GE Healthcare
Dash 2500
The standard of excellence for sub-acuity monitoring
The Dash® 2500 monitor from GE Healthcare allows you
to deliver a new standard of clinical excellence to patients
in sub-acuity settings. Because it leverages the powerful
capabilities of the clinically-advanced family of Dash
monitors, there’s no need to sacrifice performance in the
interest of cost. The Dash 2500 monitor is a reliable, affordable
bedside monitor that gives you the clinical intelligence
you need to assess and treat your patients with speed,
accuracy and precision.
Rugged, ergonomic design
Clinically advanced
• Easily configurable display to suit your viewing preferences
and information requirements
The Dash 2500 monitor includes sophisticated clinical
parameters to capture vital patient measurements and
exceptional cardiac monitoring to help accurately detect
arrhythmias.
• GE EK-Pro arrhythmia program
™
• GE DINAMAP® SuperSTAT ™ non-invasive blood pressure
• Masimo® SET® or Nellcor® OxiMax® SpO2
• Alaris® Turbo Temp®
The Dash 2500 monitor is designed for demanding clinical
environments. Built with outstanding durability, it is also
totally portable and easy to use.
• Meets the same rigorous environmental requirements as
all Dash monitors
• Built with chemically-resistant LEXAN® plastic for
excellent resilience and enduring performance
• Compact size and long battery life allow patient transport
without data gaps or interruption
• Compatible with CARESCAPE™ CIC Pro for centralized viewing
Product specifications
Physical specifications
Masimo SpO2
Height22.0 cm (8.7 in)
Range Width
35.8 cm (14.1 in)
Depth
17.0 cm (6.7 in)
Weight
5.5 kg (12 lb)
Accuracy and motion tolerance
Without motion 25 to 240 bpm ± 3 digits
With motion normal physiologic range,
25 to 240 bpm ± 5 digits
Environmental
Operating temperature 5°C to 40°C (41°F to 104°F)
Storage temperature -20°C to 60°C (-4°F to 140°F)
Operating humidity 5 to 95% non-condensing
Storage humidity 5 to 95% non-condensing
Operating atmospheric pressure
700 to 1060 hPa
Storage atmospheric pressure
500 to 1060 hPa
Electrical
AC input voltage 100 to 240 V
AC input frequency 50/60 HZ
AC input power 120 VA
Internal battery 8.4 V nickel metal-hydride (NiMH)
Power supply The Dash 2500 Patient Monitor
can be powered from the internal
battery or AC power
Battery
Capacity 8.4 V; 7.0 amp-hr
Battery lifeGreater than 180 minutes using
fully charged internal battery (NIBP:
five min auto cycle with adult cuff.
ECG, RESP, SpO2: Active. TEMP:
predictive mode. Printer: printing
two waveforms for one min every
20 min at 25 mm/sec)
Charge time4 hours maximum with the
Monitor switched OFF
8 hours maximum with the
Monitor switched ON
Heart Rate/Pulse
25 to 240 bpm
Non-invasive blood pressure (NIBP)
Adult/pediatric range 30 to 200 bpm
Neonate range 30 to 220 bpm
Accuracy ± 3.5% or 3bpm, whichever is greater
ECG
Leads available
3-electrode configuration 5-electrode configuration I, II, III
I, II, III, aVR, aVL, aVF, and VA
Heart rate accuracy 30 to 300 bpm, ± 3 bpm or 3% of
reading, whichever is greater
Heart rate resolution 1 bpm
Bandwidth
0.5 to 40 Hz +1/-6 dB
0.05 to 40 Hz +1/-6 dB
0.05 to 100 Hz +1/-6 dB
Standardizing voltage 1 mV marker
Common mode rejection 1 mV RTI or 10 mm p-p max
displayed noise allowed with 20
Vrms, 50-60 Hz input
Input impedance
Common mode Differential > 2.5 MΩ at 10 Hz
> 2.5 MΩ from DC to 60 Hz
60 Hz tolerance up to 10 mV
Pacemaker detection/rejection
Input voltage range ± 2 to ± 700 mV
Lead off sensing current < 0.1 μA DC signal leads,
< 1 μA DC driven lead
Respiratory
ECG-Derived respiration rate
Electrocardiography (ECG)
Leads available Heart rate accuracy 30 to 300 bpm, ± 3 bpm or 3%
of reading, whichever is greater
Range6 to 120 breaths/min (adult/pediatric)
6 to 180 breaths/min (neonate)
Nellcor SpO2
Accuracy ± 2 breaths/min or ± 3% of reading,
whichever is greater
Range
20 to 250 bpm
Accuracy and tolerance 20 to 250 bpm ± 3 digits
Low perfusion 20 to 250 bpm ± 3 digits
I or II
Resolution 1 breath/min
Base impedance 100 to 2000 Ω
Detection sensitivity 0.2 Ω at 30 breath/min with
500 Ω baseline impedance
NIBP
Masimo SET SpO2
Method Oscillometric with step deflation
Measurement range
Modes Manual, automatic, stat
SpO2 1 to 100%
BP Measurement ranges
Pulse rate 25 to 240 bpm
Systolic
30 to 290 mmHg (adult/pediatric)
4.0 to 38.7 kPa (adult/pediatric)
30 to 140 mmHg (neonate)
4.0 to 18.7 kPa (neonate)
Accuracy and motion tolerance
MAP
20 to 260 mmHg (adult/pediatric)
2.7 to 34.7 kPa (adult/pediatric)
20 to 125 mmHg (neonate)
2.7 to 16.7 kPa (neonate)
Diastolic
10 to 220 mmHg (adult/pediatric)
1.3 to 29.3 kPa (adult/pediatric)
10 to 110 mmHg (neonate)
1.3 to 14.7 kPa (neonate)
Resolution
1 mmHg
Accuracy Meets AAMI/ANSI standard
SP10:2002
Initial cuff inflation pressure
Maximum determination time
135 ± 15 mmHg default;
user selectable (adult/pediatric)
100 ± 15 mmHg default;
user selectable (neonate)
120s (adult/pediatric)
85s (neonate)
Over pressure monitor
300 to 330 mmHg (adult/pediatric)
150 to 165 mmHg (neonate)
Hose/cuff interface
Compatible with current Dash hoses
Pulse rate When NIBP is the source, HR values
are derived from the pulse rate that
is determined by the oscillometric
technique of measuring blood
pressure. The rate source field is
labeled NIBP.
