TDSUSB2 Universal Serial Bus Measurement Package

TDSUSB2
Universal Serial Bus Measurement Package
077-0015-03
This document supports software version 3.0.3 and above.
www.tektronix.com
Copyright © Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its suppliers
and are protected by United States copyright laws and international treaty provisions.
Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication
supercedes that in all previously published material. Specifications and price change privileges reserved.
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
TDSUSB2 Compliance Test Software Online Help, 077-0015-03, Version 3.0.6.
Contacting Tektronix:
Tektronix, Inc.
14200 SW Karl Braun Drive or P.O. Box 500
Beaverton, OR 97077 USA
For product information, sales, service, and technical support:
•
In North America, call 1-800-833-9200.
•
Worldwide, visit www.tektronix.com to find contacts in your area.
Table of Contents
General Safety Summary................................................................xi
Introduction .................................................................................... 1
Welcome to the TDSUSB2 Universal Serial Bus Measurements Package .....1
Online Help and Related Documentation ........................................................2
Using Online Help ...........................................................................................2
Printing from the Online Help .........................................................................4
Conventions .....................................................................................................4
Related Documentation....................................................................................5
Feedback ..........................................................................................................6
General information ..................................................................................6
Application specific information...............................................................6
Getting Started ................................................................................ 7
Introduction and Product Description..............................................................7
Compatibility ...................................................................................................8
Requirements and Restrictions ........................................................................8
Accessories ......................................................................................................8
Optional Accessories.................................................................................8
Installation .................................................................................. 11
Setting up the Instrument to Take Measurements..........................................11
Recommended Instruments and Probes .........................................................11
Installing the Application...............................................................................11
Deskewing .....................................................................................................12
Deskewing the Probes and Channels.......................................................12
Deskewing Probes and Channels on the supported Instruments .............12
Operating Basics........................................................................... 19
About Basic Operations .................................................................................19
Application Interface .....................................................................................19
Application Interface Menu Controls ............................................................19
Basic Instrument Functions ...........................................................................20
Using Basic Oscilloscope Functions .......................................................20
Minimizing and Maximizing the Application .........................................20
TDSUSB2 Universal Serial Bus Measurement Package
i
Table of Contents
Application Directories and File Names .................................................21
File Name Extensions..............................................................................21
Returning to the Application ...................................................................21
Exiting the Application ...........................................................................22
Setting up the Software to take Measurements..............................................22
Description of the Test Fixture................................................................22
Setting Up the Device Details .................................................................23
Setting up the Software ...........................................................................24
Setting Preferences ..................................................................................24
Table of Measurements and Options .............................................................27
Low Speed and Full Speed Measurements..............................................27
High Speed Measurements......................................................................27
Selecting a Measurement ...............................................................................28
Signal Quality Check .....................................................................................29
Inrush Current Check.....................................................................................30
Droop Measurement ......................................................................................30
Receiver Sensitivity Measurement ................................................................31
Chirp Measurement........................................................................................33
Monotonic Property .......................................................................................34
Configuring a Measurement ..........................................................................34
Configure Limits............................................................................................35
Configuring Signal Quality Measurements ...................................................35
Configuring Inrush Current Measurements ...................................................36
Configuring Droop Measurements ................................................................36
Configuring Receiver Sensitivity Measurements ..........................................37
Configuring Chirp Measurement ...................................................................37
Configuring Reset from High Speed Measurement.......................................39
Configuring Resume Measurement ...............................................................40
Configuring Suspend Measurement...............................................................41
Configuring Reset from Suspend Measurement ............................................42
Configuring Packet Parameter Measurement ................................................43
Packet Parameter Measurement.....................................................................44
Suspend Measurement ...................................................................................45
ii
TDSUSB2 Universal Serial Bus Measurement Package
Table of Contents
Resume Measurement....................................................................................45
Reset from Suspend Measurement.................................................................46
Reset from High Speed Measurement ...........................................................48
Taking Measurements................................................................... 49
Acquiring Data...............................................................................................49
Control Menu Options ...................................................................................51
Viewing Results.............................................................................................51
Eye Diagram............................................................................................52
Waveform Plot ........................................................................................54
Report Generation..........................................................................................57
Generating Reports..................................................................................57
Viewing Reports......................................................................................58
TSV File Generator .................................................................................58
TSV file format description.....................................................................59
CSV file format description ....................................................................59
Saving and Recalling Setups..........................................................................60
How to Save and Recall a Setup .............................................................60
Saving a Setup.........................................................................................60
Recalling a Saved Setup..........................................................................60
Recalling the Default Setup.....................................................................61
Recently Saved Setup..............................................................................61
Recently Recalled Setup..........................................................................61
Tutorial
.................................................................................. 63
Introduction to the Tutorial............................................................................63
Starting the Application .................................................................................63
Recalling a .tsv File .......................................................................................63
Taking a Full Speed Signal Quality Measurement ........................................64
Taking a Low Speed Inrush Current Measurement .......................................67
Taking a High Speed Signal Quality Measurement.......................................69
Taking a Low Speed Signal Quality Measurement .......................................72
Application Examples.................................................................... 77
Full Speed Signal Quality Tests for Downstream Testing.............................77
TDSUSB2 Universal Serial Bus Measurement Package
iii
Table of Contents
Specifying the Equipment-Full Speed Signal Quality
Downstream Tests ...................................................................................77
Typical Equipment Setup-Full Speed Signal Quality
Downstream Tests ...................................................................................77
Selecting and Configuring Measurements-Full Speed Signal
Quality Tests ..........................................................................................78
Configuring the Measurement.................................................................79
Viewing Results-Full Speed Signal Quality Tests ..................................80
Generating Reports-Full Speed Signal Quality Tests..............................81
Droop Test for Ports of Hub ..........................................................................82
Specifying the Equipment-Low Speed Droop Tests ...............................82
Typical Equipment Setup-Low Speed Droop Tests ................................82
Selecting and Configuring Measurements-Droop Tests..........................84
Configuring the Measurement.................................................................84
Viewing Results-Droop Tests .................................................................84
Generating Reports-Droop Tests.............................................................85
Full Speed Signal Quality Tests for Upstream Testing..................................86
Specifying the Equipment-Full Speed Signal Quality Tests for
Upstream Testing ....................................................................................86
Typical Equipment Setup-Full Speed Signal Quality Tests for
Upstream Testing ....................................................................................86
Selecting and Configuring Measurements-Full Speed Signal
Quality Tests for Upstream Testing ........................................................87
Configuring the Measurement.................................................................88
Viewing Results-Full Speed Signal Quality Tests for
Upstream Testing ....................................................................................90
Generating Reports-Full Speed Signal Quality Tests for
Upstream Testing ...................................................................................91
Inrush Current Test for a Device ...................................................................91
Specifying the Equipment-Full Speed Inrush Current Test ....................91
Typical Equipment Setup-Inrush Current Test........................................91
Selecting and Configuring Measurements-Inrush Current Test..............93
Configuring the Measurement.................................................................93
iv
TDSUSB2 Universal Serial Bus Measurement Package
Table of Contents
Viewing Results-Inrush Current Test......................................................93
Generating Reports-Inrush Current Test .................................................94
Signal Quality Tests for High Speed Devices................................................95
Specifying the Equipment-Signal Quality Tests for High Speed
Devices for Upstream Testing.................................................................95
Typical Equipment Setup-Signal Quality Tests for High Speed
Devices for Upstream Testing.................................................................95
Selecting and Configuring Measurements-Signal Quality Tests for
High Speed Devices for Upstream Testing .............................................96
Configuring the Measurement.................................................................97
Viewing Results-Signal Quality Tests for High Speed Devices for
Upstream Testing ....................................................................................97
Generating Reports-Signal Quality Tests for High Speed Devices for
Upstream Testing ..................................................................................100
Packet Parameter Measurement...................................................................101
Specifying the Equipment-Packet Parameter Measurement .................101
Typical Equipment Setup-Packet Parameter Measurement ..................101
Set up the Oscilloscope for High Speed Host .......................................101
Selecting and Configuring Measurement-Packet Parameter
Measurement .........................................................................................111
Configuring the Measurement...............................................................112
Viewing Results-Packet Parameter Measurement.................................112
Generating Reports-Packet Parameter Measurement............................113
Resume Measurement..................................................................................115
Specifying the Equipment-Resume Measurement ................................115
Typical Equipment Setup-Resume Measurement .................................116
Selecting and Configuring Measurement-Resume................................117
Configuring the Measurement...............................................................117
Viewing Results-Resume Measurement ...............................................118
Generating Reports-Resume Measurement...........................................118
Reset from Suspend Measurement...............................................................121
Specifying the Equipment-Reset from Suspend Measurement .............121
Typical Equipment Setup-Reset from Suspend Measurement ..............121
TDSUSB2 Universal Serial Bus Measurement Package
v
Table of Contents
Selecting and Configuring Measurement-Reset from Suspend
Measurement ........................................................................................122
Configuring the Measurement...............................................................123
Viewing Results-Reset from Suspend Measurement ............................123
Generating Reports-Reset from Suspend Measurement........................124
Suspend Measurement .................................................................................126
Specifying the Equipment-Suspend Measurement................................126
Typical Equipment Setup-Suspend Measurement.................................127
Selecting and Configuring Measurement-Suspend Measurement ........128
Configuring the Measurement...............................................................128
Viewing Results-Suspend Measurement...............................................129
Generating Reports-Suspend Measurement ..........................................130
Reset from High Speed Measurement .........................................................132
Specifying the Equipment-Reset from High Speed Measurement........132
Typical Equipment Setup-Reset from High Speed Measurement.........132
Selecting and Configuring Measurement-Reset from High Speed
Measurement .........................................................................................133
Configuring the Measurement...............................................................134
Viewing Results-Reset from High Speed Measurement .......................134
Generating Reports-Reset from High Speed Measurement ..................136
Chirp Measurement......................................................................................138
Specifying the Equipment-Chirp...........................................................138
Typical Equipment Setup-Chirp............................................................138
Selecting and Configuring Measurement-Chirp....................................140
Configuring the Measurement...............................................................141
Viewing Results-Chirp..........................................................................141
Generating Reports-Chirp .....................................................................143
R-GPIB Commands .................................................................... 147
About the R-GPIB Program.........................................................................147
R-GPIB Reference Materials .......................................................................147
Introduction to the R-GPIB Commands ......................................................148
Guidelines to R-GPIB Programming ...........................................................148
Launching the Application Using R-GPIB ..................................................149
vi
TDSUSB2 Universal Serial Bus Measurement Package
Table of Contents
Variable: Value Command ..........................................................................150
Variable: Value TDSUSB2 Command Arguments and
Queries- Application....................................................................................150
Variable: Value TDSUSB2 Command Arguments and
Queries- Sequencer ......................................................................................150
Variable: Value TDSUSB2 Command Arguments and
Queries- Save/Recall....................................................................................151
Variable: Value TDSUSB2 Command Arguments and
Queries- Report Generation .........................................................................151
Variable: Value TDSUSB2 Command Arguments and Queries- Results ...152
Sample Program...........................................................................................156
Reference
................................................................................ 165
Shortcut Keys...............................................................................................165
USB2.0 Specifications .................................................................................166
Low Speed.............................................................................................166
Full Speed..............................................................................................167
High Speed ............................................................................................168
Inrush Current Check ............................................................................168
Droop Test.............................................................................................169
Error Codes and Warnings...........................................................................169
Settings for the supported Instruments ........................................................176
Upstream Setups for Low Speed signals...............................................176
Upstream Setups for Full Speed signals................................................176
Downstream Setups for Low Speed signals ..........................................177
Downstream Setups for Full Speed signals ...........................................177
Upstream Setup for High Speed signals................................................177
Inrush Setup...........................................................................................177
Droop Setup...........................................................................................178
Chirp Device EL_28, EL_29 EL_31 setup ...........................................178
Chirp Host EL_33, EL_34 setup ...........................................................178
Chirp Host EL_35 setup........................................................................178
Suspend Setup .......................................................................................179
Resume Setup........................................................................................179
Reset from High Speed .........................................................................179
TDSUSB2 Universal Serial Bus Measurement Package
vii
Table of Contents
Reset from Suspend...............................................................................179
Packet Parameter ...................................................................................180
Parameters ................................................................................ 181
About Application Parameters.....................................................................181
File Menu Parameters ..................................................................................181
Save and Recall Menu Parameters...............................................................181
Preferences Parameters ................................................................................181
Measurement Menus....................................................................................182
Measurement Parameters.............................................................................182
Configure Menu...........................................................................................183
Configure Signal Quality Measurements Parameters ..................................183
Configure...............................................................................................183
Source....................................................................................................184
Configure Inrush Current Measurement Parameters ...................................184
Configure...............................................................................................184
Source....................................................................................................184
Configure Droop Test Parameters ...............................................................185
Configure...............................................................................................185
Source....................................................................................................185
Configure Receiver Sensitivity ....................................................................185
Configure Suspend, Reset from High Speed, Resume, Reset
from Suspend ...............................................................................................185
Configure Packet Parameter ........................................................................186
Configure Chirp ...........................................................................................186
Results Menus..............................................................................................186
Utilities Menus.............................................................................................187
Deskew ........................................................................................................187
Report Generator Menu Parameters.............................................................187
TSV File Generator Menu Parameters.........................................................188
Help Menu ...................................................................................................188
Control Menu...............................................................................................188
Control Menu Parameters......................................................................188
viii
TDSUSB2 Universal Serial Bus Measurement Package
Table of Contents
Measurement Algorithms ............................................................ 189
About Measurement Algorithms .................................................................189
Cross-Over Voltage for Low Speed and Full Speed Signals .......................189
Cross-Over Voltage for High-Speed Signals...............................................191
Signal Rate...................................................................................................192
EOP Width Calculation ...............................................................................193
Consecutive and Paired Jitter.......................................................................194
Eye Diagram ................................................................................................195
Eye Violation ...............................................................................................195
Inrush Current ..............................................................................................196
Droop Test ...................................................................................................196
Receiver Sensitivity Test .............................................................................197
Chirp Test for a Device................................................................................197
Host Chirp Test............................................................................................198
Monotonic Property Test .............................................................................198
Rise Time.....................................................................................................199
Fall Time......................................................................................................199
Packet Parameter..........................................................................................200
Resume ........................................................................................................201
Suspend........................................................................................................201
Reset from Suspend .....................................................................................202
Reset from High Speed ................................................................................203
Equipment Setup for Tests...........................................................................204
Low Speed Downstream Signal Quality Host Equipment Setup ..........204
Low Speed Signal Quality for HUB Downstream ................................205
Full Speed Signal Quality HUB Downstream Setup.............................206
Full Speed Downstream Host Equipment Setup ...................................208
Full Speed Upstream Signal Quality Setup...........................................209
High Speed Device Signal Quality Setup..............................................211
Chirp Test Equipment Setup .................................................................212
Receiver Sensitivity Setup ....................................................................214
Procedural steps.....................................................................................215
Resume Test Equipment Setup..............................................................218
Reset from Suspend Test Equipment Setup ..........................................219
TDSUSB2 Universal Serial Bus Measurement Package
ix
Table of Contents
Suspend Test Equipment Setup.............................................................220
Reset from High Speed Test Equipment Setup .....................................221
Packet Parameter Test Equipment Setup .....................................................222
Test Fixture Setup .................................................................................222
Set up the Oscilloscope for High Speed Host .......................................223
Glossary
x
................................................................................ 233
TDSUSB2 Universal Serial Bus Measurement Package
General Safety Summary
Review the following safety precautions to avoid injury and prevent damage to
this product or any products connected to it. To avoid potential hazards, use this
product only as specified.
Only qualified personnel should perform service procedures.
While using this product, you may need to access other parts of the system. Read
the General Safety Summary in other system manuals for warnings and cautions
related to operating the system.
To Avoid Fire or Personal Injury:
Connect and Disconnect Properly: Do not connect or disconnect probes or test
leads while they are connected to a voltage source.
Observe All Terminal Ratings: To avoid fire or shock hazard, observe all
ratings and markings on the product. Consult the product manual for further
ratings information before making connections to the product.
Do Not Operate With Suspected Failures: If you suspect there is damage to
this product, have it inspected by qualified service personnel.
Symbols and Terms: The following terms and symbols may appear in the online
help.
WARNING: Warning statements identify conditions or practices that could
result in injury or loss of life.
:
CAUTION: Caution statements identify conditions or practices that could
result in damage to this product or other property.
Terms on the Product: The following terms may appear on the product:
DANGER indicates an injury hazard immediately accessible as you read the
marking.
WARNING indicates an injury hazard not immediately accessible as you
read the marking.
CAUTION indicates a hazard to property including the product.
TDSUSB2 Universal Serial Bus Measurement Package
xi
General Safety Summary
Symbols on the Product: The following symbol (s) may appear in the product:
CAUTION Refer to Help
xii
TDSUSB2 Universal Serial Bus Measurement Package
Introduction
Welcome to the TDSUSB2 Universal Serial Bus Measurements Package
TDSUSB2 Universal Serial Bus Measurements Package is a Sun Java-based
application that runs on the supported instruments connected to a PC. You can
use the application software with the compliance test fixture to take the following
measurements:
Signal Quality
Inrush
Droop
Drop (supported by the test fixture only)
Receiver Sensitivity
Impedance Measurement using Time Domain Reflectometry (TDR),
supported by the application only with the test fixture
Chirp measurements
Packet Parameter
Resume
Reset from Suspend
Reset from High Speed
Suspend
TDSUSB2 Universal Serial Bus Measurement Package
1
Introduction
The application runs on the oscilloscope and displays on the lower part of the
screen. The oscilloscope application runs on the upper part of the screen.
Online Help and Related Documentation
You can access the information on how to operate the application and the
oscilloscope through the following related documents and online help.
Click Start > Programs > TekApplications > TDSUSB2 > Help to access the
Online Help.
Click Start > Programs > TekApplications > TDSUSB2 > Help (PDF
Version) to access the PDF version of the Online Help.
Using Online Help
Online help has many advantages over a printed manual because of advanced
search capabilities. Select Help on the right side of the application menu bar to
bring up the help file.
The main (opening) Help screen shows a series of book icons and three tabs
along the top menu, each of which offers a unique mode of assistance:
Tables of Contents (TOC) tab -organizes the Help into book-like sections.
Select a book icon to open a section; select any of the topics listed under the
book.
Index tab - enables you to scroll a list of alphabetical keywords. Select the
topic of interest to bring up the appropriate help page.
Search tab - allows a text-based search. Follow these steps:
1.
Type the word or phrase you want to find in the search box
2.
Select some matching words in the next box to narrow your search
3.
Choose a topic in the lower box, and then select the Display button
To print a topic, select the Print button from the Help Topics menu bar.
Select Options from the menu bar for other commands, such as
Refresh.
Select the Back button to return to the previous help window. Use the
Hyperlink to jump from one topic to another. If the Back button is
grayed out or a jump is not available, choose the Help Topics button to
return to the originating help folder.
2
TDSUSB2 Universal Serial Bus Measurement Package
Introduction
Browse buttons (Next >> and Previous <<) allow you to move forward
and backward through topics in the order of the Table of Contents
(TOC).
Sometimes you will see the word, Note, in the topic text. This indicates
important information.
Certain aspects of the online help are unique to applications that run on
the oscilloscope.
Note: Green-underlined text indicates a jump (hyperlink) to another topic. Select
the green text to jump to the related topic. For example, select the green text to
jump to the topic on Online Help and Related Documentation and the Back
button to return to this topic page.
You can tell when the cursor is over an active hyperlink (button, jump, or
popup), because the cursor arrow changes to a small hand.
Note: The light bulb and the word Tip in the graphic
. This graphic
indicates additional information to help you function faster or more efficiently.
TDSUSB2 Universal Serial Bus Measurement Package
3
Introduction
Printing from the Online Help
Some online help topics have color in the examples of the displayed application.
If you want to print this type of topic on a monochrome printer, some
information may not print because of certain colors. Instead, you should print the
topic from the PDF (portable document format) file that corresponds to the
Online Help. You can find the file in the Documents directory on the Optional
Applications Software on Windows-Based Oscilloscopes CD–ROM. The figures
of application menus in the PDF file are gray scale and all of the information
prints.
Conventions
Online help uses the following conventions:
Refers to the software part of the TDSUSB2 Universal Serial Bus
Measurements Package as the TDSUSB2 application or as the application
When steps require a sequence of selections using the application interface,
the ">" delimiter marks each transition between a menu and an option. For
example, one of the steps to recall a setup file would appear as File> Recall
Unit under test (UUT) refers to the USB2.0 device under test, hub under
test, host under test, and port under test
SOF refers to the Start of Frames exchanged between the host controller and
the device when the device is connected to the host and enumerated by the
test mode software
4
TDSUSB2 Universal Serial Bus Measurement Package
Introduction
Related Documentation
In addition to the Online Help, you can access other information on how to
operate the oscilloscope through the following related documents:
Instrument Information: The user manual for your oscilloscope provides
general information on how to operate the oscilloscope
Programmer Information: The online help for your oscilloscope provides
details on how to use GPIB commands to control the oscilloscope
You can also download the following files, which contain programmer
information and examples, from the Tektronix Web site.
Optional Applications Software on Windows-Based Oscilloscopes
Installation Manual
TDSUSB2 Universal Serial Bus Measurements Package Reference
TDSUSBF USB2.0 Compliance Test Fixture Instructions
For more information on USB2.0 specifications, visit www.usb.org.
Refer to the Optional Applications Software on Windows-Based Oscilloscopes
Installation Manual for the following information:
Software warranty
List of all available applications, compatible instruments, and relevant
software and firmware version numbers
Applying a new label
Installing an application
Enabling an application
Downloading updates from the Tektronix Web site
You can find a PDF (portable document format) file of this document in the
Documents directory on the Optional Applications Software on Windows-Based
Oscilloscopes CD –ROM. The CD booklet only contains information on
installing the application from the CD and on how to apply a new label.
TDSUSB2 Universal Serial Bus Measurement Package
5
Introduction
Feedback
Tektronix values your feedback on our products. To help us serve you better,
please send us your suggestions, ideas, or comments about your oscilloscope.
Direct your feedback via email to techsupport@tektronix.com or FAX at (503)
627-5695 and include the following information. Please be as specific as
possible.
General information
Instrument model number and hardware options, if any
Probes used
Your name, company, mailing address, phone number, FAX number
Please indicate if you would like to be contacted by Tektronix about your
suggestion or comments
Application specific information
Software version number
Description of the problem such that technical support can duplicate the
problem
If possible, save the oscilloscope and application setup files as .set files
If possible, save the waveform on which you are performing the
measurement as a .wfm file
Once you have gathered this information, you can contact technical support by
phone or through e-mail. If using e-mail, be sure to enter in the subject line
"TDSUSB2 Problem," and attach the .set and .wfm files.
To include screen shots from the oscilloscope menu bar, select File> Export
Setup> Image tab or File > Save > Screen Capture. Save the screen shot in the
default directory or you can choose a directory of your choice. If you want the
screen shots in color, select Edit> Copy setup> Images tab or select File > Save >
Screen Capture and select Options to get the Screen Capture Save Option dialog
box. Select the Color, Full Screen and the Normal option buttons. Select Copy.
Copy the picture to any Paint editor software.
You can then attach the file to your email (depending on the capabilities of your
email editor).
6
TDSUSB2 Universal Serial Bus Measurement Package
Getting Started
Introduction and Product Description
The TDSUSB2 Universal Serial Bus Measurements is a package that runs on the
supported instruments. The solution consists of a Java-based application and a
comprehensive test fixture. You can use the application software with the
compliance test fixture to take the following measurements:
Signal Quality
Inrush
Droop
Drop (supported by the test fixture only)
Receiver Sensitivity
Impedance Measurement using Time Domain Reflectometry (TDR),
supported by the application only with the test fixture
Chirp measurements
Packet Parameter
Resume
Reset from Suspend
Reset from High Speed
Suspend
The application performs tests that measure the test signals for USB2.0
compliance, displays eye diagrams and plots, displays the results as a summary
or as details, and generates reports in different formats.
TDSUSB2 Universal Serial Bus Measurement Package
7
Getting Started
Compatibility
For information on oscilloscope compatibility, refer to the *Optional
Applications Software on Windows-Based Oscilloscopes Installation Manual,
Tektronix part number 071-1888-XX. The manual is also available as a PDF file.
The dynamic range of the probes used for Low Speed and Full Speed testing
should be at least ± 8 volts. For High Speed testing, the dynamic range should be
± 2 volts.
*For a current list of compatible instruments, see the Software and Drivers
category in the Tektronix Web site
Requirements and Restrictions
You must install Java Run-Time Environment V1.4.2_02 on the supported
oscilloscope.
The application uses the Math1 channel for Low Speed and Full Speed mode of
operation. You will lose any information that you have stored in the Math1
channel.
You cannot restore the oscilloscope settings if you select File> Exit from the
application.
Accessories
The application includes the following standard accessories:
Optional Application Software on Windows-Based Instruments CD–ROM
Optional Application Software on Windows-Based Instruments Installation
Manual
TDSUSB2 Universal Serial Bus Measurements Package Reference
TDSUSBF USB2.0 Compliance Test Fixture Instructions
TDSUSBF USB2.0 Compliance Test Fixture
Optional Accessories
DG2040 Data Generator (for receiver sensitivity)
AWG610 Arbitrary Waveform Generator
P6248 Differential probes
8
TDSUSB2 Universal Serial Bus Measurement Package
Getting Started
P6245, P6243, TAP1500 Single-ended active probes
TCP202/TCP0030 Current probes
TDS8000 Instrument with 80E04 Time Domain Reflectometry sampling
module
Deskew Fixture: for supported instrument-Probe Calibration and Deskew
Fixture, Tektronix part number (067-0405-xx)
TDSUSB2 Universal Serial Bus Measurement Package
9
Getting Started
10
TDSUSB2 Universal Serial Bus Measurement Package
Installation
Setting up the Instrument to Take Measurements
To set up the oscilloscope, follow these steps:
1.
You must power on the oscilloscope for twenty minutes before you can start
to take measurements.
2.
You must run the compensation signal path on the oscilloscope.
3.
You must make sure that the default factory setup is recalled before you
start using the application in the oscilloscope. To do so, push the recall
default setup button on the front-panel of the oscilloscope to recall the
default factory settings.
4.
You should always use calibrated probes and degauss the current probes.
Recommended Instruments and Probes
For information on recommended instruments and probes, refer to the Optional
Application Software on Windows-Based Oscilloscopes Installation Manual.
Installing the Application
Refer to the Optional Applications Software on Windows-Based Instruments
Installation Manual for the following information:
Installing an application
Applying a new label
Enabling an application
Downloading updates from the Tektronix Web site
You can find a PDF (portable document format) file of this document in the
Documents directory on the Optional Applications Software on Windows-Based
Instruments CD–ROM. The CD booklet only contains information on installing
the application from the CD and on how to apply a new label.
TDSUSB2 Universal Serial Bus Measurement Package
11
Installation
Deskewing
Deskewing the Probes and Channels
To ensure accurate measurements, you must deskew the probes before you take
measurements from your unit under test. The deskew process is where the
oscilloscope adjusts the relative delay between the signals to accurately time
correlate the displayed waveforms.
The application includes an automated deskew utility that you can use to deskew
any pair of oscilloscope channels.
Note 1: It is recommended that you use the deskew fixture specified in the
accessories section to perform deskew. The deskew source can be the built-in
probe compensation signal in the oscilloscope or an external signal source.
Note 2: The oscilloscope has a deskew range of 50 ns.
