ADVANCED PHOTO M ETER SYSTEM
DIGITAL ACCURACY FOR BETTER POOL CARE
QUICK START GUIDE
Sensafe.com
2
CONTENTS
Welcome. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Pool eXact ® EZ Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Install “AAA” Batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Fill, Dip, Read
Total Alkalinity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
pH . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-7
Free Chlorine . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-9
Combined Chlorine . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-11
eXact ® Strip Micro CL Interferences . . . . . . . . . . . . . . . . . . . . 11
Phosphate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12-13
Calcium Hardness . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-15
Cyanuric Acid. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16-17
Copper . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18-19
Chloride (Salt). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20-22
For Best Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Pool eXact ® EZ Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Technical Support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
About
Built-in Sample Cell . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Warranty (2 years) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Method Verification Ready Snap® 3. . . . . . . . . . . . . . . . . . . . . 25
Compliance Testing (USEPA) . . . . . . . . . . . . . . . . . . . . . . . . . 26
Carrying Case . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
eXact ® Micro Tests & Reagents . . . . . . . . . . . . . . . . . . . . . . . . 28
Patent Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
WELCOME TO YOUR NEW POOL EXACT® EZ
PHOTOMETER.
Thank you for your Pool eXact ® EZ purchase! This guide will quickly
walk you through the technical details of your new photometer. After
initial set-up, test procedures, and tips, you will be on your way to
digital water testing! Each test will require the use of one or more of
the testing methods outlined in this manual.
YOUR POOL EXACT® EZ COMES WITH:
• Cleaning Brush
• Quick Start Guide (this booklet)
• 48 foil packet strips (6 of each): Free Chlorine (DPD-1),
Combined Chlorine (DPD-3), pH, Total Alkalinity, Calcium
Hardness, Phosphate, Chloride, and Copper
WHAT YOU WILL NEED TO GET STARTED:
• Four (4) AAA batteries
• #4 Phillips head screwdriver
Note: This system has been calibrated for use with only our
eXact ® Micro reagents.
©2016, Industrial Test Systems, Inc. Printed in USA
POOL EXACT® EZ OVERVIEW
3
YOUR NEW POOL EXACT EZ PHOTOMETER IS IDEAL FOR
TESTING AND MAINTAINING YOUR POOL AND SPA WATER.
SAMPLE CELL - 4 ML
Built-in plastic 4mL cell
for collecting your water
sample
LIGHT SOURCE
Uses 525nm
wavelength and
11mm path-length
for testing
LCD DISPLAY
Displays results
and test name
LCD DISPLAY
READ BUTTON
Starts test timer
LOW
BATTERY
MENU BUTTON
Cycles through
available tests
RESULT
UNITS
BASE
Install/replace batteries here
(IP67 rated waterproof)
ZERO/ON BUTTON
Turns the meter
on and creates a
baseline for your
water testing
CELL COVER
Covers the cell for
mixing and bright
light situations
Protected against the effects
of immersion in water to depth
between 15cm and 1m
Part no. 486201
INSTALL “AAA” BATTERIES (NOT INCLUDED)
1. Use a #4 Phillips head screwdriver to remove the screw
from the base of your Pool eXact ® EZ.
2. Remove the base.
3. Install Four (4) new AAA batteries as illustrated inside
your photometer’s battery compartment. We recommend
using high quality batteries.
4. Replace the base firmly with pressure while tightening the
screw. The meter will turn on automatically.
5. Tighten the screw with #4 Phillips head screwdriver. Be
sure not to over tighten.
Sensafe.com
SCREW
Unscrew to
remove base
4
1
2
FILL, DIP, READ
TOTAL ALKALINITY
POWER ON PHOTOMETER
Press the
button to power on the Pool eXact® EZ.
FILL CELL
Rinse the cell 3 times with the water sample to
be tested and FILL to the top to begin test.
