Dissolved oxygen monitor : BAMOPHAR 451 LOG

Dissolved oxygen monitor
BAMOPHAR 451 LOG
USER MANUAL
INTERNATIONAL
22, Rue de la Voie des Bans · Z.I. de la gare · 95100 ARGENTEUIL
Tel +33 (0)1 30 25 83 20 Web
www.bamo.eu
Fax +33 (0)1 34 10 16 05 E-mail export@bamo.fr
Dissolved oxygen monitor
BAMOPHAR 451 LOG
09-03-2018
M-451.03-EN-AB
OXY
451-03 /1
SUMMARY
1.
2.
3.
4.
4.1
4.1.1
5.
5.1
5.2
6.
7.
7.1
7.2
7.3
7.4
8.
9.
9.1
9.2
9.2.1
9.2.2
9.2.3
9.3
9.4
9.5
9.6
9.7
9.8
9.9
9.10
9.11
9.12
9.13
DESCRIPTION
TECHNICAL FEATURES
DIMENSIONS
ELECTRICAL CONNECTIONS
VIEWS OF TERMINALS
Connection of AQUAPLUS probe
MAIN DISPLAY
WITHOUT EXTENSION UNIT
WITH AN EXTENSION UNIT
DESCRIPTION OF ICONS
DISPLAY SETTINGS
SCREEN INFORMATION
SCREEN SAVER
LANGUAGE SELECTION
DESIGNATION
CONSULTATION / MODIFICATION
DISSOLVED OXYGEN SETTINGS
AQUAPLUS PROBE CALIBRATION
ADJUSTMENT OF THRESHOLDS S1, S2 AND S3 IN MODE ON/OFF REGULATION
REGULATION WITH BAMOPHAR, PRELIMINARIES
SETTING OF THRESHOLDS S1, S2 AND S3 IN MODE ON/OFF REGULATION
ASSIGNMENT OF S3 THRESHOLD ON AN EXTERNAL SIGNAL
SETTINGs OF OPERATING MEASUREMENT
SETTINGS OF ANALOGUE OUTPUT FOR MEASUREMENT
SETTINGS OF ANALOGUE OUTPUT FOR TEMPERATURE
FORCED RELAY
SETTING OF THE RELAY REGULATION MODE
SETTING ALARMS
SETTINGS OF CLEANING ROUTINE
SETTING A PID REGULATION
FORCED MEASUREMENT - MEASUREMENT SIMULATION
ATMOSPHERIQUE PRESSURE
DATA LOGGER
INTERNATIONAL
22, Rue de la Voie des Bans · Z.I. de la gare · 95100 ARGENTEUIL
Tel +33 (0)1 30 25 83 20 Web
www.bamo.eu
Fax +33 (0)1 34 10 16 05 E-mail export@bamo.fr
3
3
3
4
5
6
7
7
7
8
9
9
9
9
9
9
10
10
10
10
11
11
11
12
12
12
12
13
13
13
14
14
14
Dissolved oxygen monitor
BAMOPHAR 451 LOG
09-03-2018
M-451.03-EN-AB
OXY
451-03 /2
1.
DESCRIPTION
BAMOPHAR 451 offers a reliable and accurate measurement of dissolved oxygen (DO) concentration in water treatment process.
It has a color touch screen for the display of a multilingual menu friendly and intuitive.
The unit also displays the menu for setting the analog output, setting the thresholds and setting the operating mode.
In order to facilitate its commissioning, a programming menu can simulate the measurement, acting on analogue outputs and thresholds
2.
TECHNICAL FEATURES
End-user interface
Measuring scales
Accuracy
Sensor signal input
Relay outputs
Configurable thresholds
Contact initial resistance
Switching power
Switching capacity (min.)
Impulse regulation
PID Regulation
Calibration routine
Measurement output
Temperature output / PID
Programme Testing
Main power supply
Models
Data Logger
CE Conformity
3.
