TI232 CUM740 - Axon Automation Inc.

Technical
Information
TI 232C/24/ae
Turbidity and solids content transmitter
CUM 740
Transmitter for optical solids measurement for
turbidity and solids content
The CUM 740 is used for optical solid
matter content measurement in clear
and turbid water and in sludge.
The transmitter covers a wide range of
meauring sensors from low to high solids
concentrations.
Applications
• Closed sewage treatment plant areas
such as inflow, preclarifier, and sludge
removal
• Process monitoring in hightemperature and hazardous areas
in the chemical industry, waste
incenerators and steam generation
plants
Features and benefits
• Large selection of sensors for fourbeam pulsed light systems
• Large two line LCD for set up and
measured value display
• Large LED display for separate
measured value display
• Large concentration range from
2 FNU to 150 ppm (150 g/l)
• Measuring units in ppm, g/l, mg/l,
TEF, %
• Menu controlled set up and calibration
in plain text
• Measured value processing is done
in the sensor, giving low transmission
sensitivity
• Five relay outputs (sensor cleaning,
error message, two configurable
limit relays, Hold)
• Battery-powered measured value
storage
• One or two channel versions
The Power of Know How
Measuring system
The complete measuring system
consists of:
• Turbidity transmitter CUM 740
• Turbidity sensor, e.g. TurbiMax W
CUS 65 with measuring cable
• Sensor installation holder or
immersion assembly
CUM 740 Transmitter
ENDRESS+HAUSER
CUM 740
Measuring cable
CYY 105 Immersion
assembly
CUS 65 Turbidity sensor
Complete measuring system
Measuring principle
Signal processing
Measured value preprocessing takes
place in the sensor electronics. The
connectable sensors operate using the
four-beam pulsed light method. Depending on the sensor version connected,
turbidity is determined by different optical
measuring methods:
• Absorption light method
• Backscatter light method
• 90° scattered light method
Turbidity sensors
Endress+Hauser offers the following
turbidity sensors for connection to the
CUM 740 transmitter:
• CUS 31, 90° scattered light method,
measuring range of 0.00 to 9999 FNU,
0.0 to 300 ppm (g/l), 0.0 to 200.%
concentration
• CUS 41, 90° scattered light method,
measuring range of 0.00 to 9999 FNU,
0.0 to 300 ppm (g/l), 0.0 to 200.0%
concentration
2
The sensor generates a turbidity or
solids concentration signal which is
converted into a frequency signal. the
frequency signals are assigned to
corresponding turbidity units and solids
concentration content, and are indicated
on the transmitter display.
• CUS 65 - A, absorption light method,
measuring range 0 to 12 ppm (g/l)
• CUS 65 - B, absorption light method,
measuring range 0 to 40 ppm (g/l)
• CUS 65 - C, absorption light method,
measuring range 0 to 50 ppm (g/l)
• CUS 65 - D, 90° scattered light
method, measuring range 1 to
1000 FNU
• CUS 65 - E, backscattered light
method, measuring range 10 to
150 ppm (g/l)
The CUM 740 is fully set up and
calibrated in a menu-assisted software
using a dirt-proof membrane keypad.
The operator is guided interactively
through the operating muen. The
interface is a two-line plain text display.
Operation
1
Programming levels which go beyond
everyday operation processes are only
accessible by entering a password.
All the calibration data and parameters
are retained if there is a power failure or
when the device is shut down (nonvolatile RAM).
3
2
ENDRESS+HAUSER
CUM 740
Operating panel
1 LED display
(measrued value display)
2 LCD (plain text display)
3 Main power switch
4 Membrane keypad
Functions
4
The 16-bit processor offers the following
possibilities for signal evaluation:
• Measured value detection, display
and evaluation
• Menu control with alph-numeric LCD
• Measuring system monitoring
including sensor
• User parameter storage and
management
Instrument control functions are
arranged in a logical menu structure.
Operating panel
MEASUREMENT
PARAMETER ENTRY
CALIBRATION
ASSIGN
FREQUENCY
CONFIGURATION
LANGUAGE
ERROR DISPLAY
3
Function
Detection, evaluation and display of
sensor signal, analog current and
sensor frequency
Measuring range selection, limit
setting, measured value damping
setting, cleaning interval setting
Sensor calibrated using stored
calibration curves or using application
specific customer standards
Calibration value assignment to
appropriate sensor signals
Retrieval and option for manually
editing measuring frequencies
determined during calibration
Sensor type selection, measuring unit
selection, calibration factor setting,
analog output configuration, alarm relay
configuration
User interface in your own language
Error message display
Dimensions
9.49”
(241)
0.75”
(19)
8.35”
(212)
0.39”
(10)
4.50”
(114,5)
ENDRESS+HAUSER
CUM 740
7.28”
(185)
5.98”
(152)
2.36”
(60)
2.72”
(69)
Dimensions are in
inches (mm)
Installation
The CUM 740 can be installed indoors or
outdoors. For outdoor installations, a
weather protection cover is recommended (see accessories). The ambient
temperature limits must be observed for
proper instrument operation.