TIP
Use the included cleaning brush to clean the cell before/
after each test to prevent reagent interferences and
achieve the most accurate result.
3
4
SELECT TEST
Press and re-press the
display shows AL1.
button until the
ZERO METER
Press
and the LCD Display reads 0PPM,
indicating the meter is ready for testing.
TIP
In bright light situations use cell cover for zeroing and reading.
©2016, Industrial Test Systems, Inc. Printed in USA
5
6
FILL, DIP, READ
TOTAL ALKALINITY
5
REMOVE STRIP
Remove one eXact® Strip Micro AL Total Alkalinity
and set in a dry, convenient place. Replace cap
on bottle.
DIP STRIP AND PRESS READ
Press
to initiate a 20 second countdown and
simultaneously DIP the eXact® Strip by submerging all
pads in the sample then use a gentle, constant, back
and forth motion (2 strokes per second) until the timer
displays ‘1’. Remove and discard the strip.
TIP
For water temperatures above 95°F/35°C (hot tubs), remove
the strip when the timer displays “10”, or use the specially
calibrated SPA eXact® EZ photometer.
7
READ RESULTS
READ result displayed as Total Alkalinity. This result is
automatically stored in the AL1 menu. After testing is
complete, rinse the sample cell immediately.
TIP
If result reads "HI", repeat steps 5-7 with a Total Alkalinity Range
Extender Strip (part no. 486665) until a numerical value is
displayed; each strip adds 130 ppm to the final result.
Sensafe.com
FILL, DIP, READ
PH
6
1
2
POWER ON PHOTOMETER
Press the
button to power on the Pool eXact® EZ.
FILL CELL
Rinse the cell 3 times with the water sample to
be tested and FILL to the top to begin test.
TIP
Use the included cleaning brush to clean the cell before/
after each test to prevent reagent interferences and
achieve the most accurate result.
3
4
SELECT TEST
Press and re-press the
display shows PH2.
button until the
ZERO METER
Press
and the LCD Display reads 0.0PH,
indicating the meter is ready for testing.
TIP
In bright light situations use cell cover for zeroing and reading.
©2016, Industrial Test Systems, Inc. Printed in USA
5
6
7
FILL, DIP, READ
PH
7
REMOVE STRIP
Remove one eXact® Strip Micro PH and set in a dry,
convenient place. Replace cap on bottle.
DIP STRIP AND PRESS READ
Press
to initiate a 20 second countdown and
simultaneously DIP the eXact® Strip by submerging all
pads in the sample then use a gentle, constant, back
and forth motion (2 strokes per second) until the timer
displays ‘1’. Remove and discard the strip.
READ RESULTS
READ result displayed as pH. This result is
automatically stored in the PH2 menu. After testing is
complete, rinse the sample cell immediately.
Sensafe.com
8
1
2
FILL, DIP, READ
FREE CHLORINE AND BROMINE
POWER ON PHOTOMETER
Press the
button to power on the Pool eXact® EZ.
FILL CELL
Rinse the cell 3 times with the water sample to
be tested and FILL to the top to begin test.
TIP
Use the included cleaning brush to clean the cell before/
after each test to prevent reagent interferences and
achieve the most accurate result.
3
4
SELECT TEST
Press and re-press the
display shows CL3.
button until the
ZERO METER
Press
and the LCD Display reads 0.00PPM,
indicating the meter is ready for testing.
TIP
In bright light situations use cell cover for zeroing and reading.
©2016, Industrial Test Systems, Inc. Printed in USA
5
6
7
FILL, DIP, READ
FREE CHLORINE AND BROMINE
9
REMOVE STRIP
Remove one eXact® Strip Micro CL (DPD-1) Free
Chlorine and set in a dry, convenient place. Replace
cap on bottle
DIP STRIP AND PRESS READ
Press
to initiate a 20 second countdown and
simultaneously DIP the eXact® Strip by submerging all
pads in the sample then use a gentle, constant, back
and forth motion (2 strokes per second) until the timer
displays ‘1’. Remove and discard the strip.
