RISH Clamp ES1000 AC/ES400 AC 3 ¾ Digits digital Clamp meter

RISH Clamp ES1000 AC/ES400 AC 3 ¾ Digits digital Clamp meter
Data Sheet
3 ¾ Digits digital
Clamp meter
RISH Clamp ES1000 AC
RISH Clamp ES400 AC
AUTO/MAN
AUTO/MAN HOLD
REL
HOLD
REL
OFF
OFF
NCV
NC
/%
/%
V
RISHABH INSTRUMENTS PVT.LTD.
F-31, MIDC, Satpur, Nashik-422 007,India.
Tel.: +91 253 2202028, 2202202 Fax : +91 253 2351064
E-mail : India :- marketing@rishabh.co.in
International :- exp.marketing@rishabh.co.in
www.rishabh.co.in
www.rishabh.co.in
Page 1 of 5
Version: A 29/10/2014
RISH Clamp ES1000 AC/ES400 AC 3 ¾ Digits digital Clamp meter
Application
Hold Function
RISH Clamp Es1000 AC/ES400 AC measures important electrical
parameters like AC Current, AC Voltage, and DC Voltage. It also
features Capacitance, Ohm & Continuity, frequency, and Duty cycle
and temperature measurement.
By pressing HOLD key reading on the display can be
latched. Simultaneously HOLD is displayed on display.
Product Features
Hz / Duty
The instrument can measure frequency (Hz) and Duty cycle (%) of AC
voltage by pressing yellow key in VAC function.
Unique Design
RISH Clamp ES1000 AC/ES400 AC is a highly innovative design
for features those increases safety and comfort of user.
Rotating clamp jaws facilitate the measurement at physically
awkward positions, vertical bus bars, conductors placed at
positions difficult to access.
Clamp jaws can be opened or closed with the trigger placed at
bottom side away from the jaws. This allows the user to place
his/her hand at safer distance from live conductor. This greatly
reduces exposure of human beings to electrical shocks
Location and design of trigger eliminates fatigues caused by
single finger operation. It allows spreading the force required
to open the jaws over more than one finger to ensure
comfortable operation.
Comfortable operation of push buttons and function selector
switch, in adverse field conditions.
NULL ZERO Correction for Resistance

For Low ohm measurement, the lead resistance can be
compensated by pressing REL key.

Non contact voltage (NCV) detection
Presence of AC voltage >75 V AC 50/60 Hz can be detected
by keeping jaws near voltage carrying conductor.
It is indicated by beep sound.


AUTO and MANUAL ranging modes
In AUTO ranging mode the instrument automatically selects the
range with best resolution depending on the applied input.
In MANUAL ranging mode range is user selectable using MAN key.
Large Jaw Opening
For RISH Clamp ES1000 AC Jaw opening of 51mm for standard
wire diameter of 50mm and for RISH Clamp ES400 AC Jaw
opening of 41mm for standard wire diameter of 40mm for 400A
Diode and continuity testing
For testing diode and transistors, diode measurement function is
available. Continuity test generates beep sound if resistance is less
than 75 ohm
Narrow Body
Narrow housing for firm grip and easy to carry.
High Accuracy for low current measurement
Protection from dust and water
The clamp meter can measure accurately at not only the High
currents but also Low current ranges.
IP20 for terminals as per IEC60529
User selectable Backlit
Applicable International Safety standards
It is possible to conduct measurement using the clamp meter during
night time in darkness with the help of Backlit.
600 V CAT III/1000V CAT II as per International Safety standard
IEC 61010-1- 2010
Temperature measurement
Double molded Cover for soft touch and firm
grip of the Instrument
Temperature from 0 to 1300 °C using K type thermocouple
sensors.
AUTO POWER OFF
In order to save the power of the Batteries, the clamp meter will
automatically shut OFF if it detects no activity for 15 minutes.
Relative Measurement
By pressing REL key, the zero correction is made and relative
value is measured. All functions can measure Relative value
except Hz/Duty.
RISHABH INSTRUMENTS PVT.LTD.
F-31, MIDC, Satpur, Nashik-422 007,India.
Tel.: +91 253 2202028, 2202202 Fax : +91 253 2351064
E-mail : India :- marketing@rishabh.co.in
International :- exp.marketing@rishabh.co.in
www.rishabh.co.in
www.rishabh.co.in
Page 2 of 5
Version: A 29/10/2014
RISH Clamp ES1000 AC/ES400 AC 3 ¾ Digits digital Clamp meter
Specifications
Meas.
Function
Measuring
Range
Resolution
400.0mV
4.000V
40.00V
400.0V
1000V
400.0mV
4.000V
40.00V
400.0V
1000V
100µV
1mV
10mV
100mV
1V
100mV
1mV
10mV
100mV
1V
V
V
A
Clamp meter 400A
V(AC) / V(DC)
40.00A
10mA
400.0A
100mA
400.0A
100mA
A
Clamp meter 1000A
Input Impedance
1000A
1A
400.0
4.000k
40.00k
400.0k
4.000M
40.00M
100m
1
10
100
1k
10k
400.0
1.000V
5.000nF
50.00nF
500.0nF
5.000µF
50.00µF
200.0µF
100m
1mV
1pF
10pF
100pF
1nF
10nF
100nF
10.000Hz
100.00Hz
1.0000kHz
10.000kHz
100.00kHz
500.0kHz
0.001Hz
0.01Hz
0.1Hz
1Hz
10Hz
100Hz
Intrinsic error of digital display
at reference conditions
+ (...% of rdg + ...digits)
>20G
11M
10M
10M
10M
11M
11M
10M
10M
10M
Overload capacity
1)
0.75+2
Overload value
Overload
duration
0.5+2
1050V(DC)
Continuous
1050V(AC)
rms
Continuous
1.5+5
1+5
1+10
1.5 % of range +5 digits
480 A
Continuous
1.5 % of range +5 digits
1100A
Continuous
500V
DC/AC
rms
10 min
Open-circuit voltage
F
Hz
%
0
C
3)
2.0...98.0%
0
0...+1300 C
0.8+5
0.8+2
approx 0.45V
Acoustic signal for
0...<75 (approx)
approx 1V
0
2+10
3+402)
2+10 2)
0.5+3
1+2
1.5+2
5+10 4)
fmin
10Hz
10Hz
10Hz
10Hz
10Hz
10Hz
0.1%
1C
1+5
2+5
500V
DC/AC
rms
10 min
<1kHz : 1000V
0.2+2
<10kHz : 400V
Continuous
<500kHz : 40V
10Hz...1kHz : +5D
1kHz...10kHz : +5D/kHz
except 400mV
2+3 5)
500V DC/AC
rms
Sensor
K-type
NiCr-Ni
10 min
1) At 0º …. + 40 ºC
2) With zero adjustment, using REL key.
3) Indication of frequency measurement expanded to 9999 Digits.
4) Time required for measurement approximately 60 secs
5) Without sensor
Reference conditions for Accuracy
Reference temperature
Relative Humidity
Waveform of measured quantity
Input frequency
Battery Voltage
23°C ± 2K
45%...55% RH
Sinusoidal
50 or 60 Hz ±2%
3 V ± 0.1 V
RISHABH INSTRUMENTS PVT.LTD.
F-31, MIDC, Satpur, Nashik-422 007,India.
Tel.: +91 253 2202028, 2202202 Fax : +91 253 2351064
E-mail : India :- marketing@rishabh.co.in
International :- exp.marketing@rishabh.co.in
www.rishabh.co.in
www.rishabh.co.in
Page 3 of 5
Version: A 29/10/2014
RISH Clamp ES1000 AC/ES400 AC 3 ¾ Digits digital Clamp meter
Influence Quantities and Variations
Influence of
frequency on
Intrinsic
4V, 40V, 400V
400 mV,1000V
6
6
Battery voltage influence:
(Without display) - all ranges except capacitance: 8 Digits
- For capacitance 60 D at battery voltage 2.6V
RISHABH INSTRUMENTS PVT.LTD.
F-31, MIDC, Satpur, Nashik-422 007,India.
Tel.: +91 253 2202028, 2202202 Fax : +91 253 2351064
E-mail : India :- marketing@rishabh.co.in
International :- exp.marketing@rishabh.co.in
www.rishabh.co.in
www.rishabh.co.in
Page 4 of 5
Version: A 29/10/2014
RISH Clamp ES1000 AC/ES400 AC 3 ¾ Digits digital Clamp meter
Environmental
Operating temperature
Storage temperature
Relative humidity
Terminal Protection
-10 to +50°C
-25 to +70°C
45...75% non condensing
IP 52 for Housing and IP20
for terminals
Battery
Battery Voltage
1.5 x 2 V AAA size batteries
OR
Battery type
IEC 6LR 03
Battery Life
200 hrs
400 hrs
Display
3
Applicable Standards
EMC
Immunity
IEC 61326: Class B
IEC 61000-4-2
8 KV atmosphere discharge,
4 KV contact discharge
IEC 61000-4-3 : 3 V/m
Safety
IP for water & dust
Pollution degree
Installation category
High Voltage Test
IEC 61010-1-2010
IEC60529
2
600V CATIII / 1000V CATII
4.4 kV AC, 50Hz for 1 minute
between housing and input.
Mechanical configuration
Dimensions
Weight
90mm (W) x 270mm (L) x 70mm (H)
0.6 Kg
RISHABH INSTRUMENTS PVT.LTD.
F-31, MIDC, Satpur, Nashik-422 007,India.
Tel.: +91 253 2202028, 2202202 Fax : +91 253 2351064
E-mail : India :- marketing@rishabh.co.in
International :- exp.marketing@rishabh.co.in
www.rishabh.co.in
www.rishabh.co.in
Page 5 of 5
Version: A 29/10/2014
RISH Clamp-on CT 1000/1A/, CT 1000/5A, CT 300/0.3A
Clamp-on CT for AC current measurement
Data Sheet
Clamp- on CT for AC
current measurement
Rish clamp on CT 300 A / 0.3A
Rish clamp on CT 1000/ 5A
RISH Clamp -on CT 300 A should be used as
accessories for RiSH Multi 12s which directly
displays the Current in 300 A range.
RISH Clamp -on CT 1000 A should be used as
accessories for RiSH Multi series.
Rish clamp on CT 1000 A / 1A
RISH Clamp -on CT 1000 A should be used as
accessories for RiSH Multi series.
www.rishabh.co.in
Page 1 of 2
Version: A 03/01/2012
Application
Periodic Check-up
Rish Clamp-on CT 300 A / 1000A measures current without
interrupting the conducting line. They are used for electrically
isolated measurement of alternating current within a broad range
,without interrupting the current conducting cable and bus bars.
The clamp on CT does not require any specific maintenance. While
cleaning, the probes must not be immersed into any liquid. No
special maintenance is required for housing. The surface between
the opening jaws should be cleaned with dry cloth before operating.
Avoid use of cleansers, abrasives or solvents.
Features of Rish Clamp on
CT 300 A / CT 1000 A
Safety Features & Precautions
- Application around & removal from
HAZARDOUS LIVE conductor is permitted
Safety in accordance with IEC 61010-1.
Minimumtransformation ratio error.
n
Compact, handy design.
n
Permanently connected safety cables.
n
n
Measurement
Rish Clamp on CT measures AC current without interrupting the
conducting line. The opening contact of both the core which
become visible when trigger is pressed, must be always kept clean.
Even small outside particles can deviate measuring accuracy. The
measuring secondary long cables with contact- protected
connection plugs, which electrically isolates the measurement of
AC current without interruption.
The Rish Clamp on current transformers are constructed with
safety rules of IEC 61010-1. When in operation the safety of user
and unit is assured. Also when the secondary circuit is interrupted,
the Clamp has a Protection circuit which prevents high voltage to be
generated. Before operating read completely the instruction
manual. These instruction must be followed by users in all respects.
In case of incorrect use or careless handling, the safety of both user
and clamp is not assured.
Specifications
Clamp on current transformer
Type
Descrition/
Appllication
CT 1000A / 1A
CT 300A / 0.3A
1000A /5A
The current transformers are used for electrically isolated measurement
of alternating current within broad range.It is possible to measure AC
current without interrupting the current conducting cable and bus bars
via hard wired leads with contact protected plugs.
Nominal Primary
Current
1-1000 A AC
4-300 A AC
1-1000 A AC
Output signal
1mA AC / A AC
0.3 mA AC / A AC
1mA AC / A AC
Used with
Transmission ratio
Input:Output
Accuracy /
max.error
Max. operating
Voltage
Rish multi series
1000:1
Rish multi 12s
Rish multi series
1000:1
200:1
Class 1 ( 1.25 VA)
Class 1 (2.5 VA)
(for current r ange f rom (for current range from
4A to 300A, 50Hz)
1A t o 1000A, 50Hz)
CAT III 600 V
as per IEC-61010
CAT III 60 0 V
as p er IE C- 61010
Frequency range 40H z. ...1kHz
40Hz....1kHz
40H z. ...1kHz
Jaw opening /
Conductor
Diameter
35mm /
Diameter 35mm
43mm /
Diam eter 43 mm
Dimensions
CAT III 60 0 V
as p er IE C- 61010
Class 1 (2.5 VA)
(for current r ange f rom
1A t o 1000A, 50Hz)
43mm /
Diam eter 43 mm
(W X H X D)
( 92 X 39 X 220) mm (84 X 39 X 202)mm
Weight
A ppro x 650 gms
Approx 500 gms
A ppro x 650 gms
Cable length
1.5 me ter
1.5 meter
1.5 me ter
( 92 X 39 X 220) mm
RISHABH INSTRUMENTS PVT.LTD.
F-31, MIDC, Satpur, Nashik-422 007,India.
Tel.: +91 253 2202160, 2202202 Fax : +91 253 2351064
E-mail : India :- marketing@rishabh.co.in
International :- exp.marketing@rishabh.co.in
www.rishabh.co.in
www.rishabh.co.in
Page 2 of 2
Version: A 03/01/2012
RISH Clamp POWER
1000A / 400A AC-DC
Technical Datasheet
1000 A AC-DC
RISH Clamp POWER is specially designed for
Measurement of AC-DC POWER and
Power quality parameters.
Up to 49th Individual Harmonics
Non contact voltage detection.
Inbuilt Three Phase power Measurement. Energy Measurement
LPF Mode for VFD
Horse Power Measurement
F-31, MIDC, Satpur, Nashik-422 007,India.
Tel.: +91 253 2202160, 2202202 Fax : +91 253 2351064
E-mail : India :- marketing@rishabh.co.in
International :- exp.marketing@rishabh.co.in
www.rishabh.co.in
www.rishabh.co.in
Page 1 of 6
Version: A 06/04/2015
RISH Clamp POWER
1000A / 400A AC-DC
Application
RISH Clamp POWER 1000A/400A measures, calculate and displays
important electrical parameters of single phase or three phase
power system. It also features Resistance, continuity, diode and
non contact voltage detection.
Product Features
Measures following parameters
Ÿ
AC & DC Voltage up to 1000V
Ÿ
AC & DC Current up to 1000A /400 A
Ÿ
Inrush/Peak Value Measurement
Ÿ
Active, Reactive and Apparent Power
Ÿ
Horse Power Measurement
Ÿ
kWh
Ÿ
Measure up to 49th Harmonics
Ÿ
Phase Angle
Ÿ
THD
Ÿ
DF
Ÿ
Power Factor
Ÿ
Crest Factor
Ÿ
LPF Mode for VFD Application
RISH Clamp POWER
400 A AC-DC
REL
INRUSH
HOLD
MIN/MAX
Non Contact Voltage Detection
Unique Design
RISH Clamp POWER1000A/400A is a highly innovative design for
features those increases safety and comfort of user.