Color touch screen 4.3'', resolution 480x272 pixels
Display of measurements (D.O. concentration and temperature) and relay status
Configuration - Keyword protected
0... 500 % or 0.01 to 50 mg/l
According to probe AQUAPLUS features,
0to 20 mg/l: ±1 % of reading,
20 to 50 mg/l: ± 10 % of reading
Screw terminals - digital signal
4 contacts N.O., voltage free
S1, independent threshold, to set up for measurement or temperature
S2, independent threshold, to set up for measurement or temperature
S3, independent threshold, to set up for measurement or temperature or an external function
S4, threshold, to set up for alarming function, out of range, temperature sensor defect,
or probe cleaning mode
100 mΩ max. (voltage drop 6 V DC 1 A)
831 VA AC / 3 A / 277 V AC
90 W / 3 A / 30 V DC
100 mA, 5 V DC (variable according to switching frequency, environmental conditions, accuracy)
Cycle time setting from 0 to 9999 s, high and low proportional bandwidths, dead zones
Adjustable proportionality from 0 to 200%, Integrant and Derivative: 0 to 999 s
Relay outputs inhibited; Analogue outputs on standby at latest values
0/4 - 20 mA (max. 600 Ω) proportional to measurement
0/4-20 mA (max. 600 Ω), scaling 0...100°C
This temperature output is not available when PID regulation function is operating.
Simulation through the menu on measurement, temperature, PID and thresholds
230 V AC / 50-60 Hz (others on request) - Consumption 10 VA
Panel mounting, 96x144 mm; screw terminals, Front IP65, rear IP40
Wall mounting, IP65, screw terminals, cable glands
Record of cycle average measurement (programmable)
150 000 records max. on memory card.
The instrument meets the legal requirements of the current European Directives
DIMENSIONS
INTERNATIONAL
22, Rue de la Voie des Bans · Z.I. de la gare · 95100 ARGENTEUIL
Tel +33 (0)1 30 25 83 20 Web
www.bamo.eu
Fax +33 (0)1 34 10 16 05 E-mail export@bamo.fr
Dissolved oxygen monitor
BAMOPHAR 451 LOG
09-03-2018
M-451.03-EN-AB
OXY
451-03 /3
4.
ELECTRICAL CONNECTIONS
PANEL MOUNT VERSION
Measurement signal input
WALL MOUNT VERSION
Measurement signal input
INTERNATIONAL
22, Rue de la Voie des Bans · Z.I. de la gare · 95100 ARGENTEUIL
Tel +33 (0)1 30 25 83 20 Web
www.bamo.eu
Fax +33 (0)1 34 10 16 05 E-mail export@bamo.fr
Dissolved oxygen monitor
BAMOPHAR 451 LOG
09-03-2018
M-451.03-EN-AB
OXY
451-03 /4
4.1
VIEWS OF TERMINALS
Description
Terminal
1
Output mA, measurement
2
3
Output mA, temperature or PID regulation
4
5
Input for blocking the regulation
External sensor
6
7
Power supply to sensor
Power to AQUAPLUS probe
8
9
10
Input for Pt 100 sensor (3-wire)
11
12
13
14
Connections for an extension terminal (blind unit)
15
16
17
Limit value S4 (N.O. contact) for Fault on Regulation, Temperature or for Cleaning routine
18
19
Limit value S2 (N.O. contact) for measurement/ temperature
20
Limit value S1 (N.O. contact) for measurement/ temperature
21
22
Limit value S3 (N.O. contact) for measurement/ temperature
23
Can be assigned to external function: becomes N.C. contact (see 9.2.3)
24
Not used
25
26
Main power supply
27
28
31
32
Input signal from AQUAPLUS probe
33
34
Connection
+ mA
- mA
- mA / PID
+ mA / PID
+ 24 V
- 24 V
- 24 V
+ 24 V
+
+
Shielding
Y
Z
B
A
S4
S2
S1
S3
PE = Ground (equipotential)
L = Neutral
L = Live
Raccordement au boîtier d'extension
Main unit
Cable max. length
Cable type
Extension unit
500 m
Network cable or 4-wire shielded cable, ≥ 0,25 mm²
INTERNATIONAL
22, Rue de la Voie des Bans · Z.I. de la gare · 95100 ARGENTEUIL
Tel +33 (0)1 30 25 83 20 Web
www.bamo.eu
Fax +33 (0)1 34 10 16 05 E-mail export@bamo.fr
Dissolved oxygen monitor
BAMOPHAR 451 LOG
09-03-2018
M-451.03-EN-AB
OXY
451-03 /5
4.1.1 Connection of AQUAPLUS probe
Terminal on
AQUAPLUS probe
Power
24 V
BAMOPHAR
Cable of 4 shielded pairs
(8-wire)
Connections in head housing on top of PVC holder
INTERNATIONAL
22, Rue de la Voie des Bans · Z.I. de la gare · 95100 ARGENTEUIL
Tel +33 (0)1 30 25 83 20 Web
www.bamo.eu
Fax +33 (0)1 34 10 16 05 E-mail export@bamo.fr
Dissolved oxygen monitor
BAMOPHAR 451 LOG
09-03-2018
M-451.03-EN-AB
OXY
451-03 /6
5.
