HomeFill Oxygen Conserving Cylinder

Operator’s Manual
HomeFill™
Oxygen Conserving
Cylinder
HF2PCL6
HF2PC9
HF2PCL4
Dealer: This manual MUST be given to the end
user.
User: Before using this cylinder, read this
manual and save for future reference.
For more information regarding
Invacare products, parts, and services,
please visit www.invacare.com
WARNING
WARNING
DO NOT OPERATE THIS PRODUCT OR ANY
AVAILABLE OPTIONAL EQUIPMENT WITHOUT
FIRST COMPLETELY READING AND
UNDERSTANDING THESE INSTRUCTIONS AND
ANY ADDITIONAL INSTRUCTIONAL MATERIAL
SUCH AS OWNER’S MANUALS, SERVICE
MANUALS OR INSTRUCTION SHEETS SUPPLIED
WITH THIS PRODUCT OR OPTIONAL
EQUIPMENT. IF YOU ARE UNABLE TO
UNDERSTAND THE WARNINGS, CAUTIONS, OR
INSTRUCTIONS, CONTACT A HEALTHCARE
PROFESSIONAL, DEALER OR TECHNICAL
PERSONNEL BEFORE ATTEMPTING TO USE
THIS EQUIPMENT - OTHERWISE INJURY OR
DAMAGE MAY OCCUR.
ACCESSORIES WARNING
Invacare products are specifically designed and
manufactured for use in conjunction with Invacare
accessories. Accessories designed by other manufacturers
have not been tested by Invacare and are not
recommended for use with Invacare products.
NOTE:Updatedversionsofthismanualareavailableon
www.invacare.com.
HomeFill™
2
Part No 1112648
TABLE OF CONTENTS
TABLE OF CONTENTS
SPECIAL NOTES ..........................................................4
RECEIVING INSPECTION .............................................5
INTENDED USE ...........................................................5
SECTION 1—GENERAL GUIDELINES .............................6
Operating Information.............................................................................................6
Maintenance ................................................................................................................8
SECTION 2—FEATURES ...............................................9
SECTION 3—TYPICAL PRODUCT PARAMETERS ........10
SECTION 4—INSPECTION AND COMPRESSOR
INTERFACE ..............................................................11
Cylinder Prefill Inspection.................................................................................... 11
Connecting/Disconnecting the Cylinder to/from the Compressor....... 14
SECTION 5—OPERATING INSTRUCTIONS .................19
SECTION 6—MAINTENANCE ....................................21
SECTION 7—CYLINDER FILL TIMES ............................22
Troubleshooting...................................................................................................... 22
Replacement Parts - Accessories ...................................................................... 22
LIMITED WARRANTY ...............................................23
Part No 1112648
3
HomeFill™
SPECIAL NOTES
SPECIAL NOTES
Signalwordsareusedinthismanualandapplytohazardsor
unsafepracticeswhichcouldresultinpersonalinjuryor
propertydamage.Refertothefollowingtablefordefinitions
ofthesignalwords.
SIGNAL WORD
MEANING
DANGER
Danger indicates an imminently hazardous
situation which, if not avoided, will result in
death or serious injury.
WARNING
Warning indicates a potentially hazardous
situation which, if not avoided, could result
in death or serious injury.
CAUTION
Caution indicates a potentially hazardous
situation which, if not avoided, may result in
property damage or minor injury or both.
NOTICE
Theinformationcontainedinthisdocumentissubjectto
changewithoutnotice.
DANGER
DONOTSMOKEwhileusingthisdevice.Keepall
matches,litcigarettesorothersourcesofignitionoutofthe
roominwhichthisproductislocatedandawayfromwhere
oxygenisbeingdelivered.
NOSMOKINGsignsshouldbeprominentlydisplayed.
Textilesandothermaterialsthatnormallywouldnotburn
areeasilyignitedandburnwithgreatintensityinoxygen
enrichedair.Failuretoobservethiswarningcanresultin
severefire,propertydamageandcausephysicalinjuryor
death.
HomeFill™
4
Part No 1112648
RECEIVING INSPECTION
CAUTION
“Caution:Federallawrestrictsthisdevicetosalebyorrental
orontheorderofaphysician,oranyotherpractitioner
licensedbythelawofthestateinwhichhe/shepracticesto
useorordertheuseofthisdevice.”
