutes).
Check that the operating environment such as ambient temperature and
humidity is within the normal operating conditions (see section 12.6).
Apply appropriate input signals corresponding to 0%, 50%, and 100% of the
input range and calculate the errors from the readings.
If the error is outside the accuracy specifications, contact your nearest
YOKOGAWA dealer.
Note
For thermocouple inputs, you must measure the temperature of the input terminal and apply a
voltage taking into account the reference junction temperature.
DC Voltage Measurement
–
+
+
–
DC voltage standard
Input terminals
11-4
IM 04P01B01-01E
11.4 Calibrating the Recorder
Temperature Measurement When Using an RTD
The resistance of three lead wires must be equal.
/b −/B +/A
Decade resistance box
(Model 2793-01 from Yokogawa M&C)
Input terminals
Temperature Measurement When Using a Thermocouple
Copper wires
Thermocouple wires or TC extension wires
−
+
+
–
DC voltage standard
Copper
wires
Thermocouple
Input terminals
wires
(0°C standard temperature device ZC-114/ZA-10 by Coper Electronics)
IM 04P01B01-01E
11-5
11
Maintenance
RJC of TC input
As the measurement terminal of the recorder is generally at room temperature, the
actual output of the thermocouple is different from the values given on the
thermoelectromotive force table based on 0°C. The recorder performs compensation by
measuring the temperature at the input terminal and adding the corresponding
thermoelectromotive force to the actual output of the thermocouple. Therefore, when the
measurement terminal is shorted (equivalent to the case when the detector tip is 0°C),
the measured value indicates the temperature of the input terminal.
When calibrating the recorder, this compensation voltage (thermoelectromotive force of
0°C reference corresponding to the input terminal temperature) must be subtracted from
the output of the DC voltage standard before application. As shown in the figure, by
using the 0°C standard temperature device to compensate the reference junction at 0 °C,
you can input the thermoelectromotive force of 0 °C reference from the DC voltage
standard and perform the calibration.
11.5 Adjusting the Pen Position (Pen Model)
This section explains the details of adjusting the pen position on the chart paper.
It is recommended that the pen be adjusted once a year to assure its recording accuracy.
Note
Pen position is adjusted under standard operating conditions when the recorder is shipped from
the factory. Because the chart paper may expand or contract depending on the environment in
which the recorder is used, it is recommended that the position be readjusted before use.
Procedure
1.
2.
3.
4.
5.
6.
7.
8.
Warm up the recorder for at least 30 minutes.
Check that the operating environment such as ambient temperature and
humidity is within the normal operating conditions (see section 12.6).
Hold down the MENU key for 3 seconds to enter Setting mode.
Hold down the
and
keys simultaneously for 3 seconds to enter Basic
Setting mode.
Press the
key or SHIFT+
key to select P_Adj and then press the
key.
Select the Pen No.
key to select Zero (left edge of the chart paper) or Full (right
Press the
edge of the chart paper) and then press the
key.
Adjust Zero first and then Full.
Increase or decrease the setting to align the pen position with the scale line on
the chart paper. Then, press the
key. To change the setting, press the
key to move the cursor and
key to change the value.
The pen moves after the
key is pressed.
A value change of 1 corresponds to 0.033 mm change in the pen position.
Decreasing the value moves the pen to the left and increasing the value moves
the pen to the right.
Press the
key to set adjust other pens.
Press the ESC key to finish the adjustment. The screen returns to the
Basic=P_Adj screen.
Basic=P_Adj
P_Adj=Zero
Select Zero or Full.
Pen
Select the pen No.
No.=1
Change the value to adjust the pen
position.
Zero=40
ESC/?
P_Adj
Setting
The pen moves.
complete
Applying the Changes and Returning to Operation Mode
key or SHIFT +
key to select End and then press the
key. Press
Press the
the
key to select Store and then press the
key. The changes are applied, and
key, the
the screen returns to Operation mode. If you select Abort and press the
changes are discarded, and the screen returns to Operation mode. Press the ESC key
to return to the Basic= screen.
11-6
IM 04P01B01-01E
11.6 Adjusting the Dot Printing Position (Dot Model)
This section explains the details of adjusting the dot printing position on the chart paper.
It is recommended that the position be adjusted once a year to assure its recording
accuracy.
Note
Dot printing position is adjusted under standard operating conditions when the recorder is
shipped from the factory. Because the chart paper may expand or contract depending on the
environment in which the recorder is used, it is recommended that the position be readjusted
before use.
Procedure
1.
2.
3.
4.
5.
6.
Warm up the recorder for at least 30 minutes.
Check that the operating environment such as ambient temperature and
humidity is within the normal operating conditions (see section 12.6).
Hold down the MENU key for 3 seconds to enter Setting mode.
Hold down the
and
keys simultaneously for 3 seconds to enter Basic
Setting mode.
key or SHIFT +
key to select P_Adj and then press the
Press the
key.
Press the
key to select Hysteresis (center of the chart paper), Zero (left
edge of the chart paper), or Full (right edge of the chart paper) and then press
the
key.
Adjust Hysteresis first followed by Zero and then Full.
Adjusting the Hysteresis
A line is drawn at the center of the chart paper. If the line appears as shown in Figure
a, increase the displayed setting. Then press the
key. If the line appears as shown
in Figure b, decrease the displayed setting. Then, press the
key.
To change the setting, press the
key to move the cursor and
key to change
the value.
Repeat this procedure until the line becomes straight.
A value change of 1 corresponds to 0.1 mm change in the dot printing position.
Up
Up
Down
Down
11
Figure b
Adjusting Zero and Full
A line is drawn to the left edge for Zero and right edge for Full.
Increase or decrease the displayed setting so that the drawn line is aligned with the
scale line on the chart paper. Then, press the
key. The procedure of changing the
setting is the same as with the adjustment of hysteresis.
The line moves after the
key is pressed.
A value change of 1 corresponds to 0.1mm change in the dot printing position.
Decreasing the value moves the line to the left and increasing the value moves the
line to the right.
IM 04P01B01-01E
11-7
Maintenance
Figure a
11.6 Adjusting the Dot Printing Position (Dot Model)
7.
Press the ESC key to finish the adjustment. The screen returns to the
Basic=P_Adj screen.
Basic=P_Adj
ESC/?
P_Adj=Hysteresis
Select Zero or Full.
Hysteresis=-2
Change the value to adjust the pen
position.
P_Adj
Setting
The dot printing position moves.
complete
Applying the Changes and Returning to Operation Mode
Press the
key or SHIFT +
key to select End and then press the
key. Press
key to select Store and then press the
key. The changes are applied, and
the
the screen returns to Operation mode. If you select Abort and press the
key, the
changes are discarded, and the screen returns to Operation mode. Press the ESC key
to return to the Basic= screen.
11-8
IM 04P01B01-01E
11.7 Recommended Replacement Periods for Worn
Parts
To preserve the reliability of the recorder and to use the recorder in good condition for an
extended time, it is recommended that periodic replacements be made on parts.
The table below shows the recommended replacement period for expendable parts.
The replacement period shown here applies when the recorder is used under standard
operating conditions. For the actual replacement period, consider the actual conditions
of use.
Replacement of parts other than the chart paper, pen, ribbon cassette, and internal light
LED will be carried out by a YOKOGAWA engineer or an engineer certified by
YOKOGAWA. Contact your nearest YOKOGAWA dealer when such replacement is
necessary.
Note
The replacement period of the display and the internal light LED is the half life of the
brightness. The deterioration of brightness varies depending on the condition of use, and its
determination is subjective. Consider these facts for determining the actual replacement
period.
Pen Model
Replacement
Period
Item
Part No.
Note
Z-fold chart paper
33 days
CHART
B9565AW
When used at 20 mm/h
Felt pen
2 km
PEN ASSY
B9902AM
B9902AN
B9902AP
B9902AQ
Red
Green
Blue
Violet
PEN ASSY
B9902AR
When printing continuously
100,000
characters
Plotter pen
Quantity
Used
1
1 each
At a pen speed of
10 cm/s
1
5 years*
DISPLAY ASSY
B8800CA
1
Chart paper feed
motor
5 years
MOTOR ASSY
B9962EJ
1
Plotter carriage
5 years
CARRIAGE ASSY B8800DJ
1
Plotter motor
5 years
MOTOR ASSY
B8800DT
For the X-axis
1
Lever bearing
5 years
BEARING
B9900RP
For the plotter
1
Pen servo
5 years
SERVO ASSY
B8800FG
Shared by all pens
(excludes the pen arm ASSY)
Internal light LED
2 years*
LED ASSY
B8800CR
1 to 4
11
1
Maintenance
Display
*
IM 04P01B01-01E
Part Name
The half life of the brightness at the factory default brightness setting.
11-9
11.7 Recommended Replacement Periods for Worn Parts
Dot Model
Replacement
Period
Item
Part No.
Note
Quantity
Used
Z-fold chart paper
33 days
CHART
B9565AW
Ribbon cassette
3 months
RIBBON CASSETTE
B9901AX
1
Display
5 years*
DISPLAY ASSY
B8800CA
1
Chart paper feed
motor
5 years
MOTOR ASSY
B9962EJ
1
Lever
3 years
LEVER ASSY
B9901EK
For the carriage
1
Pulley
3 years
PULLY
B9963CJ
For the carriage
2
Carriage motor
5 years
MOTOR ASSY
B9963CF
1
Carriage
5 years
CARRIAGE ASSY
B9963CL
1
Ribbon shift motor
5 years
MOTOR ASSY
B9962EJ
1
Ribbon shift gear
5 years
GEAR
B8801BX
B8801BW
1 each
Ribbon feed motor 5 years
MOTOR ASSY
B9962EJ
1
Ribbon feed gear
5 years
GEAR
B9901HL
B9901HM
B9901HN
1 each
Internal light LED
2 years*
LED ASSY
B8800CR
1
*
11-10
Part Name
When used at 20 mm/h
1
The half life of the brightness at the factory default brightness setting.
