Armatures for analytical probes

8200
Armatures for analytical probes
•For many different types of installations and
applications
•Large range of probe holders
•General purpose, water treatment, food & beverage,
pharmaceutical applications
Type 8200 can be combined with...
Type S020
Pipe and fitting
INSERTION fitting
T
he holder range for 120 mm analytical
probes (pH/ORP/conductivity...) covers many
types of applications:
- General purpose
- Water treatment
- Food & Beverage
- Pharmaceutical/Biotechnology.
A wide range of process connections are
available for:
General purpose holders mounted on
pipes or tanks
Hygienic holders designed for hygienic applications:
- 3-rod holders which ensure good sensor
protection and allow easy cleaning. Due to
a sanitary design steam sterilisation, autoclavation, CIP are possible.
- The direct welding holders save space
and are designed for installation in fermenters and many other applications with
tanks and pipes. Steam sterilisation, autoclavation, CIP are possible.
- Special holders for the 2" Clamp
connection or for DN50 thread process
connections satisfy highest demands, for
example, in CIP-applications A Pt1000
temperature probe is available as an option.
- The 15° version with 2" (DN50/40) connection adapted for GEA Tuchenhagen
VARINLINE process connections enables
positioning of the probe in relation to the
flow or in vertical pipes.
A
special holder, the analytical measurement
chamber Type 8200, is designed to be used
with the chlorine sensor Type 8232.
Type 8203
Type 8221
Type 8232
Type BBS- 11
pH/ORP probe
Conductivity probe
Chlorine sensor
Safety sockets
General data
Process connection
General purpose
G 2" for use with INSERTION fitting Type S020
G 1" for use with T-Fitting
Solvent union for use with T-Fitting d32 x d32 up to
d32 x d110
Immersion fitting with fixing kit for use on tanks
Hygienic holder
G 1¼ (28 or 46 mm O-ring position)
Clamp 1½" (Ø 50.5 mm) or 2" (Ø 64 mm - acc. to ISO2852)
for DN50 thread (acc. to SMS1145) process connection
2" (DN50/40) connection adapted for GEA Tuchenhagen
VARINLINE process connections
Direct welding on pipe
Measurement chamber
Medium temperature
Medium pressure
Screw connections (straight for inlet, elbow for outlet) ¼" external thread to 6/8 hose; mounting nut (to fasten the sensor);
hose sleeve (to sample)
Temperature limits may depend on the inserted probe.
Refer to the relevant instruction manual and technical
data on next page. If the temperature ranges given for the
holder and the inserted probe are different, use the most
restrictive range.
Pressure limits may depend on the inserted probe. Refer
to the relevant instruction manual and technical data on
next page. If the pressure ranges given for the holder and
the inserted probe are different, use the most restrictive
range.
Environment
Ambient temperature
Temperature limits may depend on the inserted probe.
Refer to the relevant instruction manual or datasheet for
more details
www.burkert.com
p. 1/14
8200
General purpose holder
G 2" connection
Materials
Body
Seal
Medium temperature
Stainless steel 316L 1.4404, PVC
FKM (EPDM option)
with S020 fitting in
PVC: 0 … + 50 °C (+ 32 … + 122 °F),
Stainless steel: - 20 … + 130 °C
Hygienic holder
2" clamp (Ø 64 mm - acc. to ISO2852) connection or
for DN50 thread (acc. to SMS1145) process connection
Materials
Body
Seal
Medium temperature
Stainless steel (316L/1.4404)
EPDM
- 20 … + 140 °C (+ 5 … + 284 °F)
Medium pressure
PN16 (232 PSI)
(+ 5 … + 266 °F)
Medium pressure
with S020 fitting in
PVC: PN10 (145 PSI), Stainless steel: PN16 (232 PSI)
General purpose holder
G 1" or solvent union connection
Materials
Body
Seal
Medium temperature
PVC
FKM
0 … + 50 °C (+ 32 … + 122 °F)
Medium pressure
PN10 (145 PSI)
General purpose holder
Immersion fitting with fixing kit
Materials
Sensor holder
Extension tube