Deskewing Probes and Channels on the supported Instruments
To deskew probes and channels on the supported instruments, follow these steps:
12
1.
Connect the probes to Ch1 and Ch2 on the oscilloscope.
2.
Connect the probe compensation signal to the deskew fixture. You can use
the probe compensation signal from the oscilloscope as the source for the
deskew fixture.
3.
Follow the on-screen prompts for the deskew operation with an external
source.
TDSUSB2 Universal Serial Bus Measurement Package
Installation
The following figure shows signals before performing the deskew procedure with
a single edge:
Figure 1: Signals before deskew
4.
To start the application from a supported oscilloscope, select File> Run
Application> USB2.0 Test Package or Analyze>USB2.0 Test Package from
the menu bar of the oscilloscope.
5.
Select Utilities> Deskew.
6.
Set the channel Source in the From area to Ch1. The Source waveform is the
reference point to which the remaining channels are deskewed.
7.
Set the channel Source in the To area to Ch2, to deskew the channel.
8.
Select the Reference Level for Ch1 and set the reference value. The
reference level is the percentage level of the waveform from which to take
the edges to deskew.
9.
Select the Reference Level in the To area for Ch2 and set the reference
value.
10. Select the Hysteresis in the From area for Ch1. Hysteresis helps to ignore
the noise level in the waveform.
TDSUSB2 Universal Serial Bus Measurement Package
13
Installation
11. Select the Hysteresis in the To area for Ch2 and set the hysteresis value.
12. Select the Slope, Rise or Fall, on which to perform the deskew operation.
13. Select the number of edges used for deskew.
14. To start the deskew utility, select Utilities> Perform Deskew and confirm
the operation.
15. Without changing the From: Ch1 channel, deskew the remaining channels.
The following figure shows the results after performing the deskew operation
with a single edge. The reference level is set to 50% for Ch1 and Ch2, the
hysteresis is set to 5%, the number of edges is set to 1, and the slope is set to the
rising edge.
Note: The probe compensation signals vary from 0.8 to 1 volt (for all supported
TDS instruments) and from 0.35 to 1 volt (for all supported DPO instruments)
rather than swinging from 0 to 1 volt in a traditional oscilloscope. Set the
Reference level and the Hysteresis level appropriately so that the Ref level is at
the middle of the swing on the signal.
Figure 2: Signals after deskew
14
TDSUSB2 Universal Serial Bus Measurement Package
Installation
To perform the deskew operation with an external source, follow these steps:
1.
Connect the probes to Ch1 and Ch2 on the oscilloscope.
2.
Connect an external source to the deskew fixture.
3.
Follow the instructions of the Probe Calibration and Deskew fixture
(Tektronix part number: 067-0405-xx) to make the connections.
4.
Set up the oscilloscope as follows:
Use the Horizontal Scale knob to set the oscilloscope to an acquisition
rate so that there are two or more samples on the deskew edge.
Use the Vertical Scale and Position knobs to adjust the signals to fill
the display without missing any part of the signals.
Set the Record Length so that there are more samples for the edges in
the acquisition. It is recommended that you set the record length to
25000 points.
5.
To start the application from a supported oscilloscope, select File> Run
Application> USB2.0 Test Package or Analyze> USB2.0 Test Package from
the menu bar of the oscilloscope.
6.
Select Utilities> Deskew.
7.
Set the channel Source in the From area to Ch1. The remaining channels are
deskewed to the Source waveform, which is the reference point.
8.
Set the channel Source in the To area to Ch2, the channel to be deskewed.
9.
Select the Reference Level for Ch1. The reference level is the percentage
level of the waveform from which to take the edges to deskew.
10. Select the Reference Level in the To area for Ch2 and set the reference
value.
11. Select the Hysteresis in the From area for Ch1. Hysteresis helps to ignore
the noise level in the waveform.
12. Select the Hysteresis in the To area for Ch2 and set the hysteresis value.
13. Select the Slope, Rise or Fall, on which to perform the deskew operation.
14. Select the number of edges used for deskew.
15. To start the deskew utility, select Utilities> Perform Deskew and confirm
the operation.
16. Without changing the From: Ch1 channel, deskew the remaining channels.
17. The setup is an acquisition of a square signal at 100 KHz, with the Record
length set to 25000 points to achieve the sample resolution of 1.6 ns.
TDSUSB2 Universal Serial Bus Measurement Package
15
Installation
The following figure shows an example of a deskew setup:
Figure 3: Example of deskew setup
18. Set the Reference levels for Ch1 and Ch2 in the application to 50%, the
hysteresis to 10%, the number of edges to 4, and the type to rising edge. The
following figure shows the display after performing the deskew for the
multiple edge:
16
TDSUSB2 Universal Serial Bus Measurement Package
Installation
Figure 4: Signals after deskew for multiple edges
19. The zoomed section of the waveform available on the lower part of the
screen shows the results of the deskew operation clearly. You will see that
after performing the deskew operation the skew is automatically removed.
TDSUSB2 Universal Serial Bus Measurement Package
17
Installation
18
TDSUSB2 Universal Serial Bus Measurement Package
Operating Basics
About Basic Operations
This section contains information on the following topics and tasks:
Application interface
Using basic oscilloscope functions
Setting up the software
Selecting a measurement
How to Save and Recall setups
Exiting the application
Application Interface
The application uses a Windows interface. You should refer to your oscilloscope
user manual for the operating details of other controls, such as the front-panel
buttons.
Application Interface Menu Controls
Table 1. Application menu description
Item
Menu bar
Tab
Area
Option button
List box
Box
Check Boxes
Scroll bar
Browse
Command button
Keypad
MP/GPknob
Description
Located at the top of the application display and contains application menus
Labeled group of options containing similar items
Visual frame that encloses a set of related options
Button that defines a particular command or task
Box that contains a list of items from which you can select one item
Box that you can use to type in text, or to enter a value with the Keypad or a Multipurpose knob
Square box that you can use to select or clear preferences
Vertical or horizontal bar at the side or bottom of a display area that can be used for moving around in that area
Displays a window where you can look through a list of directories and files
Button that initiates an immediate action
Keypad appears when you select the box; select and use it to enter a value
A line that appears between the knob and the box when the MP or GP knob is selected; turn the knob and select a
value
TDSUSB2 Universal Serial Bus Measurement Package
19
Operating Basics
Basic Instrument Functions
Using Basic Oscilloscope Functions
You can use the Help menu to access information about the oscilloscope. You
can also use other oscilloscope functions and easily return to the application.
The TDSUSB2 application includes Online Help about the application menus
and controls.
To display the Online Help, follow these steps:
1.
Choose Help from the TDSUSB2 menu bar.
2.
Use the Contents, Table of contents, or Index tabs to navigate through the
help.
3.
The touch screen mode of operation is enabled by default. You can disable
this by using the Touch Screen OFF button on the oscilloscope front panel.
Minimizing and Maximizing the Application
To minimize the application, select File> Minimize. When you minimize the
oscilloscope, the application is continually displayed.
To maximize the application, select the TDSUSB2 icon in the Windows taskbar.
20
TDSUSB2 Universal Serial Bus Measurement Package
Operating Basics
Application Directories and File Names
The application uses the directories for several functions, such as save and recall
setup files, and uses the extensions appended to the file names to identify the file
type.
The following table lists default directory names for the supported instruments:
Table 2. Application Directories and File names
Directory
C:\TekApplications\tdsusb2
C:\TekApplications\tdsusb2\report
C:\TekApplications\tdsusb2\setup
C:\TekApplications\tdsusb2\tsvfilegenerator
C:\TekApplications\tdsusb2\datagen
C:\TekApplications\tdsusb2\images
C:\TekApplications\tdsusb2\temp
Used for
Home location
Report files
Setup files
Default directory for the tsv file generated by the file generator and for other csv
files.
Digital signal generator pattern files for use in Receiver Sensitivity Test
Images of the eye diagram and waveform plots
Temporary files used in the application
File Name Extensions
Table 3. Description of File names
Extensions
.csv
.ini
.set
.tsv
Description
Input file is in the .csv (comma separated variable) format and is used to generate a .tsv (tab separated variable) file.
An exported waveform that may be used as a source to generate a tsv format file
Application setup file
Instrument setup file with the same name, saved and recalled with .ini and .set file name extension
Input file in a Tab Separated Variable format
Returning to the Application
When you access oscilloscope functions, the oscilloscope fills the display. You
can access oscilloscope functions in the following ways:
1.
Select the Hide button in the application display.
2.
Select the Menu bar or the Toolbar mode on the oscilloscope and access the
menus.
3.
To return to the application, click the App button on the menu bar of the
oscilloscope or click on Analyze > Restore Application to restore the
application.
TDSUSB2 Universal Serial Bus Measurement Package
21
Operating Basics
Exiting the Application
To exit the application, select the Exit button or File> Exit or select the Exit
button from the control panel.
When you exit the application, you can keep the oscilloscope setup currently in
use or restore the oscilloscope setup that was present before you started the
application.
Figure 5: Exiting the application
Setting up the Software to take Measurements
Description of the Test Fixture
The TDSUSBF USB2.0 Compliance Test Fixture is a break-out board that
enables you to test live USB2.0 signals, exchanged between the device and the
host. For more information on the compliance test fixture, refer to the TDSUSBF
USB2.0 Compliance Test Fixture Instructions shipped with the test fixture.
The compliance test fixture consists of the following sections:
SQ Test Section
Receiver Sensitivity Test
Impedance Measurement Test (TDR)
Disconnect Detect Test
Adjacent Trigger and Droop Test
Host SQ
Inrush Test
Downstream Signal Quality
22
TDSUSB2 Universal Serial Bus Measurement Package
Operating Basics
Figure 6: Test fixture
Setting Up the Device Details
You can enter a unique identifier (ID) and description for the unit under test
(device). The identifier and the description appear in the generated reports for the
tests performed on the unit under test. You can either type in the text directly in
the boxes for these fields, or use the keypad or a keyboard. You can also choose
to generate automatic IDs for the unit or enter a prefix using the keypad.
Figure 7: Setting up the device details
Note: If you check the Generate Automatic ID, the Device ID field is disabled
and the Prefix field is enabled. The report for Inrush Measurement displays
details of the Signal Setup with the device description.
TDSUSB2 Universal Serial Bus Measurement Package
23
Operating Basics
Setting up the Software
You can set up the application to take one or more measurements at the same
time. In addition, you can view the results as a summary, details, an eye diagram,
and a waveform plot for Signal Quality checks. You can also generate the reports
in one of the three formats: Tektronix Specific, Plug-Fest and CSV formats.
The application runs on the oscilloscope. It is recommended that you connect the
keyboard and the mouse to use the application effectively.
Setting Preferences
The Preferences menu consists of two tabs: General and Advanced. To access the
Preferences menu, select File> Preferences.
Figure 8: Setting general preferences
24
TDSUSB2 Universal Serial Bus Measurement Package
Operating Basics
Figure 9: Setting advanced preferences
In the General tab, you can select the following options:
Set a warning to indicate that the configuration options were not changed
since the last time you ran the application. If you set this option, the
application prompts you to configure the measurements before acquiring
data
Automatically display the Eye Diagram without having to select the Eye
Diagram button in the Results screen
Automatically display the Waveform Plot without having to select the
Waveform Plot in the Results screen
Prompt for a File name or Device ID before automatically generating reports
or data files
Automatically generate .tsv files
Display the generated report
Always display the waiver mask for High-Speed Near End and Low Speed
eye diagrams
Do Autoset Every Time.
Note: Autoset will happen only for Full-Speed/High Speed SQC measurements.
TDSUSB2 Universal Serial Bus Measurement Package
25
Operating Basics
In the Advanced tab, you can select the following options:
Test a Low or Full Speed device with USB2.0 silicon
Prompt during report generation for the Droop test readout. When you
enable this option, the application disables report generation in the
Automatic mode
Configure the test limits. The result: PASS or FAIL is determined by the
limits you have defined
Place the vertical cursors between the start and end of the single USB2.0
packet. This is used when more than one USB2.0 packet is displayed on the
oscilloscope for Signal Quality measurements
Set the trigger holdoff value to capture the USB2.0 high-speed test packet
Note: If you enable the Configure Test Limits options, the Results Summary and
Report Generation in Plug-Fest format is disabled.
26
TDSUSB2 Universal Serial Bus Measurement Package
Operating Basics
Table of Measurements and Options
Low Speed and Full Speed Measurements
Table 4. Low and Full Speed measurements
Area
Signal Quality Check
Option
Eye Diagram
Signal Quality Check
Signal Rate
Signal Quality Check
Signal Quality Check
Signal Quality Check
Paired JK Jitter
Paired KJ Jitter
Consecutive Jitter
Signal Quality Check
Signal Quality Check
Signal Quality Check
EOP Width
Cross-Over voltage
Rise Time
Signal Quality Check
Fall Time
Inrush Current Check
Inrush Current
Droop Test
Droop Test
Description
Checks whether the USB signal is aligned with its corresponding eye
diagram
Measures the inverse of the average bit-time that gives the transmission
rate of the USB signal
Measures the jitter time for paired (JK next to KJ) differential data transition
Measures the jitter time for paired (KJ next to JK) differential data transition
Measures the jitter at every consecutive data bit calculated using the signal
rate
Measures the width of the end-of-packet of a USB signal
Measures the voltage at which the D+ voltage crosses the D– voltage
Measures the elapsed time from when a rising edge crosses the low
reference voltage level and then the high reference voltage level
Measures the elapsed time from when a falling edge crosses the high
reference voltage level and then the low reference voltage level
Measures the amount of electrical charge drawn by a device as soon as it
is connected to a USB network
Measures the difference in the VBUS voltage when the load switch is open
to the lowest value of the voltage and the load switch
High Speed Measurements
Table 5. Low and Full Speed measurements
Area
Signal Quality Check
Option
Eye Diagram
Signal Quality Check
Signal Rate
Signal Quality Check
Monotonic Property
Signal Quality Check
Signal Quality Check
EOP Width
Rise Time
Signal Quality Check
Fall Time
Inrush Current Check
Inrush Current
Droop Test
Droop Test
Description
Checks whether the USB signal is aligned with its corresponding eye
diagram
Measures the inverse of the average bit-time that gives the transmission
rate of the Test_Packet
Detects when the Signal is Monotonic if and only if data [i+1] ≥ data [i] in
case of consistently increasing (rising slope), never decreasing data [i+1] ≤
data [i] when consistently decreasing (falling slope) and never increasing in
value where i ranges from 0 to n
Measures the width of the end-of-packet pattern of a Test_Packet
Measures the elapsed time from when a rising edge crosses the low
reference voltage level and then the high reference voltage level
Measures the elapsed time from when a falling edge crosses the high
reference voltage level and then the low reference voltage level
Measures the amount of electrical charge drawn by a device as soon as it
is connected to a USB network
Measures the difference in the VBUS voltage when the load switch is open
to the lowest value of the voltage and the load switch
TDSUSB2 Universal Serial Bus Measurement Package
27
Operating Basics
Table 5. Low and Full Speed measurements (cont.)
Area
Option
Description
Receiver Sensitivity Test Receiver Sensitivity
Chirp Test
Chirp
Suspend Test
Suspend
Resume Test
Resume
Rest from High-Speed
Test
Reset
Reset from Suspend
Reset from Suspend
Packet Parameter
Packet Parameter
Is an indicator of Receiver Sensitivity. A High Speed capable device must
indicate 'packet (Data) not received' (squelch) when the input of the
receiver falls below 100 mV differential amplitude. Similarly the device must
not indicate squelch if the differential amplitude is greater than 150 mV
Checks a part of handshake that occurs during Reset Protocol for highspeed capable hubs and devices. In this handshake, the hub/host and the
device are required to detect chirp J and K of a specified minimum duration
and amplitude
Measures the time between the end of the last Start of Frame (SOF) and
the rising edge transition to the Full Speed J state. The acceptable range
should be between 3 ms to 3.125 ms. This is applicable if you have
selected a Host, Device, or a Hub-Upstream device
Resumes the High-speed operation in a device or a hub, which is indicated
by the presence of High-speed SOF packets (with 400 mV nominal
amplitude) following the K state driven by the host controller. For the Host
this is the time between the falling edge of D+ to the First SOF. This should
not exceed 3.0 ms
Measures the time between the beginning of the last SOF and before the
reset and the beginning of Chirp-K. This is between 3.1 ms and 6 ms. This
test will be applicable for Device and HUB upstream
Measures the time between the falling edge of D+ signal and the start of
Device chirp-K. This is between 2.5 us and 3 ms. This test is applicable for
Device and HUB upstream
Measures the parameter of the packet such as Sync Bits, EOP, and InterPacket Gap depending on the selected measurement
Selecting a Measurement
To take a measurement, select Measurement from the Measurements menu,
which is also the default opening screen in the application. To access the Select
option in the Measurement menu, select Measurements> Select.
There are three categories of measurements: Low Speed, Full Speed, and High
Speed measurements. The measurements for Low, Full and High Speed signals
are Signal Quality Checks, Inrush Current Check, and Droop Test. The additional
measurements for High Speed tests are Receiver Sensitivity, Monotonic
Property, and Chirp Test, Packet Parameter, Suspend, Resume, Reset from
Suspend, and Reset from High Speed.
You need to select the measurements for a particular signal speed of the unit
under test. After selecting the tests, you must configure the application based on
Signal Source, Tier, Test Point, and Signal Direction.
You can test the units for the following:
Devices for Upstream Signal Quality Check
28
TDSUSB2 Universal Serial Bus Measurement Package
Operating Basics
Hubs and Hosts for Downstream Signal Quality Check
Ports of a Hub for Droop Test
Devices for Inrush Current
Signal Quality Check
The application performs Signal Quality tests that include the following tests:
Table 6. Signal Quality checks
Area
Signal Quality Check
Signal Quality Check
Signal Quality Check
Signal Quality Check
Signal Quality Check
Signal Quality Check
Signal Quality Check
Signal Quality Check
Signal Quality Check
Signal Quality Check
Option
Eye Diagram
Description
Checks whether the USB signal is aligned with its corresponding
eye diagram
Signal Rate
Measures the inverse of the average bit-time that gives the
transmission rate of the USB signal
Paired JK Jitter
Measures the jitter time for paired (JK next to KJ) differential
data transition
Paired KJ Jitter
Measures the jitter time for paired (KJ next to JK) differential
data transition
Consecutive Jitter
Measures the jitter at every consecutive data bit calculated
using the signal rate
EOP Width
Measures the width of the end-of-packet of a USB signal
Cross-Over voltage
Measures the voltage at which the D+ voltage crosses the D–
voltage
Rise Time
Measures the elapsed time from when a rising edge crosses the
low reference voltage level and then the high reference voltage
level
Fall Time
Measures the elapsed time from when a falling edge crosses the
high reference voltage level and then the low reference voltage
level
Monotonic Property (For High Speed only) Measures the Monotonic Signal if the data [i] < data [i-1] where
data[l] is the record point in the acquired waveform. This is in the
case of consistently increasing (rising Slope) only. If the data [i]
> data [i-1] in case of consistently decreasing) falling Slope) and
never increasing in value where i ranges from 0 to n, the signal I
is said to be monotonic
Note: The USB2.0 specifications recommend that you should test the signal
quality for upstream and downstream traffic. In upstream testing, the application
captures the signals transmitted from the device to the host. In downstream
testing, the application captures the signals transmitted from the host to the
device.
Downstream testing is performed on ports of a hub. When testing a hub, you need
to connect the USB2.0 unit to Tier 6 to ensure the worst case. Each hub level is
referred to as a Tier. The hub under test is connected to the Tier 5, so that you
can test the hub on the Tier 6.
TDSUSB2 Universal Serial Bus Measurement Package
29
Operating Basics
Inrush Current Check
The application can perform an Inrush current check to verify that the unit under
test does not draw current higher than that specified in the USB2.0 specifications
when connected to a USB2.0 system. If the measured current drawn is higher
than a specified value, the other USB2.0 devices connected to the bus may not be
able to function properly.
When a unit is connected, there is a sharp intake of current followed by a
comparatively less steep decay. Small humps or perturbations are noticed in the
current trace, depending on when the unit resets.
The TDSUSB2 application automatically sets up the oscilloscope. The
application gives a direct readout of Charge (µC) and Capacitance (µF) values.
The application displays the details of the results after comparing the test results
with the USB2.0 specifications.
Figure 10: Results of the Inrush current check
Droop Measurement
The Droop voltage is the difference in the VBUS voltage when you apply a no load
condition and a 100 mA load to the port under test (all other ports are fully
loaded).
The Droop test evaluates the worst case droop by alternately applying a droop
load and no load to the port under test while all other ports are supplying the
maximum load possible. All the VBUS measurements are relative to local ground.
The TDSUSB2 application automatically sets up the oscilloscope for the
specified test configuration. When you start the application, it acquires the signal,
and provides the VDROOP measurement, and displays PASS or FAIL.
30
TDSUSB2 Universal Serial Bus Measurement Package
Operating Basics
Figure 11: Results of the Droop measurement
Note: The TDSUSB2 application helps to report the Drop test. You can do this by
enabling an option in the File> Preferences> Advanced menu. You can enter the
multimeter reading for the Drop test in the TDSUSB2 application during report
generation for a consolidated report.
Receiver Sensitivity Measurement
To improve the performance of the application in a noisy environment, the
USB2.0 high-speed device should respond to IN tokens with NAKs when the
signaling level is at or above the specified level.
The Receiver Sensitivity test requires a high-speed data simulator, such as the
Tektronix DG2040 USB Data Generator or a DTG5000 series instrument, to
transmit IN tokens of varying amplitude. The test requires the unit under test to
be placed in the Test_SE0_NAK mode. The host is then replaced by the data
simulator to continue to transmit IN tokens. The signaling amplitude is presented
to the device under test at a level at or above 150 mV. At these levels, the unit
under test must not be in the squelched mode, responding to IN packets with
NAKs. The amplitude of the signals from data simulator varies less than 100 mV.
The unit under test must be squelched and not respond to IN tokens with NAKs.
This tests the receiver capability of high-speed units (device) to respond to the
particular data pattern generated by the USB2.0 data simulator. The unit under
test responds to the data pattern level above the squelch level (>150 mV) and
should not respond when the data pattern level is below the squelch level
(<100 mV).
The TDSUSB2 application provides the procedural steps to perform this
measurement. It also provides Digital signal generator pattern files. Pattern files
for other Tektronix data simulators are available from www.tektronix.com.
TDSUSB2 Universal Serial Bus Measurement Package
31
Operating Basics
Figure 12: Results of the Receiver Sensitivity measurement
32
TDSUSB2 Universal Serial Bus Measurement Package
Operating Basics
Chirp Measurement
To perform a Chirp test, connect the unit under test and the single-ended probes
to acquire data. You can measure the Data for Chirp K amplitude, Chirp K
duration, and Reset duration. You need to manually verify that there are three K–
J pairs in less than 500 µs.
Figure 13: Results of the Chirp measurement
The TDSUSB2 application automates this process and automatically generates
the results and reports for the results of Chirp-K duration, Chirp-K amplitude,
and Reset duration.
Figure 14: Summary results of the Chirp measurement
TDSUSB2 Universal Serial Bus Measurement Package
33
Operating Basics
Monotonic Property
While performing a USB2.0 High Speed compliance test, you need to verify that
the signal to be tested is monotonic.
The following graph shows the monotonic behavior of a USB2.0 high-speed
signal with a rise time of 500 ps.
To verify the monotonic behavior of a signal, the oscilloscope must have a
sample rate high enough to capture as many sample points as possible on a rising
or falling edge. In addition, the oscilloscope should have enough bandwidth to
ensure that the high frequency non-monotonic transition is not attenuated.
The application coupled with a high performance Tektronix oscilloscope
automates the process and ensures repeatability of test results.
Configuring a Measurement
To access the Configure: Measurement, go to Measurements> Configure. The
application also provides a Configure option with each measurement area to
allow you to configure the selected measurements.
Note: If you select the Run button before configuring a measurement, the
application displays the message ‘The selected measurements have not been
configured. Do you want to continue? In this case, the application runs with the
default settings. If you select the Run button before configuring the measurement
and have enabled the File> Preferences> General option, the application
displays the warning message ‘Warn if configuration was not changed since last
run, when the run button is pressed.
34
TDSUSB2 Universal Serial Bus Measurement Package
Operating Basics
Configure Limits
The application displays the maximum and minimum values for the selected
tests. You can use the ‘>’ sign on the keypad to configure limits for these
options.
Table 7. Configure limits description
Option
Set
Default
Cancel
Description
Sets the values you enter
Restore the default values
Cancels all the changes you enter
Note: The application enables the Configure Limits values when you select the
option File> Preferences> Advanced menu.
Configuring Signal Quality Measurements
To access the Measurement: Configure, go to Measurements> Configure. Be sure
to select the relevant measurements before you configure them. There are two
tabs for the Signal Quality Measurements: Configure and Source.
The Configure tab allows you to select and set the Tier, Signal Direction and the
Test Point options. You must select the Tier (Tier 1 through 6), the direction of
signal (Upstream or Downstream) and the Test Point (Near End or Far End) at
which the unit will be tested.
For Low Speed and High Speed signals, you can set Test Point to Near End or
Far End. For Full Speed signals, you can set the Test Point to Far End only.
For Rise Time, Fall Time, and Monotonic Property, you can configure the
measurement levels.
The Source tab allows you to select the Source of the signal: a live signal or the
signal from a file. For a live signal there are two options: Differential and SingleEnded. For Low Speed and Full Speed devices, you can test only single-ended
signals (D+ and D–). For High Speed devices, it is recommended that you use a
differential probe.
TDSUSB2 Universal Serial Bus Measurement Package
35
Operating Basics
Configuring Inrush Current Measurements
To access the Measurements: Configure menu, go to the Measurements>
Configure.
The Configure tab allows you to set the voltage value on VBUS and the unit under
test. The VBUS can be entered manually or probed from a channel that is captured
from the test fixture. Select this option from the Source tab of Inrush
Configuration menu.
The Tier 1 is always used for testing. The unit under test can be any one of the
following types:
Hot Plug Attach
Low Powered Configure
Low Powered Resume
High Powered Configure
High Powered Unconfigure
High Powered Resume
The Hot Plug Attach is the most common unit under test. The Source tab allows
you to select the source of the signal: a live signal or the signal taken from a tsv
file. If you choose to manually enter the VBUS voltage, you cannot select the VBUS
voltage source.
Configuring Droop Measurements
Select the Droop measurement and configure it. To access the Configure:
Measurements menu, go to Measurements> Configure.
The Configure tab allows you to select the Port (Port 1 through 7), Hub and
Source of the signal. The source of the signal can be live or from a .tsv file. The
hubs can either be Self-powered or Bus-powered.
36
TDSUSB2 Universal Serial Bus Measurement Package
Operating Basics
Configuring Receiver Sensitivity Measurements
To configure receiver sensitivity measurements, follow these steps:
1.
Select the High Speed measurement tab> More button> Receiver Sensitivity
measurement.
2.
To access the Configure Measurements menu, select Measurements>
Configure.
Figure 15: Configuring Receiver Sensitivity measurement
3.
Select the Source from Ch1 to Ch4.
4.
Select the Run button.
5.
Select OK and follow the on-screen prompts to perform the tests.
6.
To generate reports, select Utilities> Report Generator.
Configuring Chirp Measurement
To configure chirp measurements, follow these steps:
1.
Select the High Speed measurement tab> More button> Chirp measurement.
2.
To access the Configure Measurements menu, go to Measurements>
Configure.
3.
In the Configure tab, you can select the DUT (Host or Device) and perform
selected tests associated with it.