Rotating clamp jaws facilitate the measurement at physically
awkward positions, vertical bus bars, conductors placed at
positions difficult to access.
Clamp jaws can be opened or closed with the trigger placed at
bottom side away from the jaws. This allows the user to place
his/her hand at safer distance from live conductor. This greatly
reduces exposure of human beings to electrical shocks.
Location and design of trigger eliminates fatigues caused by
single finger operation. It allows spreading the force required
to open the jaws over more than one finger to ensure
comfortable operation.
Comfortable operation of push buttons and function selector
switch, in adverse field conditions.
n
n
n
n
The clamp meter can detect the presence of AC Voltage between
100 to 1000 V 50hz/60Hz without any electrical connection and
give acoustic signal as an indication.
Three Phase Power Measurement
Clamp meter can measure power in 3 phase 3 wire or 3 phase 4
wire (Symmetric as well as Asymmetric) network without any
manual calculation like other clamp meters.
Dual Display
User friendly dual display shows the simultaneous parameters of
measuring input quantity.
LPF Mode
LPF mode is available for voltage and current for true measurement
of VFD Application
Large Jaw Opening
TRMS Measurement
Jaw opening of 51mm and 41 mm for standard wire diameter of
50mm and 40mm for 1000A and 400A respectively.
Narrow Body
In order to calculate true value of distorted waveform due to
presence of high crest factor or harmonics, TRMS measurements is
done for AC voltage and current
Narrow housing for firm grip and easy to carry.
Auto Power OFF
DATA Hold Function
In order to save the power of the Batteries, the clamp meter will
automatically shut OFF if it detects no activity for 10 minutes.
By pressing DATA HOLD button, reading on the display can be
latched for Hands free operation.
Continuous ON Mode
MIN,MAX Function
In this mode, AUTO POWER OFF is disabled.
By pressing MIN/MAX button, the clamp meter will start recording
latest Minimum and Maximum readings
Low Battery Indication
User Selectable Backlit
Double molded Cover for soft touch and firm grip
of the Instrument
It is possible to conduct measurement using the clamp meter during
www.rishabh.co.in
Page 2 of 6
Version: A 06/04/2015
Reference conditions for Accuracy
Protection from dust and water
Reference temperature
Relative Humidity
Input frequency
Power Factor
Battery Voltage
IP20 for terminals as per IEC60529
23°C ± 2°C
45%...55% RH
50 or 60 Hz
0.5L....1....0.5C
8 V ± 0.1 V
Applicable International Safety standards
600 V CAT IV/1000V CAT III as per International Safety standard
IEC 61010-1- 2010
Technical Specification
Measuring
function
Measuring
range
Resolution
Intrinsic error of digital display
at reference condition
V DC
999.9 V
0.1 V
±(0.5% of rdg + 5 dgt)
999.9 V
0.1 V
±(0.75% of rdg+5 dgt)
V ACDC
999.9 V
0.1 V
±(1.25% of rdg+10dgt)
LPF V~
999.9 V
0.1 V
V~
POWER CLAMP
1000A ADC-AAC
POWER CLAMP
400A ADC - AAC
POWER CLAMP
1000A A ACDC
POWER CLAMP
400A A ACDC
POWER CLAMP
LPF 1000A A AC
POWER CLAMP
LPF 400A A AC
Active
2)
Power
Reactive
Power 2)
Apparent
Power
2)
Horse
Power 2)
kWh 2)
www.rishabh.co.in
Over load capacity
50.....60 Hz
±(0.75% of rdg + 5dgt)
61...400Hz
±(5.0% of rdg + 5dgt)
999.9A
0.1 A
±(1.5% of rdg+5 dgt)1)
99.99 A
0.01 A
400 A
0.1 A
display value ±(1.5% of rdg+0.2A)1)
<1000 add 10
1)
±(1.5% of rdg+5 dgt)
dgt
999.9A
0.1 A
±(3% of rdg+10 dgt)1)
99.99 A
0.01 A
400 A
0.1 A
999.9A
0.1 A
99.99 A
0.01 A
400 A
0.1 A
9.999 kW
1W
99.99 kW
10 W
999.9 kW
100 W
9999 kW
1 kW
9.999 kVAr
1 VAr
99.99 kVAr
10 VAr
999.9 kVAr
100 VAr
9999 kVAr
1 kVAr
9.999 kVA
1 VA
99.99 kVA
10 VA
999.9 kVA
100 VA
9999 kVA
1 kVA
9.999 hp
0.001 hp
99.99 hp
0.01 hp
999.9 hp
0.1 hp
9999 hp
1 hp
9.999 kWh
0.001 kWh
99.99 kWh
0.01kWh
999.9 kWh
0.1 kWh
9999 kWh
1 kWh
display value
<1000 add 10
dgt
50....60 Hz
61...400Hz
50....60 Hz
61...400Hz
50....60 Hz
61...400Hz
±(3% of rdg+0.4A)
Over load value
Overload
duration
1000 V DC/AC
eff/rms Sine wave
Continuously
1100 A AC/DC
for power clamp
1000A
1)
Continuously
±(3% of rdg+10 dgt)1)
±(1.5% of rdg + 5dgt)
±(5.0% of rdg + 5dgt)
±(1.5% of rdg + 0.3A)
±(5.0% of rdg + 5dgt)
±(1.5% of rdg + 5dgt)
±(5.0% of rdg + 5dgt)
±(2% of rdg+5 dgt) 1)
440 A AC/DC
for power clamp
400A
1000 V DC/AC
1100 A AC/DC for
Power Clamp
1000A
440 A AC/DC for
Power Clamp
400A
Continuously
±(3% of rdg+5 dgt)
Page 3 of 6
Version: A 06/04/2015
Technical Specification
Measuring
function
Measuring
range
Resolution
Intrinsic error of digital display
at reference condition
Ahr
999.9 Ahr
0.1 Ahr
±(3% of rdg+5 dgt)
0.0°…360.0°
0.1°
-1…0…1
0.001
Harmonics
1…13
±(3% of rdg+10 dgt)
(RMS & %)3)
14…49
0.1V
0.1A
0.1%
THD
0…99.9%
0.1%
±(3% of rdg+20 dgt)
3)
0…99.9%
0.1%
±(3% of rdg+20 dgt)
1.0…2.9
0.1
±(2% of rdg+3 dgt)
3.0…5.0
0.1
±(3% of rdg+5 dgt)
1400 A /
1400V
1A
±(3% of rdg+3 dgt)
0.1 A
±(3% of rdg+10 dgt)
1A/1V
±(3% of rdg+3 dgt)
999.9A
0.1 A
±(3% of rdg+5 dgt)
99.99 A
0.01 A
±(3% of rdg+0.3A)
400 A
0.1 A
±(3% of rdg+5 dgt)
Resistance
4000 Ohm
1 Ohm
±(0.5% of rdg+5 dgt)
Continuity
Below 40 Ohm
1 Ohm
±(0.5% of rdg+5 dgt)
Diode
0…2.2V
0.001 V
±(0.5% of rdg+5 dgt)
2)
Phase angle
Pow er Factor
2)
3)
DF
Crest Factor3)
POWER CLAMP
1000A Peak
100 A
POWER CLAMP
400A Peak
560 A/ 1000 V
POWER CLAMP
4)
1000A INRUSH
POWER CLAMP
4)
400A INRUSH
Over load capacity
Over load value
Overload
duration
1000 V DC/AC
1100 A AC/DC for
Power Clamp
1000A
440 A AC/DC for
Power Clamp
400A
Continuously
±3°
±(5% of rdg+20 dgt)
1000 V DC/AC
eff/rms Sine wave
10 Secs
Note:- Accuracy claimed for Power and Current when conductor is positioned at the center of the jaw.
1)For DC A make auto zero correction by long pressing the HOLD key
For Power Clamp 1000A
2) Accuracy Defined for V ≧ 10V and I ≧ 5A
Add 10 digit to accuracy when power is
< 5.000 kW/kVAr/kVA or <6.700 hp
3) Accuracy Defined for V ≧ 10V and I ≧ 10A
4) Accuracy Defined for I ≧ 10A
For Power Clamp 400A
2) Accuracy Defined for V ≧ 10V and I ≧ 4A
Add 10 digit to accuracy when power is
< 5.000 kW/kVAr/kVA or <6.700 hp
3) Accuracy Defined for V ≧ 10V and I ≧ 10A
4) Accuracy Defined for I ≧ 5A
For Power Clamp 1000A
In 1P2W mode maximum power meter can measure is, 1000 kVA / 1000 kVAr / 1000 kW / 1341 hp
In 3P4W mode maximum power meter can measure is,3000 kVA / 3000 kVAr / 3000 kW / 4023 hp
In 3P3W mode maximum power meter can measure is,1732 kVA / 1732 kVAr / 1732 kW / 2322 hp
For Power Clamp 400A
In 1P2W mode maximum power meter can measure is, 400 kVA / 400 kVAr / 400 kW / 536 hp
In 3P4W mode maximum power meter can measure is,1200 kVA / 1200 kVAr / 1200 kW / 1608 hp
In 3P3W mode maximum power meter can measure is,693 kVA / 693 kVAr / 693 kW / 928 hp
www.rishabh.co.in
Page 4 of 6
Version: A 06/04/2015
Influence Quantity
Infulence
quantity
Range of Infuence
Temperature
Frequecy of
the measured
quantity
0 ºC… 21 ºC
and
25 ºC….50 ºC
40 Hz… 50 Hz
and
60 Hz….400 Hz
2)
Crest Factor
1)
Supply
Voltage
Relative
humidity
45 Hz….65 Hz
1.4…2
2…2.5
2.5...5
When Low Battery
symbol is ON
75%
Measured
quantity /
Measuring Range
V AC
V DC
V ACDC
A AC
A DC
A ACDC
AC Power
DC Power
Resistance/ Diode/
Continuity
V AC
V ACDC
A AC
A ACDC
AC Power
Variation
0.15 X Intrinsic Error / ºC
1 X Intrinsic Error
V AC
A AC
1% + Intrinsic Error
2.5% + Intrinsic Error
4% + Intrinsic Error
All Ranges
1 X Intrinsic Error
All Ranges
1 X Intrinsic Error
1) Except SineWave
CF 2 @ 690V, 690Afor Power Clamp Meter 1000 A ACDC
CF 3 @ 690V, 186A for Power Clamp Meter 400 AACDC
CF 4 @ 345V, 345A for Power Clamp Meter 1000 A ACDC
CF 4 @ 345V, 140A for Power Clamp Meter 400 A ACDC
CF 2 @ 690V, 280A for Power Clamp Meter 400 AACDC
CF 5 @ 280V, 280A for Power Clamp Meter 1000 A ACDC
CF 3 @ 460V, 460A for Power ClampMeter 1000 A ACDC
CF 5 @ 280V, 112A for Power Clamp Meter 400 A ACDC
2) Except for 50 or 60 Hz
www.rishabh.co.in
Page 5 of 6
Version: A 06/04/2015
Display
Environmental
Operating temperature
Storage temperature
Temp. Coefficient
0 to +55°C
-20 to +70°C
0.15 X(Intinsic Error) / °C
Relative humidity
0... 75% non condensing
Terminal Protection
for terminals
IP50 for Housing
and IP20
Applicable Standards
EMC
Emission
Immunity
Number of digits
Maximum count
Over range indication
Polarity indication
Seven Segment
Main Display Character : 11.5 mm
Sub Display Character : 7.2 mm
4 digits.
9999 counts For V, I and Power
4000 counts For Resistance
“OL” is displayed
“─” sign is displayed for
negative values.
Battery
Electro magnetic compatibility
IEC 61326: 2012 ClassB
IEC 61326: 2012
IEC61000-4-2 :8 KV air discharge,
4 KV contact discharge
IEC 61000-4-3 :- 3 V/m
Battery Voltage
Battery type
Consuption
Battery Life
9 V DC
Manganese Dioxide Cell as per
IEC6F22
Alkaline manganese cell as per
IEC 6LR 61
20 mA Avg. (Without Backlight)
48 Hrs Approx.
Scope of delivery
Safety
IEC 61010-1-2010 -06
IP for water & dust
IP 50 for housing
IP 20 for terminal
2
III
IV
1000V
600V
Pollution degree
Installation category
Display
Character Height
Ÿ
Ÿ
Ÿ
Ÿ
Ÿ
Ÿ
Ÿ
Clamp Meter
Probe Set
Instruction Manual/Warranty card
Clamp Carrying Case
Test Certificate
Battery
Two crocodile clips
High Voltage Test
between housing and input.
7.4 kV AC, 50Hz for 1 minute
between housing with jaws and 4.26 kV AC, 50Hz for 1 minute
input.
Rishabh Instruments always tries for innovation and therefore product specifications are subject to change without notice
RISHABH INSTRUMENTS PVT.LTD.
F-31, MIDC, Satpur, Nashik-422 007,India.
Tel.: +91 253 2202028, 2202202 Fax : +91 253 2351064
E-mail : India :- marketing@rishabh.co.in
International :- exp.marketing@rishabh.co.in
www.rishabh.co.in
www.rishabh.co.in
Page 6 of 6
Version: A 06/04/2015
RISH Clamp1000A/300A AC-DC
TRMS 3 ¾ Digits digital Clamp meter with VFD measurement feature
Data Sheet
3 ¾ Digits digital
Clamp meter
RISH Clamp 300 AC-DC
www.rishabh.co.in
Page 1 of 5
Version: E 08/04/2014
Application
CONTINOUS ON MODE
RISH Clamp 1000A/300A measures important electrical parameters like
AC Current (TRMS), DC Current, AC Voltage(TRMS), and DC Voltage.
It also features Capacitance, Ohm & Continuity, frequency, and Duty
cycle and temperature measurement.
In this mode, AUTO POWER OFF is disabled.
DATA Hold Function
Product Features
By pressing DATA HOLD button, reading on the display can be
latched for Hands free operation.
Unique Design
MIN,MAX Function
Rish Clamp 1000A/300A is a highly innovative design for features those
increases safety and comfort of user.
Rotating clamp jaws facilitate the measurement at physically
awkward positions, vertical bus bars, conductors placed at
positions difficult to access.
Clamp jaws can be opened or closed with the trigger placed at
bottom side away from the jaws. This allows the user to place
his/her hand at safer distance from live conductor. This greatly
reduces exposure of human beings to electrical shocks
Location and design of trigger eliminates fatigues caused by
single finger operation. It allows spreading the force required
to open the jaws over more than one finger to ensure
comfortable operation.
Comfortable operation of push buttons and function selector
switch, in adverse field conditions.




Large Jaw Opening
By pressing MIN/MAX button, the clamp meter will start recording
latest Minimum and Maximum readings
NULL ZERO Correction for Resistance
For Low ohm measurement, the lead resistance can be
compensated by pressing the shift key (Yellow Key)
NULL ZERO Correction for Capacitance
Null zero connection for capacitance. For nF range, stray
capacitance can be compensated by shift key (Yellow Key)
AUTO and MANUAL ranging modes
In AUTO ranging mode the instrument automatically selects the
range with best resolution depending on the applied input.
In MANUAL ranging mode range is user selectable using MAN key.
For RISH CLAMP 1000A AC-DC Jaw opening of 51mm for standard
wire diameter of 50mm and for RISH CLAMP 300A AC-DC Jaw
opening of 41mm for standard wire diameter of 40mm for 300A
Diode Measurement
Narrow Body
For testing diode and transistors, diode measurement function is
available.
