PAT850-9.4T (200 V input) - kikusui electronics corp.

PAT850-9.4T (200 V input) specification_KIKUSUI
Unless specified otherwise, the specifications are for the following settings and conditions.
The load is a pure resistance.
The warm-up time is 30 minutes (with current flowing).
After warm-up is complete, the PAT-T series must be calibrated correctly according to the procedures given in the operation
manual in a 23 °C ± 5 °C environment.
TYP (typical) values do not guarantee the performance.
rtg: Indicates the rated output.
rdng: Indicates the reading.
Rated load and no load are defined as follows.
During constant voltage operation (set the output current at the rated voltage output greater than equal to the rated
output current)
Rated load: Refers to a load with a resistance that makes the current that flows when the rated output voltage is applied to
be 95 % to 100 % of the rated output current at the rated output voltage.
No load: Refers to a load through which no output current flows or an open output terminal condition with no load connected.
During constant current operation (set the output voltage at the rated output current greater than equal to the rated
output voltage)
Rated load: Refers to a load with a resistance that makes the voltage drop when the rated output current is supplied to be 95
% to 100 % of the maximum output voltage at the rated output current.
The output voltage of the PAT-T series including the voltage drop in the load cable must not exceed the maximum output
voltage at the rated output current.
No load: Refers to a load with a resistance that makes the voltage drop when the rated output current is supplied to be 10 %
of the maximum output voltage or 1 V, whichever is greater, at the rated output current.
AC input
Nominal input rating
200 V to 240 V, 50 Hz to 60 Hz, three-phase
Input voltage range
180 V to 250 V
Input frequency range
47 Hz to 63 Hz
Hold-up time for power interruption (MIN)
20 ms or greater (at 50 % load),
10 ms or greater (at the rated load)
Current (MAX) *1
32 A
Inrush current (MAX) *2
100 Apeak
Wattage (MAX) *1
10 kVA
Power factor (TYP) *3
0.95
Efficiency (MIN) *3
85 %
*1. At the rated load.
*2. Excludes the charge current component that flows through the capacitor of the internal EMC filter circuit immediately after the POWER
switch is turned on (within approximately 1 ms).
*3. Input voltage of 200 Vac, at the rated load.
1/7
PAT850-9.4T (200 V input) specification_KIKUSUI
Output
Rating
Output power
8 kW
Output voltage
850.0 V
Output current
Voltage
9.40 A
Preset range
0 % to 105 % of rtg *1
Setting accuracy
± (0.2 % of rtg +50 mV) *2
Line regulation
±(0.05 % of rtg + 5 mV) *3
Load regulation
±(0.1 % of rtg + 5 mV) *4
Transient response
Ripple noise
Rise time (MAX)
Fall time (MAX)
5 ms *5
(p-p)
600 mV *6
(rms)
100 mV *7
Rated load
100 ms *8
No load
100 ms *8
Rated load
200 ms *9
No load
4000 ms *9
Temperature coefficient (MAX)
Current
100 ppm/°C (during external control) *10
Preset range
0 % to 105 % of rtg *1
Setting accuracy
±(1.0 % of rtg + 100 mA) *2
Line regulation
±(0.1 % of rtg + 30 mA) *3
Load regulation
±(0.2 % of rtg +30 mA) *4
Ripple noise (rms)
120 mA *7
Temperature coefficient (TYP)
200 ppm/°C (during external control) *10
*1. The maximum preset voltage and current are used to determine constant voltage or constant current operation when the activation
point of the constant voltage or constant current operation is set to the rated output voltage or current. It does not guarantee power
supply to the load exceeding the rated output voltage or current.
To establish a constant voltage operation at the activation point (rated output voltage or current), set the output current (I Set) so that
rated output current < I Set ≤ maximum preset current. Likewise, to establish constant current operation, set the output voltage (V
Set) so that rate output voltage < V Set ≤ maximum preset voltage.
*2. The difference between the actual output voltage or output current and the preset value under constant voltage or constant current
operation.
*3. Output voltage or output current fluctuation with respect to ±10 % fluctuation of the nominal input voltage under constant voltage or
current operation.
*4. Output voltage or output current fluctuation when the output voltage or output current is set to the rated output voltage or current and
the load is changed from rated load to no load under constant voltage or current operation.
*5. The time it takes for the output voltage fluctuation to recover from outside 0.1 % + 10 mV of the output voltage setting to within 0.1 %
+ 10 mV when the output current is changed from 100 % to 50 % or 50 % to 100 % of the rated output current under constant voltage
operation. The output voltage when the output current is 100 % is used as a reference. When the remote sensing function is used.
*6. When the measurement frequency bandwidth is 10 Hz to 20 MHz.
*7. Output voltage is 10 % to 100 % of the rating when the measurement frequency bandwidth is 5 Hz to 1 MHz.
*8. The time it takes for the output voltage to rise from 10 % to 90 % of the rating when the output is turned on.
*9. The time it takes for the output voltage to fall from 90 % to 10 % of the rating when the output is turned off.
*10. When the ambient temperature is within the range of 0 °C to 50 °C.
