Printing recorder with text printing and
24-character LED dot-matrix display
Brief description
The printing recorder is equipped with 3 or (optionally) 6 measurement inputs, which are
electrically isolated from one another. The evaluation of the measurement traces of the
printing recorder can be assisted by extensive text printing. The recorder can be programmed either by using the 8 keys on the front of the instrument or through a PC setup
program.
Thermocouples, resistance thermometers, resistance transmitters, potentiometers, voltages or currents (standard signals) are possible as input signals. The appropriate linearizations are carried out automatically, but can also be adapted to customer-specific
linearizations with the help of the PC setup program.
Further outstanding features of the LOGOPRINT 500, which are already included in the
basic version, are four open-collector outputs for signaling infringements of limits and
faults, eight event traces, as well as peak value recording.
The color assignments (measurement traces and texts) are freely programmable via the
PC setup program, which is available as an accessory.
Overview of functions
Analog inputs
(configurable and
isolated)
3 or 6 inputs for:
- thermocouples
- resistance thermometers
- resistance transmitters
- potentiometers
- voltage and current
8 logic inputs
available as extra code
Outputs
- 4 open-collector outputs
available as extra code:
- interface for 8 relay outputs
- supply for 2-wire transmitter
Recording
- measurement traces
- text printing
- event traces
Setup interface
for configuration and parameter setting via PC
RS422/RS485 interface
extra code for the data transfer from and to the recorder
Supply voltage
- 110 — 240V AC +10/-15%, 48 — 63Hz
- 20 — 53V AC/DC ±0%, 48 — 63Hz
Type 706030
Features
k Limit monitoring
k Event traces
k Four open-collector outputs
k Peak value recording
k Extensive text printing
k Statistics (report) with minimum,
maximum and mean values
k Event- and time-controlled chart speed
k Math and logic module
(PC setup program is required)
k Universal chart cassette
Approvals/marks of conformity (see Technical data)
2013-01-29/00355277
JUMO GmbH & Co. KG • 36035 Fulda, Germany
Data Sheet 706030
Page 2/12
Technical data
Thermocouple input
Designation
Fe-Con
Fe-Con
Cu-Con
Cu-Con
NiCr-Ni
NiCr-Con
NiCrSi-NiSi
Pt10Rh-Pt
Pt13Rh-Pt
Pt30Rh-Pt6Rh
L
J
U
T
K
E
N
S
R
B
Linearisation accuracy1
Range
2
DIN 43710
EN 60584
DIN 437102
EN 60584
EN 60584
EN 60584
EN 60584
EN 60584
EN 60584
EN 60584
Shortest span
-200 to
-210 to
-200 to
-270 to
-270 to
-270 to
-270 to
-50 to
-50 to
0—
±0.2%
±0.2% above -200°C
±0.3%
±0.5% above -200°C
±0.2% above -150°C
±0.2% above -200°C
±0.2% above -150°C
±0.5% above 0°C
±0.5% above 0°C
±0.5% above 500°C
+ 900°C
+1200°C
+ 600°C
+ 400°C
+1372°C
+1000°C
+1300°C
+1768°C
+1768°C
1820°C
Types L, J, U, T, K, E, N:
Types S, R, B:
Range start/end
100°C
500°C
freely programmable within the limits in 0.1°C steps
Cold junction
Pt 100 internal or thermostat as external constant
Cold junction accuracy (internal)
± 1°C
Cold junction temperature (external)
-50 to +100°C, adjustable through setup software
Measurement time
for 3 channels < 2sec; for 6 channels < 4sec
Input filter
2nd order digital filter; filter constant adjustable from 0 — 50.0sec
Features
1
2
also programmable in °F; customer-specific linearizations
The linearization accuracy refers to the maximum span. The linearization accuracy is reduced for shorter spans.
invalid DIN since 1995
Resistance thermometer input
Connection
Range
Linearisation accuracy
Meas.
