Keysight 16048A Test Leads
Operation and
Service Manual
3
Notices
The information contained in this document is subject to change without notice.
This document contains proprietary information that is protected by copyright.All
rights are reserved. No part of this document may be photocopied, reproduced, or
translated to another language without the prior written consent of the Keysight
Technologies.
Keysight Technologies Japan, Ltd.
Component Test PGU-Kobe
1-3-2, Murotani, Nishi-Ku, Kobe-shi, Hyogo, 651-2241 Japan
Copyright © 1990 - 2014 Keysight Technologies Japan, Ltd.
Manual Printing History
The manual’s printing date and part number indicate its current edition. The
printing date changes when a new edition is printed. (Minor corrections and
updates that are incorporated at reprint do not cause the date to change.) The
manual part number changes when extensive technical changes are incorporated.
1990
First Edition (part number: 16048-90000)
November 1998 Second Edition (part number: 16048-90001)
August 1999
Third Edition (part number: 16048-90001)
March 2000
Fourth Edition (part number: 16048-90001)
July 2010
Fifth Edition (part number: 16048-90001)
November 2014
Sixth Edition (part number: 16048-90001)
October 2016
Seventh Edition (part number: 16048-90001)
Safety Summary
The following general safety precautions must be observed during all phases of
operation, service, and repair of this instrument. Failure to comply with these
precautions or with specific WARNINGS elsewhere in this manual may impair the
protection provided by the equipment. In addition it violates safety standards of
design, manufacture, and intended use of the instrument.
4
Keysight Technologies assumes no liability for the customer’s failure to comply
with these requirements.
16048A comply with INSTALLATION CATEGORY I and POLLUTION
DEGREE 2 in IEC61010-1. 16048A are INDOOR USE product.
•
Ground The Instrument
To avoid electric shock hazard, the instrument chassis and cabinet must be
connected to a safety earth ground by the supplied power cable with earth
blade.
•
DO NOT Operate In An Explosive Atmosphere
Do not operate the instrument in the presence of flammable gasses or fumes.
Operation of any electrical instrument in such an environment constitutes a
definite safety hazard.
•
Keep Away From Live Circuits
Operating personnel must not remove instrument covers. Component
replacement and internal adjustments must be made by qualified maintenance
personnel. Do not replace components with the power cable connected. Under
certain conditions, dangerous voltages may exist even with the power cable
removed. To avoid injuries, always disconnect power and discharge circuits
before touching them.
•
DO NOT Service Or Adjust Alone
Do not attempt internal service or adjustment unless another person, capable of
rendering first aid and resuscitation, is present.
•
DO NOT Substitute Parts Or Modify Instrument
Because of the danger of introducing additional hazards, do not install
substitute parts or perform unauthorized modifications to the instrument.
Return the instrument to a Keysight Technologies Sales and Service Office for
service and repair to ensure that safety features are maintained.
•
Dangerous Procedure Warnings
Warnings, such as the example below, precede potentially dangerous
procedures throughout this manual. Instructions contained in the warnings must
be followed.
Dangerous voltages, capable of causing death, are presenting this instrument.
Use extreme caution when handling, testing, and adjusting this instrument
5
Certification
Keysight Technologies certifies that this product met its published specifications at
the time of shipment from the factory. Keysight Technologies further certifies that
its calibration measurements are traceable to the United States National Institute of
Standards and Technology, to the extent allowed by the Institution’s calibration
facility, or to the calibration facilities of other International Standards Organization
members.
Warranty
This Keysight Technologies instrument product is warranted against defects in
material and workmanship for a period corresponding to the individual warranty
periods of its component products. Instruments are warranted for a period of one
year. Fixtures and adapters are warranted for a period of 90 days. During the
warranty period, Keysight Technologies Company will, at its option, either repair
or replace products that prove to be defective.
For warranty service or repair, this product must be returned to a service facility
designated by Keysight Technologies. Buyer shall prepay shipping charges to
Keysight Technologies and Keysight Technologies shall pay shipping charges to
return the product to Buyer. However, Buyer shall pay all shipping charges, duties,
and taxes for products returned to Keysight Technologies from another country.
Keysight Technologies warrants that its software and firmware designated by
Keysight Technologies for use with an instrument will execute its programming
instruction when property installed on that instrument. Keysight Technologies does
not warrant that the operation of the instrument, or software, or firmware will be
uninterrupted or error free.
Limitation of Warranty
The foregoing warranty shall not apply to defects resulting from improper or
inadequate maintenance by Buyer, Buyer-supplied software or interfacing,
unauthorized modification or misuse, operation outside the environmental
specifications for the product, or improper site preparation or maintenance
No other warranty is expressed or implied. Keysight Technologies specifically
disclaims the implied warranties of merchantability and fitness for a particular
purpose.
6
Exclusive Remedies
The remedies provided herein are buyer’s sole and exclusive remedies. Keysight
Technologies shall not be liable for any direct, indirect, special, incidental, or
consequential damages, whether based on contract, tort, or any other legal theory.
Assistance
Product maintenance agreements and other customer assistance agreements are
available for Keysight Technologies products.
For any assistance, contact your nearest Keysight Technologies Sales and Service
Office. Addresses are provided at the back of this manual.
Safety Symbol
General definitions of safety symbols used on the instrument or in manuals are
listed below.
!
Instruction Manual symbol: the product is marked with this symbol when it is
necessary for the user to refer to the instrument manual.