Seal
Screws
Medium temperature
PVDF
PP
FKM (EPDM option)
Stainless steel
0 … + 80 °C (+ 32 … + 176 °F)
Hygienic holder
G 1¼ connection
(28 or 46 mm O-ring position)
Materials
Body
Seal
Medium temperature
Stainless steel (316L/1.4435)
EPDM (FDA)
- 10 … + 135 °C (+ 14 … + 275 °F)
Medium pressure
Max. 6 bar (max. 87 PSI)
Hygienic holder - short immersion depth
1½" clamp (Ø 50.5 mm) connection
Materials
Body
Seal
Medium temperature
Stainless steel (316L/1.4435)
EPDM (FDA)
- 10 … + 135 °C (+ 14 … + 275 °F)
Medium pressure
Max. 6 bar (max. 87 PSI)
Hygienic holder
2" (DN50/40) connection adapted for GEA Tuchenhagen VARINLINE process connections
Materials
Body
Seal
Medium temperature
Stainless steel (316L/1.4435)
EPDM (FDA)
- 10 … + 135 °C (+ 14 … + 275 °F)
Medium pressure
Max. 6 bar (max. 87 PSI)
Hygienic holder
Direct welding connection
Materials
Body
Seal
Medium temperature
Stainless steel (316L/1.4435)
EPDM (FDA)
- 10 … + 140 °C (+ 14 … + 284 °F)
Medium pressure
Max. 16 bar (max. 232 PSI)
Analytical measurement chamber
Materials
Body
Mounting nut
O-ring holder
Slide ring (30 x 25.5 x 4)
O-ring (30 x 2.6)
O-ring (25 x 2.5)
Hose connection
Inlet ball valve
Float
O-ring flat (12 x 6 x 2)
Sample valve
Ball valve (with hose sleeve)
Elbow screw connection
Medium temperature
Medium pressure
PMMA polished, beveled edges
PVC-U grey
PVDF nature
PETP black
FPM
Silicone
PA
PVC-U
PEEK, steel 1.0037
Silicone transparent
PVC-U
Stainless steel
Max. 45 °C (max. 113 °F)
Max. 4 bar (max. 58 PSI) - Permitted operating pressure of the sensor has to be
respected
Hygienic holder - long immersion depth
1½" clamp (Ø 50.5 mm) connection
Materials
Body
Seal
Medium temperature
Stainless steel (316L/1.4404)
FKM
- 10 … + 135 °C (+ 14 … + 275 °F)
Medium pressure
Max. 6 bar (max. 87 PSI)
p. 2/14
8200
Installation example of Type 8200
Type 8200 probe holder for installation on S020 Bürkert fitting
A complete pH/ORP/conductivity sensor consists of
5
4
1. Complete Type 8200 probe holder with nut and seals
2. pH/O.R.P probe or conductivity probe with PG 13.5 connection
3. Pt1000 temperature probe/liquid earth rod (option, if needed)
4. Shielded cable for pH/ORP or conductivity cable varioPin (6.0)
5.Shielded cable for Pt1000 temperature probe/liquid earth rod (option,
1
if needed)
6. S020 Bürkert fitting (G 2" connection)
3
2
6
Type 8200 G 1" or solvent probe holder for installation on T-Fitting
Ø 32
d
d
G1"
Type 8200 G 1¼ probe holder with o-ring position of 28 or 46 mm for installation on sockets
The G 1¼ probes holder with O-ring position of 28 mm or 46 mm has to
be mounted into sockets which are welded on pipes or tanks.
Coupling nut
loosened
Sockets 15°
Robust weld-in socket with 15° angle to be mounted on tanks.
The sockets have a safety construction. The socket seals only if the o-ring
of the holder is exactly in the right position.
Otherwise the medium leaks through the G 1¼ coupling nut.
Process medium
leaks before the holder is dismounted
Type 8200 probe holders with 2" clamp (acc. to ISO2852) connection or for DN50 thread (acc. to SMS1145) process connection
p. 3/14
8200
Installation example of Type 8200
Direct welding Type 8200 probes holder for installation on pipe
The steel mantle can be welded into a hole in the tank wall at virtually any
depth.
As a result the probe is always immersed to exactly the desired position
in the tank.
The o-ring can be easily replaced thanks to the « seal pusher » part.
By screwing in an adapter instead of a pH sensor, conductivity sensors
can be mounted practically flush with the holder and is then designed in
accordance with EHEDG guidelines.
Type 8200 analytical measurement chamber
A complete chlorine sensor consists of
1. T
ype 8200 analytical measurement chamber with nut and seals
ype 8232 chlorine sensor
2. T
Do not install the sensor in the main pipe.
