Wiring Diagrams - Booth Medical Equipment

Refer To:
Page
Operation & Troubleshooting ................ A-1
Component Testing / Repair ................. B-1
Nex
Go To Table O
Bac
Wiring Diagrams
TP2 Grd.
15 13 11 9 7 5 3 1
16 14 12 10 8 6 4 2
2
4
6
8
10
12
14
(Motor Lever Arm
Depresses Switch
to open position
)
WH
NC
RD
2
BK
2 1
D1
© Midmark Corporation 2004
SF-1854
J3
(To J13 on
Main
P.C. Board)
Rev. 10/10
Line**
VAC
Vent
Solenoid
Coil (NO)
Fill
BK
BK
Line**
VAC
Fill
Solenoid
Coil
J4
NO
11
BR/WH
Line
VAC**
12
LCD
Contrast
Adjustment
Wiring Diagrams
D-2
Line VAC
(When Door
is Open)
BR
Steam Heating
Element
015-1639-00
Display P.C. Board
RD
Hi Limit Jumper
015-1640-00
Note
Test point locations may vary.
Models:
Serial Numbers:
M9
(-020/-021/-022)
Early Models
Steam Heating
Element
6
5
4
3
2
1
1
6
L1
High
L1 Limit
High
Limit In
5
Main Harness
015-1554-00
BK
3
W3
YL
5
Air
Solenoid
Coil
W4
Heater
L1 N/L2
Lt BL
7
6
Vent
W6
J3
J2
RD
W5
RD
8
4
K1
Heater
W2
Door Interlock
Switch
Line**
VAC
10
F2
1
Line
VAC
(Energized for
apprx. 12 secs)
9
VI/WH
J6
F1
WH
J2 (To Switch Touch Pad)
VI
Air
W1
3
Door CC
Interlock
Switch
R1
2 3 4 5 6 7 8 9 10 11 12 13
Air Solenoid Coil
015-1638-00
J7
WH
VR1
1
RD
K4
K5
Vent
Air
Valve Valve
TP1
+12VDC*
Water
Level
Sensor
Probe
015-1552-00
J5
Note: Disconnect plug connector
when checking voltage.
12 11
10 9
+4.5*
VDC
2
1
C
* Constant Voltage
** Voltage Present Only
During Component Operation
*** Rectified DC Voltage Only
Present During Operation
14 13
Pressure
Transducer
T1
Transformer
K2
K3
Fill
Valve Door
J1
16 15
J12
TP4
+5VDC*
J8
Chamber
Level
ON
C
Switch
Motor
(Line Voltage)
230 VAC models:
F1 ................ 0.125 amp, 250 V, Slo-Blo, 5mm x 20mm
F2 ................ 8 amp, 250 V, Fast-Acting, 5mm x 20mm
+5VDC*
TP3
+5VDC*
PS1
WH
Line
VAC
Steam Temp
Sensor
1
+5
VDC*
Door
J15
BK
Fuses:
115 VAC models:
F1 ................ 0.250 amp, 250 V, Slo-Blo, 1/4" x 1-1/4"
F2 ................ 15 amp, 250 V, Fast-Acting, 1/4" x 1-1/4"
+4.5*
VDC
RD
U7
Door Opening
Motor Assembly
Door Motor Harness
015-1672-00
WH
2
DS1
PSD
JTAG
Port
J9
GN
3
TP5 Printer
+5VDC*
ON
1
2
3
4
5
6
7
8
Equipment Alert
Line voltage is present between
Ground and Heater terminals
L1 and L/N2 PCB connections at all times.
J1
J14 Printer
J10
RS232
Interface
1
3
5
7
9
11
13
Exploded Views / Part Numbers ........... E-1
8 7 6 5 4 3 2 1
J13 Display
1
2
3
4
+5
VDC*
Access Procedures .............................. C-1
Wiring Diagrams ................................... D-1
Steam Temp
Sensor Harness
015-1680-00
+5
+5
VDC* VDC*
3
4
2
Grd.