IM 04P01B01-01E
Chapter 12 Specifications
12.1 Input Specifications
Number of Inputs and Scan Interval on the Pen Model
Item
Specifications
Number of inputs
1, 2, 3, or 4
Scan interval
125 ms
Number of Inputs and Scan Interval on the Dot Model
Item
Specifications
Number of inputs
6
Scan interval
1 s (when the integration time of the A/D converter is 20 ms or 16.7 ms)
2.5 s (when the integration time of the A/D converter is 100 ms)
Input Type
Item
Specifications
Input type
DC voltage, 1-5V, TC, RTD, DI (ON/OFF input), DC current (by adding an external shunt resistor)
Input Range and Measurable Range
Input
Range Type
Measurable Range
Volt
20 mV
60 mV
200 mV
2V
6V
20 V
50 V
1-5V
–20.00 to 20.00 mV
–60.00 to 60.00 mV
–200.0 to 200.0 mV
–2.000 to 2.000 V
–6.000 to 6.000 V
–20.00 to 20.00 V
–50.00 to 50.00 V
(0.800 to 1.200 V range) to (4.800 to 5.200 V range)
TC
R1
S1
B1
K1
E1
J1
T1
N1
W2
L3
U3
WRe4
0.0 to 1760.0 °C
0.0 to 1760.0 °C
0.0 to 1820.0 °C
−200.0 to 1370.0 °C
−200.0 to 800.0 °C
−200.0 to 1100.0 °C
−200.0 to 400.0 °C
0.0 to 1300.0 °C
0.0 to 2315.0 °C
−200.0 to 900.0 °C
−200.0 to 400.0 °C
0.0 to 2400.0 °C
32 to 3200 °F
32 to 3200 °F
32 to 3308 °F
−328 to 2498 °F
−328.0 to 1472.0 °F
−328.0 to 2012.0 °F
−328.0 to 752.0 °F
32 to 2372 °F
32 to 4199 °F
−328.0 to 1652.0 °F
−328.0 to 752.0 °F
32 to 4352 °F
RTD
PT (Pt100)5
JPT (JPt100)5
–200.0 to 600.0°C
–200.0 to 550.0°C
−328.0 to 1112.0 °F
−328.0 to 1022.0 °F
DI
Level
Contact
0: Less than 2.4 V, 1: Greater than or equal to 2.4 V
0: Open, 1: Closed
IM 04P01B01-01E
12
R, S, B, K, E, J, T, N: IEC584-1 (1995), DIN IEC584, JIS C1602-1995
W: W-5% Rd/W-26% Rd (Hoskins Mfg. Co.), ASTM E988
L: Fe-CuNi, DIN43710, U: Cu-CuNi, DIN43710
WRe: W-3%Re/W-25%Re (Hoskins Mfg Co.)
Pt100: JIS C1604-1997, IEC751-1995, DIN IEC751-1996
JPt100: JIS C1604-1989, JIS C1606-1989
Measurement current: i = 1 mA (Pt100, JPt100)
Specifications
1
2
3
4
5
12-1
12.1 Input Specifications
Item
Specifications
Input type
Floating unbalanced input.
Isolation between channels (except, b terminal is shared for RTD input).
Burnout Detection of TC
Can be set for each channel.
Upscale/Downscale switchable.
RJC of TC input
A/D converter
Resolution
Integration time
Input Type
Operating Conditions
TC input
2 kΩ or less: normal, 10 MΩ or more: burnout, detection current: approx. 10 µA
1-5V input
0.2 V or less: burnout
Set whether to use the recorder RJC function or an external RJC function for each channel.
Set the compensation voltage when using an external RJC function.
Compensation voltage range: –20000 µV to 20000 µV.
16 bits
Pen model: Selectable from 20 ms (50 Hz), 16.7 ms (60 Hz), and Auto (auto switch between 20 ms
and 16.7 ms according to the power supply frequency).
Dot model: Selectable from 20 ms (50 Hz), 16.7 ms (60 Hz), 100 ms and Auto (auto switch
between 20 ms and 16.7 ms according to the power supply frequency).
Filter function (pen model) Filter ON/OFF switchable for each channel.
Time constant selectable from 2, 5, and 10 s.
Moving average function
(dot model)
Moving average ON/OFF switchable for each channel.
The number of samples of the moving average is selectable between 2 and 16.
Input Computation (Standard Function)
Item
Specifications
Delta computation
Calculates the difference between two channel inputs using the following equation.
Delta = (the input value of the channel set to delta computation) – (the input value of the
reference channel)
where (reference channel NO.) < (channel No. set to delta computation).
Computable input type: DC voltage, TC, and RTD. The channel set to delta computation and the
reference channel is set to the same range type.
Linear scaling
Scales the input value to a value with a unit suitable for the application.
Computable input type: DC voltage, TC, RTD, DI
Value after scaling
Mantissa: –20000 to 30000
Decimal position: Can be set arbitrarily
Unit: Can be set arbitrarily (up to 6 characters)
Displayable and printable range
–19999 to 31500
Computation on the 1-5V input
Linear scaling
Linearly scales the 1-5V signal.
Value after scaling: Same as linear scaling.
Displayable and printable range: Same as linear scaling.
Low-cut function
Sets values below 0% of the recording span to the leftmost value of the scale. The leftmost value of
scale must be less than the rightmost value of scale.
Square root computation
Linear scaling
Low-cut function
Bias
12-2
Takes the square root of the input value and linearly scales the result.
Computable input type: DC voltage
Value after scaling: Same as linear scaling.
Displayable and printable range: Same as linear scaling.
Sets values below the low-cut point to the leftmost value of scale. The leftmost value of scale must
be less than the rightmost value of scale.
Low-cut point range: 0.0 to 5.0% of the recording span (0.1% steps)
Adds a bias to the measuring input value.
Bias value range: ±10% of the span of the measurable range at the specified range type.
±10% of the scaling span if linear scaling is used.
IM 04P01B01-01E
12.2 Alarm Function Specifications
Item
Specifications
Number of alarms
Up to four alarms (level) for each measurement channels.
Alarm types
High limit (H), low limit (L), difference high limit (h), difference low limit (l), high limit on rate-ofchange (R), low limit on rate-of-change (r), delay high limit (T), delay low limit (t). The symbol
indicating the alarm is given in parentheses.
Rate-of-change calculation interval of rate-of-change alarms (can be set individually for the high limit
on rate-of-change and low limit on rate-of-change alarms)
Rate-of-change calculation interval = Scan interval × the number of sampled data
where the number of sample data is a value between 1 and 15
Alarm delay time of delay high and low limit alarms (can be set for each channel, common within a
channel)
1 to 3600 s (1 hour)
However, if the alarm delay time setting is not an integer multiple of the scan interval, the alarm
delay time is set to a value that is the first integer multiple of the scan interval exceeding the
delay time setting.
Hysteresis
Set a width to the value for detecting alarm occurrence/release (common to all channels and all levels)
Applied to high limit alarm and low limit alarm.
Hysteresis range: Approx. 0.0% to 1.0% of the recording span (0.1% steps)
Display
Indicates the alarm occurrence status on the display.
Indication for each channel: Displays the alarm type (display precedence: H, L, T, t, h, l, R, and r).
Alarm summary display: Displayed using an icon.
Displays the alarm point mark and status on the bar graph.
Non-hold/hold operation of alarm indication
Turn ON/OFF the indication along with the alarm occurrence/release (non-hold) or hold the
indication until the alarm ACK operation.
Alarm relay contact output (optional specifications)
See section 12.5
12
Specifications
IM 04P01B01-01E
12-3
12.3 Recording Function Specifications
Trend Recording (Pen Model)
Item
Specifications
Recording pen
Disposable felt pen
Step response time
Approx. 1 s (using the IEC61143 measurement method)
Number of pens
Up to 4
Recording color
Pen 1: Red, pen 2: Green, pen 3: Blue, pen 4: Violet
Trend recording
Updates the data at the scan interval. Continuous recording.
Pen offset compensation
Records by compensating for the offset of each pen along the time axis.
Chart speed
5 to 12000 mm/h (82 steps).
Zone recording
Recording zone can be specified for each channel.
Recording width: 5 mm or more (1 mm steps)
Partial expanded recording Records by expanding the right or left side of the boundary position (the other section is reduced).
Boundary position: 1 to 99%
Boundary value: Within the recording span range
Trend Recording (Dot Model)
Item
Specifications
Recording method
Trend recording by a 6-color wire dot printer
Recording color
Channel 1: Purple, channel 2: Red, channel 3: Green, channel 4: Blue, channel 5: Brown,
channel 6: Black
Trend recording interval
Records at the following interval through a raster scan system.
Auto: Synchronizes to the chart speed and automatically sets to the interval
Fix:
10 s/6 channels (see section 12.5 for the recording interval on models with the
computation function (/M1 option)
Chart speed
1 to 1500 mm/h (1 mm steps)
Recording ON/OFF
Recording can be turned ON/OFF for each channel.
Zone recording
Same as the pen model.
Partial expanded recording Same as the pen model.
Chart Paper
Item
Specifications
Printing width
100 mm
Type/Length
Z-fold type. Approx. 16 m.
Chart paper feeding accuracy
Within ±0.1% After feeding 1000 mm and with respect to the print scale on the chart paper.
12-4
IM 04P01B01-01E
12.3 Recording Function Specifications
Printouts (Pen Model)
Item
Specifications
Recording pen (color)
Plotter pen (purple)
Alarm printout
Printout contents
Prints alarm occurrence/release.
Occurrence ( )/release ( ) marks, channel No. or tag, alarm type, alarm level, time, printout
buffer overflow mark.
Selectable from hour:minute, hour:minute:second, month:day:hour:minute,
month:day:hour:minute:second, and year:month:day:hour:minute:second.
Stores 8 occurrences of alarm information waiting to be printed.
Time printout format
Alarm printout buffer
Periodic printout
Interval
Mode
Printout contents
Message printout
Number of messages
Printout contents
Time printout format
Message printout buffer
Prints at specified time intervals from the reference time.
Reference time: 00 hour 00 minute to 23 hour 00 minute (1 hour steps, minute fixed)
Auto:
Automatically determined from the chart speed.
Manual: Select 10 min, 12 min, 15 min, 20 min, 30 min, 1 hour, 2 hours, 3 hours, 4 hours, 6 hours,
8 hours, 12 hours, or 24 hours.
Selectable from disable periodic printout, print instantaneous values, and print report data (report mode).
Report data: Selectable from the average, minimum, maximum, minimum/maximum/average, sum
value, and instantaneous value.
For instantaneous values: Date/Time, offset compensation ON/OFF, channel information
(measured value, alarm status, unit, delta computation ON/OFF, scale,
channel number or tag, recording color), chart speed (with time ticks)
For report data: Report start/end date/time, offset compensation ON/OFF, events, channel
information (value, unit, range change occurrence, scale, recording color), chart
speed (with time ticks)
Prints preset message strings.
5
Time, message (up to 16 characters), printout buffer overflow mark.
Selectable from hour:minute, hour:minute:second, month:day:hour:minute,
month:day:hour:minute:second, year:month:day:hour:minute:second, and not print.
Stores 5 messages waiting to be printed.
Recording start printout
Printout contents
Time printout format
Prints at the start of recording.
Time and chart speed, printout buffer overflow mark.
Selectable from hour:minute, hour:minute:second, month:day:hour:minute,
month:day:hour:minute:second, and year:month:day:hour:minute:second.