Narrow housing for firm grip and easy to carry.
Protection from dust and water
High Accuracy for low current measurement
IP20 for terminals as per IEC60529
The clamp meter can measure accurately at not only the High
currents but also Low current ranges.
Applicable International Safety standards
True Root Mean Square (TRMS) measurement
Clamp meter measures AC signal’s root-mean-square value
accurately irrespective of the shape of input waveform.
600 V CAT IV/1000V CAT III as per International Safety standard
IEC 61010-1- 2010
Double molded Cover for soft touch and firm
grip of the Instrument
Measurement on Variable Frequency Drives
The clamp meter can measure accurately on variable frequency
drives (VFD) and UPS.
User selectable Backlit : (Optional)
It is possible to conduct measurement using the clamp meter during
night time in darkness with the help of Backlit. The back lit can be
switched ON or OFF by pressing a single key.
Temperature measurement
Temperatures from -200 to 800 °C using Pt 100 and Pt 1000
sensors.
AUTO POWER OFF
In order to save the power of the Batteries, the clamp meter will
automatically shut OFF if it detects no activity for 10 minutes.
Analog Scale
Analog scale that updates at the rate 20 times/sec to observe
fluctuations in input.
www.rishabh.co.in
Page 2 of 5
Version: E 08/04/2014
Measuring
function
V dc
V~
Ω
Rish clamp 1000A~/Adc
Rish clamp 300A~/Adc
Measuring
Function
F
Hz
Measuring
range
Resoluti
on
30.00 mV
300.0 mV
3.000 V
30.00 V
300.0 V
1000 V
3.000 V
30.00 V
300.0 V
1000 V
10 µV
100 µV
1 mV
10 mV
100 mV
1V
1 mV
10 mV
100 mV
1V
>10 GΩ // <40pF
>10 GΩ // <40pF
11 MΩ // <40pF
10 MΩ // <40pF
10 MΩ // <40pF
10 MΩ // <40pF
11 MΩ // <40pF
10 MΩ // <40pF
10 MΩ // <40pF
10 MΩ // <40pF
No load voltage
30.00
300.0
3.000
30.00
300.0
3.000
30.00
Ω
Ω
KΩ
KΩ
KΩ
MΩ
MΩ
10 mΩ
100 mΩ
1Ω
10 Ω
100 Ω
1 KΩ
10 KΩ
Max. 3.2 V
Max. 3.2 V
Max. 1.25 V
Max. 1.25 V
Max. 1.25 V
Max. 1.25 V
Max. 1.25 V
2.000 V
1 mV
Max. 3.2 V
0.1 A
1A
0.1 A
1A
-----------------
2 to 300.0 A
1000 A
0.2 to 30.0A
300.0 A
Measuring range
Input
impedance
Resolution
Intrinsic error of digital display
± (…% of rdg + …digit) at
reference condition
Over load capacity
Over load value
Overload duration
1000 V
DC
AC
eff / rms
Sine wave
Continuously
1000 V
DC
AC
eff / rms
Sine wave
10 min
1100 A
Continuously
0.5 + 3 2)
0.5 + 3
0.25 + 1
0.25 + 1
0.25 + 1
0.35 + 1
0.75 + 2 (10….300 Digit)
0.75 + 1 ( > 300 Digit )
0.5 + 3 2)
0.5 + 3
0.4 + 1
0.4 + 1
0.4 + 1
0.6 + 1
2.0 + 1
0.25 + 1
1.5 % of range + 5 Digits
Discharge
resistance
U0 max.
30.00 nF
300.0 nF
3.000 µF
30.00 µF
10 pF
100 pF
1 nF
10 nF
250 KΩ
250 KΩ
25 KΩ
25 KΩ
2.5 V
2.5 V
2.5 V
2.5 V
300.0 Hz
3.000 KHz
30.00 KHz
100.0 KHz
0.1 Hz
1 Hz
10 Hz
100 Hz
f min V dc
1 Hz
1 Hz
10 Hz
100 Hz
f min V ~
45 Hz
45 Hz
45 Hz
100 Hz
Intrinsic error of digital display
± (…% of rdg + …digit) at
reference condition
2)
1.0 + 3
1.0 + 3
1.0 + 3
3.0 + 3
0.5 + 1
3)
4)
%
2.0....98.0%
Pt
100
ºC
Pt
1000
-200.0...
+200.0 ºC
+200.0...
+850.0 ºC
-100.0...
+200.0 ºC
+200.0...
+850.0 ºC
0.1 %
2 Hz
-
0.1 ºC
2 Hz… 1kHz ± 5 Digit
1 kHz … 10 kHz; ± 5 Digit / kHz
2 Kelvin + 5 Digit
-
1.0 + 5
0.1 ºC
2 Kelvin + 2 Digit
-
0.1 ºC
1.0 + 2
4)
Over load capacity
1)
Over load
value
Over load
duration
1000 V
DC
AC
eff / rms
Sine
10 min
3 kHz
1000 v
30 kHz;
300 V
100 kHz
30 V
Continuously
5)
5)
0.1 ºC
-
1)
5)
1000 V
DC
AC
eff / rms
Sine
10 min
5)
1) At 0º …. + 40 ºC
2) With zero adjustment, without zero adjustment + 35 digits
3) Range :
3 V ac/dc:
Ue = 1.5 V eff/rms … 100 V eff/rms
30 V ac/dc: Ue = 15 V eff/rms … 300 V eff/rms
300 V ac/dc: Ue = 150 V eff/rms … 1000 V eff/rms
4) On the range 3 V dc, square – wave signal positive on one side 5 … 15 V,
f = const., not 163.84 Hz or integral multiple.
5) Without sensor
Reference conditions for Accuracy
Reference temperature
Relative Humidity
Waveform of measured quantity
Input frequency
Battery Voltage
www.rishabh.co.in
23°C ± 2K
45%...55% RH
Sinusoidal
50 or 60 Hz ±2%
8 V ± 0.1 V
Page 3 of 5
Version: E 08/04/2014
Display
Environmental
Operating temperature
Storage temperature
Relative humidity
Terminal Protection
-10 to +55°C
-20 to +70°C
0... 90% non condensing
IP50 for Housing and IP20
for terminals
Number of digits
Maximum count
Over range indication
3 ¾ digits.
3100 counts.
“OL” is displayed.
Polarity indication
“─” sign is displayed for DC
functions, if positive pole is at“ ”.
Battery
Battery Voltage
Battery type
Battery Life
9 V DC
Manganese Dioxide Cell as per
IEC6F22 , alkaline manganese
cell as per IEC 6LR 61
Minimum 220 hours on Vdc,
Adc, 80 hours on Vac, Aac.
Influence Quantities and Variations
Influence Quantity
Temperature
Range of Influence
0 ºC
+21 ºC
and
+25 ºC…+40ºC
1)
Measured Quantity/
Measuring Range
Variation
± (….% of rdg. + ….digits)
30/300 mV dc
3...300 V dc
1000 V dc
V~
30 Ω 2)
300 Ω
3 KΩ – 3 MΩ
30 MΩ
2)
30 nF – 3 μF
30 μF
Hz
%
-200…+200 ºC
+200…+850ºC
1.0 + 3
0.15 + 1
0.2 + 1
0.4 + 2
0.15 + 2
0.25 + 2
0.15 + 1
1.0 + 1
0.5 + 2
2.0 + 2
0.5 + 1
± 5 digits
0.5 K + 2
0.5 + 2
30 A ~/ A DC
0.2 X Specified accuracy
300 A ~/ A DC
300 A ~/ A DC
0..1 X Specified accuracy
0.2 X Specified accuracy
1000 A ~/ A DC
0..1 X Specified accuracy
Rish clamp1000A AC-DC
Rish clamp300A AC-DC
> 65 Hz...400 Hz
>400 Hz…1 KHz
Frequency of the
measured quantity
>65 Hz … 1 KHz
3…300 V ~
2.0 + 3
1000 V ~
3.0 + 3
15Hz …<45 Hz
>66 Hz… 400 Hz
www.rishabh.co.in
A~
Page 4 of 5
1.0 % of range + 1
Version: E 08/04/2014
Influence Quantity
Range of Influence
Measured Quantity/
Measuring Range
± 1 % of rdg
1....3
Wave form of the
3)
measured quantity
V ~ 4) A ~ 4)
Crest factor CF
1....5
± 3 % of rdg
Battery Voltage
V DC
V~
AAC/ADC
30Ω / 300 Ω/˚C
3 kΩ – 30MΩ
nF, μF
Hz
%
V~,VDC
A~, ADC
Ω
F
Hz
%
0
C
-V ac/dc , A ~ , ADC
5)
…< 7.9 V
> 8.1 V …10.0 V
75%
Relative humidity
3 Days
Meter off
-
HOLD
MIN/MAX
1)
Variation
± (….% of rdg. + ….digits)
2 Digit
4 Digit
8 Digit
4 Digit
3 Digit
10 Digit
10 Digit
10 Digit
1 х intrinsic error
± 1 digits
± 2 digits
1) With temperature: Error data apply per 10 K change in temperature.
For Aac/Adc error data apply per K change in temperature.
With frequency: Error data apply to a display from 300 digits onwards.
2) With zero adjustment.
3) With unknown waveform (crest factor CF > 2), measure with manual range selection
4) With the exception of sinusoidal waveform.
5) After the “
” symbol is displayed
Applicable Standards
EMC
Immunity
IEC 61326: Class B
IEC 61000-4-2
8 KV atmosphere discharge,
4 KV contact discharge
IEC 61000-4-3 : 3 V/m
Safety
IEC 61010-1-2010
IP for water & dust
Pollution degree
Installation category
High Voltage Test
IEC60529
2
IV
6.7 kV AC, 50Hz for 1 minute
between housing and input.
3.7 kV AC, 50Hz for 1 minute
between housing with jaws and
input.
Weight
0.6 Kg
Warranty
1 years
RISHABH INSTRUMENTS PVT.LTD.
F-31, MIDC, Satpur, Nashik-422 007,India.
Tel.: +91 253 2202028, 2202202 Fax : +91 253 2351064
E-mail : India :- marketing@rishabh.co.in
International :- exp.marketing@rishabh.co.in
www.rishabh.co.in
www.rishabh.co.in
Page 5 of 5
Version: E 08/04/2014
1000A/ 400A AC-DC
PDF created with FinePrint pdfFactory trial version www.softwarelabs.com
PDF created with FinePrint pdfFactory trial version www.softwarelabs.com
1
PDF created with FinePrint pdfFactory trial version www.softwarelabs.com
(1)
Liquid crystal display
(6)
Function selector switch.
(2)
Push button for data hold
and MIN/MAX storage functions
(7)
(8)
Terminal sockets
Rotary mechanism for clamp jaws.
(3)
Push button for Down and Inrush
(9)
Safe trigger mechanism.
(4)
Push button for Up and Relative
Function
Multifunction push button
(10)
Limit of safe access for hand held.
(5)
Contents
Page
1.
2.
Introduction , Application and Features ...................................
.......................................3
Safety features and safety precautions............... ............................3
3.
Switching the Power Clamp meter "ON".......................
..............................5
4
5.
Liquid crystal display and backlit..............................
.........................................6
Advanced Data “HOLD” Facility......................................................7
6.
Minimum value and maximum value"MIN/MAX storage facility. ...........8
7.
Relative Function ............................... .............................................9
8.
Voltage measurement ....................................................................10
8.1
8.2
8.3
8.4
8.5
8.6
8.7
9.
THD .................... ....................................................................10
DF ...................... ....................................................................10
CF ...................... ....................................................................11
Peak Min/ Peak Max ...............................................................11
Frequency........... ....................................................................11
Individual Harmonic measurement.........................................11
LPF.......................... ...............................................................11
Current measurement.....................................................................13
9.1 THD .................... ....................................................................14
9.2 DF ...................... ....................................................................15
9.3 CF ...................... ....................................................................15
9.4 Peak Min/ Peak Max ...............................................................15
9.5 Frequency........... ....................................................................15
9.6 Individual Harmonic measurement .........................................15
9.7 LPF.......................... ...............................................................15
9.8 Inrush Current measurement .................................................17
9.9 Ampere hour measurement..... ...............................................18
10. 1Ph Power measurement .......................... ...................................19
10.1
10.2
10.3
10.4
10.5
kVA,kW ,kVAr .......................................................................19
PF & Ф.............. ....................................................................19
HP ..........................................................................................19
DC Power ............................................................................19
kWh measurement ...............................................................22
11. 3-Phase 4 were Power measurement ........... ................................23
11.1 3 Ph 4W unbalance Load Power ..........................................23
11.2 3 Ph 4W balance. Load Power...............................................25
12. 3 Phase 3 Wire Power measurement ........... ................................27
12.1 3 Ph 3W unbalance Load Power..........................................27
12.2 3 Ph 3W balance. Load Power..............................................29
13. NCV Detection..................... ............................
...................................31
14. Resistance, continuity & Diode measurement..................................31
15. Empty Positions................................................................................32
16. Specifications................................................... ................................33
17. Maintainance .................................. .............................................. 37
17.1 Battery ........................................ .....................................37
17.2 Periodic Check-up..............................
.....................................37
18. Servicing ....................................... ...............................................37
2
PDF created with FinePrint pdfFactory trial version www.softwarelabs.com
1. Introduction , Application and Features:
Power Clamp. meter is a Portable Digital multi functional measuring instrument
designed for Measuring selected power network parameters, AC/DC Voltage,
AC/DC current, Resistance, Continuity, Diode and Frequency. Apart From
basic quantities meter enables the measurement of additional quantities
calculated from values of voltage and current.
Power clamp meter is innovatively designed for increasing safety and comfort of
user. Rotating clamp jaws facilitate the measurement at physically awkward
position, vertical bus bars , conductors placed at position difficult to access.
Clamp jaws can be opened or closed with the triggered placed at bottom side
placed away from the jaws . This allows the user to place his/her hand at safer
distance from live conductor.
Location and. design of trigger eliminates fatigues caused by single finger
operation. It allows spreading the force required to open the jaws over more
than one finger to ensure comfortable operation.
Power Clamp meter have following features
2. Safety features and safety precautions
Power Clamp meter provides a very high degree of safety.
The digital power clamp meter 400A/1000A manufactured and tested
in compliance with the safety standard IEC 61010-1:2010.
In case of incorrect use or careless handing,the safety of both user and power
clamp meter is not assured.
For proper use and safe handling,it is absolutely necessary to read and
understand the operating instructions before using the clamp meter.
Please note the following safety precautions:
Operators must use individual protective equipment if
hazardous live parts of installation could be accessible .
Keep hands/fingers behind the edge that separates rotating
jaws with hand held part(10). This is the limits of the hand
held part during measurement .
The clamp meter must be operated only by persons who understand
the danger of shock hazards and are aware of the necessary safety
precautions.Shock hazards exist wherever voltages of more than
30V(TRMS) are present.
Do not work alone in shock hazardous environment while carrying
out measurement.
The maximum permissible voltage between any of the terminal
sockets (7) and ground is 1000V.
3
PDF created with FinePrint pdfFactory trial version www.softwarelabs.com
Take into account that unexpected voltages can occur on devic e
under test(e.g.defective instrument).For example, capacitors may
be charged to a dangerously high voltage.
Verify that the test leads are in good condition, e.g.no crac ked
insulation, no open circuits in the leads or connectors.
This clamp meter must not be used for measurements on circuit s with
corona discharge (high voltage).