TDSUSB2 Universal Serial Bus Measurement Package
37
Operating Basics
Figure 16: Configuring Chirp measurement for host
4.
Select the Host option to display the different tests for the DUT. The
available tests are:
EL_33, EL_34: You can perform two separate measurements on an
acquired waveform
EL_35: You can perform a single measurement on an acquired
waveform
Figure 17: Configuring Chirp measurement for device
5.
Select the Device option to display the different tests for the DUT. The
available tests are:
EL_28, EL_29: You can perform two separate measurements on an
acquired waveform
EL_31: You can perform a single measurement on an acquired
waveform
38
TDSUSB2 Universal Serial Bus Measurement Package
Operating Basics
6.
Select the channel source in the D+ and D– fields. Use the drop down arrow
in the D+, D- field to set the source. The available options are: Ch1- Ch4
and Ref1-Ref4.
7.
Select the Run button.
8.
To generate reports, select Utilities> Report Generator.
Configuring Reset from High Speed Measurement
To configure Reset from High Speed measurement, follow these steps:
1.
Select the High Speed measurement tab> More button> Reset from High
Speed measurement. This is the time between the beginning of the last SOF
and before the reset and the beginning of Chirp-K. This test will be
applicable for Device and HUB upstream.
2.
To access the Configure Measurements menu, go to Measurements>
Configure.
Figure 18: Configuring Reset from High Speed measurement
3.
The Upstream signal direction is enabled by default. This is because you can
measure Reset from High Speed only on an upstream signal. The
Downstream option is disabled here.
4.
Select the channel source in the D+ and D– fields. Use the drop down arrow
in the D+, D– field to set the source. The available options are: Ch1- Ch4
and Ref1-Ref4.
5.
Select the Run button.
6.
To generate reports, select Utilities> Report Generator.
TDSUSB2 Universal Serial Bus Measurement Package
39
Operating Basics
Configuring Resume Measurement
To configure Resume measurement, follow these steps:
1.
Select the High Speed measurement tab> More button> Resume
measurement. The device/HUB resumes the High-Speed operation, which is
indicated by the presence of High-speed SOF packets (with 400 mV
nominal amplitude) following the K state driven by the host controller. For
the Host, this is the time between the falling edge of D+ to the First SOF.
This should not exceed 3.0 ms.
2.
To access the Configure Measurements menu, go to Measurements>
Configure.
Figure 19: Configuring Resume measurement
40
3.
Set the input Signal Direction to either Upstream or Downstream.
4.
Select the channel source in the D+ and D– fields. Use the drop down arrow
in the D+, D– field to set the source. The available options are: Ch1- Ch4
and Ref1-Ref4.
5.
Select the Run button.
6.
To generate reports, select Utilities> Report Generator.
TDSUSB2 Universal Serial Bus Measurement Package
Operating Basics
Configuring Suspend Measurement
To configure Suspend measurement, follow these steps:
1.
Select the High Speed measurement tab> More button> Suspend
measurement. This measures the time between the end of the last SOF and
the rising edge transition to the Full Speed J state. The acceptable range is
between 3 ms to 3.125 ms.
2.
To access the Configure Measurements menu, go to Measurements>
Configure.
Figure 20: Configuring Suspend measurement
3.
Set the input Signal Direction to either Upstream or Downstream.
4.
Select the channel source in the D+ and D– fields. Use the drop down arrow
in the D+, D– field to set the source. The available options are: Ch1- Ch4
and Ref1-Ref4.
5.
Select the Run button.
6.
To generate reports, select Utilities> Report Generator.
TDSUSB2 Universal Serial Bus Measurement Package
41
Operating Basics
Configuring Reset from Suspend Measurement
To configure Reset from Suspend measurement, follow these steps:
1.
Select the High Speed measurement tab> More button> Reset from Suspend
measurement. This measures the time between the falling edge of D+ signal
and the start of Device chirp-K. It is between 2.5 us and 3 ms. This test is
applicable for Device and HUB upstream.
2.
To access the Configure Measurements menu, go to Measurements>
Configure.
Figure 21: Configuring Reset from Suspend measurement
42
3.
In the Configure tab, you can set the channel Source.
4.
The signal direction is always set to Upstream. This is because you can
measure Reset from Suspend only on an upstream signal. The Downstream
option is disabled.
5.
Select the channel source in the D+ and D– fields. Use the drop down arrow
in the D+, D– field to set the source. The available options are: Ch1- Ch4
and Ref1-Ref4.
6.
Select the Run button.
7.
To generate reports, select Utilities> Report Generator.
TDSUSB2 Universal Serial Bus Measurement Package
Operating Basics
Configuring Packet Parameter Measurement
To configure Packet Parameter measurement, follow these steps:
1.
Select the High Speed measurement tab> More button> Packet Parameter
measurement.
2.
To access the Configure Measurements menu, go to Measurements>
Configure.
3.
In the Configure tab, you can select the DUT (Host or Device) and perform
selected tests associated with it.
Figure 22: Configuring Packet Parameter measurement for host
4.
Select the Host option to display the different tests for the DUT. The
available tests are:
EL_21, EL_23, EL_25: You can perform three separate measurements
on a single acquired waveform
EL_22: You can perform a single measurement on an acquired
waveform
EL_55: You can perform a single measurement on an acquired
waveform
TDSUSB2 Universal Serial Bus Measurement Package
43
Operating Basics
Figure 23: Configuring Packet Parameter measurement for device
5.
Select the Device option to display the different tests for the DUT. The
available tests are:
EL_21, EL_22, EL_25: You can perform three separate measurements
on an acquired waveform
EL_22: You can perform a single measurement on an acquired
waveform
6.
Use the drop down arrow in the Differential field to set the channel source.
The available options are: Ch1-Ch4, Ref1-Ref4.
7.
Select the Run button.
8.
To generate reports, select Utilities> Report Generator.
Packet Parameter Measurement
Packet parameter measurement is for high-speed Host controller and Device.
There are several important packet characteristics for upstream and downstream
signaling. The measurement calculates the SYNC field length, EOP length and
inter packet gap. The acceptable range of EOP for all transmitted packets (except
SOFs) must be between 7.5 and 8.5 bits. The packet parameter algorithm
calculates the EOP depending on this range.
The acceptable range of inter packet gap is between 88 bits to 192 bits. The
SYNC field for all transmitted packets is calculated by counting the bits to check
for 32-bit SYNC field. The inter packet delay and SYNC status is calculated
between the EOP indexes. The EOP width in time is calculated by dividing the
EOP width by 480 Mbps.
44
TDSUSB2 Universal Serial Bus Measurement Package
Operating Basics
Suspend Measurement
This test calculates the time between the end of last SOF and the rising edge
transition to Full-speed J state for Host / Device / Hub- upstream. This time must
be between 3 ms and 3.125 ms. To get the Suspend signal, hot-plug the unit
under test (device) and measure the signaling with single-ended probes on both
lines. The application analyzes data for the Suspend Time.
Resume Measurement
This test calculates the device/HUB resume High-speed operation, indicated by
High-speed SOF packets (with 400 mV nominal amplitude) following the K state
driven by the host controller. For the Host, this is the time between the falling
edge of D+ and the First SOF. This should not exceed 3.0 ms. To get the Suspend
signal, hot-plug the unit under test (device) and measure the signaling with
single-ended probes on both lines. The application analyzes data for the
following:
Resume Time
Amplitude
TDSUSB2 Universal Serial Bus Measurement Package
45
Operating Basics
Reset from Suspend Measurement
This test calculates the time between the falling edge of D+ signal and the start of
Device chirp-K for the Device/HUB upstream. This must be between 2.5 us and
3 ms. To get the Reset from Suspend Measurement signal, hot-plug the unit
under test (device), and measure the signaling with single-ended probes on both
lines. The application analyzes data for the Reset From Suspend Time.
46
TDSUSB2 Universal Serial Bus Measurement Package
Operating Basics
TDSUSB2 Universal Serial Bus Measurement Package
47
Operating Basics
Reset from High Speed Measurement
This test calculates the time between the beginning of the last SOF and before the
reset and the beginning of Chirp-K for Device and HUB upstream. This must be
between 3.1 ms and 6 ms. To get the Reset from High-Speed Measurement
signal, hot-plug the unit under test (device) and measure the signaling with
single-ended probes on both lines. The application analyzes data for Reset From
High Speed Time.
48
TDSUSB2 Universal Serial Bus Measurement Package
Taking Measurements
Acquiring Data
The application automatically sets the oscilloscope settings for the selected
measurements. To acquire data from the oscilloscope, follow these steps:
command button to run the application.
1.
Select the
2.
The application displays the message “Please press OK when correct
waveform is acquired” for live signals.
Figure 24: Confirm waveform screen
3.
If you select the Cursor Mode, the application displays the message “Place
the Cursors at the start and end position of the USB packet.” Press OK when
done.
Figure 25: Place cursors screen
4.
If you select the Inrush measurement, the application displays the message
in the following figure:
TDSUSB2 Universal Serial Bus Measurement Package
49
Taking Measurements
Figure 26: Confirm waveform for Inrush measurement
5.
If you select the Droop measurement, the application displays the message
in the following figure:
Figure 27: Confirm waveform for Droop measurement
6.
If you select the Packet Parameter measurement, the application displays the
message in the following figure:
Figure 28: Confirm waveform screen
50
7.
The application automatically displays the result after acquiring the data.
8.
The application automatically displays the eye diagram and the waveform
plot for the acquired signals. This is possible only if you enable this option
in the Preferences> General.
TDSUSB2 Universal Serial Bus Measurement Package
Taking Measurements
Note: You can modify the automatic oscilloscope settings if there is no valid
waveform on the oscilloscope screen. To do so, select the
to run the application and select OK to complete the process.
command button
You may need to adjust the inrush setups, as inrush currents have a wide
variety of durations and peak currents. Use the vertical division settings between
100 mA/division to 5 A/division and time bases between 500 ns/div and 20
µs/division. If you set a time base higher than 20 µs/division, adjust the
instrument’s record length to have minimum acquisition duration of 20 µs.
Control Menu Options
Table 8. Control options
Option/button
Description
Run icon
Runs the application
Hide button
Hides the application and displays the oscilloscope
application on the entire screen
Exit button
Exits the application
Viewing Results
You can view the results in a summary form or in a detailed form. To access the
Results: Summary, select Results> Summary. The application displays one of the
following results.
A PASS indicates that the results of the measurements comply with
the USB2.0 specifications.
A FAIL indicates that the measured values of the measurements are
beyond the waiver limits and do not comply with the USB2.0 specifications.
A Conditional PASS indicates that the limits of the tests are within
the USB2.0 waiver limits.
Note: The application displays PASS or FAIL based on the limits you set.
TDSUSB2 Universal Serial Bus Measurement Package
51
Taking Measurements
To access the Details: select Results> Details. Use the scroll bar to view the
results that are not visible within the display window. The report contains
statistical values for the following:
Standard deviation (StdDev)
Mean
Peak-to-peak (Pk–Pk)
Root mean square (RMS)
Maximum (Max) and minimum (Min) values
Population (the number of cycles used to calculate the statistics). The
population used for signal rates is number of bits
Status (PASS/FAIL/Conditional PASS) for the selected tests
You can also view the eye diagram and the waveform plot for signal quality
check.
Eye Diagram
You can select the Eye Diagram option to view the eye diagram for the unit
under test. The eye diagram has the Zoom and Cursor features that you can use to
view the results. You can also use the Reset button to reset the default eye
pattern.
The Eye Diagram menu contains the following options:
icon to zoom
Zoom: You can select the zoom from the menu. Select the
into the area of interest. Define the area of interest using the mouse and selecting
the zoom in area. The selected area is displayed in the entire upper half of the
oscilloscope. Select the
icon to zoom out. You can use the Zoom In and
Zoom Out icons until the application reaches the maximum and the minimum
zoom limits.
52
TDSUSB2 Universal Serial Bus Measurement Package
Taking Measurements
Figure 29: Eye diagram
Figure 30: Eye diagram (zoomed)
Cursors: You can select the Cursors and the vertical cursors
appear. You can drag the cursors to change the
positions, and read the time values on the X-axis.
Use the mouse to select the Horizontal cursor
. You
can drag the cursors to change the positions, and read the voltage values on the
Y-axis.
TDSUSB2 Universal Serial Bus Measurement Package
53
Taking Measurements
Waveform Plot
You can select the Waveform Plot option to view the Waveform Plot for the unit
under test. The Waveform Plot has a Zoom and the Cursor feature that you can
use to view the results.
The Waveform Plot has features that enable you to zoom in and out on the
waveform, use Vertical and Horizontal Cursors, Save the Plot, Reset the original
plot, and select to display the signals.
To display any or all the signals for the waveform plots menu, you can use the
following options:
icon to zoom
Zoom: You can select the Zoom from the menu. Select the
into the area of interest. Define the area of interest using the mouse and selecting
the zoom in area. The entire upper half of the oscilloscope displays the selected
icon to zoom out. You can use the Zoom In and Zoom Out
area. Select the
icons until the application reaches the maximum and the minimum zoom limit.
Figure 31: Using the zoom feature
Cursor: You can select the Cursors and the vertical cursors
appear. Drag the cursors to change the positions of
the cursors, and read the time values on the X-axis.
54
TDSUSB2 Universal Serial Bus Measurement Package
Taking Measurements
Figure 32: Using the cursors and viewing cursor readout
Select the Horizontal cursors
. You can drag the
horizontal cursors to change the positions, and read the voltage values on the Yaxis.
Note: The Zoom and Cursor options are mutually exclusive.
Save: You can use the Save button to save the zoomed in or zoomed out diagram
as a .jpg file.
Reset: You can use the Reset button to restore the waveform plot to its original
display.
For waveform plots, any or all the signals can be displayed using the following
buttons:
D+: Use this button
to turn on or turn off the display of the D+ signal
D–: Use this button
to turn on or turn off the display of the D– signal
CMD: Use this button
Common Mode voltage
Diff: Use this button
Differential signal
TDSUSB2 Universal Serial Bus Measurement Package
to turn on or turn off the display of the
to turn on or turn off the display of the
55
Taking Measurements
You can use any combination of the Zoom or Cursors, D+, D–, CMD (Common
Mode Voltage) and Diff buttons to view and save the waveform plot.
The annotations at the lower edge of the Waveform Plot classify the signal pulse
into different bus states: J, K, E (EOP), I (IDLE) and Sync (C).
There are two Results menu options.
Summary: The application summarizes the results of the measurements. The
results are classified as PASS, FAIL and Conditional PASS. Click on PASS,
FAIL and Conditional PASS. The dialog box shows the Measured Mean, USB2.0
Specifications.
Figure 33: Details popup
Details: The application displays the results of the completed tests in a tabular
form.
Figure 34: Results details
Note: The application displays the user column in the Results Details only if the
User Configurable Limits are selected in File> Preferences> Advanced tab.
The application disables the Results> Summary and Report generation in PlugFest Specific format when you enable Configure Test Limits.
56
TDSUSB2 Universal Serial Bus Measurement Package
Taking Measurements
Report Generation
Generating Reports
To access the Report Generator menu, select Utilities> Report Generator. You
can generate the USB IF report in any one of the three formats:
Tektronix Specific is the default format used by the TDSUSB2 application
Plug-Fest format is the format used by the USB-IF
CSV format report is a user-defined report generated in a comma separated
variable format
You can generate reports in one of the following modes:
Manual: Select the report format, the directory name, and the file to which
the report is saved. Select the Generate option to generate the report. Specify
the file name for the report
Automatic: The application generates reports automatically without user
intervention as soon as the results are calculated. Select Generate Automatic
Ids to enable this option while selecting measurements. The application uses
the prefix that you enter to generate the file. If you select the Ask for the
Filename/Device ID before automatically generating Reports/Data files
check box in the File> Preferences> General tab, you can enter the Device
ID/File Name before generating the report. If the File already exists, the
application displays a message box to confirm whether to overwrite the
existing report. If you select Yes, the application overwrites the report; if
you select NO, you can enter a new Device ID/File Name or cancel the
report generation. The reports are generated in the following path:
C:\TekApplications\tdsusb2\report
The file name is prefix_001_reportformat.csv or efix_001_reportformat.htm.
The report contains the device ID, device description, date and result of the test.
The report also contains the following statistical values:
Maximum (Max) and Minimum (Min) values
Mean
Peak-to-peak (Pk–Pk)
Standard deviation (StdDev)
Root mean square (RMS)
Population
TDSUSB2 Universal Serial Bus Measurement Package
57
Taking Measurements
Status (PASS/FAIL/Conditional PASS) for the selected tests
User- The application displays this value if you select the "Allow user
configuration of test limits" in the Preferences menu and set the test limits
before running the measurement
The report includes the eye diagram and the waveform plots for the Signal
Quality Check. The display on the oscilloscope screen is embedded in the Inrush
or the Droop report.
Note: Automatic Report generation is not available if you select the Drop Test
readout option in File> Preferences> Advanced tab.
Viewing Reports
You can use an HTML viewer or a browser to view the Tektronix specific and
Plug-Fest format reports. You can view the .csv (Comma Separated Variable)
report in a text editor, spreadsheet, database, or a data analysis program for
further analysis. You can edit the .csv file to suit your needs.
Note: All the reports contain the eye diagram and the waveform plot except for
the reports generated in the .csv file format.
TSV File Generator
You can use the TSV File Generator Utility to convert a .csv (Comma Separated
Variable) file to a .tsv (Tab Separated Variable) file. To access TSV File
Generator, go to the main menu and select Utilities> TSV File Generator.
If the Input is a .csv file, then the application enables the CSV Waveform Source
area. You can select the sources based on the Signal type. For a single-ended
signal, specify the D+ and D– inputs.
You can specify the directory C:\TekApplications\tdsusb2\tsvfilegenerator for
the differential signal from where the .csv file is taken.
If the Input is a Live signal, you cannot access the CSV Waveform Source area.
Select the Live or Ref as the source file to enable the Live Input option. You can
enable the Convert option only if you take the measurements and display the
results. The default directory for the .tsv files is
C:\TekApplications\tdsusb2\tsvfiles. You can also specify the location where the
tsv file is to be generated.
58
TDSUSB2 Universal Serial Bus Measurement Package
Taking Measurements
TSV file format description
For Low Speed and Full Speed Signal Quality Check, the input TSV file is an m
x 3 matrix with m rows and three columns. The file has Time values and Voltage
values for D+ and D– in the first, second and third columns respectively.
The input TSV file for High Speed Signal Quality Check is a m x 2 matrix with
m rows and two columns. The file has Time and Voltage values for Differential
signals in the first and second columns.
The input TSV files for Inrush and Droop tests are an m x 2 matrix with m rows
and 2 columns. For Inrush tests, the file has Time and Current values in the first
and second columns. For Droop test, the file has Time and Voltage values in the
first and second columns.
CSV file format description
Low Speed and Full Speed single-ended signals for Signal Quality Check require
two .csv files to generate the corresponding TSV file. The .csv files have two
columns. One .csv file has time and voltage values for D+ signals, while the
second column has the time and the current values for D– signals. Both files must
have the same time values.
Differential signals require one .csv file with two columns, one column for time
and another for voltage values.
If you select Differential for Inrush and Droop signals, the .csv file has two
columns: one column for time values, and another for current and voltage values.
You require only one .csv file for Inrush and Droop tests. This .csv file has two
columns: one column for the Time Values and the other for Current (Inrush) and
Voltage (Droop) values. You must use the Differential option for conversion.
TDSUSB2 Universal Serial Bus Measurement Package
59
Taking Measurements
Saving and Recalling Setups
How to Save and Recall a Setup
You can use the Save and Recall menus to save and recall the various
configuration setups. To access the Save and Recall menus, go to the File menu
in the menu bar and choose Save or Recall.
Note: Do not edit the .ini or the .set files, or the recall setup files generated by
the application. This can cause instability to the application. If you try to edit
the.ini of the .set files, you get an error message ‘The saved file has been
corrupted and cannot be recalled.’
Saving a Setup
You can save various configuration setups and recall them when needed. To save
the application settings to a setup file, follow these steps:
1.
Select File> Save
2.
Select the Save button
3.
Browse or enter a file name. The application appends an .ini extension to the
name of setup files
4.
Choose Save to save the setup or Cancel to cancel the action
Note: The application saves the oscilloscope setup to a .set file. The application
.ini files and the oscilloscope .set files have the same file name.
Recalling a Saved Setup
To recall the application settings from a saved setup file, follow these steps:
1.
Select File> Recall
2.
Browse the directory C:\TekApplications\tdsusb2\setup to recall the .ini files
or select the directory where you have saved the setup file
3.
Select or enter an .ini file name
4.
Choose Open to recall the setup or Cancel to cancel the operation
Note: The application recalls the oscilloscope setup from a .set file.
60
TDSUSB2 Universal Serial Bus Measurement Package
Taking Measurements
Recalling the Default Setup
To recall the application settings from the Default setup file, select File> Recall
Default. The application recalls the default setup and displays the message
‘Default setup recalled successfully.’
Figure 35: Recalling a default setup
Recently Saved Setup
The application stores the last four saved setups. If you need to modify any of
them, follow these steps:
1.
Select File> Recently Saved.
2.
Select the setup from the list of recently saved setups that are displayed.
Note: When you select a file from the list of recently saved files menu, the file
becomes the first element in the list of selections.
Recently Recalled Setup
The application stores the last four recalled setups. If you need to recall any of
them, follow these steps:
1.
Select File> Recently Recalled.
2.
Select the setup from the list of setups displayed.
Note: When you select a file from the list of recently saved files popup menu, the
file becomes the first element in the list of selections.
TDSUSB2 Universal Serial Bus Measurement Package
61
Taking Measurements
62
TDSUSB2 Universal Serial Bus Measurement Package
Tutorial
Introduction to the Tutorial
The tutorial teaches you how to set up the application by recalling a .tsv file, take
measurements, and view the results. More operating information is available in
the Operating Basics section. Before you begin the tutorial, you must do the
following tasks:
Start the application
Recall a .tsv file
Starting the Application
To start the application, select File> Run Application> USB2.0 Test Package or
App > USB2.0 Test Package or Analyze > USB2.0 Test Package.
Recalling a .tsv File
The application distribution includes the .tsv files used with this tutorial. The
table below shows the types of signals that represent these waveforms.
Table 9. .tsv files
Tsv File name
Signal type
C:\TekApplications\tdsusb2\tsvfiles\LS_SQC.tsv
C:\TekApplications\tdsusb2\tsvfiles\FS_SQC.tsv
C:\TekApplications\tdsusb2\tsvfiles\HS_sqc.tsv
Low Speed
Full Speed
High Speed
C:\TekApplications\tdsusb2\tsvfiles\inrush.tsv
Low/Full/High Speed
TDSUSB2 Universal Serial Bus Measurement Package
63
Tutorial
Taking a Full Speed Signal Quality Measurement
This section discusses how to take a Full Speed Signal Quality measurement,
view the results, and generate a report. To perform these tasks, the application
must be installed and enabled on a supported oscilloscope. View Installing the
application to install the application.
To take a Full Speed Signal Quality measurement, follow these steps:
1.
To set the application to default values, select File> Recall Default. The
application displays the message ‘Default setup recalled successfully.’
Select OK to view a screen image showing the default settings.
2.
Select all the Signal Quality measurements in the Signal Quality Check area
using Select All button.
3.
Select the Configure tab or select Measurements> Configure. The
Measurements: Configure default settings display.
4.
Select the Measurements: Source. If the source is Live or Ref, configure the
appropriate channels for D+, D– and Qualifier. You have to always select
D+ first. If you select any one of the Live channels CH1-CH4 for D+
source, the remaining Live channels are selected from the D– and the
Qualifier channel source. If you select any one of the Ref channels Ref1Ref4 for D+ source, the remaining Ref channels are selected from the D–.
There is no Qualifier channel source for the Ref signals.
If the source is from a file, use the browse button to look for the file.
Select the file C:\TekApplications\tdsusb2\tsvfiles\FS_SQC.tsv.
5.
64
command button to run the application. The application
Select the
displays the Eye Diagram and the Waveform Plot.
TDSUSB2 Universal Serial Bus Measurement Package
Tutorial
Figure 36: Eye diagram
Figure 37: Waveform plot
6.
Minimize the eye diagram and waveform plot to view the summary results.
7.
The application displays the Results Summary as PASS
can also select the result to view the details of the selected test.
8.
To view the Results Details, select Results> Details in the application menu
bar or PASS command button in the Overall Result area to view the details
of each measurement.
TDSUSB2 Universal Serial Bus Measurement Package
. You
65
Tutorial
Figure 38: Results details for Signal Quality Check
9.
Select Utilities> Report Generator in the application menu bar to generate
the Report.
10. You can view the default screen with the Tektronix Specific Format
enabled. The report directory appears with a default file name. You can
change the file name if you want. Click on the Generate button.
11. Click the Generate button. The application generates an HTML file in
C:\TekApplications\tdsusb2\report. To view this report, open it in an HTML
viewer or a browser.
66
TDSUSB2 Universal Serial Bus Measurement Package
Tutorial
Figure 39: Report in Tektronix-Specific format for Signal Quality Test
Taking a Low Speed Inrush Current Measurement
This section discusses how to take a Low Speed Inrush Current Measurement,
view the results, and generate a report. To perform these tasks, the application
must be installed and enabled on a supported oscilloscope.
1.
To set the application to default values, select File> Recall default. The
application displays the message ‘Default setup recalled successfully.’
2.
Select Measurements> Low Speed tab.
3.
Select Inrush Current measurement.
4.
Select Measurements> Configure tab and use the default settings set by the
application.
5.
Select the Source tab. Select From File.
6.
Select file C:\TekApplications\tdsusb2\tsvfiles\LS_inrush.tsv.
7.
Select the
8.
The application displays the Results Summary as Conditional PASS.
9.
You can also select the result to view the details of the selected tests.
command button to run the application.
TDSUSB2 Universal Serial Bus Measurement Package
67
Tutorial
10. Select Results> Details from the application menu bar to view Results
Details.
Figure 40: Results details for Inrush Current Check
11. Select Utilities> Report Generator in the application menu bar to generate a
report.
12. Select the CSV Specific Report and use the default file name. Select the
Generate button.
13. The application displays the sample report file in a browser.
68
TDSUSB2 Universal Serial Bus Measurement Package
Tutorial
Figure 41: Report for Inrush Current in CSV format
Taking a High Speed Signal Quality Measurement
This section discusses how to take a High Speed Signal Quality measurement,
view the results, and generate the report. To perform these tasks, the application
must be installed and enabled on a supported oscilloscope.
To take a High Speed Signal Quality measurement, follow these steps:
1.
To set the application to default values, select File > Recall Default. The
application displays the message ‘Default setup recalled successfully.’
2.
Select Measurements> High Speed and select all the Signal Quality
measurements in the Signal Quality Check area using Select All button.
3.
Select the Configure option or go to Measurements> Configure.
4.
Select the Source tab> From File>Browse and locate the file.
5.
Select the file C:\TekApplications\tdsusb2\tsvfiles\HS_sqc.tsv.
6.
Select the
command button to run the application. The application
automatically displays the Eye Diagram and the Waveform Plot.
TDSUSB2 Universal Serial Bus Measurement Package
69
Tutorial
Figure 42: Eye diagram for a High Speed Signal Quality Measurement
Figure 43: Waveform plot for a High Speed Signal Quality Measurement
70
7.
Minimize the eye diagram and the waveform plot button to view the
summary results.
8.
Select Results> Details in the application menu bar to view the Results
Details. Click the Additional Information button to display the additional
information.