GN/YL
Plug
Outlet
13
GN/YL
L1
VAC
Out
Fan
L1
(Line VAC)
BK
N / L2
WH
WH
8
Lt BL
WH/BL
1
14
Hi Limits (N/C)
Opens 450 F 25
Closes 350 F 25
M11
(-020/-021/-022)
Early Models
WH
Fan Thermostat (N/O)
Fan Motor
Closes 130 F 8
Opens 101 F 6
SA1764i
Refer To:
Page
Operation & Troubleshooting ................ A-1
Component Testing / Repair ................. B-1
Nex
Go To Table O
Bac
Wiring Diagrams
TP2 Grd.
15 13 11 9 7 5 3 1
16 14 12 10 8 6 4 2
Exploded Views / Part Numbers ........... E-1
2
4
6
8
10
12
14
BK
WH
NC
RD
Line
VAC
2
BK
2 1
D1
Models:
Serial Numbers:
M9
Fill
BK VAC / VDC BK or WH
(Rectifier)
Steam Heating
Element
015-1639-00
LCD
Contrast
Adjustment
Hi Limit Jumper
015-1640-00
J3
(To J13 on
Main
P.C. Board)
NO
11
BR/WH
Line
VAC**
BR
Steam Heating
Element
(-020/-021/-022)
Later Models
VDC**
Fill
Solenoid
Coil
BK or WH
6
5
4
3
2
1
1
6
L1
High
L1 Limit
High
Limit In
5
Main Harness
015-1554-00
3
4
2
Grd.
GN/YL
Plug
Outlet
13
GN/YL
L1
VAC
Out
Fan
L1
(Line VAC)
BK
N / L2
WH
WH
8
Lt BL
WH/BL
1
14
Hi Limits (N/C)
Opens 450 F 25
Closes 350 F 25
M11
Vent
Solenoid
Coil (NO)
BK
J4
12
Display P.C. Board
(-020/-021/-022)
Later Models
Line VAC
(When Door
is Open)
VDC**
Vent RD VAC / VDC RD or WH
(Rectifier)
BK
3
W3
Lt BL
5
J3
J2
RD or WH
W4
Heater
L1 N/L2
YL
7
6
W6
Air
Solenoid
Coil
RD
W5
W2
Door Interlock
Switch
Line**
VAC
10
1
Line
VAC
(Energized for
apprx. 12 secs)
9
VI/WH
F2
K1
Heater
RD
8
4
VI
Air
F1
WH
J2 (To Switch Touch Pad)
2 3 4 5 6 7 8 9 10 11 12 13
Air Solenoid Coil
015-1638-00
J6
W1
3
Water
Level
Sensor
Probe
015-1552-00
RD
J7
TP1
+12VDC*
Door CC
Interlock
Switch
R1
1
+4.5*
VDC
2
1
C
WH
VR1
12 11
10 9
Pressure
Transducer
J5
Note: Disconnect plug connector
when checking voltage.
14 13
TP4
+5VDC*
K4
K5
Vent Air
Valve Valve
ON
C
Switch
Motor
(Line Voltage)
* Constant Voltage
** Voltage Present Only
During Component Operation
*** Rectified DC Voltage Only
Present During Operation
+5VDC*
J12
T1
Transformer
K2
K3
Fill
Valve Door
J1
16 15
TP3
+5VDC*
PS1
WH
230 VAC models:
F1 ................ 0.125 amp, 250 V, Slo-Blo, 5mm x 20mm
F2 ................ 8 amp, 250 V, Fast-Acting, 5mm x 20mm
Steam Temp
Sensor
1
+5
VDC*
J8
Chamber
Level
Door
J15
Fuses:
115 VAC models:
F1 ................ 0.250 amp, 250 V, Slo-Blo, 1/4" x 1-1/4"
F2 ................ 15 amp, 250 V, Fast-Acting, 1/4" x 1-1/4"
+4.5*
VDC
RD
U7
Door Opening
Motor Assembly
Door Motor Harness
015-1672-00
WH
2
DS1
PSD
JTAG
Port
J9
GN
3
TP5 Printer
+5VDC*
ON
1
2
3
4
5
6
7
8
(Motor Lever Arm
Depresses Switch
to open position
)
J1
J14 Printer
J10
RS232
Interface
1
3
5
7
9
11
13
Access Procedures .............................. C-1
Wiring Diagrams ................................... D-1
8 7 6 5 4 3 2 1
J13 Display
1
2
3
4
+5
VDC*
Equipment Alert
Line voltage is present between
Ground and Heater terminals
L1 and L/N2 PCB connections at all times.
