05304 QRG 9-27-05 10/27/05 12:13 PM Page 1

®

ARROW AutoCAT

®

2 WAVE

IABP Abbreviated

Operation and Troubleshooting Guide

®

ARROW 24-HOUR INTRA-AORTIC

BALLOON PRODUCT SUPPORT:

US and Canada (800)-447-IABP

(800)-447-4227

®

05304 QRG 9-27-05 10/27/05 12:13 PM Page 2

05304 QRG 9-27-05 10/27/05 12:13 PM Page 3

Modes of Operation

AutoPilot

Pump Selects ECG / AP source lead

Continuously monitors all available ECG signals and selects an alternate lead if current ECG signal is lost or noisy

Pump Sets Timing

Automatically sets inflation and deflation based on current patient condition and timing method selected

Automatic Trigger Selection

Selects and changes trigger mode based on patient condition.

Selects next best trigger source if present one is lost

Operator

User selects ECG and AP sources and adjusts as needed

User sets Timing

User selects Trigger

ECG Patient Connections

Modes of Operation

05304 QRG 9-27-05 10/27/05 12:13 PM Page 4

Power On

Power on.

LED indicates the pump is plugged in.

LED indicates 80% charge on the battery.

05304 QRG 9-27-05 10/27/05 12:13 PM Page 5

ECG Patient Connections

BEST Connection Method

5 lead ECG Cable

(offers 12 lead capability)

Skin lead connection

Alternate ECG Signal Option

SKIN selection is preset

Lead selected is displayed

Standard 1/4'' phono cable

Phono to Phono connection

AutoPilot ™ will automatically select if first ECG source available

Monitor source displayed

Patient Connections

05304 QRG 9-27-05 10/27/05 12:13 PM Page 6

Arterial Pressure Patient Connections

BEST OPTION via FiberOptix

IAB

Light bulb Icon Legend

Black light bulb with blue square

FiberOptix IAB not connected

Blue light bulb

FiberOptix IAB not zero'd

prior to insertion

Connect blue sensor and Cal Key PRIOR to insertion

To perform manual Zero of FiberOptix

IAB

Verify Auto Zero via audible tone, green light bulb and AP FOS

Zero’d message appears

Press AP select once

Green light bulb

FiberOptix IAB zero'd prior

to insertion

White light bulb

FiberOptix IAB cal value

manually adjusted

Press soft key under fiber optic sensor for Zero

Verify green light display and

AP FOS Zero’d message appears

X

Red “X” through light bulb

FiberOptix unavailable

05304 QRG 9-27-05 10/27/05 12:13 PM Page 7

Arterial Pressure Patient Connections (cont.)

Alternate arterial pressure signal options:

Standard transduced central lumen

Select transducer cable

Connect to pump

Press AP select to highlight xducer

Level as usual, open stopcock, press transducer zero

Close stopcock; verify waveform on screen

05304 QRG 9-27-05 10/27/05 12:13 PM Page 8

Arterial Pressure Patient Connections (cont.)

Alternate arterial pressure signal options:

To slave arterial pressure from remote monitor

Standard 1/4'' phono cable

Connect to pump

Press AP select to highlight monitor

Zeroing procedure performed at primary monitor

NOTE: In OFF and STANDBY modes, unassisted AP values appear in white on right side of screen.

In ON mode in 1:1 pumping, assisted AP values appear in white.

In ON mode in 1:2, 1:4 and 1:8 pumping, assisted AP values appear in white and unassisted AP values appear in yellow subscript.

05304 QRG 9-27-05 10/27/05 12:13 PM Page 9

Start up

Balloon Connection

Push firmly, right side up or upside down; it does not matter

Start Up

05304 QRG 9-27-05 10/27/05 12:13 PM Page 10

Start up

Verify:

Flashing Heart Icon

White Overlay on ECG

Both denote valid trigger

Correct IAB Volume display

Adequate Helium display

RED BAR

indicates less than

125 PSI in tank

Press Pump ON

05304 QRG 9-27-05 10/27/05 12:13 PM Page 11

Control and Function Keys

ECG Select

ECG SELECT provides selection choices for desired LEAD, input source and gain mode

To change a skin lead, press ECG SELECT once. The current lead is highlighted in white. Press the soft key under the desired LEAD label to switch leads.

AUTOGAIN is the preset gain mode. To manually gain the ECG lead, press the soft key to change the white highlight to manual and adjust gain as desired.

To select an ECG that is brought in from another monitor source, press the ECG SELECT key twice.

NOTE: Illuminated LED indicates source selected.

Control and

Function keys

05304 QRG 9-27-05 10/27/05 12:13 PM Page 12

Control and Function Keys

AP Select

AP SELECT provides selection for source, scale, alarm parameter and limit as well as zero and calibration functions.