READ RESULTS
READ result displayed as Free Chlorine (for result
as Bromine, multply value by 2.6). This result is
automatically stored in the CL3 menu. After testing is
complete, rinse the sample cell immediately.
TIP
If result is 6.0 or greater, repeat steps 5-7 above
using same sample and a fresh strip.
Important! Do not empty the CELL if you are also
testing for Combined Chlorine. You will need the sample.
Sensafe.com
10
FILL, DIP, READ
COMBINED CHLORINE
This procedure is a continuation from Free Chlorine Test Method
on previous pages. DO NOT DISCARD SAMPLE FROM FREE
CHLORINE TEST.
8
ZERO METER
With the Free Chlorine sample still in the CELL,
press
and the LCD Display reads 0.00PPM,
indicating the meter is ready for testing.
TIP
Use the included cleaning brush to clean the cell before/
after each test to prevent reagent interferences and
achieve the most accurate result.
9
10
REMOVE STRIP
Remove one eXact® Strip Micro CL (DPD-3) Combined
Chlorine and set in a dry, convenient place. Replace
cap on bottle.
DIP STRIP AND PRESS READ
Press
to initiate a 20 second countdown and
simultaneously DIP the eXact® Strip by submerging all
pads in the sample then use a gentle, constant, back
and forth motion (2 strokes per second) until the timer
displays ‘1’. Remove and discard the strip.
©2016, Industrial Test Systems, Inc. Printed in USA
11
FILL, DIP, READ
COMBINED CHLORINE
11
READ RESULTS
READ result displayed as Combined Chlorine. This
result is automatically stored in the CL3 menu. After
testing is complete, rinse the sample cell immediately.
eXact® Strip Micro CL (DPD-1, DPD-3, DPD-4) Interferences
Interfering
Substance
Interfering Levels & Treatments
Acidity
If sample has acidity above 150mg/L CaCO3
test may not develop full color. Neutralize to
pH 6.0 to 7.0 with 0.5N Sodium hydroxide.
Alkalinity
If sample has alkalinity above 200mg/L
CaCO3 test may not develop full color.
Neutralize to pH 6.0 to 7.0 with 0.5N
Sulfuric acid.
Bromine &
Bromamines, Br2
Color similar to free chlorine reaction at all levels.
Chlorine Dioxide, ClO2
Color similar to free chlorine reaction at all levels.
Copper, Cu+2
Color development is reduced above 10
ppm (mg/L).
Iodine, I2
Color similar to free chlorine reaction at all levels.
Manganese, Oxidized
(Mn+4, Mn+6) or
Chromium,
Oxidized (Cr+6)
See AWWA procedure 4500-CL F, 1(d) for
removal of interferences.
Monochloramines
Monochloramine interferences are known
(NH2Cl)
to occur in free chlorine
(applies to DPD-1 only) DPD methods. This interference is
dependent on temperature and
monochloramine concentration.
Ozone, O3
Color similar to free chlorine reaction at all levels.
Peroxides
Interference is possible.
pH
Typical pH samples of potable water with
a pH of 6.0 to 9.0 are OK. If outside this
range adjust to pH 6.0 to 7.0 using acid
(0.5N Sulfuric acid) or base (0.5N Sodium
hydroxide).
Sensafe.com
FILL, DIP, READ
PHOSPHATE
12
1
POWER ON PHOTOMETER
Press the
button to power on the Pool eXact® EZ.
TIP
Clean CELL with 0.1N HCl, Distilled Vinegar (5%), or Muriatic
Acid before filling the meter with the sample to be tested.
2
FILL CELL
Rinse the cell 3 times with the water sample to
be tested and FILL to the top to begin test.
TIP
Use the included cleaning brush to clean the cell before/
after each test to prevent reagent interferences and
achieve the most accurate result.