Be particularly careful when measuring on HF circuits.Dangerous
composite voltages may exist there.
Measurements under moist environmental conditions are not
permitted.
Do not overload the measuring ranges beyond their allowable
capacities. Limit values are given in specifications. Ref. Chapter 16.
Please verify the performance of clamp meter after repairing
of the meter, before using it for actual measurement.
Do not use the clamp meter if obvious wear in jaw opening
is visible.
Protection provided by the digital clamp meter may be
impaired if the clamp meter is not used in a manner
specified in this user manual.
Meaning of categories and their significance as per IEC61010-1
CAT I: Measurements in electrical circuits which are not directly
connected to the mains: for example electrical systems in motor
vehicles and aircraft, batteries etc.
CAT II: Measurements in electrical circuits which are electrically
connected to the low-voltage mains:with plugs, e.g. at home, in the
office or laboratory etc.
CAT III: Measurements in building installations, stationary power
consumers, distributor terminals, devices connected permanently
to the distributor.
CAT IV: Measurements at power sources for low-voltage installations,
meters, mains terminals, primary over voltage protection devices.
4
PDF created with FinePrint pdfFactory trial version www.softwarelabs.com
Meaning of the symbols on the device
Warning of a danger point
(Attention, refer to the user
manual)
Earth (ground)terminal.
Double or reinforced insulation
Instrument for over voltage
category III OR IV
CAT III/IV
Repair, replacement of parts:
When opening the meter, live parts may be exposed. Therefore , the
meter must be disconnected from the measuring circuit prior to opening
its case for repair or replacement of parts. If repair cannot be avoided
unless the meter is opened and live, this work must only be performed
by a qualified person who understands the danger involved.
Faults and abnormal stress:
When it is realized that the safe operation is no longer possible, take the
meter out of service and secure it against accidental use.
Safe operation may not be possible,
when the meter shows obvious signs of damage,
when the meter no longer functions correctly,
after prolonged storage under adverse conditions,
due to severe stress during transportation.
3. Switching the Power Clamp meter "ON"
Battery
We have already fitted your meter with a 9V flat cell battery according to
IEC 6F22 or IEC 6R61. It is ready for operation. Before you use the meter
for the first time or after storage, refer to section “ 17.1 Maintenance battery”
Switching the meter "ON"
Rotate the function selector switch or knob from “OFF" position to another.
Switch-"ON" is acknowledged by a sound signal.
Note:
Electric discharges and high-frequency influence may cause incorrect
information to be displayed and block the measuring process .Reset the
meter by switching it OFF and ON again otherwise, check the battery
connections.
5
PDF created with FinePrint pdfFactory trial version www.softwarelabs.com
Automatic TURN-OFF
The meter turns off automatically, when neither a push button nor the
function selector switch is operated for about 10 minutes.
How to prevent automatic TURN-OFF
In order to prevent automatic "TURN OFF" select "CONTINUOUSLY
ON" mode. For this, press “Down and Inrush” push button and the “Min/Max
and Hold " pushbutton together. The function " CONTINUOUSLY ON"
is shown on the LCD by the symbol ON
Turning the clamp meter OFF
Rotate the function selector switch to “OFF” position
4. LCD & Backlit
INRUSH REL LPF unbal 3W 4W
%
ON
hp
kVA h
kW h
AC
DC
Σ L123
Ф
MIN MAX THD F
Peak PF CF
kV r
% Hz
h :mm:s
LCD of Clamp meter is as shown above. The digital display shows measured
value with sign. Description of remaining symbol is as follow.
V
A
W
VA
Volts
Amperes
Active Power
Apparent Power
Ah
Reactive Power
Ampere Hour
Wh
Watt Hour
VAr
INRUSH
AC Inrush Current Measurement
REL
Relative Measurement
LPF
Low Pass Filter
unbal
Un Balance
bal
Balance
4W
Hold function
3W
Three Wire
4W
Four Wire
Continuous ON
Continuity Mode
AC
Alternating current
DC
Direct current
hp
Horse Power
Resistance Measurement
6
PDF created with FinePrint pdfFactory trial version www.softwarelabs.com
L1
L2
L3
Σ L123
MIN
MAX
Phase 1 Readings
Phase 2 Readings
Phase 3 Readings
Three phases system parameters
Minimum Value
Maximum Value
Low Battery
Diode Measurement
Inductive
THD
DF
PF
CF
Ф
Peak
Hz
k
h :mm:s
Capacitive
Total Harmonic Distortion
Distortion Factor
Power Factor
Crest Factor
Phase Angle
Peak Value of Current or voltage
Hertz
Multiplied by 1000(Kilo)
Hours:Minutes:Seconds
Motor application
Backlit
The instrument is provided with Back-lit for taking
measurements in poor lighting conditions/dark areas.
Switching the Backlit ON
by pressing "MULTIFUNCTION (YELLOW)" and "UP / REL" keys
simultaneously the Backlit can be switched ON.
Switching the Backlit OFF
by pressing "MULTIFUNCTION (YELLOW)" and "UP / REL" keys
simultaneously the Backlit can be switched OFF.
5. Advanced Data “HOLD” facility
The HOLD function allows to automatically hold the measured values.
The meter holds the measured value on the digital display with a sound
signal and displays “ HOLD ” on LCD display. The probes or clamp
can now be removed from the measuring point and the measured value
on the digital display can be read.
Why We Call this as ADVANCED HOLD Function ?
armonics Up to 49.
7
PDF created with FinePrint pdfFactory trial version www.softwarelabs.com
Now if we press hold key on any one of the parameter screen then not
only that parameter but all other measured parameter gets hold. including
all 49 harmonics.
So Advanced Hold feature benefits user to make data analysis of different
parameters even after removal of input .
To activate “HOLD ” function, momentarily press (short press)
the “HOLD / MIN MAX” key.
The data HOLD function is switched OFF, when
The “HOLD” pushbutton is pressed for approx. 2sec (long press).
This is acknowledged by sound signals.
The function selector switch is operated.
The Clamp meter is turned OFF and ON again.
The “HOLD / MIN MAX” key will not be enabled in 3 phase power
measurement unless and until measurement of power is done.
The “HOLD / MIN MAX” key will be disabled in NCV, kWh, Ahr functions.
Note : HOLD function not applicable for 3P4W, 3P3W unbalanced system, Diode
and Continuity .
Minimum value and Maximum value "MIN / MAX"
storage facility.
With the MIN/MAX function, we can hold the minimum and the maximum
measured value which was applied to the input of the clamp meter after
activating MIN/MAX function.The most important application is the
determination of the minimum and the maximum value for long-term
monitoring of measured quantities.
To activate “MIN MAX ” function, press the “HOLD / MIN MAX” key twice.
The data “MIN MAX ” function is switched OFF, when
The “HOLD / MIN MAX” pushbutton is pressed for approx.2s.
(long press) This is acknowledged by sound signals.
The function selector switch is operated
The Clamp meter is turned OFF and ON again.
“MIN MAX” is not applicable for Harmonics and the parameters on sub display
“MIN MAX” is also not applicable for 3P3W and 3P4W unbalanced load.
Figure 6.1 will show procedure for entering in hold and min max functions
8
PDF created with FinePrint pdfFactory trial version www.softwarelabs.com
Fig 6.1: Screens for HOLD/MIN/MAX
ON
DC
A
Short pressing of
“HOLD / MIN MAX” key.
ON
HOLD
Function
DC
A
Long pressing of
“HOLD / MIN MAX” key.
Short pressing of
“HOLD / MIN MAX” key.
ON
MIN/MAX
Function
DC
A
MIN
A
Short pressing of
“HOLD / MIN MAX” key.
ON
MIN/MAX
Function
DC
A
MAX
Short pressing of
“HOLD / MIN MAX” key.
A
Long press
Hold /Min /Max key
7. Relative Function(REL)
With the REL function, we can see the measured value with respect to
some value.
REL function is applicable for Voltage, Current and Resistance functions only.
To activate “Relative ” function, press the “UP/ REL” key .
The “REL ” function is switched OFF, when
The “UP / REL” pushbutton is pressed (short press)
This is acknowledged by sound signals.
The function selector switch is operated
The Clamp meter is turned OFF and ON again.
9
PDF created with FinePrint pdfFactory trial version www.softwarelabs.com
8. Voltage measurement
According to the voltage to be measured, set the function selector
switch to V
Connect the test leads as shown in fig 8.1. The “ socket
should be connected to the lowest potential ground available.
Select the appropriate working mode i.e. AC or DC or ACDC by long
pressing Yellow key (Function key).
REL
INRUSH
HOLD
MIN/MAX
100 0V
CAT III
600 V
CAT IV
Fig 8.1 Voltage measurement on electrical systems up to 1000V
For AC voltage mode following parameters measured.
8.1 THD (Total Harmonic Distortion)
th
Power Clamp meter can measure THD up to 49 Harmonic.
By default meter is present on THD measurement screen.
49
th
( n order harmonic voltage RMS value)
THD =
2
n=2
2
X 100 %
( Fundamental wave voltage RMS value)
8.2 DF ( Distortion Factor)
th
Power Clamp Power can measure DF up to 49 Harmonic.
49
th
( n order harmonic voltage RMS value)
DF =
n=2
2
2
X 100 %
( Voltage RMS value)
10
PDF created with FinePrint pdfFactory trial version www.softwarelabs.com
8.3 CF ( Crest Factor)
Power Clamp meter can measure CF. CF is the ratio between the value
the peak voltage and corresponding RMS voltage.
CF =
Vm (Peak Voltage Value)
VRMS (RMS Voltage Value)
8.4 Peak Max / Peak Min
Peak Max /Min is the Positive / Negative peak value of measuring
waveform. It updates continuously as per measuring waveforms peak.
8.5 Frequency
Power Clamp meter measures frequency from 45 to 65 Hz.
8.6 Individual Harmonic measurement
th
Power Clamp meter can measure individual harmonic voltage up to 49
Harmonic. Measured harmonic can also be seen in percent (%) w.r.t.
fundamental voltage.
8.7 LPF measurement mode
Power Clamp meter have LPF mode for voltage measurement.
In LPF mode meter measures voltage bellow cut off frequency. The
cutoff frequency of clamp meter for LPF mode is 400 Hz. This means
that meter will measure voltage having frequency bellow 400 Hz.
Note : 1) If display shows OL then it indicates Voltage Overload
i.e > 1020 V .
To enter in LPF mode Simultaneously press “ yellow function key”
and “Hold/Min/Max” keys.
Fig 8.2 Low Pass Filter mode
Note : 1) “----” is displayed for THD, DF, CF, Freq and Harmonics if voltage level
is below measurement band or applied voltage is OL value i.e. 1020 V
2) “----” is displayed for THD, DF and Harmonics if signal frequency is
out of measuring band i.e. Freq Except 45Hz....65Hz
3) Meter will display 0 V if V <0.5V and in LPF mode meter will display
0 V if V < 1V
11
PDF created with FinePrint pdfFactory trial version www.softwarelabs.com
Voltage measurement
Knob Position
Yellow key
Long Press
AC Voltage / THD
DC Voltage
AC
V
Yellow key
Long Press
TH D
V
DC
%
Yellow key
Yellow key
Short Press
AC
AC
V
V
DC
DF
Distortion
Factor
%
Yellow key
Long Press
Short Press
AC DC Voltage
AC
V
Crest Factor
CF
Yellow key
Short Press
AC
V
MIN MAX
Peak
Yellow key
Max Peak
V
Short Press
AC
Min Peak
V
MIN MAX
Peak
Yellow key
V
Short Press
AC
V
Frequency
Hz
Yellow key
Individual Harmonics Amplitude
Short Press
REL key
AC
AC
V
V
REL
Short Press
Yellow key
Short Press
Individual Harmonics in %
REL key
%
%
AC
REL
AC
Short Press
Yellow key
Short Press
Fig 8.3 Voltage measurement screens
12
PDF created with FinePrint pdfFactory trial version www.softwarelabs.com
9. Current measurement
Set the function selector switch to A .
Connect the Clamp as shown in figure 9.1 for current measurement.
To measure current through cable push the trigger (present at back )
to open the jaws and clamp the jaws around the cable as shown in fig 9.1
Figure 9.1
Select the appropriate working mode i.e. AC or DC or ACDC by long
pressing yellow key (Function key).
Unique design for safety and comfort
Rotary mechanism for clamp jaws:
In conventional clamp meters display, keys and clamp jaws are in the same plane.
When current measurement is to be done on vertical bus bars, over head cables,
cables in congested places user connect the clamp meter but the keys and display
may not be visible, hence not able to take the readings or operate the keys.
To over come the above mentioned problem Clamp meter 400A/1000A has a
unique feature called“Rotary mechanism for clamp jaws”. In this, the clamp jaws
are rotating. Hence it is possible to align the clamp jaws as the orientation of
bus bar/conductor while keeping Display and keys facing the user, so that user can
take the readings and operate the keys.
Rotary clamp jaws can be rotated at different angles with the step of 30º, maximum
up to 90º in both clock-wise as well as anti-clock wise direction as shown in figure 9.2
Note :- For better accuracy of current, position conductor at the marking on
the jaw (center of the jaw)
13
PDF created with FinePrint pdfFactory trial version www.softwarelabs.com
Figure 9.2
For DC and ACDC current measurement mode , Auto Zero adjustment
functionality is available for DC current up to + / - 5 A.
To make zero adjustment , long Press HOLD key on meter.
The meter acknowledge zero setting by a sound signal,
For AC Current mode following parameters are measured.
9.1 THD (Total Harmonic Distortion)
th
Power Clamp meter can measure THD up to 49 Harmonic.
By default meter is present on THD measurement screen.
49
th
( n order harmonic Current RMS value)
THD =
2
n=2
( Fundamental wave Current RMS value)
2
X 100 %
14
PDF created with FinePrint pdfFactory trial version www.softwarelabs.com
9.2 DF ( Distortion Factor)
Power Clamp Meter can measure DF up to 49 th Harmonic.
49
2
th
( n order harmonic Current RMS value)
DF =
n=2
( Current RMS value)
2
X 100 %
9.3 CF ( Crest Factor)
Power Clamp meter can measure CF. CF is the ratio between the value
the peak current and corresponding RMS current.
CF =
Im (Peak Current Value)
IRMS (RMS Current Value)
9.4 Peak Max / Peak Min
Peak Max /Min is the Positive / Negative peak value of measuring
waveform. It updates continuously as per measuring waveforms peak.
9.5 Frequency
Power Clamp meter measures frequency from 45 to 65 Hz.
9.6 Individual Harmonic measurement
th
Power Clamp meter can measure individual harmonic current up to 49
Harmonic. Measured harmonic can also be seen in percent (%) w.r.t.
fundamental current.
9.7 LPF measurement mode
Power Clamp meter have LPF mode for current measurement.
In LPF mode meter measures current bellow cutoff frequency. The
cutoff frequency of clamp meter for LPF mode is 400 Hz. This means
that meter will measure current having frequency bellow 400 Hz.
Note : 1) If display shows OL then it indicates Current Overload.
Overload. i.e. > 1020A
To enter in LPF mode Simultaneously press “ yellow function key”
and “Hold/Min/Max” keys.