To change input source, press AP SELECT twice.

Note: Illuminated LED indicates source selected.

AP scaling is automatic based on patient waveform. To change AP scale, press AP SELECT once. Then press the soft key under AP SCALING, and select desired scale.

To enable an AP alarm, press AP SELECT once, then press the soft key under AP Alarm Off. This will turn the alarm ON.

Press the soft key under AP ALARM MAP to toggle between a mean AP or augmentation alarm. Select the desired alarm limit by pressing the soft key to raise (>) or lower (<) the limit. A yellow box will appear on the display to the left of the arterial pressure waveform indicating the selected AP alarm source and the limit.

05304 QRG 9-27-05 10/27/05 12:13 PM Page 13

Control and Function Keys

Pump Status

ON

Starts pumping.

If pressed before STNDBY, pumping starts after one purge cycle and the pneumatic system fills with helium to 2.5mmHg.

NOTE: The first time pump ON is pressed a special purge cycle is initiated. This is a purge beat followed by 9 beats, and is repeated automatically up to three times to optimize helium concentration.

STANDBY

If pump is ON, immediately stops pumping but does not vent the pneumatic system.

If the pump is OFF, completes a four beat purge cycle and pressurizes the pneumatic system to 2.5mmHg.

After three minutes in the standby position, the system will issue an audible alert.

OFF

Immediately stops pumping, deflates the balloon and vents the pneumatic system to atmosphere.

05304 QRG 9-27-05 10/27/05 12:13 PM Page 14

Control and Function Keys

Inflate

Adjusts the inflation point in Operator mode. Inflation occurs earlier when the left arrow is depressed and later when the right arrow is depressed. Allows the operator to optimize timing by monitoring the hemodynamic changes produced on the AP waveform.

1) Move the inflation point to the right until you can clearly see the dicrotic notch (DN) on the arterial pressure waveform (green arrow).

2) Slowly move the inflation point to the left until the dicrotic notch

(40ms in front of the DN) is no longer visible.

3) Check Augmentation. Generally

AUG should be greater than SYS.

If not, additional assessment of the patient or IABP may be required.

05304 QRG 9-27-05 10/27/05 12:13 PM Page 15

Control and Function Keys

Deflate

Adjusts the deflation point in Operator mode. Deflation occurs earlier when the left arrow is depressed and later when the right arrow is depressed. Allows the operator to optimize timing by monitoring the hemodynamic changes produced on the AP waveform.

1) Move the deflation point to the left to see the effect on the AP waveform (yellow arrow). Note the rise in the ASYS and the “U” shape of the ADIA.

2) Move the deflation point to the right to lower the ASYS while simultaneously keeping the ADIA lower than or equal to the DIA.

3) Return the ASSIST RATIO to 1:1.

05304 QRG 9-27-05 10/27/05 12:13 PM Page 16

Control and Function Keys

Arrhythmia Timing

The arrhythmia timing key activates or deactivates the arrhythmia timing and/or R wave deflation function of the pump when in

AutoPilot mode. The AutoCAT 2 system will automatically recognize arrhythmias and implement real-time R wave deflation when the arrhythmia timing function is ON and conditions are acceptable for

R wave deflation. This will be indicated by the LED light on.

If automatic R wave deflation is NOT desired, press the

ARRHYTHMIA TIMING ON/OFF key, then the soft key under ARR.

TIMING until OFF is displayed. ARRHYTHMIA TIMING OFF will be displayed on the screen (when an arrhythmia is present) and an alternate algorithm will be utilized for deflation.

To select R wave deflation ON at all times, press the ARRHYTHMIA

TIMING ON/OFF key, then the soft key under R-WAVE DEFL until ON is displayed. In this case, R wave deflation (AFIB trigger) will be selected whether an arrhythmia is present or not.

05304 QRG 9-27-05 10/27/05 12:13 PM Page 17

Control and Function Keys

Trigger Modes (Operator Mode Only)

Select Operator mode and then Press the trigger key to change trigger modes. The selections will appear on the bottom of the display.

PATTERN

Preset trigger; triggers on ECG. Recommended for regular rhythms with normal QRS complex and HR <130. Rejects pacer spikes.

PEAK

Triggers on ECG. Recommended for regular rhythms with HR >130.

Recommended for wide or abnormal QRS complexes. Recommended for irregular rhythms when R-wave deflation is not desired.

Rejects pacer spikes.

05304 QRG 9-27-05 10/27/05 12:13 PM Page 18

Control and Function Keys

Trigger Modes (Operator Mode Only)

AFIB

Triggers on ECG. Recommended for irregular rhythms. Provides

REAL TIME (R-wave) deflation. Rejects pacer spikes.