RECEIVING INSPECTION
Removeproductfrompackageandinspectfordamage.If
thereisanydamage,DONOTuse.Contactyourdealer/
HomeCareequipmentprovider.
INTENDED USE
Tostore,regulate,anddeliversupplementaloxygento
patientswhomayhavedifficultyextractingoxygenfromthe
airtheybreathe.Itisforpatientswhowouldnormallyreceive
theoxygenviaanasalcannula.ThedevicedeliversUSP93%
oxygenatanumberofdifferentflow/pulsesettings.Itis
intendedtobeusedasanoxygensavingdevicethatmay
reducethedryingoutoftheairways.
Part No 1112648
5
HomeFill™
SECTION 1—GENERAL GUIDELINES
SECTION 1—GENERAL GUIDELINES
Inordertoensurethesafeinstallation,assemblyand
operationoftheHomeFillOxygenConservingCylinderthese
instructionsMUSTbefollowed.
WARNING
This section contains important information for the safe operation
and use of this product.
Operating Information
Onlypersonnelinstructedandtrainedintheuseofthe
HomeFillO2ConservingCylindershouldoperatethis
product.
Productcontainsmagnetic,ferrousmaterialthatmayaffect
theresultsofanMRI.
Besureallconnectionsaretightandleakfree.
DONOTautoclave.
DONOTgassterilizewithethyleneoxide.
DONOTcleanwitharomatichydrocarbons.
DONOTimmerseproductinanykindofliquid.
Storeproductinacleanareawhennotisuse.
OnlyuseInvacareCorporationHF2PCL6carryingbag,
HF2PC9carryingbag,orHF2PCL4carryingbag.
NEVERsmokeinanareawhereoxygenisbeing
administered.
ALWAYSconfirmprescribeddosebeforeadministeringto
patientandmonitoronafrequentbasis.
DONOTuseifdirtorcontaminantsarepresentonoraround
cylinder,valve,productorconnectingdevices.
HomeFill™
6
Part No 1112648
SECTION 1—GENERAL GUIDELINES
NEVERusenearanytypeofflameorflammable/explosive
substances,vaporsoratmosphere.
ALWAYSfollowANSIandCGAstandardsforMedicalGas
ProductsandFlowmeters(E7)andOxygenHandling(G4).
NOOXYGENisdeliveredwhenthepointerisalignedwith
“Off”.
DONOTblocktheoutletfittingorkinkthecannulawhenthe
productisinuse,thismaydamagetheunit.
Theuseofthisdeviceislimitedtotheoxygenpatient.
CylindersMUSTbeusedonlybytheoxygenpatientandare
nottobedistributedtoanyotherindividualforanypurpose.
ThisproductisNOTtobeusedbypatientswhobreathe
throughtheirmouths.
DONOTuseoils,greases,organiclubricantsorany
combustiblematerialsonornearthisproduct.Washanddry
handsproperlypriortousage.
DONOTallowoilfromyourhandsorothersourcestocome
intocontactwiththecompressorcouplingorbottlefittings.
Thesesubstancesmaybecomeflammableinthepresenceof
oxygenandcauseinjury.Avoidtouchingthebottlefitting
wheneverpossible.Exampleofitemstoavoidarehand
creamsorlotions,cookingoils,suntanoilsorsunblocksand
similaroilyproducts.Ifyouusethesetypesofproducts,wash
yourhandspriortousingtheHomeFillunit.Ifthebottle
fittingshouldcomeincontactwithsuchsubstancesitshould
becleanedoffwithadampclothpriortoconnectingittothe
compressor.
NEVERoilorlubricatethecompressorcouplerorbottle
fittings.NEVERuseanypenetratingoilorlubricantsuchas
WD40®or3in1oil®.Ifconnectiondifficultyisexperienced
verifythatthebottlefittingisbeinginsertedstraightdown
intothecompressorcoupler.Ifcontinueddifficultyis
experiencedcontactyoursupplierforassistance.
Part No 1112648
7
HomeFill™
SECTION 1—GENERAL GUIDELINES
DONOTuseahumidifierwiththisproduct.
DONOTallowcylinderstotiporfall.Securegascylindersso
theycannotfall.Ensurecylinderisalwaysinuprightposition.
DONOTstorecylindersnearsourcesofheatorflame.
