User Manual Lx 500

Add to my manuals
25 Pages

advertisement

User Manual Lx 500 | Manualzz

LABOMED

Lx 500

User Manual

Research Microscopy

To ensure proper use of this instrument as well as to avoid injury while operating instrument, understanding this manual completely before use is highly recommended.

CONTENTS

1

2

INTRODUCTION

SAFETY INFORMATION

3

4

5

6

7

Lx 500 BINOCULAR

Lx 500 TRINOCULAR

Lx 500 ERGONOMIC

STANDARD COMPONENTS

OPTIONAL ACCESSORIES

8

9

INITIAL SETUP

ASSEMBLY

10 SUMMARY OF BRIGHTFIELD OBSERVATION PROCEDURE

11 DETAILED OBSERVATION PROCEDURE

12 TROUBLESHOOTING GUIDE

13 SPECIFICATIONS

1

2-4

8

9-10

5

6

7

11

12

13

14-18

19-20

21

1

1

INTRODUCTION

The Lx 500 is a research microscope reflecting a modern design as well as the latest in optical and mechanical advancements. Designed for professionals as well as clinicians, this microscope offers many features and functions for a diverse set of applications. Here are a few points highlighting the benefits of the Lx500:

- Extra clarity and contrast is provided through a 360° rotatable Binocular body inclined at 30° with IPD adjustments.

- The pressure die cast stand consists of ball bearing, friction based focusing mechanism to avoid any loss in motion.

- The sturdy new stylish design provides a high degree of comfort as well as stability.

- The high powered objectives are spring loaded to prevent accidental damage to specimen slides.

- The reverse angle quadruple nosepiece has a comfortable ribbed grip for easy rotation that also safeguards the turret

system against any damage. All positions are par-centered and par-focalised ensuring the highest level of accuracy.

- The ball bearing mechanical stage allows smooth travel over a 76 x 50mm area with spring loaded stage clamps for retaining

the specimen at the exact position desired. A 0.1mm vernier scale provides accurate location of specimen area.

- High power illumination is delivered through our well crafted Universal Power Supply operatable within range of 100V- 240V

AC input. This ensures continuous operation even under fluctuating voltages.

- Our Halogen bulb (6V-30W) has an average life span of up to 500 hours.

- The Lx 500 comes equipped with a removable flip top N.A. 1.25 Abbe condenser with aspheric lens for brighter illumination

levels, and an iris diaphragm for better resolution and contrast control.

Lx 500

2

SAFETY INFORMATION

1. After the microscope has been used for observation of a specimen containing bacteria, clean all parts coming in contact with the specimen to prevent infection.

to clean all surfaces that may have come in contact with the specimen.

Fig. 1

Applicable bulb 6V30W Halogen bulb P/N EL-455

3. Install microscope on a sturdy, level table or bench and avoid any restriction of air vents in the base of the unit.

Do not place microscope on a flexible surface, as this could result in blocking the air vents and cause overheating.

4. Always use the power cord provided by LABOMED. If the proper power cord is not used, product safety performance cannot be warranted.

5. When installing the microscope, route the power cord away from the microscope frame. Should the power cord come in contact with the microscopes base, the power cord could melt due to overexposure to heat.

6. Always ensure that the grounding terminal of the microscope and that of the wall outlet are properly connected. If the unit is not grounded, LABOMED can not warrant electrical safety.

7. Never allow metallic objects to penetrate the air vents of the microscope frame as this could result in user injury and damage to the microscope.

8. After operation of microscope, be sure to disconnect power cord from connector socket of the microscope or from the wall power outlet.

Safety Symbols

The following symbols are found on the microscope. For optimal use, it is recommended that users understand these symbols and always use the equipment as prescribed.

Symbol

!

Explanation

This surface has a tendency to heat up and should not be touched unless system has completely cooled down.

Before use, carefully read the instruction manual. Improper use could result in injury to the user and/or damage to the equipment.

Warning against risk of electric shock.

|

Main switch is ON.

Main switch is OFF.

2

3

!

Caution

If the microscope is used in a manner not specified by this manual, the safety of the user may not be warranted. In addition, the equipment may also suffer damage. Always use the equipments as outlined in this instruction manual.

1

Getting Started

Fig. 2

1. A microscope is a precision instrument with delicate glass components.

Please handle with care.

2. Do not use the microscope where it is subjected to direct sunlight, high temperature , humidity, dust and vibrations. (For the operating conditions, see chapter 14, "SPECIFICATIONS")

3. The microscope is ventilated by natural convection. Be sure to leave enough space (10 cm or more) around body when installing the unit.