Figure 9.3
Note : 1) “----” is displayed for THD, DF, CF, Freq and Harmonics if current level
is below measurement band or applied current is OL Value,
1020 A for Power clamp 1000A and 415 A for Power Clamp 400A
2) “----” is displayed for THD, DF and Harmonics if signal frequency is
out of measuring band i.e. Freq Except 45Hz....65Hz
3) Meter will display 0 A if I <0.5 A and in LPF mode meter will display
0 A if I < 1A
4) For better accuracy of harmonic do not apply voltage at input terminal
15
PDF created with FinePrint pdfFactory trial version www.softwarelabs.com
Current measurement
Knob Position
AC Current / THD
Yellow key
Long Press
DC Current
AC
A
THD
%
Yellow key
A
DC
Yellow key
Long Press
Short Press
Yellow key
Long Press
AC
A
%
Yellow key
AC
Distortion
Factor
DF
A
DC
Short Press
AC DC Current
AC
A
Crest Factor
CF
Yellow key
Short Press
ACAC
A
MIN MAX
Peak
Yellow key
Short Press
AC
VA
MIN MAX
Peak
Yellow key
Max Peak
A
Min Peak
VA
Short Press
AC
A
Frequency
Hz
Yellow key
Individual Harmonics Amplitude
Short Press
REL key
AC
A
AC
A
REL
Short Press
Yellow key
Short Press
Individual Harmonics in %
REL key
%
AC
%
REL
AC
Short Press
Yellow key
Short Press
Fig 9.4 Current measurement screens
16
PDF created with FinePrint pdfFactory trial version www.softwarelabs.com
9.8 Inrush current measurement
Power Clamp meter can measure AC Inrush current . This function allows
us to easily measure inrush current that occurs when starting a motors
Just clamp the clamp meter to line of motor and set meter in inrush
mode before starting of motor so that after starting of motor meter will
automatically measure and hold inrush Current value that occurred in
100 ms (measurement period ). Refer fig 9.5
Step1: Keep Knob Position on A AC
Step2: Clamp the meter around live conductor of motor
Step3: Press INRUSH key to set the meter in inrush mode
Now meter will wait for trigger to occur and display
would look like
INRUSH
AC
A
Step4: Start the motor. Clamp meter will be triggered by inrush
current >5 A. Inrush current for 100 ms is measured.
INRUSH
AC
A
M
Note: In harmonics screen INRUSH key is used to scroll harmonics
Inrush mode can not be activated in harmonics screen
Inrush Current
Motor
A
100A
10A
RE L
I NRUS H
HOLD
MIN/MA X
0.0A
100 0V
CAT III
600 V
CAT IV
100 ms
Motor ON
Trigger
Point
Fig 9.5
Time
17
PDF created with FinePrint pdfFactory trial version www.softwarelabs.com
9.9 Ampere hour measurement
Power Clamp Meter can measure Ahr for AC and DC current .
Clamp meter can accumulate Ahr for max time of 23:59 Hrs .
Maximum value of accumulated Ahr can be 999.9 Ahr . After
999.9 Ahr meter will display OL.
To measure Ahr go through following steps
Step1: Keep Knob position on AAC
Step2: For Ahr in AC current Press REL and INRUSH key simultaneously
clamp meter will start measuring AC Ampere hours.
AC
Ah
mm:s
Note : For Ahr in DC current, long press (>2 Secs) yellow function key
clamp meter enters In DC current mode. Now Press REL and INRUSH
key simultaneously, clamp meter will start measuring DC Ampere hours.
Ah
mm:s
18
PDF created with FinePrint pdfFactory trial version www.softwarelabs.com
10. Single Phase Power Measurement
According to the single phase Power to be measured, set the function
selector switch to
position.
Connect the voltage test leads as shown in fig 10.2.
Connect the Clamp as shown in fig. 10.2.
Select the appropriate working mode i.e. AC or DC power by long
pressing yellow key (function key).
For AC Power mode following parameters are measured.
10.1 KVA , KW , KVAr (Apparent Power, Active Power , Reactive Power)
Power Clamp Meter can measure KVA, KW and KVAr for single phase.
KVA = V * I
KW = V * I * Cos(ø)
KVAr = V * I * Sin(ø)
10.2 PF and ø ( Power Factor and phase angle)
Power Clamp Meter can measure PF and ø .
PF = KW / KVA
ø = COS
-1
(PF)
10.3 hp ( horse power)
Power Clamp meter can measure Horse power (hp).
hp = KW 0.7456
*
For DC Power mode following parameters are measured.
10.4 DC Power (Active Power )
Power Clamp Meter can measure DC KW for single phase.
KW = V * I
For DC Power measurement mode , Auto Zero adjustment
functionality is available. To make zero adjustment , long Press hold
key on meter. The meter acknowledge zero setting by a sound signal,
By Auto Zero Function maximum + / - 5 A DC can be nullified.
Note :
1) If display shows OL.U then it indicates Voltage Overload.(>1020V)
2) If display shows OL.I then it indicates Current Overload.(>1020A)
Overload.
3) If display shows OL then it indicates both current and voltage
Overload.
4) If display shows +ve active power then the power flows from
power source to load.
5) If display shows -ve active power then the power flows from
load to power source.
6) If display shows +ve power factor then the phase of current
is lagging behind the voltage (Inductive load).
7) If display shows -ve power factor then the phase of current
is leading before the voltage (Capacitive load).
8) OL.U, OL.I and OL logic will remain same for 3P3W and 3P4W
balanced system.
19
PDF created with FinePrint pdfFactory trial version www.softwarelabs.com
POWER Display Ranging For 1P2W, 3P3W and 3P4W
For measurement of power total 4 Ranges are available 9.999 kVA,
99.99 kVA, 999.9kVA and 9999 kVA same applicable for hp, kVAr and kW.
Meter will select range of all powers i.e. kW, hp and kVAr based on
kVA reading.
For example :V = 230V, I = 102A and Angle = 0 then kVA = 230*102 = 23.46 kVA
kW = 230*102 * COS 0 = 23.46 kW and kVAr = 230*102*SIN 0 = 0 kVAr
So Meter will display power like 23.46 kW, 00.00 kVAr, 23.46 kVA
With same V & I now if Angle = 70 then,
kW = 230*102 * COS70 = 8.023kW and kVAr = 230*102*SIN70=22.04 kVAr
Now Meter will display kW like 8.02 and not 8.023 as kVA is in 99.99 range.
hp Range is also based on kVA range except some cases where it is
displayed one range higher than kVA.
For example :kVA = 82.53 so at maximum value of kW = kVA at PF = 1,
hp = kW / 0.7456 = 110.0 so in this case hp is display in 999.9 range.
9.999 kVA range is only applicable for V <100 V and I < 100 A.
If V > 500 V and I < 25 A then, In this kind of situation a very small amount
of current can cause a large amount of fluctuation in power reading.
To avoid this irritation of fast display fluctuation because of small current
change meter will display power in 999.9 kVA range.
All this ranging related logic is equally applicable for DC power ( In DC
POWER ranging is based on kW as there is no kVA), 3P4W balance as
well as un balance and 3P3W balance as well as un balance system type.
20
PDF created with FinePrint pdfFactory trial version www.softwarelabs.com
Knob Position
Yellow key
AC
AC
kVA
DC
kW
Long Press
Yellow key
A
MIN MAX
PF
Peak
Short Press
Yellow key
hp
AC
Short Press
AC
V
DC
kVA
A
MIN MAX
Peak
Yellow key
Short Press
Yellow key
Short Press
AC
kVA
kVA
Yellow key
Short Press
AC
kW
Ф
Yellow key
Short Press
ACAC
V
A
MIN MAX
Peak
Yellow key
Short Press
Fig 10.1 Power measurement screens
L
N
RE L
I NRUS H
HOLD
MIN/MA X
100 0V
CAT III
600 V
CAT IV
Fig 10.2 Connection diagram for 1P2W Power measurement
21
PDF created with FinePrint pdfFactory trial version www.softwarelabs.com
10.5 KWH Measurement in 1-phase 2-wire system
Power clamp meter accumulates AC or DC power to measure the energy
up to 9999kWh. Beyond this limit it shows OL. Clamp meter can accumulates
energy max up to 23.59 hrs. To measure Energy in AC power
system follow the below steps
Step1: keep knob position as shown below
Step2: For kWh in AC Power Press REL and INRUSH key simultaneously
clamp meter will start measuring kWh
AC
kW h
h:mm
Note: For kWh in DC power system, long press (>2 Secs)
yellow
function key,clamp meter enters In DC power mode.
Now Press REL and INRUSH key simultaneously clamp meter will
start measuring DC energy.
DC
kW h
h :mm
1) kWh Measurement is also possible in 3P3W and 3P4W balanced load
by using simultaneous pressing REL and INRUSH key.
2)To reset energy measurement exit from energy mode and re enter into
energy mode or restart the meter and enter into energy mode again.
3) If display shows -ve Energy then the power flows from load to power
source.
22
PDF created with FinePrint pdfFactory trial version www.softwarelabs.com
11. 3-Phase 4- wire Power Measurement
According to the 3 phase 4 wire Power to be measured, set the function
selector switch to
position.
Using power Clamp meter, we can measure power of both balance
and unbalance systems. After positioning of selector switch to
position, meter will toggle between unbalance and balance system options
using Up and Down keys. After pressing Yellow key, we can select any
one type of system type and proceed for measurement.
Note :
1) System Current and System Voltage readings will not be shown for
unbalanced system(3P3W as well as 3P4W).
2) If display shows OL.U then it indicates Voltage Overload. (>1020 V)
3) If display shows OL.I then it indicates Current Overload. (>1020 A)
4) If display shows OL then it indicates both current and voltage
Overload.
5) If display shows +ve active power then the power flows from
power source to load.
6) If display shows -ve active power then the power flows from
load to power source.
7) If display shows +ve power factor then the phase of current
is lagging behind the voltage (Inductive load).
8) If display shows -ve power factor then the phase of current
is leading before the voltage (Capacitive load).
11.1 3-Phase 4- wire unbalance Load Power Measurement
Note that using single Power Clamp meter we can measure 3Ph
power. Steps for measurement of 3 ph power are as follows.
1) Configure Clamp meter to unbalance power measurement mode
2) Meter will show L1-n and I -1 on display . Connect meter to phase L1
as shown in fig 11.1.2 (step 1).
3) After connection press function (Yellow) key, meter will display
measured apparent Power and PF for phase L1 on LCD.
4) Wait for 5-10 sec so that measured value gets stable and then
press function(Yellow) key to store readings of phase L1. After
pressing function key meter will show L2-n and I-2 on display,
which indicates that connect meter to phase L2.
5) Now disconnect meter from phase L1 and connect to phase L2.
Repeat Steps 2, 3 and 4 for phase L2 and L3.
6) After storing readings of phase L3 , meter will show System apparent
apparent power and PF . At this point all measurements are finished.
Now we can remove all input connections of meter and analyze all
measured data.
7) Measure data includes Active , Reactive, Apparent Power, Power
Factor, Phase angle, horse power , voltage, Current. All these
parameters are measured for individual phase and for system also.
After step 6 if we press function (Yellow) key then we can scroll
through all above measured parameter one by one . And if we press
Up or Down key we can see measured data for individual phases
(L1 or L2 or L3 ) and for system (Σ L123). Refer Fig 11.1.1
System Power (Σ L123) = L1 Power + L2 Power + L3 Power
23
PDF created with FinePrint pdfFactory trial version www.softwarelabs.com
System power
unbal
4W
unbal
AC
4W
AC
kVA
Σ L12 3
PF
Yellow key
Yellow key
Short Press
unbal
4W
Short Press
unbal
AC
4W
hp
AC
V
Σ L12 3
kVA
A
Yellow key
Yellow key
Short Press
unbal
4W
Short Press
unbal
AC
4W
AC
Σ L12 3
kVA
L1
kVA
kVA
PF
Yellow key
Yellow key
Short Press
unbal
Short Press
unbal
4W
4W
AC
AC
V
Σ L12 3
Yellow key
kW
Ф
A
Short Press
unbal
Yellow key
Short Press
4W
UP key
Short Press
AC
kVA
L 2
unbal
12
L1
Yellow key
4W
ACAC
PF
Short Press
kW
MIN MAX
Peak
unbal
Ф
4W
AC
V
Note: After Completion of
measurement of all three phases,
If we press function key(Yellow)
then screen will scroll through
measurement parameters. and if
we press Up or Down key then
screen will scroll through phase
L1,L2,L3 and system ΣL123.
A
Yellow key
Short Press
unbal
4W
AC
kVA
Note: If we long press function
key (Yellow) then meter comes on
unbalance system selection screen
again. that is it will reset all
measured data and start power
measurement again. applicable for
both balance and unbalance
systems
L 3
PF
Yellow key
Short Press
Fig 11.1.1 Power measurement 3-phase 4-wire Unbalanced load
Note :- In 3-phase 4-wire unbalanced system power measurement
in every screen wait for the stable reading to store the reading
without stabilization of power reading, if readings are stored by
pressing yellow key then this can cause a false reading.
If any phase voltage or any line current value is OL value then
system power will be displayed as OL for 3P4W unbalanced system.
24
PDF created with FinePrint pdfFactory trial version www.softwarelabs.com
Keep Knob Position
L1
N
L2
L3
Fig 11.1.2 Connection diagram for 3ph4w unbalance load
11.2 3-Phase 4- wire balance Load Power Measurement
In 3Ph balance load power measurement , we need to measure power
1 phase only. Steps for 1 ph power measurement are as follows.
1) Configure Clamp meter to balance power measurement mode
2) Meter will show L1-n and I -1 on display.Connect meter to phase 1
as shown in fig 11.2.1.
3) After connection press function (Yellow) key, meter will display
measured apparent Power and PF on LCD.
4) In balance load, measured parameters updates continuously. Hold
function can be used to hold all measured values. After hold we can
remove all input connections of meter and analyze all data
5) Measure data includes Active , Reactive, Apparent Power, Power
Factor, Phase angle, horse power , voltage, Current. All these
parameters are measured for system ( Σ L123 ) only.
If we press function (Yellow) key then we can scroll through all
above measured parameter one by one . Refer Fig 11.2.2.
System Power (Σ L123) = L1 Power * 3
25
PDF created with FinePrint pdfFactory trial version www.softwarelabs.com
L3
N
L2
L1
Fig 11.2.1 Connection diagram for 3ph4w balance load
unbal
4W
bal
Up key
AC
R EL
4W
AC
Short Press
Yellow key
Short Press
bal
Yellow key
Long Press
4W
AC
V
A
Yellow key
Short Press
bal
4W
AC
kVA
Σ L123
PF
Yellow key
Short Press
bal
Yellow key
4W
ACAC
V
Σ L123
Short Press
bal
4W
hp
AC
Σ L123
A
Yellow key
kVA
Yellow key
Short Press
bal
Short Press
bal
4W
4W
AC
AC
Σ L123
kW
Σ L12 3
Ф
kVA
kVA
Yellow key
Short Press
Fig 11.2.2 Power measurement 3-phase 4-wire balanced load
26
PDF created with FinePrint pdfFactory trial version www.softwarelabs.com
12. 3-Phase 3- wire Power Measurement
According to the 3 phase 3 wire Power to be measured, set the function
selector switch to
position.
Using Power Clamp Meter, we can measure power of both balance
and unbalance systems. After positioning of selector switch to
position, meter will toggle between unbalance and balance system options
using Up and Down keys. After pressing Yellow key, we can select any
one type of system type and proceed for measurement.
Note :
1) System Current and System Voltage readings will not be shown for
unbalanced system(3P3W as well as 3P4W).
2) If display shows OL.U then it indicates Voltage Overload. (>1020 V)
3) If display shows OL.I then it indicates Current Overload. (>1020 A)
4) If display shows OL then it indicates both current and voltage
Overload.