current
Pt 100 EN 60751
2/3-wire
2/3-wire
4-wire
4-wire
-200 to +250°C
-200 to +850°C
-200 to +250°C
-200 to +850°C
±0.6°C
±1.0°C
±0.5°C
±0.8°C
500A
250A
500A
250A
Pt 100 JIS
2/3-wire
2/3-wire
4-wire
4-wire
-200 to +260°C
-200 to +649°C
-200 to +260°C
-200 to +649°C
±0.6°C
±1.0°C
±0.5°C
±0.8°C
500A
250A
500A
250A
Pt 500 DIN
2/3-wire
2/3-wire
4-wire
4-wire
-200 to +150°C
-200 to +850°C
-200 to +150°C
-200 to +850°C
±0.6°C
±1.0°C
±0.5°C
±0.8°C
250A
250A
250A
250A
Pt 1000 DIN
2/3-wire
2/3-wire
4-wire
4-wire
-200 to +250°C
-200 to +850°C
-200 to +250°C
-200 to +850°C
±0.6°C
±1.0°C
±0.5°C
±0.8°C
500A
250A
500A
250A
Ni 100
2/3-wire
2/3-wire
4-wire
4-wire
-60 to +125°C
-60 to +180°C
-60 to +125°C
-60 to +180°C
±0.6°C
±1.0°C
±0.5°C
±0.8°C
500A
250A
500A
250A
Connection type
Shortest span
Probe lead resistance
Range start/end
Measurement time
Input filter
Features
2013-01-29/00355277
2-, 3- or 4-wire circuit
15°C
max. 30  per core for 4-wire circuit
max. 20 per core for 2- and 3-wire circuit
for Pt 100 up to 260°C max. 10  per core in 2-wire and 3-wire circuit
freely programmable within the limits in 0.1°C steps
for 3 channels < 2sec; for 6 channels < 4sec
2nd order digital filter; filter constant adjustable from 0 — 50sec
also programmable in °F; customer-specific linearizations
JUMO GmbH & Co. KG • 36035 Fulda, Germany
Data Sheet 706030
Resistance transmitter and potentiometer input
Range
Accuracy
±300m
up to 200
±600m
up to 400
±1
up to 800
±2
up to 2000
±3
up to 4000
Connection type
Shortest span
Probe lead resistance
Resistance values
Measurement time
Input filter
Input for DC voltage or DC current
Basic range
-25 to +75mV
0 — 100mV
-100 to +100mV
0 — 200mV
-500 to +500mV
0 —
1V
-1 to
+1V
-5 to
+5V
0 — 10V
-10 to +10V
Shortest span
Range start/end
4 — 20mA
0 — 20mA
-20 to +20mA
Shortest span
Range start/end
Measurement time
Input filter
Features
Measuring current
500A
250A
250A
500A
250A
resistance transmitter: 3-wire circuit
potentiometer: 2-, 3- or 4-wire circuit
6
max. 30 per core in 4-wire circuit
max. 20 per core in 2-and 3-wire circuit
up to 200  range: max. 10  per core in 2-and 3-wire circuit
freely programmable within the limits in 0.1 steps
for 3 channels < 2sec; for 6 channels < 4sec
2nd order digital filter; filter constant adjustable from 0 — 50.0sec
Accuracy
±100V
±100V
±150V
±150V
±1mV
±1mV
±2mV
±10mV
±10mV
±15mV
±20A
±20A
±40A
Input resistance
RE > 10 M
RE > 10 M
RE > 10 M
RE > 10 M
RE > 10 M
RE > 10 M
RE > 10 M
RE > 0.5 M
RE > 0.5 M
RE > 0.5 M
5mV
freely programmable within the limits
(up to 999mV in 0.01mV steps, from 1V in 1mV steps)
burden voltage2.6V max.
burden voltage 2.6Vmax.
burden voltage 2.6Vmax.
0.5mA
freely programmable within the limits in 0.1mA steps
for 3 channels < 2sec; for 6 channels < 4sec
2nd order digital filter; filter constant adjustable from 0 — 50.0sec
adjustable linearizations for thermocouples and resistance thermometers
(for connection to transmitters without linearization)
Transducer short-circuit/break
Thermocouple
Resistance thermometer
Resistance transmitter
Potentiometer
Voltage up to ± 1V
Voltage above ± 1V
Current
1
2
Short-circuit1
not recognized
recognized
recognized
not recognized
not recognized
not recognized
not recognized
Break1
recognized
recognized 2
recognized
recognized2
recognized
not recognized
not recognized
The print head is positioned to 0%, “>>>>>>” appears in the LED dot-matrix display.