This warning sign denotes a hazard. It calls attention to a procedure, practice,
condition or the like, which, if not correctly performed or adhered to, could
result in injury or death to personnel.
This Caution sign denotes a hazard. It calls attention to a procedure, practice,
condition or the like, which, if not correctly performed or adhered to, could
result in damage to or destruction of part or all of the product.
Note denotes important information. It calls attention to a procedure, practice,
condition or the like, which is essential to highlight.
7
8
Contents
1. Operation
Product Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Typical Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Compensation for Fixture Residual Impedance Error . . . . . . . . . . . . . . . . . . . . .
12
13
14
15
2. Service
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
9
Contents
10
1
Operation
This operating note provides complete information on the 16048A Test Leads. The
16048A is shown pictorially in Figure 1-1, its physical dimensions are given in
Table 1-1, and typical characteristics related to offset error are given in Table 1-2.
To order additional copies of this operating note, use the part number listed on the
rear cover.
11
Operation
Product Description
Product Description
The 16048A consists of a direct attachment, 4-terminal pair type fixture which is
equipped with four BNC (m) terminated-coaxial test leads. These test leads are
used to attach user-fabricated test fixtures. DC bias levels of up to ±42V can be
applied to the 16048A. Cable length is 1 meter. The 16048A is shown in Figure
1-1.
Figure 1-1
Product Overview
12
Chapter 1
Operation
Specifications
Specifications
Table 1-1
Specifications of the 16048A
Function:
For use with 4 terminal-pair LCR Meters and
Impedance Analyzers.
Connector Type:
BNC male
Cable Length:
Approximately 1m
Maximum Voltage:
± 42V peak max. (AC+DC)
Frequency Range:
DC to 30 MHz
Weight:
315 g
Safety Standards:
EN61010-1:1993 +A2:1995
IEC61010-1:1990 +A1:1992 +A2:1995
CSA C22.2 No.1010.1:1992
INSTALLATION CATEGORY I
POLLUTION DEGREE 2
INDOOR USE
Chapter 1
13
Operation
Typical Characteristics
Typical Characteristics
Table 1-2
Typical Characteristics
Incremental error at freq. 1 MHz
Parameter reading
error (%)
Offset value for D
f 2
 0.02   ------
10
f 2
 5   ------
 10
f is the measurement frequency in megahertz. The incremental errors calculated
from the equation in the table for measurements at frequencies above 1 MHz are
additive.
Figure 1-2
14
Chapter 1
Operation
Compensation for Fixture Residual Impedance Error
Compensation for Fixture Residual Impedance Error
The 16048A has inherent stray capacitance, residual inductance, and residual
resistance that affect the accuracy of measured values. To compensate for, or
negate, these residuals to minimize measurement error, the instrument’s
Open/Short compensation procedure should be performed. The procedure is given
in the instrument’s operating manual.
Chapter 1
15
Operation
Compensation for Fixture Residual Impedance Error
16
Chapter 1
2
Service
17
Service
Maintenance
Maintenance
An exploded view of the 16048A (for parts identification) is shown in Figure 2-1.
Do not disassemble any further than shown. Maintenance consists principally of
cleaning contacts and replacing worn or damaged parts. Take special care when
cleaning contacts. To order parts, use the Keysight Technologies part numbers
listed in Table 2-1. If a faulty part is located in an assembly that cannot be
disassembled, order the next higher assembly or return the whole device to the
nearest Keysight Technologies Sales/Service Office for repair or replacement.
18
Chapter 2
Service
Maintenance
Figure 2-1
Parts Identification
Chapter 2
19
Service
Maintenance
Table 2-1
Parts Identification
Keysight Part
No.
Reference
20
Qty.
Description
1
*
4
SLEEVE
2
*
4
NUT
3
*
4
NUT
4
*
4
WASHER
5
*
4
WASHER
6
2360-0192
2
SCREW
7
*
1
COVER-BOTTOM
8
*
4
INSULATOR
9
*
2
CONNECTOR-BNC
10
*
2
CONNECTOR-BNC
11
2200-0103
1
SCREW
12
2190-0206
1
WASHER
13
16047-40000
1
STOPPER
15
16048-04000
1
COVER-TOP
16
1400-0719
2
TIE RAP
17
*
1
GROMMET
18
*
1
CABLE
19
*
1
BUSHING
21
2360-0115
1
SCREW
22
3050-0010
1
WASHER
23
16021-50021
1
CABLE CRAMP
24
16021-50022
1
CABLE CRAMP
25
*
4
BOOT-BNC
26
*
4
SLEEVE
27
*
4
SLEEVE
28
*
4
INSULATOR
29
1250-0089
4
CONTACT
30
1250-0052
4
CONNECTOR-BNC
16048-60010
1
TEST LEAD
Note
1 thru 30
Chapter 2
Service
Maintenance
Reference
Keysight Part
No.
Qty.
Description
31
1250-0118
4
CONNECTOR-BNC
FEMALE
32
2360-0115
2
SCREW
33
16032-10021
1
PLATE
34
2190-0016
4
WASHER
35
2950-0001
4
NUT
36
16032-10022
1
PLATE
37
16032-60001
1
BNC BRACKET
Note
1 thru 30
31thru36
*: Not separately replaceable. Order 16048-60010
Chapter 2
21
Service
Maintenance
22
Chapter 2
This information is subject to change without notice.
© Keysight Technologies 1990 - 2014
Edition 7, October 2016
*16048-90001*
16048-90001
www.keysight.com
Download PDF