TDSUSB2 Universal Serial Bus Measurement Package
Tutorial
Figure 44: Results details for a High Speed Signal Quality Measurement
9.
Select Utilities> Generate Report to generate the report.
10. Select the Report format as Plug-Fest Specific and use the default file name.
Select Generate.
11. The application displays the sample report file in a browser.
Figure 45: Report in Plug-Fest format for High Speed Signal Quality Measurement
TDSUSB2 Universal Serial Bus Measurement Package
71
Tutorial
Taking a Low Speed Signal Quality Measurement
This section discusses how to take a Low Speed Signal Quality measurement,
view the results, and generate a report. To perform these tasks, the application
must be installed and enabled on a supported oscilloscope. View Installing the
application to install the application.
To take a Low Speed Signal Quality measurement, follow these steps:
1.
To set the application to default values, select File> Recall Default. The
application displays the message ‘Default setup recalled successfully.’
Select OK to view a screen image showing the default settings.
2.
Select the Low Speed tab and select the Signal Quality measurements in the
Signal Quality Check area using Select All button.
3.
Select the Configure tab or select Measurements> Configure. The
Measurements: Configure default settings display.
4.
Select the Measurements: Source. If the source is Live or Ref, configure the
appropriate channels for D+, D– and Qualifier. You have to always select
D+ first. If you select any one of the Live channels CH1-CH4 for D+
source, the remaining Live channels are selected from the D– and the
Qualifier channel source. If you select any one of the Ref channels Ref1Ref4 for D+ source, the remaining Ref channels are selected from the D–.
There is no Qualifier channel source for the Ref signals.
If the source is from a file, use the browse button to look for the file.
Select the file C:\TekApplications\tdsusb2\tsvfiles\LS_SQC.tsv.
5.
72
command button to run the application. The application
Select the
displays the Eye Diagram and the Waveform Plot.
TDSUSB2 Universal Serial Bus Measurement Package
Tutorial
Figure 46: Eye diagram for Low Speed Signal Quality Measurement
Figure 47: Waveform plot for Low Speed Signal Quality Measurement
6.
Minimize the eye diagram and waveform plot to view the summary results.
7.
The application displays the Results Summary as PASS
can also select the result to view the details of the selected test.
8.
To view the Results Details, select Results> Details in the application menu
bar or PASS command button in the Overall Result area to view the details
of each measurements.
TDSUSB2 Universal Serial Bus Measurement Package
. You
73
Tutorial
Figure 48: Results details for Low Speed Signal Quality Measurement
9.
Select Utilities> Report Generator in the application menu bar to generate
the Report.
10. You can view the default screen with the Tektronix Specific Format
enabled. The report directory appears with a default file name. You can
change the file name if you want. Click the Generate button.
11. Click the generate button. The application generates an HTML file in
C:\TekApplications\tdsusb2\report. To view this report, open it in an HTML
viewer or a browser.
74
TDSUSB2 Universal Serial Bus Measurement Package
Tutorial
Figure 49: Report in Tektronix-Specific format for Low Speed Signal Quality Measurement
TDSUSB2 Universal Serial Bus Measurement Package
75
Tutorial
76
TDSUSB2 Universal Serial Bus Measurement Package
Application Examples
This section presents the application examples. The simplified examples
highlight the application measurements and show how to use the application to
solve your test problems.
To use these examples, you must have the TDSUSB2 application installed and
enabled on the oscilloscope. Connect the probes to your unit under test (UUT),
and perform the configuration tasks. For information, see Installing the
application and Starting the Application.
Full Speed Signal Quality Tests for Downstream Testing
Specifying the Equipment-Full Speed Signal Quality Downstream Tests
The following equipment is needed for downstream signal quality check on a
Full Speed device testing:
Tektronix digital oscilloscope
TDSUSB2 application
TDSUSBF compliance test fixture
Three P6245 or TAP1500 single-ended voltage probes
Typical Equipment Setup-Full Speed Signal Quality Downstream Tests
To set up the equipment for Full Speed Signal Quality Downstream test, follow
these steps:
1.
Set the S6 switch to the Init position.
2.
Use the adapters to connect the A receptacle from Device SQ test section
(marked DUT) of the test fixture to the USB2.0 Low Speed device.
3.
Connect Ch1 of the D+ probe to the D+ pins on the Device SQ section of the
test fixture.
4.
Connect Ch2 of the D– probe to the D– pins on the Device SQ section of the
test fixture.
5.
Connect the Init port of the Device SQ section of the test fixture to any port
of the unit under test (hub) using the USB cable.
TDSUSB2 Universal Serial Bus Measurement Package
77
Application Examples
6.
Select the measurement and select the
application.
command button to run the
7.
Select OK after acquiring a waveform. Verify that it is a correct waveform.
Note: Make sure the acquired signal is a valid waveform.
Figure 50: Equipment setup for Full Speed Signal Quality Measurement for downstream testing
Selecting and Configuring Measurements-Full Speed Signal Quality Tests
Follow the steps to select measurements for Full Speed Signal Quality check:
1.
From the application menu, select Measurement > Select> Full Speed tab.
2.
Select the following signal quality checks:
Eye Diagram Test
Signal Rate
Paired JK Jitter
Paired KJ Jitter
Consecutive Jitter
EOP Width
78
TDSUSB2 Universal Serial Bus Measurement Package
Application Examples
Cross-Over Voltage
Rise Time
Fall Time
3.
Select the Select All toggle button to select all the measurements
simultaneously. Click on any measurement button to deselect it.
Configuring the Measurement
Follow the steps to configure the selected measurements:
1.
From the application menu, select Measurement> Configure> Configure.
2.
Configure the following options:
Table 10. Configuring Measurements
Option
Tier
Direction
Test Point
Set to
Tier6
Downstream
Set the test point to Far End
3.
Select the Source tab.
4.
Configure the following options:
Table 11. Configure options
Option
Live/Ref
Set to
Single-ended
5.
If you select Single-ended, you must select two channels for D+ and D–.
6.
Select
7.
The application automatically displays the eye diagram and waveform plot
of the signal acquired from the unit under test.
to acquire the data.
TDSUSB2 Universal Serial Bus Measurement Package
79
Application Examples
Viewing Results-Full Speed Signal Quality Tests
To view the results of the tests, follow these steps:
1.
From the application menu, select Results> Summary.
Figure 51: Results summary for Full Speed Signal Quality Measurement
2.
Click on any of the test result
test.
buttons to get the details of that
3.
From the application menu, select Results> Details.
Figure 52: Results details for Full Speed Signal Quality Measurement
80
4.
Click the Eye Diagram option in Results Summary or the Details to view the
Eye Diagram.
5.
Click the Waveform Plot option to view the annotated waveform plot.
TDSUSB2 Universal Serial Bus Measurement Package
Application Examples
Generating Reports-Full Speed Signal Quality Tests
To generate reports of the test results, follow these steps:
1.
From the application menu, select Utilities> Report Generator.
2.
Select any one of the Report Formats: Tektronix Specific, Plug-Fest Specific
or CSV format.
Figure 53: Report in Plug-Fest format for Full Speed Signal Quality Measurement
3.
Select the manual generation mode.
4.
Select the Generate button to display and view the selected report format.
Note: You can view the Plug-Fest specific report format as HTML pages and the
.csv format in Microsoft Excel.
TDSUSB2 Universal Serial Bus Measurement Package
81
Application Examples
Droop Test for Ports of Hub
Specifying the Equipment-Low Speed Droop Tests
The following equipment is needed to perform a Droop Test for Low Speed
device:
Tektronix digital oscilloscope
TDSUSB2 application
TDSUSBF compliance test fixture
Two P6245or TAP1500 single-ended probes
Typical Equipment Setup-Low Speed Droop Tests
To set up the equipment for Droop test, follow these steps:
82
1.
Use the Droop and Adjacent Trigger section for the Droop test. Use standard
USB cable to connect the loads on the test fixture to the ports of the unit
under test as shown in the following figure.
2.
The setup requires two channels of the oscilloscope: one for the Droop Load
Trigger Timer that is on the Droop test load section, and the other for VBUS
that is on the Adjacent Trigger and Droop Test section of the test fixture.
Use the load switch to select appropriate loads for Droop Test.
3.
In the Droop and Adjacent Trigger section on the test fixture, probe Ch1
from the VBUS and ground the pins.
4.
In the Droop Test Load section on the test fixture, probe Ch2 from the
oscillator (marked OSC) and ground the pins.
5.
Connect the Droop Trigger Test Load to the Port1 of the unit under test
(hub).
6.
Connect the Load1 to Port2 of the hub under test.
7.
Connect the Load2 to the A receptacle of the Droop and Adjacent Trigger
section on the test fixture. Connect the A pin dongle from the Droop and
Adjacent Trigger section to the Port4 of the unit under test (hub). Port4 is
the port under test of the hub.
8.
Connect the Load3 to Port3 of the unit under test (hub). Now all the ports of
the unit under test are connected (hub).
TDSUSB2 Universal Serial Bus Measurement Package
Application Examples
9.
Select the measurement and select the
application.
command button to run the
10. Select OK when the valid waveform is acquired.
11. Observe the droop in the VBUS when Droop test load is applied.
Note 1: The application automatically sets up the oscilloscope to acquire the
Droop signal. If you do not get a valid signal, set up the oscilloscope
accordingly.
Note 2: Use the Load Switch to select 500 mA test loads for Droop testing of a
self powered hub and system. Use the Load Switch to select 100 mA test loads for
Droop testing of bus powered hub and system. Use the Load Switch to select 500
mA test loads for Droop testing of self powered hub and system.
Figure 54: Equipment setup for Low Speed Droop test
TDSUSB2 Universal Serial Bus Measurement Package
83
Application Examples
Selecting and Configuring Measurements-Droop Tests
From the application menu, select Measurement> Select> Droop Test.
Configuring the Measurement
Follow the steps to configure the selected measurement:
1.
From the application menu, select Measurement> Configure.
2.
Configure the following options:
Table 12. Configuring measurement
Option
Set to
Port
Source
VBUS
Trigger
Port under test
Downstream
Ch1
Ch2
Device type
Self-powered
to acquire the data.
3.
Select
4.
After acquiring the data, the application displays the eye diagram and the
waveform plot automatically.
Viewing Results-Droop Tests
To view the results of the tests, follow these steps:
84
1.
Run the application and from the application menu, select Results>
Summary.
2.
Click on any of the test result buttons to get the details of that test.
3.
From the application menu, select Results> Details.
4.
Click on the Eye Diagram option in Results Summary or Results Details to
view the Eye Diagram.
5.
Click on the Waveform Plot to view the annotated waveform plot.
TDSUSB2 Universal Serial Bus Measurement Package
Application Examples
Generating Reports-Droop Tests
To generate reports of the test results, follow these steps:
1.
From the application menu, select Utilities> Report Generator.
2.
Select any one of the Report Formats: Tektronix Specific, Plug-Fest Specific
or CSV format.
3.
Select Manual option to generate the report.
4.
Select the Generate button to display and view the selected report format.
Note: You can view the Plug-Fest specific report format as HTML pages and the
.csv format in Microsoft Excel.
TDSUSB2 Universal Serial Bus Measurement Package
85
Application Examples
Full Speed Signal Quality Tests for Upstream Testing
Specifying the Equipment-Full Speed Signal Quality Tests for Upstream Testing
The following equipment is needed for upstream signal quality check on a low or
full speed device:
Tektronix digital oscilloscope
TDSUSB2 application
TDSUSBF compliance test fixture
Three P6245 or TAP1500 single-ended voltage probes
Typical Equipment Setup-Full Speed Signal Quality Tests for Upstream Testing
To set up the equipment for Full Speed Signal Quality test for Upstream, follow
these steps:
1.
Use the A receptacle to connect the USB unit under test (device) to the
Inrush Droop section of the test fixture.
2.
Connect the Qualifier device to the Adjacent Trigger and Droop section of
the test fixture as shown in the following figure.
3.
Connect Ch1 of the D+ probe to the D+ pins on the Inrush Droop section of
the test fixture.
4.
Connect Ch2 of the D– probe to the D– pins on the Inrush Droop section of
the test fixture.
5.
Connect the D+ (D- for Low speed) pin of the Adjacent Trigger and Droop
Section of the test fixture to Ch3 as shown in the following figure.
6.
Use the connectors to connect the A pin dongle from the Adjacent Trigger
and Droop section of the test fixture to one port of the Hub 5. Use the A pin
dongle from the Inrush Droop section of the test fixture to another port of
the Hub 5.
7.
Select the measurement and select the
application.
8.
Select OK after acquiring a waveform. Verify that it is a correct waveform.
command button to run the
Note: Use the standard USB cables to connect between the hubs. Keep the
Discharge switch in the Inrush Droop section in the ON position.
86
TDSUSB2 Universal Serial Bus Measurement Package
Application Examples
Figure 55: Equipment setup for Full Speed Signal Quality Measurement for upstream testing
Selecting and Configuring Measurements-Full Speed Signal Quality Tests for
Upstream Testing
Follow the steps to select measurements for Full Speed Signal Quality check:
1.
Select Measurements> Select > Full Speed tab.
2.
Select Signal Quality tests:
Eye Diagram Test
Signal Rate
Paired JK Jitter
Paired KJ Jitter
Consecutive Jitter
EOP Width
Cross-Over Voltage
Rise Time
Fall Time
TDSUSB2 Universal Serial Bus Measurement Package
87
Application Examples
3.
Select the Select All toggle button to select all the measurements
simultaneously. Click on any measurement button to deselect it.
Configuring the Measurement
Follow the steps to configure the selected measurement:
1.
From the application menu, select Measurement> Configure> Configure tab.
2.
Configure the following options:
Table 13. Configuring measurement
Option
Tier
Direction
Test Point
Set to
Tier6
Upstream
Far End
3.
Select the Source tab.
4.
Configure the following options:
Table 14. Configuring options
Option
Live/Ref
Qualify Channel
88
Set to
Single-ended Ch1, Ch2
Ch3
5.
Select
to acquire the data.
6.
The application automatically displays the Eye Diagram and the Waveform
Plot of the signal acquired from the unit under test.
TDSUSB2 Universal Serial Bus Measurement Package
Application Examples
Figure 56: Eye diagram for Full Speed Signal Quality Measurement for upstream testing
Figure 57: Waveform plot for Full Speed Signal Quality Measurement for upstream testing
TDSUSB2 Universal Serial Bus Measurement Package
89
Application Examples
Viewing Results-Full Speed Signal Quality Tests for Upstream Testing
To view the results of the tests, follow these steps:
1.
Run the application and from the application menu, select Results>
Summary.
2.
Click any of the test result buttons to get the details of that test.
3.
From the application menu, select Results> Details.
4.
Click the Eye Diagram option in Results Summary or Results Details to
view the Eye Diagram.
Figure 58: Eye diagram for Full Speed Signal Quality Measurement for upstream testing
5.
90
Click the Waveform Plot to view the annotated waveform plot.
TDSUSB2 Universal Serial Bus Measurement Package
Application Examples
Generating Reports-Full Speed Signal Quality Tests for Upstream Testing
To generate reports of the test results, follow these steps:
1.
From the application menu, select Utilities> Report Generator.
2.
Select any one of the Report Formats: Tektronix, Plug-Fest Specific or CSV
format.
3.
Select the Manual option to generate the report.
4.
Select the Generate button to display and view the selected report format.
Note: You can view the Plug-Fest specific report format as HTML pages and the
.csv format in Microsoft Excel.
Inrush Current Test for a Device
Specifying the Equipment-Full Speed Inrush Current Test
The following equipment is needed for Inrush Current Check on a Full Speed
Device:
Tektronix digital oscilloscope
TDSUSB2 application
TDSUSBF compliance test fixture
One TCP202 current probe
Typical Equipment Setup-Inrush Current Test
To set up the equipment for Inrush test, follow these steps:
1.
Use the dongles on the Inrush section of the test fixture to connect it to the
host system.
2.
Connect the current probe between the VBUS loop wire on the Inrush section
on the Test fixture and Ch1 of the oscilloscope.
3.
Configure the measurement and select the
application.
TDSUSB2 Universal Serial Bus Measurement Package
command button to run the
91
Application Examples
4.
Connect the unit under test to the A Receptacle of the Inrush section of the
test fixture and observe the Inrush current signal.
5.
Select OK after acquiring a waveform. Verify that it is a correct waveform.
Note: To avoid the triggering of Inrush signals and false inrush current by the
discharge switch, place the inrush discharge switch in the ON position and hotplug the unit under test (device). In case a valid Inrush signal is not acquired,
use the cursor mode in File> Preferences or set up the oscilloscope to get a valid
waveform.
Figure 59: Equipment setup for Full Speed Inrush Current test
92
TDSUSB2 Universal Serial Bus Measurement Package
Application Examples
Selecting and Configuring Measurements-Inrush Current Test
From the application menu, select Measurement> Select> Inrush Current.
Configuring the Measurement
Follow the steps to configure the selected measurement:
1.
From the application menu, select Measurement> Configure> Configure tab.
2.
Configure the following options:
Table 15. Configuring measurement
Option
Set to
Tier
Tier is always set to 1
VBUS
Enter voltage measured across VBUS manually
Device Type
Hot Plug Attach
3.
Select the Source tab.
4.
Configure the following options:
Table 16. Configuring options
Option
Live/Ref
5.
Set to
Channel acquiring the signal
Select Run.
Viewing Results-Inrush Current Test
To view the results of the tests, follow these steps:
1.
Run the application and from the application menu, and select Results>
Summary.
2.
Click any of the test result buttons to get the details of that test.
3.
From the application menu, select Results> Details.
4.
Click the Eye Diagram option in Results Summary or Results Details to
view the Eye Diagram.
5.
Click the Waveform Plot to view the annotated waveform plot.
TDSUSB2 Universal Serial Bus Measurement Package
93
Application Examples
Generating Reports-Inrush Current Test
To generate reports of the test results, follow these steps:
1.
From the application menu, select Utilities> Report Generator.
2.
Select any one of the Report formats – Tektronix Specific, Plug-Fest
Specific or CSV format.
Figure 60: Report in CSV format for Inrush Current test
3.
Select the manual report generation mode.
4.
Select the Generate button to display and view the selected report format.
Note: You can view the Plug-Fest specific report format as HTML pages and the
.csv format in Microsoft Excel.
94
TDSUSB2 Universal Serial Bus Measurement Package
Application Examples
Signal Quality Tests for High Speed Devices
Specifying the Equipment-Signal Quality Tests for High Speed Devices for
Upstream Testing
The following equipment is needed for signal quality tests on a High Speed
device for upstream testing:
Tektronix digital oscilloscope
TDSUSB2 application
TDSUSBF compliance test fixture
One P6248 differential probe
Typical Equipment Setup-Signal Quality Tests for High Speed Devices for Upstream
Testing
To set up the equipment for the High Speed Signal Quality test, follow these
steps:
1.
Set the S6 switch to the Init position.
2.
Connect the standard USB cable between the Device SQ Init port and the
host port.
3.
Connect the A receptacle from the Device SQ test port of the test fixture to
the unit under test (device).
4.
Configure the measurement and select the
application.
5.
Place the device in the test mode Test_Packet from the host controller.
6.
Set the test Init switch to the test position to isolate the unit under test while
maintaining the bus power.
7.
Select OK after acquiring a waveform. Verify that it is a correct waveform.
8.
If the signal is clipped, follow these steps to increase the vertical scale:
9.
In the oscilloscope menu, select Vertical>Vertical Setup to display the
Channel screen.
command button to run the
10. In the Scale field, increase the vertical scale values until the waveform is
completely displayed in the screen.
TDSUSB2 Universal Serial Bus Measurement Package
95
Application Examples
Figure 61: Equipment setup for Signal Quality Measurement for high-speed devices for upstream testing
Selecting and Configuring Measurements-Signal Quality Tests for High Speed
Devices for Upstream Testing
Follow the steps to select measurements for High Speed Signal Quality check:
1.
From the application menu, select Measurement> Select> High speed (tab).
2.
Select the tests:
Eye Diagram Test
Signal Rate
EOP Width
Rise/Fall Time
Monotonic Property
3.
96
Select the Select All toggle button to select all the measurements
simultaneously. Click on any measurement button to deselect it.
TDSUSB2 Universal Serial Bus Measurement Package
Application Examples
Configuring the Measurement
Follow the steps to configure the selected measurements:
1.
From the application menu, select Measurement> Configure> Configure tab.
2.
Configure the following options:
Table 17. Configuring measurements
Option
Set to
Tier
Tier1
Direction
Upstream
Test Point
Set the test point to Near End
3.
Configure the following options in the Source field:
Table 18. Configuring options
Option
Set to
Live/Ref
Differential
to acquire the data.
4.
Select
5.
The application automatically displays the Eye Diagram and Waveform Plot
of the signal acquired from the unit under test. This is possible if you enable
the automatic display of the eye diagram and the waveform plot in the
Preferences menu.
Viewing Results-Signal Quality Tests for High Speed Devices for Upstream Testing
To view the results of the tests, follow these steps:
1.
Run the application and from the application menu, select Results>
Summary.
TDSUSB2 Universal Serial Bus Measurement Package
97
Application Examples
Figure 62: Results summary for Signal Quality Measurement for high-speed devices for upstream testing
2.
Click any one of the test result buttons to get the details of that test.
3.
Click the Additional Information button to display the additional
information.
4.
From the application menu, select Results> Details.
Figure 63: Results details for Signal Quality Measurement for high-speed devices for upstream testing
5.
98
Click the Eye Diagram in Results Summary or the Results Details to view
the eye diagram.
TDSUSB2 Universal Serial Bus Measurement Package
Application Examples
Figure 64: Eye diagram for Signal Quality Measurement for high-speed devices for upstream
testing
6.
Click the Waveform Plot to view the waveform plot.
Figure 65: Waveform plot for Signal Quality Measurement for high-speed devices for
upstream testing
TDSUSB2 Universal Serial Bus Measurement Package
99
Application Examples
Generating Reports-Signal Quality Tests for High Speed Devices for Upstream
Testing
To generate reports of the test results, follow these steps:
1.
From the application menu, select Utilities> Report Generator.
2.
Select any one of the Report formats: Tektronix Specific, Plug-Fest Specific
or CSV format.
Figure 66: Report in Tektronix-Specific format for Signal Quality Measurement for high-speed devices for upstream testing
3.
Select the manual generation mode.
4.
Select the Generate button to display and view the selected report format.
Note: You can view the Plug-Fest specific report format as HTML pages and the
.csv format in Microsoft Excel.
100
TDSUSB2 Universal Serial Bus Measurement Package
Application Examples
Packet Parameter Measurement
Specifying the Equipment-Packet Parameter Measurement
The following equipment is needed for a Packet Parameter measurement:
Tektronix digital oscilloscope
TDSUSB2 application
TDSUSBF compliance test fixture
One P6248 differential probe
Host Controller (Host controller card with the test mode software on a
Microsoft Windows PC)
Typical Equipment Setup-Packet Parameter Measurement
Test Fixture Setup
To set up the test fixture, follow these steps:
1.
Set the S6 switch to the Init position.
2.
Connect the standard USB cable between the Device SQ Init port and the
host port.
3.
Connect the A receptacle (marked DUT) from the Device SQ test port to the
device under test.
4.
Apply the power to the test fixture.
5.
Apply the power to the DUT.
6.
Attach the differential probe near the device connector on the test fixture.
Set up the Oscilloscope for High Speed Host
1.
Select the measurement and run the application. The application
automatically sets the oscilloscope parameters (Horizontal, Vertical, and
Trigger) and displays the message, "Press OK when correct waveform is
acquired”. You can see the SOFs on the oscilloscope screen.
2.
If you are not able to acquire the waveform automatically, perform
Autosetup in the oscilloscope to display the SOF.
3.
If your DUT is EL_21, EL_23, EL_25, select the Single Step Set Feature
option from the HS electrical test tool.
TDSUSB2 Universal Serial Bus Measurement Package
101
Application Examples
4.
If your DUT is EL_22, select the Step button in the HS electrical test tool to
acquire the waveform as shown in the following figure:
Figure 67: Setting the trigger level
You can use the application to perform the following measurements:
1.
High Speed Host: EL_21, EL_23, EL_25 (EOPII-InterPacketI&II)
Complete procedures 1 and 2 to set up the test fixture and the
oscilloscope
The oscilloscope acquires and displays the waveform
"Press OK when correct waveform is acquired”
The application measures the synchronous bits (32) of the first and
second packets. This is EL_21
The application measures the EOP of the second packet (8bits). This is
EL_25. As the signal is differential, the EOP can be a positive or a
negative pulse
102
TDSUSB2 Universal Serial Bus Measurement Package
Application Examples
The application measures the inter-packet gap of the first two packets
(88-192 bits). This is EL_23 as shown in the following figure:
Figure 68: Measuring EL_23
2.
High Speed Host: EL_22 (InterPacketII&III)
Complete procedures 1 and 2 to set up the test fixture and the
oscilloscope
Select the Step button in the HS Electrical Test Tool
The oscilloscope acquires and displays the waveform as shown in the
following figure:
TDSUSB2 Universal Serial Bus Measurement Package
103
Application Examples
Figure 69: Acquiring the waveform
“Press OK when correct waveform is acquired”
Figure 70: Setting the trigger level
The application displays the waveform
104
TDSUSB2 Universal Serial Bus Measurement Package
Application Examples
The application measures the inter packet gap of the second and the
third packet (88-192 bits). This is EL_22
3.
High speed Host: EL_55 (SOF-EOP)
Complete the procedure 1
Run the test mode software and select the Host option
Select the Enumerate button in the HS electrical test tool to identify the
device connected to the host controller
Select the measurement and run the application. The application
automatically sets the oscilloscope parameters (Horizontal, Vertical,
and Trigger). You can see the SOFs on the oscilloscope screen
Figure 71: Viewing SOFs
TDSUSB2 Universal Serial Bus Measurement Package
105
Application Examples
To set up the oscilloscope for the High Speed Device EL_21, EL_23, EL_25
(Sync-EOPIII-InterPacketII&III) and a High Speed Device EL_22
(InterPacketI&II), follow these steps:
1.
Select the measurement and run the application. The application
automatically sets the oscilloscope parameters (Horizontal, Vertical, and
Trigger) and displays the message, Press OK when correct waveform is
acquired. You can see the SOFs on the oscilloscope screen.
2.
If you are not able to acquire the waveform automatically, perform
Autosetup in the oscilloscope to display the SOF.
3.
If your DUT is EL_21, EL_23, EL_25, select the Single Step Set Feature
option from the HS electrical test tool.
4.
If your DUT is EL_22, select the Step button in the HS electrical test tool to
acquire the waveform as shown in the following figure:
Figure 72: Setting the trigger level
You can use the application to perform the following measurements:
106
TDSUSB2 Universal Serial Bus Measurement Package
Application Examples
1.
High Speed Device:EL_21, EL_23, EL_25 (Sync-EOPIII-InterPacketII&III)
Complete procedures 1 and 3 to set up the oscilloscope
The oscilloscope acquires and displays the waveform
“Press OK when correct waveform is acquired”
The application measures the synchronous bits (32) of the third packet.
This is EL_21
The application measures the EOP of the third packet (8bits). This is
EL_25. As the signal is differential, the EOP can be a positive or a
negative pulse
TDSUSB2 Universal Serial Bus Measurement Package
107
Application Examples
The application measures the inter-packet gap between the second and
the third packets (88-192 bits). This is EL_23 as shown in the
following figure:
Figure 73: Measuring EL_23
108
TDSUSB2 Universal Serial Bus Measurement Package
Application Examples
2.