2/7
PAT850-9.4T (200 V input) specification_KIKUSUI
Display function
Voltmeter
Ammeter
Operation display
Maximum display
999.9 (fixed decimal point)
Display accuracy
± (0.2 % of rdng +5 digits) at 23 °C ± 5 °C
Maximum display
99.99 (fixed decimal point)
Display accuracy
± (0.5 % of rdng +5 digits) at 23 °C ± 5 °C
OUTPUT ON/OFF
ON: OUTPUT LED illuminates (Green LED).
OFF: OUTPUT LED turns off.
ALM operation
ALARM LED illuminates (Red LED). *1
CV operation
CV LED illuminates (Green LED).
CC operation
RMT operation
EXT operation
LOCK operation
CC LED illuminates (Red LED).
RMT LED illuminates during remote control (Green LED).
EXT LED illuminates during external control (Green LED).
LOCK LED illuminates when the keys are locked (Green LED).
*1. Illuminates when the overvoltage protection (OVP), overcurrent protection (OCP), overheat protection (OHP), input open-phase
protection (PHASE), fan failure protection (FAN), incorrect sensing connection protection (SENSE), overheat protection of the
bleeder circuit (BOHP) and so on are activated.
Illuminates even when the breaker trips (even when the POWER switch is turned off). In this case, the LED illuminates for
approximately 10 to 15 seconds.
Protection functions
Overvoltage protection (OVP)
Turns off the output or trips the breaker (turns the POWER switch off).
ALARM LED illuminates.*1
Selectable range
10 % to 111.5 % of the rated output voltage
Setting accuracy
Overcurrent protection (OCP)
± 2 % of rtg
Turns off the output or trips the breaker (turns the POWER switch off).
ALARM LED illuminates.*1
Selectable range
10 % to 111.5 % of the rated output current
Setting accuracy
± 3 % of rtg
Overheat protection (OHP)
Turns the output off. ALARM LED illuminates.
Input open-phase protection (PHASE)
Turns the output off. ALARM LED illuminates.
Fan failure protection (FAN)
Turns the output off. ALARM LED illuminates.
Incorrect sensing connection protection (SENSE)
Turns the output off. ALARM LED illuminates.
Overheat protection of the bleeder circuit (BOHP)
Shut down (SD)
Turns the output off. ALARM LED illuminates.
Turns off the output or trips the breaker (turns the POWER switch off).
ALARM LED illuminates.*1
*1. Illuminates even when the breaker trips (even when the POWER switch is turned off). In this case, the LED illuminates for
approximately 10 to 15 seconds.
3/7
PAT850-9.4T (200 V input) specification_KIKUSUI
Output signals
Monitor signal
output*1
VMON
(Voltage)
IMON
(Current)
At rated voltage
output
10.00 V ± 0.25 V
At 0 V output
0.00 V ± 0.25 V
At rated current
output
10.00 V ± 0.25 V
At 0 A output
Status signal output
*1, *2
0.00 V ± 0.25 V
OUTON STATUS
Turns on when the output is on.
CV STATUS
Turns on during CV operation.
CC STATUS
Turns on during CC operation.
ALM STATUS
Turns on when an alarm (OVP, OCP, OHP, BOHP, input open-phase protection, fan
failure protection, incorrect sensing connection protection, or shutdown) is detected.
PWR OFF STATUS
Stays on for approximately 10 to 15 seconds after the POWER switch turns off.
PWR ON STATUS
Turns on when the POWER switch is on.
*1. J1 connector on the rear panel.
*2. Photocoupler open collector output, maximum voltage 30 V, maximum current (sink) 8 mA, insulated from the output and control
circuits, and status signals are not mutually insulated.
Control functions
External
control*1
EXT-V CV CONT *2 *3
(CV external voltage control)
0 % to 100 % of the rated output voltage in the range of 0 V to 10 V.
EXT-V (FAST) CV CONT *2 *3
(CV external voltage control FAST)
0 % to 100 % of the rated output voltage in the range of 0 V to 10 V.
EXT-R CV CONT *2
(CV external resistance control)
0 % to 100 % of the rated output voltage in the range of 0 kΩ to 10 kΩ.
EXT-R (FAIL SAFE) CV CONT *2
(CV external resistance control FAIL
SAFE)
100 % to 0 % of the rated output voltage in the range of 0 kΩ to 10 kΩ.
EXT-V CC CONT *2 *3
(CC external voltage control)
0 % to 100 % of the rated output current in the range of 0 V to 10 V.
EXT-V (FAST) CC CONT *2 *3
(CC external voltage control FAST)
0 % to 100 % of the rated output current in the range of 0 V to 10 V.
EXT-R CC CONT *2
(CC external resistance control)
0 % to 100 % of the rated output current in the range of 0 kΩ to 10 kΩ.
EXT-R (FAIL SAFE) CC CONT *2
(CC external resistance control FAIL
SAFE)
100 % to 0 % of the rated output current in the range of 0 kΩ to 10 kΩ.
OUTPUT ON/OFF CONT*4
Output on with a low TTL level signal/output on with a high TTL level signal.