New chart speed printout
Printout contents
Time printout format
Prints when the chart speed is changed.
Chart speed and time, printout buffer overflow mark.
Selectable from hour:minute, hour:minute:second, month:day:hour:minute,
month:day:hour:minute:second, and year:month:day:hour:minute:second.
Manual printout
Printout contents
Prints the measured values of all channels.
Measured values of all channels at that point.
Trend recording is temporarily suspended.
Setting printout (List)
Printout contents
Prints the settings of items in Setting mode.
Range setting, alarm setting, etc.
Trend recording is temporarily suspended.
Setting printout (Setup)
Printout contents
Prints the settings of items in Basic Setting mode.
Basic alarm specifications, basic recording specifications, etc.
Trend recording is temporarily suspended.
Items dependent on the chart speed
Periodic printout
5 to 9 mm/h
10 to 1500 mm/h
1600 mm/h or higher
Not printed
Printed
Not printed
Chart speed
10 to 18 mm/h
20 to 36 mm/h
40 to 72 mm/h
75 to 135 mm/h
150 to 180 mm/h
200 to 320 mm/h
350 to 1500 mm/h
Date printout format
IM 04P01B01-01E
Alarm printout
Message printout
New chart speed printout
Printed
Printed
Not printed
12
Specifications
Chart speed
Periodic printout interval
8 hours
4 hours
2 hours
1 hour
30 minutes
20 minutes
10 minutes
Select the printout format of the date from Y/M/D, M/D/Y, D/M/Y, D.M.Y, and M.D.Y. The format
applies to the date printout of all printout items (version 1.11 and later).
12-5
12.3 Recording Function Specifications
Printout (Dot Model)
Item
Specifications
Recording
Dot printing.
Channel printout
Prints the channel number beside the trend recording every approx. 25 mm of chart paper.
Channel printout ON/OFF selectable.
Alarm printout
Printout contents
Prints alarm occurrence/release.
Occurrence ( , red)/release ( , blue) marks, channel No. or tag, alarm type, alarm level, time,
printout buffer overflow mark.
Selectable from hour:minute, hour:minute:second, month day hour:minute,
month day hour:minute:second, and month day year hour:minute:second.
Stores 12 occurrences of alarm information waiting to be printed.
Time printout format
Alarm printout buffer
Periodic printout
Interval
Measured value
Printout contents
Prints at specified time intervals from the reference time.
Same as the pen model.
Same as the pen model.
For instantaneous values: Same as the pen model except offset compensation ON/OFF and
recording color are not printed.
For report data:
Same as the pen model except offset compensation ON/OFF and
recording color are not printed.
Message printout
Same as the pen model.
Recording start printout
Same as the pen model. With time ticks.
New chart speed printout
Same as the pen model. With time ticks.
Manual printout
Same as the pen model.
Setting printout (List)
Same as the pen model.
Setting printout (Setup)
Same as the pen model.
Items dependent on the chart speed
Chart speed
Channel printout
Periodic printout
1 to 9 mm/h
10 to 100 mm/h
101 to 1500 mm/h
Printed
Printed
Not printed
Not printed
Printed
Not printed
Chart speed
10 to 19 mm/h
20 to 39 mm/h
40 to 79 mm/h
80 to 100 mm/h
Date printout format
Alarm printout
Message printout
New chart speed printout
Printed
Printed
Not printed
Periodic printout interval
8 hours
4 hours
2 hours
1 hour
Same as the pen model.
Values for Special Measured Values
Same as “Special values” for the digital display on page 12-8.
12-6
IM 04P01B01-01E
12.4 Display Function Specifications
Display and Displayed Contents
This section explains the specifications of the display functions of measurement channels and computation channels (/M1 option).
Display examples are illustrations used to explain the displayed contents and differ in appearance from the actual displays.
Item
Specifications
Display
VFD (101 × 16, dot matrix)
Up to 11 characters using large fonts. 17 characters × 2 lines using normal fonts.
Number of screens
15 (switched using keys)
Displayed information
Display type
1-channel digital display
2-channel digital display
4-channel digital display
6-channel digital display (dot model)
1-channel digital + 1-channel bar graph
1-channel digital + 4-channel bar graph (pen model)
2-channel digital + 2-channel bar graph
4-channel bar graph (pen model)
6-channel bar graph (dot model)
Flag display
Channel alarm status display
Date, time, and chart speed display
DI/DO status display
1-channel digital display (tag display)
2-channel digital display (tag display)
1-channel digital + 1-channel bar graph (tag display)
1-channel digital + 4-channel bar graph (tag display) (pen model)
Status display
System display
Lights off
Skip
Multiple display
The following screens can be assigned arbitrarily to the top and bottom sections of the display.
1-channel digital display
2-channel digital display
Date/Time
Chart speed display
Channel alarm status display
Status display
Lights off
1-channel digital display (tag display)
Display update of measured values / computed values
When displayed channels are fixed
Dot model: Updated at the scan interval
Pen model: Updated every 2 s
When displayed channels are automatically switched
Switches the channel and measured values at 1, 2, 3, 4, or 5 second intervals.
Status display
RECORD
ALARM
KEYLOCK
CHART END
IM 04P01B01-01E
12-7
12
Specifications
MATH
This indicator turns ON when recording is started and turns OFF when stopped.
The indicator turns ON when an alarm occurs and turns OFF when the alarm is released.
If the alarm indication is set to hold operation, the indicator blinks when an alarm occurs. When
alarm ACK operation is performed, the indicator turns ON or OFF.
This indicator turns ON when computation on the computation function (/M1 option) is started and
turns OFF when stopped.
This indicator turns ON when key lock is activated and turns OFF when released.
This indicator turns ON when the remaining amount of chart paper falls to approximately 2 cm on
models with the FAIL/chart end detection and output function (/F1 option).
12.4 Display Function Specifications
Item
Specifications
Display type specifications
Digital display
Measured value or computed value
• DC voltage input: Displays values between “the min. measurable value – 5% of the span of the
measurable range” and “the max. measurable value + 5% of the span of the measurable range”
of the range (example: –2.200 to 2.200 V for the 2 V range).
• TC and RTD input: Displays values for input values between “the min. measurable value – 10°C”
and “the max. measurable value + 10°C” (example: –10.0°C to 1770.0°C for TC R)
• ON/OFF input (DI): Displays the input value 0 or 1.
• Linear scaling (1-5V, scaling, and square root): Displays value for input values between “the
leftmost value of scale – 5% of scaling span” and “the rightmost value of scale + 5% of scaling
span” (example: –5.0 to 105.0 when scaled to 0.000 to 100.0). However, the displayable range
during scaling is from –19999 to 31500 (excluding the decimal). Values below –19999 and
values above 31500 are displayed as –Over and +Over, respectively.
• Computation channel (/M1 option): Displays values in the range of –9999999 to 99999999.
Channel No.
Displayed with 1 character as follows:
Measurement channels: 1, 2, 3, 4, 5, 6
Computation channels: A, B, C, D, E, F, G, J, K, M, N, P
Alarm
Alarms are displayed using the symbols below. If multiple alarms are occurring simultaneously, the
alarm with higher precedence is shown.
(Higher precedence) H, L, T, t, h, l, R, r (Lower precedence)
Special values
Measurement channel
Status
Display
Description
+Over
+Over
See below.
–Over
–Over
See below.
Skip
SKIP
A value for channels set to skip.
Error
ERROR
Values such as when both the reference channel and measurement
channel are +Over or –Over in delta computation.
+Burnout
B.out
A value displayed when a burnout is detected on a channel set to
burnout up scale.
–Burnout
B.out
A value displayed when a burnout is detected on a channel set to
burnout down scale.
+Over and –Over of measurement channels
A condition in which the value is outside the display range indicated in “Measured value or
computed value” above. For example,
• For 2 V range, values below –2.200 V are –Over and values over 2.200 V are +Over.
• For TC R, values below –10.0°C are –Over and values over 1770.0°C are +Over.
• If scaled to 0.0 to 100.0, values below –5.0 are –Over and values above 105.0 are +Over.
However, the displayable range during scaling is from –19999 to 31500 (excluding the decimal).
Values below –19999 and values above 31500 are displayed as –Over and +Over, respectively.
Computation channel
Status
Display
+Over
+Over
Description
When the computed result exceeds 99999999.
–Over
–Over
When the computed results is less than –9999999.
OFF
SKIP
A value for channels set to Off.
Channel switching interval The switching interval of displayed channels can be set to 1 s, 2 s, 3 s, 4 s, or 5 s.
12-8
IM 04P01B01-01E
12.4 Display Function Specifications
Item
Specifications
Bar graph display
Displays the measured value on a bar graph.
Relationship between the digital display and bar graph
Displays the same channel on the digital display and bar graph. On the 2-channel digital + 2channel bar graph display, the top bar graph is the left digital display channel and the bottom graph
is the right digital display channel.
Scale
Displays 10 equally spaced scale marks.
Alarm point mark
For a channel set to delta computation, the alarm point marks of h and l are displayed. For all other
channels, the alarm point marks of H, L, T, and t are displayed.
• Blinks on the bar graph when an alarm occurs. Stops blinking when the alarm is released.
Bar graph display mode Selectable from the two modes below.
• Set the base point of the bar graph to the left or right of the recording span whichever is smaller in value.
• Set the base point of the bar graph to the 50% position of the recording span.
Flag display
Display channel
Scale
Displays measured values using flags.
The flag can be turned ON/OFF for each channel.
Flag (If flags overlap, the channel with the smallest number is displayed on top)
Displays 10 equally spaced scale marks.
Channel alarm status display
Lists the alarm status of all channels using the symbols below.
Symbol
Description
•
Alarm is not set.
–
Alarm is not occurring.
Channel No.
Alarm is occurring.
Displays the channel number when an alarm occurs and clears when the alarm is
released.
If the alarm indication is set to hold operation, the indicator blinks when an alarm
occurs. When alarm ACK operation is performed, the indicator turns ON (stop
blinking) or OFF.
Date/time and chart speed display
Displays the date/time on the left and chart speed on the right.
The date display format can be selected. The format is common with the “Date printout format.”
DI/DO status display
DI: Displays the ON/OFF state of the remote control input terminal (/R1 option).
DO: Displays the relay operation (active) of the alarm output terminal (/A1, /A2, or /A3 option).
Status display
Displays the status below. The display condition is the same as the status display under the main
display (see page 12-7).
Alarm and chart end (/F1 option)
Lights off
Displays nothing.
Skip
Skips the screen during screen switching.
Tag display
Displays 7-digit tags in place of channel numbers.
See the digital display and bar graph display sections.
Multiple display (display in which different screens can be assigned to the top and bottom sections)
8 display types (see page 12-7) can be assigned to the top and bottom sections of the display
(VFD).