In 4-wire circuit: only recognized at terminals 1 and 2.
Outputs
Three open-collector outputs
One open-collector output
Printing system
Drive
Sensitivity
Reproducibility
Display and recording accuracy
2013-01-29/00355277
Page 3/12
to signal over/underlimit
to signal faults (e. g. end of chart)
stepper motor
0.2% or better referred to 100mm writing width
0.25% or better referred to 100mm writing width
Class 0.5 referred to range limits and basic ranges
JUMO GmbH & Co. KG • 36035 Fulda, Germany
Print head
Print colors
Over/underrun
Chart speed
Paper feed
Chart cassette
Chart
overall width / writing width
sprocket roller spacing
visible diagram length
overall length
Electrical data
Supply (SMPS)
Electrical safety
Data Sheet 706030
Page 4/12
print head with penlift function - sufficient for approx. 1 million dots
(depending on the ambient temperature)
Violet, red, black for 3-channel printing recorder and
violet, red, black, green, blue, brown for 6-channel printing recorder.
The color assignment can be changed at will, through the setup program.
electronically limited to 0 — 100mm writing width
programmable in the steps 0, 5, 10, 20, 60, 120, 240, 300, 360, 600, 720mm/h
by stepper motor and gearing
cassette for roll chart and fanfold chart (with tear-off edge and paper-end switch)
roll or fanfold chart to DIN 16320
120mm / 100mm
110mm
roll chart: 60mm; fanfold chart: 30 — 60mm
roll chart: 16m or 32m; fanfold chart: 15.6m
110 — 240V AC +10/-15% AC 48 — 63Hz, or 20 — 53V ±0% AC/DC 48 — 63Hz
to EN 61010, Part 1, March 1994
overvoltage Category II, contamination Grade 2
Test voltages (type test)
mains supply to measurement circuit
mains supply to housing
between measurement circuits
measurement circuits to housing
electrical isolation between the
analog inputs
Supply voltage sensitivity
Power consumption
Data buffering
Electrical connection
Electromagnetic compatibility (EMC)
interference emission
immunity to interference
with AC supply 2.3kV 50Hz, 1min; with AC/DC 510V 50Hz, 1min
with AC supply 1.5kV 50Hz, 1min; with AC/DC 510V 50Hz, 1min
200V 50Hz, 1min
500V 50Hz, 1min
up to 30V AC and 50V DC
less than 0.1% of range span
35VA max.
More than 4 years through lithium battery in RAM or 2 days with storage capacitor at
5 — 25°C ambient temperature. Additional backup in EEPROM.
At rear through plug-in screw terminals,
max. conductor cross-section 2.5mm2 or 2x 1.5mm2 with core end sleeves.
Setup connector at the front behind flip-up dot-matrix display.
EN 61326-1
Class B
to industrial requirements
Housing
Housing type
housing door
Transport mechanism
Chart cassette
Bezel size
Depth behind panel
Housing for front-panel mounting to IEC 61554, galvanized sheet steel
cast zink
in corrosion-resistant chrome-nickel steel
in plastic (polycarbonate)
144mm x 144mm
212mm without screw terminals; 227mm with screw terminals plugged in
138+1.0 mm x 138+1.0 mm
in control panel to DIN 43834
0 to 50°C
0.2%/10°C
-20 to +70°C (without print head), -20 to +55°C (with print head)
20 — 70% relative humidity, no condensation
normal position: vertical ± 30° (NL 90 ± 30, to DIN 16257)
to EN 60529 Category 2,
front IP54 (IP65 with extra code 266),
rear IP20
3.5 kg max.
Panel cut-out
Housing mounting
Ambient temperature range
Ambient temperature error
Storage temperature range
Climatic conditions
Operating position
Protection
Weight
Approvals/marks of conformity
Mark of
conformity
Testing
laboratory
Certificates /
certification numbers
Test basis
c UL us
Underwriters
Laboratories
E 201387
UL 3111-1
the flush-mounted instrument;
CAN/CSA C22.2 No. 1010.1-92 not in conjunction with any housing extra code
2013-01-29/00355277
valid for
JUMO GmbH & Co. KG • 36035 Fulda, Germany
Data Sheet 706030
Page 5/12
Operating modes
Graphic print-out
Extra codes
Limit monitoring
Eight limit comparators are available to
monitor the limits.