1
ENDRESS+HAUSER
CUM 740
4
3
2
1
2
3
4
CUM 740 turbidity transmitter
CUA 451 retractable holder
CUS 65-A turbidity sensor
Process pipe
1
2
3
1
2
3
4
4
CUM 740 turbidity transmitter
CYY 105 immersion holder assembly
CUS 65-C turbidity sensor
Process basin or channel
4
ENDRESS+HAUSER
CUM 740
Electrical connection
Sensor 1
Sensor 2
Connection
X2
Terminals voltage
to ground even
when the unit is
switched off !!!
X3
_
+
33
30
32
29
Shield 31
28
24 VDC / 100 mA
OUT
1
2
4 to 20 mA
27
0V
26
+24 V
25
+24 V
24
0V
IN
23
TxD
22
RxD
CTS
21
20
GND
19
Shield
RS 232
F1
NO
NC
C
18
15
12
9
6
17
14
11
8
5
16
13
10
7
4
Error
Hold
Cleaning
Alarm A
3
2
N
1
L
T 500 mA
Supply 230 / 115 VAC
Alarm B
Technical data
General data
Materials
Input
Manufacturer
Endress+Hauser
Product designation
CUM 740 turbidity and solids content transmitter
Dimensions
Refer to top of page 4
Weight
3.5 lb (1.6 kg)
Display
LED display, 0.5” (12 mm) height for measured
value display, 2-line LCD 0.2” (5 mm) height for
set up
Housing
Polycarbonate
Cover
Plexiglas®
Parameters
Turbidity and solids content measurement
Measuring principle
Four-beam pulsed light method
Measuring light
Infrared light
Wavelength
880 nm (absorption maximum)
Measuring range
Dependent on sensor, see page 2
Accuracy
≤ 1% from measuring range end value
Reproducibility
0.5%
Signal output
0/4 to 20 mA
Number of signal outputs
Maximum two
Load
Maximum 500 Ω
Switching outputs
1 relay contact for sensor cleaning, 1 relay
contact for Hold function, 1 relay contact for
error messages, 2 limit relays freely
configurable
Switching power
3A at 115/230 VAC, 1A at 24 VAC/VDC
Interfaces
RS 232, port for bus extension
Electrical connection
Power supply
115 / 230 VAC, 50/60 Hz, +6 to -10%,
24 VAC / VDC
Ambient conditions
Ambient temperature
-4° to +140°F (-20 to +60°C)
Ingress protection
NEMA 4 (IP 65)
Output
5
Accessories
Weather protection cover for wall
mounting, 316 SS
Order number: CYY101-A
Upright mounting post, 316 SS, for cover
and CUM 740 transmitter mounting
Order number: CYY102-A
2.36” x 2.36” x 0.78”
(60 x 60 x 2)
11.8”
(300)
12.6”
(320)
10.6”
(270)
58.9”
(1495)
37.4”
(950)
21.6”
(550)
Æ 0.47”
(12)
5.91”
(150)
4.72”
(120)
Supplemental documentation
TurbiMax W CUS 31
TI 176C/24/ae
TurbiMax W CUS 41
TI 177C/24/ae
TurbiMax W CUS 65
TI 379C/24/ae
6
Ordering information
CUM 740 transmitter for turbidity and
solids content measurement
1
2
3
4
CUM 740 1
2
3
4
Power supply
0
230 VAC
1
115 VAC
8
24 VAC / VDC
Communication
A
RS 232 and 0/4 to 20 mA
Type
1
Single channel
2
Two channel
Additional equipment
A
Standard version
7
For application and selection assistance,
in the U.S. call 888-ENDRESS
For total support of your installed base, 24 hours
a day, in the U.S. call 800-642-8737
Visit us on our web site, www.us.endress.com
United States
Canada
Mexico
Endress+Hauser, Inc.
2350 Endress Place
Greenwood, IN 46143
Phone: (317) 535-7138
888-ENDRESS
FAX: (317) 535-8498
Endress+Hauser
Canada Ltd.
1440 Graham’s Lane
Unit 1, Burlington
ON, L7S 1W3
Phone: (905) 681-9292
800-668-3199
FAX: (905) 681-9444
Endress+Hauser
Paseo del Pedregal No. 610
Col. Jardines del Pedregal
01900, Mexico D.F.
Mexico
Phone: (525) 568-2405
FAX: (525) 568-7459
Endress+Hauser
The Power of Know How
TI 232C/24/ae/02.03