4. Arm handle is provided for carrying the microscope.

!

To prevent damage, do not hold the microscope by the stage or observation tube.

Be sure to remove the specimen from the stage clip while transporting unit to avoid damage to the specimen slide.

2

Maintenance and Storage

1. Clean all glass components by wiping gently with cleaning cloth provided. To remove fingerprints or oil smudges,

!

wipe with cleaning cloth slightly moistened with a mixture of petroleum (85%) and isopropanol (15%).

Since solvents such as petroleum and isopropanol are highly flammable, they must be handled carefully.

Be sure to keep these chemicals away from open flames or potential sources of electrical sparks - for example, electrical equipment that is being switched “ON” or “OFF”. Also remember to always use these chemicals only in a well and ventilated room.

2. Do not attempt to use organic solvents to clean the microscope components other than the glass components. To clean non-glass components, use a lint-free, soft cloth slightly moistened with a diluted neutral detergent.

3. Do not disassemble any part of the microscope as this could result in malfunction or mitigated performance.

4. When not using the microscope, ensure that the frame is fully cooled before storing the unit in a dry locker or covering with a dust cover

(provided).

5. To clean the condenser, fully loosen the securing knobs and remove the condenser by lowering condenser bracket then, wipe the front lens of the condenser with optical cleaning solution (mixture suggested above) and lens tissue.

The condenser can be re-attached by replacing the condenser in its seat, tightening securing screws, and raising condenser bracket to desired position.

Fig. 3

Lx 500

4

Care & Maintenance

Your microscope has been engineered for a long and safe operational life with the least amount of maintenance required. In general, routine maintenance is limited to keeping the microscopes working parts lubricated and optics clean. Always cover the microscope with the provided dust cover when not in use.

Optical Cleaning:

1. The objectives have been adjusted for a tight fit to prevent any damage during transportation. To remove an objective, rotate it counterclockwise while gripping it with a rubber sheet, etc. to avoid any slippage.

2. To clean the lens surfaces, remove dust using a soft brush or compressed air (cans available at your local electronics store). For removing finger marks or grease, soft cleaning cloth or lens tissue lightly moistened with cleaning solution (85% petroleum ether and 15% isopropanol) should be used. For cleaning the objective optics, use Methanol. Observe sufficient caution in handling Methanol. Place the objectives and/or eyepieces on a dustfree surface (e.g. aluminum foil). All other optical components to be cleaned should be as accessible as possible.

3.. Blow all loose dust particles away with compressed air or mini dust blower.

4. Remove all water-soluble dirt with distilled water. If this is unsuccessful repeat using a solution of diluted hand soap liquid. Remove any remaining residue with a dry cotton swab.

5. To remove oil, use a solution of diluted hand-soap liquid initially. If this does not produce a satisfactory result, repeat the cleaning using a solvent (Optical Cleaning Solution 85% petroleum ether and 15% isopropanol).

6. Grease must always be removed using a solvent.

7. Cleaning is achieved by using a spiral motion from the center to the rim. Never wipe using zig-zag movements as this will only spread the dirt. With larger optical surfaces (e.g. tube lenses) the spiral motion starts initially at the rim before moving to the middle and is then followed by a center to rim cleaning motion. Normally several spiral wipes are recommended.

We recommend pure, volatile petroleum ether or Optical Cleaning Solution as explained in point 3 above.

zig-zag motion (X) spiral motion ( )

Wipe using a spiral movement. Do not use a zig-zag motion!

2.

Cleaning of painted surfaces :

Avoid the use of any organic solvent ( e.g. thinner, xylene, ether, alcohol etc.) for cleaning of painted surfaces of the instrument. Painted surfaces can be cleaned with a very lightly moistened micro fiber cloth. Loose dust and other dirt particles can be removed using a soft bristle brush used exclusively for this purpose.