MAIN DISPLAY
5.1
WITHOUT EXTENSION UNIT
Switching status of relays S1, S2, S3, S4
(Open or closed)
Measurements display
mg/l or % and T°
Designation - Identification of your
instrument
Display settings
Parameter settings of
BAMOPHAR
5.2
Measurements details
Date and time
WITH AN EXTENSION UNIT
Switching status of relays S1, S2, S3, S4
(Open or closed)
Measurements display
OD & Temperature + Parameter of extension unit (Example)
Designations - Identifications of your instruments
Display settings
Parameter settings of
both BAMOPHAR
INTERNATIONAL
22, Rue de la Voie des Bans · Z.I. de la gare · 95100 ARGENTEUIL
Tel +33 (0)1 30 25 83 20 Web
www.bamo.eu
Fax +33 (0)1 34 10 16 05 E-mail export@bamo.fr
Main unit A is shown in yellow
Extension unit B is shown in green
Measurements details
Date and time
Dissolved oxygen monitor
BAMOPHAR 451 LOG
09-03-2018
M-451.03-EN-AB
OXY
451-03 /7
6.
DESCRIPTION OF ICONS
The touch screen allows you to navigate within the menus and set up the display according your application.
To each yellow icon dedicated to main unit A corresponds a green icon dedicated to the extension unit B.
HOME
Return to main display
e.g. Green icon for extension unit
DISPLAY SETTINGS
Access to parameters of display (language and identification of the monitor)
LANGUAGES
Language selection
MENU
Access to parameter settings of the monitor
INFORMATION
Access to serial number and version of your BAMOPHAR
PADLOCK
Open: means MODIFICATION mode
Closed: means CONSULTATION (review) mode
RETURN
To go back to the previous screen
ARROWS
Cursors to navigate within the menus
SELECTION
Scroll through the selection list
CONFIRMATION
Confirm setting and go to the next parameter
SAVE
To save parameters of current setting
INTERNATIONAL
22, Rue de la Voie des Bans · Z.I. de la gare · 95100 ARGENTEUIL
Tel +33 (0)1 30 25 83 20 Web
www.bamo.eu
Fax +33 (0)1 34 10 16 05 E-mail export@bamo.fr
Dissolved oxygen monitor
BAMOPHAR 451 LOG
09-03-2018
M-451.03-EN-AB
OXY
451-03 /8
7.
DISPLAY SETTINGS
Note : To access to the display settings, BAMOPHAR must be in MODIFICATION mode (See § CONSULTATION/MODIFICATION)
7.1
SCREEN INFORMATION
Screen identification number and its version are available in this menu.
7.2
SCREEN SAVER
Screen saver brightness is adjusted by moving the cursor: decreased (to the left) or increased (to the right).
7.3
LANGUAGE SELECTION
Select the flag according to your choice.
The display automatically returns to previous view.
Press icon "HOME" to go back to the main display.
7.4
DESIGNATION
it is recommended to name each channel (main unit A and Extension unit B) for an easiest reading:
1) Press on icon A or B according to the instrument you wish to rename.
2) A keyboard appears, enter the new name.
3) Save the new designation.
8.
CONSULTATION / MODIFICATION
CONSULTATION mode let the operator to check out all working parameters.
This mode is represented by the closed padlock icon.
To change the settings of the BAMOPHAR, you need to enter in the MODIFICATION mode.
This mode is protected by a password identical to the last 4 digits of the serial number.