5) If display shows +ve active power then the power flows from
power source to load.
6) If display shows -ve active power then the power flows from
load to power source.
7) If display shows +ve power factor then the phase of current
is lagging behind the voltage (Inductive load).
8) If display shows -ve power factor then the phase of current
is leading before the voltage (Capacitive load).
12.1 3-Phase 3- wire unbalance Load Power Measurement
Note that using single Power Clamp Meter we can measure 3Ph
power. Steps for measurement of 3 ph power are as follows.
1) Configure Clamp meter to unbalance power measurement mode
2) Meter will show L1-2 and I -1 on display. Now connect meter to
phase 1 as shown in fig 12.1.2.
3) After connection press function (Yellow) key, meter will display
measured apparent Power and PF of phase 1-2 on LCD.
4) Wait for 5-10 sec so that measured value gets stable and then
press function(Yellow) key to store readings of phase 1-2. After
pressing function key meter will show L3-2 and I-3 on display,
which indicates that connect meter to phase 3-2
5) Now disconnect meter from phase 1-2 and connect to phase 3-2.
Repeat Step 2,3 and 4 for phase 3-2.
6) After storing readings of phase 3-2 , meter will show System apparent
power and PF . At this point all measurements are finished.
Now we can remove all input connections of meter and analyze all
measured data.
7) Measure data includes Active , Reactive, Apparent Power, Power
Factor, Phase angle, horse power . All these parameters are measured
for phase 1-2 , phase 3-2 and for system (Σ L123).
After step 6 if we press function (Yellow) key then we can scroll
through all above measured parameter one by one . And if we press
Up or Down key we can see measured data for phase1-2 , phase 3-2
and for system (Σ L123). Refer Fig 12.1.1
System Power (Σ L123) = L1-2 Power + L2-3 Power
27
PDF created with FinePrint pdfFactory trial version www.softwarelabs.com
Fig 12.1.1 Power measurement 3-phase 3-wire Unbalanced load
unbal 3W
unbal 3W
AC
AC
kVA
Σ L123
PF
Yellow key
Short Press
Yellow key
unbal 3W
Short Press
unbal 3W
AC
hp
AC
V
Σ L123
kVA
A
Yellow key
Yellow key
Short Press
Short Press
unbal 3W
unbal3W
AC
AC
kVA
L12
kVA
Σ L123
kVA
PF
Yellow key
Yellow key
Short Press
Short Press
unbal 3W
unbal3W
AC
AC
V
Σ L123
kW
Ф
A
Yellow key
Short Press
Yellow key
Short Press
unbal3W
AC
kVA
Up key
L 23
Short Press
PF
Yellow key
unbal 3W
Short Press
ACAC
L12
kW
MIN MAX
Peak
Ф
Note :- After completion of measurement in individual phases, if we
press YELLOW key then screen will scroll through measurement
parameters and if we press UP and DOWN key then screen will
scroll through phase L1-2, L2-3 and system Σ L123.
If we long press YELLOW key then meter will come to Un-bAl and
bal system type selection screen and this will reset all measured
data to again restart measurement. It is applicable for both system
type.
In 3-phase 3-wire unbalanced system power measurement
in every screen wait for the stable reading to store the reading
without stabilization of power reading, if readings are stored by
pressing yellow key then this can cause a false reading.
If any phase voltages or any line current value is OL value then
system power will be displayed as OL for 3P3W unbalanced system.
28
PDF created with FinePrint pdfFactory trial version www.softwarelabs.com
Keep Knob Position
L1
L3
L2
Fig 12.1.2 Connection diagram for 3ph3w unbalance load
12.2 3-Phase 3- wire balance Load Power Measurement
In 3Ph balance load power measurement , we need to measure power
1-2 phase only. Steps for 1-2 ph power measurement are as follows.
1) Configure Clamp meter to balance power measurement mode
2) Meter will show L1-2 and I -1 on display. Connect meter to phase
1-2 as shown in fig 12.2.1.
3) After connection press function (Yellow) key, meter will display
measured apparent Power and PF on LCD.
4) In balance load, measured parameters updates continuously. Hold
function can be used to hold all measured values. After hold we can
remove all input connections of meter and analyze all data
5) Measure data includes Active , Reactive, Apparent Power, Power
Factor, Phase angle, horse power , voltage, Current. All these
parameters are measured for system (Σ L123 ) only.
If we press function (Yellow) key then we can scroll through all
above measured parameter one by one .Refer Fig 12.2.2
System Power (Σ L123) = VL1-2 * I1 * 1.732
29
PDF created with FinePrint pdfFactory trial version www.softwarelabs.com
L1
L3
L2
Fig 12.2.1 Connection diagram for 3ph3w balance load
unbal 3W
Up key
AC
REL
bal 3W
AC
Short Press
Yellow key
Short Press
bal 3W
Yellow key
AC
Long Press
V
A
Yellow key
Short Press
bal 3W
AC
kVA
Σ L123
PF
Yellow key
Short Press
Yellow key
bal 3W
ACAC
hp
AC
V
Σ L123
Short Press
bal 3W
Σ L123
A
Yellow key
kVA
Short Press
Yellow key
bal 3W
Short Press
bal 3W
AC
AC
Σ L123
kW
Σ L12 3
Ф
kVA
kVA
Yellow key
Short Press
Fig 12.2.2 Power measurement 3-phase 3-wire balanced load
30
PDF created with FinePrint pdfFactory trial version www.softwarelabs.com
13. NCV Detection (Non Contact Voltage Detection)
For NCV detection mode place selector switch to NCV position..
NCV allows detection of AC voltage from >100 V 50/60 Hz.
For NCV detection,touch the left jaw to voltage carrying conductor
Presence of voltage will be indicated by buzzer sound along with
backlit blinking
Note :- Even if voltage presence is not detected do not touch bare wire
or cable.
14. Resistance, Continuity and Diode measurement
For Resistance or continuity or diode measurement place function
selector switch to position .
For Resistance and continuity measurement connect DUT as shown bellow
The clamp meter can measure resistance up to 4000 ohm . For continuity
measurement , meter will generate continuous beep sound if measured
resistance is from 0 to 40 ohm approx.
For Diode measurement connect DUT as shown bellow
The measuring instrument displays forward voltage in volts. As long as
voltage drop does not exceeds the maximum display value of 2.2V ,we can
test several instruments in series.
OL will be displayed on display when diode is connected in reverse bias.
31
PDF created with FinePrint pdfFactory trial version www.softwarelabs.com
15.Empty Positions
Empty positions on dial indicates no function is available on these positions.The
digital display will look like as in figure below.
32
PDF created with FinePrint pdfFactory trial version www.softwarelabs.com
16. Specifications
Reference Conditions
Ambient temperature
+ 23 0C +2K
:
Relative humidity:
45% ... 55 % RH
Frequency of measured
quantity
Power factor
0.5L...1....0.5C
50Hz or 60 Hz
Waveform of the measured
quantity
Sinusoidal
Battery voltage
8V +0.1V
Technical Specification
Measuring
func tion
Measuring
range
Resolution
Intrins ic error of digital display
at referenc e c ondition
V DC
999.9 V
0.1 V
±(0.5% of rdg + 5 dgt)
V~
999.9 V
0.1 V
±(0.75% of rdg+5 dgt)
V ACDC
999.9 V
0.1 V
±(1.25% of rdg+10dgt)
LPF V~
999.9 V
0.1 V
50.....60 Hz ±(0.75% of rdg + 5dgt)
61...400Hz
POWER CLAMP
1000A ADC-AAC
POWER CLAMP
400A ADC - AAC
POWER CLAMP
1000A A ACDC
POWER CLAMP
400A A ACDC
POWER CLAMP
LPF 1000A A AC
POWER CLAMP
LPF 400A A AC
Active
Power2)
Reactive
Power 2)
Apparent
Power 2)
±(5.0% of rdg + 5dgt)
1)
999.9A
0.1 A
±(1.5% of rdg+5 dgt)
99.99 A
0.01 A
400 A
0.1 A
display value ±(1.5% of rdg+0.2A)1)
<1000 add
1)
±(1.5% of rdg+5 dgt)
10 dgt
999.9A
0.1 A
±(3% of rdg+10 dgt)
99.99 A
0.01 A
400 A
0.1 A
display value ±(3% of rdg+0.4A) 1)
<1000 add
1)
±(3% of rdg+10 dgt)
10 dgt
999.9A
0.1 A
50....60 Hz ±(1.5% of rdg + 5dgt)
61...400Hz ±(5.0% of rdg + 5dgt)
99.99 A
0.01 A
50....60 Hz ±(1.5% of rdg + 0.3A)
61...400Hz ±(5.0% of rdg + 5dgt)
400 A
0.1 A
50....60 Hz ±(1.5% of rdg + 5dgt)
61...400Hz ±(5.0% of rdg + 5dgt)
9.999 kW
1W
99.99 kW
10 W
999.9 kW
100 W
9999 kW
1 kW
9.999 kVAr
1 VAr
99.99 kVAr
10 VAr
999.9 kVAr
100 VAr
9999 kVAr
1 kVAr
9.999 kVA
1 VA
99.99 kVA
10 VA
999.9 kVA
100 VA
9999 kVA
1 kVA
9.999 hp
0.001 hp
Horse
99.99 hp
0.01 hp
Power 2)
999.9 hp
0.1 hp
9999 hp
1 hp
1)
±(2% of rdg+5 dgt) 1)
33
PDF created with FinePrint pdfFactory trial version www.softwarelabs.com
Technical Specification
Measuring function
2)
kWh
Ahr
2)
Phase angle
2)
Measuring
r ange
Resolution
9.999 kWh
99.99 kWh
0.001 kWh
0.01kWh
999.9 kWh
9999 kWh
0.1 kWh
1 kWh
999.9 Ahr
0.1 Ahr
Intr ins ic error of digital dis play
at reference condition
±(3% of rdg+5 dgt)
0.0°…360.0°
0.1°
-1…0…1
±(3% of rdg+10 dgt)
±3°
14…49
0.001
0.1V
0.1A
0.1%
THD
0…99.9%
0.1%
±(3% of rdg+20 dgt)
3)
0…99.9%
0.1%
±(3% of rdg+20 dgt)
1.0…2.9
0.1
±(2% of rdg+3 dgt)
3.0…5.0
0.1
±(3% of rdg+5 dgt)
1400 A / 1400V
1A
±(3% of rdg+3 dgt)
100 A
0.1 A
±(3% of rdg+10 dgt)
560 A/ 1000 V
1A/1V
±(3% of rdg+3 dgt)
±(3% of rdg+5 dgt)
Pow er Factor
1…13
Harmonics
(RMS & %)
3)
3)
DF
Crest Factor
3)
POWER CLAMP
1000A Peak
POWER CLAMP
400A Peak
POWER CLAMP
1000A INRUSH 4)
±(5% of rdg+20 dgt)
999.9A
0.1 A
99.99 A
0.01 A
±(3% of rdg+0.3A)
400 A
0.1 A
±(3% of rdg+5 dgt)
Resistance
4000 Ohm
1 Ohm
±(0.5% of rdg+5 dgt)
Continuity
Below 40 Ohm
1 Ohm
±(0.5% of rdg+5 dgt)
Diode
0…2.2V
0.001 V
±(0.5% of rdg+5 dgt)
POWER CLAMP
400A INRUSH
4)
Note:- Accuracy claimed for Power and Current when conductor is positioned at
the center of the jaw.
1) For DC A make auto zero correction by long pressing the HOLD key
For Power Clamp 1000A
2) Accuracy Defined for V ≧ 10V and I ≧ 5A
Add 10 digit to accuracy when power is < 5.000 kW/ kVA/ kVAr or < 6.700 hp
3) Accuracy Defined for V ≧ 10V and I ≧ 10A
4) Accuracy Defined for I ≧ 10A
For Power Clamp 400A
2) Accuracy Defined for V ≧ 10V and I ≧ 4A
Add 10 digit to accuracy when power is < 5.000 kW/ kVA/ kVAr or < 6.700 hp
3) Accuracy Defined for V ≧ 10V and I ≧ 10A
4) Accuracy Defined for I ≧ 5A
For Power Clamp 1000A
In 1P2W mode maximum power meter can measure is,
1000 kVA / 1000 kVAr / 1000 kW / 1341 hp
In 3P4W mode maximum power meter can measure is,
3000 kVA / 3000 kVAr / 3000 kW / 4023 hp
In 3P3W mode maximum power meter can measure is,
1732 kVA / 1732 kVAr / 1732 kW / 2322 hp
For Power Clamp 400A
In 1P2W mode maximum power meter can measure is,
400 kVA / 400 kVAr / 400 kW / 536 hp
In 3P4W mode maximum power meter can measure is,
1200 kVA / 1200 kVAr / 1200 kW / 1608 hp
In 3P3W mode maximum power meter can measure is,
693 kVA / 693 kVAr / 693 kW / 928 hp
34
PDF created with FinePrint pdfFactory trial version www.softwarelabs.com
Over Load Capacity
Over load capacit y
Meas ur ing f unction
Meas ur ing r ange
VDC - VAC - VACDC
999.9 V
1000 V DC/AC
999.9A
1100 A AC/DC
POWER CLAMP 1000A
ADC - AAC -AACDC
POWER CLAMP 400A
ADC - AAC - A ACDC
99.99 A
Act ive Pow er
Reactive Pow er
Appar ent Pow er
Hor s e Pow er
k Wh
9.999 kW/ kVA/ kVAr/ hp/ kWh
Ahr
Inr us h
440A AC/DC
400 A
Continuously
1000 V DC/AC
99.99 kW/ kVA/ kVAr/ hp/ kWh 1100 A AC/DC
999.9 kW/ kVA/ kVAr/ hp/ kWh f or Pow er Clamp
1000A
999.9 kW/ kVA/ kVAr/ hp/ kWh 440 A AC/DC for
Pow
er Clamp
999.9 Ahr
400A
1000 A / 400A
Res is t ance /
Continuit y
4000 Ohm
Diode
2.2 V
Over load
dur ation
Over load value
Continuously
Continuously
1000 V DC/AC
eff /rms
Sine w ave
10 Secs
Influence quantity and variations
Infulence
quantity
Range of Infuence
Temperature
Frequecy of
the measured
quantity
Crest Factor
Supply
Voltage
Relative
humidity
V AC
V DC
V ACDC
A AC
A DC
A ACDC
AC Power
DC Power
Resistance/ Diode/
Continuity
V AC
V ACDC
A AC
A ACDC
0 ºC… 21 ºC
and
25 ºC….50 ºC
40 Hz… 50 Hz
and
60 Hz….400 Hz
45 Hz…….65 Hz
1)
2)
1.4…2
2…2.5
2.5...5
When Low Battery
symbol is ON
75%
Measured
quantity /
Measuring Range
Variation
0.15 X Intrinsic Error / ºC
1 X Intrinsic Error
AC Power
V AC
A AC
1% + Intrinsic Error
2.5% + Intrinsic Error
4% + Intrinsic Error
All Ranges
1 X Intrinsic Error
All Ranges
1 X Intrinsic Error
1) Except Sine Wave
CF 2 @ 690V, 690A for Power Clamp Meter 1000 A ACDC
CF 2 @ 690V, 280A for Power Clamp Meter 400 A ACDC
CF 3 @ 460V, 460A for Power Clamp Meter 1000 A ACDC
CF 3 @ 690V, 186A for Power Clamp Meter 400 A ACDC
CF 4 @ 345V, 345A for Power Clamp Meter 1000 A ACDC
CF 4 @ 345V, 140A for Power Clamp Meter 400 A ACDC
CF 5 @ 280V, 280A for Power Clamp Meter 1000 A ACDC
CF 5 @ 280V, 112A for Power Clamp Meter 400 A ACDC
2) Except for 50 or 60 Hz
35
PDF created with FinePrint pdfFactory trial version www.softwarelabs.com
Digital Display :
Display
Character Height
Number of digits
Maximum Count
Over range Indication
Polarity indication
Seven Segment Display
Main Display character : 11.5 mm
Sub Display character : 7.2 mm
4 digits
9999 counts for V, I & P and 4000 counts for Resistance
"OL"is displayed.