SHUT DOWN*5
POWER switch off with a low TTL level signal.
*1. J1 connector on the rear panel.
*2. Set EXT-V, EXT-V (FAST), EXT-R, and EXT-R (FAIL SAFE) in the CONFIG settings. The selected function is enabled.
*3. The input impedance EXT-V CV CONT, EXT-V (FAST) CV CONT and EXT-V CC CONT, EXT-V (FAST) CC CONT is approximately
100 kΩ.
The setting accuracy is ±5 % of the rated output voltage or ±5 % of the maximum output current.
*4. Set the logic low/high using CONFIG settings.
*5. The output turns off even if the breaker trip setting of the CONFIG parameter is set so that the POWER switch does not turn off.
4/7
PAT850-9.4T (200 V input) specification_KIKUSUI
Interface
Common specifications
Software protocol
Command language
RS232C
IEEE Std 488.2-1992
Complies with the SCPI Specification 1999.0 specifications.
Hardware
Complies with EIA232D.
D-SUB 9-pin connector (male) *1
Baud rate: 1200, 2400, 4800, 9600, 19200, and 38400 bps
Data length: 7 bits or 8 bits. Stop bit: 1 bit or 2 bits. No parity.
Flow control X-Flow or none.
Program message terminator
GPIB*2
LF during reception, CR/LF during transmission.
Hardware
Complies with IEEE Std 488.1-1987.
SH1, AH1, T6, L4, SR1, RL1, PP0, DC1, DT1, C0, and E1.
24 pin connector (receptacle)
Program message terminator
LF or EOI during reception, LF+EOI during transmission.
Primary address
USB*2
Hardware
1 to 30
Complies with USB 2.0. Data rate: 12 Mbps (full speed).
Socket B type
Program message terminator
Device class
LAN*2
LF or EOM during reception, LF+EOM during transmission.
Complies with the USBTMC-USB488 device class specifications.
Hardware
IEEE 802.3 100Base-TX/10Base-T Ethernet
Complies with the LXI Class C, Specification 1.2.
IPv4, RJ-45 connector *3
Communications protocol
Program message terminator
VXI-11
LF or END during reception, LF+END during transmission.
*1. Use a cross cable (null modem cable).
*2. Factory option.
*3. Category 5, use a straight cable.
Other functions
Master-Slave parallel operation
Output status setting at power-on setting
Output on / off delay setting
Preset memory
Lock
Fine adjustment
Voltage / current setting limit
Communication error display
5/7
PAT850-9.4T (200 V input) specification_KIKUSUI
General specifications
Weight
Approx. 23 kg (50.71 lb)
Dimensions
Environmental
conditions
See the outline drawing
Operating conditions
Indoor use, Overvoltage Category II
Operating temperature
0 °C to +50 °C (32 °F to +122 °F)
Operating humidity
20 %rh to 85 %rh (no condensation)
Storage temperature
-25 °C to +70 °C (-13 °F to +158 °F)
Storage humidity
90 %rh or less (no condensation)
Altitude
Up to 2000 m
Cooling system
Forced air cooling using a fan. (With fan control)
Grounding polarity
Negative grounding or positive grounding possible.
Isolation voltage
Withstand voltage
± 1000 Vmax
Across the input terminals and
chassis
No abnormalities at 1500 Vac for 1 minute.
Across the input terminals and the
output terminals
No abnormalities at 1500 Vac for 1 minute.
Across the output terminals and
chassis
No abnormalities at 1000 Vdc for 1 minute.
Insulation resistance Across the input terminals and
chassis
Across the input terminals and the
output terminals
1000 Vdc, 30 MΩ or more. (at 70 %rh or less)
Across the output terminals and
chassis
1000 Vdc, 30 MΩ or more. (at 70 %rh or less)
Safety
Accessories
1000 Vdc, 30 MΩ or more. (at 70 %rh or less)
Complies with the requirements of the following standard.
IEC61010-1:2001 (Class I *1, Pollution degree 2 *2)
Output terminal cover
2 sets (with screws)
Output terminal bolt set
2 sets (M8 × 20 mm bolts, nuts, and spring washers)
Heavy object warning label
J1/J2 connector kit
1 pc.
1 set (2 sets of protection covers, 2 sokets, and 30 pins)
Chassis connection wire
1 set (with screws)
Quick Reference (base model)
1 pc. (english), 1 pc. (japanese)
Setup Guide
1 copy
Safety Information
1 copy
CD-ROM
1 pc.
*1. This is a Class I equipment. Be sure to ground the PAT-T series’s protective conductor terminal. The safety of this product is only
guaranteed when the product is properly grounded.
*2. Pollution is addition of foreign matter (solid, liquid or gaseous) that may produce a reduction of dielectric strength or surface resistivity.
Pollution Degree 2 assumes that only non-conductive pollution will occur except for an occasional temporary conductivity caused by
condensation.
6/7
PAT850-9.4T (200 V input) specification_KIKUSUI
Dimensions
Unit: mm (inch)
PAT-T Series Outline Drawing
© KIKUSUI Electronics Corp.
7/7