High Speed Device:EL_22 (InterPacketI&II)
Complete procedures 1 and 3 to set up the oscilloscope
Select the Step button in the HS electrical test tool to acquire the
waveform as shown in the following figure:
Figure 74: Acquiring the waveform
TDSUSB2 Universal Serial Bus Measurement Package
109
Application Examples
Figure 75: Setting the trigger level
The application measures the number of bits (88-192) between the
packets
110
TDSUSB2 Universal Serial Bus Measurement Package
Application Examples
Figure 76: Equipment setup for Packet Parameter measurement
Selecting and Configuring Measurement-Packet Parameter Measurement
Follow these steps to select measurements for Packet Parameter measurement:
1.
From the application menu, select Measurement> Select> High speed tab.
2.
Select the More button to display the following tests:
Receiver Sensitivity
Suspend
Reset from High Speed
Packet Parameter
Chirp
Resume
Reset Suspend
3.
Select Packet Parameter measurement.
TDSUSB2 Universal Serial Bus Measurement Package
111
Application Examples
Configuring the Measurement
Follow the steps to configure the selected measurements:
1.
From the application menu, select Measurement> Configure> Configure tab.
2.
Configure the following options:
Table 19. Configuring measurements
Option
Set to
Select DUT
Host, Device
3.
Configure the following options:
Table 20. Configuring options
Option
Set to
Host
EL_21, EL_23, EL_25
EL_22,EL_55
EL_21, EL_22, EL_25
EL_22
Ch1-Ch4, Ref1-Ref4
Device
Source
4.
Select
to acquire the data.
Viewing Results-Packet Parameter Measurement
To view the results of the tests, follow these steps:
1.
Run the application and from the application menu, select Results>
Summary.
Figure 77: Results summary of the Packet Parameter Measurement
112
TDSUSB2 Universal Serial Bus Measurement Package
Application Examples
2.
Click any of the test result buttons to get the details of that test.
Figure 78: Results summary and details
3.
From the application menu, select Results> Details.
Generating Reports-Packet Parameter Measurement
To generate reports of the test results, follow these steps:
1.
From the application menu, select Utilities> Report Generator.
2.
Select any one of the Report formats: Tektronix Specific, Plug-Fest
Specific, or CSV format.
TDSUSB2 Universal Serial Bus Measurement Package
113
Application Examples
Figure 79: Report in CSV format
Figure 80: Packet Parameter test results
114
TDSUSB2 Universal Serial Bus Measurement Package
Application Examples
Figure 81: Report in Tektronix Specific format
3.
Select the manual generation mode.
4.
Select the Generate button to display and view the selected report format.
Note: You can view the Plug-Fest specific report format as HTML pages and the
.csv format in Microsoft Excel.
Resume Measurement
Specifying the Equipment-Resume Measurement
The following equipment is needed for Resume measurement:
Tektronix digital oscilloscope
TDSUSB2 application
TDSUSBF compliance test fixture
Two P6245 or TAP1500 single-ended probes
TDSUSB2 Universal Serial Bus Measurement Package
115
Application Examples
Typical Equipment Setup-Resume Measurement
To set up the Device SQ in the test fixture for the Resume test, follow these
steps:
1.
Set the S5 switch to the Init position.
2.
Use a standard length of the USB cable with an A plug on one end and a B
plug on the other end. Connect one end of the cable to the B receptacle
socket connector on the Init port of the Device SQ section and the other end
to the host port A socket.
3.
Connect the A receptacle from the Device SQ test port to the unit under test
(device).
4.
Connect the single-ended probes of the oscilloscope to the D+ and D– pins.
5.
Select the Resume measurement in the application, configure its options and
select the Run button to run the application.
6.
Select OK after acquiring a waveform. Verify that it is a correct waveform.
Figure 82: Equipment setup for Resume Measurement
116
TDSUSB2 Universal Serial Bus Measurement Package
Application Examples
Selecting and Configuring Measurement-Resume
Follow these steps to select measurements for the Resume measurement:
1.
From the application menu, select Measurement> Select> High speed (tab).
2.
Select the More button to display the following tests:
Receiver Sensitivity
Suspend
Reset from High Speed
Packet Parameter
Chirp
Resume
Reset Suspend
3.
Select the Resume measurement.
Configuring the Measurement
Follow the steps to configure the selected measurements:
1.
From the application menu, select Measurement> Configure> Configure tab.
2.
Configure the following options:
Table 21. Configuring measurements
Option
Set to
Signal Direction
Upstream, Downstream
Source
D+ Ch1-Ch4, Ref1-Ref4
D– Ch1-Ch4, Ref1-Ref4
3.
Select
to acquire the data.
TDSUSB2 Universal Serial Bus Measurement Package
117
Application Examples
Viewing Results-Resume Measurement
To view the results of the tests, follow these steps:
1.
Run the application and from the application menu, and select Results>
Summary.
Figure 83: Results summary for Resume Measurement
2.
Click any of the test result buttons to get the details of that test.
3.
From the application menu, select Results> Details.
Figure 84: Results details for Resume Measurement
Generating Reports-Resume Measurement
To generate reports of the test results, follow these steps:
118
1.
From the application menu, select Utilities> Report Generator.
2.
Select any one of the Report formats: Tektronix Specific, Plug-Fest
Specific, or CSV format.
TDSUSB2 Universal Serial Bus Measurement Package
Application Examples
Figure 85: Report for Resume Measurement in CSV format
Figure 86: Report in Plug-Fest format for Resume Measurement
TDSUSB2 Universal Serial Bus Measurement Package
119
Application Examples
Figure 87: Report in Tektronix-Specific format for Resume Measurement
3.
Select the manual generation mode.
4.
Select the Generate button to display and view the selected report format.
Note: You can view the Plug-Fest specific report format as HTML pages and the
.csv format in Microsoft Excel.
120
TDSUSB2 Universal Serial Bus Measurement Package
Application Examples
Reset from Suspend Measurement
Specifying the Equipment-Reset from Suspend Measurement
The following equipment is needed for Reset from Suspend measurement:
Tektronix digital oscilloscope
TDSUSB2 application
TDSUSBF compliance test fixture
Two P6245 or TAP1500 single-ended probes
Typical Equipment Setup-Reset from Suspend Measurement
To set up the Device SQ in the test fixture for the Reset from Suspend test,
follow these steps:
1.
Set the S6 switch to the Init position.
2.
Use a standard length of the USB cable with an A plug on one end and a B
plug at the other end.
3.
Connect one end of the cable to the B receptacle socket connector on the Init
port of the Device SQ section and the other end to the host port A socket.
4.
Connect the A receptacle from the Device SQ test port to the unit under test
(device).
5.
Connect the single-ended probes of the oscilloscope to the D+ and D– pins.
6.
Select the Reset from Suspend measurement from the application, configure
its options, and select the Run button to run the application.
7.
Select OK after acquiring a waveform. Verify that it is a correct waveform.
TDSUSB2 Universal Serial Bus Measurement Package
121
Application Examples
Figure 88: Equipment for Reset from Suspend Measurement
Selecting and Configuring Measurement-Reset from Suspend Measurement
Follow these steps to select measurements for Reset from Suspend measurement:
1.
From the application menu, select Measurement> Select> High Speed tab.
2.
Select the More button to display the following tests:
Receiver Sensitivity
Suspend
Reset High Speed
Packet Parameter
Chirp
Resume
Reset From Suspend
3.
122
Select the Reset from Suspend measurement.
TDSUSB2 Universal Serial Bus Measurement Package
Application Examples
Configuring the Measurement
Follow the steps to configure the selected measurements:
1.
From the application menu, select Measurement> Configure> Configure tab.
2.
Configure the following options:
Table 22. Configuring measurements
Option
Set to
Signal Direction
Upstream
Source
D+ Ch1-Ch4, Ref1-Ref4
D– Ch1-Ch4, Ref1-Ref4
3.
Select
to acquire the data.
Viewing Results-Reset from Suspend Measurement
To view the results of the tests, follow these steps:
1.
Run the application and from the application menu, and select Results>
Summary.
Figure 89: Results summary for Reset from Suspend Measurement
2.
Click any of the test result buttons to get the details of that test.
TDSUSB2 Universal Serial Bus Measurement Package
123
Application Examples
Figure 90: Results summary for Reset from Suspend Measurement
3.
From the application menu, select Results> Details.
Figure 91: Results details for Reset from Suspend Measurement
Generating Reports-Reset from Suspend Measurement
To generate reports of the test results, follow these steps:
124
1.
From the application menu, select Utilities> Report Generator.
2.
Select any one of the Report formats: Tektronix Specific, Plug-Fest
Specific, or CSV format.
TDSUSB2 Universal Serial Bus Measurement Package
Application Examples
Figure 92: Report in CSV format for Reset from Suspend Measurement
Figure 93: Report in Plug-Fest format for Reset from Suspend Measurement
TDSUSB2 Universal Serial Bus Measurement Package
125
Application Examples
Figure 94: Report in Tektronix-Specific format for Reset from Suspend Measurement
3.
Select the manual generation mode.
4.
Select the Generate button to display and view the selected report format.
Note: You can view the Plug-Fest specific report format as HTML pages and the
.csv format in Microsoft Excel.
Suspend Measurement
Specifying the Equipment-Suspend Measurement
The following equipment is needed for Suspend measurement:
Tektronix digital oscilloscope
TDSUSB2 application
TDSUSBF compliance test fixture
Two P6245 or TAP1500 single-ended probes
126
TDSUSB2 Universal Serial Bus Measurement Package
Application Examples
Typical Equipment Setup-Suspend Measurement
To set up the Device SQ in the test fixture for the Suspend test, follow these
steps:
1.
Set the S6 switch to the Init position.
2.
Use a standard length of USB cable with an A plug on one end and a B plug
on the other end.
3.
Connect one end of the cable to the B receptacle socket connector on the Init
port of the Device SQ section and the other end to the host port A socket.
4.
Connect the A receptacle from the Device SQ test port to the unit under test
(device).
5.
Connect the single-ended probes of the oscilloscope to the D+ and D– pins.
6.
Select the Suspend measurement, configure its options and select the Run
button to run the application.
7.
Select OK after acquiring a waveform. Verify that it is a correct waveform.
Figure 95: Equipment setup for Suspend Measurement
TDSUSB2 Universal Serial Bus Measurement Package
127
Application Examples
Selecting and Configuring Measurement-Suspend Measurement
Follow these steps to select measurements for Suspend measurement:
1.
From the application menu, select Measurement> Select> High Speed tab.
2.
Select the More button to display the following tests:
Receiver Sensitivity
Suspend
Reset High Speed
Packet Parameter
Chirp
Resume
Reset From Suspend
3.
Select the Suspend measurement.
Configuring the Measurement
Follow the steps to configure the selected measurements:
1.
From the application menu, select Measurement> Configure> Configure tab.
2.
Configure the following signal direction options:
Table 23. Configuring measurements
Option
Set to
Signal Direction
Upstream, Downstream
Source
D+ Ch1-Ch4, Ref1-Ref4
D– Ch1-Ch4, Ref1-Ref4
3.
128
Select
to acquire the data.
TDSUSB2 Universal Serial Bus Measurement Package
Application Examples
Viewing Results-Suspend Measurement
To view the results of the tests, follow these steps:
1.
Run the application and from the application menu, and select Results>
Summary.
Figure 96: Results summary for Suspend Measurement
2.
Click any of the test result buttons to get the details of that test.
Figure 97: Results summary/details for Suspend Measurement
3.
From the application menu, select Results> Details.
TDSUSB2 Universal Serial Bus Measurement Package
129
Application Examples
Figure 98: Results details for Suspend Measurement
Generating Reports-Suspend Measurement
To generate reports of the test results, follow these steps:
1.
From the application menu, select Utilities> Report Generator.
2.
Select any one of the Report formats: Tektronix Specific, Plug-Fest
Specific, or CSV format.
Figure 99: Report in CSV format for Suspend Measurement
130
TDSUSB2 Universal Serial Bus Measurement Package
Application Examples
Figure 100: Report in Plug-Fest format for Suspend Measurement
Figure 101: Report in Tektronix-Specific format for Suspend Measurement
3.
Select the manual generation mode.
4.
Select the Generate button to display and view the selected report format
Note: You can view the Plug-Fest specific report format as HTML pages and the
.csv format in Microsoft Excel.
TDSUSB2 Universal Serial Bus Measurement Package
131
Application Examples
Reset from High Speed Measurement
Specifying the Equipment-Reset from High Speed Measurement
The following equipment is needed for Reset from High Speed measurement:
Tektronix digital oscilloscope
TDSUSB2 application
TDSUSBF compliance test fixture
Two P6245 or TAP1500 single-ended probes
Typical Equipment Setup-Reset from High Speed Measurement
To set up the Device SQ in the test fixture for the Reset from High Speed test,
follow these steps:
132
1.
Set the S6 switch to the Init position.
2.
Use a standard length of the USB cable with an A plug on one end and a B
plug on the other end.
3.
Connect one end of the cable to the B receptacle socket connector on the Init
port of the Device SQ section and the other end to the host port A socket.
4.
Connect the A receptacle from the Device SQ test port to the unit under test
(device).
5.
Connect the single-ended probes of the oscilloscope to the D+ and D– pins.
6.
Select the Reset from High Speed from the application, configure its options
and select the Run button to run the application.
7.
Select OK after acquiring a waveform. Verify that it is a correct waveform.
TDSUSB2 Universal Serial Bus Measurement Package
Application Examples
Figure 102: Equipment setup in Reset from High Speed Measurement
Selecting and Configuring Measurement-Reset from High Speed Measurement
Follow these steps to select measurements for Reset from High Speed
measurement:
1.
From the application menu, select Measurement> Select> High Speed tab.
2.
Select the More button to display the following tests:
Receiver Sensitivity
Suspend
Reset High Speed
Packet Parameter
Chirp
Resume
Reset From Suspend
3.
Select the Reset from High Speed measurement.
TDSUSB2 Universal Serial Bus Measurement Package
133
Application Examples
Configuring the Measurement
Follow these steps to configure the selected measurements:
1.
From the application menu, select Measurement> Configure> Configure tab.
2.
Configure the following options:
Table 24. Configuring measurements
Option
Set to
Signal Direction
Upstream
Source
D+ Ch1-Ch4, Ref1-Ref4
D– Ch1-Ch4, Ref1-Ref4
3.
Select
to acquire the data.
Viewing Results-Reset from High Speed Measurement
To view the results of the tests, follow these steps:
1.
Run the application and from the application menu, and select Results>
Summary.
Figure 103: Results summary for Reset from High Speed Measurement
2.
134
Click any of the test result buttons to get the details of that test.
TDSUSB2 Universal Serial Bus Measurement Package
Application Examples
Figure 104: Results summary/detail for Reset from High Speed Measurement
3.
From the application menu, select Results> Details.
Figure 105: Results details for Reset from High Speed Measurement
TDSUSB2 Universal Serial Bus Measurement Package
135
Application Examples
Generating Reports-Reset from High Speed Measurement
To generate reports of the test results, follow these steps:
1.
From the application menu, select Utilities> Report Generator.
2.
Select any one of the Report formats: Tektronix Specific, Plug-Fest Specific
or CSV format.
Figure 106: Report in CSV format for Reset from High Speed Measurement
136
TDSUSB2 Universal Serial Bus Measurement Package
Application Examples
Figure 107: Report in Plug-Fest format for Reset from High Speed Measurement
Figure 108: Report in Tektronix-Specific format for Reset from High Speed Measurement
3.
Select the manual generation mode.
4.
Select the Generate button to display and view the selected report format.
Note: You can view the Plug-Fest specific report format as HTML pages and the
.csv format in Microsoft Excel.
TDSUSB2 Universal Serial Bus Measurement Package
137
Application Examples
Chirp Measurement
Specifying the Equipment-Chirp
The following equipment is needed to test Chirp measurement:
Tektronix digital oscilloscope
TDSUSB2 application
TDSUSBF compliance test fixture
Two P6245 or TAP1500 single-ended probes
Typical Equipment Setup-Chirp
The section used for this device test is Device SQ in the test fixture. To set up the
equipment for the Chirp test, follow these steps:
138
1.
Set the S6 switch to the Init position.
2.
Use a standard length of the USB cable with an A plug on one end and a B
plug on the other end.
3.
Connect one end of the cable to the B receptacle socket connector on the
Init port of Device SQ section and the other end to the host port A socket.
4.
Connect the A receptacle from the Device SQ test port to the unit under test
(device).
5.
Connect the single-ended probes of the oscilloscope to the D+ and D– pins.
6.
Select the measurement and select the
7.
Run the HS Electrical Test Tool on the connected host. Enumerate the unit
under test (device) and observe the chirp signal on the oscilloscope. Rather
than enumerating the device, an alternative method to generate the chirp
signal is to disconnect and reconnect the unit under test (device) to the port.
button to run the application.
TDSUSB2 Universal Serial Bus Measurement Package
Application Examples
Figure 109: Chirp signal
8.
Select OK after acquiring a waveform. Verify that it is a correct waveform.
Note: To avoid false triggering for the chirp signals while operating the test
fixture, it is recommended that you place the switch in the Init position and
connect the unit under test. This disables the switch bounce to the trigger.
TDSUSB2 Universal Serial Bus Measurement Package
139
Application Examples
Figure 110: Equipment setup for Chirp Measurement
Selecting and Configuring Measurement-Chirp
Follow these steps to select measurements for Chirp measurement:
1.
From the application menu, select Measurement> Select> High Speed tab.
2.
Select the More button to display the following tests:
Receiver Sensitivity
Suspend
Reset from High Speed
Packet Parameter
Chirp
Resume
Reset Suspend
3.
140
Select Chirp measurement.
TDSUSB2 Universal Serial Bus Measurement Package
Application Examples
Configuring the Measurement
Follow the steps to configure the selected measurements:
1.
From the application menu, select Measurement> Configure> Configure tab.
2.
Configure the following options:
Table 25. Configuring measurements
Option
Set to
Select DUT
Host, Device
Host
Device
EL_33, EL_34
EL_35
EL_28, EL_29, EL_31
Source
Ch1-Ch4, Ref1-Ref4
3.
Select
to acquire the data.
Viewing Results-Chirp
To view the results of the tests, follow these steps:
1.
Run the application and from the application menu, and select Results>
Summary. The following figures show the result of a Chirp Device
measurement:
Figure 111: Results summary for Chirp Measurement
2.
Click any of the test result buttons to get the details of that test.
TDSUSB2 Universal Serial Bus Measurement Package
141
Application Examples
Figure 112: Results summary/details for Chirp Measurement
3.
From the application menu, select Results> Details.
Figure 113: Results details for Chirp Measurement
4.
Click here to view the results of the Chirp Host measurement.
Figure 114: Results summary for Chirp Host Measurement
142
TDSUSB2 Universal Serial Bus Measurement Package
Application Examples
Figure 115: Results detail for Chirp Host Measurement
Figure 116: Results details for Chirp Host Measurement
Generating Reports-Chirp
To generate reports of the test results, follow these steps:
1.
From the application menu, select Utilities> Report Generator.
2.
Select any one of the Report formats: Tektronix Specific, Plug-Fest Specific
or CSV format.
TDSUSB2 Universal Serial Bus Measurement Package
143
Application Examples
Figure 117: Report in SCV format for Chirp Measurement
Figure 118: Report in Plug-Fest format for Chirp Measurement
144
TDSUSB2 Universal Serial Bus Measurement Package
Application Examples
Figure 119: Report in Tektronix-Specific format for Chirp Measurement
3.
Select the manual generation mode.
4.
Select the Generate button to display and view the selected report format.
Note: You can view the Plug-Fest specific report format as HTML pages and the
.csv format in Microsoft Excel.
TDSUSB2 Universal Serial Bus Measurement Package
145
Application Examples
146
TDSUSB2 Universal Serial Bus Measurement Package
R-GPIB Commands
About the R-GPIB Program
An example of an R-GPIB program that can execute the TDSUSB2 measurement
is included with the application. The oscilloscope hard disk and optional
applications compact disc both contain the file, TDSUSB2_rgpib.c. On the hard
drive, the file resides in the C:\Program Files\TekApplications\tdsusb2 directory.
This example shows how an R-GPIB program executes the application to do the
following tasks:
1.
Start up the application
2.
Recall High Speed Signal Quality setup
3.
Run the measurement
4.
Generate the reports
5.
Exit the application
R-GPIB Reference Materials
To use the R-GPIB commands with your oscilloscope, refer to the following
materials:
The TDSUSB2_rgpib.c file on the oscilloscope hard drive (located in the
(located in the C:\Program Files\TekApplications\tdsusb2 directory) and
optional application compact disc for an example of an R-GPIB command
that can execute the application.
The R-GPIB Program Example section for guidelines to use while designing
the R-GPIB program.
The programmer information is in the online help of your oscilloscope.
TDSUSB2 Universal Serial Bus Measurement Package
147
R-GPIB Commands
Introduction to the R-GPIB Commands
With the knowledge of R-GPIB command syntax, you can design an R-GPIB
program to do the following tasks:
Start the TDSUSB2 application
Recognize an active application with R-GPIB protocol
Program and read the application setup parameters
Sequence measurements
Generate reports
Guidelines to R-GPIB Programming
The application includes an example of an R-GPIB program for your reference as
a program example. Your R-GPIB program should comply with the following
guidelines:
The application startup must complete before sending additional R-GPIB
commands to the application (see example).
To generate reports, first check whether the sequencer state is "Ready".
Appropriate delay (for example: 2 secs) should be maintained between
commands.
The status variable should be checked to ensure that an error has not
occurred because of a measurement command problem.
R-GPIB event queue needs to be monitored. Make sure the event queue is
clear before sending the next R-GPIB command to prevent event queue
overflow.
Commands are case and space sensitive. Your program will not operate
correctly if you do not follow the capitalization and spacing precisely.
148
TDSUSB2 Universal Serial Bus Measurement Package
R-GPIB Commands
Launching the Application Using R-GPIB
You must manually set up the oscilloscope to launch the application. To start the
TDSUSB2 application, you must send the oscilloscope the following R-GPIB
command:
application:activate"USB2.0 Test Package"
The application uses the R-GPIB VARIABLE: VALUE command with
arguments to execute some features. The set of R-GPIB commands includes the
variable names and variable values necessary to select, configure, and run the
measurements and to generate reports in the R-GPIB program.
You can select and configure the measurements that you want to use with your
R-GPIB program.
Note: When using R-GPIB commands, the reports are saved in
C:\TekApplications\tdsusb2\reports directory.
TDSUSB2 Universal Serial Bus Measurement Package
149
R-GPIB Commands
Variable: Value Command
Variable: Value Command
Description
This command accepts strings arguments for a control or data variable and a
value to which to set the argument.
Syntax
VARIABLE:VALUE “<variable name>”,“<variable value>”
The arguments <variable name> and <variable value> are required in the order
indicated.
VARIABLE:VALUE? “<variable name>” for query.
Note: Commands are case and space sensitive. Your program will not operate
correctly if you do not follow the capitalization and spacing precisely.
Variable: Value TDSUSB2 Command Arguments and Queries- Application
Table 26. Variable: Value TDSUSB2 Command Arguments and Queries- Application
Variable Name
Valid values
Function
Querying Form
exit
Setting the value will terminate Returns the name of the currently
the running application
running application
Terminating the application
application
Variable: Value TDSUSB2 Command Arguments and Queries- Sequencer
Table 27. Variable: Value TDSUSB2 Command Arguments and Queries- Sequencer
Variable Name
Valid values
Function
Querying Form
{Sequencing}-for Sequencing
{Ready, Sequencing}-for Query
Sets the sequencer state
Returns the sequencer state
Running measurements
seqeuncerState
150
TDSUSB2 Universal Serial Bus Measurement Package
R-GPIB Commands
Variable: Value TDSUSB2 Command Arguments and Queries- Save/Recall
Table 28. Variable: Value TDSUSB2 Command Arguments and Queries- Save/Recall
Variable Name
Valid values
Function
setup
{Default, Recall, Save}
Sets the save/recall/default
action
recallName
It is recommended to have any
string of length 1-8 chars,
comprising of A-Z, a-z, 0-9
saveName
It is recommended to have any
string of length 1-8 chars,
comprising of A-Z, a-z, 0-9
Querying Form
The default value for this variable
is an empty string. The variable is
set to the selected value
momentarily and after completion
of the task, it returns to its default
value
Sets the setup recall file name Returns the setup recall file name
Sets the setup save file name
Returns the setup save file name
Variable: Value TDSUSB2 Command Arguments and Queries- Report Generation
Table 29. Variable: Value TDSUSB2 Command Arguments and Queries- Report Generation
Variable Name
Valid values
Function
Querying Form
reportFormat
{tek, plug-fest, csv}
Returns the current report format
reportName
Sets the Report format to
Tektronix specific, plug-fest
specific or CSV format
Sets the report file name
It is recommended to have any
string comprising of A-Z, a-z, 0-9 in
the manual mode. Do not add file
name extension to the report name
It is recommended to have any
Sets the directory to save the
string comprising of A-Z, a-z, 0-9 report
and '\'. End the directory name with
'\'
{auto, manual}
Sets the report generation
mode to automatic or manual
reportDirectory
reportMode
reportGenerate
generate
TDSUSB2 Universal Serial Bus Measurement Package
Returns the report file name
Returns the directory to save
report
Returns the report generation
mode
Generates the report in the
manual mode
151
R-GPIB Commands
Variable: Value TDSUSB2 Command Arguments and Queries- Results
Variable Name
Valid values
Function
Querying Form
resultFor
{eye, sigrt, jk, kj, con, eop,
eopbit, cross, rt, ft, mon, in,
pphost, pphost2, pphost3,
ppdevice1,ppdevice2, resume,
suspend, rfr, rfs,
chirphost1, chirphost2,
chirpdevice1}
{Result}
Sets the result variables with
appropriate result values for
that particular test
After being set to a measurement
value, the query returns "Busy"
until the results are refreshed.