Limits, differential values (hysteresis), limit
comparator functions (lk), texts and the
channels to be monitored can be programmed. The result of the monitoring is
fed to the open-collector outputs (1 — 3)
and the optional relay module (1 — 8) as
limit state (logic 0 or 1).
The different lk functions are:
lk 7:
Measurement traces
RS422/RS485 interface
This interface is intended for communication with higher-level systems (e. g. bus
system or PC). It can be used to read out
measurements, to monitor operating states
and to transmit texts and values to the recorder.
The limit state is set to 1 when:
measurement > limit + upper differential.
lk8:
Zoom (plot area)
In zoom operation, an enlarged recording is
made of a section of the full range of a
trace.
Presentation range (offset)
This parameter is used to define the presentation range of a trace on the chart.
This assists the evaluation, in particular of
traces which are close to each other or
which overlap.
Peak value recording
The peak value recording can be switched
on or off for each channel.
In the switched-off state, the present value
of a channel is printed.
Since more values can be measured than
can be printed, the minimum and maximum values measured between two lines
to be printed are stored when the peak value recording is switched on. These minimum and maximum values are printed
when peak value recording is activated.
Event traces
Eight event traces can be printed. Limit
monitoring (limit comparators) or the state
of the optional logic inputs can thus be
documented on the chart.
Text printing
as lk7, but function is reversed.
Chart speeds
The LOGOPRINT 500 can be programmed
with four different operating modes for the
chart speed:
1. Normal operation
2. Limit operation
If the measurement goes above/below the
programmed limits, the recorder switches
to the speed which has been programmed
under “limit operation”.
3. External operation
A signal on one of the logic inputs at the
back of the recorder switches to the speed
which has been programmed under “external speed”.
4. Timed operation
The chart speed which is operative within a
programmable time span.
Text printing is used for comments on the
recorded trace and for event recording.
Printing priorities can be assigned to texts,
to serve as abort criteria during simultaneous text printing requests.
Text printing can be separately configured
for each text, either time-optimized or during continued recording of traces.
Text printing facilities:
1
2013-01-29/00355277
Time, date
Scaling of the channels
Channel numbers
Change of chart speed
Recording start/end text
“Power on” and “power off” text
Print text to check the printing system
and the service print
16 limit comparator texts1
(eight for underlimit and eight for overlimit)
2 reports (calculate and print minimum,
maximum and mean values)
Eight external texts1 (extra code)
16 binary-linked external texts1
(extra code)
Event counter1 (extra code)
These texts are buffered through a queue. As long
as the queue is not full, complete documentation
is assured.
Logic inputs
Eight logic inputs can be operated either
through floating contacts or by the following voltage levels:
inactive: 0 — 5V / active: 20 — 35V
The voltage levels must be applied for
0.4sec.
Functions available include:
-
External start/stop
Activate external chart speed
Text printing
Start/stop external report
Start scaling print
Increment two event counters
Key inhibit
Event traces
Supply for 2-wire transmitter
An electrically isolated supply for a 2-wire
transmitter is available.
24V DC 45mA ± 5%
ER8 external relay module
The LOGOPRINT 500 can be equipped
with an ER8 external relay module (eight
relay outputs) to monitor infringements of
upper and lower limits. The relay outputs
are permanently assigned to the limit comparators. The assignment to the measurement channels can be made freely through
the limit comparator parameter.
Contact rating:
3 A 250V AC 50Hz , or 3 A 30V DC
resistive load
JUMO GmbH & Co. KG • 36035 Fulda, Germany
Data Sheet 706030
Display and controls
Fault signal LED
24-character green LED
dot-matrix display
Print head
Keys for operating
and programming
Setup interface
(behind the swing-up
LED dot-matrix display)
Housing door
Chart paper
Channel label
Panel-mounting housing to IEC 61554 in galvanized sheet steel
Setup program
Page 6/12
Operation and
configuration
At the recorder
The eight keys on the instrument are used
for operating the instrument, and to configure all the parameters essential to the operation.
The 24-character dot-matrix display is
available for indicating and monitoring the
measurements and parameters.
Via the setup program for PC
(accessory)
The recorder can be configured using the
setup program for PC (see diagram, bottom left) more conveniently than by using
the instrument keys.