4

5

Lx 500

3

Lx 500 Binocular

Eyepieces

Revolving nosepiece

Objectives

Specimen holder

Aperture iris diaphragm ring

Abbe condensor

Filter holder

Binocular viewing tube, 30° inclined

Observation tube clamping screw

Mechanical stage

Coarse and fine focus knob

X-Y movement control knob

4

Lx 500 Trinocular

Eyepieces

Revolving nosepiece

Objectives

Specimen holder

Aperture iris diaphragm ring

Abbe condensor

Filter holder

Lx 500

Trinocular Port

Trinocular viewing tube, 30° inclined

Observation tube clamping screw

Mechanical stage

Coarse and fine focus knob

X-Y movement control knob

6

7

Lx 500

5

Lx 500 Ergonomic Binocular

Eyepieces

Revolving nosepiece

Objectives

Specimen holder

Aperture iris diaphragm ring

Abbe condensor

Filter holder

Ergonomic Binocular viewing tube,

0° to 25° inclined

Observation tube clamping screw

Mechanical stage

Coarse and fine focus knob

X-Y movement control knob

Lx 500

6

STANDARD COMPONENTS

“Please note that the contents of your microscope may vary as the optional configuration, contrasting method or viewing body opted for may not be of the standard configuration highlighted here”

Binocular viewing tube, 30° inclined

Observation tube clamping screw

Eyepieces

Revolving nosepiece

Objectives

Specimen holder

Aperture iris diaphragm ring

Abbe condensor

Filter holder

Mechanical stage

Coarse and fine focus knob

Daylight (blue) filter

Paired Eyepieces Allen Wrench 3mm

Power Cord

8

7

OPTIONAL ACCESSORIES

System Diagram of Optional Accessories

Video adapter1/2"&1/3" Digital camera adapter iVu 1500 iVu 3000 iVu 5000 iVu 7000

Binocular head Trinocular head Ergonomic head WF 10x WF 16x WF 20x

RP 4x RP 10x RP 20x RP 40x (SL) RP 50x (SL, Oil) RP 100x (SL, Oil) RP PH 10x RP PH 20x RP PH 40x (SL) RP PH 100x (SL, Oil)

RP FL 4x RP FL 10x RP FL 20x RP FL 40x (SL) RP FL 50x (SL, Oil) RP FL 100x (SL, Oil) RP 100x,(Oil, Iris)

Flip top condenser Polarizer Mount Analyzer Mount

Blue filter Green filter

Halogen bulb

9

Installation and Operation of Optional Accessories

1 iVu Camera Module System

1.Mount the Video adapter 1/2” (part # 3143300-912) on Trinocular

observation head.

2. Mount iVu Camera Module System on video adapter.

Fig. 4

2

Optional Eyepieces

Lx 500

10X eyepieces are provided. To replace:

1. Pull out the 10x eyepieces out from the observation heads ocular tube.

2. Insert desired eyepieces in empty ocular tube.

3

Phase Turret

Fig. 5

1

2

Fig. 6

1. Remove the condenser from its position by loosening the two thumb screws (1) on both sides of the condenser as shown in figure 6.

2. Place the phase annulus turret in position from where the condenser has been removed.

3. The long “push in to engage” thumbscrews (2) on both sides of the phase turret are provided for proper alignment of the phase ring. By disengaging these screws, the phase turret can be adjusted at any desired phase annulus (10x, 20x, 40x & 100x) by turning the thumbwheel to the desired position. There is also an ‘0’ setting on the thumbwheel for bright field application

Note: Refer to the user manual provided with the Phase Contrast kit for centering and alignment.

10

11

8

INITIAL SETUP

1

Objectives

Objectives are factory set. Objectives are par-centered and parfocalised during assembly.

All objectives have been secured for a tight fit to prevent them from coming loose during transit. To remove an objective, rotate it counterclockwise while holding it with a rubber grip to avoid any slippage.

Fig. 7

2

Observation Head

1

Install the observation head using the following procedure:

1. Using allen wrench 3mm (provided), loosen the Head Locking

Screw (1) and remove the dust cover cap provided in dovetail cavity as well as on observation head dovetail.

2. Mount the Observation Head by engaging the dovetail provided at the bottom of the head into the dovetail cavity provided in the microscopes arm.

3. Tighten the Head Locking Screw (1) after positioning the

Observation Head as desired. See figure 8.

Fig. 8

3

Eyepieces

Insert the eyepieces into the ocular tube of Observation Head using following procedure:

1.

Remove the protective caps from the observation tube.

2.

Insert 10x eyepieces into the spring loaded and to prevent undesired rotation of eye pieces. See figure 9.

Fig.9

Lx 500

9

ASSEMBLY

Each standard set can be assembled by simply attaching the bulb, filter and power cord.

1

Installing or Replacing the Lamp Bulb

Before attaching the lamp bulb, remove the parts that may drop such as the filter and specimen from the microscope frame, and place the microscope on its back so that the bottom plate is exposed.