System display
12-9
12
Specifications
IM 04P01B01-01E
Displays system information by switching screens every 3 seconds. The main displayed contents
are as follows:
Analog: Number of inputs
Option1, Option2: Options that are installed
Math: Number of computation channels
Periodic: Periodic printout interval
Interval: Trend recording interval (dot model)
POC: Pen offset compensation ON/OFF (pen model)
MAC address: MAC address
Version: Firmware version number
12.4 Display Function Specifications
Name
1-channel digital display
Display Example
Measurement channel
Description
1H3000.0ABC
Alarm
Channel No.
Measured
value
Unit
Computation channel
AH9999999.9
2-channel digital display
4-channel digital display
Channel No. (1 digit), alarm (1 digit),
measured value (6 digits), unit (3 digits)
Displays using 5 × 7 dots when the unit
exceeds 3 digits.
Channel No. (1 digit), alarm (1 digit),
computed value (9 digits)
Measurement channel
1H-1999.9ABCDEF
2L 2000.0PQRSTU
Channel No. (1 digit), alarm (1 digit),
measured value (7 digits), unit (6 digits)
Computation channel
AH-999999.9ABCDEF
BL9999999.9PQRSTU
Channel No. (1 digit), alarm (1 digit),
computed value (9 digits), unit (6 digits)
Measurement channel
1H3000.0 1H3000.0
3H3000.0 4H3000.0
Channel No. (1 digit), alarm (1 digit),
measured value (6 digits)
Computation channel
AH-1999.9 BH-1999.9
CH9999.9 DH9999.9
6-channel digital display
(Dot model)
999.9
999.9
1-channel digital display
+ 1-channel bar graph display
1H3000.0ABC
999.9
999.9
999.9
999.9
Measurement channel
2-channel digital display
+ 2-channel bar graph display 1H3000.0 2L-999.9
Computation channel
AH-1999.9 BH9999.9
Measurement channel
1-channel digital display
+ 4-channel bar graph display 1H-1999.9ABCDEF
(Pen model)
Computation channel
AH9999999.9ABCDEF
Channel No. (1 digit), alarm (1 digit),
computed value (6 digits)
Displayable values are –19999 to 99999 and
a decimal. If this range is exceeded, the value
is displayed as“–####” or “+####.”
Measured value (5 or 6 digits)
Channel No. (1 digit), alarm (1 digit),
measured value (6 digits), unit (3 digits)
Bar graph, scale, alarm (bar graph)
Value and bar graph of the same channel.
Channel No. (1 digit), alarm (1 digit),
measured value (6 digits)
Bar graph, scale
Channel No. (1 digit), alarm (1 digit),
computed value (6 digits)
Bar graph, scale
Displayable values are –19999 to 99999 and a
decimal. If this range is exceeded, the value
is displayed as“–####” or “+####.”
Channel No. (1 digit), alarm (1 digit),
measured value (7 digits), unit (6 digits)
Channel No. (1 digit), alarm (1 digit),
computed value (9 digits), unit (6 digits)
4-channel bar graph display
(Pen model)
Bar graph, scale
6-channel bar graph display
(Dot model)
Bar graph, scale
Flag display
Flag, scale
Channel ID alarm status
display
•
–
Channel No.
Date/time and chart speed
display
12-10
Sep.21.2004
11:26:37
1200
mm/h
Alarm is not set.
Alarm is not occurring.
Alarm is occurring.
Displays the date/time on the left and chart
speed on the right.
IM 04P01B01-01E
12.4 Display Function Specifications
Name
Display Example
Description
DI/DO status display
DI (remote control inputs) status.
DO (alarm output relays) status.
Status display
Displays the appropriate mark.
Chart end
Alarm
Lights off
1-channel digital display
(Tag display)
No display.
Measurement channel
abcdefg
H-1999.9ABCDEF
Alarm
Tag
Measured
value
Tag (7 digits), alarm (1 digit),
measured value (7 digits), unit (6 digits)
Unit
Computation channel
abcdefg
H9999999.9ABCDEF
2-channel digital display
(Tag display)
Tag (7 digits), alarm (1 digit),
computed value (9 digits), unit (6 digits)
Measurement channel
Tag (7 digits), alarm (1 digit),
measured value (6 digits), unit (3 digits)
The top section is measurement channel;
Computation channel
bottom section is computation channel
Tag (7 digits), alarm (1 digit),
computed value (9 digits)
abcdefgH-1999.9ABC
abcdefgH9999999.9
Measurement channel
1-channel digital display
+ 1-channel bar graph display abcdefgH-1999.9ABC
(Tag display)
Computation channel
abcdefgH9999999.9
The digital display is the same as
the 2-channel digital display (tag display).
Bar graph, scale, alarm (bar graph)
Value and bar graph of the same channel.
1-channel digital display
Measurement channel
+ 4-channel bar graph display abcdefgH-1999.9ABC
(Tag display) (Pen model)
The digital display is the same as
the 2-channel digital display (tag display).
Bar graph
Computation channel
abcdefgH9999999.9
12
Specifications
IM 04P01B01-01E
12-11
12.4 Display Function Specifications
Multiple display (Different screens assigned to the top and bottom sections)
Name
1-channel digital display
Display Example
Computation channel
1H-1999.9ABCDEF
Measurement channel
AH9999999.9ABCDEF
2-channel digital display
Computation channel
1H-1000.0 2H3000.0
Measurement channel
AH-1999.9 BH9999.9
Date/Time
Chart speed
Sep21.04
11:26:37
25mm/h
Channel ID alarm status
display
Alarm summary display
of computation channels
Status display
Lights off
1-channel digital display
(Tag display)
Description
Channel No. (1 digit), alarm (1 digit),
measured value (7 digits), unit (6 digits)
Channel No. (1 digit), alarm (1 digit),
computed value (9 digits), unit (6 digits)
Channel No. (1 digit), alarm (1 digit),
measured value (6 digits)
Channel No. (1 digit), alarm (1 digit),
measured value (6 digits)
Date/Time
Chart speed
•
–
Channel No.
Alarm is not set.
Alarm is not occurring.
Alarm is occurring.
Displays the appropriate mark.
No display.
abcdefgH-1999.9ABC
abcdefgH9999999.9
The digital display is the same as
the 2-channel digital display (tag display).
The top section is measurement channel;
bottom section is computation channel
12-12
IM 04P01B01-01E
12.5 Specifications of Optional Functions
Alarm Output Relay (/A1, /A2, and /A3)
Item
Specifications
Operation
Number of outputs
Relay contact rating
Outputs relay contact signals from the dedicated terminals on the rear panel when alarms occur.
2 outputs (/A1), 4 outputs (/A2), 6 outputs (/A3)
250 VDC/0.1 A (for resistance load)
250 VAC (50/60 Hz)/3 A
NO-C-NC
Energized/De-energized operation selectable (common to all relays)
Hold/Non-hold operation selectable (common to all relays)
AND/OR operation selectable
Reflash operation (Fixed to relay I01, I02, and I03. Relay release time is approx. 500 ms.)
Relay I01 can be used for diagnosis output.
Diagnosis output: Outputs a relay signal when an error in the recording section (plotter on the
pen model), a burnout, or an error in the A/D converter is detected. Relay operation is fixed to
de-energized and non-hold.
Output format
Auxiliary functions
Diagnosis output
RS-422A/485 Communication Interface (/C3)
Item
Specifications
Connection
Connection type
Transmission mode
Synchronization
Baud rate
Start bit
Data length
Stop bit
Parity
Received buffer length
Escape sequence
Protocol
Communication mode
Complies with EIA RS-422A/485
Multi-drop connection 1:32 (four-wire system), 1:31 (two-wire system, version 1.11 and later)
Half-duplex
Start-stop synchronization
1200, 2400, 4800, 9600, 19200, or 38400 bps
Fixed to 1 bit
7 bits or 8 bits
Fixed to 1 bit
Odd, even, or none (no parity)
2047 bytes
Open and close
Dedicated protocol, Modbus slave protocol (version 1.11 and later)
ASCII for input/output for control and setting
ASCII or binary for input/output of measured data
1.2 km
Communication distance
Ethernet Communication Interface (/C7)
Item
Specifications
Function
Setting/Measurement server
Miscellaneous
User authentication
Keepalive
IM 04P01B01-01E
Function
Administrators
Maximum Number of Number of
Connections Users
Simultaneous Users
Port Number
Setting/Measurement server
3
1
2
Maintenance/Test server
1
1
1
34261/tcp
Instrument information server
-
-
-
34264/udp
34260/tcp
Administrator: 1, users: 6 (limitations placed on users)
Drops the connection if there is no response to an inspection packet. Use or not use is
selectable.
12-13
12
Specifications
Uses a dedicated protocol
Measured/computed data, setup data, status byte
Maintenance/Test server
Connection information, network statistics, etc.
Instrument information server Model, serial number, etc.
Electrical and mechanical specifications
Conforms to IEEE 802.3.
Interface
Basic specifications
Connection
Ethernet (10BASE-T)
Protocol
TCP, IP, UDP, ICMP, and ARP
The maximum number of simultaneous connections and the number of simultaneous use
12.5 Specifications of Optional Functions
FAIL/Chart End Detection and Output (/F1)
Item
Specifications
FAIL output
Relay output on CPU error
Relay operation
Energized during normal operation and de-energized on CPU error. Non-hold operation.
Chart end detection output Stops recording and outputs a relay signal when the remaining amount of chart paper is approx.
2 cm. Also, turns ON the words CHART END on the display.
Relay operation
The relay is energized when the above conditions are met. Non-hold operation.
Contact rating
250 VAC (50/60 Hz)/3 A, 250 VDC/0.1 A (for resistor load)
Clamped Input Terminal (/H2)
Item
Specifications
Measuring input terminal
Uses clamped input terminals.
Non-Glare Door Glass (/H3)
Item
Specifications
Door glass
Uses a door glass with minimal reflection.
Computation Function (/M1)
Item
Specifications
Number of computation channels
Pen model: 8 channels, dot model: 12 channels
Operation
General arithmetic operations: Four arithmetic operations, square root, absolute, common
logarithm, exponential, power, relational operations (<, ≤ , >, ≥ , =,
≠), logical operations (AND, OR, NOT, XOR)
Statistical operations (TLOG): Average, max., min., sum, and max. – min. for each interval
Constants
Up to 30 constants can be defined
Range: –9.9999E+29 to –1.0000E–30, 0, 1.0000E–30 to 9.9999E+29
Symbol: K01 to K30
Communication input data Digital data via communication can be used in computations other than statistical computation (TLOG)
Number of communication input data points: 8 channels (pen model), 12 channels (dot model)
Range: –9.9999E+29 to –1.0000E–30, 0, 1.0000E–30 to 9.9999E+29
Symbol: C01 to C12
Remote input
Use remote input status (0/1) in equations
Number of inputs: 5
Symbol: D01 to D05
Computation accuracy
32-bit floating-point computation (excluding TLOG computation)
Computation range
Computed values throughout the computation remains within approx. ±3.4×1038.