Steam Temp
Sensor Harness
015-1680-00
+5
+5
VDC* VDC*
WH
Fan Thermostat (N/O)
Fan Motor
Closes 130 F 8
Opens 101 F 6
SA1762i
Wiring Diagrams
© Midmark Corporation 2004
SF-1854
D-3
Rev. 10/10
J14 Printer
J10 J13 Display
RS232
Interface
1
2
3
4
+5
VDC*
8 7 6 5 4 3 2 1
+4.5*
VDC
Steam Temp
Sensor
1
J12
TP3
+5VDC*
TP4
+5VDC*
PS1
J8
Chamber
Level
+4.5*
VDC
2
1
K4
K5
Vent
Air
Valve Valve
3
RD
2
1
BK
Air
J6
W1
F1
Door
J15
WH
Air Solenoid Coil
015-1638-00
VI
J7
TP1
+12VDC*
Chamber
Sensor
Harness
015-1552-00
RD
T1
Transformer
K2
K3
Fill
Valve Door
Exploded Views / Part Numbers ........... E-1
Pressure
Transducer
+5
VDC*
9
Line**
VAC
Air
Solenoid
Coil
Line**
VAC
Vent
Solenoid
Coil (NO)
VI/WH
10
F2
K1
Heater
RD
W5
Vent
W2
W6
RD
Refer To:
Page
Operation & Troubleshooting ................. A-1
Component Testing / Repair ................. B-1
Access Procedures .............................. C-1
Wiring Diagrams ................................... D-1
RD
PSD
JTAG
Port
J9
ON
1
2
3
4
5
6
7
8
WH
2
TP5 Printer
+5VDC* DS1
U7
2
4
6
8
10
12
14
1
3
5
7
9
11
13
GN
3
Equipment Alert
Line voltage is present between
Ground and Heater terminals
L1 and L/N2 PCB connections at all times.
Fuses:
115 VAC models:
F1 ................ 0.250 amp, 250 V, Slo-Blo, 1/4" x 1-1/4"
F2 ................ 15 amp, 250 V, Fast-Acting, 1/4" x 1-1/4"
230 VAC models:
F1 ................ 0.125 amp, 250 V, Slo-Blo, 5mm x 20mm
F2 ................ 8 amp, 250 V, Fast-Acting, 5mm x 20mm
* Constant Voltage
** Voltage Present Only
During Component Operation
*** Rectified DC Voltage Only
Present During Operation
W4
Door Interlock
Switch
J1
Heater
L1 N/L2
J3
J2
W3
Fill
BK
Line**
VAC
BK
Fill
Solenoid
Coil
J4
Note: Disconnect plug connector
when checking voltage.
J5
WH
BR
Steam Heating
Element
015-1639-00
Hi Limit Jumper
015-1640-00
Steam Heating
Element
3
2
1
VR1
1
6
L1
High
L1 Limit
High
Limit In
5
Main Harness
015-1554-00
BK
12
4
YL
11
5
Lt BL
NO
6
RD
Door CC
Interlock
Switch
Line
VAC**
BR/WH
WH
Line VAC
(When Door
is Open)
3
4
BK
2
Grd.
GN/YL
13
GN/YL
L1
VAC
Out
Fan
L1
(Line VAC)
16 15
12 11
10 9
J2 (To Switch Touch Pad)
1
2 3 4 5 6 7 8 9 10 11 12 13
WH
8
7
6
5
4
3
2 1
8
D1
LCD
Contrast
Adjustment
14 13
+5VDC*
Plug
Outlet
N / L2
WH
R1
J1
J3
(To J13 on
Main
P.C. Board)
Display P.C. Board
WH/BL
Lt BL
1
14
Hi Limits (N/C)
Opens 450 F 25
Closes 350 F 25
Wiring Diagrams
D-4
© Midmark Corporation 2004
SF-1854
Rev. 10/10
WH
Fan Thermostat (N/O)
Closes 130 F 8
Opens 101 F 6
Models:
Serial Numbers:
M9D
Fan Motor
(-020/-022)
Early Models
Note
Test point locations may vary.
SA1765i
M11D
(-020/-022)
Early Models
Nex
Go To Table O
Bac
TP2 Grd.