1. Pull out the lamp house cover (1) to detach it.

2. Pull up the lamp from the lamp holder & replace with lamp.

3. Replace the lamp house cover.

Always use the designated bulb. Using a bulb other than those specified by LABOMED may lead to a fire hazard and decentering of light. Fingerprints or stains on the lamp bulb reduce its life. If contamination occurs, wipe bulb surface with a cloth slightly moistened with alcohol.

Fig. 10

!

Caution: For Bulb Replacement During Use or Right After Use

The bulb, lamp socket and areas near these will be extremely hot during and right after use. Set the main switch to" O" (OFF), disconnect the power cord from the wall outlet, and allow the bulb and lamp socket to cool before replacing the bulb with a new bulb of the designated type. Cooling time may vary to users discretion.

2

Mounting the Daylight (Blue) Filter

2

1

This filter modifies the color of observation light into a natural (daylight color).

• Fit the filter (1) into the bottom of the condenser (2) until it clicks into place. (Figure 11).

Fig. 11

12

13

10

SUMMARY OF BRIGHTFIELD OBSERVATION PROCEDURE

Flip the main switch to “ON”

Place the specimen on the stage.

Engage the 10X objective in the light path.

Bring the specimen in focus.

Adjust the observation tube and eyepieces

Adjust the interpupillary distance.

Adjust the dioptric setting.

On/Off

A/C Inlet

Adjust the aperture iris diaphragm.

Engage the objective to be used in the light path and bring the specimen in focus.

Engage the required filters.

Adjust the brightness.

Observe Specimen.

Lx 500

11

DETAILED OBSERVATION PROCEDURE

1

Turning the Lamp ON

1. Flip the main switch to ” ” as shown in Figure 12.

2. Rotating the light intensity adjustment knob (Fig. 13) in the direction of the arrow increases brightness and rotating knob in the opposite direction decreases brightness. The intensity bar next to the knob indicates the direction of intensity level.

1

Fig. 12

2

Fig. 13

2

Placing specimen on the stage

3

1

Fig. 14

5

4

2

1. Rotate the coarse adjustment knob (2) in anticlockwise direction to fully lower the stage.

2. Open the bow-shaped lever (3) outward by pulling on lever handle (1), place the specimen by sliding the specimen glass plate(s) on the stage from the front toward the slide seat at the rear.

3. After positioning your specimen slides,(2 max) return the bow-shaped lever (3) gently by slowly releasing control knob (1).

4. Rotating the upper co-axial knob controlling the Y-axis movement (4) moves the specimen in the vertical direction. Rotating the lower knob

(5) moves the specimen in the X- axis or horizontal direction.

!

Do not adjust the specimen holder directly by hand. This will damage the rotary mechanisms.

force of the X/Y knobs become stiff. Stop rotating the at this point.

14

15

2

1

Slide glass

Fig. 15

Cover glass

Cover glass

This is the glass plate placed on the specimen. For optimum optical performance, the cover glass thickness, which is the distance from its surface to the specimen surface, should be 0.17 mm.

Slide glass

This glass plate should ideally have a length of 76 mm, width of 26 mm

±1 mm and thickness between 0.9 and 1.4 mm.

“Vernier Scales”

These scales allow for easy identification of the specimen’s position

(coordinates), making it easy to return to a particular region of interest after scanning the slide. (Figure 16.)

1. The horizontal coordinate can be read at position (1) on the specimen

holder.

2. The vertical coordinate can be read at the index line (2).

Fig. 16

3

Adjusting the Focus

WD

3

Fig. 17

1

2

Focusing Procedure

1. Rotate the coarse adjustment knob (1) clockwise so that the objective

(3) is as close as possible to the specimen (We recommend starting with

10X).

2. While observing the specimen through the eyepieces, slowly rotate the

coarse adjustment knob (1) counterclockwise to lower the stage.

3. When coarse focusing of the specimen is obtained (a blurr image is

observed), rotate the fine adjustment knob (2) for fine detail focusing in anticlockwise direction.

Working Distance (WD)

The WD refers to the distance between each objective and the specimen,

Objective

Magnification

WD (mm)

4X

30.5

10X

4.82

40X

0.55

100X

0.11

Lx 500

4

Adjusting the Interpupillary Distance (IPD)

Fig. 18

The inter-pupillary distance adjustment consists of moving the two eyepieces to align with both eyes’ pupils so that you can observe a single microscopic image through two eyepieces in stereo vision. This greatly helps to reduce fatigue and discomfort during observation.