Thisproductisequippedwithareliefvalve.Ifyouhearaloud
hissingorpoppingsoundcomingfromtheproduct,
discontinueuseofproduct,turntheflowselectortothe“Off”
position,andcontactyourdealer/HomeCareequipment
provider.
DONOTusewhilesleepingwithoutconsultingyour
healthcareprovider.
Thisproductisdesignedtooperatewithastandardsingle
lumen,adultoxygennasalcannulawithamaximumlengthof
sevenfeet.
Thecannulaisforsinglepatientuseonly.
CannulaMUSTbeinstalledonoutletconnectionpriortouse.
DONOTplacetheunitunderclothingwhileinuse.When
thisproductisinuseasmallamountofoxygenisvented.
Wearingtheunitunderclothingmaysaturatefabricswith
oxygenandcausethemtoburnrapidlyifexposedtosparks
orflame.Itmaytakeseveralhoursforoxygenlevelsinfabrics
toreturntonormal.
Maintenance
DONOTusealiquidleakdetectortotestforleaks.
DONOTattempttorepairproduct.
HomeFill™
8
Part No 1112648
SECTION 2—FEATURES
SECTION 2—FEATURES
DANGER
Thisproductisnotintendedasalifesustainingor
lifesupportingdevice.
Outlet
Connection
Cannula
Flow Dial
Cylinder Fill
Port
Cylinder
Gauge
FIGURE 2.1 Product Diagram
Part No 1112648
9
HomeFill™
SECTION 3—TYPICAL PRODUCT PARAMETERS
SECTION 3—TYPICAL PRODUCT
PARAMETERS
CONSERVER S/N RANGE
1-69,999
75,000 +
INLET PRESSURE RANGE:
300 - 2100 psig
300 - 2100 psig
50 psig
50 psig
HF2PCL6: Weight
4.3 lbs (1.95 kg)
4.1 lbs (1.86 kg)
HF2PCL6: Overall Length
11.1 in (28 cm)
10.9 in (27.7 cm)
HF2PCL6: Diameter
4.4 in (11.1 cm)
4.4 in (11.1 cm)
HF2PC9: Weight
5.3 lbs (2.4 kg)
5.1 lbs (2.31 kg)
HF2PC9: Overall Length
14.2 in (36 cm)
14 in (35.5 cm)
HF2PC9: Diameter
4.4 in (11.1 cm)
4.4 in (11.1 cm)
HF2PCL4: Weight
N/A
3.6 lbs (1.63 kg)
HF2PCL4:Overall Length
N/A
8.9 in (22.2 cm)
HF2PCL4:Diameter.
N/A
4.4 in (11.1 cm)
INTERNAL REGULATED
PRESSURE:
DIMENSIONS:
PULSE SETTINGS:
1,2,3,4, & 5
FLOW SETTINGS:
2 LPM Continuous
SAVINGS RATIO:
Up to 4.3:1
Up to 5.7:1
CANNULA
REQUIREMENT:
4 ft recommended (MS0204) (7 ft long max)
std. adult single lumen oxygen nasal cannula.
OPERATING CONDITIONS TEMPERATURE:
32°F (0°C) to 105°F
(40.6°C)
35°F (1.7°C) to 105°F
(40.6°C)
0 to 5280 ft (0
to1600 m)
0 to 10,000 ft (0 to
3071 m)
-40°F (-40°C) to
140°F (60°C)
-40°F (-40°C) to 140°F
(60°C)
ALTITUDE:
STORAGE CONDITIONS
TEMPERATURE:
MAXIMUM HUMIDITY:
HomeFill™
95%
10
Part No 1112648
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE
SECTION 4—INSPECTION AND
COMPRESSOR INTERFACE
NOTE:RefertotheHomeFillCompressorOwner’sManualfor
compressoroperatinginstructions.
Cylinder Prefill Inspection
WARNING
AllcylindersMUSTbeinspectedbeforeattemptingtofill;
otherwise,injuryordamagemayoccur.
NEVERsmokeinanareawhereoxygenisbeing
administered.
NEVERusenearanytypeofflameorflammable/explosive
substances,vaporsoratmosphere.
NOTE:Forthisprocedure,refertoFIGURE4.1.
The Hydrostatic Testing Date
AluminumcylindersMUSTundergotestingeveryfiveyears.
WARNING
DONOTfillcylindersthathavenotbeentestedinthepast
fiveyears.Otherwise,seriouspersonalinjurymayresult.