Equation
Defined using up to 120 characters.
Execution interval
Scan interval
TLOG timer
Performs statistical calculation for each interval specified by the timer.
Number of timers
3 (periodic printout timer, timer 1, and timer 2).
Timer 1 and 2 modes
Absolute or relative
Reset
TLOG data can be reset at each interval.
Printout
Printout at each interval is possible.
Trend recording
Recording span
Unit
Recording methods
Pen
Dot printing
–9999999 to 99999999, decimal position arbitrary.
Up to 6 characters
Zone recording and partial expanded recording available.
Measurement channels and computation channels can be assigned arbitrarily to pens.
Trend recording of measurement and computation channels. The fastest dot printing interval is as follows:
Number of channels being trend recorded Fastest dot printing interval
10 s
6 or less
15 s
7 to 12
20 s
13 to 18
12-14
IM 04P01B01-01E
12.5 Specifications of Optional Functions
Alarms
Number of alarms
Alarm types
Hysteresis
Miscellaneous
Display
Up to four alarms (level) for each measurement channels.
High limit (H), low limit (L), delay high limit (T), delay low limit (t). The alarm symbol is shown in parentheses.
Alarm delay time of delay high and low limit alarms (can be set for each channel, common within a
channel) See section 12.2
Applied to high limit alarm and low limit alarm.
Hysteresis range: Approx. 0.0% to 1.0% of the recording span (0.1% steps)
See section 12.2.
See section 12.4.
Cu10, Cu25 RTD Input (/N1)
Item
Specifications
Cu10, Cu25 RTD
This option allows Cu10 and Cu25 inputs to be added to the standard input types.
Input type and measurable range
Measurable Range
Input Type
Cu10 (GE)
Cu10 (L&N)
RTD
Cu10 (WEED)
–200 to 300°C
(Measurement
Cu10 (BAILEY)
–328.0 to 572.0°F
current i: 2 mA)
Cu10: α = 0.00392 at 20°C
Cu10: α = 0.00393 at 20°C
Cu25*: α = 0.00425 at 0°C
* Measurement current i = 1 mA
Measurement and recording accuracy
Input Type
Cu10 (GE)
Cu10 (L&N)
Cu10 (WEED)
Cu10 (BAILEY)
Cu10: α = 0.00392 at 20°C
Cu10: α = 0.00393 at 20°C
Cu25: α = 0.00425 at 0°C
Measurement Accuracy
±(0.4% of rdg + 1.0°C)
Recording Accuracy
Measurement accuracy
± (0.3% of recording span)
±(0.3% of rdg + 0.8°C)
3 Legs Isolated RTD (/N2)
Item
Specifications
Input terminal isolation
Applies to the dot model.
The RTD input terminals are isolated between channels (on the standard dot model, input terminal b
are common to all channels)
Expansion Inputs (/N3)
Specifications
Input type
Adds the input types shown in the table below.
Input Type
PR40-20
PLATINEL
Thermocouple
NiNiMo
W/WRe26
Type N (AWG14)
Kp vs Au7Fe
Pt25
Pt50
Ni100 (SAMA)
RTD
Ni100 (DIN)
(Measuring current
Ni120
i = 1 mA)
J263*B
Cu53
Cu100: α = 0.00425 at 0°C
IM 04P01B01-01E
Measurement Range
0.0 to 1900.0°C
32 to 3452°F
0.0 to 1400.0°C
32 to 2552°F
0.0 to 1310.0°C
32 to 2390°F
0.0 to 2400.0°C
32 to 4352°F
0.0 to 1300.0°C
32 to 2372°F
0.0 to 300.0 K
–200.0 to 550.0°C
–328.0 to 1022.0°F
–200.0 to 600.0°C
–328.0 to 1112.0°F
–200.0 to 250.0°C
–328.0 to 482.0°F
–60.0 to 180.0°C
–76.0 to 356.0°F
–70.0 to 200.0°C
–94.0 to 392.0°F
0.0 to 300.0 K
–50.0 to 150.0°C
–58.0 to 302.0°F
–50.0 to 150.0°C
–58.0 to 302.0°F
12-15
12
Specifications
Item
12.5 Specifications of Optional Functions
Item
Specifications
Measurement and recording accuracy
PR40-20
Input Type
0 to 450°C
450 to 750°C
750 to 1100°C
1100 to 1900°C
PLATINEL
NiNiMo
W/WRe26
Type N(AWG14)
Kp vs Au7Fe
Pt25
Pt50
Ni100(SAMA)
Ni100(DIN)
Ni120
J263*B
0 to 400°C
400 to 2400°C
0 to 20 K
20 to 300 K
Measurement Accuracy
Not warranted
±(0.9% of rdg + 3.2°C)
±(0.9% of rdg + 1.3°C)
±(0.9% of rdg + 0.4°C)
±(0.25% of rdg + 2.3°C)
±(0.25% of rdg + 0.7°C)
±15.0°C or less
±(0.2% of rdg + 2.0°C)
±(0.2% of rdg + 1.3°C)
±4.5 K
±2.5 K
±(0.15% of rdg + 0.6°C)
±(0.3% of rdg + 0.6°C)
Recording Accuracy
Measurement
accuracy
±(0.3% of
recording span)
±(0.15% of rdg + 0.4°C)
±3.0 K
±1.0 K
Cu53
±(0.15% of rdg + 0.8°C)
Cu100
±(0.2% of rdg + 1.0°C)
Note) No RJC for PR40-20 (fixed to 0°C)
0 to 40 K
40 to 300 K
Remote Control 5 Points (/R1)
Item
Specifications
Number of input terminals 5
Input type
Isolated from the main circuitry through a photocoupler, built-in isolated power supply for the input
terminals, and shared common
Input type and signal level
Voltage-free contact
Contact closed at 200 Ω or less and contact open at 100 kΩ or greater
Open collector
ON voltage: 0.5 V or less (sink current 30 mA or more), leakage current when OFF: 0.25 A or less
Signal type
Edge, level, or trigger (250 ms or more)
Operation
Perform the following control operations by applying a given signal to the remote signal input
terminal.
• Start/stop recording (edge)
• Alarm ACK (trigger)
• Set the clock (adjust the clock to the nearest hour with the contact input, trigger)
Time When Signal Is Input
Procedure
00 min 00 s to 01 min 59 s
Truncates the minutes and seconds. Example: 10:00:50 → 10:00:00
58 min 00 s to 59 min 59 s
Rounds up the minutes and seconds. Example: 10:59:50 → 11:00:00
02 min 00 s to 57 min to 59 s No operation
•
•
•
•
•
12-16
Manual printout (trigger)
Message 1 to 5 printout (trigger)
Switch the chart speed (level)
Start/stop computation (/M1 option) (edge)
Reset computation (/M1 option) (trigger)
IM 04P01B01-01E
12.6 General Specifications
Construction
Item
Specifications
Mounting
Mounting angle
Allowable panel thickness
Material
Flush panel mounting (on a vertical plane)
Inclined backward up to 30 degrees from a horizontal plane.
2 to 26 mm
Case: drawn steel
Front door: Aluminum die-cast
Case: Charcoal gray light (Munsell 10B3.6/0.3 or equivalent)
Front door: Charcoal gray light (Munsell 10B3.6/0.3 or equivalent)
Water and dust-proof (complies with DIN40050-IP54)
144(W) × 144(H) × 220(D) mm (D: depth from the panel mounting plane)
1-pen: approx. 2.1 kg, 2-pen: approx. 2.2 kg, 3-pen: approx. 2.3 kg, 4-pen: approx. 2.4 kg
6-dot: approx. 2.5 kg
Color
Front panel
Dimensions
Weight
Normal Operating Conditions
Item
Specifications
Ambient temperature
0 to 50°C
Ambient humidity
20 to 80%RH (at 5 to 40°C)
Allowable power supply voltage range
90 to 132, 180 to 264 VAC
Power supply frequency
50 Hz±2%, 60 Hz±2%
Vibration
10 to 60 Hz 0.2 m/s2 or less
Shock
Not acceptable.
Magnetic field
400 A/m or less (DC and 50/60 Hz)
Noise
Normal mode (50/60 Hz)
Volt: The peak value including the signal must be less than 1.2 times the rated value of the range
(Example: 2 V for 2 V range).
TC: The peak value including the signal must be less than 1.2 times the rated value of the range.
RTD: 50 mV or less
Common mode noise (50/60 Hz): 250 VACrms or less for all ranges
Maximum noise voltage between channels (50/60 Hz): 250 VACrms or less. 6-dot model with the 3
legs isolated RTD (/N2 option) installed: 200 VACrms or less.
Mounting position
Up to 30° of backward tilt angle allowed, no tilt to the left or right
Warm-up time
At least 30 minutes after power on
Altitude
2000 m or less above sea level
Power Supply
Item
Specifications
Rated supply voltage
100 to 240 VAC
Allowable power supply voltage range
90 to 132, 180 to 264 VAC
Rated power supply frequency
50 Hz, 60 Hz
Rated power
Pen model: 40 VA, dot model: 40 VA
Power consumption
100 VAC
240 VAC
Maximum
1- to 4-pen model
Approx. 12 VA*
Approx. 17 VA*
Approx. 40 VA
6-dot model
Approx. 13 VA*
Approx. 18 VA*
Approx. 40 VA
Specifications
Model
*
IM 04P01B01-01E
12
When balanced
12-17
12.6 General Specifications
Isolation
Item
Specifications
Insulation resistance
Dielectric strength
Each terminal to ground terminal: 20 MΩ or more (at 500 VDC)
Power supply to ground terminal: 1500 VAC (50/60 Hz), 1 minute
Contact output terminal to ground terminal: 1500 VAC (50/60 Hz), 1 minute
Measuring input terminal to ground terminal: 1000 VAC (50/60 Hz), 1 minute
Between measuring input terminals: 1000 VAC (50/60 Hz), 1 minute (except for RTD input terminal)
Remote input terminal to ground terminal: 500 VDC, 1 minute
Grounding resistance: 100 Ω or less
Grounding
Transport and Storage Conditions
Item
Specifications
Ambient temperature
Ambient humidity
Vibration
Shock
–25 to 60°C
5 to 95%RH (no condensation)
10 to 60 Hz, 4.9 m/s2 maximum
392 m/s2 maximum (in packaged condition)
Supported Standards
Item
Specifications
CSA
Certified by CSA22.2 No. 61010-1, Installation category II1, Measurement category II2, Pollution
degree 23
Certified by UL61010B-1 (CSA NRTL/C)
UL
CE
EMC
Low voltage directive
C-Tick
EN61326(Emission: Class A, Immunity: Annex A)
EN61000-3-2
EN61000-3-3
EN55011 Class A Group 1
EN61010-1, Installation category II, Measurement category II, Pollution degree 2
AS/NZS CISPR 11 Class A Group 1
1 Installation category (overvoltage category) describes a number which defines a transient overvoltage condition. It implies
the regulation for impulse withstand voltage. “II” applies to electrical equipment which is supplied from the fixed installation
like a distribution board.