On the main display, press MENU icon.
Press the padlock icon and type the 4 last digits of Serial Number.
To validate it, press "OK"; Device is now in MODIFICATION mode (Open padlock)
If the entered keyword is wrong, an error message appears for 3 seconds.
After 30 minutes the mode MODIFICATION switches back to CONSULTATION mode.
Where to find the serial number?
The serial number (SN) is written on the identification label of the BAMOPHAR.
It appears as well in MENU, icon " i " (INFORMATION).
INTERNATIONAL
22, Rue de la Voie des Bans · Z.I. de la gare · 95100 ARGENTEUIL
Tel +33 (0)1 30 25 83 20 Web
www.bamo.eu
Fax +33 (0)1 34 10 16 05 E-mail export@bamo.fr
Dissolved oxygen monitor
BAMOPHAR 451 LOG
09-03-2018
M-451.03-EN-AB
OXY
451-03 /9
9.
DISSOLVED OXYGEN SETTINGS
9.1
AQUAPLUS PROBE CALIBRATION
Press MENU and choose ADJUST PROBE
(This program step is accessible only in MODIFICATION mode)
This function is used to calibrate the AQUAPLUS probe.
In order not to disturb the process, the measurement and regulation are frozen on BAMOPHAR when entering the calibration menu.
This blocking is effective during the whole time of the calibration, plus an adjustable delay entered at the end of the menu (DELAY).
This delay allows to install back all components and commissiononing of complete system.
CAUTION: Check and modify if necessary, the atmospheric pressure set in BAMOPHAR if it is different from the current
pressure.
To modify it, go to menu AIR PRESSURE
ZERO ADJUST YES
CAL BUFFER ZERO
OXY. +000.0
CALIBRATION ON
PROBE IN AIR
OXY. +000.0
CALIBRATION ON
DELAY 0015 Sec
9.2
Confirm YES (after replacemnt of the sensor end)
Note : ZERO of dissolved oxygen is set in factory; However a new calibration is necessary when
replacing the sensor end.
Confirm to access to next parameter.
Plunge the bottom of AQUAPLUS probe in a calibration buffer solution (sodium sulfite).
Confirm to access to next parameter.
Wait until display keeps steady on 000.0%
Confirm to access to next parameter.
Probe and BAMOPHAR exchange data together.
Wait until appears new display CALIBRATION OK
Zero of DO concentration is done.
Confirm to access to next parameter.
Rinse the probe, wipe out water drops on the sensor end; Then, surround AQUAPLUS probe with a
clean, damp cloth.
Confirm to access to next parameter.
Wait until display keeps steady on 100.0%
This step must delay at least more than 20 seconds.
Confirm to access to next parameter.
Probe and BAMOPHAR exchange data together.
Wait until display CALIBRATION OK appears
100,0 % of oxygen in the atmosphere is done.
Confirm to access to next parameter.
Enter the delay in seconds, during which measurements, outputs and regulationctions, keep on stand-by
on values existing before the calibration; Then confirm.
To record the settings, press the icon SAVE.
ADJUSTMENT OF THRESHOLDS S1, S2 AND S3 IN MODE ON/OFF REGULATION
9.2.1 REGULATION WITH BAMOPHAR, PRELIMINARIES
Before setting the thresholds or regulation mode, the operating mode must be set.
BAMOPHAR 451 LOG allows 3 operating modes:
1) On/OFF regulation through relays, thresholds S1, S2 and/ or S3
2) Impulse regulation with cycle time, high and low proportional bandwidths, dead zones (S1 and S2)
3) PID regulation with an output signal of 4-20 mA
Thresholds S3, specificities:
S3 may be assigned to an external signal from a sensor (e.g. inductive flow controller) to block the regulation mode (terminals 5 et 6).
In this case the relay S3 becomes N.C. and it opens when the regulation blocking is necessary.
In other words:
- Relay S3 is closed when no voltage applies between terminals 5 & 6
- Relay S3 opens when a voltage applies between terminals 5 & 6 (To block the regulation mode)
Blocking the regulation has the following consequences:
• For impulse regulation mode: relays S1 and S2 keep open (not actuating).
• Inhibition of regulation. – With PID mode, signal 4-20mA will frozen.