3
4
SELECT TEST
Press and re-press the
display shows PO4.
button until the
ZERO METER
Press
and the LCD Display reads 0.00PPM,
indicating the meter is ready for testing.
TIP
In bright light situations use cell cover for zeroing and reading.
©2016, Industrial Test Systems, Inc. Printed in USA
5
6
7
FILL, DIP, READ
PHOSPHATE
13
REMOVE STRIP
Remove one eXact® Strip Micro PO4 Phosphate and set
in a dry, convenient place. Replace cap on bottle
DIP STRIP AND PRESS READ
Press
to initiate a 20 second countdown and
simultaneously DIP the eXact® Strip by submerging all
pads in the sample then use a gentle, constant, back
and forth motion (2 strokes per second) until the timer
displays ‘1’. Remove and discard the strip. The display
will automatically start to count up for 120 seconds.
READ RESULTS
READ result displayed as Phosphate. This result is
automatically stored in the PO4 menu. After testing is
complete, rinse the sample cell immediately.
Sensafe.com
14
1
2
FILL, DIP, READ
CALCIUM HARDNESS
POWER ON PHOTOMETER
Press the
button to power on the Pool eXact® EZ.
FILL CELL
Rinse the cell 3 times with the water sample to
be tested and FILL to the top to begin test.
TIP
Use the included cleaning brush to clean the cell before/
after each test to prevent reagent interferences and
achieve the most accurate result.
3
4
SELECT TEST
Press and re-press the
display shows CA5.
button until the
ZERO METER
Press
and the LCD Display reads 0PPM,
indicating the meter is ready for testing.
TIP
In bright light situations use cell cover for zeroing and reading.
©2016, Industrial Test Systems, Inc. Printed in USA
5
6
7
FILL, DIP, READ
CALCIUM HARDNESS
15
REMOVE STRIP
Remove one eXact® Strip Micro CA Calcium
Hardness and set in a dry, convenient place. Replace
cap on bottle.
DIP STRIP AND PRESS READ
Press
to initiate a 20 second countdown and
simultaneously DIP the eXact® Strip by submerging all
pads in the sample then use a gentle, constant, back
and forth motion (2 strokes per second) until the timer
displays ‘1’. Remove and discard the strip.
READ RESULTS
READ result displayed as Calcium Hardness
(CaCO3). This result is automatically stored in the CA5
menu. After testing is complete, rinse the sample cell
immediately.
Sensafe.com
FILL, DIP, READ
CYANURIC ACID
16
1
2
POWER ON PHOTOMETER
Press the
button to power on the Pool eXact® EZ.
FILL CELL
Rinse the cell 3 times with the water sample to
be tested and FILL to the top to begin test.
TIP
Use the included cleaning brush to clean the cell before/
after each test to prevent reagent interferences and
achieve the most accurate result.
3
4
SELECT TEST
Press and re-press the
display shows CY6.
button until the
ZERO METER
Place cell cover on sample cell before zeroing.
Press
and the LCD Display reads 0PPM,
indicating the meter is ready for testing.
©2016, Industrial Test Systems, Inc. Printed in USA
5
6
7
FILL, DIP, READ
CYANURIC ACID
17
SHAKE BOTTLE & ADD DROPS
Tilt the meter to discard about 0.2mL water to leave
room for liquid reagent. Shake the bottle of eXact®
Reagent CY to mix the suspension in the bottle. Add
5 drops of the eXact® Reagent CY to the sample cell.
Keep reagent bottle vertical while adding drops.
CAP & PRESS READ
Place the cell cover onto the sample cell. Press
to initiate a 20 second countdown. Place thumb or
finger over cap, and mix the sample by turning the
meter upside-down repetitively during the 20 second
countdown. When timer displays 1, place the meter on a
flat surface (upright). The display will automatically start
to count up for 60 seconds.