2 Applies to measuring circuits connected to low voltage installation, and electrical instruments supplied with power from fixed
equipment such as electric switchboards.
3 “Pollution degree” describes the degree to which a solid, liquid, or gas which deteriorates dielectric strength or surface
resistivity is adhering. “2” applies to normal indoor atmosphere. Normally, only non-conductive pollution occurs.
12-18
IM 04P01B01-01E
12.6 General Specifications
Standard Performance
Item
Specifications
Measurement and recording accuracy
The following specifications apply to operation of the recorder under standard operation conditions:
Temperature:
23 ± 2°C
Humidity:
55% ± 10%RH
Power supply voltage:
90 to 132 or 180 to 264 VAC
Power supply frequency: 50/60 Hz ± 1%
Warm-up time:
At least 30 minutes.
Other ambient conditions such as vibration should not adversely affect the operation
Input Type
DC voltage
Range Type
Measurement (Digital Display)
Measurement Accuracy
20 mV
60 mV
200 mV
2V
±(0.1% of rdg + 2 digits)
6V
DC voltage
TC
(excludes
RJC
accuracy)
20 V
50 V
1-5 V
R
S
B
except R,S: 0 to 100°C, ±3.7°C
100 to 300°C, ±1.5°C
B: 400 to 600°C, ±2°C
Accuracy not warranted for
values less than 400°C
K
±(0.15% of rdg + 0.7°C)
except ±(0.15% of rdg + 1°C)
for –200 to –100°C
E
±(0.15% of rdg + 0.5°C)
J
T
±(0.15% of rdg + 0.5°C)
except ±(0.15% of rdg + 0.7°C)
for J: –200 to –100°C
N
±(0.15% of rdg + 0.7°C)
W
±(0.15% of rdg + 1°C)
L
U
±(0.15% of rdg + 0.5°C)
except ±(0.15% of rdg + 0.7°C)
for L: –200 to –100°C
WRe
±(0.2% of rdg + 1°C)
Pt100
JPt100
ON/OFF input Level
Contact
RTD
±(0.1% of rdg + 2 digits)
±(0.1% of rdg + 2 digits)
±(0.15% of rdg + 1)
Recording (Trend)
Highest Res. Recording Accuracy
Resolution
10 µV
Measurement
Pen model:
accuracy
Dead band 0.2%
10 µV
± (0.3% of
of recording span
100 µV
recording span)
Dot model:
1 mV
Resolution 0.1 mm
1 mV
10 mV
10 mV
1 mV
0.1°C
±(0.15% of rdg + 0.3°C)
Threshold level (2.4 V)
accuracy ± 0.1 V
≤ 1 KΩ: ON, ≥ 100 kΩ: OFF
(Parallel capacitance
0.01 µF or less)
–
–
IM 04P01B01-01E
12-19
12
Specifications
Measuring accuracy in case of linear scaling
Accuracy during scaling (digits) = measuring accuracy (digits) × multiplier + 2 digits (rounded up)
where the multiplier = scaling span (digits)/measuring span (digits).
Example: Assuming that
• range type:
1-5 V
• measuring span:
1.000 to 5.000 V
• scaling span:
0.000 to 2.000 Then,
Measuring accuracy = ±(0.1% × 5 V + 2 digits)
= ±(0.005 V [5 digits] + 2 digits)
= ±7 digits
Multiplier = {2000 digits (0.000 to 2.000)}/4000 digits (1.000 to5.000) = 0.5
Accuracy during scaling = ±(7 × 0.5 + 2) digits = 6 digits (rounded up)
Reference junction compensation
Internal/External selectable for each channel
12.6 General Specifications
Item
Specifications
Reference junction compensation accuracy
Above 0°C with input terminal temperature balanced (60 minutes after power on)
Type R, S, B, W, WRe: ±1.0°C
Type K, J, E, T, N, L, U: ±0.5°C
Maximum input voltage
±10 VDC (continuous) for ranges of 200 mV or less, TC, RTD, and DI ranges
±60 VDC (continuous) for 2 VDC or higher ranges
Input resistance
Approximately 10 MΩ or more for ranges of 200 mV or less and TC
Approximately 1 MΩ for 2 VDC or higher ranges
Input source resistance
Volt, TC:
2 kΩ or less
RTD input:
10 Ω or less per wire (The resistance of all three wires must be equal).
Bias current
10 nA or less (except when burnout detection function is enabled)
Common mode rejection ratio 120 dB (50/60 Hz ± 0.1%, 500 Ω unbalanced, between the minus terminal and ground)
Normal mode rejection ratio 40 dB or more (50/60 Hz ± 0.1%)
Noise rejection
By the integrating A/D converter: The integration time is 16.7 ms, 20 ms, or 100 ms (dot model)
Low-pass filtering (pen model): Filter time constant is 2 s, 5 s, or 10 s
Moving average (dot model):
The number of samples is 2 to 16.
Effects of Operating Conditions
Item
Specifications
Ambient temperature
With temperature variation of 10°C:
Measurement: ±(0.1% of rdg + 1 digit) or less
Recording:
Measurement fluctuation + ±0.2% of the recording span or less
* Excluding the error of reference junction compensation
With variation within 90 to 132 V and 180 to 264 VAC (50/60 Hz):
Measurement: ±1 digit or less
Recording:
±0.1% of the recording span or less
With variation of ±2 Hz from rated power frequency (power supply voltage 90 to 132, 180 to 264 VAC):
Measurement: ±(0.1% of rdg + 1 digit) or less
Recording:
Same as the measurement fluctuation
AC (50/60 Hz) and DC 400 A/m fields:
Measurement: ±(0.1% of rdg 10 digit) or less
Recording:
±0.5% of the recording span or less
Power supply fluctuation
Magnetic field
Input source resistance
DC voltage range
TC range
RTD range
Mounting position
Vibration
With variation of +1 kΩ:
Ranges of 200 mV or less: Within ±10 µV
Ranges of 2V or more:
±0.1% of rdg
With variation of +1 kΩ: ±10 µV or less
With variation of 10 Ω per wire (resistance of all three wires must be equal):
Measurement: ±(0.1% of rdg + 1 digit) or less
Recording:
Measurement fluctuation + ±0.1% of the recording span or less
With maximum difference of 40 mΩ between wires: Approx. 0.1°C (for Pt100)
With a backward tilt of 30° or less
Measurement: ±(0.1% of rdg + 1 digit) or less
Recording:
Indication fluctuation + ±0.1% of the recording span or less
When a sinusoidal vibration along all three axis at a frequency between 10 to 60 Hz and an
acceleration of 0.2 m/s2 is applied for 2 hours
Measurement: ±(0.1% of rdg + 1 digit) or less
Recording:
Indication fluctuation + ±0.1% of the recording span or less
Other Specifications
Item
Specifications
Clock
With a calendar function
Accuracy of clock
±100 ppm, excluding a delay (of 1 second, maximum) caused each time the power is turned on.
DST
The date/time for switching between standard time and DST can be specified.
Memory backup
A built-in lithium battery backs up the setup parameters and runs the clock
(battery life: approximately ten years at room temperature).
Key lock function
Key lock items
Password for releasing the key lock can be set.
RCD key, MENU key, DISP MENU key, FEED key, and function under the FUNC key (Alarm ACK, computation start/
stop, print, message, printout buffer clear, periodic printout reset, and pen or ribbon cassette exchange)
Internal light
Lights up the recording area of the chart paper. Uses a dedicated LED.
Noise
Machine Noise Information Ordinance 3.GSGV, Jan 18, 1991
Maximum noise level: 60 dB (A) or less (complies with ISO7779)
12-20
IM 04P01B01-01E
12.7 Dimensional Drawings
Unit: mm (approx. inch)
Unless otherwise specified, tolerance is ±3%
(however, tolerance is ±0.3 mm when below 10 mm).
220 (8.66)
(1.08) 27.5
151.5 (5.96)
178 (7.01)
Mounting panel
thickness
144 (5.67)
144
(5.67)
151.5
(5.96)
2 to 26
9.4 (Dimensions before attaching
(0.37) the mounting bracket)
136.5
(5.37)
+2
0
7.5 (Dimensions after attaching
(0.30) the mounting bracket)
Note
When mounting the recorder to a panel, use two brackets, top and bottom or left and right.
12
Specifications
IM 04P01B01-01E
12-21
Appendix
Appendix 1 Periodic Printout and Printout Using the
TLOG Timer (/M1 Option)
Periodic Printout
This section explains the contents of the periodic printout. The periodic printout contents
vary between the pen model and dot model. The contents also vary depending on
whether the periodic printout is in instantaneous mode, report mode, or Off.
Instantaneous mode or report mode is specified using Print1 > Mode in Basic Setting
mode.
The figure below illustrates the printout contents. The font used in the figure differs from that
of the actual printout. The printout positions may also differ from those of the actual printout.
Pen Model
Instantaneous Mode
Channel No.
or tag
Date
Time
Channel data
Scale
Recording color
Chart speed
Mar.31.2004!
15:50*
TAG-001
1.000V
2
-1.000V
3
H-2.1050UNIT03
4
d
2.000V
TAG-00A
1UNIT0A
C
L
1.00
D
-10.00UNIT0D
E
1.0000%
J
-1.00UNIT0J
0.000
TAG-001
RED
50mm/h_
Time tick
Alarm
Time tick cancel mark
Offset compensation mark
2.000
V
Prints the letter “d” for channels set to
delta computation.
• Date/Time
The date/time when the periodic printout was executed.
• Time tick
Indicates the pen 1 position at the time the periodic printout was executed.
• Time tick cancel mark
Prints an exclamation point (!) when the time tick is not printed correctly.
• Offset compensation mark
Prints an asterisk (*) when offset compensation is enabled.
• Channel data
Prints the channel numbers (1 to 4 or A to J) or tags, measured/computed values, and unit.
Prints the letter “d” on channels set to delta computation.
• Alarm
Prints the alarm that is occurring. If multiple alarms are occurring, the alarm with the
highest precedence is printed.