• With ON/OFF mode no action occurs on S1 and/ or S2.
INTERNATIONAL
22, Rue de la Voie des Bans · Z.I. de la gare · 95100 ARGENTEUIL
Tel +33 (0)1 30 25 83 20 Web
www.bamo.eu
Fax +33 (0)1 34 10 16 05 E-mail export@bamo.fr
Dissolved oxygen monitor
BAMOPHAR 451 LOG
09-03-2018
M-451.03-EN-AB
OXY
451-03 /10
9.2.2 SETTING OF THRESHOLDS S1, S2 AND S3 IN MODE ON/OFF REGULATION
Below is a detailed description of the steps to set the threshold S1:
Enter to the menu ADJUST ALARM 1
ALARM 1 ON / OFF
To set up output S1, select ON, then confirm.
To disable the relay S1 select OFF mode, then confirm and save the setting.
In ON mode, threshold S1 can be assigned to MEASUREMENT or to TEMPERATURE signal
MEASURE = Threshold dedicated to mg/l or % D.O.
TEMP. = Threshold dedicated to temperature measurement
Select the appropriate mode, then confirm.
HIGH = Triggering when measurement rises over threshold S1
LOW = Triggering when measurement decreases below threshold S1
Select the appropriate actuating mode, then confirm.
Enter the value at which relay S1 will actuate (closing), then confirm.
Enter the value at which relay S1 will be open, then confirm.
With or without delay to actuate the relay S1.
Enter the duration of delay before actuating S1, then confirm.
With or without delay to open relay S1
Enter the duration of delay before opening S1, then confirm.
To record the settings, press the icon SAVE.
ALARM 1 MEASURE / TEMP.
ALARM 1 HIGH / LOW
ON 000.0 mg / % / °C
OFF 000.0 mg / % / °C
DELAY UP ON / OFF
TIME 0000 Sec
DELAY DOWN ON / OFF
TIME 0000 Sec
SAVE ?
Settings for S2 and S3 are available respectively in ADJUST ALARM 2 and ADJUST ALARM 3 menus.
9.2.3 ASSIGNMENT OF S3 THRESHOLD ON AN EXTERNAL SIGNAL
Go to the menu ADJUST ALARM 3
Select the ON mode to set up the triggering of S3, then confirm.
If NO : the sequence of settings are same as for S1 or S2 in ON/ OFF regulation mode
If YES = Relay S3 is dedicated to input signal from external sensor
Relay S3 is then a normally closed contact (N.C.) and will open when blocking regulation mode operates.
Confirm the selection.
To record the settings, press the icon SAVE.
SAVE ?
Measurement
RISING signal
Relay ACTUATED
Trigger point RELAY OFF
Time
Measurement
ADJUST ALARM 3 ON / OFF
EXTERN. YES / NO
Trigger point RELAY ON
Time
Delay before relay switches ON
DECREASING signal
Trigger point RELAY OFF
Trigger point RELAY ON
Measurement
Measurement
Trigger point RELAY ON
Time
Time
Actuated relay
Actuated relay
Delay before relay switches ON
9.3
SETTINGs OF OPERATING MEASUREMENT
Go to menu MEASURE DATA
DISPLAY: mg/l %
SAVE ?
Select the unit mg/l or %, then confirm
To record the settings, press the icon SAVE.
INTERNATIONAL
22, Rue de la Voie des Bans · Z.I. de la gare · 95100 ARGENTEUIL
Tel +33 (0)1 30 25 83 20 Web
www.bamo.eu
Fax +33 (0)1 34 10 16 05 E-mail export@bamo.fr
Dissolved oxygen monitor
BAMOPHAR 451 LOG
09-03-2018
M-451.03-EN-AB
OXY
451-03 /11
9.4
SETTINGS OF ANALOGUE OUTPUT FOR MEASUREMENT
Go to menu OUTPUT mA OXY.
HIGHER 10.00 mg
LOWER 00,00 mg
OUTPUT 4/20 mA 0/20 mA
SAVE ?
9.5
Enter the highest value at 20 mA output signal ; Then confirm.
Enter the lowest value for the output signal at 4 mA (or 0 mA); Then confirm.
Enter the range of output signal; Then confirm.