READ RESULTS
READ result displayed as Cyanuric Acid. This result is
automatically stored in the CY6 menu. After testing is
complete, rinse the sample cell immediately.
Sensafe.com
FILL, DIP, READ
COPPER
18
1
2
POWER ON PHOTOMETER
Press the
button to power on the Pool eXact® EZ.
FILL CELL
Rinse the cell 3 times with the water sample to
be tested and FILL to the top to begin test.
TIP
Use the included cleaning brush to clean the cell before/
after each test to prevent reagent interferences and
achieve the most accurate result.
3
4
SELECT TEST
Press and re-press the
display shows CU7.
button until the
ZERO METER
Press
and the LCD Display reads 0.00PPM,
indicating the meter is ready for testing.
TIP
In bright light situations use cell cover for zeroing and reading.
©2016, Industrial Test Systems, Inc. Printed in USA
5
6
7
FILL, DIP, READ
COPPER
19
REMOVE STRIP
Remove one eXact® Strip Micro CU Copper and set in a
dry, convenient place. Replace cap on bottle
DIP STRIP AND PRESS READ
Press
to initiate a 20 second countdown and
simultaneously DIP the eXact® Strip by submerging all
pads in the sample then use a gentle, constant, back
and forth motion (2 strokes per second) until the timer
displays ‘1’. Remove and discard the strip. The display
will automatically start to count up for 120 seconds.
READ RESULTS
READ result displayed as Copper. This result is
automatically stored in the CU7 menu. After testing is
complete, rinse the sample cell immediately.
Sensafe.com
20
1
CHLORIDE (Salt)
PREPARE SAMPLE
Prepare sample for testing using Mini Dilution Kit
II #487202.
1. Rinse 50mL graduated conical tube with
distilled or deionized (salt-free) water.
2. Rinse the 3.0mL syringe with saltwater sample
to be tested. Finally, fill the 3.0mL syringe to the
2.0mL line very precisely (plunger ring should
line up at the 2.0mL line and little or no air
bubble should be present).
3. Add the syringe content (2.0mL salt water
sample) to a clean 50mL graduated conical
tube by pushing plunger all the way down to
expel sample.
4. Now, fill the graduated conical tube to the 40mL
line with distilled or deionized (salt-free) water.
Cap graduated conical tube.
5. Mix content of graduated conical tube by turning
up side down at least three times. 1:20 Dilution
Sample is ready for testing.
©2016, Industrial Test Systems, Inc. Printed in USA
1
2
FILL, DIP, READ
CHLORIDE (Salt)
21
POWER ON PHOTOMETER
Press the
button to power on the Pool eXact® EZ.
FILL CELL
Rinse the cell 3 times with the water sample to
be tested and FILL to the top to begin test.
TIP
Use the included cleaning brush to clean the cell before/
after each test to prevent reagent interferences and
achieve the most accurate result.
3
4
SELECT TEST
Press and re-press the
display shows CHH.
button until the
ZERO METER
Press
and the LCD Display reads 0.00PPM,
indicating the meter is ready for testing.
TIP
In bright light situations use cell cover for zeroing and reading.
Sensafe.com
22
5
6
7
FILL, DIP, READ
CHLORIDE (as NaCl)
REMOVE STRIP
Remove one eXact® Strip Micro CH Chloride III and set
in a dry, convenient place. Replace cap on bottle
DIP STRIP AND PRESS READ
Press
to initiate a 20 second countdown and
simultaneously DIP the eXact® Strip by submerging all
pads in the sample then use a gentle, constant, back
and forth motion (2 strokes per second) until the timer
displays ‘1’. Remove and discard the strip.
READ RESULTS
READ result displayed as Chloride. Due to the limit
of three digits on the Pool eXact® EZ photometer, you
should add a “0” to the end of the result on the LCD
(ex. 121 = 1210 ppm). This result is automatically stored
in the CHH menu. After testing is complete, rinse the
sample cell immediately with cleaning brush.