" - " sign is displayed for negative values.
Power supply
Battery
9V flat cell battery , manganese-di oxide cell according
to IEC6F22, alkaline-manganese cell according to
IEC6LR61
14 mA average (without backlit)
Approx 48 Hrs
Consumption
Battery life
Electrical Safety:
As per IEC 61010-1 2010-06
Measuring Category
Pollution Degree
High Voltage (HV)
III
IV
1000V 600V
2
7.4 kV between housing and input
4.26 kV between housing with jaws and input
Ingress Protection
IP 50
IP 20
EMC
Electromagnetic compatibility
Emission
Immunity
IEC61326: 2012 ClassB
IEC61326:2012
IEC61000-4-2 8 kV Air discharge
4kV contact discharge
IEC61000-4-3 3V/m
Environmental
Conditions
Operating temperature
range
Storage temperature
Range
Relative humidity
Altitude
for housing
for terminal
0 to +55° C
0
0
-20 C...+70 C
0 to 75 % non condensing
up to 2000m
Mechanical configuration
Dimensions
Weight
90mm(W)x270mm(L)x70mm(H)
600g approx.,including battery
36
PDF created with FinePrint pdfFactory trial version www.softwarelabs.com
17. Maintenance
17.1.Battery
Prior to initial start-up,or after storage of power clamp meter, verify that the battery
of power clamp meter does not leak. Repeat this check in regular short intervals.
If the battery leaks, completely remove the battery electrolyte carefully
with a moist cloth and install a new battery before you operate power clamp
meter again.
When the symbol "
" is ON (Battery Voltage < 6.5 V) on the LCD replace
the battery as soon as possible. Measurement can be done, but a reduced measuring
accuracy must be taken into account. “ bAtt ” will be displayed on LCD when
battery voltage drops below 5 V, after that measurement is not possible with meter.
The Power clamp meter operates with a 9V flat cell battery according to IEC6F22
or IEC6LR61 or with a suitable NiCd storage battery .
Disconnect the instrument from the measuring circuit
Attention! before opening battery cover to replace the batteries.
Replacing the battery
Place the power clamp meter on its face. Loosen the screw of battery cover.
Which is at rear bottom side of meter.Remove battery cover by
Sliding It to bottom side.
Remove the battery from the battery compartment and careful ly
disconnect battery connectors.
Snap the battery connectors to a new 9 V battery and insert the
battery into the battery compartment.
Replace the battery cover by fitting it into slots on battery compartment
Tighten the battery cover with the screw.
Please destroy the battery in an environment friendly way
.
17.2.Periodic Check-up:
The power clamp meter does not require any specific maintenance. The surface
between opening jaws should be cleaned with dry cloth before operating.
Avoid use of cleansers, abrasives or solvents.
18.Servicing
When you need service, refer the cover page for company addre ss.
37
PDF created with FinePrint pdfFactory trial version www.softwarelabs.com
WARRANTY
Dear Customer,
You are now the privileged owner of Digital Power Clamp meter / Accessories, a product that ranks
the first of its kind in the world.
Company provides 12 months warranty from the original date of purchase against defective
material and workmanship.
In the unlikely event of failure of this meter / accessories within the warranty period,
Company will repair the meter / accessories free of charges. Please hand over the meter /
accessories to the dealer / Stockiest from whom you have purchased along with this card
and relevant cash memo / Invoice. This warranty entitles you to bring the meter /
accessories at your cost to the nearest stockiest / dealer and collect it after repairs.
NO TRANSPORTATION CHARGES WILL BE REIMBURSED
The warranty is not valid in following cases:
1) Warranty card duly signed / stamped and original cash memo / invoice are not sent
along with the meter / accessories.
2) Complete warranty card is not presented to authorized person at the time of
repairs.
3) Meter / accessories not used as per the instruction in the user manual.
4) Defect caused by misuse, negligence, accidents, tampering and acts of god.
5) Improper repairing by any person not authorized by the company.
6) Any sort of modification, alteration is made in electrical circuitry.
7) Seal provided inside is broken.
Warranty of Digital Clamp meter / Accessories does not cover fuses and battery. In case of
dispute to the validity of warranty, the decision of Company’s service center will be final.
If you bought this meter / accessories directly from the company, and if you notice transit
Damage, then you must obtain the insurance surveyors report and forward it to Company.
Thank you.
(To be lled by authorized dealer)
Model No.
:
Serial Number
:
Date of Purchase
:
Cash Memo / Invoice No.
:
Dealer’s Signature
:
Dealer’s Stamp
:
Scope of
Supply:
1) Meter
3) Battery
5) Warranty Card
7) Carrying Case
2) Cable set
4) User Manual
6) Test Certicate
8) Crocodile Clips
38
PDF created with FinePrint pdfFactory trial version www.softwarelabs.com
39
PDF created with FinePrint pdfFactory trial version www.softwarelabs.com
40
PDF created with FinePrint pdfFactory trial version www.softwarelabs.com
41
PDF created with FinePrint pdfFactory trial version www.softwarelabs.com
RISHABH INSTRUMENTS PVT. LTD.
Factory: F-31, MIDC, Satpur, Nashik-422 007 India
Phone(0253) 2202202, 2202162
Fax : 91 - 0253 - 2351064, 2202302
E-mail : International: exp.marketing@rishabh.co.in
India: marketing@rishabh.co.in
Website:www.rishabh.co.in
Printed in India , Subject to change without Notice
Rev A: 18/03/2015
PDF created with FinePrint pdfFactory trial version www.softwarelabs.com
RISH Clamp 1000A AC/DC 3 ¾ (TRMS) Digits digital Clamp meter
Data Sheet
3 ¾ Digits digital
Clamp meter
www.rishabh.co.in
Page 1 of 5
Version: A 21/02/2013
Application
NULL ZERO Correction for Capacitance
RISH Clamp 1000A measures important electrical parameters like
AC Current, DC Current, AC Voltage, and DC Voltage. It also
features Capacitance, Ohm & Continuity, frequency, and Duty cycle
and temperature measurement.
Null zero connection for capacitance. For nF range, stray
capacitance can be compensated by shift key (Yellow Key)
Product Features
Unique Design
Rish Clamp 1000A is a highly innovative design for features those
increases safety and comfort of user.
Rotating clamp jaws facilitate the measurement at physically
awkward positions, vertical bus bars, conductors placed at
positions difficult to access.
Clamp jaws can be opened or closed with the trigger placed at
bottom side away from the jaws. This allows the user to place
his/her hand at safer distance from live conductor. This greatly
reduces exposure of human beings to electrical shocks
Location and design of trigger eliminates fatigues caused by
single finger operation. It allows spreading the force r e q u i r e d
to open the jaws over more than one finger to ensure
comfortable operation.
Comfortable operation of push buttons and function selector
switch, in adverse field conditions.


AUTO and MANUAL ranging modes
In AUTO ranging mode the instrument automatically selects the
range with best resolution depending on the applied input.
In MANUAL ranging mode range is user selectable using MAN key.
Diode Measurement
For testing diode and transistors, diode measurement function is
available.
Protection from dust and water
IP20 for terminals as per IEC60529
Applicable International Safety standards


600 V CAT IV/1000V CAT III as per International Safety standard
IEC 61010-1- 2010
Double molded Cover for soft touch and firm
grip of the Instrument
Large Jaw Opening
Jaw opening of 55mm for standard wire diameter of 50mm
Narrow Body
Narrow housing for firm grip and easy to carry.
High Accuracy for low current measurement
The clamp meter can measure accurately at not only the High
currents but also Low current ranges.
User selectable Backlit : (Optional)
It is possible to conduct measurement using the clamp meter during
night time in darkness with the help of Backlit. The back lit can be
switched ON or OFF by pressing a single key.
Temperature measurement
Temperatures from -200 to 800 °C using Pt 100 and Pt 1000
sensors.
AUTO POWER OFF
In order to save the power of the Batteries, the clamp meter will
automatically shut OFF if it detects no activity for 10 minutes.
Analog Scale
Analog scale that updates at the rate 20 times/sec to observe
fluctuations in input.
CONTINOUS ON MODE
In this mode, AUTO POWER OFF is disabled.
DATA Hold Function
By pressing DATA HOLD button, reading on the display can be
latched for Hands free operation.
MIN,MAX Function
By pressing MIN/MAX button, the clamp meter will start recording
latest Minimum and Maximum readings
NULL ZERO Correction for Resistance
For Low ohm measurement, the lead resistance can be
compensated by pressing the shift key (Yellow Key)
www.rishabh.co.in
Page 2 of 5
Version: A 21/02/2013
Measuring
function
V dc
V~
Ω
Measuring
range
Resoluti
on
30.00 mV
300.0 mV
3.000 V
30.00 V
300.0 V
1000 V
3.000 V
30.00 V
300.0 V
1000 V
10 µV
100 µV
1 mV
10 mV
100 mV
1V
1 mV
10 mV
100 mV
1V
>10 GΩ // <40pF
>10 GΩ // <40pF
11 MΩ // <40pF
10 MΩ // <40pF
10 MΩ // <40pF
10 MΩ // <40pF
11 MΩ // <40pF
10 MΩ // <40pF
10 MΩ // <40pF
10 MΩ // <40pF
No load voltage
30.00
300.0
3.000
30.00
300.0
3.000
30.00
Ω
Ω
KΩ
KΩ
KΩ
MΩ
MΩ
10 mΩ
100 mΩ
1Ω
10 Ω
100 Ω
1 KΩ
10 KΩ
Max. 3.2 V
Max. 3.2 V
Max. 1.25 V
Max. 1.25 V
Max. 1.25 V
Max. 1.25 V
Max. 1.25 V
2.000 V
1 mV
Max. 3.2 V
300.0 A
1000 A
300.0 A
1000 A
0.1 A
1A
0.1 A
1A
-----------------
A~
A dc
Measuring
Function
F
Hz
Measuring range
Input
impedance
Resolution
Intrinsic error of digital display
± (…% of rdg + …digit) at
reference condition
Over load capacity
Over load value
Overload duration
1000 V
DC
AC
eff / rms
Sine wave
Continuously
1000 V
DC
AC
eff / rms
Sine wave
10 min
1100 A
Continuously
2)
0.5 + 3
0.5 + 3
0.25 + 1
0.25 + 1
0.25 + 1
0.35 + 1
0.75 + 2
(10….300 Digit)
0.75 + 1
> 300 Digit
2)
0.5 + 3
0.5 + 3
0.4 + 1
0.4 + 1
0.4 + 1
0.6 + 1
2.0 + 1
0.25 + 1
1.5 % of range
1.5 % of range
1.5 % of range
1.5 % of range
Discharge
resistance
U0 max.
30.00 nF
300.0 nF
3.000 µF
30.00 µF
10 pF
100 pF
1 nF
10 nF
250 KΩ
250 KΩ
25 KΩ
25 KΩ
2.5 V
2.5 V
2.5 V
2.5 V
300.0 Hz
3.000 KHz
30.00 KHz
100.0 KHz
0.1 Hz
1 Hz
10 Hz
100 Hz
f min V dc
1 Hz
1 Hz
10 Hz
100 Hz
f min V ~
45 Hz
45 Hz
45 Hz
100 Hz
+ 5 Digits
+ 5 Digits
+ 5 Digits
+ 5 Digits
Intrinsic error of digital display
± (…% of rdg + …digit) at
reference condition
2)
1.0 + 3
1.0 + 3
1.0 + 3
3.0 + 3
0.5 + 1
3)
4)
%
2.0....98.0%
Pt
100
ºC
Pt
1000
-200.0...
+200.0 ºC
+200.0...
+850.0 ºC
-100.0...
+200.0 ºC
+200.0...
+850.0 ºC
0.1 %
1)
2 Hz
-
0.1 ºC
2 Hz… 1kHz ± 5 Digit
4)
1 kHz … 10 kHz; ± 5 Digit / kHz
2 Kelvin + 5 Digit
-
1.0 + 5
0.1 ºC
2 Kelvin + 2 Digit
-
-
0.1 ºC
1.0 + 2
1)
Over load
value
Over load
duration
1000 V
DC
AC
eff / rms
Sine
10 min
≤ 3 kHz
1000 v
≤ 30 kHz;
300 V
≤100 kHz
30 V
Continuously
1000 V
DC
AC
eff / rms
Sine
10 min
5)
5)
0.1 ºC
Over load capacity
5)
5)
1) At 0º …. + 40 ºC
2) With zero adjustment, without zero adjustment + 35 digits
3) Range :
3 V ac/dc:
Ue = 1.5 V eff/rms … 100 V eff/rms
30 V ac/dc: Ue = 15 V eff/rms … 300 V eff/rms
300 V ac/dc: Ue = 150 V eff/rms … 1000 V eff/rms
4) On the range 3 V dc, square – wave signal positive on one side 5 … 15 V,
f = const., not 163.84 Hz or integral multiple.
5) Without sensor
Reference conditions for Accuracy
Reference temperature
Relative Humidity
Waveform of measured quantity
Input frequency
Battery Voltage
www.rishabh.co.in
23°C ± 2K
45%...55% RH
Sinusoidal
50 or 60 Hz ±2%
8 V ± 0.1 V
Page 3 of 5
Version: A 21/02/2013
Response time
Influence Quantity
Range of Influence
Measuring Ranges
Noise quantity max. 1000 V
Common
Mode
interference
voltage
Normal Mode
interference
voltage
Noise quantity max. 1000 V ~
50 Hz, 60 Hz sinusoidal
> 120 dB
30 V~
300 V~
> 70 dB
1000 V~
> 60 dB
Noise quantity V ~
Value of the measuring range
at a time
Max. 1000V~ ,50Hz, 60Hz
sinusoidal
V dc
> 50dB
Noise quantity max. 1000 V-
V~
>110dB
Display
Environmental
Operating temperature
Storage temperature
Relative humidity
Terminal Protection
V dc
3V~
Attenuation
-10 to +55°C
-20 to +70°C
0... 90% non condensing
IP50 for Housing and IP20
for terminals
Number of digits
Maximum count
Over range indication
3 ¾ digits.
3100 counts.
“OL” is displayed.
Polarity indication
“─” sign is displayed for DC
functions, if positive pole is at“ ”.
Battery
Battery Voltage
Battery type
Battery Life
9 V DC
Manganese Dioxide Cell as per
IEC6F22 , alkaline manganese
cell as per IEC 6LR 61
Minimum 220 hours on Vdc,
Adc, 80 hours on Vac, Aac.