15 13 11 9 7 5 3 1
16 14 12 10 8 6 4 2
Wiring Diagrams
Steam Temp
Sensor Harness
015-1680-00
+5
+5
VDC* VDC*
TP2 Grd.
15 13 11 9 7 5 3 1
16 14 12 10 8 6 4 2
U7
2
4
6
8
10
12
14
1
3
5
7
9
11
13
GN
3
+4.5*
VDC
Steam Temp
Sensor
1
J12
TP3
+5VDC*
TP4
+5VDC*
PS1
J8
Chamber
Level
+4.5*
VDC
2
1
Air Solenoid Coil
015-1638-00
K4
K5
Vent
Air
Valve Valve
VI
J7
TP1
+12VDC*
3
RD
2
1
BK
Air
F1
9
Line**
VAC
10
F2
K1
Heater
W6
RD or WH
RD
W5
W2
Air
Solenoid
Coil
VI/WH
J6
W1
Door
J15
WH
Equipment Alert
Line voltage is present between
Ground and Heater terminals
L1 and L/N2 PCB connections at all times.
Chamber
Sensor
Harness
015-1552-00
RD
T1
Transformer
K2
K3
Fill
Valve Door
Exploded Views / Part Numbers ........... E-1
Pressure
Transducer
+5
VDC*
Refer To:
Page
Operation & Troubleshooting ................. A-1
Component Testing / Repair ................. B-1
Access Procedures .............................. C-1
Wiring Diagrams ................................... D-1
RD
PSD
JTAG
Port
J9
ON
1
2
3
4
5
6
7
8
WH
2
TP5 Printer
+5VDC* DS1
Wiring Diagrams
Steam Temp
Sensor Harness
015-1680-00
J14 Printer
J10 J13 Display
RS232
Interface
1
2
3
4
+5
VDC*
8 7 6 5 4 3 2 1
Nex
Go To Table O
Bac
+5
+5
VDC* VDC*
Vent RD VAC / VDC RD or WH
(Rectifier)
VDC**
Vent
Solenoid
Coil (NO)
Fuses:
115 VAC models:
F1 ................ 0.250 amp, 250 V, Slo-Blo, 1/4" x 1-1/4"
F2 ................ 15 amp, 250 V, Fast-Acting, 1/4" x 1-1/4"
230 VAC models:
F1 ................ 0.125 amp, 250 V, Slo-Blo, 5mm x 20mm
F2 ................ 8 amp, 250 V, Fast-Acting, 5mm x 20mm
* Constant Voltage
** Voltage Present Only
During Component Operation
*** Rectified DC Voltage Only
Present During Operation
W4
Door Interlock
Switch
J1
J2
Heater
L1 N/L2
J3
W3
Fill
BK VAC / VDC BK or WH
(Rectifier)
J4
Steam Heating
Element
015-1639-00
Hi Limit Jumper
015-1640-00
BR
Steam Heating
Element
3
2
1
VR1
1
6
L1
High
L1 Limit
High
Limit In
5
3
4
L1
(Line VAC)
BK
2
Grd.
GN/YL
13
GN/YL
L1
VAC
Out
Fan
Note: Disconnect plug connector
when checking voltage.
Main Harness
015-1554-00
BK
12
4
YL
11
5
Lt BL
NO
6
RD
Door CC
Interlock
Switch
WH
BR/WH
WH
Line
VAC**
Fill
Solenoid
Coil
BK or WH
J5
Line VAC
(When Door
is Open)
VDC**
BK
N / L2
WH
R1
J1
16 15
12 11
10 9
J2 (To Switch Touch Pad)
1
2 3 4 5 6 7 8 9 10 11 12 13
WH
8
7
6
5
4
3
2 1
8
D1
Lt BL
LCD
Contrast
Adjustment
14 13
+5VDC*
Plug
Outlet
J3
(To J13 on
Main
P.C. Board)
Display P.C. Board
WH/BL
1
14
Hi Limits (N/C)
Opens 450 F 25
Closes 350 F 25
Models:
Serial Numbers:
M9D
(-020/-022)
Later Models
WH
M11D
Fan Thermostat (N/O)
Closes 130 F 8
Opens 101 F 6
(-020/-022)
Later Models
Fan Motor
SA1763i
Wiring Diagrams
© Midmark Corporation 2004
SF-1854
D-5
Rev. 10/10