Contactyourdealerforreplacement.
Part No 1112648
11
HomeFill™
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE
External Examination
1. Examinetheoutsideofthecylinderforthefollowing
conditions,andreplacethecylinderiftheyexist:
• Dentsordings
• ArcBurns
• OilorGrease
• Anyothersignsofdamagethatmightcauseacylinder
tobeunacceptableorunsafeforuse.
2. Examinethecylinderforevidenceoffireorthermal
damage.Evidenceincludescharringorblisteringofthe
paint,orotherprotectivecoatingorheatsensitive
indicator.Iffireorthermaldamageisfound,replacethe
cylinder.
3. Inspectthecylinderfillportforthefollowing,andreplace
iffound:
• Debris,oilorgrease
• Noticeablesignsofdamage
• Signsofcorrosioninsidethevalve
• Signsofexcessiveheatorfiredamage
HomeFill™
12
Part No 1112648
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE
Improper filling or abuse of this cylinder, or failure to follow these safety
instructions may cause serious bodily injury or death. High pressure
oxidizing gas vigorously accelerates combustion.
DO NOT alter or modify cylinder or related components.
DO NOT use oil or grease. Keep cylinder, cylinder valve, all equipment
and connections free of oil and grease to avoid violent ignition resulting
in serious bodily injury or death.
For medical applications. Use only as directed by a licensed
practitioner. Uninterrupted use of high concentrations of oxygen over
a long duration, without monitoring its effect on oxygen content of
arterial blood, may be harmful.
Oxygen Cylinders filled by the Invacare HomeFill oxygen
system shall be used for personal use only. "Not to be
filled for resale or use by professional users."
This cylinder is to be used with the Invacare HomeFill
oxygen system. This cylinder contains 93 ± 3% oxygen
produced by an oxygen concentrator using the pressure swing
adsorption process.
DO NOT handle cylinder or use contents until you are professionally
trained, including emergency procedures.
Use in accordance with Invacare HomeFill oxygen system Operator’s
Manual. Open valve slowly. Close valve after each use and when
empty. Secure cylinder during storage and use. No smoking in cylinder
area. Keep away from heat, flame and spark. Keep out of reach of
children. DO NOT drop.
Filling of this gas is performed by Invacare HomeFill oxygen system
Only.
Cylinders that have been refinished and/or exposed to elevated
temperatures must be hydrostatically tested before filling. Cylinder
temperature should not exceed 125°F (52°C). Aluminum cylinders
subject to the action of fire or heated to temperatures in excess of
350°F must be withdrawn from service and condemned by trained
personnel. DO NOT use caustic paint strippers. Valve and safety
relief devices must be removed and replaced by trained and
authorized personnel. DO NOT alter or change valve assembly.
Invacare Corporation
2101 E. Lake Mary Blvd.
Sanford, FL 32773
1-800-832-4707
DO NOT REMOVE THIS PRODUCT LABEL
Not for Commercial Resale
Part Number 1125090 Rev D
FIGURE 4.1 External Examination
Part No 1112648
13
HomeFill™
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE
Connecting/Disconnecting the Cylinder
to/from the Compressor
Connecting the Cylinder to the Compressor
WARNING
NEVERusetoolsofanykindtoconnect/disconnectthe
cylinderandthecompressor.Otherwise,severeinjuryand/or
damagemayoccur.
DONOTdropoxygencylinders.Usetwohandswhen
handling/transportingoxygencylinders.Otherwise,injuryor
damagemayoccur.
DONOTusealiquidleakdetectortotestforleaks.
NOTE:Forthisprocedure,refertoFIGURE4.2andFIGURE4.3.
1. Examinethecylindergauge.Ifthecylinderpressureisless
than1500psig.,proceedtoSTEP2(FIGURE4.3).
NOTE:Ifthecylinderpressureisgreaterthan1500psig.,DO
NOTattempttotopoffthiscylinder.Itmaynotfill.
2. SettheregulatorflowdialonthecylindertoOFF(FIGURE
4.3).
3. Removethecylinderandconnectorfillportcovers(if
present).
CAUTION
DONOTconnecttheoutlet(cannula)connectiontothe
HomeFillunit,otherwisetheconservingcylinderwillnot
workproperly.
4. MomentarilypushDOWNontheouterring(sleeve)ofthe
connectorfillporttoresettheconnectorasshownin
FIGURE4.2.