Measure only in bypass with use of Type 8200 analytical
measurement chamber.
Respect the sensor pressure and temperature ranges.
C
lose the water inlet ball valve of Type 8200 analytical measurement chamber.
A
void installations that allow air bubbles to enter the measuring
water.
S
top the circulation of fluid, cut off the pressure and drain the
pipe before loosening the process connections.
Installing Type 8200 analytical measurement chamber on the support.
Snap Ring
Slide Ring
Seal
Union nut
Sampling
valve for DPDmeasurement
Sample water
outlet
Holes for fixing
screws
W
hen choosing of the installation place for Type 8200 analytical
measurement chamber, please consider the max. height of the
chlorine sensor (approx. 220 mm without cable connected), so it can be
set up in this analytical measurement chamber.
1
Drilling for flow
switch (option)
Float
Inlet ball valve
(Flow indicator)
Sample water
inlet
2
A.Drill holes in the support according to the dimensions indicated on the dimension drawing on page 9.
B.Mount Type 8200 analytical measurement chamber with two
screws (recommendation: M4 x 60 mm pan head screw or hexagon socket
head screw. The screws are not provided) onto the support.
C.Connect the water inlet of the analytical measurement chamber with a 6/8 hose to the sample water source.
D.Connect the water outlet of the analytical measurement chamber with a 6/8 hose to the drain for example.
E.Install the chlorine sensor (see operating instruction manual)
F. Establish the circulation of the fluid.
G.Open the water inlet ball valve.
p. 4/14
8200
Dimensions [mm]
General purpose holder
G 2" connection* for installation on Bürkert fitting S020 PVC, stainless steel
G 1" connection* for installation on T-Fitting PVC
44
Ø 42
G1”
133
154
64
(71)
133
154
30
53
Ø 56
Ø 76
Ø 26
Ø 26
Stick connection* for installation on T-Fitting PVC
64
44
155
Immersion fitting with fixing kit for installation on Tanks
L
Ø 145
Ø 32
133
154
Ø 75
Ø 104
Ø 42
L
500
1000
1500
2000
Ø 26
* with analytical probe and Pt1000 temperature probe/liquid earth rod - have to be ordered separately
p. 5/14
8200
Dimensions [mm]
Hygienic holder
G 1¼ connection stainless steel 316L/ DIN1.4435
H
133
PG13.5
24
Ø24.97
H = 28 or 46
2
(according to ordered version)
Ø30
PG13.5
ø 42.4
39.5
1½" clamp connection (Ø 50.5 mm)* stainless steel 316L
18
1½" clamp connection (Ø 50.5 mm)* stainless steel 316L / DIN1.4435
117
117
98.5
ø 50.5
ø 16.7
24
33
ø5
Long immersion depth
Short immersion depth
Ø24
Ø50.5
(Clamp1.5")
ø 30.1
* with analytical probe and Pt1000 temperature probe/liquid earth rod - have to be ordered separately
p. 6/14
8200
Dimensions [mm]
Hygienic holder
2" clamp connection (Ø 64 mm - acc. to ISO2852) stainless steel 1.4404
For thread process connection - DN50
stainless steel 1.4404
ø60
ø5.2
PG13.5
ø5.2
PG13.5
(DN50/40)
80
16
10
14
11
16
80
7
12.8
7
D
19
ø64
connection*
11
M8
1.7
ø51.6
19
2"
-
ø11.5
M8
10
14
1.7
ø11.5
(acc. to SMS1145)
adapted for GEA Tuchenhagen VARINLINE process connections
12.8
ø65
- stainless steel 316L/ DIN1.4435
3.5
(27)
Ø30
102
PG1
15°
DN 50/40 (Ø84)
* with analytical probe - have to be ordered separately
p. 7/14
8200
Dimensions [mm]
Hygienic holder
Direct welding connection - stainless steel 316L/ DIN1.4435
without adapter
with adapter
PG13.5
128.5
63
63
107
107
Standard EHEDG approved design:
with adapter and conductivity probe
Type 8221
(10)
(10)
PG13.5
Ø28
Ø28
p. 8/14
8200
Dimensions [mm]
Analytical measurement chamber
130
51
60
30
4.5
M18x1
9
50
260
100
175
8
8
55
Flow switch for analytical measurement chamber
(option)
56
50
M18x1
ø 4.6
LED
24
14
25 Nm
36
35 Nm
p. 9/14
8200
Ordering information for complete pH/ORP sensor using Type 8200
Combining Type 8200 probe holder with Type S020 fitting
Available fitting DN
DN15
DN200
DN50 DN65
DN100
DN200
T-Fitting
Weld-in socket
Fusion spigot
Analytical measurement using probe
holder G 2" connection for S020
Note A
Note A: Only use plastic fitting in analytical version with true union acc. to DIN8063 (PVC), to DIN16962 (PP) or to ISO10931 (PVDF), see datasheet, Type S020.