While looking through the eyepieces, move both eyepieces laterally until the left and right fields of view coincide completely. The position of index dot (•) indicates the inter-pupiliary distance value.

Note your interpupillary distance so that it can be quickly referred to in the future. This may need when multiple users work with the microscope.

5

Adjusting the Diopter

Procedure for adjusting the diopter:

1. Rotate the right eyepiece to match the markings of your IPD (If your

IPD is 64, rotate the eyepiece to 64 mark).

2. While looking through the right eyepiece with your right eye, rotate

the coarse and fine adjustment knobs to bring the specimen into

focus.

3. While looking through the left eyepiece with your left eye, rotate only

diopter adjustment ring on the eyepiece until specimen is at its best

possible focus.

Fig. 19

Using the Eye Guards

When Wearing Eyeglasses

Use with the eye guards in the normal, folded-down position. This will prevent the eyeglasses from being scratched.

When Not Wearing Eyeglasses

Extend the folded eye guards outwards (direction of the arrow) to prevent ambient light from entering into your line of vision.

Fig. 20

16

6

Centering the Condenser

Fig. 21

There is a possibility of having the condenser de-centered during shipping. it is mandatory to check the centering before observing the specimen as it is required to distribute even light to the specimen.

For models with Koehler illumination:

1. Bring the 10x objective (1) in the front position.

2. Move the condenser to the highest point.

3. Close the koehler. (2) Focus the light spot as show in the illustration below.

4. Use the centering screws (3) provided on left and right sides of the condenser bracket to bring the light spot to center of the view field.

17

6

Adjusting the Condenser Position and Aperture Iris Diaphragm

The condenser is most often used in the highest position. If the observed field of view is not bright enough, brightness may be improved by lowering the condenser slightly

2

1

1. Rotate the condenser height adjustment knob (2) to move the

condenser to the highest or desired position.

2. The aperture iris diaphragm ring (1) has an objective magnification scale

(4X, 10X, 40X, 100X). Rotate the ring so that the magnification of the

objective is clearly visible while using the microscope.

Fig. 22

7

Switching the Objectives

1

Rotate the revolving nosepiece (1) so that the objective to be used is in line above the specimen. Always use the ribbed grip(1) to rotate the objective nosepiece.

Fig. 23

Lx 500

8

Using the 100X Immersion Objective

Fig. 24

The designated immersion oil should be in contact with the cover lens of the 100X immersion objective. If not, the specimen will appear distorted and dull. It is recommended that LABOMED immersion oil is always used.

Immersion Process:

1. Bring the specimen in focus using first the 10x, then 40x objective.

2. Disengage the 40x cycling towards 100x, and place a drop of immersion

oil on the center point of the specimen.

3. Rotate the revolving nosepiece to engage the immersion objective and

rotate the fine adjustment knob to bring the specimen into focus

(Since air bubbles in the oil will affect the image quality, make sure that the oil Is free of bubbles. To remove bubbles, rotate the revolving nosepiece slightly to agitate the oil).

4. The condenser of this microscope manifests the full performance when

oil is placed between the slide glass and the front lens of condenser. If oil

is not placed there, the observed image may appear dark.

5. After use, remove oil from the objective front lens by wiping with lens

tissue slightly moistened with an ether (70%) alcohol (30%) mixture.

Caution

If immersion oil makes contact with your eyes, rinse eyes out thoroughly with fresh water. If immersion oil makes contact with skin, wash affected areas with soap and water.

If prolonged discomfort is experienced, consult your physician immediately.

18

19

12

TROUBLESHOOTING GUIDE

Under certain conditions, performance of the unit may be adversely affected by factors other than defects. If problems occur, please review the following list and take corrective action as needed. If problem persists, please contact LABOMED or your local LABOMED dealer.

Observation Cause

1. Uneven brightness in observation The objective is not engaged in the

field light path

2. Dust or stains are visible in

observation field

3. Glare visible in field of View

4. Observation image is hazy

or unclear

Remedy

Engage the objective into position until the nose turret clicks

The condenser is too low

The objective, eyepiece, condenser and/or window lens are dirty

Raise up to achieve more light (Pg#18)

Clean them thoroughly as previously prescribed in “Optical Cleaning”(pg#4)

The eyepiece, condenser, window lens Clean glass parts thoroughly with lens and/or specimen glass is dirty tissue and cleaning solution

prescribed in“Optical Cleaning”(Pg#4)

The condenser is too low

The condenser iris diaphragm ring is

closed

The objective is not engaged in the light path

The objective, eyepiece, condenser and/or specimen glass is dirty

Raise condenser light

Adjust the aperture according to the objective magnification (Pg# 18)

Engage the objective into position until it clicks (Pg# 18)

Clean glass parts thoroughly with lens tissue and cleaning cloth

Immersion oil is not used with an immersion objective.