HomeFill™
14
Part No 1112648
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE
NOTE:Iftheouterring(sleeve)isintheUPposition
(GREENdotsnotvisible),theconnectorfillportwillnot
beabletoacceptthecylinderfillport.PushingDOWN
momentarilywillresettheconnectorfillport(GREEN
dotsvisible)toacceptthecylinderfillport.
PUSH
DOWN
Outer Ring (Sleeve)
Connector Fillport
GREEN Dots
BEFORE
COUPLING
CYLINDER,
PUSH DOWN
ON SLEEVE
UNTIL
GREEN DOTS
ARE VISIBLE.
FIGURE 4.2 Resetting Connector Fillport
5. Graspthecylinder/regulatorassemblyintheareabehind
thecylindergauge(FIGURE4.3).
6. Positionthecylinderinthecompressorcradle(FIGURE
4.3).
7. Alignthecylinderfillportwiththeconnectorfillport
(FIGURE4.3).
8. PullUPontheouterring(sleeve)oftheconnectorfillport
whilepushingDOWNonthecylinder/regulatorassembly
tocouplethecylinderfillportintotheconnectorfillport
(FIGURE4.3).
NOTE:Thecylinderisproperlyconnectedwhenanaudible“click”
isheard.
Part No 1112648
15
HomeFill™
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE
Compressor
Cylinder/Regulator
Assembly
Cylinder
Fillport
Connector
Fillport
Outer Ring
(Sleeve)
Cylinder
Gauge
PULL UP
Regulator
Cylinder
Cylinder
Gauge
Regulator
Flow Dial
Outlet (Cannula)
Connection
Cylinder
Fillport
Power
Switch
Compressor
Cradle
Outer Ring
(Sleeve)
Connector
Fillport
NOTE:Fillportcoversnotshownforclarity.
FIGURE 4.3 Connecting the Cylinder to the Compressor
HomeFill™
16
Part No 1112648
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE
Disconnecting the Cylinder from the Compressor
WARNING
NEVERusetoolsofanykindtoconnect/disconnectthe
cylinderandthecompressor.Otherwise,severeinjuryand/or
damagemayoccur.
NOTE:Forthisprocedure,refertoFIGURE4.4.
1. PressthecompressorpowerswitchtotheOff(O)position.
2. Graspthecylinder/regulatorassemblyintheareabehind
thecylindergauge.
3. Withtheotherhand,grasptheouterring(sleeve)ofthe
connectorfillportandpushDOWN.
4. Liftuponthecylinder/regulatorassemblytoremovefrom
theconnectorfillport.
Cylinder/Regulator
Assembly
Cylinder
Gauge
Compressor
Cylinder
Fillport
PUSH DOWN
Outer Ring
(Sleeve)
Connector
Fillport
FIGURE 4.4 Disconnecting the Cylinder from the Compressor
Part No 1112648
17
HomeFill™
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE
5. Whenthecylinderfillportisdisconnectedfromthe
connectorfillport,releasetheouterring(sleeve)ofthe
connectorfillportandusetwohandstoremovethe
cylinderfromthecompressorcradle.
6. Placethefillportcoversontotheconnectorandcylinder
fillports.
NOTE:Thefillportcoversshouldbeusedwhenevercylindersare
notbeingfilled.
HomeFill™
18
Part No 1112648
SECTION 5—OPERATING INSTRUCTIONS
SECTION 5—OPERATING
INSTRUCTIONS
WARNING
NEVERsmokeinanareawhereoxygenisbeing
administered.
NEVERusenearanytypeofflameorflammable/explosive
substances,vaporsoratmosphere.
NOTE:Forthisprocedure,refertoFIGURE5.1.
1. Priortoeachuse,inspectproductforvisibledamage.DO
NOTuseifanydamageisfound.
NOTE:Ifforanyreason,anylabelbecomesillegibleorlost,contact
yourdealer/HomeCareequipmentprovider.
2. Besureregulatorisinthe“Off”position.Align“Off”with
theindicatingpointer.
3. Positionthecylindersotheoxygencylindervalveoutletis
pointingawayfromtheuserandanyotherperson(s).
WARNING
DONOTdirectflowofoxygenatanyperson,orflammable
materialwhenadjustingtheflowdial.