VPACE

Uses V-pacer spikes to trigger. MUST BE 100% PACED. For V or

AV sequential pacers. Recommended for V or AV paced rhythms with poor QRS complex.

APACE

Uses A-pacer spikes to trigger. MUST BE 100% PACED. For Atrial pacers only. Recommended for A paced rhythms with poor QRS complex.

05304 QRG 9-27-05 10/27/05 12:14 PM Page 19

Control and Function Keys

Trigger Modes (Operator Mode Only)

AP

Uses AP waveform to trigger. Recommended when ECG is not available or too noisy. AP NOT RECOMMENDED FOR IRREGULAR

RHYTHMS.

INTERNAL

Uses IABP internal signal for triggering. Used only when patient has no ECG and no Cardiac Output. ASYNCHRONOUS TO PATIENT

CARDIAC ACTIVITY. Press INT again to confirm.

Not available in AutoPilot mode.

05304 QRG 9-27-05 10/27/05 12:14 PM Page 20

Control and Function Keys

Assist Ratio

ASSIST RATIO 1:1

Initiates one inflation-deflation cycle for each cardiac cycle. Preset assist ratio at start-up.

Maximizes counterpulsation support.

ASSIST RATIO 1:2

Initiates one inflation-deflation cycle for every other cardiac cycle. Generally used to assess and optimize timing in Operator mode and to wean patient from IABP support.

ASSIST RATIO 1:4

Initiates one inflation-deflation cycle for every fourth cardiac cycle. Used to wean patient from

IABP support.

ASSIST RATIO 1:8

Initiates one inflation-deflation cycle for every eighth cardiac cycle. Used to wean patient from

IABP support.

05304 QRG 9-27-05 10/27/05 12:14 PM Page 21

Control and Function Keys

Balloon Volume

Press BALLOON VOLUME key.

Select INCREASE / DECREASE until desired volume is displayed. Press FULL VOLUME to return to volume based on balloon connector.

Press APPLY key to initiate change. Press CANCEL key to cancel change.

05304 QRG 9-27-05 10/27/05 12:14 PM Page 22

Control and Function Keys

Recorder

To change recorder settings:

Press HOME

Press RECORDER SETUP

Select desired waveforms or set recorder automatic print interval

Recorder will print waveforms highlighted in white. Press soft key under desired waveforms to select or de-select.

05304 QRG 9-27-05 10/27/05 12:14 PM Page 23

Control and Function Keys

Alarm Control

RESET

Silences the audible alarm tone and clears the alarm message.

If pumping was interrupted, the alarm message is not cleared until STNDBY or ON is pressed. If there is more than one alarm condition, one message is cleared at a time.

ON/OFF

Turns pneumatic alarms audio, recording, drain and refill on or off.

Alarm message will be displayed. To select alarm time off, press key under desired setting. Time remaining for alarms off is displayed above

AP Scale. Press ON/OFF again to turn on alarms.

NOTE: DO NOT leave pump unattended with pneumatic alarms disabled.

05304 QRG 9-27-05 10/27/05 12:14 PM Page 24

Control and Function Keys

Home

AP SCALING

Selects AP Scaling mode and scales for display and recorder.

RECORDER SETUP

Selects waveforms to be recorded and time interval for automatic recordings.

WEANING SETUP

Allows the user to make changes in volume and/or assist ratio and set the time interval for the weaning step. It allows the user to start the weaning step and cancel the weaning step.

SHOW STATUS

This key will display a summary of all current operational settings as well as selected information which is tracked by the AutoCAT 2.

05304 QRG 9-27-05 10/27/05 12:14 PM Page 25

Control and Function Keys

Home- (continued)

AUDIO SETUP

Allows the user to set the key volume, and turn it ON/OFF and independently set alarm volume.

HEMODYNAMICS

Automatically calculates:

1. AUG-SYS

2. AUG-ADIA

Freezes the hemodynamic numbers for 30 seconds and displays the assisted pressures in white and the unassisted pressures in yellow.

CLOCK SETUP

Allows the user to set the time and date for the pump.

05304 QRG 9-27-05 10/27/05 12:14 PM Page 26

Control and Function Keys

Cursor

Moves horizontal cursor to desired assessment point. Value is displayed above cursor on the right hand side.

05304 QRG 9-27-05 10/27/05 12:14 PM Page 27

Alarms/Alerts

NOTE: An alarm may cause the pump to stop pumping. The pump will display a message on the screen to assist in troubleshooting. Once the condition is corrected:

To resume pumping

Press Alarm RESET

Press Pump ON

If the alarm condition repeatedly appears, refer to the Operator’s Manual for further information or call Arrow’s IABP support line at 800/447-IABP.