To record the settings, press the icon SAVE.
SETTINGS OF ANALOGUE OUTPUT FOR TEMPERATURE
Go to menu OUTPUT mA TEMP.
HIGHER +000.0 °C
LOWER 00,00 °C
OUTPUT 4/20 mA 0/20 mA
SAVE ?
9.6
Enter the highest value at 20 mA output signal ; Then confirm.
Enter the lowest value for the output signal at 4 mA (or 0 mA); Then confirm.
Enter the range of output signal; Then confirm.
To record the settings, press the icon SAVE.
FORCED RELAY
This menu allows to test manually the relays S1, S2, S3 and S4.
Status of last one (S4) is OFF (non actuated) by default. Test begins with relay S1.
To test it, press the arrows SELECTION to switch the relay from OFF to ON.
Confirm to go to next relay.
9.7
SETTING OF THE RELAY REGULATION MODE
Go to the menu RELAY REGULATION
There is no access to settings of relay regulation mode if relays S1 and/ or S2 are operating.
IMPULSE MODE:
To access the setting, first disable the relays S and/ or S2 through the menu ADJUST ALARM (1) and/or
(2).
IMPULSE MODE is not available when PID mode is operating.
PID MODE:
To access to impulse regulation mode, go to to the menu REGUL. PID to switch it OFF.
Regulation mode is thus accessible if, and only if, the THRESHOLD and PID modes are disabled.
REGULATION ON / OFF
SET VAL 00.00 mg
T. CYCLE 0060 Sec
HIGH PB 00.00 mg
LOW PB 00.00 mg
HIGH DZ 00.00 mg
LOW DZ 00.20 mg
SAVE ?
Select ON mode to set up parameters, then confirm.
Enter the threshold value, then confirm.
Enter the cycle time according to the process, then confirm.
Enter the differential for upper bandwidth, then confirm.
Note: Output S2 is dedicated to upper bandwidth.
Enter the differential for lower bandwidth, then confirm.
Note: Output S1 is dedicated to lower bandwidth.
Enter the differential value for the upper dead zone, then confirm.
Enter the differential value for the lower dead zone, then confirm.
To record the settings, press the icon SAVE.
NOTE : Relay S1 corresponds to lower bandwidth and S2 to the upper bandwidth.
Measurement
Cycle time
High limit
Dead zone
Set point
Low limit
Time
S1 keeps OFF
INTERNATIONAL
22, Rue de la Voie des Bans · Z.I. de la gare · 95100 ARGENTEUIL
Tel +33 (0)1 30 25 83 20 Web
www.bamo.eu
Fax +33 (0)1 34 10 16 05 E-mail export@bamo.fr
Dissolved oxygen monitor
BAMOPHAR 451 LOG
09-03-2018
M-451.03-EN-AB
OXY
451-03 /12
9.8
SETTING ALARMS
This function activates relay S4 when the switch-on of relays S1 or S2 is delayed (too long time).
Go to the menu ADJUST ALARM
WITH / WITHOUT ALARM
TIME S1 0005 Sec
TIME S2 0005 Sec
SAVE ?
9.9
Select WITH ALARM to set the relay S4, then confirm.
Enter the holding time max. for relay S1, then confirm.
Enter the holding time max. for relay S2, then confirm.
To record the settings, press the icon SAVE.
SETTINGS OF CLEANING ROUTINE
In order not to disturb the regulation, measurements are frozen during the cleaning routine.
This blocking is effective throughout the cleaning time, plus a delay.
This delay allows the dilution of cleaner agent without interfering in the process.
Go to the menu CLEANING PROBE.
CLEANING YES / NO
PERIODE 0000 Sec
TIME 0000 Sec
DELAY 0000 Sec
SAVE ?
Select YES to set up the cleaning routine, then confirm.
Enter the cycle time between two cleaning routines, then confirm.
Enter the duration of cleaning routine, then confirm.
Enter the delay before the measurement is resumed; Then confirm.
To record the settings, press the icon SAVE.
9.10 SETTING A PID REGULATION
It allows setting of a PID regulation through analogue output (0...20 or 4...20 mA) between terminals 3 and 4.
This output is no more dedicated to the copy of temperature measurement.