Returns the selected
measurement for result querying
reset
Clears the active
measurements results
Statistical results-Basic statistics applicable for all SQC measurements
measUnits
Units string for the measurement
(for example “s” - seconds is the
unit for Period)
Maximum value of the
measurement
Mean value of the result
max
mean
min
Minimum value of the
measurement
pkpk
Peak-to-peak value of the
measurement
RMS value of the result
rms
population
stdDev
status
{Pass, Fail, Conditional Pass}
Population (number of)
measurements used to calculate
the current statistics
Standard deviation measurement
set
Returns Pass/Fail result for the
selected SQC measurement
Pass/Fail Status
eyeStatus
{Pass, Fail, Conditional Pass}
sqcOverallStatus
{Pass, Fail, Conditional Pass}
Returns the eye diagram
measurement
Returns the overall SQC
measurement result
Inrush
inCharge
Returns the inrush charge value
inCap
Returns the inrush capacitance
value
Returns the Pass/Fail status of
inrush measurement
inStatus
152
{Pass, Fail, Conditional Pass}
TDSUSB2 Universal Serial Bus Measurement Package
R-GPIB Commands
Variable Name
Valid values
Function
Querying Form
Packet Parameter
ppOverallStatus
{Pass, Fail}
ppHostEL21Status
{Pass, Fail}
ppHostEL23
ppHostEL23Status
{Pass, Fail}
ppHostEL25
ppHostEL25Status
{Pass, Fail}
ppHostEL22
ppHostEL55
ppDeviceEL21Status
{Pass, Fail}
Returns the sync field Pass/Fail
status
Returns the inter packet gap
{Pass, Fail}
Returns the inter packet gap
Pass/Fail status
Returns the EOP width in bits
{Pass, Fail}
Returns the EOP width Pass/Fail
status
ppDeviceEL22
ppDeviceEL22Status
ppDeviceEL25
ppDeviceEL25Status
Returns the overall Pass/Fail
status of the selected packet
parameter measurement
result
Returns the Sync Field Pass/Fail
status
Returns the inter packet gap
between first two packets
Returns the inter packet gap
Pass/Fail status
Returns the EOP width in bits of
the second packet
Returns the EOP width Pass/Fail
status
Returns the inter packet gap
between second and third packet
Returns the EOP width in bits
Reset from Suspend/ Reset from High-Speed/Suspend/ Resume
resetSusTime
resetSusStatus
{Pass, Fail}
Returns the reset from suspend
time
Returns the reset from suspend
status
Returns the suspend time
{Pass, Fail}
Returns the suspend status
susTime
susStatus
resetHSTime
resetHSStatus
{Pass, Fail}
resumeTime
resumeAmp
resumeStatus
{Pass, Fail}
TDSUSB2 Universal Serial Bus Measurement Package
Returns the reset from high
speed time
Returns the reset from high
speed status
Returns the resume time for
upstream
Returns the resume amplitude for
downstream
Returns the resume status
153
R-GPIB Commands
Variable Name
Valid values
Function
Querying Form
Chirp
chirpHostEL33
chirpHostEL33Status
Returns the host response time
{Pass, Fail}
Returns the host response time
status
chirpHostKEL34
Returns the host K duration
chirpHostJEL34
Returns the host J duration
ChirpHostEL34Status
{Pass, Fail}
Returns the host KJ duration
status
Returns the SOF time
{Pass, Fail}
Returns the SOF time status
chirpHostEL35
chirpHostEL35Status
chirpDeviceEL28
chirpDeviceEL28Status
Returns the device reset time
{Pass, Fail}
Returns the device reset time
status
Returns the device K duration
time
Returns the device K duration
time status
chirpDeviceEL29
chirpDeviceEL29Status
{Pass, Fail}
chirpOverallStatus
{Pass, Fail}
Returns the overall Pass/Fail
status of the selected chirp
measurement result
Additional Result
additionalResult
Returns the additional result if
applicable for selected
measurement using resultFor
ResultFor Commands
The resultFor commands are of the format:
variable:value “<variable_name>”,“<value>”
Get or query commands are of the format:
variable:value? “<variable_name>”
Ensure that the order mentioned below is followed for the command script to
work correctly. A sample script is listed here. It recalls a signal quality test
setting, selects the test, runs the test, and queries the result. For example:
Measured mean value for rise time test.
It is recommended to give a delay of at least one second between commands.
154
TDSUSB2 Universal Serial Bus Measurement Package
R-GPIB Commands
variable:value “recallName”, “sqc”
variable:value “setup”,”Recall”
variable:value “sequencerState”, “Sequencing”
…..keep polling until it returns a “Ready”
variable:value? “sequencerState”
“Ready”
variable:value “resultFor”, “rt”
variable:value? “resultFor”
“Busy”
…..keep polling until it returns a “rt”, the current test for which the result is
queried
variable:value? “resultFor”
“rt”
variable:value? “mean”
1.88n
TDSUSB2 Universal Serial Bus Measurement Package
155
R-GPIB Commands
Sample Program
/* TDSUSB2
* This is a reference program to illustrate how to communicate with TDSUSB2
* using Remote GPIB facilities.
Typically, the application does the following steps:
1.
Start up the application
2.
Recall Full Speed Signal Quality setup
3.
Run the measurement
4.
Generate the report
5.
Exit the application
For the current program, we will recall a setup file named as sqcsetup.
You can save setup files according to your own needs using the GUI based
interface of the application.
*/
#include <windows.h>
#include <stdio.h>
#include <stdlib.h>
#include "decl-32.h"
/* Forward Declarations */
int start_application(int scope);
int exit_application(int scope);
int do_single_test (int scope);
int recall_setup( int scope, char *filename);
156
TDSUSB2 Universal Serial Bus Measurement Package
R-GPIB Commands
void display_results(int scope);
/* parameters needed to access the device driver
handler */
#define BDINDEX 0
// Board Index
#define PRIMARY_ADDR_OF_DMM 1
address of device
// Primary
#define NO_SECONDARY_ADDR 0
address of device
// Secondary
#define TIMEOUT T10s
= 10 seconds
// Timeout value
#define EOTMODE 1
message
#define EOSMODE 0
// Enable the END
// Disable the EOS
mode
char
ErrorMnemonic[21][5] = {"EDVR", "ECIC",
"ENOL", "EADR", "EARG",
"ESAC", "EABO", "ENEB", "EDMA", "",
"EOIP", "ECAP", "EFSO", "", "EBUS",
"ESTB", "ESRQ", "", "", "", "ETAB"};
/*
* After each GPIB call, the application checks
whether the call
* succeeded. If an NI-488.2 call fails, the GPIB
driver sets the
* corresponding bit in the global status
variable. If the call
* failed, this procedure prints an error message,
takes
* the device offline and exits.
*/
TDSUSB2 Universal Serial Bus Measurement Package
157
R-GPIB Commands
void GPIBCleanup(int ud, char* ErrorMsg)
{
printf("Error : %s\nibsta = 0x%x iberr =
%d (%s)\n",
ErrorMsg, ibsta, iberr,
ErrorMnemonic[iberr]);
if (ud != -1)
{
printf("Cleanup: Taking device
offline\n");
ibonl(ud, 0);
}
exit(0);
}
int start_application( int scope )
{
char write_buffer[100];
char read_buffer[100];
char app_name[] = "\"USB2.0 Test
Package\"\n";
int status, timer;
/* Start the TDSUSB2 application */
sprintf(write_buffer, "%s",
"Application:activate \"USB2.0 Test Package\"");
status = ibwrt(scope, write_buffer,
strlen(write_buffer));
if (ibsta & ERR)
{
GPIBCleanup(scope, "Unable to start
the application");
return 0;
}
158
TDSUSB2 Universal Serial Bus Measurement Package
R-GPIB Commands
timer = 1;
while (1)
{
/* Check whether application has
started */
sprintf(write_buffer, "%s",
"Variable:value? \"application\"");
status = ibwrt(scope, write_buffer,
strlen(write_buffer));
status = ibrd(scope, read_buffer,
sizeof(read_buffer));
read_buffer[ibcnt] = '\0';
if (strcmp(app_name, read_buffer) ==
0)
{
return 1;
}
timer++;
if (timer > 60)
{
return 0;
}
Sleep(1000);
}
return 1;
}
int exit_application(int scope)
{
char write_buffer[100];
printf("Exit Application ...\n");
sprintf(write_buffer, "%s",
"Variable:value \"application\",\"exit\"");
TDSUSB2 Universal Serial Bus Measurement Package
159
R-GPIB Commands
ibwrt(scope, write_buffer, strlen(write_buffer));
return 1;
}
int recall_setup(int scope, char* filename)
{
char write_buffer[100];
int status;
/* Set Recall file name */
sprintf(write_buffer, "%s%s%s",
"Variable:value \"recallName\",\"", filename, "\"");
status = ibwrt(scope, write_buffer,
strlen(write_buffer));
if (ibsta & ERR)
{
GPIBCleanup(scope, "Unable to
communicate with Scope");
return 0;
}
Sleep(1000);
/* Recall setup */
sprintf(write_buffer, "%s",
"Variable:value \"setup\",\"Recall\"");
status = ibwrt(scope, write_buffer,
strlen(write_buffer));
if (ibsta & ERR)
{
GPIBCleanup(scope, "Unable to
communicate with Scope");
return 0;
}
160
TDSUSB2 Universal Serial Bus Measurement Package
R-GPIB Commands
return 1;
}
int run_test (int scope)
{
char write_buffer[100];
char read_buffer[100];
int timer;
sprintf(write_buffer, "%s",
"Variable:value \"sequencerState\",\"Sequencing\"");
ibwrt(scope, write_buffer,
strlen(write_buffer));
printf("Executing Test...\n");
Sleep(100);
/* Wait for application to come to Ready
State */
timer = 1;
while (1)
{
timer++;
if (timer > 90)
{
printf("******Test Time Out
******\n");
return 0;
}
sprintf(write_buffer, "%s",
"Variable:value? \"sequencerState\"");
ibwrt(scope, write_buffer,
strlen(write_buffer));
ibrd(scope, read_buffer, 99);
if (ibsta & ERR)
{
TDSUSB2 Universal Serial Bus Measurement Package
161
R-GPIB Commands
GPIBCleanup(scope, "Unable to
write to device");
}
read_buffer[ibcnt] = '\0';
if
(strcmp(read_buffer,"\"Ready\"\n") == 0)
{
printf("Test Complete
...\n");
return 1;
}
Sleep(1000);
}
}
void report_generate(int scope,char*
reportfilename)
{
char write_buffer[100];
char read_buffer[100];
int status;
/* Set report file name */
sprintf(write_buffer, "Variable:value
\"reportName\",\"%s\"",reportfilename);
printf("%s",write_buffer);
status = ibwrt(scope, write_buffer,
strlen(write_buffer));
if (ibsta & ERR)
{
GPIBCleanup(scope, "Unable to
communicate with Scope");
162
TDSUSB2 Universal Serial Bus Measurement Package
R-GPIB Commands
return;
}
Sleep(1000);
/*Generate the report*/
sprintf(write_buffer, "variable:value
\"reportGenerate\",\"generate\"");
printf("%s",write_buffer);
ibwrt(scope, write_buffer,
strlen(write_buffer));
if (ibsta & ERR)
{
GPIBCleanup(scope, "Unable to start
the application");
return;
}
Sleep(2000);
printf("\t Full Speed Signal Quality
Result stored\n");
}
void main()
{
int Dev;
char write_buffer[100];
int status;
Dev = ibdev (BDINDEX, PRIMARY_ADDR_OF_DMM,
NO_SECONDARY_ADDR,
TIMEOUT, EOTMODE, EOSMODE);
if (ibsta & ERR)
{
TDSUSB2 Universal Serial Bus Measurement Package
163
R-GPIB Commands
GPIBCleanup(Dev, "Unable to open
device");
}
else
{
printf("My device id - %i", Dev);
}
Sleep(1000);
sprintf(write_buffer, "%s", "header off");
status = ibwrt(Dev, write_buffer,
strlen(write_buffer));
if (start_application(Dev))
{
printf("\nApplication
started....\n");
}
Sleep(10000);
recall_setup(Dev,"sqcsetup");
Sleep(2000);
run_test(Dev);
Sleep(4000);
report_generate(Dev,"fs_sqc");
Sleep(2000);
exit_application(Dev);
/* leave the device back elegantly */
printf("Cleanup: Taking device
offline\n");
ibonl(Dev, 0);
}
164
TDSUSB2 Universal Serial Bus Measurement Package
Reference
Shortcut Keys
This table lists the shortcut keys that you can use for different tasks.
Table 30. Shortcut keys
Action
Shortcut Keys
File Menu
Recall Default
AltF+D
Recall
AltF+R
Save
AltF+S
Preferences
AltF+P
Recently Recalled
AltF+C
Recently Saved
AltF+A
Minimize
AltF+N
Exit
AltF+X
Measurements Menu
Select
AltM+S
Configure
AltM+C
Results Menu
Summary
AltR+S
Details
AltR+D
Utilities Menu
Deskew
AltU+K
Report Generator
AltU+R
TSV File Generator
AltU+T
Help Menu
Help Topics
AltH+T
About TDSUSB2
AltH+A
TDSUSB2 Universal Serial Bus Measurement Package
165
Reference
Note: Use the Alt key with only the first keystroke. Release the Alt key before you
press the final key. For example, to use the shortcut key AltF+D, press the Alt
and F keys together. Release the keys and then press D.
USB2.0 Specifications
This section gives the USB2.0 values for measurements for Low, Full and High
speed signals.
Low Speed
Table 31. Low Speed measurements
USB Limits
User Configure Limits
References*
Measurements
Max
Min
Max. range
Min. range
Cross-Over
Voltage
2.0 V
1.3 V
1.6-2.4 V
1.04-1.56 V Chapter 7, Table 7-7,
Section 7.1.2.1
Signal Rate
1.5225 Mbps 1.4775 Mbps 1.5675-1.881
Mbps
Consecutive
Jitter
<25 ns
N/A
Paired Jitter
<10 ns
Rise Time
Chapter 7, 7.1.11
20-30 ns
1.1461.4325
Mbps
N/A
N/A
8-12 ns
N/A
Chapter 7, 7.1.13.1
300 ns
75 ns
240-360 ns
60-90 ns
Chapter 7, 7.1.2.1
Fall Time
300 ns
75 ns
1240-360 ns
60-90 ns
Chapter 7, 7.1.2.1
EOP Width
1.5e-6 s
1.25e-6 s
1.2-1.8
1-1.5 ns
Chapter 7, 7.1.13.2.1
Chapter 7, 7.1.13.1
*The References Section refers to the chapter numbers in the ‘Universal Serial
Bus Specifications Revision 2.0-2000’.
Note 1: Results within the USB limit lead to a PASS condition. Results within the
waiver limits, but outside USB limits lead to a Conditional PASS. Results within
the user configured limits lead to PASS. When user configure limits is selected,
you can view only the result details.
Note 2:Testing at Tier 6 leads to a PASS. Testing at Tier 5 and 4 leads to a
Conditional PASS that is a Pass within the waiver limits. Testing at Tier 1 to 3
leads to a FAIL.
166
TDSUSB2 Universal Serial Bus Measurement Package
Reference
Full Speed
Table 32. Full Speed measurements
USB Limits
User Configure Limits
References*
Measure-ments Max
Min
Max range
Min range
Cross-Over
Voltage
Signal Rate
2.0 V
1.3 V
1.6-2.4 V
12.03 Mbps
Consecutive Jitter <2 ns
11.97 Mbps 12.15 to 14.8
Mbps
N/A
1.6-2.4 ns
1.04-1.56 V Chapter 7, Table 7-7,
Section 7.1.2.1
9.48-11.85 Chapter 7,7.1.13.1
Mbps
N/A
Chapter 7, 7.1.13.1
Paired Jitter
<1 ns
N/A
.8-1.2 ns
N/A
Chapter 7, 7.1.13.1
Rise Time
20 ns
4 ns
16-24 ns
3.2-4.8 ns
Chapter 7, 7.1.2.1
Fall Time
20 ns
4 ns
16-24 ns
3.2-4.8 ns
Chapter 7, 7.1. 2.1
EOP Width
175 ns
160 ns
140-210 ns
128-192 ns
Chapter 7, 7.1.13.2.1
References Section refers to the chapter numbers in the ‘Universal Serial Bus
Specifications Revision 2.0-2000’.
Note 1: Results within the USB limits lead to a PASS. Results within the waiver
limits, but outside USB limits lead to a Conditional PASS. Results within the user
configured limits lead to a PASS result. If you select configured limits, you can
view only the result details.
Note 2: Testing at Tier 6 leads to a PASS. Testing at Tier 5 and 4 leads to a
Conditional PASS that is a Pass within the waiver limits. Testing at Tier 1 to 3
leads to a FAIL.
Note 3: The application uses the USB signal rate limits of Low Speed or Full
Speed devices that are not capable of High Speed. The same limits for Low Speed
or Full Speed that are High Speed capable to decide the result - PASS or FAIL.
TDSUSB2 Universal Serial Bus Measurement Package
167
Reference
High Speed
Table 33. High Speed measurements
USB Limits
User Configure Limits
MeasureMax
ments
Signal Rate 480.24
Mbps
Rise Time
N/A
Min
Fall Time
500ps
N/A
400-600 ps
YES
Monotonicity 0
0
N/A
N/A
N/A
EOP Width
7.5 bit times
8.5-10.2 bit
times
6-7.5 bit times YES
N/A
8.5 bit
times
Max range
Min range
80% to
120%
USB
Pass/Fail
YES
479.76 Mbps 480.264383.7888576.3168 Mbps 479.736 Mbps
500ps
N/A
400-600 ps YES
References*
Chapter 7,7.1.11
Chapter 7, 7.1.22
Chapter 7, 7.1.
22
N/A
Chapter 7,
7.1.13.2.2
*The References Section refers to the chapter numbers in the ‘Universal Serial
Bus Specifications Revision 2.0-2000’.
Note 1: Results within the USB limits lead to a PASS. Results within the waiver
limits, but outside USB limits lead to a Conditional PASS. Results within the user
configured limits lead to PASS. If you select your configured limits, you can view
only the result details.
Note 2:Whenever a high speed device is used in the low speed and full speed
mode, the USB limits of the low speed and full speed devices are used to decide
PASS or FAIL.
Note 3: USB2.0 specifications have not specified the maximum rise time and fall
time. If rise or fall times are greater than 0.5 bits, then a wrong, (long) rise or
fall time will be captured as an eye diagram failure.
Inrush Current Check
Table 34. Inrush Current measurements
USB Limits
User Configure Limits
80% to 120%
USB Pass/Fail
Measurements Max
Hot Plug
Attach
VBUS Default 5.25V
Volts 5.15
Min
Max range
4.4 V
N/A
N/A
YES
Chapter 7, 7.22
Inrush Level
N/A
80-120 mA
N/A
YES
Current
Draw.pdf
168
100 mA
Min range
References*
TDSUSB2 Universal Serial Bus Measurement Package
Reference
Droop Test
Table 35. Inrush Current measurements
Measurements
USB Limits
User Configure Limits
80% to 120% References*
Max
Max range
Min range
USB
Pass/Fail
N/A
YES
Min
Droop Compliance <330 mV 70.0 mV 264-396 mV
Voltage
Chapter 7, 7.2.4.1
*References Section refers to the chapter numbers in ‘Universal Serial Bus
Specifications Revision
2.0-2000’.
Error Codes and Warnings
This section gives a list of error codes that the application displays and their
descriptions.
Table 36. Error Code descriptions
Error Codes
Error Message
Description
E101
Error importing waveform from
the oscilloscope
The application is trying to import the
waveform(s) from the selected
source(s), but is not able to import the
waveform to the application
E102
Error in accessing .tsv file
E103
Mismatch in the .tsv file format
E104
Record length is more. Set the
record length to less than
XXXXXX data points
Possible Solutions
Sequencer Errors
Make sure that the application settings are
the same as the selected oscilloscope
channels. Check the probe connections
and the compliance test fixture connections
for the live signal(s)
This error occurs whenever .tsv file is Check whether the file name extension is
selected as the source file and this file .tsv. The selected .tsv source file may be
is not a valid .tsv file
corrupted. Try running the application with
a different tsv file
Try generating the .tsv file using .tsv File
The .tsv file should match the file
Generator utility that is available in the
structure mentioned in Reference
section of this help. For Low and Full Utilities menu
Speed signals, each data record
should have three values (Timestamp,
D+, D–). For a High Speed .tsv file, the
file should have two columns (Time,
Diff). Use a valid file
The record length is too high for the Click the "Horiz" button or "Horiz/Acq"
waveform
Menu from the oscilloscope UI. Go to
Horizontal tab and decrease the record
length
TDSUSB2 Universal Serial Bus Measurement Package
169
Reference
Table 36.Error Code descriptions (cont.)
Error in acquiring waveform from oscilloscope
E111
Timeout occurred while
acquiring a waveform
E112
Error in turning on a channel
E113
Error in importing waveform
from instrument
This is the GPIB timeout information
that appears when the application is
acquiring the waveform
The application failed to turn on the
live channel selected as source(s)
Make sure the probe and compliance test
fixture connections are proper. Reacquire
the new waveform
Make sure the probe and compliance test
fixture connections are proper. Reacquire
the new waveform
The operation of importing the
Make sure the probe and compliance test
waveform from the acquisition memory fixture connections are proper. Reacquire
failed. This will fail whenever there is the new waveform
no valid waveform in the acquisition
memory
Packet Detection Warning
E201
E202
E203
E204
Try acquiring the signals again by pressing
Testing will be performed on an full
Run button
USB2.0 packet, which includes an
EOP region. If no EOP region is
found, the results might be incorrect
since testing will be performed on a
incomplete packet. For Upstream
signal quality testing, EOP must be
present
Number of record points (data At least 500 record points are required Try acquiring the signals again the
points) are too few to process to perform an operation. The
pressing Run button
application automatically sets the
required record length. The user
should not change the record length
Idle region is not found after
As part of the USB2.0 packet
Try acquiring the signals again by pressing
EOP
description, an EOP region follows an the Run button
idle region. If no idle region is found, it
may lead to incorrect results
Reconfirm acquired data before As part of the USB2.0 packet
Try acquiring the signals again by pressing
computing the results
description, a minimum of three J-K or the Run button
K-J state transitions are expected. An
absence of these transitions may lead
to incorrect results
No EOP region found
Signal rate measurement errors or warning
E301
E306
170
There should be at least two
Cross-Over time values to
calculate the signal rate
Device/ File doesn't match with
the selected speed
For signal rate calculations, the
application expects a minimum of two
Cross-Over points
The signal rate should be within 30%
of the specified signal rate of the Low,
Full and High Speed signals
Try acquiring the signals again by pressing
the Run button
Please check the speed of the connected
device. Ensure that the selection in the
application is the same. For example, if a
Low Speed unit is under test, make sure
Low Speed is selected in the application
TDSUSB2 Universal Serial Bus Measurement Package
Reference
Table 36.Error Code descriptions (cont.)
JK and KJ Jitter Measurement Errors
E304
Number of bits is not sufficient
to calculate JK jitter
E305
Number of bits is not sufficient
to calculate KJ jitter
More than four Cross-Over points are
expected by the algorithm
More than four Cross-Over points are
expected by the algorithm
Try acquiring the signal again by pressing
Run button
Try acquiring the signal again by pressing
Run button
Report Generation Errors
E401
E402
E403
E404
E405
E406
Error generating report in Plug This error is generated if there are no Check if the valid directory and file name is
Fest format
results to process
selected for report generation. Check if the
file name extension is .htm. Try generating
the report again by pressing Run button
Error Generating report in Tek This error is generated if there are no Check if the valid directory and file name
format
results to process
are selected for report generation. Check if
the file name extension is .htm. Try
generating the report again by pressing
Run button
Error Generating report in CSV This error is generated if there are no Check if the valid directory and file name s
format
results to process
are selected for report generation. Check if
the file name extension is .htm. Try
generating the report again by pressing
Run button.
No results present to generate This error is generated if there are no Press the Run button to get the results and
report
results present for the USB Inrush,
try again
Droop and Signal Quality Check
Reports
Could not generate a report as This error is generated if there is no Press the Run button and perform the
the Eye diagram is missing
eye diagram found in
measurements again
C:\TekApplications\TDSUSB2\temp
Could not generate report as
This error is generated if there is no Press the Run button and perform the
Waveform plot is missing
waveform plot found in
measurements again
C:\TekApplications\TDSUSB2\temp
Eye Measurements
E601
E602
The maximum zoom factor has N/A
Press the Reset button or use the Zoom
been reached
out feature.
Signal not proper, check the
This is a check to confirm the acquired Try acquiring the correct waveform
signal
waveform
Inrush Measurements
E501
Final data point is still above the The signal has not fallen below the
inrush threshold
Inrush threshold level
TDSUSB2 Universal Serial Bus Measurement Package
Try acquiring the correct waveform
171
Reference
Table 36.Error Code descriptions (cont.)
Utilities
E701
E702
E701
E702
Number of edges found in the
waveform is less than required
edges (user input) to perform the
deskew operation
Before the deskew operation can
start, the edge(s) available for
deskew operation must be greater
than or equal to the number of
edges needed
Propagation delay is more than The deskew operation is performed
one cycle
on the same signal source. Check
whether dissimilar signals or any
external sources (like different
probes) are used
Number of edges found in the
Before starting the deskew
waveform is less than required
operation, the application tries to
edges (user input) to perform the find out if the edge(s) available for
deskew operation
deskew operation is greater than or
equal to the number of edges
needed
Propagation delay is more than The deskew operation is performed
one cycle
on the same signal source. Check
whether dissimilar signals or any
external sources (like different
probes) are used
Adjust the timebase and (or) increase the
record length. Perform the deskew operation
by entering less number of edges
Check whether the same signal source is
used for the "From" and "To" channels.
Check whether the bandwidth of the probes
is sufficient for the signal bandwidth
Adjust the timebase and (or) increase the
record length. Perform the deskew operation
by entering less number of edges
Check whether the same signal source is
used for the "From" and "To" channels.
Check whether the bandwidth of the probes
is sufficient for the signal bandwidth
TSV File Generator
E721
E722
E725
Error in generating a .tsv file
This message is generated if there
is an error while generating the .tsv
file for the given inputs
Timestamp of the two .csv files There will be two separate .csv files,
did not match. Conversion failed one for D+ and another for D-. Each
of these files consists of the time
values of the D+ and D- signal. It is
expected that both the files should
have Data at same time stamp
Invalid CSV file format
This happens if the user does not
use a valid CSV file format for the
CSV-TSV file conversion
E723
.csv file(s) not found
E724
tsv file not found
172
Check the correctness of the input .csv file(s)
and the output .tsv file
Try generating the .tsv file with a new set of
.csv files
Check whether the CSV files are generated
using the supported instruments and ensure
that "Include waveform scale factors" is
selected while generating a CSV file
The .csv file(s) are needed for the Check that the extension of the file(s) is .csv.
.csv source mode selection is not
Make sure to press the Enter key while
found. Two valid .csv files are
entering the file name in the file selection
needed for the Single-ended signals field to update old file names
and one valid .csv file is needed for
Differential signals
A valid .tsv file is needed as
Make sure to press the Enter key while
destination file
entering the file name in the file selection
field to update old file names. Enter the .tsv
extension while entering the file name
TDSUSB2 Universal Serial Bus Measurement Package
Reference
Table 36.Error Code descriptions (cont.)
Save and Recall
E751
E754
E752
E753
The file name has invalid
characters and could not be
saved
The file name has invalid
characters and could not be
recalled
The file doesn't exist
The saved file has been
corrupted and not recalled
The application expects the file
names in alphanumeric characters
Make sure the valid characters are used in
the file names
The application expects the file
names in alphanumeric characters
Make sure that valid characters are used in
the file names
The file name selected for recall
Recall an existing file name
does not exist
The application saves the
Try recalling a correct file. Store the settings
application settings with the .ini file to a new file and recall it when needed
extension. The user should not
modify the data inside the file
Save and Recall
E801
E802
E803
E-804
E-805
The current signal is not a Chirp The acquired signal is not a correct Try acquiring the signals again by pressing
Signal. Acquire the correct
Chirp Signal. Correct Chirp Signal is Run button
signal and proceed
required to perform chirp
measurements
No Chirp-K or Chirp-J state
As part of the USB2.0 Chirp
Press the Run button to acquire the signals
found
description, a minimum of 1 Chirp-K
and 1 Chirp-J state are expected
after Chirp Response timing
calculation
There should be at least 3
As part of the USB2.0 Chirp
Press the Run button to acquire the signals
Chirp-K and 3 Chirp-J state
description, a minimum of three
Chirp-K and three Chirp-J state are
expected after Chirp Response
timing .An absence of this may lead
to incorrect results
No Chirp-J state found
As part of the USB2.0 Chirp
Press the Run button to acquire the signals
description, a minimum of one
Chirp-J state is expected after Chirp
Response timing. An absence of this
may lead to incorrect results
No Chirp-K state found
As part of the USB2.0 Chirp
Press the Run button to acquire the signals
description, a minimum of one
Chirp-K state is expected after Chirp
Response timing. An absence of this
may lead to incorrect results
TDSUSB2 Universal Serial Bus Measurement Package
173
Reference
Table 36.Error Code descriptions (cont.)
