Troubleshooting
The most important principle in troubleshooting the SMART System 5 is to check each component of the system separately and trace any malfunction to its source.
Verify each of these factors:
• Balance Calibration
• Microwave Power Measurement
• Standard Solution Test
• Analytical Method (verify analysis method) – If necessary, contact the CEM
Applications Department or the nearest subsidiary or distributor.
A qualified service technician, using standard troubleshooting techniques, should be able to identify and resolve operational problems. If a problem persists, call the
CEM Service Department or the nearest subsidiary or distributor.
Troubleshooting Guide
Symptom
Instrument Inoperative
Possible Cause/Remedy
Ensure that instrument is properly connected to power outlet and that the power switch is in the
“on” position. Check fuses and replace if necessary.
Balance overloaded. Use less sample.
Ensure that balance stem is installed and centered.
Balance Will Not Tare
Balance Not Operating
No Microwave Drying
Sample Burning
Instrument cover not closed and latched properly.
Check dual voltage supply and fuses (10A fuses for 60 Hz and 5A for 50 Hz) and correct dual voltage supply and/or replace fuses, if required.
Sample too large and/or unevenly applied to pad.
Sample not homogeneous. Sample not positioned under temperature sensor. Power level too high.
Results Not
Reproducible
Low Power
Improper Heating
Discoloration or Oil
Buildup on Air Shield
Sample applied unevenly to pad. Sample not homogeneous and absorbs microwaves unevenly.
Keep sample container closed when not extracting sample. Work quickly when applying sample to pads and placing pads on balance pan.
Incorrect Results Verify proper sample handling, spread techniques and method parameters.
“Data Results Invalid” Final weight greater than initial weight. Repeat
Displayed at End of Test test. Check balance calibration.
“Cover Open” Message Ensure that the instrument cover is properly closed.
Check for faulty relay and replace if required.
Ensure that the air shield is installed and not clogged.
Replace the air shield.
142
Troubleshooting Guide
Symptom
Balance Over Load
Error
Balance Under Load
Error
Thermal Overload
Error
Fuses Blow
Balance Error
Instrument Will Not
Calibrate
Unstable Balance
Readings
Balance Displays
Incorrect Weight
IR Reads Negative
Number(s)
Instrument Top Cover
Difficult to Close or
Noisy when Opened
Garbled Display
Intermittent “clicking” when Microwaves On
Possible Cause/Remedy
Weight on balance is too large to be recognized by balance.
Weight on balance is too small to be recognized by balance.
Ensure that instrument cover is closed properly.
Ensure that switches on all three interlocks are adjusted properly.
Check dual voltage supply and fuses (10A fuses for 60 Hz and 5A for 50 Hz) and correct dual voltage supply and/or replace fuses, if required.
Balance stem not installed. Turn instrument off, install balance stem and turn instrument on.
Ensure that instrument is located on a stable work surface.
Balance cable not installed. Properly install cable.
Balance not calibrated. Perform calibration using
10g or 50g weight.
Instrument turned on before balance stem and pan were installed. Turn instrument off, install stem and pan, and turn instrument on.
Faulty balance. Contact CEM Service.
Faulty balance. Contact CEM Service.
Faulty IR sensor. Contact CEM Service.
Gas spring installed improperly or faulty gas spring. Contact CEM Service.
Cold start instrument. Contact CEM Service.
Arcing at microwave waveguide slot. Contact
CEM Service.
143