pH/O.R.P sensor for tank or pipe installation
A complete pH/ORP sensor consists of Type 8200 probe holder with seals, Type 8203 pH/O.R.P probe, a Pt1000 temperature probe/liquid earth rod
(option) and a fitting according to the selected holder.
The following information is necessary for the selection of a complete device:
•Article no. of the desired Type 8200 probe holder (see ordering chart, p. 13)
•Article no. of the selected Type 8203 pH or O.R.P probe (see separate datasheet)
•Article no. of the Pt1000 temperature probe/liquid earth rod if needed (see separate datasheet Type 8203)
•Article no. of the selected Type S020 fitting (DN15 - DN200) only if probe holder has a G 2" connection (see separate datasheet)
or of the selected socket only if probes holder has a G 1¼ connection with o-ring position of 28 or 46 mm (see ordering chart, p. 14)
You have to order the components separately.
When you click on the orange box “More info.”, you will come to our website for the resp. product where you can download the datasheet.
Type 8200 probe holder
Type 8203 probe
pH or ORP probes
+
Fitting S020
Pt1000/liquid earth rod
+
Socket
p. 10/14
8200
Ordering information for complete pH/ORP sensor using Type 8200
(continued)
pH/O.R.P sensor for tank installation with the immersion fitting.
A complete pH/ORP sensor for tank installation consists of an immersion fitting, an extension kit for immersion fitting, a fixing kit (flange DN65 with stainless steel
screws), a probe holder with Pt1000 temperature probe/liquid earth rod, Type 8203 pH/ORP probe and a seal.
The following information is necessary for the selection of a complete device:
•Article no. of the immersion fitting (see ordering chart of probe holders, p. 13)
•Article no. of the extension kit for the immersion fitting (see ordering chart of accessories, p. 14)
•Article no. of the fixing kit (flange DN65 with stainless steel screws - see ordering chart of accessories, p. 14)
•Article no. of the probe holder for immersion fitting with Pt1000 temperature probe/liquid earth rod (see ordering chart of probe holder, p. 13)
•Article no. of Type 8203 120 mm pH/ORP probe (see separate datasheet)
•Article no. of the seal set if EPDM desired (see ordering chart of accessories, p. 14)
You have to order the components separately.
When you click on the orange box “More info.”, you will come to our website for the resp. product where you can download the datasheet.
Type 8200 probe holder
Immersion fitting Extension kit
Fixing kit
Sensor holder with
Type 8203 pH/ORP probes
Seal set
Pt1000/liquid earth rod
for immersion fitting
Probe of 120 mm
p. 11/14
8200
Ordering information for complete conductivity sensor using Type 8200
Conductivity sensor for tank or pipe installation
A complete conductivity sensor consists of a probes holder Type 8200 and a conductivity probe.
The following information is necessary for the selection of a complete device:
•Article no. of Type 8200 probes holder (see ordering chart of probes holder, p. 13)
•Article no. of the selected conductivity probe (see separate datasheet Type 8221)
•Article no. of the selected Type S020 fitting (DN15 - DN200) only if probe holder has a G 2" connection (see separate datasheet)
You have to order the components separately.
When you click on the orange box “More info.”, you will come to our website for the resp. product where you can download the datasheet.
Type 8200 probe holder
Type 8221 probe
Conductivity probe
+
Fitting S020
+
Socket
Ordering information for complete chlorine sensor using Type 8200
A complete chlorine sensor consists of Type 8200 analytical measurement chamber and a chlorine sensor.
The following information is necessary for the selection of a complete device:
•Article no. of Type 8200 analytical measurement chamber (see ordering chart of analytical measurement chamber, p. 13)
•Article no. of the selected chlorine sensor (see separate datasheet, Type 8232)
•Article no. of the flow switch for continuous monitoring of the flow (optional) (see ordering chart of analytical measurement chamber, p. 13)
You have to order the components separately.