4. Attachastandardadultsinglelumenoxygennasal
cannula,nolongerthansevenfeettotheproduct’soutlet
accordingtothecannulamanufacturer’sinstructions.DO
NOTusepediatric,lowflownasalcannulasoroxygen
maskswiththisproduct.
NOTE:Aunitwithoutcannulaattachedmaypulseifsetatposition
15.
Part No 1112648
19
HomeFill™
SECTION 5—OPERATING INSTRUCTIONS
5. Placethecannulaover
yourearsandposition
theprongsinyournose
asinstructedbyyour
HealthCareprovideror
cannulamanufacturer.
RefertoFIGURE5.1.
6. Turndialuntilindicating
pointerisalignedwith
theprescribedsetting.
FIGURE 5.1 Cannula
Placement
WARNING
Whenpointeroncontentsgaugeentersredzone,itis
recommendedtorefillthecylinderontheInvacareHomeFill
Compressor.
NOOXYGENisdeliveredinbetweensettings.Toobtain
desiredflow,theindicatingpointermustalignwithaspecific
numberonthedial.
Toavoidinjurytopatient,ALWAYSconfirmprescribed
settingbeforeadministeringtopatientandmonitorflowona
frequentbasis.
OnlyuseInvacareCorporationHF2PCL6carryingbag,the
HF2PC9carryingbag,ortheHF2PCL4carryingbag.
7. Breathethroughthenoseandfeelapulseofoxygenasyou
inhale.
NOTE:Whenonapulsesetting,thereisfloworapulseonlyatthe
beginningofeachbreath.Ifyoudonotfeelthepulseatthebeginning
ofeachbreath,checkthesetting.
NOTE:Becauseeachpatient’sbreathingpatternisdifferentandthe
environmentvaries,itmaybedifficulttofeelsomelowsetting
pulses.
HomeFill™
20
Part No 1112648
SECTION 6—MAINTENANCE
SECTION 6—MAINTENANCE
1. Aftereachuse,cleanexterioroftheproductwithadry,lint
freeclothonly.
2. Storeproductinacleanareafreefromgrease,oil,and
othersourcesofcontamination.
CAUTION
DONOTusecleaningsolutions.
DONOTimmerseproductinanykindofliquid.
AllrepairsMUSTbedonebyInvacareCorporation.
Part No 1112648
21
HomeFill™
SECTION 7—CYLINDER FILL TIMES
SECTION 7—CYLINDER FILL TIMES
NOTE:Allfillingtimesareapproximateandmayvaryon
environmentalconditions.
*
Cylinder
Filling
Times
Concentrator Flow Rate to Patient:
(Platinum 5 up to 3 L/min. and Platinum 10 up to 5
L/min)
HF2PC9
HF2PCL4
HF2PCL6
NORMAL
2 hr 20 min
60 min
1 hr 25 min
MAXIMUM
2 hr 45 min
1 hr 20 min
1 hr 50 min
Troubleshooting
IfConservingCylinderfailstofunction,consultthe
Troubleshootingguide.Ifproblemcannotbecorrected,
consultyourDealer/HomeCareequipmentprovider.
PROBLEM
No flow.
POSSIBLE CAUSE
SOLUTION
1. Regulator in “Off”
position.
1. Set to prescribed
setting.
2. Cylinder empty.
2. Replace/fill cylinder.
3. Dial not set at increment.
3. Set dial so indictor
points to a setting.
4. Conserver not sensing breath.
4. Check position of
cannula in nose. Do not
breathe through mouth.
5. Check to see if the
cannula is kinked.
Replacement Parts - Accessories
HF2PCL6CarryingBag,HF2PC9CarryingBag,orHF2PCL4
CarryingBag
HomeFill™
22
Part No 1112648
LIMITED WARRANTY
LIMITED WARRANTY
PLEASE NOTE: THE WARRANTY BELOW HAS BEEN DRAFTED TO
COMPLY WITH FEDERAL LAW APPLICABLE TO PRODUCTS
MANUFACTURED AFTER JULY 4, 1975.
This warranty is extended only to the original purchaser/user of our products.
This warranty gives you specific legal rights and you may also have other legal
rights which vary from state to state.
Invacare warrants this product to be free from defects in materials and
workmanship for a period of three years from date of purchase. If within such
warranty period any such product shall be proven to Invacare’s satisfaction to
be defective, such product shall be repaired or replaced, at Invacare’s option.