Go to the menu REGUL. PID
REGULATION ON / OFF
REGUL. AUTO / MANU
SET VAL 05.00 mg
GAIN: 4.800
T.i : 0050 Sec
Td : 0012 Sec
ACTION DIRECT / REVERSE
OUTPUT 4-20mA / 0-20mA
SAVE ?
Select ON to set up parameters of PID; Then confirm.
Select AUTO to access to parameters, then confirm.
Enter the value of set point, then confirm.
Enter the required proportional gain value, then confirm.
Enter the value for integral timer, then confirm.
Enter the required value for derivative timer, then confirm.
Enter the right action, then confirm.
Select the required output signal range, then confirm.
To record the settings, press the icon SAVE.
Note: To block the PID regulation, set a voltage of 24V DC (=20 mA) between terminals 5 (+) and 6 (0).
METHOD FOR ADJUSTMENT OF PID PARAMETERS
In order to setup values for start-up on PID, we recommend to use the Ziegler-Nichols open loop method.
Proceed as follows:
- Connect a recorder to the analogue output or write the displayed values, then draw the diagram vs. time.
- Set the menu PID on MANUAL mode.
- Reach and keep the measurement close to the set point by adjusting the PID output.
- Reach and keep the measurement close to the set point by adjusting the PID output.
- Apply a step (ΔCde) of 10 % for instance, on the analogue output.
(e.g. If preset value was 30%, apply 40%).
- Note on the diagram the corresponding time for this event (step of +10%).
- Find out on this diagram both times t and T:
t = delay of response and T = Time corresponding to a variation Δm of measurement equal to percentage of variation ΔCde of output
(Δm = ΔCde)
This value may be found out from the slope at the beginning of measurements.
-
Set up the PID parameters with the help of following table:
Regulation
PID
PI
PP
Gain
1.2 x T/t
0.9 x T/t
T/t
INTERNATIONAL
22, Rue de la Voie des Bans · Z.I. de la gare · 95100 ARGENTEUIL
Tel +33 (0)1 30 25 83 20 Web
www.bamo.eu
Fax +33 (0)1 34 10 16 05 E-mail export@bamo.fr
Ti (s)
2xt
3.3 x t
9999
Td (s)
0.5 x t
00
00
Dissolved oxygen monitor
BAMOPHAR 451 LOG
09-03-2018
M-451.03-EN-AB
OXY
451-03 /13
∆ Cde = 10%
PID
output
In manual mode
(REGUL. MANU)
Time
∆ m = 10 %
Measurement (mg/l or
%)
Reached set point
Time
9.11 FORCED MEASUREMENT - MEASUREMENT SIMULATION
Go to menu FORCED MEASUREMENT
It is possible to force the measurement or the PID.
The first line shows the current measurement.
Confirm to access to the keyboard.
Enter the value you need to simulate.
Note: The value is immediately considered by BAMOPHAR (thresholds, regulation, 4-20mA outputs ...)
To exit this menu, press icon RETURN KEY.
Note : It is possible to force the output 4-20 mA for PID when PID regulation is in manual mode (REGUL. MANU).
If PID regulation is operating, the display of analogue output is in %.
9.12 ATMOSPHERIQUE PRESSURE
This menu allows you to change the value of the atmospheric pressure, if it is different from the value of the calibration day.
Go to the menu AIR PRESSURE
P = 1013 hPa
SAVE ?
(in hectoPascal); Keeps this value by pressing RETURN icon, or
Enter the new value (in hPa); then confirm.
To record the settings, press the icon SAVE.
9.13 DATA LOGGER
Last 3 menus: CLOCK, RECORD PERIOD, FORMAT MMC, correspond to parameters of the data logger.
For operating and to set up the data logger, please use the specif end-user manual "PHOXLOG".
INTERNATIONAL
22, Rue de la Voie des Bans · Z.I. de la gare · 95100 ARGENTEUIL
Tel +33 (0)1 30 25 83 20 Web
www.bamo.eu
Fax +33 (0)1 34 10 16 05 E-mail export@bamo.fr
Dissolved oxygen monitor
BAMOPHAR 451 LOG
09-03-2018
M-451.03-EN-AB
OXY
451-03 /14