©2016, Industrial Test Systems, Inc. Printed in USA
FOR BEST ACCURACY
23
The Pool eXact® EZ photometer has a 5 minute auto-shutoff timer.
Each test menu can store 20 results. To retrieve the stored results,
go to the desired test using the MENU key. When the desired test is
displayed, press and hold down the MENU key. Continue holding down
the MENU key to scroll the stored results for that test, starting with the
most recent result. The meter will display, from memory, the last 20
readings in sequence beginning with -20, which is the latest result,
followed by -19, which is the 2nd latest result, etc; and finally -01,
which is the oldest result retained. Only the last 20 readings are stored
in each menu. This meter is able to store 160 results in memory (20 in
each menu).
Before testing, rinse the meter sample cell with the sample water 3 times.
Always fill the cell to capacity (4mL); be careful not to splash liquid over
the side.
Test immediately after filling the cell with the water sample.
To obtain optimal accuracy when testing outdoors (sunlight), use the
cell cover when zeroing and reading the sample.
Due to the strip slitting process, you may find one or two strips that are
noticeably smaller or larger in width than the normal strips in the bottle.
These should be discarded. Using these strips may give unreliable results.
Meter is not compatible for use with powder pillows, tablets, or liquids
from other manufacturers.
Dip strip for entire countdown.
Each eXact® Strip Micro is valid for ONLY one test. Discard strip after use.
Dry the outside of the meter before storage.
Remove batteries before storing for prolonged periods.
Store the meter and test materials out of direct sunlight and away from
chemical storage areas.
Minimize exposure of meter and test reagents to heat above 90°F (32°C).
When installing batteries, verify the O-ring is still attached to the screw
before tightening.
Cleaning the cell with water and brush after each test is recommended
for best accuracy and prior to storage of unit.
Sensafe.com
24
TROUBLESHOOTING
Listed below are possible situations that may arise while testing.
Please contact one of our knowledgeable customer service
representatives if you require further assistance.
Subject
Cause
Solution
Dim screen or no
Low battery
response from meter
“LO” on LCD while Low battery
zeroing
Dirty cell
Cloudy sample
Bad LED
“HI” on LCD while
Result above
reading
detection level
“LO” on LCD while Result below
reading
detection level
“LO” flashes on
Improper test
LCD, then "Err"
procedure for
Combined Chlorine
"HI" flashes on LCD, Combined Chlorine
then "Err"
result above
detection limit
Replace batteries
Replace batteries
Clean cell
Dilute sample or use filter
Contact ITS
Re-run test to verify result
Re-run test to verify result
Re-read test procedure
and follow directions
carefully
Re-run test to verify result
Dilute and re-run test
POOL EXACT® EZ ACCURACY
All tests have been calibrated using certified reference standards
and analytical spectrophotometric methods. The Pool eXact ®
EZ has been factory calibrated and will stay valid because of its
exceptional quality. We are so confident in the Pool eXact ® EZ,
we offer an industry leading 2-year warranty.
We built the Pool eXact ® EZ to be easy, accurate and
environmentally friendly. We have achieved this by utilizing our
patented eXact ® Strip Micro Technology, which uses 60% less
water and chemistry than alternative methods. Instead of using
a 10mL water sample, eXact ® Strip Micro uses a 4mL water
sample. The accuracy of the meter is maintained by designing
the sample cell with an 11mm path-length.
TECHNICAL SUPPORT
ITS Europe, Ltd.
The UK Centre for Homeland Security
Building 7, Chilmark, Salisbury,
Wiltshire SP3 5DU UK +44 1722 717911
itseurope@sensafe.com | itseurope.co.uk
l Test System
s,
tria
us
26
c.
In
Industrial Test Systems, Inc.
1875 Langston Street, Rock Hill, SC 29730
800-861-9712 or 803-329-9712
its@sensafe.com | sensafe.com
Ind
Please visit sensafe.com/photometers/pool-exact-ez/
for the latest technical information and how-to-videos.