This warranty does not apply to problems arising from normal wear and tear
or failure to adhere to these instructions, nor does it include shipping charges
incurred in replacement part installation or repair of any such product.
Invacare’s sole obligation and your exclusive remedy under this warranty
shall be limited to such repair and/or replacement.
For warranty service, please contact the Invacare service department during
normal business hours at the toll-free telephone number identified on the
back cover of this manual.
Upon receiving notice of an alleged defect in a product, Invacare will issue a
serialized return authorization. It shall be the responsibility of the purchaser
to pack the product(s) or part(s) in an appropriate manner to avoid shipping
damage and return the product(s) or part(s), at the purchaser’s expense, to
either Invacare’s plant or service center as specified by Invacare in advance.
The defective product(s) or component part(s) must be returned for
warranty inspection using the serial number as identification within 30 days of
the return authorization date. DO NOT return any product(s) or part(s) to
our factory or service center without prior consent. C.O.D. shipments will be
refused – please pre-pay shipping charges.
LIMITATIONS AND EXCLUSIONS: THE FOREGOING WARRANTY
SHALL NOT APPLY TO SERIAL NUMBERED PRODUCTS IF THE SERIAL
NUMBER HAS BEEN REMOVED OR DEFACED; PRODUCTS SUBJECTED
TO NEGLIGENCE, ACCIDENT, IMPROPER OPERATION,
MAINTENANCE, OR STORAGE; PRODUCTS MODIFIED WITHOUT
INVACARE’S EXPRESS WRITTEN CONSENT (INCLUDING, BUT NOT
LIMITED TO, MODIFICATION THROUGH THE USE OF
UNAUTHORIZED PARTS OR ATTACHMENTS); PRODUCTS DAMAGED
BY REASON OF REPAIRS MADE TO ANY COMPONENT WITHOUT THE
SPECIFIC CONSENT OF INVACARE; PRODUCTS DAMAGED BY
CIRCUMSTANCES BEYOND INVACARE’S CONTROL; OR PRODUCTS
REPAIRED BY ANYONE OTHER THAN INVACARE OR AN INVACARE
SERVICE CENTER. SUCH EVALUATION SHALL BE SOLELY
DETERMINED BY INVACARE.
Part No 1112648
23
HomeFill™
THE FOREGOING EXPRESS WARRANTY IS EXCLUSIVE AND IN LIEU
OF ANY OTHER WARRANTIES WHATSOEVER, WHETHER EXPRESSED
OR IMPLIED, INCLUDING THE IMPLIED WARRANTIES OF
MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE,
AND THE SOLE REMEDY FOR VIOLATIONS OF ANY WARRANTY
WHATSOEVER, SHALL BE LIMITED TO REPAIR OR REPLACEMENT OF
THE DEFECTIVE PRODUCT PURSUANT TO THE TERMS CONTAINED
HEREIN. THE APPLICATION OF ANY IMPLIED WARRANTY
WHATSOEVER SHALL NOT EXTEND BEYOND THE DURATION OF
THE EXPRESS WARRANTY PROVIDED HEREIN. INVACARE SHALL
NOT BE LIABLE FOR ANY CONSEQUENTIAL OR INCIDENTAL
DAMAGES WHATSOEVER.
SOME STATES DO NOT ALLOW THE EXCLUSION OR LIMITATION OF
INCIDENTAL OR CONSEQUENTIAL DAMAGE, OR LIMITATION OF
HOW LONG AN IMPLIED WARRANTY LASTS, SO THE ABOVE
EXCLUSION AND LIMITATION MAY NOT BE APPLICABLE.
THIS WARRANTY SHALL BE EXTENDED TO COMPLY WITH
STATE/PROVINCIAL LAWS AND REQUIREMENTS.
Invacare Corporation
www.invacare.com
USA
One Invacare Way
Elyria, Ohio USA
44036-2125
440-329-6000
800-333-6900
Technical Services
440-329-6593
800-832-4707
© 2011 Invacare Corporation. All rights
reserved. Republication, duplication or
modification in whole or in part is
prohibited without prior written
permission from Invacare. Trademarks
are identified by ™ and ®. All trademarks are owned by or licensed to
Invacare Corporation or its subsidiaries
unless otherwise noted.
3-in-1 Oil is a registered trademark of
American Home Products Corporation.
WD-40 is a registered trademark of the
WD-40 Company.
Part No 1112648