Rise Time/ Fall Time Measurement
E901
E902
E903
E904
E905
E906
174
The High level values needs to
be decreased
The Low level values needs to
be increased
Both the High and Low level
values needs to be changed
This message is displayed if no
edge is found in the high level
If no edge is found in the low level
the message is displayed
If no edge is found in the high and
low level the message is displayed
The calculated Rise Time is
This warning appears when the
more than the calculated bit time calculated Rise Time is more than
of the input signal
the calculated bit time of the input
signal
The calculated Fall Time is more This warning appears when the
than the calculated bit time of calculated Fall Time is more than
the input signal
the calculated bit time of the input
signal
The calculated Rise Time and This warning appears when the
Fall Time is more than the
calculated Rise Time and Fall Time
calculated bit time of the input is more than the calculated bit time
signal
of the input signal
Decrease the percentage of the high level
values of Rise Time/Fall Time
Decrease the percentage of the low level
values of Rise Time/Fall Time
Increase and decrease the percentage of the
high and low level values of Rise Time/Fall
Time
Increase the lower level and decrease the
upper level of the configure limits of Rise
Time/Fall Time
Increase the lower level and decrease the
upper level of the configure limits of Rise
Time/Fall Time
Increase the lower level and decrease the
upper level of the configure limits of Rise
Time/Fall Time
TDSUSB2 Universal Serial Bus Measurement Package
Reference
Table 36.Error Code descriptions (cont.)
Miscellaneous
E1000
Invalid device Chirp-K for this
test
E1001
No SOF found
E1002
E1003
E1004
E1005
E1006
E1007
E1008
As part of the USB2.0 description, a
minimum of 3 Chirp-K and 3 Chirp-J
states are expected after Chirp
Response timing. An absence of this
may lead to incorrect results
No SOF in D+ and D-
Try acquiring the signals again by pressing
Run button
Power off the DUT, check the probe
connection and try acquiring the signal
The current signal is not a Reset Acquired signal is not a correct Reset Try acquiring the signals again by pressing
From Suspend signal. Acquire From Suspend signal. Correct Reset Run button
the correct signal and proceed From Suspend Signal is required to
perform the measurement
The current signal is not a
Acquired signal is not a correct
Try acquiring the signals again by pressing
Resume signal acquire the
Resume signal. Correct Resume
Run button
correct signal and proceed
signal is required to perform the
measurement
The current signal is not a
Acquired signal is not a correct
Try acquiring the signals again by pressing
Suspend signal acquire the
Suspend signal. Correct Suspend
Run button
correct signal and proceed
signal is required to perform the
measurement
There is no falling edge
If there is no J state found
Try acquiring the signals again by pressing
transition for this test.
Run button
There is no start of Chirp-K state If there is no K state found
Try acquiring the signals again by pressing
for this device
Run button
There is no K State transition
If there is no K state found
Try acquiring the signals again by pressing
driven by the host controller
Run button
The current signal is not a Reset The acquired signal is not a correct
Try acquiring the signals again by pressing
From Resume signal. Acquire Reset From Resume signal. Correct Run button
the correct signal and proceed Reset From Resume signal is required
to perform the measurement
TDSUSB2 Universal Serial Bus Measurement Package
175
Reference
Settings for the supported Instruments
This section gives a list of default oscilloscope settings the application uses for
supported instruments.
The recommended voltage probes for single-ended signals are the P6245,
TAP1500 probes and for differential signals is the P6248 probe. The next table
shows the oscilloscope settings for Low Speed and Full Speed signals for the
supported instruments.
Upstream Setups for Low Speed signals
Table 37. Oscilloscope settings for low speed signals
Ch.
Vertical Setup Pos.
Scale Offset Coupling
Horizontal Setup Record
Main Trigger Length
Scale Position
Trigger Setup Trigger Logic Pattern Threshold Repeat
Trigger Type Function Limit Mode
State
Ch1(D-)
0
1V
0
DC
5000
2.00 E07s
74%
Logic
AND
>100 ns 800 mV Normal
Off
Ch2(D+)
Ch3
Qualify
0
0
1V
1V
0
0
DC
DC
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
800 mV N/A
2.7 V
N/A
Upstream Setups for Full Speed signals
You can test the Full Speed devices at Tier 6. This testing is also known as
legacy testing.
When Full Speed and Low Speed device are High Speed capable, the application
uses the signal rate limits of Full Speed/Low Speed devices that are not High
Speed capable to determine PASS or FAIL.
Table 38. Oscilloscope settings for full speed signals
Ch.
Vertical Setup Pos.
Scale Offset
Coupling
Ch1 (D-)
0
1V
0
DC 5000
25 E-9s 83%
Logic
AND
>100 ns 800 mV
Normal Off
Ch2 (D+) 0
Ch3
0
Qualify
1V
1V
0
0
DC N/A
DC N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
176
Horizontal Setup
Record Main Trigger
Length Scale Pos.
N/A
N/A
Trigger Setup Trigger Logic Pattern
Threshold
Trigger Type Function Limit Mode
800 mV
2.7 V
Repeat
State
N/A
N/A
TDSUSB2 Universal Serial Bus Measurement Package
Reference
Downstream Setups for Low Speed signals
Table 39. Downstream settings for low speed signals
Ch.
Vertical Setup Position Scale
Offset Coupling
Ch1 (D-) 0
1V
0
DC
Horizontal Setup Record Main
Trigger length scale pos.
2500
5.00 E-07s 6%
Trigger Setup Trigger Slope Threshold Trigger
type mode
Edge
Rise
1.65 V
Normal
Ch2 (D+) 0
N/A
N/A
1V
0
DC
N/A
N/A
N/A
800 mV
N/A
Trigger Setup Trigger Slope
Threshold Trigger type mode
N/A
N/A
N/A
N/A
Edge
Rise
1.65 V Normal
Repeat
State
N/A
Off
Downstream Setups for Full Speed signals
Table 40. Downstream settings for low speed signals
Ch.
Ch1 (D-)
Ch2 (D+)
Vertical Setup Position Scale
Offset Coupling
-1.5
1V
0
DC
-1.5
1V
0
N/A
Horizontal Setup Record Main
Trigger length scale pos.
5000
50.0 E-9 s 5%
N/A
N/A
N/A
Hold Off
Mode
5s
N/A
Upstream Setup for High Speed signals
Probe Function External Attenuation 1.0
Table 41. Upstream settings for high-speed signals
Ch.
Vertical Setup
Horizontal Setup Record Main
Position Scale Offset Trigger Fit to Length scale pos.
Differential 0
0.1 V 0
20000 2.00 E-07 10%
On
Trigger Setup Trigger
Threshold Trigger Type
mode
Edge
176 mV Normal
Repeat Hold Off Display
State
Mode Mode
Off
1.20
E-06 s
Sin(x)/x
Inrush Setup
Table 42. Inrush current settings
Vertical Setup Pos. Scale Offset
Coupling
-3
1A
0
DC
Horizontal Setup Record Main Trigger Setup Trigger Slope
Display
Trigger length scale pos.
Threshold Trigger type mode
Mode
25000
1.00E-05 10%
Edge
Rise 20 mA
Normal Sin(x)/x
TDSUSB2 Universal Serial Bus Measurement Package
Hold Off
Mode
100 E-09
177
Reference
Droop Setup
Table 43. Droop settings
Ch.
Vertical Setup Pos. Scale
Offset Coupling
Ch1
(Vbus)
2
Ch2
-3
Droop
Trigger
load
200 mV
5
DC
Horizontal Setup Record
Main
Trigger length scale pos.
5000
2.00 E-05 50%
2V
0
DC
N/A
N/A
Trigger Setup Trigger
Slope
Threshold Type
N/A
N/A
N/A
Display
Mode
Acquisition
Mode
N/A
Average 16
Points
Edge
2.5 V
Sin(x)/x
N/A
Trigger Setup Trigger Slope
Threshold Trigger type mode
Rpt
State
Edge
Fall
1.35 V Normal Off
N/A
N/A
N/A
N/A
Rise
Chirp Device EL_28, EL_29 EL_31 setup
Table 44. Chirp Device settings
Ch.
Vertical Setup Pos. Scale
Offset Coupling
Ch1 (D+) -3.4
.5V
0
DC
Horizontal Setup Record
Main Trigger length
scale pos.
5000 4.00E-04 5%
Ch2 (D-) -3.4
.5 V
0
DC
N/A
N/A
N/A
N/A
N/A
Hold Off Display
Mode
1.00 E04
N/A
Sin(x)/x
N/A
Hold
off
Display
mode
Chirp Host EL_33, EL_34 setup
Table 45. Chirp Host settings
Ch.
Vertical Setup Pos. Scale
Offset Coupling
Ch1 (D+) -3.4
.5V
0
DC
Horizontal Setup Record Trigger Setup Trigger Slope
Rpt
Main Trigger length scale Threshold Trigger type mode
state
pos.
5000 4.00E-04 5%
Edge
Fall
1.35 V Normal Off
Ch2 (D-) -3.4
.5 V
0
DC
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
1.00 E- Sin
04
(x)/x
N/A
N/A
Chirp Host EL_35 setup
Table 46. Chirp Host El_35 settings
Ch.
Vertical Setup Pos.
Scale
Offset Coupling
-3.40 .5V 0
DC
Ch1
(D+)
Ch2 (D- -3.4 .5 V 0
)
178
DC
Horizontal Setup
Record Main Trigger
length scale pos.
12500 1.00E-03 95%
N/A
N/A
N/A
Trigger Setup Trigger Polarity
Threshold Trigger type Mode
Pulse
(Glitch)
N/A
Repeat Width Filter Hold Display
state
off mode
Positive 0.30 V Normal Off
N/A
N/A
N/A
N/A
9.0E-9 Accept 5.00 Sin(x)/x
E-5
N/A N/A N/A N/A
TDSUSB2 Universal Serial Bus Measurement Package
Reference
Suspend Setup
Table 47. Suspend settings
Ch.
Vertical Setup
Pos Scal Offset Cpling
Horizontal Setup
Reclen Main
scale Trigpos
Ch1
(D+)
Ch2
(D-)
-2.96 .5V
0
DC
20000
4.00E-04
-1.72 .5 V
0
DC
N/A
N/A
Trigger Setup
Rpt
Trigtype Slope Threshold Trigmd state
Hold
off
Display
mode
90%
Edge
Rise
2.69V
Normal Off
Auto
N/A
N/A
N/A
N/A
N/A
N/A
Sin
(x)/x
N/A
N/A
Resume Setup
Table 48. Resume settings
Ch.
Vertical Setup
Pos Scale Offset Cpling
Horizontal Setup
Reclen Mainscale Trig
pos
Trigger Setup
Trigtype Slope Threshold
Trigmd
Ch1 (D+)
Ch2 (D-)
-3.40
-3.40
50000
N/A
Edge Fall
N/A N/A
.5V 0
.5 V 0
DC
DC
2.00E-05 10%
N/A
N/A
Repeat Hold off Display
state
mode
0.690 V Normal Off
N/A
N/A
N/A
1.00 E-5 Sin(x)/x
N/A
N/A
Reset from High Speed
Table 49. Reset from high speed settings
Ch.
Vertical Setup
Pos Scale Offset Cpling
Horizontal Setup
Reclen Mainscale Trigpos
Trigger Setup
Trigtype Slope Threshold Trigmd
70%
Edge
Fall
0.690 V Normal Off
N/A
N/A
N/A
N/A
Ch1 (D+) -3.40
.5V
0
DC
25000
Ch2 (D-) -3.40
.5 V 0
DC
N/A
1.00E03
N/A
N/A
Repeat Hold off Display
state
mode
N/A
1.00 E-5 Sin
(x)/x
N/A
N/A
Reset from Suspend
Table 50. Reset from Suspend settings
Ch.
Vertical Setup
Pos Scale Offset Cpling
Horizontal Setup
Trigger Setup
Repeat Hold
Reclen Mainscale Trigpos Trig type Slope Threshold Trigmd state off
Ch1 (D+) -3.40
.5V
0
DC
50000
1.00E-03 15%
Edge
Fall
1.8 V
Normal Off
Ch2 (D-) -3.40
.5 V
0
DC
N/A
N/A
N/A
N/A
N/A
N/A
N/A
TDSUSB2 Universal Serial Bus Measurement Package
N/A
Display
mode
1.00 E- Sin(x)/x
5
N/A
N/A
179
Reference
Packet Parameter
Table 51. Packet Parameter settings
Ch.
Vertical Setup Pos Scale
Offset Cpling
Horizontal Setup
Reclen Mainscale
Trigpos
Trigger Setup Trigtype Slope
Threshold Trigmd
Repeat Hold
state off
Ch1 (D+) 0.0
0.2V
0
DC
50000
50
50
Width
Fall
N/A
Normal Off
Ch2 (D-) 0.0
0.2V
0
DC
N/A
N/A
N/A
N/A
N/A
N/A
N/A
180
N/A
Display
mode
1.00 E- Sin(x)/x
5
N/A
N/A
TDSUSB2 Universal Serial Bus Measurement Package
Parameters
About Application Parameters
This section describes the TDSUSB2 application parameters, and includes the
menu default settings. You should refer to the user manual for your oscilloscope
for operating details of other controls, such as front-panel buttons.
The parameters for the menus and options list the selections or range of values
available for each, and include the default values.
File Menu Parameters
There are no parameters for the File menu items.
Save and Recall Menu Parameters
There are no parameters for the Save and Recall menu.
Preferences Parameters
Table 52. Preferences parameters
Parameters
Warn if the configuration was not changed since last run, when Run button is
pressed
Show the Eye Diagram automatically, without having to press the Eye Diagram
button in the Results screen
Show the Waveform Plot automatically without having to press Waveform Plot
button in the Results screen
Ask for Filename or Device ID before automatically generating Reports or Data files
Automatically generate Data files (TSV)
Show the report when generated
Always display the waiver mask for High Speed Near-end and Low Speed Eye
Diagrams
Do Autoset Every Time
LS or FS device with USB2.0 silicon
Allow the user configuration of test limits
Prompt for the Droop Test readout during Report Generation
Packet identification by user using cursors
TDSUSB2 Universal Serial Bus Measurement Package
Selections
On, Off
Default setting
On
On, Off
On
On, Off
On
On, Off
On, Off
On, Off
On, Off
Off
Off
On
Off
On, Off
On, Off
On, Off
On, Off
On, Off
On
Off
Off
Off
Off
181
Parameters
Parameters
Modify HoldOff
Selections
250 ns to 12 s
Default setting
1.2 µs
Measurement Menus
The options available under the Measurements menu are as follows:
Select
Configure
Measurement Parameters
The next table lists the options in the Select Measurements menu by area:
Table 53. Measurement parameters
Parameters
Low Speed
Full Speed
182
Selections
Eye Diagram
Signal Rate
EOP Width
Rise Time
Fall Time
Cross-Over Voltage
Consecutive Jitter
Paired JK Jitter
Paired KJ Jitter
Inrush Current
Droop test
Eye Diagram
Signal Rate
EOP Width
Rise Time
Fall Time
Cross-Over Voltage
Consecutive Jitter
Paired JK Jitter
Paired KJ Jitter
Inrush Current
Droop test
Default setting
None
Eye Diagram
TDSUSB2 Universal Serial Bus Measurement Package
Parameters
Table 53. Measurement parameters (cont.)
High Speed
Device ID
Device
Description
Eye Diagram
Signal Rate
EOP Width
Rise Time
Fall Time
Monotonic Property
Chirp
Receiver Sensitivity
Inrush Current
Droop test
Packet Parameter
Suspend
Resume
Reset from High Speed
Reset from Suspend
User defined or fsfe_001
User defined or Dummy Device
None
fsfe_001
Dummy Device
Configure Menu
You can configure the parameters for the selected measurements. The
configurations for the Signal Quality, Inrush, and Droop tests are available as
different menus.
Configure Signal Quality Measurements Parameters
Configure
Table 54. Configure Signal Quality parameters
Parameters
Tier
Direction
Test Point
Selections
Tier 1, Tier 2, Tier 3, Tier 4, Tier 5, Tier 6
Upstream, Downstream
Near End, Far End
TDSUSB2 Universal Serial Bus Measurement Package
Default setting
Tier 6
Upstream
Far End
183
Parameters
Source
Table 55. Configure Signal Quality Source parameters
Parameters
Live/Ref
Selections
Differential Ch1, Ch2, Ch3, Ch4
Ref1, Ref2, Ref, Ref4
Live/Ref Single Ch1, Ch2, Ch3, Ch4
Ended D+
Ref1, Ref2, Ref, Ref4
Single Ended D– Ch1, Ch2, Ch3, Ch4
Qualifier
Ch1, Ch2, Ch3, Ch4
Ref1, Ref2, Ref, Ref4
File
None
Default setting
Ch1
Ch1
Ch2
Ch3
C:\TekApplications\tdsusb2\tsvfi
les\FS_SQC.tsv
Configure Inrush Current Measurement Parameters
Configure
Table 56. Configure Inrush Current Measurement parameters
Parameters
Tier
Vbus
Device Type
Selections
None
Range
Hot Plug Attach, Low Power Configure Low Power
Resume, High Power Configure, High Power
Unconfigure, High Power Resume
Default setting
Tier 1
5.15V
Hot Plug Attach
Source
Table 57. Configure Inrush Current Source Measurement parameters
Parameters
Live/Ref
File
184
Selections
Ch1, Ch2, Ch3, Ch4Ref1,
Ref2, Ref, Ref4
None
Default setting
Ch4
C:\TekApplications\tdsusb2\tsvfiles\LS_INRUSH.tsv
TDSUSB2 Universal Serial Bus Measurement Package
Parameters
Configure Droop Test Parameters
Configure
Table 58. Configure Droop Test parameters
Parameters
Port
Device Type
Selections
Default setting
Port 1, Port 2, Port 3, Port 4, Port 5, Port 6, Port 7 Port 1
Self Powered Hub
Bus Powered Hub
Bus Powered Hub
Source
Table 59. Droop Test Source parameters
Parameters
Live/Ref
File
Selections
Ch1, Ch2, Ch3, Ch4
Ref1, Ref2, Ref, Ref4
Ch1, Ch2, Ch3, Ch4
Ref1, Ref2, Ref, Ref4
Ch1, Ch2, Ch3, Ch4
None
VBUS source
Ch1, Ch2, Ch3, Ch4
VBUS channel
Trigger
Default setting
Ch1
Ch2
C:\TekApplications\tdsusb2\tsvfiles\FS_SQ
C.tsv
Ch1
Configure Receiver Sensitivity
Table 60. Receiver Sensitivity parameters
Parameters
Source
Selections
Ch1, Ch2, Ch3, Ch4
Default setting
Ch 1
Configure Suspend, Reset from High Speed, Resume, Reset from Suspend
Table 61. Configure Reset, Resume, Suspend parameters
Parameters
Selections
Signal Direction Upstream, Downstream
Source
D+ Ch1, Ch2, Ch3, Ch4
Ref1, Ref2, Ref3, Ref4
D-Ch2, Ch3, Ch4
Ref1, Ref2, Ref3, Ref4
TDSUSB2 Universal Serial Bus Measurement Package
Default setting
Upstream
Ch1
Ch2
185
Parameters
Configure Packet Parameter
Table 62. Configure Packet parameters
Parameters
Selections
Default setting
DUT
Host, Device
Host
Host
EL_21, EL_23, EL_25
EL_55
EL_22
EL_21, EL_23, EL_25
Source
Ch1, Ch2, Ch3, Ch4
Ref1, Ref2, Ref3, Ref4
EL_21, EL_22, EL_25
EL_22
Ch1, Ch2, Ch3, Ch4
Ref1, Ref2, Ref3, Ref4
Ch1
Device
Source
EL_21, EL_22, EL_25
Ch1
Configure Chirp
Table 63. Configure Chirp parameters
Parameters
Selections
Default setting
DUT
Host, Device
Host
Host
EL_33, EL_34
EL_35
D+-Ch1, Ch2, Ch3, Ch4
Ref1, Ref2, Ref3, Ref4
D--Ch2, Ch3, Ch4
Ref1, Ref2, Ref3, Ref4
EL_28, EL_29, EL_31
EL_33, EL_34
D+-Ch1, Ch2, Ch3, Ch4
Ref1, Ref2, Ref3, Ref4
D–-Ch2, Ch3, Ch4
Ref1, Ref2, Ref3, Ref4
Ch1
Source
Device
Source
Ch1
Ch2
EL_28, EL_29, EL_31
Ch2
Results Menus
There are no parameters for the Results Menus.
186
TDSUSB2 Universal Serial Bus Measurement Package
Parameters
Utilities Menus
There are three Utilities menu items:
Deskew
Report Generator
TSV File Generator
Deskew
Table 64. Deskew parameters
Parameters
Source1
Ref level1
Hysteresis1
Source2
Ref level2
Hysteresis2
Slope
Edges
Selections
Ch1, Ch2, Ch3, Ch4
0 -100 %
0-25%
Ch1, Ch2, Ch3, Ch4
0 -100 %
0-25%
Rise, Fall
1 to 50
Default setting
Ch1
50%
5%
Ch2
50%
5%
Fall
1
Report Generator Menu Parameters
Table 65. Report Generator Menu parameters
Parameters
Report Format
Generation
File name
Selections
Tektronix specific, Plug-Fest format,
CSV format
Automatic, Manual
fsfe_001-tek.htm
Report Directory
None
TDSUSB2 Universal Serial Bus Measurement Package
Default setting
Tektronix specific
Manual
C:\TekApplications\tdsusb2\report\fsfe_001tek.htm
C:\TekApplications\tdsusb2\report
187
Parameters
TSV File Generator Menu Parameters
Table 66. TSV File Generator Menu parameters
Parameters
Selections
Input
CSV, Live
CSV Waveform Source Single-Ended
(D+
D-)
Differential
TSV File name
TSV Source File
Default setting
C:\TekApplications\tdsusb2\csv files\csvFile001.csv
Single-Ended
C:\TekApplications\tdsusb2\tsvfilegenerator\Dplus.csv
C:\TekApplications\tdsusb2\tsvfilegenerator\Dminus.csv
C:\TekApplications\tdsusb2\tsvfilegenerator\Differential.csv
C:\TekApplications\tdsusb2\tsvfilegenerator\TsvFile001.tsv
Help Menu
Help Topics - Displays the help file for the TDSUSB2 application.
About TDSUSB2 - Displays a dialog box with information about the current
TDSUSB2 application.
Control Menu
Control Menu Parameters
Table 67. Control parameters
Parameters
Run
Hide
Exit
188
Selections
None
None
None
TDSUSB2 Universal Serial Bus Measurement Package
Measurement Algorithms
About Measurement Algorithms
The TDSUSB2 package performs measurements for USB2.0 compliance. This
section contains information about the algorithms used by the application to
perform each measurement.
Cross-Over Voltage for Low Speed and Full Speed Signals
You can define the Cross-Over point as the point where the Data+ line voltage
crosses the Data– line voltage. The voltage value at this point is called the CrossOver Voltage and the time value is called the Cross-Over time.
In the following figure, C1, C2, C3, C4 and C5 are called Cross-Over Points and
R1, R2, R3, R4, R5, R6 and R7 are called Reference points.
The following figure explains the interpolation technique used to find the actual
Cross-Over where p1 and p2 are the adjacent data points after q2:
TDSUSB2 Universal Serial Bus Measurement Package
189
Measurement Algorithms
The intersection of the four voltage points p1, p2, q1, and q2 gives the CrossOver point for Voltage level (y) and time (x). x and y coordinates are obtained by
solving the following two equations:
Where: p1, p2, q1, and q2 are the consecutive data points of a single-ended
signal.
x and y are the Cross-Over coordinates x and y that is given by the intersection of
p1, p2, q1, and q2.
t1 and t2 are the time values for the data points p1 and p2.
190
TDSUSB2 Universal Serial Bus Measurement Package
Measurement Algorithms
Cross-Over Voltage for High-Speed Signals
The Cross-Over is defined as that point where the Differential line voltage
becomes zero. The voltage value at this point is called the Cross-Over Voltage
and the time value is called the Cross-Over Time.
The following algorithms are used for High Speed signals for Cross-Over
voltage.
for High Speed y=0
If,
If,
Where: p1 and p2 are the consecutive data points of a single-ended signal, x and
y are the coordinates of the point where the Differential line voltage becomes
zero and t1 and t2 are the time values for the data points p1 and p2.
Note 1: Low Speed and Full Speed Signals are acquired with single-ended
probes that enable finding the Cross-Over Voltage.
Note 2: For High Speed devices, the signal is differential, Cross-Over voltage is
zero and cross-Over time is the interpolation of two data points at zero crossings.
TDSUSB2 Universal Serial Bus Measurement Package
191
Measurement Algorithms
Signal Rate
You can define the signal rate for Low or Full speed signals as the inverse of the
average bit time that gives the transmission rate of the USB2.0 signal. For high
speed signals, the signal rate is defined as the inverse of the average bit time that
gives the transmission rate of the Test_Packet.
The average signal rate is the calculated average of number of bits divided by
sum of all periods.
Where: SumWeights is the total number of bits in the packet.
SumPeriod is the sum of all time periods between the Cross-Overs.
C1 and C2 are the Cross-Over points.
Bit time is the difference of the time values of C2 and C1.
192
TDSUSB2 Universal Serial Bus Measurement Package
Measurement Algorithms
EOP Width Calculation
You can define the EOP width for Low or Full speed signals as the width of the
end-of-packet of a USB2.0 signal. For high speed signals, it is defined as the
width of the end-of-packet pattern of a Test_Packet.
EOP Width is calculated as shown below.
Where:
EOP1 and EOP2 are the data points on a USB2.0 signal crossing the Mean
Cross-Over level. The Cross-Over level is the mean value of all Cross-Overs in
the USB2.0 packet.
TDSUSB2 Universal Serial Bus Measurement Package
193
Measurement Algorithms
Consecutive and Paired Jitter
You can test two types of Jitter measurements: Consecutive and Paired Jitter.
Consecutive jitter measures the consecutive data bit calculated using the signal
rate.
Consecutive jitter is calculated as follows.
Where: C1, C2, C3, and C4 are the Cross -Over points.
R1, R2, R3, and R4 are the reference points.
Consecutive jitter is the difference between the time values at Rx and Cx.
Consecutive jitter is calculated for all Cross-Overs of the USB2.0 signal.
Paired JK Jitter is defined as the jitter time for paired (JK next to KJ) differential
data transition. Paired KJ Jitter is defined as the jitter time for paired (KJ next to
194
TDSUSB2 Universal Serial Bus Measurement Package
Measurement Algorithms
JK) differential data transition. They are calculated for all the consecutive jitters
as:
and so on till the last Cross-Over.
Eye Diagram
The Eye Diagram checks whether the USB signal is aligned with its
corresponding eye diagram. It represents the whole signal by splicing it into a
number of waveforms of unit interval (each waveform is of one bit time interval)
scaled and represented on the eye masks.
Eye Violation
An eye violation occurs when an USB signal crosses an eye mask. The results of
the eye violation are PASS if the waveform does not violate the eye masks. The
result is FAIL if the waveform violates the Eye Masks. The result is
CONDITIONAL PASS if the waveform violates the eye mask, but is within the
waiver mask.
By default, the application displays the waiver mask if the result is either
CONDITIONAL PASS or FAIL. If you select the option "Always show the
waiver mask in the eye diagram" in File> Preferences menu, the waiver mask is
displayed irrespective of the result.