GAS ALARMS

(CLASS 1 ALARMS)

• Pump to OFF

• IAB deflated

• Helium vented to atmosphere

• Audible alarm

• Freeze display

• Alarm/troubleshooting message displayed

• Prints 10 second strips

(3 prior to alarm/7 after alarm)

TRIGGER LOSS ALARMS

(CLASS 2 ALARMS)

• Pump to Stand-by

• IAB deflated

• Helium maintained in bellows

• Audible alarm

• Alarm/troubleshooting message displayed

• Automatic reset of some alarms when condition is corrected

AUDIBLE ALERTS

(CLASS 3 ALERTS)

• Pump remains ON

• Audible alarm

• Alarm/troubleshooting message displayed

• Automatic reset of some alarms when condition is corrected

MESSAGE ALERTS

(CLASS 4 ALERTS)

• Pump remains ON

• Alarm/troubleshooting message displayed

• Automatic reset of some alarms when condition is corrected

Alarms / Alerts

05304 QRG 9-27-05 10/27/05 12:14 PM Page 28

Alarms/Alerts

GAS ALARMS

(CLASS 1 ALARMS)

• System error alarms 1-8

• Unable to Refill

• Possible Helium loss

2 and 3

• High Pressure

• High Baseline

• Large Helium Leak

• Purge Failure

TRIGGER LOSS ALARMS

(CLASS 2 ALARMS)

• Stand-by alarm disabled

• Pump in Stand-by for greater than 3 minutes

• ECG trigger loss

(Operator mode)

• AP trigger loss

(Operator mode)

• ECG lead fault

(Operator mode)

• Trigger loss (AutoPilot mode)

AUDIBLE ALERTS

(CLASS 3 ALERTS)

• Drain failure

• Deflation beyond 100%

(Operator mode)

• Insufficient Time to Inflate

• Insufficient time to Deflate

• Battery Inoperative

• System Running on

Battery

• Battery time less than

20, 10 and 5 minutes

• Possible ECG detected

• Weaning Step Complete

• Arterial Pressure alarm

• Low Helium Tank Pressure

• AP FOS Signal Weak

• AP FOS CAL key missing or corrupt

• AP FOS Sensor out of range

MESSAGE ALERTS

(CLASS 4 ALERTS)

• Possible Late Deflation

• Erratic Trigger: R-Wave

Deflation

• Erratic Triggering

• ECG Lead Fault (Autopilot mode)

• No ECG Signal Available

• No AP Signal Available

• Arrhythmia Timing not available

• Bad Real time clock

Battery

• Bad Ram Battery

05304 QRG 9-27-05 10/27/05 12:14 PM Page 29

Alarms/Alerts

GAS ALARMS

(CLASS 1 ALARMS)

• Reset alarm

• Review alarm message and troubleshooting information

• Correct condition if known

• Restart pump

• Review alarm strip

• If alarm persists refer to Chapter 8 of the

Operator’s manual for further information.

TRIGGER LOSS ALARMS

(CLASS 2 ALARMS)

• Correct condition if known. Many alarms will reset automatically when the condition is cleared

-OR-

• Reset alarm

• Review alarm message and troubleshooting information

• Correct condition if known

• Restart pump

• If alarm persists refer to Chapter 8 of the

Operator’s manual for further information.

AUDIBLE ALERTS

(CLASS 3 ALERTS)

• Correct condition if known. Many alarms will reset automatically when the condition is cleared

-OR-

• Reset alarm

• Review alarm message and troubleshooting information

• Correct condition if known

• If alarm persists refer to

Chapter 8 of the

Operator’s manual for further information.

MESSAGE ALERTS

(CLASS 4 ALERTS)

• Correct condition if known. Many alarms will reset automatically when the condition is cleared

-OR-

• Reset alarm

• Review alarm message and troubleshooting information

• Correct condition if known

• If alarm persists refer to

Chapter 8 of the

Operator’s manual for further information.

05304 QRG 9-27-05 10/27/05 12:14 PM Page 30

Notes:

–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––

–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––

–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––

–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––

–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––

–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––

05304 QRG 9-27-05 10/27/05 12:14 PM Page 31

05304 QRG 9-27-05 10/27/05 12:14 PM Page 32

www.arrowintl.com

Caution: Federal law restricts this device to sale by or on order of a physician.

Contents of unopened, undamaged package are sterile. Disposable.

Refer to package insert for current warnings, indications, contraindications, precautions, and instructions for use. U.S. Patent Nos. 6,258,035; 6,569,103; and 6,887,206.

Applicable to software version 2.22 and greater.

A2W-QRG 9/05 5M CRB-05-0288 ©2005 Arrow International, Inc. All rights reserved. Printed in the Czech Republic.