please fill out - SpectraSensors

ReliablePrecise Rapid
SS2100i-1 Single Box
IECEx/ATEX Zone 1 Datasheet
TDLAS gas analyzer
Key Features
•Touch keypad interface, no tools
required
•Simple design, trouble-free
operation
•No routine maintenance
required
•Field calibration not needed
•No drift or interference from
contaminants
•Reliable in harsh environments
•Available for the following
measurements:
H2O (moisture)
CO2 (carbon dioxide)
H2S (hydrogen sulfide)
NH3 (ammonia)
C2H2 (acetylene)
SpectraSensors SS2100i-1 Process Gas
Analyzers are exceptionally reliable for
measuring trace gas components using
tunable diode laser absorption spectroscopy
technology (TDLAS). TDLAS is a highresolution infrared technique that enables
the measurement of specific gases with
precision while avoiding interferences that
are common with traditional infrared
analyzers. The SS2100i-1 is certified for
ATEX, IECEx, CNEx, KC, CCOE, and EAC*.
Simple operation The operation of the
analyzer is very straightforward. +perate
the system in a very brief time. When
coupled with the fact the analyzer has very
little maintenance requirements, the end
result is an extremely low cost of
ownership.
At the same time, technical support
capability is a crucial element of the product
design. There are several health monitoring
•ATEX, IECEx, CNEx KC, CCOE,
and EAC Certifications
SS2100i-1
parameters and remote access is available
using service software or directly through
the touch sensitive keypad.
Reliable Trustworthy measurements are
vital in process analytical applications. The
TDLAS sensor is unaffected by
contaminants and corrosives since the gas
stream never touches the laser or detector.
The SS2100i-1 requires little regular
maintenance and does not need
recalibration or periodic replacement parts
due to the inherent stability of TDLAS
technology.
Trouble-free installation The SS2100i-1
is easy to install; connect the power, data
link and measured gas line and the analyzer
begins working without the need for
extensive calibrations or setup.
* EAC (Formerly GOST-R)
Specifications
Application Data
Target Components
H2O, H2S, CO2, NH3, C2H2 (Ranges from low ppmv to %)*
Principle of Measurement
Tunable Diode Laser Absorption Spectroscopy
Measurement Time
Typically less than 20 seconds*
Environmental Temperature Range
-20°C to +50°C (-4°F to 122°F) - standard
Sample Cell Operating Pressure Range
800-1200 mbara - standard, or 950-1700 mbara - optional*
Pressure to Sample Cabinet
Typically between 140-350 kPaG (20-50 PSIG)*
Sample Flow Rate
0.5-4 SLPM (0.02-0.1 SCFM)*
Electrical & Communications
Input Power, Maximum
120 or 240 VAC ±10%, 50-60 Hz; ~300W*
Analog Communication
Isolated Analog channels, 1200 ohms @ 24 VDC max
Outputs: Qty 2 4-20 mA (measurement value)
Input: Qty 1 4-20 mA (process pressure)*
Serial Communication
Ethernet & RS485 half-duplex
Digital Signals
Outputs: Qty 5 Hi/Lo Alarm, General Fault, Validation Fail*, Validation 1 Active*,
Validation 2 Active*
Inputs: Qty 2 Flow Alarm*, Validation Request*
Protocol
Modbus Gould RTU or Daniel RTU or ASCII
Diagnostic Value Examples
Detector Power (Optics Health), Spectrum Reference Comparison and
Peak Tracking (Spectrum Quality), Cell Pressure and Temperature
(Overall System Health)
LCD Display
Concentration, Cell Pressure and Temperature & Diagnostics
Physical
Electronics Enclosure
IP66 Copper-Free Aluminum with Weather Resistant Polyeurothane Powder
Coating, 80-120 micron thickness
Analyzer Dimensions
670 H x 470 W x 377 D mm (26.3 H x 18.5 W x 14.8 D inches)
(not including Sample System)
Analyzer Weight
Approximately 86 kg (190 lbs) (not including Sample System)
Sample Cell Construction
316L Series Polished Stainless Steel - standard
Number of Sample Cells
1 per Analyzer
Analyzer (Electronics & Laser)
CE, ATEX and IECEx:
II 2 G Ex db IIB+H2 T4 Gb; LCIE 11 ATEX 3071,
IECEx LCI 11.0057x; TC RU C-US.ГБ05.B.00787, CNEx 14.3396, CCE
P353439/1, KTL15KB4B0-0082
Analyzer Assembly
Assembled using electrical components certified for:
II 2 G IIB+H2 T3 Gb
Manage
m
ity
al
t
en
Qu
*Application dependant.
LASER RADIATION–AVOID
DIRECT EXPOSURE TO BEAM
Contact
4333 W. Sam Houston Pkwy N.
Suite 100
Houston, TX 77043
Tel +1 713 300 2700
+1 800 619 2861
Fax +1 713 856 6623
sales@spectrasensors.com
service@spectrasensors.com
www.spectrasensors.com
10 mW MAX CW 800-3000 nm
ISO
CLASS IIIb LASER PRODUCT
9 0 01
DS SS2100i-1 ATEX IECEX Zone1 EN (03/18)
Certification