The configuration data of a configured instrument can be read out and altered using
the setup program.
For a further instrument with the same configuration, the data can be copied through
the setup program. The configuration data
can be archived on data media and printed
out.
In addition to the programming possibilities
from the keys of the recorder, the setup
program offers the following extra functions:
- Setting different print colors
- Customer-specific linearizations
- Setting the printing mode for the texts
(printing mode: “Overwrite measurement trace” or “Interrupt measurement
trace”)
- Printing priorities
- Math and logic module editor
- Various settings can be managed
Customer-specific linearizations
The setup program offers a choice between linear, square law and cube law linearization. There can be up to 41 calibration
points for linear and square law linerarization, and up to 61 calibration points for
cube law linearization. These calibration
points are used to determine the coefficients for polynomials which are defined
for each section, so that even a few calibration points produce a smooth graph.
Accuracy: depends on the shape of the
graph and the selected linearization.
Language
The language setting (English, German,
French) appears in the print-out and on the
LED dot-matrix display.
2013-01-29/00355277
JUMO GmbH & Co. KG • 36035 Fulda, Germany
Example of a recording
with text print-out
(1) Recording start/end text
(2) Speed change to 120 mm/h through a logic input
(3) Printing the time (with every fourth print-out, the current chart
speed, the programmed instrument name, or the date are
printed alternately).
(4) If a selectable logic input is closed, the count of an event
counter is incremented and documented together with the
programmed text. Altogether, two event counters are available.
(5) The scaling of all active channels can be printed, either by
pressing the F key (hold down for at least 4 sec), or through
a logic input.
Data Sheet 706030
Page 7/12
(1)
(6)
(2)
(6)
(3)
(4)
(6)
(3)
(5)
(6) The channel number can be printed in the selected channel
color so that individual traces can be more easily differentiated.
(6)
(7) Documentation of over/underlimit conditions in the limit
comparators.
(1)
(8) A total of eight event traces can be printed out. They can be
used either to document the state of the limit comparators,
or that of the logic inputs. The position of the event traces on
the chart can also be programmed.
(7)
(9) Additional texts (external texts) can be printed if one logic input, or a combination of up to four logic inputs, is switched.
(10)Print-out of a report. The print-out shows the period of time
in which the measurements were acquired and calculated, as
well as the minimum, maximum and mean values of all active
channels (including channel name and unit).
(9)
(11)As opposed to (5), the scaling of the channels can be performed alternately in a programmable spacing.
All texts which relate to a logic input can only be printed if the extra code “8 logic inputs” (code 259) has been implemented in the
recorder.
In the example, the traces which were printed above the report
(10), are printed out in the normal mode, i.e. all traces share the
entire chart width (0 — 100mm). The presentation range can be
selected freely on the chart for each trace. This assists the evaluation, in particular of traces which are close to one another or
which overlap. The traces below the report have thus been arranged over two sections of the chart (0 — 50mm and
50 — 100mm).
(8)
(9)
(7)
(10)
(3)
(11)
(3)
(3)
2013-01-29/00355277
JUMO GmbH & Co. KG • 36035 Fulda, Germany
Data Sheet 706030
Connection diagram for maximum terminal assignment
Rear view of 3/6-channel version with plug-in screw-clamp connectors
Connection for 3/6-channel version
Supply as on label
Diagram
N neutral
L1 line
PE protective earth
Analog inputs
Connector
10.
Inputs
1
2
3
4
5
6
Connector
Thermocouple
Resistance thermometer
in 2-wire circuit
1. 2. 3. 4. 5. 6.
Resistance thermometer
in 3-wire circuit
Resistance thermometer
in 4-wire circuit
2013-01-29/00355277
Page 8/12
JUMO GmbH & Co. KG • 36035 Fulda, Germany
Analog inputs
Data Sheet 706030
Inputs
1
2
3
4
Diagram
5
6
Connector
Resistance transmitter
in 3-wire circuit
A = start
S = slider
E = end
Potentiometer
in 2-wire circuit
Potentiometer
in 3-wire circuit
Potentiometer
in 4-wire circuit
1. 2. 3. 4. 5. 6.