Alarm printout precedence: (Higher) H, L, T, t, h, l, R, r (Lower)
IM 04P01B01-01E
App-1
Periodic Printout and Printout Using the TLOG Timer (/M1 Option)
Computation channel
Measured
value
Unit
App
Appendix 1 Periodic Printout and Printout Using the TLOG Timer (/M1 Option)
• Scale
Prints the leftmost and rightmost values of the recording span and the channel
number or tag for channels that have scale printout specified. The scale is printed for
one channel at each periodic printout. The channel whose scale is printed changes in
ascending order. The scale of channels that are being zone recorded is printed within
the recording range of the zone.
• Recording color
Prints the recording colors of channels that have scale printout and recording color
printout specified.
• Chart speed
Prints the chart speed.
Report Mode
Channel No. or tag
Range change
occurrence mark
Measured value
Unit
Report start date/time
Report end date/time
Channel data
Computation channel
Scale
Recording color
Chart speed
Mar.31.2004 15:00!
-Apr.01.2004 15:00*TPR
TAG-001*1.000V
2
-1.000V
3
-2.105500E+03UNIT03
4
2.000V
TAG-00A 1UNIT0A
C
*1.00
D
-10.00UNIT0D
150.00UNIT0D
100.00UNIT0D
E
1.0000%
J
-1.00UNIT0J
0.000
TAG-001
RED
50mm/h_
Time tick cancel mark
Offset compensation mark/event
Sum value
Min./Max./Ave.
2.000
V
Time tick
• Report start date/time and report end date/time
The start date/time and end date time of the report (statistical calculation) (periodic
printout interval).
• Time tick and time tick cancel mark
These are the same as with the instantaneous mode.
• Offset compensation mark
This is the same as with the instantaneous mode.
• Event
Prints the events that occurred in the interval.
T: Time change
P: Power failure or power OFF/ON
R: Data reset
• Channel data
Prints the channel numbers (1 to 6 or A to J) or tags, report data, and unit.
An asterisk (*) is printed for channels whose range type was changed in the interval.
• Scale, recording color, and chart speed
These are the same as with the instantaneous mode.
App-2
IM 04P01B01-01E
Appendix 1 Periodic Printout and Printout Using the TLOG Timer (/M1 Option)
Dot Model
Instantaneous Mode
Channel No.
or tag
Date
Time
Channel data
Computation channel
Scale
Chart speed
Measured
value
Unit
Prints the letter “d” for channels set to
delta computation.
Mar.31.2004
15:50
TAG-001
1.000V
2
-1.000V
3
H-2.1050UNIT03
4
d
2.000V
TAG-00A
1UNIT0A
J
L
-1.00UNIT0J
0.000
TAG-001
50mm/h_
Time tick
2.000
V
Alarm
• Date/Time
The date/time when the periodic printout was executed.
• Time tick
Indicates the position at the time the periodic printout was executed. If the time tick
cannot be printed correctly, it is not printed.
• Channel data, alarm, scale printout, and chart speed
These are the same as with the instantaneous mode of the pen model.
Channel No. or tag
Range change occurrence mark
Measured value
Unit
Report start date/time
Report end date/time
Channel data
Computation channel
Scale
Chart speed
Mar.31.2004 15:00
-Apr.01.2004 15:00*TPR
TAG-001*1.000V
2
-1.000V
3
-2.105500E+03UNIT03
4
2.000V
TAG-00A 1UNIT0A
C
*1.00
D
-10.00UNIT0D
150.00UNIT0D
100.00UNIT0D
E
1.0000%
J
-1.00UNIT0J
0.000
TAG-001
50mm/h_
Event
Sum value
Min./Max./Ave.
,
2.000
V
Time tick
• Report start date/time and report end date/time
The start date/time and end date time of the report (statistical calculation) (periodic
printout interval).
• Time tick
This is the same as with the instantaneous mode.
• Event
This is the same as with the report mode of the pen model.
• Channel data
This is the same as with the report mode of the pen model. However, the channel
numbers are 1 to 6 or A to P.
• Scale and chart speed
These are the same as with the instantaneous mode of the pen model.
IM 04P01B01-01E
App-3
Periodic Printout and Printout Using the TLOG Timer (/M1 Option)
Report Mode
App
Appendix 1 Periodic Printout and Printout Using the TLOG Timer (/M1 Option)
Printout Using the TLOG Timer (/M1 Option)
Prints the computed values of computation channels for each interval determined by the
TLOG timer.
Pen Model
Channel No. or tag
Date
Time
Channel data
Scale
Recording color
Timer
Mar.31.2004
15:50*TPR
TAG-00A
B
C
L
G
J
-100
TAG-00A
RED
TLOG1
Measured
value
Unit
Offset compensation mark/event
Alarm
1UNIT0A
1.0
1.00
-1.0
-1.00UNIT0J
100
UNIT0A
• Date/Time
The date/time when the printout using the TLOG timer was executed.
• Offset compensation mark, scale, and recording color
These are the same as with the periodic printout of the pen model.
• Event
This is the same as with the report mode of the periodic printout of the pen model.
• Channel data
Prints the computation channel numbers (A to J) or tags, computed values, and unit.
• Alarm
This is the same as with the instantaneous mode of the periodic printout of the pen model.
• Timer
Prints the timer number being used.
Dot Model
Channel No. or tag Event
Measured value
Unit
Date
Time
Channel data
Scale
Timer
Mar.31.2004
15:50*TPR
TAG-00A
C
L
D
E
G
J
-100
TAG-00A
TLOG2
Alarm
1UNIT0A
1.00
100.00UNIT0D
1.0000%
-1.0
-1.00UNIT0J
100
UNIT0A
• Date/Time
The date/time when the printout using the TLOG timer was executed.
• Event
This is the same as with the report mode of the periodic printout of the pen model.
• Channel data
Prints the computation channel numbers (A to P) or tags, computed values, and unit.
• Alarm
This is the same as with the instantaneous mode of the periodic printout of the pen model.
• Scale
These are the same as with the periodic printout of the pen model.
• Timer
Prints the timer number being used.
App-4
IM 04P01B01-01E
Appendix 1 Periodic Printout and Printout Using the TLOG Timer (/M1 Option)
Special Cases
This section explains the operation of the periodic printout report (statistical calculation)
and the TLOG computation for special cases. In the explanation below, statistical
calculation represents computation in periodic printout report mode or TLOG
computation.
Operation during Power Failures
If a power failure occurs in the middle of statistical calculation, the operation is as follows
when the power recovers.
The statistical calculation resumes when the power recovers. The exact operation will
vary depending on whether the power is restored before or after the printout time.
Time of Recovery
Statistical Calculation Operation
After the printout time
Statistical calculation data is created immediately after power is
restored. The measured/computed data up to the time of the power
failure is used. At the next printout time, data after the power recovery
is used to create the statistical calculation data.
Before the printout time The data is created at the printout time after the power recovers. The
measured/computed data excluding the power failure period is used.
Procedure of Handling “Over” Values during Periodic Printout
When an “over” value* is input for sum or average computation, that value is not used in
the computation.
An “over” value on a measurement channel refers to +Over or –Over. See section 12.4. An
“over” value on a computation channel refers to a value that is less than or equal to
–100000000 or greater than or equal to 100000000 excluding the decimal.
Periodic Printout and Printout Using the TLOG Timer (/M1 Option)
*
App
IM 04P01B01-01E
App-5
Index
Symbols
Page
1-5V .................................................................................... 1-2
1-5V input ............................................................................ 5-6
1-5V low-cut function ........................................................ 7-24
A
Page
absolute time mode ........................................................... 9-21
absolute value ................................................................... 1-24
accuracy of clock ............................................................ 12-20
activating the key lock ....................................................... 3-19
adjusting the dot printing position ..................................... 11-7
adjusting the hysteresis .................................................... 11-7
adjusting the pen position ................................................. 11-6
alarm ........................................................................... 1-7, 5-9
alarm ACK ............................................................... 1-12, 3-18
alarm acknowledge operation ................................. 1-12, 3-18
alarm delay function .......................................................... 7-24
alarm hysteresis ................................................ 1-8, 1-26, 7-3
alarm indication ................................................................... 1-9
alarm information .............................................................. 7-11
alarm output relays ............................................................. 1-9
alarm output terminals ...................................................... 2-10
alarm printout .................................................................... 1-18
alarm type ..................................................... 5-10, 9-12, 12-3
alarm value ............................................................. 5-10, 9-12
alarms on the delta computation channel ......................... 5-10
AND operation (alarm output relay) .................................. 1-11
AND/OR .............................................................................. 7-3
assigned function (remote control) .................................... 7-29
assigning channels to recording pens ............................... 9-24
chart speed ............................................................. 1-14, 5-13
cleaning ............................................................................. 11-2
clearing alarm information ................................................. 3-15
comment line ....................................................................... 4-3
common logarithm ............................................................ 1-24
communication input data ................................................... 9-4
compensation voltage ......................................................... 7-7
computation errors ............................................................ 9-29
computation-dedicated channels ...................................... 1-24
computing equation ............................................................. 9-2
computing equations ........................................................... 9-4
constants ................................................................... 9-4, 9-10
construction ..................................................................... 12-17
conventions used in this manual ........................................... ix
copying & pasting a character string ........................... 4-4, 9-3
current input ........................................................................ 1-2
cursor .......................................................................... 4-3, 4-4
D
Page
bar graph ........................................................................... 1-22
base point of the bar graph ..................................... 7-15, 9-27
basic setting mode .............................................................. 4-1
bias ........................................................................... 1-4, 6-12
bias function ...................................................................... 7-24
brightness (display) ........................................................... 6-11
brightness (internal light) ................................................... 6-11
buffer memory ......................................................... 1-18, 1-19
buffer overflow mark ............................................... 1-18, 1-19
burnout detection ........................................................ 1-4, 7-5
data display setup screen ................................................... 8-1
data types (computation) .......................................... 1-24, 9-4
date format ........................................................................ 7-30
date/time ................................................................. 2-14, 5-14
DC voltage .......................................................................... 5-2
de-energized ............................................................. 1-11, 7-3
default values (basic setting mode) .................................. 4-10
default values (setting mode) .............................................. 4-9
delay high limit alarm .......................................................... 1-8
delay low limit alarm ............................................................ 1-8
deleting a character .................................................... 4-4, 9-3
deleting an entire character string ............................... 4-4, 9-3
delta computation ........................................................ 1-3, 5-4
diagnosis output .......................................................... 1-9, 7-2
difference high limit alarm ................................................... 1-7
difference low limit alarm .................................................... 1-7
dimensional drawings ..................................................... 12-21
display ............................................................................... 1-22
display and key panel ......................................................... 3-2
display examples .............................................................. 1-23
display type ............................................................. 1-22, 12-7
display update rate .............................................................. 8-2
display/recording language ............................................... 7-22
displayed contents ............................................................ 12-7
dot printing interval .................................................... 1-14, 6-1
duration (alarm delay) ............................................. 6-10, 9-19
C
E
B
Page
Page
IM 04P 01B01-01E
Page
edge .................................................................................. 1-29
effects of operating conditions ........................................ 12-20
EMC standards ..................................................................... iii
energized .................................................................. 1-11, 7-3
entering basic setting mode ................................................ 4-2
entering characters ............................................................. 4-4
entering setting mode ......................................................... 4-2
entering values .................................................................... 4-3
error messages ................................................................. 10-1
ESC key .............................................................................. 4-3
exponent ........................................................................... 1-24
external dimensions ............................................................ 2-4
external reference compensation ........................................ 1-4
Index-1
Index
Index
calibration .......................................................................... 11-4
calibration instrument ........................................................ 11-4
cassette holder .................................................................. 3-10
changing the settings .......................................................... 4-3
channel auto switching ...................................................... 3-12
channel number ................................................................ 7-11
channel printout ................................................................ 1-17
channel range ......................................................... 5-10, 6-12
channel switching interval ................................................... 8-3
character type ............................................................. 4-4, 9-3
chart cassette ...................................................................... 3-4
chart end output ................................................................ 1-27
chart holder ......................................................................... 3-5
chart paper ................................................................ 3-5, 12-4
chart paper guide ................................................................ 3-5
Index
F
Page
FAIL output ....................................................................... 1-27
fastest recording interval ................................................... 1-26
feeding the chart paper ....................................................... 3-6
felt pen ................................................................................ 3-7
filter ................................................................... 1-5, 6-2, 7-20
format of the date .............................................................. 1-19
four arithmetic operation ........................................... 1-24, 9-5
front cover tab of the chart cassette .................................. 3-11
FUNC key operation ........................................................... 4-5
H
Page