Use immersion oil as suggested (Pg# 19)

Bubbles are present in immersion oil Remove the bubbles by agitation

(Pg# 19)

The specified immersion oil is not used Use the immersion oil supplied by

LABOMED

5. Part of image is defocused The objective is not properly engaged Engage the objective into position in the light path until the nose turret clicks

The specimen is not set properly on the stage

Set the specimen correctly on the stage and secure using the specimen holder (Pg# 15)

Raise the condenser 6. Coarse focus adjustment cannot

lower the stage low enough

The condenser is too low

7. Fields of view through both The interpupillary distance is not

eyepieces is inconsistent adjusted properly

Adjust IPD to the appropriate setting

(Pg# 17)

Dioptric compensation for the two eyes Adjust diopter settings (Pg# 17) is not set

The left and right eyepieces are of different magnification

Ensure that both eyepieces are of same magnification. LABOMED does not recommend using third party eyepieces in conjunction with

LABOMED microscopes.

Lx 500

Observation

12. Objective hits the specimen when

an objective is switched to a higher

magnification objective

13. Bulb does not turn On

Cause

The specimen slide is upside down

The cover glass is too thick

The stage is raised too high

The slide has slipped from the slide holder

Slide is of excessive thickness

Bulb is not mounted

Bulb is blown

Remedy

Set the specimen correctly with the cover glass facing upwards

Use a cover glass with thickness of

0.17mm

Lower the stage

Re-position the slide in the slide holder

Use slides with thickness between 0.9 and 1.4mm

Attach a bulb

Replace the bulb

The power cord is unplugged / Not firmly Ensure power cord is securilly plugged secured into the box socket + wall outlet

14. Bulb blows easily

Fuse is blown

Battery is low

The specified bulb is not used

Check and replace with live fuse

Charge battery

Replace with the specified bulb

20

21

13

SPECIFICATIONS

1. Illumination

2. Focusing mechanism

3. Revolving nosepiece

4. Observation tube

5. Stage

6. Condenser

7. Dimensions & Weight

8. Electrical

9. Operating environment

Built-in illumination system Halogen

Stage height adjustment mechanism

Fine adjustment scale: 0.003µm per graduation

Fine adjustment stroke: 0.3mm per turn

Total stroke: 12.7mm

Co-axial coarse and fine focusing on ball drive

Quintuple positions fixed (Reverse angle)

Binocular

Field number

Tube tilting angle

22 (Standard)

30°

Trinocular Ergonomic

22 (Standard) 22 (standard)

30° 0°-25°

48-75 48-75 Interpupillary distance adjustment range

Size

Movement range

Specimen holder

Type

48-75

200 x 160mm (with mechanical stage)

76 x 50mm

Double specimen

Fliptop Abbe condenser (daylight filter detachable)

N. A.

Aperture iris diaphragm

1.25

Built-in

301.0mm (L) x 227.0mm (W) x 420.mm (H); 9 kg net

Halogen 6V-30W

Bulb life Upto 500 hours

Indoor use

Altitude: Max. 2000 meters

Ambient temperature: 5° to 40°C (41° to 104° F)

Maximum relative humidity: 80% for temperature up to 31°C (88°F), decreasing linearly through 70% at 34°C (93°F),to 50% relative humidity at 40°C

(104°F)

Supply voltage fluctuations: Not to exceed ±10% of the normal voltage.

Pollution degree: 2 (in accordance with IEC60664)

Installation/Overvoltage category: II (in accordance with IEC60664)

www.laboamerica.com

Our policy is one of continuous development. Labo America, Inc., reserves the right to change design and specifications without prior notice.

Labo America Inc.

920 Auburn Court

Fremont

CA 94538

U.S.A.

Telephone: 510 445 1257

Fax: 510 445 1317 [email protected]

LABOMED and Lx 500 are registered trademarks of Labo America, Inc.

With a policy of continuous development, Labo America, Inc. reserves the right to change design and specifications without prior notice.

© 2009 Labo America, Inc. | 9144000-990A 02-2009

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Related manuals

Download PDF

advertisement