For additional technical support, call (803) 329-9712.
th
5
Manufacturing quality products since 1989
©2016, Industrial Test Systems, Inc. Printed in USA
25
ABOUT
BUILT IN SAMPLE CELL
The built-in sample cell is made of transparent plastic;
the sturdy cell design will last for over 20,000 readings.
Our studies have shown that scratches on the cell will not
compromise the accuracy of your results because of the
cell’s fixed position.
WARRANTY (2 YEARS)
Registration of your Pool eXact® EZ photometer must be received within 30
days from date of purchase to activate the warranty. Registration is available
over the phone (+1-803-329-9712 Ext. 0) or online at sensafe.com/micro/
warranty/ (Personal data is kept confidential). The Pool eXact® EZ photometer
is warranted to be free from defects in materials and workmanship for a
period of two (2) years from the date of purchase by the customer. ITS will
repair or replace, at its discretion, product which is deemed to be faulty due
to manufacturing defect. Warranty does not cover product damage caused
by abuse (such as crushing a tablet in the cell), battery corrosion damage, or
improper use. If the meter is faulty or otherwise defective contact ITS by phone
(+1-803-329-9712 Ext. 0) or email (its@sensafe.com) to describe the problem
and obtain a return authorization form before returning the photometer to ITS.
Damage caused by improper packing of the photometer for return shipment to
ITS will not be covered by the warranty. Customer is responsible for shipping
charges to ITS. ITS pays postage when photometer is returned to customer. A
maximum processing fee of $75 will be charged for repair or replacement of
non-registered photometers and damages not covered by this warranty. The
repair or replacement of the photometer will not extend or renew the period of
guarantee. This warranty does not affect your statutory rights. The warranty is
not transferable.
METHOD VERIFICATION
Ready Snap® 3 is a calibration verification
solution with predetermined color to verify the
accuracy of your Pool eXact ® EZ photometer.
The easy 3 step procedure (snap, fill, and
test) allows for quick verification of test
parameters. Each box contains 10 ampoules
of 10mL solution with no dilution necessary.
POOL EXACT® EZ ASSIGNED VALUE FOR READY SNAP® 3
READY SNAP® 3
FREE CHLORINE
DESIRED VALUE
FREE CHLORINE
ACCEPTABLE RANGE
Red dye # 505
1.5 ppm
1.46 – 1.59
Note: Values indicated are specific to Pool eXact® EZ readings and current
concentrations as found at time of manufacture.
Sensafe.com
26
ABOUT
COMPLIANCE TESTING
This DPD test system for Chlorine is accepted
for reporting by most health departments
because the tests are USEPA (DIN Standard
38 408 G4/G5, ISO 7393/2) accepted for
testing requirements for Free Chlorine and Total Chlorine.
The compliance requirement is a photometer wavelength to
measure between 490 and 530nm. The Pool eXact ® EZ smart
photometer uses a 525nm wavelength and 11 mm path-length.
The eXact ® Strip Micro CL (DPD-1) use the same reagents and
proportions, and the resulting solution pH is maintained between
6.2 and 6.5 as specified by AWWA method 4500-Cl G/ClO2-D.
The USEPA does not “approve” commercial DPD delivery
systems. The eXact ® Strip Micro CL (DPD-1) for Free Chlorine,
and the eXact ® Strip Micro CL (DPD-3) for Combined Chlorine,
and the eXact ® Strip Micro CL (DPD-4) for Total Chlorine meet
your reportable testing requirements because the eXact ® Strip
Micro strips deliver the same chemicals in identical proportions.
Consult with your local health department for official regulations.
COMPONENT (FREE CHLORINE)
Anhydrous DPD sulfate
Anhydrous Na2HPO4
Anhydrous KH2PO4Na2
EDTA
AWWA 4500-CL G
POOL EXACT® EZ
1.5%
33.4%
64.0%
1.1%
1.5%
33.4%
64.0%
1.1%
TIP
Store all your necessary reagents together with your
Pool eXact ® EZ in our convenient carrying case!