half life of the brightness ................................................... 11-9
handling precautions ........................................................... 2-1
heat sink .............................................................................. 3-3
high limit alarm .................................................................... 1-7
high limit on rate-of-change alarm ...................................... 1-7
hold ........................................................................... 1-9, 1-12
hold operation ..................................................................... 7-3
hysteresis ............................................................................ 1-8
I
Page
initializing the recorder settings ......................................... 7-27
input computation .............................................................. 12-2
input range .......................................................................... 5-1
input type .......................................................................... 12-1
inserting a character ........................................................... 9-3
inserting characters ............................................................. 4-4
installation location .............................................................. 2-2
instantaneous mode ....................................................... App-1
integration time of the A/D converter .......................... 1-6, 7-4
internal clock adjustment .................................................. 1-28
internal light LED assembly .............................................. 11-3
interval (periodic printout) ................................................. 1-18
interval (rate-of-change alarm) .................................... 1-7, 7-3
isolation ........................................................................... 12-18
K
Page
key lock ................................................................... 3-19, 7-18
keys ..................................................................................... 3-2
keys to be key-locked ....................................................... 7-18
L
Page
language ........................................................................... 7-22
level ................................................................................... 1-29
linear scaling ....................................................................... 5-3
List .................................................................................... 3-14
loading the chart paper ....................................................... 3-4
logical computation ................................................... 1-24, 9-6
low limit alarm ..................................................................... 1-7
low limit on rate-of-change alarm ........................................ 1-7
low-cut ......................................................................... 5-6, 5-7
M
Page
main display ...................................................................... 1-22
manual printout ....................................................... 1-19, 3-13
measurable range ............................................................... 1-2
measuring input terminal ..................................................... 2-6
memory backup .............................................................. 12-20
menu (basic setting mode) .................................................. 4-8
Index-2
menu (data display setup) ................................................... 8-2
menu (setting mode) ........................................................... 4-7
message ............................................................................. 6-8
message printout .............................................................. 1-19
model ..................................................................................... v
mounting brackets ............................................................... 2-3
mounting hole ..................................................................... 3-1
moving average ........................................................ 1-5, 7-19
N
Page
names of parts .................................................................... 3-1
new chart speed printout ................................................... 1-19
non-hold .................................................................... 1-9, 1-12
non-hold operation .............................................................. 7-3
normal operating conditions ............................................ 12-17
number of samples of moving average ............................... 6-3
O
Page
offset compensation mark .............................................. App-1
ON/OFF input ...................................................................... 5-5
opening the display .............................................................. vii
operation menus using the FUNC key ................................ 4-5
operation mode ................................................................... 4-1
operations on the panel keys .............................................. 4-2
optional accessories ............................................................. vi
optional functions ............................................................ 12-13
optional terminal .................................................................. 2-9
order of precedence of computation ................................... 9-4
over value ......................................................................... 9-29
over values during periodic printout ............................... App-5
P
Page
packing materials .................................................................. vii
panel ................................................................................... 2-3
panel cutout ........................................................................ 2-4
parallel connection .............................................................. 2-6
partial expanded recording ................... 1-15, 6-5, 7-21, 9-16
password (key lock) .......................................................... 3-19
password for releasing the key lock .................................. 7-18
pen caps ............................................................................. 3-7
pen holder ........................................................................... 3-7
pen offset compensation ........................................... 1-15, 7-9
periodic inspection ............................................................ 11-1
periodic printout ................................. 1-18, 6-6, 9-17, App-1
periodic printout interval .................................................... 7-14
plotter pen ........................................................................... 3-8
POC .................................................................................... 7-9
power (computation) ......................................................... 1-24
power and other computations ............................................ 9-5
power failure .................................................................. App-5
power supply ......................................................... 2-12, 12-17
power supply terminals ..................................................... 2-13
power switch ..................................................................... 2-14
precautions (input signal wiring) ......................................... 2-5
precautions (optional terminal wiring) ................................. 2-9
precautions (power supply wiring) .................................... 2-12
printing a message ............................................................ 3-16
printing the computed value .............................................. 9-22
printing the recorder settings ............................................ 3-14
printout contents ............................................................... 1-16
printout using the TLOG timer ........................................ App-4
printouts ............................................................................ 12-5
protective ground terminal ................................................ 2-13
IM 04P 01B01-01E
Index
R
Page
range type ........................................................................... 1-2
rear panel ............................................................................ 3-3
recorder ............................................................................... 1-1
recording ........................................................................... 1-13
recording (computation channels) ..................................... 1-25
recording color ...................................... 1-14, 7-8, 7-11, 9-23
recording span .................................................................... 1-2
recording start printout ...................................................... 1-19
recording zone .......................................................... 6-4, 9-15
reference junction compensation ........................................ 1-4
reference pen .................................................................... 1-15
reference time ................................................................... 7-14
reflash alarm ............................................................. 1-10, 7-3
registering display types to screens .................................... 8-3
relational computation ............................................... 1-24, 9-6
relative time mode ............................................................. 9-21
relay contact input ............................................................. 2-11
relay contact output ........................................................... 2-11
relay numbers ......................................................... 5-11, 9-12
releasing the alarm output ................................................ 3-18
releasing the key lock ....................................................... 3-19
remaining chart length ...................................................... 11-1
remote control ......................................................... 1-28, 7-28
remote control input terminals ........................................... 2-10
remote signal .................................................................... 1-29
replacement period for expendable parts ......................... 11-9
replacing felt pens ............................................................... 3-7
replacing the internal light LED ......................................... 11-3
replacing the plotter pen ..................................................... 3-8
report data ............................................................... 7-14, 9-26
report mode .................................................................... App-2
resetting the computed values ............................................ 9-1
resetting the report data .................................................... 3-17
resetting the TLOG computed value ................................. 9-22
revisions .................................................................................. i
ribbon cassette .................................................................. 3-10
ribbon feeding knob .......................................................... 3-10
ribbon is loose ................................................................... 3-10
RJC ..................................................................................... 7-7
RTD ..................................................................................... 5-2
run modes ........................................................................... 4-1
S
Page
IM 04P 01B01-01E
T
Page
tag ................................................................... 6-7, 7-11, 9-18
TC ....................................................................................... 5-2
temperature unit ................................................................ 7-31
thermocouple ...................................................................... 5-2
thermocouple input ............................................................. 2-6
time constant ....................................................................... 6-2
time printout format ................................................. 1-18, 7-26
time tick ................................................................ 1-16, App-1
time tick cancel mark ........................................... 1-16, App-1
timer .................................................................................. 9-21
TLOG computation .................................................... 1-24, 9-8
transistor input .................................................................. 2-11
trend recording ...................................... 1-14, 6-6, 9-17, 12-4
trend recording interval ....................................................... 6-1
trigger ................................................................................ 1-29
troubleshooting ................................................................. 10-4
type of measured values ................................................... 7-14
type of report data ............................................................. 9-26
types of computations ....................................................... 1-24
types of inputs ..................................................................... 1-2
U
Page
unit ............................................................................ 5-12, 9-9
unit in computations .......................................................... 1-25
V
Page
VFD ................................................................................... 1-22
Volt ...................................................................................... 5-2
W
Page
wiring (measuring input ) ..................................................... 2-8
wiring (optional terminal) ..................................................... 2-9
wiring (power supply) ........................................................ 2-12
Z
Page
zone recording .................................................................. 1-14
Index
Index
safety precautions ................................................................. iii
safety standards .................................................................... iii
safety symbols ...................................................................... iii
scale printout .................................................................. App-2
scaling ................................................................................. 1-3
scan interval ........................................................................ 1-2
secondary chart speed ........................................................ 6-9
Setting complete screen ..................................................... 4-3
setting mode ....................................................................... 4-1
settings necessary to use various functions ..................... 4-13
Setup ................................................................................. 3-14
setup guide ....................................................................... 4-13
setup items (basic setting mode) ...................................... 4-10
setup Items (setting mode) ................................................. 4-9
shunt resistor ...................................................................... 1-2
skip ...................................................................................... 5-8
sprocket teeth ..................................................................... 3-5
square root ........................................................................ 1-24
square root computation ............................................. 1-3, 5-7
square root low-cut function .............................................. 7-24
standard accessories ............................................................ vi
standard performance ..................................................... 12-19
standard temperature device ............................................ 11-5
start recording ................................................................... 3-11
starting the computation ...................................................... 9-1
status display .................................................................... 1-22
status of the remote control input terminal .......................... 9-5
stop recording ................................................................... 3-11
stoppers ...................................................................... 3-4, 3-6
stopping the computation .................................................... 9-1
suffix code .............................................................................. v
SUM scale ..................................................... 7-14, 9-14, 9-26
supported standards ....................................................... 12-18
switching the display screen ............................................. 3-12
Index-3

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Related manuals

Download PDF

advertisement