When you click on the orange box “More info.”, you will come to our website for the resp. product where you can download the datasheet.
Type 8200 analytical measurement chamber
Type 8232 chlorine sensor
p. 12/14
8200
Stainless steel 316L/1.4404
Article no.
PVC
Protection
tube
standard
Boring for
Pt1000
temperature
probe/liquid
earth rod
Material
G 2" connection for installation on
Bürkert fitting S020
Version
Specifications
Ordering chart - Type 8200 probe holders
No
Yes
429224 
Yes
Yes
429228 
No
Yes
429227 
Yes
Yes
429231 
No
No
429220 
G 1" connection
short
PVC
Yes
No
429221 
Stick connection
short
PVC
No
No
564236 
Yes
No
563475 
Pt1000 temperature probe/liquid
earth rod in stainless steel
Pt1000 temperature probe/
liquid earth rod in titanium
L = 0.5 m
PVDF
Yes
Yes
418889 
PVDF
Yes
Yes
418890 
PP
No
Yes
419567 
L = 1.0 m
PP
No
Yes
419568 
L = 1.5 m
PP
No
Yes
419569 
L = 2.0 m
PP
No
Yes
419570 
High = 28
Stainless steel 316L/1.4435
No
Yes
562431 
High = 46
Stainless steel 316L/1.4435
No
Yes
562432 
Short immersion depth
Stainless steel 316L/1.4435
No
Yes
558885 
Long immersion depth
Stainless steel 316L/1.4404
Yes
Yes
429235 
standard
Stainless steel 316L/1.4404
No
No
567197 
standard
Stainless steel 316L/1.4404
Yes
No
567198 
standard
Stainless steel 316L/1.4404
Yes
No
566501 
standard
Stainless steel 316L/1.4404
No
No
566502 
15°
Stainless steel 316L/1.4435
No
Yes
562433 
Standard
Stainless steel 316L/1.4435
No
No
561728 
For immersion fitting
Immersion fitting
Hygienic G 1¼ connection
1½" clamp connection - (Ø 50.5 mm)
2" clamp connection (Ø 64 mm - ISO2852)
For DN50 thread (acc. to SMS1145)
process connection
2" (DN50/40) connection adapted for GEA
Tuchenhagen VARINLINE process connections
Hygienic direct welding connection
Article no.
Description
Ordering chart - Type 8200 analytical measurement chamber
Analytical measurement chamber
566054 
Flow switch for analytical measurement chamber, PNP, 2 m cable (optional)
772858 
Article no.
Description
Ordering chart for accessories
Pt1000 temperature probe/liquid earth rod - in stainless steel 1.4571
427023 
Pt1000 temperature probe/liquid earth rod - in titanium
560317 
p. 13/14
8200
Article no.
Description
Ordering chart for accessories
Set with FKM seal for the general purpose holder G 2" connection
429264 
Set with 1 green FKM + 1 black EPDM seal for the general purpose holder G 2" connection
552111 
Extension kit for the immersion fitting L = X m
562573 
Fixing kit - flange DN65 with stainless steel screws for the immersion fitting L = X m
413615 
Weld-in socket 15° L=28 for holder G 1¼
737241 
Weld-in socket 15° L=46 for holder G 1¼
737260 
Adapter for hygienic direct welding connection and conductivity probe
563477 
Interconnection possibilities with other Bürkert devices
Type 8619 multiCELL
transmitter/controller
Wall-mounted
Panel-mounted
Type 8200 Probe holder
Type 8232 chlorine sensor
with
Type 8203 pH/ORP probe
or
Type 8221 conductivity probe
Type 8203 pH/ORP probe
or
Type S020 INSERTION Fitting
Type 8221 conductivity
probe
T-Fitting
(see corresp. datasheet)
Weld-in
socket
Measuring
chamber
Pipe or tank
Type 8200 Analytical
measurement
chamber (flow switch
optional available)
To find your nearest Bürkert facility, click on the orange box
In case of special application conditions,
please consult for advice.
www.burkert.com
Subject to alteration.
© Christian Bürkert GmbH & Co. KG
1803/12_EU-en_00895158
p. 14/14
Download PDF