TDSUSB2 Universal Serial Bus Measurement Package
195
Measurement Algorithms
Inrush Current
Inrush Current is calculated by recording all the current values above the 100 mA
current level throughout the signal. This signal is integrated to get the total
charge greater than 100 mA. This is the charge in Coulombs and is calculated
using the equation:
i is the current waveform above 100 mA.
Capacitance is calculated by using the equation,
Where: C is the capacitance.
Q is the charge.
V is the voltage.
The supply voltage is one of the inputs to Inrush measurement.
Droop Test
Droop voltage is the difference between the VBUS value when the droop load is
off and the lowest voltage of the VBUS value when the droop load is powered on.
196
TDSUSB2 Universal Serial Bus Measurement Package
Measurement Algorithms
Receiver Sensitivity Test
You can test the Receiver sensitivity of a high speed device to respond to the
particular data pattern generated by the digital signal generator. Receiver
Sensitivity responds whenever the data pattern level of the voltage level is greater
than 150 mV and does not respond when the voltage level is equal to or less than
100 mV below the squelch level. The application provides the procedural steps to
this test.
For more details, refer to the procedural steps.
Chirp Test for a Device
This test examines the basic timing and voltages of both upstream and
downstream ports during the speed detection protocol. To get the Chirp signal,
hot-plug the unit under test (device) and measure the signaling with single-ended
probes on both lines. The application analyzes data for the following:
Reset duration
Chirp-K amplitude
Chirp-K duration
TDSUSB2 Universal Serial Bus Measurement Package
197
Measurement Algorithms
Note: There must be three K-J pairs after chirp-K that are less than 500 µs. This
is to check the response of the device. The response time must be less than 100
µs. You can verify this manually.
Host Chirp Test
This test examines the basic timing and voltages of both upstream and
downstream ports during the speed detection protocol. To get the Chirp signal,
hot-plug the unit under test (Host) and measure the signaling with single ended
probes on both lines. The application analyzes data for the following:
Chirp Response Timing
Chirp-K and Chirp-J duration
The application analyzes data for the Chirp J/K to first SOF Time.
Monotonic Property Test
This algorithm calculates the number of monotonic violations that are present
from of the start of the signal to the end of packet (EOP). For the signal to be
monotonic, the signal is checked from 15% to 85% of the Peak to Peak
differential signal levels or user specified levels.
Monotonicity is calculated for Rising and Falling slopes as follows.
For Rising slope, f(x) is a differential signal if,
signal is monotonic for all values of x till the last data point.
198
TDSUSB2 Universal Serial Bus Measurement Package
Measurement Algorithms
For Falling slope, f(x) is a differential signal if
for all values of x till the last data point.
Rise Time
The Rise Time measurement is the time difference between when the VRefHi
reference level is crossed and the VRefLo reference level is crossed on the rising
edge of the waveform. The Rise Time algorithm uses the VRef values as the
reference voltage level. Each edge is defined by the slope, voltage reference level
(threshold) and hysteresis.
The application calculates this measurement using the following equation:
Where: TRise is the rise time,
THi+ is the VRefHi crossing on the rising edge
TLo+ is the VRefLo crossing on the rising edge
Fall Time
The Fall Time measurement is the time difference between when the VRefLo
reference level is crossed and the VRefHi reference level is crossed on the falling
edge of the waveform. The Fall Time algorithm uses the VRef values as the
reference voltage level. The slope, voltage reference level (threshold), and
hysteresis define each edge.
The application calculates this measurement using the following equation:
Where: TFall is the fall time,
TLo- is the VRefLo crossing on the falling edge
THi- is the VRefHi crossing on the falling edge
TDSUSB2 Universal Serial Bus Measurement Package
199
Measurement Algorithms
Packet Parameter
The algorithm calculates the SYNC field length, EOP length, and the InterPacket gap. The acceptable range of EOP for all transmitted packets (except
SOFs) must be between 7.5 and 8.5 bits. The packet parameter algorithm
calculates and verifies the EOP depending on this range.
The acceptable range of inter-packet gap should be between 88 bits to 192 bits.
The SYNC field for all transmitted packets is calculated by counting the bits to
check for 32-bit SYNC field. The inter-packet delay and SYNC status is
calculated between the respective EOP indexes of the packets. The EOP width in
time is calculated by dividing the EOP width by the signaling rate of 480 Mbps.
Oscilloscope setup details
The application sets the oscilloscope automatically to the following values:
Horizontal Scale: 400 ns
Record Length: 50 K
Vertical Scale: 200 mV
Trigger Type Pulse Width
Trigger Hold Off: 5 s
200
TDSUSB2 Universal Serial Bus Measurement Package
Measurement Algorithms
Resume
This test calculates the device/HUB resume High-speed operation, indicated by
High-speed SOF packets (with 400 mV nominal amplitude) following the K state
driven by the host controller. For the Host, this is the time between the falling
edge of D+ and the First SOF. This should not exceed 3.0mS. To get the Suspend
signal, hot-plug the unit under test (device) and measure the signaling with
single-ended probes on both lines. The application analyzes data for the
following:
Resume Time
Amplitude
Suspend
This test calculates the time between the end of last SOF and the rising edge
transition to Full-speed J state for Host / Device / Hub- upstream. This time must
be between 3 ms and 3.125 ms. To get the Suspend signal, hot-plug the unit
under test (device) and measure the signaling with single-ended probes on both
lines. The application analyzes data for the Suspend Time.
TDSUSB2 Universal Serial Bus Measurement Package
201
Measurement Algorithms
Reset from Suspend
This test calculates the time between the falling edge of D+ signal and the start of
Device chirp-K for the Device/HUB upstream. This must be between 2.5 us and
3 ms. To get the Reset from Suspend Measurement signal, hot-plug the unit
under test (device), and measure the signaling with single-ended probes on both
lines. The application analyzes data for the Reset From Suspend Time.
202
TDSUSB2 Universal Serial Bus Measurement Package
Measurement Algorithms
Reset from High Speed
This test calculates the time between the beginning of the last SOF and before the
reset and the beginning of Chirp-K for Device and HUB upstream. This must be
between 3.1 ms and 6 ms. To get the Reset from High-Speed Measurement
signal, hot-plug the unit under test (device) and measure the signaling with
single-ended probes on both lines. The application analyzes data for Reset From
High Speed Time.
TDSUSB2 Universal Serial Bus Measurement Package
203
Measurement Algorithms
Equipment Setup for Tests
Low Speed Downstream Signal Quality Host Equipment Setup
To set up the equipment for Low Speed Downstream Signal Quality test, follow
these steps:
1.
Connect a cable between the A receptacle from the Inrush test section of the
test fixture and the USB device.
2.
Connect Ch1 of the D+ probe to the D+ pins on the Inrush section of the test
fixture.
3.
Connect Ch2 of the D– probe to the D– pins on the Inrush section of the test
fixture.
4.
Use the connectors to connect the A pin dongle from the Inrush section of
the test fixture to any port of the unit under test (host or PC).
5.
Select the measurement and select the
application.
6.
Select OK after acquiring a waveform. Verify that it is a correct waveform.
command button to run the
If the signal is clipped, follow these steps to increase the vertical scale:
204
1.
In the oscilloscope menu, select Vertical>Vertical Setup to display the
Channel screen.
2.
In the Scale field, increase the vertical scale values until the waveform is
completely displayed on screen.
TDSUSB2 Universal Serial Bus Measurement Package
Measurement Algorithms
Figure 120: Equipment setup for Low Speed Downstream Signal Quality Host
Low Speed Signal Quality for HUB Downstream
To set up the equipment for Low Speed Signal Quality (hub) test, follow these
steps:
1.
Connect the A plug dongle from the Inrush section of the test fixture to the
port of the unit under test (hub).
2.
Connect the low speed test device to the A receptacle on the Inrush section
of the test fixture. Keep the Discharge switch always in the ON position.
3.
Connect Ch1 of the D+ probe to the D+ pins on the Inrush section of the test
fixture.
4.
Connect Ch2 of the D– probe to the D– pins on the Inrush section of the test
fixture.
5.
Configure the measurement and select the
application.
6.
Select OK after acquiring a waveform. Verify that it is a correct waveform.
command button to run the
If the signal is clipped, follow these steps to increase the vertical scale:
1.
In the oscilloscope menu, select Vertical>Vertical Setup to display the
Channel screen.
2.
In the Scale field, increase the vertical scale values until the waveform is
completely displayed on screen.
TDSUSB2 Universal Serial Bus Measurement Package
205
Measurement Algorithms
Figure 121: Equipment setup for Low Speed Signal Quality for HUB Downstream
Full Speed Signal Quality HUB Downstream Setup
To set up the equipment for Full Speed Signal Quality Downstream test, follow
these steps:
206
1.
Set the S6 switch to the Init position.
2.
Use the adapters to connect the A receptacle from Device SQ test section
(marked DUT) of the test fixture to the USB2.0 Low Speed device.
3.
Connect Ch1 of the D+ probe to the D+ pins on the Device SQ section of the
test fixture.
4.
Connect Ch2 of the D– probe to the D– pins on the Device SQ section of the
test fixture.
5.
Connect the Init port of the Device SQ section of the test fixture to any port
of the unit under test (hub) using the USB cable.
6.
Select the measurement and select the
application.
7.
Select OK after acquiring a waveform. Verify that it is a correct waveform.
command button to run the
TDSUSB2 Universal Serial Bus Measurement Package
Measurement Algorithms
Note: Make sure the acquired signal is a valid waveform.
If the signal is clipped, follow these steps to increase the vertical scale:
1.
In the oscilloscope menu, select Vertical>Vertical Setup to display the
Channel screen.
2.
In the Scale field, increase the vertical scale values until the waveform is
completely displayed on screen.
Figure 122: Equipment setup for Full Speed Signal Quality HUB Downstream
TDSUSB2 Universal Serial Bus Measurement Package
207
Measurement Algorithms
Full Speed Downstream Host Equipment Setup
To set up the equipment for Full Speed Downstream (host) test, follow these
steps:
1.
Set the S6 switch to the Init position.
2.
Connect the A receptacle from Device SQ test section (marked DUT) of the
test fixture to the hub system. Connect the Full Speed test device to Hub 5.
3.
Connect Ch1 of the D+ probe to the D+ pins on the Device SQ section of the
test fixture.
4.
Connect Ch2 of the D– probe to the D– pins on the Device SQ section of the
test fixture.
5.
Use the standard USB cable to connect the Device SQ section of test fixture
to the test port of the unit under test (host).
6.
Select the measurement and select the
application.
7.
Select OK after acquiring a waveform. Verify that it is a correct waveform.
command button to run the
If the signal is clipped, follow these steps to increase the vertical scale:
208
1.
In the oscilloscope menu, select Vertical>Vertical Setup to display the
Channel screen.
2.
In the Scale field, increase the vertical scale values until the waveform is
completely displayed on screen.
TDSUSB2 Universal Serial Bus Measurement Package
Measurement Algorithms
Figure 123: Equipment setup for Full Speed Downstream Host
Full Speed Upstream Signal Quality Setup
To set up the equipment for Full Speed Upstream test, follow these steps:
1.
Use the A receptacle to connect the USB unit under test (device) to the
Inrush section of the test fixture.
2.
Connect the Qualifier device to the Adjacent Trigger and Droop section of
the test fixture as shown in the following figure.
3.
Connect Ch1 of the D+ probe to the D+ pins on the Inrush section of the test
fixture.
4.
Connect Ch2 of the D– probe to the D– pins on the Inrush section of the test
fixture.
5.
Connect the D+ (D– for Low speed) pin of the Adjacent Trigger and Droop
Section of the test fixture to Ch3 as shown in the following figure.
6.
Use the connectors to connect the A pin dongle from the Adjacent Trigger
and Droop section of the test fixture to one port of Hub 5. Use the A pin
dongle from the Inrush Droop section of the test fixture to another port of
Hub 5.
7.
Select the measurement and select the
application.
8.
Select OK after acquiring a waveform. Verify that it is a correct waveform.
TDSUSB2 Universal Serial Bus Measurement Package
command button to run the
209
Measurement Algorithms
Note: Use the standard USB cables to connect between the hubs. Keep the
Discharge switch in the Inrush Droop section in the ON position.
If the signal is clipped, follow these steps to increase the vertical scale:
1.
In the oscilloscope menu, select Vertical>Vertical Setup to display the
Channel screen.
2.
In the Scale field, increase the vertical scale values until the waveform is
completely displayed on screen.
Figure 124: Equipment setup for Full Speed Upstream testing
210
TDSUSB2 Universal Serial Bus Measurement Package
Measurement Algorithms
High Speed Device Signal Quality Setup
To set up the equipment for the High Speed Signal Quality test, follow these
steps:
1.
Set the S6 switch to the Init position.
2.
Connect the standard USB cable between the Device SQ Init port and the
host port.
3.
Connect the A receptacle (marked DUT) from the Device SQ test port of the
test fixture to the unit under test (device).
4.
Configure the measurement and select the
application.
5.
Place the device in the test mode Test_Packet from the host controller.
6.
Set the S6 switch to the test position to isolate the unit under test while
maintaining the bus power.
7.
Select OK after acquiring a waveform. Verify that it is a correct waveform.
command button to run the
Figure 125: Equipment setup for High Speed Device testing
TDSUSB2 Universal Serial Bus Measurement Package
211
Measurement Algorithms
Chirp Test Equipment Setup
The section used for this device test is Device SQ in the test fixture. To set up the
equipment for Chirp test, follow these steps:
1.
Set the S6 switch to the Init position.
2.
Use a standard USB cable with an A plug on one end and B plug on the
other end. Connect one end of the cable to the B socket on the Init port of
Device SQ section and the other end to the host port A socket.
3.
Connect the A receptacle from the Device SQ test port to the unit under test
(device).
4.
Connect the single-ended probes of the oscilloscope to the D+ and D– pins.
5.
Select the measurement and select the
application.
6.
Disconnect and connect the unit under test (device) to the port and observe
the chirp signal on the oscilloscope.
7.
Select OK after acquiring a waveform. Verify that it is a correct waveform.
command button to run the
Note: To avoid false triggering for the chirp signals while operating the test
fixture, it is recommended that you place the switch in the Init position and
connect the unit under test. This disables the switch bounce to the trigger.
212
TDSUSB2 Universal Serial Bus Measurement Package
Measurement Algorithms
Figure 126: Equipment setup for Chirp Test
TDSUSB2 Universal Serial Bus Measurement Package
213
Measurement Algorithms
Receiver Sensitivity Setup
To set up the equipment for Receiver Sensitivity test, follow the procedural steps
as given below in the application.
Figure 127: Equipment setup for Receiver Sensitivity Test
214
TDSUSB2 Universal Serial Bus Measurement Package
Measurement Algorithms
Procedural steps
Figure 128: Receiver Sensitivity analysis
Select View Procedure to view the procedural steps. Click the Overall Result to
display the following figure:
TDSUSB2 Universal Serial Bus Measurement Package
215
Measurement Algorithms
216
TDSUSB2 Universal Serial Bus Measurement Package
Measurement Algorithms
Inrush setup
To set up the equipment for Inrush test, follow these steps:
1.
Use the dongle on the Inrush section of the test fixture to connect it to the
host system.
2.
Connect the current probe between the VBUS loop wire on the Inrush section
on the Test fixture and Ch1 of the oscilloscope.
3.
Select the measurement and select the
application.
4.
Connect the unit under test to the A Receptacle of the Inrush section of the
test fixture and observe the Inrush current signal.
5.
Select OK after acquiring a waveform. Verify that it is a correct waveform.
command button to run the
Note: To avoid the triggering of Inrush signals and false inrush current by the
discharge switch, place the inrush discharge switch in the ON position and hotplug the unit under test (device). If a valid Inrush signal is not acquired, use the
cursor mode in File> Preferences or setup the oscilloscope to get a valid
waveform.
TDSUSB2 Universal Serial Bus Measurement Package
217
Measurement Algorithms
Figure 129: Equipment setup for Inrush test
Resume Test Equipment Setup
To set up the Device SQ in the test fixture for the Resume test, follow these
steps:
218
1.
Set the S5 switch to the Init position.
2.
Use a standard length of the USB cable with an A plug on one end and a B
plug on the other end. Connect one end of the cable to the B socket on the
Init port of the Device SQ section and the other end to the host port A
socket.
3.
Connect the A receptacle from the Device SQ test port to the unit under test
(device).
4.
Connect the single-ended probes of the oscilloscope to the D+ and D– pins.
5.
Select the Resume measurement in the application, configure its options and
select the Run button to run the application.
6.
Select OK after acquiring a waveform. Verify that it is a correct waveform.
TDSUSB2 Universal Serial Bus Measurement Package
Measurement Algorithms
Figure 130: Equipment setup for Resume test
Reset from Suspend Test Equipment Setup
To set up the Device SQ in the test fixture for the Reset from Suspend test,
follow these steps:
1.
Set the S6 switch to the Init position.
2.
Use a standard length of the USB cable with an A plug on one end and a B
plug on the other end.
3.
Connect one end of the cable to the B socket on the Init port of the Device
SQ section and the other end to the host port A socket.
4.
Connect the A receptacle from the Device SQ test port to the unit under test
(device).
5.
Connect the single-ended probes of the oscilloscope to the D+ and D– pins.
6.
Select the Reset from Suspend measurement from the application, configure
its options, and select the Run button to run the application.
7.
Select OK after acquiring a waveform. Verify that it is a correct waveform.
TDSUSB2 Universal Serial Bus Measurement Package
219
Measurement Algorithms
Figure 131: Equipment setup for Reset from Suspend test
Suspend Test Equipment Setup
To set up the Device SQ in the test fixture for the Suspend test, follow these
steps:
220
1.
Set the S6 switch to the Init position.
2.
Use a standard length of USB cable with an A plug on one end and a B plug
on the other end.
3.
Connect one end of the cable to the B socket on the Init port of the Device
SQ section and the other end to the host port A socket.
4.
Connect the A receptacle from the Device SQ test port to the unit under test
(device).
5.
Connect the single-ended probes of the oscilloscope to the D+ and D– pins.
6.
Select the Suspend measurement, configure its options and select the Run
button to run the application.
7.
Select OK after acquiring a waveform. Verify that it is a correct waveform.
TDSUSB2 Universal Serial Bus Measurement Package
Measurement Algorithms
Figure 132: Equipment setup for Suspend test
Reset from High Speed Test Equipment Setup
To set up the Device SQ in the test fixture for the Reset from High Speed test,
follow these steps:
1.
Set the S6 switch to the Init position.
2.
Use a standard length of the USB cable with an A plug on one end and a B
plug on the other end.
3.
Connect one end of the cable to the B socket on the Init port of the Device
SQ section and the other end to the host port A socket.
4.
Connect the A receptacle from the Device SQ test port to the unit under test
(device).
5.
Connect the single-ended probes of the oscilloscope to the D+ and D– pins.
6.
Select the Reset from High Speed from the application, configure its options
and select the Run button to run the application.
7.
Select OK after acquiring a waveform. Verify that it is a correct waveform.
TDSUSB2 Universal Serial Bus Measurement Package
221
Measurement Algorithms
Figure 133: Equipment setup for Reset from High Speed test
Packet Parameter Test Equipment Setup
Test Fixture Setup
To set up the test fixture, follow these steps:
222
1.
Set the S5 switch to the Init position.
2.
Connect the standard USB cable between the Device SQ Init port and the
host port.
3.
Connect the A receptacle (marked DUT) from the Device SQ test port to the
B receptacle device.
4.
Apply the power to the test fixture.
5.
Apply the power to the DUT.
6.
Attach the differential probe near the device connector on the test fixture.
TDSUSB2 Universal Serial Bus Measurement Package
Measurement Algorithms
Set up the Oscilloscope for High Speed Host
1.
Select the measurement and run the application. The application
automatically sets the oscilloscope parameters (Horizontal, Vertical, and
Trigger) and displays the message, "Press OK when correct waveform is
acquired”. You can see the SOFs on the oscilloscope screen.
2.
If you are not able to acquire the waveform automatically, perform
Autosetup in the oscilloscope to display the SOF.
3.
If your DUT is EL_21, EL_23, EL_25, select the Single Step Set Feature
option from the HS electrical test tool.
4.
If your DUT is EL_22, select the Step button in the HS electrical test tool to
acquire the waveform as shown in the following figure:
Figure 134: Setting the trigger level
TDSUSB2 Universal Serial Bus Measurement Package
223
Measurement Algorithms
You can use the application to perform the following measurements:
High Speed Host: EL_21, EL_23, EL_25 (EOPII-InterPacketI&II)
a.
Complete procedures 1 and 2 to set up the test fixture and the
oscilloscope.
b.
The oscilloscope acquires and displays the waveform as shown in the
following figure.
c.
Press OK when correct waveform is acquired.
d.
The application measures the synchronous bits (32) of the first and
second packets. This is EL_21.
e.
The application measures the EOP of the second packet (8bits). This is
EL_25. As the signal is differential, the EOP can be a positive or a
negative pulse.
f.
The application measures the inter-packet gap of the first two packets
(88-192 bits). This is EL_23 as shown in the following figure:
Figure 135: Measuring EL_23
224
TDSUSB2 Universal Serial Bus Measurement Package
Measurement Algorithms
High Speed Host: EL_22 (InterPacketII&III)
a.
Complete procedures 1 and 2 to set up the test fixture and the
oscilloscope.
b.
Select the Step button in the HS Electrical Test Tool.
c.
The oscilloscope acquires and displays the waveform as shown in the
following figure:
d.
Press OK when correct waveform is acquired.
Figure 136: Acquiring the waveform
TDSUSB2 Universal Serial Bus Measurement Package
225
Measurement Algorithms
Figure 137: Setting the trigger level
e.
The application displays the waveform.
f.
The application measures the inter packet gap of the second and the
third packet (88-192 bits). This is EL_22.
High speed Host: EL_55 (SOF-EOP)
226
a.
Complete the procedure 1.
b.
Run the test mode software and select the Host option.
c.
Select the Enumerate button in the HS electrical test tool to identify the
device connected to the host controller.
d.
Select the measurement and run the application. The application
automatically sets the oscilloscope parameters (Horizontal, Vertical,
and Trigger). You can see the SOFs on the oscilloscope screen.
TDSUSB2 Universal Serial Bus Measurement Package
Measurement Algorithms
Figure 138: Viewing SOF
TDSUSB2 Universal Serial Bus Measurement Package
227
Measurement Algorithms
To set up the oscilloscope for the High Speed Device EL_21, EL_23, EL_25
(Sync-EOPIII-InterPacketII&III) and a High Speed Device EL_22
(InterPacketI&II), follow these steps:
1.
Select the measurement and run the application. The application
automatically sets the oscilloscope parameters (Horizontal, Vertical, and
Trigger) and displays the message, "Press OK when correct waveform is
acquired”. You can see the SOFs on the oscilloscope screen.
2.
If you are not able to acquire the waveform automatically, perform
Autosetup in the oscilloscope to display the SOF.
3.
If your DUT is EL_21, EL_23, EL_25, select the Single Step Set Feature
option from the HS electrical test tool.
4.
If your DUT is EL_22, select the Step button in the HS electrical test tool to
acquire the waveform as shown in the following figure:
Figure 139: Setting the trigger level
228
TDSUSB2 Universal Serial Bus Measurement Package
Measurement Algorithms
You can use the application to perform the following measurements:
High Speed Device:EL_21, EL_23, EL_25 (Sync-EOPIII-InterPacketII&III)
a.
Complete procedures 1 and 3 to set up the oscilloscope.
b.
The oscilloscope acquires and displays the waveform.
c.
“Press OK when correct waveform is acquired".
d.
The application measures the synchronous bits (32) of the third packet.
This is EL_21.
e.
The application measures the EOP of the third packet (8bits). This is
EL_25. As the signal is differential, the EOP can be a positive or a
negative pulse.
f.
The application measures the inter-packet gap between the second and
the third packets (88-192 bits). This is EL_23 as shown in the
following figure:
Figure 140: Measuring EL_23
High Speed Device:EL_22 (InterPacketI&II)
TDSUSB2 Universal Serial Bus Measurement Package
229
Measurement Algorithms
a.
Complete procedures 1 and 3 to set up the oscilloscope.
b.
Select the Step button in the HS electrical test tool to acquire the
waveform as shown in the following figure:
Figure 141: Acquiring the waveform
230
TDSUSB2 Universal Serial Bus Measurement Package
Measurement Algorithms
Figure 142: Setting the trigger level
c.
The application measures the number of bits (88-192) between the
packets.
TDSUSB2 Universal Serial Bus Measurement Package
231
Measurement Algorithms
Figure 143:Equipment setup for Packet Parameter Measurement
232
TDSUSB2 Universal Serial Bus Measurement Package
Glossary
Cross-Over Points
The Cross-Over point is defined as the intersection of the D+ and D– singleended signals. For differential signals, the Cross-Over is zero crossings of the
differential signal.
Downstream
The direction of data flow from the host or away from the host. A downstream
port is the port on a hub electrically farthest from the host that generates
downstream data traffic from the hub. Downstream ports receive upstream data
traffic.
EOP
End-of-Packet.
EOP (E), Idle (I), J, K
The different bus states of the USB signal.
Eye Pattern
A representation of the USB signal that provides minimum and maximum
voltage levels, as well as the signal jitter.
Eye Violation
Any part of the waveform that crosses the defined eye mask.
Full-speed
USB operation at 12 Mb/s.
High-speed
USB operation at 480 Mb/s.
Host
The host computer system where the USB Host Controller is installed. This
includes the host hardware platform (CPU, bus, etc.) and the operating system in
use.
Host Controller
The host’s USB interface.
TDSUSB2 Universal Serial Bus Measurement Package
233
Glossary
Hot-Plug
It is the technology that supports automatic configuration of the PC hardware and
the attached device. You can attach a device or hot plug and start working
without having to manually configure the device. This is how it is referred to in
the application.
HUT
A Host/Hub Under Test.
Hub
A USB device that provides additional connections.
Low-speed
USB operation at 1.5 mb/s.
mb/s
It is the transmission rate expressed in megabits per second.
NAK
A handshake packet indicating a negative acknowledgment.
Pulse Diagram
The Plot of the USB signals showing Annotations of J,K, EOP (E) , Idle (I),
Cross-Over Points (Cov) and Reference Points (Ref).
Reference point
The simulation of ideal Cross-Over points, which helps calculate jitter.
Reflectometer
An oscilloscope capable of measuring impedance characteristics of the USB
signal lines.
Sample Rate
It is the number of samples per second, expressed in Hertz (Hz).
Signal Direction Downstream
It is defined as the direction of data flow away from the host. A downstream port
is the port on a hub farthest from the host that generates downstream data traffic
from the hub. Downstream ports receive upstream data traffic.
234
TDSUSB2 Universal Serial Bus Measurement Package
Glossary
Signal Direction Upstream
It is defined as the direction of data flow towards the host. An upstream port is
the port on a device closest to the host that generates upstream data traffic from
the hub. Upstream ports receive downstream data traffic.
Tier
The position in the hub where the device is connected to the system.
Test Point
A device is classified as far end or near end depending upon the captive cable. A
device with captive cable is usually called as a far end device, otherwise as a near
end device.
Test Fixture
It is the break-out board that helps in probing signals.
UUT
Called as the Unit Under Test. The unit can be a USB device, hub, port or a host.
Upstream
The direction of data flow is towards the host. An upstream port is the port on a
device electrically closest to the host that generates upstream data traffic from the
hub. Upstream ports receive downstream data traffic.
VBUS
It is the supply voltage which a function or hub requires to work.
TDSUSB2 Universal Serial Bus Measurement Package
235
Glossary
236
TDSUSB2 Universal Serial Bus Measurement Package
Download PDF