CARRYING CASE FOR POOL EXACT® EZ
Made of sturdy material lined with
foam, the carrying case offers storage
for a Pool eXact ® EZ photometer and
eXact ® Micro reagents.
INCLUDES
ITEM
Standard
Carrying case
Blue carrying case with foam inserts
(holds up to 7 bottles)
Black carrying case with foam inserts
(holds up to 14 bottles)
©2016, Industrial Test Systems, Inc. Printed in USA
XL
Carrying case
PART NO.
486111
486001
NOTES
Sensafe.com
27
Pool
28
ADVANCED PHOTO M ETER SYSTEM
EXACT® MICRO TESTS & REAGENTS
PARAMETER / TEST 1
PART #
RANGE
ppm
Alkalinity, Total
486641
Alkalinity, Total Range Extender 2
486665
Chloride (Salt)
486757
80 – 6700
5
25
Chlorine, Free (DPD–1) 3
486637
0.01 – 12.0
5
100
Chlorine, Combined (DPD–3) 4
486638
0.01 – 12.0
5
100
Copper
486632
0.06 – 9
2
50
481652-II
3 – 110
12
60
Cyanuric Acid
Hardness, Calcium
pH
10 – 200
% BEST † # OF
ACCURACY TESTS
10
Each strip adds 130 ppm
100
100
486629
3 – 700
5
50
486639-II
6.4 – 8.4
0.2 pH
100
486814
0.02 – 4
8
50
Phosphate
EXACT REFILL BOX
PART #
Pool Refill Box
486211
PARAMETER / TESTS
Total Alkalinity, pH, Free Chlorine,
Combined Chlorine, Calcium
Hardness, Cyanuric Acid
†
Value provided represents best possible accuracy under laboratory conditions, but may vary throughout the detection
range. For a complete list of accuracies throughout all ranges, please visit sensafe.com/photometers/pool-exact-ez/.
1
Performance verified with various salt systems and water samples with optimal water temperature at 10-40˚C / 50-104˚F.
Optimal water temperature for Phosphate test is 15-40˚C / 59-104˚F.
2
For detailed instructions on using this product, contact us (see below) or visit sensafe.com/photometers/pool-exact-ez/.
3
Requires the use of 2 strips if reading is above 6 ppm.
4
Combined Chlorine DPD-3 Test requires Free Chlorine DPD-1 (486637) to be run first.
For resellers and distributors - products sold in case quantities (12 units per case).
PATENT INFORMATION
US Patent #7,333,194; Euro Pat No. 1 725 864 DE FR UK; South Africa Pat No 2007/0628
by Industrial Test Systems, Inc., 1875 Langston Street, Rock Hill, SC USA.
EXACT® is a registered trademark of Industrial Test Systems, Inc. Rock Hill, SC USA.
Visit us online:
sensafe.com/photometers/pool-exact-ez/
for up–to–date product information.
SENSAFE
MANUFACTURED BY
Industrial Test Systems, Inc.
1875 Langston Street
Rock Hill, SC 29730
1–800–861–9712—Inside the U.S.A.
1–803–329–9712—Outside the U.S.A.
Fax:1–803–329–9743
Email: ITS@SENSAFE.COM
WWW.SENSAFE.COM
ITSSENSAFE
ITS EUROPE, LTD
The UK Centre for Homeland Security
Building 7, Chilmark
Salisbury, Wiltshire, SP3 5DU, UK
Phone: +44 (0)1722 717911
Fax: +44 (0) 1722 717941
Email: ITSEUROPE@SENSAFE.COM
WWW.ITSEUROPE.CO.UK
©2016, Industrial Test Systems, Inc. Printed in USA
R072016-PEZQG
Download PDF
Similar pages