Influence Quantities and Variations
Influence Quantity
Temperature
Frequency of the
measured quantity
www.rishabh.co.in
Range of Influence
0 ºC
+21 ºC
and
+25 ºC…+40ºC
15 Hz....<30 Hz
30 Hz....<45 Hz
> 65 Hz...400 Hz
>400 Hz…1 KHz
>1 KHz…20 Khz
15 Hz…<30 Hz
30Hz …<45 Hz
>65 Hz … 1 Khz
15 Hz…<30 Hz
30Hz …<45 Hz
>66 Hz… 1 kHz
Measured Quantity/
Measuring Range
Variation 1)
± (….% of rdg. + ….digits)
30/300 mV dc
3...300 V dc
1000 V dc
V~
30 Ω 2)
300 Ω
3 KΩ – 3 MΩ
30 MΩ
30 nF 2) – 3 μF
30 μF
Hz
%
-200…+200 ºC
+200…+850ºC
A ~, A DC
1.0 + 3
0.15 + 1
0.2 + 1
0.4 + 1
0.15 + 2
0.25 + 2
0.15 + 1
1.0 + 1
0.5 + 2
2.0 + 2
0.5 + 1
± 5 digits
0.5 K + 2
0.5 + 2
0.75% of range + 1
3…300 V ~
1000 V ~
A~
Page 4 of 5
--2.0 + 3
2.0 + 3
---3.0 + 3
0.5 % of range + 1
Version: A 21/02/2013
Influence Quantity
Range of Influence
Measured Quantity/
Measuring Range
Variation 1)
± (….% of rdg. + ….digits)
1....3
Wave form of the
measured quantity 3)
V ~ 4) A ~ 4)
Crest factor CF
1....5
Battery Voltage
5)
…< 7.9 V
> 8.1 V …10.0 V
75%
Relative humidity
3 Days
Meter off
HOLD
MIN/MAX
-
V DC
V~
A~
30Ω / 300 Ω/˚C
3 kΩ – 30MΩ
nF, μF
Hz
%
V~,VDC
A~, ADC
Ω
F
Hz
%
0
C
-V ac/dc , A ~ , ADC
2 Digit
4 Digit
6 Digit
4 Digit
3 Digit
1 Digit
1 Digit
1 Digit
1 х intrinsic error
± 1 digits
± 2 digits
1) With temperature: Error data apply per 10 K change in temperature.
With frequency: Error data apply to a display from 300 digits onwards.
2) With zero adjustment.
3) With unknown waveform (crest factor CF > 2), measure with manual range selection
4) With the exception of sinusoidal waveform.
5) After the “
” symbol is displayed
Applicable Standards
EMC
Immunity
IEC 61326: Class B
IEC 61000-4-2
8 KV atmosphere discharge,
4 KV contact discharge
IEC 61000-4-3 : 3 V/m
Safety
IEC 61010-1-2010
IP for water & dust
Pollution degree
Installation category
High Voltage Test
IEC60529
2
IV
6.7 kV AC, 50Hz for 1 minute
between housing and input.
3.7 kV AC, 50Hz for 1 minute
between housing with jaws and
input.
Weight
0.6 Kg
Warranty
1 years
RISHABH INSTRUMENTS PVT.LTD.
F-31, MIDC, Satpur, Nashik-422 007,India.
Tel.: +91 253 2202028, 2202202 Fax : +91 253 2351064
E-mail : India :- marketing@rishabh.co.in
International :- exp.marketing@rishabh.co.in
www.rishabh.co.in
www.rishabh.co.in
Page 5 of 5
Version: A 21/02/2013
RISH Clamp 1000A / 300A 3 ¾ Digits digital Clamp meter
Data Sheet
3 ¾ Digits digital
Clamp meter
www.rishabh.co.in
Page 1 of 5
Version: G 28/12/2011
Application
NULL ZERO Correction for Resistance
RISH Clamp 1000A /300A measures important electrical parameters
like AC Current, AC Voltage, DC Voltage. It also features
Capacitance, Ohm & Continuity, frequency, Dutycycle and
temperature measurement.
For Low ohm measurement, the lead resistance can be
compensated by pressing the shift key (Yellow Key)
Product Features
Null zero connection for capacitance. For nF range, stray
capacitance can be compensated by shift key (Yellow Key)
Unique Design
Rish clamp 1000A / 300Ais a highly innovative design for
features those increases safety and comfort of user.
Rotating clamp jaws facilitate the measurement at physically
awkward positions, vertical bus bars, conductors placed at
positions difficult to access.
Clamp jaws can be opened or closed with the trigger placed at
bottom side away from the jaws. This allows the user to place
his/her hand at safer distance from live conductor. This greatly
reduces exposure of human beings to electrical shocks
Location and design of trigger eliminates fatigues caused by
single finger operation. It allows spreading the force r e q u i r e d
to open the jaws over more than one finger to ensure
comfortable operation.
Comfortable operation of push buttons and function selector
switch, in adverse field conditions.
n
n
n
n
Large Jaw Opening
Rish Clamp 1000A: Jaw opening of 55mm for standerd wire diameter
of 50mm
Rish Clamp 300A : Jaw opening of 44mm for standerd wire diameter
of 40mm
NULL ZERO Correction for Capacitance
AUTO and MANUAL ranging modes
In AUTO ranging mode the instrument automatically selects the
range with best resolution depending on the applied input.
In MANUAL ranging mode range is user selectable using MAN key.
Diode Measurement
For testing diode and transistors, diode measurement function is
available.
Protection from dust and water
IP20 for terminals as per IEC60529
Applicable International Safety standards
600 V CAT IV/1000V CAT III as per International Safety standard
IEC 61010-1- 2001
Double molded Cover for soft touch and firm
grip of the Instrument
Narrow Body
Narrow housing for firm grip and easy to carry.
High Accuracy for low current measurement
The clamp meter can measure accurately at not only the High
currents but also Low current ranges.
User selectable Backlit : (Optional)
It is possible to conduct measurement using the clamp meter during
night time in darkness with the help of Backlit. The back lit can be
switched ON or OFF by pressing a single key.
Temperature measurement
Temperatures from -200 to 800 °C using Pt 100 and Pt 1000
sensors.
AUTO POWER OFF
In order to save the power of the Batteries, the clamp meter will
automatically shut OFF if it detects no activity for 10 minutes.
Analog Scale
Analog scale that updates at the rate 20 times/sec to observe
fluctuations in input.
CONTINOUS ON MODE
In this mode, AUTO POWER OFF is disabled.
DATA Hold Function
By pressing DATA HOLD button, reading on the display can be
latched for Hands free operation.
MIN,MAX Function
By pressing MIN/MAX button, the clamp meter will start recording
latest Minimum and Maximum readings
www.rishabh.co.in
Page 2 of 5
Version: G 28/12/2011
Measuring
function
V dc
V~
Ω
A~
Measuring
range
Resolution
30.00 mV
300.0 mV
3.000 V
30.00 V
300.0 V
1000 V
3.000 V
30.00 V
300.0 V
1000 V
10 μV
100 μV
1 mV
10 mV
100 mV
1V
1 mV
10 mV
100 mV
1V
30.00 Ω
300.0 Ω
3.000 KΩ
30.00 KΩ
300.0 KΩ
3.000 MΩ
30.00 MΩ
2.000 V
300.0 A
1000A*
10 mΩ
100 mΩ
1Ω
10 Ω
100 Ω
1 KΩ
10 KΩ
1 mV
0.01 A
0.1 A
Measuring
function
F
Hz
Measuring
range
Input
impedance
>10 GΩ // <40pF
>10 GΩ // <40pF
11 MΩ // <40pF
10 MΩ // <40pF
10 MΩ // <40pF
10 MΩ // <40pF
11 MΩ // <40pF
10 MΩ // <40pF
10 MΩ // <40pF
10 MΩ // <40pF
No load voltage
Max. 3.2 V
Max. 3.2 V
Max.1.25 V
Max. 1.25 V
Max. 1.25 V
Max. 1.25 V
Max. 1.25 V
Max. 3.2 V
---------
Resolution
Discharge
resistance
Intrinsic error of digital display
± (…% of rdg + …digit) at
reference condition
0.5 + 3 2)
0.5 + 3
0.25 + 1
0.25 + 1
0.25 + 1
0.35 + 1
0.75 + 2
(10….300 Digit)
0.75 + 1
> 300 Digit
0.5 + 3 2)
0.5 + 3
0.4 + 1
0.4 + 1
0.4 + 1
0.6 + 1
2.0 + 1
0.25 + 1
1.5 % of range + 5 Digits
1.5 % of range + 5 Digits
U0 max.
Over load capacity 1)
Over load value Overload duration
1000 V
DC
AC
eff / rms
Sine wave
Continuously
1000 V
DC
AC
eff / rms
Sine wave
10 min
1100* A / 360A
Continuously
Intrinsic error of digital display
± (…% of rdg + …digit) at
reference condition
30.00 nF
300.0 nF
3.000 μF
30.00 μF
10 pF
100 pF
1 nF
10 nF
250 KΩ
250 KΩ
25 KΩ
25 KΩ
2.5 V
2.5 V
2.5 V
2.5 V
1.0 + 3 2)
1.0 + 3
1.0 + 3
3.0 + 3
300.0 Hz
3.000 KHz
30.00 KHz
100.0 KHz
0.1 Hz
1 Hz
10 Hz
100 Hz
f min V dc
1 Hz
1 Hz
10 Hz
100 Hz
f min V ~
45 Hz
45 Hz
45 Hz
100 Hz
0.5 + 1 3)
1000 V
DC
AC
eff / rms
Sine
2.0....98.0%
ºC
Pt
100
Pt
1000
-200.0...
+200.0 ºC
+200.0...
+850.0 ºC
-100.0...
+200.0 ºC
+200.0...
+850.0 ºC
0.1 %
2 Hz
--
0.1 ºC
0.1 ºC
–
--
2 Kelvin + 2 Digit 5)
0.1 ºC
0.1 ºC
2 Hz… 1kHz ± 5 Digit
1 kHz … 10 kHz; ± 5 Digit /
kHz 4)
2 Kelvin + 5 Digit 5)
1.0 + 5 5)
–
--
10 min
≤ 3 kHz
4)
%
Over load capacity 1)
Over load Overload
value
duration
1000 v
≤ 30 kHz;
300 V
≤100 kHz
30 V
1000 V
DC
AC
eff / rms
Sine
Continuously
10 min
1.0 + 2 5)
1) At 0º …. + 40 ºC
2) With zero adjustment, without zero adjustment + 35 digits
3) Range :
3 V ac/dc:
Ue = 1.5 V eff/rms … 100 V eff/rms
30 V ac/dc:
Ue = 15 V eff/rms … 300 V eff/rms
300 V ac/dc: Ue = 150 V eff/rms … 1000 V eff/rms
4) On the range 3 V dc, square – wave signal positive on one side 5 … 15 V,
f = const., not 163.84 Hz or integral multiple.
5) Without sensor
*applicable for 1000A
Reference conditions for Accuracy
Reference temperature
Relative Humidity
Waveform of measured quantity
Input frequency
Battery Voltage
www.rishabh.co.in
23°C ± 2K
45%...55% RH
Sinusoidal
50 or 60 Hz ±2%
8 V ± 0.1 V
Page 3 of 5
Version: G 28/12/2011
Response time
Influence Quantity
Range of Influence
Measuring Ranges
Noise quantity max. 1000 V
Common
Mode
interference
voltage
Normal Mode
interference
voltage
Noise quantity max. 1000 V ~
50 Hz, 60 Hz sinusoidal
> 120 dB
30 V~
300 V~
> 70 dB
1000 V~
> 60 dB
Noise quantity V ~
Value of the measuring range
at a time
Max. 1000V~ ,50Hz, 60Hz
sinusoidal
V dc
> 50dB
Noise quantity max. 1000 V-
V~
>110dB
Display
Environmental
Operating temperature
Storage temperature
Relative humidity
Terminal Protection
V dc
3V~
Attenuation
-10 to +55°C
-20 to +70°C
0... 90% non condensing
IP50 for Housing and IP20
for terminals
Number of digits
Maximum count
Over range indication
3 ¾ digits.
3100 counts.
“OL” is displayed.
Polarity indication
“─” sign is displayed for DC
functions, if positive pole is at“ ”.
Battery
Battery Voltage
Battery type
Battery Life
9 V DC
Manganese Dioxide Cell as per
IEC6F22 , alkaline manganese
cell as per IEC 6LR 61
Minimum 220 hours on Vdc,
Adc, 80 hours on Vac, Aac.
Influence Quantities and Variations
Influence Quantity
Temperature
Frequency of the
measured quantity
www.rishabh.co.in
Range of Influence
0 ºC
+21 ºC
and
+25 ºC…+40ºC
15 Hz....<30 Hz
30 Hz....<45 Hz
> 65 Hz...400 Hz
>400 Hz…1 KHz
>1 KHz…20 Khz
15 Hz…<30 Hz
30Hz …<45 Hz
>65 Hz … 1 Khz
15 Hz…<30 Hz
30Hz …<45 Hz
>66 Hz… 1 kHz
Measured Quantity/
Measuring Range
Variation 1)
± (….% of rdg. + ….digits)
30/300 mV dc
3...300 V dc
1000 V dc
V~
30 Ω 2)
300 Ω
3 KΩ – 3 MΩ
30 MΩ
30 nF 2) – 3 μF
30 μF
Hz
%
-200…+200 ºC
+200…+850ºC
A~
1.0 + 3
0.15 + 1
0.2 + 1
0.4 + 1
0.15 + 2
0.25 + 2
0.15 + 1
1.0 + 1
0.5 + 2
2.0 + 2
0.5 + 1
± 5 digits
0.5 K + 2
0.5 + 2
0.75% of range + 1
3…300 V ~
1000 V ~
A~
Page 4 of 5
--2.0 + 3
2.0 + 3
---3.0 + 3
--2.0% of range + 1
Version: G 28/12/2011
Influence Quantity
Range of Influence
Measured Quantity/
Measuring Range
Variation 1)
± (….% of rdg. + ….digits)
1....3
Wave form of the
measured quantity 3)
V ~ 4) A ~ 4)
Crest factor CF
1....5
Battery Voltage
5)
…< 7.9 V
> 8.1 V …10.0 V
75%
Relative humidity
3 Days
Meter off
HOLD
MIN/MAX
-
V DC
V~
A~
30Ω / 300 Ω/˚C
3 kΩ – 30MΩ
nF, μF
Hz
%
V~,VDC
A~
Ω
F
Hz
%
0
C
-V ac/dc , A ~
2 Digit
4 Digit
6 Digit
4 Digit
3 Digit
1 Digit
1 Digit
1 Digit
1 х intrinsic error
± 1 digits
± 2 digits
1) With temperature: Error data apply per 10 K change in temperature.
With frequency: Error data apply to a display from 300 digits onwards.
2) With zero adjustment.
3) With unknown waveform (crest factor CF > 2), measure with manual range selection
4) With the exception of sinusoidal waveform.
5) After the “
” symbol is displayed
Applicable Standards
EMC
Immunity
IEC 61326: Class B
IEC 61000-4-2 8 KV
atmosphere discharge, 4 KV
contact discharge
IEC 61000-4-3 : 3 V/m
Safety
IEC 61010-1-2001
IP for water & dust
Pollution degree
Installation category
High Voltage Test
IEC60529
2
IV
6.7 kV AC, 50Hz for 1 minute
between housing and input.
3.7 kV AC, 50Hz for 1 minute
between housing with jaws and
input.
Weight
0.6 Kg
Warranty
1 years
RISHABH INSTRUMENTS PVT.LTD.
F-31, MIDC, Satpur, Nashik-422 007,India.
Tel.: +91 253 2202160, 2202202 Fax : +91 253 2351064
E-mail : India :- marketing@rishabh.co.in
International :- exp.marketing@rishabh.co.in
www.rishabh.co.in
www.rishabh.co.in
Page 5 of 5
Version: G 28/12/2011