Adult/pediatric range 30 to 200 bpm (± 3.5% or 3 bpm)
Neonate range 30 to 220 bpm (± 3.5% or 3 bpm)
Nellcor OxiMax SpO2
Measurement range
SpO2 1 to 100%
Pulse rate 20 to 250 bpm
Accuracy
Saturation
Adult Neonate Low perfusion 70 to 100% ±2 digits
70 to 100% ±3 digits
70 to 100% ±2 digits
Pulse Rate
Adult and neonate 20 to 250 bpm ±3 digits
Low perfusion 20 to 250 bpm ±3 digits
Saturation
Without motion–
70 to 100% ±2 digits
adult/pediatric
Without motion–
70 to 100% ±3 digits
neonate
With motion–
70 to 100% ±3 digits
adult/pediatric/neo
Low perfusion 70 to 100% ±2 digits,
0 to 69% unspecified
Pulse Rate
Without motion 25 to 240 bpm ±3 digits
With motion normal physiologic range
25 to 240 bpm ±5 digits
Alaris Turbo Temp
Scale
°Fahrenheit (F)
°Celsius (C)
Predictive mode
Range 35.6°C to 41.1°C (96.0°F to 106.0°F)
Resolution 0.1°C (0.1°F)
Monitor mode
Range 26.7°C to 42.2°C (80.0°F to 108.0° F)
Accuracy ± 0.1°C (± 0.2°F) (when tested in a
calibrated liquid bath; meets ASTM
E1112, Table 1, in range specified)
Resolution 0.1°C (0.1°F)
Probes Use only Alaris Turbo Temp probes
and probe covers. The size, shape,
and thermal characteristics of the
probe covers can affect the performance of the instrument. Inaccurate
readings or retention problems may
occur unless Alaris Turbo Temp
probes and probe covers are used.
Determination time Approximately 10 seconds, typical
© 2008 General Electric Company – All rights reserved.
General Electric Company reserves the right to make changes in
specifcations and features shown herein, or discontinue the product
described at any time without notice or obligation. Contact your GE
Representative for the most current information.
GE and GE Monogram are trademarks of General Electric company.
Masimo and SET are trademarks of Masimo Corporation.
Nellcor and OxiMax are trademarks of Nellcor Puritan Bennett, Inc.
Alaris and Turbo Temp are trademarks of Cardinal Health, Inc.
LEXAN is a trademark of SABIC Innovative Plastics IP BV.
Dash, DINAMAP, EK-PRO and SuperSTAT are trademarks
of GE Medical Systems, Information Technologies Inc.
GE Medical Systems Information Technologies, Inc.,
a General Electric company, doing business as GE Healthcare.
Healthcare Re-imagined
GE is dedicated to helping you transform healthcare
delivery by driving critical breakthroughs in biology
and technology. Our expertise in medical imaging
and information technologies, medical diagnostics,
patient monitoring systems, drug discovery, and
biopharmaceutical manufacturing technologies is
enabling healthcare professionals around the world
to discover new ways to predict, diagnose and treat
disease earlier. We call this model of care “Early Health.”
The goal: to help clinicians detect disease earlier,
access more information and intervene earlier with
more targeted treatments, so they can help their
patients live their lives to the fullest. Re-think,
Re-discover, Re-invent, Re-imagine.
GE Healthcare
8200 West Tower Ave.
Milwaukee, WI 53223
U.S.A.
www.gehealthcare.com
GE Healthcare Oy
Kuortaneenkatu 2
00510 Helsinki
Finland
GE China Technology Park
No.1 Huatuo Road
Shanghai PRC
201203
Tel: 86-21-3877-7888
Fax: 86-21-3877-7451