Voltage input up to ± 1V
Voltage input above ± 1V
Current input ± 20mA
Current input (shunt)  ±20mA
(when using transducers with
changeable internal resistance; extra
code “terminal with shunt” is required)
2013-01-29/00355277
Page 9/12
JUMO GmbH & Co. KG • 36035 Fulda, Germany
Inputs/outputs
Connector
ER8 external relay module
Communication
with external
relay module
Voltage source for external
2-wire transmitter
24V 45mA ± 5%
Logic operating inputs
Contact operation
LOW = ROFF  100k
HIGH = RON  50
min. pulse length:
HIGH 400msec
LOW 400msec
Voltage operation
LOW = 0 — 5V DC
(inactive)
HIGH = 20 — 35V DC
(active)
Open-collector outputs
RS422/RS485 serial interface
Overview of the electrical isolation
2013-01-29/00355277
Data Sheet 706030
DOUT1 — DOUT4
Umax = 32V DC
Imax = 100mA
Residual voltage DOUT
active
UDOUTactive =
0.4 — 1.2V
Communication
with higher-level
systems
Diagram
7.
contact no. 7 = logic input 1
.
.
.
contact no. 14 = logic input 8
8.
9.
Page 10/12
JUMO GmbH & Co. KG • 36035 Fulda, Germany
Data Sheet 706030
Dimensions
Panel-mounting housing
373
226
314
Extra code 350
Portable recorder housing
for varying applications in
mobile use
326
Extra code 351
Housing with carrying handle,
rubber feet and terminal
cover, also 3m mains cable
with earthed plug
2013-01-29/00355277
366
Page 11/12
JUMO GmbH & Co. KG • 36035 Fulda, Germany
Data Sheet 706030
Page 12/12
Order matrix:
(1) Basic version
706030/14
LOGOPRINT 500 with 3 universal inputs
706030/15
LOGOPRINT 500 with 6 universal inputs
x x
x x
888
999
(2) Inputs 1 — 3 (programmable)
factory-set
configuration to customer specification1
x
x
x
000
888
999
(3) Inputs 4 — 6 (programmable)
not assigned
factory-set
configuration to customer specification1
x x
x x
x x
00
52
53
(4) Interface
not assigned
RS422, Jbus, Modbus
RS485, Jbus, Modbus
x x
x x
22
23
(5) Supply
20 — 53V AC/DC +0/-0% 48 — 63Hz
110 — 240V AC +10/-15% 48 — 63Hz
x
x
x
x
x
x
x
x
x
x
x
x
x
x
020
021
030
259
266
350
351
(6) Extra codes
lithium battery for RAM buffer (ex-factory)
storage capacitor for RAM buffer (on request)
terminal with shunt (6 items)
8 logic inputs, interface for external relay module (ER8), voltage output 24V 50mA DC
IP65 seal, wide fixing brackets
universal carrying case TG-352
housing with carrying handle3
(1)
Order code
Order example
706030/14
(2)
-
888
(3)
-
000
(4)
-
00
(5)
-
23
(6)
/
/
, ...4
020
1
2
Please specify probe types and measurement ranges in plain text.
This extra code is available in combination with supply voltage 110—240V AC, not with low supply voltage. UL approval is not available.
The protection type in the carrying case corresponds to IP20, outside IP20D.
3
UL approval is not available.
4
Extra codes are listed in sequence and separated by a comma.
Standard accessories
-
1 Operating Instructions
2 mounting brackets
cable-tie with foot (can be released), for tension relief of the sensor leads connected
1 print head, 3 colors (each color is available twice) or 1 print head, 6 colors
1 roll chart 32m long and 1 fanfold chart pack 15.6m long
Accessories (Data Sheet 709700)
-
For further accessories, see data sheet 709700.
Consumables
Part no.
-
Roll chart (5 rolls), no name, % graduation, linear; overall length 16m, overall width 120mm
Roll chart (5 rolls), no name, % graduation, linear; overall length 32m, overall width 120mm
Roll chart (5 rolls), no name, special graduation, linear; marked as specified
00331497
00331499
-
-
Fanfold chart (5 packs), no name, % graduation, linear; overall length 15.6m, overall width 120mm
Fanfold chart (5 packs), no name, special graduation, linear; marked as specified
00331490
-
-
Print head (2 items), 3 colors
Print head (2 items), 6 colors
00355244
00355255
2013-01-29/00355277